Open this leaflet for calibration/testing/maintenance

instructions.
Changing Batteries
1. Open battery INC
compartment

++
lid (with
attached
lanyard loop).
2. Remove old
batteries;
replace
with fresh DEC
ones. Note
Diagram 1
polarity as
shown in battery compartment. See Diagram 1.
3. Recalibrate after battery change.
Warranty
Eutech Instruments warrants its meter free from
manufacturing defects for 2 years and electrode
module for 6 months. If repair, adjustment or
replacement is necessary and has not been the result
of abuse or misuse within the time period, please
return the tester – freight prepaid – and correction
will be made without charge. Out of warranty products
will be repaired on a charge basis.
Return Of Items
Authorization must be obtained from Eutech’s
distributor before returning items for any reason.
When applying for authorization, please include
information regarding the reason the item(s) are to • Waterproof & Floats
be returned.
Note: We reserve the right to make improvements in
• Replaceable Electrode
design, construction and appearance of products • Auto-off
without notice. Prices are subject to change without
• Automatic Temperature
notice.
Compensation (ATC)
Waterproof ECScan High Instructions

Before You Begin Conductivity Testing


Remove electrode cap. Soak electrode for a few 1. Remove electrode cap. Switch unit on (ON/OFF
minutes in alcohol to remove oils. key).
Calibration 2. Dip electrode into test solution. Make sure sensor
is fully covered.
Tester is factory calibrated. However, to ensure
accuracy, calibrate conductivity on a regular basis. 3. Wait for reading to stabilize (Automatic
Select a calibration standard appropriate for the Temperature Compensation cor rects for
ECScan High tester (between 2.00 mS and temperature changes). Note reading.
19.90 mS). It is best to select a standard close
to the test solution value. 4. Press ON/OFF to turn off tester. Replace electrode
cap. Note: Tester automatically shuts off after
1. Open battery compartment lid (end with 8.5 minutes of non-use.
lanyard loop).
HOLD Function
2. Orientate the battery compartment as
shown in Diagram 1. Press HOLD key to freeze display. Press HOLD again
to release.
3. Rinse electrode in deionized water, then
rinse it in calibration standard, then dip it Tester Maintenance
into a container of calibration standard. • To improve performance, clean the electrode
4. Switch unit on (ON/OFF key). Wait by rinsing them in alcohol for 10 to 15 minutes.
several minutes for display to stabilize. • Replace all batteries if low battery indicator
5. Press the INC or DEC key to adjust appears, or if readings are faint or unstable.
reading to match the calibration • If you experience drift, periodically let electrode
standard value. fully dry.
6. After 3 seconds without a key press, When you need a new electrode, see “Electrode
the display flashes 3 times, then Replacement” on insert in back of box.
shows ‘ENT’. The tester accepts
calibration value; returns to Self-Diagnostic Messages
measurement mode. bAt Weak batteries–replace with fresh ones specified
7. Replace battery cap. by manufacturer.
Err Wrong keypress.
Or, Ur Over range/Under range signal, or electrode
is not in contact with solution, or electrode is failing.
---- Calibration mode activated but not performed.

68X247755 11/00 Rev 0


Open this leaflet for calibration/testing/maintenance
instructions.
Changing Batteries
1. Open battery INC
compartment

+
lid (with
attached
lanyard loop).
2. Remove old
batteries;
replace with
fresh ones. DEC
Note polarity
Diagram 1
as shown in
battery compartment. See Diagram 1.
3. Recalibrate after battery change.
Warranty
Eutech Instruments warrants its meter free from
manufacturing defects for 2 years and electrode
module for 6 months. If repair, adjustment or
replacement is necessary and has not been the result
of abuse or misuse within the time period, please
return the tester – freight prepaid – and correction
will be made without charge. Out of warranty products
will be repaired on a charge basis.
Return Of Items
Authorization must be obtained from Eutech’s
distributor before returning items for any reason.
When applying for authorization, please include
information regarding the reason the item(s) are to
• Waterproof & Floats
be returned.
Note: We reserve the right to make improvements in • Replaceable Electrode
design, construction and appearance of products • Auto-off
without notice. Prices are subject to change without
notice. • Automatic Temperature
Compensation (ATC)
Waterproof ECScan Low Instructions

