Beruflich Dokumente
Kultur Dokumente
Important
Qualified service personnel must install this modification. See the Service Portal for possible updates before doing this modification. When doing the procedures in this document, you must use safe work practices and wear the correct personal protective equipment (for example, safety eyewear) according to your companys standard operating procedures.
SERVICE MANUAL
0086
Use of Manual CARESTREAM CR Systems are designed to meet international safety and performance standards. Personnel operating the unit must have a thorough understanding of the proper operation of the system. This manual has been prepared to aid medical and technical personnel to understand and operate the system. Do not operate the system before reading this manual and gaining a clear understanding of the operation of the system. If any part of this manual is not clear, please contact your Carestream Health representative for clarification.
Authorized Representatives Carestream Health France 1, rue Galilee 93192 NOISY-LE_GRAND CEDEX France
Carestream Health, Inc. Rochester, NY 14608 CARESTREAM is a trademark of Carestream Health. Carestream Health, Inc., 2011
6H486603MAR11
Table of Contents
Description Safety and Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . General Safety Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrical Hazards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Explosion and Implosion Hazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Overheating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Laser Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Recycling the Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Labelling Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IEC Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Device-Specific Safety Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . System Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Operational Principles. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Component Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Service Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Service Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Removing the Service Panel and Scanner Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . Disconnecting the Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Replacing the Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Power Inlet Module Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . USB Board Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Motion Board Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sensor Board Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . PM Tube and PM Board Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Laser Board Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Driver Motor Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Roller Motor Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Linear Slide Assembly Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Loader Stepper Motor Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Power Supply Assembly Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Erase Lamps Assembly Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Erase Lamps Inverters Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Erase Lamps Sensors Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Left Limit Sensor Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Right Limit Sensor Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Screen Size Sensor Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Roller Sensor Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z0 Sensor Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . W0 Sensors Replacement (Top/Receiver and Bottom/Transmitter). . . . . . . . . . . . . . Back Rollers Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Screen Guide Replacement for Non-Auto-Loop Systems . . . . . . . . . . . . . . . . . . . . . Auto-loop Key Assembly Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Auto-loop Solenoid Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Tray Assembly Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Tray Assembly Replacement for Auto-loop Systems . . . . . . . . . . . . . . . . . . . . . . . . . Calibrations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Nominal Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Facility Calibrations for KODAK QC Software Version 2.1.2/2.4.2 Software . . . . . . . Gain Tuning Calibration Process for KODAK QC Software Version 2.5 and 3.0 . . . . Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Auto-loop Solenoid Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6H4866
SERVICE MANUAL
Auto-Loop Adjustment Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Screen Guide Adjustment (Non-Auto-Loop) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Loader Pin and Loader Pusher Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Roller Sensor Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electric Schematic Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Preventive Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cleaning the Rollers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cleaning the Phosphor Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Publication History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6H486603MAR11
Caution
Cautions point out procedures that you must follow precisely to avoid damage to the system or any of its components, yourself or others, loss of data or corruption of files in software applications.
Note
Notes provide additional information, such as expanded explanations, hints, or reminders.
Important
Important highlights critical policy information that affects how you use this guide and this product
SERVICE MANUAL
Electrical Hazards
Caution
Do not remove or open system covers or plugs. Internal circuits use high voltage capable of causing serious injury. Fuses blown within 36 hours of being replaced by a qualified technician may indicate malfunctioning electrical circuits within the system. Have the system checked by qualified service personnel. Do not attempt to replace any fuse. Fluids that seep into the active circuit components of the system may cause short circuits that can result in electrical fires. Therefore, do not place any liquid or food on any part of the system.
Overheating
Do not block the air circulation around the unit. Always maintain at least 6 in. (15 cm) clearance around the unit to prevent overheating and damage to the system.
Laser Warning
When a service operation is taking place with the cover removed, disconnect the laser according to the procedure in Disconnecting the Laser on Page 32. If the laser must be activated during the service procedure, wear protective safety glasses at all times. The required laser safety eyewear must be intended for HeliumNeon/PDT lasers, have an optical density of 4-5 wavelengths of 610-695 nm, and be marked as having CE approval.
6H486603MAR11
Labelling Summary
Safety Labels Laser Laser-emitting product Class 3B laser product inside scanner High voltage Consignes de Scurit Laser Appareil metant de laser Rayonnement de laser vitez lexposition au faisceau laser de la classe 3B. Appareil laser de classe 3B a lintrieur du Scanner. Haut voltage
Power OFF
6H4866
SERVICE MANUAL
The KODAK Point-of-Care 120/140 Scanner is a Le Scanner KODAK Point-of-Care 120/140 est CLASS 1 Laser product. un produit laser de la Classe 1. Do not remove the scanner cover. Ne pas retirer le couvercle du scanner. Cover removal shall be done only by Le retrait du couvercle doit seffectuer authorized service personnel! uniquement par un personnel comptent.
Regulatory Information
Introduction
This Product conforms to the following safety standards: IEC 601-1 Medical Electrical Equipment General Requirements for Safety, EN60601-1-2 Medical Electrical Equipment Electro-Magnetic Compatibility Requirements and Tests, IEC 60825-1 Safety of Laser Products. This device complies with 21CFR 1040.10.
CE Conformity
This product conforms to the requirements of council directive 93/42/EEC. The Point-of-Care CR 120/140 is a Class I medical device. The Point-of-Care CR 120/140 bears the following mark of conformity.
0086
The name and address of the CE representative appears on the back of the front page of this manual.
USA Regulations
The FDA cleared the system for sale in the USA.
Caution
Federal US law restricts this device for sale by or on the order of a physician.
6H486603MAR11
System Description
Operational Principles
The Point-of-Care CR 120/140 is a digital imaging system for image acquisition and processing of static projection radiography that uses a phosphor screen with energy storage capability as an X-ray image receptor. After exposure, a laser beam, which stimulates luminescence proportional to the local X-ray exposure, reads the screen. The luminescence signal is digitized. The data is then subjected to digital image processing. The Point-of-Care CR 120/140 enables the user to read a screen quickly, and then erase it to be ready for the next scan. The unit is compact and easy to use. Using the Point-of-Care CR 120/140 enables medical professionals to go digital without changing their work practices or X-ray equipment.
System Overview
System Components
The Point-of-Care CR 120/140 consists of 17 major assemblies, as well as sensors, which may be replaced in the field: USB Board Motion Board Sensor Board Photo Multiplier Assembly Optic Head Assembly Roller Motor Assembly Linear Slide Assembly Loader Stepper Motor Assembly Key Assembly and Tray Assembly Power Supply Assembly Cassette Lock Auto-loop Erase Lamps Assembly Erase Lamps Inverter Assembly Sensors: W0 Sensor Z0 Sensor Rollers Sensor Screen Size Sensors R-Limit (Home) Sensor L-Limit Sensor Loader Back Sensor Cassette Present
6H4866
SERVICE MANUAL
Power Supply Loader Stepper Motor Assembly Roller Motor Assembly Erase Lamps Assembly Erase Lamps Inverters Assembly Sensor Board Sensors
Eraser Lamps Sensor The erase lamp sensor detects an erase lamp failure. Left and Right Limit Sensors Roller Sensor The left and right limit sensors indicate the end of travel of the linear assembly movement. Detects when the screen enters and exits the rollers.
W0 Sensors Top and The W0 sensor determines the activation and deactivation of the rollers according to Bottom screen presence at the entrance to the drum. Z0 Sensor The Z0 sensor detects the presence of the screen in the drum. Screen Size Sensors Four screen size sensors determine the size of the CR screen by the location of the screen guide that is positioned by the screen as it is inserted into the drum. 15 x 30 Presence Sensor Installed in the key assembly, this sensor detects the presence of a 15 x 30 in. cassette adapter.
10
6H486603MAR11
Rotation Driver
Rotation Motor
Encoder
Laser
6H4866
1 2 3 4 5
PC
J501
Sensor Sensor
. . .
J502
P1
USB Board
J300 J305
CB090064
Sensor Board
J304 J500
. . .
J515
Sensor Sensor
J516
U211
J200
J208
J205 J211
Motion Board
J213 J203 J206 J201 J204 J212 J202
J1
C
J401
DC
SM
SM
J2
J4 J5
PM Board
Carrier - Laser
J403 J404 J402 Roller Motor Linear Load/Unload
Lamps Board
J1 J3
LP1-8
LP9-14
Solenoid
D
System Description
11
SERVICE MANUAL
Motor Driver
Linear Slide Assembly
Laser Board
Inverters
LED Panel
On/Off Switch
USB Board
12
6H486603MAR11
System Description
Roller Sensor
Laser Unit
Power Supply
Unit
6H4866
13
SERVICE MANUAL
Component Description
USB Board
The USB board receives operational commands from the host PC workstation via the USB port, and sends the commands to the appropriate unit component. It also transmits image data from the unit to the PC. 1 4 5 6 2
USB Board Connections Connector Connector 1 J7 Destination 5 V/12 V Power Supply Pin 1 Pin 3 Pin 4 Connector 2 J3 Connector 3 J1 Connector 4 J11 Green / Red Brown / Black White 5V GND 12 V
Flat cable to J1 on PM Board Flex cable to J200 on Motion Board Barcode Connection Pin 1 Pin 2 Red Black 5V Gnd
Connector 5 J6 Connector 6 J4
Not in use Barcode Connection Pin 1 Pin 2 Purple Brown Signal Signal
14
6H486603MAR11
System Description
Motion Board
The motion board has the following functions: Controls the loader and roller motors Controls the erase lamps Reads the unit sensors and passes the information to the USB board Motion Board Connectors 1 2 3 4 5 7
8 6 9 11 Connector Connector 1 J206 Destination Connectors Panel (Rollers Motor) Pin 1 Pin 2 orange pink 12 V GND 10 12
Rollers forward +12 V Rollers backwards -12 V Rollers stop 0 V Connector 2 J212 Connector 3 J213 Connector 4 J205 Connector 5 J203 Connector 6 J208 Connector 7 J202 Connector 8 J204 J212 on the Auto-loop Assembly 12 V Power Supply Pin 1 J401 on Laser Board J500 on Sensor Board J2 on Erase Lamps Inverters Assembly Loader carrier Orange + blue Red + yellow Connector 9 J201 Stepper Motor Yellow + green Brown + white Connector 10 J211 15 V Power Supply Pin 1 Pin 2 Connector 11 J215 5 V Power Supply Pin 1 Pin 2 Connector 12
6H4866
red
12 V
130 Hz load/unload 130 Hz load/unload 11 KHz moving left/right; 0 KHz stop 11 KHz moving left/right; 0 KHz stop black 1 black 2 brown white GND 15 V GND 5V
SERVICE MANUAL
Sensor Board
Sensor Board Connections The sensor board is a through board used to connect the sensors to the motion board.
