Beruflich Dokumente
Kultur Dokumente
- Blood-Flow Control . . . . . . . . . . . . . . . . . . . . . . 81
6.1.3 Adaptive Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
6.2 Arterial Pressure Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
6.2.1 Total Arterial Pressure Control . . . . . . . . . . . . . . . . . . . . . . . 87
6.2.2 Arterial Pressure Boundary Control . . . . . . . . . . . . . . . . . . . . . 88
6.3 Blood-Gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
6.3.1 State Space Substitution . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
6.3.2 Linearisation by State Feedback . . . . . . . . . . . . . . . . . . . . . . . 96
iii
Contents
6.3.3 Robust External Linear pO
2
-Controller Design . . . . . . . . . . . . . . . 103
6.3.4 pCO
2
-Controller Design . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
6.3.5 Blood-Gas Control Interconnection . . . . . . . . . . . . . . . . . . . . . 109
7 Simulation and In-vitro Control Study 111
7.1 Arterial Blood-Flow Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
7.1.1 Stationary Perfusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
7.1.2 Pulsatile Perfusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
7.2 Total Arterial Pressure Control . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
7.3 Arterial Pressure Boundary Control . . . . . . . . . . . . . . . . . . . . . . . . . 120
7.4 Blood-Gas Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
7.4.1 Stationary Blood-Gas Control (Step-Response) . . . . . . . . . . . . . . 124
7.4.2 Stationary Blood-Gas Control (Disturbance Rejection) . . . . . . . . . . 134
8 Conclusion and Discussion 141
A Abbreviations i
B Constants ii
C Notation and Symbols vi
D Experimental Setup x
D.1 Hydrodynamic Vascular System Simulator . . . . . . . . . . . . . . . . . . . . . x
D.2 Pulsatile Control Setpoint . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiii
D.3 In-vitro Blood-Gas Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiv
D.3.1 Experimental Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiv
D.3.2 Materials and Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . xvi
Bibliography xxi
iv
List of Figures
1.1 Controlled variables and controller structure for CPB . . . . . . . . . . . . . . . 3
2.1 Diagram of the simplied human circulation . . . . . . . . . . . . . . . . . . . . 6
2.2 Perfusion in the human pulmonary and systemic circulation . . . . . . . . . . . 9
2.3 Nonlinear oxygen-binding curve . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.4 Carbon dioxide transport and reaction . . . . . . . . . . . . . . . . . . . . . . . 15
2.5 Nonlinear carbon dioxide-dissociation curve . . . . . . . . . . . . . . . . . . . . 16
3.1 Components of the extracorporeal circuit . . . . . . . . . . . . . . . . . . . . . . 19
3.2 Roller pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.3 Rotational blood pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.4 Pathophysiological factors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.5 Viscosity increase due to hypothermia . . . . . . . . . . . . . . . . . . . . . . . . 26
3.6 Haemodynamic response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.7 Application of cardiopulmonary bypass . . . . . . . . . . . . . . . . . . . . . . . 32
4.1 Equivalent electro-mechanical network diagram for the BLDC motor . . . . . . . 38
4.2 Nonlinear static pressure output . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
4.3 Blockdiagram for the nonlinear state space system . . . . . . . . . . . . . . . . . 41
4.4 Electric analogue for a single vascular element . . . . . . . . . . . . . . . . . . . 47
4.5 MATLAB/Simulink implementation for a basic compartment . . . . . . . . . . . 48
4.6 Block diagram of the modelled system for haemodynamic control . . . . . . . . 52
4.7 Blood-gas diusion exchange over a membrane . . . . . . . . . . . . . . . . . . . 55
4.8 Block diagram of the oxygenator system . . . . . . . . . . . . . . . . . . . . . . 62
5.1 Frequency response for Eq. (5.3) . . . . . . . . . . . . . . . . . . . . . . . . . . 66
v
List of Figures
5.2 Static and dynamic simulation and experimental results for q = 0 lmin
1
of Eq.
(5.3) and (5.4) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
5.3 Experimental measurement setup . . . . . . . . . . . . . . . . . . . . . . . . . . 68
5.4 Polynomial nonlinear pressure tting for the arterial cannula . . . . . . . . . . . 69
5.5 Impedance spectra of the simulated vascular models . . . . . . . . . . . . . . . . 70
5.6 Time series of the vascular models . . . . . . . . . . . . . . . . . . . . . . . . . . 71
5.7 Frequency response comparison of model and vascular system simulator . . . . . 72
5.8 Response to a propofol injection impulse with pressure and ow time series. . . . 75
5.9 Response to a sodium nitroprusside injection impulse with pressure and ow
time series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
5.10 Frequency response variations of the linearised system with uncertainty . . . . . 77
5.11 Simulation and experimental step-response of the blood-gas process . . . . . . . 78
5.12 Simulation, experimental and corrected step-response of the blood-gas process . 79
6.1 Principal control structure for aortic blood-ow control . . . . . . . . . . . . . . 81
6.2 Root locus of the open-loop compensated system G
BF
(s)C(s) . . . . . . . . . . 82
6.3 Augmented system for robust control . . . . . . . . . . . . . . . . . . . . . . . . 83
6.4 Sensitivity functions for blood-ow control . . . . . . . . . . . . . . . . . . . . . 84
6.5 Structure of the adaptive control system . . . . . . . . . . . . . . . . . . . . . . 85
6.6 Total arterial pressure control . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
6.7 Root locus of the open-loop compensated system C(s)G
BPC
(s) . . . . . . . . . . 89
6.8 Pressure boundary control structure . . . . . . . . . . . . . . . . . . . . . . . . . 89
6.9 Mean arterial pressure dierence mapping to a control error . . . . . . . . . . . 90
6.10 Root locus of the open-loop compensated system C(s)G
cBFC
(s) . . . . . . . . . 91
6.11 pO
2
-pressure controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
6.12 pCO
2
-pressure controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
6.13 Linearisation loop for the nonlinear O
2
-plant . . . . . . . . . . . . . . . . . . . . 103
6.14 PI-controller sensitivity functions . . . . . . . . . . . . . . . . . . . . . . . . . . 106
6.15 PI- and H
-pO
2
blood-gas control disturbance rejection experiment . . . . . . . . . . . 140
D.1 Hydrodynamic System Simulator Elements . . . . . . . . . . . . . . . . . . . . . xi
D.2 Hydrodynamic System Circuit Control Setup . . . . . . . . . . . . . . . . . . . . xii
D.3 Pulsatile control setpoint . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiii
D.4 Blood-gas analysis control setup . . . . . . . . . . . . . . . . . . . . . . . . . . . xv
D.5 De-oxygenator serial connection . . . . . . . . . . . . . . . . . . . . . . . . . . . xvi
D.6 Blood-ow - FiCO
2
relationship . . . . . . . . . . . . . . . . . . . . . . . . . . . xix
vii
List of Tables
2.1 Haemodynamics during physiological and extracorporeal circulation . . . . . . . 10
2.2 Blood-gas- and pH-values of an healthy adolescent under physical rest . . . . . . 13
3.1 Percentile O
2
-consumption and time of HLM shutdown until tissue damage oc-
curs under hypothermia . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.1 Haemodynamic variables and conditions for control . . . . . . . . . . . . . . . . 36
7.1 Simulation and experimental results . . . . . . . . . . . . . . . . . . . . . . . . . 111
7.2 Stationary blood-ow control performance . . . . . . . . . . . . . . . . . . . . . 114
7.3 Pulsatile blood-ow control performance . . . . . . . . . . . . . . . . . . . . . . 119
7.4 Blood-gas analysis control conditions . . . . . . . . . . . . . . . . . . . . . . . . 125
7.5 Simulation performance switch-on . . . . . . . . . . . . . . . . . . . . . . . . . . 126
7.6 Experimental in-vitro performance switch-on . . . . . . . . . . . . . . . . . . . . 128
7.7 Simulation performance step-response . . . . . . . . . . . . . . . . . . . . . . . . 129
7.8 Experimental in-vitro performance step-response . . . . . . . . . . . . . . . . . . 132
7.9 Simulation performance disturbance rejection . . . . . . . . . . . . . . . . . . . 136
7.10 Experimental in-vitro performance disturbance rejection . . . . . . . . . . . . . 138
7.11 pCO
2
simulation and experimental performance disturbance rejection . . . . . . 139
D.1 Experimental BGA protocol sample . . . . . . . . . . . . . . . . . . . . . . . . . xviii
viii
1 Introduction
Extracorporeal circulation (ECC) in heart surgery has been established as a routine treatment
for several decades. In the case of a cardio-pulmonary bypass (CPB), ECC with the use of the
heart-lung machine (HLM) allows the surgeon to operate on the resting heart.
