Sie sind auf Seite 1von 6

Installation & Operational Qualification

Myr Rotary Viscometer

START UP
Date
Performed by
Accepted by (customer)
Company
Location

ROTARY VISCOMETER
Type
V1H
V1R
V1L

V2H
V2R
V2L

Serial Number

Installation Qualification
SCOPE OF DELIVERY - PROTOCOLS
1. Scope of delivery according to the delivery note
Refer to the User Manual section supplied items to check that all parts of the viscometer are there. Check
the completeness of optional accessories according to the User Manual section accessories- .
complete

incomplete

If incomplete, missing parts:


Copy of delivery note supplied:

Yes

No. / date of delivery note:

2. Documentation
Viscometer User Manual

Document number

Calibration sheet

Fulfills specifications

Yes

No

CE symbol

Available

Yes

No

Viscotech Hispania, SL
Vidriers, 21 43700 El Vendrell - Spain
Tel: +34 977 66 80 20 Fax: +34 977 66 80 30
e-mail: viscotech@viscotech.es
Web: www.viscotech.es

Installation & Operational Qualification

Myr Rotary Viscometer

INSTALLATION
1. Voltage: 100-240 V/ 50/60 Hz
Explaining the power consumption: 0,2 A

Explaining the required fuses: 1 x T2A

2. Viscometer installation
Explaining the space required:
stable, vibration-free laboratory table. Dimensions of viscometer mounted on the stand:
350x300x550 mm (LxWxH). Weight of viscometer plus stand: approx. 10 kg

Explaining the ambient conditions according to the User Manual:


indoors use only, no direct sunlight, no draughts of air or ventilation. Room temperature: 10
C to 40 C . Humidity: max 80 % RH. Altitude: max. 2000 m above sea level

Explaining the cable connections

Viscometer set up ready for use

Explaining the recommended standard vessel (600 ml laboratory beaker)

Accessory/ies mounted if applicable

not applicable

Type of accessory:
Adapter for low viscosity materials (LCP)

Adapter for low viscosity materials by high temperature (AHT)

Adapter for low viscosity materials with embedded PT 100 (LCP/T)

Adapter for small sample volumes (APM)

TR spindle set

TL spindle set

Adapter for small sample volumes with PT 100 (APM/T)

TR spindle set

TL spindle set

Adapter for helicoidal movement

Thermal printer

Viscometer levelled

Visual inspection of viscometer


Visual inspection of standard
spindles
Visual inspection of accessories

OK

NOK

Faults:

OK

NOK

Faults:

OK

NOK

Faults:

Viscotech Hispania, SL
Vidriers, 21 43700 El Vendrell - Spain
Tel: +34 977 66 80 20 Fax: +34 977 66 80 30
e-mail: viscotech@viscotech.es
Web: www.viscotech.es

Installation & Operational Qualification

Myr Rotary Viscometer

3. Software installation
Installation successful

No software ordered not applicable

4. Funktional check of the entire equipment


Power up and
OK
check the viscometer display

NOK

Faults:

Communication with PC if
applicable

NOK

Faults:

OK

not applicable

Remarks:

Participants:

Name

Date

Signature

Name

Date

Signature

Name

Date

Signature

The equipment has been delivered as ordered and has been installed correctly.

Customer
:
Name / Signature

Date

Name / Signature

Date

Supplier:

Viscotech Hispania, SL
Vidriers, 21 43700 El Vendrell - Spain
Tel: +34 977 66 80 20 Fax: +34 977 66 80 30
e-mail: viscotech@viscotech.es
Web: www.viscotech.es

Installation & Operational Qualification

Myr Rotary Viscometer

Operational Qualification
HOW TO OPERATE THE ROTARY VISCOMETER
1. General Instruction
Pointing out the safety instructions according to the instruction handbook

Explaining the correct handling of the viscometer shaft and spindles

How to mount accessories

How to prepare a measurement, important rheological hints

Viscometer menu structure

Measurement setup (speed-spindle combination)

Measuring

2. Software instruction if applicable

Not applicable

Done

a) How to use the Viscosoft Basisc software

b) Viscosoft PLUS software

Structure

Creating a test

Evaluating a test

Results database

Viscotech Hispania, SL
Vidriers, 21 43700 El Vendrell - Spain
Tel: +34 977 66 80 20 Fax: +34 977 66 80 30
e-mail: viscotech@viscotech.es
Web: www.viscotech.es

Installation & Operational Qualification

Myr Rotary Viscometer

3. Measuring

Measurement No. 1
Sample
Measurement settings:
Speed
Spindle
Temperature
[mPas] or [cP]

RESULT:

Measurement accepted

Yes

No

Torque:

[%]

Torque:

[%]

Reason:

Measurement No. 2 (optional)


Sample
Measurement settings:
Speed
Spindle
Temperature
[mPas] or [cP]

RESULT:

Measurement accepted

Yes

No

Reason:

Remarks:

Viscotech Hispania, SL
Vidriers, 21 43700 El Vendrell - Spain
Tel: +34 977 66 80 20 Fax: +34 977 66 80 30
e-mail: viscotech@viscotech.es
Web: www.viscotech.es

Installation & Operational Qualification

Myr Rotary Viscometer

Participants:

Name

Date

Signature

Name

Date

Signature

Name

Date

Signature

Name

Date

Signature

The installed equipment fulfils the required specifications, any identified deviation or problems have
been removed.

Customer
:
Name

Date

Name

Date

Supplier:

Viscotech Hispania, SL
Vidriers, 21 43700 El Vendrell - Spain
Tel: +34 977 66 80 20 Fax: +34 977 66 80 30
e-mail: viscotech@viscotech.es
Web: www.viscotech.es

Das könnte Ihnen auch gefallen