J513 Cassette Presence Sensor J509 Roller Sensor J505 Sensor 1 on the Drum J501 Sensor 4 on the Drum J500 To J208 on Motion Board Typical Sensor Connections Connector
1 (Brown) 5V 2 (White) Signal 3 (Blue) GND
J514 Cassette Lock Sensor J510 J506 Sensor 2 on the Drum J502 Sensor 5 on the Drum
J515 Right Limit Sensor J511 W0 Sensor J507 Sensor 3 on the Drum J503
J516 Left Limit Sensor J512 15 x 30 Cassette Sensor J508 Loader Sensor in Back Position J504 Auto-loop Sensor
Sensor
4 Not in use
16
6H486603MAR11
System Description
Sensor Wiring Colors Sensor Sensor On (Flag in Sensor) Wire No. 1 2 3 4 Sensor Off (Flag not in sensor) 1 2 3 4 Measuring voltage: 1. All sensors except J511: Measure between pin 2 & 3 and insert an object between the sensors. 2. For J511 only: Measure between pin 2 and pin 3, and insert a screen to the system. The voltage should be between 4.5 and 5.0V. Wire Color Brown White Blue Not in use Brown White Blue Not in use 5 V always 5V GND always Signal 5 V always 0V GND always
PM Assembly
The photo multiplier (PM) tube collects the photons emitted from the screen. PM and PM Board
6H4866
17
SERVICE MANUAL
Connector 1 2
Destination J401 on Motion Board CB080008 J402 Encoder Reader CB090014 Pin 1 Pin 2 Pin 3 Pin 4 Red Black Blue Yellow Blue Brown Red Black GND Index Encoder 5V GND 12 V Rotation Motor - On Laser On 3.3 V GND 41 0.5 Hz 41 0.5 KHz
Laser Tube The laser tube emits the light to energize the phosphorus screen.
18
6H486603MAR11
System Description
Note
Measure the voltage when the rollers are turning at full speed. Connector Panel to the Roller Motor Pin 1 Orange Rollers forward +15V Rollers backwards -15 V Rollers stop 0 V GND
Pin 2
Brown
6H4866
19
SERVICE MANUAL
Connector Panel to the Slide Stepper Motor CB 090077 Pin 1 Pin 2 Pin 3 Pin 4 White Orange Yellow Black Phase A Phase A Phase B Phase B
20
6H486603MAR11
System Description
J204 on Motion Board Orange + blue Red + yellow Black + white 130 Hz load/unload 130 Hz load/unload not in use
6H4866
21
SERVICE MANUAL
Pin Number
System Type ACL 2/4 Source: AC input power CB090065 CB090137 Black 1 Black 2 CB090071 CB 090141_a Black 1 CB090064 CB090136-a Brown White CB 090064_b White Brown Green CB 090065_b Black 1 Black 3 Black 2 Black 4 PoC 120/140
Destination
Function
J211 on Motion Board 15 V GND 0 V J213 on Motion Board 12 V J215 on Motion Board 5V GND 0 V J305 on USB Board (power input) 12 V 5V GND 0 V Fuses Connector 15 V GND 15 V GND 0 V
Red
22
6H486603MAR11
System Description
6H4866
23
SERVICE MANUAL
Destination W0 Top to Connectors panel CB 090138 Pin 1 Pin 2 Brown White White Brown Green Yellow 5V GND 0 V 5V Lamp on 0 V Lamp off 2.5-5 V Lamp on 0 V Lamp off 2.5-5 V GND
3 4
24
6H486603MAR11
System Description
Inverter Assembly Voltages Connector 1 Destination J1 ON Lamp Sensor Board Pin 1 Pin 2 Pin 3 Pin 4 2 3 White Brown Green Yellow 5V Lamp on 0 V Lamp off 2.5 - 5 V Lamp on 0 V Lamp off 2.5 - 5 V GND
J2 to J202 on Motion Board P2 on 15 V power supply Pin 1 Pin 2 Red Black 15 V GND
6H4866
25
SERVICE MANUAL
Erase Lamp Sensor Voltages 1 Connected to Erase Lamp Board on the Inverter Assembly Pin 1 Pin 2 Pin 3 Pin 4 2 Not in use White Brown Green Yellow 5V Sensor 1 Lamp on 0 V Sensor 2 Lamp on 0 V Ground Lamp off 2.5 - 5 V Lamp off 2.5 - 5 V
26
6H486603MAR11
System Description
SERVICE MANUAL
Roller Sensor
The roller sensor detects when the screen enters and exits the rollers.
Z0 Sensor
The Z0 sensor detects the presence of the screen in the drum. Refer to Sensor Board on Page 16 for information on wiring connections.
Z0 Sensor
28
6H486603MAR11
System Description
WO Sensor
The W0 sensor determines the activation and deactivation of the rollers according to screen presence at the entrance to the drum. The sensor has two parts: the transmitter below the screen path, which is accessible from the bottom service panel, and the receiver above the screen path which is inside the drum. Refer to Sensor Board on Page 16 for information on wiring connections. Top WO Sensor
Bottom W0 Sensor
6H4866
29
SERVICE MANUAL
[5] Remove the 8 screws securing the service panel on the rear of the unit. [6] Remove the service panel.
30
6H486603MAR11
Service Procedures
Caution
To remove screws from the bottom of the unit, do not turn it on its back. Move the unit to the edge of the worktable to gain access from below. [1] Pull the scanner to the edge of the table so that one side is extended over the table edge. This allows you to gain access to the cover screws on that side. [2] Remove the screws on that side. Cover Screws
[3] Pull the scanner to another side of the table, and remove the screws on that side. [4] Repeat Step 3 for the remaining 2 sides. [5] Lift the cover off.
6H4866
31
SERVICE MANUAL
[2] Disconnect the laser connector from the laser board. Depress the latch of the connector firmly to release it. [3] Reconnect the laser only after all repairs are complete. [4] Reinstall the main cover.
32
6H486603MAR11
Service Procedures
Procedure
[1] Switch OFF the power. [2] Remove the power connector from the system. [3] Locate the fuse drawer on the power inlet module.
[4] Open the drawer using a flat screwdriver, prying gently from the plastic tab. [5] Remove the blown fuse. [6] Install a new fuse. [7] Close the fuse drawer. [8] Turn on the system. [9] Verify that the system functions properly.
6H4866
33
SERVICE MANUAL
Procedure
Laser Warning
To avoid possible eye damage, before performing this procedure, disconnect the laser as described in Laser Board Replacement on Page 42. For additional laser precautions, see Disconnecting the Laser on Page 32.
Important
Before performing this procedure: Switch off the scanner. Remove the power cord from the main socket. Detach the power cord from the back of the scanner. [1] Remove the back panel and the main cover. (See Removing the Service Panel and Scanner Cover on Page 30.)
[2] Open the nuts on the side of the module. [3] Open the nut on the ground stud. [4] Remove the line filter wire from the stud. Do not try to detach it from the power module. (See the illustration above.)
34
6H486603MAR11
Service Procedures
[5] Disconnect the black numbered wires from the line filter on the input module. [6] Remove the module and install the new one. Side View of Power Module Showing Connection Tabs
[7] Connect the black numbered wires to the connection tabs of the new module. [8] Connect the yellow ground wire to the ground stud. [9] Tighten the nut. [10] Secure the input module to its location with the two nuts. [11] Close the covers. [12] Connect the system to the main power. [13] Confirm that the system is operating properly.
6H4866
35
SERVICE MANUAL
[4] Pull the USB board out. [5] Disconnect the cables from the bottom of the board. [6] Slide the USB board out of the scanner.
36
6H486603MAR11
Service Procedures
Test Procedure
Laser Warning
When a service operation is taking place with the cover removed, disconnect the laser according to the procedure in Laser Board Replacement on Page 42. If the laser must be activated during the service procedure, wear protective safety glasses at all times. [1] Confirm that the USB board functions by performing the following test: (a) Make sure that the computer recognizes that a USB device has been connected. The USB device icon should appear in the system tray in the lower right corner of the screen. (b) Perform: a scan without the cover in place. The preview image should be black. a scan with the cover in place, without tightening the screws. The preview image should be white. (c) Tighten the screws. (d) Perform a scan with an actual image and check the image quality and position. (e) Open the About screen in Setup. (f) Check that the version number of the hardware and software are correct.
6H4866
37
SERVICE MANUAL
[3] Remove the two 2.5 mm Allen screws securing the motion board bracket. [4] Pull the motion board forward.
Test Procedure
[1] Confirm that the motion board functions properly by operating the system.
38
6H486603MAR11
Service Procedures
[2] Remove the electrical connectors from the sensor board. Sensor Board Screws
[3] Remove the four 2.5 mm Allen screws securing the sensor board.
Note
Each connector is marked with a number to indicate its location. [3] Install the service cover.
Test Procedure
[1] Confirm that the motion board functions properly by operating the system.
6H4866 39
SERVICE MANUAL
PM Ground Wire
[4] Disconnect the 3 mm ground wire connector to the slide body from the PM tube.
PM Board Connector
[5] Disconnect the flex cable from its connector at the bottom of the PM board.
