Magnetic resonance imaging


From Wikipedia, the free encyclopedia

(Redirected from MRI) Magnetic resonance imaging (MRI), nuclear magnetic resonance imaging (NMRI), or magnetic resonance tomography (MRT) is a medical imaging technique used in radiology to visualize internal structures of the body in detail. MRI makes use of the property of nuclear magnetic resonance (NMR) to image nuclei of atoms inside the body. MRI can create more detailed images of the human body than possible with Xrays. An MRI scanner is a device in which the patient lies within a large, powerful magnet where the magnetic field is used to align the magnetization of some atomic nuclei in the body, and radio frequency magnetic fields are applied to systematically alter the alignment of this magnetization.[1] This causes the nuclei to produce a rotating magnetic field detectable by the scannerand this information is recorded to construct an image of the scanned area of the body.[2]:36 Magnetic field gradients cause nuclei at different locations to precess at different speeds, which allows spatial information to be recovered using Fourier analysis of the measured signal. By using gradients in different directions, 2D images or 3D volumes can be obtained in any arbitrary orientation. MRI provides good contrast between the different soft tissues of the body, which makes it especially useful in imaging the brain, muscles, the heart, and cancers compared with other medical imaging techniques such as computed tomography (CT) or X-rays. Unlike CT scans or traditional X-rays, MRI does not use ionizing radiation.[3]

Sequential sections from an MRI of the brain, concurrently showing slices through transverse, sagittal, and coronal planes (left to right).

Para-sagittal MRI of the head, with aliasing artifacts (nose and forehead appear at the back of the head)

Contents
1 How MRI works 1.1 Magnetic field 1.2 Contrast agents and implants 1.3 Prepolarized MRI 2 History 2.1 2003 Nobel Prize 3 Applications 3.1 Basic MRI scans 3.1.1 T1-weighted MRI 3.1.2 T2-weighted MRI 3.1.3 T*2-weighted MRI 3.1.4 Spin density weighted MRI
en.wikipedia.org/wiki/MRI 1/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

3.2 Specialized MRI scans 3.2.1 Diffusion MRI 3.2.2 Magnetization transfer MRI 3.2.3 T1rho MRI 3.2.4 Fluid attenuated inversion recovery (FLAIR) 3.2.5 Magnetic resonance angiography 3.2.6 Magnetic resonance gated intracranial CSF dynamics (MR-GILD) 3.2.7 Magnetic resonance spectroscopy 3.2.8 Functional MRI 3.2.9 Real-time MRI 3.2.10 Interventional MRI 3.2.11 Radiation therapy simulation 3.2.12 Current density imaging 3.2.13 Magnetic resonance guided focused ultrasound 3.2.14 Multinuclear imaging 3.2.15 Susceptibility weighted imaging (SWI) 3.2.16 Other specialized MRI techniques 3.3 Molecular imaging of disease biomarkers by MRI 3.4 Portable instruments 3.5 MRI versus CT 3.6 Economics of MRI 4 Safety 4.1 Use when indicated 4.2 Magnetic field 4.3 Peripheral nerve stimulation (PNS) 4.4 Heating caused by absorption of radio waves 4.5 Acoustic noise 4.6 Cryogens 4.7 Contrast agents 4.8 Pregnancy 4.9 Claustrophobia and discomfort 4.10 Guidance 4.11 The European Physical Agents Directive 5 See also 6 References 7 Further reading 8 External links

How MRI works


Main article: Physics of magnetic resonance imaging

en.wikipedia.org/wiki/MRI

MRI machines make use of the fact that body tissue contains lots of water, and hence protons (1H nuclei), which 2/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

MRI machines make use of the fact that body tissue contains lots of water, and hence protons (1H nuclei), which get aligned in a large magnetic field.[4] Each water molecule has two hydrogen nuclei or protons. When a person is inside the powerful magnetic field of the scanner, the average magnetic moment of many protons becomes aligned with the direction of the field. A radio frequency current is briefly turned on, producing a varying electromagnetic field. This electromagnetic field has just the right frequency, known as the resonance frequency, to be absorbed and flip the spin of the protons in the magnetic field. After the electromagnetic field is turned off, the spins of the protons return to thermodynamic equilibrium and the bulk magnetization becomes realigned with the static magnetic field. During this relaxation, a radio frequency signal (electromagnetic radiation in the RF range) is generated, which can be measured with receiver coils. Information about the origin of the signal in 3D space can be learned by applying additional magnetic fields during the scan. These additional magnetic fields can be used to generate detectable signal only from Medical MRI scanner specific locations in the body (spatial excitation) and/or to make magnetization at different spatial locations precess at different frequencies, which enables k-space encoding of spatial information. The 3D images obtained in MRI can be rotated along arbitrary orientations and manipulated by the doctor to be better able to detect tiny changes of structures within the body.[5] These fields, generated by passing electric currents through gradient coils, make the magnetic field strength vary depending on the position within the magnet. Because this makes the frequency of the released radio signal also dependent on its origin in a predictable manner, the distribution of protons in the body can be mathematically recovered from the signal, typically by the use of the inverse Fourier transform. Protons in different tissues return to their equilibrium state at different relaxation rates. Different tissue variables, including spin density, T1 and T2 relaxation times, and flow and spectral shifts can be used to construct images.[6][7] By changing the settings on the scanner, this effect is used to create contrast between different types of body tissue or between other properties, as in fMRI and diffusion MRI. MRI is used to image every part of the body, and is particularly useful for tissues with many hydrogen nuclei and little density contrast, such as the brain, muscle, connective tissue and most tumors.

Magnetic field
MRI scans require a magnetic field with two properties, uniform field density and strength. The magnetic field cannot vary more than 1/10,000 of 1% and field strength ranges (depending on the scanner) from 0.2 to 3 teslas in strength in currently clinically used scanners, with research scanners investigating higher field strengths such as seven teslas. The lower field strengths can be achieved with permanent magnets, which are often used in "open" MRI scanners, for claustrophobic patients.[8] Higher field strengths can be achieved only with superconducting magnets. An MRI with a 3.0 tesla strength magnet may be referred to as a "3-T MRI" or "3-tesla MRI" Since the gradient coils are within the bore of the scanner, there are large forces between them and the main field coils, producing most of the noise that is heard during operation. Without efforts to damp this noise, it can approach 130 decibels (dB) with strong fields [9] (see also the subsection on acoustic noise).

Contrast agents and implants


en.wikipedia.org/wiki/MRI 3/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

MRI contrast agents may be injected intravenously to enhance the appearance of blood vessels, tumors or inflammation. Contrast agents may also be directly injected into a joint in the case of arthrograms: MRI images of joints. Unlike CT, MRI uses no ionizing radiation and is generally a very safe procedure. Nonetheless the strong magnetic fields and radio pulses can affect metal implants, including cochlear implants and cardiac pacemakers. There are many electronically activated devices that have approval from the US FDA to permit MRI procedures in patients under highly specific MRI conditions (see www.MRIsafety.com). In the case of cochlear implants, the US FDA has approved some implants for MRI compatibility. In the case of cardiac pacemakers, the results can sometimes be lethal,[10] so patients with such implants are generally not eligible for MRI.

Prepolarized MRI
In 2001, a research team at Stanford invented a new technique which came to be called "Prepolarized MRI" or PMRI.[11] The team demonstrated that the magnets do not have to be both uniform and strong, rather two magnets can be used together, where one is strong and the other one is uniform.[12] The first magnet in a PMRI scanner is strong, but not uniform. This magnet creates a very strong magnetic field which varies in uniformity by as much as 40%. This is the "prepolarize" component. A second much weaker (requiring only the electric power necessary to run two hairdryers) but far more precise magnet then creates a homogeneous magnetic field. These two magnets can be ordinary copper wound magnets, greatly lowering the cost of an MRI scanner.[12] Because the magnetic field is "tuned" by the second magnet, a PMRI scan can be obtained immediately adjacent to a metal prosthetic, unlike an MRI scan.[13]

History
In 1952, Herman Carr produced a one-dimensional MRI image as reported in his Harvard PhD thesis.[14][15][16] In the Soviet Union, Vladislav Ivanov filed (in 1960) a document with the USSR State Committee for Inventions and Discovery at Leningrad for a Magnetic Resonance Imaging device,[17] although this was not approved until the 1970s.[18] In a 1971 paper in the journal Science,[19] Raymond Damadian, an Armenian-American physician, scientist, and professor at the Downstate Medical Center State University of New York (SUNY), reported that tumors and normal tissue can be distinguished in vivo by nuclear magnetic resonance ("NMR"). He suggested that these differences could be used to diagnose cancer, though later research would find that these differences, while real, are too variable for diagnostic purposes. Damadian's initial methods were flawed for practical use,[20] relying on a Raymond Damadian's "Apparatus and point-by-point scan of the entire body and using relaxation rates, which method for detecting cancer in turned out not to be an effective indicator of cancerous tissue.[21] While tissue." researching the analytical properties of magnetic resonance, Damadian created the world's first magnetic resonance imaging machine in 1972. He filed the first patent for an MRI machine, U.S. patent #3,789,832 on March 17, 1972, which was later issued to him on February 5, 1974.[22]

en.wikipedia.org/wiki/MRI

4/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

The National Science Foundation notes, "The patent included the idea of using NMR to 'scan' the human body to locate cancerous tissue."[23] However, it did not describe a method for generating pictures from such a scan or precisely how such a scan might be done.[24][25] Meanwhile, Paul Lauterbur expanded on Carr's technique and developed a way to generate the first MRI images, in 2D and 3D, using gradients. In 1973, Lauterbur published the first nuclear magnetic resonance image[26][27] and the first cross-sectional image of a living mouse in January 1974.[28] In the late 1970s, Peter Mansfield, a physicist and professor at the University of Nottingham, England, developed a mathematical technique that would allow scans to take seconds rather than hours and produce clearer images than Lauterbur had. Damadian, along with Larry Minkoff and Michael Goldsmith, performed the first MRI body scan of a human being on July 3, 1977,[29][30] studies which they published in 1977.[31][32] and in 1979 Richard S. Likes filed patent *4,307,343 (http://patft.uspto.gov/netacgi/nph-Parser? Sect1=PTO1&Sect2=HITOFF&d=PALL&p=1&u=%2Fnetahtml%2FPTO%2Fsrchnum.htm&r=1&f=G&l=50& s1=4,307,343.PN.&OS=PN/4,307,343&RS=PN/4,307,343). In 1980 Paul Bottomley joined the GE Research Center in Schenectady NY, and his team ordered the highest field-strength magnet then available a 1.5T system and built the first high-field and overcame problems of coil design, RF penetration and signal-to-noise, to build the first whole-body MRI/MRS scanner.[33] The results translated into the highly successful 1.5T MRI product-line, with over 20,000 systems in use today. Bottomley performed the first localized MRS in human heart and brain. After starting a collaboration on heart applications with Robert Weiss at Johns Hopkins, Bottomley returned to the university in 1994, as Russell Morgan Professor and Director of the MR Research Division.[34] Although MRI is most commonly performed at 1.5T, higher fields such as 3T are gaining more popularity because of their increased sensitivity and resolution. In research laboratories, human studies have been performed at up to 9.4T[35] and animal studies have been performed at up to 21.1T[36]

