Beruflich Dokumente
Kultur Dokumente
DL31/DL38 KF Titrators
Contents
Page
4 Setup ....................................................................................... 11
4.1 KF titrants ................................................................................... 12
4.2 Br No. titrants (DL38 only) ........................................................... 14
4.3 KF standards .............................................................................. 15
4.4 Main methods (DL38 only) .......................................................... 17
4.5 Peripherals ................................................................................. 18
4.6 Settings ...................................................................................... 21
5 Methods ................................................................................... 25
5.1 The KF standard method .............................................................. 26
5.2 KF method parameters ................................................................. 29
5.3 Storage procedures ...................................................................... 35
5.4 DL38: User and METTLER methods ............................................... 37
5.5 DL38: Bromine number method No. 908 ....................................... 38
DL31/DL38 1
2 Safety measures
The titrators have been tested for the experiments and intended
purposes documented in these Operating Instructions. However, this
does not absolve you from the responsibility of performing your own
tests of the products supplied by us regarding their suitability for the
methods and purposes you intend to use them for. You should there-
fore observe the following safety measures.
2 DL31/DL38
2.2 Measures for operational safety
DL31/DL38 3
3 The titrator
Titration stand
Burette
Keypad
Magnetic stirrer
Rear view
RS232C interface
for balance
RS232C
Attachment interface for
DV705 GA42 Printer
RS232C interface
Stirrer output for computer
Pump unit
Connector for On/off switch
power cable
4 DL31/DL38
3.1 Putting into operation
– Plug the power cable supplied into the connector of the titrator,
connect to the power supply and switch on the titrator: A self-test is
performed before the name of the titrator appears (example DL38).
,,,,,,,,,,
Display (mask))
,,,,,,,,,,
Arrow keys
,,,,,,,,,,
Reset
i
,,,,,,,,,,
,,,,,,,,,, 7
(
8
)
9
,,,,,,,,,,
METTLER TOLEDO
[ ]
,,,,,,,,,,
4 5 6
DL38
Entry
,,,,,,,,,,
Auxiliary Stirrer Burette
The command key <F3>, "Hello" (see Section 3.2.1), the i key and all
Menu and Auxiliary function keys can be activated after the titrator has
been switched on.
DL31/DL38 5
You can now also activate the arrow key (∇), Reset and Command
keys <F2>, Print and <F3>, OK.
▼: The arrow in the display means that the list contains more
resources. If you press the arrow key (∇), the lines are
scrolled upward, the selection bar is fixed. The commands
you can execute always refer to the line that is selected.
Print: Press <F2>: The list of KF titrants is printed out (if a printer
is selected and defined, see Section 4.5.1).
OK: Press <F3>: The list of KF titrants appears.
6 DL31/DL38
If a value can be modified or entered only with the keypad,
the "Modify" command is missing.
ABC This command always appears when you can select let-
ters or characters, e.g. for a name. In this case, you can
not enter numbers until you have confirmed ABC (press
<F1>).
DL31/DL38 7
The sample is determined with the aid of a method. All
Method methods necessary for the different analyses are stored in
the Method menu and can be changed. All resources
defined in the Setup menu are available for these meth-
ods.
DL31: Only one method can be stored!
The titration is performed in the Analysis menu. The meth-
Run od called up determines the analysis sequence.
Note
Each menu is subdivided into task areas (submenus)
which are designated menus, lists or masks in these
Operating Instructions.
You can switch the stirrer on or off and change the stirring
speed.
Stirrer
Pump
You can rinse the burette and the tip of the dispensing
tubing and add a specified volume.
Burette
8 DL31/DL38
You can view the result list of analyzed samples, perform
*
Results
recalculations, change the statistical evaluation of a se-
ries (DL38 only!) and print out additional reports.
Shift
With the Shift key, you can activate the characters above the
numbers.
Shift You can move the cursor to the right or left (only if
+
7 letters/characters are selected), see example:
Shift + User a
4
abcdefghijklm
Cursor nopqrstuvwxyz
ABCDEFGHIJKLM ▼
Help Select OK
DL31/DL38 9
3.3 Changing the language
All texts in the titrator are stored in English, German, French, Spanish
and Italian. If you have not run the Tutorial and understand one of the
available languages better than the one displayed, you should select
it:
Setup
Resources SETUP
KF titrants
Br No. titrants – Press the Shift and Arrow (∇)
KF standards ▼
once, then press only the Arrow
Print OK key to select "Settings".
Resources SETUP
Peripherals
Settings
OK – Press <F3>.
Settings SETUP
Language English
User list
– Press <F2> repeatedly until,
Date 20-08-1998 ▼
Modify OK e.g. "Español" appears.
Settings SETUP
10 DL31/DL38
4 Setup
In order to perform titrations, the titrator must know the titrants and
their concentration. It must know the settings of the attached instru-
ments to transfer the data. In this menu you define and store these
resources: You set up your titrator.
When you press the Setup key, the list of resources appears:
KF titrants
Br No. titrants (DL38 only)
KF standards
Main methods (DL38 only)
Peripherals
Settings
DL31/DL38 11
4.1 KF titrants
Name 1-comp 5
Conc.[mg/mL] 5.0000
Type 1 component
Burette 5 mL
Fill rate [%] 100
Batch
Check after [day] 0
Date 00-00-0000
Time 0:00
Name Press <F1>, "ABC" to change the name. The titrant with
its parameters will be stored under this name.
Conc. You can change the concentration or its range. You
should change the concentration only if you know the
correct value!
12 DL31/DL38
Type Select one of the preset titrant types: 1 component, 2
components, Pyridine and Others. The correct selection is
important (see " Control parameters", Section 5.2).
Burette Select the burette (volume) which is used to dispense the
titrant.
Batch Press <F1>, "ABC" to enter the batch number specified
on the bottle.
Storage procedure
The titrant with its changed parameters is stored when you confirm the
parameter list with <F3>, OK. If you press the Esc key, no changes
will be made.
New KF titrant
You add a titrant to the titrant list when you confirm the free line with
<F3>, OK . The parameters listed on page 12 always appear, only the
name of the titrant is missing.
– Change the parameters as described above.
NOTICE The name of the new titrant must differ from those
already stored, otherwise an existing titrant will be over-
written!
Note
In the DL31 you can store 10 titrants, in the DL38 20.
DL31/DL38 13
4.2 Br No. titrants (DL38 only)
With <F3>, "OK" the stored titrants for determination of the bromine
number appear:
Name Br-/BrO3-
Conc. [mol/L] 0.05
Burette 5 mL
Fill rate [%] 100
Batch
Name Press <F1>, "ABC" to change the name. The titrant and
its parameters will be stored under this name.
Burette Select the burette (volume) which is used to dispense the
titrant.
Batch Press <F1>, "ABC" to enter the batch number specified
on the bottle.
Storage procedure
The titrant with its changed parameters is stored when you confirm the
parameter list with <F3>, OK. If you press the Esc key, no changes
will be made.
