Model EJA110A, EJA120A and EJA130A Differential Pressure Transmitters

IM 1C21B1-01E

IM 1C21B1-01E
5th Edition

Blank Page

CONTENTS

CONTENTS
1. INTRODUCTION ............................................................................................ 1-1
WARRANTY .................................................................................................. 1-2

2.

HANDLING CAUTIONS ................................................................................ 2-1


2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 2.9 Model and Specifications Check ......................................................... 2-1 Unpacking ........................................................................................... 2-1 Storage ................................................................................................ 2-1 Selecting the Installation Location ...................................................... 2-2 Pressure Connection ........................................................................... 2-2 Waterproofing of Cable Conduit Connections .................................... 2-2 Restrictions on Use of Radio Transceiver .......................................... 2-2 Insulation Resistance and Dielectric Strength Test ............................ 2-3 Installation of Explosion Protected Type ............................................ 2-3 2.9.1 FM Approval ................................................................................. 2-3 2.9.2 CSA Certification .......................................................................... 2-5 2.9.3 SAA Certification .......................................................................... 2-6 2.9.4 CENELEC (KEMA)/IEC (KEMA) Certification .............................. 2-7 2.10 EMC Conformity Standards ................................................................ 2-8

3. 4.

COMPONENT NAMES.................................................................................. 3-1 INSTALLATION ............................................................................................. 4-1


4.1 4.2 4.3 4.4 Precautions ......................................................................................... 4-1 Mounting .............................................................................................. 4-1 Changing the Process Connection ..................................................... 4-2 Swapping the High/Low-pressure Side Connection ........................... 4-3 4.4.1 Rotating Pressure-detector Section 180 ..................................... 4-3 4.4.2 Using the BRAIN TERMINAL BT200 ........................................... 4-3 4.5 Rotating Transmitter Section .............................................................. 4-4

5.

INSTALLING IMPULSE PIPING ................................................................... 5-1


5.1 Impulse Piping Installation Precautions .............................................. 5-1 5.1.1 Connecting Impulse Piping to the Transmitter ............................. 5-1 5.1.2 Routing the Impulse Piping .......................................................... 5-2 5.2 Impulse Piping Connection Examples ................................................ 5-4

6.

WIRING .......................................................................................................... 6-1


6.1 6.2 6.3 Wiring Precautions .............................................................................. 6-1 Selecting the Wiring Materials ............................................................ 6-1 Connections of External Wiring to Terminal Box ................................ 6-1 6.3.1 Power Supply Wiring Connection ................................................ 6-1 6.3.2 External Indicator Connection ...................................................... 6-1 6.3.3 BRAIN TERMINAL BT200 Connection ........................................ 6-1 6.3.4 Check Meter Connection .............................................................. 6-2

FD No. IM 1C21B1-01E 5th Edition: Sep. 2000(YK) All Rights Reserved, Copyright 1997, Yokogawa Electric Corporation

IM 1C21B1-01E

CONTENTS

6.4

Wiring .................................................................................................. 6-2 6.4.1 Loop Configuration ....................................................................... 6-2 (1) General-use Type and Flameproof Type ..................................... 6-2 (2) Intrinsically Safe Type ................................................................. 6-2 6.4.2 Wiring Installation ......................................................................... 6-2 (1) General-use Type and Intrinsically Safe Type ............................. 6-2 (2) Flameproof Type ......................................................................... 6-3 6.5 Grounding ............................................................................................ 6-3 6.6 Power Supply Voltage and Load Resistance ..................................... 6-3

7.

OPERATION .................................................................................................. 7-1


7.1 7.2 7.3 7.4 7.5 Preparation for Starting Operation ...................................................... 7-1 Zero Point Adjustment ........................................................................ 7-2 Starting Operation ............................................................................... 7-3 Shutting Down Operation .................................................................... 7-3 Venting or Draining Transmitter Pressure-detector Section ............... 7-4 7.5.1 Draining Condensate .................................................................... 7-4 7.5.2 Venting Gas .................................................................................. 7-4 7.6 Setting the Range Using the Range-setting Switch ........................... 7-4

8.

BRAIN TERMINAL BT200 OPERATION ..................................................... 8-1


8.1 BT200 Operation Precautions ............................................................. 8-1 8.1.1 Connecting the BT200 ................................................................. 8-1 8.1.2 Conditions of Communication Line .............................................. 8-1 8.2 BT200 Operating Procedures ............................................................. 8-1 8.2.1 Key Layout and Screen Display ................................................... 8-1 8.2.2 Operating Key Functions .............................................................. 8-2 (1) Alphanumeric Keys and Shift Keys ............................................. 8-2 (2) Function Keys ............................................................................. 8-2 8.2.3 Calling Up Menu Addresses Using the Operating Keys .............. 8-3 8.3 Setting Parameters Using the BT200 ................................................. 8-4 8.3.1 Parameter Summary .................................................................... 8-4 8.3.2 Parameter Usage and Selection .................................................. 8-6 8.3.3 Setting Parameters ....................................................................... 8-7 (1) Tag No. Setup ............................................................................. 8-7 (2) Calibration Range Setup ............................................................. 8-7 (3) Damping Time Constant Setup ................................................... 8-8 (4) Output Mode and Integral Indicator Display Mode Setup ........... 8-9 (5) Output Signal Low Cut Mode Setup ............................................ 8-9 (6) Integral Indicator Scale Setup ................................................... 8-10 (7) Unit Setup for Displayed Temperature ...................................... 8-11 (8) Unit Setup for Displayed Static Pressure .................................. 8-12 (9) Operation Mode Setup .............................................................. 8-12 (10) Impulse Line Connection Orientation Setup ............................. 8-12 (11) Output Status Display/Setup when a CPU Failure .................... 8-12 (12) Output Status Setup when a Hardware Error Occurs ............... 8-12 (13) Bi-directional Flow Measurement Setup ................................... 8-13 (14) Range Change while Applying Actual Inputs ............................ 8-13 (15) Zero Point Adjustment ............................................................... 8-14 (16) Test Output Setup ..................................................................... 8-15 (17) User Memo Fields ..................................................................... 8-15

ii

IM 1C21B1-01E

CONTENTS

8.4

Displaying Data Using the BT200 ..................................................... 8-16 8.4.1 Displaying Measured Data ......................................................... 8-16 8.4.2 Display Transmitter Model and Specifications ........................... 8-16 8.5 Self-Diagnostics ................................................................................ 8-16 8.5.1 Checking for Problems ............................................................... 8-16 (1) Identifying Problems with BT200 .............................................. 8-16 (2) Checking with Integral Indicator ................................................ 8-17 8.5.2 Errors and Countermeasures ..................................................... 8-18

9.

MAINTENANCE ............................................................................................. 9-1


9.1 9.2 9.3 9.4 Overview ............................................................................................. 9-1 Calibration Instruments Selection ....................................................... 9-1 Calibration ........................................................................................... 9-1 Disassembly and Reassembly ............................................................ 9-3 9.4.1 Replacing the Integral Indicator ................................................... 9-3 9.4.2 Replacing the CPU Board Assembly ........................................... 9-4 9.4.3 Cleaning and Replacing the Capsule Assembly .......................... 9-4 9.4.4 Replacing the Process Connector Gaskets ................................. 9-5 9.5 Troubleshooting ................................................................................... 9-6 9.5.1 Basic Troubleshooting .................................................................. 9-6 9.5.2 Troubleshooting Flow Charts ....................................................... 9-6

10. GENERAL SPECIFICATIONS .................................................................... 10-1


10.1 10.2 10.3 10.4 Standard Specifications .................................................................... 10-1 Model and Suffix Codes .................................................................... 10-3 Optional Specifications ...................................................................... 10-6 Dimensions ........................................................................................ 10-8

Customer Maintenance Parts List


DPharp EJA Series Transmitter Section ............................ CMPL 1C21A1-02E Model EJA110A, EJA120A and EJA130A Differential Pressure Transmitter ........................................ CMPL 1C21B0-01E

REVISION RECORD

iii

IM 1C21B1-01E

1. INTRODUCTION

1.

INTRODUCTION
Safety Precautions For the protection and safety of the operator and the instrument or the system including the instrument, please be sure to follow the instructions on safety described in this manual when handling this instrument. In case the instrument is handled in contradiction to these instructions, Yokogawa does not guarantee safety. For the intrinsically safe equipment and explosionproof equipment, in case the instrument is not restored to its original condition after any repair or modification undertaken by the customer, intrinsically safe construction or explosionproof construction is damaged and may cause dangerous condition. Please contact Yokogawa for any repair or modification required to the instrument. The following safety symbol marks are used in this Manual:

Thank you for purchasing the DPharp electronic pressure transmitter. The DPharp Pressure Transmitters are precisely calibrated at the factory before shipment. To ensure correct and efficient use of the instrument, please read this manual thoroughly and fully understand how to operate the instrument before operating it. Regarding This Manual This manual should be passed on to the end user. The contents of this manual are subject to change without prior notice. All rights reserved. No part of this manual may be reproduced in any form without Yokogawas written permission. Yokogawa makes no warranty of any kind with regard to this manual, including, but not limited to, implied warranty of merchantability and fitness for a particular purpose. If any question arises or errors are found, or if any information is missing from this manual, please inform the nearest Yokogawa sales office. The specifications covered by this manual are limited to those for the standard type under the specified model number break-down and do not cover custom-made instruments. Please note that changes in the specifications, construction, or component parts of the instrument may not immediately be reflected in this manual at the time of change, provided that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint.

WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.

CAUTION

Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices.

IMPORTANT

NOTE
For FOUNDATION Fieldbus and HART protocol versions, please refer to IM 1C22T2-01E and IM 1C22T1-01E respectively, in addition to this IM.

Indicates that operating the hardware or software in this manner may damage it or lead to system failure.

NOTE
Draws attention to information essential for understanding the operation and features.

1-1

IM 1C21B1-01E

1. INTRODUCTION

WARRANTY
The warranty shall cover the period noted on the quotation presented to the purchaser at the time of purchase. Problems occurred during the warranty period shall basically be repaired free of charge. In case of problems, the customer should contact the Yokogawa representative from which the instrument was purchased, or the nearest Yokogawa office. If a problem arises with this instrument, please inform us of the nature of the problem and the circumstances under which it developed, including the model specification and serial number. Any diagrams, data and other information you can include in your communication will also be helpful. Responsible party for repair cost for the problems shall be determined by Yokogawa based on our investigation. The Purchaser shall bear the responsibility for repair costs, even during the warranty period, if the malfunction is due to: - Improper and/or inadequate maintenance by the purchaser. - Failure or damage due to improper handling, use or storage which is out of design conditions. - Use of the product in question in a location not conforming to the standards specified by Yokogawa, or due to improper maintenance of the installation location. - Failure or damage due to modification or repair by any party except Yokogawa or an approved representative of Yokogawa. - Malfunction or damage from improper relocation of the product in question after delivery. - Reason of force majeure such as fires, earthquakes, storms/floods, thunder/lightening, or other natural disasters, or disturbances, riots, warfare, or radioactive contamination.

WARNING

Instrument installed in the process is under pressure. Never loosen or tighten the process connector bolts as it may cause dangerous spouting of process fluid. During draining condensate or venting gas in transmitter pressure-detector section, take appropriate care to avoid contact with the skin, eyes or body, or inhalation of vapors, if the accumulated process fluid may be toxic or otherwise harmful. Since draining condensate or bleeding off gas gives the pressure measurement disturbance, this should not be done when the loop is in operation. Since the accumulated process fluid may be toxic or otherwise harmful, take appropriate care to avoid contact with the body, or inhalation of vapors even after dismounting the instrument from the process line for maintenance.

CAUTION

This instrument is tested and certified as intrinsically safe type or explosionproof type. Please note that the construction of the instrument, installation, external wiring, maintenance or repair is strictly restricted, and non-observance or negligence of this restriction would result in dangerous condition.

1-2

IM 1C21B1-01E

2. HANDLING CAUTIONS

2.

HANDLING CAUTIONS
2.1 Model and Specifications Check
The model name and specifications are indicated on the name plate attached to the case. If the reverse operating mode was ordered (reverse signal), REVERSE will be inscribed in field *1; if square root display mode was ordered, SQRT is inscribed in field *2; if square root output mode was ordered, SQRT is inscribed in field *3.

This chapter describes important cautions regarding how to handle the transmitter. Read carefully before using the transmitter. The EJA-A Series pressure transmitters are thoroughly tested at the factory before shipment. When the transmitter is delivered, visually check them to make sure that no damage occurred during shipment. Also check that all transmitter mounting hardware shown in Figure 2.1 is included. If the transmitter was ordered without the mounting bracket or without the process connector, the transmitter mounting hardware is not included. After checking the transmitter, repack it in the way it was delivered until installation.
Bolt Process connector Process connector Gasket U-bolt

F0202.EPS

Figure 2.2 Name Plate

2.2 Unpacking
When moving the transmitter to the installation site, keep it in its original packaging. Then, unpack the transmitter there to avoid damage on the way.

Mounting bracket (L type) U-bolt nut Transmitter mounting bolt

2.3 Storage
The following precautions must be observed when storing the instrument, especially for a long period. (a) Select a storage area which meets the following conditions: It is not exposed to rain or water. It suffers minimum vibration and shock. It has an ambient temperature and relative humidity within the following ranges. Ambient temperature: 40 to 85C without integral indicator 30 to 80C with integral indicator Relative humidity: 5% to 100% R.H. (at 40C) Preferred temperature and humidity: approx. 25C and 65% R.H. (b) When storing the transmitter, repack it as nearly as possible to the way it was packed when delivered from the factory.

Mounting bracket (Flat type)


F0201.EPS

Figure 2.1 Transmitter Mounting Hardware

2-1

IM 1C21B1-01E

2. HANDLING CAUTIONS

(c) If storing a transmitter that has been used, thoroughly clean the chambers inside the cover flanges, so that no measured fluid remains in it. Also make sure before storing that the pressure-detector and transmitter section are securely mounted.

The following precautions must be observed in order to safely operate the transmitter under pressure. (a) Make sure that the four process connector bolts are tightened firmly. (b) Make sure that there are no leaks in the impulse piping. (c) Never apply a pressure higher than the specified maximum working pressure.

2.4 Selecting the Installation Location


The transmitter is designed to withstand severe environmental conditions. However, to ensure stable and accurate operation for years, observe the following precautions when selecting an installation location. (a) Ambient Temperature Avoid locations subject to wide temperature variations or a significant temperature gradient. If the location is exposed to radiant heat from plant equipments, provide adequate thermal insulation and/or ventilation. (b) Ambient Atmosphere Avoid installing the transmitter in a corrosive atmosphere. If the transmitter must be installed in a corrosive atmosphere, there must be adequate ventilation as well as measures to prevent intrusion or stagnation of rain water in conduits. (c) Shock and Vibration Select an installation site suffering minimum shock and vibration (although the transmitter is designed to be relatively resistant to shock and vibration). (d) Installation of Explosion-protected Transmitters Explosion-protected transmitters can be installed in hazardous areas according to the types of gases for which they are certified. See Subsection 2.9 Installation of Explosion Protected Type Transmitters.

CAUTION

Maximum working pressure of the model EJA120A differential pressure transmitter is 50 kPa {0.5 kgf/cm2}. Should the pressure exceed 50 kPa {0.5 kgf/ cm2}, it is possible to break the sensor. Proceed with caution when applying pressure.

2.6 Waterproofing of Cable Conduit Connections


Apply a non-hardening sealant to the threads to waterproof the transmitter cable conduit connections. (See Figure 6.4.2a, 6.4.2b and 6.4.2c.)

2.7 Restrictions on Use of Radio Transceiver


IMPORTANT

2.5 Pressure Connection


WARNING
Instrument installed in the process is under pressure. Never loosen the process connector bolts to avoid the dangerous spouting of process fluid. During draining condensate from the pressuredetector section, take appropriate care to avoid contact with the skin, eyes or body, or inhalation of vapors, if the accumulated process fluid may be toxic or otherwise harmful.

Although the transmitter has been designed to resist high frequency electrical noise, if a radio transceiver is used near the transmitter or its external wiring, the transmitter may be affected by high frequency noise pickup. To test for such effects, bring the transceiver in use slowly from a distance of several meters from the transmitter, and observe the measurement loop for noise effects. Thereafter, always use the transceiver outside the area affected by noise.

2-2

IM 1C21B1-01E

2. HANDLING CAUTIONS

2.8 Insulation Resistance and Dielectric Strength Test


Since the transmitter has undergone insulation resistance and dielectric strength tests at the factory before shipment, normally these tests are not required. However, if required, observe the following precautions in the test procedures. (a) Do not perform such tests more frequently than is absolutely necessary. Even test voltages that do not cause visible damage to the insulation may degrade the insulation and reduce safety margins. (b) Never apply a voltage exceeding 500 V DC (100 V DC with an internal lightning protector) for the insulation resistance test, nor a voltage exceeding 500 V AC (100 V AC with an internal lightning protector) for the dielectric strength test. (c) Before conducting these tests, disconnect all signal lines from the transmitter terminals. Perform the tests in the following procedure: Insulation Resistance Test 1) Short-circuit the + and SUPPLY terminals in the terminal box. 2) Turn OFF the insulation tester. Then connect the insulation tester plus (+) lead wire to the shorted SUPPLY terminals and the minus () leadwire to the grounding terminal. 3) Turn ON the insulation tester power and measure the insulation resistance. The voltage should be applied short as possible to verify that the insulation resistance is at least 20 M. 4) After completing the test and being very careful not to touch exposed conductors disconnect the insulation tester and connect a 100 k resistor between the grounding terminal and the shortcircuiting SUPPLY terminals. Leave this resistor connected at least one second to discharge any static potential. Do not touch the terminals while it is discharging. Dielectric Strength Test 1) Short-circuit the + and SUPPLY terminals in the terminal box. 2) Turn OFF the dielectric strength tester. Then connect the tester between the shorted SUPPLY terminals and the grounding terminal. Be sure to connect the grounding lead of the dielectric strength tester to the ground terminal. 3) Set the current limit on the dielectric strength tester to 10 mA, then turn ON the power and gradually increase the test voltage from 0 to the specified voltage. 4) When the specified voltage is reached, hold it for one minute.

5) After completing this test, slowly decrease the voltage to avoid any voltage surges.

2.9 Installation of Explosion Protected Type


NOTE

For FOUNDATION Fieldbus explosion protected type, please refer to IM 1C22T2-01E.

WARNING
To pressure the safety of explosionproof equipment requires great care during mounting, wiring, and piping. Safety requirements also place restrictions on maintenance and repair activities. Please read the following sections very carefully.

2.9.1 FM Approval
a. FM Intrinsically Safe Type Caution for FM intrinsically safe type. (Following contents refer DOC. No. IFM012-A12 P.1 and 2.) Note 1. Model EJA-A Series pressure transmitters with optional code /FS1 are applicable for use in hazardous locations. Intrinsically Safe for Class I, Division 1, Groups A, B, C & D. Class II, Division 1, Groups E, F & G and Class III, Division 1 Hazardous Locations. Nonincendive for Class I, Division 2, Groups A, B, C & D. Class II, Division 2, Groups E, F & G and Class III, Division 1 Hazardous Locations. Outdoor hazardous locations, NEMA 4X. Temperature Class: T4 Ambient temperature: 40 to 60C Note 2. Entity Parameters Intrinsically Safe Apparatus Parameters [Groups A, B, C, D, E, F and G] Vmax = 30 V Ci = 22.5 nF Imax = 165 mA Li = 730 H Pmax = 0.9 W * Associated Apparatus Parameters (FM approved barriers) Voc 30 V Ca > 22.5 nF Isc 165 mA La > 730 H Pmax 0.9W

2-3

IM 1C21B1-01E

2. HANDLING CAUTIONS

Intrinsically Safe Apparatus Parameters [Groups C, D, E, F and G] Vmax = 30 V Ci = 22.5 nF Imax = 225 mA Li = 730 H Pmax = 0.9 W * Associated Apparatus Parameters (FM approved barriers) Voc 30 V Ca > 22.5 nF Isc 225 mA La > 730 H Pmax 0.9 W Entity Installation Requirements Vmax Voc or Vt, Imax Isc or It, Pmax (IS Apparatus) Pmax (Barrier) Ca Ci + Ccable, La Li + Lcable Note 3. Installation Barrier must be installed in an enclosure that meets the requirements of ANSI/ISA S82.01. Control equipment connected to barrier must not use or generate more than 250 V rms or V dc. Installation should be in accordance with ANSI/ISA RP12.6 Installation of Intrinsically Safe Systems for Hazardous (Classified) Locations and the National Electric Code (ANSI/NFPA 70). The configuration of associated apparatus must be FMRC Approved. Dust-tight conduit seal must be used when installed in a Class II, III, Group E, F and G environments. Associated apparatus manufacturers installation drawing must be followed when installing this apparatus. The maximum power delivered from the barrier must not exceed 0.9 W. Note a warning label worded SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY, and INSTALL IN ACCORDANCE WITH DOC. No. IFM012-A12 P.1 and 2. Note 4. Maintenance and Repair The instrument modification or parts replacement by other than authorized representative of Yokogawa Electric Corporation is prohibited and will void Factory Mutual Intrinsically safe and Nonincendive Approval.

[Intrinsically Safe] Hazardous Location Class I, II, III, Division 1, Groups A, B, C, D, E, F, G EJA Series Pressure Transmitters + Supply Safety Barrier + + Nonhazardous Location General Purpose Equipment +

[Nonincendive] Hazardous Location Class I, II, Division 2, Groups A, B, C, D, E, F, G Class III, Division 1. EJA Series Pressure Transmitters + Supply Not Use Safety Barrier
F0203.EPS

Nonhazardous Location

General Purpose Equipment +

b. FM Explosionproof Type Caution for FM explosionproof type. Note 1. Model EJA-A Series differential, gauge, and absolute pressure transmitters with optional code /FF1 are applicable for use in hazardous locations. Explosionproof for Class I, Division 1, Groups B, C and D. Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G. Outdoor hazardous locations, NEMA 4X. Temperature Class: T6 Ambient Temperature: 40 to 60C Supply Voltage: 42 V dc max. Output signal: 4 to 20 mA Note 2. Wiring All wiring shall comply with National Electrical Code ANSI/NEPA70 and Local Electrical Codes. When installed in Division 1, FACTORY SEALED, CONDUIT SEAL NOT REQUIRED. Note 3. Operation Keep the CAUTION nameplate attached to the transmitter. CAUTION: OPEN CIRCUIT BEFORE REMOVING COVER. FACTORY SEALED, CONDUIT SEAL NOT REQUIRED. INSTALL IN ACCORDANCE WITH THE INSTRUCTION MANUAL IM 1C22. Take care not to generate mechanical sparking when accessing to the instrument and peripheral devices in a hazardous location.