At the present time thousands of heart surgeries are performed in Europe every year. Numbers
for heart surgeries in Germany exceed about 100,000 per year and are still increasing. For
example from 1979 to 2001 the number of cardiovascular operations and procedures in the
U.S.A. increased about 417 percent [8]. Among the main reasons for cardiovascular surgery
are coronary heart disease, congestive heart failure, hypertensive disease, cardiac arrhythmias,
rheumatic heart disease, cardiomyopathy, pulmonary heart disease, and others [68, 129]. Most
of these diseases need surgical treatment, while the heart is resting. In the resting condition
the human heart and body are no longer subject to oxygenated blood perfusion, therefore heart
and brain tissue damage would occur in the course of minutes.
As a major advance in medicine, the HLM made it possible for surgeons to operate on the
resting heart. During cardiopulmonary heart-lung support, the HLM takes over the work of
the heart and lungs, that is perfusion and oxygenation. The number of heart surgeries with
HLM support is, like the total number of heart surgeries, rising. Heart surgeries with HLM
support rose in Germany from 36,000 in 1990 to 95,000 in 2003
1
. This extreme increase in
surgical operations can be explained with the expansion of HLM services to older and high risk
patients and to new surgical procedures, such as heart transplantations. Lower mortality rates
and continuously rising surgical experiences made this development possible. Major factors
that contributed to this development include the following advances [29]:
- Myocardial protective arrangements (e.g. cardioplegia solutions).
- ECC procedures (e.g. integrated circuits and priming solutions).
- Surgical procedures (e.g. heart valve protheses).
1
German Society for Thorax-, Heart-, and Vascular-Surgery
1
1 Introduction
- Knowledge of the specic action of anaesthetics and analgesics.
- Pathophysiological knowledge of acute cardiovascular diseases and the introduction of
pharmacological concepts for circulatory support.
Today, the highly complex heart-lung machine needs to be controlled by specialised perfusion
technicians, who continually monitor important patient variables and adjust manually the con-
trol input variables of the HLM in agreement with the surgical team. This procedure can lead
to errors, which in turn can increase the risk of post operative damage or the mortality rate.
In order to get a higher degree of reproducibility, increased patients safety and less workload
for the perfusion technician the need for automation and control arises.
Even up to today, no automatically controlled heart-lung-apparatus is available for clinical use.
Automatic control of ECC, whether haemodynamics or blood-gases, will still be a challenging
goal and research topic over the next years.
1.1 Extracorporeal Circulation: A Brief Historical Overview
Since its rst clinical application by Gibbon, dating back to 1953 [62], many improvements were
made in ECC with HLM machine support.
With the detection of heparin and its anticoagulant properties (Howell, 1900), CPB was pos-
sible right from the start in 1953. During surgery, the blood is exposed to the extracorporeal
circuit and a catastrophic clotting is avoided by the administration of heparin.
Today dierent HLM components are produced in various forms and from various biomedical
companies, but the trend is moving from modular to integrated machines with extended mon-
itoring and control functions. New developments in HLM systems are comprised of advances
in haemodilution, blood perfusion and monitoring techniques. Even in todays advanced stage,
CPB with HLM support remains an invasive procedure carrying numerous risks.
1.2 Goals of this Work
The introduction of automatic control is suggested to hold several advantages over manually
controlled HLMs. On the one hand, the well-known properties of automation and automatic
2
1.2 Goals of this Work
control, such as the avoidance of large manual control errors and overshoots and the introduc-
tion of safety mechanisms [73, 94, 136], provide additional reduction of infection risk after ECC
and the prevention of potential organ or tissue damage [4, 124, 125, 133]. With the design
and development of, for example, special perfusion strategies (pulsatile perfusion) or fast and
reliable oxygen partial pressure (pO
2
) disturbance rejection, the automatic control strategy can
guarantee a more physiological perfusion or respond quickly to changes in patient status.
On the other hand, a feedback control strategy reduces the workload for the sta. Over the
course of a normal heart surgery, where surgeons, anesthetists and perfusion technicians have
to make quick decisions and rely on extensive background knowledge, the reduced amount of
workload is suggested decrease the number of incorrect decisions and thus provide more patient
safety.
The variables to be controlled can be deduced from the requirements to provide a stable phys-
iological perfusion and a sucient oxygen delivery. Controlled variables should be the arterial
blood-ow q
art
, the arterial pressure p
art
, the arterial oxygen partial pressure pO
2,a
and the ar-
terial carbon dioxide partial pressure pCO
2,a
. Figure 1.1 shows an overview of the four control
circuits, with controlled variables and control actuating principle. The system to be controlled
consists of the HLM, coupled with the human vascular system. During the last 50 years of
q /p
art art
pO
2
pCO
2
Blood
pump
FiO
2
Gas-flow/
FiCO
2
Control
actuating
principle
q , pO , pCO , p
art 2 2 art
ECC
control
Operator
Upper systemic
circulation
Lower systemic
circulation
Right
heart
Left
heart
Lung
Oxy-
genator
HLM
Patient
Figure 1.1: Controlled variables and controller structure for CPB.
3
1 Introduction
CPB, dierent kinds of perfusion strategies and HLMs were developed. Variations can be,
for example, the type of oxygenation device (oxygenator) or the venous withdrawal strategy
[62, 68, 129]. The controlled variables strongly depend on the perfusion strategy and the type
of HLM. Since the strategy and management of CPB can vary even between heart centres on
a national level, the guidelines of the Heart- and Diabetes Centre Bad Oeynhausen (Univer-
sity Hospital of the Ruhr-University Bochum) were preferred in this work [68]. This perfusion
strategy is a simple, often used approach to CPB and easily adaptable to other strategies of
CPB. Chapter 3 deals with the dierent HLM applications and their consequences for control.
Controller design objectives in medical man-machine systems have to full certain more re-
strictive requirements and constraints. These requirements and constraints are due to the
physiological properties or depend on the articially generated environment during ECC.
Special emphasis has to be laid to the physiological requirements and constraints, which if vi-
olated could lead to unphysiological conditions, causing instability or unwanted damage. The
patients vascular system, metabolic/circulatory regulations and the organ blood are highly
complex systems, consisting of various mechanisms and are coupled to the human body and
the HLM. Nonlinearities, parameter uncertainties, articially induced disturbances and time
variant properties or parameter drift are inherent in these systems. In the case of the arti-
cially generated environment during ECC, HLM component nonlinearities and dynamics have
to be taken into account. In addition, uncertainties exist for HLM components of dierent
manufacturers. For the vascular system, parameter uncertainties depending on dierent perfu-
sion strategies or applied vasoactive drugs are common from patient to patient. In case of the
extracorporeal circuit, special care has to be taken of artifacts and articial disturbances.
These eects lead to a system which is extremely dicult to control. For the control design,
extensive system modelling and the use of modern robust and nonlinear control theory will be
necessary. Since only the technical parts of the system can be validated on an experimental
basis (due to ethical and safety reasons), the physiological system will require special methods
of modelling. Therefore, a reasonably large chapter of this work will consider technical and
physiological system modelling and will use the modelled system as a basis for control.
Control in the case of CPB has to be compatible with various patients, stable and robust in
the presence of parameter uncertainties, unexpected disturbances, nonlinearities, time-varying
parameters and variable time-delays. The main goal of this work is to design suitable open-
and closed-loop control algorithms which satisfy these conditions and help to raise CPB with
HLM support to a higher state of reliability and eciency.