14 DL31/DL38
New Br No. titrant
You add a new titrant to the titrant list when you confirm the free line
with <F3>, OK. The parameters listed on page 14 always appear,
only the name of the titrant is missing.
– Change the parameters as described above.
NOTICE The name of the new titrant must differ from the one
already stored, otherwise an existing titrant will be over-
written!
Note
You can store 5 titrants for the bromine number.
4.3 KF standards
With <F3>, "OK" the stored Karl Fischer standards appear which you
can use to determine the concentration of the KF titrants.
DL31/DL38 15
Name di-Na tartrate
H2O content 15.660
Unit %
Batch
Type Mass
Weight [g] 0.0500
Storage procedure
The standard with its changed parameters is stored when you confirm
the parameter list with <F3>, OK. If you press the Esc key, no
changes are made.
16 DL31/DL38
New KF standard
You can add a standard to the list of KF standards if you confirm the
free line with <F3>, OK. The parameters listed on page 16 always
appear, only the name of the standard is missing.
– Change the parameters as described above.
NOTICE The name of the new standard must differ from those
already stored, otherwise an existing standard will be
overwritten!
Notes
1. In the DL31 you can store 10 standards, in the DL38 20.
2. The Application brochure offers tips on sample preparation, sam-
pling and the optimum amount of sample (see Section 5.4).
F1: Free
You can select a method from the METTLER methods (or from the
user methods if available), e.g. "900 Standard 10":
DL31/DL38 17
With <F3>, "OK" the following appears:
F2: Free
F3: Hello!
4.5 Peripherals
With <F3>, "OK" the names of the instruments which you can attach to
the titrator via the RS232C interfaces, the Centronics interface, the
stirrer output and the DV705 connection appear. You must define
these and possibly change the preset settings.
Printer None
Balance None
Titrn stand Internal
Computer Not def.
4.5.1 Printer
18 DL31/DL38
Standard settings for the EPSON printer
Type LX/SX
Paper Fanfold
Format A4
Automatic feed No
4.5.2 Balance
Data to titrator
Titrator Balance Configuration: "Send Cont."
DL31/DL38 19
Bidirectional: On request for the weight, the balance transfers the
current weight value when ordered by the titrator.
Command from titrator
4.5.4 Computer
– If you attach a computer, you must
Defined
select "Defined" and possibly change
Not defined
the settings.
20 DL31/DL38
Baud rate Select 1200, 2400, 4800 or 9600.
Parity Select Even, Odd or None.
Data bits Select 7 or 8 bits.
Stop bits Select 1 or 2 bit(s).
Character Select ASCII or DL.
set Notice: If you select DL, you must specify 8 for the
number of data bits!
Start/end Select '{'/CR or STX/ETX.
Note
You will find additional information on the communication between
titrator and computer in the Operating !nstructions for the computer
interface.
4.6 Settings
You can or must select or define the following settings for the titrator:
Language English
User list
Date 26-08-1998
Time 10:20
Titrator ID
Access for Expert
Beep Yes
Solv. capac.[mL] 80
Pumping time [s] 60
DL31/DL38 21
User free line for the first or
list ... a new name
Date Day 26
Month 8
Year 1998
Format 31-08-1998
Time Hour 10
Minute 20
Format 24 h
Access for You can decide whether all persons who work with the
titrator should also have access to the Setup and Method
menus, in other words have the right to modify resources
or methods.
Expert: When the titrator is delivered, you have access to
all menus. To change this, select "Routine".
22 DL31/DL38
Routine: You can
• in the Setup menu only print out the defined resources
and view the list of settings.
• in the Method menu only print out the list of methods
(DL38 only) and each method.
• in the Results menu not perform any recalculation and
not change any statistical data.
• under the i key, select neither the Tutorial, the KF learn
titration nor the Hardware tests menu.
Reaccess for experts is described on the last page of
these Instructions.
DL31/DL38 23
Pumping The defined time is a safety criterion: The pump should
time shut off if you, e.g. leave the lab while it is running.
24 DL31/DL38
5 Methods
The titrator automatically performs analyses with the help of
predefined methods. On delivery of the instrument
in the DL31 the KF standard in the DL38 the KF standard
method specified by us is method specified by us and 10
stored. You can modify this METTLER methods are stored.
method and save it under a You can modify these methods
number. and save them as user meth-
ods.
Methods METHOD
Methods METHOD
KF standard method
User methods
METTLER methods
Help Modify KF standard method
Help Print OK
Method METHOD
Print: The list of methods you
No. 0 have developed is printed out.
of the method OK: The user methods with their
defined numbers and titles ap-
Help OK pear (example):
Print Modify
DL31/DL38 25
A complete titration method comprises sample preparation, mixing
and wait times, the actual titration, result calculation and a report.
These partial steps are defined in the titrator as parameters which
determine the progress of an analysis.
You modify a method
by changing its parameters.
You can use the KF standard method to develop your own method. In
this you are guided from one parameter to another in the specified
order. Exceptions are sample, control, termination and report param-
eters, which are shown as a group in a mask. In the DL38 case, this
also applies to the calculation parameters.
With the user and METTLER methods, all parameters appear as a list
in a mask. Exceptions are again the above-mentioned groups.
The last parameter of a method "Author of the method" always ap-
pears in its own mask (see Section 5.3).
Note
The method is always shown with the following default parameters.
The parameters which you can modify and those which are faded in or
out are listed in Section 5.2.
26 DL31/DL38
DL31 / DL38
No. 0
of the method
Title
of the method
Sample parameters
ID
Type Mass
Minimum [g] 0.0000
Maximum [g] 5.0000
Entry Before
Speed [%] 35
Mix time [s] 10
Titrant 2-comp 5
Volume [mL] 0.0000
for predispensing
Auto start No ....... DL38 only
Control parameters
Current [µA] 20
End point[mV] 100
∆Vmin [µL] 0.5
∆Vmax [µL] 8.0
∆Vmax factor [%] 30
Start Cautious
Termination param.
Max. time [s] 600
Drift stop Rel.
Drift [µg/min] 15
Vmax [mL] 10.000
Blank [µg] 0
DL31/DL38 27
Drift Online
Result % ....... DL31 only
Calculation 1 %
R1 = x [%]*f1
f1 = 1.0000
Unit %
Decimal places 2
Statistics Yes
Max. srel [%] 0.00
Calculation 2
R2 = No calculation
f2 = 1.0000
Unit DL38 only
Decimal places 2
Statistics Yes
Max. srel [%] 0.00
Calculation 3
R3 = No calculation
f3 = 1.0000
Unit
Decimal places 2
Statistics Yes
Max. srel [%] 0.00
Standby Ja
Report
Output Print.+Comp.
Type GLP
Author
of the method
28 DL31/DL38
5.2 KF method parameters
Sample parameters
ID You can enter an identification for the sample(s) now or
just before the titration for every sample ("Request").
Type Select mass, fixed mass, volume, fixed volume, pieces or
fixed pieces.
Weight appears when "Fixed mass" is selected.
Volume appears when "Fixed volume" is selected.
Number appears when "Fixed pieces" is selected.
