Beruflich Dokumente
Kultur Dokumente
1.02 2008/02/21
Statements Our company reserves the copyright of this unpublished manual and is entitled to keep it confidential. This manual is only used for reference to operation, care and maintenance of our products. No disclosure of this manual to any others without our consent. We shall not be liable for any incidental or indirect damages resulting from mistakes herein, provision, field performance and use of hereof. It shall not provide any other party with the franchise granted by the Patent Law. The responsibility for the legal consequence arising out of breach of Patent Law and caused by any other third party shall not lie at the door of our company. The content herein can be altered without prior notification. Responsibility of the Manufacturer We only considers itself responsible for any effects on safety, reliability and performance of the equipment if:
The electrical equipment complies with the national safety standards; The instrument is used in accordance with the instructions.
Note: This device is not intended for home use. WARNING Warning: it may cause personal danger not to observe the correct operating steps. WARNING It is a must to observe all listed instructions for safe use of this equipment, which cant take the place of the executed medical steps. Please dont supervise and care patents only by dependence upon the alarm system. It is possible to cause accidents to the patents if you set the volume too low or turn off the volume on supervising and nursing patents. It is the most reliable method of supervising patients to combine the correct operation of the monitor with the close personal supervision of patients. This monitor is provided only for the use by the well-trained health care professionals in the health-care organizations. Please dont open any covers of the equipment for fear of electric shock. In case of any failures, please have the qualified servicemen trouble-shoot the monitor. This monitor may interfere with the ultrasonic image-forming system, showing as the interference signal on the ultrasonic display screen, so it is better to keep them far from each other. It is dangerous to put the electric contacts or the connecting device in the normal saline or other liquid and conductive paste. It is a must to keep clean and dry the electric contacts and the connectors including cable connector, connectors for power and parameter module, and
2
rack connectors. Where they are polluted by liquid, please dry them completely without fail. For detailed pollution, please contact your bio-medical department or our company.
WARNING This monitor isnt intended for medical treatment. Failure of all hospital or organizations responsible for use of this monitor in implementation of a satisfactory repair plan may give rise to the abnormal failure of the machine and even endanger personal health. Nowadays we will conditionally provide the circuit diagram, and the tabular description on calibrating method and other information if users require, in order to have the qualified technicians repair the instruments for them, which should be repaired by users as classified by this patient monitor. Warranty Manufacturing Process and Raw Materials This patient monitor warrants that no failures will happen to the manufacturing process and raw materials within one year as from the shipment date in the state of normal operation and maintenance for other new instruments except for the multi-probe and the blood-oxygen probe. The warranty period for the multi-probe and the blood-oxygen probe is six months. The monitor shall only be liable for maintenance under such a warranty. Free Service Our company shall not be liable for freight and other expenses under the above warranty, and the direct, indirect or final losses and delay arising out of the following situations, neither, such as improper use, replacements of parts unauthorized by our company or maintenance by the personnel unauthorized by our company. This warranty doesnt apply to the following items: abnormal use; seldom maintenance or damaged machine; Our original SN tag or manufacturing marker replaced or ripped off; the products made by other manufacturers. Safety, Reliability and Runnability In the following cases, Our Company shall not be responsible for safety, reliability and performance of the monitor: Parts are dismantled, stretched and readjusted;
Foreword
This manual details the performance, operating methods and other safety information of STAR8000D portable-type multi-parameter monitor, which should be deemed as the best starting point for new users to use such a monitor. The following symbols mean some important prompts uses should give their attention to:
WARNING
means the information you should know about how to avoid hurts to patients
means the information you should know about how to protect the equipment
Attention means the important information that has been laid emphasis on. This manual is intended for both the familiarity with various measurements and the perusal by the personnel experienced in operation of the monitoring equipment.
Contents
CONTENTS .....................................................................................................................................5 CHAPTER 1 GENERAL INFORMATION .......................................................................................8
1.1 General Information ................................................................................................ 9 1.2 Screen display ...................................................................................................... 10 1.3 Button Functions and Basic Operations ............................................................... 13 1.4 External Interfaces ............................................................................................... 16 1.5 Built-in Chargeable Battery .................................................................................. 18
CHAPTER 2 MONITOR ASSEMBLY .......................................................................................... 19
2.1 Open Package and Check ................................................................................... 19 2.2 Connect with AC Cable ........................................................................................ 19 2.3 Power on .............................................................................................................. 20 2.4 Connect with Sensor ............................................................................................ 20 2.5 Check Recorder.................................................................................................... 20
CHAPTER 3 SYSTEM MENU ...................................................................................................... 21
3.1 PATIENT MANAGE .............................................................................................. 21 3.2 DEFAULT (Default Setting) .................................................................................. 22 3.3 Face select ........................................................................................................... 23 3.31 Work Interface Selection ............................................................................. 23 3.32 Standard Interface ....................................................................................... 23 3.33 TREND SCREEN ........................................................................................ 24 3.34 oxyCRG SCREEN....................................................................................... 25 3.35 VIEWBED SCREEN.................................................................................... 26 3.4 TREND GRAPH (TREND Diagram)..................................................................... 28 3.5 TREND TABLE (TREND Figure) .......................................................................... 30 3.6 NIBP RECALL ...................................................................................................... 31 3.7 MONITOR SETUP (Monitor Setting) .................................................................... 32 3.71 ALARM SETUP (Alarm Setting) .................................................................. 32 3.72 RECORE (Record output setting) ............................................................... 33 3.73 TIME SETUP (System Time Setting) .......................................................... 34 3.8 Selection ............................................................................................................... 34 3.81 ALRARM Volume ........................................................................................ 34 3.82 (LCD LIGHT) LCD Brightness..................................................................... 35 3.83 KEY VOL (Keyboard Vol ) ........................................................................... 35 3.84 BEAT VOL (Cardio-beating Vol) .................................................................. 35 3.9 INFO (Monitoring Information) ............................................................................. 35 3.10 Drug Calculation & Titration List ......................................................................... 36 3.101 (DRUG CALC)........................................................................................... 36 3.102 TITRATION (Titration list) ....................................................................... 38
5
3.11 MAINTAIN (Machine Maintenance) .................................................................... 39 3.111 USER KEY(key/password 2016): .............................................................. 39 3.112 FACTORY KEY(key/password 3021): ...................................................... 39 3.12 DEMO (key/password 5188) .............................................................................. 40
CHAPTER 4 RECORDING .......................................................................................................... 41
4.1 General Information on the Recorder................................................................... 41 4.2 Record Type ......................................................................................................... 41 4.3 Operation and Status Information of Recorder .................................................... 42
CHAPTER 5 PATIENT SAFETY.................................................................................................. 45 CHAPTER 6 ECG/ RESP MONITORING .................................................................................... 48
6.1 Definition of ECG Monitoring ................................................................................ 48 6.2 Monitoring Procedures ......................................................................................... 49 6.2.1 Preparation ................................................................................................. 49 6.2.2Install ECG leads ........................................................................................ 49 6.3 ECG Screen Buttons ............................................................................................ 52 6.4 ECG Menu ............................................................................................................ 54 6.4.1 ECG setting menu ...................................................................................... 54 6.5 ECG Alarms & Reminders .................................................................................... 57 6.5.1 Alarming information .................................................................................. 57 6.7 Arrhythmia Monitoring (for option) ........................................................................ 61
CHAPTER 7 RESP MEASUREMENT ......................................................................................... 67
7. 1 How to measure RESP ....................................................................................... 67 7. 2 Setting of RESP monitoring ................................................................................ 67 7. 3 Install electrodes for RESP measurement .......................................................... 67 7. 4 RESP menu ......................................................................................................... 68 7. 5 RESP Alarms & Reminders................................................................................. 69 7.6 Maintenance & Cleaning ...................................................................................... 70
CHAPTER8 SPO2 MONITORING ................................................................................................ 71
8.1 Definition of SpO2 Monitoring ............................................................................... 71 8.2 Precautions in SpO2/Pulse Monitoring ................................................................. 72 8.3 Monitoring Procedures ......................................................................................... 72 8.4 Measurement restriction ....................................................................................... 74 8.5 SpO2 Menu ........................................................................................................... 75 8.6 Alarms & Reminders ............................................................................................. 76 8.7 Maintenance & Cleaning ...................................................................................... 78
CHAPTER 9 NIBP MONITORING ............................................................................................... 79
9.1 General Information .............................................................................................. 79 9.2.1 NlBP Measurement .................................................................................... 79 9.2.2 NIBP Parameter Setting & Adjustment ...................................................... 83
6
9.3 NIBP Setting Menu ............................................................................................... 83 9.4 NIBP Alarms & Reminders ................................................................................... 87 9.5 Maintenance & Cleaning ...................................................................................... 89
CHAPTER 10 TEMP MONITORING ............................................................................................ 91
10.1 TEMP Monitoring ................................................................................................ 91 10.2 TEMP SETTING MENU ..................................................................................... 91 10.3 Alarms & Reminders........................................................................................... 92 10.4 Maintenance & Cleaning .................................................................................... 93
CHAPTER 11 CO2 MONITORING .............................................................................................. 94 CHAPTER 12 IBP MONITORING ........................................................................................... 97
12.1 General information ............................................................................................ 97 12.2 Considerations of IBP monitoring....................................................................... 97 12.3 Monitoring procedure ......................................................................................... 98 12.4 IBP Menu ............................................................................................................ 99 12.5 Alarm information and prompt information ....................................................... 103
ATTACHMENT I ......................................................................................................................... 104 DESCRIPTION OF SYSTEM ALARM PROMPT....................................................................... 104
For information about the monitor, please read the General Information on the Monitor chapter. For introduction on various information displayed on screen, please read the Screen Display chapter. For operational methods, please read the Button Functions and Basic Operations chapter. For locations of various interfaces, please read the External Interfaces chapter. For notices of using the monitor with power supply from a battery, please read the Built-in Chargeable Battery chapter. Warning This monitor is to monitor clinical patients, only for doctors and nurses use. Warning Dont open cover of the equipment to avoid possible risks in electric shock. Any maintenance or upgrading on the monitor must be conducted by service personnel trained and authorized by our Company. Warning Dont use this monitor where there are flammables such as anesthetic agent, so as to prevent from explosion. Warning Users before starting use should check whether the equipment and its accessories can work properly and safety. Warning Please make sufficient alarming setting for each patient in order to prevent from delayed therapy and make sure there is voice effect during alarming. Warning Dont use mobile phones around the monitor. Mobile phones will generate strong emission fields and disturb the monitor. Warning During defibrillation dont touch patients, tables and the machine. Warning
8
Equipments inter-connected with the monitor should form an equal-potential body (as protective effective earthing). Warning Users (doctors or nurses) should ensure safety of patients under monitoring, when the monitor is used together with electrosurgical equipments.
-500 4,600m(-1,600 15,000 feet) -50013,100m(-1,600 43,000 feet) 100~250 (V) AC, 50/60 (Hz) Pmax=80VA FUSE T 1.6A
The portable multi-parameter monitor is of rich functions (as shown in the following picture), applicable for bedside monitoring on adults, infants and newborns. This monitor can monitor main parameters including ECG, RESP, SpO2, NIBP, TEMP and IBP. It integrates parameter measurement modules, display and record output to build such a solid and light monitor. Its replaceable built-in battery makes convenience for patient movement and it will clearly display 7 waveforms and all the monitoring parameter information on the high-resolution interface. From up to down they are in turn: (1) ALARM INDICATOR; (2) SILENCE button; (3) FREEZE button; (4) PRINT button; (5) START button (for blood pressure); (6) MAIN button (for key frame return); (7) MENU button; (8) ON/OFF button; (9) JOG-DIAL; (10) WORK INDICATOR; (11) CHARGING INDICATOR. The JOG-DIAL below the ON/OFF button has such three working methods as turning left, turning right and pressing, which can respectively actualize leftward movement of cursor, rightward movement of cursor and execution confirmation, and are mainly applied to the operation of menu.
