Bus-Powered M Series Multifunction DAQ for USB - 16-Bit, up to 400 kS/s, up to 32 Analog Inputs, Isolation
Up to 32 analog inputs at 16 bits, up to 400 kS/s (250 kS/s scanning) Up to 2 analog outputs at 16 bits Up to 32 TTL/CMOS digital I/O lines Two 32-bit, 80 MHz counter/timers
NI-PGIA 2 and NI-MCal calibration technology for improved measurement accuracy NI signal streaming for 4 high-speed data streams on USB Bus-powered design Available with CAT I isolation
Overview
With recent bandwidth improvements and new innovations from National Instruments, USB has evolved into a core bus of choice for measurement and automation applications. NI M Series devices for USB deliver high-performance data acquisition in an easy-to-use and portable form factor through USB ports on laptop computers and other portable computing platforms. National Instruments designed the new and innovative patent-pending NI signal streaming technology that enables sustained bidirectional high-speed data streams on USB. The new technology, combined with advanced external synchronization and isolation, helps engineers and scientists achieve high-performance applications on USB. NI M Series bus-powered multifunction data acquisition (DAQ) devices for USB are optimized for superior accuracy in a small form factor. They provide an onboard NI-PGIA 2 amplifier designed for fast settling times at high scanning rates, ensuring 16-bit accuracy even when measuring all available channels at maximum speed. All bus-powered devices have a minimum of 16 analog inputs, digital triggering, and two counter/timers. USB M Series devices are ideal for test, control, and design applications including portable data logging, field monitoring, embedded OEM, in-vehicle data acquisition and academic. Back to Top
Driver Information
NI-DAQmx NI-DAQmx Base
Software Compatibility
ANSI C/C++ LabVIEW LabWindows/CVI Measurement Studio SignalExpress Visual Basic .NET Visual C# Back to Top
Comparison Tables
Family Connector Analog Inputs
16 16 16
Resolution (bits)
16 16 16
Analog Outputs
2 2
Resolution (bits)
16 16
Digital I/O
4 DI/4 DO 4 DI/4 DO 24 or 32 DIO -
Isolation
1/17
www.ni.com
Family
Connector
Analog Inputs
16 16 32
Resolution (bits)
16 16 16
Analog Outputs
2 2 2
Resolution (bits)
16 16 16
Digital I/O
4 DI/4 DO 24 or 32 DIO 8 DI/8 DO
Isolation
60 V, CAT I 60 V, CAT I 60 V, CAT I Back to Top
Back to Top
Ordering Information
For a complete list of accessories, visit the product page on ni.com.
Products
Part Number
Recommended Accessories
No accessories required. No accessories required. No accessories required. No accessories required. No accessories required. No accessories required.
Part Number
2/17
www.ni.com
Back to Top
Software Recommendations
LabVIEW Professional Development System for Windows
Advanced software tools for large project development Automatic code generation using DAQ Assistant and Instrument I/O Assistant Tight integration with a wide range of hardware Advanced measurement analysis and digital signal processing Open connectivity with DLLs, ActiveX, and .NET objects Capability to build DLLs, executables, and MSI installers
Quickly configure projects without programming Control over 400 PC-based and stand-alone instruments Log data from more than 250 data acquisition devices Perform basic signal processing, analysis, and file I/O Scale your application with automatic LabVIEW code generation Create custom reports or easily export data to LabVIEW, DIAdem or Microsoft Excel
Real-time advanced 2D graphs and charts Complete hardware compatibility with IVI, VISA, DAQ, GPIB, and serial Analysis tools for array manipulation, signal processing statistics, and curve fitting Simplified cross-platform communication with network variables Measurement Studio .NET tools (included in LabWindows/CVI Full only) The mark LabWindows is used under a license from Microsoft Corporation.
Customizable graphs and charts for WPF, Windows Forms, and ASP.NET Web Forms UI design Analysis libraries for array operations, signal generation, windowing, filters, signal processing Hardware integration support with native .NET data acquisition and instrument control libraries Automatic code generation for all NI-DAQmx data acquisition hardware Intelligent and efficient data-logging libraries for streaming measurement data to disk Support for Microsoft Visual Studio .NET 2012/2010/2008
Back to Top
Calibration
NI measurement hardware is calibrated to ensure measurement accuracy and verify that the device meets its published specifications. To ensure the ongoing accuracy of your measurement hardware, NI offers basic or detailed recalibration service that provides ongoing ISO 9001 audit compliance and confidence in your measurements. To learn more about NI calibration services or to locate a qualified service center near you, contact your local sales office or visit ni.com/calibration.