2003 Nobel Prize


Reflecting the fundamental importance and applicability of MRI in medicine, Paul Lauterbur of the University of Illinois at Urbana-Champaign and Sir Peter Mansfield of the University of Nottingham were awarded the 2003 Nobel Prize in Physiology or Medicine for their "discoveries concerning magnetic resonance imaging". The Nobel citation acknowledged Lauterbur's insight of using magnetic field gradients to determine spatial localization, a discovery that allowed rapid acquisition of 2D images. Mansfield was credited with introducing the mathematical formalism and developing techniques for efficient gradient utilization and fast imaging. The actual research that won the prize was done almost 30 years before, while Paul Lauterbur was at Stony Brook University in New York. The award was vigorously protested by Raymond Vahan Damadian, founder of FONAR Corporation, who claimed that he invented the MRI, and that Lauterbur and Mansfield had merely refined the technology. A group called "The Friends of Raymond Damadian" (formed by Damadian's company, FONAR[37]), took out full-page advertisements in the New York Times and The Washington Post entitled "The Shameful Wrong That Must Be Righted", demanding that he be awarded at least a share of the Nobel Prize.[38]

Applications
In clinical practice, MRI is used to distinguish pathologic tissue (such as a brain tumor) from normal tissue. One advantage of an MRI scan is that it is harmless to the patient. It uses strong magnetic fields and non-ionizing electromagnetic fields in the radio frequency range, unlike CT scans and traditional X-rays, which both use ionizing radiation.

en.wikipedia.org/wiki/MRI

5/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

While CT provides good spatial resolution (the ability to distinguish two separate structures an arbitrarily small distance from each other), MRI provides comparable resolution with far better contrast resolution (the ability to distinguish the differences between two arbitrarily similar but not identical tissues).[39] The basis of this ability is the complex library of pulse sequences that the modern medical MRI scanner includes, each of which is optimized to provide image contrast based on the chemical sensitivity of MRI.[3] For example, with particular values of the echo time (TE) and the repetition time (TR), which are basic parameters of image acquisition, a sequence takes on the property of T2-weighting. On a T2-weighted scan, water- and fluid-containing tissues are bright (most modern T2 sequences are actually fast T2 sequences) and fat-containing tissues are dark. The reverse is true for T1-weighted images. Damaged tissue tends to develop edema, which makes a T2-weighted sequence sensitive for pathology, Effects of TR and TE on MR signal. and generally able to distinguish pathologic tissue from normal tissue. With the addition of an additional radio frequency pulse and additional manipulation of the magnetic gradients, a T2-weighted sequence can be converted to a FLAIR sequence, in which free water is now dark, but edematous tissues remain bright. This sequence in particular is currently the most sensitive way to evaluate the brain for demyelinating diseases, such as multiple sclerosis.[3] The typical MRI examination consists of 520 sequences, each of which is chosen to provide a particular type of information about the subject tissues. This information is then synthesized by the interpreting physician.[3]

Basic MRI scans


T1 -weighted MRI Main article: Spin-lattice relaxation time T1-weighted scans refer to a set of standard scans that depict differences in the spin-lattice (or T1) relaxation time of various tissues within the body. T1 weighted images can be acquired using either spin echo or gradient-echo sequences. T1-weighted contrast can be increased with the application of an inversion recovery RF pulse. Gradient-echo based T1-weighted sequences can be acquired very rapidly because of their ability to use short inter-pulse repetition times (TR). T1-weighted sequences are often collected before and after infusion of T1shortening MRI contrast agents. In the brain T1-weighted scans provide appreciable contrast between gray and white matter. In the body, T1 weighted scans work well for differentiating fat from waterwith water appearing darker and fat brighter.[40] T2 -weighted MRI Main article: Spin-spin relaxation time T2-weighted scans are another basic type. Like the T1-weighted scan, fat is differentiated from water, but in this case fat shows darker, and water lighter. For example, in the case of cerebral and spinal study, the CSF (cerebrospinal fluid) will be lighter in T2-weighted images. These scans are therefore particularly well suited to
en.wikipedia.org/wiki/MRI 6/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

imaging edema, with long TE and long TR. Because the spin echo sequence is less susceptible to inhomogeneities in the magnetic field, these images have long been a clinical workhorse. T2-weighted MRI T2 (pronounced "T 2 star") weighted scans use a gradient echo (GRE) sequence, with long TE and long TR. The gradient echo sequence used does not have the extra refocusing pulse used in spin echo so it is subject to additional losses above the normal T2 decay (referred to as T2), these taken together are called T2. This also makes it more prone to susceptibility losses at air/tissue boundaries, but can increase contrast for certain types of tissue, such as venous blood. Spin density weighted MRI Spin density, also called proton density, weighted scans try to have no contrast from either T2 or T1 decay, the only signal change coming from differences in the amount of available spins (hydrogen nuclei in water). It uses a spin echo or sometimes a gradient echo sequence, with short TE and long TR.
* * *

Specialized MRI scans


Diffusion MRI Main article: Diffusion MRI Diffusion MRI measures the diffusion of water molecules in biological tissues.[41] Clinically, diffusion MRI is useful for the diagnoses of conditions (e.g., stroke) or neurological disorders (e.g., Multiple Sclerosis), and helps better understand the connectivity of white matter axons in the central nervous system.[42] In an isotropic medium (inside a glass of water for example), water molecules naturally move randomly according to turbulence and Brownian motion. In biological tissues however, where the Reynolds number is low enough for flows to be laminar, the diffusion may be anisotropic. For example, a molecule inside the axon of a neuron has a low probability of crossing the myelin membrane. Therefore the molecule moves principally along the axis of the neural fiber. If it is known that molecules in a particular voxel diffuse principally in one direction, the assumption can be made that the majority of the fibers in this area are going parallel to that direction.

DTI image

The recent development of diffusion tensor imaging (DTI)[27] enables diffusion to be measured in multiple directions and the fractional anisotropy in each direction to be calculated for each voxel. This enables researchers to make brain maps of fiber directions to examine the connectivity of different regions in the brain (using tractography) or to examine areas of neural degeneration and demyelination in diseases like multiple sclerosis. Another application of diffusion MRI is diffusion-weighted imaging (DWI). Following an ischemic stroke, DWI is highly sensitive to the changes occurring in the lesion.[43] It is speculated that increases in restriction (barriers) to water diffusion, as a result of cytotoxic edema (cellular swelling), is responsible for the increase in signal on a DWI scan. The DWI enhancement appears within 510 minutes of the onset of stroke symptoms (as compared with
en.wikipedia.org/wiki/MRI 7/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

computed tomography, which often does not detect changes of acute infarct for up to 46 hours) and remains for up to two weeks. Coupled with imaging of cerebral perfusion (http://www.trauma.org/archive/neuro/cpp.html), researchers can highlight regions of "perfusion/diffusion mismatch" that may indicate regions capable of salvage by reperfusion therapy. Like many other specialized applications, this technique is usually coupled with a fast image acquisition sequence, such as echo planar imaging sequence. Magnetization transfer MRI Main article: Magnetization transfer Magnetization transfer (MT) refers to the transfer of longitudinal magnetization from free water protons to hydration water protons in NMR and MRI. In magnetic resonance imaging of molecular solutions, such as protein solutions, two types of water molecules, free (bulk) and hydration (bound), are found. Free water protons have faster average rotational frequency and hence less fixed water molecules that may cause local field inhomogeneity. Because of this uniformity, most free water protons have resonance frequency lying narrowly around the normal proton resonance frequency of 63 MHz (at 1.5 teslas). This also results in slower transverse magnetization dephasing and hence longer T2. Conversely, hydration water molecules are slowed down by interaction with solute molecules and hence create field inhomogeneities that lead to wider resonance frequency spectrum. In free liquids, protons, which may be viewed classically as small magnetic dipoles, exhibit translational and rotational motions. These moving dipoles disturb the surrounding magnetic field however on long enough timescales (which may be nanoseconds) the average field caused by the motion of protons is zero. This is known as "motional averaging" or narrowing and is characteristic of protons moving freely in liquid. On the other hand, protons bound to macromolecules, such as proteins, tend to have a fixed orientation and so the average magnetic field in close proximity to such structures does not average to zero. The result is a spatial pattern in the magnetic field that gives rise to a residual dipolar coupling (range of precession frequencies) for the protons experiencing the magnetic field. The wide frequency distribution appears as a broad spectrum that may be several kHz wide. The net signal from these protons disappears very quickly, in inverse proportion to the width, due to the loss of coherence of the spins, i.e. T2 relaxation. Due to exchange mechanisms, such as spin transfer or proton chemical exchange, the (incoherent) spins bound to the macromolecules continually switch places with (coherent) spins in the bulk media and establish a dynamic equilibrium. Magnetization transfer: Although there is no measurable signal from the bound spins, or the bound spins that exchange into the bulk media, their longitudinal magnetization is preserved and may recover only via the relatively slow process of T1 relaxation. If the longitudinal magnetization of just the bound spins can be altered, then the effect can be measured in the spins of the bulk media due to the exchange processes. The magnetization transfer sequence applies RF saturation at a frequency that is far off resonance for the narrow line of bulk water but still on resonance for the bound protons with a spectral linewidth of kHz. This causes saturation of the bound spins which exchange into the bulk water, resulting in a loss of longitudinal magnetization and hence signal decrease in the bulk water. This provides an indirect measure of macromolecular content in tissue. Implementation of magnetization transfer involves choosing suitable frequency offsets and pulse shapes to saturate the bound spins sufficiently strongly, within the safety limits of specific absorption rate for RF irradiation. T1rho MRI
en.wikipedia.org/wiki/MRI 8/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

T1 (T1rho): Molecules have a kinetic energy that is a function of the temperature and is expressed as translational and rotational motions, and by collisions between molecules. The moving dipoles disturb the magnetic field but are often extremely rapid so that the average effect over a long time-scale may be zero. However, depending on the time-scale, the interactions between the dipoles do not always average away. At the slowest extreme the interaction time is effectively infinite and occurs where there are large, stationary field disturbances (e.g. a metallic implant). In this case the loss of coherence is described as a "static dephasing". T2* is a measure of the loss of coherence in an ensemble of spins that include all interactions (including static dephasing). T2 is a measure of the loss of coherence that excludes static dephasing, using an RF pulse to reverse the slowest types of dipolar interaction. There is in fact a continuum of interaction time-scales in a given biological sample and the properties of the refocusing RF pulse can be tuned to refocus more than just static dephasing. In general, the rate of decay of an ensemble of spins is a function of the interaction times and also the power of the RF pulse. This type of decay, occurring under the influence of RF, is known as T1. It is similar to T2 decay but with some slower dipolar interactions refocused as well as the static interactions, hence T1T2.[44] Fluid attenuated inversion recovery (FLAIR) Main article: Fluid attenuated inversion recovery Fluid Attenuated Inversion Recovery (FLAIR)[45] is an inversion-recovery pulse sequence used to null signal from fluids. For example, it can be used in brain imaging to suppress cerebrospinal fluid (CSF) so as to bring out the periventricular hyperintense lesions, such as multiple sclerosis (MS) plaques. By carefully choosing the inversion time TI (the time between the inversion and excitation pulses), the signal from any particular tissue can be suppressed. Magnetic resonance angiography Main article: Magnetic resonance angiography Magnetic resonance angiography (MRA) generates pictures of the arteries to evaluate them for stenosis (abnormal narrowing) or aneurysms (vessel wall dilatations, at risk of rupture). MRA is often used to evaluate the arteries of the neck and brain, the thoracic and abdominal aorta, the renal arteries, and the legs (called a "run-off"). A variety of techniques can be used to generate the pictures, such as administration of a paramagnetic contrast agent (gadolinium) or using a technique known as "flow-related enhancement" (e.g. 2D and 3D time-of-flight sequences), where most of the signal on an image is due to blood that recently moved into that plane, see also FLASH MRI. Techniques involving phase accumulation (known as phase contrast angiography) can also be used to Magnetic resonance angiography generate flow velocity maps easily and accurately. Magnetic resonance venography (MRV) is a similar procedure that is used to image veins. In this method, the tissue is now excited inferiorly, while signal is gathered in the plane immediately superior to the excitation planethus imaging the venous blood that recently moved from the excited plane.[46] Magnetic resonance gated intracranial CSF dynamics (MR-GILD)

en.wikipedia.org/wiki/MRI

9/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

Magnetic resonance gated intracranial cerebrospinal fluid (CSF) or liquor dynamics (MR-GILD) technique is an MR sequence based on bipolar gradient pulse used to demonstrate CSF pulsatile flow in ventricles, cisterns, aqueduct of Sylvius and entire intracranial CSF pathway. It is a method for analyzing CSF circulatory system dynamics in patients with CSF obstructive lesions such as normal pressure hydrocephalus. It also allows visualization of both arterial and venous pulsatile blood flow in vessels without use of contrast agents.[47][48] Diastolic time data acquisition (DTDA). Systolic time data acquisition (STDA).