9
The outlet of the recorder is located on the left side of the machine and its jacks and AC line cord sockets are on the rear panel of the monitor.
Figure 1-1 Portable Multi-parameter Patient Monitor This monitor has the following monitoring functions: HR Dual channel ECG waveform Analysis of arrhythmia and S-T section (for option) and, pulse analysis (PACE) Systolic pressure (NS); diastolic pressure (ND) and mean blood pressure (NM) Blood oxygen saturation (SpO2) and pulse rate (PR) SpO2 plethysmogram Temperature Data (TEMP centigrade/Fahrenheit) RR RR waveform
This monitor has rich functions, able to provide various functions such as voice-light alarming, TREND storage & output, NIBP measurement save and review, medicine calculation, ST analysis, pulse analysis, heart rate turbulence analysis, etc. This monitor has a friendly operation interface, able to provide all functions with the keys and buttons on the front panel (as and in fig. 1-1), refer to Function Keys part for details.
This monitor has a color LED screen, able to concurrently display collected patient parameters, waveforms, and alarming information provided by the monitor, bed marks, clocks, monitor status and other reminder information. The main screen is divided into 3 sub-areas, i.e., Information Area & , Waveform Area , and Parameter Area . (As shown in Figure 0-2)
Information area()
BED NO: patient's bed number. The bed number of new patient's is blanket. Other prompt message in the information area appears and disappears together with the reported state. In accordance with the content, it is divided into: The prompt message from the monitor reports the state of monitor or sensor, which appears regularly in the alarm message of regional monitor on the right of the clock (for the detailed setting method, please see Alarm Function).
means the sign of alarm pause time. When you press the silence button the prompt is displaying on the top of information area. Press the SILENCE button for a short time, which means all alarm sound is temporarily shut down until the SILENCE button is pressed for a short time or that the system can restore sounding after the alarm pause time ends. Such three kinds of time as 1 Minutes, 2 Minutes and 3 Minutes are selectable in the alarm pause time. The upper waveform on the screen being frozen, FROZEN prompt will appear at the lower part of the screen. Where it is in the demo mode, DEMO prompt will appear on the upper and lower
11
Parameter area()
Parameters can display at the fixed position (as shown in ~ of the following picture, which are separately:
ECG Heart Rate or Pulse Rate,unitbpm ST-segment of channel 1 and channel 2 (, unit: mv) PVCs times ( Unit: times/minute) NIBP (From left to right) Systolic, Diastolic, Mean (, Unit: mmHg or kPa) SpO2 SpO2,unit% Pulse Rateunitbpm TEMP Temperature,unit: or RESP Respiration Rate,unitbpm The above monitored results will show in the parameter area. The parameters are refreshed once per second, but NIBP value, once per measurement. Users can select the monitoring parameters and the main screen will display the relative content.
Warning With the monitor power on, the system will automatically detect whether the voice-light alarming system works.
FREEZE
(FREEZE button): Press this button and all waveforms on the screen can be frozen in the normal mode. Press this button once again and the frozen waveforms can be freed. (PRINT button): Press this button to start a real-time recording. The recording length is determined according to the real-time recording time in the Recorder submenu of Setting menu. Press this button once again in the recording process and the recording will stop at once. Each interval of pressing this button should be greater than 2 seconds and the frequency should not be excessively high. To start/stop NIBP measurement, press to inflate the cuff to start a blood pressure measurement. When measuring, press to stop the measurement and deflate the cuff. (The key to go back to main screen)press to back to main screen, close window if it is opened, hide the system menu bar. (MENU button): Press this button to have the main menu pop up when the interface is in the state of non-window setting. (ON/OFF button): Press this button to control the startup and shutdown of the monitor. Trimming knob to select and change the settings. Operation can be performed by turning it clockwise, counterclockwise or pressing it down. The knob is mainly used in menu and window operation.
START
MAIN MENU
KNOB
Basic operation
14
Display all required waveforms: Press the MENU button on the screen to have System Menu window pop up. SYSTEM MENU>>FACE SELECT. Modify the waveform speed: ECG: Move the cursor to ECG, and press the knob to enter ECG SETUP menu. Modify the SWEEP item. PLETH: Move the cursor to SPO2, and press the knob to enter SPO2 SETUP menu. Modify the SWEEP item. RESP: Move the cursor to RESP, and press the knob to enter RESP SETUP menu. Modify the RESP item. Modify the alarm limit: ECG: Move the cursor to ECG, and press the knob to enter ECG SETUP menu. Modify the ALM HI or ALM LO. SpO2: Move the cursor to SPO2, and press the knob to enter SPO2 SETUP menu. Modify the SPO2 ALM HI or SPO2 ALM LO.
PR: Move the cursor to SPO2, and press the knob to enter SPO2 SETUP menu. Modify the PR ALM HI or PR ALM LO. NIBP systolic pressure Move the cursor to NIBP, and press the knob to enter NIBP SETUP menu. Modify the SYS ALM HI or SYS ALM LO. NIBP MEAN: Move the cursor to NIBP, and press the knob to enter NIBP SETUP menu. Modify the MEAN ALM HI or MEAN ALM LO. NIBP DIA: Move the cursor to NIBP, and press the knob to enter NIBP SETUP menu. Modify the DIA ALM HI or DIA ALM LO. RESP: Move the cursor to RESP, and press the knob to enter RESP SETUP menu. Modify the ALM HI or ALM LO. TEMP: Move the cursor to TEMP, and press the knob to enter TEMP SETUP menu. Modify the T1 ALM HI or T1 ALM LO.
Record the real-time waveform: Press the REC/STOP button to start the real-time recording of recorder. For details, please refer to the Recording Function. Modify the volume: Alarm Vol: Press the menu button to enter the system menu. MONITORSETUP>>SELECTION>>ALARM VOL
15
Keyboard Vol: Press the MENU button to enter system menu. SELECTION>>KEY VOL Heartbeat Vol: Press the MENU button to enter system menu. SELECTION>>BEAT VOL Set the system time: Press the MENU button to enter system menu. MONITOR SETUP>>TIME SETUP Restore the default setting: For details, please refer to the Default Setting.
Figure 1-5 Left Surface On the right side of the monitor are the patients cable and the sensor jack as shown in Figure 0-6: ECG cable jack Spo2 sensor jack TEMP probe jack NIBP cuff jack
This symbol means this application part is of CF type, designed with special protection from electric shock (especially provided with F-type floating insulation apparatus for permissible leakage current) and suitable for the defibrillation process. Other symbols will be introduced in the Patient Safety chapter.
Figure 1-7 Rear Panel There are the following holes in the rear panel: 1) Power socket() 2) Network interface 3) Fuse holder() 4) USB interface() 5) Serial interface 6) Nurse call port 7) (hole ) Warning
This network port can only be connected with our central monitoring system. Warning All the simulated or digital equipments connected with this monitor must be certified under the designated IEC standards (such as IEC 60950 Date Processing Equipment Standard and IEC 60601-1 Medical Equipment Standard). And all configurations must comply with effective versions of IEC 60601-1-1 system standards. Persons in charge of connecting additional equipments with the input/ output signal terminals should configure the medical system and be responsible for compliance of the system to IEC 60601-1-1 standard. For any enquiries, please contact the supplier.
17
18
Open Package and Check Connect with AC Cable Power on Connect with sensor Check Recorder Attention For normal work of the monitor, before use please read this chapter and the Patient Safety chapter and assemble in accordance with the requirements.
booted or not.
2.3 Power on
The logo displays when the power is on, and appears the processing screen. After the 3~5 seconds checking process, the system enters the monitoring main screen and the users can start operations. Attention In case of any fatal errors found during the self-detection process, the system will alarm. Attention Check all the available monitoring functions and make sure they work properly. Attention If a battery is configured, users must charge the battery after each time of use so as to ensure sufficient power storage. Attention If any monitoring functions are found with damage or there are any error reminders, dont use this monitor to monitor patients and quickly contact with the biomedical engineers of your hospital or maintenance engineers of our Company Attention Reboot the equipment at least 1 minute after shut down.
20
21
Figure 3-2 Patient MANAGE Users can set the following content: Bed NO 1-200 beds for option SEX Patient gender( female, male) PAT TYPE Adult, Pediatric Neonate. PACE ON/OFF NEW PATIENT Config to update patient Attention PACE: The default setting is OFF after you restart the monitor. In this menu, user can select the NEW PATIENT and enter the CONFIRM TO UPDATE PATIENT dialog box to confirm whether delete the monitoring now or updating, as shown in the following figure:
Figure 3-3 Confirm Updating of Patient Select Yes and you can delete all information on the currently monitored patient and quit the menu. Select No and you can refuse updating of patient, save the information on the current patient and quit the menu. Attention Select Yes and you can cancel all information on the currently monitored patient.
Attention After any item is selected in the submenu, the selected one will take the place of the current system configuration and become the default setting of the system. The setting is depending on the patients type. (PAT TYPE)
Figure 3-4 Confirm Saving of Current Setting Select Yes and you can save all configuration of the current patient type as the users default setting. Select No and you can abandon the current operation so that the system can continuously keep unchanged the original setting. Attention After any item is selected in the Default Setting menu, the dialog box Confirm Default Setting will pop up. You can select Yes to determine selection, or select No to give up the selection.
Figure3-6 TREND Concurrence Interface Location of trend diagrams TREND diagrams are located in the right to the waveforms, with the same colors to the corresponding parameters. TREND length Dynamic trend length is 2 hours; in a trend diagram, the right side of the horizontal axis is 0 hour, and the left side is 2 hours.
24
End of trend concurrence interface Out of the Interface Selection options, select any other work interface to end the trend concurrence interface.
Figure 3-7 oxyCRG Dynamic Refresh Interface TREND Diagram in oxyCRG Interface oxyCRG Dynamic Refresh Interface is made up of such two trends as compressed respiratory wave and PR, and located on the lower part of Screen Waveform Area. OxyCRG Trend Length Selection In the oxyCRG Dynamic Refresh Interface there are such three screen hot buttons as Trend Time Length, Respiratory Rate/Wave Trend Selection and Record. TREND Diagram for 1 minute and 2 minute can be selected through the trend time buttons. Compressed Respiratory Wave The function of Respiratory Rate/Compressed Respiratory Wave means the operator can select the PR Trend or Compressed Respiratory Wave as needed, under which the displayed content occupies the same position. Select the PR Trend and this position will show the TREND Diagram of the respiratory rate; select the Compressed Respiratory Wave and this position will show the respiratory wave after compression. In the Work Interface Selection menu, select other work interface to end the OxyCRG work
25
interface.