40
6H486603MAR11
Service Procedures
[6] Remove the 4 mm retaining screws securing the PM assembly to the bracket.
6H4866
41
SERVICE MANUAL
Laser Warning
To avoid possible eye damage, before operating the unit without the cover, disconnect the laser (see Disconnecting the Laser on Page 32). For additional laser precautions see Laser Safety Instructions on Page 6. Only authorized personnel may remove the cover.
[4] Remove the three 3 mm Allen screws securing the laser board. [5] Remove the board.
42
6H486603MAR11
Service Procedures
Test Procedure
Laser Warning
If the laser must be activated during the service procedure, wear protective safety glasses at all times. [1] Perform the following test to confirm that the laser tube is functioning: (a) Check that the red line on the tube is in line with the red sign on the rotation motor bracket (b) Log in as Technician. (c) Select Setup>Diagnostics. (d) Perform a scan with the cover open. Make sure that the laser unit is functioning properly and that a steady beam is produced. [2] Perform the following test to confirm that the board is functioning properly. (a) Log in as Technician. (b) Select Setup>Diagnostics.
(c) Click On to activate the laser unit. (d) Click On to activate the rotational motor. (e) Make sure that the value 40 0.5 appears in the speed window.
6H4866
43
SERVICE MANUAL
Laser Warning
To avoid possible eye damage, before operating the unit without the cover, disconnect the laser (see Disconnecting the Laser on Page 32). For additional laser precautions see Laser Safety Instructions on Page 6. Only authorized personnel may remove the cover.
Cable Connectors
Shortcut Jumper
[3] Disconnect the cables from the driver motor board: (a) Disconnect the cable connectors using the 2 mm screwdriver. (b) Disconnect the orange flex cable connected to the rotation motor. [4] Remove the four 3 mm screws securing the driver board. [5] Remove the driver board.
44
6H486603MAR11
Service Procedures
[5] Remove the four 2 mm screws from the base of the roller motor and pull back the roller motor.
6H4866
45
SERVICE MANUAL
[6] Remove the drive adapter from between the motor drive and the roller drive. Motor Drive Retaining Screw
[7] Loosen the 3 mm retaining screw on the motor drive, and pull the roller motor out from the motor drive.
[1] Install: roller motor into the motor bracket roller motor drive completely into the roller motor [2] Tighten the 3 mm retaining screw. Drive Adapter
Alignment Channel
[3] Place the drive adapter on the roller drive. [4] Turn the roller drive to align the roller drive, adapter, and motor drive. [5] Attach the top 2 mm roller motor retaining screw to the top of the bracket. [6] Realign the roller drive. [7] Insert the remaining three 2 mm motor retaining screws and tighten them. [8] Reconnect the roller motor cable to the connector panel.
46 6H486603MAR11
Service Procedures
[9] Fasten the cable ties. [10] Perform the following test to confirm that the roller motor is functioning properly: (a) Select the Diagnostics tab.
(b) Operate the rollers forward and backward. (c) Load a cassette into the scanner. (d) Check that the scanner loads and unloads the screen correctly. [11] Install: Scanner cover Service panel
6H4866
47
SERVICE MANUAL
[5] Disconnect the laser board flex cable from the motion board.
48
6H486603MAR11
Service Procedures
[6] Release the two flex cables, until they can be taken out freely along with the slide assembly.
[7] Disconnect the slide stepper motor cable and the left limit sensor cable from the connector panel. Right Limit Sensor Connector
[8] Disconnect the right limit sensor cable from its connector. [9] Remove the connector from its bracket.
[10] Disconnect the ground wire from the slide by removing the screw.
6H4866
49
SERVICE MANUAL
Slide Assembly Screws [12] Remove the 4 screws securing the slide assembly to the system (two on each side). [13] Remove the slide assembly from the system by pulling it to the left until it is out of the drum.
Note
While pulling the slide assembly out, note the location of the flex cables. Slide Assembly
50
6H486603MAR11
Service Procedures
(b) Click the left and right arrows to move the slide from side to side. (c) Check that the optical head moves smoothly and that the wires are not stuck during movement. [7] Install: Scanner cover Service panel
6H4866
51
SERVICE MANUAL
[4] Disconnect connector J505 on the sensors board, making sure that the wire is free.
[5] Disconnect the stepper motor cable from the motion board J204 (loader) connector.
52
6H486603MAR11
Service Procedures
Bottom of Unit
[6] Place the unit on its back. [7] Open the door on the bottom.
Carriage
[9] Remove the 4 screws attaching the carriage assembly to the system. [10] Pull the carriage assembly out.
6H4866
53
SERVICE MANUAL
Coupling Screw [11] Loosen the coupling screw closest to the motor. [12] Remove the 4 screws that attach the motor to the loader assembly. [13] Carefully remove the motor from the assembly.
54
6H486603MAR11
Service Procedures
Ground Wires
Tools Required
Replacement power supply assembly 3 mm Allen wrench 7 mm socket wrench 2.5 x 100 mm cable ties Cutter
6H4866
55
SERVICE MANUAL
[4] Pull the system to the edge of the table. [5] Remove the 6 screws that attach the power supply assembly to the base of the unit.
56
6H486603MAR11
Service Procedures
J211 J215
J213
[7] Disconnect connectors J211, J213, and J215 from the motion board.
[8] Remove the 3.5 mm screw that attaches the ground wire to the slide.
SERVICE MANUAL
Fuse Connector
[10] Disconnect the fuse connector. [11] Remove the power supply assembly.
58
6H486603MAR11
Service Procedures
[4] Disconnect the erase lamp sensor cable from the inverter assembly.
6H4866
59
SERVICE MANUAL
[5] Disconnect the W0 upper sensor cable from the connector panel, and remove the connector from the panel.
Erase Lamp Assembly Screws [6] Remove the 3 screws (two on the left side, one on the right side) that attach the erase lamp assembly.
[7] Remove the erase lamp assembly through the drum front opening. The erase lamp assembly includes all of the wiring, as shown in the above image. [8] Reverse the above steps to install the erase lamp assembly. [9] Connect the W0 upper sensor cable to the connector panel. [10] Connect the erase lamp sensor cable to the inverter assembly. [11] Connect the lamp cables to the inverters. [12] Replace the slide. (See Linear Slide Assembly Replacement on Page 48.) [13] Replace the main cover. (See Removing the Service Panel and Scanner Cover on Page 30.)
60
6H486603MAR11
Service Procedures
[3] Disconnect the erase lamp connectors from the inverters. [4] Disconnect the cables from the erase lamp board.
[5] On the bottom of the unit, remove the screws that attach to the inverter assembly. [6] Remove the inverter assembly.
6H4866 61
SERVICE MANUAL
Laser Warning
When a service operation is taking place with the cover removed, disconnect the laser according to the procedure in Laser Board Replacement on Page 42. If the laser must be activated during the service procedure, wear protective safety glasses at all times. [7] Perform the following test: (a) Log in as Technician. (b) Select the Diagnostics tab.
(c) Check that the erase lamps are on. (d) Check that the indicator lamps are on. [8] Close the scanner cover.
62
6H486603MAR11
Service Procedures
Flex Cable Bracket Screws [2] Open the 5 screws that attach the flex cable bracket to the system.
6H4866
63
SERVICE MANUAL
Erase Lamp Sensor Screw [4] Remove the two screws securing each of the sensors. [5] Remove the sensor(s).
Laser Warning
When a service operation is taking place with the cover removed, disconnect the laser according to the procedure in Laser Board Replacement on Page 42. If the laser must be activated during the service procedure, wear protective safety glasses at all times. [4] Perform the following test: (a) Log in as Technician. (b) Select the Diagnostics tab.
(c) Check that the erase lamps are on. (d) Check that the indicator lamps are on. [5] Close the scanner cover.
64
6H486603MAR11
Service Procedures
[3] Disconnect the sensor cable from the connector panel. [4] Mark the location of the sensor.
[5] Remove the 2 screws attaching the sensor to the sensor bracket. [6] Remove the sensor.
6H4866
65
SERVICE MANUAL
66
6H486603MAR11
Service Procedures
[4] Mark the position of the sensor on the sensor bracket. [5] Disconnect the cable from the connector on the sensor bracket. [6] Remove the two 2 mm Allen screws. [7] Remove the sensor.
SERVICE MANUAL
[3] Locate the sensor and remove the two Allen screws. [4] Cut the cable ties securing the sensor cable, and follow the cable up to the sensor board. [5] Disconnect the cable from the sensor board.
68
6H486603MAR11
Service Procedures
(a) Make sure that the screen guide passes under the sensor without interference.
Screen Guide
(b) Select the Diagnostics tab. (c) Move the screen guide up and down and check that the Z0 light in the Diagnostics tab turns on and off.
Z0 Light
6H4866
69
SERVICE MANUAL
Roller Sensor
[3] Mark the position of the sensor. [4] Remove the 2 screws securing the sensor. [5] Remove the sensor. [6] Disconnect the sensor cable J509 from the sensor board.
70
6H486603MAR11
Service Procedures
Z0 Sensor Replacement
Procedure
[1] Disconnect the scanner from the main power. [2] Remove the service panel and scanner cover. (See Removing the Service Panel and Scanner Cover on Page 30.)
Z0 Sensor
[3] Locate the Z0 sensor on the back of the drum. [4] Disconnect the cable from sensor board J505. [5] Remove the two 2 mm retaining screws from the sensor. [6] Remove the sensor from the slide. [7] Attach the Z0 sensor on the back of the drum. [8] Loosely insert the two 2 mm retaining screws. [9] Slide the sensor up or down till the screen is in the middle of the bracket. [10] Reconnect the sensor connector to J505 on the sensor board.
