Eddy Current Array Tutorial

Eddy current array (ECA) is a nondestructive testing technology that provides the ability to electronically drive multiple eddy current coils, which are placed side by side in the same probe assembly. Each individual eddy current coil in the probe produces a signal relative to the phase and amplitude of the structure below it. This data is referenced to an encoded position and time and represented graphically as a C-scan image. Most conventional eddy current flaw detection techniques can be reproduced with ECA inspections; however, the remarkable advantages of ECA technology allow improved inspection capabilities and significant time savings.

ECA technology includes the following advantages:


A larger area can be scanned in a single-probe pass, while maintaining a high resolution. Less need for complex robotics to move the probe; a simple manual scan is often enough. C-scan imaging improves flaw detection and sizing. Complex shapes can be inspected using probes customized to the profile of the part being inspected.

1.0 Introduction
1.1 General Introduction to Eddy Current (EC) Testing 1.2 History of Eddy Current Testing 1.3 Equipment

2.0 What is Eddy Current Array (ECA) Testing?


2.1 2.2 2.3 2.4 2.5 Basic Concepts Depth of Penetration Multiplexing Normalization Benefits of Eddy Current Testing

3.0 Probes
3.1 3.2 3.3 3.4 3.5 3.6 EC Probes Eddy Current Array Probes Probes Design Probes Parameters Custom Probes Calibration Standards

4.0 Typical Applications


4.1 Rivet Inspection 4.2 Corrosion Detection

1.0 Introduction
1.1 General Introduction to Eddy Current (EC) Testing
Eddy current (EC) testing is a no contact method for the inspection of metallic parts. Eddy currents are fields of alternating magnetic current that are created when an alternating electric current is passed through one or more coils in a probe assembly. When the probe is linked with the part under inspection, the alternating magnetic field induces eddy currents in the test part. Discontinuities or property variations in the test part change the flow of the eddy current and are detected by the probe in order to make material thickness measurements or to detect defects such as cracks and corrosion. Over the years, probe technology and data processing have advanced to the point where eddy current testing is recognized as being fast, simple, and accurate. The technology is now widely used in the aerospace, automotive, petrochemical, and power generation industries for the detection of surface or near-surface defects in materials such as aluminum, stainless steel, copper, titanium, brass, Inconel, and even carbon steel (surface defects only).

1.2 History of Eddy Current Testing

The phenomenon of eddy currents was discovered by French physicist Leon Foucault in 1851, and for this reason eddy currents are sometimes called Foucault currents. Foucault built a device that used a copper disk moving in a strong magnetic field to show that eddy currents (magnetic fields) are generated when a material moves within an applied magnetic field. Eddy current testing began largely as a result of the English scientist Michael Faraday's discovery of electromagnetic induction in 1831. Faraday discovered that when a magnetic field passes through a conductor (a material in which electrons move easily)-or when a conductor passes through a magnetic fieldan electric current will flow through the conductor if there is a closed path through which the current can circulate. In 1879, another breakthrough was made when another English scientist, David Hughes, demonstrated how the properties of a coil change when placed in contact with metals of different conductivity and permeability. However, it was not until the Second World War that these developments in the transmitting and receiving of electromagnetic waves were put to practical use for materials testing. Beginning in 1933, in Germany, while working for the Kaiser-WilhelmInstitute, Professor Friedrich Frster adapted eddy current technology to industrial use, developing instruments for measuring conductivity and for sorting mixed-up ferrous components. In 1948, Frster founded his own company in Reutlingen, a business based on eddy current testing that continues to this day. Other companies soon followed. Many advances were made throughout the 1950s and 1960s, especially in the aircraft and nuclear industries. There have been many recent developments in eddy current testing, leading to improved performance and the development of new applications. Eddy current testing is now a widely used and well-understood inspection technique for flaw detection as well as for thickness and conductivity measurements.

1.3 Equipment
With thousands of units used throughout the world, the R/D Tech OmniScan MX is Olympus NDT's most successful modular and portable test unit. The OmniScan family includes the innovative phased array and eddy current array test units, as well as the eddy current and conventional ultrasound modules, all designed to meet the most demanding requirements of NDT. The OmniScan MX offers a high acquisition rate and powerful software features in a portable, modular mainframe to efficiently perform manual and automated inspection. The OmniScan ECA test configuration supports 32 sensor coils (up to 64 with an external multiplexer) working in bridge or transmit-receive mode. The operating frequency ranges from 20 Hz to 6 MHz with the option of using multiple frequencies in the same acquisition.