Figure 3-8 Bed to Bed Observation Interface VIEWBED SCREEN Button (reserved feature) There are two buttons in the VIEWBED SCREEN Interface, i.e. Bed Selection Button and Waveform Selection Button. Press the Bed Selection button and the number of other bed networked presently and the patients name. You can select other monitors to observe according to the patients name and the bed number. Provided that no monitors are connected to this monitor on the same network, the bed selection area will prompt the N/A message, which means no other monitors are networked at this time. You can use the specific bed selected by use of this button first, and then the system will turn to the selected bed for Bed to Bed observation. The waveform selected means a certain waveform in the area of waveform selection button. Waveform Selection button is used to select a waveform on the observed monitor. Where
26
the waveform name display area shows N/A, it means there are no waveforms for observation on the monitor beside the bed. You can select different waveforms of the observed bed for observation by use of this button. VIEWBED Alarm Indication The alarm indicator on the upper right part of the Bed to Bed observation interface is mainly used to indicate the alarm situation of the observed bed. This indicator is the same as the one on the keyboard panel of the observed bed in operation. Where there is mid-level/lo-level alarm from the observed bed, this indicator will turn yellow; where there is hi-level alarm from the bed, this indicator will turn red. If no alarm happens to the observed bed or the alarm is shielded, the alarm indication icon will not appear. Bed to Bed Observation Parameter Area All parameters of the observed bed will display in the parameter display area on the lower part of the bed selection button. Bed to Bed Observation Waveform Area Waveform Display Area is on the lower part of the waveform selection button. The scanning method (refresh or rolling) of this waveform is the same as the waveform scanning method of this bed. The related waveform description to the displayed waveform is on the upper part of the waveform. The scanning speed of waveform is the same as that set for this bed. Technical Information Prompt Area The Technical Information Prompt Area is on the right side of the patients name in the Bed to Bed observation interface and mainly used to prompt the observed technical information about other beds in such situations as network failure or connection timed out so that no information about other observed beds is prompted. Bed to Bed Observation In the Work Interface Selection menu select the other work interface to end the Bed to Bed observation interface. Principles for Automatic Selection of Observed Bed and Waveform When you start the monitor or enter the Bed to Bed observation work interface, the system will automatically select an onlined beside-bed monitor for you and a waveform on the monitor for observation. If the bed is off-lined, the observation bed will automatically close and delete all alarms, parameter and waveform, but the Bed to Bed observation interface still remains. If you want to observe other beds, you must make selections once again by use of the bed selection button. If the waveform being observed disappears due to the cancellation or close of the observed bed measurement module, the waveform area will not be waveform-refreshed and will become a blank area. If you want to observe the other waveforms of this bed, you have to make reselection.
27
Figure 3-9 TREND Diagram Menu Vertical axis is for measured values and horizontal axis for measurement time. The symbol is the cursor for TREND diagrams, and the measured value at the position it arrows is displayed below the TREND diagram while its corresponding time is displayed above the TREND diagram. Except NIBP values, all the other TRENDs are displayed in continuous curves; for NIBP TREND diagrams, refers to systolic pressure, refers diastolic pressure, and * refers to average pressure.
observe 1-hour TREND, or select 1 minute, 5 minutes or 10 minutes if you want to observe 72-hour TREND.
Obtain the TREND data at certain time in the current TREND diagram
Select Cursor and rotate the knob to control movement of the cursor; with the cursor moves, its arrowed time also changes, and the parameter value at such time will be displayed below the horizontal axis. If there is a indication in the right side of the window, when the cursor moves onto this indication the TREND diagram will automatically page down to display later TREND curves; and if there is a indication in the left side of the window, when the cursor moves onto this indication the TREND diagram will automatically page up to display earlier TREND curves.
Operation sample
Observe the NIBP TREND diagram within the latest 1 hour: Press the TREND button on the control panel to pop up the Main menu; Select the TREND Diagram Review option in the menu; Select the parameter: rotate the knob in the Parameter Selection item until NIBP is shown in the dropdown box; Select 1 or 5 seconds in the Resolution item; Select Move left/ right and rotate the knob, while observing changes in the TREND diagram time and TREND curves; Stop at the period to be carefully observed; in case the vertical axis is out of proper size, for example, some TREND values exceed the highest value of the current vertical axis, select Adjust amplitude to adjust; If users want to know the measured value at certain time, just select move cursor and move the cursor to where they wants, then time will be displayed above the curve and measured values below the curve; If users need output the TREND diagrams to the recorder, just select the record button so as to let the recorder output NIBP TREND of the current review window;
29
Figure 3-10 TREND Figure setting menu Time corresponding to various groups of TREND data is displayed at the left column, with dates braced. What are listed is the cases that have once been marked, which corresponds to the time of the marked cases. Parameters in the TREND figures can be categorized into the following 5 groups: HR, RR, PVC ST1, ST2 TEMP SPO2, PR NIBP NS/ND (NM) NIBP TREND data has its own characteristics; besides of measured values, below each measurement point there is time for this NIBP measurement.
Operation sample
To observe a NIBP TREND figure: Press the TREND button on the control panel to pop up the Main menu; Select the TREND Figure Review option in the menu; Select the parameter: select Shift left/ right and rotate the knob until NIBP is shown in the dropdown box; Select the resolution: click the left item and select the expected data interval; Select Page up/ down and rotate the knob, while observing NIBP TREND data over various time; Press exit to exit observation on TREND figure.
Data is sorted in time sequence, from early to late, and each screen can display 10 times of measurement data, while users can select Page up/ down to view later or earlier data. Maximally 400 measurement results can be displayed, and when the measure times are over 400, only the latest 400 will be displayed.
Figure 3-13 Alarm Setup Alarm Setup has the following options: ALM REC TIME (Alarm Record Time): In case of physiological alarm, the system can record the information before and after the alarm time. The system can give three kinds of time, i.e., 8 seconds, 16 seconds. The seconds in the options are the sum of seconds before and after the alarm time. For example, 8 seconds mean the
32
information within 4 seconds before the alarm time and within 4 seconds after such a time. ALM PAUSE TIME (Alarm pause time)1 minute, 2 minutes or 3 minutes PARA ALM TYPE (Parameter alarming method)UNLATCH or LATCH LAM (Display alarm limits)ON andOFF EXIT Note: LATCH means the system will raise the alarm until someone interferes it in case of alarm; UNLATCH means the alarm will automatically disappear with the alarm factor vanishing.
In this menu users can set the output of record waveforms whose options are: REC WAVE1 SPO2 REC RATE REC GRID RT REC TIME ECG1 SpO2 volume graphic waveform 20.0/50.0 mm/s record speed. ON/OFF 3S/5S/8S/CONTINUAL (real-time record time)
Printing recording time is 10 seconds within the operator freezing the screen. Alarm recording was setting each of parameters cursor. Attention The recorder is a component for option.
33
Attention Where two similar waveforms are selected, the system will automatically adjust the other one into a different one.
WARINING Where the alarm volume of the system is set at 0, the monitor will not give alarm sound in case of alarm, so you should use this function carefully. Attention Where the Alarm Vol is closed at 0, the state can only work after this startup. When the monitor is started up once again, this setting shall return to the previous set value of the system.
35
Figure3-17 Drug Calculation The following formulae are used for medicine dosage calculation: Medicine contents = Total medicine volume / Liquid volume Infusion speed = Medicine dosage / Medicine contents Continued time = Total medicine volume / Medicine dosage Medicine dosage = Infusion speed Medicine contents
Operation method:
In the medicine calculation window, operators should firstly select names of the medicines to be
36
calculated, and then confirm patient weight, and input other known values. Subsequently, operators move the cursor to the various calculation items in the calculation formulae, press the knob and rotate it, so as to select the calculation value. After the calculation value is selected, value of the items to be calculated will be displayed at the corresponding position. Values for each calculation item have their limits, if the calculated results exceed such limits, the system will display ---.--. Attention Under this medicine calculation function, other menu items are available for input only after operators input patient weight and medicine names. The values firstly given in the system are only a random group of initial values, and operators should not take such values as calculation standard, instead, should re-input a group of values suitable for the current patient, based on the comments by doctors. Attention Each kind of medicine is subject with fixed units or unit series, and operators must select proper unit based on comments by doctors. Under the same unit series, numbering system of the units will be automatically adjusted with the current input values, and when the input value exceed out of expression of the relevant unit, the system will display ---. Attention After operators input a certain value, the system will give a clear reminder in the menu, reminding operators to check correctness of the inputted value; only inputted values are guaranteed to be correct, the calculated values will be reliable and safe. Attention In case of newborns, dropping speed and volume of an infusion drop make no sense. Attention The system gives a reminder for each inputted value, asking operators to confirm. Operators must be serious with every such reminder, as only valid and correct inputs can get reliable calculation results. Select medicine type: move the cursor onto Medicine name, rotate the knob and select one from aminophylline, dobutamine, dopamine, epinephrine, heparin, isuprel, lidocaine, nipride, nitroglycerin, pitocin, Medicine A, Medicine B, Medicine C, Medicine D and Medicine E, altogether 15 types. At one time, only one type of medicine can be selected for calculation. Attention The above introduced A, B, C, D, and E are not actual medicine names but only codes for medicines. Units for these five types of medicines are fixed, and operators can select proper units based on general practice of medicines. The expression rules of their units are as follows: Medicines A, B, and C are fixed under the mg unit series, including g, mg, and mcg; Medicine D is fixed under the unit unit series, including unit, k unit, and m unit; and Medicine E is fixed under the mEq unit. Patient weight: When entering the medicine calculation window, operators should firstly or secondly input patient weight, which will be taken as independent information for calculation of medicine contents.
37
Attention This function of medicine calculation is only to provide a medicine calculator, while values in the list should not be related with the patient under monitoring. Thus the patient weight under this menu is different from the patient weight in the system; when the system refresh with a new patient, values in this menu will not be affected.
Figure 3-18Titration List The specific operations are as follows: In the Titration list, move the cursor to the Reference Item with the knob first, and then press the knob to select the required item. Dosage and Injection Speed are two options. Move the cursor to the Step Length and press the knob to select the step length in the range of 1~10. Move the cursor to the Dosage Type and press the knob to select the dosage unit. Move the cursor to the Page Up/Down and press and turn the knob to check the previous and next pages of the list. Move the cursor to theRecord and press the knob to output the Titration list data on the current display interface. Move the cursor to theQuit and press the knob to return to the Medicine Calculation menu.
38
Figure 3-19 Users Maintenance Language Selection: English or other language. Network Bed (MONITORS IP): means the IP bed number supervised by the monitor in the network. This IP is 200.200.200.X. (X means the network bed number.) CONNECT TRY: This function is used in the situation that the central machine cant be connected. Use this function for trial connection 4 minutes and you can see the connecting result. If the status bar on the bottom appears, the network bed number X OK means successful connection and X NO means the unsuccessful connection. COLOR SELF-DEFINE: It was used to setting the colors of waveform and parameters on the screen. There are 5 colors for option, i.e. red, yellow, green, cyan and white.
WAVE MODE: MONO/COLOR MONO: arithmetic-ladder for wave-drawing. COLOR: arithmetic-color for wave-drawing. It smoothes the wave. REC SETUP:ON/OFF It is the switch for printer. Turn it on if 8000D installed printer. NIBP FACTORY CAL (NBIP FACTORY CALIBRATE) This item is special for calibrating. Adjust it to the suitable value when cuff connect to the NIBP simulator. The default value of pressure is 150.
40
Chapter 4 Recording
General Information on the Recorder Configuration and method for recording Record information
Recorder compatibility:
Outputted waveforms run at 25mm /sec. or 50mm / sec; Maximally record two waveforms; Grid output function is optional English output; Real-time record time and waveforms are selected by users through menus; Automatic record interval is selected by users through menus, while waveforms are identical to real-time records;
Real-time record
Start recording waveforms from the moment you press the REC/ STOP button. Waveforms for real-time continuous record or real-time 8-second record are set by the system (normally only the first two waveforms are displayed) or set by users through menu. Please refer to relevant chapters for details. Under the Record output setting menu, users can select to concurrently print two waveforms or close one of the two, printing the other waveform only. If the print functions for both the two waveforms are disabled, the real-time record will only output the measured parameter values.
41
Attention During output process, the next parameter alarming output will be outputted after completion of the current output.
Remark record
Bed number Real-time record Alarm parameter, alarming time, and FREEZE time Parameter name and value Record time Waveform name Waveform amplitude (only for ECG waveforms) ECG lead, ruler, and gain
Normal service
While the recorder is under normal service with record paper outputted in uniform speed, users should not pull paper in order to protect the recorder. Dont use the recorder without paper loaded.
Insufficient paper
Dont boot the recorder when there is a reminder of add paper to the recorder in the information area. Please load qualified heat-sensitive record paper.
during paper loading or trouble shooting, the recorder door must be kept open.