Magnetic resonance spectroscopy Main articles: In vivo magnetic resonance spectroscopy and Nuclear magnetic resonance spectroscopy Magnetic resonance spectroscopy (MRS) is used to measure the levels of different metabolites in body tissues. The MR signal produces a spectrum of resonances that correspond to different molecular arrangements of the isotope being "excited". This signature is used to diagnose certain metabolic disorders, especially those affecting the brain,[49] and to provide information on tumor metabolism.[50] Magnetic resonance spectroscopic imaging (MRSI) combines both spectroscopic and imaging methods to produce spatially localized spectra from within the sample or patient. The spatial resolution is much lower (limited by the available SNR), but the spectra in each voxel contains information about many metabolites. Because the available signal is used to encode spatial and spectral information, MRSI requires high SNR achievable only at higher field strengths (3 T and above). Functional MRI Main article: Functional magnetic resonance imaging Functional MRI (fMRI) measures signal changes in the brain that are due to changing neural activity. The brain is scanned at low resolution but at a rapid rate (typically once every 23 seconds). Increases in neural activity cause changes in the MR signal via T2 changes;[51] this mechanism is referred to as the BOLD (blood-oxygen-level dependent) effect. Increased neural activity causes an increased demand for oxygen, and the vascular system actually overcompensates for this, increasing the amount of oxygenated hemoglobin relative to deoxygenated hemoglobin. Because deoxygenated hemoglobin attenuates the MR signal, the vascular response leads to a signal increase that is related to the neural activity. The precise nature of the relationship between neural activity and the BOLD signal is a subject of current research. The BOLD effect also allows for the generation of high resolution 3D maps of the venous vasculature within neural tissue.
*

en.wikipedia.org/wiki/MRI

10/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

While BOLD signal is the most common method employed for neuroscience studies in human subjects, the flexible nature of MR imaging provides means to sensitize the signal to other aspects of the blood supply. Alternative techniques employ arterial spin labeling (ASL) or weight the MRI signal by cerebral blood flow (CBF) and cerebral blood volume (CBV). The CBV method requires injection of a class of MRI contrast agents that are now in human clinical trials. Because this method has been shown to be far more sensitive than the BOLD technique in preclinical studies, it may potentially expand the role of fMRI in clinical applications. The CBF method provides more quantitative information than the BOLD signal, albeit at a significant loss of detection sensitivity. Real-time MRI Main article: Real-time MRI Real-time MRI refers to the continuous monitoring ("filming") of moving objects in real time. While many different strategies have been developed over the past two decades, a recent development reported a real-time MRI technique based on radial FLASH and iterative reconstruction that yields a temporal resolution of 20 to 30 milliseconds for images with an in-plane resolution of 1.5 to 2.0 mm. The new method promises to add important information about diseases of the joints and the heart. In many cases MRI examinations may become easier and more comfortable for patients. Interventional MRI Main article: Interventional MRI

A fMRI scan showing regions of activation in orange, including the primary visual cortex (V1, BA17).

Real-time MRI of a human heart at a resolution of 50 ms

The lack of harmful effects on the patient and the operator make MRI well-suited for "interventional radiology", where the images produced by a MRI scanner are used to guide minimally invasive procedures. Of course, such procedures must be done without any ferromagnetic instruments. A specialized growing subset of interventional MRI is that of intraoperative MRI in which the MRI is used in the surgical process. Some specialized MRI systems have been developed that allow imaging concurrent with the surgical procedure. More typical, however, is that the surgical procedure is temporarily interrupted so that MR images can be acquired to verify the success of the procedure or guide subsequent surgical work. Radiation therapy simulation Because of MRI's superior imaging of soft tissues, it is now being utilized to specifically locate tumors within the body in preparation for radiation therapy treatments. For therapy simulation, a patient is placed in a specific, reproducible, body position and scanned. The MRI system then computes the precise location, shape and orientation of the tumor mass, correcting for any spatial distortion inherent in the system. The patient is then marked or tattooed with points that, when combined with the specific body position, permits precise triangulation for radiation therapy.[citation needed ]
en.wikipedia.org/wiki/MRI 11/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

Current density imaging Current density imaging (CDI) endeavors to use the phase information from images to reconstruct current densities within a subject. Current density imaging works because electrical currents generate magnetic fields, which in turn affect the phase of the magnetic dipoles during an imaging sequence.[citation needed ] These experiences have highlighted the importance of whole body magnetic resonance imaging in early diagnosis of pathologies, because when patients show manifestations through X-ray and CT, their prognosis already may be unfavorable or grim.Enero 2012 Magnetic resonance guided focused ultrasound In MRgFUS therapy, ultrasound beams are focused on a tissueguided and controlled using MR thermal imaging and due to the significant energy deposition at the focus, temperature within the tissue rises to more than 65 C (150 F), completely destroying it. This technology can achieve precise ablation of diseased tissue. MR imaging provides a three-dimensional view of the target tissue, allowing for precise focusing of ultrasound energy. The MR imaging provides quantitative, real-time, thermal images of the treated area. This allows the physician to ensure that the temperature generated during each cycle of ultrasound energy is sufficient to cause thermal ablation within the desired tissue and if not, to adapt the parameters to ensure effective treatment.[52] Multinuclear imaging Hydrogen is the most frequently imaged nucleus in MRI because it is present in biological tissues in great abundance, and because its high gyromagnetic ratio gives a strong signal. However, any nucleus with a net nuclear spin could potentially be imaged with MRI. Such nuclei include helium-3, lithium-7, carbon-13, fluorine-19, oxygen-17, sodium-23, phosphorus-31 and xenon-129. 23Na and 31P are naturally abundant in the body, so can be imaged directly. Gaseous isotopes such as 3He or 129Xe must be hyperpolarized and then inhaled as their nuclear density is too low to yield a useful signal under normal conditions. 17O and 19F can be administered in sufficient quantities in liquid form (e.g. 17O-water) that hyperpolarization is not a necessity.[citation needed ] Multinuclear imaging is primarily a research technique at present. However, potential applications include functional imaging and imaging of organs poorly seen on 1H MRI (e.g. lungs and bones) or as alternative contrast agents. Inhaled hyperpolarized 3He can be used to image the distribution of air spaces within the lungs. Injectable solutions containing 13C or stabilized bubbles of hyperpolarized 129Xe have been studied as contrast agents for angiography and perfusion imaging. 31P can potentially provide information on bone density and structure, as well as functional imaging of the brain. Multinuclear imaging holds the potential to chart the distribution of lithium in the human brain, this element finding use an important drug for those with conditions such as bipolar disorder.[citation needed ] Susceptibility weighted imaging (SWI) Main article: Susceptibility weighted imaging Susceptibility weighted imaging (SWI), is a new type of contrast in MRI different from spin density, T1, or T2 imaging. This method exploits the susceptibility differences between tissues and uses a fully velocity compensated, three dimensional, RF spoiled, high-resolution, 3D gradient echo scan. This special data acquisition and image processing produces an enhanced contrast magnitude image very sensitive to venous blood, hemorrhage and iron
en.wikipedia.org/wiki/MRI 12/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

storage. It is used to enhance the detection and diagnosis of tumors, vascular and neurovascular diseases (stroke and hemorrhage, multiple sclerosis, Alzheimer's), and also detects traumatic brain injuries that may not be diagnosed using other methods[53] Other specialized MRI techniques New methods and variants of existing methods are often published when they are able to produce better results in specific fields. Examples of these recent improvements are T2-weighted turbo spin-echo (T2 TSE MRI), double inversion recovery MRI (DIR-MRI) or phase-sensitive inversion recovery MRI (PSIR-MRI), all of them able to improve imaging of the brain lesions.[54][55] Another example is MP-RAGE (magnetization-prepared rapid acquisition with gradient echo),[56] which improves images of multiple sclerosis cortical lesions.[57]
*

Molecular imaging of disease biomarkers by MRI


Main article: Molecular imaging MRI has the advantages of having very high spatial resolution and is very adept at morphological imaging and functional imaging. MRI does have several disadvantages though. First, MRI has a sensitivity of around 103 mol/L to 105 mol/L which, compared to other types of imaging, can be very limiting. This problem stems from the fact that the difference between atoms in the high energy state and the low energy state is very small. For example, at 1.5 teslas, a typical field strength for clinical MRI, the difference between high and low energy states is approximately 9 molecules per 2 million. Improvements to increase MR sensitivity include increasing magnetic field strength, and hyperpolarization via optical pumping or dynamic nuclear polarization. There are also a variety of signal amplification schemes based on chemical exchange that increase sensitivity. To achieve molecular imaging of disease biomarkers using MRI, targeted MRI contrast agents with high specificity and high relaxivity (sensitivity) are required. To date, many studies have been devoted to developing targeted-MRI contrast agents to achieve molecular imaging by MRI. Commonly, peptides, antibodies, or small ligands, and small protein domains, such as HER-2 affibodies, have been applied to achieve targeting. To enhance the sensitivity of the contrast agents, these targeting moieties are usually linked to high payload MRI contrast agents or MRI contrast agents with high relaxivities.[58]

Portable instruments
Portable magnetic resonance instruments are available for use in education and field research. Using the principles of Earth's field NMR, they have no powerful polarizing magnet, so that such instruments can be small and inexpensive. Some can be used for both EFNMR spectroscopy and MRI imaging.[59] The low strength of the Earth's field results in poor signal to noise ratios (SNR), requiring long scan times to capture spectroscopic data or build up MRI images. However, the extremely low noise floor of SQUID-based MRI detectors (http://www.freshpatents.com/-dt20110901ptan20110210738.php), and the low density of thermal noise in the low-frequency operating range (tens of kiloHertz) may result in usable SNR approaching that of mid-field conventional instruments. Further, the ultra-low field technologies enable electron spin resonance detection, and potentially imaging, at safe operating frequencies (NASA Technical Brief (http://www.techbriefs.com/component/content/article/6598)).

