Sie sind auf Seite 1von 56

Installation, Operation and Maintenance Manual TWC Series

Water Treatment System

Manual
Controller

Installation Maintenance Repair Manual

Advantage Controls P.O. Box 1472 Muskogee, OK 74402 Phone: 800-743-7431 Fax: 888-686-6212 www.advantagecontrols.com email: support@advantagecontrols.com

1/2011

MegaTronSS Controller Instruction & Maintenance Manual Table of Contents


Contents Page I. Introduction. ................................................................................. 3 Model Numbering........................................................................ 3 Description of Unit....................................................................... 3 II. III. IV. V. VI. VII. VIII. Installation................................................................................... 4 Electrical Wiring. .......................................................................... 4 Wiring Diagrams.......................................................................... 5 Mounting Instructions.................................................................. 8 Electrode Installation................................................................... 8 A. Cooling Tower........................................................................ 8 B. Boiler.................................................................................... 10 Front Panel Drawing..................................................................................... 12 Description. ................................................................................ 12 System Operation Overview. ..................................................... 13 Start-Up..................................................................................... 13 Calibration................................................................................. 13 Menu Navigation....................................................................... 14 Maintenance.............................................................................. 27 Electrode Cleaning Procedures. ................................................ 27 Troubleshooting. ........................................................................ 28 Warranty & 30 Day Billing Memo Policy.................................... 29

I.

Introduction

The MegaTronSS microprocessor based controllers are designed to provide a wide range of control functions for recirculating water treatment systems. The controller is programmed through a front panel keypad and can be configured to provide a customized control system for your application. Your particular units functions can be determined by comparing the units model number to the Model Numbering table listed below.

Model Numbering
MegaTronSS units have several base system control functions and unit optional features. Your unit may be supplied with one or more of the features described in this manual. To determine what features apply to your unit check the model number label located on the controller enclosure. Base System Control Functions C - Tower Conductivity Control B - Boiler / Condensate Conductivity B1- Boiler RTD Conductivity Control B2- Boiler No Temp Conductivity P - pH Control Q - Dual pH Control R - ORP Control F1 to F5- Chemical Feed Timers E - Flow Switch Whole Unit Optional Features A - Conduit Connections D - Dry Contact Relays H1- Internet Connect Communications H4- Internet Phone Modem Communications H5- Internet with Cell Modem Router N1-2 - 4-20mA inputs O1-3 - 4-20mA outputs S - Saturation Index (must have pH) W3 - Auxiliary Flow Meter Inputs Y - ETL Agency Listing / Approvals

Model numbers start SS followed by the System Control functions. A dash separates the whole unit options listed after all base system control functions. Example: SSCPF3E-N2. Notice: Your unit may not have all features and functions described in this manual. This list represents our most popular options, additional option codes are available.

Description of Unit
MegaTronSS controllers control a single recirculating water system including cooling tower and boiler applications and may have various features depending on the model number.

Control Functions
Each of these control functions are based on an analog input from a probe and will include user settable relay control settings along with a High and Low Alarm setting and Limit Timer. Each control function will include a control relay output. When the reading reaches the Set Point the control relay is activated until the reading changes by the Differential amount. 1. System Conductivity - The conductivity function of the controller is designed to monitor and control Total Dissolved Solids (TDS) in a recirculating system like a cooling tower or boiler in terms of electrical conductivity measured in MicroSiemens/cm. This control function is also referred to as Bleed. pH - The pH function monitors and controls pH on a scale of 0-14 pH units. ORP - The ORP function monitors and controls ORP on a scale of +/- 1000 mV.

2. 3.

Chemical Feed Timers


Chemical feed timers are designed to automate the addition of various chemicals by activating a relay output. Multiple timers can be supplied depending upon the model number and each timer will include a relay output. All timers can be programmed to be one of the following types. 1. 2. 3. 4. 5. Pulse Time - Accepts pulses from a make-up water meter (supplied separately). It can accumulate 1-9999 gallons before activating the timer to run. Feed with Bleed - Activates the relay output simultaneously with the bleed and limits the amount of time the relay output will be on during the bleed cycle. Feed after Bleed - Activates the relay output based on a user defined percentage of time based on the bleed off, another timer or digital input. The relay is activated after a bleed cycle and runs for the set percentage of that bleed cycle. Percentage - The relay is on for a percentage of a continuously repeating cycle time that can be set from 1 minute to 99 hours and 59 minutes. 28 Day - The timer is based on a 28 day cycle with four independent programmable feed cycles with prebleed and bleed lockout settings.

II.

Installation

Electrical Wiring
The MegaTron SS controller has an internal regulated fused power supply that will operate off of 90 to 250 VAC at 47 to 63 Hz on the incoming wiring. Each output relay is individually protected with a replaceable fuse. Relay outputs will equal incoming line voltage. WARNINGS: 1. The controller should be connected to its own isolated circuit breaker, and for best results, the ground should be a true earth ground, not shared. Wiring must be done according to all applicable local codes. 2. Power (line voltage) must be disconnected while making any connections. If power is supplied to the unit, line voltage will be present on the relay cards located in the lower section of the enclosure, even with the power (ON/OFF) switch off. 3. Low voltage signal wires (probes, flow switch, water meter, etc.) should never be run in conduit with high voltage (like 115VAC) wires.

NOTE: Liquid tight fittings and some labeled signal leads are provided for all signal (low voltage) connections for both pre-wired and conduit units.

Pre-Wired
Pre-wired units are supplied with a 16 AWG cable(s) with 3-wire grounded USA 115 volt plug for incoming power and 3-wire grounded receptacle cords for all control relay outputs also 16 AWG.

Conduit
Conduit units are predrilled at the factory and supplied with conduit knockouts for easy hard wiring to supplied detachable connectors on the relay card(s) located in the lower section of the controller. Remove the screws of the lower panel for access. NOTES: 1. 2. 3. 4.

Do not drill holes in upper section of enclosure. All relays provide a N.O. and N.C. output. The control function that activates each relay output is pre-configured at the factory based on the options selected. To change relay activation see on page 24. See page 30 for common bleed / blowdown valve wiring.

Relay Card Wiring


2

Ribbon Cable 10
9

NET GRD N.O. N.C. FUSE NET GRD N.O. N.C. FUSE

Relay 1

Relay Card Rev. D


Dry contact 115volt relay See note below
NET GRD

Relay 2

N.O. N.C.

Relay 4

FUSE

NET GRD N.O. N.C. FUSE

NET

Relay 3

GRD N.O. N.C.

Relay 5

FUSE

Jumpers above relay can change it to a dry contact relay. Use GRD and POWER INPUT N.O. for a N.O. dry contact relay.

H N G

If a whole card is configured for dry contact relays outputs each relay will be like the top one as shown on the Rev. C card.

Motherboard Connections
Aux Flow PIC Chip W W W R All 3

B Ground B Ground

ORP ORP pH pH Ground Drum Level 5 Drum Level 4 Drum Level 3 Ground Drum Level 2 Drum Level 1 Water Meter Ground Water Meter 2 Signal Water Meter 1 Signal

Signal Aux Flow 3 Signal Aux Flow 2 Signal Aux Flow 1 Aux Flow +5 vdc

Aux Flow Aux Flow

Ground Ground Ground

+5 VDC

Conductivity Temperature

Red Black Green White Red Black Black Red

Note: Use 22 AWG (.76 mm) twisted pair shielded wire for all of these low voltage signal connections.

Communications Program / Memory Card Connection Card Connection Key Pad Ribbon Cable Connection Incoming Power +5 +5

Processor

-12 +12

Display Ribbon Cable Connection

LED Cable

Relay Card Ribbon

White Black Red

RTD Probe Connections


pH pH Ground Drum Level 5 Drum Level 4 Drum Level 3 Ground Drum Level 2 Drum Level 1 Water Meter Ground Water Meter 2 Signal Water Meter 1 Signal Temp
Note: For probes with Red & Green RTD temp wires. To check probes RTD, measure across red & green with an ohm meter. At room temp it should read around 1097 ohms.

In + In Black Red Green

White Black Red Green

Conductivity Conductivity

Temp

WARNING! 6 wires plus a shield must be run from the controller to the probe. Make the 4 temperature wire connection at the probe for correct operation.

Flow Switch

Flow Switch

4-20mA Output Card Wiring


A. Isolated Configuration For isolated 4-20mA outputs an external power source for the loop must be supplied. JP1 and JP2 on the board must not be jumpered when an external power source is being supplied. The external power source must not exceed 24 volts DC and will connect onto the card on connector J1 on the postion marked Loop Supply. B. Non-isolated Configuration For non-isolated 4-20mA outputs the controller will supply the power for the loop. JP1 and JP2 must be jumpered and no connections are made to the Loop Supply points. Note: If multiple outputs are being connected to a single energy management system the negatives or commons MUST remain isolated from each other.

4-20mA Input Card Wiring


The 4-20mA input card requires that the external device sending the 4-20mA input signal(s) supply the power for the loop. The external power source must not exceed 24 volts DC.

4-20mA Output Isolation Jumper Settings

J6

J5

External VDC Power Source Not needed if MegaTron SS is supplying power. Adjust jumpers shown at left.

Part # SS-MAO-3
Non-Isolated Output (MegaTron SS Powered)

Loop Supply GND Loop Supply + (24 DVC max) Return 3 Out 3 + Return 2 Out 2 + Return 1 Out 1 +
Passive 4-20mA Loop Loop voltage supplied by External Power Source +24VDC Max or +12VDC by MegaTron SS

+12-24VDC

Isolated Output (External VDC Powered)

MegaTron SS 4-20mA Out

+
Building Automation Terminal

Part # SS-5VO-2

Rev C

Part # SS-MAI-2

Input 1
J1

Ground Signal + 12 VDC

Input 1 Input 1 + Input 2 Input 2 +


Rev D

Input 2
J3

Ground Signal + 12 VDC

Rev A

MegaTron SS 0-5 VDC Input

Input 3 Input 3 + MegaTron SS 4-20mA Input

Mounting Instructions
Select a mounting location that provides the operator easy access to the unit and a clear view of the controls through the cover of the controller. The location should be convenient to grounded electrical connections, the needed sample line plumbing and is on a stable vertical surface. WARNING: Avoid locations that expose the controller to direct sunlight, vapors, vibration, liquid spills or extreme temperatures; less than 0F (-17.8C) or greater than 120F (50C). EMI (electromagnetic interference) from radio transmissions and electric motors can also cause damage or interference and should be avoided.