Density appears when "Volume" or "Fixed volume" is selected.
Minimum appears when "Mass", "Volume" or "Pieces" is selected.
Maximum The unit depends on the type of sample: g, mL or pcs.
<F1>, "Help": For a fixed or variable amount of sample
you can have the titrator calculate and enter values if you
know the approximate water content of your sample.
Entry of the amount of sample before or during the actual
titration. "Entry" is faded out when "Fixed mass", "Fixed
volume" or "Fixed pieces" is selected.
DL38: If you select "Yes" for the parameter "Auto start"
(see next page), on storage "Before" is automatically set
to "During".
DL31/DL38 29
Speed
You should not set the stirring speed higher than 60% to prevent the
magnetic stirring bar "jumping".
Mix time
The time is an empirical value and can be entered specific to the
sample ("Request")..
Titrant
You can select from the titrants stored in the Setup menu.
Wait time
The parameter does not appear in the KF standard method if you have
entered "0" for the predispensing. (This parameter always appears in
the user methods.)
30 DL31/DL38
Control parameters
The following holds for all control parameters: You should not modify
the preset default values for the different reagent types until you have
evaluated the titration curve of your sample (see Application brochure,
Section 4).
Note
∆Vmin and ∆Vmax depend on the burette volume; the values listed
apply to the 5 mL burette.
In the Tutorial and KF standard method these default values are
automatically entered for the selected titrant and hence the type. This
does not apply to user methods: When you select a titrant whose type
differs from the previous one, you should also adapt the control
parameters.
Pol. current A Karl Fischer titration can be indicated by a positive or
negative polarization current. Both determinations result
in equivalent measured values. With the DL31/DL38 the
entry possibility is thus limited to positive values for polar-
ization current and end point.
Termination parameters
To terminate the titration you must define "Drift stop" (or "Delay") and
"Vmax" as a condition; the maximum time is an additional termination
condition. The titration is terminated when the first of the selected
conditions is met.
DL31/DL38 31
Drift stop Select Rel.(ative), Abs.(olute) or No.
No: Drift [µg/min] is faded out and Delay [s] superim-
posed.
Delay If you enter "0" for the delay time, the titration is terminat-
ed as soon as the end point is reached.
Vmax The maximum volume is intended as a safety precaution:
It avoids dispensing an unnecessarily large amount of
titrant if the titration is faulty.
Blank
For the entry of a correct value, you should perform a determination.
You can enter the value specific to the sample ("Request").
Drift
The drift value is needed for calculation of the result (see Section
6.2.3). You can enter a value in the method ("µg/min") or just before
every titration ("Request") or have it entered:
"Online": The moisture which enters the titration vessel is always
measured, evaluated and displayed in the standby mode.
"Determination": Some standard operating procedures require inser-
tion of the value of a drift determination. If you have performed a drift
determination, the result is stored under the i key and used in the
calculation (see Sections 6.2.3 and 8).
32 DL31/DL38
"mg or µg" / "% or ppm": With the following limit values, the result is
automatically set from one unit to the other:
mg µg % ppm
less than 0.1 99... less than 0.1 999...
0.1.. greater than 100 0.1.. greater than 1000
DL31/DL38 33
Calculation 2/Calculation 3 (DL38 only)
The available parameters are the same as for calculation 1. Instead of
a formula, you can select "No calculation". The following units are
missing: ppm (ext.extr), % (ext. extr.), ppm (ext. soln), % (ext. soln)
Note
You can not determine samples from an external extraction or dissolu-
tion in the routine mode as you must enter the amount of sample for
the extraction/dissolution and the amount of solvent under calculation
2 and 3 (see Section 4.6: Access for).
Standby
"Yes": After a titration, the titration vessel is kept anhydrous. We
recommend allowing the standby titration to run even during lengthy
pauses so that you can start the next titration at any time.
"No": After a titration the start mask of the Analysis menu reappears
(see Section 6.1).
"[min]": After the sample determination, a switch is made to the
standby mode and the subsequent rate of consumption is displayed
and printed out during the defined time (see Section 6.2.2).
Report
Output Select printer, computer or Print .+Comp. as output
device.
Type DL31: Select GLP, Short-form or None
DL38: Select GLP, GLP+curves, Short-form or None.
Curves can be printed out only on an EPSON-compati-
ble printer.
(GLP report: All data which are required by the rules of
Good Laboratory Practice)
34 DL31/DL38
Author of the method
The selection of your name documents that you are responsible for
the development of the method (see Section 4.6: User list).
DL31 / DL38
The following always appears as the last parameter of a method:
Method METHOD
Author
of the method The modified method is stored when
you confirm OK (examples):
Help Modify OK
DL31 DL38
Methods METHOD Methods METHOD
If you have not entered any or a wrong method No., the following
appears:
Error METHOD Method METHOD
DL31/DL38 35
DL38
If you have entered a method No. which already exists, the following
appears:
While you are modifying a user or a METTLER method, you can press
the Run key (the Setup and Auxiliary functions keys are blocked).
a. If you have not yet modified any parameter of the method, the start
mask of the Analysis menu appears (see Section 6.1).
b. If you have already modified a parameter of the method, the mask
"Author of the method" appears (see page 35). The method is
stored with OK and the start mask of the Analysis menu is dis-
played.
36 DL31/DL38
5.4 DL38: User and METTLER methods
You can print out the selected method with <F2>. You can delete user
methods, but not METTLER methods.
Methods METHOD Methods METHOD
OK
DL31/DL38 37
5.5 DL38: Bromine number method No. 908
38 DL31/DL38
Calculation 1
R1 = x[Br No.]*f1
f1 = 1.0000
Unit
Decimal places 2
Statistic Yes ....... see Section 5.2
Max. srel [%] 0.00 ....... see Section 5.2
Calculation 2 mL
R2 = x[mL]*f2
f2 = 1.0000
Unit mL
Decimal places 3
Statistics No
Max. srel [%] 0.00
Calculation 3 see Section 5.2
R3 = No calculation
f3 = 1.0000
Unit
Decimal places 2
Statistics Yes
Max. srel [%] 0.00
Report
Output Print.+Comp. see Section 5.2
Type GLP
Author
of the method ........see Section 5.2
Note
"Author of the method" only appears if you change the method.
DL31/DL38 39
Sample parameters
Type Select Mass, Fixed mass, Volume or Fixed volume
Weight appears when "Fixed mass" is selected.
Volume appears when "Fixed volume" is selected.
Density appears when "Volume" of "Fixed volume" is selected.
Minimum appear when "Mass" or "Volume" is selected. The unit
Maximum depends on the type of sample: g or mL.
Entry of the amount of sample before or during the actual
titration. "Entry" is faded out when "Fixed mass" or "Fixed
volume" is selected.
Calculation 1
R1 Result of the first calculation. The bromine number or
bromine index is calculated using the appropriate formu-
la by selection of "Br No." or Br index" (see Section 6.5).
mL: titrant consumption up to the end point .
min: duration of the titration
Unit Bromine number and bromine index are characteristics
without dimension: The bromine number is the calculated
number of grams of bromine which react with 100 g
sample, the bromine index is the calculated number of
milligrams of bromine which react with 100 g sample.