en.wikipedia.org/wiki/MRI

13/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

Research with atomic magnetometers have discussed the possibility for cheap and portable MRI instruments without the large magnet.[60][61]

MRI versus CT
The use of X-rays, a type of ionizing radiation, by a computed tomography (CT) allows for examination of tissues composed of elements of a higher atomic number than the surrounding tissues. MRI, in contrast, uses non-ionizing radio frequency (RF) signals to acquire images and is best suited for soft tissue (although MRI can also be used to visualize bones, teeth[62] and even fossils[63]). Since CT scans use ionizing radiation (X-rays) to produce images, there is a risk of damage to DNA that can subsequently cause cancer. In 2007, it was estimated that 0.4% of current cancers in the United States are due to CTs performed in the past, although in the future this may increase to as high as 1.52% based on past rates of CT usage.[64] Unlike CT, MRI does not use ionizing radiation, although it is associated with other risks.[65] Contrast in CT images is generated purely by X-ray attenuation, while a variety of properties may be used to generate contrast in MR images. By variation of scanning parameters, tissue contrast can be altered to enhance different features in an image (see Applications for more details). Both CT and MR images may be enhanced by the use of contrast agents. Contrast agents for CT contain elements of a high atomic number relative to tissue, such as iodine or barium, while contrast agents (such as gadolinium and manganese) for MRI have paramagnetic properties that are used to alter tissue relaxation times. Commonly used MRI contrast agents may be contraindicated in people with significant permanent or transient kidney dysfunction.[66] CT and MRI scanners are able to generate multiple two-dimensional cross-sections (tomographs, or "slices") of tissue and three-dimensional reconstructions. MRI can generate cross-sectional images in any plane (including oblique planes). In the past, CT was limited to acquiring images in the axial plane (or near axial), and so the images were called Computed Axial Tomography scans (CAT scans). However, the development of multi-detector CT scanners with near-isotropic resolution allows the CT scanner to produce data that can be retrospectively reconstructed in any plane with minimal loss of image quality. For purposes of tumor detection and identification in the brain, MRI is generally superior.[67][68] However, in the case of solid tumors of the abdomen and chest, CT is often preferred as it suffers less from motion artifacts. Furthermore, CT usually is more widely available, faster, and less expensive. MRI is also best suited for cases when a patient is to undergo several exams in the short term, since it does not expose the patient to the hazards of ionizing radiation. However MRI is usually contraindicated if the patient has any type of medical implant, such as vagus nerve stimulators, implantable cardioverter-defibrillators, loop recorders, insulin pumps, cochlear implants, deep brain stimulators, metallic foreign bodies (e.g. shrapnel or shell fragments), or metallic implants such as surgical prostheses. These devices can malfunction or heat up during an MRI scan, so CT scans are considered the safer option for these patients.

Economics of MRI
Standard 1.5 tesla MRI scanners used to cost between US$1 million and US$1.5 million. Standard 3.0 tesla MRI scanners would often cost between US$2 million and US$2.3 million. Construction of MRI suites could cost up to US$500,000, or more, depending on project scope. PMRI scanners today now cost a great deal less, around US$50,000.[69]

en.wikipedia.org/wiki/MRI

14/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

MRI scanners have been significant sources of revenue for healthcare providers in the US. This is because of favorable reimbursement rates from insurers and federal government programs. Insurance reimbursement is provided in two components, an equipment charge for the actual performance of the MRI scan and professional charge for the radiologist's review of the images and/or data. In the US Northeast, an equipment charge might be $3,500 and a professional charge might be $350 [70] although the actual fees received by the equipment owner and interpreting physician are often significantly less and depend on the rates negotiated with insurance companies or determined by the Medicare Fee Schedule. For example, an orthopedic surgery group in Illinois billed a Looking through an MRI scanner. charge of $1,116 for a knee MRI in 2007 but the Medicare reimbursement in 2007 was only $470.91.[71] Many insurance companies require preapproval of an MRI procedure as a condition for coverage. In the US, the Deficit Reduction Act of 2005 significantly reduced reimbursement rates paid by federal insurance programs for the equipment component of many scans, shifting the economic landscape. Many private insurers have followed suit.[citation needed ] In the United States, an MRI of the brain with and without contrast billed to Medicare Part B entails, on average, a technical payment of US$403 and a separate payment to the radiologist of US$93.[72] In France, the cost of an MRI exam is approximately 150 Euros. This covers three basic scans including one with an intravenous contrast agent, as well as a consultation with the technician and a written report to the patient's physician.[citation needed ] In Japan, the cost of a MRI examination (excluding the cost of contrast material and films) ranges from US$155 to US$180, with an additional radiologist professional fee of US$17.[73] In India, the cost of an MRI examination including the fee for the radiologist's opinion comes to around Rs 30004000 (US$5060) excluding the cost of contrast material.

Safety
A number of features of MRI scanning can give rise to risks. These include: Powerful magnetic fields Radio waves Cryogenic liquids Noise Claustrophobia. In addition, in cases where MRI contrast agents are used, these also typically have associated risks.

Use when indicated


Medical societies issue guidelines for when physicians should use MRI on patients, and recommend against overuse. MRI can detect health problems or confirm a diagnosis, but often medical societies recommend that MRI not be the first procedure for creating a plan to diagnose or manage a patient's complaint. A common case is the
en.wikipedia.org/wiki/MRI 15/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

use of MRI to seek a cause of low back pain; the American College of Physicians, for example, recommends against this for being unlikely to result in a positive outcome for the patient.[74][75]

Magnetic field
Some types of medical implants are generally considered contraindications for MRI examinations, while others may be acceptable for patients under high specific MRI conditions. Patients are therefore always asked for complete information about all implants before entering the room for an MRI scan. Several deaths have been reported in patients with pacemakers who have undergone MRI scanning without appropriate precautions.[76] To reduce such risks, implants are increasingly being developed to make them able to be safely scanned,[77] and specialized protocols have been developed to permit the safe scanning of selected implants and pacing devices. Cardiovascular stents are considered safe, however.[78] Ferromagnetic foreign bodies such as shell fragments, or metallic implants such as surgical prostheses and ferromagnetic aneurysm clips are also potential risks. Interaction of the magnetic and radio frequency fields with such objects can lead to trauma due to movement of the object in the magnetic field or thermal injury from radio-frequency induction heating of the object.[79] Titanium and its alloys are safe from attraction and torque forces produced by the magnetic field, though there may be some risks associated with Lenz Effect forces acting on titanium implants in sensitive areas within the subject, such as Stapes implants in the inner ear. In the United States a classification system for implants and ancillary clinical devices has been developed by ASTM International and is now the standard supported by the US Food and Drug Administration:
MR-Unsafe sign

MR-Safe sign

MR-Conditional sign

MR-Safe The device or implant is completely non-magnetic, non-electrically conductive, and non-RF reactive, eliminating all of the primary potential threats during an MRI procedure. MR-Conditional A device or implant that may contain magnetic, electrically conductive or RF-reactive components that is safe for operations in proximity to the MRI, provided the conditions for safe operation are defined and observed (such as 'tested safe to 1.5 teslas' or 'safe in magnetic fields below 500 gauss in strength'). MR-Unsafe Nearly self-explanatory, this category is reserved for objects that are significantly ferromagnetic and pose a clear and direct threat to persons and equipment within the magnet room. The very high strength of the magnetic field can also cause "missile-effect" accidents, where ferromagnetic objects are attracted to the center of the magnet, and there have been incidences of injury and death.[80][81] To reduce the risks of projectile accidents, ferromagnetic objects and devices are typically prohibited in proximity to the MRI scanner and patients undergoing MRI examinations are required to remove all metallic objects, often by changing into a gown or scrubs and ferromagnetic detection devices are used by some sites.[82][83] There is no evidence for biological harm from even very powerful static magnetic fields.[84]

Peripheral nerve stimulation (PNS)


en.wikipedia.org/wiki/MRI 16/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

The rapid switching on and off of the magnetic field gradients is capable of causing nerve stimulation. Volunteers report a twitching sensation when exposed to rapidly switched fields, particularly in their extremities.[85][86] The reason the peripheral nerves are stimulated is that the changing field increases with distance from the center of the gradient coils (which more or less coincides with the center of the magnet).[87] Although PNS was not a problem for the slow, weak gradients used in the early days of MRI, the strong, rapidly switched gradients used in techniques such as EPI, fMRI, diffusion MRI, etc. are indeed capable of inducing PNS. American and European regulatory agencies insist that manufacturers stay below specified dB/dt limits (dB/dt is the change in field per unit time) or else prove that no PNS is induced for any imaging sequence. As a result of dB/dt limitation, commercial MRI systems cannot use the full rated power of their gradient amplifiers.

Heating caused by absorption of radio waves


Every MRI scanner has a powerful radio transmitter to generate the electromagnetic field which excites the spins. If the body absorbes the energy, heating occurs. For this reason, the rate at which energy is absorbed by the body has to be limited (see Specific absorption rate).

Acoustic noise
Switching of field gradients causes a change in the Lorentz force experienced by the gradient coils, producing minute expansions and contractions of the coil itself. As the switching is typically in the audible frequency range, the resulting vibration produces loud noises (clicking or beeping). This is most marked with high-field machines[88] and rapid-imaging techniques in which sound pressure levels can reach 120 dB(A) (equivalent to a jet engine at takeoff),[89] and therefore appropriate ear protection is essential for anyone inside the MRI scanner room during the examination.[90]

Cryogens
As described in Physics of Magnetic Resonance Imaging, many MRI scanners rely on cryogenic liquids to enable superconducting capabilities of the electromagnetic coils within. Though the cryogenic liquids used are non-toxic, their physical properties present specific hazards. An unintentional shut-down of a superconducting electromagnet, an event known as "quench", involves the rapid boiling of liquid helium from the device. If the rapidly expanding helium cannot be dissipated through an external vent, sometimes referred to as 'quench pipe', it may be released into the scanner room where it may cause displacement of the oxygen and present a risk of asphyxiation.[91] Oxygen deficiency monitors are usually used as a safety precaution. Liquid helium, the most commonly used cryogen in MRI, undergoes near explosive expansion as it changes from liquid to a gaseous state. The use of an oxygen monitor is important to ensure that oxygen levels are safe for patient/physicians. Rooms built in support of superconducting MRI equipment should be equipped with pressure relief mechanisms[92] and an exhaust fan, in addition to the required quench pipe. Because a quench results in rapid loss of cryogens from the magnet, recommissioning the magnet is expensive and time-consuming. Spontaneous quenches are uncommon, but a quench may also be triggered by an equipment malfunction, improper cryogen fill technique, contaminants inside the cryostat, or extreme magnetic or vibrational disturbances.[93][94]
en.wikipedia.org/wiki/MRI 17/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

