GenEon Lab Testing Reports

Table 1.
The virucidal efficacy of the solution generated by the provided Trio RX Unit (BCS1402294) unit
against Human herpesvirus 1 Virus. Test was conducted as per AOAC Official Method 961.02;
Germicidal Spray Products as Disinfectants (2005) and ASTM E1053 Standard Test Method for
Efficacy of Virucidal Agents Intended for Inanimate Environmental Surfaces. Free Available
Chlorine: 530ppm at pH 7.0. Viruses are enumerated by cell culture as Infectious Units (IU).

16026 University Oak, San Antonio, TX 78249

sales@geneontechnologies.com

866.217.0205

Table 2.

The bactericidal

efficacy

of the liquid

generated

by GenEons On-Site

Generators; Germicidal Spray Products as Disinfectants (2005) using a 30 second contact time
with a ppm of between 350-380.

Number of
Sprayed
Inoculated Slides
(number of
replicates tested)

Average
microorganism
cfu/mL
inoculated per
slide#

Average
cfu/mL
recovered from
each of slides
sprayed*

Bacteriophage
MS-2

2.4 x 105

< .91

>99.9996%

>5.4

Pseudomonas
aeruginosa

3.0 x 105

< .91

>99.9997%

>5.5

Microorganism

Percent
Reduction

Log10
reduction

BCS Laboratories, Inc. - Gainesville

4609 NW 6th Street, Ste A, Gainesville, Florida 32609


Tel. (352) 377-9272, Fax. (352) 377-5630
www.microbioservices.com
FL DOH Laboratory #E82924, EPA# FL01147

File: Geneon Human HIV -poliovirus virucidal efficacy BCS 1411090 December 5 2014.doc
This report shall not be reproduced, except in full, without the written consent of BCS Laboratories.

Table 3.
The virucidal efficacy of the solution generated by the provided Trio RX Unit (BCS1402294) unit against Human Poliovirus 1 (CHAT) in a
protein load inoculum. Test was conducted as per AOAC Official Method 961.02; Germicidal Spray Products as Disinfectants (2005) and
ASTM E1053 Standard Test Method for Efficacy of Virucidal Agents Intended for Inanimate Environmental Surfaces. Free Available Chlorine:
500ppm at pH 7.0. Viruses are enumerated as Infectious Units (IU). Poliovirus served as a model for Human Immunodeficiency Virus (HIV)
inactivation*

Microorganism

Number of
replicates
tested

Human Poliovirus 1
VR 1562

Contact
Time
(minutes)

Average iu/mL
Average iu/mL
Average
Percent
Recovered From
Recovered From
Percent
Reduction
Control Slides# Each Sprayed Slide*
Reduction

0.5

3.1 x 104

56

99.8%

46

99.9%

68

99.8%

16

99.95%

12

99.96%

12

99.96%

0.4

99.999%

1.2

99.996%

2.2

99.993%

BCS Laboratories, Inc. - Gainesville

4609 NW 6th Street, Ste A, Gainesville, Florida 32609


Tel. (352) 377-9272, Fax. (352) 377-5630
www.microbioservices.com
FL DOH Laboratory #E82924, EPA# FL01147

File: Geneon Human HIV -poliovirus virucidal efficacy BCS 1411090 December 5 2014.doc
This report shall not be reproduced, except in full, without the written consent of BCS Laboratories.

Log10
reduction

99.8%

2.7

99.96%

3.4

99.996%

4.6

Table 4. The virucidal efficacy of the solution generated by the provided Trio RX Unit (BCS1402294) unit against Bovine
Diarrhea Virus 1. Test was conducted as per AOAC Official Method 961.02; Germicidal Spray Products as Disinfectants (2005)
and ASTM E1053 Standard Test Method for Efficacy of Virucidal Agents Intended for Inanimate Environmental Surfaces. Free
Available Chlorine: 510 ppm at pH 6.77. Viruses are enumerated by cell culture as Infectious Units (IU).

Microorganism

Number of
replicates
tested

Bovine Diarrhea
Virus 1
VR-534

Contact
Average iu/mL
Average iu/mL
Average
Percent
Time
Recovered From
Recovered From
Percent
Reduction
(Seconds) Control Slides# Each Sprayed Slide*
Reduction

30

60

180

2.9 x 103

0.5

99.98%

0.8

99.97%

0.5

99.98%

<0.2

>99.994%

<0.2

>99.994%

<0.2

>99.994%

<0.2

>99.994%

<0.2

>99.994%

<0.2

>99.994%

BCS Laboratories, Inc. - Gainesville

4609 NW 6th Street, Ste A, Gainesville, Florida 32609


Tel. (352) 377-9272, Fax. (352) 377-5630
www.microbioservices.com
FL DOH Laboratory #E82924, EPA# FL01147

File: Geneon HSV-1, Enterovirus 69, and BDV virucidal efficacy BCS1402294 December 5 2014.doc
This report shall not be reproduced, except in full, without the written consent of BCS Laboratories.

