Number T10112 revision 8 Project number SO12200027 Page 1 of 1 Issued by NMi Certin B.V., designated and notified by the Netherlands to perform tasks with respect to conformity modules mentioned in article 9 of Directive 2004/22/EC, after having established that the Measuring instrument meets the applicable requirements of Directive 2004/22/EC, to: S.C. AEM S.A Calea Buziasului 26 300693 Timisoara Romania A Diaphragm Gas Meter Type Destined for the measurement of Accuracy class Environment classes Temperature range Location Further are described in the annexes Description T10112 revision 8 Documentation folder T10112-7 : : : : : : GN x / GNM x Gas volume Class 1,5 M2 / E1 See description T10112 revision 8 Closed
Manufacturer
Measuring instrument
14 April 2018 This revision replaces the earlier versions, including its documentation folder.
Issuing Authority
500 101
Description
Number T10112 revision 8 Project number SO12200027 Page 1 of 3
1.1
Essential parts Description Construction Diaphragm Valve and valve seat Temperature compensation Documentation 10112/2-02 10112/3-01 10112/0-06 10112/0-07 10112/0-08 10112/2-04 Remarks Description Material Synthetic rubber, manufacturer EFFBE France.
1.2 1.2.1
Essential characteristics See EC Type-examination certificate no. T10112 Revision 8 and the characteristics mentioned below: Type GN G6(E/ET) GNM G6(E/ET) GN G4(E/ET) GNM G4(E/ET) GN G4CT(E) GN G4TC(E) GN G4T(E/ET) GN G2,5(E/ET) GNM G2,5(E/ET) Qmax [m3/h] 10 10 6 6 6 6 6 4 4 Qmin [m3/h] 0,060 0,060 0,040 0,040 0,040 0,040 0,040 0,025 0,025 Qt [m3/h] 1,0 1,0 0,6 0,6 0,6 0,6 0,6 0,4 0,4 V [dm3] 3,5 3,5 1,2 1,2 1,2 2,0 2,0 1,2 1,2 Tmin [C] -10 -25 -10 -25 -25 -25 -25 -10 -25 Tmax [C] +40 +55 +40 +55 +40 +40 +55 +40 +55 Maximum pmax [bar] 0,5 0,5 0,5 0,5 0,5 0,5 0,5 0,5 0,5
Essential shapes The nameplate is bearing at least, good legible, the information as mentioned in the EN1359/A1. An example of the markings is shown in document no. 10112/2-03. Sealing: see chapter 2.
1
500
Description
Number T10112 revision 8 Project number SO12200027 Page 2 of 3 1.4 1.4.1 Conditional parts Construction In addition to the essential parts as mentioned at 1.1, the meter contains at least the following conditional parts: - housing; - transmission; - register. The meter can also be provided with low frequency impulse output. Housing The gas meter has a housing, which has sufficient tensile strength. The cover is made of steel sheet, the lower and upper case are connected with each other by a clamp. The register case is also connected to the upper case by screws. Examples of the housing are stated on documents no.s 10112/2-01, 10112/0-03, 10112/4-01 and 10112/6-01. 1.4.2 Transmission The transmission between the measuring part and the register is executed via a fixed mechanical (GNM) coupling or a magnetic (GN) coupling. When an electronic register is used the pulse transmission is checked using a three-pulse system. In document 10112/7-04 an example of the pulse transmission is shown. Register The indication can take place by two different means: - A mechanical register which indicates the volume in m3, by at least 5 drums before the comma and 3 drums after the comma. There can be two different versions of the mechanical register, see drawings no. 10112/2-05 and 10112/8-01 where examples of the counters are presented. - An electronic index register which indicates the volume in m3 on an LCD display with 9 digits of which at least 5 digits before the comma and 3 digits after the comma. The electronic index can optionally be equipped with temperature compensation. The use of the electronic register will results in an E or ET marking in the meter type. In documents no. 10112/7-01, 10112/7-02 and 10112/7-03 the drawings of the electronic register and its PCBs are presented. Adjustment The gas meter is adjustable by using the adjustment screw, see the adjustment instructions on document no. 10112/0-09. When equipped with the mechanical register as described in document no. 10112/8-01, the gas meter is also adjustable by choosing different gearwheel combinations as listed in document 10112/8-01, page 3. Software Measuring instrument type P Risk Class C; while extension L, T, D and S are not applicable or excluded. Software part Electronic register Version 1.2 Checksum 33 Remark The software version 1.2 and checksum 33 are presented on the display every 10 minutes for 6 seconds each.
1.4.3
1.4.4
1.4.5
2
500
Description
Number T10112 revision 8 Project number SO12200027 Page 3 of 3
1.5
Conditional shapes Connections Type Pipe configuration Minimum connection diameter [mm] 27 27 20 20 20 20 50 20 50 20 20 Maximum connection distance [mm] 250 250 165 165 165 250 250 165 165
GN G6(E/ET) GNM G6(E/ET) GN G4(E/ET) GNM G4(E/ET) GN G4CT(E) GN G4TC(E) GN G4TC(E) GN G4T(E/ET) GN G4T(E/ET) GN G2,5(E/ET) GNM G2,5(E/ET) 1.6 1.6.1 1.6.2 1.6.3 1.6.4 1.6.5
Double Double Double Double Double Double Single Double Single Double Double
Non-essential parts Reverse stop for preventing registration in reversed flow direction Pulse generator Serial communication through optical port Wired M-bus Radio M-bus
Seals
The following items of the meter are sealed: The entrance to the register is sealed with one or more seals. If a separate nameplate is used to show the pulse value this nameplate has to be sealed.
See drawings no. 10112/0-03 and 10112/8-01 page 1 for an example of the sealing.
3
500