40 DL31/DL38
6 Analysis (Run)
You perform titrations with a selected method. The required data are
stored in the Method and Setup menu. You prepare your sample,
titrate and obtain the result.
While an analysis is running, all menus and auxiliary functions are
blocked.
Preparations
1. Press the Burette key. Rinse the burette and fill ensuring that
there are no air bubbles in the burette and tubing (see Section
7.2).
2. Press the Pump key and fill the titration vessel with approx. 40 mL
solvent (see Section 7.3).
When you press the "Run" key, the start mask appears:
DL31 DL38
Start RUN Start RUN
As method number
• "0" is entered if no method (DL31) or no user method (DL38) is
stored.
• in the DL31 case always that of the stored method is automatically
entered together with the title.
DL31/DL38 41
• in the DL38 case the number which you enter directly or accept it
with <F2> from the METTLER or user methods is entered.
Notes
1. You can select the user name. The name which was selected in
the analysis last performed remains stored.
2. If you have defined an expiry date for the titrant and this has been
passed, an appropriate message appears before the pretitration
starts. You should then first perform a concentration determina-
tion (see Section 6.3).
3. DL38: The start mask always shows the number of the method
with which an analysis was last performed, e.g. Method No. 22. If
you delete this, "Method not available" appears when you confirm
the mask with OK (see next point).
4. The titrator also accepts methods which are stored in the attached
computer: When you enter a number whose method is not stored
in the titrator, the method is requested from the computer (see
Operating Instructions of the com- Start RUN
puter interface). If it is available, the
pretitration is started immediately Data transfer...
following transfer; if it is not avail-
able, "Method not available" ap-
pears.
42 DL31/DL38
6.1.1 Menu sequence up to standby mode (KF methods)
DL31/DL38 (example)
Start RUN
Method No. 22
Title Coffee
Benutzer C. De Caro
Help Modify OK DL31: <F2>, Modify is missing.
The pretitration starts with OK: It is
PRETITRATION RUN always executed with the control pa-
rameters of the selected method.
The length of the solid bar reflects
the addition of the titrant. The verti-
cal bar represents the threshold for
Help
the transition to the standby mode.
STANDBY RUN
→: This arrow indicates that you can
Active method: 22 start a determination.
→ :Wait, drift still decreasing!
Drift [µg/min] 2→
DRIFT CONC. SAMPLE →:Wait, drift increasing again!
Note
If the pretitration lasts longer that 10 minutes, a message appears.
After confirmation with OK, a pretitration again starts. In such a case,
you should check whether, e.g. all connections of the titration stand
are tight.
DL31/DL38 43
6.1.2 Standby mode
Stirrer: You can change the speed. This is reported for the subse-
quent sample determination.
Results: The menu is not shown until you have performed the first
measurement or a result is available (see Section 7.4).
Method: You can modify all methods. With the active method you can
not modify the following parameters:
• Titrant
• Polarization current
• End point
The modified parameters are adopted for the next sample. For docu-
mentation conforming with GLP, you should print out the method
before you switch back to the Analysis menu.
44 DL31/DL38
6.2 Sample determination
ANALYSIS RUN
2. the masks for the sample entry if H2O exp. [%] 2.00
a variable amount of sample is Weight [g] 0
defined with "Entry: Before". Minimum [g] 0.3750
Maximum [g] 0.8750
Help Bal.* OK
3. the mask for the sample addition
ANALYSIS RUN
if a variable amount of sample is
defined with "Entry: During" (ex- Please add sample
ample: mass). 0.3750 g -
0.8750 g
Help OK
Notes
1. If you take longer than 5 minutes for the entry of samples or
sample-specific parameters, the titrator switches back to the
standby mode.
2. If you add a sample while the titrator is in the standby mode
without first pressing the Run key, the titrator will begin a pretitra-
tion to titrate all the water in the sample. Immediately pressing the
Reset key and thereafter pressing the Shift and Run keys simul-
taneously can prevent this. By doing this, the pretitration and
standby mode is skipped and one of the above masks appears. In
this manner your sample is not “lost”.
DL31/DL38 45
Sequence of the sample determination (with point 2)
ANALYSIS RUN
The titration is ended in accordance
Content: with the defined termination param-
eters. In the background the titrator
42.0012 mg switches to the standby mode or the
Sample**Consum Graph start mask .
46 DL31/DL38
ANALYSIS RUN
Consumpn: 1.2010 mL
Duration: 0:31 min
If you have not yet entered the
Sample** Drift Graph amount of sample, the parameter
"Weight [g]" appears.
Report RUN
Otherwise, the result is reported be-
Printing... fore it is displayed.
DL31: R = 1.98 %
Result RUN
Method 22
Sample 1
R1 = 1.98 %
OK
Note
If, after the sample addition, the elapsing mixing time is shown and no
amount of sample has been entered, you can also press <F1> "Sam-
ple" to enter the amount of sample (see titration masks).
DL31/DL38 47
6.2.1 Weight transfer from a balance
6.2.3 Calculations
Confirm
SAMPLE Requests Mix time Predisp.. Wait time Titration
Measured value acquisition
Sample
addition
48 DL31/DL38
DL38 (Auto start)
In the case of an automatic start of the titration, the titrator does not
start the time measurement until you have entered the sample.
Confirm
SAMPLE Requests Mix time Predisp.. Wait time Titration
Calculation formulae
DL31/DL38 49
6.2.4 Sample series with statistics (DL38 only)
Note
You can perform a series with maximum 99 samples. See Sections
7.4.1 and 7.4.5.
50 DL31/DL38
After, e.g. the third sample determination, the following appears:
Result RUN
Method 22
Sample 3
R1 = 1.91 % <F1>: You can switch between the
result and statistics displays.
x OK
Note
Statistics RUN The mean value is calculated after
Method 22 two samples, the absolute and rela-
R1 n = 3 tive standard deviation after three
x = 1.95 %
samples.
s = 0.021095 % ▼
Result OK
H2 O content ⋅ mass
CONC in mg/mL =
drift ⋅ duration
consumption –
conc. (old)
DL31/DL38 51
With <F2>, CONC appears:
CONC RUN You can select one of the standards
KF standard: stored in the Setup menu with <F2>
di-Na tartrate (see Section 4.3).
Help Modify OK
OK
This mask appears when "Entry: Be-
CONC RUN fore" is defined in the method. If "En-
di-Na tartrate: try: During" is defined, the sample
Weight [g] 0 addition mask appears immedi-
Proposal [g] 0.0500 ately..
* Appears only if a balance is defined (see
Help Bal.* OK Sections 4.5.2 and 6.2.1).
52 DL31/DL38
DL31
Each result is entered in the parameter list of the titrant, in other words
the second overwrites the first etc.
DL38
The first result is entered in the parameter list of the titrant and is
overwritten by the calculated mean value of two then three results
etc..
