Vacuum Drying
HS-60VD
HS-60VD
BASKET
HANSHIN 28
Vacuum Drying
Vertical STEAM STERILIZER
HANSHIN MEDICAL
HS-60VD
THE MOST EFFECTIVE VERTICAL STEAM STERILIZER FOR LIQUIDS AT CLINICAL LABORATORY. VACUUM DRYING SYSTEM MINIMIZE THE WET PACK PROBLEM.
FEATURE
Though this sterilizer employs the same control system as the tabletop, it may be used more effectively and conveniently according to purpose, because of following features.
1. Liquid sterilization require special condition than other sterilization. So, this sterilizer adopted liquid cycle program with dual temperature sensor (chamber and liquid sensor) which controls sterilization temperature (121) and exhaust temperature (80) of the liquid contained bottle. 2. User doesn't need to open the door when drying phase due to its use of vacuum pump, hence moisture retained by the fabric package shall not exceed 2% than original weight after drying. 3. This sterilizer, adopting vertical chamber, may enlarge its capacity more than tabletop sterilizer. It may be used conveniently even at narrow place for featuring large capacity. 4. It can process most safely culture medium or the glassware in the clinical examination or the laboratory of the research institute; especially, it is useful when processing the liquids contained in a large glass bottle.
CYCLE PROGRAM
CYCLE NO.
LOADS TYPE
5
Liquids 30 min Refer to note 1
Unwrapped Wrapped Unwrapped Wrapped 5 min 15 min 15 min 30 min 20 min 30 min 30 min 30 min
EXPOSURE TEMPERATURE 134.0 134.0 121.0 121.0 121.0 EXPOSURE TIME DRY TIME VENT
SPECIFICATIONS
MODEL
OVERALL SIZE CABINET CHAMBER
HS-60VD
W575 X H1,080 X D531 mm W575 X H807 X D494 mm
340 X D723 mm, Volume 60 , Vertical Type, SUS 316
Fast Exhaust
Note 1 : Chamber will be exhausted when the temperature inside of bottle decrease less than 80.
CYCLE GRAPH
AC230 V, 50/60 Hz 2,950 W Saturated Steam of max. 2.10 bar Pressure Microprocessor 121.0 ~ 134.0 Flow Rate : 40 /min Min. filtration efficiency : 99.97% for 0.3 particles Dial Type, 760 mmHg ~4 bar Optional : Model STP-103, Thermal Printing, 200 x 200 dpi, W58 x 60 mm Paper Roll 108 kg Altitudes up to 2,000 meters. Temperature range of +5 to +40 . Maximum relative humidity of 80 %. Main supply voltage fluctuation of 10 % of normal. Installation Category (Overvoltage Category) II, Pollution Degree 2.
STERILIZING AGENT CONTROL EXPOSURE TEMPERATURE VACUUM PUMP AIR FILTER PRESSURE GAUGE
DRYING ABILITY Less Than 2.0% of the amount of moisture within porous loads
ENVIRONMENT CONDITIONS
CYCLE No.4 WRAPPED CYCLE No.5 LIQUID