Flow measurement

Selection guide

12.06/I.I.

Instruments International Endress+Hauser Instruments International AG Kaegenstrasse 2 4153 Reinach Switzerland Tel. +41 61 715 81 00 Fax +41 61 715 25 00 http://www.endress.com info@ii.endress.com

CP001D/06/en/04.07 71024143 INDD CS2

An example of using this selection guide


Selecting the measuring principle
D  efine all the primary requirements for your measuring point: fluid type, nominal diameter, accuracy, pressure, temperature, etc. Example

Requirements

Fluids (examples)
Electromagnetic

Measuring principle
Differential pressure Ultrasonic Thermal Coriolis Vortex

Fluid: drinking water (urban area) Pipe diameter: DN 1600 Required accuracy: <0.3% o.r. Measured variable: Volume flow
Liquids in general Water Additives, flavors Oil, hydrocarbons Molasse, shower gel Liquid O2 , Ar, N2 Water, milk, solvents Chemicals

I  n the table on Page 4, select the category for the type of fluid in question: Liquid Gas Steam Special applications. I  n the table, specify all the requirements (see Point ) you defined for your measuring point. N  ow select the measuring principles to which all the requirements apply ().

Liquids

Flow rates < 2 l/h Non-conductive liquids Viscous liquids > 500 mPas Cryogenic fluids

Liquids

Liquids in general Volume flow Nominal diameter DN > 500 Electromagnetic Ultrasonic Differential pressure

Volume flow Mass flow Accuracy 0.2% Nominal pipe diameter DN > 500 Operable flow range > 1:30 Pressure loss negligible Density, concentration, viscosity metering Process temperature > 200 C Process pressure > 100 bar

Foods, shower gel Crude oil

Selecting the flowmeter


N  ow open out this brochure completely. The overview contains all the measuring principles you selected under Point and the corresponding Endress+Hauser flowmeters.

Electromagnetic Ultrasonic Differential pressure Promag W Prosonic Flow W (clamp-on) Prosonic Flow W (insertion) Prosonic Flow P Prosonic Flow C Deltatop/Deltaset DPP 50 (W) 53 (W) Promag 50W (standard) Promag 53W (high-end)

Promag W Promag P

Diameter
DN 25 to 2000 (1" to 78")

Measuring range
0 to 10 m/s 0 to 110,000 m/h Basic

Combinable with
Sensor W, P, H

(best possible accuracy) 0.5%

Accuracy

Proline Promag

Water/Wastewater water Water/Waste

10

Now mark the sensors that can be installed in the pipe of your choice
(DN = 1600 mm).

Electromagnetic (EMF)

Chemical/Process applications

DN 15 to 600 (" to 24")

0 to 10 m/s 0 to 9600 m/h Standard Standard 0 to 10 m/s 0 to 282 m/h High-end High-end 0 to 10 m/s 0 to 9600 m/h Custody transfer

50

Sensor W, P, H

0.5% Optional: 0.2%

Promag H
Hygiene

DN 2 to 100 (1/12" to 4")

53 51

Sensor W, P, H

0.2%

Promag S

Inhomogeneous fluids High solid content

DN 15 to 600 (" to 24")

Sensor W, P

0.5% Optional: 0.2%

M  ark those transmitters that can be combined with the sensors in question and that meet the required accuracy specifications (<0.3% o.r.)


Prosonic Flow U
DN 15 to 100 (" to 4")

23 55
Clamp-on sensor Pharma, semiconductor ind. 0 to 10 m/s 0 to 424 m/h Basic

Two-wire

Sensor P, H

0.5%

 ou have found the two flowmeters that suit your measuring point. Y

Special applications

Sensor S

0.2%

91 90 93

Sensor W

Onsite: <2.0% Under reference conditions: 0.5% Onsite: <2.0% (U, W, P) Under reference conditions: 0.5% 0.5% (C) Onsite: <2.0% Under reference conditions: 0.5% Liquids: For 2 or 3 beams 0.5% For 4 beams 0.3% Onsite: <2.0% Under reference conditions: 0.5%

Prosonic Flow W
Clamp-on sensor Standard applications

DN 50 to 4000 (2" to 156")

0 to 15 m/s 0 to 678,550 m/h Standard 0 to 15 m/s 0 to 678,550 m/h High-end 0 to 15 m/s 0 to 678,550 m/h Inline

Sensor U, W, P

Proline Prosonic Flow

Ultrasonic

Prosonic Flow W
Insertion sensor Standard applications

DN 200 to 4000 (8" to 156")