You must determine the drift when a value is required for the calcula-
tion of the water content which has been obtained by a titration (see
Section 5.2: Drift). With <F1> "DRIFT" appears:
DL31/DL38 53
6.5 Bromine number determination (DL38 only)
Start RUN
OK
54 DL31/DL38
Calculation formula
If you are unsure what parameters should be entered for your sample,
select the learn titration stored under the i key. The titrator calculates
the range for the amount of sample you need to enter as well as
control and termination parameters, stores these and prints them out
after the titration.
The learn titration is blocked when an analysis or auxiliary function is
running.
If you confirm "KF learn titration" with OK, the following appears:
Modify OK
DL31/DL38 55
Method 55 KF-L
STANDBY KF-L
Weight [g] 0
Help Bal.* OK
* Appears only if a balance is defined (see
Sections 4.5.2 and 6.2.1).
ANALYSIS KF-L
Add sample or
Appears if you have already entered
confirm addition:
1.2345 g the amount of sample. Otherwise,
"Please add sample" appears.
Help OK
56 DL31/DL38
Result KF-L After the end of the titration, the re-
sult and the "learned" parameters
R1 = 0.61% ! are printed out before the result is
Calculated parame- displayed. The result is always
ters are stored marked with an exclamation mark.
OK
The standby mask of the Analysis
STANDBY RUN menu with the active method
number appears so that you can de-
Active method: 55 termine the next sample with the
"learned" method.
Drift [µg/min] 2→
DL31/DL38 57
7 Auxiliary functions
With the aid of the Symbol keys you can execute functions which are
independent of a titration method, but which can support or expand it.
While an analysis is running, all auxiliary functions are blocked; in
standby mode (KF titrations) you can activate only the functions of the
Stirrer or Results key.
7.1 Stirrer
You can switch the stirrer on and off and change its speed.
Stirrer STIRRER Start and stop the stirring with <F3>.
Start: The stirrer stirs at the defined
Speed [%] 35 rate.
Start
58 DL31/DL38
7.2 Burette
Rate: With <F2> you can change the rate at which the burette should
be filled. For the two-component reagents we advise always setting
60%. They outgas easier than the one-component reagents.
Start: The titrator delivers 1 burette volume. The piston of the burette
moves below the zero position to ensure that air bubbles "hidden" in
the stopcock are siphoned up and then expelled.
Rinse tip
Start: 5% of the burette volume are delivered to eliminate any diffusion
of solvent in the dispensing tubing tip.
You should also place the dispensing tubing in a waste beaker during
this operation to ensure the titrated solvent is not "overtitrated".
Dispense
Start: You can dispense the displayed volume immediately.
DL31/DL38 59
7.3 Pump
Filling
– Push the dispensing tubing up to the 60 mL mark of the titration
vessel so that the solvent does not splash during filling. The solvent
starts to flow into the titration vessel as soon as a sufficiently high
vacuum which siphons up the solvent has built up in the titration
vessel.
– Add approx. 40 mL to the titration vessel (the platinum pins of the
electrode must be immersed in the solvent!), release the pump
button and press <F3>.
– Pull up the tubing.
Emptying
– Push the suction tubing to the base of the titration vessel so that the
solvent can be siphoned off.
– Pull up the tubing when the pump is running as far as it will go, then
press <F3>.
Notes
1. The pump switches off automatically when the pumping time
defined in the Setup menu is reached (see Section 4.6).
2. Each pumping operation resets the solvent capacity to the value
defined in the Setup menu (see Sections 4.6 and 8).
3. In the case of the pump of the attached DV705 Titration Stand,
you must press the lower arrow key during emptying and the
upper arrow key during siphoning up.
60 DL31/DL38
7.4 Results
DL31/DL38 61
DL38: You can perform a sample series with 99 samples. The results
of the last 60 samples remain stored, in other words the
results of sample 1 are eliminated when you start to titrate
sample 61.
Statistical data of the last 60 samples are always calculated.
Notes
1. As "Result" the sample whose determination was terminated with
the error message "No titration!" is also stored.
2. A result is also stored if you terminate a determination with Reset.
The result is calculated with the titrant volume consumed up to
that time.
7.4.2 Recalculation
62 DL31/DL38
Recalculn RESULTS If you change one of the values, you
Probe 1 can nullify it with "Undo". With <F1>
Drift [µg/min] 6 the changed values are always reset
Conc. [mg/mL] 4.9875 to the original values.
Blank [µg] 0▼
Undo OK
Notes
1. The values obtained in the determination always remain stored so
that you can also nullify changes later.
2. Under "All results" you can print out the recalculated results. Here,
the changed values and the original are reported.
3. DL38: The changed results are also taken into account for the
statistical data (see Section 7.4.5).
4. DL38: In KF conc. determinations the calculated result or its mean
value is not entered in the parameter list of the KF titrant if you
change a value, e.g. after three determinations. However, if you
perform a conc. determination after this change, the recalculated
result is taken into account in the statistics calculation.
DL31/DL38 63
7.4.3 Report
64 DL31/DL38
Example of a statistical evaluation of 3 samples:
Statistics RESULTS Using the result list (press <F1>) you
Method 11 decide what sample result is an
R1 n = 3 outlier and then delete it.
x = 0.77 %
▼
If you delete "Sample 1" with <F1> ,
s = 0.102194 %
Result Print OK "Undo" is superimposed to allow you
nullify the command. At the same
time, sample 1 is marked as "delet-
Statistics RESULTS
Method 11 ed".
Sample 1
R1 = 0.77 %
Sample 2 ▼ Confirm OK to show the following:
Delete OK (Example: Sample 1 is deleted)
As long as the result list of this sample series is stored, you can nullify
the deletion operation (see also Note in Section 7.4.2).
DL31/DL38 65
8 Information menu
The measured values and menus below are stored under the i key:
Solv. capac.[mL] 80
Drift [µg/min] 0
KF learn titration
Hello! (Tutorial)
Hardware tests
Solv. capac. The solvent capacity defined in the Setup menu is en-
tered as an initial value (see Section 4.6). After each
titration, the difference between the initial value and the
amount of titrant consumed is displayed. If "0" is reached
or exceeded, an error message appears. As soon as you
switch on the pump, the value defined in the Setup menu
is reentered.
In bromine number determinations, the solvent capacity
is set to "0".
Drift Here, the value of the drift determination last performed
is stored and automatically adopted for the drift compen-
sation if you have defined "Drift: Determination" as a
method parameter (see Section 5.2).
KF learn t. The course of this titration is described in Section 6.6..
Tutorial As you get to know the titrator step by step, the proce-
dure is not described.
Hardware The functions of this menu which you can use to test the
tests hardware of the titrator are described in Section 9.3.
66 DL31/DL38
9 Error messages and malfunctions
9.1 Error messages from the titrator
Here, the error messages which refer to this section or are not self-
explanatory are explained.
1. After the titrator has been switched on, the following errors can be
displayed:
• ERROR
• RAM x test ERROR(Q) "x" represents a number
• ERROR –> HW version
• ERROR –> DCB
• ERROR –> application
• No application found
The messages refer to a faulty hardware component.