2-4

IM 1C21B1-01E

2. HANDLING CAUTIONS

Note 4. Maintenance and Repair The instrument modification or parts replacement by other than authorized representative of Yokogawa Electric Corporation is prohibited and will void Factory Mutual Explosionproof Approval. c. FM Intrinsically Safe Type/FM Explosionproof Type Model EJA-A Series pressure transmitters with optional code /FU1 can be selected the type of protection (FM Intrinsically Safe or FM Explosionproof) for use in hazardous locations. Note 1. For the installation of this transmitter, once a particular type of protection is selected, any other type of protection cannot be used. The installation must be in accordance with the description about the type of protection in this instruction manual. Note 2. In order to avoid confusion, unnecessary marking is crossed out on the label other than the selected type of protection when the transmitter is installed.

* Associated apparatus (CSA certified barriers) Maximum output voltage (Voc) 30 V Maximum output current (Isc) 165 mA Maximum output power (Pmax) 0.9 W Note 3. Installation All wiring shall comply with Canadian Electrical Code Part I and Local Electrical Codes. The instrument modification or parts replacement by other than authorized representative of Yokogawa Electric Corporation and Yokogawa Corporation of America is prohibited and will void Canadian Standards Intrinsically safe and nonincendive Certification.
[Intrinsically Safe] Hazardous Location Class I, II, III, Division 1, Groups A, B, C, D, E, F, G EJA Series Pressure Transmitters + Supply Safety Barrier + + Nonhazardous Location General Purpose Equipment +

[Nonincendive]

2.9.2 CSA Certification


a. CSA Intrinsically Safe Type Caution for CSA Intrinsically safe type. (Following contents refer to DOC No. ICS003-A12 P.1-1 and P.1-2.) Note 1. Model EJA-A Series differential, gauge, and absolute pressure transmitters with optional code /CS1 are applicable for use in hazardous locations Intrinsically Safe for Class I, Division 1, Groups A, B, C & D. Class II, Division 1, Groups E, F & G and Class III, Division 1 Hazardous Locations. Nonincendive for Class I, Division 2, Groups A, B, C & D, Class II, Division 2, Groups F & G, and Class III, Hazardous Locations. (not use Safety Barrier) Encl. Type 4X Temperature Class: T4 Ambient temperature: 40 to 60C Process Temperature: 120C max. Note 2. Entity Parameters Intrinsically safe ratings are as follows: Maximum Input Voltage (Vmax) = 30 V Maximum Input Current (Imax) = 165 mA Maximum Input Power (Pmax) = 0.9 W Maximum Internal Capacitance (Ci) = 22.5 nF Maximum Internal Inductance (Li) = 730 H

Hazardous Location Class I, II, Division 2, Groups A, B, C, D, E, F, G Class III, Division 1. EJA Series Pressure Transmitters + Supply

Nonhazardous Location

General Purpose Equipment + Not Use Safety Barrier


F0204.EPS

b. CSA Explosionproof Type Caution for CSA explosionproof type. Note 1. Model EJA-A Series differential, gauge, and absolute pressure transmitters with optional code /CF1 are applicable for use in hazardous locations: Explosionproof for Class I, Division 1, Groups B, C and D. Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G. Encl Type 4X Temperature Class: T6, T5, and T4 Process Temperature: 85C (T6), 100C (T5), and 120C (T4) Ambient Temperature: 40 to 80C Supply Voltage: 42 V dc max. Output Signal: 4 to 20 mA

2-5

IM 1C21B1-01E

2. HANDLING CAUTIONS

Note 2. Wiring All wiring shall comply with Canadian Electrical Code Part I and Local Electrical Codes. In hazardous location, wiring shall be in conduit as shown in the figure. CAUTION: SEAL ALL CONDUITS WITHIN 50 cm OF THE ENCLOSURE. UN SCELLEMENT DOIT TRE INSTALL MOINS DE 50 cm DU BTIER. When installed in Division 2, SEALS NOT REQUIRED. Note 3. Operation Keep the CAUTION label attached to the transmitter. CAUTION: OPEN CIRCUIT BEFORE REMOVING COVER. OUVRIR LE CIRCUIT AVANT DNLEVER LE COUVERCLE. Take care not to generate mechanical sparking when accessing to the instrument and peripheral devices in a hazardous location. Note 4. Maintenance and Repair The instrument modification or parts replacement by other than authorized representative of Yokogawa Electric Corporation and Yokogawa Corporation of America is prohibited and will void Canadian Standards Explosionproof Certification.
Non-Hazardous Hazardous Locations Division 1 Locations Non-hazardous Location Equipment

c. CSA Intrinsically Safe Type/CSA Explosionproof Type Model EJA-A Series pressure transmitters with optional code /CU1 can be selected the type of protection (CSA Intrinsically Safe or CSA Explosionproof) for use in hazardous locations. Note 1. For the installation of this transmitter, once a particular type of protection is selected, any other type of protection cannot be used. The installation must be in accordance with the description about the type of protection in this instruction manual. Note 2. In order to avoid confusion, unnecessary marking is crossed out on the label other than the selected type of protection when the transmitter is installed.

2.9.3 SAA Certification


a. SAA Intrinsically Safe Type Caution for SAA Intrinsically safe type and Type n. Note 1. Model EJA-A Series differential, gauge, and absolute pressure transmitters with optional code /SU1 are applicable for use in hazardous locations. Type of Protection and Marking Code: Ex ia IIC T4 (Tamb = 60C) IP67 Class I Zone 0 Type of Protection and Marking Code: Ex n IIC T4 (Tamb = 60C) IP67 Class I Zone 2 Ambient Temperature: 40 to 60C Note 2. Entity Parameters Intrinsically safe rating of the transmitters are as follows. Maximum Input Voltage (Ui) = 30 V Maximum Input Current (Ii) = 165 mA Maximum Input Power (Pi) = 0.9 W Maximum Internal Capacitance (Ci) = 0.02 F Maximum Internal Inductance (Li) = 0.73 mH Note 3. Wiring All Wiring shall comply with the Australian Standard. Note 4. Maintenance and Repair The instrument modification or parts replacement by other than authorized representative of Yokogawa Electric Corporation is prohibited and will void Standards Association of Australia Intrinsically safe and Type n Certification.

50 cm Max.

42 V DC Max. 4 to 20 mA DC Signal

Sealing Fitting

Conduit EJA Series

Non-Hazardous Hazardous Locations Division 2 Locations Non-hazardous Location Equipment

42 V DC Max. 4 to 20 mA DC Signal

Sealing Fitting EJA Series


F0205.EPS

2-6

IM 1C21B1-01E

2. HANDLING CAUTIONS
[Intrinsic Safety] Hazardous Location (Zone 0) EJA Series Pressure Transmitter + Nonhazardous Location

+ Safety Barrier *1

Note 4. Maintenance and Repair The instrument modification or parts replacement by other than authorized representative of Yokogawa Electric Corporation is prohibited and will void Standards Association of Australia Flameproof Certification.

F0206.EPS

2.9.4 CENELEC (KEMA)/IEC (KEMA) Certification


a. CENELEC (KEMA) Intrinsically Safe Type Caution for CENELEC (KEMA) intrinsically safe type. Note 1. Model EJA-A Series differential, gauge, and absolute pressure transmitters with optional code /KS1 for potentially explosive atmospheres: Type of Protection and Marking code: EEx ia IIC T4 Temperature Class: T4 Process Temperature: 120C max. Ambient Temperature: 40 to 60C Note 2. Electrical Data In type of explosion protection intrinsic safety EEx ia IIC only for connection to a certified intrinsically safe circuit with following maximum values: Ui = 30 V Ii = 165 mA Pi = 0.9 W Effective internal capacitance; Ci = 22.5 nF Effective internal inductance; Li = 730 H Note 3. Installation All wiring shall comply with local installation requirements. (Refer to the installation diagram) Note 4. Maintenance and Repair The instrument modification or parts replacement by other than authorized representative of Yokogawa Electric Corporation is prohibited and will void KEMA Intrinsically safe Certification.
[Installation Diagram] Hazardous Location Transmitter + Supply + Safety Barrier *1 Nonhazardous Location

*1: Any safety barriers used for the output current must be limited by a resistor R such that Imaxout-Uz/R.

[Type n] Hazardous Location (Zone 2) EJA Series Pressure Transmitter + + Power Supply *2

F0207.EPS

*2: The voltage of the power supply is not exceed 30V dc.

b. SAA Flameproof Type Caution for SAA flameproof type. Note 1. Model EJA-A Series differential, gauge, and absolute pressure transmitters with optional code /SU1 are applicable for use in hazardous locations: Type of Protection and marking Code: Ex d II C T* IP67 Class I Zone 1 (T* see schedule) Temperature Class: T6, T5, and T4 Process Temperature: 85C (T6), 100C (T5), and 120C (T4) Supply voltage: 42 V dc max. Output Signal: 4 to 20 mA Ambient Temperature: 40 to 80C Note 2. Wiring All wiring shall comply with the Australian Standard. Note 3. Operation Keep the CAUTION label attached to the transmitter. CAUTION: AMBIENT TEMPERATURE ABOVE 75 DEG C SELECT SUITABLE CABLE. DISCONNECT POWER AND WAIT 1 MINUTE BEFORE REMAKING COVER Take care not to generate mechanical sparking when accessing to the instrument and peripheral devices in a hazardous location.

F0208.EPS

*1: In any safety barriers used the output current must be limited by a resistor R such that Imaxout-Uz/R.

2-7

IM 1C21B1-01E

2. HANDLING CAUTIONS

b. CENELEC (KEMA) Flameproof Type Caution for CENELEC (KEMA) flameproof type. Note 1. Model EJA-A Series differential, gauge, and absolute pressure transmitters with optional code /KF1 for potentially explosive atmospheres: Type of Protection and Marking Code: EEx d IIC T6T4 Temperature Class: T6, T5, and T4 Maximum Process Temperature: 85C (T6), 100C (T5), and 120C Ambient Temperature: T6; 40 to 75C, T4 and T5; 40 to 80C Note 2. Electrical Data Supply voltage: 42 V dc max. Output signal: 4 to 20 mA Note 3. Installation All wiring shall comply with local installation requirement. The cable entry devices shall be of a certified flameproof type, suitable for the conditions of use. Note 4. Operation Keep the CAUTION label to the transmitter. CAUTION: WAIT 10 MIN. AFTER POWER-DISCONNECTION, BEFORE OPENING THE ENCLOSURE. WHEN THE AMBIENT TEMP.70C, USE HEAT-RESISTING CABLES90C. Take care not to generate mechanical sparking when accessing to the instrument and peripheral devices in a hazardous location. Note 5. Maintenance and Repair The instrument modification or parts replacement by other than authorized representative of Yokogawa Electric Corporation is prohibited and will void KEMA Flameproof Certification. c. IEC (KEMA) Type of Protection n Caution for IEC (KEMA) Type of Protection n. Note 1. Model EJA-A Series pressure transmitters with optional code /KU1 for potentially explosive atmospheres. Type of Protection and Marking Code: Ex nA IIC T4 Temperature Class: T4 Process Temperature: 120C max. Ambient Temperature: 40 to 60C Note 2. Electrical Data Supply and output circuit 30 V dc, 165 mA (terminals + and ) Note 3. Installation All wiring shall comply with local installation requirements. (refer to the installation diagram)

Note 4. Maintenance and Repair The instrument modification or parts replacement by other than authorized representative of Yokogawa Electric Corporation is prohibited and will void KEMA Type of Protection n Certification.
[Installation Diagram] Hazardous Location (Zone 2 only) Transmitter + Supply Nonhazardous Location

+ Power Supply

F0209.EPS

Ratings of the Power Supply as follows; Maximum Voltage: 30 V Maximum Current: 165 mA

d. CENELEC (KEMA) Intrinsically Safe Type/ CENELEC (KEMA) Flameproof Type/IEC (KEMA) Type of Protection n Model EJA-A Series pressure transmitters with optional code /KU1 can be selected the type of protection (CENELEC (KEMA) Intrinsically Safe or CENELEC (KEMA) Flameproof or IEC (KEMA) Type of Protection n) for use in hazardous locations. Note 1. For the installation of this transmitter, once a particular type of protection is selected, any other type of protection cannot be used. The installation must be in accordance with the description about the type of protection in this instruction manual. Note 2. In order to avoid confusion, unnecessary marking is crossed out on the label other than the selected type of protection when the transmitter is installed.

2.10 EMC Conformity Standards


For EMI (Emission): EN55011, AS/NZS 2064 1/2 For EMS (Immunity): EN500822

NOTE

YOKOGAWA recommends customer to apply the Metal Conduit Wiring or to use the twisted pair Shield Cable for signal wiring to conform the requirement of EMC Regulation, when customer installs the EJA Series Transmitters to the plant. 2-8
IM 1C21B1-01E

3. COMPONENT NAMES

3.

COMPONENT NAMES
Vertical impulse piping type Process connection Pressure-detector section Process connector (Note 1) Cover flange Horizontal impulse piping type External indicator conduit connection (Note 1) Terminal box cover Conduit connection

CPU assembly Integral indicator (Note 1) Mounting screw Zeroadjustment screw Transmitter section Range-setting switch (Note 1) (See Subsection 7.6) Amplifier Cover Setting Pin (CN4) Position (Note 2) H HIGH L H LOW L Burn-Out Direction Output at Burn-Out 110% or higher -5% or lower
F0301.EPS

Setting pin (CN4)

Vent plug Drain plug

Note 1: See Subsection 10.2, Model and Suffix Codes, for details. Note 2: Insert the pin (CN4) as shown in the figure above to set the burn-out direction. The pin is set to the H side for delivery (unless option code /C1 is specified in the order). The setting can be confirmed by calling up parameter D52 using the BRAIN TERMINAL. Refer to Subsection 8.3.3 (11). Figure 3.1 Component Names

Table 3.1 Display Symbol


Display Symbol Meaning of Display Symbol The output signal being zero-adjusted is increasing. The output signal being zero-adjusted is decreasing. %, Pa, kPa, MPa, kgf/cm2, gf/cm2, mbar, bar, atm, mmHg, mmH2O, inH2O, inHg, ftH2O, psi, Torr Select one of these sixteen available engineering units for the display.
T0301.EPS

3-1

IM 1C21B1-01E

4 . INSTALLATION

4.

INSTALLATION
Vertical pipe mounting

4.1 Precautions
Before installing the transmitter, read the cautionary notes in Section 2.4, Selecting the Installation Location. For additional information on the ambient conditions allowed at the installation location, refer to Subsection 10.1 Standard Specifications.

Transmitter mounting bolt

When welding piping during construction, take care not to allow welding currents to flow through the transmitter. Do not step on this instrument after installation.

U-bolt nut Mounting bracket

U-bolt

50 mm(2-inch) pipe Horizontal pipe mounting

4.2 Mounting
The distance between the impulse piping connection ports is usually 54 mm (Figure 4.2.1). By changing the orientation of the process connector, the dimension can be changed 51 mm or 57 mm. The transmitter is shipped with the process connection, according to the ordering specifications. To change the orientation of the process connections, refer to Section 4.3. The transmitter can be mounted on a nominal 50 mm (2-inch) pipe using the mounting bracket supplied, as shown in Figure 4.2.2 and 4.2.3. The transmitter can be mounted on either a horizontal or a vertical pipe. When mounting the bracket on the transmitter, tighten the (four) bolts that hold the transmitter with a torque of approximately 39 Nm {4kgfm}.
Transmitter mounting bolt

U-bolt nut Mounting bracket

U-bolt
F0402.EPS

50 mm(2-inch) pipe

Figure 4.2.2 Transmitter Mounting (Horizontal Impulse Piping Type)

57 mm

54 mm

51 mm
F0401.EPS

Figure 4.2.1 Process Connector Impulse Piping Connection Distances

4-1

IM 1C21B1-01E

4 . INSTALLATION
Vertical pipe mounting (Process connector upside)

4.3 Changing the Process Connection


The transmitter is shipped with the process connection specified at the time of ordering. To make a change such as modifying the drain (vent) plug(s) attached to the upside of the cover flange on shipment to be attached to the downside follow the procedure below. To begin, use a wrench to slowly and gently unscrew the drain (vent) plug(s). Then, remove and remount them on the opposite side. Wrap sealing tape around the drain (vent) plug threads (*1 in the figure below), and apply a lubricant to the threads of the drain (vent) screw(s) (*2 below) to screw it (them) in. To tighten the drain (vent) plugs, apply a torque of 34 to 39 Nm {3.5 to 4 kgfm}. Tighten the process connector bolts uniformly to a torque of 39 to 49 Nm {4 to 5 kgfm}.
Vertical impulse piping type Bolt Process connector gasket Horizontal impulse piping type

U-bolt nut

Mounting bracket

Transmitter mounting bolt Vertical pipe mounting (Process connector downside)

U-bolt

50 mm(2-inch) pipe

Transmitter mounting bolt Mounting bracket

U-bolt nut U-bolt

50 mm(2-inch) pipe
F0403.EPS

Drain/vent plug

Figure 4.2.3 Transmitter Mounting (Vertical Impulse Piping Type) Figure 4.3

Note: For a horizontal impulse piping type, moving the process connectors from the front side to the back is not allowed.
F0404.EPS

Changing Process Connection

4-2

IM 1C21B1-01E

4 . INSTALLATION

4.4 Swapping the High/Low-pressure Side Connection


4.4.1 Rotating Pressure-detector Section 180
This procedure can be applied only to a transmitter with a vertical impulse piping type. The procedure below can be used to turn the pressuredetector assembly 180. Perform this operation in a maintenance shop with the necessary tools laid out and ready for use, and then install the transmitter in the field after making the change. 1) Use a Allen wrench (JIS B4648, nominal 5 mm) to remove the two Allen screws at the joint between the pressure-detector section and transmitter section. 2) Leaving the transmitter section in position, rotate the pressure-detector section 180. 3) Tighten the two Allen screws to fix the pressuredetector section and transmitter section together (at a torque of 5 Nm).

4.4.2 Using the BRAIN TERMINAL BT200


This method is applicable only to the Model EJA110AL, EJA110A-M, EJA110A-H, EJA120A-E, EJA130A-M and EJA130A-H. With a BRAIN TERMINAL, you can change which process connection is used as the high-pressure side without mechanically rotating the pressure-detector section 180 as described in Subsection 4.4.1. To change, call parameter D45: H/L SWAP and select REVERSE (right side: low pressure; left side: high pressure) or select NORMAL to change back to normal (right side: high pressure; left side: low pressure).
NORMAL Output

Input

Do not rotate the pressure-detector section more than 180.


Figure 4.5 Input/Output Relationship

REVERSE
F0406.EPS

Pressure-detector Section Allen screw

Before

After rotating 180


F0405.EPS

Figure 4.4

Before and After Modification

Since the H/L label plate on the capsule assembly will remain unchanged even if this function is used, use this function only if you cannot switch the impulse piping. If the D45: H/L SWAP setting is changed, the input/output relationship is reversed as shown in Figure 4.5; be sure this is understood by all. Whenever possible, use the procedure in Subsection 4.4.1.

4-3

IM 1C21B1-01E

4 . INSTALLATION

4.5 Rotating Transmitter Section


The DPharp transmitter section can be rotated in 90 segments. 1) Remove the two Allen screws that fasten the transmitter section and capsule assembly, using the Allen wrench. 2) Rotate the transmitter section slowly in 90 segments. 3) Tighten the two Allen screws to a torque of 5 Nm.

Do not rotate the transmitter section more than 180.

Vertical impulse piping type Pressure-detector section Rotate 90 or 180 segments

Conduit connection

Transmitter section

Horizontal impulse piping type Transmitter section Rotate 90 or 180 segments

Conduit connection Zero-adjustment screw

Pressure-detector section
F0407.EPS

Figure 4.6

Rotating Transmitter Section

4-4

IM 1C21B1-01E

5. INSTALLING IMPULSE PIPING

5.

INSTALLING IMPULSE PIPING


(4) Removing the Impulse Piping Connecting Port Dustproof Cap The impulse piping connecting port of the transmitter is covered with a plastic cap to exclude dust. This cap must be removed before connecting the piping. (Be careful not to damage the threads when removing this cap. Never insert a screwdriver or other tool between the cap and port threads to remove the cap.) (5) Connecting the Transmitter and 3-Valve Manifold A 3-valve manifold consists of two stop valves to block process pressure and an equalizing valve to equalize the pressures on the high and low pressure sides of the transmitter. Such a manifold makes it easier to disconnect the transmitter from the impulse piping, and is convenient when adjusting the transmitter zero point. There are two types of 3-valve manifold: the pipemounting type and the direct-mounting type; care should be taken with respect to the following points when connecting the manifold to the transmitter. Pipe-Mounting Type 3-Valve Manifold (Figure 5.1.1b) 1) Screw nipples into the connection ports on the transmitter side of the 3-valve manifold, and into the impulse piping connecting ports on the process connectors. (To maintain proper sealing, wind sealing tape around the nipple threads.) 2) Mount the 3-valve manifold on the 50 mm (2-inch) pipe by fastening a U-bolt to its mounting bracket. Tighten the U-bolt nuts only lightly at this time. 3) Install the pipe assemblies between the 3-valve manifold and the process connectors and lightly tighten the ball head lock nuts. (The ball-shaped ends of the pipes must be handled carefully, since they will not seal properly if the ball surface is scratched or otherwise damaged.) 4) Now tighten the nuts and bolts securely in the following sequence: Process connector bolts transmitter-end ball head lock nuts 3-valve manifold ball head lock nuts 3-valve manifold mounting bracket U-bolt nuts

5.1 Impulse Piping Installation Precautions


The impulse piping that connects the process outputs to the transmitter must convey the process pressure accurately. If, for example, gas collects in a liquidfilled impulse piping, or the drain of a gas-filled impulse piping becomes plugged, the impulse piping will not convey the pressure accurately. Since this will cause errors in the measurement output, select the proper piping method for the process fluid (gas, liquid, or steam). Pay careful attention to the following points when routing the impulse piping and connecting the impulse piping to the transmitter.

5.1.1 Connecting Impulse Piping to the Transmitter


(1) Check the High and Low Pressure Connections on the Transmitter (Figure 5.1.1a) Symbols H and L are shown on a capsule assembly to indicate high and low pressure side. Connect the impulse piping to the H side, and the low impulse piping to the L side.
H and L are shown Pressure connection Process connection Process connector Bolt
Figure 5.1.1a H and L Symbols on a Capsule Assembly
F0501.EPS

(2) Changing the Process Connector Piping Connections (Figure 4.1) The impulse piping connection distances can be changed between 51 mm, 54 mm and 57 mm by changing the orientation of the process connectors. This is convenient for aligning the impulse piping with the process connectors when connecting the piping. (3) Tightening the Process Connector Mounting Bolts After connecting the impulse piping, tighten the process connector mounting bolts uniformly.