Sensor U, W, P, C

Prosonic Flow P
Clamp-on sensor Process applications

DN 50 to 4000 (2" to 156")

92 F 92 T

Sensor F

Prosonic Flow C
Inline sensor Water/Waste water

DN 300 to 2000 (12" to 78")

0 to 10 m/s 0 to 47,400 m/h Portable 0 to 10 m/s 0 to 600 m/h Liquids: 0 to 18,280 m/h Gases: 0 to 169,650 m/h Steam: 0 to 169,650 m/h

Sensor U, W, P

Prosonic Flow F
Inline sensor Two-wire technology

DN 25 to 150 (1" to 6")

Diff. pressure

Deltatop/-set DPO
Orifice plate

DN 10 to 1000 (" to 40")

Deltatop/-set DPP
Pitot tube

DN 40...12000 (1" to 480")

Liquids: 0 to 327,500 m/h Gases: 0 to 9,767,000 m/h Steam: 0 to 4,400,000 m/h

Deltabar S PMD 70/75

Deltatop PMD 70/75 Deltaset PMD 70/75

Orifice plate DPO: 1.0%

Pitot tube DPP: 1.5%

Selection table Measuring principles


Requirements Fluids (examples)
Electromagnetic

Supplementary documentation
Differential pressure Ultrasonic Thermal

Measuring principle
Coriolis Vortex

Liquids in general Flow rates < 2 l/h Non-conductive liquids Viscous liquids > 500 mPas Cryogenic fluids

Water Additives, flavors Oil, hydrocarbons Molasse, shower gel Liquid O2 , Ar, N2 Water, milk, solvents Chemicals

Liquids

Volume flow Mass flow Accuracy 0.2% Nominal pipe diameter DN > 500 Operable flow range > 1:30 Pressure loss negligible Density, concentration, viscosity metering Process temperature > 200 C Process pressure > 100 bar Gas flow in general Contaminated gas Volume flow Mass flow


Flow handbook A practical guide: Measurement technologies applications solutions ISBN 3-9520220-4-7 (Order number 50104766) Flow measurement For liquids, gases and steam Products and services at a glance FA005D/06/en

Foods, shower gel Crude oil Natural gas, air Biogas, exhaust gas Exhaust gas CO2 Rare gas (e.g. argon)

Gas

Accuracy 0.5% Nominal pipe diameter DN > 500 Operable flow range > 1:30 Pressure loss negligible Process temperature > 200 C Process pressure > 100 bar Volume flow Mass flow Nominal pipe diameter DN > 300 Operable flow range > 1:10 Pressure loss negligible Process temperature > 300 C Process pressure > 100 bar Hygienic applications (SIP/CIP) Slurries, suspended solids

Air Exhaust gas CNG

Steam

Overheated steam Overheated steam Milk, beer Thick sludge Oil/water, emulsions Ore slurry

Special applications

Liquid/liquid mixtures Highly abrasive fluids (e.g. in mining) Bidirectional metering (forward/reverse) Measurement from outside without process interruption/retrofitting/temporary metering Filling/bottling applications Flow monitoring

Metering flow the cost-optimised way

Flow measurement in utilities Energy management in your hands customized solutions for compressed air, gas, steam and water CP007D/06/en

Comprehensive metering point solutions for planning, commissioning and maintenance CP006D/06/en

see Flow measurement OEM (inside) see Flow monitoring (inside)

Please note: The selection table above provides a rough overview and is used for quick clarification of suitable measuring principles. The table is not exhaustive. If further information has to be clarified, feel free to contact Endress+Hauser any time. Well be glad to be of assistance.

Applicator Optimal planning and sizing


The first steps are always the most important when it comes to planning metering points. An optimally dimensioned flowmeter at the right location provides many advantages: Consistent process quality Diminished risk of instrument failure at a later time Product savings due to error-free, precise measurement Proven since more than 20 years Applicator is a software for selecting and specifing flowmeters that has proven its abilities over many years. A wealth of industry experience and specialized know-how is embedded in Applicator. It enables people with limited technical knowledge to specify metering points quickly and easily, without timeconsuming research. The most recent version of Applicator has a raft of new functions, including selection functions for all the other areas covered by Endress+Hauser namely level, pressure, temperature, liquid analysis, etc. Youll find that it opens up an enormous range of application possibilities in metrology and control.

W@M

W@M Life cycle management from Endress+Hauser is an open information system which provides technical and operative management with complete

data flow and archival from commissioning on through to maintenance and service within a plant, at any time or location. This is attained within an open

system based on Internet/intranet technology, and incorporating all software programs, products and services from Endress+Hauser.