4
1.3 Outline
1.3 Outline
This thesis is organised as follows.
Chapter 2: Physiological Background describes the basic knowledge of the human cir-
culatory and vascular system, the organ blood and the coupling to the HLM, necessary to
understand the system modelling in Chapter 4.
Chapter 3: Technical Background introduces the extracorporeal perfusion system, its
components, perfusion concepts, advances in technology and the suggested control strategy.
Chapter 4: System Modelling. In this chapter the systems to be controlled are modelled
in the state of extracorporeal circulation, divided into technical and physiological subsystems
and later on connected to give the models for CPB.
Chapter 5: Simulation and Experimental Model Validation validates the technical
subsystems presented in Chapter 4 in an experimental study and compares physiological sub-
systems to literature.
Chapter 6: Control Design addresses robust, nonlinear and adaptive control design, based
on the modelling results presented in Chapter 4 and validated in Chapter 5.
Chapter 7: Simulation and In-vitro Control Study presents feedback control validation
and performance in simulation and in-vitro experimental test conditions.
Chapter 8: Conclusion and Discussion. Finally, this Chapter ends up with a discussion,
summarising achieved goals, limits and contributions. Conclusions are drawn and directions for
future research are outlined.
Appendix: The appendix closes with constants for modelling and control, some conventions
about notations, symbols and the experimental setup for haemodynamic and blood-gas control.
5
2 Physiological Background
2.1 The Circulatory System
The human circulatory system in the undisturbed condition can be regarded as a continuous
ow circuit made up of distinct parts, satisfying a number of functions. The circulatory system
can be divided into two separated circuits, the systemic (body) and the pulmonary (lung)
circulation, connected by the heart. Figure 2.1 shows the interconnection between the two
circuits and the separated organ heart as the natural blood pump. Blood pumped by the left
heart ows through the aorta and the vascular system and passes the dierent organ areas, like
brain or muscles. The blood-ows from the arterial to the venous systemic system and back to
the right heart, where it is pumped to the pulmonary (lung) system. There, the gas exchange
Left atrium
Left ventricle
Aorta Pulmonary
artery
Right ventricle
Right atrium
Pulmonary
valve
Tricuspid valve
Aortic valve
Mitral valve
Pulmonary circulation
Systemic circulation
Septum
Figure 2.1: Diagram of the simplied human circulation, with the heart, connected to pulmonary
(lung) and systemic (body) circulation.
6
2.2 The Human Heart
of oxygen (O
2
) and carbon dioxide (CO
2
) is accomplished. Finally the oxygen enriched blood
is transported back to the left heart and again to the organs by the systemic circulation.
The haemodynamics of the circulatory system have to follow the present status of the human
body, which of course is subject to various disturbances such as, for example, physical stress,
change of body posture or blood loss. The perfusion of the organs and dierent tissue areas
with blood guarantees [43]
- Oxygen delivery.
- Delivery of nutrients, e.g. glucose or amino acids.
- Carbon dioxide removal.
- Removal of hydrogen ions.
- Maintenance of proper concentrations of other ions.
- Transport of various hormones and other specic substances.
- Distribution of body heat.
To achieve these goals under natural and articial disturbances, the circulatory system is pro-
vided with complex autoregulation systems for haemodynamics. Some of these control mech-
anisms are well understood, while others are still subject to extensive research [112] and are
described in more detail below.
2.2 The Human Heart
The human heart is a hollow muscular organ that serves a principal purpose: Pumping the
blood through the systemic and the pulmonary circuit. For this purpose the heart is divided
in a left and a right heart which are separated by a thick muscular wall, the septum (see
Figure 2.1). Each part of the heart is again separated into a blood collection (atrium) and a
blood ejection (ventricle) chamber. Blood-ow in the heart is achieved by rhythmic contraction
(systole) of the left and right ventricles. A number of heart valves prevent backow during
relaxation (diastole). The heart is coupled to the vascular system by the aorta ascendens and
the arteria pulmonalis on the arterial side and by the venae cavae superior/inferior and venae
pulmonalis on the venous side. The blood supply for the heart itself is carried out by the
coronary arteries.
7
2 Physiological Background
Heart diseases and disorders, which are the main reason for initiation of cardiopulmonary bypass
(CPB) procedures exist at the time of birth (congenital heart defects) or become established
on a long-term base. Some heart diseases can be treated with minimal invasive procedures, e.g.
the implantation of a cardiac pacemaker. However, for most of the more severe heart diseases
surgical procedures which require HLM support are necessary.
During CPB the heart is decoupled or closed from the circulatory system and perfused and
cooled with blood and a protective solution. Therefore the pulmonary vascular system is no
longer perfused nor are the lungs for the time in which the CPB is applied.
2.3 The Vascular System
The vascular system can be divided into a systemic circulation or peripheral circulation and
a pulmonary (lung) circulation. The vascular system consists of serial and parallel connected
blood vessels (arteries, capillaries and veins) which transport the blood to separated tissue
areas. Depending on the need for blood, the dierent organs of the human body, shown in
Figure 2.2, are perfused with blood. Long- and short-term regulation mechanisms are able to
change the resistance of a certain organ area, thereby changing the perfusion rate of that area.
In the case of the pulmonary circulation the lung is perfused with 100 % of the cardiac output.
For the purpose of haemodynamic regulation the blood vessels are endowed with a muscular
wall that is capable of contraction and dilation. In principle, the pulmonary vascular system is
similar to the systemic vascular system.
2.3.1 The Systemic Circulation
Arteries, capillaries and veins make up the systemic circulation and can be further separated in
arterioles and venules belonging on the arterial and venous system respectively. These vessels
are the functional parts of the systemic circulation and all fulll a certain role.
Arteries transport the blood from the heart to the arterioles and the tissue.
Arterioles consist of a strong muscular wall that can completely close or dilate the arteriole. One
can think of an adjustable resistance to control local tissue perfusion. The arterioles transport
the blood to the capillaries.
8
2.3 The Vascular System
Lung
Coronary vessels
Brain
Muscles
Intestine
Skin, etc.
Right heart Left heart
Kidney
Liver
100 %
100 % 100 %
5 %
15 %
20 %
7 %
23 %
20 %
10 %
Figure 2.2: Perfusion in the human pulmonary and systemic circulation, in normal (physiological)
conditions and without physical strain [112].
In capillaries uid, blood-gases, nutrients, electrolytes, hormones and other substances are
exchanged over thin and permeable vascular walls between blood and interstitial spaces.
Venules are small veins that collect the blood from the capillaries and pass it on to the veins.
Veins function as conduits for storage and transport of blood back to the heart. Veins are
muscular blood vessels, possessing the ability to contract and expand.
2.3.2 The Pulmonary Circulation
In the pulmonary circulation, the blood-ows from the right heart through the lung to the left
heart. A small fraction of oxygen rich blood ows from the left heart (arterial) to the right
heart (venous) and supplies the supporting tissues of the lung with oxygen. This small fraction
of only 1 to 2 % is not shown in Figure 2.2. The main blood supply, however, ows from
the right heart through the pulmonary arteries and arterioles (98 to 99 %). The walls of the
pulmonary arteries and arterioles are very thin and distensible, allowing it to accommodate
most of the stroke volume output of the right ventricle. The gas exchange takes place between
9
2 Physiological Background
the capillaries and the alveolar walls of the pulmonary alveoli (see Section 2.6). During ECC
the pulmonary circulation is, like as the heart, temporarily separated or closed from the extra-
corporeal circuit. Therefore the pulmonary vascular system will be disregarded in the system
modelling, in Chapter 4.
2.3.3 Haemodynamics
Haemodynamics in the vascular system follows a complex mathematical relationship and can
be described with the Navier-Stokes equations (NSE). These complexities comprise mainly non-
linearities in distensible tubes including turbulent ows in vessel branches, vessel collapsibility
and the medium blood as a non-Newtonian uid. Chapter 4 deals with the system modelling
of the vascular system under ECC.