5-1

IM 1C21B1-01E

5. INSTALLING IMPULSE PIPING


Impulse piping Nipple Vent plug (optional) Stop valve (low pressure side) Equalizing valve (balancing) Stop valve (high pressure side) Pipes Ball head lock nut Nipple 50 mm(2-inch) pipe Process connector bolts Process connector
F0502.EPS

NOTE

3-valve manifold

Ball head lock nut Pipe

After completing the connection of the transmitter and 3-valve manifold, be sure to CLOSE the low pressure and high pressure stop valves, OPEN the equalizing valve, and leave the manifold with the equalizing valve OPEN. You must do this in order to avoid overloading the transmitter from either the high or the low pressure side when beginning operation. This instruction must also be strictly followed as part of the startup procedure (Chapter 7).

Figure 5.1.1b 3-Valve Manifold (Pipe-Mounting Type)

Direct-Mounting Type 3-Valve Manifold (Figure 5.1.1c) 1) Mount the 3-valve manifold on the transmitter. (When mounting, use the two gaskets and the four bolts provided with the 3-valve manifold. Tighten the bolts evenly.) 2) Mount the process connectors and gaskets on the top of the 3-valve manifold (the side on which the impulse piping will be connected).
Bolts Process connector Gasket Impulse piping Stop valve Equalizing valve 3-valve manifold

5.1.2 Routing the Impulse Piping


(1) Process Pressure Tap Angles If condensate, gas, sediment or other extraneous material in the process piping gets into the impulse piping, pressure measurement errors may result. To prevent such problems, the process pressure taps must be angled as shown in Figure 5.1.2 according to the kind of fluid being measured.

NOTE

Stop valve

Equalizing valve Stop valve Impulse piping Gasket Process connector Bolts Stop valve 3-valve manifold
F0503.EPS

If the process fluid is a gas, the taps must be vertical or within 45 either side of vertical. If the process fluid is a liquid, the taps must be horizontal or below horizontal, but not more than 45 below horizontal. If the process fluid is steam or other condensing vapor, the taps must be horizontal or above horizontal, but not more than 45 above horizontal.
[Gas] 45 45 Pressure taps Process piping 45 45 45
F0504.EPS

[Liquid]

[Steam]

45

Figure 5.1.1c 3-Valve Manifold (Direct-Mounting Type)

Figure 5.1.2 Process Pressure Tap Angle (For Horizontal Piping)

5-2

IM 1C21B1-01E

5. INSTALLING IMPULSE PIPING

(2) Position of Process Pressure Taps and Transmitter If condensate (or gas) accumulates in the impulse piping, it should be removed periodically by opening the drain (or vent) plugs. However, this will generate a transient disturbance in the pressure measurement, and therefore it is necessary to position the taps and route the impulse piping so that any extraneous liquid or gas generated in the leadlines returns naturally to the process piping. If the process fluid is a gas, then as a rule the transmitter must be located higher than the process pressure taps. If the process fluid is a liquid or steam, then as a rule the transmitter must be located lower than the process pressure taps. (3) Impulse Piping Slope The impulse piping must be routed with only an upward or downward slope. Even for horizontal routing, the impulse piping should have a slope of at least 1/10 to prevent condensate (or gases) from accumulating in the pipes. (4) Temperature Difference Between Impulse Piping If there is a temperature difference between the high and low impulse piping, the density difference of the fluids in the two lines will cause an error in the measurement pressure. When measuring flow, impulse piping must be routed together so that there is no temperature difference between them. (5) Condensate Pots for Steam Flow Measurement If the liquid in the impulse piping repeatedly condenses or vaporizes as a result of changes in the ambient or process temperature, this will cause a difference in the fluid head between the high pressure and low pressure sides. To prevent measurement errors due to these head differences, condensate pots are used when measuring steam flow.

(6) Preventing Wind Speed Effects in Very Low Differential Pressure Measurement

IMPORTANT

When using a differential pressure transmitter to measure very low pressures (draft pressure), the low pressure connection port is left open to atmospheric pressure (the reference pressure). Any wind around the differential pressure transmitter will therefore cause errors in the measurement. To prevent this, it will be necessary either to enclose the transmitter in a box, or to connect a impulse piping to the low pressure side and insert its end into a windexcluding pot (cylindrical with a base plate).

(7) Preventing Freezing If there is any risk that the process fluid in the impulse piping or transmitter could freeze, use a steam jacket or heater to maintain the temperature of the fluid.

NOTE

After completing the connections, close the valves on the process pressure taps (main valves), the valves at the transmitter (stop valves), and the impulse piping drain valves, so that condensate, sediment, dust and other extraneous material cannot enter the impulse piping.

5-3

IM 1C21B1-01E

5. INSTALLING IMPULSE PIPING

5.2 Impulse Piping Connection Examples


Figure 5.2 shows examples of typical impulse piping connections. Before connecting the transmitter to the process, study the transmitter installation location, the process piping layout, and the characteristics of the process fluid (corrosiveness, toxicity, flammability, etc.), in order to make appropriate changes and additions to the connection configurations. Note the following points when referring to these piping examples. The high pressure connecting port on the transmitter is shown on the right (as viewed from the front). The transmitter impulse piping connection is shown for a vertical impulse piping connection configuration in which the direction of connection is either upwards or downwards. If the impulse piping is long, bracing or supports should be provided to prevent vibration. The impulse piping material used must be compatible with the process pressure, temperature, and other conditions. A variety of process pressure tap valves (main valves) are available according to the type of connection (flanged, screwed, welded), construction (globe, gate, or ball valve), temperature and pressure. Select the type of valve most appropriate for the application.
Tap valve

Liguid Orifice

Gas

Steam Condensate pot

Union or flange

Tee 3-valve manifold Drain valve Drain plug

Liquid Level - Open Tank

Liquid Level - Closed Tank Vent plug

Open tank

Closed tank

Tap valve Union or flange Pipe Tee 3-valve manifold

Drain valve Drain plug


F0505.EPS

Figure 5.2 Impulse Piping Connection Examples

5-4

IM 1C21B1-01E

6. WIRING

6.

WIRING
6.3 Connections of External Wiring to Terminal Box
6.3.1 Power Supply Wiring Connection
Connect the power supply wiring to the SUPPLY + and terminals.
Transmitter terminal box
+

6.1 Wiring Precautions


IMPORTANT

Lay wiring as far as possible from electrical noise sources such as large capacity transformers, motors, and power supplies. Remove electrical connection dust cap before wiring. All threaded parts must be treated with waterproofing sealant. (A non-hardening silicone group sealant is recommended.) To prevent noise pickup, do not pass signal and power cables through the same ducts. Explosion-protected instruments must be wired in accordance with specific requirements (and, in certain countries, legal regulations) in order to preserve the effectiveness of their explosionprotected features. The terminal box cover is locked by an Allen head bolt (a shrouding bolt) on CENELEC and SAA flameproof type transmitters. When the shrouding bolt is driven clockwise by an Allen wrench, it is going in and cover lock is released, and then the cover can be opened by hand. See Subsection 9.4 Disassembly and Reassembly for details.

Power supply

F0601.EPS

Figure 6.3.1 Power Supply Wiring Connection

6.3.2 External Indicator Connection


Connect wiring for external indicators to the CHECK + and terminals.
(Note) Use a external indicator whose internal resistance is 10 or less.

External indicator
Power supply
+

Transmitter terminal box


Figure 6.3.2 External Indicator Connection

F0602.EPS

6.2 Selecting the Wiring Materials


(a) Use stranded leadwires or cables which are the same as or better than 600 V grade PVC insulated wire (JIS C3307) or equivalent. (b) Use shielded wires in areas that are susceptible to electrical noise. (c) In areas with higher or lower ambient temperatures, use appropriate wires or cables. (d) In environment where oils, solvents, corrosive gases or liquids may be present, use wires or cables that are resistant to such substances. (e) It is recommended that crimp-on solderless terminal lugs (for 4 mm screws) with insulating sleeves be used for leadwire ends.

6.3.3 BRAIN TERMINAL BT200 Connection


Connect the BT200 to the SUPPLY + and terminals (Use hooks).
Transmitter terminal box
+

Power supply

BT200
Figure 6.3.3 BT200 Connection

Ignore the polarity since the BT200 is AC-coupled to the terminal box.
F0603.EPS

6-1

IM 1C21B1-01E

6. WIRING

6.3.4 Check Meter Connection


Connect the check meter to the CHECK + and terminals (use hooks). A 4 to 20 mA DC output signal from the CHECK + and terminals.
(Note) Use a check meter whose internal resistance is 10 or less.

(2) Intrinsically Safe Type For intrinsically safe type, a safety barrier must be included in the loop.
Hazardous Location Transmitter terminal box Distributor (Power supply unit) Nonhazardous Location

Power supply
+

Receiver instrument

Check meter

Transmitter terminal box

F0604.EPS

Safety barrier

F0606.EPS

Figure 6.3.4 Check Meter Connection

Figure 6.4.1b Connection between Transmitter and Distributor

6.4 Wiring
6.4.1 Loop Configuration
Since the DPharp uses a two-wire transmission system, signal wiring is also used as power wiring. DC power is required for the transmitter loop. The transmitter and distributor are connected as shown below. For details of the power supply voltage and load resistance, see Section 6.6; for communications line requirements, see Subsection 8.1.2. (1) General-use Type and Flameproof Type
Hazardous Location Transmitter terminal box Nonhazardous Location

6.4.2 Wiring Installation


(1) General-use Type and Intrinsically Safe Type Make cable wiring using metallic conduit or waterproof glands. Apply a non-hardening sealant to the terminal box connection port and to the threads on the flexible metal conduit for waterproofing.

Flexible metal conduit Apply a non-hardening sealant to the threads for waterproofing.

Wiring metal conduit

Distributor (Power supply unit)

Tee Drain plug


F0607.EPS

Receiver instrument
F0605.EPS

Figure 6.4.2a Typical Wiring Using Flexible Metal Conduit

Figure 6.4.1a Connection between Transmitter and Distributor

6-2

IM 1C21B1-01E

6. WIRING

(2) Flameproof Type Wire cables through a flameproof packing adapter, or using a flameproof metal conduit. Wiring cable through flameproof packing adapter. Apply a nonhardening sealant to the terminal box connection port and to the threads on the flameproof packing adapter for waterproofing.

6.5 Grounding
(a) Grounding should satisfy JIS Class 3 requirements (grounding resistance, 100 or less). Grounding is required for JIS flameproof type and intrinsically safe type.
(Note) If equipped with built-in Lightning Protector, grounding should satisfy Special JIS class 3 requirements (grounding resistance, 10 or less).

Flameproof packing adapter Flexible metal conduit Wiring metal conduit Tee Apply a non-hardening sealant to the threads for waterproofing.

(b) There are ground terminals on the inside and outside of the terminal box. Either of these terminals may be used. (c) Use 600 V grade PVC insulated wires for grounding.
Transmitter terminal box

Ground terminal

Drain plug
F0608.EPS

Figure 6.4.2b Typical Cable Wiring Using Flameproof Packing Adapter

Flameproof metal conduit wiring A seal fitting must be installed near the terminal box connection port for a sealed construction. Apply a non-hardening sealant to the threads of the terminal box connection port, flexible metal conduit and seal fitting for waterproofing.
Gas sealing device Non-hazardous area Flameproof flexible metal conduit Hazardous area Flameproof heavy-gauge steel conduit Tee Drain plug Apply a non-hardening sealant to the threads of these fittings for waterproofing Seal fitting After wiring, impregnate the fitting with a compound to seal tubing.
F0609.EPS

Ground terminal

Figure 6.5 Ground Terminals

6.6 Power Supply Voltage and Load Resistance

When configuring the loop, make sure that the external load resistance is within the range in the figure below.
(Note) In case of an intrinsically safe transmitter, external load resistance includes safety barrier resistance.

600

Figure 6.4.2c Typical Wiring Using Flameproof Metal Conduit

External load resistance R () 250

Figure 6.6 Relationship between Power Supply Voltage and External Load Resistance

    ,
F0610.EPS

R=

E10.5 0.0236

Communication applicable range BRAIN and HART

10.5

16.4

24.7

42

Power supply voltage E (V DC)

F0611.EPS

6-3

IM 1C21B1-01E

7. OPERATION

7.

OPERATION
(b) Turn ON power and connect the BT200. Open the terminal box cover, and connect the BT200 to the SUPPLY + and terminals. (c) Using the BT200, confirm that the transmitter is operating properly. Check parameter values or change the setpoints as necessary. See Chapter 8 for BT200 operation. If the transmitter is equipped with an integral indicator, its indication can be used to confirm that the transmitter is operating properly.

7.1 Preparation for Starting Operation


The Model EJA110A, EJA120A and EJA130A pressure transmitter measures the flow rates and the pressure of liquids, gases, and steam, and also liquid levels. This section describes the operation procedure for the EJA110A as shown in Figure 7.1 (vertical impulse piping type, high-pressure connection: right side) when measuring a liquid flow rate.

NOTE

Orifice Tap valve (high pressure)

Check that the process pressure tap valves, drain valves, and 3-valve manifold stop valves on both low pressure and high pressure sides are closed, and that the 3-valve manifold equalizing valve is opened. (a) Follow the procedures below to introduce process pressure into the impulse piping and transmitter. 1) Open the low pressure and high pressure tap valves to fill the impulse piping with process liquid. 2) Slowly open the high pressure stop valve to fill the transmitter pressure-detector section with process liquid. 3) Close the high pressure stop valve. 4) Gradually open the low pressure stop valve and completely fill the transmitter pressure-detector section with process liquid. 5) Close the low pressure stop valve. 6) Gradually open the high pressure stop valve. At this time, equal pressure is applied to the low and high pressure sides of the transmitter. 7) Check that there are no liquid leaks in the impulse piping, 3-valve manifold, transmitter, or other components. Venting Gas from the Transmitter Pressuredetector Section Since the piping in the example of Figure 7.1 is constructed to be self-venting, no venting operation is required. If it is not possible to make the piping selfventing, refer to Subsection 7.5 for instructions. Leave the equalizing valve open even after venting gas.

Tap valve (low pressure)

Stop valve (low pressure)

3-valve manifold

Equalizing valve Stop valve (high pressure) Drain valve (high pressure)
F0701.EPS

Figure 7.1 Liquid Flow Measurement

7-1

IM 1C21B1-01E

7. OPERATION

Confirming that Transmitter is Operating Properly Using the BT200 If the wiring system is faulty, communication error appears on the display. If the transmitter is faulty, SELF CHECK ERROR appears on the display.
PARAM C60:SELF CHECK ERROR communication error

7.2 Zero Point Adjustment


Adjust the zero point after operating preparation is completed.

IMPORTANT
Do not turn off the power to the transmitter immediately after a zero adjustment. Powering off within 30 seconds after a zero adjustment will return the adjustment back to the previous settings.

DATA

DIAG

PRNT

ESC

Communication error (Faulty wiring)

Self-diagnostic error (Faulty transmitter)


F0702.EPS

Using the integral indicator If the wiring system is faulty, the display stays blank. If the transmitter is faulty, an error code will appear on the display according to the nature of the error.

The zero point adjustment can be made in either way: using the zero-adjustment screw of the transmitter or the BT200 operation. For output signal checking, display the parameter A10: OUTPUT (%) in the BT200.
BT200
PARAM A10:OUTPUT(%) 0.0 % A11:ENGR OUTPUT

Output signal (%) display

Self-diagnostic error on the integral indicator (Faulty transmitter)


F0703.EPS

A20:AMP TEMP

DATA

DIAG

PRNT

ESC

Zero-adjustment Screw

NOTE
If any of the error indications above appears on the display of the integral indicator or BT200, refer to Subsection 8.5.2 for corrective action.

Zero-adjustment screw
F0704.EPS

Verify and Change Transmitter Parameter Setting and Values The following parameters are the minimum settings required for operation. The transmitter has been shipped with these parameters. To confirm or change the values, see Subsection 8.3.3. Measuring range ....... See Subsection 8.3.3 (2) Output/integral indicator mode ......... See Subsection 8.3.3 (4) Operation mode ......... See Subsection 8.3.3 (9)

Using the Transmitter Zero-adjustment Screw Before adjusting a screw, check that the parameter J20: EXT ZERO ADJ displays ENABLE. See Subsection 8.3.3 (15) for the setting procedure. Use a slotted screwdriver to turn the zero-adjustment screw. Turn the screw clockwise to increase the output or counterclockwise to decrease the output. The zero point adjustment can be made with a resolution of 0.01% of the setting range. Since the degree of zero adjustments varies with the screw turning speed, turn the screw slowly for fine adjustment and quickly for coarse adjustment.

7-2

IM 1C21B1-01E

7. OPERATION

Using the BT200 Zero point can be adjusted by simple key operation of the BT200. Select parameter J10: ZERO ADJ, and press the ENTER key twice. The zero point will be adjusted automatically to the output signal 0% (4 mA DC). Confirm that the setting value displayed for the parameter is 0.0% before pressing the ENTER key. See Subsection 8.3.3 (15) for BT200 operating procedures.
SET J10:ZERO ADJ 0.0 % + 000.0

IMPORTANT

A display when parameter J10 is selected. Press key

twice for 0% output 4 mA DC.


CLR ESC
F0705.EPS

Remove the BT200 from the terminal box, and confirm that none of the terminal screws are loosened. Close the terminal box cover and the amplifier cover. Screw each cover in tightly until it will not turn further. Two covers are required to be locked on the CENELEC and SAA Flameproof type transmitters. An Allen head bolts (shrouding bolts) are provided under edge of the each cover for locking. When a shrouding bolts are driven counterclockwise by an Allen wrench, it is coming out and locks up a cover. (See page 93) After locking, the covers should be confirmed not to be opened by hand. Tighten the zero-adjustment cover mounting screw to fix the cover in position.

7.3 Starting Operation


After completing the zero point adjustment, follow the procedure below to start operation. 1) Close the equalizing valve. 2) Gradually open the low pressure stop valve. This places the transmitter in an operational condition. 3) Confirm the operating status. If the output signal exhibits wide fluctuations (hunting) due to periodic variation in the process pressure, use BT200 to dampen the transmitter output signal. Confirm the hunting using a receiving instrument or the integral indicator, and set the optimum damping time constant. See Subsection 8.3.3 (3), Damping Time Constant Setup. 4) After confirming the operating status, perform the following.

7.4 Shutting Down Operation


Shut down the transmitter operation as follows. 1) Turn off the power. 2) Close the low pressure stop valve. 3) Open the equalizing valve. 4) Close the high pressure stop valve. 5) Close the high pressure and low pressure tap valves.

NOTE
Whenever shutting down the transmitter for a long period, remove any process fluid from the transmitter pressure-detector section. The equalizing valve must be left OPEN.

7-3

IM 1C21B1-01E

7. OPERATION

7.5 Venting or Draining Transmitter Pressure-detector Section


Since this transmitter is designed to be self-draining and self-venting with vertical impulse piping connections, neither draining nor venting will be required if the impulse piping is configured appropriately for selfdraining or self-venting operation. If condensate (or gas) collects in the transmitter pressure-detector section, the measured pressure may be in error. If it is not possible to configure the piping for self-draining (or self-venting) operation, you will need to loosen the drain (vent) screw on the transmitter to completely drain (vent) any stagnated liquid (gas). However, since draining condensate or bleeding off gas gives the pressure measurement disturbance, this should not be done when the loop is in operation.

7.5.2

Venting Gas

1) Gradually open the vent screw to vent gas from the transmitter pressur-detector section. (See Figure 7.5.2.) 2) When the transmitter is completely vented, close the vent screw. 3) Tighen the vent screw to a torque of 10 Nm.

Vent screw

Vent screw
When you loosen the vent screw, the gas escpes in the direction of the arrow.
F0707.EPS

WARNING

Figure 7.5.2 Venting the Transmitter

Since the accumulated liquid (or gas) may be toxic or otherwise harmful, take appropriate care to avoid contact with the body, or inhalation of vapors.

7.6 Setting the Range Using the Range-setting Switch


With actual pressure being applied to the transmitter, the range-setting switch (push-button) located on the optional integral indicator plate and the external zeroadjustment screw allow users to change (re-range) the low- and high-limit values for the measurement range (LRV and HRV) without using BT200. However, other changes in the display settings (scale range and engineering unit) for the integral indicator requires BT200. Follow the procedure below to change the LRV and HRV settings. [Example] Rerange LRV to 0 and HRV to 3 MPa. 1) Connect the transmitter and apparatus as shown in Figure 9.3.1 and warm up for at least five minutes. 2) Press the range-setting push-button. The integral indicator then displays LSET. 3) Apply a pressure of 0 kPa (atmospheric pressure) to the transmitter. (Note 1) 4) Turn the external zero-adjustment screw in the desired direction. The integral indicator displays the output signal in %. (Note 2) 5) Adjust the output signal to 0% (1 V DC) by rotating the external zero-adjustment screw. Doing so completes the LRV setting.

7.5.1 Draining Condensate


1) Gradually open the drain screw or drain plug and drain the transmitter pressure-detector section. (See Figure 7.5.1.) 2) When all accumulated liquid is completely removed, close the drain screw or drain plug. 3) Tighten the drain screw to a torque of 10 Nm, and the drain plug to a torque of 34 to 39 Nm.

Drain plug Drain screw


When you loosen the drain screw or drain plug, the accumulated liquid will be expelled in the direction of the arrow.
F0706.EPS

Figure 7.5.1 Draining the Transmitter

7-4

IM 1C21B1-01E

7. OPERATION

6) Press the range-setting push-button. The integral indicator then displays HSET. 7) Apply a pressure of 3 MPa to the transmitter. (Note 1) 8) Turn the external zero-adjustment screw in the desired direction. The integral indicator displays the output signal in %. (Note 2) 9) Adjust the output signal to 100% (5 V DC) by rotating the external zero-adjustment screw. Doing so completes the HRV setting. 10) Press the range-setting push-button. The transmitter then switches back to the normal operation mode with the measurement range of 0 to 3 MPa.
Note 1: Wait until the pressure inside the pressure-detector section has stabilized before proceeding to the next step. Note 2: If the pressure applied to the transmitter exceeds the previous LRV (or HRV), the integral indicator may display error number Er.07 (In this case, the output signal percent and Er.07 are displayed alternately every two seconds). Although Er.07 is displayed, you may proceed to the next step. However, should any other error number be displayed, take the appropriate measure in reference to Subsection 8.5.2, Errors and Countermeasures.