Fieldcheck  Simulation and testing of alarm/ process conditions directly in the plant (without flow)  Verification of flowmeters directly in the piping (without removal)

Applicator  Optimal planning and sizing of flowmeters  Administration and documentation for entire plant projects

Selection straight to your objective When starting out to select the ideal flowmeter, Applicator guides through all the important parameters. Just enter the type of fluid, requisite process data or metering point attributes and Applicator already displays the measuring principles and families of meters best suited to the case in hand.

FieldCare  Easy and dependable commissioning and configuring of measuring devices via PC  Complete documentation of device settings  Storing device data (upload/download)  Simplified trouble-shooting

 Process-optimization through the visualization of measurement data in realtime (line plotting function)  Analysis of diagnostic data for maintenance purposes  Administration and analysis of verification results from the Fieldcheck test instrument

Project documentation data collection, archival, export The use of additional expansion modules permits cost-saving administration and documentation of data for entire plant projects:  Clear presentation in a tree structure of any desired number of metering points (e.g. by building, floor, metering point), according to ISA 88.  Uncomplicated inclusion of measuring devices directly from the Selection and Sizing functions. Also possible for instruments from other manufacturers or projects.  Detailed listing of project management data (responsible person, costing and budget data, etc.)  Multiple data exchange possibilities in Excel or XML format: data export to other programs, data import from other projects, etc.

Sizing safe specification By entering further application-specific values, such as process pressure, process temperature or measuring range, Applicator immediately determines the ideal flowmeter type and along with the suitable nominal diameter, working range, pressure loss, accuracy, material loads, conformity with the Pressure Equipment Directives (PED) or corrosion resistance. Messages and warnings make it easy to select the right meter, help avoid common mistakes, and enable optimization of the measuring point to suit specific requirements.

Flow measurement
Promag W Promag P
Water/Waste water

Sensors
Diameter
DN 25 to 2000 (1" to 78")

Transmitters
Measuring range
0 to 10 m/s 0 to 110,000 m/h

Pressure PN
PN 6 to 40, Cl 150 to 300 JIS 10 to 20 K AWWA Class D AS Table E, PN 16 PN 10 to 40 Cl 150 to 300 JIS 10 to 20 K AS Table E, PN 16 PN 16 to 40 Cl 150 JIS 10 to 20 K PN 10 to 40 Cl 150 to 300 JIS 10 to 20 K AS Table E, PN 16

Process temperature
20 to +80 C

Materials (wetted parts)


Hard rubber liner Polyurethane liner

Process connection

Protection
IP67/IP68 (NEMA 4X/6P)

Hygienic certification

Combinable with

Housings
Compact Alu Remote Alu Compact SS Compact Alu Remote Alu Compact SS Compact Alu Remote Alu Compact Alu Remote Alu

Display/Operation
Two-line LCD/ push buttons Operation via FieldCare Two-line, backlit LCD/ push buttons Operation via FieldCare Four-line, backlit LCD/ touch control Operation via FieldCare Two-line, backlit LCD/ push buttons Operation via FieldCare Four-line LCD/ touch control Operation via FieldCare Four-line, backlit LCD/ touch control Operation via FieldCare Two-line, backlit LCD/ Operation via HART resp. FieldCare Two-line, backlit LCD/ push buttons Operation via FieldCare Four-line, backlit LCD/ touch control Operation via FieldCare Four-line, backlit LCD/ touch control Operation via FieldCare

Ambient temperature
20 to +60 C

Power supply
85 to 250 V AC (45 to 65 Hz) 20 to 28 V AC (45 to 65 Hz) or 11 to 40 V DC

Galvanic isolation
Electronical connections are galvanically isolateted from each other

Inputs

Outputs

Communication

(best possible accuracy) 0.5%

Accuracy

Ex-approvals

Protection
IP67 (Nema 4X)

Flange: DIN, ANSI, JIS, AWWA, AS

10
Basic

Sensor W, P, H

4 to 20 mA, pulse, status

HART

FM, CSA Div. 2

Proline Promag

Electromagnetic (EMF)

Chemical/Process applications

DN 15 to 600 (" to 24")

0 to 10 m/s 0 to 9600 m/h

40 to +180 C

PFA liner PTFE liner

Flange: DIN, ANSI, JIS, AS Flange: DIN, ANSI, JIS, external thread, internal thread, weld connection, clamp connection, couplings Flange: DIN, ANSI, JIS, AS

IP67/IP68 (NEMA 4X/6P)

50 53 51 23 55

Standard

Sensor W, P, H

20 to +60 C Optional: 40 to +60 C

85 to 260 V AC (45 to 65 Hz) 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC

