Carbon Dioxide, DT, 8205

Carbon Dioxide
Digital Titrator Method Using Sodium Hydroxide 10 to 1000 mg/L CO2
Scope and Application: For water and seawater.

DOC316.53.01167 Method 8205 Digital Titrator

Test preparation
Before starting the test:
Avoid excess agitation when collecting and swirling the sample to prevent the loss of carbon dioxide from the sample. The sample is measured directly in the Erlenmeyer flask to avoid agitation.. For more accurate results, check the calibration of the Erlenmeyer flask. Fill a graduated cylinder with the sample volume of deionized water. Pour the water into the Erlenmeyer flask and make a mark at the correct level. Four drops of Phenolphthalein Indicator Solution1 can be substituted for the Phenolphthalein Indicator Powder Pillow. A pH meter can be used in place of the indicator. The end point pH is 8.3. For added convenience when stirring, use the TitraStir stirring apparatus1.
1

See Optional reagents and apparatus.

Collect the following items:


Description Phenolphthalein Indicator Powder Pillow Sodium hydroxide titration cartridge (see Range-specific information) Digital titrator Delivery tube for digital titrator Erlenmeyer flask, 250-mL See Consumables and replacement items for reorder information. Quantity 1 pillow 1 cartridge 1 1 1

Carbon dioxide

See Table 1

1. Select a sample volume and titration cartridge from the Rangespecific information table.

2. Insert a clean delivery tube into the titration cartridge. Attach the cartridge to the titrator.

3. Turn the delivery knob to eject air and a few drops of titrant. Reset the counter to zero and wipe the tip.

4. Collect a water sample directly into the titration flask by filling to the appropriate mark.

Carbon Dioxide Page 1 of 4

Carbon Dioxide Carbon dioxide (continued)

5. Add the contents of one Phenolphthalein Indicator Powder Pillow. Swirl to mix. If a pink color forms, no carbon dioxide is present.

6. Place the delivery tube into the solution and swirl the flask. Turn the knob on the titrator to add titrant to the solution. Continue to swirl the flask and add titrant until the color changes from colorless to a light pink color that persists for 30 seconds (pH 8.3). Write down the number of digits displayed on the counter.

7. Use the multiplier in the Range-specific information table to calculate the concentration: digits x multiplier = mg/L as CO2 Example: 100 mL of sample was titrated with the 0.3636 N cartridge and 250 digits were used to reach the endpoint. The concentration is 250 x 0.2 = 50 mg/L CO2

Table 66 Range-specific information


Range (mg/L as CO2) 1050 20100 100400 2001000 Sample volume (mL) 200 100 200 100 Titration cartridge (N NaOH) 0.3636 0.3636 3.636 3.636 Multiplier 0.1 0.2 1.0 2.0

Interferences
Interfering substances lists substances that can interfere with this test.

Table 67 Interfering substances


Interfering substance Other acids Color or turbidity Interference level Other acid components in the sample will be titrated and interfere directly in this test. Color or turbidity can mask the color change of the end point. Use a pH meter instead of the color indicators and titrate to a pH of 8.3.

Carbon Dioxide Page 2 of 4

Carbon Dioxide

Sample collection, preservation and storage


Collect samples in clean plastic or glass bottles. Fill completely and cap tightly. Prevent excessive agitation or prolonged exposure to air. Complete the test procedure as soon as possible after collection for best accuracy. The sample can be stored for at least 24 hours if cooled to 4 C (39 F) or below. Warm to room temperature before the test is started.

Accuracy check
Standard additions method (sample spike) Required for accuracy check: Carbon Dioxide Voluette Ampule Standard Solution, 10,000 mg/L CO2 Ampule breaker TenSette Pipet, 0.11.0 mL and Pipet Tips

1. Open the standard solution ampule. 2. Use the TenSette Pipet to add 0.1 mL of the standard to the titrated sample. Swirl to mix. 3. Titrate the spiked sample to the end point. Write down the amount of titrant that was used to reach the end point. 4. Use the TenSette Pipet to add 0.2 mL of standard to the titrated sample. Swirl to mix. 5. Titrate the spiked sample to the end point. Write down the amount of titrant that was used to reach the end point. 6. Use the TenSette Pipet to add 0.3 mL of standard to the titrated sample. Swirl to mix. 7. Titrate the spiked sample to the end point. Write down the amount of titrant that was used to reach the end point. 8. Each 0.1 mL of standard that was added will use approximately 50 digits of the 0.3636 N titration cartridge or 5 digits of the 3.636 N titration cartridge to reach the endpoint. If more or less titrant was used, there may be an interference (see Interferences) or the concentration of the titrant has changed.

Summary of method
Acidity due to carbon dioxide in a sample is titrated with sodium hydroxide to a phenolphthalein end point. Strong acids are assumed to be absent or of insignificant concentration.

Carbon Dioxide Page 3 of 4

Carbon Dioxide

Consumables and replacement items


Required reagents
Description Carbon Dioxide Reagent Set (approximately 100 tests) (1) Phenolphthalein Indicator Powder Pillows (1) Sodium Hydroxide Titration Cartridge, 0.3636 N (1) Sodium Hydroxide Titration Cartridge, 3.636 N 1 pillow varies varies 100/pkg each each Quantity/Test Unit Catalog number 2272700 94299 1437801 1438001

Required apparatus
Description Digital Titrator Flask, Erlenmeyer, graduated, 125-mL Flask, Erlenmeyer, graduated, 250-mL Quantity/Test 1 1 1 Unit each each each Catalog number 1690001 50543 50546

Recommended standards
Description Carbon Dioxide Standard Solution, Voluette Ampule, 10,000-mg/L as CO2, 10-mL Unit 16/pkg Catalog number 1427510

Optional reagents and apparatus


Description Phenolphthalein Indicator Solution, 5-g/L Potassium Acid Phthalate Standard Solution, 100-mg/L as CO2 Stir bar, octagonal 28.6 mm x 7.9 mm TenSette Pipet, 0.1 to 1.0 mL Tips for Tensette Water, deionized Cylinder, graduated, 250 mL Bottle, sampling, poly, 500 mL pH meter Thermometer, -10 to 220 C Ampule Breaker TitraStir Apparatus, 115 Vac TitraStir Apparatus, 230 Vac Digital Titrator delivery tips Unit 100 mL MDB 100 mL each each 50/pkg 500 mL each each each each each each each 5/pkg Catalog number 16232 226142 2095352 1970001 2185696 27249 108146 2087079 2635700 2196800 1940000 1940010 1720500

FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free 800-227-4224 Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web www.hach.com; E-mail techhelp@hach.com

HACH COMPANY WORLD HEADQUARTERS Telephone: (970) 669-3050 FAX: (970) 669-2932

Hach Company, 2007, 2010, 2012. All rights reserved. Printed in the U.S.A.

Edition 7