– Press any key: The titrator attempts to continue the start proc-
ess; this may be successful, but does not guarantee 100%
functioning.
Measures
1. Note the displayed text.
2. Note the instrument number and software version of the titrator
(shown on every report).
3. Please then inform METTLER TOLEDO service of these points
and the attached peripheral units with the defined configuration.
DL31/DL38 67
2. If the following three displays appear, parts of the hardware may
be faulty.
ERROR: x
SYSTEM IS BLOCKED!
Please call METTLER "x": represents letters and/or
TOLEDO service numbers.
Measures
1. Note the text or error, e.g. "ERROR: 1000" and the operation
that triggered it.
2. Note the instrument number and software version of the titrator
(shown on every report).
4. Then switch off the titrator.
3. Please inform METTLER TOLEDO service of these points and
the attached peripheral units with the defined configuration.
3. If the following warning appears, you must press any key: The
titrator can continue to function.
You should regard the executed action as failed and repeat it.On
no account consider the result of a determination terminated in
this manner!
68 DL31/DL38
Warning: x
"x": represents a numberl
Press any key
to continue
If the warning appears frequently,
you should take the measures de-
scribed under point 2.
DL31/DL38 69
5. Method not stored!
The titrator can no longer store a method if the user data memory
is full.
Measure
1. Confirm the message.
2. Delete one or more methods from the memory or transfer to the
compute .
70 DL31/DL38
Error Possible cause Rectification
DL31/DL38 71
9.3 Checking hardware
The menu "Hardware tests" is stored under the i key and you can use
to check whether, e.g. the stirrer and pump of the titrator are function-
ing correctly. If you order the connector set for the interface check, you
can also test the interface and the TTL inputs and outputs of the
titrator.
Display test
5 displays appear in succession, 4 with different patterns, the last
without a pattern. This allows you to detect faulty pixels of the display.
The test mode automatically switches off after approx. 20 seconds.
Keypad test
When you press a key, a numeric code assigned to this key appears
which allows you to recognize a faulty key (see following Table).
– Press the Esc key to quit the test mode.
Key codes
Key Code Key Code Key Code
Reset .......... 23 F1 ............ 17 3 ................ 3
i ................. 30 F2 ............ 18 4 ................ 4
Stirrer .......... 28 F3 ............ 19 5 ................ 5
Pump .......... 27 ∇ .............. 11 6 ................ 6
Burette ........ 29 ∆ .............. 12 7 ................ 7
Results ....... 26 Shift ......... 10 8 ................ 8
Setup .......... 25 0 ................. 0 9 ................ 9
Method ....... 24 1 ................. 1 Del ........... 31
Run ............. 21 2 ................. 2 . ............... 33
72 DL31/DL38
Beep tests
You test the internal beeper.
Interface test
– Before the test
• plug the test socket 51107288 into the balance, computer or GA42
interface.
• plug the Centronics cable into the Centronics interface and the test
plug 51107287 into the Centronics socket of the cable.
After the test phase, either "Test successful" or "Test failed" appears.
Balance test
The menu option is faded out if no balance is defined in the Setup
menu.
– The weight displayed on the balance must appear in the titrator
display.
DL31/DL38 73
9.3.3 Inputs and outputs
Internal stirrer
You test the rotational movement and change in the speed of the
magnetic stirring bar.
External stirrer
You test the rotational movement and change in the speed of the
magnetic stirring bar with the attached DV705 Titration Stand.
Internal pump
You check whether the pump functions:
1. A "noise" must be heard.
2. It must be able to dispense and siphon off when solvent and waste
bottle are correctly attached.
External pump
You check whether the pump of the attached DV705 Titration Stand
functions (see "Internal pump").
3. For dispensing, you must always press the lower button (⇓ arrow),
for siphoning off the upper button (⇑ arrow).
74 DL31/DL38
10 Installation and maintenance
10.1 Equipping the titration stand
On delivery, the titration vessel comprises O-ring, cover and threaded
ring. The seal is fastened in the titration stand. If you have to change
the cover or replace the O-ring, the illustration will serve as an aid. The
titration arm can be turned in both directions.
Silicon tubing (133 mm) 1. Carefully slide the magnetic stir-
ring bar supplied into the titration
vessel.
2. Insert the titration vessel with the
lug of the threaded ring into the
Threaded ring middle groove of the titration arm
with lug
and push downward towards the
Connector stirrer. Fasten the screw.
Stopper ST 7.5
3. Place the three-hole adapter, the
3-hole adapter
ST stopper and the connector in
O-ring
the cover openings.
4. Fill a drying tube with molecular
Titration vessel sieves, close with the cover and
"push" into the titration stand to
Cover ensure tightness. Connect the
and dry- cover and the connector by sili-
ing tube
Seal
con tubing.
Screw
Groove of
titration arm
Titration stand
DL31/DL38 75
10.2 Installing the burette
1. Screw the burette with holder onto the titrant bottle, see Section 2
in the DV100x operating instructions.
WARNING Never inhale the vapors of Karl Fischer reagents
and avoid skin contact. The reagents are toxic.
2. Place the drying tube holder in the opening of the burette holder,
fill a drying tube with molecular sieves and position in the holder.
Hole for 3. Push the burette onto the titrator:
dispensing
Driver cam tubing The slot of the driver cam and piston
rod must be parallel to the insert. (If
this is not the case, press the Reset
key.)
4. Turn arrestment knob in the oppo-
site direction to the arrow and when
pushing in check whether the piston
rod of the burette is correctly posi-
tioned in the push rod.
5. Turn the arresting knob in the direc-
Push rod of the Arrestment
burette drive knob tion of the arrow to fasten the bu-
rette.
6. Place the dispensing tubing in the
cover opening.
NOTICE When you rinse burette and tubing, place the dispensing
tubing in a waste beaker; do not dispense in the titration
vessel!
76 DL31/DL38
10.3 Installing the solvent and waste bottles
DL31/DL38 77
7. Push the silicon tubing (450 mm)
over the drying tube cover of the
waste bottle and connect to the suc-
tion fitting of the pump.
8. You can connect the output fitting to
tubing to lead off the solvent vapor to
a suitable place.
Output Suction
fitting fitting
Check all drying tubes and screw fittings once again for firm seating to
assure the tightness of the system!
RS232C interface
for balance
RS232C
DV705 interface for
connection GA42 printer
RS232C interface
Stirrer output for computer
TTL-I/O interface
Electrode
input (KF) Centronics interface
for EPSON-com-
patible printers
Pump unit
78 DL31/DL38
1. Magnetic stirrer
On delivery, the magnetic stirrer
is pushed into the slot for the
Magnetic
burette drive. Attach its cable to stirrer
the stirrer output.
2. Electrode
Unscrew the protective cover of the DM143-SC and connect the
electrode to the triaxial cable. Plug the cable into the KF input and
insert the electrode in the corresponding cover opening of the
titration vessel.
DL31/DL38 79
10.5 Using TTL inputs and outputs (DL38 only)
The TTL-I/O socket has two inputs and four outputs ((In 1, In 2 and
Out 1...Out 4). The outputs are occupied by the titrator for commands.