Technical Support
Get answers to your technical questions using the following National Instruments resources. Support - Visit ni.com/support to access the NI KnowledgeBase, example programs, and tutorials or to contact our applications engineers who are located in NI sales offices around the world and speak the local language. Discussion Forums - Visit forums.ni.com for a diverse set of discussion boards on topics you care about. Online Community - Visit community.ni.com to find, contribute, or collaborate on customer-contributed technical content with users like you.
Repair
While you may never need your hardware repaired, NI understands that unexpected events may lead to necessary repairs. NI offers repair services performed by highly trained technicians who quickly return your device with the guarantee that it will perform to factory specifications. For more information, visit ni.com/repair.
Extended Warranty
NI offers options for extending the standard product warranty to meet the life-cycle requirements of your project. In addition, because NI understands that your requirements may change, the extended warranty is flexible in length and easily renewed. For more information, visit ni.com/warranty.
OEM
NI offers design-in consulting and product integration assistance if you need NI products for OEM applications. For information about special pricing and services for OEM customers, visit ni.com/oem.
Alliance
3/17
www.ni.com
Our Professional Services Team is comprised of NI applications engineers, NI Consulting Services, and a worldwide National Instruments Alliance Partner program of more than 700 independent consultants and integrators. Services range from start-up assistance to turnkey system integration. Visit ni.com/alliance. Back to Top
Detailed Specifications
Specifications listed below are typical at 25 C unless otherwise noted. Refer to the NI USB-621x User Manual for more information about USB-621x devices. Caution The input/output ports of this device are not protected for electromagnetic interference due to functional reasons. As a result, this device may experience reduced measurement accuracy or other temporary performance degradation when connected cables are routed in an environment with radiated or conducted radio frequency electromagnetic interference. To ensure that this device functions within specifications in its operational electromagnetic environment and to limit radiated emissions, care should be taken in the selection, design, and installation of measurement probes and cables.
Analog Input
Number of channels USB-6210/6211/6212/6215/6216 USB-6218 ADC resolution DNL INL Sampling rate Maximum USB-6210/6211/6215/6218 USB-6212/6216 Minimum Timing accuracy Timing resolution Input coupling Input range Maximum working voltage for analog inputs (signal + common mode) CMRR (DC to 60 Hz) Input impedance Device on AI+ to AI GND AI- to AI GND Device off AI+ to AI GND AI- to AI GND Input bias current Crosstalk (at 100 kHz) Adjacent channels Non-adjacent channels Small signal bandwidth (-3 dB) USB-6210/6211/6215/6218 USB-6212/6216 Input FIFO size Scan list memory Data transfers Overvoltage protection (AI <0..31>, AI SENSE) Device on 30 V for up to two AI pins 450 kHz 1.5 MHz 4,095 samples 4,095 entries USB Signal Stream, programmed I/O -75 dB -90 dB 1200 1200 100 pA >10 G in parallel with 100 pF >10 G in parallel with 100 pF 250 kS/s single channel, 250 kS/s multichannel (aggregate) 400 kS/s single channel, 400 kS/s multichannel (aggregate) 0 S/s 50 ppm of sample rate 50 ns DC 10 V, 5 V, 1 V, 0.2 V 10.4 V of AI GND 100 dB 8 differential or 16 single ended 16 differential or 32 single ended 16 bits No missing codes guaranteed Refer to the AI Absolute Accuracy Tables