2.0 What is Eddy Current Array (ECA) Testing?


2.1 Basic Concepts

Eddy current array and conventional eddy current technology share the same basic principle. Alternating current injected into a coil creates a magnetic field (in blue). When the coil is placed over a conductive part, opposed alternating currents (eddy currents, in red) are generated. Defects in the part disturb the path of the eddy currents (in yellow). This disturbance can be measured by the coil. Eddy current array (ECA) technology provides the ability to electronically drive multiple eddy current coils placed side by side in the same probe assembly. Data acquisition is performed by multiplexing the eddy current coils in a special pattern to avoid mutual inductance between the individual coils. Most conventional eddy current flaw detection techniques can be reproduced with an ECA inspection. With the benefits of single-pass coverage, and enhanced imaging capabilities, ECA technology provides a remarkably powerful tool and significant time savings during inspections.

Major advantages of ECA testing are the following: Larger area can be scanned in a single-probe pass, while maintaining a high resolution Reduced need for complex robotics to move the probe; a simple manual scan is often enough Improved flaw detection and sizing with C-scan imaging Inspection of complex shapes using probes customized to the profile of the part being inspected The OmniScan ECA test configuration supports 32 sensor coils (up to 64 with an external multiplexer) working in bridge or transmit-receive mode. The operating frequency ranges from 20 Hz to 6 MHz with the option of using multiple frequencies in the same acquisition.

2.2 Depth of Penetration

Eddy current density does not remain constant across the depth of a material. The density is greatest at the surface and decreases exponentially with depth (the "skin effect"). The standard depth of penetration equation (shown to the right) is used to explain the penetration capability of eddy current testing, which decreases with increasing frequency, conductivity, or permeability. For a material that is both thick and uniform, the standard depth of penetration is the depth at which the eddy current density is 37% of the material surface value. To detect very shallow defects in a material, and also to measure the thickness of thin sheets, very high frequencies are used. Similarly, in order to detect subsurface defects, and to test highly conductive, magnetic, or thick materials, lower frequencies must be used.

Where: d = Standard depth of penetration (mm) p = 3.14 = Test frequency (Hz) m = Magnetic permeability (H/mm) s = Electrical conductivity (% IACS) + Formula poster

2.3 Multiplexing

Multiplexing is the process by which multiple analog message signals are combined into one digital signal on a shared medium. When eddy current array data is multiplexed, the individual eddy current coils are excited at different times, allowing the system to excite all of the coils in the probe without ever exciting any two adjacent coils at the same time. An undesirable effect known as mutual inductance (magnetic coupling between coils in close proximity) is minimized with the use of an internal multiplexing system to carefully program the exact time that each coil is excited to transmit its eddy current signal. The signals are then reassembled before being displayed as an image. In addition to the enhanced imaging capabilities of multiplexed data, multiplexing allows any individual coil (data) channel to be analyzed after inspection. Multiplexing allows an increased channel resolution, increased coil sensitivity (through the reduction of mutual inductance), and a reduced noise level. This ultimately leads to an improved signal-to-noise ratio.

Mutual Inductance is avoided by multiplexing

2.4 Normalization

The main purpose of performing normalization is to standardize sensitivity for an ECA probe. To do this, the operator scans a sample containing a calibration defect in order to generate the same eddy current signal for each channel. For most applications, a defect such as a long transversal notch will

suffice (see illustration to the right). Using the calibration defect, the operator adjusts the gain and rotation of each channel so that the same phase and amplitude response is obtained for all channels.

2.5 Benefits of Eddy Current Testing


Benefits of Eddy Current Testing Eddy current offers the following capabilities: Quick, simple, and reliable inspection technique to detect surface and nearsurface defects in conductive material Can be used to measure material electrical conductivity Measurement of nonconductive coating Hole inspection with the use of high-speed rotating scanner and surface probe

Benefits of Eddy Current Array Testing Compared to single-channel eddy current technology, eddy current array technology provides the following benefits: Drastically reduces inspection time

Covers a large area in one single pass Reduces the complexity of mechanical and robotic scanning systems Provides real-time cartography of the inspected region, facilitating data interpretation Is well suited for complex part geometries Improves reliability and probability of detection (POD)

3.0 Probes
3.1 EC Probes

Olympus NDT`s standard R/D Tech eddy current probes are available in different configurations:

