Beruflich Dokumente
Kultur Dokumente
SERVICE BULLETIN
FUEL INJECTION PUMP
TECHNICAL HINT
DENSO AUTHORIZED ECD SERVICE DEALER ONLY
Table of Contents
1. Adjustment Preparation 2. New Pump Adjustment Procedure with DST-1 2 2
QC1113
1. Adjustment Preparation
(1) Connect DST-1 to the conventional adjustment system with ECD controller on the bench.
DST-1
QC1114
QC1115
QC1116
QC1117
(4) Select the part number of injection pump to be calibrated. ( In this case, the following specifications are all for 098000-0300.)
QC1118
(5) Confirm the part number again and press + key if its OK.
QC1119
(6) Select 1.PUMP ADJUST. NOTE: At this moment, ROM cannot be writable due to no calibration. NOTE: 3.PUMP ROM DATA LIST is directly displayed the data in the ROM.
QC1120
2-1.Pump Adjustment
The test condition is displayed as SETUP CONDITION, and press enter key after confirming each item.
QC1121
2-1-1.BASIC CHECK
(1) Select 1.BASIC CHECK.
QC1122
(2) The calibration starts at this procedure RUNNING without referring to the conventional test specifications. Follow all procedures to complete the calibration.
QC1123
QC1124
QC1125
QC1126
(5) The display shows the completion after the running. Proceed to the next by pressing + key.
QC1127
(6) Press enter key to check fuel temperature sensor displayed as 2.FUEL TEMP. NOTE: The mark * means the completion of the procedure. It always appear every time when completing each calibration procedure.
QC1128
(7) Refer to the condition for the sensor check. Press enter key and check the resistance.
QC1129
QC1130
QC1131
QC1132
QC1133
QC1134
QC1135
QC1136
QC1137
QC1138
QC1139
QC1140
QC1141
QC1142
QC1143
(14) Press return key when completing all checkings of BASIC CHECK as shown in the figure.
QC1144
QC1145
10
QC1146
QC1147
QC1148
QC1149
11
QC1150
(5) After finishing all points, press + key and proceed to the next ( In this case, check points are 14 as shown in the display.).
QC1151
2-1-3.CAM ANGLE
(1) Press enter key to adjust the cam angle along the display guidance.
QC1152
QC1153
12
(2) Confirm the conditions for No.1 Cam Angle Preset & No.2 Cam Angle Adjustment.
QC1154
(3) Check the cam angle preset. If its out of the specification, change the command value with using + & - keys. NOTE: At first, the displayed command value is the original one of the pump. + key increases the command value and - key decreases it.
QC1155
QC1156
(5) DST-1 checks the cam angle automatically. If the display doesnt show OK but NG, return to the cam angle preset procedure and adjust the command value again.
QC1157
13
QC1158
QC1159
Reference: When both cam angle preset and cam angle adjustment are completed, the mark * appears at the marked position on the display.
QC1160
2-1-4.Q ADJUST
(1) Select 4.Q ADJUST.
QC1161
14
QC1162
(3) The first adjustment point and the following points are displayed. And, press enter key.
QC1163
(4) Check the delivery quantity. If its out of the specification, change the command value with using + & - keys.
QC1164
(5) The following points are also adjusted in the same way as the first point.
QC1165
15
(6) Confirm that the delivery quantity is OK or not at the check points as shown in the figure.
QC1166
QC1167
2-1-5.TIMER ADJUST
(1) Select 5.TIMER ADJUST.
QC1168
QC1169
16
(3) The first adjustment point and the following points are displayed. And, press enter key.
QC1170
(4) Initialize the DST-1 & ECD controller before the adjustment. Wait for a while to stabilize the pump angle ( cam angle).
QC1171
(5) Confirm that the pump angle ( cam angle ) is within the specification.
QC1172
QC1173
17
QC1174
QC1175
(8) Continue the initializing and confirming alternately along displayed procedures.
QC1176
QC1177
18
(10) Press return key when completing all checkings including TIMER ADJUST as shown in the right figure.
QC1178
QC1179
(2) Select 1.WRITE RECORDING DATA.. NOTE: 2.WRITE TO NEW PUMP ROM is selected when copying ROM data simply.
QC1180
QC1181
19
QC1182
(4) The completion message of writing the data in the ROM appears.
QC1183