A NSI Code
Maximum Temperature Useful Range 0 to 750C (32 to 1382F) The rm . Grade 0 to 200C (32 to 392F) Ex t. Grade -200 to 1250C (-328 to 2282F) The rm . Grade 0 to 200C (32 to 392F) Ex t. Grade 0 to 80C (32 to 176F) Ex t. Grade -200 to 350C (-328 to 662F) The rm . Grade -60 to 100C (-76 to 212F) Ex t. Grade
IR O N Fe (m agne tic)
2.2C or 0.75%
1.1C or 0.4%
ALO MEGA NIC KELALUMINIUM Ni-Al (m agne tic) C O NSTANTAN C O PPER NO NE NO NE NIC KEL ESTABLISHED ESTABLISHED C u-Ni C O NSTANTAN C O PPER NIC KEL C u-Ni
-6.458 to 54.886
-200 to 1250C (-328 to 2282F) 2.2C or 0.75% Above 0C 2.2C or 2.0% Be low 0C 1.1C or 0.4%
V*
C O PPER Cu
C O PPER Cu
-6.528 to 20.872
-200 to 350C (-328 to 662F) 1.0C or 0.75% Above 0C 1.0C or 1.5% Be low 0C 0.5C or 0.4%
International A NSI International IEC 584-3 CODE IEC 584-3 Intrinsically Safe J
Comments Environment Bare Wire R e ducing, Vacuum , Ine rt. Lim ite d Use in O x idising at High Te m pe rature s Not R e com m e nde d for Low Te m pe rature s C le an O x idising and Ine rt. Lim ite d Use in Vacuum or R e ducing. W ide Te m pe rature R ange . Most Popular C alibration Alte rnative to KX Type Ex te nsion W ire for Low Te m pe rature s; Not R e com m e nde d for Ge ne ral Use Mild O x idising, R e ducing Vacuum or Ine rt. Good W he re Moisture is Pre se nt, Low Te m pe rature and C ryoge nic Applications
V*
A NSI Code
Limits of Error** (Whichever is Greater) Standard Special -200 to 900C (-328 to 1652F) 1.7C or 0.5% Above 0C
www.omega.com/techref/thermcolorcodes.html
-9.835 to 76.373
1/3
9/6/12
-9.835 to 76.373
0C 1.7C or 1.0% Be low 1.0C 0C or 0.4% 2.2C or 0.75% Above 0C 2.2C or 2.0% Be low 0C
-270 to 1300C (-450 to 2372F) The rm . Grade 0 to 200C (32 to 392F) Ex t. Grade
-4.345 to 47.513
1.1C or 0.4%
PLATINUM PT
NO NE ESTABLISHED
-0.226 to 21.101
1.5C or 0.25%
0.6C or 0.1%
PLATINUM PT
NO NE ESTABLISHED
1.5C or 0.25%
0.6C or 0.1%
International A NSI International IEC 584-3 CODE IEC 584-3 Intrinsically Safe E
Comments Environment Bare Wire O x idising or Ine rt. Lim ite d Use in Vacuum or R e ducing. Highe st EMF C hange pe r De gre e
Alte rnative To Type K More Stable at High Te m ps O x idising or Ine rt. Do Not Inse rt in Me tal Tube s. Be ware of C ontam ination. High Te m pe rature O x idising or Ine rt. Do Not Inse rt in Me tal Tube s. Be ware of C ontam ination. High Te m pe rature
A NSI Code
A lloy Combination +Lead C O PPER Cu -Lead C O PPER -LO W NIC KEL C u-Ni
Maximum Temperature Useful Range 0 to 50C (32 to 122F) Ex t. Grade 0 to 1700C (32 to 3092F) The rm . Grade 0 to 100C (32 to 212F) Ex t. Grade 0 to 2320C (32 to 4208F) The rm . Grade 0 to 260C (32 to 500F) Ex t. Grade 0 to 2320C (32 to 4208F) The rm . Grade 0 to 870C (32 to 1600F) Ex t. Grade 0 to 2320C (32 to 4208F) The rm . Grade 0 to 260C
Limits of Error** EMF (mV) Over Max. (Whichever is Greater) Temperature Range Standard Special
U*
NO NE ESTABLISHED
G* (W )
TUNGSTEN W
0 to 38.564
Not Establishe d
C* (W 5)
0 to 37.066
Not Establishe d
www.omega.com/techref/thermcolorcodes.html
D* (W 3)
NO NE ESTABLISHED
0 to 39.506
Not Establishe d
2/3
9/6/12
(W 3)
0 to 39.506 0 to 260C 1.0% to (32 to 500F) 2320C Ex t. Grade **ASTM Standard E230-87, 1992 Annual Book of ASTM Standards
Establishe d
International A NSI International IEC 584-3 CODE IEC 584-3 Intrinsically Safe U*
Comments Environment Bare Wire Ex te nsion Grade C onne cting W ire for R & S The rm ocouple s; Also Known as R X and SX Ex te nsion W ire
O x idising or Ine rt. Do Not Inse rt in Me tal Tube s. Be ware of C ontam ination. High Te m pe rature . C om m on Use in Glass Industry Vacuum , Ine rt, Hydroge n. Be ware of Em brittle m e nt. Not Practical Be low 399C (750F). Not for O x idising Atm osphe re Vacuum , Ine rt, Hydroge n. Be ware of Em brittle m e nt. Not Practical Be low 399C (750F). Not for O x idising Atm osphe re Vacuum , Ine rt, Hydroge n. Be ware of Em brittle m e nt. Not Practical Be low 399C (750F). Not for O x idising Atm osphe re
G* (W )
C* (W 5)
D* (W 3)
Copyright 2003-2012 | OMEGA ENGINEERING, Inc. | OMEGA Priv acy Statement | Terms and Conditions
MXV MMV | 72
www.omega.com/techref/thermcolorcodes.html
3/3