Contrast agents
Main article: MRI contrast agent The most commonly used intravenous contrast agents are based on chelates of gadolinium. In general, these agents have proved safer than the iodinated contrast agents used in X-ray radiography or CT. Anaphylactoid reactions are rare, occurring in approx. 0.030.1%.[95] Of particular interest is the lower incidence of nephrotoxicity, compared with iodinated agents, when given at usual dosesthis has made contrast-enhanced MRI scanning an option for patients with renal impairment, who would otherwise not be able to undergo contrast-enhanced CT.[96] Although gadolinium agents have proved useful for patients with renal impairment, in patients with severe renal failure requiring dialysis there is a risk of a rare but serious illness, nephrogenic systemic fibrosis, that may be linked to the use of certain gadolinium-containing agents. The most frequently linked is gadodiamide, but other agents have been linked too.[97] Although a causal link has not been definitively established, current guidelines in the United States are that dialysis patients should only receive gadolinium agents where essential, and that dialysis should be performed as soon as possible after the scan to remove the agent from the body promptly.[98][99] In Europe, where more gadolinium-containing agents are available, a classification of agents according to potential risks has been released.[100][101] Recently a new contrast agent named gadoxetate, brand name Eovist (US) or Primovist (EU), was approved for diagnostic use: this has the theoretical benefit of a dual excretion path.[102]

Pregnancy
No effects of MRI on the fetus have been demonstrated.[103] In particular, MRI avoids the use of ionizing radiation, to which the fetus is particularly sensitive. However, as a precaution, current guidelines recommend that pregnant women undergo MRI only when essential. This is particularly the case during the first trimester of pregnancy, as organogenesis takes place during this period. The concerns in pregnancy are the same as for MRI in general, but the fetus may be more sensitive to the effectsparticularly to heating and to noise. However, one additional concern is the use of contrast agents; gadolinium compounds are known to cross the placenta and enter the fetal bloodstream, and it is recommended that their use be avoided. Despite these concerns, MRI is rapidly growing in importance as a way of diagnosing and monitoring congenital defects of the fetus because it can provide more diagnostic information than ultrasound and it lacks the ionizing radiation of CT. MRI without contrast agents is the imaging mode of choice for pre-surgical, in-utero diagnosis and evaluation of fetal tumors, primarily teratomas, facilitating open fetal surgery, other fetal interventions, and planning for procedures (such as the EXIT procedure) to safely deliver and treat babies whose defects would otherwise be fatal.

Claustrophobia and discomfort


Despite being painless, MRI scans can be unpleasant for those who are claustrophobic or otherwise uncomfortable with the imaging device surrounding them. Older closed bore MRI systems have a fairly long tube or tunnel. The part of the body being imaged must lie at the center of the magnet, which is at the absolute center of the tunnel. Because scan times on these older scanners may be long (occasionally up to 40 minutes for the entire procedure), people with even mild claustrophobia are sometimes unable to tolerate an MRI scan without management. Some modern scanners have larger bores (up to 70 cm) and scan times are shorter. This means that claustrophobia could be less of an issue, and additional patients may now find MRI to be a tolerable procedure.
en.wikipedia.org/wiki/MRI 18/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

Nervous patients may still find the following strategies helpful: Advance preparation visiting the scanner to see the room and practice lying on the table visualization techniques chemical sedation general anesthesia Coping while inside the scanner having a loved one in the room to hold hand, reassure them, etc. holding a "panic button" closing eyes as well as covering them (e.g. washcloth, eye mask) Many newer MRI systems place a diagonal mirror above the eyes to allow the patient to look down the tunnel rather than at the bore wall immediately above their face. Alternative scanner designs, such as open or upright systems, can also be helpful where these are available. Though open scanners have increased in popularity, they produce inferior scan quality because they operate at lower magnetic fields than closed scanners. However, commercial 1.5 tesla open systems have recently become available, providing much better image quality than previous lower field strength open models.[104] For babies and other young children chemical sedation or general anesthesia are the norm, as these subjects cannot be instructed to hold still during the scanning session. Obese patients and pregnant women may find the MRI machine to be a tight fit. Pregnant women may also have difficulty lying on their backs for an hour or more without moving.

Guidance
Safety issues, including the potential for biostimulation device interference, movement of ferromagnetic bodies, and incidental localized heating, have been addressed in the American College of Radiology's White Paper on MR Safety, which was originally published in 2002 and expanded in 2004. The ACR White Paper on MR Safety has been rewritten and was released early in 2007 under the new title ACR Guidance Document for Safe MR Practices (http://www.acr.org/SecondaryMainMenuCategories/quality_safety/MRSafety/safe_mr07.aspx). In December 2007, the Medicines in Healthcare product Regulation Agency (MHRA), a UK healthcare regulatory body, issued their Safety Guidelines for Magnetic Resonance Imaging Equipment in Clinical Use (http://www.mhra.gov.uk/home/idcplg? IdcService=GET_FILE&dDocName=CON2033065&RevisionSelectionMethod=LatestReleased). In February 2008, the Joint Commission, a US healthcare accrediting organization, issued a Sentinel Event Alert #38 (http://www.jointcommission.org/SentinelEvents/SentinelEventAlert/sea_38.htm), their highest patient safety advisory, on MRI safety issues. In July 2008, the United States Veterans Administration, a federal governmental agency serving the healthcare needs of former military personnel, issued a substantial revision to their MRI Design Guide,[105] which includes physical or facility safety considerations.

The European Physical Agents Directive

en.wikipedia.org/wiki/MRI

19/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

The European Physical Agents (Electromagnetic Fields) Directive is legislation adopted in European legislature. Originally scheduled to be required by the end of 2008, each individual state within the European Union must include this directive in its own law by the end of 2012. Some member nations passed complying legislation and are now attempting to repeal their state laws in expectation that the final version of the EU Physical Agents Directive will be substantially revised prior to the revised adoption date. The directive applies to occupational exposure to electromagnetic fields (not medical exposure) and was intended to limit workers acute exposure to strong electromagnetic fields, as may be found near electricity substations, radio or television transmitters or industrial equipment. However, the regulations impact significantly on MRI, with separate sections of the regulations limiting exposure to static magnetic fields, changing magnetic fields and radio frequency energy. Field strength limits are given, which may not be exceeded. An employer may commit a criminal offense by allowing a worker to exceed an exposure limit, if that is how the Directive is implemented in a particular member state. The Directive is based on the international consensus of established effects of exposure to electromagnetic fields, and in particular the advice of the European Commissions's advisor, the International Commission on Non-Ionizing Radiation Protection (ICNIRP). The aims of the Directive, and the ICNIRP guidelines it is based on, are to prevent exposure to potentially harmful fields. The actual limits in the Directive are very similar to the limits advised by the Institute of Electrical and Electronics Engineers, with the exception of the frequencies produced by the gradient coils, where the IEEE limits are significantly higher. Many Member States of the EU already have either specific EMF regulations or (as in the UK) a general requirement under workplace health and safety legislation to protect workers against electromagnetic fields. In almost all cases the existing regulations are aligned with the ICNIRP limits so that the Directive should, in theory, have little impact on any employer already meeting their legal responsibilities. The introduction of the Directive has brought to light an existing potential issue with occupational exposures to MRI fields. There are at present very few data on the number or types of MRI practice that might lead to exposures in excess of the levels of the Directive.[106][107] There is a justifiable concern amongst MRI practitioners that if the Directive were to be enforced more vigorously than existing legislation, the use of MRI might be restricted, or working practices of MRI personnel might have to change. In the initial draft a limit of static field strength to 2 T was given. This has since been removed from the regulations, and whilst it is unlikely to be restored as it was without a strong justification, some restriction on static fields may be reintroduced after the matter has been considered more fully by ICNIRP. The effect of such a limit might be to restrict the installation, operation and maintenance of MRI scanners with magnets of 2 T and stronger. As the increase in field strength has been instrumental in developing higher resolution and higher performance scanners, this would be a significant step back. This is why it is unlikely to happen without strong justification. Individual government agencies and the European Commission have now formed a working group to examine the implications on MRI and to try to address the issue of occupational exposures to electromagnetic fields from MRI.

See also
Earth's field NMR (EFNMR) Electron-spin resonance (spin physics) History of brain imaging
en.wikipedia.org/wiki/MRI 20/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

Jemris (open source MRI simulator) Magnetic immunoassay Magnetic particle imaging (MPI) Magnetic resonance elastography Magnetic Resonance Imaging (journal) Magnetic resonance microscopy Medical imaging Molecular breast imaging Nephrogenic fibrosing dermopathy Neuroimaging software Nobel Prize controversies Nuclear magnetic resonance (NMR) Relaxation (NMR) Robinson oscillator Rabi cycle Virtopsy

References
1. ^ Sheil, W. C. "Magnetic Resonance Imaging (MRI Scan)" (http://www.medicinenet.com/mri_scan/article.htm). MedicineNet.com. Retrieved 27 April 2012. 2. ^ Squire LF, Novelline RA (1997). Squire's fundamentals of radiology (5th ed.). Harvard University Press. ISBN 0-674-83339-2. 3. ^ a b c d "Magnetic Resonance" (http://www.magnetic-resonance.org). European Magnetic Resonance Forum/TRTF. Retrieved 01 May 2013. 4. ^ "Ct scan vs mri: Comparison chart" (http://www.diffen.com/difference/CT_Scan_vs_MRI). Diffen. Retrieved 27 April 2012. 5. ^ Shiel, W. C. "Magnetic Resonance Imaging (MRI Scan)" (http://www.medicinenet.com/mri_scan/article.htm). MedicineNet.com. Retrieved 25 April 2012. 6. ^ "MRI Contrast Agent" (http://www.oceannanotech.com/nav.php?qid=9). Ocean NanoTech. Retrieved 27 April 2012. 7. ^ Hendee, William R; Morgan, Christopher J (1984). "Magnetic Resonance Imaging Part IPhysical Principles" (http://www.ncbi.nlm.nih.gov/pmc/articles/PMC1021860). West J Med. 141 (4): 491500. PMC 1021860 (//www.ncbi.nlm.nih.gov/pmc/articles/PMC1021860). PMID 6506686 (//www.ncbi.nlm.nih.gov/pubmed/6506686). 8. ^ "MR of the shoulder with a 0.2-T permanent-magnet unit." (http://www.ajronline.org/content/154/4/777.full.pdf). 9. ^ Allen Counter, ke Olofsson, S.; Olofsson, A.; Borg, E.; Bjelke, B.; Hggstrm, A.; Grahn, H. (2000). "Analysis of Magnetic Resonance Imaging Acoustic Noise Generated by a 4.7 T Experimental System". Acta OtoLaryngologica 120 (6): 739743. doi:10.1080/000164800750000270 (http://dx.doi.org/10.1080%2F000164800750000270). PMID 11099151 (//www.ncbi.nlm.nih.gov/pubmed/11099151). 10. ^ Luechinger, R.; Duru, F.; Candinas, R.; Boesiger, P. (2004). "Safety considerations for magnetic resonance imaging of pacemaker and ICD patients". Herzschrittmachertherapie und Elektrophysiologie 15: 73. doi:10.1007/s00399-004-0401-5 (http://dx.doi.org/10.1007%2Fs00399-004-0401-5). 11. ^ Scott, Greig; Chronik, Blaine et al. (2001), "A Prepolarized MRI Scanner" (http://cds.ismrm.org/ismrm2001/PDF3/0610.pdf), Proc. Intl. Soc. Mag. Reson. Med 9: 610
en.wikipedia.org/wiki/MRI 21/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

