12CA64992-010 2013-09-23
Test Location Information Name Address UL International Germany GmbH Admiral Rosendahl Strasse 23 63263 Zeppelinheim Neu-Isenburg
Client Information Name SIKKERHEDSSTYRELSEN Address NRREGADE 63 ESBJERG , 6700, DK Report Information Report Number 12CA64992-010 Report Date Standard References Type Product Test Technician Name/Signature Joachim Freis, Andreas Richter, Stefan Scheller Laboratory Review Name/Signature Marijo Cosic 2013-09-23 IEC/EN 62109-1, Edition 1.0 2010-04, IEC/EN 62109-2, Edition 1.0 2011-06 SMA SB 3000TL-21, 3 kW Grid connected photovoltaic inverter
Product Information
Page 1 of 32
12CA64992-010 2013-09-23
General Information Information conveyed by this Report applies only to the test sample(s) actually tested. Underwriters Laboratories Inc. (UL) did not select the sample(s), determine whether the sample(s) was representative of production sample(s), nor was UL provided with information relative to the formulation or identification of component materials used in the test sample(s). UL has not established a factory Follow-Up Service Program to determine the conformance of subsequently produced products, nor has any provision been made to apply any registered mark of UL to such product(s). The issuance of this Report in no way implies any Listing, Classification or Recognition by UL and does not authorize use of ULs marks, or other reference to UL, on the product or system. ULs name and marks cannot be used in any packaging, advertising, promotion or marketing relating to the product and data in this Report, without ULs prior written permission. ULs services are provided independently of ULs Conformity Assessment services such as preliminary product investigations, product certification, or field evaluation services. UL does not provide specific product design solutions, including prescriptive revisions to existing or prospective product designs. UL does not guarantee, warranty or provide an assurance (express or implied) that a positive test result, compliance report, or issuance of a UL certification mark will result from the testing performed or the information contained in this Report. In no event shall UL be responsible to anyone for whatever use or non-use is made of the information contained in this Report and in no event shall UL, its employees, or its agents incur any obligation or liability for damages including, but not limited to, consequential damage arising out of or in connection with the use or inability to use the information contained in this Report.
Page 2 of 32
12CA64992-010 2013-09-23
Page 5
List of tests Visual inspection as received Marking and Documentation Opening in the enclosure according to fire hazards Protection against direct contact Access probe test Protective class I - Protective bonding and earthing Input output test Electrical ratings test Input output test Electrical ratings test Verification of Touch current followed by dielectric test Dielectric voltage Test Array insulation resistance detection for inverters for ungrounded and functionally grounded arrays Humidity preconditioning Dielectric voltage test after Humidity preconditioning Verification of Touch current followed by dielectric test after humidity preconditioning Mechanical resistance to deflection, impact, or drop DC injection test Cord anchorages and strain relief test Input output test Electrical ratings test
Table 2 List of tests
Page 6 7 11 12 13 14 15 16 17 18 19 20 21 22 23 25 27
Page 28 30 31
Page 3 of 32
12CA64992-010 2013-09-23
Sample Identification
Sample No. Sample Identification Number 1580006 1710326 Date Received Product Description /Serial Number SMA SB 3000TL-21, 3 kW/ 2130110507 SMA SB 3000TL-21, 3 kW/ 2130111294
10 10.1
2013-02-15 2013-09-06
Page 4 of 32
12CA64992-010 2013-09-23
For project 12CA64992 10 grid connected photovoltaic inverters were provided for specific tests according to IEC/EN 62109-1 and IEC/EN 62109-2. The test program was designed to test various properties which could have an impact to safety and performance. For each inverter a report with the extended project number from 1 10 is provided where each report is covering one of the provided inverters.
Test result summary The following table is a summary of all performed tests with an indicator of the impact of the test result. Green indicates a pass of the standard requirement yellow indicates that the standard requirement is not fulfilled but considered as a minor issue where red indicates a non-conformity with a major impact to safety. Due to the damage of the first delivered sample during normal operation some tests were repeated with a new provided sample of same type. Tests Name Visual inspection as received Marking and Documentation Opening in the enclosure according to fire hazards Protection against direct contact Access probe test Protective class I - Protective bonding and earthing Input output test Electrical ratings test Verification of Touch current followed by dielectric test Dielectric voltage Test Array insulation resistance detection for inverters for ungrounded and functionally grounded arrays Humidity preconditioning Dielectric voltage test after Humidity preconditioning Verification of Touch current followed by dielectric test after humidity preconditioning Mechanical resistance to deflection, impact, or drop DC injection test Cord anchorages and strain relief test
Table 5 Summary of test results
Test Result
Page 5 of 32
12CA64992-010 2013-09-23
Results
Outcome
The sample does not show any sign of visual damage.
Page 6 of 32
12CA64992-010 2013-09-23
Sample No.