Electrode Installation
MegaTronSS controllers may come configured for various circulating water systems. Listed below are instructions for cooling tower and boiler typical installations. Your specific installation requirements may differ but should conform to these instructions as much as possible for proper operation. A. Cooling Tower

The standard probe(s) and/or flow assembly for cooling tower installations is constructed of schedule 80 PVC and supplied with 3/4 slip fittings for installing into a sample line. To insure proper operation the sample line must have a flow rate of 3-10 gpm. Inlet pressure must be higher than outlet pressure in order for water to flow past the electrode(s) to achieve the required rate. The probes are temperature compensated for increased accuracy. NOTES: 1. 2. 3. 4. Install an isolation valve on either side of the flow assembly so electrodes can be easily isolated for removal and cleaning. A line strainer is recommended upstream from the probes to protect against fouling and damage. Mount pH electrodes vertically. Systems with a flow switch require 2-3 gpm flow rate to operate outputs.

WARNINGS: 1. Electrodes are O-ring sealed, which if damaged will cause a leak. 2. Do not allow pH sensor tips to dry out, damage will occur. 3. Do not exceed a water temperature range of 32F to 140F. 4. Do not exceed a maximum pressure of 150 psi.

Typical Cooling Tower Installation Diagram


COOLING TOWER

1 2 3 4 5
SET UP RUN
ENTER HOME

6 7 8 9 0

CANCEL

HELP

BACK

BLOWDOWN
BLEED VALVE

MAKE-UP WATERMETER

CIRCULATION

PUMP

CHILLER

Typical Cooling Tower Installation Diagram

FS-C FS-OC
Flow Switch

FS-F2 E-30 OR

TFS-OC

PE-NUT
pH Probe

FS-T

PE-2 -orOE-2
ORP Probe

T-3

E-30

E-4A
Conductivity Probe

Standard Tower Probes


Conductivity..........................E-4A pH.........................................PE-2 ORP..................................... OE-2

E-30

FS-SP
Sample Valve

B.

Boiler

Standard boiler electrodes have a MNPT stainless steel bushing and are supplied with a FNPT cross designed for mounting in the skimmer (surface) blowdown line. Sampling of the boilers water can be achieved using one of two typical plumbing configurations (continuous sampling or timed and/or hold sampling). For a successful installation, it is critical to observe the recommended distances and pipe sizes provided in the installation drawings. The probes are temperature compensated for increased accuracy. For best results, the electrode cross should be mounted in a 1skimmer blowdown line within 4 of the boiler. Smaller line sizes and greater distances may affect the response time and accuracy of the electrode. A flow-throttling device down stream from the probe (within 12-24 inches) is required to ensure that the electrode is exposed to water and not steam. Properly installed and adjusted, this device will prevent flashing in the electrode chamber. NOTES: 1. 2. 3. Install a fully ported type valve between the electrode and the boiler. This allows the electrode to be isolated for removal and cleaning. A flushing line and 1/4 turn type ball valve should be installed in the bottom of the cross to periodically flush sediment from the electrode chamber. Make sure the alignment arrows on the probe end up parallel to the flow for best performance.

WARNINGS: 1. The probe must be fully immersed in the system water to read correctly. Steam flashing will result in incorrect readings. 2. Do not exceed a maximum water temperature of 436F (224C) 3. Do not exceed a maximum pressure of 350 psi (24.1 bar) 4. A throttling device must be installed down stream from the elecrode.

Boiler Conductivity Electrodes


BE-4RTD and BE-32

Wiring Note: BE-4RTD probes require a 6 conductor cable from controller to probe, BE-32 probes only require a 2 conductor cable.

10

Boiler Conductivity Electrodes


1 2 3 4 5
SET UP RUN
ENTER HOME

6 7 8 9 0

Throttling device

Fully ported gate or ball valve 1" Continuous blow down line

CANCEL

HELP

BACK

Electrically actuated blow down valve

1/2" pipe between probe and throttle device and flush valve

Electrode within a 1" cross with electrical connection vented

Warning - Do not use on bottom blowdown lines, only continuous or surface blowdown lines.

Typical Timed Sampling and Sample and Hold Boiler Installation


Electrode within a 1" cross with electrical connection vented 1" Continuous blow down line Fully ported gate or ball valve Electrically actuated blow down valve 1/2" pipe between probe and throttle device and to the flush valve

1 2 3 4 5
SET UP RUN
ENTER HOME

6 7 8 9 0

CANCEL

HELP

BACK

Throttling device approximately 18" past probe

To drain

Warning - Do not use on bottom blowdown lines, only continuous or surface blowdown lines.

11

III. Front Panel Description

1 2 3 4 5
TIMERS

<HOME SETUP>

6 7 8 9 0

SET POINTS CALIBRATION

DATE/TIME CONFIGURE HISTORY WATER METER RELAYS

CUSTOMIZE ALARMS

SET UP RUN

ENTER

HOME

CANCEL

HELP

BACK

NUMBER Keys- UP/DOWN - LEFT/RIGHT - SET UP/RUN - ENTER - HOME - CANCEL - HELP - BACK -

Used to enter new values in the SET UP mode and to access desired sub menus. Used to cycle through text options to find desired setting. Used to cycle through text or setting options to find desired setting. System initializes into RUN mode. Press this key to put the controller in SET UP Mode and see HOME menu page. Used to log a changed value into program. Used to go back to the HOME menu page. Used to cancel a pop-up screen if no change is desired. Used to access help screens. Used to go back to last menu screen viewed or clear values keyed in that are not wanted.

12

IV. System Operation Overview


Operation
MegaTronSS controllers have two modes of operation, RUN and SET-UP.

RUN - This mode is for normal operation. In the RUN mode the display will show each systems parameters.
If an alarm is present the ALARM box will flash how many alarms are present. No settings may be entered or changed in the RUN mode. Readings are updated every 6 seconds on the screen while in the RUN mode.

SET-UP - This mode is used to make adjustments to settings and readings on the controller. To access the
SET UP mode from the RUN screen, press the SETUP/RUN key.

Typical Start Up Instruction


Complete all installation steps before beginning this procedure. Ensure that all controlled devices (pumps, solenoid valves, etc.) are operational and connected to the controller. Open the isolation valves to allow water to pass through the sample stream assembly. Before beginning the start up procedure, familiarize yourself with the programming and operation of the system by reviewing the menus available. Use the keys of the controller to skim through all your options. A. Calibration

All MegaTronSS controllers are factory calibrated for temperature, conductivity, pH and ORP. All units are shipped with the date preset, and the clock set to your current time. These readings and settings should be verified for accuracy, and adjusted as per the instructions listed below. 1. Conductivity - To calibrate the conductivity reading, remove the electrode from the line and wipe the flat surface with a clean cloth. Re-install the electrode and open the isolation valves to allow a sample across the probe. Be sure to allow the reading to stabilize for one minute. Select SYSTEM COND from the CALIBRATION menu. Then key in the corrected conductivity value. Press ENTER to log in that reading. 2. pH - Under normal operating conditions, pH calibration is achieved using the following steps. From the CALIBRATION menu, select SYSTEM PH. Enter the correct pH value. Press ENTER to log in the new reading. 3. ORP - To calibrate ORP, from the CALIBRATION menu select SYSTEM ORP. Compare the displayed reading to the actual system ORP. If these readings do not match, key in the correct ORP value, and push ENTER. There are limits to how much the calibration can be adjusted. The instrument will only accept new conductivity values which are from 1/3 to 8x the present reading. Any entry outside this range will cause a default to the original reading. If this happens, call 1-800-743-7431 for technical assistance.

13

V. Menu Navigation
To access the menus press the Set Up / Run key on the front panel. This takes you to the Home menu. MegaTronSS controllers menus are easily navigated by pressing the associated number key next to a menu box on the screen. Once you have stepped through the sub menus to reach a point at which a value or selection is made a Pop-up window will appear prompting you to enter a desired value or selection. NOTE: When entering new numeric values, enter all available digits (characters). 1. Home Menu

>HOME SETUP< SETPOINTS CALIBRATION TIMERS CUSTOMIZE ALARMS


SET POINTS - CALIBRATION - TIMERS - CUSTOMIZE - ALARMS - DAY/TIME - CONFIGURE - HISTORY - WATER METER - RELAYS -

DATE/TIME CONFIGURE HISTORY WATER METER RELAYS

From the HOME menu select the desired menu. The menu name explains what parameters can be programmed in the menu.

Setting control set points for conductivity, temperature, pH and ORP. Calibrating conductivity, temperature, pH and ORP if present. Menu for selecting type and settings for all present feed timers. Giving the controller, each system and all relays a user defined name. View current alarms. Menu for setting date and time. Menus for configuring passwords, relay activations, setting history interval, flow switch, contrast, temperature scale. Allows for view history on board in a graph form. Menu for configuring water meter totalizing. Menu for resetting accumulated ON times and manual activation.

2. Set Points The same basic format is used for defining each available analog probe inputs control parameters. Boiler conductivity systems will also have an additional menu step for programming the sampling method desired from continuous, timed or sample and hold.

>SYSTEM 1 SET POINTS< SYSTEM SYSTEM SYSTEM SYSTEM COND TEMP pH ORP mA OUT mA IN AUX INPUTS

SET POINTS - For setting the relay set points for the available analog probe readings such as conductivity, pH, ORP or temperature.

