Beruflich Dokumente
Kultur Dokumente
INDEX
1. General Methods for Measuring Pressure 2. Direct-reading Meters U-tube manometers, Bourdon gauges, Diaphragm gauges 3. Indirect-reading Meters Thermocouple gauges, hot and Cold cathode gauges 4. Election of pressure meters 5. Pressure meters for Partial Pressures
2. Direct-reading gauges
2.1. U-tube manometers
U-tube manometer p1 p2 = g(h1 h2) Hg, different oils
U-tube manometer
McLeod gauge (McLeod 1874) more sophisticated hydrostatic gauge: Sample of gas is compressed (typically 1000:1) pressure is easily measured P1V1= P2V2 Primary standard gauge in the 10-3 102 Pa range
C-type
optical electrical: resistive (strain gauge), inductive, capacitive sensor - from 10-3 Pa up
small diaphragm deflection, 10-11 m temperature change large source of error sensor with temperature stabilization
3. Indirect-reading gauges
Calculate pressure by measuring a pressure dependent parameter
Ion gauges are calibrated for nitrogen the reading changes depending on the gas: P(x) = P(measured)/(relative sensitivity of gas x)
Triode gauge 1) Cathode (filament), potential over ground +20->+50 V and current 10 -5 ->10 -2 A. 2) Anode (grid), potential over ground +150V 3) Ion collector, grounded (or biased 30V respect to the filament) 4) Cover 5) Path of emitted electrons
X-ray influence
electrons striking the anode (grid) generate x-rays (bremsstrhlung) x-rays produce photoelectrons in the collector photocurrent to the anode measured current grows too large pressure reading due to the photocurrent, there is a minimum reading value, even though the actual pressure is lower
2) Bayard-Alpert Gauge (BAG) They overcome the x-rays problem by replacing the large area collector with a fine wire 10-8 10-9 Pa Most popular design for measurements of high vacuum pressures 0.05/Pa<Sensitivity<0.15/Pa At high pressures the sensitivity diminishes (space charge effect) 3) Schulz-Phelps Gauge Close spacing between electrodes one ionization per electron Large collector surface complete ions collection Ion generation is reduced respect to BAG because the chance for an ionizing collision is proportional to the path length Smaller sensitivity, about 4x10-3/Pa Used at higher pressures: 10-4 100 Pa
VC = - 30 V VG = + 150 V VF = 0 V
Penning gauge diagram. 1. Anode, V=2-10kV 2. Cathode (grounded) 3. Permanent magnet (0.1-0.2 T) 4. Bleeder resistor
2 - 10 kV
0 kV
1) Electrons leaving the anode 2) Electron going back and forth through the anode 3) Electron colliding a molecule and ionising it
IONIZATION CROSS-SECTION
IONIZATION CROSS-SECTION IS GENERALLY GREATEST FOR MOST GASES IN THE 50-to-150 eV RANGE
2mVa q
1/2
5.2.2. RF Quadrupole
Selected charge-to-mass ratio ions in a varying quadrupole field Four rods, which are connected to U + V cos(t) positive and negative voltage, form a band pass filter that allows ions of a particular mass range to go through Discrimination of (m/q) values: 1) changing the RF frequency, U and V constant 2) changing U and V, frequency constant
-Measurement of ion flow: +Faraday-cup +Faraday-cup+ FET-amplifier (high pressures), current>10-12 A +Faraday-cup and electron multiplier or channeltron electron multiplier combination (low pressures)
Electron multiplier
Channeltron
In the analysis it is important to take care of the different ionization cross-sections of the gases: each gas produces different peaks even though the partial pressure is the same