The status message can be displayed in the result table of the Rheoplus / US 200 software. It is shown, if e.g. a certain measurement option has been used or if a limit value has been reached. If a message indicates that a limit value has been reached, the measurement results should be checked carefully.
Description The pre-set frequency has been limited to the maximum value. The pre-set frequency has been limited to the minimum value. The pre-set torque has been limited to the maximum value. The pre-set speed has been limited to the maximum value. The maximum pre-set temperature of the Peltier element has been reached (SW)*. The minimum pre-set temperature of the Peltier element has been reached (SW). The maximum pre-set temperature of the temperature controller has been reached (SW). The minimum pre-set temperature of the temperature controller has been reached (SW). The pre-set normal force has been limited to the maximum value. The pre-set normal force has been limited to the minimum value. The pre-set gap speed has been limited to the maximum value (MCR only). The pre-set gap speed has been limited to the minimum value (MCR only). The pre-set gap speed has been limited to the maximum value (MCR only). The pre-set voltage output for M1 / M2 has been limited to the maximum or minimum value. The maximum electrical torque has been reached during the adjusting time. The minimum electrical torque has been reached during the adjusting time. The adjustment has been stopped because the pre-set adjusting time has been reached. The adjustment has been stopped because the maximum adjusting time has been reached (automatic settings).
Pe-
Th+
Th-
Nf+ NfVMa
GMi
GMa
DAC
WE+
WE-
Wtn
WMa
MD+
taD
DSO
Dy xxx%
DyAuto Sdy
TGC
TR
CS_xxx
* SW - Software