Inputs
A level change from "high" to "low" is detected at the inputs.
In 1 In the Run menu
• the pretitration in KF titrations or the titration in bromine
determinations is started,
• the sample determination in KF titrations is started (al-
ways corresponds to the <F3> key),
• the actual analysis is started (corresponds to confirming
the sample addition),
• the result display is confirmed.
In 2 corresponds to the Reset command on the keypad.
Outputs
At output 1 a pulse is given with a length of 100-200 ms. The other
outputs assume either an active ("low") or inactive ("high") condition.
Out 1 outputs a pulse before the display of "Add sample" and
before display of the result list.
Out 2 is inactive when an analysis can be started.
Out 3: is active in the case of an error message; the condition does
not change until the message has been confirmed with OK.
Out 4 is active only during the mix time defined in the method.
For this condition, the connection of the TBox DR42 is
suitable, it can be used to control an homogenizer (see
TBox Operating Instructions).
80 DL31/DL38
10.6 Maintenance and servicing
10.6.1 Cleaning
– You can clean the titrator with a cloth moistened with alcohol.
– To clean the titration vessel, turn the titration arm to the right and
unscrew the vessel from the cover.
– To clean the cover, loosen the fastener screw, take the titration
vessel with cover up and out and unscrew cover.
– You should clean the electrode when their response time deterio-
rates owing to certain Karl Fischer reagents and samples (the end
point drifts!): Either place in an ultrasound bath with deionized water
for a few minutes or for 60 seconds in a chromic acid bath and then
rinse with water or ethanol.
10.6.2 O-rings
You should replace the O-rings of the screw caps when the system is
no longer tight (see Section 11.2):
– Loosen the screw sleeve, take out the pump tubing and remove the
O-ring using tweezers (see also illustration in Section 10.3).
You should replace the O-ring of the threaded ring for the same
reason:
– Loosen the fastener screw of the titration vessel, take titration
vessel with cover up and out and unscrew (see illustration in
Section 10.1).
DL31/DL38 81
10.6.4 Molecular sieves
If the drift of the KF solution remains too high, you should fill the drying
tubes with fresh molecular sieves.
You can regenerate molecular sieves in a drying oven at 160 - 300 °C
for at least 24 hours. As the molecular sieves used on the waste bottle
also contain SO2 , you should rinse them with deionized water before
regeneration.
10.6.8 Battery
The battery installed in the titrator has a service life of approx. 5 years.
The battery may be changed only by METTLER TOLEDO service. (See
also Section 9.1, point 4.)
82 DL31/DL38
10.6.9 Disposal
Contact your METTLER TOLEDO dealer if you have to dispose of the
titrator.
DL31/DL38 83
11 Standard and optional equipment
11.1 Standard equipment
You can order each part listed with an order number from METTLER
TOLEDO.
Note
The standard and optional equipment for the 5 mL burette are listed in
the Operating Instructions for the interchangeable burettes enclosed
in the burette package.
84 DL31/DL38
4 drying tubes 1 dispensing tubing
with cover with Teflon seal
1 suction tubing
with Teflon seal
1 drying 1 silicon
tube holder tubing
1 connector
A 5 mL
interchangeable 1 seal
burette incl. 1 L 1 stopper ST
brown glass 7.5
bottle
1 siphoning
1 titration tubing
vessel with
cover incl. 3- 2 screw caps
hole adapter with flat seal
with stoppers
1 pump button
DL31/DL38 85
Order No.
Threaded ring 51107459
Cover with 51107462
• three-hole adapter each with 1 ST 10
ST 7.5 and needle hole stopper 23982
• Stopper ST 7.5 (set of 5) 23452
• Connector 23957
O-Ring 51190366
Titration vessel (nominal volume: 250 mL) 51107463
Seal 51107492
(drying tube of titration stand)
86 DL31/DL38
Order No.
1-L brown glass bottle 71296
Gasket 23981
(min. order quantity: set of 5)
DL31/DL38 87
Order No.
Operating Instructions
- German 51709849
- English 51709850
- French 51709851
- Spanish 51709852
- Italian 51709853
Application brochure
- German 51709854
- English 51709855
- French 51709856
- Spanish 51709857
- Italian 51709858
TEMPERATURE
°C
HEATING
140 160
100120 180200
80
220
60
240
40
260
OFF
280
300
88 DL31/DL38
Order No.
Septum stoppers
- ST 10: set of 10. 23949
- ST 24: set of 10. 23950
Weighing boat
- glass, ø 20 x 60: set of 5 23951
- glass, ø 30 x 80: set of 5 23952
- PE, ø 25 x 65: set of 5 23953
- PE, ø 38 x 100: set of 5 23954
Syringes
- 1 mL: package of 100 71492
- 10 mL: package of 100 71482
Injection needles
- 80 x 1.2 mm: package of 12 71483
- 80 x 0.8 mm: package of 100 71484
DL31/DL38 89
Order No.