Bolt hole probes Surface probes, in various shapes and configurations Low-frequency Spot and Ring type probes Sliding probes Wheel probes Conductivity probes Specialty probes made for specific applications

Reference standards with EDM notches can be manufactured according to the application specifications. Probes used to perform eddy current inspections are made with a copper wire wound to form a coil. The coil shape can vary to better suit specific applications. a-The alternating current flowing through the coil at a chosen frequency generates a magnetic field around the coil. b-When the coil is placed close to an electrically conductive material, eddy current is induced in the material. c-If a flaw in the conductive material disturbs the eddy current circulation, the magnetic coupling with the probe is changed and a defect signal can be read by measuring the coil impedance variation. Surface preparation is minimal. Unlike liquid penetrant or magnetic particle inspection, it is unnecessary to remove the paint from the surface to inspect the parts.

3.2 Eddy Current Array Probes

Olympus NDT manufactures R/D Tech ECA probes for a wide range of applications. Probes can be designed to detect a specific type of flaw or to follow the shape of the part being inspected. Standard designs are available to detect defects such as cracks and pitting, and subsurface defects like cracks in multilayer structures as well as corrosion.

3.3 Probes Design


Surface

Surface probes are made with coils designed to be driven at relatively high frequencies (typically 50 kHz to 500 kHz). Using higher frequencies results in less penetration of the eddy current field into the test part, allowing the area directly below the surface of the part to be inspected. In addition, the higher frequencies provide a higher resolution for the detection of smaller defects.

Subsurface

The coils in subsurface probes are designed to be driven at relatively low frequencies (typically 1 kHz to 20 kHz). Using lower frequencies results in greater penetration of the eddy current field into the test part, allowing cracks or corrosion to be detected in thicker structures or in multilayer structures such as aircraft lap joints. Subsurface probes provide greater penetration; however, sensitivity to small defects decreases as the frequency and penetration are increased.

High Frequency, High Resolution

High-frequency, high-resolution probes provide both a high frequency and a high resolution in order to detectsurface-breaking defects on aluminum aircraft skins. These probes are made with 32 absolute coils positioned for the complete coverage of a 26 mm scan area. Although these probes are used for a specific application, they are very flexible and suitable for various high-frequency surface-breaking applications being developed.

3.4 Probes Parameters


To achieve optimal inspection performance, there are several important parameters to consider when designing an effective eddy current probe. Key factors include inspection coverage, sensitivity, frequency and, of course, cost. To optimize performance, it is important to carefully balance the various probe parameters. For example, high-sensitivity probes require small, high-frequency coils (providing less coverage); probes capable of greater coverage require larger, lower-frequency coils (resulting in decreased sensitivity to small defects). As is the case with conventional eddy current inspections, choosing the correct probe characteristics is essential to a successful inspection.

Where: n = Number of channels r = Resolution (also depends on the coil configuration) C = Coverage

Probe Structure

Eddy current array probes can be optimized for a specific application by

changing the probe shape and the coil configuration. Most coil configurations can be expanded into an array configuration.

Single coil probe


The earliest instruments used in aircraft inspection included the Magnaflux ED-500 and ED520, and the Foerster Defectometer (although not a bridge type instrument but a resonant circuit type), all of which used single-coil probes. The probes contain a single coil that is wound to a specific value. No other coil is needed. More recently, the introduction of the Hocking Locator and newer models of the Foerster Defectometer have kept this kind of instrument as a popular option for many users. When these probes are used with a bridge circuit type instrument, a balance coil is also required. Balance coils are normally placed in the cable connector or a separate adapter (see Fig. 1).

Fig. 1

It sometimes creates a problem when the probe inductance value is not close enough to the value of the balance coil, and the instrument does not balance correctly. This happens more often when they are not made by the same manufacturer. The result is poor performance (noisy or insensitive) or no response at all (signal saturation).

Bridge Type probes


In this configuration the probe coils are located in an electrical "bridge" (see fig. 2). The instrument balances the bridge and any change in balance is displayed as a signal.

Fig. 2 In this arrangement, the same coil produces the eddy currents and detects the impedance changes caused by the defects (or any other variables). Almost all instruments are able to operate with this type of coil arrangement.

Reflection type probe


These probes are also known as send-receive or driver-pickup. In this configuration, the eddy currents are produced by a coil connected to the instrument's oscillator (driver).