12. ^ a b Levy, Dawn (20 Mar 2001). "Two magnets are cheaper than one: Stanford engineers construct an inexpensive MRI scanner" (http://news.stanford.edu/pr/01/mri321.html). Stanford University. Retrieved 26 April 2012. 13. ^ Venook RD, Matter NI, Ramachandran M, Ungersma SE, Gold GE, Giori NJ, Macovski A, Scott GC, Conolly SM (2006). "Prepolarized magnetic resonance imaging around metal orthopedic implants" (http://www.ncbi.nlm.nih.gov/pubmed/16724303). Magnetic Resonance in Medicine (Society of Magnetic Resonance in Medicine) 56 (1): 177186. doi:10.1002/mrm.20927 (http://dx.doi.org/10.1002%2Fmrm.20927). PMID 16724303 (//www.ncbi.nlm.nih.gov/pubmed/16724303). Retrieved 26 April 2012. 14. ^ Carr, Herman (1952). "Free Precession Techniques in Nuclear Magnetic Resonance". PhD thesis (Cambridge, MA: Harvard University). 15. ^ Carr, Herman Y. (July 2004). "Field Gradients in Early MRI". Physics Today (American Institute of Physics) 57 (7). Bibcode:2004PhT....57g..83C (http://adsabs.harvard.edu/abs/2004PhT....57g..83C). doi:10.1063/1.1784322 (http://dx.doi.org/10.1063%2F1.1784322). 16. ^ Encyclopedia of Nuclear Magnetic Resonance (http://books.google.com/books?id=OBvMgEACAAJ&dq=Encyclopedia%20of%20Nuclear%20Magnetic%20Resonance&source=gbs_book_other_ver sions) 1. Hoboken, NJ: Wiley and Sons. 1996. p. 253. 17. ^ MacWilliams, Bryon (2003). "News & Views: Russian claims first in magnetic imaging". Nature 426 (6965): 375. doi:10.1038/426375a (http://dx.doi.org/10.1038%2F426375a). PMID 14647349 (//www.ncbi.nlm.nih.gov/pubmed/14647349). 18. ^ "Best Regards to Alfred Nobel" (http://www.inauka.ru/english/article36919.html). Retrieved 2009-10-16. 19. ^ Damadian, R. V. "Tumor Detection by Nuclear Magnetic Resonance," Science, 171 (March 19, 1971): 1151 1153 20. ^ "The man who did not win" (http://www.smh.com.au/articles/2003/10/16/1065917548433.html). Sydney Morning Herald. 2003-10-17. Retrieved 2007-08-04. 21. ^ "Scan and Deliver" (http://opinionjournal.com/taste/?id=110001844l). Wall Street Journal. 2002-06-14. Retrieved 2007-08-04. 22. ^ "Apparatus And Method For Detectin Cancer In Tissue" (http://www.google.com/patents?id=-AZAAAAEBAJ&printsec=abstract#v=onepage&q&f=false). United States Patent and Trademark Office. 23. ^ "NSF history" (http://www.sri.com/policy/csted/reports/techin/mri3.html). Sri.com. Retrieved 2011-11-28. 24. ^ "Scientist Claims Exclusion From Nobel Prize for MRI" (http://articles.latimes.com/2003/oct/11/science/sciflap11). Los Angeles Times. 2003-11-08. Retrieved 2013-02-13. 25. ^ "Does Dr. Raymond Damadian Deserve the Nobel Prize for Medicine?" (http://www.highbeam.com/doc/1P189000528.html). The Armenian Reporter. 2003-11-08. Retrieved 2007-08-05. 26. ^ Lauterbur PC (1973). "Image Formation by Induced Local Interactions: Examples of Employing Nuclear Magnetic Resonance". Nature 242 (5394): 190191. Bibcode:1973Natur.242..190L (http://adsabs.harvard.edu/abs/1973Natur.242..190L). doi:10.1038/242190a0 (http://dx.doi.org/10.1038%2F242190a0). 27. ^ a b Filler, Aaron G. (2010). "The History, Development and Impact of Computed Imaging in Neurological Diagnosis and Neurosurgery: CT, MRI, and DTI." (http://www.ispub.com/journal/the-internet-journal-ofneurosurgery/volume-7-number-1/the-history-development-and-impact-of-computed-imaging-in-neurologicaldiagnosis-and-neurosurgery-ct-mri-and-dti.html#sthash.S16OjZPs.dpbs). The Internet Journal of Neurosurgery. 7 (1). doi:10.5580/23c6 (http://dx.doi.org/10.5580%2F23c6). Retrieved 3/11/2013. 28. ^ Lauterbur PC (1974). "Magnetic resonance zeugmatography". Pure and Applied Chemistry 40: 149157. doi:10.1351/pac197440010149 (http://dx.doi.org/10.1351%2Fpac197440010149). 29. ^ "First MRI and ultrasound scanning" (http://benbeck.co.uk/firsts/scanning.htm). Benjamin S. Beck. 30. ^ "The "Indomitable" MRI" (http://www.smithsonianmag.com/science-nature/object_jun00.html?c=y&page=2). Smithsonian Institution. 31. ^ Damadian R, Goldsmith M, Minkoff L (1977). "NMR in cancer: XVI. Fonar image of the live human body". Physiological Chemistry and Physics 9: 97100. 32. ^ Hinshaw DS, Bottomley PA, Holland GN (1977). "Radiographic thin-section image of the human wrist by nuclear magnetic resonance". Nature 270 (5639): 722723. Bibcode:1977Natur.270..722H (http://adsabs.harvard.edu/abs/1977Natur.270..722H). doi:10.1038/270722a0
en.wikipedia.org/wiki/MRI 22/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

33.

34. 35. 36. 37. 38. 39. 40. 41.

42. 43.

44.

45.

46. 47.

48. 49.

50.

51.

52.

(http://dx.doi.org/10.1038%2F270722a0). PMID 9065906 (//www.ncbi.nlm.nih.gov/pubmed/9065906). ^ Sijbers J et al (1996). "Quantification and improvement of the signal-to-noise ratio in a magnetic resonance image acquisition procedure" (http://webh01.ua.ac.be/visielab/papers/sijbers/mri96.pdf). Magnetic resonance imaging. 14(10): 11571163. PMID 9065906 (//www.ncbi.nlm.nih.gov/pubmed/9065906). ^ http://icmic.rad.jhmi.edu/icmicpersonnel/BIOBottomleyP.pdf ^ 9.4T human MRI: Preliminary results (http://onlinelibrary.wiley.com/doi/10.1002/mrm.21073/abstract) ^ A volume birdcage coil with an adjustable sliding tuner ring for neuroimaging in high field vertical magnets: Ex and in vivo applications at 21.1 T (http://www.sciencedirect.com/science/article/pii/S1090780712002017) ^ "FONAR press release" (http://www.fonar.com/news/101603.htm). ^ H.F. Judso (20 October 2003). "No Nobel Prize for whining" (http://www.nytimes.com/2003/10/20/opinion/nonobel-prize-for-whining.html). New York Times. ^ [Medical imaging explanations "Medical imaging explanations"] Check | u r l =scheme (help). The Institution of Engineering and Technology. Retrieved 28 April 2012. ^ "T1 Weighted" (http://www.mr-tip.com/serv1.php?type=db1&dbs=T1%20Weighted), www.mr-tip.com ^ Le Bihan D, Breton E, Lallemand D, Grenier P, Cabanis E, Laval-Jeantet M. (November 1986). "MR imaging of intravoxel incoherent motions: Application to diffusion and perfusion in neurologic disorders". Radiology. 161 (2): 4017. PMID 3763909 (//www.ncbi.nlm.nih.gov/pubmed/3763909). ^ "Diffusion Inaging" (http://www-mrsrl.stanford.edu/research.html). Stanford University. Retrieved 28 April 2012. ^ Moseley ME, Cohen Y, Mintorovitch J, Chileuitt L, Shimizu H, Kucharczyk J, Wendland MF, Weinstein PR. (1990). "Early detection of regional cerebral ischemia in cats: Comparison of diffusion- and T2-weighted MRI and spectroscopy". Magn Reson Med 14 (2): 33046. doi:10.1002/mrm.1910140218 (http://dx.doi.org/10.1002%2Fmrm.1910140218). PMID 2345513 (//www.ncbi.nlm.nih.gov/pubmed/2345513). ^ Borthakur, A; Mellon, E, Niyogi, S, Witschey, W, Kneeland, JB, Reddy, R (2006 Nov). "Sodium and T1rho MRI for molecular and diagnostic imaging of articular cartilage." (http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2896046). NMR in biomedicine 19 (7): 781821. doi:10.1002/nbm.1102 (http://dx.doi.org/10.1002%2Fnbm.1102). PMC 2896046 (//www.ncbi.nlm.nih.gov/pmc/articles/PMC2896046). PMID 17075961 (//www.ncbi.nlm.nih.gov/pubmed/17075961). ^ De Coene B, Hajnal JV, Gatehouse P, Longmore DB, White SJ, Oatridge A, Pennock JM, Young IR, Bydder GM. (November 1992). "MR of the brain using fluid-attenuated inversion recovery (FLAIR) pulse sequences". Am J Neuroradiol 13 (6): 155564. PMID 1332459 (//www.ncbi.nlm.nih.gov/pubmed/1332459). ^ Haacke, E Mark; Brown, Robert F; Thompson, Michael; Venkatesan, Ramesh (1999). Magnetic resonance imaging: Physical principles and sequence design. New York: J. Wiley & Sons. ISBN 0-471-35128-8. ^ Ridgway JP, Smith MA (June 1986). "A technique for velocity imaging using magnetic resonance imaging". Br J Radiol 59 (702): 6037. doi:10.1259/0007-1285-59-702-603 (http://dx.doi.org/10.1259%2F0007-1285-59-702603). PMID 3708269 (//www.ncbi.nlm.nih.gov/pubmed/3708269). ^ Njemanze PC, Beck OJ (1989). "MR-gated intracranial CSF dynamics: Evaluation of CSF pulsatile flow". AJNR Am J Neuroradiol 10 (1): 7780. PMID 2492733 (//www.ncbi.nlm.nih.gov/pubmed/2492733). ^ Rosen Y, Lenkinski RE (2007). "The Recent advances in magnetic resonance neurospectroscopy". Neurotherapeutics 27 (3): 33045. doi:10.1016/j.nurt.2007.04.009 (http://dx.doi.org/10.1016%2Fj.nurt.2007.04.009). PMID 17599700 (//www.ncbi.nlm.nih.gov/pubmed/17599700). ^ Golder W (2007). "Magnetic resonance spectroscopy in clinical oncology". Onkologie 27 (3): 3049. doi:10.1159/000077983 (http://dx.doi.org/10.1159%2F000077983). PMID 15249722 (//www.ncbi.nlm.nih.gov/pubmed/15249722). ^ Thulborn KR, Waterton JC, Matthews PM, Radda GK (February 1982). "Oxygenation dependence of the transverse relaxation time of water protons in whole blood at high field" (http://linkinghub.elsevier.com/retrieve/pii/0304-4165(82)90333-6). Biochim. Biophys. Acta 714 (2): 26570. doi:10.1016/0304-4165(82)90333-6 (http://dx.doi.org/10.1016%2F0304-4165%2882%2990333-6). PMID 6275909 (//www.ncbi.nlm.nih.gov/pubmed/6275909). ^ H. E. Cline, J. F. Schenck, K. Hynynen, R. D. Watkins, S. P. Souza, and F. A. Jolesz. "MR-guided focused ultrasound surgery," J Comput Assist Tomogr 16(6), 95665 (1992).
23/28