IMPORTANT

Do not turn off the power to the transmitter immediately after completion of the change in the LRV and/or HRV setting(s). Note that powering off within thirty seconds after setting will cause a return to the previous settings. Changing LRV automatically changes HRV to the following value.
HRV = previous HRV + (new LRV previous LRV)

If the range-setting push-button and external zero-adjustment screw are not touched during a range-change operation, the transmitter automatically switches back to the normal operation mode.
Integral indicator Note : Use a thin bar which has a blunt tip, e.g., a hexagonal wrench, to press the rangesetting push-button

Range-setting switch (Push-button)

F0708.EPS

Figure 7.6

Range-setting Switch

7-5

IM 1C21B1-01E

8. BRAIN TERMINAL BT200 OPERATION

8.

BRAIN TERMINAL BT200 OPERATION


8.2 BT200 Operating Procedures
8.2.1 Key Layout and Screen Display

The DPharp is equipped with BRAIN communications capabilities, so that range changes, Tag No. setup, monitoring of self-diagnostic results, and zero point adjustment can be handled by remote control via BT200 BRAIN TERMINAL or CENTUM CS console. This section describes procedures for setting parameters using the BT200. For details concerning the BT200, see IM 1C0A10-E, BT200 Users Manual.

Figure 8.2.1a shows the arrangement of the operating keys on the BT200 keypad, and Figure 8.2.1b shows the BT200 screen component.

8.1 BT200 Operation Precautions


8.1.1 Connecting the BT200

LCD (21 character 8 lines)

Connection to the transmitter with the BT200 can be made by either connecting to the BT200 connection hooks in the transmitter terminal box or by connecting to a relaying terminal board.

Function keys

Movement keys ENTER key

Relaying terminals

Control room Terminal board

Power ON/OFF key Alphanumeric keys

Distributor

Shift keys
F0801.EPS

F0803.EPS

Figure 8.2.1a BT200 Key Layout Figure 8.1.1 Connecting the BT200

8.1.2

Conditions of Communication Line


Cable resistance Rc
Screen title

MENU SCREEN
MENU A:DISPLAY B:SENSOR TYPE BATTERY

Messages

Menu choices
HOME SET ADJ ESC

Power supply

cc

PARAMETER SCREEN
PARAM A10:OUTPUT 100.0 % A11:ENGR. OUTPUT 1000 mmH20 A20:AMP TEMP 23 deg C DATA DI AG

Load resistance R

Cable resistance Rc

DPharp

Parameters

Function commands

PRNT
F0804.EPS

Loop resistance = R + 2Rc = 250 to 600 Loop capacitance = 0.22 F max.

BT200
F0802.EPS

Figure 8.2.1b BT200 Screen Component

Figure 8.1.2 Conditions of Communication Line

8-1

IM 1C21B1-01E

8. BRAIN TERMINAL BT200 OPERATION

8.2.2

Operating Key Functions

(1) Alphanumeric Keys and Shift Keys You can use the alphanumeric keys in conjunction with the shift keys to enter symbols, as well as alphanumeric keys.

Use the function key [F1] CODE to enter symbols. The following symbols will appear in sequence, one at a time, at the cursor each time you press [F1] CODE: / . , + * ) ( & % $ # ! To enter characters next to these symbols, press [ > ] to move the cursor.
Entry Key-in Sequence symbol command

Alphanumeric keys

l/m (I) (/) (m)


T0803.EPS

Shift keys

F0805.EPS

a. Entering Digits, Symbols, and Spaces Simply press the alphanumeric keys.
Entry 4 0.3 1 9
T0801.EPS

(2) Function Keys The functions of the function keys depend on the function commands on display.
MENU A:DISPLAY B:SENSOR TYPE

Key-in Sequence

HOME

SET

ADJ

ESC

Function commands Function keys


F0808.EPS

b. Entering Letters (A through Z) Press an alphanumeric key following a shift key to enter the letter shown on that side which the shift key represents. You must press the shift key before entering each letter.
Left-side letter on the alphanumeric key Right-side letter on the alphanumeric key
F0806 .EPS

Function Command List


Command ADJ Function Displays the ADJ menu

CAPS/caps Selects uppercase or lowercase CODE CLR DATA Selects symbols Erases input data or deletes all data Updates parameter data Deletes one character Calls the self-check panel Returns to the most recent display Displays the menu panel Quits setup and returns to the previous display Proceeds to the next panel Enters the parameter number setup mode Displays the SET menu Returns to the slot selection panel Calls the utility panel Prints out parameters on display Paper feed Lists all parameters in the menu

Entry W IC J. B

Key-in Sequence

DEL DIAG ESC HOME NO


T0802.EPS

Use the function key [F2] CAPS to select between uppercase and lowercase (for letters only). The case toggles between uppercase and lowercase each time you press [F2] CAPS.
Entering uppercase CODE CAPS CLR ESC CODE Entering lowercase caps CLR ESC

OK PARM SET SLOT UTIL *COPY *FEED *LIST

*PON/POFF Automatic printout mode on or off


Entry Boy (B) (o) (y)
F0807.EPS

Key-in Sequence to lower case

*PRNT *GO *STOP

Changes to the print mode Starts printing Cancels printing


T0804.EPS

* Available on BT200-P00 (with printer).

8-2

IM 1C21B1-01E

8. BRAIN TERMINAL BT200 OPERATION

8.2.3

Calling Up Menu Addresses Using the Operating Keys

WELCOME BRAIN TERMINAL ID: BT200


STARTUP SCREEN check connection

push ENTER key UTIL FEED

UTILITY 1.ID 2.SECURITY CODE 3.LANGUAGE SELECT 4.LCD CONTRAST 5.PRINTER ADJUST esc

(UTIL)
PARAM 01:MODEL EJA110A-DM 02:TAG NO. YOKOGAWA 03:SELF CHECK GOOD OK

The utility screen contains the following items. 1. BT200 ID settings 2. Security code settings 3. Switching language of messages (Japanese or English) 4. LCD contrast setting 5. Adjusting printout tone (BT200-P00 only)

INITIAL DATA SCREEN

(ESC)

FUNC 1.MENU 2.UPLOAD TO BT200 3.DOWNLOAD TO INST 4.PRINT ALL DATA HOME SET ADJ ESC

(SET)
(HOME MENU SCREEN) MENU A.DISPLAY B.SENSOR TYPE
MENU SCREEN

(ADJ)
(ADJ MENU SCREEN) MENU J.ADJUST K.TEST M.MEMO P:RECORD ADJ ESC HOME SET ADJ ESC

(SET MENU SCREEN) MENU C.SETTING D.AUX SET 1 E.AUX SET 2 H:AUTO SET

HOME

SET

ADJ

ESC

HOME

SET

PARAM A60:SELF CHECK GOOD PARAM A21:CAPSULE TEMP 26.5 deg C DATA DIAG PRNT

PARAM C60:SELF CHECK GOOD PARAM C22:HIGH RANGE 100 kPa DATA DIAG PRNT PARAM J60:SELF CHECK GOOD PARAM J10:ZERO ADJ 0.0 % DATA DIAG ESC J11:ZERO DEV PRNT 22.2 % J20:EXT. ZERO ADJ ENABLE DATA DIAG PRNT ESC

PARAMETER SCREEN

ESC

ESC

PARAM A10:OUTPUT(%) 50.0 % DATA DIAG PRNT ESC A11:ENGR, OUTPUT 20.0 M A20:AMP TEMP 24.5 deg C DATA DIAG PRNT ESC

PARAM C10:TAG NO. YOKOGAWA DATA DIAG ESC C20:PRESS UNITPRNT kPa C21:LOW RANGE 0 kPa DATA DIAG PRNT ESC

SETUP SCREEN

SET C10:TAG NO. YOKOGAWA YOKOGAWA

See BT200 Instruction Manual for details concerning uploading and downloading parameters and printouts (BT200-P00).

CODE

CAPS

CLR

ESC
F0809.EPS

8-3

IM 1C21B1-01E

8. BRAIN TERMINAL BT200 OPERATION

8.3 Setting Parameters Using the BT200


8.3.1 Parameter Summary
EJA110A, EJA120A, and EJA130A EJA310A, EJA430A, EJA440A, EJA510A, and EJA530A EJA210A and EJA220A
Rewritability

Instruments to which applicable: F: Differential pressure transmitters P: Pressure transmitters L: Liquid level transmitters
No. Item

Description Model+capsule type Tag number Measured data display Output (in %) Output (in engineering units) Amplifier temperature Capsule temperature Static pressure

Remarks

Default Value

Applicability

F P L 16 alphanumerics GOOD/ERROR Menu name 5 to 110%*3 19999 to 19999 Unit specified in D30 Unit specified in D30 Unit specified in D31*1 32000 to 32000 GOOD/ERROR, CAP MODULE FAULT, AMP MODULE FAULT, OUT OF RANGE, OUT OF SP RANGE*, OVER TEMP (CAP), OVER TEMP (AMP), OVER OUTPUT, OVER DISPLAY, ILLEGAL LRV, ILLEGAL HRV, ILLEGAL SPAN, and ZERO ADJ OVER Menu name 16 uppercase alphanumerics 32000 to 32000 32000 to 32000 32000 to 32000

01 MODEL 02 TAG NO. A DISPLAY

03 SELF CHECK Self-diagnostic result A10 OUTPUT (%) A11 ENGR. OUTPUT A20 AMP TEMP A21 CAPSULE TEMP A30 STATIC PRESS A40 INPUT

Input (indicated in engineering DP unit) A60 SELF CHECK Self-diagnostic messages

B B10 B11 B20 B21 B30 B40

SENSOR TYPE MODEL STYLE NO. LRL URL MIN SPAN MAX STAT.P.

Sensor type

Model+span Style number Lower range-limit Upper range-limit Minimum span Maximum static pressure B60 SELF CHECK Self-diagnostic messages

Same as A60 Menu name 16 alphanumerics Selected from mmH2O, mmAq, mmWG, mmHg, Torr, Pa, hPa, kPa, MPa, mbar, bar, gf/cm2, kgf/cm2, inH2O, inHg, ftH2O, psi, or atm 32000 to 32000(but within measurement range) 32000 to 32000(but within measurement range) Selected from 0.2*2, 0.5, 1.0, 2.0, 4.0, 8.0, 16.0, 32.0, or 64.0 sec. Selected from OUT:LIN; DSP:LIN, OUT:LIN; DSP:SQR, OUT:SQR; DSP:SQR Same as A60 Menu name 0.0 to 20.0% LINEAR/ZERO NORMAL %/USER SET, USER & %/INP PRES, PRES & % 8 uppercase alphanumerics 10.0% LINEAR As specified when ordered. As specified when ordered. As specified when ordered.

C SETTING Setting data C10 TAG. NO. Tag number C20 PRESS UNIT Measurement range units

C21 LOW RANGE Measurement range, lower range value C22 HIGH Measurement range, RANGE higher range value C30 AMP Damping time constant DAMPING C40 OUTPUT Output mode and integral indicator mode MODE C60 SELF CHECK Self-diagnostic messages D AUX SET 1 Auxiliary setting data 1 D10 LOW CUT Low cut D11 LOW CUT Low cut mode MODE D20 DISP SELECT Display selection D21 DISP UNIT Engineering unit for display

As specified when ordered. As specified when ordered. 2.0 s As specified when ordered. If not specified, OUT: LIN; DSP: LIN.

T0805.EPS

8-4

IM 1C21B1-01E

8. BRAIN TERMINAL BT200 OPERATION


Applicability

No. D

Item AUX SET 1

Description Auxiliary setting data 1 Engineering range, lower range value Engineering range, higher range value Temperature setting units Static pressure setting units

Rewritability

Remarks Menu name 19999 to 19999 19999 to 19999 deg C/deg F Selected from mmH2O, mmAq, mmWG, mmHg, Torr, Pa, hPa, kPa, MPa, mbar, bar, gf/cm2, kgf/cm2, inH2O, inHg, ftH2O, psi, or atm NORMAL/REVERSE NORMAL/REVERSE

Default Value

F P L As specified when ordered. As specified when ordered. deg C As specified when ordered. If not specified, MPa.

D22 DISP LRV D23 DISP HRV D30 TEMP UNIT D31 STAT. P. UNIT

D40 REV OUTPUT Output reversal D45 H/L SWAP Impulse piping accessing direction D52 BURN OUT CPU error D53 ERROR OUT Hardware error D60 SELF CHECK Self-diagnostic messages E AUX SET 2 Auxiliary setting data 2 E30 BI DIRE Bidirectional mode MODE E60 SELF CHECK Self-diagnostic messages H AUTO SET Automatic setup H10 AUTO LRV Automatic measurement range lower range value setup H11 AUTO HRV Automatic measurement range higher range value setup H60 SELF CHECK Self-diagnostic messages J ADJUST Adjustment data Automatic zero adjustment J11 ZERO DEV. Manual zero adjustment J20 EXT. ZERO External zeroADJ adjustment screw permission J60 SELF CHECK Self-diagnostic messages K TEST Tests K10 OUTPUT in % Test output % setting K60 SELF CHECK Self-diagnostic messages M M10 M20 M30 M40 M50 M60 P P10 P11 P12 P13 P60 MEMO MEMO 1 MEMO 2 MEMO 3 MEMO 4 MEMO 5 SELF CHECK RECORD ERROR REC 1 ERROR REC 2 ERROR REC 3 ERROR REC 4 SELF CHECK Memo Memo Memo Memo Memo Memo Self-diagnostic messages History of the errors Last error One time before Two time before Three time before Self-diagnostic messages J10 ZERO ADJ

If not specified, NORMAL. NORMAL HIGH HIGH

HIGH/LOW, 5 to 110%*3 HOLD/HIGH/LOW, 5 to 110%*3 Same as A60 Menu name OFF/ON

OFF

Same as A60 Menu name 32000 to 32000 Displays the same data as C21. Displays the same data as C22.

32000 to 32000

Same as A60 Menu name 5 to 110.0%*3

ENABLE/INHIBIT

Same as A60 Menu name 5 to 110.0%*3 Displays ACTIVE while executing Same as A60 Menu name 8 uppercase alphanumerics 8 uppercase alphanumerics 8 uppercase alphanumerics 8 uppercase alphanumerics 8 uppercase alphanumerics Same as A60

Display the error Display the error Display the error Display the error Same as A60

*1: In case of Model EJA120A, static pressure cannot be measured. The display is always 0 MPa, but this is not a measured value. *2: When Optional code /F1 is specified, substitute the value with 0.1. *3: When Optional code /F1 is specified, substitute the value 5 with 2.5. T0806.EPS

8-5

IM 1C21B1-01E

8. BRAIN TERMINAL BT200 OPERATION

8.3.2 Parameter Usage and Selection


Before describing the procedure for setting parameters, we present the following table showing how the parameters are used and in what case. If the transmitter is turned off within 30 seconds after parameters have been set, the set data will not be stored and the terminal returns to previous settings.

Table 8.3.1 Parameter Usage and Selection Setup Item Tag No. setup P.8-7 Calibration range setup P.8-7 Description Sets the Tag No. (using 16 alphanumeric characters). Note: Up to 8 alphanumerics (upper case letters) can be used in the BT100. Sets the calibration range for 4 to 20 mA DC. Sets three data items: range unit, input value at 4 mA DC (LRV), and input value at 20 mA DC (HRV). Note: LRV and HRV can be specified with range value specifications up to 5 digits (excluding any decimal point) within the range of 32000 to 32000. Adjusts the output response speed for 4 to 20 mA DC. Can be set in 9 increments from 0.2 to 64 s. Sets modes for output signal and integral indicator to Linear mode (proportional to input differential pressure) or to Square root mode (proportional to flow). Used mainly to stabilize output near 0% if output signal is the square root mode. Two modes are available: forcing output to 0% for input below a specific value, or changing to proportional output for input below a specific value. Sets the following 5 types of integral indicator scale ranges and units: % scale indicator, user set scale indicator, alternate indication of user set scale and % scale, input pressure display, alternate indication of input pressure and % scale When using the user set scale, 4 types of data can be set: user set scale setting, unit (BT200 only), display value at 4 mA DC (LRV), and display value at 20 mA DC (HRV). Note: LRV and HRV can be specified with range value specifications up to 5 digits (excluding any decimal point) within the range of 19999 to 19999. Sets a unit for temperatures displayed on the BT200. Sets a unit for static pressure displayed on the BT200. Reverses the direction for 4 to 20 mA DC output relative to input. Reverse mode is used for applications in which safety requires that output be driven toward 20 mA if input is lost. Used where installation conditions make it imperative to connect high pressure side impulse line to low pressure side of transmitter. Reversal of orientation should be dealt with by reversing impulse line wherever possible. Use this function only where there is no alternative. Displays the status of 4 to 20 mA DC output when a CPU failure. The parameter of the standard unit is fixed to the high limit value.

Damping time constant setup P.8-8 Output and integral indicator display mode setup P.8-9 Output signal low cut mode setup P.8-9 Integral indicator scale range and unit setup P.8-10

Unit setup for displayed temperature P.8-11 Unit setup for displayed static pressure P.8-11 Operation mode (normal/reverse signal) setup P.8-12 Impulse line connection orientation (higher pressure on right/left side) setup P.8-12 Output status display/setup when a CPU failure P.8-12

Output status setup when a hardware Sets the status of the 4 to 20 mA DC output when an abnormal status is detected error occurs with the capsule or the amplifier as the result of self-diagnosis. One of the P.8-12 following statuses; last held, high limit, and low limit values, can be selected. Bi-directional flow measurement P.8-13 Used to measure bi-directional flows. Output at zero flow is 12 mA DC, with output range equally divided between forward and reverse flow. Can be used with square root mode. Range for 4 to 20 mA DC signal is set with actual input applied. Sets 20 mA DC output precisely with respect to users reference instrument output. Note that DPharp is calibrated with high accuracy before shipment, so span should be set using the normal range setup. Adjusts zero point. This can be done either using the external zero-adjustment screw on the transmitteror using the BT200. Used for loop checks. Output can be set freely from 5% to 110% in 1% steps. Allows user to enter up to 5 items of any desired text in up to 8 uppercase alphanumeric characters per item.
T0807.EPS

Range change (while applying actual inputs) P.8-13 Zero point adjustment P.8-14 Test output (fixed current output) setup P.8-15 User memo fields P.8-15

8-6

IM 1C21B1-01E

8. BRAIN TERMINAL BT200 OPERATION

8.3.3 Setting Parameters


Set or change the parameters as necessary. After completing these, do not fail to use the DIAG key to confirm that GOOD is displayed for the selfdiagnostic result at _60: SELF CHECK. (1) Tag No. Setup (C10: TAG NO) Use the procedure below to change the Tag No. Up to 16 alphanumeric characters can be entered.

SET C10:TAG NO. YOKOGAWA FIC-1a PRINTER OFF F2:PRINTER ON FEED POFF NO

This is the panel for confirming set data. The set data items flash. When all items have been confirmed, press the again. (To go back to the setting panel, press the (NO) key.

SET C10:TAG NO. FIC-1a

The DPharp TAG NO. was overwritten. Press the (OK) key to return to the parameter panel. Press the (NO) key to return to the setting panel.

FEED

NO

OK

Example: Set a Tag No. to FIC-1a


Press the the BT200.
<When power is off>

key to turn on

PARAM C10:TAG NO. FIC-1a C20:PRESS UNIT kPa C21:LOW RANGE 0 kPa DATA DIAG PRNT

ESC
F0811.EPS

WELCOME BRAIN TERMINAL ID: BT200 check connection push ENTER key UTIL FEED

Connect DPharp and BT200 using a communication cable and press the key.

(2) Calibration Range Setup a. Setting Calibration Range Unit (C20: PRESS UNIT) The unit is set at the factory before shipment if specified at the time of order. Follow the procedure below to change the unit.
Example: Change the unit from mmH2O to kPa.
SET C20:PRESS UNIT mmH20 < mmWG > < mmHG > < Torr > < kPa > ESC SET C20:PRESS UNIT kPa

PARAM 01:MODEL EJA110A-DM 02:TAG NO. YOKOGAWA 03:SELF CHECK GOOD OK MENU A:DISPLAY B:SENSOR TYPE

Displays the name of connected DPharp model, TAG NO. and diagnostics information. Press the (OK) key after confirmation. Press the (SET) key to display the SET menu panel.

Use the Press the

or key twice

key to select kPa. to enter the setting. Press the (OK) key.

HOME

SET

ADJ

ESC

MENU C:SETTING D:AUX SET 1 E:AUX SET 2 H:AUTO SET HOME SET ADJ ESC

Select C: SETTING and press the key.

MENU C10:TAG NO. YOKOGAWA C20:PRESS UNIT kPa C21:LOW RANGE 0 kPa DATA DIAG PRNT SET C10:TAG NO. YOKOGAWA YOKOGAWA

Select C10: TAG NO. and press the key.

FEED

NO

OK

mmH2O
ESC

Set the new TAG NO. (FIC-1a). FOKOGAWA FIKOGAWA FICOGAWA FIC-GAWA FIC-1AWA FIC-1aWA FIC-1a

mmAq mmWG mmHg Torr kPa MPa mbar bar gf/cm2 kgf/cm2 inH2O inHg ftH2O psi atm

CODE

CAPS

CLR

ESC

SET C10:TAG NO. YOKOGAWA FIC-1a _

Set TAG NO. and press the key.

F0812.EPS

CODE

caps

CLE

ESC

When you have made an entry mistake, return the cursor using the key, then reenter.
F0810.EPS

8-7

IM 1C21B1-01E

8. BRAIN TERMINAL BT200 OPERATION

b. Setting Calibration Range Lower Range Value and Higher Range Value (C21: LOW RANGE, C22: HIGH RANGE) These range values are set as specified in the order before the instrument is shipped. Follow the procedure below to change the range. The measurement span is determined by the high and low range limit values. In this instrument, changing the low range value also automatically changes the high range value, keeping the span constant.

Example 2: With present settings of 0 to 30 kPa, set the Higher range value to10 kPa.
SET C22:HIGH RANGE 30 kPa + 10

Set 10. Press the to enter the setting.


ESC

key twice

DEL

CLR

SET C22:HIGH RANGE 10 kPa

Press the

(OK) key.

FEED

NO

OK

Example 1: With present settings of 0 to 30 kPa, set the lower range value to 0.5 kPa.
SET C21:LOW RANGE 0 kPa + 0.5

PARAM C20:PRESS UNIT kPa C21:LOW RANGE 0 kPa C22:HIGH RANGE 10 kPa DATA DIAG PRNT

The low range value is not changed, so the span changes.

ESC

Set 0.5. Press the enter the setting.


ESC

key twice to

F0814.EPS

DEL

CLR

SET C21:LOW RANGE 0.5 kPa

Press the

(OK) key.