The haemodynamics of physiological and extracorporeal circulation can be classied by a num-
ber of variables. These will be summarised in brief in this section, if further needed for control.
Table 2.1 compares the variables of physiological and extracorporeal circulation. The values of
Table 2.1 were summarised from normal physiology and ECC literature, [43, 68, 112, 129] or
from papers of detailed ECC experiments [30, 41, 42, 45, 67, 69, 90, 105].
The Mean Arterial Pressure (MAP) is the mean value of the arterial pressure curve over time,
lying between systolic (maximum) and diastolic pressure (minimum).
The Cardiac Output (CO) is the Heart Rate (HR) times Stroke Volume (SV )
CO = HR SV. (2.1)
In a healthy heart the CO is the same as Mean Arterial Flow (MAF), since a backow of blood
to the heart is stopped by the heart valves.
Central Venous Pressure (CV P) is a measurement of the pressure in the right atrium. CV P
Table 2.1: Haemodynamics during physiological and extracorporeal circulation (extracorporeal refers
to a standard CPB procedure). BS is the patients body surface in m
2
.
MAP CO (MAF) HR CV P TPR C
art
[mmHg] [l/min] [BPM] [mmHg] [mmHg/(l/min)] [ml/mmHg]
Physiological 70-90 4-6 60-90 2-8 15-20 0.5-1.3
Extracorporeal 40-60 2.4BS - 0 ( 10) 7-15 1-2
10
2.4 The Blood
reects the ability of the right heart to pump blood and is important in CPB to keep the
venous systemic blood vessels from collapsing.
The Total Peripheral Resistance (TPR) refers to the cumulative resistance of the systemic
vascular system. The TPR is a uid resistance calculated by
TPR =
MAP
CO
. (2.2)
The Arterial Compliance (C
art
) is the ability of the systemic vascular arterial tree to bend to
pressure increases on ow/pressure wave (Windkessel).
2.4 The Blood
Blood is a viscous nontransparent uid composed of plasma and suspended cells. The blood cells
consist of red (erythrocyte) and white (leukocyte) blood cells and of platelets (thrombocyte).
Most of the blood cells (99 %) are red blood cells and determine the physical characteristics of
the blood. Cellular fraction of the blood is called haematocrit and given in percent.
Blood cells and plasma accomplish versatile functions such as transport, homeostasis, resistance
to body infection and protection from blood loss. Blood plasma consists of water, protein
and other molecular substances and plays an important role in the regulation of a constant
osmolar pressure. The main function of the erythrocytes is to transport haemoglobin, which
in turn serves as an oxygen carrier. Leukocytes are part of the bodys protective system and
thrombocytes are important for blood coagulation.
Before the application of CPB, the extracorporeal circuit is primed with a uid. This priming
solution (refer to Chapter 3) necessarily expands the total body water and the extracellular
uid compartments. This process called haemodilution has signicant eects on the transport
function of blood-gases, the uid resistance in terms of blood viscosity, and even indirect eects
on the vascular system. The organ blood under ECC is modelled in Chapter 4.
2.5 Regulation Mechanics
Regulation of the haemodynamics is accomplished by a highly complex system with various
cascaded control structures and can be generally divided in local tissue, nervous and humoral
11
2 Physiological Background
control. A short overview on these regulation mechanisms will be given below.
Local tissue blood-ow control can be further dierentiated into rapid and long-term control.
Local control can be a matter of seconds to minutes in the case of acute control, or is achieved
over a period of hours, days or even weeks. The muscle bers of the small blood vessels re-
act to local concentration factors in the tissues, like oxygen, carbon dioxide, hydrogen-ions,
electrolytes and other substances. Eects of local tissue perfusion cannot be inuenced or con-
trolled during ECC. Since the eects of rapid local tissue perfusion regulation can be caused by
the changes in the concentration factors of substances, they have to be taken into account as
uncertainty in system modelling. Long-term local tissue perfusion regulation can be neglected.
Nervous regulation of haemodynamics is rapid response and superimposed on local tissue
haemodynamic control. Nervous regulation is achieved mainly by the autonomous nervous
system that can cause vasoconstrictive or -dilative vessel action. General anaesthesia, as used
in CPB, is the state of unconsciousness produced by anaesthetic agents, with the absence of
pain sensation over the entire body and a greater or lesser degree of muscular relaxation [100].
In the state of anaesthesia the functions of the central nervous system and the autonomous
nervous system are damped or disconnected. This has a signicant inuence on e.g. TPR,
which can change to more than 100 % of its original value. Chapter 4 refers to this issue, where
changes in the vasculature invoked by dierent anaesthetic agents are modelled as uncertainties.
Humoral regulation of haemodynamics can be either rapid or long-term based and is superim-
posed on local tissue haemodynamic control. In the humoral regulation system substances are
formed in special glands and are distributed by the blood over the circulatory system. These
substances can be hormones, ions or various chemical factors. Substances can be divided into
vasoconstrictor and vasodilator agents. Due to the articially generated environment during
CPB, such as the priming of blood and the low temperature (hypothermia), the hormonal con-
centrations are changed and some substances are reperfused or released in greater quantities to
the circulatory system, which has to be considered in modelling, refer to Chapter 3.
2.6 Transport of Blood-Gases and Acid-Base Management
In physiological circulation the de-oxygenated blood is circulated through the lung, where car-
bon dioxide is removed from and oxygen is added to the blood. In CPB, where the heart and
lung are resting, lung function is taken over by an oxygenation device. The general principle
12
2.6 Transport of Blood-Gases and Acid-Base Management
of blood-gas exchange, in either lung or oxygenation device, however, remains the same. Gases
are driven by partial pressure dierences and diuse over membranes between the alveoli and
capillaries in the physiological (lung) case or between gas and blood compartment in the oxy-
genator (CPB).
The transport of oxygen and carbon dioxide is mainly accomplished by the haemoglobin which
is contained in the erythrocytes. Besides this transport function the haemoglobin and other
buer systems of the blood play a certain role in the regulation of the acid-base management.
Blood-gas- and pH-values are given in Table 2.2, where S
O
2
is the oxygen saturation of the
blood (see below). Changes to the blood due to haemodilution during ECC mean a change in
the transport of O
2
and CO
2
and a change in the acid-base management. This will be covered
in detail in Chapter 3.
2.6.1 O
2
-Transport
Oxygen (O
2
) in the blood is transported in a physically dissolved or chemically bound condi-
tion. About 30 to 100 times as much oxygen can be transported in chemical binding to the
haemoglobin than physically dissolved oxygen in the blood plasma. After the diusion process
over the membrane of the lung cells (pulmonary alveoli), the O
2
-molecule becomes physically
dissolved in the water of the blood and then can react to the haemoglobin.
The amount of physically dissolved oxygen is dependent on the partial pressure of the oxygen
in the gas (refer to Chapter 4).
The amount of chemically bound oxygen in the blood is nonlinearly dependent on various
factors. The O
2
-binding curve, shown in Figure 2.3, describes this nonlinearity as the O
2
-
saturation of the haemoglobin (S
O
2
), depending on the O
2
-partial pressure. O
2
-saturation is
the ratio of chemically bound O
2
-concentration ([HbO
2
]), to the total haemoglobin concentra-
Table 2.2: Blood-gas- and pH-values of an healthy adolescent under physical rest [112].
pO
2
S
O
2
[O
2
] pCO
2
[CO
2
] pH
[mmHg] [%] [l
O
2
/l
Blood
] [mmHg] [l
CO
2
/l
Blood
]
Arterial blood 90 97 0.2 40 0.48 7.4
Venous blood 40 73 0.15 46 0.52 7.37
13
2 Physiological Background
tion ([Hb
total
]).
S
O
2
=
[HbO
2
]
[Hb
total
]
(2.3)
S
O
2
is usually given in %. At an O
2
-saturation of 0 % all of the haemoglobin is deoxygenated,
where at an O
2
-saturation of 100 % every haemoglobin molecule carries its full O
2
-load. The
S
O
2
-saturation curve depends on a number of other factors, which are temperature, pH, pCO
2
and 2,3-diphosphoglycerate (control mechanism for oxygen movement to and from the erythro-
cytes), see Figure 2.3.