NOTE: In the Setpoint pop-up screen the direction (Rising or Falling) of the setpoint can also be set. Rising setpoints will activate the control relay when the particular probe reading rises above the setpoint and will stay activated until the reading comes down by the amount of the differential. If set for Falling the relay is activated when the probe reading falls below the setpoint and stays on until the probe reading comes back up by the amount of the differential.

14

2.1

Set Point Options

>SYSTEM 1 COND SETPOINT< SET POINT DIFFERENTIAL HIGH ALARM LOW ALARM LIMIT TIME

SET POINT - What reading turns the relay on DIFFERENTIAL - Amount reading changes by before the relay is turned off HIGH ALARM - What reading generates a High alarm notification. LOW ALARM - What reading generates a Low alarm notification. LIMIT TIMER - What amount of continuous bleeding will generate a time alarm notice.

Note: Each alarm value can also be set-up to be displayed or not on the front screen as will as remotely notified or not if connected to the Web Advantage server. 2.2 pH and ORP The pH and ORP set point settings follow the same format as shown above in section 2.1.

>SYSTEM 1 pH SETPOINT< SET POINT DIFFERENTIAL HIGH ALARM LOW ALARM LIMIT TIME INTERRUPT SET POINT 2

INTERRUPT - Only applies to pH settings and allows the Interruption of pH control during bleed, other chemical feed or both. SET POINT 2 - Is only present on systems with Dual pH control option. It is a second set point for applications when an acid and caustic are required

Note: When the Limit Time value is met it will force off the control relay being driven by pH or ORP. To reset the limit timer and get the control relay active again enter a new Limit Timer value.

15

2.3

Boiler Conductivity

Conductivity on boiler systems can be configured for Timed Sampling, Sample and Hold or Continuous for the conductivity sampling method. Timed sampling incorporates a sample timer which allows the boiler to be sampled at periodic intervals. Sample intervals are adjustable from 1 minute to 99 hours, 59 min. Sample duration (on-time) is adjustable from 1 second to 99 minutes, 59 seconds. Sample and hold uses a sample timer for periodic sampling intervals. The unit will sample for its duration then hold the blowdown valve closed for a settable period (hold time). The conductivity is checked at the end of the hold period, if additional blowdown is required the blowdown valve is held open for a preset amount of time (blowdown time). Then sample cycle is repeated. Continuous sample has a sample of boiler or condensate water go past the probe continuously. If the reading is above the set point, blowdown will continue until the set point has been satisfied.

>SYSTEM 1 COND SETPOINT< SET POINT DIFFERENTIAL HIGH ALARM LOW ALARM SAMPLE INT SAMPLE DUR HOLD TIME BLOWDOWN

SET POINT - What reading turns the relay on DIFFERENTIAL - Amount reading changes by before the relay is turned off HIGH ALARM - What reading generates a High alarm notification. LOW ALARM - What reading generates a Low alarm notification. SAMPLE INT - Period between samples. SAMPLE DUR - How long a sample lasts. HOLD TIME - How long a hold lasts. BLOWDOWN - How long to blowdown if reading is above the set point after the hold. Hold repeated after BLOWDOWN time.

2.4

Aux Inputs

Auxiliary inputs are the digital inputs for optional Flow Switch and other digital inputs such as low drum level alarms. From these menus the user can set if they want each of the alarms Displayed, Remote Notification, both or none.

>SYSTEM 1 DIGITAL INPUTS ALARM< ALARM NOTIFICATION FLOW SW = DISPLAY DIGITAL 1 = DISPLAY FLOW ALARM DIGITAL DIGITAL DIGITAL DIGITAL DIGITAL 1 2 3 4 5

16

2.5

4-20mA Out

Units with a 4-20mA output option will have a menu for setting up the 4-20mA output. The 4mA and 20mA values can be defined by giving the output proportioning capability. i.e. 4mA = a pH of 6.0 and 20mA = a pH of 8.0.

>OUT 1 SETUP< SIGNAL SOURCE 4mA VALUE 20mA VALUE


2.6 4-20mA Input

SIGNAL SOURCE - Select which probe reading the mA will use as its reading source. 4 mA VALUE - What the 4mA signal equals 20mA VALUE - What the 20mA signal equals on the assigned signal sources scale.

>mA INPUT 1 SETPOINT< SET POINT DIFFERENTIAL HIGH ALARM LOW ALARM

SET POINT - What reading turns the relay on DIFFERENTIAL - Amount reading changes by before the relay is turned off HIGH ALARM - What reading generates a High alarm notification. LOW ALARM - What reading generates a Low alarm notification.

3.

Calibration

Calibration is for adjusting the displayed value of a probes reading to match your tester or known solution. Pick the system or mA input first. From a particular system pick the probe to calibrate.

>SYSTEM 1 CALIBRATION< SYSTEM SYSTEM SYSTEM SYSTEM COND TEMP pH ORP

CALIBRATION - For adjusting the actual reading values of the available analog probe inputs, such as conductivity, pH, ORP or temperature.

3.1

Conductivity Calibration

>SYSTEM 1 COND CALIBRATION< CALIBRATE RESET ZERO Cal Factor 1.10

CALIBRATE - With a clean probe on-line seeing the systems water enter the known (tested from a calibrated hand held tester) value. RESET ZERO - With the probe out of solution and dry enter a new zero point. Note: Probe must be dry!

NOTE: In the calibration pop-up window the raw analog to digital (A/D) value will be displayed. A new calibration value should only be entered when the probe is sensing a stable A/D value. If the system is a boiler the associated relay can be forced on in the pop-up to allow for a fresh hot sample. Boiler probes should have fresh hot water blown past them for a minimum of 2 minutes before calibrating.The A/D reading has a range of 0 to 32,767. If it is at one end of the range or the other when trying to calibrate something is wrong with the probe or wire run.

17

3.2

pH and ORP Calibration

>SYSTEM 1 pH CALIBRATION< 1 POINT CAL 2 POINT CAL

1 POINT - With a clean probe on-line seeing the systems water enter the known (tested from a calibrated hand held tester) value. 2 POINT - Enter a known Low value with a clean probe in a buffer solution. Then enter a known High value with the clean probe in a buffer.

Notes: 1. 2. 3.3

Probes must be rinsed between buffers when doing a 2 Point calibration and given at least 30 seconds in solution to adjust to the buffer. There must be at least 2 full pH points between buffer solutions. Using a buffer of 4 and 10 is the best procedure. ORP buffers should be at least 200 points apart.

4-20mA Output Calibration

>CURRENT LOOP CALIBRATION< OUTPUT 1 OUTPUT 2

4-20mA outputs can be calibrated to insure that the output generated by the controller and received by the external device match. With a volt meter connected across the out and return wires (see page 7) of the 4-20mA output channel to be calibrated go into the outputs Low or High calibration.

>mA OUTPUT 1 CALIBRATION< HIGH 4030 LOW 800

The number displayed in the Calibration dialog box can range from 0-4095 with 800 equal to 0 mA output and 4030 equal to 20 mA. This number range of 04095 is the raw digital to analog (D/A) values and is strictly used for a reference. The D / A numbers you get will vary based on your installation conditions.

>mA OUTPUT 1 CALIBRATION< OUTPUT 1 CAL HIGH 4000 Use Up/Down arrows to change Use Enter to save value

While in the High or Low calibration pop-up screen use the up and down arrows to change the output value being read with the volt meter. Adjust the High value for the 20 mA reading and the Low value for the 4 mA value.

18

3.4

4-20mA Input Calibration

4-20mA inputs can be calibrated to insure that the input seen by the controller from the external device match. It also allows for setting the 4-20mA input into a number range that relates to the value being read.

>CURRENT LOOP CALIBRATION< INPUT 1 INPUT 2


Select the Input to be calibrated

>mA INPUT 1 CALIBRATION< 20mA 4mA MAX LOW OFFSET 14500 2900 200 0 DISABLED

>mA INPUT 1 20mA CAL< INPUT 1 CAL HIGH 14500 A/D PUT 20mA ON INPUT THEN PRESS ENTER

The 20mA and 4mA values are where the controllers raw analog to digital value is adjusted to match a 20mA (full scale) and 4mA (bottom of scale) signal from the external device inputting the 4-20mA input. The external device must be connected to the controller and showing either full scale or bottom of scale when calibrating each. The number shown along with either the 20mA or 4mA while calibrating is the raw A/D value and is only a reference. A 20mA input should be around 14500 and 4mA around 2900. If the A/D numbers are not in this range check input device. The MAX and LOW calibration inputs are for telling the controller what to display for a 20mA input and a 4mA input. For example if the input is a drum level sensor monitoring a 55 gallon drum the value for MAX should be 55 and LOW should be 0. The controller then displays a number automatically ranging between 55 and 0 based on the input value. The units of measure (gallons for example) is set in the Customize menu from the Home page. OFFSET - Offsets current displayed value of the 4-20mA input reading to allow for a manual 1pt calibration. 4. Timers

A unit may have up to 5 selectable timers for each system on a controller. All timers are associated with their system, so for a % of post bleed timer looks at the bleed of that system.

>SYSTEM 1 TIMERS< TIMER TIMER TIMER TIMER TIMER 1 2 3 4 5

TIMERS - Select the type (28-day, pulse, limit, percent or percent of post blowdown) as well as the run times of each timer available per system.

19

4.1

Timer Type Selection

A pop-up screen lets you scroll through the various timer types available.

>SYSTEM 1 TIMER 1 SET UP< >SET TIMER TYPE (PULSE)< -> PULSE USE UP/DOWN KEYS TO CHANGE PRESS ENTER TO ACCEPT SET UP TIMER TYPE

Pulse - A water meter activated timer Limit - Feed with bleed with a maximum run time or limit for one bleed cycle. Percent - A continuous cycle timer that sets a percentage On time of a cycle. Percent Post Bleed - For feed after bleed for a settable percentage of the bleed time with a maximum run time. 28-Day - A biocide or event timer.