FEED

NO

OK

(3) Damping Time Constant Setup (C30: AMP DAMPING) When the instrument is shipped, the damping time constant is set at 2.0 seconds. Follow the procedure below to change the time constant. Example: Change from 2.0 sec to 4.0 sec.
SET C30:AMP DAMPING 2.0 sec < 2.0 sec < 4.0 sec < 8.0 sec < 16.0 sec ESC
F0813.EPS

SET C20:PRESS UNIT kPa C21:LOW RANGE 0.5 kPa C22:HIGH RANGE 30.5 kPa DATA DIAG PRNT

The higher range value is changed while the span remains constant.
ESC

Span = Higher range value Lower range value

> > > >

Use the or select 4.0 sec. Press the enter the setting. Press the

key to key twice to

SET C30:AMP DAMPING 4.0 sec

(OK) key.

Note, however, that changing the higher range value does not cause the lower range value to change. Thus, changing the higher range value also changes the span. Calibration range can be specified with range value specifications up to 5 digits (excluding any decimal point) for low or high range limits within the range of 32000 to 32000.

FEED

NO

OK

0.2sec 0.5sec 1.0sec 2.0sec 4.0sec 8.0sec 16.0sec 32.0sec 64.0sec


F0815.EPS

Note: The damping time constant set here is the damping time constant for the amplifier assembly. The damping time constant for the entire transmitter is the sum of the values for the amplifier assembly and for the capsule assembly. For the capsule assembly damping time constant (fixed), see the General Specifications found at the end of this manual. (See Chapter 10.)

8-8

IM 1C21B1-01E

8. BRAIN TERMINAL BT200 OPERATION

(4) Output Mode and Integral Indicator Display Mode Setup (C40: OUTPUT MODE) The mode setting for the output signal and the integral indicator coordinate as shown in the table below.
BT200 Display OUT: LIN DSP: LIN OUT: LIN DSP: SQR OUT: SQR DSP: SQR Output Mode Linear Linear Square root Integral Indicator Display Mode Linear Square root Square root
T0808.EPS

Example: Change the low cut setting range from 10% to 20%, and the low cut mode from LINEAR to ZERO.
SET D10:LOW CUT 10.0 % + 20.0

Set 20. Press the enter the setting.


ESC

key twice to

CLR SET D10:LOW CUT 20.0 %

Press the

(OK) key.

Next, the [D11: LOW CUT MODE] setting panel is displayed.


NO OK

This mode is set as specified in the order when the instrument is shipped. Follow the procedure below to change the mode. If the instrument is equipped with an integral indicator and the display mode is square root, is displayed on the integral indicator. For details, see Chapter 3.

FEED

SET C11:LOW CUT MODE LINEAR < LINEAR > < ZERO > ESC SET D11:LOW CUT MODE ZERO

Use the

or

key

to select ZERO. Press the enter the setting. Press the (OK) key. key twice to

Example: Set output mode to Linear and display mode to Square root.
SET C40:OUTPUT OUT:LIN <OUT:LIN <OUT:LIN <OUT:SQR MODE DSP:LIN DSP:LIN > DSP:SQR.> DSP:SQR > ESC SET C40:OUTPUT MODE OUT:LIN DSP:SQR

FEED

NO

OK

Use the

or

key

to select OUT: LIN, DSP: SQR. Press the enter the setting. Press the (OK) key. key twice to

PARAM D10:LOW CUT 20.0 % D11:LOW CUT MODE ZERO D20:DISP SELECT NORMAL % DATA DIAG PRNT

ESC

F0818.EPS

FEED

NO

ESC

F0816.EPS

(5) Output Signal Low Cut Mode Setup (D10: LOW CUT, D11: LOW CUT MODE) Low cut mode can be used to stabilize the output signal near the zero point. The low cut point can be set in a range from 0 to 20% of output. (Hysteresis of cut point: 1%) Either LINEAR or ZERO can be selected as the low cut mode.
Low cut mode LINEAR (%) 50 Output 20 20 LOW CUT at 20% Low cut mode ZERO (%) 50

Input

50 (%) 0

Input

50 (%)
F0817.EPS

8-9

IM 1C21B1-01E

8. BRAIN TERMINAL BT200 OPERATION

(6) Integral Indicator Scale Setup The following 5 displays are available for integral indicators.
D20: DISP SELECT and Display NORMAL % Description and Related parameters Indicates 5 to 110% range depending on the Measurement range (C21, C22).
A10:OUTPUT (%) 45.6 %

% indication and input pressure indication


D20: DISP SELECT NORMAL % INP PRES PRES & %

User-set engineering unit display

D20: DISP SELECT USER SET USER & %

Transmitter is set for % display when shipped.

Set for user-set engineering unit display.

USER SET

Indicates values depending on the Engineering range (D22, D23).(Note 1) Units set using Engineering unit (D21) are not indicated.
A11:ENGR.OUTPUT 20.0 M

D21: DISP UNIT Set a unit to be displayed on the BT200.

D22: DISP LRV

USER & %

Indicates user set and % alternately in 3 second intervals.


A10:OUTPUT (%) 45.6 % A11:ENGR. OUTPUT 20.0 M

Set a numeric value for engineering unit for 4 mA output (LRV).

D23: DISP HRV Set a numeric value for engineering unit for 20 mA output (HRV).
F0818.EPS

INP PRES

Indicates input pressure. Indication limits 19999 to 19999.


A40:INPUT 456 kPa

PRES & %

Indicates input pressure and % alternately in 3 second intervals.


A10:OUTPUT (%) 45.6 % A40:INPUT 456 kPa

a. Display Selection (D20: DISP SELECT) Follow the instructions given to the below to change the range of integral indication scales. When USER SET is selected, the user set values of integral indication and A11: ENGR. OUTPUT parameter are indicated. Example: Set the integral indicator scale to engineering units display.

(Note 1) Scale range can be specified with range limit specifications up to 5 digits (excluding any decimal point) for low or high range limits within the range of 19999 to 19999.
T0808.EPS

See (a.) through (c.) for each setting procedure.


SET D20:DISP SELECT NORMAL % <NORMAL %> <USER SET> <USER & %> <INP PRES> ESC SET D20:DISP SELECT USER SET

Use the

or

key

to select USER SET. Press the enter the setting. Press the (OK) key. key twice to

FEED

NO

OK

The % disappears from the integral indicator display.


F0820.EPS

8-10

IM 1C21B1-01E

8. BRAIN TERMINAL BT200 OPERATION

b. Setting User-set Engineering Unit (D21: DISP UNIT) This parameter allows entry of the engineering units to be displayed on the BT200. When the instrument is shipped, this is set as specified in the order. Follow the procedure below to change this setting. Since these units are not displayed on the integral indicator, use the adhesive labels provided. This parameter need not be set for % display.
Example: Set an engineering unit M.
SET D21:DISP UNIT M_

c. Lower and Higher Range Value Setup in Engineering Unit (D22: DISP LRV, D23: DISP HRV) These parameter items are used to set the lower and higher range values for the engineering unit display. When the instrument is shipped, these are set as specified in the order. Follow the procedure below to change these settings. Note that these parameters need not be set for % display.

Example: Set lower range value (LRV) to 50 and higher range value (HRV) to 50.
Setting LRV
SET D22:DISP LRV 0M 50

Set M. Press the enter the setting. key twice to

Set 50. Press the enter the setting. key twice to

CODE

CAPS

CLR

ESC

SET D21:DISP UNIT M

Press the

(OK) key.

DEL

CLR

ESC

Setting HRV
FEED NO OK

SET D23:DISP HRV 100M + 50


F0821.EPS

Set 50. Press the enter the setting.


CLR ESC

key twice to

DEL

SET D23:DISP HRV 50M

Press the

(OK) key.

FEED PARAM D21:DISP M D22:DISP D23:DISP DATA

NO

OK

UNT

LRV 50M HRV 50M DIAG PRNT

ESC
F0822.EPS

(7) Unit Setup for Displayed Temperature (D30: TEMP UNIT) When the instrument is shipped, the temperature units are set to degC. Follow the procedure below to change this setting. Note that changing the unit here changes the unit for A20: AMP TEMP (amplifier temperature) and A21: CAPSULE TEMP (capsule temperature).

Example: Change the unit for the temperature display.


SET D30:TEMP UNIT deg C < deg C > < deg F > ESC

Use the select deg F. Press the enter the setting.

or

key to

key twice to

F0823.EPS

8-11

IM 1C21B1-01E

8. BRAIN TERMINAL BT200 OPERATION

(8) Unit Setup for Displayed Static Pressure (D31: STAT.P.UNIT) Follow the procedure below to change the static pressure units. Changing this parameter changes the unit for the A30: STATIC PRESS (static pressure) display. (9) Operation Mode Setup (D40: REV OUTPUT) This parameter allows the direction of the 4 to 20 mA output to be reversed with respect to input. Follow the procedure below to make this change. Example: Change 4 to 20 mA output to 20 to 4 mA output.
SET D40:REV OUTPUT NORMAL < NORMAL > < REVERSE> ESC

Optional code/C1 The parameter is set to LOW. If a failure, output which is 5% or lower is generated. The parameter D53: ERROR OUT is set to LOW from the factory.

Example: Standard specifications


D52: BURN HIGH OUT

pin (CN4) position: H

Example: Optional code/C1


D52: BURN LOW OUT

pin (CN4) position: L

F0827.EPS

Use the

or

key

to select REVERSE. Press the enter the setting.


F0825.EPS

key twice to

(12) Output Status Setup when a Hardware Error Occurs (D53: ERROR OUT) This parameter allows the setting of the output status when a hardware error occurs. The following three selections are available. (a) HOLD; Outputs the last value held before the error occurred. (b) HIGH; Outputs an output of 110% when an error has occurred. (c) LOW; Outputs an output of 5% when an error has occurred.
Note: A hardware error means CAP MODULE FAULT of Er.01 or AMP MODULE FAULT of Er. 02 which are shown in 8.5.2 Errors and Countermeasures.)

(10) Impulse Line Connection Orientation Setup (D45: H/L SWAP) This parameter allows the impulse line connections to be reversed at the transmitter. Follow the figure below to make this change.
Example: Change the impulse line connection from high pressure on right to high pressure on left.
SET D45:H/L SWAP NORMAL < NORMAL > < REVERSE> ESC

Example: Set the output status to LOW when a hardware error occurs.
SET D53:ERROR OUT HIGH < HIGH> < LOW> < HOLD> ESC

Use the

or

key

to select REVERSE. Press the enter the setting.


F0826.EPS

Use the to select LOW. Press the enter the setting.

or

key

key twice to

key twice to

F0828.EPS

(11) Output Status Display/Setup when a CPU Failure (D52: BURN OUT) This parameter displays the status of 4 to 20 mA DC output if a CPU failure occurs. In case of a failure, communication is disabled. Setting of HIGH or LOW is enabled. This is done with the pin (CN4) on the CPU assembly. See Chapter 3 for details. Standard specifications The parameter is set to HIGH. If a failure, the transmitter outputs the signal of 110% or higher. The parameter D53: ERROR OUT is set to HIGH from the factory.

8-12

IM 1C21B1-01E

8. BRAIN TERMINAL BT200 OPERATION

(13) Bi-directional Flow Measurement Setup (E30: BI DIRE MODE) (a) This parameter enables selection of 50% output at an input of 0 kPa. Procedure is shown in the figure below. (b) Combining this with C40: OUTPUT MODE provides a square root output computed independently for 0% to 50% output and for 50% to 100% output. Example: If measurement range is 0 to 10 kPa (LRV=0 kPa, HRV=10 kPa)
SET E30:BI DIRE MODE OFF < OFF > < ON > ESC

(14) Range Change while Applying Actual Inputs (H10: AUTO LRV, H11: AUTO HRV) This feature allows the lower and higher range values to be set up automatically with the actual input applied. If the lower and higher range values are set, C21: LOW RANGE and C22: HIGH RANGE are changed at this same time. Follow the procedure in the figure below. The measurement span is determined by the higher and lower range values. Changing the lower range value results in the higher range value changing automatically, keeping the span constant. Example 1: When changing the lower range value to 0.5 kPa for the present setting of 0 to 30 kPa, take the following action with input pressure of 0.5 kPa applied.
SET H10:AUTO LRV 0 kPa + 0

Use the to select ON. Press the enter the setting.

or

key

key twice to

The measurement range changes to 10 to 0 to 10 kPa (output 0% to 50% to 100). Note that C21: LOW RANGE and C22: HIGH RANGE are not changed.
F0829.EPS

Press the

key twice.

The lower range value is changed to 0.5 kPa.


ESC

Output mode LINEAR 20 mA (100% display)

SET H10:AUTO LRV 0.5000 kPa

Press the

(OK) key.

FEED

NO

OK

LRV

HRV

4 mA (100% display)

PARAM H10:AUTO LRV 0.5000 kPa H11:AUTO HRV 30.500 kPa H60:SELF CHEC GOOD DATA DIAG PRNT

The higher range value is changed keeping the span constant. Parameters C21 and C22 are
ESC

Output mode SQUARE ROOT Low Cut 20 mA (100% display)

changed at the same time.


F0831.EPS

LRV

HRV

4 mA (100% display)
F0830.EPS

8-13

IM 1C21B1-01E

8. BRAIN TERMINAL BT200 OPERATION

Note that changing the higher range value does not cause the lower range value to change but does change the span.
Example 2: When the higher range value is to be changed to 10 kPa with the present setting of 0 to 30 kPa, take the following action with an input pressure of 10 kPa applied.
SET H10:AUTO HRV 30 kPa + 30

(a) Follow the procedure below when setting the present output to 0% (4 mA).

A10:OUTPUT (%) 0.5 %

Output is 0.5%.

SET J10:ZERO ADJ 0.0 % + 000.0

Press the

key twice.

CLR

ESC

Press the

key twice.

The higher range value is changed to 10 kPa.


ESC

SET J10:ZERO ADJ 0.0 %

Zero adjustment is completed. Press the


NO OK

(OK) key.

FEED
SET H11:AUTO HRV 10.000 kPa

Press the

(OK) key.
A10:OUTPUT (%) 0.0 %

Output is 0%.

FEED

NO

OK

F0833.EPS

PARAM H10:AUTO LRV 0 kPa H11:AUTO HRV 10.000 kPa H60:SELF CHECK GOOD DATA DIAG PRNT

The lower range value is not changed, so the span changes. Parameter C22 is changed at the
ESC

same time.
F0832.EPS

(b) In tank level measurement, if the actual level cannot be brought to zero for zero adjustment, then the output can be adjusted to correspond to the actual level obtained using another measuring instrument such as a glass gauge.
Present level: 45% Present output: 41% 100%

(15)

Zero Point Adjustment (J10: ZERO ADJ, J11: ZERO DEV, J20: EXT ZERO ADJ) The DPharp supports several adjustment methods. Select the method best suited for the conditions of your application. Note that output signal can be checked by displaying parameter A10:OUTPUT (%) on the BT200.
Adjustment Method Using the BT200 Description Set the present input to 0%. Adjust for 0% output at input level of 0%. Adjust output to the reference value obtained using other means. If the input level cannot easily be made 0% (because of tank level, etc.), adjust output to the reference value obtained using other means, such as a sight glass. Using the external zero-adjustment screw Adjust zero point using the zeroadjustment screw on the transmitter. This permits zero adjustment without using the BT200. Accurately adjust the output current to 4 mA DC or other target output value using an ammeter that accuratly reads output currents.
T0810.EPS

Present level 45% EJA 0%


F0834.EPS

(b)-1 Follow the procedure below to use J10: ZERO ADJ.


Present output is 41.0%.

A10:OUTPUT (%) 41.0 %

SET J10:ZERO ADJ 0.0 % + 040.0

Enter the present actual level, 40%. Press the


ESC

key twice.

CLR

A10:OUTPUT (%) 40.0 %

The output is changed to 40%.

F0835.EPS

8-14

IM 1C21B1-01E

8. BRAIN TERMINAL BT200 OPERATION

(b)-2 Follow the procedure below to use J11: ZERO DEV.


Present output is 41.0%. Output error = 45.0 41.0 = 4.0%. Since J11: ZERO DEV. contains the previous correction, obtain the correction value by adding 4.0% to
ESC

(16) Test Output Setup (K10: OUTPUT X%) This feature can be used to output a fixed current from 3.2 mA (5%) to 21.6 mA (110%) for loop checks.

A10:OUTPUT (%) 41.0 %

Example: Output 12 mA (50%) fixed current.


SET K10:OUTPUT X % 0.0 % + 050.0

SET J11:ZERO DEV. 2.50 % 0

Set 50.0%. Press the


ESC

key twice to

it. (2.50% + 4.0% = 6.50%) Set the correction value, 6.50. Press the key twice.
FEED NO

output a fixed current at 50%.


SET K10:OUTPUT X % 50.0 % ACTIVE

SET J11:ZERO DEV. 2.50 % 6.50

Active is displayed while this is being executed. Press the (OK) key to cancel

ECS

OK

the fixed current output.


F0838.EPS

A10:OUTPUT (%) 45.0 %

The output is changed to 45%.

F0836.EPS

(c) Zero Point Adjustment Using the External Zero Adjustment Screw Enabling/inhibiting of zero point adjustment using the external zero-adjustment screw on the transmitter (J20: EXT ZERO ADJ) Follow the procedure below to enable or inhibit zero point adjustment from the zero-adjustment screw on the transmitter. This is set to ENABLE when the instrument is shipped.
Example: Inhibiting zero adjustment by the external zero-adjustment screw
SET J20:EXIT ZERO ADJ ENABLE < ENABLE > < INHIBIT> ESC

1. Test output is held for approximately 10 minutes, and then released automatically after the time has elapsed. Even if the BT200 power supply is turned off or the communication cable is disconnected during test output, it is held for approximately 10 minutes. 2. Press the (OK) key to release test output immediately.

Use the select INHIBIT. Press the enter the setting.

or

key to

(17) User Memo Fields (M: MEMO) This feature provides 5 user memo fields, each holding up to 8 alphanumeric characters. Up to 5 items such as inspection date, inspector, and other information can be saved in these fields.

key twice to

F0837.EPS

Example: Save an inspection date of January 30, 1995.


PARAM M10:MEMO 1 M20:MEMO 2 M30:MEMO 3

Zero point adjustment using external zero-adjustment screw on the transmitter Turn the zero-adjustment screw on the outside of the transmitter case using a slotted screwdriver. Turn the screw to the right to increase the zero point or to the left to decrease the zero output; the zero adjusts in increments of 0.01% of the range setting. Note that the amount of adjustment to the zero point changes according to the speed at which the screw is turned. To make fine adjustments, turn the screw slowly; to make coarse adjustments, turn the screw quickly.
Note: When a zero point adjustment has been made, do not turn off the transmitter less than 30 seconds after adjustment.

Set 95.1.30 in the order of year, month, and day. Press the
PRNT ESC

key twice to

DATA

DIAG

enter the setting.

SET M10:MEMO 1 95.1.30_

ESC

F0839.EPS

8-15

IM 1C21B1-01E

8. BRAIN TERMINAL BT200 OPERATION

8.4 Displaying Data Using the BT200


8.4.1 Displaying Measured Data
The BT200 can be used to display measured data. The measured data is updated automatically every 7 seconds. In addition, the display can be updated to the present data value at any time by pressing the (DATA) key. For parameters associated with the display of measured data, see Subsection 8.3.1, Parameter Summary.

8.5 Self-Diagnostics
8.5.1 Checking for Problems
(1) Identifying Problems with BT200 The following four areas can be checked. (a) Whether connections are good. (b) Whether BT200 was properly operated. (c) Whether settings were properly entered. (d) History of the errors. See examples below.
Example 1: Connection errors

Example: Display output.


MENU A:DISPLAY B:SENSOR TYPE

WELCOME BRAIN TERMINAL ID: BT200 check connection push ENTER key UTIL FEED

Press the

key.

When the panel shown on the left appears, press the key.

Since communications will be


HOME SET ADJ ESC
communication error

unsuccessful if there is a problem in the connection to the BT200, the

PARAM A10:OUTPUT (%) XX.X % A11:ENGR.OUTPUT YY.Y % A20:AMP TEMP ZZ deg C DATA DIAG PRNT PARAM A10:OUTPUT (%) A11:ENGR.OUTPUT A20:AMP TEMP

Display A10: OUTPUT (%).


ESC

display at the left will appear. Recheck the connection.

ESC

communi

Data is updated automatically at 7-second intervals.

Press the

(OK) key.

Example 2: Setting entry errors


PARAM 01:MODEL EJA110A-DM 02:DAG NO. YOKOGAWA 03:SELF CHECK ERROR OK PARAM C20:PRESS UNIT kPa C21:LOW RANGE 600 kPa C22:HIGH RANGE 600 kPa DATA DIAG PRNT DIAG C60:SELF CHECK ERROR < ERROR > < ILLEGAL LRV > FEED PRNT ESC

The initial data panel shows the result of current transmitter diagnostics.

F0840.EPS

8.4.2

Display Transmitter Model and Specifications

Press the

(DIAG) key in the

The BT200 can be used to display the model and specifications of the transmitter.

parameter panel to go to the diagnostics panel


ESC

(C60: SELF CHECK). An error message is displayed when an error occurs in the diagnostics panel.

Example: View transmitter model name.


MENU A:DISPLAY B:SENSOR TYPE

Press

F0842.EPS

HOME

SET

ADJ

ESC

PARAM B10:MODEL EJA110A-DM B11:STYLE NO. S1.01 B20:LRL 98.07 kPa DATA DIAG PRNT

For the associated parameters, see Subsection 8.3.1, Parameter Summary.

ESC
F0841.EPS

8-16

IM 1C21B1-01E

8. BRAIN TERMINAL BT200 OPERATION

Example 3: Checking the history of the errors


MENU J:ADJUST K:TEST M:MEMO P:RECORD HOME SET ADJ ESC

(2) Checking with Integral Indicator

Connect the BT200 to the transmitter, and call item P.

PARAM P10:ERROR REC 1 ERROR P11:ERROR REC 2 ERROR P12:ERROR REC 3 GOOD DATA DIAG PRNT

If an error is detected in the self-diagnostic, an error number is displayed on the integral indicator. If there is more than one error, the error number changes at two-second intervals. See Table 8.5.1 regarding the error numbers.