100
80
60
40
20
0
0 20 40 60 80 100 120
pO [mmHg]
2
S [%]
O
2
100
80
60
40
20
0
0 20 40 60 80 100 120
pO [mmHg]
2
S [%]
O
2
100
80
60
40
20
0
0 20 40 60 80 100 120
pO [mmHg]
2
S [%]
O
2
100
80
60
40
20
0
0 20 40 60 80 100 120
pO [mmHg]
2
S [%]
O
2
T= pH=
pCO =
2
pH=7.4 T=37C
v
a
20 30 37 42C 7.6 7.4 7.2
20 40 60mmHg
Decrease
2.3 DPG Increase
2.3 DPG
T=37C
Figure 2.3: Nonlinear oxygen-binding curve, with dependencies on temperature, pH-value, carbon
dioxide and 2,3-DPG. The dotted line between point a and v corresponds to arterial (a) and
venous (v) blood under resting conditions [112].
14
2.6 Transport of Blood-Gases and Acid-Base Management
2.6.2 CO
2
-Transport
Carbon dioxide (CO
2
) is transported in the blood as physically dissolved CO
2
, as chemically
bound bicarbonate (HCO
3
) and as carbamate (Hb CO
2
). The chemical binding process for
carbon dioxide is far more complex than that for oxygen, as it also inuences the acid-base
balance, and vice versa. Transport of carbon dioxide, even in abnormal conditions is not a
problem because much greater quantities of carbon dioxide than oxygen can be transported.
Figure 2.4 shows the carbon dioxide transport process. The carbon dioxide diuses from the
tissue cells in gaseous form through the cell membrane. From there it enters the capillary and
the blood, where it initiates the following physical and chemical reactions.
Capillary
Red blood cell
Hb
.
CO
2
Carbonic
anhydrase
Hb
H CO
2 3
H O + CO
2 2
HCO
-
+ H
3
+
H O
2
Hb
-
H Hb
H O
2
Cl
-
Cl
-
HCO
-
3
Plasma
CO
2
CO
2
Interstitial
fluid
Cell
CO
2
+
+
CO transported as:
1. CO = 7 %
2. Hb
.
CO =23 %
3. HCO
-
= 70 %
2
2
2
3
Figure 2.4: Carbon dioxide transport and reaction [43].
Dissolved CO
2
: Only a small portion of carbon dioxide (about 7 % in physiological circulation)
is transported in the dissolved state. Most of the dissolved CO
2
in the blood plasma enters the
erythrocyte.
Bicarbonate: The dissolved carbon dioxide reacts with water to form carbonic acid (H
2
CO
3
).
In the erythrocyte this reaction is about 5000 times faster, because of the catalysation enzyme
carbonic anhydrase. The resulting time constant for this reaction lies in the range of a small
fraction of a second, which allows enormous amounts of CO
2
to be transformed into carbonic
acid. The carbonic acid then is dissociated in hydrogen and bicarbonate ions. Hydrogen ions
15
2 Physiological Background
combine with the haemoglobin in the erythrocytes, a powerful acid-base buer. Bicarbonate
ions diuse over the erythrocyte membrane into the blood plasma and chloride ions from the
plasma take their places. The transport of CO
2
in bicarbonate form accounts for at least 70 %
of the total CO
2
transport in physiological circulation.
Carbamate: In the erythrocytes carbon dioxide also reacts with the haemoglobin, forming the
compound carbamino haemoglobin (Hb CO
2
) or carbamate. This is a reversible reaction and
the carbon dioxide is released in the alveoli or the oxygenation device where the carbon dioxide
partial pressure is lower than that of the blood. The quantity of carbon dioxide transported by
the carbamate reaction is approximately 23 %.
The total quantity of carbon dioxide in the blood of all of the above-named forms depends on
the CO
2
-partial pressure. Figure 2.5 shows the so-called carbon dioxide dissociation curve for
oxygenated and de-oxygenated blood. The dierence in the binding of the carbon dioxide in
both cases is due to the Haldane eect [112].
0.7
0.6
0.5
0.4
0.3
0 10 20 30 40 50 60 70
30
25
20
15
CO -content [mmol / l]
2
CO -partial-pressure [mmHg]
2
a
v
CO -content [ml CO / ml blood]
2 2
De-oxygenated blood
Oxygenated blood
Figure 2.5: Nonlinear carbon dioxide-dissociation curve for oxygenated and de-oxygenated blood.
The dotted line between point a and v corresponds to arterial (a) and venous (v) blood under
resting conditions [112].
16
2.6 Transport of Blood-Gases and Acid-Base Management
2.6.3 Acid-Base Management
Acid-base balance is described as the regulation of the concentration of hydrogen ions (H
+
),
which can vary from less than 10
14
up to 10
0
equivalents per litre. The hydrogen ion concen-
tration is expressed by the pH-value and dened as the negative decadic logarithm
pH = log
1
[H
+
]
= log[H
+
]. (2.4)
A low pH-value corresponds to a high hydrogen ion concentration and is called acidosis, in
contrast to a low hydrogen ion concentration, which corresponds to a high pH-value and is called
alkalosis. In Table 2.2, the pH-values for arterial (oxygenated) and venous (de-oxygenated)
blood are given. Acidosis and alkalosis pH-values are considered to be lethal if below 6.8 or
above 8.0 for a longer time.
To prevent the body uids from acidosis and alkalosis, several regulation systems of hydrogen
ion concentration are available.
- Acid-base buer systems prevent excessive changes in the hydrogen ion concentration.
This occurs in fractions of a second.
- The respiratory system is immediately stimulated to overcome changes in the hydrogen
ion concentration. By changing the rate of breathing, the carbon dioxide and therefore
the hydrogen ion content is changed. Regulation of hydrogen ion concentration with the
respiratory system is achieved over the course of minutes.
- On a medium- to long-term scale the kidneys regulate the hydrogen ion concentration by
excreting either acid or alkaline urine.
Chapter 4 covers the regulation systems of hydrogen ion concentration that play a role under
CPB with HLM support.
17
3 Extracorporeal Circulation
Extracorporeal circulation (ECC) refers to the pumping of the blood outside the human body.
In general, blood is taken from a blood vessel, for example for dialysis (minimally invasive)
and is pumped back into another vessel of the circulation system. Invasive procedures of ECC
include extracorporeal membrane oxygenation (ECMO), ventricular assist devices (VAD) or
cardiopulmonary bypass (CPB). In an ECMO, the function of the lung is totally or partially
served by an articially extracorporeal oxygenation device; the VAD partially takes over the
work of the left or right heart. In CPB, the function of heart and lungs is taken over by an
articial device, the heart-lung machine (HLM). Even though, a partial CPB with the HLM
is possible, this work refers to the total CPB, where the whole function of heart and lungs is
taken over by the HLM.
3.1 Principles and Components of the Extracorporeal
Circuit
CPB circuits consist of several components, of which a few satisfy the most important functions.
The essential components of an CPB circuit can be seen in Figure 3.1 and are blood pumps (ar-
ticial hearts), oxygenators (articial lungs) and the tubing system (articial vascular system).
Additional components are heat exchangers (in most cases included with the oxygenator), a
venous reservoir, a cardioplegia line for myocardial protection, as well as gas, bubble detection
and arterial lters. Partially decoupled from the ECC system is the control and monitoring
system, which consists of dierent sensors and devices for manual control.
The CPB circuit always has a main line, the tubing system for blood transport. The main line
is called venous return line on the pre-oxygenator and arterial line on the post-oxygenator side.
Added to the main venous return line are the suction lines, a venting line for the heart and a
bypass line for collecting the shed blood. The structure of an CPB circuit may vary, dependent
18
3.1 Principles and Components of the Extracorporeal Circuit
on hospitals. The structure of the HLM for this work was adapted from the University Hospital
of the Ruhr-University Bochum (Heart and Diabetes Center, Bad Oeynhausen, Germany) and
is shown in Figure 3.1.