Before You Begin Conductivity Testing


Remove electrode cap. Soak electrode for a few 1. Remove electrode cap. Switch unit on (ON/OFF
minutes in alcohol to remove oils. key).
Calibration 2. Dip electrode into test solution. Make sure
Tester is factory calibrated. However, to ensure sensor is fully covered.
accuracy, calibrate conductivity on a regular basis.
3. Wait for reading to stabilize (Automatic
Select a calibration standard appropriate for the Temperature Compensation corrects for
ECScan Low tester (between 200 µS and temperature changes). Note reading.
1990 µS). It is best to select a standard close
4. Press ON/OFF to turn off tester. Replace
to the test solution value.
electrode cap. Note: Tester automatically
1. Open battery compartment lid (end with shuts off after 8.5 minutes of non-use.
lanyard loop). HOLD Function
2. Orientate the battery compartment as Press HOLD key to freeze display. Press HOLD again
shown in Diagram 1. to release.
3. Rinse electrode in deionized water, then Tester Maintenance
rinse it in calibration standard, then dip it • To improve performance, clean the electrode
into a container of calibration standard. by rinsing them in alcohol for 10 to 15 minutes.
4. Switch unit on (ON/OFF key). Wait • Replace all batteries if low battery indicator
several minutes for display to stabilize. appears, or if readings are faint or unstable.
5. Press the INC or DEC key to adjust • If you experience drift, periodically let electrode
reading to match the calibration fully dry.
standard value.
When you need a new electrode, see “Electrode
6. After 3 seconds without a key Replacement” on insert in back of box.
press, the display flashes 3 times,
then shows ‘ENT’. The tester accepts Self-Diagnostic Messages
calibration value; returns to
bAt Weak batteries–replace with fresh ones
measurement mode.
specified by manufacturer.
7. Replace battery cap. Err Wrong keypress.
Or, Ur Over range/Under range signal, or electrode
is not in contact with solution, or electrode is failing.
---- Calibration mode activated but not performed.

68X247754 11/00 Rev 0


Open this leaflet for calibration/testing/maintenance
instructions.
Changing Batteries
1. Open battery
compartment INC
lid (with

++
attached
lanyard loop).
2. Remove old
batteries;
replace with
fresh ones.
Note polarity DEC
as shown
in battery Diagram 1
compartment. See Diagram 1.
3. Recalibrate after battery change.
Tester Maintenance
• To improve performance, clean the electrode
by rinsing them in alcohol for 10 to 15 minutes.
• Replace all batteries if low battery indicator
appears, or if readings are faint or unstable.
• If you experience drift, periodically let electrode
fully dry.
When you need a new electrode, see “Electrode
Replacement” on insert in back of box.
Warranty
Eutech Instruments warrants its meter free from
manufacturing defects for 2 years and electrode
module for 6 months. If repair, adjustment or
replacement is necessary and has not been the result
of abuse or misuse within the time period, please
return the tester – freight prepaid – and correction
will be made without charge. Out of warranty products • Waterproof & Floats
will be repaired on a charge basis. • Replaceable Electrode
Return Of Items • Auto-off
Authorization must be obtained from Eutech’s • Adjustable TDS Factor
distributor before returning items for any reason.
When applying for authorization, please include • Automatic Temperature
information regarding the reason the item(s) are to
be returned.
Compensation (ATC)
Note: We reserve the right to make improvements in
design, construction and appearance of products
without notice. Prices are subject to change without
notice.
WP TDScan High Instructions