4/17
www.ni.com
Device off Input current during overvoltage condition Settling Time for Multichannel Measurements Accuracy, full scale step, all ranges USB-6210/6211/6215/6218 90 ppm of step (6 LSB) 30 ppm of step (2 LSB) 15 ppm of step (1 LSB) USB-6212/6216 90 ppm of step (6 LSB) 30 ppm of step (2 LSB) 15 ppm of step (1 LSB) Typical Performance Graphs
Analog Output
Number of channels USB-6210 USB-6211/6212/6215/6216/6218 DAC resolution DNL Monotonicity Maximum update rate 1 channel 2 channels Timing accuracy 250 kS/s 250 kS/s per channel 50 ppm of sample rate 0 2 16 bits 1 LSB 16 bit guaranteed
5/17
www.ni.com
Timing resolution Output range Output coupling Output impedance Output current drive Overdrive protection Overdrive current Power-on state Power-on glitch Output FIFO size Data transfers AO waveform modes: Non-periodic waveform Periodic waveform regeneration mode from onboard FIFO Periodic waveform regeneration from host buffer including dynamic update Settling time, full scale step 15 ppm (1 LSB) Slew rate Glitch energy Magnitude Duration
50 ns 10 V DC 0.2 2 mA 30 V 2.4 mA 20 mV 1 V for 200 ms 8,191 samples shared among channels used USB Signal Stream, programmed I/O
32 s 5 V/s
100 mV 2.6 s
AbsoluteAccuracy = Reading (GainError) + Range (OffsetError) + NoiseUncertainty GainError = ResidualAIGainError + GainTempco (TempChangeFromLastInternalCal) + ReferenceTempco (TempChangeFromLastExternalCal) OffsetError = ResidualAIOffsetError + OffsetTempco (TempChangeFromLastInternalCal) + INL_Error
Absolute accuracy at full scale on the analog input channels is determined using the following assumptions:
TempChangeFromLastExternalCal = 10 C TempChangeFromLastInternalCal = 1 C number_of_readings = 100 CoverageFactor = 3 For example, on the 10 V range, the absolute accuracy at full scale is as follows: GainError = 75 ppm + 7.3 ppm 1 + 5 ppm 10 GainError = 132 ppm OffsetError = 20 ppm + 34 ppm 1 + 76 ppm OffsetError = 130 ppm
Sensitivity is the smallest voltage change that can be detected. It is a function of noise.
Accuracies listed are valid for up to one year from the device external calibration.
6/17
www.ni.com
AbsoluteAccuracy = Reading (GainError) + Range (OffsetError) + NoiseUncertainty GainError = ResidualAIGainError + GainTempco (TempChangeFromLastInternalCal) + ReferenceTempco (TempChangeFromLastExternalCal) OffsetError = ResidualAIOffsetError + OffsetTempco (TempChangeFromLastInternalCal) + INL_Error
Absolute accuracy at full scale on the analog input channels is determined using the following assumptions:
TempChangeFromLastExternalCal = 10 C TempChangeFromLastInternalCal = 1 C number_of_readings = 100 CoverageFactor = 3 For example, on the 10 V range, the absolute accuracy at full scale is as follows: GainError = 75 ppm + 7.3 ppm 1 + 5 ppm 10 GainError = 132 ppm OffsetError = 20 ppm + 34 ppm 1 + 76 ppm OffsetError = 130 ppm
Sensitivity is the smallest voltage change that can be detected. It is a function of noise.
Accuracies listed are valid for up to one year from the device external calibration.