The main line extracts the carbon dioxide rich blood from the venous side of the human vascular
system, stores it in a small reservoir (venous bag) and pumps the blood through an oxygenator
back to the arterial side of the human vascular system. In the oxygenator, carbon dioxide is
removed from the blood, oxygen is added and before entering the vascular system, the blood is
ltered.
All components that are in direct contact with the blood are single-use sterile components.
There follows a short description of the main components used in a modern ECC circuit.
Mixture of
gases
Venous
bag
Bubble
detector
Filter
Drugs
Level
sensor
main
line
Cardioplegia line Arterial
line
Cardioplegia
line
Cardio-
tomy
reservoir
Sucker
Blood substitute
substances
Ventline
Oxygenator
+ heat
exchanger
BGA (art.)
Pressure
Flow
Pressure
Heat
exchanger
Control
variable
blood flow
HLM Patient
Blood pump
and
rotary speed controller
w(z)
p
Aort.
q
Aort.
Blood flow
Influenced by
vasoactive
substances
Cannula
p
out
Drugs
Figure 3.1: Components of the extracorporeal cardiopulmonary bypass circuit, with the HLM to the
left and the patients vascular system to the right (BGA: blood-gas-analysis (arterial), Ventline:
drainage of the ventricle, Cardioplegia line: cooling, suspension of the heart and drug delivery).
19
3 Extracorporeal Circulation
3.1.1 The Oxygenator
During extracorporeal circulation the oxygenator acts as the articial lung. The oxygenator
refreshes the de-oxygenated blood with new oxygen and removes carbon dioxide. Modern
oxygenator systems are exclusively membrane oxygenators. As additional types of oxygenators,
bubble and lm oxygenators are only of historical importance.
In a membrane oxygenator, the gas ow is separated from the blood-ow. The separating
membrane is a semi-permeable membrane, consisting of polypropylene or silicone rubber. Gas
exchange is accomplished by the diusion across the membrane, driven by partial pressure
dierences. This process is very similar to the physiological diusion process in the human
lung. Heat exchangers are often incorporated in modern membrane oxygenators in the form of
heating or cooling by separated water ow. Membrane oxygenators are safe and ecient HLM
oxygenation/carbon dioxide removal components and currently they dominate the market. New
technical developments comprise integrated oxygenator/blood pump systems [22].
3.1.2 Blood Pumps
Arterial blood pumps take over the work of the heart and pump the blood from the venous to
the arterial systemic system. The requirements for blood pumps are:
- Flow range up to 7 l/min.
- Minimum turbulence and blood stagnation.
- Minimum blood cell damage (haemolysis).
- Simplicity and safety of use.
- High reliability.
- Cost vs. eectiveness.
In modern HLMs mainly two dierent types of blood pumps are available: Roller pumps and
centrifugal (circulatory) pumps.
Roller pumps consist of a semi-circular stator, which is mounted on a rotor, see Figure 3.2.
The blood tubing is compressed between rotor and stator and due to the rotary movement of
the rotor, the blood is displaced in the direction of the rotation. If the rotation of the rotor
is stopped, the ow is reduced to zero and because of the compressed tubing, backow is not
possible. Roller pumps are simple, low cost devices. One of the disadvantages is the haemolysis
20
3.1 Principles and Components of the Extracorporeal Circuit
Adjustment
nut
Tube
guides
Tubing
bushing
Roller
Backing
plate
Rotation
direction
Outflow Inflow
Figure 3.2: Roller pump.
caused by the compression of the tubing. Furthermore a line restriction upstream will create
an excessive vacuum, leading to a degassing of the blood and a generation of a bubble train
inside the tubing. Conversely, a line restriction downstream will lead to an immediate pressure
build-up, with possible dire consequences depending on the source of obstruction. A roller
pump displaces air and blood in the same way, which could lead to severe organ and tissue
damage when massive amounts of air bubbles are passed towards the patient.
In rotational blood pumps, a rotating impeller moves the blood in the desired direction by
centrifugal forces. In a centrifugal blood pump, the blood is drawn axially to the rotating axis
of the impeller and ejected tangentially. Due to the advanced design process (nite element
simulation methods), used for most modern centrifugal blood pumps, shear stress and turbulent
blood-ow are minimised. Amongst the most prominent advantages of these blood pumps are
the reduced haemolysis, the practical implementation, the long-life time, an only moderate
pressure rise on the occlusion of the arterial line and the small time constants (varying of course
on pump type). Disadvantages are the certainly higher costs (single-use product, pump head
or whole pump), a possible backow at impeller cessation and the lack of a possible pulsatile
perfusion. Figure 3.3 shows the DeltaStream blood pump as an example for a rotary blood
pump with diagonally streamed rotor. The black arrows in Figure 3.3 indicate the direction of
blood-ow and the direction of the rotational speed of the pump impeller.
21
3 Extracorporeal Circulation
1
2
3
4
Figure 3.3: The DeltaStream blood pump as an example for a rotational blood pump. 1 is the
direction of blood inow, 2 is the rotating impeller, 3 is the blood stream owing around the
impeller, and 4 is the rotation direction of the impeller.
3.1.3 Tubing
Considering the fact that during ECC the blood is in contact with a large articial surface
area (several meters of tubing), the defensive system of the human body may initiate multi-
ple biological reactions. Such defense reaction systems include for example the coagulation,
the brinolytic, the complement, the kallikrein and the kinin system [43, 112]. Systemic re-
sponse may be highly inammatory and can aect heart, lungs, brain and other organs. Since
haemostatic mechanisms within the vascular endothelium are quite complex (and up to now
subject to research), a tubing coated with healthy vascular endothelium, would be the ultimate
biocompatible surface. State-of-the-art is the heparinisation of the blood and heparin-coated
biosurfaces. Clinical and research results of biosurfaces outline benecial mitigating body de-
fensive system and coagulation response eects [132].
Although dierent tubing materials are available on the market, the tubing of choice is polyvinyl
chloride (PVC). Today heparin-coated biosurfaces are not only available for the tubing system
but for all other HLM components in contact with blood.
22
3.2 Pathophysiology of Extracorporeal Circulation
3.1.4 Other Components
Besides the main components, described above, various other components are used during
CPB. These include blood reservoirs, heat exchangers, arterial and venous cannulae, sensors,
measurement devices, device drivers, blood-gas analysers, infusion rate controllers and surgical
instruments. If needed for system modelling and automatic control, the components will be
described in detail in Chapter 4.
3.2 Pathophysiology of Extracorporeal Circulation
Pathophysiology of ECC can have dierent meanings. Physiology refers to the organ function
and regulation under normal conditions. Pathophysiology on the one hand is the abnormal
organ function, the degradation and the inadequate reaction of body organs in association with
the HLM. These are of course invoked by the articial environment directly inuencing hep-
atic, neurologic, renal, haemodynamic and other functions, but primarily by the mechanical
and pharmacological situation, inherent to the machine. On the other hand pathophysiology
can be the dysfunction of the extracorporeal articial organ caused by a mechanical system
failure or by the inadequate managing of an operator.
The application of the articial organ heart-lung machine in conjunction with a number of
cardiosurgical and anaesthetical procedures means a major alteration to body and organ func-
tions. Beyond that, CPB is applied in common to patients with cardiovascular or even multiple
diseases, which means a coupling with the pre-CPB pathophysiological situation.
The pathophysiological factors with reference to ECC are generally divided into certain medical
areas, as shown in Figure 3.4. In contrast to that, physiology and pathophysiology of ECC was
divided in the following points with respect to the automatic control aspect of this work.
- The articial environment, with the major pathophysiological alterations.
- Pathophysiological response to ECC of transport functions of blood and blood loss.
- Blood component dysfunction, foreign surface interaction, haemolysis inuences, etc.
- Organ changes and dysfunctions.
- Changes to the vascular system.
23
3 Extracorporeal Circulation
With regard to the main question what changes are introduced by CPB circulation and what
changes may occur on a particular dysfunction of the HLM, the rest of the subsection is organ-
ised. Physiological and pathophysiological changes that occur during CPB and inuence the
automatic control system are highlighted, where other non-inuential eects may be neglected.