Heat-sensitive head out of the recording position Recorder out of paper Communication error Continuously for over 30m record
Restart recorder Restart the recorder Restart the recorder Restart the recorder Restart the recorder Restart the recorder Restart the recorder
Recorder initialization error 1 Recorder initialization error 2 Recorder initialization error 3 Recorder initialization error 4 Recorder initialization error 7 Recorder initialization error 8
Too many alarms Low happens at the same time Recorder initialization in Low
Close alarm record Wait until completion initialization Restart the recorder of
43
Low Over high voltage Over high voltage Low Over low voltage Recorder error Over low voltage
Stop using the record until recovery of proper voltage Stop using the record until recovery of proper voltage Restart the recorder
communication Serial port Low communication error Error with RAM, ROM Low or CPU Guard In printing Low
If errors are still there after the recorder restarts, please contact with our sales engineer.
44
This symbol means the application part is of IEC 60601-1 type CF equipment, and designed with special electric shock resistant apparatus (especially with an F-type floating insulation apparatus for permissible leakage current), especially recommended for use during defibrillation period.
Warning During defibrillation period dont touch the relevant patients, beds or equipments.
Environment
Users should follow the following guides to ensure absolute safety of electricity installation. For an environment where the portable monitor is located, users should reasonably avoid vibration, dusts, corrosive or explosive gases, extreme temperature and moisture. In case installed inside a chamber, the front side must be given sufficient space for convenient operations, and while the chamber door is open, the rear side must be given sufficient space for easy repair. Besides, must make sure of air flow inside the chamber. The monitor, when working in an ambient temperature between 040, can meet the technical indexes, otherwise may have equipment accuracy affected or parts or circuits damaged. Moreover, there should be at least 2 inch (5 cm) of space reserved surrounding the monitor to ensure air flow.
Power Source
Please refer to the Product Specification chapter.
Monitor earthing
To protect patients and medical staffs, the portable monitor must has its cover connected with the earth; for such reason the monitor is equipped with a dismountable 3-line cable, which should be
45
plugged into a matching 3-line socket and further connected with the earth through the ground line of the power supply cable. In case of no 3-line socket, please consult with the electricity staffs of your hospital. Warning Dont connect the 3-line cable of this monitor with a 2-line socket. Connect the ground line with the equal-potential earthing terminal of the monitor. If unaware whether a certain equipment combination is risky in terms of equipment specification, for example, whether gathered leakage current is dangerous, users should consult with relevant manufacturers or specialists, so as to make sure the necessary safety of the relevant equipment will not be damaged by the proposed combination.
Equal-potential earthing
First level protection on the equipment has been contained in the house protective earthing system through earthing of the power socket. For heart or head internal check, this portable monitor must be individually connected with an equal-potential earthing system. One side of the equal-potential cable (potential balanced cable) should be connected with the equal-potential earthing terminal on the rear panel of the monitor, while the other side connected with one interface of the equal-potential system. In case of any damage to the protective earthing system, the equal-potential earthing system will take the safety function of protecting the earthing cable. Heart or head checks should be conducted within houses for medical use installed with protective earthing systems. Before each time of use, users should check whether the equipment is under good work status and pay attention the cable connecting patients and the equipment must be free from electrolytes pollution. Warning If the protective earthing system is instable, the monitor should be applied with internal power supply.
Condensation
During work period the equipment must be made sure of no condensation. When the equipment is shifted from one room to another room, condensation may be formed as the equipment is exposed in moistured atmosphere and different temperature. Warning If the monitor is used where there are flammable anesthetic agents, there may be explosion.
46
This means the applied component is Model CF and designed with the special anti-electroshock device (especially in the allowed leakage current is a F-shape ground-free isolation device) that is a durable defibrillator.
Protective earthing.
Power-up Shutdown
47
European standard Lead name R L F N C Color Red Yellow Green Black White
Figure 6-1
Attention For patient safety, all the leads must be connected with patient body. For 5-lead device, put the breast (V) electrode at one of the following positions (Fig. 0-4): V1, around the 4th frame at right side to the breast bone V2, around the 4th frame at left side to the breast bone V3, between V2 and V4 V4, around the 5th frame along middle line of the left clavicle V5, at front line of the left axilla, at the same horizontal position of V4 V6, at middle line of the left axilla, at the same horizontal position of V4 V3R-V7R, at right side of the breast, identical to those positions at left side VE, at apophysis of the xiphoid process; in case V leads are put on the back, the electrodes must be put at one of the following position: V7, around the 5th frame at back line of the left axilla on the back V7R, around the 5th frame at back line of the right axilla on the back
50
Figure 6-2
Figure 6-3
ECG leads
Attention In case electrodes are pasted correctly but ECG waveforms are not accurate, users need change the leads. Attention Disturbance from non-earthing equipments near patients or ESU distance may make troubles to waveforms. Characteristics of a good signal: High and narrow, without incisura High R-wave, fully above or below the baseline Pacing signal no higher than R-wave T-wave lower than 1/3 of R-wave P-wave significantly lower than T-wave To obtain 1mV of calibrated ECG waveform, users should conduct ECG calibration, in such case the screen will display Dont monitor patients during calibration. R
Figure 6-4
52
Figure 6-5 ECG Buttons Name of the First ECG Lead: The selectable leads include I, II, III, aVR, aVL, aVF and V. ECG using 5-lead, the selectable leads include I, II, III, aVR, aVL, aVF and V; ECG using 3-lead, the selectable leads include I, II and III. The leads on the ECG waveform should not have the same names, otherwise the system will automatically change the similar name into another. The 1st-ECG Waveform Gain: used to adjust the amplitude of ECG waveform. The gain of each calculation channel can be selected, which has such columns as 0.25, 0.5, 1 and 2 as well as auto mode. Auto mode means that the monitor can automatically adjust the gain. On the right side of each ECG waveform there is a 1-mv rod of which height and amplitude are proportional.
Attention The input signal being too strong, the wave crest may be truncated. At this time users can manually change the gain column of ECG waveform by reference to the actual waveform for fear of incompleteness of waveform. Filtering Mode: The cleaner or precise waveform can be obtained through filtering. There are three filtering modes for option. The unfiltered ECG waveform is shown in the diagnostic mode; the monitoring mode will possibly lead to the artifact filtering; the operation mode used in the surgery can reduce the artifact and interference from the electrosurgery unit. The filtering mode can be used in two channels and displayed on the upper part of the first ECG waveform. Warning Only in the diagnostic mode can the system provide the real signal that has not been
53
treated. In the filtering modes such as Monitoring and Operation, the ECG waveform will abnormally occur to the different extents. At this time the system can only provide the basic ECG status, and will produce greater influence on the analysis result of ST Segment. The analysis result of ARR may partially be affected in the operating mode, so it is suggested that efforts are made to monitor patients in the diagnostic mode when the interference is small. The Name of 2nd-ECG Waveform Gain: for details, please refer to . The 2nd-ECG Waveform Gain: for details, please refer to .
Attention The detected pacing signal displays on the upper part of the ECG waveform in the waveform area, which is expressed as .
Figure 6-6 ECG setting menu ECG Alarm Setting 1.1.1.5 Heart Rate (HR) Alarm: Select ON to give alarm prompt and storage when the heart rate alarm happens. will be prompted beside ECG. 1.1.1.6 Alarm level: three options: High, MED and Low, and High is for the most serious alarm. 1.1.1.7 Alarm record: Users can select On to print HR alarms when they happen 1.1.1.8 Alarm upper limit: used for users to set the upper limit for HR alarms 1.1.1.9 Alarm lower limit: used for users to set the lower limit for HR alarms
54
Alarms will happen once the HR values exceed the upper or lower limit. Adjustable ranges for HR alarm upper & lower limits are as follows: Highest upper limit HR adult HR infant HR newborn 300 350 350 Lowest lower limit 15 15 15 Adjustment length 1 1 1 step
Attention Users should set the alarm upper & lower limits based on the clinical conditions of every patient. Setting of the HR alarm upper limit is very important, and users should not set it too high but consider revision factors. The set HR alarm upper limit should not be over 20 beats/ minutes than patient HR. HR Source Users can select to check HR through ECG or PLETH (blood-oxygen volume recording waveform); if users select Automatic, the monitor will decide HR source based on signal quality; if users select All, the monitor will concurrently display HR and PR. In case PLETH is taken as the HR source, the PULSE reminder will be displayed together with pulse voice. In case PLETH is taken as HR source, no alarm judgment on HR but alarm judgment on PR will be conducted. In case All is selected, PR measurement values will be displayed in the right to SpO2 on the main screen, and HR & PR make alarms at the same time. Pulse voice will give priority to HR, as long as there is HR data, voice reminder will be there; only when there is no HR data, voice reminder will be subject with PR. Selection of HR calculation channel Channel 1 means the HR is calculated according to the first ECG waveform data. Channel 2 means the HR is calculated according to the second ECG waveform data. Auto means the monitor will automatically select the channel of calculating HR. Lead type: 5-lead or 3-lead Waveform speed: ECG Waveform scanning wave has three levels for option, such as 12.5, 25.0 and 50.0mm/s. ST Segment Analysis Select this item and enter the ST Segment Analysis menu. For details, please refer to the related content as follows. Arrhythmia Analysis Select this option and enter the Arrhythmia Analysis menu. For details, please refer to the related content as follows.
55
Other Setting Select this option and enter the ECG Setting>>OTHER SETUP menu as shown in the following figure.
Figure 6-7 ECG Setting Menu In this submenu there are the following functions: 1.1.1.10 ECG Monitoring Type: select the Normal Display to display the two ECG waveforms in 5-lead (it is possible to display one waveform in 3-lead). Select the Full-screen Multi-lead Display and six ECG waveforms will display in the screen waveform area and occupy six waveform positions; Attention In case that 3-lead type is set in the ECG Setting Menu, only Normal Display can be selected in the ECG Monitoring Type. 1.1.1.11 Pacing Analysis: when the pacemaker in the patient information management is set at ON, this option will automatically be ON, otherwise it will be OFF. Attention Where the patient with the pacemaker is monitored, the pacemaker should be started; where the patient with the pacemaker is not monitored, the pacemaker should be closed; when the pacemaker is started, the system will not start the ARR analysis of some types. For details, please refer to the introduction in the Arrhythmia Analysis. Warning To the patient with the pacemaker, it is a must to start the pacemaker pulse analysis function, otherwise it is possible to use the pacemaker pulse for calculation of normal QRS waveform so as to make the ECG Signal Weak alarm unable to be detected. Provided that the monitor is equipped with ST Segment Analysis and Arrhythmia Analysis, please refer to ST Segment Monitoring and Arrhythmia Analysis for the using method of this content. Attention The pacemaker analysis being started, no detection will be given to the arrhythmia (including PVCs enumeration) related to PVB (premature ventricular beats) and no ST segment analysis, either.