Before You Begin TDS Testing


Remove electrode cap. Soak electrode for a few 1. Remove electrode cap. Switch unit on (ON/OFF
minutes in alcohol to remove oils. key).
Calibration 2. Dip electrode into test solution. Make sure sensor
is fully covered.
Tester is factory calibrated. However, to ensure
accuracy, calibrate TDS on a regular basis. 3. Wait for reading to stabilize (Automatic
Select a calibration standard appropriate for the Temperature Compensation corrects for
TDScan High tester (between 2.00 ppt and temperature changes). Note reading.
10.00 ppt). It is best to select a standard close
to the test solution value. 4. Press ON/OFF to turn off tester. Replace electrode
cap. Note: Tester automatically shuts off after
1. Open battery compartment lid (end with 8.5 minutes of non-use.
lanyard loop).
HOLD Function
2. Orientate the battery compartment as
shown in Diagram 1. Press HOLD key to freeze display. Press HOLD
again to release.
3. Rinse electrode in deionized water, then
rinse it in calibration standard, then dip it Setting TDS Factor
into a container of calibration standard. This tester lets you select a TDS factor of 0.4 to 1.0.
4. Switch unit on (ON/OFF key). Wait 1. Open battery compartment. With meter on,
several minutes for display to stabilize. press the HOLD key, then press the INC key
5. Press the INC or DEC keys to adjust See Diagram 1.
reading to match the calibration 2. Press the INC or DEC key to adjust the TDS factor.
standard value.
3. After 3 seconds without a key press, the display
6. After 3 seconds without a key press, flashes 3 times, then shows ‘ENT’. Tester accepts
the display flashes 3 times, then shows TDS factor and returns to measurement mode.
‘ENT’. The tester accepts calibration
value; returns to measurement mode. 4. Replace battery cap.

7. Replace battery cap. Self-Diagnostic Messages


bAt Weak batteries–replace with fresh ones
specified by manufacturer.
Err Wrong keypress.
Or, Ur Over range/Under range signal, or electrode
is not in contact with solution, or electrode is failing.
---- Calibration mode activated but not performed.

68X247753 11/00 Rev 0


Open this leaflet for calibration/testing/maintenance
instructions.
Changing Batteries
1. Open battery
compartment INC
lid(with

+
attached
lanyard loop).
2. Remove old
batteries;
replace with
fresh ones.
Note polarity DEC
as shown
in battery Diagram 1
compartment. See Diagram 1.
3. Recalibrate after battery change.
Tester Maintenance
• To improve performance, clean the electrodes
by rinsing them in alcohol for 10 to 15 minutes.
• Replace all batteries if low battery indicator
appears, or if readings are faint or unstable.
• If you experience drift, periodically let electrode
fully dry.
When you need a new electrode, see “Electrode
Replacement” on insert in back of box.
Warranty
Eutech Instruments warrants its meter free from
manufacturing defects for 2 years and electrode
module for 6 months. If repair, adjustment or
replacement is necessary and has not been the result
of abuse or misuse within the time period, please
return the tester – freight prepaid – and correction
will be made without charge. Out of warranty products • Waterproof & Floats
will be repaired on a charge basis. • Replaceable Electrode
Return Of Items • Auto-off
Authorization must be obtained from Eutech’s • Adjustable TDS Factor
distributor before returning items for any reason.
When applying for authorization, please include • Automatic Temperature
information regarding the reason the item(s) are to Compensation (ATC)
be returned.
Note: We reserve the right to make improvements in
design, construction and appearance of products
without notice. Prices are subject to change without
notice.
WP TDScan Low Instructions

Before You Begin TDS Testing


Remove electrode cap. Soak electrode for a few 1. Remove electrode cap. Switch unit on (ON/OFF
minutes in alcohol to remove oils. key).
Calibration 2. Dip electrode into test solution. Make sure sensor
is fully covered.
Tester is factory calibrated. However, to ensure
accuracy, calibrate TDS on a regular basis. 3. Wait for reading to stabilize (Automatic
Select a calibration standard appropriate for the Temperature Compensation corrects for
TDScan Low tester (between 200 ppm and temperature changes). Note reading.
1990 ppm). It is best to select a standard
close to the test solution value. 4. Press ON/OFF to turn off tester. Replace electrode
1. Open battery compartment lid (end with cap. Note: Tester automatically shuts off after
lanyard loop). 8.5 minutes of non-use.