Fig. 3 The signals received back in the probe are detected by separate coils called pickups (see Fig. 3 and Fig. 4). All new impedance plane instruments and also many older models are able to operate in both bridge and reflection modes. If you are in doubt, call the manufacturer or give us a call.

Fig. 4

Bridge or reflection
This is a common question asked by those involved in trying to select the best probe for an inspection. The answer is "It depends." Let us consider both systems. Gain. Reflection probes will give a higher gain, particularly if they are "tuned" to a specific frequency, but normally the difference is on average about 6 dB. It is true that t his doubles the signal, but if you consider that the instruments are able to give this increase of gain easily, it is not so important. Nevertheless, in critical applications this increase is very welcomed. Frequency range. Reflection probes do not need to balance the driver to the pickup coils. This means that they will give a wider frequency range. As long as the driver produces eddy currents, the pickup will detect them and some signal will be displayed. This may not provide good information at certain frequencies, but the probe is still working! Bridge type probes used to give a limited frequency span in the older instruments, as these had to balance an electrical bridge using its other arms (X and R controls). In modern instruments, the bridge is normally formed with fixed precision resistors, or a fixed transformer inside it. The signals detected in this manner are electronically processed without any "mechanical" adjustments, and this means a greater ability to balance over a wider frequency range. Drift: Probe drift is mostly caused by temperature change in the coils. This may be caused by varying ambient temperature, or the heat produced by the oscillator current, or both. There are design parameters that can be optimized to reduce drift, such as wire diameter and ferrite selection, but

reflection probes are normally a good choice to avoid this problem even more. In a reflection probe, the driver current does not flow through the pickup coils; in fact, the magnetic field received back from the specimen is normally much smaller and, consequently, the current flowing in the pickups is also reduced. Most probe types (pencil, spot, ring, bolt hole, etc.) can be made as bridge or reflection. Keep in mind that a reflection probe is almost invariably more difficult to manufacture and therefore more expensive.

Absolute and differential probe


This is an area where some confusion exists. Many users have called a probe "differential" when the signal displayed gives an up and down movement or a figure 8 type signal. This is caused by the two coils sensing the defect in sequence. When both sensing coils are on the probe surface, they compensate for lift-off and as a result no line is visible (see Fig. 5).

Fig. 5 In contrast, an absolute display is produced by a single sensing coil (see Fig. 1 through Fig. 4), giving a single, upward movement with a near horizontal lift-off line. Others have called a probe "differential" simply when the coils were connected differentially such as in a bridge circuit. The problem with this definition is that probes can be connected differentially in a reflection system as well as when using two pickups (such as most scanner-driven bolt hole probes). In this case, the two pickup coils are positioned close to one another and contained within a driver coil (see Fig. 6). The best way out of this confusion is often to specify the probe as bridgedifferential, absolute, or reflection-differential-absolute as needed. It seems to make more sense to qualify the description according to the displayed

signal, since this is what really matters. Not many people are concerned as to how the coils are connected internally.

Shielded and unshielded probes


Probes are normally available in both shielded and unshielded versions; however, there is an increasing demand for the shielded variety. Shielding restricts the magnetic field produced by the coils to the physical size of the probe or even less. A shield can be made of various materials, but the ones mostly used are: ferrite (like a ceramic made of iron oxides), Mumetal, and mild steel. Ferrites make the best shields because they provide an easy path for the magnetic field but have poor conductivity. This means that there are few eddy current losses in the shield itself. Mild steel has more losses but is widely used for spot probes and ring probes due to its machinability and when ferrites are not available in certain sizes or shapes. Mumetal is used sometimes for pencil probes as it is available in thin sheet; however, it is less effective than ferrite. Shielding has several advantages: first, it allows the probe to move in (or close to) geometry changes, such as edges, without giving false indications; next, it allows the probe to touch ferrous fastener heads with minimal interference; last, it allows the detection of smaller defects due to the stronger magnetic field concentrated in a smaller area. On the other hand, unshielded probes allow somewhat deeper penetration due to the larger magnetic field. They are also slightly more tolerant to liftoff. Unshielded probes are recommended for the inspection of ferrous materials (steel) for surface cracks, and in particular with meter instruments. The reason for this is that the meter response is too slow to allow the signal from a shielded probe to be displayed at normal scanning speeds due to the smaller sensitive area.