56
Technical alarms: Reminders ECG lead disconnected ECG V-lead disconnected or ECG C-lead disconnected ECG LL-lead disconnected or ECG L-lead disconnected ECG RA-lead disconnected or ECG R-lead disconnected ECG module initialization error ECG module initialization error 1 ECG module initialization error 2 ECG module initialization error 3 ECG module initialization error 4 ECG module initialization error 5
57
Causes
Alarming levels
Solution
ECG electrodes drop off from patient body or ECG cables drop off from the monitor
Low
Make sure electrodes, leads and cables are all connected properly
ECG High
Stop using the measurement function provided from the ECG module and inform biochemical engineers or our maintenance
ECG module initialization error 6 ECG module initialization error 7 ECG module initialization error 8 ECG module stopped communication ECG measurement module failure or communication failure Casual communication error Function error / safety
personnel
High
Ditto
High
Ditto if continues
the
failure
High
Stop using HR alarming and inform biochemical engineers or our maintenance personnel Keep patients silent, electrode connection reliable, and AC supply earthing well
ECG over-disturbance
Low
Reminders (including general alarming information): Reminders HR measurement out of scope Causes HR measurement values out of measurement scope Alarming levels High
when the switch is ON, ST Segment Analysis can proceed. Alarm Switch: Where ON is selected, the alarm prompt and saving will proceed when ST analysis result is alarmed; where OFF is selected, alarming will not happen, but will be prompted beside ST in the screen parameter area. ST Alarm will be triggered only when its measured value exceeds ST Alarm Upper Limit or ST Alarm Lower Limit. Alarm Level: used to set the ST Alarm Level according to three options such as High, Middle and Low. Alarm Record: when it is set at ON, the system will start the recorder for alarm record. Alarm Upper Limit: used to set the alarm upper limit of ST Segment, of which maximum upper limit is 2.0 and of which minimum upper limit must be more than -1.6 of the set lower limit. Alarm Lower Limit: used to set the alarm lower limit of ST Segment, of which minimum lower limit is 1.8 and of which maximum lower limit must be lower than -2 of the set upper limit. The adjustable range for upper limit and lower limit of alarm as follows: Max. Upper Limit ST 2.0mv Min. Lower Limit -2.0mv Single Adjustable Quantity 0.1mv
Determine the ST Segment AP (Analysis Point): Select this option to enter the Determine the ST Segment AP window and set the values at ISO and ST. ISO (BP: base point): set the baseline point. Power on time is ste at 78 ms. ST (SP: starting point): set the measuring point. Power on time is set at 109ms.
Figure6-9 Determine the ST Segment AP ISO and ST are two measuring points of ST Segment, which are adjustable.
59
R wave crest point is the reference point in setting of ST measuring point (as shown in the following figure): R Wave
BP: ISO
} ST Value }
ST Measuring Point Initial Value +109 +109 ms
-78 ms
ms
Figure 6-10 ST Analysis Point ST measured value of each HB composite waveform is the vertical distance between this waveform and the crossing of two measuring points. Attention Where obvious changes happen to the patients HR or ECG Waveform, it is necessary to adjust the ST measuring point in the following methods. Method of Adjusting ISO and ST Adjust the values by turning the knob. Setting the measuring point of ST Segment, please open the Determine Analysis Point window and the window will show QRS wave-group module (if the channel is not opened, ST Analysis Switch OFF will be prompted). The location for high-brightness line in the window can be adjusted. Select the ISO or ST first, and then turn the knob both leftward and rightward to move such a line in parallel so as to determine the reference point or measuring point. Attention The abnormal QRS wave group will not be taken into consideration when ST segment is analyzed.
Physical alarms:
Prompt Messages ST1 too high The measured value for ST Segment of Channel 1 is higher than the set alarm upper limit. ST1 too low The measured value for ST Segment of Channel 1 is lower than the set alarm lower limit. ST2 too high The measured value for ST Segment of Channel 2 is higher than the set alarm upper limit. ST2 too low The measured value for ST Segment of Channel 2 is lower than the set alarm lower limit. Selectable by users Selectable by users Selectable by users Selectable by users Causes Alarm Levels
Technical alarms
Prompt Message ST Alarm Limit is wrong. Causes Function failure safety Alarm Level High Solutions Stop use of ST SEGMENT alarm and advise the biomedicine engineer or our companys servicemen.
The measured value for ST Segment of Channel 1 exceeds the measured range.
The measured value for ST Segment of Channel 2 exceeds the measured range.
High
The functional default of arrhythmia monitoring in the system is the off state. Users can start this function as needed.
Arrhythmia monitoring can arouse the doctors attention to the patients cardiac rhythm and give an alarm through test and classification of arrhythmia and HB abnormality.
61
This monitor can perform 13 kinds of arrhythmia analysis. In arrhythmia analysis, the system will save the latest 60 alarm events (the single-channel ECG waveform four seconds before and after alarm). The operator can edit the arrhythmia events through this menu.
Figure 6-11 Arrhythmia Analysis Menu Arrhythmia Analysis: during monitoring it can be set at ON and during default, OFF; Alarm Switch: select the ON and the alarm prompt and saving will proceed; select the OFF do beside PVCs in the screen parameter area. and PVCs alarm wont start, but prompting Alarm Level: there are such three options as high, middle and low. High means the most serious PVCs alarm. Alarm Record: select the ON and the recorder will output during PVCs alarm. Alarm Upper Limit: PVCs alarm is based upon the set alarm upper limit. The alarm will happen when PVCs exceeds the upper limit.
Technical alarms
PROMPT MESSAGE PVCs alarm limit is wrong. Function safety failure CAUSES ALARM LEVEL High Stop use of PVCs alarm and advise the biomedicine engineer or our companys servicemen. SOLUTIONS
ARR self-learning: press this button to trigger a learning course and the screen information area will show Learning ARR. ARR Alarm Setting: Set the Arrhythmia Alarm. In this menu ALM means the alarm switch; REC means the alarm record switch; LEV means the alarm level.
Figure 6-12 Arrhythmia Alarm Setting Users can select the Alarm Full-On to set the Arrhythmia Alarm at ON. Selecting the Alarm Full-Off to set the Arrhythmia Alarm at OFF. Besides, Record Full-On can be selected to set the alarm record switch fully at ON and Record Full-Off can be selected to set the alarm record switch fully at OFF. Uniform change can happen to Arrhythmia Alarm Level due to change of Alarm Level. Arrhythmia Review: Select this option and you can view and edit the patients arrhythmia information.
The latest saved arrhythmia events are listed in the window (one page can show 10 events and at most 6 pages can display). Page Up/Down: to observe the list of arrhythmia events in other pages. Cursor Movement: to move the cursor to select the arrhythmia events in the list. Waveform: press this button and the window will show the waveforms, occurrence time and parameters of the selected arrhythmia events . Arrhythmia Waveform Review Window: Page Up/Down: to observe waveforms of other arrhythmia events. Movable Waveform: to observe all waveforms of arrhythmia events in 8 seconds. Waveform: to output the arrhythmia event waveform through the recorder. Quit: to return to the window of arrhythmia event list.
Attention In the event that the number of arrhythmia event is more than 200, the monitor will retain the latest instead of the earliest. As for the monitor with the power-fail saving function, it can save 200 arrhythmia events with power-fail.
Arrhythmia Alarm
In case of arrhythmia, the system will give an alarm sound. If Alarm Switch is ON, the system will give an alarm sound and the indicator light; if Alarm Record Switch is ON, the alarm record will be outputted (ECG waveform of channels analyzed 4 seconds before and after alarming). The alarm or prompt message related to arrhythmia analysis is as shown in the following table. Physical alarms:
Prompt Message Applied Patient Type Conditions Alarm Level 64
Asystole
Full
No QRS waveforms successivefully for 4 seconds F Wave successively for 4 seconds or the continuous ventricular beats are
VFIB/VTAC
No pacing
greater than the number upper limit of serial ventricular beats (5pcs), of which R-R time is smaller than 600ms.
Multiple PVS
No pacing
The number of serial PVS greater than or equal to 3 and smaller than 5.
Two PVS
No pacing
PVS bigeminy
No pacing
PVS bigeminy
PVS trigeminy
No pacing
PVS trigeminy HR<100, RR interval < 1/3of average interval, after which there is a
R ON T
No pacing
compensation interval of which length is 1.25 x R-R mean (next R waveform is moved up to the previous T wave.)
Single PVS
No pacing
Tachycardia
Full
Bradycardia
Full
Successively 5 QRS waveforms and RR mean greater than 1.5s Where HR is smaller than 100times/minute, HB is not detected in the
1.75 times of time for RR mean; or where HR is greater than 100times/minute, pulsation is not detected within 1 second.
In the 1.75 times of time for RR mean there are no QRS waveform and Selectable by users pacing pulse (only for the patient with the pacemaker). In case of pacing pulse there is no QRS wave in the 1.75 times of time for Selectable RR mean (only for the patient with the pacemaker). by users
Applied Patient Type: Full means the arrhythmia analysis that can be given to the patient with the pacemaker as well as that to the patient without the pacemaker. Non-pacing: means the arrhythmia analysis that can be given to the patient without the pacemaker. Pacing: means the arrhythmia analysis that can be given to the patient with the pacemaker. Prompt Message:
Prompt Message ARR Learning Cause QRS wave template necessary in ARR analysis is forming. 65 Alarm Level No Alarm
66
N F
Attention Install the white and red electrodes in a diagonal line so as to obtain the best RESP wave. Need keep the liver and heart area out of the line formed by such electrodes, so as to avoid artificial discrepancy generated from heart cover or pulsatile blood, which is very important for newborns.
amount
RR infant/ newborn 150 0 1 Choke alarm: Users can set the time to judge patient choke; 10-40 seconds are optional, each rotation of knob will increase/ decrease 5 seconds. Waveform speed: Three optional speeds, 6.25mm/s, 12.5mm/s and 25.0mm/s Waveform amplitude: Users can set enlarged display of RESP waveforms under four optional enlargement rates: 0.25, 0.5, 1.0, 2.0, 3.0 4.0, and 5.0. Calculation mode: there are AUTO and MANUAL for option. When AUTO is selected, such menu items as Adjust Upper Dashed Position and Adjust Lower Dashed Position are not available and the monitor automatically calculates the RR value. When MANUAL is selected, users can use such menu items as Adjust Upper Dashed Position and Adjust Lower Dashed Position to adjust the two dashed lines in the respiratory waveform area respectively. The position indicated by the dashed line will be used by the monitor as the upper and lower limit values of RR. Adjust the positions for upper and lower dashed lines: when the calculation mode is manual, users can pitch on the Adjust Upper Dashed Position or Adjust Lower Dashed Position by the knob first, and then turn the knob to adjust the two dashed positions in the RESP waveform area respectively. The positions indicated by two dashed lines will be used to calculate the upper and lower limit values of RR. Default Setting: select this option to enter the RESP Default Setting dialog box. Users can select the Factory Default Setting or User Default Setting. On quitting the dialog box, the system will eject a determined dialog box to have users confirm their selection.
RESP choke
High
Technical alarms Reminder RESP alarm limit error Cause Function/ failure safety Alarming level High Solution Stop using the RESP alarm function and inform biochemical engineers or our maintenance personnel
Reminders (including general alarming information): Reminder RR measurement out of scope Cause RR measurement value out of measurement scope Alarming level High
70
BOS/Pulse Monitoring
Warning The cable for the equipment of electrosurgery cant be twisted together with the sensor cable. Warning Please dont place the senor on the limb with arterial duct or vein injection syringe.
71
Attention Please dont place SpO2 detector and cover on the same limb for measurement of blood pressure, because in the course of measuring blood pressure the vascular obstruction will affect the BOS reading.
Insert the connector at the other side of the sensor cable into the SpO2 hole of the SpO2 module.
Figure 8-1 Sensor Positioning SpO2 volume recording measurement for newborns The SpO2 volume recording measurement process for newborns is basically same as that for adults. The following is introduction on the SpO2 detector for newborns and its positioning method. 1. SpO2 detector for newborns A SpO2 detector for newborns consists of a Y-shape SpO2 probe and a cover. Respectively insert the LED terminal and PD terminal of the Y-shape SpO2 probe into the upper and lower slots of the cover (as shown in Pic 13-2), and the completed combination is shown in Pic 13-3.
Y-shape
SpO2
Figure8-3 SpO2 detector for newborns (2) 2. Positioning of a SpO2 detector for newborns
73
Clamp the SpO2 detector onto a hand/ foot of a newborn (as shown in Pic 13-4). Hold the SpO2 detector, pull on the tape and put the V-edge along one side of the tape inside the V-slot along the corresponding side of the cover, properly extend the tape (for about 20mm) and put the V-edge along the other side of the tape inside the V-slot along the corresponding side of the cover, release the tape, make sure the two sides of V-edges of the tape have been matched with the cover, then insert the tape into the first latch for securing, as shown in Pic 13-4. In case the tape is too long, it may be inserted into the second latch. Users must follow such a method to secure the SpO2 detector so as to make sure the opto-electronic device stat at a position to be correctly faced. Meanwhile, users must pay attention to control pulled length of the tape, otherwise measurement will be not so accurate and blood cycling may be seriously blocked.