10
N/A
N/A
N/A
Page 7 of 32
Identification (5.1.3)
Pass
Fail
N/A
Results
N/A
General (5.3.1)
Visibility and legibility requirements for warning markings (5.2.1) Content for warning markings (5.2.2) Sonic hazard markings and instructions (5.2.3)
Equipment with multiple sources of supply (5.2.4) Excessive touch current (5.2.5)
12CA64992-010 2013-09-23
Outcome
Page 9 of 32
12CA64992-010 2013-09-23
The durability of marking test was conducted on two different labels where both tests lead a result not in compliance with the standard.
According to IEC/EN 62109-1 sub clause Marking and documentation the following items arent stated out or provided: Pollution degree (according to sub clause 5.3.1) a warning that when the photovoltaic array is exposed to light, it supplies a d.c. voltage to the PCE (according to sub clause 5.3.1) explanation of the pin-out of connectors for external connections, unless the connector is used for a standard purpose (e.g. RS 232) (according to sub clause 5.3.2) PV modules for non-isolated inverters: Non-isolated inverters shall be provided with installation instructions that require PV modules that have an IEC 61730 Class A rating (according to sub clause 5.3.2.5)
Page 10 of 32
12CA64992-010 2013-09-23
Pass
Fail
N/A
Results
Sample No. 10
Page 11 of 32
12CA64992-010 2013-09-23
Protection against direct contact Access probe test Description and Setup
Access probe test was performed according to IEC/EN 62109-1, Edition 1.0 2010-04 (sub clause 7.3.4). Protection against direct contact is provided to prevent persons from touching live parts. The test was performed with the test finger and test pin according to Annex D, IEC/EN 62109-1.
P Pass F Fail N/A Standard clause not applicable
Results
Sample No. 10
Result P
Page 12 of 32
12CA64992-010 2013-09-23
Results
Sample No. 10 Measured impedance / [m] 9
Table 15 Protective bonding results
Outcome
Measured protective bonding impedance is acceptable, measured value is below the maximum allowed permissible value.
Page 13 of 32
12CA64992-010 2013-09-23
2013-03-12 11054/C/%RH/mBar 11029/400A 11043/1000:1 11048/600V/20mA/W 1013/5A/10A//50mA 500V/1000V/50mV/500mV/5V 287/25A 35/50A Date 2013-03-12
Results
Sample 10 U DC [Vdc] 175.19 500.75 I DC [Adc] 17.95 6.24 P DC [kW] 3.14 3.13 U grid [V] 230.69 230.70 I AC [A] 13.01 13.02 P AC [kW] 3.00 3.00
Outcome
Sample did not exceed the power by more than 10%.
Page 14 of 32
12CA64992-010 2013-09-23
2013-09-12 11054/C/%RH/mBar; 131/300V/25A; 1009/1000Vdc; 1003/5A/10A/50mA/500V/1000V/50mV/ 500mV/5V; 287/25A Ambient / [C] 22.9 Date 2013-09-12
Results
Sample 10.1
Outcome
Sample did not exceed the power by more than 10%.
Page 15 of 32
12CA64992-010 2013-09-23
2013-09-12 11054/C/%RH/mBar 131/600V/25A 1013/5A/10A//50mA 500V/1000V/50mV/500mV/5V 64/Leakage current network Date 2013-09-12
Results
Measured Value / [mAac] Measured Value / [Adc] Result
Sample 10.1
4.52
220
Outcome
The measured touch current is not in the allowed limits.
Page 16 of 32
12CA64992-010 2013-09-23
DIELECTRICVOLTAGE TEST
Results
Sample
Test voltage Location Result
DC pins to dead metal AC pins to dead metal AC pins to communication line DC pins to communication line
P P N/A N/A
Outcome
The sample withstood the applied dielectric voltage The dielectric voltage was not applicable to communication line as there is no communication port accessible.
Page 17 of 32
ARRAY INSULATION RES DETECTION FOR INV FOR UNGROUNDED AND FUNCTIONALLY GROUNDED ARRAYS
12CA64992-010 2013-09-23
Array insulation resistance detection for inverters for ungrounded and functionally grounded arrays Description and Setup
The Array insulation resistance detection for inverters for ungrounded and functionally grounded arrays were conducted according to IEC/EN 62109-2, Edition 1.0 2011-06. Inverters for use with ungrounded arrays shall have means to measure the DC insulation resistance from the PV input (array) to ground before starting operation.
P Pass F Fail N/A Standard clause not applicable
Results
Sample No. 10.1 Result P
Table 23 Array insulation test results
Outcome
Display shows: ISO-Resistance ERROR CODE 3501. Inverter does not start to generate power. Internal insulation monitor is able to detect an insulation fault.