Reference Tempco 5
Absolute INL Error Accuracy at Full (ppm of Range) Scale1 (V) 3,512 118.0
Absolute Accuracy at full scale numbers is valid immediately following internal calibration and assumes the device is operating within 10 C of the last external calibration. Accuracies listed are valid for up to one year from the device external calibration. AbsoluteAccuracy = OutputValue (GainError) + Range (OffsetError) GainError = ResidualGainError + GainTempco (TempChangeFromLastInternalCal) + ReferenceTempco (TempChangeFromLastExternalCal) OffsetError = ResidualOffsetError + AOOffsetTempco (TempChangeFromLastInternalCal) + INL_Error
Digital I/O/PFI
Static Characteristics Number of channels Digital input USB-6210/6211/6215 USB-6218 Digital output USB-6210/6211/6215 USB-6218 Digital input or output USB-6212/6216 Screw Terminal 32 total, 16 (P0.<0..15>), 16 (PFI <0..7>/P1.<0..7>, PFI <8..15>/P2.<0..7>) 4 (PFI <4..7>/P1.<0..3>) 8 (PFI <4..7>/P1.<0..3>, PFI <12..15>/P1.<4..7>) 4 (PFI <0..3>/P0.<0..3>) 8 (PFI <0..3>/P0.<0..3>, PFI <8..11>/P0.<4..7>)
7/17
www.ni.com
USB-6212/6216 Mass Termination/BNC Ground reference Pull-down resistor USB-6210/6211/6215/6218 USB-6212/6216 Input voltage protection1 PFI Functionality USB-6210/6211/6215/6218 PFI <0..3>, PFI <8..11>/Port 0 Functionality Debounce filter settings PFI <4..7>, PFI <12..15>/Port 1 Functionality Timing output sources USB-6212/6216 PFI <0..15> Functionality Timing output sources Debounce filter settings
Static digital input, timing input 125 ns, 6.425 s, 2.56 ms, disable; high and low transitions; selectable per input
Static digital output, timing output Many AI, AO, counter timing signals
Static digital input, static digital output, timing input, timing output Many AI, AO, counter timing signals 125 ns, 6.425 s, 2.56 ms, disable; high and low transitions; selectable per input
Maximum Operation Conditions Level IOL output low current IOH output high current Min Max 16 mA -16 mA
Digital Input Characteristics (USB-6210/6211/6215/6218) Level VIL input low voltage VIH input high voltage IIL input low current (Vin = 0 V) IIH input high current (Vin = 5 V) Min 0V 2V Max 0.8 V 5.25 V -10 A 120 A
Digital Input Characteristics (USB-6212/6216) Level VIL input low voltage VIH input high voltage IIL input low current (Vin = 0 V) IIH input high current (Vin = 5 V) Positive-going threshold (VT+) Negative-going threshold (VT-) Delta VT hysteresis (VT+ - VT-) Min 0V 2.2 V 0.8 V 0.2 V Max 0.8 V 5.25 V -10 A 250 A 2.2 V
8/17
www.ni.com
VOL
0.6 V 2.7 V
6 mA -16 mA -6 mA
VOH 3.8 V
General-Purpose Counter/Timers
Number of counter/timers Resolution Counter measurements Position measurements Output applications Internal base clocks External base clock frequency Base clock accuracy Inputs Routing options for inputs USB-6210/6211/6215/6218 USB-6212/6216 FIFO Data transfers PFI <0..3>, PFI <8..11>, many internal signals PFI <0..15>, many internal signals 1,023 samples USB Signal Stream, programmed I/O 2 32 bits Edge counting, pulse, semi-period, period, two-edge separation X1, X2, X4 quadrature encoding with Channel Z reloading; two-pulse encoding Pulse, pulse train with dynamic updates, frequency division, equivalent time sampling 80 MHz, 20 MHz, 0.1 MHz 0 MHz to 20 MHz 50 ppm Gate, Source, HW_Arm, Aux, A, B, Z, Up_Down
Frequency Generator
Number of channels Base clocks Divisors Base clock accuracy Output can be available on any output PFI terminal. 1 10 MHz, 100 kHz 1 to 16 50 ppm
9/17
www.ni.com
Counter/timer functions
Bus Interface
USB USB Signal Stream (USB) USB 2.0 Hi-Speed or Full-Speed2 4, can be used for analog input, analog output, counter/timer 0, counter/timer 1
Power Limits
+5 V terminal as output3 Voltage Current (internally limited) +5 V terminal as input3 Voltage Current Caution Do not exceed 16 mA per DIO pin. Protection 10 V 4.75 to 5.35 V 350 mA max, self-resetting fus 4.6 to 5.2 V 50 mA max, shared with digital outputs
Power Requirements
Input voltage on USB-621x USB port Maximum inrush current No load typical current Maximum load Typical current Suspend current 400 mA at 4.5 V 260 A, typical 4.5 to 5.25 V in configured state 500 mA 320 mA at 4.5 V
Physical Characteristics
Enclosure dimensions (includes connectors) USB-621x Screw Terminal USB-621x Mass Termination USB-621x BNC Weight USB-621x Screw Terminal USB-6212 Mass Termination USB-6216 Mass Termination USB-6212/6216/6218 BNC USB-6210 OEM USB-6212/6216/6218 OEM I/O connectors USB-6210/6211/6215 USB-6212/6216/6218 Screw Terminal USB-6212/6216 Mass Termination USB-6212/6216/6218 BNC USB connector Screw terminal wiring Torque for screw terminals Two 16-position combicon Four 16-position combicon One 68-pin SCSI 19 BNCs and 26 screw terminals Series B receptacle 16 to 28 AWG 0.22-0.25 N m (2.0 -2.2 lb in.) 206 g (7.2 oz) 227 g (8.0 oz) 231 g (8.1 oz) 950 g (33.5 oz) 73 g (2.5 oz) 76 g (2.6 oz) 16.9 9.4 3.1 cm (6.65 3.70 1.20 in.) 19.3 9.4 3.1 cm (7.61 3.68 1.20 in.) 23.5 11.2 6.4 cm (9.25 4.40 2.50 in.)