Electronical connections are galvanically isolateted from each other

Status

4 to 20 mA, pulse/frequency (1 kHz passive), status

HART, PROFIBUS PA, PROFIBUS DP HART, PROFIBUS PA, PROFIBUS DP, FOUNDATION Fieldbus, MODBUS RS485

0.5% Optional: 0.2%

ATEX, FM, CSA, TIIS, NEPSI In preparation: IEC

IP67 (Nema 4X)

Promag H
Hygiene

DN 2 to 100 (1/12" to 4")

0 to 10 m/s 0 to 282 m/h

20 to +150 C

PFA liner

IP67 (NEMA 4X)

3-A, EHEDG

High-end

Sensor W, P, H

20 to +60 C Optional: 40 to +60 C

85 to 260 V AC (45 to 65 Hz) 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC

Electronical connections are galvanically isolateted from each other

Status, current input

4 to 20 mA, pulse/frequency (10 kHz, active/passive), relays, status

0.2%

ATEX, FM, CSA, TIIS, NEPSI In preparation: IEC

IP67 (Nema 4X)

Promag S

Inhomogeneous fluids High solid content

DN 15 to 600 (" to 24")

0 to 10 m/s 0 to 9600 m/h

40 to +180 C

PTFE, PFA, polyurethane, natural rubber liner

IP67/IP68 (NEMA 4X/6P)

Custody transfer

Sensor W, P

20 to +60 C

85 to 260 V AC (45 to 65 Hz) 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC Two-wire 12 to 30 V DC 13.9 to 30 V DC (Ex) (45 to 65 Hz) 20 to 260 V AC/DC (45 to 65 Hz)

Electronical connections are galvanically isolateted from each other

Status

4 to 20 mA, pulse/frequency (1 kHz passive), status

HART

0.5% Optional: 0.2%

ATEX In preparation: IEC

IP67 (Nema 4X)

Two-wire

Sensor P, H

Compact Alu

20 to +60 C

Electronical connections are galvanically isolateted from each other

4 to 20 mA, pulse/frequency, status

HART

0.5%

ATEX, FM, CSA

IP67 (Nema 4X)

Special applications

Sensor S

Compact Alu Remote Alu

20 to +50 C Optional: 40 to +50 C

Electronical connections are galvanically isolateted from each other

Status, current input

4 to 20 mA, pulse/frequency (10 kHz, active/passive), relays, status

HART, PROFIBUS PA, FOUNDATION Fieldbus

0.2%

FM, CSA Div. 2 In preparation: ATEX II 3G, IEC Zone 2

IP67 (Nema 4X)

Promass F

Universal flowmeter

DN 8 to 250 (5/16" to 10") DN 25, 50, 80 (HT) DN 8 to 80 (5/16" to 3") DN 8 to 25 (HP) DN 1 to 4 (1/24" to 5/32")

0 to 2,200,000 kg/h

PN 16 to 100 Cl 150 to 600 JIS 10 to 63 K PN 40 to 100, PN 350 (HP) Cl 150 to 600 JIS 10 to 63 K PN 16 to 400 Cl 150 to 600 JIS 10 to 63 K PN 40 to 100 Cl 150 to 600 JIS 10 to 63 K PN 40 Cl 150 to 300 JIS 10 to 20 K PN 40 to 100 Cl 150 to 600 JIS 10 to 63 K PN 40 to 63 Cl 150 to 300 JIS 20 to 40 K PN 40 to 63 Cl 150 to 300 JIS 20 to 40 K No limit Non-invasive

50 to +200 C 50 to +350 C (HT)

904L (1.4539)/ 316L (1.4404) Alloy C-22 (2.4602) DN 8...50: Titanium grade 9 DN 80: Titanium grade 2

Flange: DIN, ANSI, JIS, Tri-Clamp, couplings, VCO

IP67 (NEMA 4X)

3-A, EHEDG

40
Basic

Sensor E Sensor F, M, A, I, H, E, S, P Sensor F, M, A, I, H, E, S, P Sensor F, M, A

Compact Alu

20 to +60 C Optional: 40 to +60 C

85 to 260 V AC (45 to 65 Hz) 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC

Electronical connections are galvanically isolateted from each other

Status

4 to 20 mA, frequency, status

HART

Liquids: 0.5% Gases: 1.0%

ATEX, IEC, FM, CSA, TIIS, NEPSI

IP67 (Nema 4X)

Promass M
High pressure

0 to 180,000 kg/h

50 to +150 C

Flange: DIN, ANSI, JIS, Tri-Clamp, couplings

IP67 (NEMA 4X)