Page 18 of 32
12CA64992-010 2013-09-23
HUMIDITY PRECONDITIONING
Results
Sample 10.1 Conditions 42 C for 4h / 40C, 92%RH for 48h / 23C for 2h
Table 24: Humidity preconditioning
Page 19 of 32
12CA64992-010 2013-09-23
Results
Sample
Test voltage Location Result
DC pins to dead metal AC pins to dead metal AC pins to communication line DC pins to communication line
P P N/A N/A
Outcome
The sample withstood the applied dielectric voltage after humidity. The dielectric voltage was not applicable to communication line as there is no communication port accessible.
Page 20 of 32
12CA64992-010 2013-09-23
2013-09-16 11054/C/%RH/mBar 131/600V/25A 1013/5A/10A//50mA 500V/1000V/50mV/500mV/5V 64/Leakage current network Date 2012-09-16
Verification of Touch current followed by dielectric test after humidity preconditioning Description and Setup
The touch current followed by a dielectric test was verified according to IEC/EN 62109-1, Edition 1.0 2010-04. The touch current shall not be greater than 3.5 mA a.c. or 10 mA d.c.
P Pass F Fail N/A Standard clause not applicable
Results
Measured Value / [mAac] Measured Value / [Adc] Result
Sample 10.1
4.48
220
Outcome
The measured touch current is not in the allowed limits.
Page 21 of 32
12CA64992-010 2013-09-23
Sample Tested:
Results
250 N deflection test for metal enclosures 13.7.2 P 7 J impact test for polymeric enclosures 13.7.3 N/A Drop Test 13.7.4
Sample No. 10
N/A
Page 22 of 32
12CA64992-010 2013-09-23
DC INJECTION TEST
2013-09-12 1054/C/%RH/mBar 11048/600V/20mA/W 1013/5A/10A//50mA 500V/1000V/50mV/500mV/5V 64/Leakage current network 287/25A Shunt 35/50A Shunt Date 2012-09-12
Results
Sample No. 50% (Idc) PE connected / [mA] 10mAdc 10.1 100% (Idc) PE connected / [mA] 15mAdc 50% (Idc) / [A] 207
Table 32: Results - DC injection test
Page 23 of 32
12CA64992-010 2013-09-23
DC INJECTION TEST
Outcome
It is not allowed to use an RCD type A, as the maximum d.c. current is over 6 mA. In this case a RCD Type B has to be applied.
Page 24 of 32
12CA64992-010 2013-09-23
Tested by:
Sample Tested:
Test Date:
2013-03-22
Instrument Code/Range:
Results
Sample No. 10
Test Passed
F
Table 33 Strain relief test results
Page 25 of 32
12CA64992-010 2013-09-23
Outcome
The connectors did not withstand strain relief test.
Page 26 of 32
12CA64992-010 2013-09-23
2013-03-12 11054/C/%RH/mBar 11029/400A 11043/1000:1 11048/600V/20mA/W 1013/5A/10A//50mA 500V/1000V/50mV/500mV/5V 287/25A 35/50A Date 2013-03-12
Results
Sample 10 U DC [Vdc] 175.19 500.75 I DC [Adc] 17.95 6.24 P DC [kW] 3.14 3.13 U grid [V] 230.69 230.70 I AC [A] 13.01 13.02 P AC [kW] 3.00 3.00
Outcome
Sample did not exceed the power by more than 10%.
Page 27 of 32
12CA64992-010 2013-09-23
Due
2013-01-11 2014-01-11 2013-04-05
11029
Current Transducer
11054
11043
2012-12-07
2013-12-07
11048
2012-10-02
2013-10-12
1013
Digital Multimeter
2012-09-25
2013-09-25
1009
Digital Multimeter
Fluke 289
2012-12-03
2013-12-03
1003
Digital Multimeter
Fluke 289 Power Analyzer WT 200 Shunt 25A-4mV Leakage Current Network IEC 60950-1 / IEC 61010-1 A1 Combined Insulation Resistance and Dielectric Tester HA 2201G
2012-10-30
2013-10-30
131
Power analyzer
2013-08-29
2014-08-29
284
2013-05-21
2015-05-21
64
2013-06-11
2014-06-11
11026
Dielectric tester
2013-01-24
2014-01-24
Page 28 of 32
12CA64992-010 2013-09-23
Due
2015-02-17
233
Caliper
CSV/2
183
Test Finger
Standard Test Finger Test Pin IEC60601 Fig. 8 / Test Pin IEC60950 Fig. 2B Shunt 25A-4mV
2012-04-18
2013-04-18
599
2011-05-13
2014-05-13
287
Shunt
2012-11-27
2013-11-27
35 2101
Page 29 of 32
12CA64992-010 2013-09-23
APPENDIX - LABELS
Appendix Labels
Picture 4 Label
Page 30 of 32
12CA64992-010 2013-09-23
APPENDIX - PICTURES
Appendix Pictures
Page 31 of 32
12CA64992-010 2013-09-23
Page 32 of 32