6H4866
71
SERVICE MANUAL
Laser Warning
When a service operation is taking place with the cover removed, disconnect the laser according to the procedure in Laser Board Replacement on Page 42. If the laser must be activated during the service procedure, wear protective safety glasses at all times. [11] Perform the following tests:
Screen Guide
(a) Select the Diagnostics tab. (b) Move the screen guide up and down and check that the Z0 light in the Diagnostics tab turns on and off.
Z0 Light
72
6H486603MAR11
Service Procedures
W0 Sensor
The W0 sensor: has two partsthe receiver on top and the transmitter below identifies when a screen is inserted or ejected from the system
Tools Required
Replacement top receiver W0 sensor Replacement bottom transmitter W0 sensor 2.5 mm Allen wrench 2.5 x 100 mm cable ties Cutter
6H4866
73
SERVICE MANUAL
[3] Reveal the W0 sensor in the drum by moving the PM assembly to the right using the back of the linear motor. Sensor Connector
W0 Top Sensor
[4] From the inside of the drum, disconnect the cable from the W0 sensor. [5] Remove the two 3 mm retaining screws from the sensor. [6] Remove the sensor.
Laser Warning
When a service operation is taking place with the cover removed, disconnect the laser according to the procedure in Laser Board Replacement on Page 42. If the laser must be activated during the service procedure, wear protective safety glasses at all times. Black Probe Attachment Point
74
6H486603MAR11
Service Procedures
[3] Confirm that the top W0 sensor functions by the following test: (a) Connect a DVM as follows: Red probeConnect to the middle pin of the 511 connector on the sensor board. Black probeAttach to the yellow ground wire ground (see figure above). (b) Check that the initial voltage is zero and does not exceed 0.5V. (c) With the DVM still attached, insert a cassette into the scanner. (d) When the screen is between the two parts of the W0 sensor, the voltage should be 4.5-5.0V. [4] Install: Scanner cover Service panel
[3] Place the scanner on its back. [4] Slide out the bottom service access panel.
Sensor Connector
W0 Bottom Sensor
[5] From the left side of the stepper motor assembly, disconnect the cable from the W0 sensor. [6] Remove the 3 mm retaining screws from the W0 sensor on the stepper motor carriage. [7] Remove the W0 sensor.
6H4866 75
SERVICE MANUAL
Laser Warning
When a service operation is taking place with the cover removed, disconnect the laser according to the procedure in Laser Board Replacement on Page 42. If the laser must be activated during the service procedure, wear protective safety glasses at all times. Black Probe Attachment Point
[5] Confirm that the top W0 sensor functions by the following test: (a) Connect a DVM as follows: Red probeConnect to the middle pin of the 511 connector on the sensor board. Black probeAttach to the yellow ground wire ground (see figure above). (b) Check that the initial voltage is zero and does not exceed 0.5V. (c) With the DVM still attached, insert a cassette into the scanner. (d) When the screen is between the two parts of the W0 sensor, the voltage should be 4.5-5.0V. [6] Install: Scanner cover Service panel
76
6H486603MAR11
Service Procedures
Laser Warning
When a service operation is taking place with the cover removed, disconnect the laser according to the procedure in Disconnecting the Laser on Page 32. If the laser must be activated during the service procedure, wear protective safety glasses at all times. The required laser safety eyewear must be intended for HeliumNeon/PDT lasers, have an optical density of 4-5 wavelengths of 610-695 nm, and be marked as having CE approval.
Tools Required
Allen wrench set: 2.5, 3, 4 mm Open key: 7 mm Cutter Dynamometer
Note
When removing the top back roller bracket, make sure that you do not lose the spring. Top Back Roller Left Side View
Spring
[3] Remove the 2.5 mm Allen screw securing the spring on the right side of the system.
6H4866 77
SERVICE MANUAL
[4] Remove the 2.5 mm Allen screw securing the top back roller bracket. [5] Gently remove the top back roller located outside of the drum. Top Back Roller Right Side View
[7] Install the new roller to its components in the same order.
78
6H486603MAR11
Service Procedures
[5] Remove the USB board. (See Removing the USB Board on Page 36.) [6] Remove the four 2.5 mm Allen screws (two on the left and two on the right) securing the USB bracket. [7] Remove the 2.5 mm grounding screw located at the center of the bracket. Screws Securing the USB Board Bracket
[8] Loosen the 7 mm nut securing the roller belt tension pulley; do not remove it entirely.
6H4866
79
SERVICE MANUAL
[9] Lift up the wheel tension pulley. [10] Move the belt on top of the left wheel. [11] Roll the right wheel counter clockwise until the belt is released from the wheels.
Roller Belt
[12] Remove the wheel connected to the left side of the bottom back roller by unscrewing the 2.5 mm Allen screw.
80
6H486603MAR11
Service Procedures
Roller Screws
[14] Remove the power supply assembly from its location by removing the six 3 mm Allen screws. Power Supply Assembly
Note
There is no need to disconnect any cables.
6H4866 81
SERVICE MANUAL
[16] Remove the screw securing the cogwheel on the right side of the system.
Cogwheel Screw
[17] Remove the 3 screws securing the roller on its right side. Roller Right Side View
[18] Open the service panel (located on the bottom cover). [19] Remove the loader assembly. (See Removing the Loader Stepper Motor Carriage on Page 52.) [20] Remove the two 2.5 mm Allen screws securing the Mylar strip located around the bottom back roller.
82
6H486603MAR11
Service Procedures
[21] Slide the roller out from the left side. The roller is shown below.
[4] Slide the roller into the cogwheel and secure roller bracket. [5] Secure the roller bracket on the right side with its screws. (See Roller Right Side View on Page 82). [6] Secure the roller bracket on the left side with its screws.
6H4866
83
SERVICE MANUAL
[7] Carefully insert the Mylar strip around the roller until it is completely in position, as shown above. [8] Insert and secure the left cogwheel. [9] Insert the belt around the 3 cogwheels located on the left side of the system.
(a) Lift the belt on the left cogwheel. (b) Roll the right cogwheel clockwise until the belt is located around the cogwheels. (c) Check that the two cogwheels are properly aligned. If they are not aligned, adjust them by locating the left cogwheel. [10] Adjust the belt tension using a dynamometer. (a) Press down the dynamometer using 1 kg of force. The belt should not stretch more than 5-6 mm. (b) Close the cogwheel tension pulley nut screw after you receive the definition tension. (See Roller Belt Tension Pulley Nut Screw on Page 80.) Belt Tension Adjustment
[11] Replace the coupling on the left side. (See View of Outer Left Coupling on Page 79.) [12] Secure the Mylar strip around the roller. (See Screws Securing the Mylar Strip on Page 83.)
84
6H486603MAR11
Service Procedures
Bracket Spring
[4] Secure the left side of the top back roller with its 2.5 mm screw. [5] Connect the roller spring to the lower hole.
Spring Screw
[6] Replace the roller motor assembly. [7] Secure the roller motor assembly with its three 3 mm screws.
6H4866
85
SERVICE MANUAL
[8] Secure the roller spring located on the right side. [9] Connect the roller spring to the top hole. [10] Replace the power supply assembly. (See Power Supply Assembly on Page 81.)
86
6H486603MAR11
Service Procedures
Equipment Required
Screen Guide (SK25008) Screen Guide Testing Tool (PGTL) (TO09252)
Tools Required
Phillips screwdriver 2.5 mm Allen wrench
Flex Cable
6H4866
87
SERVICE MANUAL
[4] Remove the two Allen screws attaching the L bracket and securing the screen guide.
Flex Cable
[6] Remove: (a) 2 Allen screws securing the key assembly on the left (b) 2 Allen screws securing the key assembly on the right Removing the Key Assembly Screws
88
6H486603MAR11
Service Procedures
Caution
The screen guide is situated between the rollers. Pull the screen guide carefully toward you. [8] Remove the key assembly.
[9] Remove the 2 Allen screws attaching the screen guide bracket to the key assembly, releasing the screen guide.
6H4866
89
SERVICE MANUAL
Bracket
Screen Guide
U-Plate
[2] Place a piece of adhesive tape on the end of the screen guide.
[3] Install the key assembly while fastening the adhesive tape to the roller. [4] Rotate the rollers manually so that the screen guide is pulled underneath the rollers until it is visible from the upper hole.
90
6H486603MAR11
Service Procedures
[5] Place the screen guide in its location under the black bracket.
Note
There are two oval holes on the screen guide. Do not tighten the screws before performing Step 7. [6] Install the L bracket and the L bracket screws, positioning the screen guide. Installing the L Bracket
[7] Perform the screen guide adjustment procedure. (See Screen Guide Adjustment (Non-Auto-Loop) on Page 128.) [8] Insert the silver bracket and install the 2 Allen screws. [9] Install: (a) 2 Allen screws securing the key assembly on the left (b) 2 Allen screws securing the key assembly on the right [10] Connect the white flex cable back to the LED panel connector. [11] Install the scanner cover.
6H4866 91
SERVICE MANUAL
Tools Required
Allen wrenches Wire cutter
Flex Cable
[7] Disconnect: (a) Two 4-pin connectors (J512 and J504) from the sensor board (b) 2-pin connector (J212) from the motion board
Connectors
[8] Cut the cable ties securing the 3-grouped cable leading from the auto-loop key assembly to the right of the scanner to the motion board.
92 6H486603MAR11
Service Procedures
[9] Release the cable. [10] Remove the 4 screws securing the auto-loop key assembly: 2 on the left side and 2 on the right side.
Left Side [11] Remove the flex cable that is secured to the auto-loop assembly.
Right Side
Caution
Do not attempt to lift the auto-loop key assembly. Carefully pull it straight toward you. [12] Remove the auto-loop key assembly.