en.wikipedia.org/wiki/MRI

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

53. ^ "Radiology" (http://radiology.rsna.org/content/204/1/272.long). Retrieved 2 August 2010. (subscription required) 54. ^ Wattjes MP, Lutterbey GG, Gieseke J, et al. (1 January 2007). "Double inversion recovery brain imaging at 3T: Diagnostic value in the detection of multiple sclerosis lesions" (http://www.ajnr.org/cgi/pmidlookup? view=long&pmid=17213424). AJNR Am J Neuroradiol 28 (1): 549. PMID 17213424 (//www.ncbi.nlm.nih.gov/pubmed/17213424). 55. ^ Nelson F, Poonawalla AH, Hou P, Huang F, Wolinsky JS, Narayana PA (October 2007). "Improved identification of intracortical lesions in multiple sclerosis with phase-sensitive inversion recovery in combination with fast double inversion recovery MR imaging". AJNR Am J Neuroradiol 28 (9): 16459. doi:10.3174/ajnr.A0645 (http://dx.doi.org/10.3174%2Fajnr.A0645). PMID 17885241 (//www.ncbi.nlm.nih.gov/pubmed/17885241). 56. ^ Nelson F, Poonawalla A, Hou P, Wolinsky JS, Narayana PA (November 2008). "3D MPRAGE improves classification of cortical lesions in multiple sclerosis" (http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2650249). Mult Scler. 14 (9): 12149. doi:10.1177/1352458508094644 (http://dx.doi.org/10.1177%2F1352458508094644). PMC 2650249 (//www.ncbi.nlm.nih.gov/pmc/articles/PMC2650249). PMID 18952832 (//www.ncbi.nlm.nih.gov/pubmed/18952832). 57. ^ Brant-Zawadzki M, Gillan GD, Nitz WR (March 1992). "MP RAGE: A three-dimensional, T1-weighted, gradientecho sequenceinitial experience in the brain" (http://radiology.rsnajnls.org/cgi/pmidlookup? view=long&pmid=1535892). Radiology 182 (3): 76975. PMID 1535892 (//www.ncbi.nlm.nih.gov/pubmed/1535892). 58. ^ Shenghui, Xue; Jingjuan Qiao, Fan Pu, Mathew Cameron, Jenny J. Yang (17). "Design of a novel class of protein-based magnetic resonance imaging contrast agents for the molecular imaging of cancer biomarkers." (http://www.ncbi.nlm.nih.gov/pubmed/23335551). Wiley Interdiscip Rev Nanomed Nanobiotechnol. 5 (2). doi:10.1002/wnan.1205 (http://dx.doi.org/10.1002%2Fwnan.1205). PMID 23335551 (//www.ncbi.nlm.nih.gov/pubmed/23335551). 59. ^ "Terranova-MRI Earth's Field MRI teaching system" (http://www.magritek.com/terranova.html). Magritek.com. Retrieved 2010-08-02. 60. ^ I. M. Savukov and M. V. Romalis (2005). "NMR Detection with an Atomic Magnetometer" (http://www.atomic.princeton.edu/romalis/magnetometer/papers/Savukov%20and%20Romalis%20%20NMR%20Detection%20with%20an%20Atomic%20Magnetometer.pdf) (PDF). Physical Review Letters 94. arXiv:physics/0411163 (http://arxiv.org/abs/physics/0411163). Bibcode:2005PhRvL..94l3001S (http://adsabs.harvard.edu/abs/2005PhRvL..94l3001S). doi:10.1103/PhysRevLett.94.123001 (http://dx.doi.org/10.1103%2FPhysRevLett.94.123001). Blog comment: "Hi-res, cheap & portable MRI" (http://neurophilosophy.wordpress.com/2006/09/06/hi-res-cheap-portablemri/). Neurophilosophy (blog). 61. ^ Raftery D (August 2006). "MRI without the magnet" (http://www.ncbi.nlm.nih.gov/pmc/articles/PMC1568902). Proc Natl Acad Sci USA. 103 (34): 126578. Bibcode:2006PNAS..10312657R (http://adsabs.harvard.edu/abs/2006PNAS..10312657R). doi:10.1073/pnas.0605625103 (http://dx.doi.org/10.1073%2Fpnas.0605625103). PMC 1568902 (//www.ncbi.nlm.nih.gov/pmc/articles/PMC1568902). PMID 16912110 (//www.ncbi.nlm.nih.gov/pubmed/16912110). 62. ^ Wu Y, Chesler DA, Glimcher MJ, et al. (February 1999). "Multinuclear solid-state three-dimensional MRI of bone and synthetic calcium phosphates" (http://www.pnas.org/cgi/pmidlookup?view=long&pmid=9990066). Proc. Natl. Acad. Sci. U.S.A. 96 (4): 15748. Bibcode:1999PNAS...96.1574W (http://adsabs.harvard.edu/abs/1999PNAS...96.1574W). doi:10.1073/pnas.96.4.1574 (http://dx.doi.org/10.1073%2Fpnas.96.4.1574). PMC 15521 (//www.ncbi.nlm.nih.gov/pmc/articles/PMC15521). PMID 9990066 (//www.ncbi.nlm.nih.gov/pubmed/9990066). 63. ^ Mietchen, D.; Aberhan, M.; Manz, B.; Hampe, O.; Mohr, B.; Neumann, C.; Volke, F. (2008). "Threedimensional Magnetic Resonance Imaging of fossils across taxa" (http://direct.sref.org/1726-4189/bg/2008-5-25). Biogeosciences 5 (1): 2541. doi:10.5194/bg-5-25-2008 (http://dx.doi.org/10.5194%2Fbg-5-25-2008). Retrieved 2008-04-08. 64. ^ Brenner DJ, Hall EJ (November 2007). "Computed tomographyan increasing source of radiation exposure" (http://www.columbia.edu/~djb3/papers/nejm1.pdf). N. Engl. J. Med. 357 (22): 227784. doi:10.1056/NEJMra072149 (http://dx.doi.org/10.1056%2FNEJMra072149). PMID 18046031
en.wikipedia.org/wiki/MRI 24/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

65.

66.

67.

68.

69.

70. 71. 72.

73.

74.

75.

76. 77. 78. 79. 80.

(//www.ncbi.nlm.nih.gov/pubmed/18046031). ^ Detection of DNA double-strand breaks using ?h2AX after MRI exposure at 3 Tesla: An in vitro study (http://onlinelibrary.wiley.com/doi/10.1002/jmri.21138/abstract;jsessionid=CC3D9CAA0C505CB441378CB24102B CA8.d02t03?userIsAuthenticated=false&deniedAccessCustomisedMessage=) "Neither exposure to the static magnetic field alone nor to the applied imaging sequences showed significant differences in ?H2AX expression between exposed and sham-exposed cells. X-raytreated cells as positive control showed a significant increase in ? H2AX expression." ^ "Nephrogenic Systemic Fibrosis" (http://www.acr.org/~/media/ACR/Documents/PDF/QualitySafety/Resources/Contrast%20Manual/Nephrogenic%2 0Systemic%20Fibrosis.pdf). ACR Manual on Contrast Material. American College of Radiology. Retrieved 13 October 2012. ^ Allen ED, Byrd SE, Darling CF, Tomita T, Wilczynski MA. (1993). "The clinical and radiological evaluation of primary brain neoplasms in children, Part II: Radiological evaluation" (http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2568155). J Natl Med Assoc. 85 (7): 54653. PMC 2568155 (//www.ncbi.nlm.nih.gov/pmc/articles/PMC2568155). PMID 8350377 (//www.ncbi.nlm.nih.gov/pubmed/8350377). ^ Deck MD, Henschke C, Lee BC, Zimmerman RD, Hyman RA, Edwards J, Saint Louis LA, Cahill PT, Stein H, Whalen JP. (March 1989). "Computed tomography versus magnetic resonance imaging of the brain. A collaborative interinstitutional study". Clin Imaging 13 (1): 215. doi:10.1016/0899-7071(89)90120-4 (http://dx.doi.org/10.1016%2F0899-7071%2889%2990120-4). PMID 2743188 (//www.ncbi.nlm.nih.gov/pubmed/2743188). ^ "Research Areas: Prepolarized MRI for Imaging Near Metal Implants" (http://wwwmrsrl.stanford.edu/research.html). Stanford University. Retrieved 26 April 2012. "Prepolarized MRI is a novel method of MRI that uses two inexpensive electromagnets to make an MRI study of the human body. By using inexpensive electromagnets, this new method allows for high quality imaging at nearly an order-of-magnitude cost reduction relative to conventional MRI. The total scanner hardware cost is about $50,000, including manufacturer's materials, labor and profit." ^ Stamford Hospital price quotation October 2008, Stamford CT US ^ Gordon AC, et al "Over-utilization of MRI in the osteoarthritis patient" AAOS meeting 2008; P145. ^ Current Procedural Terminology code #70553 "2010 Medicare Part B National Summary Data File" (https://explore.data.gov/Health-and-Nutrition/Part-B-National-Summary-Data-File-CY2010/5vgw-8z3q). Data.gov, An Official Website of the United States Government. ^ Ehara S, Nakajima Y, Matsui O. "Radiology in Japan in 2008" (http://www.ajronline.org/content/191/2/328.full). AJR Am J Roentgenol. 2008 Aug;191(2):328-9. PubMed PMID 18647897 (http://www.ncbi.nlm.nih.gov/pubmed? term=18647897). ^ Consumer Reports; American College of Physicians, "Five Things Physicians and Patients Should Question" (http://choosingwisely.org/wp-content/uploads/2012/04/5things_12_factsheet_Amer_College_Phys.pdf), presented by ABIM Foundation, Choosing Wisely (Consumer Reports), retrieved August 14, 2012 ^ Consumer Reports; American College of Physicians (April 2012), "Imaging tests for lower-back pain: Why you probably dont need them." (http://consumerhealthchoices.org/wp-content/uploads/2012/04/High-Value-Care-BackPain-ACP.pdf), presented by Annals of Internal Medicine, High Value Care (Consumer Reports), retrieved August 14, 2012 ^ "Physics of magnetic resonance imaging" (http://www.my-ms.org/mri_physics.htm). My-MS.org. Retrieved 27 April 2012. ^ "Nano-coated implants cut MRI scan dangers 01 June 2003" (http://www.newscientist.com/article.ns? id=dn3771). New Scientist. Retrieved 2011-11-28. ^ Jost, C.; Kumar V, V. (1998). "Are Current Cardiovascular Stents MRI Safe?". The Journal of invasive cardiology 10 (8): 477479. PMID 10762825 (//www.ncbi.nlm.nih.gov/pubmed/10762825). ^ "Magnetic resonance safety policy of ucsf" (http://www.radiology.ucsf.edu/patient-care/patientsafety/mri/policy). University of California, San Francisco. Retrieved 28 April 2012. ^ Randal C. Archibold, "Hospital Details Failures Leading to M.R.I. Fatality (http://query.nytimes.com/gst/fullpage.html?res=9C04E4D91731F931A1575BC0A9679C8B63)," The New York
25/28

en.wikipedia.org/wiki/MRI

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

81. 82.

83. 84.

85.

86. 87.

88. 89.

90.

91.

92.

93. 94. 95.

96. 97.