Log10
reduction

99.98%

3.7

>99.994%

>4.2

>99.994%

>4.2

Table 5. The virucidal efficacy of Trio RX unit (BCS1411090) generated solution against Human Poliovirus 1 (CHAT)
suspension following a 30 second contact. Test was conducted as per ASTM E1052 Standard Test Method for
Efficacy of Antimicrobial Agents Against Viruses in Suspension. Free Available Chlorine: 1268ppm. Viruses are
enumerated as Infectious Units (IU). Poliovirus served as a model for Human Immunodeficiency Virus (HIV) inactivation*

Microorganism

Human
Poliovirus 1
ATCC VR 1562

Number of
replicates
tested

Contact Time
(seconds)

30

Average iu/mL
Average iu/mL
Recovered From Recovered From Each
Control Slides#
Sprayed Slide*

7.7 x 105

Percent
Reduction

< 1.0

> 99.9998%

< 1.0

> 99.9998%

< 1.0

> 99.9998%

BCS Laboratories, Inc. - Gainesville

4609 NW 6th Street, Ste A, Gainesville, Florida 32609


Tel. (352) 377-9272, Fax. (352) 377-5630
www.microbioservices.com
FL DOH Laboratory #E82924, EPA# FL01147

File: Geneon 30 second virucidal efficacy BCS 1411090 April 01 2015.doc


This report shall not be reproduced, except in full, without the written consent of BCS Laboratories.

Average
Percent
Reduction

Log10
reduction

> 99.9998%

> 5.6

Table 5. The virucidal efficacy of Trio RX unit (BCS1411090) generated solution against Murine Norovirus MNV-1 suspension
following a 30 second contact period. Test was conducted as per ASTM E1052 Standard Test Method for Efficacy of Antimicrobial
Agents Against Viruses in Suspension. Free Available Chlorine: 1224 ppm. Viruses are enumerated as Infectious Units (IU).
Norovirus served as a model for Hepatitis B virus*

Microorganism

Murine
Norovirus 1
ATCC PTA-5935

Number of
replicates
tested

Average iu/mL
Average iu/mL
Contact Time
Recovered From Recovered From
(seconds)
Control Slides# Each Sprayed Slide*

30

2.3 x 105

Percent
Reduction

0.2

99.99992%

0.5

99.9998%

1.1

99.99995%

BCS Laboratories, Inc. - Gainesville

4609 NW 6th Street, Ste A, Gainesville, Florida 32609


Tel. (352) 377-9272, Fax. (352) 377-5630
www.microbioservices.com
FL DOH Laboratory #E82924, EPA# FL01147

File: Geneon 30 second virucidal efficacy BCS 1411090 April 01 2015.doc


This report shall not be reproduced, except in full, without the written consent of BCS Laboratories.

Average
Percent
Reduction

Log10
reduction

99.9997%

5.6

Table 6. The virucidal efficacy of Trio RX unit (BCS1411090) generated solution against Bovine Diarrhea Virus suspension following
a 30 second contact. Test was conducted as per ASTM E1052 Standard Test Method for Efficacy of Antimicrobial Agents Against
Viruses in Suspension. Free Available Chlorine: 1264ppm. Viruses are enumerated as Infectious Units (IU). Bovine Diarrhea Virus
served as a surrogate for Hepatitis C virus as per the US EPA protocols*

Microorganism

Bovine Diarrhea
Virus
ATCC VR-534

Number of
replicates
tested

Average iu/mL
Average iu/mL
Contact Time
Recovered From Recovered From
(seconds)
Control Slides# Each Sprayed Slide*

30

7.3 x 104

Percent
Reduction

<0.2

>99.99997%

<0.2

>99.99997%

<0.2

>99.99997%

BCS Laboratories, Inc. - Gainesville

4609 NW 6th Street, Ste A, Gainesville, Florida 32609


Tel. (352) 377-9272, Fax. (352) 377-5630
www.microbioservices.com
FL DOH Laboratory #E82924, EPA# FL01147

File: Geneon 30 second virucidal efficacy BCS 1411090 April 01 2015.doc


This report shall not be reproduced, except in full, without the written consent of BCS Laboratories.

Average
Percent
Reduction

Log10
reduction

>99.9997%

>5.6

Table 7. The virucidal efficacy of Trio RX unit (BCS1411090) generated solution against Human Herpesvirus 1 (HSV)
suspension following a 30 second contact. Test was conducted as per ASTM E1052 Standard Test Method for Efficacy of Antimicrobial
Agents Against Viruses in Suspension. Free Available Chlorine: 1264ppm. Viruses are enumerated as Infectious Units (IU).