6H4866
93
SERVICE MANUAL
[8] Connect: (a) the 2-pin connector (J212) from the motion board (b) Two 4-pin connectors (J512 and J504) from the sensor board [9] Attach cable ties to secure the cable where possible. [10] Install the protective sponge strip on the top of the auto-loop key assembly. [11] Install: (a) scanner cover (b) scanner door
94
6H486603MAR11
Service Procedures
Tools Required
2.5 mm Allen wrench 2.5 x 75 mm flat screwdriver Long-nosed pliers
Solenoid bracket
Retaining screws
Solenoid
[2] Install the 2 M3 screws. [3] Attach connectors J212 and J504. [4] Install the auto-loop solenoid on the system. (See Installing the Auto-loop Key Assembly on Page 94.) [5] Perform the procedure described in Auto-loop Solenoid Adjustment on Page 123.
6H4866
95
SERVICE MANUAL
If you are not sure what configuration your scanner has, send the serial number of the scanner to Carestream Health Service to verify. The serial number is on the back of the scanner cover.
Tools Required
Phillips screwdriver Allen wrenches: 2.5, 3, 4 mm Flat screwdriver 2.5 x 75 mm Small cutter 2.5 x 100 mm cable ties
96
6H486603MAR11
Service Procedures
[3] Cut the cable ties on the sensor cables. [4] Turn the system on its back. [5] Cut the cable ties that hold the sensor cable.
Cable Ties
6H4866
97
SERVICE MANUAL
[2] Remove the two M3 screws attached to the black L-bracket securing the Mylar strip.
[3] Remove the black L-bracket. [4] Release the flex cable from the key assembly connector.
Flex Cable
98
6H486603MAR11
Service Procedures
[5] Remove the 4 screws securing the key assembly: 2 on the left side and 2 on the right side.
Left Side
Right Side
Caution
Do not attempt to lift the key assembly. Carefully pull it straight toward you. [6] Remove the key assembly.
6H4866
99
SERVICE MANUAL
[3] Cut the cable ties on the sensor cables. [4] Remove the cable from the cable clamps and set it beside the machine. [5] Remove the 4 screws securing the key assembly: 2 on the left side and 2 on the right side.
Left Side
Right Side
100
6H486603MAR11
Service Procedures
Caution
Do not attempt to lift the key assembly. Carefully pull it straight toward you. [6] Remove the key assembly.
6H4866
101
SERVICE MANUAL
[3] Remove the 4 M3 screws securing the plate sides: 2 on the left side and 2 on the right side.
[4] Remove the four 2.5 mm screws securing the tray assembly: 2 on the left side and 2 on the right side.
Caution
Do not attempt to lift the tray assembly. Carefully pull it straight toward you. [5] Remove the tray assembly.
102
6H486603MAR11
Service Procedures
6H4866
103
SERVICE MANUAL
Section 4: Calibrations
Nominal Calibration
The Nominal Calibration procedure is identical for KODAK QC Software versions 2.1.2, 2.4.2, 2.5, and 3.0. The GUI is different from one version to another. In this section, the GUI for software version 2.5 is shown. Although the screens for the other software versions look different, they contain exactly the same information and instructions.
Origin Calibration
Important
The machine is calibrated at the factory before it is shipped. Sometimes it is necessary to re-calibrate after a parts replacement procedure. [1] Enter the KODAK QC Software configuration screen. [2] Click the Calibration tab.
Note
A message with the exposure setting for the origin calibration appears.
104
6H486603MAR11
Calibrations
[6] Expose the cassette with the correct settings and insert it into the scanner.
Note
A message with the new X and Y coordinates will appear at the end of the scanning process.
Note
The calibration process ends. A green check mark should appear next to the Origin calibration field.
Offset Calibration
[1] Select Set Offset.
6H4866
105
SERVICE MANUAL
Note
A message instructing you to insert a cassette appears.
Note
The cassette will be erased. At the end of the erase process, a message instructs you to push the cassette into the scanner.
Note
The offset calibration ends. A green check mark should appear next to the offset calibration field.
106
6H486603MAR11
Calibrations
PM Nominal Calibration
Parts Needed: Description 0.5 mm Copper Sheet Quantity 1 Photo Catalog Number
1 mm Aluminum Sheet
Calibrated Dosimeter
14 x 17 in. CR Cassette
Preliminary X-ray Requirements The 14 x 17 in. exposed area must be verified for a uniform lateral exposure level of 15% at a dose of 1.4-1.6Gy. Pre-Calibration X-ray Setup - X-ray Dose Adjustment [1] Using the X-ray machine setup panel, verify on the X-ray monitor screen, the following parameters: Large focal spot SID=2.00m 80 kVp mAs as follows: KV 80 mAs 1.5 (150 mA, 10 ms)
6H4866
107
SERVICE MANUAL
The red frame indicates the focal point of the beam. The dosimeter should be in the beam center (cross).
[2] Slide the X-ray source head to a SID 2.00 m. [3] Move the table aside and clear the path for the X-ray beam from the source down to the floor to achieve the required SID. [4] Place the dosimeter on the floor, at the beam center. Placing the Copper and Aluminum Sheets Blocking the X-ray beam with the opper and aluminum sheets
[5] Place the copper and aluminum sheets at the X-ray collimator window.
Note
No filtration is used for RT calibration. [6] Execute an exposure and read the dosimeter readout. The value should be in the range of 1.4-1.6 g. [7] If the dosimeter readout is out of the required range, slide the X-ray head slightly above or below the nominal SID to achieve the required dose. [8] Remove the dosimeter detector when the dose requirement has been applied.
108
6H486603MAR11
Calibrations
Note
Origin-find and Offset-find calibrations must have been performed previously. (See Nominal Calibration on Page 104.) [1] Slide the copper and aluminum sheets out of the collimator window. [2] Place the cassette, loaded with a well-erased CR screen, on the floor, with its center located at the X-ray beam center. [3] Slide back the copper and aluminum sheets to cover the collimator window.
Note
The aluminum sheet must be facing down with the copper sheet above it, facing the X-ray tube. [4] Expose the cassette setup to X-ray radiation. [5] Enter the KODAK QC Software setup configuration screen. [6] Select the Calibration tab. [7] Select the Set PM check box.
6H4866
109
SERVICE MANUAL
Note
A message with the X-ray dose settings appears.
[9] Click OK. [10] Insert the cassette which has been exposed with the required dose into the scanner.
Note
When the calibration process has been completed successfully, a green check mark appears next to the PM nominal calibration field along with the date of the calibration procedure.
110
6H486603MAR11
Calibrations
Note
If the calibration was not successful, a red X appears next to the calibration field. Calibration Failed
[11] Make sure that the scanner is not located near a window or other strong light source and repeat the calibration process. [12] If the calibration still fails, contact Carestream Health Service.
6H4866
111
SERVICE MANUAL
Introduction
The unit parameter that controls the calibration of the image is called PM Gain. The PM Gain value is usually similar for all organs when a specific X-ray technique (e.g., 200/400 ASA) is used. However, the gain can be set separately for each sub-organ in order to achieve optimum image quality. The calibration process consists of several steps: Gain tuning: Finding a gain value for a specific X-ray technique (200-400 ASA) that can be used for thick and thin anatomy at the specific clinics settings. Fine tuning during clinical work: Tuning the gain value of a specific image type in order to obtain optimal image quality. Image enhancement: Fine tuning of the CONTEXT VISION filters to obtain optimal image processing.
Required Equipment
Software: KODAK QC Software v2.1 or later 14 x 17 in. general purpose cassette Filters: 0.5 mm copper and 1 mm aluminum Lead sheet for dose measurement
Setup
[1] In the KODAK QC Software window, click Setup. [2] Select the Anatomical tab. [3] In the organ box, select an organ. [4] In the Global PM Setting text box, type: 100 [5] Click Set.
[6] Select the Setup tab. [7] In the Image area, verify that the Data correction box is set to Linearization. [8] In the Diagnostics tab, select the Load image viewer after scan check box. [9] Click Apply. [10] Click OK.
112
6H486603MAR11
Calibrations
6H4866
113
SERVICE MANUAL
[7] When the scan is completed, the image is displayed in the Image Viewer screen.
[8] Select: Chest in the Organ box Chest AP in the Sub-Organ box [9] Verify: No Filter is selected in the Filter box Linearization is selected in the Data Correction box [10] Measure the pixel value on the center of the field. [11] Use the Simulate PM Gain field to change the panel pixel value on the center of the image to 2300 100. (a) To increase the pixel value, type a lower PM gain value in the Simulate PM GAIN text box. (b) To decrease the pixel value, type a higher PM gain value in the Simulate PM GAIN text box. [12] Click Process. [13] Measure the pixel value on the center of the image. [14] Continue changing the gain value in Step 10 to Step 13 until a pixel value of 2300 100 is measured. [15] Click the Floppy Disk icon to save. [16] In the pop-up box, check save PM and save to all organs. [17] Click Reject & Exit. [18] Perform an additional exposure and verify that the target pixel value of 2300 100 is measured. (a) If the pixel value is correct, click Reject & Exit to exit the viewer. (b) If the pixel value is not correct, repeat the tuning procedure, Step 8 to Step 17. [19] Enter Setup and select the Setup tab. (a) In the Image frame, verify that the Data Correction box is set to Linearization. (b) In the Diagnostics tab, uncheck the Load image viewer after scan check box. [20] Click Apply. [21] Click OK.
Note
If you have more than one X-ray, the calibration procedure should be done for each X-ray using the same method but under a different room. Four rooms are available.
114
6H486603MAR11
Calibrations
Note
The appearance of the Histogram window may vary depending on the software. Centered Histogram
[3] No fine tuning is necessary if the histogram is approximately centered. Histogram Too Much to the Left (Over Exposed)
- PM Gain
6H4866
115
SERVICE MANUAL
[4] Decrease the Gain value if the histogram is too much to the left.
[5] Change the Simulate PM Gain value (decrease the PM gain by 10). [6] Click Process until the histogram appears in the middle. [7] Click the Floppy Disk icon to save the new PM gain value directly to the sub-organ or to all organs. [8] Save the PM setting as described in Step 11 to Step 15 of Gain Tuning Procedure on Page 113. Histogram Too Much to the Right (Under Exposed)
+ PM Gain
116
6H486603MAR11
Calibrations
[9] Increase the Gain value if the histogram is too much to the right.