Adapters
To connect a probe with a connector different from the type used in the instrument, it is necessary to use an adapter. An adapter consists of two different connectors joined and wired to match the inputs and outputs as necessary. It is normally housed in a short body that can be positioned at the instrument's input. Sometimes, it is also possible to have a "cable adapter," which is made to match a connector located at the probe body. Depending on the instrument's wiring, it may be possible to have a single adapter for both bridge and reflection. In other cases, it is necessary to have two separate adapters or use a switchable type.

3.5 Custom Probes

Custom probes can be ordered to suit specific geometries or applications. Custom probes are designed and manufactured specifically for the task

required using coils for low-frequency or high-frequency inspections. Your local Olympus NDT representative will help to answer any questions regarding custom solutions. Olympus manufactures R/D Tech ECA probes for a wide range of applications. Probes can be designed to detect a specific type of flaw or to conform to the shape of the part under inspection. Standard designs are available to detect defects such as cracks and pitting, and subsurface defects such as cracks in multilayer structures, as well as corrosion. Probes can be made in different shapes and sizes to better conform to the contour of the part under inspection.

3.6 Calibration Standard

Olympus NDT has the capability and experience to manufacture a wide range of calibration standards for eddy current array applications. Whether these calibration standards are defined in an aircraft manual or are entirely custom-made, Olympus NDT can manufacture to any requirements. Sample parts can also be manufactured, or real parts can have artificial defects inserted into the material using spark-erosion and wire-cutting technologies.

4.0 Typical Applications

4.1 Rivet Inspection

Each individual eddy current coil in the probe produces a signal relative to the phase and amplitude of the structure below it. This data is referenced to an encoded position and time and represented graphically as a C-scan image. For rivet applications, the eddy current coils that pass over a defective rivet generate a unique signal response. For coils that are affected by a crack initiating from the rivet hole, an amplitude change is represented in the C-Scan display. For coils that detect no change, the color representation remains constant in the C-scan display.

4.2 Corrosion Detection

Corrosion detection using eddy current array technology offers major advantages over conventional eddy current inspection methods. Because each individual eddy current coil generates a unique electrical signal in relation to the structure below it, the coils can detect very small changes in material thickness, along with other parameters, and display these changes as a color-coded C-scan image. Imaging using eddy current array allows easy interpretation of the data generated from the probe coils. After it has been collected, the inspection data can be stored, transmitted, and analyzed. Color palettes play a very important role in the imaging of eddy current array data. Color palettes determine how the data will be displayed. Color palettes are often linked to the amplitude of the eddy current signals; however, when required, color palettes can also be linked to the phase angle of the signals. Color palettes range from a gradual rainbow palette to a precise, two-color "go/no-go" palette. Black and white palettes are also often used.

Eddy Current Probes and Application Guide

Introduction
This paper is intended to provide information to help the user in selecting the right E.C. probe(s) for a given inspection. Using this data, best results will be achieved by optimizing the frequency and choosing a suitable instrument. The subject is divided into three sections -

1. Coil types available 2. Typical applications 3. New developents

1. COIL TYPES AVAILABLE

The early E.C. coils generally had either no core ( air core) or a ferrite center

core only. Meter instruments were used almost exclusively and sensitivity was comparatively low, but in most cases adequate. Many inspections are still being done using this type of comparatively large, low sensitivity coils.

Absolute Type Coils


A very widely used type coil is a 100 kHz with a diameter of .1" (2.5mm) to . 2" (5mm) that fits most bridge type older meter instruments (Figure 1). The sensitivity is acceptable for long cracks that exceed the probe diameter by a factor of approximately 2, as the field is considerably larger than the coil size. This also produces a large edge effect.

Figure 1 A suitable alternative now offered is a smaller coil approximately .060" (1.5mm) diameter with a shield (preferably ferrite) around it. This gives improved sensitivity particularly to short cracks and good isolation from edges, bolt heads, etc. 1 The older probes were normally calibrated using an infinitely long (to the coil) notch, .040" (1mm)or ,020" (.5mm), and their sensitivity will drop off rapidly with shorter notches. The new probes maintain their sensitivity with a notch approximately the internal diameter of the ferrite shield and will still detect a shorter one. Even meter instruments benefit from this type of coil, but probing speed is

limited as the needle needs time to respond. Display-type instruments allow for much faster scanning. Larger surface type probes respond in a similar way to the above. Shielding produces similar improvements when looking for subsurface cracks or corrosion.