Microorganism

Human
Herpesvirus 1
ATCC VR-733

Number of
replicates
tested

Average iu/mL
Average iu/mL
Contact Time
Recovered From Recovered From
(seconds)
Control Slides# Each Sprayed Slide*

30

1.3 x 105

Percent
Reduction

<1.0

>99.99992%

<1.0

>99.99992%

<1.0

>99.99992%

BCS Laboratories, Inc. - Gainesville

4609 NW 6th Street, Ste A, Gainesville, Florida 32609


Tel. (352) 377-9272, Fax. (352) 377-5630
www.microbioservices.com
FL DOH Laboratory #E82924, EPA# FL01147

File: Geneon 30 second virucidal efficacy BCS 1411090 April 01 2015.doc


This report shall not be reproduced, except in full, without the written consent of BCS Laboratories.

Average
Percent
Reduction

Log10
reduction

>99.9992%

>5.1

Table 8. The mycobacteriocidal efficacy of the liquid generated by the provided Trio RX (BCS1403223)1 at 30 second
contact
time at 412 ppm.

Challenge

Average
Mycobacterium
terrea cfu/mL
inoculated per
slide#

Average cfu/mL
recovered from
each of slides
sprayed*

Slide 1

<0.45

Slide 2

<0.45

Slide 3

1.9 x 106

12.7

Slide 4

<0.45

Slide 5

<0.45

Cumulative
Percent
Reduction

Cumulative
Log10 reduction

> 99.9998%

> 5.8

BCS Laboratories, Inc. - Gainesville

4609 NW 6th Street, Ste A, Gainesville, Florida 32609


Tel. (352) 377-9272, Fax. (352) 377-5630
www.microbioservices.com
FL DOH Laboratory #E82924, EPA# FL01147

File: Trio RX Disinfectant 961 02 Mycobacterium terrea time kill report 06 26 2014 draft.doc
This report shall not be reproduced, except in full, without the written consent of BCS Laboratories.

Table 9. The mycobacteriocidal efficacy of the liquid generated by the provided Trio RX (BCS1403223)1 at 60 second contact time

Challenge

Average
Mycobacterium
terrea cfu/mL
inoculated per
slide#

Average cfu/mL
recovered from
each of slides
sprayed*

Slide 1

<0.45

Slide 2

<0.45

Slide 3

1.9 x 106

5.0

Slide 4

0.45

Slide 5

0.45

Cumulative
Percent
Reduction

Cumulative
Log10 reduction

> 99.9999%

> 6.2

BCS Laboratories, Inc. - Gainesville

4609 NW 6th Street, Ste A, Gainesville, Florida 32609


Tel. (352) 377-9272, Fax. (352) 377-5630
www.microbioservices.com
FL DOH Laboratory #E82924, EPA# FL01147

File: Trio RX Disinfectant 961 02 Mycobacterium terrea time kill report 06 26 2014 draft.doc
This report shall not be reproduced, except in full, without the written consent of BCS Laboratories.

Table 10. The mycobacteriocidal efficacy of the liquid generated by the provided Trio RX (BCS1403223)1 at 120 second contact time

Challenge

Average
Mycobacterium
terrea cfu/mL
inoculated per
slide#

Average cfu/mL
recovered from
each of slides
sprayed*

Slide 1

<0.45

Slide 2

<0.45

Slide 3

1.9 x 106

0.45

Slide 4

20.0

Slide 5

<0.45

Cumulative
Percent
Reduction

Cumulative
Log10 reduction

> 99.9998%

> 5.7

BCS Laboratories, Inc. - Gainesville

4609 NW 6th Street, Ste A, Gainesville, Florida 32609


Tel. (352) 377-9272, Fax. (352) 377-5630
www.microbioservices.com
FL DOH Laboratory #E82924, EPA# FL01147

File: Trio RX Disinfectant 961 02 Mycobacterium terrea time kill report 06 26 2014 draft.doc
This report shall not be reproduced, except in full, without the written consent of BCS Laboratories.

Table 11. The mycobacteriocidal efficacy of the liquid generated by the provided Trio RX (BCS1403223)1 at 180 second contact time

Challenge

Average
Mycobacterium
terrea cfu/mL
inoculated per
slide#

Average cfu/mL
recovered from
each of slides
sprayed*

Slide 1

<0.45

Slide 2

0.9

Slide 3

1.9 x 106

<0.45

Slide 4

NA**

Slide 5

NA**

Cumulative
Percent
Reduction

Cumulative
Log10 reduction

> 99.9999%

> 6.5

BCS Laboratories, Inc. - Gainesville

4609 NW 6th Street, Ste A, Gainesville, Florida 32609


Tel. (352) 377-9272, Fax. (352) 377-5630
www.microbioservices.com
FL DOH Laboratory #E82924, EPA# FL01147

File: Trio RX Disinfectant 961 02 Mycobacterium terrea time kill report 06 26 2014 draft.doc
This report shall not be reproduced, except in full, without the written consent of BCS Laboratories.