2. Orientate the battery compartment as HOLD Function


shown in Diagram 1. Press HOLD key to freeze display. Press HOLD again
3. Rinse electrode in deionized water, then to release.
rinse it in calibration standard, then dip it Setting TDS Factor
into a container of calibration standard.
This tester lets you select a TDS factor of 0.4 to 1.0.
4. Switch unit on (ON/OFF key). Wait
several minutes for display to stabilize. 1. Open battery compar tment. With meter on,
press the HOLD key, the press the INC key
5. Press the INC or DEC keys to adjust See Diagram 1.
reading to match the calibration
standard value. 2. Press the INC or DEC key to adjust the TDS factor.

6. After 3 seconds without a key 3. After 3 seconds without a key press, the display
press, the display flashes 3 times, flashes 3 times, then shows ‘ENT’. Tester accepts
then shows ‘ENT’. The tester accepts TDS factor and returns to measurement mode.
calibration value; returns to
4. Replace battery cap.
measurement mode.
7. Replace battery cap.
Self-Diagnostic Messages
bAt Weak batteries–replace with fresh ones specified
by manufacturer.
Err Wrong keypress.
Or, Ur Over range/Under range signal, or electrode
is not in contact with solution, or electrode is failing.
---- Calibration mode activated but not performed.

68X247752 11/00 Rev 0


(Picture A)
Twist-off REMOVE THE OLD ELECTRODE
Battery A
Compartment

Lanyard
connection

(Picture B)
REPLACE THE ELECTRODE
Replaceable B
Sensor

• Push-button calibration
• Fast, stable, repeatable readings
• Available through authorized distributors

SPECIFICATIONS
Tester TDScan TDScan ECScan ECScan Electrode Replacement
Low High Low High
Range 0 to 0 to 0 to 0 to You can replace the electrode module at
1990 10.00 1990 19.90 the fraction of the cost of a new tester.
ppm ppt µS/cm mS/cm When the tester fails to calibrate or gives
Resolution 10 0.10 10 0.10 fluctuating readings in calibration
ppm ppt µS/cm mS/cm standards, you need to change the
Accuracy ±1% Full Scale electrode.
TDS factor 0.4 to 1.0 selectable - 1. With dry hands, grip the ribbed tester
Calibration 200 to 2 .00 to 200 to 2.00 to collar with electrode facing you. Twist the
Standard 1990 10.00 1990 19.90 collar counter clockwise (see picture A).
Range ppm ppt µS/cm mS/cm Save the ribbed tester collar and O-ring
Calibration 1 point (calibration range is ±30% for later use.
of factory default parameter) 2. Pull the old electrode module away from
ATC 0 to 50 °C the tester.
Temp 2% per °C 3. Align the four tabs on the new module
Coefficient so that they match the four slots on the
Operating 0 to 50 °C tester (see picture B).
Temp 4. Gently push the module onto the slots
Power Four X 1.5V to sit it in position. Push the smaller O-
Battery (Type: A76, LR 44 or V136A); ring fully onto the new electrode module.
> 140 hrs continuous use Push the collar over the module and thread
Dimension; Tester: 16.5 cm x 3.8 cm; 90 g it into place by firmly twisting clockwise.
Weight Boxed: 22 cm x6 cm x 5 cm; 170 g
Applications
Manufactured By:
Eutech Instruments Pte Ltd Water quality testing • water and
Blk 55, Ayer Rajah Crescent, Wastewater treatment • hydroponics
#04-14/24, Singapore 139949
Tel: (65)778 6876 Fax: (65)773 0836 • labs • boilers • factories
E-Mail: marketing@eutechinst.com • ecology studies • and more!
Web-Site: http://www.eutechinst.com

Made in Singapore
68X247828 11/00 Rev 0