3-A

80 83 84

Standard

Compact SS Compact Alu Remote Alu Compact SS Compact Alu Remote Alu Compact SS Compact Alu Remote Alu

20 to +60 C Optional: 40 to +60 C

85 to 260 V AC (45 to 65 Hz) 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC

Electronical connections are galvanically isolateted from each other

Status

4 to 20 mA, pulse/frequency (1 kHz passive), status

HART, PROFIBUS PA, PROFIBUS DP HART, PROFIBUS PA, PROFIBUS DP, FOUNDATION Fieldbus, MODBUS RS485 HART MODBUS RS485

Liquids: 0.15% Gases: 0.35%

ATEX, IEC, FM, CSA, TIIS, NEPSI

IP67 (Nema 4X)

Promass A
Low flow rates

0 to 450 kg/h

50 to +200 C

904L (1.4539) Alloy C-22 (2.4602)

Tri-Clamp, VCO

IP67 (NEMA 4X)

3-A, EHEDG

High-end

20 to +60 C Optional: 40 to +60 C

85 to 260 V AC (45 to 65 Hz) 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC

Electronical connections are galvanically isolateted from each other

Status, current input

4 to 20 mA, pulse/frequency (10 kHz, active/passive), relays, status 4 to 20 mA, pulse/frequency (10 kHz, active/passive), 2pulse (90/180), relays, status

Liquids: 0.1% Gases: 0.35 %

ATEX, IEC, FM, CSA, TIIS, NEPSI

IP67 (Nema 4X)

Proline Promass

Coriolis

Promass I

Hygiene, viscosity

DN 8 to 80 (5/16" to 3")

0 to 180,000 kg/h

50 to +150 C

Measuring tube: Titanium grade 9 Flange: Titanium grade 2

Flange: DIN, ANSI, JIS, Tri-Clamp, couplings

IP67 (NEMA 4X)

3-A, EHEDG

Custody transfer

20 to +60 C Optional: 40 to +60 C

85 to 260 V AC (45 to 65 Hz) 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC

Electronical connections are galvanically isolateted from each other

Status

Liquids: 0.1% Gases: 0.35 %

ATEX, IEC, FM, CSA, TIIS, NEPSI

IP67 (Nema 4X)

Promass H

Chemically aggressive fluids

DN 8 to 50 (5/16" to 2")

0 to 70,000 kg/h

50 to +200 C

Zirconium 702/R 60702

Flange: DIN, ANSI, JIS

IP67 (NEMA 4X)

Promass E

Basic applications

DN 8 to 50 (5/16" to 2")

0 to 70,000 kg/h

40 to +125 C

904L (1.4539) 316L (1.4404)

Flange: DIN, ANSI, JIS, Tri-Clamp, couplings, VCO

IP67 (NEMA 4X)

Promass S

Food & Beverages

DN 8 to 50 (5/16" to 2")

0 to 70,000 kg/h

50 to +150 C

904L (1.4539) 316L (1.4435)

Flange: DIN, ANSI, JIS, Tri-Clamp, couplings

IP67 (NEMA 4X)

3-A, EHEDG

Promass P

Life Sciences Industries (Pharma, biotechnology)

DN 8 to 50 (5/16" to 2")

0 to 70,000 kg/h

50 to +200 C

316L (1.4435)

Flange: DIN, ANSI, JIS, Tri-Clamp, couplings

IP67 (NEMA 4X)

3-A, EHEDG, ASME BPE

Prosonic Flow U

Clamp-on sensor Pharma, semiconductor ind.

DN 15 to 100 (" to 4")

0 to 10 m/s 0 to 424 m/h

20 to +80 C

not wetted (Clamp-on sensor)

Clamp-on sensor

IP52

91
Basic

Sensor W

Remote Alu

Two-line LCD/ push buttons Operation via FieldCare Two-line, backlit LCD/ push buttons Operation via FieldCare Four-line, backlit LCD/ touch control Operation via FieldCare Two-line LCD/ push buttons Operation via FieldCare Graphic display/ push buttons

20 to +60 C

85 to 250 V AC (45 to 65 Hz) 20 to 28 V AC (45 to 65 Hz) or 11 to 40 V DC

Electronical connections are galvanically isolateted from each other

4 to 20 mA, pulse, status

HART

Onsite: <2.0% Under reference conditions: 0.5% Onsite: <2.0% (U, W, P) Under reference conditions: 0.5% 0.5% (C)

FM, CSA Div. 2

IP67 (Nema 4X)

Prosonic Flow W
Clamp-on sensor Standard applications

DN 50 to 4000 (2" to 156")