- with the Options 03, 011, 040 59759
Installation disk
Software DLWin or DLWin+ 51106120
DLWin V.xx
© Mettler-Toledo AG 1995
1
2
ON
IN
OUT 1 OUT 2
24 VDC 1.2 A
100-240 V
50/60 Hz
unfused
max. 9 A
OUT 3
OUT 4
TBox DR42
external devices via TTLIO)
90 DL31/DL38
12 Technical data
Power supply
• Voltages 100 – 240 VAC ±10%
• Frequency 50/60 Hz
• Power consumption 45 VA
• Fuse T3,15H250V (not interchangea-
ble)
Ambient conditions
• Ambient temperature +5 °C to +40 °C
• Atmospheric humidity max. 80% at 40 °C
• Use indoors
• Overvoltage category II
• Pollution degree 2
Materials
• Titrator housing Polypropylene
• Titration stand Polypropylene
• Keypad housing Polyester
• Membrane keypad Polyester
• Dispensing/suction tubing FEP
• Siphoning/connection tubing Silicon
• Three-hole adapter Polypropylene
• O-ring (screw cap) EPDM
• O-ring (threaded ring) FEP/silicon
• Ring seal (bottle) PTFE/silicon
• Seal (dispensing/suction tubing) PTFE
• Connector Polypropylene
DL31/DL38 91
Measurement principle Volumetric water content determina-
tion following Karl Fischer
End point indication Voltametric: voltage measurement at a
polarized double platinum pin electrode
Control Fuzzy logic control (parameters
changeable)
Measurement range Several ppm to 100% water
Measurement system
KF input Triaxial socket
• Current source
- Range -110 µA to +110 µA
- Resolution 1 µA
- Max. permissible errors 2%
• Voltage range
- Range -2000 mV to +2000 mV
- Resolution 1 mV
- Max. permissible errors 0.5%
Outputs
Magnetic stirrer Type: solid state
• Output 8-pin socket (mini DIN)
• Speed software-controlled (1-100%)
• Power 2 W (24 V)
External titration stand Type: DV705
• Output 5-pin socket (mini DIN)
- Magnetic stirrer DC motor
Speed software-controlled (1-100%)
Voltage range 0 to 18 VDC (max. 300 mA)
- Pump
Voltage 24 VDC (max. 400 mA)
92 DL31/DL38
Burette drive module
Drive Stepper motor
Resolution 1/10000 of the burette volume
Max. permissible errors 0.3% relative to the respective
burette volume of 5, 10 and 20 mL
Fill time minimal 20 s (rate can be set
between 10 and 100%)
Delivery time minimal 20 s
Interchangeable burettes
Volume 1, 5, 10 and 20 mL
Materials which come
into contact with the Fluoroplastic, borosilicate glass,
titrant ceramic
Diaphragm pump
Voltage 24 VAC (4 W)
Pressure buildup approx. 180 mbar
Display
Liquid crystal display 6 lines. 20 characters per line,
backlit
Languages English, German, French, Spanish
and Italian
DL31/DL38 93
Memories
Database
• KF standard method 1
• Tutorial method 1
• KF learn method 1
• METTLER methods 10 (DL38 only)
User data memory
• User methods DL38: max. 50 / DL31: max. 1
• KF titrants DL38: max. 20 / DL31: max. 10
• KF standards DL38: max. 20 / DL31: max. 10
• Br number titrants DL38: max. 5 / –
94 DL31/DL38
RS232C connector (balance/GA42/computer)
Pin
assignment
5
9 RxD (In) 2 Data to titrator
4 Data from titrator
8 TxD (Out) 3
3 470 Ω Handshake from titrator
7 DTR +12 V 4 fixed to operational
2 readiness
6 GND 5
1 Signal ground, galva-
Titrator nically connected to
titrator electronics
DL31/DL38 95
Centronics socket
Pin
assignment
Titrator
96 DL31/DL38
13 Index
<F1>...<F3> 6 Calculations (auxiliary fct.) 64
Centronics socket 96
Access for expert/routine 22 Cleaning 81
All results 61 Command keys 6
Amount of sample Computer (attaching) 79
(calculating) 29 Computer (defining) 20
Analyse (Run) 41 Computer method 42
Arrow keys 9 Concentration (KF) 12
Author 35 Concentration determination
Auto start 30 (KF) 51
Auxiliary function keys 8 Control parameters (Br No.) 38
Auxiliary functions 58 Control parameters (KF) 31
DL31/DL38 97
Expiry date 42 METTLER methods 37
Mix time 30
Fill rate 59 Molecular sieves (replacing) 82
Filling (titration vessel) 60
O-rings (replacing) 81
GLP 34 Operating concept 5
Optional equipment 88
Hardware tests 72
Help texts 1 Peripheral interfaces
(checking) 73
Peripherals 18
i key 9
Polarization current 31
Information menu 66
Post consumption 48
Inputs and outputs
(ckecking) 74 Predispensing 30
Installation 75 Pretitration 43
Printer (attaching) 79
Printer (defining) 18
Keypad 5
Pump (auxiliary fct.) 60
KF learn titration 55
Pump button (replacing) 82
KF standard method 26
Pump unit (replacing) 82
KF standards 15
Pumping time 24
KF titrants 12
Putting into operation 5
Language (changing) 10
Recalculation 62
Report (auxiliary fct.) 64
Magnetic stirrer (attaching) 78 Report (method parameter) 34
Magnetic stirrer (replacing) 82 Reset key 7
Main methods 17 Resources 11
Maintenance 81 Result (DL31) 32
Malfunctions 70 Results (auxiliary fct.) 61
Max. srel 33 Rinse tip 59
Menu keys 7 Routine 23
Method parameters (Br No.) 38 RS232C connector 95
Method parameters (KF) 29 Run (analysis menu) 41
Methods 25
98 DL31/DL38
Safety measures 2 Titrant (method parameter) 30
Sample determination Titration stand 20
(Br No.) 54 Titration stand (equipping) 75
Sample determination Titrator (overview) 4
(KF, Br No.) 45 Titrator ID 22
Sample determination (KF) 46 TTL inputs & outputs 80
Sample parameters (Br No.) 40 TTL-I/O socket 95
Sample parameters (KF) 29 Tubing (replacing) 81
Sample series 50 Tutorial 6, 66
Series (terminating) 65 Type (sample) 29
Settings 21
Setup 11 User data memory 94
Solvent bottle (installing) 77 User interface (checking) 72
Solvent capacity 23,66 User list 22
Speed (stirrer) 30 User methods 37
Standard equipment 84
Standby (method parameter) 34
Wait time 30
Standby mode 44
Waste bottle (installing) 77
Starting a method 41
Weight transfer 48
Statistics (changing) 64
Statistics (method
parameter) 33
Statistics (printout) 64
Statistics series 65
Stirrer (auxiliary fct.) 58
Stirrer speed 30
Storage procedure
13, 14, 16, 35
DL31/DL38 99
Declaration of conformity
We, Mettler-Toledo GmbH, Analytical, Sonnenbergstrasse 74, CH-8603
Schwerzenbach, Switzerland, declare under our sole responsibility that the
products listed below – marked with “CE” – are in conformity with the
directives mentioned.
98
Product Directive
73/23/EEC
DL31 Low voltage
DL38
89/336/EEC
Electromagnetic compatibility
Schwerzenbach,
August 1998
The following standards have been applied for the declaration of conformity:
DL31/DL38 101
Safety testing of our products
These products have passed the required safety tests for electrical prod-
ucts for the approvals and marks mentioned.
Switzerland Canada
Standard
CAN/CSA-C22.2 No. 1010.1-92
UL Std. No. 3101-1
MG8 NRTL/C FCC, Part 15, class A
FCC rules
This equipment has been tested and found to comply with the limits for a
Class A digital device, pursuant to both Part 15 of the FCC Rules and the
radio interference regulations of the Canadian Department of Communica-
tions. These limits are designed to provide reasonable protection against
harmful interference when the equipment is operated in a commercial
environment. This equipment generates, uses and can radiate radio fre-
quency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio communica-
tions. Operation of this equipment in a residential area is like to cause
harmful interference in which case the user will be required to correct the
interference at his own expense.
102 DL31/DL38
Declaration of System Validation
DL31, DL38
DL31/DL38 103
Product Life Cycle Model
Ideas
Quality Plan
Documentation Plan
Project Plan
Review Report
Review Report
Phase:
Confirmation Review Report
Phase:
Project End Readiness Review Report
Phase out
104 DL31/DL38
DL31/DL38
Switching from routine to expert mode
DL31/DL38 105
To protect your METTLER TOLEDO product’s future:
METTLER TOLEDO Service assures the quality, measuring
accuracy and preservation of value of all METTLER TOLEDO
products for years to come. Please send for full details about
our attractive terms of service. Thank you.
*P51709850*
Mettler-ToledoGmbH,Analytical,Sonnenbergstrasse74,
CH-8603Schwerzenbach,
Tel. (01) 806 77 11, Fax (01) 806 73 50, Internet:http://www.mt.com