[10] Change the Simulate PM Gain value (increase the PM gain by 10). [11] Click Process until the histogram appears in the middle. [12] Click the Floppy Disk icon to save the new PM gain value directly to the sub-organ or to all organs. [13] Save the PM setting as described in Step 11 to Step 15 of Gain Tuning Procedure on Page 113.
6H4866
117
SERVICE MANUAL
[7] The image will appear in the Image Viewer screen. [8] Click the four way split screen icon to divide the screen into 4 identical images, in order to apply different filters to the each image and compare the results side by side.
118
6H486603MAR11
Calibrations
[9] Click an image (a red frame will appear around it) and select the desired filter from the Filter list.
[10] Click Process to apply the filter to the image. [11] Apply a different filter (3 options for each organ: light, medium, and sharp) to each image. [12] When you decide which filter gives the best results, select the image which has that filter applied to it. [13] Click the Floppy Disk icon to save the settings.
Note
The Save settings screen appears.
[14] Select Save Cv Filter. [15] Select a destination: Save to sub organ Save to organ Save to all organs [16] Click OK. [17] Repeat this procedure for all organs to achieve the best image quality with the most suitable filter. The filter chosen is saved only under the specific sub organ scanned. The filter chosen is saved for all sub organs under the organ scanned. The filter chosen is saved globally to all organs and sub organ.
6H4866
119
SERVICE MANUAL
Gain Tuning Calibration Process for KODAK QC Software Version 2.5 and 3.0
Introduction
This section describes how to calibrate the KODAK Point-of-Care CR 120/140 System with KODAK QC Software Version 2.5 and 3.0 to the X-ray machine used to achieve optimum image quality in the scanned images. The PoC scanners undergo a tuning process before leaving the factory, but should be finely tuned during installation in order to set the unit to the nominal point of operation. The unit is gain tuned using a 14 x 17 in. flat field 2mR exposure.
Origin Calibration
Important
Do not combine the two screen types (AGFA and KODAK PG-2) on the same reader. Ensure that you perform the Gain Tuning calibration procedure with the same screen as the one that will be used on the reader (e.g., do not perform a calibration for an AGFA screen on a KODAK screen and visa versa). Ensure that you use the same screen type on the reader as the one that with which you performed the calibration.
Equipment Required
14 x 17 in. general purpose cassette 0.5 mm copper plate 1 mm aluminum plate Dosimeter
[2] Adjust the SID to get a value of 2mR 0.1mR (17.4 Gy 0.4 Gy). [3] Place a 14 x 17 in. cassette in the center of the field. [4] Make an X-ray exposure.
120
6H486603MAR11
Calibrations
[5] Wait 5 minutes after exposing the cassette. [6] In the Settings>Setup tab, select: No correction Load image viewer after scan.
[7] Load the cassette and scan using the existing PM Gain value. [8] Measure the pixel value in the center of the field.
[9] Change the PM Gain to get a pixel value (APV) of 3200 50. To increase the Pixel Value, decrease the PM Gain To decrease the Pixel Value, increase the PM Gain
6H4866
121
SERVICE MANUAL
[10] Click on the icon to open the Save Settings dialog. [11] Select: Save PM Save to all organs [12] Click OK.
[13] Go to Settings>Anatomical tab. [14] Check that the PM values are set to the value that you saved in the above step. [15] If the PM Value is not updated, type the value in the Global PM Settings. [16] Click Set.
122
6H486603MAR11
Adjustments
Section 5: Adjustments
Auto-loop Solenoid Adjustment
Introduction
If an auto-loop solenoid has been installed in the system, it might require adjusting. The following error message appears if the mechanism needs adjusting.
The adjustment procedure comprises two sub-procedures which are intended to attain the correct functioning of the solenoid during its action: Electrical test (coil resistance) of the solenoid connector terminals: - If this test fails, replace the solenoid on the key assembly. - If the test passes, proceed to the solenoid adjustment procedure. Solenoid adjustment procedure.
Required Equipment
Digital voltmeter Two 50 x 21 x 0.4 mm plastic gap jigs
Adjustment Procedure
Electrical Test [1] Remove the scanner cover in order to reach the auto-loop connectors. Resistance Measuring Points on Auto-Loop Assembly
6H4866
123
SERVICE MANUAL
[2] Using the DVM check the solenoid resistance on sensor J212.
Note
The resistance value between the terminals should be 5.8 - 5.9 10% ohm. [3] Are the resistance values correct? Yes If the solenoid is not operating, go to Step 6. [4] Log in as Technician. [5] Select Settings>Diagnostics. [6] Measure the voltage peak during operation. The voltage should peak at 15VDC for approximately 1.5 seconds. Mechanical Adjustment Procedure [1] Loosen the solenoid holder screws on the key assembly. [2] Insert two (50 x 21 x 0.4 mm) plastic gap jigs between the pressure lever and the assembly wall (see below). Replace the solenoid. No
[3] Press the solenoid pin with a screwdriver so that it reaches its final stroke position. Facilitating the Final Stroke Position
[4] Adjust the position of the solenoid by moving the solenoid holder back and forth so that the pin presses the lever against the plastic gap jigs. [5] In that position, tighten the two M3 screws. [6] Remove the plastic gap jigs. [7] Verify the free movement of the pressure lever.
124
6H486603MAR11
Adjustments
Tools Required
SK Auto-loop Solenoid Height Gauge (SK000124)
[3] Slide the NO TOUCH end of the SK Auto-loop Solenoid Height Gauge up to the solenoid fork so that the prongs of the gauge slides under the solenoid fork.
Note
Shine a strong lamp into the flap so you can see the solenoid fork. If you do not have a lamp, turn on the erase lamps. (Select On under Erase lamp in the Technician Diagnostics tab.)
6H4866
125
SERVICE MANUAL
[4] Look closely at the gap between the Solenoid Height Gauge and the solenoid fork. [5] Log in as Technician. [6] In the Diagnostics tab, select On under Loop Solenoid. The solenoid pushes the fork down for about half a second.
Note
The illustration below does not show the scanner cover and key assembly. NO TOUCH End of the Solenoid Height Gauge in Place
[7] Does the solenoid fork touch the NO TOUCH end of the Solenoid Height Gauge? Yes The test has failed. [8] Click Off. No The test has passed.
126
6H486603MAR11
Adjustments
Note
Shine a strong lamp into the flap so you can see the solenoid fork. If you do not have a lamp, turn on the erase lamps. (Select On under Erase lamp in the Technician Diagnostics tab.) [2] Log in as Technician. [3] In the Diagnostics tab, select On under Loop Solenoid. The solenoid pushes the fork down for about half a second.
Note
The illustration below does not show the scanner cover and key assembly. TOUCH End of the Solenoid Height Gauge in Place
[4] Does the solenoid fork touch the TOUCH end of the Solenoid Height Gauge? Yes The test has passed. [5] Click Off. [6] Click OK. No The test has failed.
6H4866
127
SERVICE MANUAL
Tools Required
Screen Guide Testing Tool (PGTL) (SK000017) Screen Guide tool
Adjustment Procedure
[1] Remove the main cover. (See Removing the Service Panel and Scanner Cover on Page 30.) [2] Turn ON the system. [3] Select Settings>Diagnostics. [4] Select On under Erase lamp. [5] Go to Loader Control and click Backward << to move the loader to the end of its travel.
128
6H486603MAR11
Adjustments
[6] Check that the screen pusher is in backward position and that the loader pin is up.
[7] Disconnect the flex cable from the key assembly and move it aside to prevent damage.damage.
Flex Cable
6H4866
129
SERVICE MANUAL
[9] Loosen the two Allen screws attaching the L bracket and securing the screen guide. Do not remove the screws.
[10] Insert each side of the PGTL between the tray and the screen guide. Note the correct position of the screen guide: when inserted, the GO side of the PGTL touches the screen guide when inserted, the NO GO side does not touch the screen guide NO GO not touching Screen Guide GO touching Screen Guide
[11] If the screen guide is higher/lower than the height needed, move the screen guide under the L bracket to adjust the screen guide height in relation to the PGTL and then tighten the two L bracket Allen screws.
[12] Insert the silver bracket. [13] Install the 2 Allen screws. [14] Install the scanner cover.
130
6H486603MAR11
Adjustments
[3] Go to Loader Control and click Backward << to move the loader to the end of its travel.
Note
The Loader Back indicator lights. [4] Select On under Erase lamp. [5] Insert the Tray Gauge into the tray entrance and position the Go slot around the loader pin.
Note
The gauge should pass smoothly over the loader pin. This determines that the pin is not too high.
6H4866 131
SERVICE MANUAL
[6] Insert the Tray Gauge into the tray entrance and position the No Go slot next to loader pin.
Note
The gauge should not pass over the loader pin. This determines that the pin is not too low. [7] If the pin height is incorrect, remove the bottom access cover, loosen the motor frame retaining screws and adjust the frame position as necessary. (See Replacing the Stepper Motor on Page 54.) [8] Do Step 5 and Step 6 again.
[3] Insert the Trolley Roller Gauge into the tray entrance so that it will be between the pusher and the top roller.
Note
The Go side (0.3mm) should pass through smoothly. The No Go side (0.4 mm) should not pass through.
132
6H486603MAR11
Adjustments
[4] If the pusher height is incorrect, remove the bottom access cover, loosen the motor frame retaining screws and adjust the frame position as necessary. (See Replacing the Stepper Motor on Page 54.)
Note
The adjustment can be performed with or without the tray assembly installed. [5] Repeat Step 3 to verify the adjustment.
6H4866
133
SERVICE MANUAL
Adjustment Procedure
[1] Switch ON the system and wait for it to complete initialization. [2] Insert the 0.4 mm Roller Sensor Adjustment Screen into the drum.