Figure 2

Differential Type Coils


Differential coils have the attraction of built-in lift-off compensation. This has made them useful for many applications. The older types of coils had no ferrite shield and they were built just by placing two coils side-by-side (Figure 3). Later types added individual shields (Figure 4), but the greatest improvement to the sensitivity was achieved when both coils were placed within a common shield (Figure 5). Differential type probes are mostly used in small sizes for surface crack detection only.

Figure 3

Figure 4

Figure 5 In a probe of this type signals that affect both Normally the air point difference is present due both coils are wound in opposition. Consequently, simultaneously will cancel out (such as lift- off). and the working point will be close, but some to small coil variations.

Normal scan direction is as shown (Figure 6), giving the typical display presentation. The double indication is, in fact helpful, as it doubles the size of the defect in the screen (Figure 7).

Figure 6

Figure 7 Sometimes it is necessary to scan in the same direction as the cracks (Figure 8). This is permissible and the result will be similar for a very short defect. A larger defect affecting both coils will tend to cancel out because they are in opposition, but in practice there are enough differences in angle and depth for this not to happen totally. In any case, the ends of the crack will show normally.

FIgure 8

The probe body has a line to show the normal scan direction. These coils can be fitted in pencil, bore hole or molded probes to fit almost any shape.

Bridge and Reflection Coils


Older coils were normally connected to the two branches of a bridge configuration (Figure 9). Later coils have also been used in the reflection mode where separate coils are used for generating and detecting the eddy currents (Figure 10).

Figure 9

Figure 10 Bridge coils give generally good performance, particularly if the probe is designed for a specific application and frequency. Reflection coils will often give a greater gain and a wider frequency range of operation but they are more complicated to manufacture. They also have less drift. 2 Reflection probes are also used on special probe designs that rely on the transmit-receive principle to create a certain size/shape sensitive area (like the sliding type probes).

2. TYPICAL APPLICATIONS
Surface Cracks
Normal operating frequency: Aluminum - 100 kHz - 1 MHz Steel - 1 MHz - 2 MHz Inconel Titanium - 2 MHz - 5 MHz

Probes available
Pencil types - Absolute or differential . Shielded or unshielded Surface types - Larger diameter probes can sometimes be used Sliding probes - For cracks starting under fastener heads Wheel probes - For bead seat radius, molded

Subsurface Cracks
Frequency must be low enough to penetrate the required depth (use slide rule or graph attached). Minimum usable frequency is 100 HZ. 3 Probes Available : Surface probes - Normally absolute shielded. Diameter can be as large as spacing between fasteners Encircling probes - Give better penetration. I.D. must be close to fastener head diameter Sliding probes - For fast directional inspection (reflection)

Bore Hole Cracks


Frequency as for surface cracks Probes available: Absolute - (preferably shielded) in hand operated or scanner versions. Differential - (shielded or unshielded) - hand operated and scanner versions Probes can be expanding (contact type) or non-expanding (slightly below

hole size) . Contact probes can be more sensitive to cracks (no lift - off distance) but they also generate scanning noise. Low frequency types can be used to penetrate through brass bushings. Automated systems can be implemented. 4

Coatings
Frequency varies with type of coating Non-conductive coatings can be considered as lift-off measurements. Metallic coatings require good penetration, but as they are normally very thin, fairly high frequencies are usable. Standard absolute pencil and/or surface probes are suitable.

Conductivity
Using a standard absolute pencil or surface probe and a conductivity sample, it is possible to identify different alloys with a standard instrument. More accurate measurements require special conductivity meters and probes. 5,6

Corrosion/Thickness
Frequency must be able to penetrate thickness required. Use EC slide rule or graph attached. Best results are obtained using shielded surface probes.
7,8

3. NEW DEVELOPMENTS
Aluminum Angle Inspection
The problem was to inspect aluminum angle for cracks in the opposite side (not accesible) in one pass and with high sensitivity. In order to cover the required width, the coil has a lift-off of about .1" (2.5mm). A modified version of the sliding probe proved to be the best answer to several alternatives tried (Figure 11). A .020" maximum (15mm) depth notch produced with a 1" diameter blade was detectable. Lift-off variations and edge effects could be identified as such on the display.

Figure 11

Thin Probe
To inspect between two lugs spaced less than .080" (2mm) apart, the probe had to be shielded and be completely unsensitive at the back (Figure 12). Using stainless steel for strength, a shielded coil was placed at the tip . A nylon handle placed at 45 allowed for better access. Such a flat coil proved very sensitive to small cracks both in steel and aluminum.