4.2

Timer Set Up

>SYSTEM 1 TIMER 1 SET UP< TIMER TYPE: PULSE GALLONS: 10 RUN TIME (MM:SS): 01:00 INPUT: WATER METER 1 SET UP TIMER TYPE

Each timer type selected will have its own unique Set Up sub menu with additional selections specific to the type of timer selected. The page displayed before entering the Set Up menu of a timer provides an overall review of the timers current settings.

4.3

Pulse Timer

>SYSTEM 1 TIMER 1 CHANGE< ACCUMULATE RUN TIME METER INPUT

BLEED INTR

ACCUMULATE - The number of gallons or liters from water meter to count before activating timer. RUN TIME - The amount of time for the timer to run METER INPUT - Select water meter 1 or 2 for the timers activation. Water meter 1 + 2 can be selected only if the meters have the same contact value. BLEED INTR - Allows the pulse timer to be interrupted during conductivity bleed.

4.4

Percent Timer

>SYSTEM 1 TIMER 1 CHANGE< % OF CYCLE CYCLE TIME

% of CYCLE - The percent of the defined cycle time that the timer is to be on. CYCLE TIME - The amount of time in MM:SS that the cycle is going to be on.

20

4.5

Post Bleed Timer

>SYSTEM 1 TIMER 1 CHANGE< % OF BLEED LIMIT TIME SOURCE

% of BLEED - The % of the post bleed time or other source time that you want the timer to run. LIMIT TIMER - The limit timer is a safety feature that limits a single feed cycle to the amount of time set regardless of the calculated post feed %. SOURCE - Select between cond. bleed, another timer or digital input as the time souce.

4.6

With Bleed Timer

>SYSTEM 1 TIMER 1 CHANGE< LIMIT TIME

LIMIT TIMER - This timer starts when the conductivity bleed starts and turns off when the bleed stops or until the limit time has been reached. The time can be set in MM:SS.

4.7

28-Day Timer

Each 28-day timer has Program 1-4 for programming the various feed times. While the programming steps for four programs are the same each can have its own independent settings.

>SYSTEM 1 TIMER 1 CHANGE< WEEKS DAYS START TIME RUN TIME PREBLEED MIN FEED BLEED FLOW ORP COND LOCK LOCK LOCK LOCK

WEEKS - The week(s) that the timer is to feed. DAYS - The day(s) that the timer is to feed. START - The time of day for the timer to start. RUN - How long the timer is to run. PREBLEED - How long the controller is to bleed down before feeding in chemical. NOTE: The prebleed starts at the START time programmed above. MIN COND - The minimum conductivity that the unit will prebleed down to. FEED LOCK - Which other system timer to lockout during this timers run time. BLEED LOCK - How long to lock out the bleed function after the timers run time starts. FLOW LOCK - If the system has a flow switch you can ignore it for this timer.

NOTE: ORP lock is only for systems with the ORP function. If interlocked the ORP set point will control the 28-day timers relay On and OFF during the timers programmed RUN time. The ORP will need to have a 00:00 setting for its limit timer.

21

5.

Customize

This menu allows the user to define the on-screen name of the unit plus the name of each system and relay. The user can also setup the Notepad for each system and 4-20mA Inputs name and unit of measurement.

>CUSTOMIZE< UNIT NAMES RELAY NAMES SYSTEM 1 NAME NOTEPAD pH INDEX mA IN RUN SCREEN

RUN SCREEN - Allows the user to select what will be shown on the screen while the controller is in the RUN mode. Like diplaying temperature readings or water meter totals for a particular system.

NOTE: When entering values for custom names use the numerical keys for numbers and the up / down arrows to scroll through all the characters of a key board. Press the right arrow to advance the curser after setting a desired value. Press the Help button to place the last entered character into the new cursor space to speed up the process. The Help button will also jump advance through the characters. 5.1 Notepad

The Notepad function allows the user to set up a customized manually entered data field for each system with ten notepad items. The NOTEPAD is ideal for setting up and storing into the controllers history the items typically tested for reporting a service call. The Notepad items come with no names but when an individual note is selected a menu for setting it appears. Also can be used as a Reminder Timer with or without flow.

>NOTEPAD SYS 1 NOTE 1< NAME NUMBER UNITS ALARMS

NAME - Pick from a list of defined names or customize your own. NUMBER - Set the number range. UNITS - Set the units of measurement. ALARMS - Set Hi/Low alarm points and how frequently a new value is expected to be manually enter via the History menu.

5.2

mA Inputs

>mA INPUT 1 CUSTOMIZE< NAME UNITS NUMBER

NAME - Name the input. UNITS - Set the units of measurement. NUMBER - Set the number range.

5.3

pH Index

If the controller has the pH index option (must have pH control) user can select LSI or RSI indexing. The controller will automatically set some Notepad entries for manually inputting needed data to perform the calculation which will be displayed in the RUN screens.

22

6.

Alarms

>ALARMS< SYS 1 ALARMS

ALARMS - Shows any current alarms.

7.

Date and Time Set Up

>SET DATE AND TIMES< SET DATE SET TIME SET DAY SET WEEK Friday May 14, 2005 03:04:56

DATE AND TIME - For setting the date, time, day and week on the controller.

8.

Configure

Provides access to menus to set-up passwords, relay activation, temp scale, display contrast, flow switch, inputs, history time stamps, factory set-up and system information.

>CONFIGURE< PASSWORD RELAYS HISTORY FLOW SW CONTRAST TEMP SCALE NETWORK SYS INFO FACTORY

CONTRAST - This screen allows for adjusting the display contrast. FLOW SW - Defines a flow switch to be open or closed with flow. FACTORY - A factory only menu TEMP SCALE - Set Celsius or Fahrenheit HISTORY - Sets the history time stamp interval. SYS INFO - Tells unit software specifics.

8.1

Password

>CONFIGURE PASSWORDS< ADMIN PASSWORD USER PASSWORD USER SET UP

ADMIN PASSWORD - The administrator password gives access to all menus except factory set up. USER PASSWORD - The user password allows the user to access HOME menus that are made available in USER SET UP.

23

8.2

Relays

>RELAY 1 SETUP< MAIN ACTION ACTIVATOR 2 ACTIVATOR 3 ACTIVATOR 4

DISABLE 1 DISABLE 2 DISABLE 3 A pop-up screen appears with a list of all available DISABLE 4 activation functions to arrow through. DAILY MAX

CONFIGURE RELAYS - This menu lets you choose a Main Action or function (timer 1, conductivity, alarms etc...) to activates a relay.

Additional relay logic is available with up to 3 additional Activators and up to 4 Disablers allowing multiple functions to activate the same relay and multiple functions to prevent the relay from coming on. There is also a Daily Max amount of time that a relay can be on. If a relay is on for the max amount it does not let the relay come on anymore that day. (A 24 hour clock is used for the day with midnight being the start of the day). 8.3 History This menu is used to set the history time stamp interval, the water meter daily history starting hour and the alarm delay period.

>CONFIGURE HISTORY< INTERVAL W/M HOUR ALARM DELAY

INTERVAL - The amount of time between each history time stamp for probe readings. W/M HOUR - The time of day that the daily water meter history cycle is to start. ALARM DELAY - The amount of time an alarm has to be on before it is recognized as an alarm.

8.4 Flow Switch This menu allows the user to select if a flow switch signal will represent a flowing condition when a closed or open signal is seen for each systems flow switch input. User can also select if timers can work always or only with flow. 8.5 Contrast This menu is used to adjust the contrast of the display. 8.6 Temperature Scale This menu is used to select the type of temperature scale to display. 8.7 Network The Network menu is used when a controller is being remotely communicated with either a local network connection or over the internet on the Web Advantage server.

>CONFIGURE NETWORK< IP ADDRESS IP MASK GATEWAY SERVER HTTP LOCAL HTTP REMOTE FTP RESET

NETWORK - This menu is used for setting up the remote WebAdvantage communications and is covered in a separate manual.

24

8.8 System Information System information will identify the version of firmware installed in the controller along with the controllers serial number. 9. History The onboard history allows for viewing the history of the probe readings, relay activations, key-pad activity, calibrations, water meter hourly and daily logs and alarms for each system present. It is also where Notepad data is entered and reviewed. An initial overview page is displayed showing your current sample interval, the calculated number of days the unit can keep probe history for before losing the oldest. The number of sensor samples and relay/alarm events and Notepad entries currently stored is also displayed.

>HOME REVIEW< SYSTEM 1


Sample Time: 5 MIN (Length 164.62 days) Sensor Samples 882 Relay/Alarm Events 323 Notepad Entries

NOTE: The history can be reset by going to the configure menu and entering a different sample interval. After the new sample interval has been set the onboard history is reset.

9.1

Viewing History

>HISTORY< RELAY LOGS ALARM LOGS SENSOR HISTORY EVENT LOG NOTEPAD WATER METER

RELAY LOGS - Relay activations displayed in a log form. Arrow up to advance through the log. ALARM LOG - Alarm activations in log form. SENSOR HISTORY - For selecting the parameters and viewing of a given probe readings history in log or graph form. EVENT LOG - Displays various activities.

9.2

Notepad Entries

>NOTEPAD: SYS 1 NOTE 1< ENTER VALUE LOG GRAPH Total Hardness 8 Entries 517.2 Hrs to Alarm

The Notepad section under History is where the user goes to enter new values for the customized notepad items. Each individual notepad items manually entered entries are stored in the units history and can be reviewed in log or graph form after 4 or more values have been entered.

25

9.3 Water Meter History The water meter history allows the user to review both water meter one and two of a particular system in both an hourly format (for the past 24 hours) or a daily format for the past 60 days. If an evaporation calculation is being kept, a daily history of this value is also available. 10. Water Meters Each system with a timer on it will have 2 water meter inputs. Each of these can have the incoming contact defined allowing the controller to keep track of water usage. If desired the controller can calculate evaporation loss by subtracting the deference between a systems 2 water meter inputs.