0 to 15 m/s 0 to 678,550 m/h

No limit Non-invasive

20 to +80 C

not wetted (Clamp-on sensor)

Clamp-on sensor

IP67 (NEMA 4X)

90

Standard

Sensor U, W, P

Remote Alu

20 to +60 C Optional: 40 to +60 C

85 to 260 V AC (45 to 65 Hz) 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC

Electronical connections are galvanically isolateted from each other

Status

4 to 20 mA, pulse/frequency (1 kHz passive), status 4 to 20 mA, pulse/frequency (10 kHz, active/passive), relays, status 4 to 20 mA, pulse

HART, PROFIBUS PA

FM, CSA Div. 2

IP67 (Nema 4X)

Proline Prosonic Flow

Ultrasonic

Prosonic Flow W
Insertion sensor Standard applications

DN 200 to 4000 (8" to 156")

0 to 15 m/s 0 to 678,550 m/h

PN 16

20 to +80 C

316L (1.4404)

Welding connector

IP67 (NEMA 4X)

93
Inline

High-end

Sensor U, W, P, C

Remote Alu

20 to +60 C Optional: 40 to +60 C

85 to 260 V AC (45 to 65 Hz) 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC 12 to 36 V DC, EEx i: 12 to 30 V DC, EEx d: 15 to 36 V DC Rechargeable battery: 100 to 240 V AC (47 to 63 Hz)

Electronical connections are galvanically isolateted from each other

Status, current input

HART, PROFIBUS PA, FOUNDATION Fieldbus

Onsite: <2.0% Under reference conditions: 0.5% Liquids: For 2 or 3 beams 0.5% For 4 beams 0.3% Onsite: <2.0% Under reference conditions: 0.5%

ATEX, FM, CSA, TIIS, NEPSI In preparation: IEC

IP67 (Nema 4X)

Prosonic Flow P
Clamp-on sensor Process applications

DN 50 to 4000 (2" to 156")

0 to 15 m/s 0 to 678,550 m/h

No limit Non-invasive

40 to +170 C

not wetted (Clamp-on sensor)

Clamp-on sensor

IP68 (NEMA 6P)

92 F 92 T Portable

Sensor F

Compact Alu Remote Alu

Compact: 40 to +70 C Remote: 40 to +80 C

Electronical connections are galvanically isolateted from each other

HART, PFM, Profibus PA

ATEX, FM, CSA In preparation: IEC, TIIS, NEPSI

IP67 (Nema 4X)

Prosonic Flow C
Inline sensor Water/Waste water

DN 300 to 2000 (12" to 78")

0 to 10 m/s 0 to 47,400 m/h

PN 16 CI 150

20 to +60 C

Carbon steel, epoxy coated

Flange: DIN, ANSI, AWWA

IP68 (NEMA 6P)

Sensor U, W, P

Remote Alu

10 to +45 C

Electronical connections are galvanically isolateted from each other

4 to 20 mA

4 to 20 mA

Serial interface (RS 232)

IP50

Prosonic Flow F
Inline sensor Two-wire technology

DN 25 to 150 (1" to 6")

0 to 10 m/s 0 to 600 m/h

PN 40 CI 150 PN 10 to 250 Cl 150 to 1500 JIS 10 to 40 K PN 10 to 40 Cl 150 to 300 JIS 10 to 20 K

40 to +150 C

Meter body: CF3M (1.4404) Sensor: 316/3166 (1.4404) Flange: F316/F316L (1.4404) Meter body: CF3M (1.4404), Alloy C-22 (2.4602) DSC sensor: 316L (1.4435), Inconel 718, Alloy C-22 (2.4602) Meter body: CF3M (1.4404) DSC sensor: 316L (1.4435)

Flange: DIN, ANSI, JIS

IP68 (NEMA 6P)

Proline Prowirl

Prowirl F

Vortex

Flanged version

DN 15 to 300 (" to 12")

Liquids: 0.16 to 2360 m/h Gases: 3 to 19,700 m/h

200 to +400 C Optional: 200 to +450 C 200 to +400 C Optional: 200 to +450 C

Flange: DIN, ANSI, JIS Optional: with integrated reduction of line size

IP67 (NEMA 4X)

72

Standard

Sensor F, W

Compact Alu Remote Alu

Two-line LCD/ push buttons Operation via FieldCare Two-line LCD/ push buttons Operation via FieldCare Two-line, backlit LCD/ push buttons Operation via FieldCare

Compact: 40 to +70 C Remote: 40 to +80 C

Prowirl W
Wafer version

DN 15 to 150 (" to 6")

