Introduction
Air must flow through the Sensor Control Electronics Module (CEM) at a rate of at least 50 feet per minute to ensure proper cooling and avoid rack shutdown. For this reason, the CEM contains an airflow sensor that automatically shuts the rack down any time it detects an interruption in the flow of cooling air through the rack. This airflow sensor must be properly calibrated in order to do its job. The airflow sensor calibration tool allows you to calibrate the sensor. This tool consists of a small wind tunnel with a two-speed fan and a circuit to detect and alert you when the airflow sensor trips. A properly calibrated sensor will not trip at the higher air speed and will trip at the lower speed. The sensors calibration potentiometer is very sensitive, and it may take several attempts between the two settings to successfully calibrate it. The procedures outlined in this document explain how to calibrate the airflow sensor. For more information regarding the procedure, or to obtain the necessary equipment, contact ETC at 800/688-4116.
K4 CR13 F5 J13 CR12 U35 T4 U27 U28 CR11 U33 U29 R59 C18 U31 R56 C32 C25 F3 T3 T2 U25 C9 U24 CR21 CR22 R2 CR20 CR17 R3 R7 C11 CR19 CR18 U23 C14 T5 F1 T1 U42 C26 CR26 7050B4001 ELECTRONIC THEATRE CONTROLS R42 R11 R67 R12 R44 R19 R16 R45 R15 R14 R43 R10 R66 R13 P5 C29 C34 C35 C36 C52 U37 R18 R17 R5 R40 C21 CR10 C23 K
574
AC power connector Outputs 1-24 Test switch ETCLink connector (on face panel) Airflow sensor
U39
CR14
RP1
K C10
VR2
C2 C33
C62
U40
U14
U45 TP1
C55
J11
C58
VR4 C37
C59
C49 C50
VR3
U46
U17
F2
U49
C1
U47
C57
C15
R21
RP12
C54
C69
U38
C72
U48
U44
R46 TS1
C41
R23 U13
Q1
C48
C60
CR1
J12
J16
Calibration procedure
Part A
You will need the following equipment: Custom AC power cord Calibration tool Control ribbon cable ETCLink cable Jewelers flat screwdriver (not provided)
1. Plug ribbon cable into the connector on the calibration tool and outputs 1 24 on the CEM. Be certain the side of the connector marked TOP faces up and that the red edge of the ribbon cable lines up with pin 1 on the CEM. 2. Plug one end of the ETCLink cable into the ETCLink connector on the calibration tool and the other end into the CEMs face panel. 3. Place the airflow outlet of the calibration tool against the airflow intake on the right side of the CEM face panel. Line up the right side of the CEM with the right side of the tool. See illustration below.
Top Top
AC power Pin 1
Outputs 1-24
Test Norm
4. Attach the power cord connector to the connector labeled AC in the back left corner of the CEM. Be certain the side of the connector marked TOP faces up. 5. Set the test switch on the CEM to the TEST position. 6. Plug the power cord inot a standard wall outlet. This turns on the CEM.
Part B
1. Set the fan switch (S1) on the calibration tool to the high speed setting (toward the CEM). 2. Press and release the test button on the calibration tool. The test button LED turns off. 3. If the test button LED relights within one minute, go to Part D. Otherwise, go to Part C.
Fan switch
Part C
1. Set the fan switch (S1) on the calibration tool to the low speed setting (furthest from the CEM). 2. If the test button LED lights within 15 seconds, go to Part E. Otherwise, go to step 3. 3. Pry up the metal flange on the CEM airflow sensor unit.
Flange
Caution: The black center of the airflow sensor is very fragile. As you pry up the flange, take extreme care not to contact it.
4. Use the screwdriver to turn the potentiometer inside the airflow sensor slightly clockwise. 5. Wait 15 seconds, then press and release the test button on the calibration tool. 6. If the test button LED lights within 15 seconds, return to Part B. Otherwise, return to step 4.
Part D
Airflow sensor
1. Pry up the metal flange on the CEM airflow sensor.
Caution: The black center of the airflow sensor is very fragile. As you pry up the flange, take extreme care not to contact it.
2. Use the screwdriver to turn the potentiometer inside the airflow sensor slightly counterclockwise. 3. Wait 15 seconds then press and release the test button on the calibration tool. 4. If the test button LED stays off for one minute, return to Part C. Otherwise, return to step 2.
Part E
1. The airflow sensor in the CEM is calibrated to 50 feet per minute. Unplug AC power cord from power source. 2. Disconnect all three cables from CEM. 3. Close the metal flange on the CEM airflow sensor. 4. Set test switch on CEM to the NORM position.