Figure 8-4 Positioning of a SpO2 detector for newborns Attention If the test position and detector fail to be positioned correctly, SpO2 readings may be inaccurate, and possibly pulse wave will not be searched for SpO2 monitoring. In such scenario users should re-position the test position and detector. Over-movement of the measurement position may cause inaccurate measurement results. In such case users should calm down the relevant patient or change the measurement position so as to reduce effects of over-movement on measurement. Warning Over long time of continuous monitoring, users should check the end recycling and skin conditions at the measurement position every 2 hours. In case any bad change is found, users should timely change the measurement position. Over long time of continuous monitoring, users should periodically check positioning of the detector, avoiding that any movement changes the detector position and further affects measurement accuracy
Light radiation from outside; Improper installation of the sensor or improper touching position with objects; improper sensor temperature (ideal temperature should be 28-42); The sensor is put onto body with blood pressure cuff, arterial duct or vein tube; Contents of non-functional Hb such as COHb and MetHb; SpO2 over low; Bad microvascular perfusion at the test position; Shock, anemia, low temperature and application of vessel shrinking medicines, which all can reduce the arterial blood flow to a non-measurable level;
Measurement is also up to absorption of lights with special wavelengths by oxyhemoglobin and deoxygenated hemoglobin. Existence of other materials that absorbs the same wavelengths, such as carbonated hemoglobin, hemoglobin, methylene blue and indi carmine, will make artificial or low SP02 values.
Figure8-5 SP02 Setting menu Warning Setting the SpO2 alarm upper limit to be 100% means to release the upper limit. However, high SpO2 level will make early-born infants infected with retrolental fibroplasias, thus the SpO2 alarm upper limit must be carefully selected based on common acknowledged clinical practice.
75
Alarm level: used to set alarm levels and, during Sp02 alarming, for alarm reminder and saving. Options include High, Middle, and Low; High for the most serious alarm.
Alarm record: If On is selected, the recorder will output during Sp02 alarming.
SpO2 Alarm Upper/Lower Limit: according to the set upper/lower limit, alarm will happen when SpO2 is higher than upper limit or lower than lower limit.
PR Alarm Upper/Lower Limit: according to the set upper/lower limit, alarm will happen when PR is higher than upper limit or lower than lower limit.
Sp02 & PR adjustable limits: Parameter Sp02 PR Waveform Speed Max upper limit 100 254 Min lower limit 0 0 Adjustable amount each time 1 1
The scanning speed of SpO2 volume recording waveform is provided with 12.5 and 25.0mm/s for option. Pulse Volume: In volume there are 1, 2, 3 and 4 for option. Calculation Sensitivity: In calculation sensitivity there are low, middle and high for option. Default Setting: Select this option to enter SPO2 Default Setting dialog box where users can select Factory Default Setting and User Default Setting. When the dialog box is quitted after selection of any option, the system will eject a dialog box for users to select and confirm.
Possible physical alarms, technical alarms and reminders during Sp02 module measurement are as listed here: Physical alarms: Reminders SPO2 over-high SpO2 over-low PR over-high PR over-low Causes SpO2 measurement value higher than the set upper limit SpO2 measurement value lower than the set lower limit PR measurement value higher than the set upper limit PR measurement value lower than the set lower limit Alarming levels User defined User defined User defined User defined
Technical alarms
Reminders Causes Alarm Levels SpO2 sensor drops down from the patient or monitor. Remedial Measures Ensure the sensor is fixed on the patients Low finger or other locus and that the monitor is normally connected with the cable.
SPO2 module initialization wrong SPO2 module initialization wrong 1 SPO2 module initialization wrong 2
SPO2 module initialization wrong 3 Stop use of SpO2 module measurement SpO2 module wrong High function and advise the biomedicine engineer or our User Service Dept.
SPO2 module initialization wrong 4 SPO2 module initialization wrong 5 SPO2 module initialization wrong 6 SPO2 module initialization wrong 7 SPO2 module initialization wrong 8 SPO2 communication stop SPO2 module module
SpO2
module
wrong
or
Stop use of SpO2 module measurement High function and advise the biomedicine engineer or our User Service Dept.
communication mistaken
communication wrong
77
Stop use of SpO2 module measurement SPO2 alarm limit wrong Function/safety faulty High function and advise the biomedicine engineer or our User Service Dept. Stop use of SpO2 module measurement PR alarm limit wrong Function/safety faulty High function and advise the biomedicine engineer or our User Service Dept.
Cleaning:
Having cleansed the surface of the sensor with the cotton ball or cotton cloth soaked with medical alcohol, dry it with the dry cloth. The luminotron and receiver of the sensor can be cleaned in the same method. The cable can be cleaned and sterilized with 3% of hydrogen peroxide or 70% of isopropyl alcohol. Active reagent can also be used for this purpose. However, the joint cant be soaked in the above solution.
78
2. In accordance with the following method (Pic 14-1), tie the blood pressure cuff on upper arm of upper leg of a patient. Confirm the cuff is fully vented. Select a cuff in proper size for the patient, and make sure the mark is just along the proper vein and cuff tie the body non-toughly, otherwise remote body part may have color change or even ischaemia.
Attention Cuff width should be 40% of arm perimeter (50% in case of newborns) or 2/3 of upper arm length. Width of the pumping part of a cuff should be as long as to surround 50%~80% of the arm. Cuffs in improper size will generate wrong readings. In case of size problem with a cuff, users should change it with a bigger one so as to reduce mistakes. Ault/ newborn/ infant recyclable cuff Patient type Newborn Infant Adult 1 Adult 2 Leg Body perimeter 1019cm 1826cm 2535cm 3347cm 4666cm Cuff width 8cm 10.6cm 14cm 17cm 21cm Length of pumping pipe 1.5m or 3m
Newborn/ infant one-time cuff Size 1 2 3 4 Body perimeter 3.15.7cm 4.38.0cm 5.810.9cm 7.113.1cm Cuff width 2.5cm 3.2cm 4.3 cm 5.1cm Length pipe of pumping
1.5m or 3m
Check the cuff edges are between the <-> marks; otherwise users should change with a more proper cuff. 3. 1. Connect the cuff with a pumping pipe. Make sure the body part used for pressure
80
measurement is at the same horizontal level with patient heart, and if failing to realize this, users should apply the following correction method to correct the measurement results: In case horizontal level of cuff is higher than that of heart, add 0.75mmHg (0.10kPa) onto the displayed value for each cm difference. In case horizontal level of cuff is lower than that of heart, deduct 0.75mmHg (0.10kPa) onto the displayed value for each cm difference. 4. Confirm correctness of the monitoring mode (as displayed on the information area); if requiring to change the monitoring mode, users need go to the Patient Information Setting item under the Main menu and change Patient Type. 5. Select the measurement mode under the NIBP menu. Refer to the following Operational Guide for details 6. Press the START button on the front panel to start pressure measurement.
Operational guide
1. Conduct one time of Automatic measurement Enter the NIBP Setting menu, select a proper time interval at the Time Interval item, and press the START/STOP button on the front panel. Then the system will start automatic pumping measurement in the specified time interval. Warning If NIBP measurement under the Automatic mode lasts too long, body touching with the cuff may have allergic purpuras, ischemia and neural injury. During monitoring on patients, users should often check color, warmness and sensitiveness of remote body parts. Once any abnormal phenomenon is found, users should put the cuff at another location or immediately stop measuring blood pressure. 2. Stop automatic measurement At any moment during the automatic measurement process, press the START/STOP button will stop the automatic measurement. 3. Conduct one time of manual measurement Enter the NIBP Setting menu, select the Time Interval item and set its value as Manual, then press the START/STOP button on the front panel so as to start manual measurement. During spare time of an automatic measurement, press the START/STOP button will start a manual measurement; then if users press the START/STOP button again, the manual measurement will stop and the automatic measurement will continue. 4. Conduct a manual measurement during automatic measurement process Just press the START/STOP button on the control panel. 5. Stop a manual measurement Re-press the START/STOP button on the control panel.
81
6. Conduct a continuous measurement Enter the NIBP Setting menu and select the Continuous item to start a continuous measurement, which will always last 10 minutes.
Warning If NIBP measurement under the Automatic mode lasts too long, body touching with the cuff may have allergic purpuras, ischemia and neural injury. During monitoring on patients, users should often check color, warmness and sensitiveness of remote body parts. Once any abnormal phenomenon is found, users should put the cuff at another location or immediately stop measuring blood pressure. 7. Stop continuous measurement At any moment during the continuous measurement process, press the START/STOP button will stop the continuous measurement.
Attention In case of suspecting reading accuracy, users should take possible methods to check life signs of patients before checking the monitor, Warning In case any liquid is sprayed onto the equipment or its accessories, especially when the liquid may enter the tube or monitor, please contact with the maintenance department of your hospital.
Measurement restriction
Vibration measurement has its restriction subject with patient conditions. This measurement method looks for regular pulse waves generated from arterial pressure, so when patient conditions make this wave detection method hard to work, measured values are no more reliable and measurement time last longer. Users must understand the following cases will disturb the measurement method, making measured press unreliable or measurement time extended. In such cases, patient conditions disable measurement to be continued. Patient movement In case a patient is moving, shaking or convulsing, measurement will be unreliable or even impossible; as such scenarios will disturb detection of arterial pulse and extend measurement time. Arrhythmia In case a patient shows irregular heartbeats resulted from arrhythmia, measurement will be unreliable or even impossible, while measurement time will also be extended.
82
Heart-lung machine If a patient is connected with an artificial heart-lung machine, measurement cant be realized. Pressure change Within certain time if the patient blood pressure immediately changes while users are analyzing arterial pulse so as to obtain measurement values, measurement will be unreliable or even impossible. Serious shock In case a patient is under serious shock or extreme low temperature, measurement will be unreliable as reduction in blood flowing peripherally will result reduction in arterial pulse. HR limits In case of HR lower than 40bpm or higher than 240bpm, no blood pressure measurement can be done.
NIBP
Measured Value
16: 50 mmHg
BP Unit
108
Manual
84 70
Measuring Mode
NS 160 90 NS 160 90
Prompt Area
Manual Measurement
83
Figure 9-2 NIBP Setting menu NIBPAlarm Setting Where ON is selected, the alarm prompt and saving will proceed when the pressure will be is alarmed; where OFF is selected, alarming will not happen, but prompted beside NIBP in the screen parameter area. Alarming levels: Optional levels are High, Middle and Low , where High is the highest alarm. Alarm record: Users can select On to output through recorder when blood pressure alarms happen Pressure alarms are based on the specified upper & lower limits; there will be alarms once pressure is higher than the upper limits or lower than the lower limits. Systolic pressure, average pressure and diastolic pressure can have their respective alarming treatment.
Adjustable range for alarm upper & lower limits: Adult Systolic pressure 40270 mmHg Diastolic pressure 10215 mmHg Average pressure 20235 mmHg Infant Systolic pressure 40200 mmHg Diastolic pressure 10150 mmHg Average pressure 20165 mmHg Newborn
84
Systolic pressure 40135 mmHg Diastolic pressure 10100 mmHg Average pressure 20110 mmHg Reset Press this button will reset the pumping value of the blood pressure pump to be default setting. When the blood pressure pump works unwell while the monitor doesnt remind of any problem, users are recommended to use this button, as it will bring the blood pressure pump for self-detection and automatic recovery during any accidental pumping failure. Continuous measurement Start continuous measurement After this option is selected, the menu will automatically disappear and the continuous measurement can be done at once. Time interval Time interval (Unit: minute) for automatic measurement: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480 minutes, Manual, and Continuous. After users select an interval, there will be a display of Please press the START/STOP button in the NIBP reminder area, then users just press the button to start pumping for the first time of automatic measurement. To end the automatic measurement and return to the manual mode, users need only select Manual during the measurement interval. Blood pressure unit: mmHg or kPa Calibration The manufacturer recommends to use pressure meter or mercurial sphygmomanometer with calibrated precision higher than 1mmHg for calibration. Users can select the Calibration item to start calibration, while this item turns to be Stop calibration; if at such moment press the button, the system will stop calibration. Warning Calibration for NIBP measurement should be done every two years (or conducted following the maintenance plan of your hospital). Please follow the following details to check its performance.