Environmental
Operating temperature Storage temperature 0 to 45 C -20 to 70 C
10/17
www.ni.com
Caution This device is rated for Measurement Category I and the voltage across the isolation barrier is limited to no greater than 30 Vrms/60 VDC/42.4 V pk continuous. Do not use for measurements within Categories II, III, or IV. The USB-6215/6216/6218 Maximum Working Voltage figure illustrates the maximum working voltage specifications.
Safety
This product is designed to meet the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use: IEC 61010-1, EN 61010-1 UL 61010-1, CSA 61010-1 Note For UL and other safety certifications, refer to the product label or visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.
Electromagnetic Compatibility
This product is designed to meet the requirements of the following standards of EMC for electrical equipment for measurement, control, and laboratory use: EN 61326 (IEC 61326): Class A emissions; Basic immunity EN 55011 (CISPR 11): Group 1, Class A emissions AS/NZS CISPR 11: Group 1, Class A emissions FCC 47 CFR Part 15B: Class A emissions ICES-001: Class A emissions Note For the standards applied to assess the EMC of this product, refer to the Online Product Certification section. Note For EMC compliance, operate this product according to the documentation. Note For EMC compliance, operate this device with shielded cables.
CE Compliance
This product meets the essential requirements of applicable European Directives, as amended for CE marking, as follows: 2006/95/EC; Low-Voltage Directive (safety)
11/17
www.ni.com
Environmental Management
National Instruments is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial not only to the environment but also to NI customers. For additional environmental information, refer to the NI and the Environment Web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document. Waste Electrical and Electronic Equipment (WEEE) EU Customers At the end of their life cycle, all products must be sent to a WEEE recycling center. For more information about WEEE recycling centers and National Instruments WEEE initiatives, visit ni.com/environment/weee.htm.
1 2 3
Stresses beyond those listed under Input voltage protection may cause permanent damage to the device. If you are using a USB M Series device in Full-Speed mode, device performance will be lower and you will not be able to achieve maximum sampling/update rates.
USB-621x Screw Terminal/BNC devices have a self-resetting fuse that opens when current exceeds this specification. USB-621 x Mass Termination devices have a user-replaceable socketed fuse that opens when current exceeds this specification. Refer to the NI USB-621x User Manual for information about fuse replacement.
4 5 6
Maximum working voltage refers to the signal voltage plus the common-mode voltage. In the USB-6215/6216/6218 Maximum Working Voltage figure, |Va Vd|, |Vb Vd|, and |Vc Vd|.
Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.
7
In the USB-6215/6216/6218 Maximum Working Voltage figure, |Va Ve|, |Vb Ve|, and |Vc Ve|.
Back to Top
12/17
www.ni.com
USB-6210 Pinout
13/17
www.ni.com
USB-6211/6215 Pinout
14/17
www.ni.com
USB-6212/6216 Pinout
15/17
www.ni.com
16/17
www.ni.com
2008 National Instruments. All rights reserved. CVI, DIAdem, LabVIEW, Measurement Studio, National Instruments, National Instruments Alliance Partner, NI, ni.com, NI CompactDAQ, SCXI, and SignalExpress are trademarks of National Instruments. The mark LabWindows is used under a license from Microsoft Corporation. Windows is a registered trademark of Microsoft Corporation in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries. Other product and company names listed are trademarks or trade names of their respective companies. A National Instruments Alliance Partner is a business entity independent from NI and has no agency, partnership, or joint-venture relationship with NI.
My Profile | RSS | Privacy | Legal | Contact NI 2012 National Instruments Corporation. All rights reserved.
17/17
www.ni.com