>SYS1 WATER METER 1<


TOTAL VALUE: 0280500.0 GALLONS FLOW RATE: 000.0 GALLONS/MIN CONTACT VALUE: 0010.0 GAL/CONTACT CONTACT VALUE CONTACT UNIT RESET TOTAL EVAP CALC 11. Relays

CONTACT VALUE - Defines the numerical value for a contact; i.e. 10. CONTACT UNIT - Defines the units of measurement for a contact; i.e. Gallons / Contact RESET TOTAL - Resets the totalizer count. EVAP CALC - Defines which way to subtract the 2 water meter inputs for an evaporation value.

>RELAYS< STATUS RESET

STATUS - Allows for viewing accumulated relay ON times, temporary forcing relays ON or OFF or seeing which relay is on. RESET - Allows the accumulated run time of a particular relay to be reset to zero.

26

VI. Maintenance
The only required maintenance for normal uninterrupted operation of your MegaTronSS controller is cleaning of the electrode(s). After initial start up, it is a good idea to clean the electrode frequently until a schedule based on need has been developed. Since each application is unique, it is difficult to estimate the required frequency of cleaning. The first cleaning should take place after about one week of the system being on line. To determine the required cleaning frequency, record the reading on the controller before the electrode is removed for cleaning. After cleaning, record the new reading. If a change is observed in the two readings, the electrode was dirty. The more significant the change, the dirtier the electrode. If no change occurs, cleaning needs to be done less often. Conductivity Electrode Cleaning Procedure 1. 2. 3. 4. 5. Record the current conductivity reading. Turn off water flow through the electrode loop, bleed pressure from the line and remove electrode. Use a clean cloth and a mild cleaning solution to remove loose dirt etc., from the flat surface of the electrode. If the electrode has deposits such as scale attached to the electrode surface a more aggressive cleaning approach will be needed. There are several ways to do this, the preferred method being the one that is easiest for the user. a. Use a mild acid solution to dissolve deposits. b. Lay a piece of sandpaper (200 grit or finer) on a flat surface such as a bench top. Sand electrode to remove stubborn deposits. (Do not wipe surface with your finger.) Oil from your skin will foul carbon tips. Reinstall the electrode in the system. After the reading stabilizes, calibrate the unit to a reliable test reading.

Many times an electrode can appear to be clean, but the unit still cannot be calibrated. If this is the case, use one of the more aggressive electrode cleaning procedures listed in step 4 above. Recheck the calibration after completion of this procedure. If no change was observed in the reading, replace the electrode. If a change occurred but the unit still will not calibrate, repeat procedure as many times as necessary. pH and ORP Electrode Cleaning Procedure 1. 2. 3. 4. Remove the pH electrode from the system. Spray with water and/or detergent, using a soft brush to dislodge any fouling. Visually inspect the electrode for signs of damage. Calibrate the electrode while it is in a known solution.

Slow response or non-reproducible measurements are signs that the electrode has become coated or clogged. The pH glass is susceptible to mounting by many substances. The speed of response, normally 95% of the reading in less than 10 seconds, is dramatically degraded when the pH glass is coated.

27

To restore the speed of response for a pH electrode, clean the bulb with a high quality detergent, methyl alcohol or other suitable solvent using a Q-tip. Rinse well with distilled water and retest. If the electrode now responds, but erratically, soak the sensor in 0.1 Molar HCl for 5 minutes. Remove and rinse with water and place in 0.1 Molar NaOH for 5 minutes. Remove, rinse again and then place the sensor in pH 4. buffer for 10 minutes before use. ORP/REDOX electrode may be gently abraded by use of 600 grade wet silicone carbide paper, jewelers rouge or very fine steel wool. Try one of the previous cleaning methods first before using this method.

VII. Troubleshooting
The Advantage MegaTronSS controller is designed for many years of trouble free operation. Should a problem occur, refer to the following chart to help identify the problem. If replacement is required, follow the procedures listed in the Warranty and Factory Service portion of this manual. SYMPTOM POSSIBLE CAUSE False reading. .................................... Bad or dirty electrode Out of calibration Will not calibrate................................ Dirty electrode Faulty electrode Faulty wiring to electrode No system power. .............................. Check power source Check fuse Check connections Pulse timer not activating.................. Check wiring Check external device Outputs not energized....................... No flow Check fuse SOLUTION Clean, as needed Calibrate unit Clean electrode Replace electrode if needed Replace wiring if needed Plug into different receptacle Replace as needed Make sure ribbon cables are secure Repair as needed Repair/replace as needed Check sample line for clogged pipes or strainers Replace as needed

28

VIII. Advantage Controls Product Warranty


Advantage Controls warrants control systems of its manufacture to be free of defects in material or workmanship. Liability under this policy extends for 24 months from date of installation. Liability is limited to repair or replacement of any failed equipment or part proven defective in material or workmanship upon manufacturers examination. Removal and installation costs are not included under this warranty. Manufacturers liability shall never exceed the selling price of equipment or part in question. Advantage disclaims all liability for damage by its products caused by improper installation, maintenance, use or attempts to operate products beyond their intended functionality, intentionally or otherwise, or any unauthorized repair. Advantage is not responsible for damages, injuries or expenses incurred through the use of its products. The above warranty is in lieu of other warranties, either expressed or implied. No agent of ours is authorized to provide any warranty other than the above. 30 Day Billing Memo Policy Advantage Controls maintains a unique factory exchange program to ensure uninterrupted service with minimum downtime. If your controller malfunctions, call 1-800-743-7431, provide our technician with Model and Serial Number information. If they are unable to diagnose and solve your problem over the phone, a fully warranted replacement will be shipped, usually within 48 hours, on a 30 Day Billing Memo. This service requires a purchase order and the replacement is billed to your regular account for payment. The replacement will be billed at current list price for that model less any applicable resale discount. Upon return of your old panel, credit will be issued to your account at either 100% if your unit is in warranty or at 50% if your unit was out of warranty. The exchange covers only the panel. Electrode and enclosure are not included. FCC Warning This equipment generates and uses radio frequency energy and if not installed and used properly, that is, in strict accordance with the manufacturers instruction, may cause interference to radio communications. It has been type tested and found to comply with the limits for a class A computing device pursuant to subpart J of part 15 of FCC Rules, which are designed to provide reasonable protection against such interference when operated in a commercial or industrial environment. Operation of this equipment in a residential area is likely to cause interference in which case the user, at his own expense, will be required to take whatever measures necessary to correct the interference.

29

Notes Common boiler blowdown valve connections.

N.O.

GND
COM

SOLENOID

MOTORIZED BALL VALVE

M
CO M

N.O N.C

LIMIT SWITCH
MBWB 1036

N.O.

N.C.

COM
1 2 3 4 5 6 7 8 9 10 1 11 2

MBWA 1075
30

-Notes-

31

Get the Advantage in Water Treatment Equipment


Advantage Controls can give you the Advantage in products, knowledge and support on all of your water treatment equipment needs. Cooling Tower Controllers Boiler Blow Down Controllers Blow Down Valve Packages Solenoid Valves Water Meters Chemical Metering Pumps Corrosion Coupon Racks Chemical Solution Tanks Solid Feed Systems Feed Timers Filter Equipment Glycol Feed Systems Pre Fabricated Systems

Get the Advantage

32

THERMALCARE TWC-1000 CONTROL PANEL PREPROGRAMMED SETTINGS Relays Relay 1 Name: Setpoint: Deadband: Timeout: Bio-Lockout Name: By: Setpoint: Feed Time: Bio-Lockout Name: By: Day: Feed Duration: Lockout: Name: By: Activate: High: 1200 mhos Makeup: Contacts per Gal. BLOW 1000 mhos 30 mhos 3 hours Yes INHB (Inhibitor) Water Meter 50 Gallons 1 Minute No BIO (Biocide) BioSchedule Wednesday 1 Hour 2 Hours ALRM (Alarm) Alarm All Alarms Low: 750 mhos 1" Turbine 65 Central Standard Time 2" Turbine 15

Relay 2

Relay 3

Relay 4

Alarms Meters:

Clock:

Data Sheet
Controller
Microprocessor Control of:
Conductivity pH ORP Chemical Feed Biocide Feed Service Reports 4-20mA In & Outs

Key Features

Customizable LCD Display On Board History Graphs Simple ATM Style Menu One Point Calibration Internet Communications Option 5 Assignable Relays Relay Test Keys Customizable Notepad Multi-Level Security Code LSI or RSI Index Options 2 Year Replacement Warranty Email Alarm Capable MODbus and BACnet Options

Application
The MegaTron SS controllers offer all of the user friendly features of the original MegaTron at a more economical price for single water treatment systems. Units can be configured to control a wide range of digital and analog inputs. Relay activation and names plus many other features are field selectable. The Notepad allows the input and history gathering of service report parameters. MegaTron SS units can control a single cooling tower or boiler system with the most customization flexibility available.

Build a Model
The model number starts with SS followed by the code for each function needed for the first system. (SSCPF3E) Once the main control functions have been selected, insert another - and list the desired common options. (SSCPF3-H1). Conductivity Control (1 per system max) B2 = BE-32 Standard Boiler probe C = TE-4A Standard Tower probe PH Control (1 per system max) P = TPE-21 Standard Tower probe ORP Control (1 per system max) R = TOE-21 Standard ORP probe Feed Timers (max of 5 per system) F1 to F5 (F4 = Four feed timers) Flow Switch E = Flow Switch OPTIONS A Conduit connection A3 Conduit with CE approval H1 Internet Connect Communications H4 Internet card with phone modem H5 Internet card with cellular router H11 Internet card w/ CAT5 & Modbus TCP/IP H21 Internet card w/ CAT5 & BACNet TCP/IP N3 (3) 4-20mA Inputs (no option U3 if selected) O3 (3) 4-20mA Outputs S pH saturation indexes (with pH control) U3 (3) 0-5 volt inputs (no option N3 if selected) V 5 volt D.C. output with water meter wires W3 (3) Auxiliary flow meter inputs Y Agency Approval (ETL, US&C) NOTES: 1. Timers selectable (pulse, %, post bleed, limit or 28-day). Each system has 2 water meter inputs. 2. All expansion slots provided. 3. MegaTron SS controllers can have 3 probes max. 4. MegaTron SS controllers have 5 relays max. 5. Contact the factory for a complete option list.