ESC

P10: ERROR REC 1 displays the last error. P11: ERROR REC 2 displays the error one time before the last error occurred. P12: ERROR REC 3 displays the error two times before the last error occurred. P13: ERROR REC 4 displays the error three times before the last error occurred. The history of up to four errors can be stored. When the 5th error has occurred, it is stored in P10. The error stored in P13 will be deleted, and then, the error in P12 will be copied to P13. In this sequence, the history of the most previously occurred error will be removed from memory. GOOD will be displayed if there was no previous error.
SET P10:ERROR REC 1 ERROR < ERROR > < ILLEGAL LRV > < ILLEGAL HRV > ESC

F0844.EPS

Figure 8.5.1 Identifying Problems Using the Integral Indicator

Select P10: ERROR REC1 and press the the error message. key to display

<(a) SETUP PANEL> For the details of the messages listed below, see Table 8.5.1 Error Message Summary.
CAP MODULE FAULT AMP MODULE FAULT OUT OF RANGE OUT OF SP RANGE OVER TEMP (CAP) OVER TEMP (AMP) OVER OUTPUT OVER DISPLAY ILLEGAL LRV ILLEGAL HRV ILLEGAL SPAN ZERO ADJ OVER

Note 1: Press the key twice in the setting panel (panel 1) to clear all error message (P10 to P13) information. Note 2: After two hours from when an error occurs, the error message of that error will be recorded. Therefore, if you switch off the transmitter within two hours from when the error occurs, there is no history of that error stored in the transmitter, and this function is meaningless.
F0843.EPS

8-17

IM 1C21B1-01E

8. BRAIN TERMINAL BT200 OPERATION

8.5.2 Errors and Countermeasures


The table below shows a summary of error messages.
Table 8.5.1 Error Message Summary
Integral Indicator Display None ---Er. 01

BT200 Display GOOD ERROR CAP MODULE FAULT

Cause

Output Operation during Error

Countermeasure

Capsule problem.*1

Outputs the signal (Hold, High, or Low) set with parameter D53. Outputs the signal (Hold, High, or Low) set with parameter D53. Outputs high range limit value or low range limit value. Displays present output. Displays present output.

Replace capsule.*2

Er. 02

AMP MODULE FAULT

Amplifier problem.

Replace amplifier.

Er. 03

OUT OF RANGE

Input is outside measurement range limit of capsule. Static pressure exceeds specified range.*3 Capsule temperature is outside range (50 to 130C). Amplifier temperature is outside range (50 to 95C).

Check input.

Er. 04

OUT OF SP RANGE OVER TEMP (CAP)

Check line pressure (static pressure). Use heat insulation or make lagging to keep temperature within range. Use heat insulation or make lagging to keep temperature within range. Check input and range setting, and change them as needed. Check input and display conditions and modify them as needed. Check LRV and modify as needed. Check HRV and modify as needed. Check SPAN and change as needed. Readjust zero point.

Er. 05

Er. 06

OVER TEMP (AMP)

Displays present output.

Er. 07

OVER OUTPUT

Output is outside high Outputs high or low or low range limit range limit value. value. Displayed value is outside high or low range limit value. Displays high or low range limit value.

Er. 08

OVER DISPLAY

Er. 09

ILLEGAL LRV

LRV is outside setting Holds output range. immediately before error occurrence. HRV is outside setting Holds output range. immediately before error occurrence. SPAN is outside setting range. Holds output immediately before error occurrence.

Er. 10

ILLEGAL HRV

Er. 11

ILLEGAL SPAN

Er. 12

ZERO ADJ OVER Zero adjustment is too Displays present large. output.

*1: For Model EJA510A and EJA530A, this error code appears when an illegal overpressure is applied to the pressure sensor in addition to the capsule problem. The Er. 01 will remain even if the normal input pressure is regained. *2: For Model EJA510A and EJA530A, restart the power to the transmitter. Then if no error code appears, perform necessary adjustments such as zero-adjustment to continue the operation. If the error code still exists, replace the capsule assembly. *3: For Model EJA120A, static pressure cannot be measured. The display is always 0 MPa, but this is not a measured value. T0811 .EPS

8-18

IM 1C21B1-01E

9. MAINTENANCE

9.

MAINTENANCE
9.2 Calibration Instruments Selection
Table 9.2.1 shows the instruments required for calibration. Select instruments that will enable the transmitter to be calibrated or adjusted to the required accuracy. The calibration instruments should be handled carefully so as to maintain the specified accuracy.

9.1 Overview
WARNING

Since the accumulated process fluid may be toxic or otherwise harmful, take appropriate care to avoid contact with the body, or inhalation of vapors during draining condensate or venting gas in transmitter pressure-detector section and even after dismounting the instrument from the process line for maintenance. Maintenance of the transmitter is easy due to its modular construction. This chapter describes the procedures for calibration, adjustment, and the disassembly and reassembly procedures required for component replacement. Since the transmitters are precision instruments, carefully and thoroughly read the following sections for proper handling during maintenance.

9.3 Calibration
Use the procedure below to check instrument operation and accuracy during periodic maintenance or troubleshooting. 1) Connect the instruments as shown in Figure 9.3.1 and warm up the instruments for at least five minutes.
IMPORTANT

IMPORTANT

As a rule, maintenance of this transmitter should be implemented in a maintenance service shop where the necessary tools are provided. The CPU assembly contains sensitive parts that may be damaged by static electricity. Exercise care so as not to directly touch the electronic parts or circuit patterns on the board, for example, by preventing static electrification by using grounded wrist straps when handling the assembly. Also take precautions such as placing a removed CPU assembly into a bag with an antistatic coating.

To adjust the transmitter for highest accuracy, make adjustments with the power supply voltage and load resistance including leadwire resistances set close to the conditions under which the transmitter is installed. If the measurement range 0% point is 0 kPa or shifted in the positive direction (suppressed zero), the reference pressure should be applied as shown in the figure. If the measurement range 0% point is shifted in the negative direction (elevated zero), the reference pressure should be applied using a vacuum pump. 2) Apply reference pressures of 0%, 50%, and 100% of the measurement range to the transmitter. Calculate the errors (differences between digital voltmeter readings and reference pressures) as the pressure is increased from 0% to 100% and is decreased from 100% to 0%, and confirm that the errors are within the required accuracy.

9-1

IM 1C21B1-01E

9. MAINTENANCE Table 9.2.1 Instruments Required for Calibration


Name Power supply Load resistor Voltmeter Yokogawa-recommended Instrument Model SDBT or SDBS distributor Model 2792 standard resistor [250 0.005%, 3 W] Load adjustment resistor [100 1%, 1 W] Model 2501 A digital multimeter Accuracy (10V DC range): (0.002% of rdg + 1 dgt) Model MT110, MT120 precision digital manometer 1) For 10 kPa class Accuracy: (0.015% of rdg + 0.015% of F.S.) . . . . . (0.2% of rdg + 0.1% of F.S.) . . . . . . . . . 2) For 130 kPa class Accuracy: 0.02% of rdg . . . . . . . . . . . . . . . . . . . . . . .......................... 5digits (0.2% of rdg + 0.1% of F.S.) . . . . . . . . . 3) For 700 kPa class Accuracy: (0.02% of rdg + 3digits) . . . . . . . . . . . . . ......................... 5 digits (0.2% of rdg + 0.1% of F.S.) . . . . . . . . . 4) For 3000 kPa class Accuracy: (0.02% of rdg + 10 digits) . . . . . . . . . . . (0.2% of rdg + 0.1% of F.S.) . . . . . . . . . 5) For 130 kPa abs class Accuracy: (0.03% of rdg + 6 digits) . . . . . . . . . . . . Remarks 4 to 20 mA DC signal

for 0 to 10 kPa for -10 to 0 kPa for 25 to 130 kPa for 0 to 25 kPa for -80 to 0 kPa for 100 to 700 kPa for 0 to 100 kPa for -80 to 0 kPa for 0 to 3000 kPa for -80 to 0 kPa for 0 to 130 kPa abs

Digital manometer

Select a manometer having a pressure range close to that of the transmitter.

Pressure generator

Model 2657 pneumatic pressure standard for 200 kPa {2 kgf/cm2}, 25 kPa {2500 mmH2O} Requires air pressure Accuracy: 0.05% of F.S. or 0.1% setting (whichever is greater) supply. Dead weight gauge tester 25 kPa {2500mmH2O} Accuracy: 0.03% of setting Model 6919 pressure regulator (pressure pump) Pressure range: 0 to 133 kPa {1000 mmHg} Select the one having a pressure range close to that of the transmitter. Prepare the vacuum pump for negative pressure ranges.
T0901.EPS

Pressure source

Note: The above table contains the instruments capable of performing calibration to the 0.2% level. Since special maintenance and management procedures involving traceability of each instrument to higher-level standards are required for calibration to the 0.1% level, there are difficulties in calibration to this level in the field. For calibration to the 0.1% level, contact Yokogawa representatives from which the instrument was purchased or the nearest Yokogawa office.

Using pressure generator P Low pressure side open to atmosphere

Supply pressure

Pressure generator

Using pressure source with manometer P

Pressure source Model MT110, MT120 precision digital manometer Reference pressure High pressure side

Reference pressure Low pressure side open to atmosphere High pressure side

Load resistance, 250

Power supply E

Load resistance, 250

Power supply E

R Load Rc adjusting V resistance, 100 Digital voltmeter

R Load Rc adjusting V resistance, 100 Digital voltmeter

F0901.EPS

Figure 9.3.1 Instrument Connections

9-2

IM 1C21B1-01E

9. MAINTENANCE

9.4 Disassembly and Reassembly


This section describes procedures for disassembly and reassembly for maintenance and component replacement. Always turn OFF power and shut off and release pressures before disassembly. Use proper tools for all operations. Table 9.4.1 shows the tools required.
Table 9.4.1 Tools for Disassembly and Reassembly
Tool Phillips screwdriver Slotted screwdriver Allen wrenches Quantity 1 1 2 JIS B4648 One each, nominal 3 and 5 mm Allen wrenches Width across flats, 17 mm Remarks JIS B4633, No. 2

9.4.1 Replacing the Integral Indicator

CAUTION

Cautions for Flameproof Type Transmitters Users are prohibited by law from modifying the construction of a flameproof type transmitter. This would invalidate the agency approval and the transmitters use in such rated area. Thus the user is prohibited from using a flameproof type transmitter with its integral indicator removed, or from adding an integral indicator to a transmitter. If such modification is absolutely required, contact Yokogawa. This subsection describes the procedure for replacing an integral indicator. (See Figure 9.4.1) Removing the Integral Indicator 1) Remove the cover. 2) Supporting the integral indicator by hand, loosen its two mounting screws. 3) Dismount the LCD board assembly from the CPU assembly. When doing this, carefully pull the LCD board assembly straight forward so as not to damage the connector pins between it and the CPU assembly. Attaching the Integral Indicator 1) Align both the LCD board assembly and CPU assembly connectors and engage them. 2) Insert and tighten the two mounting screws. 3) Replace the cover.
Output terminal cable

Wrench Torque wrench Adjustable wrench Socket wrench Socket driver Tweezers

1 1 1 1 1 1

Width across flats, 16 mm Width across flats, 5.5 mm


T0902.EPS

CAUTION
Precautions for CENELEC and SAA Flameproof Type Transmitters Flameproof type transmitters must be, as a rule, removed to a non-hazardous area for maintenance and be disassembled and reassembled to the original state. On the flameproof type transmitters the two covers are locked, each by an Allen head bolt (shrouding bolt). When a shrouding bolt is driven clockwise by an Allen wrench, it is going in and cover lock is released, and then the cover can be opened. When a cover is closed it should be locked by a shrouding bolt without fail. Tighten the shrouding bolt to a torque of 0.7 Nm.

Press forward LCD board assembly Integral indicator

Shrouding Bolt
Cover

Flat cable CPU assembly Zero-adjustment Bracket screw pin (for zero-adjustment screw pin) Mounting screw
F0903.EPS

Boss

Shrouding Bolt
Figure 9.4 Shrouding Bolts

Figure 9.4.1 Removing and Attaching LCD Board Assembly and CPU Assembly

9-3

IM 1C21B1-01E

9. MAINTENANCE

9.4.2 Replacing the CPU Board Assembly


This subsection describes the procedure for replacing the CPU assembly. (See Figure 9.4.1) Removing the CPU Assembly 1) Remove the cover. If an integral indicator is mounted, refer to Subsection 9.4.1 and remove the indicator. 2) Turn the zero-adjustment screw to the position (where the screw head slot is horizontal) as shown in Figure 9.4.1. 3) Disconnect the output terminal cable (cable with brown connector at the end). When doing this, lightly press the side of the CPU assembly connector and pull the cable connector to disengage. 4) Use a socket driver (width across flats, 5.5mm) to loosen the two bosses. 5) Carefully pull the CPU assembly straight forward to remove it. 6) Disconnect the flat cable (cable with black connector at the end) that connects the CPU assembly and the capsule.

NOTE

Confirm that the zero-adjustment screw pin is placed properly in the groove on the bracket prior to tightening the two bosses. If it is not, the zero-adjustment mechanism will be damaged. 5) Replace the cover.

9.4.3 Cleaning and Replacing the Capsule Assembly


This subsection describes the procedures for cleaning and replacing the capsule assembly. (See Figure 9.4.2.)

CAUTION

NOTE
Be careful not to apply excessive force to the CPU assembly when removing it.

Mounting the CPU Assembly 1) Connect the flat cable (with black connector) between the CPU assembly and the capsule. 2) Connect the output terminal cable (with brown connector).
NOTE

Make certain that the cables are free of pinching between the case and the CPU assembly edge. 3) Align and engage the zero-adjustment screw pin with the groove on the bracket on the CPU assembly. Then insert the CPU board assembly straight onto the post in the amplifier case. 4) Tighten the two bosses. If the transmitter is equipped with an integral indicator, refer to Subsection 9.4.1 to mount the indicator.

Cautions for Flameproof Type Transmitters Users are prohibited by law from modifying the construction of a flameproof type transmitter. If you wish to replace the capsule assembly with one of a different measurement range, contact Yokogawa. The user is permitted, however, to replace a capsule assembly with another of the same measurement range. When doing so, be sure to observe the following. The replacement capsule assembly must have the same part number as the one being replaced. The section connecting the transmitter and capsule assembly is a critical element in preservation of flameproof performance, and must be checked to verify that it is free of dents, scratches, and other defects. After completing maintenance, be sure to securely tighten the Allen screws that fasten the transmitter section and pressure-detector section together.

9-4

IM 1C21B1-01E

9. MAINTENANCE

Removing the Capsule Assembly

Pressure-detector section Nut

IMPORTANT
Cover flange

Exercise care as follows when cleaning the capsule assembly. Handle the capsule assembly with care, and be especially careful not to damage or distort the diaphragms that contact the process fluid. Do not use a chlorinated or acidic solution for cleaning. Rinse thoroughly with clean water after cleaning. 1) Remove the CPU assembly as shown in Subsection 9.4.2. 2) Remove the two Allen screws that connect the transmitter section and pressure-detector section. 3) Separate the transmitter section and pressuredetector section. 4) Remove the nuts from the four flange bolts. 5) Hold the capsule assembly by hand and remove the cover flange. 6) Remove the capsule assembly. 7) Clean the capsule assembly or replace with a new one. Reassembling the Capsule Assembly 1) Insert the capsule assembly between the flange bolts, paying close attention to the relative positions of the H (high pressure side) and L (low pressure side) marks on the capsule assembly. Replace the two capsule gaskets with new gaskets. 2) Install the cover flange on the high pressure side, and use a torque wrench to tighten the four nuts uniformly to a torque shown below.
Model Torque(Nm) {kgfm} EJA110A EJA120A EJA130A 147 {15}

Allen screw

Capsule gasket Flange bolt

Transmitter section
F0904.EPS

Figure 9.4.2 Removing and Mounting the Pressuredetector Section

9.4.4 Replacing the Process Connector Gaskets


This subsection describes process connector gasket replacement. (See Figure 9.4.3.) (a) Loosen the two bolts, and remove the process connectors. (b) Replace the process connector gaskets. (c) Remount the process connectors. Tighten the bolts securely and uniformly to a torque of 39 to 49 Nm {4 to 5 kgfm}, and verify that there are no pressure leaks.
Bolt Process connector Process connector gasket

39{4}

3) After the pressure-detector section has been reassembled, a leak test must be performed to verify that there are no pressure leaks. 4) Reattach the transmitter section to the pressuredetector section. 5) Tighten the two Allen screws. (Tighten the screws to a torque of 5 Nm) 6) Install the CPU assembly according to Subsection 9.4.2. 7) After completing reassembly, adjust the zero point and recheck the parameters.

F0905.EPS

Figure 9.4.3 Removing and Mounting the Process Connector

9-5

IM 1C21B1-01E

9. MAINTENANCE

9.5 Troubleshooting
If any abnormality appears in the measured values, use the troubleshooting flow chart below to isolate and remedy the problem. Since some problems have complex causes, these flow charts may not identify all. If you have difficulty isolating or correcting a problem, contact Yokogawa service personnel.

9.5.2 Troubleshooting Flow Charts


The following sorts of symptoms indicate that transmitter may not be operating properly. Example : There is no output signal. Output signal does not change even though process variable is known to be varying. Output value is inconsistent with value inferred for process variable.

Connect BRAIN TERMINAL and check self-diagnostics.

9.5.1 Basic Troubleshooting


First determine whether the process variable is actually abnormal or a problem exists in the measurement system. If the problem is in the measurement system, isolate the problem and decide what corrective action to take. This transmitter is equipped with a self-diagnostic function which will be useful in troubleshooting; see Section 8.5 for information on using this function.
: Areas where self-diagnostic offers support

Does the self-diagnostic indicate problem location?

YES

NO

Refer to error message summary in Subsection 8.5.2 to take actions.

Is power supply polarity correct?

NO

YES
Abnormalities appear in measurement.

Refer to Section 6.3 to check/correct polarity at each terminal from power supply to the terminal box.

YES

Is process variable itself abnormal?

Are power supply voltage and load resistance correct?

NO

YES
Inspect the process system. NO Measurement system problem

Refer to Section 6.6 for rated voltage and load resistance.

Isolate problem in measurement system.

Are valves opened or closed correctly?

NO

YES

Fully close equalizing valve, and fully open high pressure and low pressure valves.

YES

Does problem exist in receiving instrument?

YES Is there any pressure leak?

Inspect receiver.

NO

NO

Fix pressure leaks, paying particular attention to connections for impulse piping,pressure-detector section, etc.

Environmental conditions

Transmitter itself

Check/correct environmental conditions. Operating conditions

Check transmitter.

Is there continuity through the transmitter loop wiring? Do the loop numbers match? YES

NO

Find/correct broken conductor or wiring error.

Check/correct operating conditions.


F0906.EPS

Contact Yokogawa service personnel.


F0907.EPS

Figure 9.5.1 Basic Flow and Self-Diagnostics

9-6

IM 1C21B1-01E

9. MAINTENANCE

Output travels beyond 0% or 100%.

Large output error.

Connect BRAIN TERMINAL and check self-diagnostics.

Connect BRAIN TERMINAL and check self-diagnostics.

Does the selfdiagnostic indicate problem location? NO

YES

Does the selfdiagnostic indicate problem location? NO

YES

Refer to error message summary in Subsection 8.5.2 to take actions.

Refer to error message summary in Subsection 8.5.2 to take actions.

Is power supply polarity correct?

NO

Are valves opened or closed correctly?

NO

YES

Refer to Section 6.3 to check/correct polarity at each terminal from power supply to the terminal box.

YES

Fully close equalizing valve, and fully open high pressure and low pressure valves.

Are valves opened or closed correctly?

NO

Is impulse piping connected correctly?

NO

YES

Fully close equalizing valve, and fully open high pressure and low pressure valves.

YES

Refer to individual model user manuals and connect piping as appropriate for the measurement purpose.

Is there any pressure leak?

YES

Are power supply voltage and load resistance correct? YES

NO

NO

Fix pressure leaks, paying particular attention to connections for impulse piping, pressure-detector section, etc. Is transmitter installed where there is marked variation in temperature?

Refer to Section 6.6 for rated voltage and load resistance.

Is impulse piping to high pressure and low pressure side correct?

YES

NO

YES

Refer to individual model user manuals and connect piping as appropriate for the measurement purpose.

NO

Provide lagging and/or cooling, or allow adequate ventilation.

Is zero point adjusted correctly?

NO

Were appropriate instruments used for calibration? YES

NO

YES

Adjust the zero point.

Refer to Section 9.2 when selecting instruments for calibration.

Contact Yokogawa service personnel.


F0908.EPS

Is output adjusted correctly?

NO

YES

Adjust the output.

Contact Yokogawa service personnel.


F0909.EPS

9-7

IM 1C21B1-01E

10. GENERAL SPECIFICATIONS

10. GENERAL SPECIFICATIONS


10.1 Standard Specifications
Refer to IM 1C22T2-01E for Fieldbus communication type marked with . Performance Specifications
See General Specifications sheet, GS 1C21B1-E, GS 1C21B3-E, and GS 1C21B4-E. Output : Two wire 4 to 20 mA DC output with digital communications, linear or square root programmable. BRAIN or HART FSK protocol are superimposed on the 4 to 20 mA signal. Failure Alarm: Output status at CPU failure and hardware error; Up-scale: 110%, 21.6 mA DC or more(standard) Down-scale: -5%, 3.2 mA DC
Note: Applicable for Output signal code D and E

Functional Specifications Span & Range Limits


EJA110A and EJA130A:
Measurement Span and Range

kPa 0.5 to 10

inH2O (/D1) 2 to 40

mbar (/D3) 5 to 100

mmH2O (/D4) 50 to 1000 1000 to 1000

Span L

Damping Time Constant (1st order): The sum of the amplifier and capsule damping time constant must be used for the overall time constant. Amp damping time constant is adjustable from 0.2 to 64 seconds.
Capsule (Silicone Oil) Time Constant (approx. sec) L 0.4 M 0.3 H 0.3 V 0.3 E 0.2
T1003.EPS

Range 10 to 10 40 to 40 100 to 100 Span 1 to 100 100 to 100 4 to 400 400 to 400

10 to 1000 100 to 10000 1000 to 1000 10000 to 10000 0.05 to 5 kgf/cm2 5 to 5 kgf/cm2 1.4 to 140 kgf/cm2 5 to 140 kgf/cm2
T1001.EPS

Range Span

5 to 500 20 to 2000 50 to 5000 500 to 500 0.14 to 14 MPa 0.5 to 14 MPa 2000 to 2000 20 to 2000 psi 71 to 2000 psi 5000 to 5000 1.4 to 140 bar 5 to 140 bar

H Range Span
V*

Ambient Temperature Limits: * Safety approval codes may affect limits. EJA110A, EJA130A 40 to 85C (40 to 185F) 30 to 80C (22 to 176F) with LCD Display EJA120A 25 to 80C (13 to 176F) Process Temperature Limits: * Safety approval codes may affect limits. EJA110A, EJA130A 40 to 120C (40 to 248F) EJA120A 25 to 80C (13 to 176F) Maximum Overpressure: See General Specifications sheet. Working Pressure Limits (Silicone Oil) Maximum Pressure Limit: Capsule Pressure L (EJA110A) 3.5 MPa {500 psig} M, H (EJA110A) 14 MPa {2000 psig} M, H (EJA130A) 32 MPa {4500 psig} E (EJA120A) 50 kPa {7.25 psig} Minimum Pressure Limit: Capsule Pressure L, M, H (EJA110A) See Figure 1. M, H (EJA130A) See Figure 1. E (EJA120A) 50 kPa {7.25 psig}

Range

* For Wetted parts material code other than S, the ranges are 0 to 14 MPa, 0 to 2000 psi, 0 to 140 bar, and 0 to 140 kgf/cm2.