Liquids: 0.19 to 625 m/h Gases: 4 to 5210 m/h

Wafer: DIN, ANSI, JIS

IP67 (NEMA 4X)

73

With flow computer

Sensor F, W

Compact Alu Remote Alu

Compact: 40 to +70 C Remote: 40 to +80 C

Two-wire 12 to 36 V DC, EEx i: 12 to 30 V DC, EEx d: 15 to 36 V DC Two-wire 12 to 36 V DC, EEx i: 12 to 30 V DC, EEx d: 15 to 36 V DC 85 to 260 V AC 20 to 55 V AC or 16 to 62 V DC

Electronical connections are galvanically isolateted from each other

Pressure using HART, PROFIBUS PA, FF Temperature, density using HART

4 to 20 mA, pulse/status 4 to 20 mA, pulse/status, frequency 4 to 20 mA, pulse/frequency, relays, status

HART, PFM, Profibus PA, Foundation Fieldbus

Liquids: 0.75% Volume Gases/steam: 1.0% Volume Liquids: 0.75% Volume Gases/steam: 1.0% Volume Saturated steam: 1.7% Mass

ATEX, FM, CSA, TIIS, NEPSI In preparation: IEC

IP67 (Nema 4X)

Electronical connections are galvanically isolateted from each other

HART, PFM, Profibus PA, Foundation Fieldbus

ATEX, FM, CSA, TIIS, NEPSI In preparation: IEC

IP67 (Nema 4X)

Proline t-mass

Thermal

t-mass I

Insertion sensor

DN 80 to 1500 (3" to 60")

Gases: 0 to 70 m/s

PN 20

40 to +130 C

316L (1.4404) Alloy C-22

Thread: NPT, BSP

IP67

65

Standard

Sensor I, F

Compact Alu Remote Alu

20 to +60 C Optional: 40 to +60 C

Electronical connections are galvanically isolateted from each other

HART, PROFIBUS DP, MODBUS RS485

Gases: 1.5%

ATEX, FM, CSA, NEPSI In preparation: IEC, TIIS

IP67

t-mass F

Flanged version

DN 15 to 100 (" to 4")

Gases: 0 to 70 m/s

PN 40 Cl 300

40 to +100 C

316L (1.4404) Alloy C-22 C22.8 316Ti (1.4571) 15Mo3 (1.5415) C22.8 316Ti (1.4571)

Flange: DIN, ANSI, JIS

IP67

Diff. pressure

Deltatop/-set DPO
Orifice plate

DN 10 to 1000 (" to 40")

Liquids: 0 to 18,280 m/h Gases: 0 to 169,650 m/h Steam: 0 to 169,650 m/h Liquids: 0 to 327,500 m/h Gases: 0 to 9,767,000 m/h Steam: 0 to 4,400,000 m/h

PN 160/PN 420

200 to +1000 C

Wafer (corner tapping) Flange (flange tapping) Pipe unit (meter run) Cutting ring, flange, flow tap

IP65

Deltatop/-set DPP
Pitot tube

DN 40...12000 (1" to 480")

Deltabar S PMD 70/75


IP65

PN 160/PN 420

200 to +1000 C

Deltatop PMD 70/75 Deltaset PMD 70/75

Compact SS Remote SS Compact Alu Remote Alu

Orifice plate DPO: 1.0% 4-digits LCD/ push buttons Operation via FieldCare 40 to +85 C Two-wire 11.5 to 45 V DC No galvanic isolation 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus Pitot tube DPP: 1.5% ATEX, IEC, FM, CSA, TIIS, NEPSI IP65

Flow measurement OEM


EMF
Dosimag
Filling/bottling Conductive liquids DN 4, DN 8, DN 15K, DN 15 (5/32", 5/16", ") DN 8 to 25 (5/16" to 1") 0 to 10 m/s 0 to 1.66 l/s PN 16 40 to +125 C PFA liner Tri-Clamp, weld connection IP67 (NEMA 4X) 3-A, EHEDG

Compact version

Compact SS

Operation via FieldCare

20 to +60 C

20 to 30 V DC

Electronical connections are galvanically isolateted from each other

Pulse, status

0.5% Optional: 0.25%

IP67 (NEMA 4X)

Dosimass

Filling/bottling Liquids

0 to 5 kg/s

PN 16

40 to +125 C

904L (1.4539) 316L (1.4404)

Tri-Clamp, couplings

IP67 (NEMA 4X)

3-A

Compact version

Compact SS

Operation via FieldCare

20 to +60 C

20 to 30 V DC

Electronical connections are galvanically isolateted from each other

Pulse, status

0.15%

ATEX II 3G

IP67 (NEMA 4X)