Specifications
Electrical Input: 95-240 VAC, 50/60 Hz Control: Equal to input voltage (95-240VAC) fused at 2.5A per relay Prewired units are supplied with an 8 (248.84 cm) power cord and 8 (20.32 cm) output receptacles. Operational Display: 240 X 128 Graphic LCD Conductivity Control: 0-10,000 S/cm scale Boiler sample method is selectable from continuous, timed sampling or sample and hold. pH: 0-14 scale ORP: +/- 1000 millivolts scale The ORP setpoint can be tied to an optional 28-day feed timer as an override. Accuracy: +/- 1% Feed Timers are all selectable from: Pulse - water meter activated w/accumulator Percent - 1 - 100% of a cycle time. Limit - Limits feed with bleed Post - 1-100% of post bleed or other time with an over all limit. 28 Day - biocide timer. Enclosure Heavy duty NEMA 4X style high impact thermoplastic with padlockable gasketed Lexan viewing door. Environment Ambient temperature: Relative humidity: 0o to 125oF (-17 to 52oC) 0 to 100%

Electrodes Tower Standard electrodes are supplied in 3/4 (1.91 cm) SCH 80 PVC female slip tees with quick release nut. Conductivity: TE-4A 150 psi (10.3 bar) / 140oF(58oC) Max pH: TPE-2 100 psi (6.8 bar) / 140oF(58oC) Max ORP: TOE-2 100 psi (6.8 bar) / 140oF(58oC) Max Temperature: TC-1 150 psi (10.3 bar) / 140oF (58oC) Max Boiler Conductivity electrodes are supplied with 1 (2.54 cm) MNPT SS threads. Conductivity: BE-32 350 psi (24 bar) / 436oF(224oC) Max Shipping Weight: Dimensions: Approximately 8 lbs. (3.62 kg)

W 10 (28.4 cm) x H 12 (30.5 cm) x D 7.25 (18.4 cm)

Get the Advantage


4/2010

4700 Harold Abitz Dr Muskogee, OK 74403 800-743-7431 phone 888-686-6212 fax www.advantagecontrols.com

Manual MicroTron
Series B
Chemical Metering Pump

Installation Maintenance Repair Manual

Advantage Controls P.O. Box 1472 Muskogee, OK 74402 Phone: 800-743-7431 Fax: 888-686-6212 www.advantagecontrols.com email: support@advantagecontrols.com
1

3/2010

Table of Contents
I. II. III. IV. V. VI. VII. Introduction..........................................................................................3 Model Numbering................................................................................3 Unpacking............................................................................................4 Safety Considerations. .........................................................................4 A. Chemical Compatibility.............................................................4 B. Safety Equipment and Preparation...........................................4 Installation. ...........................................................................................4 A. Location....................................................................................4 B. Electrical...................................................................................4 C. Plumbing...................................................................................5 Start-Up...............................................................................................7 A. Priming the Pump.....................................................................7 B. Adjusting Feed Rate.................................................................7 C. Stroke Length. ...........................................................................7 D. Calculating Output....................................................................7 E. Optional Features.....................................................................8 Maintenance........................................................................................8 A. Diaphragm Replacement..........................................................8 B. Suction and Discharge Check Valve Replacement..................9 C. Liquid End Diagram................................................................10 D. Pump Replacement Parts.......................................................11 Trouble Shooting Guide.....................................................................12

Manufacturers Product Warranty


Advantage Controls warrants units of its manufacture to be free of defects in material or workmanship. Liability under this policy extends for 24 months from date of installation. Liability is limited to repair or replacement of any failed equipment or part proven defective in material or workmanship upon manufacturers examination. Removal and installation costs are not included under this warranty. Manufacturers liability shall never exceed the selling price of equipment or part in question. Advantage disclaims all liability for damage caused by its products by improper installation, maintenance, use or attempts to operate products beyond their intended functionality, intentionally or otherwise, or any unauthorized repair. Advantage is not responsible for damages, injuries or expense incurred through the use of its products. The above warranty is in lieu of other warranties, either expressed or implied. No agent of ours is authorized to provide any warranty other than the above.

30 Day Billing Memo Policy

Advantage Controls maintains a unique factory exchange program to ensure uninterrupted service with minimum downtime. If your unit malfunctions, call 1-800-743-7431, and provide our technician with Model and Serial Number information. If we are unable to diagnose and solve your problem over the phone, a fully warranted replacement unit will be shipped, usually within 48 hours, on a 30 Day Billing Memo. This service requires a purchase order and the replacement unit is billed to your regular account for payment. The replacement unit will be billed at current list price for that model less any applicable resale discount. Upon return of your old unit, credit will be issued to your account if the unit is in warranty. If the unit is out of warranty or the damage not covered, a partial credit will be applied based upon a prorated replacement price schedule dependent on the age of the unit. Any exchange covers only the controller or pump. Electrodes, liquid end components and other external accessories are not covered.

I.

Introduction

This manual covers all facets of operation of the Advantage MicroTron pump, including unpacking, mounting, electrical and plumbing connection, and start-up. Safety, maintenance and repair, warranty, and factory information is also provided. Please read this manual completely before proceeding. Observe safety protocols and heed all warnings and precautions.

Model Numbering

MicroTron Series B pump model numbers define the output, pressure and control functions present on a particular pump. Your pump may be supplied with one or more of the options described in this manual. To determine what features apply to your pump, check the model number label located on the pump. Model Number Example Pressure Rating 1 = 110 150 and or 75 psi 2 = 250 psi Gallon Per Day Rating Control Options X = Stroke Length Adjust E = External Pacing D = Dual output 10-50 gph or 0-125 spm F = Hall effect meter input by 9 L = Limit Timer Voltage 1 = 120 volt - 50/60 Hz - .35A with USA plug 2 = 240 volt - 50/60 Hz - .65A no plug 3 = 240 volt - 50/60 Hz - .65A with specified plug 4 = 12 volt dc 5 = 240 volt USA plug Pump Head Material K = Kynar S = 316 Stainless Seat Material V = Viton F = Teflon H = Hypalon Check Ball C = Ceramic D = Ceramic single on discharge S = Stainless Tubing Connections 1 = 3/8 PE, models up to 150 psi 2 = PE, 250 psi models 3 = 3/8 PE on 250 psi models P = MNPT V = 3/8 Clear flexible suction tubing Special Options S = 3-function injection valve B 1 30 X 1 K F C 1 - S

NOTE: This list represents our most popular options. If you have an option not covered, contact the factory or your dealer for more details.

II.

Unpacking
The MicroTron pump has been shipped as a complete package, ready for installation. If the shipping carton shows any signs of damage, notify the shipping company immediately upon receipt. Advantage Controls cannot be held responsible for damage from shipping. Unpack the carton and insure the following items are present: Metering pump Suction, discharge and priming tubing Foot valve and weight 4. 5. Injection fitting Instruction manual

1. 2. 3.

III.

NOTE:

Safety Considerations
All MicroTron pumps are primed with water before leaving the factory. If the solution to be pumped is not compatible with water, disassemble the pump fluid end before use. After disassembly, thoroughly dry the pump head, valves, and seals before pump is reassembled and used.

A. Chemical Compatibility MicroTron metering pumps are designed to work with most liquid chemicals depending upon your pumps liquid end materials of construction. A chemical resistance chart is available for determining specific compatibility with a wide variety of chemicals. If you have further compatibility questions, contact Advantage Controls service department at 1-800-743-7431. B. Safety and Preparation Always wear the proper protective clothing and gear when working around chemicals and chemical metering pumps. Safety glasses, gloves, and aprons are critical in preventing accidental exposure to dangerous chemicals. Liquids under pressure can present a special hazard when a line or seal is punctured resulting in the spraying of chemical many yards away. If a chemical spillage occurs, consult the Material Safety Data Sheet (MSDS) for specific instructions regarding the chemical being used.

IV.

Installation
A. Location Select a mounting location convenient to the chemical supply as well as a source of power for the pump. Do not install the pump in a location where the ambient temperature exceeds 120 degrees F (50oC). Higher temperatures will affect the output as well as the useful life of the pump. While the MicroTron pump is suitable for most outdoor installations, do not use the standard poly tubing in direct sunlight. If you must mount pump in direct sunlight or under bright fluorescent lights use ultra-violet resistant tubing, consult your distributor or the factory. Accessory item R00225 (plastic mounting bracket) is recommend for a secure installation. B. Electrical 1. The Standard MicroTron pump has a voltage regulated internal power supply capable of operating in the range of approximately 95 to 135 VAC. Use a supply voltage of 100 to 120 VAC for best results. The 3-wire grounded plug must be used in a 3-wire wall plug.

2. With a 240 volt option, the MicroTron pump has a voltage regulated internal power supply capable of operating in the range of approximately 195 to 260 VAC. Use a supply voltage of 210 to 250 VAC for best results.

CAUTION: Never remove ground wire from plug!


C. Plumbing 1. Tubing Connections The MicroTron pump uses carefully matched components to achieve a predictable metering output. This predictability can only be maintained if all fitting sizes remain unaltered. Do not attempt to reduce tubing size. All tubing connections should be double checked to insure against leakage. If hazardous chemicals are being pumped, use shielding around discharge tubing. NOTES: 1. Clear flexible tubing is not intended for pressurized use. 2. When cutting lengths of tubing for your installation, ensure a clean, square cut. Use short lengths of tubing and as few connections as possible. 3. There is an approximate 2.5 psi capability lost for ever 1 foot of vertical rise of the discharge tubing to the injection point.