Note
The center of the screen must be at the center of the screen guide.
[3] Turn the rollers manually at least 360 degrees so that the screen rolls between the rollers. [4] Check that sensor 509 is activated. The sensor must remain active while the rollers turn 360 degrees.
Note
The sensor must remain inactive while the rollers turn 360 degrees.
134
6H486603MAR11
Adjustments
[5] Loosen the adjusting screws locking bolt, and adjust the screw so that J509 sensor LED becomes inactive. [6] Turn the screw in the opposite direction until the sensor LED lights up (is inactive). Sensor LED Active Sensor LED Inactive
[7] Remove the 0.4 mm Roller Sensor Adjustment Screen. [8] Insert the 0.25 mm Roller Sensor Adjustment Screen. [9] Turn the rollers manually at least 360 degrees so that the screen roles between the rollers and check that Sensor 509 remains inactive.
Note
The sensor must remain inactive while the rollers turn 360 degrees. [10] Lock the adjusting screw. [11] Check that when the rollers turn 360 degrees, the sensor remains inactive. [12] Remove the 0.25 mm Roller Sensor Adjustment Screen. [13] Insert the 0.4 mm Roller Sensor Adjustment Screen into the drum. [14] Check that when the rollers turn 360 degrees, the sensor lights up. [15] Remove the screen. [16] Install the scanner cover and service door. [17] Connect the scanner to the PC. [18] Load and unload 14 x 17 in. and 8 x 10 in. cassettes a number of times and check that no error messages are received.
6H4866
135
Rotation Driver
Rotation Motor
Encoder
Laser
136
1 2 3 4 5
Main Schematics
SERVICE MANUAL
PC
J501
Sensor Sensor
. . .
J502
P1
USB Board
J300 J305
CB090064
Sensor Board
J304 J500
. . .
J515
Sensor Sensor
J516
U211
J200
J208
J205 J211
Motion Board
J213 J203 J206 J201 J204 J212 J202
J1
C
J401
DC
SM
SM
J2
J4 J5
PM Board
Carrier - Laser
J403 J404 J402 Roller Motor Linear Load/Unload
Lamps Board
J1 J3
LP1-8
LP9-14
Solenoid
D
6H486603MAR11
+5V
GND +12V
BLK2 BLK1
BRN WHT
1 2 3 4
J1 CONN 26PIN
USB1 P300
usb[4..1]
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26
B_MOTION_ADDRESS1 B_MOTION_ADDRESS0 B_MOTION_ADDRESS3 B_MOTION_ADDRESS2 B_MOTION_DATA_BUS1 B_MOTION_DATA_BUS0 B_MOTION_DATA_BUS3 B_MOTION_DATA_BUS2 B_MOTION_DATA_BUS5 B_MOTION_DATA_BUS4 B_MOTION_DATA_BUS7 B_MOTION_DATA_BUS6 Index Ch_A B_MAX2_OE B_LASER_LINE B_MOTION_RD B_MOTION_W R B_MAX2_CLRn B_SCAN_RATE
J213 CON2
1 2
1 2
2 1
1 2 3 4 5 6 7 8 9 10
GRN WHITE
+5V COM
BRN
COM +15V
RED
J213
J204
CON12
4 3 2 1
1 2 3 4 5 6 7 8 9 10 11 12
J200
J212
2 1
CB090008
SOLENOID
CB090149
RED
AS096105
YELLOW
1 2 3 4 5 6 7 8 9 10 11 12
CON12
1 2 3 4 5 6 7 8 9 10 11 12
GND GND GND LASER_ENABLE GND GND ROT_ENABLE ENCODER_IN INDEX_IN +5V +5V ROTATION_SUPPLY
Ready_Green Ready_Red
Lock_Green Lock_Red
PC
CB070010 CB090008
GND GND
6H4866
PAGE 6
4 1
FROM PS 12V
PAGE 6
CB090136_b
CB090141_a
BLK1
FROM PS 5V /12V
FROM PS 15V
CB090016
CB090137_a CB090136_a
FUSE1.6A
PAGE 6
FROM PS 5V /12V
PAGE 6
CON10-Flat
CB090058
U211
+5V +5V
TO LAMPS BOARD
PAGE 5
GND
1 2 3 4 5 6 7 8 9 10 11 12 13 14
J202 J206
B B A A ROLL MOTOR +
GRY
PAGE 3
CB090142
FUSE 0.8A
ROLL MOTOR 3
CB090034
M1 LOAD / UNLOAD ORANGE
MOTION CARD
B_MOTION_ADDRESS1 B_MOTION_ADDRESS0 B_MOTION_ADDRESS3 B_MOTION_ADDRESS2 B_MOTION_DATA_BUS1 B_MOTION_DATA_BUS0 B_MOTION_DATA_BUS3 B_MOTION_DATA_BUS2 B_MOTION_DATA_BUS5 B_MOTION_DATA_BUS4 B_MOTION_DATA_BUS7 B_MOTION_DATA_BUS6 Index Ch_A B_MAX2_OE B_LASER_LINE B_MOTION_RD B_MOTION_W R B_MAX2_CLRn B_SCAN_RATE
BLUE
J3 CONN 12PIN
PAGE 4
FLAX1 to PM BOARD
CB090017
GND GND GND
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26
J204
TO SENSOR BOARD
J203
J205
PAGE 3
CON12
CB090008
PAGE 4
PAGE 4
137
GND
+5V
SIG
Brown (1)
White (2)
Blue (3)
1[2..0]
1..3
CON4 CON4
4 3 2 1
4 3 2 1
CB090123
CB090033
1[2..0] 1..3
1 2 3 4
4 3 2 1
1 2 3 4
CB090122
1[2..0] 1..3
1 2 3 4
1 2 3 4
CB090120
1[2..0] 1..3
SENSOR_BOARD AS099000
1 2 3 4 1 2 3 4 1 2 3 4
1 2 3 4
1 2 3 4
1 2 3 4
1 2 3 4
1 2 3 4
1 2 3 4
Loader_in_Back_Position
1 2 3 4
1 2 3 4
1 2 3 4 5 6 7 8 9 10
450mm
300mm
CB090016
CB090049
CB090022
CB090021
PAGE 2
CB090144 CB090138
BRN W HT BROW N
550mm
1 2
1 2
BRN W HT
1[2..0]
D
1..3
1..3
1[2..0]
Z1
CB090024
W HT BRN
D
Z2
1[2..0]
1..3
Z4
138
1 2 3 4 5
SERVICE MANUAL
Sensors Schematics
1..3
1[2..0]
Example
COMNNECTION PANEL
CON2 PNK YEL CON2
ROOF
CB090146
RED RED
SOLENOID
1 2
1 2
CB090139
CON4
CB090147
BRN BLK BLU
1 2 3 4
1..3
1[2..0]
Cassette In Position
CB090149
J515 CON4 BRN W HT GRN CON4 J513 J514
1..3
1[2..0]
Roller Sensor
PAGE 2
FROM PS
CB090056_a CB090130
J J +15V GROUP 1 BLK1 COM GROUP 1 BLK2 +15V GROUP 2 BLK3 COM GROUP 2 BLK4
Z0
CB090023
1[2..0] 1..3
PAGE 6
J505 J506 J507
1 2 3 4 5 6
CON6
1 2 3 4 5 6
CON6
TO LAMP BOARD
PAGE 5
Z 3 (N/A)
J501
J502
J503
CB090076
1..3 1[2..0]
CB090143
J516
CON4
CON4
CB090121
C
1[2..0]
1..3
4 3 2 1
4 3 2 1
BLUE W HT BRN
Limit Switch_L
STEPPER LINEAR M2 ORANGE W HITE
C
J508 J511
CB090142
B B A A
CON2
4 3 2 1
BLACK
CB090077
J500
TO MOTION CARD
PAGE 2
PNK GRY
2 1
ORANGE
WO
6H486603MAR11
1 2 3 4 5
ENCODER OP500012
CB090014
U6
1 2 3 4
CB090029
L[1..2]
LASER
J404
OP080005
RED BLACK
1 2
1 2 3 4 5 6 7 8 9 10 11 12
GND GND GND LASER_ENABLE GND GND ROT_ENABLE ENCODER_IN INDEX_IN +5V +5V ROTATION_SUPPLY
BROWN BLUE
J403
CB090040
BROWN BLUE BLUE
1 2 3
1 2 3 4 5 6
ROTATION MOTOR
ROTATION DRIVER
MO080010
MR1
1 2 3 4
1 2 3 4
1 2 3 4 5 6 7 8 9
VEL ADJ
RED
1 2 3 4 5 6 7 8 9 10 11 12
VEL ADJ
RED
Ready_Green Ready_Red
CB080008
1 2 3 4 5 6 7 8 9 10 11 12
6H4866
1 2 3 4
139
CON5
J1
ENCODER
J402
CON12
LS13
LASER J401
J1
U3 CON12
CB090145
J1
PM BOARD AS087000
E2
3 2 1
U4
CON12
J1
CB080008
CB080008
PAGE 2
PAGE 2
PAGE 2
E 1 2 3 4
CN1
R8
CB090056_c
CON5 CON6 Inverter Lamp_5 Inverter Lamp_1 CON6 CON5
CN2
CN1
6 5 4 3 2 1
1 2 3 4 5
1 2 3 4 5
CN2
CN1
CN2
CN1
R9
CB090056_b
1 2 3 4 5
GROUP 2 W HITE GROUP 1 CON5 W HITE CON5 CON6 Inverter Lamp_6
1 2 3 4 5
1 2 3 4 5
CN2
1 2 3 4 5 6 7 8 9 10 11 12
1 2 3 4
+5V +5V
GND
140
2 3 4 5
SERVICE MANUAL
CB090056_c CB090056_a
CONNECTION PANEL
PAGE 3
GROUP 2
CB090056_d
INV1 LP2 INV5
GROUP 1
CB090070
6 5 4 3 2 1
LP1
CB090070 CB090056_b
LP10
R3
LP9
CB090056_d CB090070
INV2 INV6 RED BLK LP4
CB090070
LP12
6 5 4 3 2 1
LP3 Inverter Lamp_2 CON6
R2
B
6 5 4 3 2 1
LP11
J4
INV3
CB090070
LP6
CB090070
INV7
2 1
J3
R1
CON5 Inverter Lamp_3 CON6 LP5
LP14
1 2 3 4 5
6 5 4 3 2 1
R10
CON5 CON6 Inverter Lamp_7
6 5 4 3 2 1
1 2 3 4 5
LP13
INV4
LP8
J2 J1
CON4
1 2 3 4 5
CON5 Inverter Lamp_4 CON6
6 5 4 3 2 1
LP7
R7
CB090140
CB090058
1 2 3 4
CON4
J1
MOTION CARD
PAGE 2
6H486603MAR11
2
J1 FUSE 3.15A J2 J3
- 1 SRC001 - 1 SRC001
L N
CB090131
GND
CB090072
GRN/YEL
BLK1
BLK3
BLACK BLACK2
BLK2
BLK1
CN2
CN2
GRN/YEL
1 JD2
GRN/YEL
CB090073
CN1
GND
CN1
LINEAR W
YEL/GRN
GND
MPS-65-15
1
+15V
+15V
COM
+15V
+15V
COM
COM
COM
CHASSIS STUD
BLK4
MPS-65-15
BLK1
COM
BLK1
BLK2
BLK1
BLK2
JD1 1 -
PS STUD
CB090132
YEL/GRN
BLK1
BLK1
GND
CN2
+12V
+12V
COM
MPS-65-12
CN1 GND 1 2 GND N L BLK1 YEL/GRN
COM
BLK2
2
C
BLK1
CB090134
BLK1
L 3
4 5 1 2
BLK2 BLK2
GND
BRN
PS1 BLK1
CN2
WHT
GRN
BLK1
WHITE
BRN
+12V
MPT-45B
BLK2 BLK1 GND 1 2
CN1
3 4 5 1
YEL/GRN
BLK2 BLK2
CB090135
BLK2 MPS65-15
GND
YEL/GRN
-12V
COM
+5V
+5V
COM
6H4866
1 2 3 4 5
AC Input ~220/110 V
FN282-2-06
FUSE 3.15A On/Off Switch
BLK1 BLK2
1 2 3 4 5 6
CON6
1 2 3 4 5 6
CON6
Feder AC
CB090130 CB090130
YEL/GRN
CB090137_a
RED BLK
TO PAGE 3
CB090137_b
BLK1
BLACK
CB090141
CB090074
BLK1
CB090141_a
CB090136_b
CB090133
BLK1
TO PS STUD JD1