Coriolis

CNGmass

Fueling Compressed Natural Gas

DN 15, DN 25 (", 1")

0 to 150 kg/min

G" and G1" for max. 350 bar PN 40 Cl 150 to 300 JIS 10 to 63 K

50 to +125 C

316L (1.4404) 316L (1.4435)

Internal thread: G" (DN 15) G1" (DN 25) Flange: DIN, ANSI, JIS, VCO

IP67 (NEMA 4X)

Compact version

Compact Alu

Operation via FieldCare

40 to +60 C

20 to 28 V AC 20 to 30 V DC

Electronical connections are galvanically isolateted from each other

Pulse/frequency, status

MODBUS RS485

0.5%

ATEX, FM, CSA In preparation: NEPSI, TIIS, IEC

IP67 (NEMA 4X)

LPGmass

Fueling Liquified Petroleum Gas

DN 8 to 50 (5/16" to 2")

0 to 70,000 kg/h

40 to +125 C

316L (1.4404) 904L (1.4539)

IP67 (NEMA 4X)

Compact version

Compact Alu

Operation via FieldCare

40 to +60 C

20 to 28 V AC 20 to 30 V DC

Electronical connections are galvanically isolateted from each other

Pulse/frequency, status

MODBUS RS485

0.2%

ATEX, FM, CSA In preparation: NEPSI, TIIS, IEC

IP67 (NEMA 4X)

Flow monitoring
EMF
Magphant
Limit switch/Trend display Conductive liquids

DN 1 to 2000 (" to 78")

0 to 5 m/s

PN 16

20 to +120 C

PVDF

Weld nipple

IP66 (NEMA 4X)

Compact version

Compact Alu

LED

20 to +60 C

20 to 30 V DC

No galvanic isolation

4 to 20 mA, relays 1 or 2PNP switching outputs, optional: 1PNP switching output and 14 to 20 mA output, active 1 or 2PNP switching outputs, optional: 1PNP switching output and 14 to 20 mA output, active

2.0%

FM, CSA Div. 2

IP66 (NEMA 4X) IP65/IP66, dependent on electrical connection IP65/IP66, dependent on electrical connection
Subject to modification

Flowphant DTT31
Standard applications Liquids

DN 8 to 2000 (5/16" to 78")

0.03 to 3 m/s as a relative value between 0 and 100% 0.03 to 3 m/s as a relative value between 0 and 100% Liquids: 0 to 3 m/s Gases: 0 to 50 Nm/s

max. PN 100

20 to +85 C

AISI 316L

Weld nipple, compression fitting, threaded connection: ANSI, ISO

IP65/ IP66

Compact version

Compact SS

One-line, backlit LCD Color change on faults, LEDs for status indication, push buttons One-line, backlit LCD Color change on faults, LEDs for status indication, push buttons LED, 4-digits, LCD (option) DIL switches

40 to +85 C

10 to 30 V DC (reverse polarity protection)

No galvanic isolation

Programmable by a PC (ReadWin 2000)

2.0%

Flowphant DTT35

Thermal

Hygienic applications Liquids

DN 8 to 2000 (5/16" to 78")

max. PN 100

20 to +85 C

AISI 316L and FKM Viton, EPDM (sealings)

Conical metal-metal G" Adapter concept: Tri-Clamp

IP65/ IP66

Compact version

Compact SS

40 to +85 C

10 to 30 V DC (reverse polarity protection)

No galvanic isolation

Programmable by a PC (ReadWin 2000)

2.0%

t-switch ATT11
Limit switch Liquids, gases

DN 40 to 1000 (1" to 40")

PN 25

10 to +80 C

316L (1.4404/1.4435)

Threaded connection: BSP, NPT, couplings, Tri-Clamp

IP66 (NEMA 4X)

Compact version

Compact SS Compact Alu

10 to +65 C without LCD

18 to 30 V AC/DC

No galvanic isolation

Relays

5.0%

IP66

t-trend ATT12
Trend display Liquids, gases

DN 40 to 1000 (1" to 40")

Liquids: 0 to 3 m/s Gases: 0 to 50 Nm/s

PN 25

10 to +80 C

316L (1.4404/1.4435)

Threaded connection: BSP, NPT, couplings, Tri-Clamp

IP66 (NEMA 4X)

Compact version

Compact SS Compact Alu

LED, 4-digits, LCD (option) push buttons

10 to +65 C without LCD

18 to 30 V AC/DC

No galvanic isolation

4 to 20 mA, pulse, open collector

5.0%

IP66

www.endress.com