Gas pipe
Metal vessel Round pump Figure 9-3 Connection for NIBP calibration Gas leakage detection This button is used to detect gas leakage from the NIBP measurement pump. Users, after connecting with a NIBP cuff, can use this button to start NIBP pumping process so as to observe whether the NIBP gas route is sealed well. If such a gas leakage test gives a good result, the system will make no reminder; otherwise there will be the corresponding error reminder in the NIBP information area. Default Setting Users can select this item to enter the NIBP Default Setting dialogue box, and then further select Manufacturer Default or User Default. After making a selection, the system will eject a dialog box for confirmation of selection by users. Warning This gas leakage test, different from as described in the EN 1060-1 Standard, is only for users to simply detect gas leakage during NIBP pumping process. In case the system shows there is NIBP gas leakage, please contact with our maintenance engineers.
In about 20 seconds, the system will automatically open the air bleeder, indicating completion of gas leakage measurement. No reminder information displaying on the NIBP parameter area doesnt mean no gas leakage within the system. Display of Pump leaking means there is possible gas leakage with the gas route, in such case operators should check whether there is any loose connection, and after making sure of no more loose connection re-do the gas leakage detection; if the error reminder is still displayed, please contact with the manufacturer for repair
Cylinder
Gas pipe
Cuff
Stop using the NIBP module alarming function and High inform biochemical engineers or our maintenance team Stop using the NIBP module alarming function and High inform biochemical engineers or our maintenance team Stop using the NIBP module alarming function and High inform biochemical engineers or our maintenance team
NIBP error
self-detection
Sensor
maintenance team If the failure continues, stop using the NIBP High measurement function and inform biochemical engineers or our maintenance team
NIBP error
communication
Cuff loose or out of connection Gas leakage with cuff pumping pipe
Cuff isnt properly wrapped or no cuff Damage with cuff, pipe or connector Fail to obtain pressure
Low
Wrap the cuff properly Check and change the part with leakage, and if
Low
necessary also inform biochemical engineers or our maintenance team Check whether there is pipe wrapping; if the
stable
Low
problem
continuous,
inform
biochemical
Low
Reset the NIBP measurement module; and if the High problem continuous, stop using the NIBP measurement module and inform biochemical engineers or our maintenance team Make sure the patient is silent, without movement Re-measure. If the problem continuous, stop High using the NIBP measurement module and inform biochemical engineers or our maintenance team Low Stop the patient from movement
exceeding upper
Big signal noise or Arm movement irregular PR due to arm movement Pressure exceeding Overpressure protection specified limit Signal saturation Significant movement upper Low
88
Pump leakage
Check and change the part(s) with leakage, and if Low necessary, inform biochemical engineers or our maintenance team Stop using the NIBP measurement function and High inform biochemical engineers or our
pump
running failure Current cuff type Wrong cuff type not suitable for Low
maintenance team
patient Measurement time over 120 seconds Measurement time out (for adults) or 90 seconds newborns) NIBP reset error Module improperly During measurement system Measurement error execute measurement analysis calculation or can the not High Check the cuff and re-measure while making sure the patient is silent during monitoring reset High Re-use the reset function (for High Re-measure methods or apply other measurement
Reminders (displayed in the reminder area below NIBP values): Reminders Causes Manual measurement Manual measurement in process Continuous measurement Continuous measurement in process Automatic measurement Automatic measurement in process Please press START After you select the measurement time interval Measurement stopped During measurement users press the START button to stop measurement Calibration Calibration in process Calibration stopped Calibration completed Gas leakage detection Gas leakage detection in process Gas leakage detection stopped Gas leakage detection is stopped Module reset Reset after NIBP module is loaded Manual reset NIBP reset (activated by users) in process Fail to reset Fail to reset
Alarming levels
No alarm
Dont compress the rubber pipe on a cuff. Keep water or cleaning liquids out of the connector socket in the front of the monitor, otherwise the equipment may be damaged. During monitor cleaning, users need only sweep the outer surface of the connector socket instead of its inner surface. In case a recyclable cuff is disconnected with the monitor or being cleaned, users should locate the cover cap above the rubber pipe so as to prevent any liquids from entering the rubber pipe and being absorbed into the module.
To put the rubber bag back into the cuff, users should put the rubber bag near the cuff opening side, making the rubber pipe aligned with the long opening of the cuff, then vertically roll the rubber bag and insert it into the long opening, hold on the rubber pipe and cuff, and then shake the whole cuff until the rubber gag is positioned exactly. Insert the rubber pipe into the cuff, letting it go through the hole liner and extend out.
90
Figure 10-1 TEMP Setup Alarm Setting Alarm Switch: Select ON and the alarm prompt and saving will proceed when TEMP is alarmed; select OFF and no alarm will happen, but will be prompted beside TEMP in the screen parameter area. Alarming levels: High, Middle or Low to be selected by users to set alarming levels. Alarm Record: it is mainly used in starting/closing the output function of TEMP alarm record. If On is selected, the present TEMP alarm will be outputted through the recorder. TEMP alarms are based on the specified upper and lower limits; upon temperature higher than the upper limits or lower than the lower limits, there will be alarms. Adjustable range of TEMP alarm upper & lower limits: Parameters TEMP TEMP unit: or Max upper limit 50 122 Min lower limit 0 Adjusted per time 0.1 amount
Default Setting: Please refer to the ECG Default Setting in the ECG/TEMP Monitoring.
Technical alarms Reminders Sensor disconnected Alarming limit error Causes TEMP cable disconnected from the monitor Function/ safety failure Alarming levels Low High Solutions Make sure the cable is reliably connected Stop using the TEMP alarm function and inform biochemical engineers or our maintenance personnel
Reminders: Reminders TEMP Measurement out of scope Causes TEMP measurement value out of measurement scope Alarming levels High
93
Select the Module On/Off Setting option in the System Setting menu and set the CO2 on-off to be on (by marking with a tick ). The following figure (the present figure is in demonstrating mode) will come out:
CO2 module adopts automatic (Autorun) command measuring modes with a waveform sampling rate of once per 31msc.
CO2 settings
CO2 switch: On; Off. Alarm switch: Select On, and alarm prompt will be provided in case of CO2 alarms; select Off, and no alarm will be given. Alarm level: Three options, High, Middle and Low, are provided. High means the most serious alarms. Gain: 0 .25, 0.5, 1, 2, 4 and Auto CO2 alarms are done according to set upper and lower limits. When CO2 is higher than the upper limit or lower than the lower limit, alarms will be sent out. Pressure unit: mmHg, kpa or %. Oxygen compensation: 5~100 Atmospheric pressure: 400~850 mmHg (may be adjusted according to local geographic positions) Zero calibration: Zero calibration should be done before monitoring CO2. Select Autozero and then Wait for 30 seconds will appear on the interface; after 30 seconds, the system will automatically cancel Wait for 30 seconds. Then CO2 monitoring data will be more precise.
Attention
To guarantee higher precision of CO2 data, Zero calibration should be done each time after the module is plugged into the monitor.
Attention
95
If CO2 is not monitored, the CO2 module should be turned to Off state so as to prolong the service life thereof.
96
Chapter 12
12.1 General information
IBP Monitoring
This chapter mainly introduces invasive blood pressure (IBP) monitoring methods and contents relevant to maintenance and cleaning of accessories. STAR8000D portable-type multi-parameter monitor can be directly used for measuring vascular pressures (diastolic pressure, systolic pressure and mean blood pressure). The following waveforms can be displayed:
Definition arterial pressure pulmonary arterial pressure central venous pressure right atrial pressure left atrial pressure intracranial pressure
Warning
In connection and application, accessories should be avoided from contacting metal parts connected with electric apparatus.
Warning
Users, when connecting the monitor with a high-frequency surgical instrument, should avoid the sensor and cable of the monitor from contacting the high-frequency surgical instrument so as to prevent patients from being burnt in case of electricity leakage.
Warning
97
Attention
Only the pressure sensor specified in the Manual can be used. The specified sensor has shock-proof function (with resistance to leakage current) and can prevent influence of cardiac defivrillators. It can be used in surgeries. When the patient is in the defibrillation period, the pressure wave may exhibit temporary disorders; but after defibrillation, the monitor will work normally and the operation mode and user configuration of it wont be affected.
Warning
Before monitoring, the sensor should be examined for normality assurance. If the sensor is pulled out of the jack, an error-warning message, IBP sensor detached, will appear on the screen and alarm sounds will be sent out.
Attention
Sensors, new or used, should be regularly calibrated according to hospital practice.
Warning
If liquid (not solution applied to the pressure pipe and the sensor) is splashed on the instrument or accessories, especially when the liquid may enter the sensor or the monitor, place contact with the maintenance department of your hospital.
1 2 3
Insert cables into corresponding sockets and check to ensure the monitor has been plugged in. Have the pressure pipe and the sensor prepared. Fill the system full with physiological saline solution to make sure no bubble exists therein. Connect the patient catheter to the pressure pipe and make sure that there is no air in the catheter and the pressure pipe or the sensor.
Warning
If bubbles are found in the pressure pipe or the sensor, flush the system with the perfusion liquid. 4 5 6 Position the sensor on the same level as the heart, approximately on the midaxiallary line Confirm correct ruler names have been chosen. See the following section for details. Do zero-adjustment of the sensor. See the following section for details.
98
anticoagulant saline pressure sensor drip pipe valve patient joint three-way joint interface cable of pressure sensor
STAR8000D
Figure 12-1 IBP Monitoring
Settings can be done on the following items: Alarming switch: Where ON is selected, the alarm prompt and saving will proceed when the IBP (invasive blood pressure) is alarmed; where OFF is selected, alarming will not happen, but will be prompted beside IBP in the screen parameter area. Alarming levels: Optional levels are High, Middle and Low. Channel 1 pressure name: Six options including ART, PA, CVP, RAP, LAP and ICP are provided. In the IBP measurement range, the waveform position in the screen is adjusted by
99
users through the Upper Scale Limit or the Lower Scale Limit item. Channel 1: Upper SP alarm limit: used for setting the upper alarm limit; Lower SP alarm limit: used for setting the lower alarm limit. Upper MP alarm limit: used for setting the upper alarm limit; Lower MP alarm limit: used for setting the lower alarm limit. Upper DP alarm limit: used for setting the upper alarm limit; Lower DP alarm limit: used for setting the lower alarm limit. Channel 2 pressure name: Six options including ART, PA, CVP, RAP, LAP and ICP are provided. In the IBP measurement range, the waveform position in the screen is adjusted by users through the Upper Scale Limit or the Lower Scale Limit item. Channel 2: Upper SP alarm limit: used for setting the upper alarm limit; Lower SP alarm limit: used for setting the lower alarm limit. Upper MP alarm limit: used for setting the upper alarm limit; Lower MP alarm limit: used for setting the lower alarm limit. Upper DP alarm limit: used for setting the upper alarm limit; Lower DP alarm limit: used for setting the lower alarm limit. Pressure unit: Two options, mmHg and kPa, are available. Zero calibration: Perform IBP zero calibration and prompt IBP Zero Calibration at the left upper corner of the screen. Exit: Select this item to return to the main screen.