2. Tubing Nuts

Do not overtighten the tubing connectors. Tighten the fittings no more than 1/4 turn after the fitting contacts the seal. Hand tighten only. Do not use a wrench or pliers as they may damage the fittings. Do not use Teflon tape except on NPT fittings. Be sure to observe applicable local plumbing codes. 3. WARNING: Clear flexible tubing is not intended for pressurized use. Tubing Connections
3/8"
TUBING MUST SLIP OVER NIPPLE

1/4"
1/4" TUBING ADAPTER

IMPORTANT THE END OF THE TUBING


MUST EXTEND 1/4" BEYOND ADAPTOR

4.

Suction Lift vs Flooded Suction Applications Suction Lift Installation Mount the MicroTron pump around the top of the solution tank, not to exceed 5 feet from pump to bottom of tank. Flooded Suction This installation is recommended for very low outputs, solutions that gasify and/or high viscosity solutions. Priming is easier and loss of prime is reduced. Failure of the pump diaphragm or rupture of the solution tubing can cause loss of solution in the tank.

Suction lift

Flooded Suction

Not recommended

5.

Wall Mounting The fluid end portion (head assembly) of the pump is set up to accommodate mounting of the pump to the chemical container, either as a flooded suction, or a suction lift. The pump head must be kept in a vertical position for proper operation. The head can be removed and rotated 90o if needed to keep the inlet and outlet valves in a vertical position. Foot valve installation A weight is provided to hold the tubing and foot valve in a vertical position at the bottom of the tank. Do not allow the foot valve to lay horizontally in the chemical container. This defeats the action of the valve and causes the pump to lose prime. Keep suction tubing reasonably short and avoid high spots or bends. Injection valve installation The injection valve is designed to prevent a back flow and to inject chemical into the line. To work properly, this valve must be mounted within 45 degrees of vertical (see drawing). One end of the injection valve is 1/2 MNPT. Install this end into the piping system. Use Teflon tape on this fitting only. Connect the pumps discharge tubing to the opposite end of the injector. Do not use Teflon tape or joint compound on this fitting. Connect tubing between this fitting and the pump discharge fitting at the pump head.

6.

7.

NOTE: When installation is made into a line with zero pressure or when pumping into an open vessel, use the optional three function injection valve which provides back pressure and anti-syphon capabilities.

USE Teflon Tape On These Threads Only

Pipe Cross Section

DO NOT Use Teflon Tape On Machine Threads

Flow

80 Variation Acceptable

Injection Check Valve

8.

Optional Three Function Valve. The optional 3-function valve injection assembly provides three functions in one injection valve assembly.

Anti-siphon feature allows metering of liquids down hill or into the suction side of a circulating pump. It provides protection against an accidental application of suction pressure at the fluid injection point. Its Teflon coated diaphragm provides a positive anti-siphon action. Back pressure function permits metering into atmospheric discharge (open container) without overpumping. Line check permits removal of discharge tubing without release of system fluid.

9.

Priming / Degassing Valve Connection Connect the clear poly tubing to the outlet of the bleed or priming valve. Position the free end of this tube in the chemical container. Standard head configurations include a priming valve built into the head. Clear poly tubing should be connected to the outlet of this valve, the other end of the tube should be placed in the chemical container above the fluid level.

V.

Start-Up
A. Priming the pump
Plug in pump, set stroke to 100% and strokes per minute to maximum speed. While pump is operating, if fluid begins moving, no further priming is required. If fluid is not moving, open bleed valve approximately one turn until fluid begins to move. When suction line fills, close bleed valve. Do not over tighten bleed valve. Damage may occur.

B.

Adjusting feed rate

The standard Model B allows for the exact setting of the pumps stroking rate (speed) on the pumps control panel. Standard strokes per minute settings available are: 1, 2, 3, 4, 5, 6, 7, 10, 15 and increase by 5 thereafter up to the maximum of 125 (B-155 models have a max speed of 160 strokes/minutes).

C.

Stroke length

The stroke length can be adjusted on all MicroTron pumps. This adjustment is a mechanical adjustment made using the large knob on the control panel. To avoid damage to the pump, this adjustment should only be made while the pump is running at a high stroking rate.

D.

Calculating Output
A pumps output per minute can be determined by dividing the maximum rated gallons per day by 1440 (minutes per day). For example, a 30 gallons per day (gpd) pump at a maximum stroke length and speed setting of 125 strokes per minute (spm) will pump 0.000167 gallons per stroke (gps). 30 1440 = 0.0208 gpm 125 spm = 0.000167

With this value and the pumps speed setting (strokes per minute) you can calculate your pumps output at its rated pressure. A 30 gpd pump set at 50 strokes per minute: 50spm x .000167gps x 1440 (minutes per day) = 12.02 gallons per day

Reducing the stroke length will reduce the pumps output again. If the example pump above had its stroke length reduced to 50% the 12.02 gallons per day output is reduced to 6.01. (example: 12.02 gpd x 0.50 = 6.01 gpd) A higher product viscosity will reduce the output. Pressures lower than the pumps rating can increase the output.

E.

Optional Features
A pumps output per minute can be determined by dividing the maximum rated gallons per day by 1440 (minutes per day). For example, a 30 gallons per day (gpd) pump at a maximum stroke length and speed setting of 125 strokes per minute (spm) will pump 0.000167 gallons 1. External Pacing The external pacing option allows the speed of the pump to be controlled by an external device like a flow meter that gives dry contact switch closures. The pump sends out a 5 volt D.C. current to read a switch closure. Each time the pump sees the switch closure it strokes once, up to its maximum strokes per minute rate. 2. Hall Effect Meter Pacing This option is like the external pacing only the pump divides the incoming switch closures by 9 before stroking on time. This is a three wire connection: D.C. voltage, signal and ground. 3. External Stop Pumps with this option are allowed to stroke as long as they see a closed condition from a dry contact source like a flow switch or drum level wand. When an open condition is seen the pump is not allowed to stroke.

VI.

Maintenance
The Advantage MicroTron pump is designed for long service life with minimum maintenance. If for any reason, maintenance is necessary or desireable, the MicroTron pump is easily maintained. Before any maintenance or service is performed, observe the following precautions: 1. 2. 3. 4. 5. Disconnect the MicroTron pump from power source. Drain chemical from discharge tubing. Disconnect discharge tubing from pump. If the MicroTron pump is used in a flooded suction application, remove foot valve from chemical container. Observe relevant safety protocols when handling parts which have been in contact with hazardous chemicals.

A. Diaphragm Replacement
1. 2. 3. 4. 5. Remove fluid end cover by lightly prying it loose from the fluid end. Remove the four screws attaching the fluid end to pump body. Remove the fluid end from the pump body. Unscrew the diaphragm from the pump shaft in a counter-clockwise direction. Be careful that diaphragm support ring does not fall out. Do not allow sharp or abrasive objects to come in contact with pump parts.

6. 7. 8. 9.

Inspect end of shaft to assure that threads are in good condition. Replace shaft bellows if necessary. No further disassembly is recommended. Screw new diaphragm onto pump shaft until it bottoms out on shoulder of shaft. It is not necessary to tighten further. Replace fluid end. Make sure that screws are evenly tightened. Reconnect plumbing and power. Prime the pump.

B. Suction and Discharge Check Valve Replacement


1. 2. 3. 4. 5. 6 7. 8. 9. Disconnect suction tubing from pump. Unscrew fitting from pump head. Remove check valve from suction fitting and replace. Remove O-ring from cavity in fluid end. Remove check valve from suction side pump and replace. Install new O-ring in cavity of fluid end. Replace valve fitting with check valve in fluid end. Replace fluid end. Make sure that screws are evenly tightened. Reconnect plumbing and power. Prime the pump. Tighten pump head screws after pumps initial week of operation. When installing check valves, remember that the seats are always installed at the bottom.

NOTES: 1. 2.

C. Liquid End Diagram

2 14
OR

OR

11 6 1 8 12 9 10

Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

Description Complete Head Kit Injection Valve Foot Valve Suction Valve Discharge Valve Bleed Valve Assembly Pump Head Pump Diaphragm Support Ring Shaft Seal Head Bolts Priming Tube Suction Tubing Discharge Tubing Weight

4
OR

Complete Head Kit includes items: 4,5,6,7,8, and 11


13 15
OR

10

D. Replacement Parts
Getting the right materials of construction for your spare parts is easy. Using positions 7-10 of the pump model number, example: B130X1-PFCV. Find the assembly needed and add the codes of your pumps liquid end after the standard prefix part number for the assembly. Note: For standard pumps using the short model number (example: B130) use the following liquid end
codes. a. 110 psi models use -PVC1. b. 250 psi models use -PVC2.

Part Assemblies
Item Description Part Number

1.. ...................... Complete Head Assembly 110 psi....................................... CKR-1- __ __ __ __ ......................... Complete Head Assembly 150 & 250 psi................................ CKR-2- __ __ __ __ 2.. ...................... Injection Valve Assembly. .............................................................. INJ- __ __ __ __ Optional........ 3-Function Injection Valve............................................................ 3FV- __ __ __ __ 3.. ...................... Foot Valve Assembly................................................................... FTV- __ __ __ __ 4.. ...................... Suction Valve Assembly.............................................................. SUC- __ __ __ __ 5.. ...................... Discharge Valve Assembly............................................................DIS- __ __ __ __ 6.. ...................... Priming Valve Assembly................................................................PRI- __ __ __ __

Body Seat Ball Connection K- Kynar V- Viton C- Ceramic 1- 3/8 Tube S- 316 Stainless F- Teflon D- Single 2- 1/4 Tube H- Hypalon S- Stainless 3- 3/8 Tube K- 3/8 Tube P- 1/4 Pipe U- 3/8 UV V- 3/8 Clear Suction

Single Parts
7.. ................ Pump Head 110 psi................................................................. R00026 - __ ................... Pump Head 150 & 250 psi....................................................... R00039 - __ 8.. ................ Diaphragm 110 psi................................................................... R00007 ................... Diaphragm 150 & 250 psi........................................................ R00006 9.. ................ Support Ring 110 psi............................................................... R00069 ................... Support Ring 150 & 250 psi..................................................... R00068 10.. .............. Shaft Seal................................................................................ R00050 11.. .............. Head Bolts............................................................................... R00045 12.. .............. Priming Tubing ........................................................................ R00255 13.. .............. Suction Tubing 3/8 Clear ....................................................... R00255 ................... Suction Tubing 3/8 PE............................................................ R00122 ................... Suction Tubing 1/4 PE ........................................................... R00097 14.. .............. Discharge Tubing 3/8 PE........................................................ R00122 ................... Discharge Tubing 1/4 PE........................................................ R00267 15.. .............. Weight, Suction Tubing............................................................ R00139

