DEFIBRILLATOR SYSTEM

Two-in-one
The Cardio-Aid 200 is especially designed for defibrillation in-hospital by all levels of user. Whether in pads or paddles versions it offers unparalleled ease of use to whoever is first at a cardiac arrest. Advanced manual features for professional users are immediately available behind a discrete flap; manual defibrillation, cardioversion, internal defibrillation and pacing functions without adding complexity and cost.

Designed for the user


The semi-automatic mode offers a straight forward and guided two-button operation for extreme simplicity and safety. By flipping the quick guide flap, professional users get full access to the dedicated buttons for manual functions. The ergonomic paddles offer complete energy, charge, and shock controls.

Always ready for rescues


Combined with the automatic wake-up self-test, the Cardio-Aid 200 offers unique reliability, availability and performance. The unit is ready to deliver a full 360J shock in just 5 seconds from pressing the charge key. The built-in power supply, battery, and quick charge facility with IBMS (Intelligent Battery Management System) charges depleted battery to full in one hour, ensuring the unit is always ready.

Features
The high quality display, integrated recorder, and memory card document the situation as the rescue process progresses. Events can be stored and later transferred to Codelog for detailed analysis.

Cardio-Aid 200

WHEN PATIENT SAFETY HAS A NAME...TM

SPECIFICATIONS
GENERAL
Size: 31 (h) x 36 (w) x 25 (d) cm Weight: 7.6 kg incl. defibrillation electrodes and battery 9 kg incl. paddles, battery, pacer and recorder Power: 100-240 VAC, 50/60 Hz, max. 270W Temperature: Operating: 0 to 40C. Storage: -10C to +60C (<168h) Humidity: Operating: 30 to 95% RH, non-condensing. Storage: 30 to 95% RH, non-condensing Operating and storage pressure: 860 kPa to 1060 kPa Classification: Class I/Internally powered equipment Water protection: Drip-proof equipment (IP22) Mode of operation: Continuous monitoring CMRR (Common mode rejection ratio): Typically 110 dB (50/60 Hz) (with notch filter) Differential input: +/-5mV Overload recovery time: < 4 sec typically CMDR (Common mode dynamic range): +/-5 V Differential dynamic range: +/-360 mV Classification: CF, defi-proof QRS detection trigger level: Typically 200 V Heart rate alarm limits (Manual/Pacer mode, adjustable in step of 1 b/min). ECG manual alarm upper limit: Max 300 b/min ECG manual alarm lower limit: Min 30 b/min Heart rate, alarm delay: Low and high limit: <10 sec. HR display: Steps of 1 b/min

RECORDER (OPTION)
Printout modes: Manual or automatic (selectable) Printout time: Selectable: 15, 20, 30 sec, continuous Speed: 25 mm/sec Paper: Thermosensitive, 50 mm (w) x approx. 25 m (l) Mode: ECG delayed by 8 sec Weight: 370 g

MEMORY CARD (OPTION)


Size: 256KByte-16MByte Data storage time: Min. 1 year Type: SRAM 68 pin Conforming to JEIDA 4.0 Card Standard PCMCIA 1.0 Storage capacity (approximate number of sequences of 28 sec): 256k=35; 512k=70; 1M=140; 2M=280; 4M=570; 8M=1140; 16M=2280

ECG VIA DEFIBRILLATION ELECTRODES


Mode: Differential Frequency response (-3 dB): Display, 2 to 25 Hz; Filter 2 to 30 Hz; Monitor 0.5 to 30 Hz; No filter 0.05 to 105 Hz at 0.5 mVpp input Impedance range: 20 to 2000 measured at 110 kHz Pace pulse limitation: Slew rate limitation without marker

LCD DISPLAY
Size: 120 (w) x 90 (h) mm Resolution: 320 x 240 pixels Mode: Non-fade, erasing bar. ECG window: 4.0 sec ECG speed: 25 mm/sec Type: Black and white LCD display with CCFL backlight

EXT. INTERFACE FOR REMOTE SERVICE (OPTION)

ECG VIA ECG ELECTRODES


Mode: Differential Frequency response (-3 dB): Display, 0.25 to 25 Hz; Recorder, 0.25 to 30 Hz Impedance: 20 to 2000 at 110 kHz Pace pulse limitation: Discriminator with marker. Max. pace rate 180 b/min

External interface via PCMCIA port: PCMCIA to RS 232 adaptor card 2 port Parity: none Data bits: 8 Stop bit: 1 Flow control: XON/XOFF Baud rate: 9600

DEFIBRILLATOR
Pulse: Monophasic pulse, 3.2 msec wide at 50% amplitude in 50 load at 360 J Energy settings: AED: #1=200 J, #2=200 J, #3=360 J or as set in configuration menu External paddles and defibrillation electrodes: Selectable: 2, 5, 7, 10, 20, 30, 40, 50, 60, 80, 100, 150, 200, 300, 360 J Internal paddles: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 40, 50 J Energy availability: Auto disarm after 30 sec Charging time (max. energy) at 15-35C : <5 sec running on mains supply. <7 sec running on battery Synchronization: Internal generation of QRS sync Synchronization shock delay: Internal pulse: < 60 msec from top of QRS Synchronization marker: S with a vertical line Classification defibrillation electrodes: BF

Specifications subject to change. CE marked according to Council Directive 93/42 EEC of 14 June 1993 concerning medical devices.

ECG FROM MONITOR


Dynamic range: +/-5V (1V/mV signal) Max ECG signal delay in monitor: 5 msec

BATTERY
Type: Built-in, quick-charge NiCd, 12 V 1.7 Ah Operating time at 20C: 2 h or 50 shocks of 360 J Charging time for fully depleted battery: 1 h Life time: Min. 18 months at normal usage (i.e. up to one charge cycle per day)

PACER (OPTION)
Mode: VVI demand or VVO fixed Pulse type: Rectangular, constant current Pulse duration: 40 msec +/-10% Pulse amplitude: Variable 20-200 mA +/-10%. Default 0 mA Pulse adjust step size: +/-1mA (+/-5 mA after 1 sec) Pacing rate: Variable 30-180 p/min +/-5%. Default 60 p/min Pace rate adjust step size: +/-1p/min (+5-20 p/min after 1 sec) Output voltage: Max. 350 V

ECG
Lead selector: Selectable: I, II, III, P, Ext QRS volume: Selectable 0 to 5 Gain: x.25, x.5, x1, x2, x4, auto (default setting)

Artema Monitoring and Emergency Care A/S Herstedvang 8 . DK-2620 Albertslund . Denmark Telephone: +45 43 62 33 66 . Fax: +45 43 62 10 01 www.artema.dk . E-mail: artema@artema.dk