EJA120A:
Measurement Span and Range

kPa 0.1 to 1 1 to 1

inH2O (/D1) 0.4 to 4 4 to 4

mbar (/D3) 1 to 10 10 to 10

mmH2O (/D4) 10 to 100 100 to 100


T1002.EPS

Span E Range

URL is define as the Upper Range Limit from the table above. Zero Adjustment Limits: Zero can be fully elevated or suppressed, within the Lower and Upper Range Limits of the capsule. External Zero Adjustment : External zero is continuously adjustable with 0.01% incremental resolution of span. Span may be adjusted locally using the digital indicator with range switch.

10-1

IM 1C21B1-01E

10. GENERAL SPECIFICATIONS

Installation
Supply & Load Requirements : * Safety approvals can affect electrical requirements. See Section 6.6, Power Supply Voltage and Load Resistance. EMC Conformity Standards: , For EMI (Emission): EN55011, AS/NZS 2064 1/2 For EMS (Immunity): EN50082-2 Communication Requirements : BRAIN Communication Distance; Up to 2 km (1.25 miles) when using CEV polyethylene-insulated PVC-sheathed cables. Communication distance varies depending on type of cable used. Load Capacitance; 0.22 F or less (see note) Load Inductance; 3.3 mH or less (see note) Input Impedance of communicating device; 10 k or more at 2.4 kHz.
Note: For general-use and Flameproof type. For Intrinsically safe type, please refer to Optional Specifications.

Physical Specifications
Wetted Parts Materials: Diaphragm, Cover flange, Process connector and Drain/Vent Plug; See Model and Suffix Codes Capsule Gasket; Teflon-coated SUS316L Process Connector Gasket; PTFE Teflon (EJA110A and EJA120A) Fluorinated Rubber (EJA110A and EJA120A with Optional code /N2 and /N3 and EJA130A with Process connection code 3 and 4) Glass reinforced Teflon (EJA130A with Process connection code 1 and 2) Non-wetted Parts Materials: Bolting; SCM435, SUS630, or SUH660 Housing; Low copper cast-aluminum alloy with polyurethane paint (Munsell 0.6GY3.1/2.0) Enclosure Classification; JIS C0920 immersion proof (equivalent to NEMA 4X and IEC IP67) Cover O-rings; Buna-N Data plate and tag; SUS304 Fill Fluid; Silicone or Fluorinated oil (optional) Weight: 3.9 kg (8.6 lbs) without mounting bracket or process connector (EJA110A) Connections: Refer to the Model and Suffix Codes to specify the process and electrical connection type. < Settings When Shipped >
Tag Number Output Mode Display Mode Operation Mode As specified in order *1 Linear unless otherwise specified in order Linear unless otherwise specified in order Normal unless otherwise specified in order

HART Communication Distance; Up to 1.5 km (1 mile) when using multiple twisted pair cables. Communication distance varies depending on type of cable used. Use the following formula to determine cable length for specific applications: L= 65 x 106 (Cf + 10,000) C (R x C)

Where: L = length in meters or feet R = resistance in (including barrier resistance) C = cable capacitance in pF/m or pF/ft Cf = maximum shunt capacitance of receiving devices in pF/m or pF/ft

100{14.5} Working pressure kPa abs {psi abs} Applicable range 10{1.4}

Atmospheric pressure

Damping Time 2 sec. Constant *2 Calibration Range Lower Range Value As specified in order Calibration Range As specified in order Higher Range Value Selected from mmH2O, mmAq, mmWG, Calibration Range mmHg, Pa, hPa, kPa, MPa, mbar, bar, gf/cm2, kgf/cm2, inH2O, inHg, ftH2O, or psi. Units (Only one unit can be specified)
T1004.EPS

*1: *2:

Up to 16 alphanumeric characters (including - and ) will be entered in the amplifier memory. If using square root output, set damping time constant to 2 sec. or more.

2.7{0.38} 1{0.14} -40 (-40)

0 (32)

40 (104)

80 (176)

120 (248)
F1001.EPS

Process temperature C (F)

Figure 1. Working Pressure and Process Temperature

10-2

IM 1C21B1-01E

10. GENERAL SPECIFICATIONS

10.2 Model and Suffix Codes


Model EJA110A
Model EJA110A Output Signal Suffix Codes -D -E (Note 1) -F (Note 5) L M H V S H M T A D B 0 1 2 3 4 5 Differential pressure transmitter 4 to 20 mA DC with digital communication (BRAIN protocol) 4 to 20 mA DC with digital communication (HART protocol) Digital communication (FOUNDATION Fieldbus protocol) 0.5 to 10 kPa {50 to 1000 mmH2O} 1 to 100 kPa {100 to 10000 mmH2O} 5 to 500 kPa {0.05 to 5 kgf/cm2} 0.14 to 14 MPa {1.4 to 140 kgf/cm2} [Body] (Note 4) SCS14A SCS14A SCS14A SCS14A Hastelloy C-276 equiv. (Note 6) Hastelloy C-276 equiv. (Note 6) Monel equivalent (Note 7) [Capsule] SUS316L Hastelloy C-276 (Note 3) Monel (Note 3) Tantalum (Note 3) Hastelloy C-276 (Note 3) Tantalum (Note 3) Monel
(Note 2)

Description

Measurement span(capsule)

Wetted parts material

[Vent plug] SUS316 SUS316 SUS316 SUS316 Hastelloy C-276 Hastelloy C-276 Monel

Process connections

without process connector (Rc1/4 female on the cover flanges) with Rc1/4 female process connector with Rc1/2 female process connector with 1/4 NPT female process connector with 1/2 NPT female process connector without process connector (1/4 NPT female on the cover flanges) [Maximum working pressure] (L capsule) (M and H capsule) 3.5 MPa {35 kgf/cm2} 14 MPa {140 kgf/cm2} 3.5 MPa {35 kgf/cm2} 14 MPa {140 kgf/cm2} 3.5 MPa {35 kgf/cm2} 14 MPa {140 kgf/cm2}

Bolts and nuts material A SCM435 B SUS630 C SUH660 Installation -2 -3 -6 -7 -8 -9 0 2 3 4 5 7 8 9 D E N A B C D N

Vertical impulse piping type, right side high pressure, process connector upside Vertical impulse piping type, right side high pressure, process connector downside Vertical impulse piping type, left side high pressure, process connector upside Vertical impulse piping type, left side high pressure, process connector downside Horizontal impulse piping type, right side high pressure Horizontal impulse piping type, left side high pressure G1/2 female, one electrical connection 1/2 NPT female, two electrical connections without blind plug Pg 13.5 female, two electrical connections without blind plug M20 female, two electrical connections without blind plug G1/2 female, two electrical connections and a blind plug 1/2 NPT female, two electrical connections and a blind plug Pg 13.5 female, two electrical connections and a blind plug M20 female, two electrical connections and a blind plug Digital indicator Digital indicator with the range setting switch (None) SECC Carbon steel SUS304 SECC Carbon steel SUS304 (None) / Optional specification 2-inch pipe mounting (flat type) 2-inch pipe mounting (flat type) 2-inch pipe mounting (L type) 2-inch pipe mounting (L type)

Electrical connection

Integral indicator

Mounting bracket

Optional codes

Example: EJA110A-DMS5A-92NN/ Note 1: Refer to GS 1C22T1-E for HART Protocol version. Note 2: Diaphragm; Hastelloy C-276. Other wetted parts materials; SUS316L Note 3: Diaphragm and other wetted parts. Note 4: Body; Material of cover flanges and process connectors. Note 5: Refer to GS 1C22T2-E for Fieldbus communication. Note 6: Indicated material is equivalent to ASTM CW-12MW. Note 7: Indicated material is equivalent to ASTM M35-2.

T1005.EPS

10-3

IM 1C21B1-01E

10. GENERAL SPECIFICATIONS

Model EJA120A
Model EJA120A Output Signal Description Suffix Codes . . . . . . . . . . . . . . . . . . . . Differential pressure transmitter (for draft application) -D . . . . . . . . . . . . . . . . . . 4 to 20 mA DC with digital communication (BRAIN protocol) -E . . . . . . . . . . . (Note 1) 4 to 20 mA DC with digital communication (HART protocol) -F . . . . . . . . . . . (Note 4) Digital communication (FOUNDATION Fieldbus protocol) E . . . . . . . . . . . . . . . . . 0.1 to 1 kPa {10 to 100 mmH2O} [Body] (Note 3) SCS14A [Capsule] SUS316L (Note 2) [Vent plug] SUS316

Measurement span (capsule) Wetted parts material

S................ 0............. Process connections 1............. 2............. 3............. 4............. 5............. Bolts and nuts material

without process connector (Rc1/4 female on the cover flanges) with Rc1/4 female process connector with Rc1/2 female process connector with 1/4 NPT female process connector with 1/2 NPT female process connector without process connector (1/4 NPT female on the cover flanges)

Installation

Electrical connection

Integral indicator

Mounting bracket

Optional codes

[Maximum working pressure] 50 kPa {0.5 kgf/cm2} A . . . . . . . . . . . . SCM435 50 kPa {0.5 kgf/cm2} B . . . . . . . . . . . . SUS630 50 kPa {0.5 kgf/cm2} C . . . . . . . . . . . . SUH660 -2 . . . . . . . . . . Vertical impulse piping type, right side high pressure, process connector upside -3 . . . . . . . . . . Vertical impulse piping type, right side high pressure, process connector downside -6 . . . . . . . . . . Vertical impulse piping type, left side high pressure, process connector upside -7 . . . . . . . . . . Vertical impulse piping type, left side high pressure, process connector downside -8 . . . . . . . . . . Horizontal impulse piping type, right side high pressure -9 . . . . . . . . . . Horizontal impulse piping type, left side high pressure 0 . . . . . . . . . . G1/2 female, one electrical connection 2 . . . . . . . . . . 1/2 NPT female, two electrical connections without blind plug 3 . . . . . . . . . . Pg 13.5 female, two electrical connections without blind plug 4 . . . . . . . . . . M20 female, two electrical connections without blind plug 5 . . . . . . . . . . G1/2 female, two electrical connections and a blind plug 7 . . . . . . . . . . 1/2 NPT female, two electrical connections and a blind plug 8 . . . . . . . . . . Pg 13.5 female, two electrical connections and a blind plug 9 . . . . . . . . . . M20 female, two electrical connections and a blind plug D . . . . . . . . Digital indicator E . . . . . . . . Digital indicator with the range setting switch N . . . . . . . . (None) 2-inch pipe mounting (flat type) A . . . . . . . SECC Carbon steel 2-inch pipe mounting (flat type) B . . . . . . . SUS304 2-inch pipe mounting (L type) C . . . . . . . SECC Carbon steel 2-inch pipe mounting (L type) D . . . . . . . SUS304 N . . . . . . . (None) / Optional specification
Example: EJA120A-DES5A-92NN/ Note 1: Refer to GS 1C22T1-E for HART Protocol version. Note 2: Diaphragm; Hastelloy C-276. Other wetted parts materials; SUS316L. Note 3: Body; Material of cover flanges and process connectors. Note 4: Refer to GS 1C22T2-E for Fieldbus communication.
T1006.EPS

10-4

IM 1C21B1-01E

10. GENERAL SPECIFICATIONS

Model EJA130A
Model EJA130A Output Signal Suffix Codes ......................... -D . . . . . . . . . . . . . . . . . . . . . . . -E . . . . . . . . . . . . . . . . (Note 1) -F . . . . . . . . . . . . . . . . (Note 4) M...................... H...................... Description Differential pressure transmitter 4 to 20 mA DC with digital communication (BRAIN protocol) 4 to 20 mA DC with digital communication (HART protocol) Digital communication (FOUNDATION Fieldbus protocol) 1 to 100 kPa {100 to 10000 mmH2O} 5 to 500 kPa {0.05 to 5 kgf/cm2} [Body] (Note 3) [Capsule] [Vent plug] SUS316 SUS316L (Note 2) SUS316 without process connector (Rc1/4 female on the cover flange) with Rc1/4 female process connector with Rc1/2 female process connector with 1/4 NPT female process connector (Note 5) with 1/2 NPT female process connector (Note 5) without process connector (1/4 NPT female on the cover flanges) [Maximum working pressure] SCM435 32 MPa {320 kgf/cm2} SUS630 32 MPa {320 kgf/cm2} SUH660 32 MPa {320 kgf/cm2} Vertical impulse piping type, right side high pressure, process connector upside Vertical impulse piping type, right side high pressure, process connector downside Vertical impulse piping type, left side high pressure, process connector upside Vertical impulse piping type, left side high pressure, process connector downside Horizontal impulse piping type, right side high pressure Horizontal impulse piping type, left side high pressure G1/2 female, one electrical connection 1/2 NPT female, two electrical connections without blind plug Pg 13.5 female, two electrical connections without blind plug M20 female, two electrical connections without blind plug G1/2 female, two electrical connections and a blind plug 1/2 NPT female, two electrical connections and a blind plug Pg 13.5 female, two electrical connections and a blind plug M20 female, two electrical connections and a blind plug Digital indicator Digital indicator with the range setting switch (None) SECC Carbon steel SUS304 SECC Carbon steel SUS304 (None) / Optional specification 2-inch pipe mounting (flat type) 2-inch pipe mounting (flat type) 2-inch pipe mounting (L type) 2-inch pipe mounting (L type)

Measurement span(capsule) Wetted parts material

S.................... 0................... Process connection 1 . . . . . . . . . . . . . . . . . . . 2................... 3................... 4................... 5................... Bolts and nuts material A . . . . . . . . . . . . . . . . . B................. C................. -2 . . . . . . . . . . . . . . . . Installation -3 . . . . . . . . . . . . . . . . -6 . . . . . . . . . . . . . . . . -7 . . . . . . . . . . . . . . . . -8 . . . . . . . . . . . . . . . . -9 . . . . . . . . . . . . . . . . 0............... Electrical connection 2............... 3............... 4............... 5............... 7............... 8............... 9............... Integral indicator D.............. E.............. N.............. A............ B............ C............ D............ N............

Mounting bracket

Optional codes

Example: EJA130A-DMS5A-92NN/ Note 1: Refer to GS 1C22T1-E for HART Protocol version. Note 2: Diaphragm; Hastelloy C-276. Other wetted parts materials; SUS316L. Note 3: Body; Material of cover flanges: SUS316, Process connectors: SCS14A. Note 4: Refer to GS 1C22T2-E for Fieldbus communication. Note 5: Lower limit of ambient and process temperature is 15C.

T1007.EPS

10-5

IM 1C21B1-01E

10. GENERAL SPECIFICATIONS

10.3 Optional Specifications


Item Description FM Explosionproof Approval Explosionproof for Class I, Division 1, Groups B, C and D Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G Hazardous (classified) locations, indoors and outdoors ( NEMA 4X ) Temperature class: T6 Amb. Temp.:40 to 60 C (40 to 140 F) Electrical connection: 1/2 NPT female FM Intrinsically safe Approval Intrinsically Safe for Class I, Division 1, Groups A, B, C & D, Class II, Division 1, Groups E, F & G and Class III, Division 1 Hazardous Locations. Nonincendive for Class I, Division 2, Groups A, B, C & D, Class II, Division. 2, Groups E, F & G, and Class III, Division 1 Hazardous Locations. Enclosure: NEMA 4X, Temp. Class: T4, Amb. Temp.: 40 to 60 C (40 to 140 F) Intrinsically Safe Apparatus Parameters [Groups A, B, C, D, E, F and G] Vmax=30 V, Imax=165 mA, Pmax=0.9 W, Ci=22.5 nF, Li=730 H [Groups C, D, E, F and G] Vmax=30 V, Imax=225 mA, Pmax=0.9 W, Ci=22.5 nF, Li=730 H Electrical connection: 1/2 NPT female Combined FF1and FS1 Electrical connection: 1/2 NPT female CENELEC (KEMA) Flameproof Approval EExd IIC T4, T5, T6 Amb. Temp.: T4 and T5; 40 to 80 C ( 40 to 176 F), T6; 40 to 75C( 40 to 176F) Max. process Temp.: T4; 120 C (248 F), T5; 100 C (212 F), T6; 85 C (185 F) Electrical connection: 1/2 NPT female, Pg 13.5 female and M20 female CENELEC (KEMA) CENELEC (KEMA) Intrinsically safe Approval EEx ia IIC T4, Amb. Temp.: 40 to 60 C(40 to 140 F) Ui=30 V, Ii=165 mA, Pi=0.9 W, Ci=22.5 nF, Li=730 H Electrical connection: 1/2 NPT female, Pg 13.5 female and M20 female Combined KF1, KS1 and Type N Approval KEMA Type N Approval Ex nA IIC T4, Amb. Temp.: 40 to 60 C(40 to 140 F) U=30 V, I=165 mA Electrical connection: 1/2 NPT female, Pg 13.5 female and M20 female CSA Explosionproof Approval Explosionproof for Class I, Division 1, Groups B, C and D Dustignitionproof for Class II/III, Division 1, Groups E, F and G Division2 SEALS NOT REQUIRED , Temp. Class : T4, T5, T6 Encl Type 4x Max. Process Temp.: T4; 120 C (248 F), T5; 100 C (212 F), T6; 85 C (185 F) Amb. Temp.:40 to 80 C (40 to 176 F) Electrical connection: 1/2 NPT female CSA Intrinsically safe Approval Class I, Groups A, B, C and D Class II and III, Groups E, F and G Encl Type 4x, Temp. Class: T4, Amb. Temp.: 40 to 60 C (40 to 140 F) Vmax=30 V, Imax=165 mA, Pmax=0.9 W, Ci=22.5 nF, Li=730 H Electrical connection: 1/2 NPT female Combined CF1 and CS1 Electrical connection: 1/2 NPT female SAA Flameproof, Intrinsically safe and Non-sparking Approval Ex d IIC T4/T5/T6, IP67 class I, Zone 1, Amb. Temp. : 40 to 80 C (40 to 176 F) Max. Process Temp.: T4; 120 C (248 F), T5; 100 C (212 F), T6; 85 C (185 F) Ex ia IIC T4, IP67 class I, Zone 0 Ex n IIC T4, IP67 class I, Zone 2 Ui=30 V DC, Ii=165 mA DC, Wi=0.9 W, Amb. Temp.: 40 to 60 C (40 to 140 F) Electrical connection: 1/2 NPT female, Pg 13.5 female and M20 female Code

FF1

Factory Mutual (FM)

FS1

FU1

KF1

KS1

KU1

CF1

Canadian Standards Association (CSA)

CS1

CU1

Standards Association of Australia (SAA)

SU1

T1008.EPS

10-6

IM 1C21B1-01E

10. GENERAL SPECIFICATIONS Item Painting Color change Coating change Lightning protector Amplifier cover only Epoxy resin-baked coating Transmitter power supply voltag: 10.5 to 32 V DC (10.5 to 30 V DC for intrinsically safe type, 9 to 32 V DC for Fieldbus communication type.) Allowable current: Max. 6000 A (1340 s), Repeating 1000 A (1340 s) 100 times Degrease cleansing treatment Oil-prohibited use Degrease cleansing treatment and with fluorinated oilfilled capsule. Operating temperature 20 to 80 C Degrease cleansing treatment and dehydrating treatment Degrease cleansing treatment and dehydrating treatment with fluorinated oilfilled capsule. Operating temperature 20 to 80 C P calibration ( psi unit ) Calibration units bar calibration ( bar unit ) M calibration Sealing treatment to SUS630 nuts Long vent (Note 1) Fast response Failure alam down-scale (Note 2) Stainless steel amplifier housing (Note 3) Gold-plate Configuration (kgf/cm2 unit ) ( See Table for Span and Range Limits.) Description Code P X1 A K1 K2 K5 K6 D1 D3 D4 Y U F1 C1 E1 A1 R1 N1 N2 N3 N4 HAC M01 M11 T01 Nitrogen(N2) Gas (Note 10) Retention time: 10 minutes T02 T04 T09
T1009.EPS

Oil-prohibited use with dehydrating treatment

Sealant ( liquid silicone rubber ) is coated on surfaces of SUS630 nuts used for cover flange mounting. Total vent plug Length: 112 mm (standard, 32 mm) Updete time: 0.125 sec or less, see GS for response time Output status at CPU failure and hardware error. When combining with Optional code F1, output signal is 2.5%, 3.6 mA DC or less. Amplifier housing material: SCS14A stainless steel (equivalent to SUS316 cast stainless steel or ASTM CF-8M) Gold-plated diaphragm Custom software configuration (Applicable only for Model EJA110A) Without drain and vent plugs N1 and Process connection on both sides of cover flange with blind kidney flanges on back N1, N2, and Mill certificate for cover flange, diaphragm, capsule body, and blind kidney flange

Body option

Stainless steel tag plate High Accuracy type Mill Certificate

JIS SUS 304 stainless steel tag plate wired onto transmitter High Accuracy (Applicable only for Model EJA120A) Cover flange (Note 4) Cover flange, Process connector (Note 5) Test Pressure: 3.5 MPa{35 kgf/cm2} (Note 6)

Pressure test/Leak test Certificate

Test Pressure: 14 MPa{140 kgf/cm2} (Note 7) Test Pressure: 50 kPa{0.5 kgf/cm2} (Note 8) Test Pressure: 32 MPa{320 kgf/cm2} (Note 9)

Note 1: Applicable only for vertical impulse piping types (Installation Code 2, 3, 6 or 7). Long vent material is SUS316. Note 2: The hardware error indicates faulty amplifier or capsule. Standarard output status (without /C1) is up-scale of 110%, 21.6 mA DC or more. Note 3: Applicable only for electrical Connection code 2, 3, 4 and 7. Not applicable for optional Code P and X1. Note 4: Applicable for Process Connections Code 0 and 5. Note 5: Applicable for Process Connections Code 1, 2, 3 and 4. Note 6: Applicable for Capsule Code L of Model EJA110A. Note 7: Applicable for Capsule Code M and H of Model EJA110A. Note 8: Applicable for Capsule Code E of Model EJA120A. Note 9: Applicable for Capsule Code M and H of Model EJA130A. Note 10: Pure nitrogen gas is used for Oil-prohibited use (Option Code K1, K2, K5 and K6).