Figure 12

Small Sliding Probes


Used to inspect between fasteners in fuselage structures, a sliding probe is easier to handle than a small round probe. They give directional response with a frequency range of 1-100 kHz and penetration up to .100" (2.5mm)

Sliding Probe Transparent Lens Adjustment


The sensitive area of this type of sliding probe is the distance between the driver and pick-up sections. The lens is therefore adjusted to the required width of inspection (Figure 13).

Figure 13

When scanning along a line of fasteners, this distance can also be optimized for best discrimination between the fastener indication and the defect . With

no lens present, the probe is at its highest gain, but because of this it also gives a large indication for the fastener (Figure 14).

Figure 14 By allowing a thin lens, it is noticeable that the fastener indication has decreased, but the defect has not. Eventually a good compromise can be reached where the gain is not too high. The defect gives an indication that is clearly seen and the fastener only causes a comparatively small movement. A too wide lens will result in a very high gain setting and a decrease in sensitivity to short defects. 9

Figure 15

Transparent Probe Holders


When calibrating a probe in a test block with small notches, it is difficult to optimize the indication with an opaque holder. The holder is needed for stability, so using a transparent material makes it possible t o see the notch position (Figure 16) and make the calibration easier. It is also useful when scanning

in confined areas or between fasteners to avoid running too close to the heads.

Figure 16

High Gain Probes


Sometimes standard types of EC probes are not able to give a large enough indication for the defect. This may be due to geometry, defect size or a higher than normal lift-off . To achieve the extra gain, it is often possible to position a very small hybrid amplifier circuit within the probe body or in a separate small box close to it (Figure 17). This pre-amplifier can be tailored to a specific inspection if needed.

Figure 17 With this extra amplification the instrument gain can be lowered to a medium setting and the resultant signal will have a better signal-to-noise ratio than before. Care should be taken to ensure that too large of a signal does not saturate the instrument input. A quick check for this is to compare the lift- off with the edge effect. If both lines superimpose on each other,

saturation is present and the gain should be lowered (there should always be a small angle separation). A typical application has been the testing of bore holes in disks with a lift-off of up t o .020" (probe .040" under size), enabling the use of a single probe size. Extra gain has been needed to inspect aluminum-titanium sandwich structures. The titanium allows eddy currents to penetrate to the aluminum layer if the gain available is sufficient. Unfortunately, when the aluminum is covering the titanium, the eddy currents do not penetrate this as the conductivity is much lower than the aluminum, and it behaves like a shield. If more than one probe is needed, it is possible to multiplex several probes or coils and use a single instrument. 10

REFERENCES
1. "Shielded E.C. Probes" Process Laboratory Report The Physical Sciences Laboratory Branch, Directorate of Maintenance, Sacramento Air Logistics Center, McClellan A.F.B. Project No. 81-497-085, Donald M. Bailey 2. "E. C. Development Activities" Walter J. Harris, Boeing Commercial Airplane Co., ATA 1981 3. "Low Frequency E.C. Inspection of Aircraft Structure" D. J. Hagemaier and A. P. Steinberg, Douglas Aircraft Company ATA 1980 4. "Update on Automated E.C. Inspection System" Art Thompson, G.E. Evendale, ATA 1983 5. "Boeing Process Specification" , BAC 5946 6. "Evaluation of Heat Damage to Aluminum Aircraft Structures" D. J. Hagemaier, Douglas Aircraft Company, ATA 1981 7. "A General Procedure for the Detection and Measurement of Corrosion in Aircraft Skins Using E.C. Equipment" J. Pellicer, Staveley NDT Technologies, Inc., NORTEC Division, ATA 1984 8. "Aircraft Corrosion and Detection Methods" D. J. Hagemaier, A. H. Wendelbo, Jr., Douglas Aircraft Co., ATA 1984 9. "Sliding Probes with Transparent Lens", Technical Bulletin J. Pellicer, Staveley NDT Technologies, Inc., NORTEC Division 10. "Application Advancement in E.C. Using Pre-Amplifiers and Multiplexers "

J. Pellicer, Staveley NDT Technologies, Inc., NORT EC Division, ASNT Spring Conference, Tulsa, OK 1986

Eddy Current Technology


An update for aircraft engine inspections

Eddy current array probes can replace one axis of a two-axis scan and offer greater flexibility in the eddy current setup.