GND
3 4 5
1
141
CASSET CONTACTS
CB090074 CB090105
PS STUD
CB090059
CB090042
CB090036
142
B C D E
SERVICE MANUAL
Grounding Schematics
CB090104
CB090073
CHASSIS STUD
CB090103
CB090108
3
INVERTER BOX
RIGHT RT CONTACT
USB SHELF
LEFT RT
CONTACT
CB090100
USB STUD
CB090101
FLAT CABLE BRACKET WHITE FLEX CABLE
CB090008
BASE PLATE
2
BRIDGE
OPTICAL HEAD PM
6H486603MAR11
B
Preventive Maintenance
Note
It is recommended that the rollers be cleaned every 1000 cycles. (See the message that appears in version 3.0.) The cleaning device includes the following items: Cleaning tray Cleaning plate with adhesive strips covered with protective paper/a protective envelope The cleaning mechanism is affected by the auto-loop mechanism. After upgrading a system to auto-loop, you cannot use the old type of cleaning tray and cleaning plate. Use only the new set supplied with the upgrade kit. New - Compatible with Auto-Loop Cleaning Tray Old - Not Compatible with Auto-Loop
Cleaning Plate
Note
The area marked in red is the difference between the two cleaning plates. It is critical to work with the correct cleaning plate and tray.
6H4866 143
SERVICE MANUAL
[5] In the User tab, click Prepare. The following message appears:
[6] Insert the cleaning tray, making sure that it locks in place. [7] Click OK. The rollers begin to turn.
144
6H486603MAR11
Preventive Maintenance
[8] Remove the protective paper from the cleaning plate to expose the adhesive.
[9] Place the cleaning plate on the tray. Make sure the cleaning plate is placed in the correct direction, as specified on the plate.
[10] Hold the plate and push it slightly into the unit. It should go in almost entirely, with approximately a quarter of the screen remaining outside. The following message appears:
6H4866
145
SERVICE MANUAL
[13] Disconnect the cleaning tray by completely pulling out the knob underneath the front tray until it comes to a stop. The unit performs a reset (homing cycle).
Note
If the homing is not performed, turn the unit off and then back on. [14] Repeat the cleaning process two more times.
The back side of KODAK phosphor screens are dark and the label on the cassette is yellow. The back side of AGFA phosphor screens are white and the label on the cassette is grey.
146 6H486603MAR11
Preventive Maintenance
Caution
Do not clean KODAK screens with ADC AGFA screen cleaner! See General Cautions on Page 146. Additional Tools and Materials Required Extraction Tool (SK250056) Clean, soft, lint-free fabric gloves Non-abrasive, lint-free cloths Low-level disinfectant (See Disinfecting the Screen or Equipment on Page 150)
Caution
Screen damage can result from contact with certain materials used in facilities performing radiography. Contact with Isopropyl alcohol, hydrogen peroxide and other peroxides, citrus based cleaners, hand lotions and water-less hand sanitizers, as well as surfactants and lubricants may cause visible or hidden damage to the screen and could result in immediate or future image artifacts. AVOID CONTACT BETWEEN THESE MATERIALS AND KODAK POINT-OF-CARE PHOSPHOR SCREENS. Cleaning materials other than those recommended can contain chemicals that cause visible or hidden damage to the screen and could result in immediate or future image artifacts. Never use isopropyl alcohol (isopropanol, rubbing alcohol) to clean screens or cassettes.
6H4866
147
SERVICE MANUAL
Extraction Tool
[2] Pull the tool down until it enters the square slots in the back of the cassette. Extraction Tool
[3] Hold the cassette horizontally. Push the slider in the middle of the extraction tool towards the open edge of the cassette until the screen is visible. Extracting the Screen
[4] Carefully pull the screen out. [5] Pull up on the release lever to remove the tool.
148
6H486603MAR11
Preventive Maintenance
Caution
Prior to use, read and follow the instructions in the Material Safety Data Sheets (MSDS) for KODAK Screen Cleaner. Failing to clean the screens routinely with KODAK Screen Cleaner may cause the following problems: The screen may not move smoothly from the cassette into the reader, resulting in an error message. The screen may not fully return to the cassette at the end of the cycle, preventing the cassette from latching. No error message will appear. Artifacts appear on radiographs. Cleaning Method [1] Fold a non-abrasive, lint-free wipe or cloth.
Caution
DO NOT POUR THE SOLUTION DIRECTLY ONTO THE SCREEN. Excessive amounts of screen cleaner may damage the screen. [2] Dampen the cloth with a small amount of the screen cleaner. [3] Wipe the screen thoroughly, one section at a time. Fold the cloth to expose a fresh area for each section of the screen. [4] Apply pressure to remove persistent dirt, if necessary. [5] Wipe the screen thoroughly dry with a clean, dry, non-abrasive, lint-free cloth to remove residual dirt. DO NOT LEAVE THE SCREEN TO AIR DRY.
Caution
Never insert a storage phosphor screen into a cassette unless the cassette and screen are thoroughly dry.
6H4866
149
SERVICE MANUAL
Caution
Dilute bleach solutions may cause eye irritation and dry skin. Wash hands with soap and water following use. Prior to use, read and follow the instructions in the manufacturers Material Safety Data Sheets (MSDS). Disinfection Materials Dilute bleach solution (dilute 1 part of 5.25% sodium hypochlorite with 10 parts water). A commercially prepared equivalent solution of dilute bleach that does not contain any of the materials listed previously that can cause screen damage. Disinfection Method for Dilute Bleach Solution 1. Wipe with a soft, lint-free cloth thoroughly dampened (but not dripping) with dilute bleach. Contact the contaminated surface for up to 15 minutes for adequate disinfection. 2. Wipe the bleach residue with a soft, lint-free cloth dampened with water to remove. 3. Wipe dry with a soft, lint-free cloth. Allow the screen to air dry thoroughly. Disinfection Method for Commercially Prepared Equivalent If a commercially prepared equivalent solution of dilute bleach is selected, it should be used according to its manufacturers instructions.
Caution
Be sure that all external and internal surfaces are thoroughly dry before reassembling and returning screens and cassettes to use.
[3] Hold the cassette vertically and make sure that the screen is fully inserted into the cassette.
150
6H486603MAR11
Preventive Maintenance
Note
If you insert the screen incorrectly, calibration and functional failures may result: The screen may not move smoothly from the cassette into the reader, resulting in an error message. The screen may fall out of the cassette. An exposed image might not be processed.
6H4866
151
SERVICE MANUAL
Notes Added new HW information Added calibration processes Added Adjustment processes Revised all current data
15FEB08 18JUN08
146-151
6H4866_PoC_120_140_Service_15 Screen cleaning Feb2008 instructions update. Added Tray assembly replacement Screen cleaning instructions update
July 2008
6H4866-03
6H4866-03_PoC_120_140_Service- Erase Lamps assy Manual_July_08 Screen cleaning instructions update. 6H4866 Added Driver motor replacement Added Back Roller replacement Tray Assembly replacement update Auto-Loop adjustment verification
07AUG09
6H4866
03MAR11
6H4866
cover, 2, 8
6H4866.fm
152
6H486603MAR11
Publication History
6H4866
153
The Kodak trademark and trade dress are used under license from Kodak.