Sie sind auf Seite 1von 1

Label Durability Test Petrol Test

IEC 60950
http://www.i-spec.com/IEC_60950/durability.html Note that references are to IEC 60950 Third Edition 1999-04. Where the references in the Second Edition differ, they appear in parentheses. REQUIREMENT Markings required by the standard must be durable and legible. In considering the durability of the marking, the effect of normal use shall be taken into account. TEST METHOD The marking is rubbed by hand for 15 seconds with a piece of cloth soaked with water, and then for a further 15 seconds with a piece of cloth soaked in petroleum spirit. After the test, the marking should be legible. There should be no curling and it should not be possible to remove the marking easily. REFERENCE Complete requirements are given in IEC 60950, Clause 1.7.13 (1.7.15).

Arabian Standard - SASO

SSA 385/1983 METHODS OF TESTING ROOM AIR CONDITIONERS UDC 621.395.92 : 621.56.59
5- CHECKING OF DURABILITY OF MARKING The durability of marking shall be checked by rubbing it by hand for IS seconds, with a piece of cloth soaked in water and then for a further 15 seconds with a piece of cloth soaked in petroleum spirit.
http://www.devicelink.com/mddi/archive/04/05/009.html

Durability of Markings IEC 60601-1 requires that both internal and external markings be clearly legible and that all required external markings be durable. IEC 60601-1 specifies rubbing external markings with water, methylated spirit, and isopropyl alcohol. Using a cloth rag, each substance is to be rubbed on the external marking for 15 seconds. The test requirements include verifying that the external markings are clearly legible and that labels do not fall off or curl after the rub treatments.