Times, August 22, 2001 ^ Donald G. McNeil Jr, "M.R.I.'s Strong Magnets Cited in Accidents (http://www.nytimes.com/2005/08/19/health/19magnet.html)," The New York Times, August 19, 2005. ^ "ACR Guidance Document for Safe MR Practices: 2007" (http://www.acr.org/SecondaryMainMenuCategories/quality_safety/MRSafety/safe_mr07.aspx). Retrieved 2 August 2010. ^ "MRI Design Guide" (http://www.Mednovus.com/downloads/VA_MRI_Design_Guide-08.pdf). Retrieved 2 August 2010. ^ "Biological effects of exposure to magnetic resonance imaging: an overview" (http://www.biomedicalengineering-online.com/content/3/1/11), Domenico Formica and Sergio Silvestr, BioMedical Engineering OnLine (2004) ^ MS Cohen, RM Weisskoff, RR Rzedzian and HL Kantor (1990). "Sensory stimulation by time-varying magnetic fields". Magnetic Resonance in Medicine 14 (2): 409414. doi:10.1002/mrm.1910140226 (http://dx.doi.org/10.1002%2Fmrm.1910140226). PMID 2345521 (//www.ncbi.nlm.nih.gov/pubmed/2345521). ^ TF Budinger, H Fischer, D Hentschel, HE Reinfelder and F Schmitt (1991). "Physiological effects of fast oscillating magnetic field gradients". J Comput Assist Tomogr 15 (6): 909914. ^ J Reilly (1989). "Peripheral nerve stimulation by induced electric currents: exposure to time-varying magnetic fields". Medical & Biological Engineering and Computing 27 (2): 101110. doi:10.1007/BF02446217 (http://dx.doi.org/10.1007%2FBF02446217). PMID 2689806 (//www.ncbi.nlm.nih.gov/pubmed/2689806). ^ "The Evolution of Magnetic Resonance Imaging: 3T MRI in Clinical Applications" (http://www.eradimaging.com/site/article.cfm?ID=426), Terry Duggan-Jahns, www.eradimaging.com ^ Price DL, de Wilde JP, Papadaki AM, Curran JS, Kitney RI (January 2001). "Investigation of acoustic noise on 15 MRI scanners from 0.2 T to 3 T". Journal of Magnetic Resonance Imaging 13 (2): 288293. doi:10.1002/1522-2586(200102)13:2<288::AID-JMRI1041>3.0.CO;2-P (http://dx.doi.org/10.1002%2F15222586%28200102%2913%3A2%3C288%3A%3AAID-JMRI1041%3E3.0.CO%3B2-P). PMID 11169836 (//www.ncbi.nlm.nih.gov/pubmed/11169836). ^ The Open University 2007: Understanding Cardiovascular Diseases, course book for the lesson SK121 Understanding cardiovascular diseases (http://www3.open.ac.uk/courses/bin/p12.dll?C01SK121), printed by Cambridge University Press, ISBN 978-0-7492-2677-0 (can be found at OUW (http://www.ouw.co.uk/)), pages 220 and 224. ^ Kanal E, Barkovich AJ, Bell C, et al. (2007). (http://www.acr.org/SecondaryMainMenuCategories/quality_safety/MRSafety/safe_mr07.aspx)dead link] "ACR Guidance Document for Safe MR Practices: 2007". AJR Am J Roentgenol. 188 (6): 127. doi:10.2214/AJR.06.1616 (http://dx.doi.org/10.2214%2FAJR.06.1616). PMID 17515363 (//www.ncbi.nlm.nih.gov/pubmed/17515363). page 22. ^ International Electrotechnical Commission 2008: Medical Electrical Equipment Part 2-33: Particular requirements for basic safety and essential performance of magnetic resonance equipment for medical diagnosis, manufacturers' trade standards [1] (http://webstore.iec.ch/Webstore/webstore.nsf/0/EC11496F487C406DC125742C000B2805), published by International Electrotechnical Commission, ISBN 2-8318-9626-6 (can be found for purchase at [2] (http://webstore.iec.ch/Webstore/webstore.nsf/0/EC11496F487C406DC125742C000B2805)). ^ "Cryogen Awareness Quenching and MRI Safety Training" (http://www.falckproductions.com/resources/mrisafety/cryogen-awareness-quenching-and-your-mri-machine/). Falck Productions. Retrieved 10 July 2012. ^ "GE Health Care" (http://www.gehealthcare.com/usen/mr/docs/SPV8_Avoiding_Quench.pdf). GE. Retrieved 10 July 2012. ^ Murphy KJ, Brunberg JA, Cohan RH (1 October 1996). "Adverse reactions to gadolinium contrast media: A review of 36 cases" (http://www.ajronline.org/cgi/pmidlookup?view=long&pmid=8819369). AJR Am J Roentgenol 167 (4): 8479. PMID 8819369 (//www.ncbi.nlm.nih.gov/pubmed/8819369). ^ "ACR guideline (http://www.guideline.gov/summary/summary.aspx?doc_id=8283), 2005" ^ H.S. Thomsen, S.K. Morcos and P. Dawson (November 2006). "Is there a causal relation between the administration of gadolinium-based contrast media and the development of nephrogenic systemic fibrosis (NSF)?". Clinical Radiology 61 (11): 9056. doi:10.1016/j.crad.2006.09.003
26/28

en.wikipedia.org/wiki/MRI

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

98.

99. 100. 101. 102.

103.

104.

105. 106.

107.

(http://dx.doi.org/10.1016%2Fj.crad.2006.09.003). PMID 17018301 (//www.ncbi.nlm.nih.gov/pubmed/17018301). ^ "FDA Drug Safety Communication: New warnings for using gadolinium-based contrast agents in patients with kidney dysfunction" (http://www.fda.gov/Drugs/DrugSafety/ucm223966.htm). Information on Gadolinium-Based Contrast Agents. U.S. Food and Drug Administration. 23 December 2010. Retrieved 12 March 2011. ^ "FDA Public Health Advisory: Gadolinium-containing Contrast Agents for Magnetic Resonance Imaging (http://www.fda.gov/cder/drug/advisory/gadolinium_agents.htm)" ^ [3] (http://www.mhra.gov.uk/home/idcplg? IdcService=GET_FILE&dID=35149&noSaveAs=0&Rendition=WEB) ^ "ismrm.org MRI Questions and Answers" (http://www.ismrm.org/special/EMEA2.pdf) (PDF). Retrieved 201008-02. ^ "Response to the FDA's May 23, 2007, Nephrogenic Systemic Fibrosis Update1 Radiology" (http://radiology.rsna.org/content/246/1/11.full? searchid=1&HITS=10&hits=10&sortspec=relevance&resourcetype=HWCIT&maxtoshow=&RESULTFORMAT= &author1=kanal&FIRSTINDEX=0). Radiology.rsna.org. 2007-09-12. Retrieved 2010-08-02. ^ Ibrahim A. Alorainy, Fahad B. Albadr, Abdullah H. Abujamea (2006). "Attitude towards MRI safety during pregnancy" (http://www.saudiannals.net/pdfs/06-201.pdf). Ann Saudi Med 26 (4): 3069. PMID 16885635 (//www.ncbi.nlm.nih.gov/pubmed/16885635). ^ "Siemens Introduces First 1.5 Tesla Open Bore MRI" (http://www.medical.siemens.com/webapp/wcs/stores/servlet/PressReleaseView~q_catalogId~e_-1~a_catTree~e_ 100005,13839,17712~a_langId~e_-1~a_pageId~e_50677~a_storeId~e_10001.htm). Medical.siemens.com. 200407-29. Retrieved 2010-08-02. ^ "MRI Design Guide" (http://www.cfm.va.gov/til/dGuide.asp). United States Department of Veterans Affairs. April 2008. Retrieved 12 October 2012. ^ Bassen, H; Schaefer, D J. ; Zaremba, L; Bushberg, J; Ziskin, M [S]; Foster, K R. (2005). "IEEE Committee on Man and Radiation (COMAR) technical information statement "Exposure of medical personnel to electromagnetic fields from open magnetic resonance imaging systems"". Health Physics 89 (6): 6849. doi:10.1097/01.HP.0000172545.71238.15 (http://dx.doi.org/10.1097%2F01.HP.0000172545.71238.15). PMID 16282801 (//www.ncbi.nlm.nih.gov/pubmed/16282801). ^ HSE 2007,RR570:Assessment of electromagnetic fields around magnetic resonance (MRI) equipment (http://www.hse.gov.uk/research/rrpdf/rr570.pdf). MCL-T Ltd, London

Further reading
Guadalupe Portal; Aliosvi Rodriguez Whole body magnetic resonance imaging in early diagnosis in Trinidad BMJ (2010) ISSN 1756-1833 url = http://www.bmj.com/rapid-response/2011/12/19/re-whole-bodymagnetic-resonance-imaging Simon, Merrill; Mattson, James S (1996). The pioneers of NMR and magnetic resonance in medicine: The story of MRI. Ramat Gan, Israel: Bar-Ilan University Press. ISBN 0-9619243-1-4. Haacke, E Mark; Brown, Robert F; Thompson, Michael; Venkatesan, Ramesh (1999). Magnetic resonance imaging: Physical principles and sequence design. New York: J. Wiley & Sons. ISBN 0471-35128-8. Lee, S. C. et al. (2001). "One Micrometer Resolution NMR Microscopy". J. Magn. Res 150 (2): 207 213. Bibcode:2001JMagR.150..207L (http://adsabs.harvard.edu/abs/2001JMagR.150..207L). doi:10.1006/jmre.2001.2319 (http://dx.doi.org/10.1006%2Fjmre.2001.2319).

External links
en.wikipedia.org/wiki/MRI 27/28

6/6/13

Magnetic resonance imaging - Wikipedia, the free encyclopedia

A Peer-Reviewed, Critical Introduction (http://www.magnetic-resonance.org/). European Magnetic Resonance Forum (EMRF)/The Round Table Foundation (TRTF); Peter A. Rinck (editor) A Guided Tour of MRI: An introduction for laypeople (http://www.magnet.fsu.edu/education/tutorials/magnetacademy/mri/) National High Magnetic Field Laboratory The Basics of MRI (http://www.cis.rit.edu/htbooks/mri/). Underlying physics and technical aspects. Video: What to Expect During Your MRI Exam (http://www.imrser.org/PatientVideo.html) from the Institute for Magnetic Resonance Safety, Education, and Research (IMRSER) International Society for Magnetic Resonance in Medicine (http://www.ismrm.org) Trends in Biotechnology Volume 28, Issue 7, July 2010, Pages 363370 (http://dx.doi.org/10.1016/j.tibtech.2010.04.002) Blue Plaque commemorating the manufacture of the first commercial MRI whole body scanner at Osney Mead, Oxford (http://www.oxfordshireblueplaques.org.uk/plaques/mri.html) Royal Institution Lecture MRI: A Window on the Human Body (http://www.vega.org.uk/video/programme/73) Animal Imaging Database (AIDB) (http://aidb.crbs.ucsd.edu/sand/index.jsp) How MRI works explained simply using diagrams (http://www.howequipmentworks.com/physics/medical_imaging/mri/magnetic_resonance_imaging.html) Retrieved from "http://en.wikipedia.org/w/index.php?title=Magnetic_resonance_imaging&oldid=557871480" Categories: American inventions Discovery and invention controversies 1973 introductions Magnetic resonance imaging This page was last modified on 1 June 2013 at 19:22. Text is available under the Creative Commons Attribution-ShareAlike License; additional terms may apply. By using this site, you agree to the Terms of Use and Privacy Policy. Wikipedia is a registered trademark of the Wikimedia Foundation, Inc., a non-profit organization.