Sie sind auf Seite 1von 2

CTS-10458

Assy, Rotor Seal Equivalent to: Agilent 0101-0626

Technical Data Sheet


1.0 Description
This Assembly Rotor Seal is designed to be compatible with Agilent 1050 modular HPLC systems. It is equivalent to Agilent part number 0101-0626.

2.0 Specifications
2.1 Operating Specifications
Pressure: Max 6000psi/400bar (with Stainless steel sample loop) Max 5000psi (with PEEK sample loop) Operating Range: 4C - 80C pH Range: 0-10 2. Remove the three set stator screws using an Allen wrench. 3. Remove the stator head, stator face, and ring. 4. Pull the rotor seal from the 4 set pins on the valve body. 5. Remove the isolation seal and replace the isolation seal if necessary (open or spring side facing the valve body). 6. Mount the new rotor seal with the grooves facing the stator. Line up the rotor seal so that the notch on the rotor seals metal band aligns with the rotor pin. 7. Re-install the stator ring, stator face and stator head assembly. 8. Replace the three stator screws. Gradually tighten to each screw finger tight, alternating among the three while tightening down and then tighten each until secure or no more than 1/4 turn past fingertight.

2.2 Wetted Materials


Vespel

3.0 Replacement Frequency


The rotor seal generally is not part of a required maintenance protocol. Indications that the rotor may need replaced are poor reproducibility of injection volume or a leak at the injection valve.

4.0 Installation Information


CTSs parts are designed to be compatible and interchangeable with the OEM parts. 1. Remove all of the capillary lines attached to injection valve ports making sure to label or identify the position of each line prior to removal.

5.0 Operating Recommendations


The typical cause of failure at the rotor seal is wear of the Vespel material. A major

cause of wear is the use of buffered mobile phases. To help improve the lifetime of the rotor seal it is recommended that the system and injection valve be flushed with either the mobile phase without the buffer addition or with an aqueous solution. Be sure to toggle the valve through both the by-pass and inject positions to flush all of the mobile phase salts from the injection port.

6.0 Warranty
We view your business with us as a partnership based on mutual trust. We extend a Lifetime Warranty against any manufacturing defects on all CTS parts. On the rare occasion that you experience a problem, we invite you to contact your Account Manager directly. Returns can only be accepted if a Return Material Authorization Number (RMA) is obtained prior to shipment. Please call (1) 952-895-8292 or email: ctssales@ctshplc.com to get an RMA.