Eddy current array probe and tooling that requires one pass to inspect the dovetail

The use of a C-scan image increases the reliability of the inspection because it guarantees full coverage of the surface inspected.

High-frequency eddy current array probe for surface crack detection.

The Omniscan MX is a modular, portable, battery operated instrument. Existing modules can drive ultrasonic phased array and conventional ultrasound as well as eddy current array. The Omniscan ECA configuration supports 32 sensor coils (up to 64 with an external multiplexer) working in bridge- or transmit-receive mode. The operating frequency ranges from 20 Hz to 6 MHz with the option of using multiple frequencies in the same acquisition. Encoders can be connected. Data files up to 200 Mb can be recorded.

Engine disk dovetail inspection

By Andre Lamarre Eddy current technology is an electromagnetic technique widely used in the aerospace, automotive, petrochemical, and power generation industries for the inspection of metallic structures. In this technique, the probe, which is excited with an alternative current, induces eddy current into the part under inspection. Any discontinuities or material property variations that change the eddy current flow in the part are detected by the probe and considered a potential defect. Recently, an important improvement of this technique was realized with the development of eddy current array technology. Eddy current arrays Eddy current array (ECA) technology electronically drives and reads several

eddy current sensors positioned side-by-side in the same probe assembly. Data acquisition is made possible through the use of multiplexing, which avoids mutual inductance between the individual sensors. Benefits of eddy current arrays Compared to single-channel eddy current technology, eddy current array technology provides the following benefits:

Drastically reduces inspection time Covers a large area in a single pass Reduces the complexity of mechanical and robotic scanning systems Provides real-time cartography of the inspected region, facilitating data interpretation Is well suited for complex part geometries Improves reliability and probability of detection (POD).

Eddy current array probes An eddy current array probe can be optimized for a specific application by changing the coil configuration and the probe shape. Different types of probes can be realized like absolute bridge, differential bridge, absolute reflection, differential reflection, transmit-receive, shielded, and cross-axis. Probes can be designed to detect a specific type of flaw or to follow the shape of the part being inspected. Standard designs are available to detect surface defects (such as cracks and pitting) and subsurface defects (such as cracks in multilayer structures and corrosion). Imaging Representation of the data plays a major role in the use of eddy current array. The Omniscan ECA allows the use of C-scan imaging, which is a colorcoded two-dimensional mapping of the inspected surface. The data from each individual coil are recorded so the impedance plane is always available. The following image demonstrates the principle of the image representation. The ECA probe moves over a flaw and each coil produces an EC signal as shown. The C-scan representation is the color coded image of the amplitude of the signal. The X axis represents the movement axis while the Y axis represents the probe axis. So, the C-scan gives the position of the flaw as well as its size.

Testing and inspection Eddy current array testing is used in a number of different fields like aircraft and engine maintenance and manufacturing, power generation, oil and gas, and tube manufacturing. This article covers the inspection of the engine disk dovetail. Aircraft engines are submitted to a lot of stress. Rupture of rotating parts can cause catastrophic failure of the engine and the aircraft. The engine disk dovetails or the blade attachment, being submitted to high stress, are inspected to make sure there is no crack initiation site within the area of the dovetail. Conventional eddy current are mainly used for this application. The operator is required to use a tooling that holds a conventional EC probe. The operator scans the probe along the dovetail length and index position of the probe approximately 40 times to assure full coverage. The operator also has to constantly monitor the screen of the conventional eddy current instrument. This method is long, tedious, and causes fatigue to the operator. An eddy current array method was developed specifically for this application resulting in a significant time-savings. A new eddy current array was developed composed of 32 sensors shaped in a way to fit the dovetail contour. This probe is attached to a holder designed for the application. The operator simply puts the probe in place and pushes it only one time through the dovetail. This one-pass inspection assures full-coverage of the dovetail areas. The C-scan mapping displayed on the instrument also helps the operator to localize position of a defect and to size it. The inspection time is reduced dramatically while the reliability of the inspection is increased resulting in an important gain for the inspectors. Conclusion Eddy current array is a new technology that is used successfully in many different fields. Portable and easy to operate, the Omniscan ECA makes the use of EC arrays easy. The main advantages of the ECA technology are increasing inspection speed, better reliability due to C-scan imaging, and better reproducibility and probability of detection due to the coverage of the whole surface assured by the array. Andre Lamarre is Business Development Director, Aerospace and Defense for Olympus NDT