Warning
When setting alarm limits, users should confirm the item to be set.
Attention
Users should guarantee that zero calibration has been done on the sensor before the measurement; otherwise the instrument has no effective zero value, which may lead to inaccuracy of measured data. Once the measured data exceed the alarm limits, the alarm will be triggered. IBP alarm limits: Pressure Scale Name Max Upper Limit (mmHg)
ART PA CVP RAP LAP ICP 300 120 40 40 40 40
1 1 1 1 1 1
100
pressure sensor
three-way joint
T-shaped connector
standard blood pressure gauge three-way stop cock pipeline with a length of about 25cm
1.
Close the three-way stop cock which is opened to the atmosphere for zero calibration.
2. 3. 4.
Connect the pipeline with the blood pressure gauge. Confirm that connection to the patient has been off. Connect a three-way stop cock with the three-way joint that hasnt been connected to the patient catheter (when the patient is being monitored). Connect a syringe to one end of the three-way stop cock and connect the blood pressure gauge and the pipeline with the other end. Open the end open to the blood pressure gauge. Select the channel to be calibrated in the pressure calibration menu and adjust the pressure values of the channel to be calibrated. Charge gas to raise the scale of the mercury column to the set pressure value in the menu. Repeatedly adjust until values in the menu equal to pressure values in the blood pressure gauge. Press once the calibration button in the calibration menu to command the instrument to begin calibration. according to prompt information.
5.
6.
7. 8.
9.
10. Wait until the calibration ends. Make corresponding countermeasures to be taken 11. Detach the pipeline of the blood pressure gauge and the added three-way stop cock
after completion of the calibration. The IBP waveform area provides scales for waveforms. Two dash line of each IBP waveform, from the upper to the lower, respectively represents the upper-limit scale and the lower-limit scale of the waveform. Values of the two scales may be set. The detailed setting method is introduced in the current menu. IBP pressure scale name: ART, RA, CVP, RAP, LAP and ICP are available for selection in the hotkey area of the IBP menu; Upper scale: The pressure value represented by the upper scale limit. The choice range is the measurement range of the current pressure.
Attention
The upper scale limit value should not be lower than the lower limit value.
Lower scale: The pressure value represented by the lower scale limit. The choice
Attention
The lower scale limit value should not be higher than the upper limit value.
Attention
The lower pressure limit, the upper pressure limit, the reference scale and the waveform are displayed simultaneously on the screen so that users can observe waveform changes after the scales are adjusted.
102
Technical alarms:
PROMPT MESSAGE CAUSE ALARM LEVEL IBP lead detached IBP cable is detached from the monitor IBP module IBP measurement High Suspend the IBP measurement function and inform biomedical engineers or COMEN maintenance workers. Low Make sure the cable is connected reliably. SOLUTION
initialization wrong
103
XX means all parameter values in such systems as HR, ST1, ST2, RR, SPO2, IBP and NIBP. The patient's ECG signal is too small, "ECG signal is too weak " so the system cant do ECG signal analysis. The patient's pulse signal is too small, "Pulse is not found " so the system cant do pulse signal analysis. The patient's RESP signal is too small, "RESP apnea" so the system cant do RESP signal analysis. The patient suffers from arrhythmia of a systole. Check whether the electrode and lead cable are connected correctly, and the current status of the patient. Check the connection of the sensor, and the current status of the patient.
Check the connection of the lead cable, and the current status of the patient. Check the current status of the patient, and whether the electrode and lead cable are connected correctly. Check the current status of the patient, and whether the electrode and lead cable are connected correctly. Check the current status of the patient, and whether the electrode and lead cable are connected correctly. Check the current status of the patient, and whether the electrode and lead cable are connected correctly. Check the current status of the patient, and whether the electrode and lead cable are connected correctly. Check the current status of the patient, and whether the electrode and lead cable are connected correctly. Check the current status of the patient, and whether the electrode and lead cable are connected correctly. Check the current status of the patient, and whether the electrode and lead cable
"ASYSTOLE "
"VFIB/VTAC "
"R ON T"
"TACHY (tachycardia)"
104
are connected correctly. Check the current status of the patient, "BRADY (bradycardia)" The patient suffers from bradycardia. and whether the electrode and lead cable are connected correctly. The patient suffers from arrhythmia of multiple PVS. Check the current status of the patient, and whether the electrode and lead cable are connected correctly. Check the current status of the patient, and whether the electrode and lead cable are connected correctly. Check the connection of the pacemaker, "PNP (Pacemaker not paced)" The pacemaker doesnt pace. the current status of the patient, and whether the electrode and lead cable are connected correctly. Check the current status of the patient, "PNC (Packer not captured)" No pacemaker signal is captured. and whether the electrode and lead cable are connected correctly. "ECG lead disconnected" "ECG V lead disconnected" ECG lead cable is not joined well. ECG V-lead cable is not joined well. Check the connection of ECG lead cable. Check the connection of ECG V-lead cable. Check the connection of ECG LL-lead cable. Check the connection of ECG LA-lead cable. Check the connection of ECG RA-lead cable. Check the connection of ECG C-lead cable. Check the connection of ECG F-lead cable. Check the connection of ECG L-lead cable. Check the connection of ECG R-lead cable.
"MISSED BEATS"
"SPO2 sensors fall off " "Searching pulse" "TEMP sensors fall off " "IBP lead disconnected" "IBP needs zeroing"
SPO2 sensor is not joined well. SPO2 sensor is not joined well, or the patient moves his arms. TEMP sensor is not joined well. IBP sensor is not joined well. IBP needs zeroing before measurement. Stronger interference signal appear in ECG signal. 105
Check the connection of SPO2 sensor. Check the connection of SPO2 sensor and the current status of the patient. Check the connection of TEMP sensor. Check the connection of IBP sensor. Zero pressure for IBP. Check the connection of ECG R-lead cable, the current status of the patient and whether more actions happen.
"XX module initialization wrong X" "XX stop" "XX module communication module communication
Error X appears when XX module is initialized. XX module cant communicate with the main system normally. XX module cant communicate with the main system normally. Restart the monitor for trial. If errors still exist, please contact the manufacturer for maintenance.
wrong "
XX means all parameter modules in the system such as ECG module, NIBP module, SPO2 module, IBP module and etc. "XX alarm limit wrong " The alarm limit of XX parameter is changed accidently. The measured value of XX parameter "XX measurement out of scope " goes out of the range of measurement by the system. Contact the manufacturer for
maintenance.
XX means the names of all parameters in the system, such as HR, ST1, ST2, RR, SPO2, IBP, NIBP and etc. "The real-time clock needs If the system time is 2000-1-1, the system will remind users that this time is wrong. No button batteries exist in the system or the batteries have no charge. Reset the system time. After resetting, please restart the monitor for fear of unnecessary time error being saved. Increase or change the button battery.
resetting. "The real-time clock doesnt exist " "System failure: software" " System failure: cmos full" " System failure: cmos err" " System failure: fpga" " System failure 2" " System failure 3" " System failure 4" " System failure 5" " System failure 6" " System failure 7" " System failure 8" " System failure 9" " System failure 10" " System failure 11" " System failure 12"
Please check whether the keys are held on to artificially or other articles. If no "Keyboard unusable "; The keys on the keyboard are unusable. abnormal pressing is found, please contact the manufacturer for
maintenance. "Communication with keyboard wrong "; "Key wrong "; " Keyboard wrong 1"; " Keyboard wrong 2"; The keyboard fails, so it is unusable. Please contact the manufacturer for maintenance.
106
"Network initialization wrong (G.)" " Network initialization wrong (Ram)" " Network initialization wrong (Reg)" " Network initialization wrong (Mii)" " Network initialization wrong (Loop)" " Network wrong (Run1)" " Network wrong (Run2)" " Network wrong (Run3)" "5V voltage too high" "5V voltage too low" "Power system wrong 3" " Power system wrong 4" "12V voltage too high " "12V voltage too low " " Power system wrong 7" " Power system wrong 8" "3.3V12V voltage too high " "3.3V voltage too low " " Button battery voltage too high" "Button battery voltage too low. Please change it." Something wrong happens to the button battery. The button battery is short of charge or it is not fixed or the connection of button battery is loose. Please replace the battery. If the fault still exists, please contact the manufacturer for maintenance. Something wrong happens to the power supply of the system. Please contact the manufacturer for maintenance if such a prompt appears frequently. Something wrong happens to the Please contact the manufacturer for maintenance.
In the recorder-setting menu, start the record-deletion function that can make "Recorder self-check wrong " The system fails in connection with the recorder module in self-check. the upper machine and the lower machine reconnected. If the fault still exists, please contact the manufacturer for maintenance. "Recorder voltage too high" " Recorder voltage too low" The recorder module voltage fails. Please contact the manufacturer for maintenance. After the recorder is fully cooled, do the "Recorder head too hot " Maybe the continuous recording time is too long. record output. If the fault still exists, please contact the manufacturer for maintenance. "The position of the recorder head wrong " The paper-pressing handle of the Depress the paper-pressing handle of the recorder.
"Recorder short of paper " "Recorder having paper jam " "Recorder communication wrong "
There is no record paper in the recorder. The recorder has a paper jam.
Install the record paper. Place the recorder correct first and then start recording. In the record setting menu, start the record-deletion function that can make
The recorder communication is not "The serial port of recorder wrong in communication "The paper position of recorder wrong " The paper is not placed correctly in the recorder. normal.
the upper machine and the lower machine reconnected. If the fault still exists, please contact the manufacturer for maintenance. Re-place the corder paper rolls at the correct location. In the record setting menu, start the record-deletion function that can make
the upper machine and the lower machine reconnected. If the fault still exists, please contact the manufacturer for maintenance.
"NIBP initialization wrong " "NIBP self-check wrong " NIBP initialization is wrong.
Select the resetting function in NIBP menu. If the error persists, please contact the manufacturer for maintenance. Please do measurement again after Abnormal resetting happens when NIBP is measured. checking the air path of NIBP and whether there is air jam. If the error persists, please contact the manufacturer for maintenance. Something wrong happens to the NIBP communication. NIBP cuff is not joined well. Select the resetting function in NIBP menu. If the error persists, please contact the manufacturer for maintenance. Please reconnect the NIBP cuff. Please check the connection of all parts
or replace a cuff. If the error persists, please contact the manufacturer for maintenance.
wrong
happens
to
the
Please check the connection of all parts or replace a cuff. If the error persists, please contact the manufacturer for maintenance.
After inspecting whether patient type is set correctly, please check the connection of all parts or replace a cuff. If the error persists, please contact the manufacturer for maintenance.
calculation. "Pressure out of scope " Something wrong happens to 108 the
or replace a cuff. If the error persists, please contact the manufacturer for maintenance. Please do measurement again after checking the connection of all parts and
calculation.
the patient's condition. If the error persists, please contact the manufacturer for maintenance. Please do measurement again after
checking whether the air path is smooth and the patients condition. If the error persists, please contact the manufacturer for maintenance.
wrong
happens to
the
Please do measurement again after checking the connection of all parts and the patient's condition. If the error persists, please contact the manufacturer for maintenance.
calculation.
wrong
happens to
the
Please do measurement again after checking the connection of all parts and the patient's condition. If the error persists, please contact the manufacturer for maintenance. After inspecting whether patient type is
calculation.
Maybe the cuff doesnt accord with the set patient type.
set correctly, Please check the connection of all parts or replace a cuff. If the error persists, please contact the manufacturer for maintenance. Please check the connection of all parts
or replace a cuff. If the error persists, please contact the manufacturer for maintenance.
wrong
happens to
the
Please do measurement again after checking the connection of all parts and the patient's condition. If the error persists, please contact the manufacturer for maintenance.
calculation.
wrong
happens to
the
Please do measurement again after checking the connection of all parts and the patient's condition. If the error persists, please contact the manufacturer for maintenance.
calculation.
109