Anti-
coagulation
Hypo-
thermia
Haemo-
dilution
Chirurgical
trauma
Anesthesia
Haemo-
dynamics
Diseases
(cardiovasc.)
ECC
Figure 3.4: Pathophysiological factors for extracorporeal circulation (ECC).
3.2.1 The Articial Environment
The articially generated environment aects the patients physiological system by means of
changed conditions, like haemodilution, hypothermia and haemodynamics.
Haemodilution is the increase in the uid content of the blood resulting from priming the HLM
with a priming solution uid. The extracorporeal circuit needs to be primed with either donor
blood, isotonic saline or colloid solutions before establishing the CPB. Priming the blood with
substances other than donor blood is done to overcome the aliated problems. An increased
viscosity, haemolysis, a transfusion reaction and the transmission of infections are the potential
risks if donor blood is used during hypothermia. These problems are overcome at the expense
of a decreased haematocrit (up to 20-50 % of the original value) and the risk of postoperative
damage, such as the formation of oedemas. On the other hand, the use of crystalloid or colloid
priming uids decreases the bloods viscosity and therefore works against the increasing viscosity
eect of hypothermia.
24
3.2 Pathophysiology of Extracorporeal Circulation
Induced hypothermia is the cooling of the blood and the human body. Common CPB ap-
plications of hypothermia range from moderate hypothermia (34
C) to complete circulatory
arrest (at about 10
C). Hypothermia is used in cardiac surgery to damp the metabolic rate,
protecting the tissue and the organs. Advantages of hypothermia are the decrease in the ow
rate due to a rise in the total peripheral resistance (TPR), which results in a lower traumatisa-
tion of the blood and in case of failure of the HLM in more patient safety because of the cooled
and protected organs, see Table 3.1.
Table 3.1: Percentile O
2
-consumption and time of HLM shutdown until tissue damage occurs under
hypothermia [68].
Temperature O
2
-consumption Time of HLM shutdown
[
C] [%] [min]
37 100 4-5
29 50 8-10
22 25 16-20
16 12 32-40
10 6 64-80
A disadvantage of hypothermia is the rise in the viscosity of the blood shown in Figure 3.5,
which can be compensated by blood priming. The eects of a left drift of the O
2
-saturation
curve due to low temperature can also be seen as a disadvantage since there is a decrease of
O
2
-delivery to the tissue. This disadvantage can be overcome by the adaption of O
2
-partial
pressures to the temperature. The sludging eect of the erythrocytes at low temperatures,
which can block the capillaries, is also lowered by the priming of blood.
The haemodynamics during CPB depend on the perfusion strategy and can be the greatest
articially induced change. The ideal qualities of a blood pump are that of a heart: minimum
haemolysis, pulsatile ow and adjustable stroke volume. However, despite the advanced design
techniques of modern blood pumps, the blood is still damaged by shear stress occurring in
turbulent ows in the pumps and the extracorporeal circuit. The elements between arterial
blood pump and aortic input cannula (oxygenator and arterial lter, see Figure 3.1) prevent
a physiological pressure curve in the aorta, even if a physiological pulsatile ow curve is gen-
erated by the blood pump. The reason for this change is the additional resistance of these
elements and the impedance that changes the dynamics. The long existing debate of pulsatile
vs. non-pulsatile and centrifugal vs. roller generated ow has produced conicting results
25
3 Extracorporeal Circulation
8
7
6
5
4
3
blood viscosity [mPa s]
at = 213 mPa
haematocrit = 40%
temperature [C]
20 25 30 35 37
Figure 3.5: Increase in the viscosity with decreasing temperature (hypothermia), at a constant shear
stress of [62].
[21, 30, 31, 89, 124, 131, 140].
In pulsatile vs. non-pulsatile ow, a key issue is the systemic vasoconstriction in certain local ar-
eas after application of CPB with the possibility of potential organ damage. It has been reported
that an increase of peripheral resistance due to multifactorial reasons, such as catecholamine
release, activation of the renin/angiotensin system, vasopressin release and local tissue factor
release occurs [126, 127, 128]. Additionally, the carotid baroreceptors are suggested to be in-
volved in the vasoconstriction process during non-pulsatile perfusion [7]. Pulsatile perfusion
increases the microcirculation and the local tissue metabolism postoperatively. However, most
patients do not suer from non-pulsatile perfusion because of the use of vasodilative agents.
In addition to the typical advantages and disadvantages resulting from technical constitution,
roller and centrifugal blood pumps dier mainly in the anity to microembolisation and haemol-
ysis. It has been reported that spallation is common with roller but minimal with centrifugal
pumps [98] and haemolysis is signicantly lower with centrifugal pumps [40].
For the haemodynamic values during onset, maintenance and weaning of CPB, see Section 3.4
below.
3.2.2 Pathophysiological Response
The important changes in haemodynamics and the vascular system will be studied in an extra
section in this chapter. Additional changes in blood volume, blood hormone levels (endocrine),
uids and electrolytes appear during CPB and may lead to signicant pathophysiological re-
sponses.
26
3.2 Pathophysiology of Extracorporeal Circulation
The abnormally high release of endocrine hormones can be attributed to the shock-like stress
conferred by ECC. Hormone levels rise on the application of ECC and continue at a high
level after weaning. Antidiuretic hormone (ADH), renin, cortisol and catecholamine levels are
known to rise with induced anaesthesia and surgery alone. ECC further accelerates the distri-
bution of prostaglandins and serotonin, as well as the levels of epinephrine and norepinephrine.
The increased level of these hormones can be traced back to the application of hypothermia,
haemodilution and low perfusion.
Body uid and electrolyte disorder have been observed during CPB [101]. After the application
of ECC the levels of extracellular uid and exchangeable sodium were found to be increased,
where the total body potassium was decreased. Hypokalemia is of most important signicance
because of the potential of developing arrhythmias.
To guarantee a satisfactory venous blood return to the HLM, lost blood volume resulting from
interstitial uid shift and renal excretion has to be replaced adequately.
3.2.3 Blood Component Dysfunction and Oxygen Transport
The priming of the blood, blood haemolysis, brin formation and clotting as well as immune
defense response aect the dierent blood components and therefore also the haemodynamics
and the oxygen transport.
Haemolysis is the excessive breakdown of red blood cells (erythrocytes) and occurs in the
HLM in areas with high shear stress rates in the blood, i.e. in regions of turbulent ows.
High shear stress rates also damage other blood cells, which is not so critical in terms of
immunological or haemodynamical response. If erythrocytes are damaged or destroyed, the
contained haemoglobin is distributed to the blood plasma, where it is bound to haptoglobin.
If the binding capacities of the haptoglobin are saturated, the haemoglobin circulates in the
blood and is excreted by renal tubulus cells. In modern HLM systems haemolysis, should not
be a problem and the development of a deciency of erythrocytes (anemia) is seldom. For
the modelling and control of blood-gas transport the decreasing haemoglobin value has to be
considered.
Foreign surface contact of the blood leads to platelet (thrombocyte) activation and the
formation of brin. Even though in modern HLM systems heparin-based and biocompatible
surfaces, as well as the application of heparin as an anticoagulant are available, the eects
27
3 Extracorporeal Circulation
induced by the complex multiple biological reactions involving whole defensive systems, such as
the coagulation, complement, brinolytic, kallikrein and kinin systems can only be reduced. If
anti-coagulation techniques are not properly applied, these systems may lead to the activation
and consumption of thrombocytes, the formation of brin and clotting and other biological
reactions, which may also aect heart, brain and other organs, known as systemic inammatory
response or post-perfusion syndrome.
The immune defense response bases on an irritation of the defensive system due to ECC.
Tissue damage to the endothelium and the organs can be the result of the defensive action,
which can be damped by hypothermia and vasoconstriction. Microbiological defense by the
leucocytes as a part of the defensive system can also lead to tissue destruction.