11

VII. Trouble Shooting


PROBLEM CAUSE REMEDY Straighten suction line so as to eliminate high spots. Inspect foot valve screen and assure that foot valve is in a vertical position below fluid level. Maximum suction lift is 5 feet with water or fluids of similar specific gravity; less with heavier liquids such as acids. Mount pump in a lower position relative to the chemical container. Check fittings. Overtightening may cause restriction. Conversely, if any leakage occurs, pump will suck air and fail to prime. Inspect check valves in fluid end for cleanliness. Clean or replace as necessary. Inspect suction tube through its full length to assure that there are no splits at the connections or other restrictions. Move any objects or equipment which impinges upon suction tube or reroute as required to assure a smooth transition from foot valve to pump. Check fluid level in chemical supply tank. Check operation of stroke limiter knob. If pump delivers too low adjustable rate, check settings. Readjust as required. Inspect, clean or replace as necessary. Check suction line for obstructions, clogging, kinks or pinch points. Clean or replace foot valve screen. Relocate the injector to a lower pressure part of the the system. Replace diaphragm, making sure that it is screwed on fully to shoulder of shaft. Consult dealer or factory. Pump does not achieve Air trapped in suction line or maintain prime Foot valve contaminated or improperly installed Excessive lift

Suction fittings not properly tightened

Worn or contaminated check valves Split or pinch in suction tube

Low chemical level Insufficient fluid Stroke adjustment set too low

Worn or contaminated check valves Obstruction in suction line Clogged foot valve screen Output (system) pressure too high Diaphragm worn or torn Electronic failure

12

Excessive fluid

Failure or lack of antisiphon valve

Inspect or add anti-siphon valve. This is caused when system is in a vacuum condition or valve in delivery applications with flooded suction which feeds systems at very low pressures. Lower the stroke rate if adjustable on your pump. Reduce stroke length. Check system pressure to assure that it is within system rated paramaters of the pressure. Make sure that diaphragm is screwed fully unto shaft. Clean or replace as required. Check outlet with meter to assure that correct or voltage is present and that power supply cord is in good condition and plugged in. Consult dealer or factory. Prime pump. Add an anti-siphon valve to provide 25 PSI restriction on pump discharge.

Excessive stroke rate Improper stroke length Pump will not pump System pressure too high

Diaphragm improperly installed Check valves worn or clogged Pump will not run not Pump not turned on or not plugged in plugged in Electronic failure Excessive noise Pump not primed No output pressure

13

Application Notes:

14

Application Notes:

15

Get the Advantage in Water Treatment Equipment


Advantage Controls can give you the Advantage in products, knowledge and support on all of your water treatment equipment needs. Cooling Tower Controllers Boiler Blow Down Controllers Blow Down Valve Packages Solenoid Valves Water Meters Chemical Metering Pumps Corrosion Coupon Racks Chemical Solution Tanks Solid Feed Systems Feed Timers Filter Equipment Glycol Feed Systems Pre Fabricated Systems

Get the Advantage

16

Data Sheet MicroTron


Series B
Chemical Metering Pump

Digitally Accurate Speed Adjustment

Key Features
Dual Output Adjustment NEMA 4X Style Enclosure Guided Double Ball Checks Circuit Voltage Protection Thermal Overload Protection Priming / Degassing Valve LED Indicating Lights Splash Cover Over Controls 2 Year Warranty

Application
MicroTron Series B metering pumps provide reliable, economical chemical addition for a wide variety of applications. Precise stroking speed is maintained with a digitally accurate speed adjustment. Metering accuracy is achieved with double ball guided checks. A NEMA 4X style housing allows for outdoor installations. Series B pumps are available in outputs up to 45 GPD and 250 PSI. The manual stroke length and digital speed adjustment provides a wide turn down capability on all models.

Build a Model
Model B_ _ _ _ _ - _ _ _ _
Build your model number by selecting one code from each step. Example: B130X1-KVC1
STEP 1............... OUTPUT (MAXIMUM) 115 = 15 GPD / 2.37 LPH at 150 PSI (10.3 bar) 130 = 30 GPD / 4.70 LPH at 110 PSI (7.5 bar) 155 = 45 GPD / 7.09 LPH at 75 PSI (5.1 bar) 217 = 17 GPD / 2.60 LPH at 250 PSI (17.2 bar) STEP 2............... CONTROL OPTIONS X = Standard D = Dual Output 10-50 sph or 1-125 spm E = External Pace L = Limit Timer STEP 3 .................... VOLTAGE 1 = 120 Volt USA Plug 2 = 240 Volt 3 = 240 Volt Specified Plug 4 = 12 Volt Operation STEP 4...............HEAD MATERIAL K = Kynar S = Stainless STEP 5............... SEAT MATERIAL V = Viton F = Teflon H = Hypalon STEP 6.................. CHECK BALL C = Ceramic S = Stainless STEP 7................ CONNECTIONS 1 = 3/8 (9.525 mm) Tubing - 110 & 150 PSI 2 = 1/4 (6.35 mm) Tubing - 250 PSI 3 = 3/8 (9.525 mm) Tubing - 250 PSI P = 1/4 (6.35 mm) MNPT Connections V = 3/8 (9.525 mm) Vinyl Suction Tubing - 110 or 250 PSI For other options, see Price List or consult factory.

Specifications
Electrical
Input - 120 VAC or 240 VAC @ 50/60 Hz

Operational

Stroke length adjustment - 10-100% Stroke frequency - 1-125 strokes/minute for 15, 30 & 17 GPD; 1 - 145 strokes / min. for the 45 GPD

Enclosure

Heavy duty NEMA 4X style high impact thermoplastic with clear lexan cover.

Environment

Ambient temperature 32o to 140oF (0 to 60oC) Relative humidity 0 to 100%

Shipping Weight: Approximately 10 lbs. (4.536 kg)


10.6" 269 mm

7.5" 190 mm

4.13" 105 mm

Accessories
Standard models include everything needed for immediate start-up: injection, foot and priming valve assemblies plus suction and discharge tubing.

5.0" 127 mm

Get the Advantage

2/2010

4700 Harold Abitz Dr Muskogee, OK 74403 800-743-7431 phone 888-686-6212 fax www.advantagecontrols.com

Warranty
Thermal Care warrants its equipment to be free from defects in material and workmanship when used under recommended operating conditions. Thermal Cares obligation is limited to repair (i.e. rewind a motor) or replacement (not adjustment or maintenance), F.O.B. the factory of any parts supplied by Thermal Care within a period as shown below from the date of shipment to the original purchaser. Model SQ, LQ EQ EQR, LQR, SQR(remote condensers) TCW, TSW, TXW TCR, TSR, TXR (remote condensers) Optional Compressor Warranty Chilled Water Systems FT or FC Tower Systems FT Cooling Tower FC Cooling Tower RA, RB All other products
1 2

Parts 18 months 12 months 12 months 12 months 12 months 5 years 3 See note 3 See note 12 months (10 years - shell) 5 years (10 years - shell) (See Warranty Sheet Form 1-415) 12 months

Labor 12 months 12 months 2 12 months 12 months 2 12 months 12 months 4 12 months


4

Continental U.S.A., Canada, and Puerto Rico only. Refrigerant and any labor associated with its evacuation or replacement are not covered for remote condenser systems. 3 See individual product listing for parts warranty coverage. 4 The labor warranty covers all equipment purchased at the same time consisting of a minimum of at least one pumping system and one cooling tower and/or chiller. This warranty does not cover the cost of labor during overtime hours (after normal working hours or during weekends and holidays). Any cost differential for overtime labor will be the responsibility of the customer. Thermal Care is not responsible for any sales, use, excise or other applicable taxes associated with the replacement of parts under this warranty. This warranty will be voided when, in Thermal Cares opinion, the equipment and/or system has been subject to misuse, negligence or operation in excess of recommended limits, including freezing, or has been altered, and/or repaired without express factory authorization. If equipment is installed in hostile environments, unless such conditions were specified at the time of purchase; or the serial number has been removed or defaced, this warranty shall not apply. This warranty is not transferable. Under no circumstances shall Thermal Care be liable for loss of prospective or speculative profits, or special, indirect, incidental or consequential damages. Thermal Care must authorize all warranty service prior to work being performed and have a Thermal Care purchase order issued. All defective parts become the property of Thermal Care and must be returned as advised by Thermal Care. Thermal Care neither assumes, nor authorizes any person to assume for it, any liability not expressed in this warranty. There is an implied warranty of merchantability and of fitness for that particular purpose; all other implied warranties, and any liability not based upon contract are hereby disclaimed and excluded by this warranty. This warranty is part of the standard conditions and terms of sale of Thermal Care.

7720 N. Lehigh Avenue Niles, Illinois 60714-3491 Phone: (847) 966-2260 (888) 828-7387 Fax: (847) 966-9358

Email: info@thermalcare.com www.thermalcare.com Customer Service Phone: (847) 966-2636 Fax: (847) 966-2906

Warranty Form 1-410.11 Effective 4/9/07

Thermal Care, Inc. 7720 North Lehigh Ave. Niles, IL 60714-3491 www.thermalcare.com

New Equipment Sales Phone (847) 966-2260 Fax (847) 966-9358 Email info@thermalcare.com

Customer Service Phone (847) 966-2636 Fax (847) 966-2906 Email service@thermalcare.com

Parts Phone Fax Email

(847) 966-8560 (847) 966-6065 tcparts@thermalcare.com

TWC Series IOM 6-315.3, March 2011

Das könnte Ihnen auch gefallen