10-7

IM 1C21B1-01E

10. GENERAL SPECIFICATIONS

10.4 Dimensions
Model EJA110A and EJA120A Vertical Impulse Piping Type Process connector upside (INSTALLATION CODE 6) (For CODE 2, 3 or 7, refer to the notes below.) Unit: mm (approx. inch)
Process connection (Optional) External indicator conduit connection Blind plug (Optional)

97 (3.82)

259(10.20) 197 (7.76) 146 (5.75)


Process connectors

110 (4.33)
High pressure side

54 (2.13)

Low pressure side (Note 1)

72 (2.83)

46 (1.81)

234(9.21)

78 (3.07)

102 (4.02)

Terminal side

Internal indicator (Optional) Ground terminal Conduit connection Zero adjustment 9(note 3) (0.35) Vent/Drain plugs 2-inch pipe (O.D. 60.5 mm)
F1002.EPS

53 ( 2.09)

Mounting bracket (L-type, Optional)

Horizontal Impulse Piping Type (INSTALLATION CODE 9) (For CODE 8, refer to the notes below.)
110 (4.33) 94 (3.70)
External indicator conduit connection Blind plug (Optional)

72 (2.83)

Conduit connection Internal indicator (Optional)

9(note3) (0.35)

Zero adjustment

78 (3.07)

197 (7.76)

Terminal side Ground terminal

148 (5.83)
Low pressure side (Note 1) Vent plugs Drain plugs Mounting bracket (Flat-type, Optional)
F1003.EPS

162 (6.38)

High pressure side

124 (4.88)

46 (1.81)

Process connections

54 (2.13)
128 (5.04)

125 (4.92)
47 (1.85)
Process connector (Optional) 2-inch pipe (O.D. 60.5mm)

Note 1: When INSTALLATION CODE 2, 3 or 8 is selected, high and low pressure side on above figure are reversed. (i. e. High pressure side is on the left side.) Note 2: When INSTALLATION CODE 3 or 7 is selected, process connetion and mounting bracket on above figure are reversed. Note 3: 15 mm (0.59 inch) for right side high pressure. (for CODE 2, 3 or 8) 12 mm (0.47 inch) for EJA120A.

10-8

IM 1C21B1-01E

146 (5.75)

10. GENERAL SPECIFICATIONS

Model EJA130A Vertical Impulse Piping Type Process connector upside (INSTALLATION CODE 6) (For CODE 2, 3 or 7, refer to the notes below.) Unit: mm (approx. inch)

259(10.20) 200(7.87) 97(3.82) 146(5.75)


Process connection Internal indicator (Optional) High pressure side

132(5.20) 54 (2.13)
Low pressure side (Note 1)

94(3.70)

68(2.68)

279(10.98)

124(4.88)

Conduit connection Ground terminal Zero adjustment

Terminal side 9 (note3) (0.35) Vent plug Drain plug

53(2.09)

Mounting bracket (L-type)

2B pipe(60.5)
F1004.EPS

Horizontal Impulse Piping Type (INSTALLATION CODE 9) (For CODE 8, refer to the notes below)
110(4.33) 116(4.57) 94(3.70)
Conduit connection Zero adjustment
(note3)

9 (0.35)

78 (3.07)

162(6.38)

200(7.87)

146(5.75)

Vent plug

Ground terminal

High pressure side

Low pressure side (Note 1)

124(4.88)

68(2.68)

Process connection

Drain plug

54(2.13) 154(6.06)

169(6.65)
2B pipe(60.5)

47 (1.85)

Mounting bracket (L-type)


F1005.EPS

Note 1: When INSTALLATION CODE 2, 3 or 8 is selected, high and low pressure side on above figure are reversed. (i. e. High pressure side is on the right side.) Note 2: When INSTALLATION CODE 3 or 7 is selected, process connection and mounting bracket on above figure are reversed. Note 3: 9 mm (0.35 inch) for right side high pressure type. (CODE 2, 3 or 8).

10-9

IM 1C21B1-01E

192(7.56)

78 (3.07)

Customer Maintenance Parts List

DPharp EJA Series Transmitter Section

2 11 4 A 10 3 12 13 1 14 A 2

1 2

6 7-2 7-1 8 9 5

Item 1 2 3

Part No. Bellow F9341RA F9341RJ F9341JP Below F9341AA F9341AC F9341AE F9341AH F9341AJ

Qty 2 2 1

Description Cover Cast-aluminum alloy SCS14A stainless steel O-ring Case Assembly (Note 1) Cast-aluminum alloy for G1/2 Cast-aluminum alloy for G1/2 (two electrical connections) Cast-aluminum alloy for 1/2 NPT (two electrical connections) Cast-aluminum alloy for M20 (two electrical connections) Cast-aluminum alloy for Pg13.5 (two electrical connections)

4 5

F9341AR F9341KA Bellow F9300AG F9303JU F9341KL Below F9342BB F9342BH F9342AF F9342AM F9342BF F9342BG Y9406ZU Y9612YU Below F9340NW F9340NX G9330DP G9612EB Bellow F9341FM F9341FJ Below F9342BL F9342BM F9342MK F9300PB

1 4

SCS14A stainless steel for 1/2 NPT (two electrical connections) Name Plate Screw For cast-aluminum alloy case assembly For SCS14A stainless steel case assembly Tag Plate CPU Assembly For BRAIN protocol version For HART protocol version For BRAIN protocol version(Optional code /F1) For HART protocol version with write protection switch (Optional code /F1) For FOUNDATION Fieldbus protocol version For FOUNDATION Fieldbus protocol version with PID/LM function (Optional code /LC1) Cap Screw Screw Plug For Pg13.5 For M20 For G1/2 For 1/2 NPT Cover Assembly Cast-aluminum alloy SCS14A stainless steel LCD Board Assembly Without range-setting switch With range-setting switch Mounting Screw Label

6 7-1

1 1

7-2 8 9 10

1 2 2 1

11 12

1 1

For integral indicator

13 14

2 2

(Note 1) Applicable for BRAIN and HART protocol versions (Output signal code D and E). For FOUNDATION Fieldbus protocol version (Output signal code F), consult Yokogawa local office.

All Rights Reserved, Copyright 1997, Yokogawa Electric Corporation.

Subject to change without notice. Printed in Japan.

CMPL 1C21A1-02E
6th Edition: Sep. 2000(YK)

Yokogawa Electric Corporation

Blank Page

Customer Maintenance Parts List

Model EJA110A, EJA120A and EJA130A Differential Pressure Transmitter (Pressure-detector Section)

Horizontal Impulse Piping Type


15

18 19 16 1 41 3

111

41

5 101 6

121 21 122 22 For EJA130A 131 14 112

102 42 133

42 123 132

All Rights Reserved, Copyright 1999, Yokogawa Electric Corporation.

CMPL 1C21B0-01E
1st Edition: Oct. 1999(YK) 3rd Edition: Sep. 2000(YK)

Yokogawa Electric Corporation

Vertical Impulse Piping Type

14 3 13 1 3 12 1 1 41 111

4 1

8 10 1 9

21

22

18 20 17 15

For EJA130A 13 3 13 2 11 2 4 2

12 3 42

10 2

Sep. 2000 Subjeject to change without notice. Printed in Japan.

CMPL 1C21B0-01E

3
Item 1 2 3 Part No. F9300AJ Below F9340GA F9340GC F9340GE F9340GF F9300FD F9300FR Below F9340VA F9340VB F9340VC F9340VD F9340VP F9340VR F9340VN F9340VQ F9340VS F9340VT F9340VU F9340VV F9340TP F9340TR F9340TN 4-2 F9340TQ Below F9340UA F9340UC F9340UH F9340UJ Below F9340SA F9340SB F9340SE F9340SF F9340SS F9340ST Below D0114PB F9340SK F9270HG Below F9200CS D0114RZ F9340SL F9340SM F9340SW D0117MS Below F9340SC F9340SD F9340SG F9340SH F9340SU 9 F9340SV Below F9270HE F9340SJ D0117MR Below F9340AB F9340AC F9340AQ Below F9340AF F9340AG F9340AS Below F9275KL F9275KH F9340BQ Below F9300GB F9300GD F9340BS 2 2 Qty 1 1 2 Description Capsule Assembly (see Table 1, Table 2 and Table 3 on page 5) (Note 1) O-Ring Gasket (for EJA110A with Wetted Teflon-coated SUS316L Stainless Steel Teflon-coated SUS316L Stainless Steel (degreased) Parts Material code S) (for EJA110A with Wetted Parts Material PTFE Teflon PTFE Teflon (degreased) code H, M, T, A, D and B and EJA120A) Teflon-coated SUS316L Stainless Steel Teflon-coated SUS316L Stainless Steel (degreased) (for EJA130A) Cover Flange for EJA110A and EJA120A (Note 2) Rc 1/4 1/4 NPT Rc 1/4 1/4 NPT Rc 1/4 1/4 NPT Rc 1/4 1/4 NPT Rc 1/4 1/4 NPT Rc 1/4 1/4 NPT Rc 1/4 1/4 NPT Rc 1/4 SCS14A Stainless Steel (for EJA110A with Wetted Parts Material code S) For Horizontal Impulse Piping Type SCS14A Stainless Steel (for EJA110A with Wetted Parts Material code S) For Vertical Impulse Piping Type SCS14A Stainless Steel (for EJA110A with Wetted Parts Material code H, M and T and EJA120A) For Horizontal Impulse Piping Type SCS14A Stainless Steel (for EJA110A with Wetted Parts Material code H, M and T and EJA120A) For Vertical Impulse Piping Type Hastelloy C-276 equivalent (for EJA110A with Wetted Parts Material code A and D) For Horizontal Impulse Piping Type Hastelloy C-276 equivalent (for EJA110A with Wetted Parts Material code A and D) For Vertical Impulse Piping Type Monel equivalent (for EJA110A with Wetted Parts Material code B) For Horizontal Impulse Piping Type Monel equivalent (for EJA110A with Wetted Parts Material code B) For Vertical Impulse Piping Type

4-1

1/4 NPT Cover Flange for EJA130A (Note 2) Rc 1/4 SUS316 Stainless Steel (for Horizontal Impulse Piping Type) Rc 1/4 SUS316 Stainless Steel (for Vertical Impulse Piping Type) 1/4 NPT SUS316 Stainless Steel (for Horizontal Impulse Piping Type)

1/4 NPT SUS316 Stainless Steel (for Vertical Impulse Piping Type) Vent Plug R 1/4 SUS316 Stainless Steel (for EJA110A with Wetted Parts 1/4 NPT Material code S, H, M and T, EJA120A and EJA130A) R 1/4 Hastelloy C-276 (for EJA110A with Wetted Parts Material code A and D) 1/4 NPT R 1/4 Monel (for EJA110A with Wetted Parts Material code B) 1/4 NPT Vent Screw SUS316 Stainless Steel (for Models except EJA110A with Wetted Parts Material codes A, D and B) Hastelloy C-276 (for EJA110A with Wetted Parts Material code A and D) Monel (for EJA110A with Wetted Parts Material code B) Drain Plug (Note 2) R 1/4 SUS316 Stainless Steel (for EJA110A with Wetted Parts 1/4 NPT Material code S, H, M and T, EJA120A and EJA130A) R 1/4 Hastelloy C-276 (for EJA110A with Wetted Parts Material code A and D) 1/4 NPT R 1/4 Monel (for EJA110A with Wetted Parts Material code B) 1/4 NPT Drain/Vent Plug R 1/4 1/4 NPT R 1/4 1/4 NPT R 1/4 SUS316 Stainless Steel (for EJA110A with Wetted Parts Material code S, H, M and T, EJA120A and EJA130A) Hastelloy C-276 (for EJA110A with Wetted Parts Material code A and D)

Monel (for EJA110A with Wetted Parts Material code B) 1/4 NPT Drain/Vent Screw SUS316 Stainless Steel (for Models except EJA110A with Wetted Parts Material code A, D and B) Hastelloy C-276 (for EJA110A with Wetted Parts Material code A and D) Monel (for EJA110A with Wetted Parts Material code B) Bolt (for EJA110A and EJA120A) SCM435 Chrome Molybdenum Steel SUS630 Stainless Steel SUH660 Stainless Steel Bolt (for EJA130A) SCM435 Chrome Molybdenum Steel SUS630 Stainless Steel SUH660 Stainless Steel Nut (for EJA110A and EJA120A) SCM435 Chrome Molybdenum Steel SUS630 Stainless Steel SUH660 Stainless Steel Nut (for EJA130A) SCM435 Chrome Molybdenum Steel SUS630 Stainless Steel SUH660 Stainless Steel

10-1

10-2

11-1

11-2

Sep. 2000 Subjeject to change without notice. Printed in Japan.

CMPL 1C21B0-01E

Item 12-1 12-2

Part No. Below D0114RB U0102XC Below F9340GN F9340GP Below F9340GN F9340GP F9202FJ F9201HA Below F9340XY F9340XW F9340XZ F9340XX F9340WY F9340WW F9340WZ F9340WX F9340TY F9340TW F9340TZ F9340TX Below F9340XT F9340XS Below F9271FD F9271FC Below X0100MN F9273DZ F9340AZ Below F9270AY F9273CZ Below F9270AW F9300TJ F9300TA Below F9340EA F9340EB F9340EC D0117XL-A Below F9270AX F9300TN F9300TE Below F9340EF F9340EG F9340EM Below F9275EC F9275ED F9275EE

Qty 2 2

4
Description Gasket (for EJA110A and EJA120A) PTFE Teflon PTFE Teflon (degreased) Gasket (for EJA110A and EJA120A with Optional code /N2 and /N3) Fluorinated Rubber Fluorinated Rubber (degreased) Gasket (for EJA130A) Fluorinated Rubber Fluorinated Rubber (degreased) Glass Reinforced Teflon

12-3

For Process connection code 3 and 4

13-1

For Process connection code 1 and 2 Glass Reinforced Teflon (degreased) Process Connector (for EJA110A and EJA120A)(Note 2) Rc 1/4 Rc 1/2 SCS14A Stainless Steel (for EJA110A with Wetted 1/4 NPT Parts Material code S, H, M and T and EJA120A) 1/2 NPT Rc 1/4 Rc 1/2 1/4 NPT 1/2 NPT

Hastelloy C-276 equivalent (for EJA110A with Wetted Parts Material code A and D)

13-2

Rc 1/4 Rc 1/2 Monel equivalent (for EJA110A with Wetted Parts Material code B) 1/4 NPT 1/2 NPT Process Connector (for EJA130A with Process connection code 3 and 4)(Note 2) 1/4 NPT SCS14A Stainless Steel 1/2 NPT Process Connector (for EJA130A with Process connection code 1 and 2)(Note 2) Rc 1/4 SUS316 Stainless Steel Rc 1/2 Bolt SCM435 Chrome Molybdenum Steel SUS630 Stainless Steel SUH660 Stainless Steel Bolt S15C Carbon Steel SUS XM7 Stainless Steel Bracket Assembly (Flat type) SECC Carbon Steel SECC Carbon Steel (for Epoxy resin-baked coating) SUS304 Stainless Steel Bracket Assembly (L type) SECC Carbon Steel SECC Carbon Steel (for Epoxy resin-baked coating) SUS304 Stainless Steel U-Bolt/Nut Assembly, SUS304 Stainless Steel Bracket (Flat type) SECC Carbon Steel SECC Carbon Steel (for Epoxy resin-baked coating) SUS304 Stainless Steel Bracket (L type) SECC Carbon Steel SECC Carbon Steel (for Epoxy resin-baked coating) SUS304 Stainless Steel Vent Plug (degreased), SUS316 Stainless Steel R 1/4 1/4 NPT Needle Assembly (degreased), SUS316 Stainless Steel

13-3

14

15 16

4 1

17

18 19

1 1

20

21

22

(Note 1) In case of degrease cleansing treatment (Optional code/K1 or K5), consult YOKOGAWA local office. (However, see Table 1, Table 2 and Table 3 in case of Optional code/K2 or K6) (Note 2) In case of degrease cleansing treatment (Optional code/K1, K2, K5 or K6), consult YOKOGAWA local office.

Sep. 2000 Subjeject to change without notice. Printed in Japan.

CMPL 1C21B0-01E

Capsule Assembly Part Number


EJA110A
Table 1. Capsule Assembly Part Number (Item 1) For General-use type, Flameproof type and Intrinsically safe type Installation of Transmitter Horizontal Impulse Piping Type High Pressure Side Right Capsule Code L M H V L M H V L M H V L M H V S(*1) F9349AA F9349BA F9349CA F9349DA F9349AB F9349BB F9349CB F9349DB F9349AC F9349BC F9349CC F9349DC F9349AD F9349BD F9349CD F9349DD Wetted Parts Material Code S(*2) T, D H, A F9352AA F9349AE F9349AJ F9352BA F9349BE F9349BJ F9352CA F9349CE F9349CJ F9352DA F9349DJ F9349DE F9352AB F9349AF F9349AK F9352BB F9349BF F9349BK F9352CB F9349CF F9349CK F9352DB F9349DF F9349DK F9352AC F9349AG F9349AL F9352BC F9349BG F9349BL F9352CC F9349CG F9349CL F9352DC F9349DG F9349DL F9352AD F9349AH F9349AM F9352BD F9349BH F9349BM F9352CD F9349CH F9349CM F9352DD F9349DH F9349DM M, B F9349AN F9349BN F9349CN F9349DN F9349AP F9349BP F9349CP F9349DP F9349AQ F9349BQ F9349CQ F9349DQ F9349AR F9349BR F9349CR F9349DR

Left

Vertical Impulse Piping Type

Right

Left

EJA120A
Table 2. Capsule Assembly Part Number (Item 1) For General-use type, Flameproof type and Intrinsically safe type Installation of Transmitter Horizontal Impulse Piping Type Vertical Impulse Piping Type High Pressure Side Right Left Right Left Capsule Code E E E E Part No. (*1) F9349EA F9349EB F9349EC F9349ED Part No. (*2) F9352EA F9352EB F9352EC F9352ED

EJA130A
Table 3. Capsule Assembly Part Number (Item 1) For General-use type, Flameproof type and Intrinsically safe type Installation of Transmitter Horizontal Impulse Piping Type High Pressure Side Right Left Right Left Capsule Code M H M H M H M H Part No. (*1) F9359AA F9359BA F9359AB F9359BB F9359AC F9359BC F9359AD F9359BD Part No. (*2) F9359EA F9359FA F9359EB F9359FB F9359EC F9359FC F9359ED F9359FD

Vertical Impulse Piping Type

*1. Silicone oil filled capsule (Standard) *2. Fluorinated oil filled capsule (for oil-prohibited use: Optional code /K2 or K6)

Sep. 2000 Subjeject to change without notice. Printed in Japan.

CMPL 1C21B0-01E

Blank Page

REVISION RECORD
Title: Model EJA110A, EJA120A and EJA130A Differential Pressure Transmitter Manual No.: IM 1C21B1-01E
Edition 1st 2nd Date Jun. 1997 Mar. 1998 Page CONTENTS 1-1 5-3 6-1 11-1 11-3 11-7 2-9+ CMPL New publication Page 3 1 Add REVISION RECORD. Add NOTE notice for FOUNDATION Fieldbus and HART protcol versions. 5.1.1 Correct the mounting procedure for Direct-Mounting Type 3valve Manifold. 6.1 Add Item to the Wiring Precautions. 11.1 Add FOUNDATOIN Fieldbus protocol. 11.2 Add Output signal code F and Wetted parts material code A and D. 11.3 Add Optional code A1. Change the figure of terminal configuration. CMPL 1C21A1-02E 1st 2nd Page 2 Add Item 7-2. CMPL 1C21B1-01E 1st 2nd Page 3 Add Part No. to Item 3, 4, 5, 6, 7, 8, 9, and 13. Page 4 Add Wetted parts material code A and D. CMPL 1C21B3-01E 1st 2nd Page 3, 4 Add Optional code K5 and K6. CMPL 1C21B4-01E 1st 2nd Page 3, 4 Delete Optional code K5 and K6. 2.10 8.3.2(11) 11.1 11.2 Delete EMC Conformity Standards Tables. Correction made in BURN OUT figure. Add Capsule code V. Add Capsule code V. Add Wetted parts material code M. Add Electrical connection code 7, 8, and 9. Add Electrical connection code 7, 8, and 9. Add Electrical connection code 7, 8, and 9. CMPL 1C21A1-02E 2nd 3rd Page 2 Add Part No. to Item 3 (For PG13.5 and M20). Add Part No. to Item 10 (For 1/2NPT, PG13.5, and M20). CMPL 1C21B1-01E 2nd 3rd Page 4 Add Capsule code V and Wetted parts material code M to Table 1. CMPL 1C21B4-01E 2nd 3rd Page 3 Add Part No. to Item 13 (For Rc1/4 and Rc1/2). Add Part No. to Item 4, 5, 7, and 8 (For Rc1/4). Revised a book in a new format. (The location of contents and the associated page numbers may not coincide with the one in old editions.) 2.10 Add AS/NZS 2064 1/2 to EMI, EMC Conformity Standards. 8.3.1 Move Parameter Summary table to Chapter 8. 10.2 Add Wetted parts material code B. CMPL 1C21A1-02E 3rd 4th Change a format. CMPL 1C21B0-01E 1st Combine CMPL 1C21B1-01E, 1C21B3-01E, and 1C21B4-01E. 2.9.4b 8.3.1 9.4.3 10.1 10.2 Change contents of NOTE 1 and 4. Add footnote 2 and 3. Add table for tightning torque for cover flange bolts. Add calibration units of Pa and hPa. Add Bolts and nuts material code C. Add Bolts and nuts material code C. Add Bolts and nuts material code C and footnote 5. Add Amb. Temp. for T6: 40 to 75C under /KF1. Add Optional code /F1, /N1, /N2, /N3, /N4, and /R1 Revised Item

3rd

Sep. 1998

2-13 8-18 11-1 11-3

11-4 11-5 CMPL

4th

Oct. 1999

2-8 8-4 10-3 CMPL

5th

Sep. 2000

2-8 8-5 9-5 10-2 10-3 10-4 10-5 10-6 10-7

10.3

REVISION RECORD.EPS

IM 1C21B1-01E

Edition 5th (Continued)

Date Sep. 2000

Page CMPL

Revised Item CMPL 1C21A1-02E 4th 5th(Manual Change) Add part numbers to 7-1 CPU Assembly. F9342AF and F9342AM CMPL 1C21A1-02E 5th 6th Add part numbers to 7-2 CPU Assembly. F9342BG Change part number of 7-1 CPU Assembly. F9342BC F9342BB Change part number of 10 Plug G9330DK G9330DP CMPL 1C21B0-01E 1st 2nd(Manual Change) Page 3 Add part number to 10-1 and 10-2 Bolt. F9340AQ and F9340AS Add part number to 11-1 and 11-2 Nut. F9340BQ and F9340BS CMPL 1C21B0-01E 2nd 3rd Page 4 Add part number to 14 Bolt. F9340AZ