Oxygen transport capabilities of the blood and of the oxygenator are changed at the onset
and degrade during CPB. Oxygen transport inuencing variables can be seen from the oxygen
binding curve (refer to Figure 2.3) and are haemoglobin and temperature. The haemoglobin
content of the blood is changed at the beginning of CPB and changes slowly during CPB
(due to haemolysis, etc.). Degradation of oxygen transport capabilities can also depend on the
oxygenator. The diusion capacities are known to degrade over the course of CPB because of
the blood clotting eect in the microporous membranes of the oxygenator [37]. This means
a change to the inherent system properties of the coupled oxygenator-blood system that is
important for automatic control and is modelled as uncertainty.
3.2.4 Pathophysiological Response of the Vascular System
The response of the vascular system to the application and maintenance of ECC is certainly
complex and constitutes a crossre of dierent, superposed reactions and interconnections be-
tween those reactions and connected systems. Vasoactive substances and their eects on the
vascular system are covered in an extra section of this chapter (see below). Figure 3.6 shows
a principal haemodynamic course of a hypothermal CPB. After onset, the TPR drops and is
almost halved. The initial drop in TPR can be explained by the drop in viscosity because of
blood priming with non-sanguine uids. Additional inuences are applied hypothermia (paral-
ysis of vascular musculature), dilution of circulating catecholamines, the increased distribution
of histamine and the application of anaesthetic or vasoactive drugs (see below). There are other
factors that inuence this process, but even the involvement of the above-named processes is not
28
3.2 Pathophysiology of Extracorporeal Circulation
clear up to now and subject to research [62]. The subsequent increase of TPR after onset can
be explained with an increased catecholamine release, a compartmental uid shift, activation of
the renin/angiotensin system, vasopressin release and local tissue factor release. Furthermore,
arteriolar reactions on local metabolic activity, the acidity state, the amplitude of pressure and
ow and the formation of shunts have been suggested inuential in TPR rise. The inuence
of microcirculation on this process remains uncertain. The TPR of the vascular system as a
characteristic value can change up to and more than 100 %, where most of the resistance
change is due to the arterioles and not due to the larger arterial blood vessels or the aorta. In
addition to the TPR, the compliance
1
of the vascular system and the inertance
2
of the uid
have been reported to change during ECC [109]. Inertance changes are due to the blood prim-
ing and the changed physical characteristics (viscosity) during ECC. The vascular compliance
changes when vasoactive substances or anesthetics are applied (see below), however, not much
is known about other eects during ECC.
3.2.5 Organ Response
ECC involves the use of a prosthetic device and its main alteration to the physiological human
system is the contact of blood with a foreign articial surface, which replaces the endothelial
surface and the reticulo-endothelial system. If properly used, that is if well-known perfusion,
pH and blood-gas management regimens are preserved, ECC keeps the functions of body or-
gans unchanged during the conduct of the extracorporeal procedure. However, decreased blood
volume (hypovolemia), caused by haemodilution and hypotension, hypothermia, non-pulsatile
perfusion, haemolysis and anesthetisation depending on pre-operative heart disease may result
in tissue damage, temporary or permanent neurological damage, single or multiple organ failure
and inammatory response, depending on the length of CPB. Errors due to manually controlled
perfusion dynamics and blood-gases may contribute to this process. Prosthetic surface interac-
tion and blood exposure with the resulting blood component damage and complement and other
mediator activation can be identied as the major physiological alteration caused by ECC.
1
The compliance in physiology is a measure for the distensibility of body structures. It is used for the charac-
terisation and quantication of the elastic properties of the considered tissues.
2
The inertance of a uid is the eective mass for the modelling of the inertia when uid acceleration is regarded.
29
3 Extracorporeal Circulation
TPR
(mmHg/( )) lmin
-1
Hct
(viscosity)
buffer base
(mmol l )
-1
BP
(mmHg)
CI
(lmin m )
-1 -2
ECC flow
(lmin m )
-1 -2
Onset Weaning
On bypass
3
2
100
60
30
20
0
-5
3
2
1
time
Figure 3.6: Principal haemodynamic response during conventional cardiopulmonary bypass with
haemodilution, hypothermia and low ow, where CI is cardiac index, BP is blood pressure, TPR
is total peripheral resistance and Hct is haematrocrit [62].
3.3 Anaesthesia for Cardiopulmonary Bypass
Anaesthetics and other drugs for CPB are administered in the commencement (pre-operative),
the running and termination (weaning) of CPB, but also post-operatively. Administration
is eected by inhalation and intra-venous injection. During ECC, the anaesthetist has to
guarantee a proper common anaesthesia by means of unconsciousness (hypnosis), painlessness
(analgesia), damping or disconnection of certain reex activities and the reversible paralysis
of skeletal muscles (muscle relaxation). In addition, haemodynamics, blood-gases and other
ECC eects (hypothermia and haemodilution) have to be monitored to predict the distribution
and the eective circulation period of these substances (pharmacokinetics). Preliminarily to a
success of CPB is the knowledge of
- the haemodynamics in terms of cardiac disorders and the inuence of the anaesthetic
procedures
30
3.4 Application of Cardiopulmonary Bypass
- the pharmacokinetics and pharmacodynamics
3
- the available monitoring methods
- dierent procedures and techniques in critical phases of the CPB.
Eects and mechanisms of cardiovascular substances and solutions are complex and often un-
predictable by theoretical models. A detailed study of each drug is necessary and no general
model for the eects and mechanisms of all drugs is known to exist. Drugs used in CPB can
be divided in dierent categories,
- anaesthetics (inhaled or injected)
- intra-operative administered drugs (e.g. electrolytes, anticoagulation, haemostasis, anti-
brinolysis, etc.)
- inotropic (heart muscle contractive) and vasoconstrictive agents
- muscle relaxants and vasodilative drugs.
Some of these substances have strong adverse eects on the compliance and the total peripheral
resistance of the vascular system. This will be covered in detail if needed in Chapter 4.
3.4 Application of Cardiopulmonary Bypass
With regard to anaesthesia, the methods for the application and management of CPB can be
distinguished for dierent phases. In the pre-operative stage the perfusion technicians work
together with the surgeon and the anaesthetist to obtain background information about the
patient, such as all perfusion relevant data (for example age, height, weight, etc.) and the
clinical history (hypertonia, renal values, diseases, etc.). During the onset of CPB, the HLM
is congured with the guidelines of the cardiosurgical hospital and primed with the priming
solution of choice (see above). A second HLM backup system should be kept in addition to the
normal HLM. The onset of the HLM system is a critical phase during CPB. After heparinisation
of the blood, aorta and vena cava are cannulated and connected to the machine, Figure 3.7. The
ow of the HLM then is continually increased while the ventricle is still beating. During the
maintaining stage, the HLM variables are maintained with extensive monitoring, corrections
3
The study of the biochemical and physiological eects of drugs and the mechanisms of their actions, including
the correlation of their actions and eects with their chemical structure.
31
3 Extracorporeal Circulation
are made and the heart-surgery can take place. In the last stage, the weaning, the patient is
weaned from the machine.
3.4.1 Onset Stage
The onset begins with the conguration and installation of the HLM. Mainly two dierent HLM
structures are possible and used in modern HLM systems. In an open system, the venous return
and the sucker lines are returned to an open reservoir, from which the blood is pumped to the
arterial line. In a closed system, a closed soft bag and an open reservoir are connected to each
other (refer to Figure 3.1). Because of the venous soft bag the system is likely to collapse on
clearance and therefore closed for air embolism. After conguration and installation the HLM
is primed with donor blood or blood substitute substances. In addition to the heparinisation of
the patients blood, donor blood or priming substance in the HLM is heparinised. Hypothermia
is induced after application of ECC, which reduces coronary and bronchial backow to the heart
Aorta
Truncus
pulmonalis
(TP)
Right ventricle
(RV)
Right
atrium
(RA)
SVC
Venous outflow (VO)
Vena cava inferior (IVC)
Vena cava superior (SVC)
IVC
RV
A
B
1
2
LV
D
C
Arterial inflow (AI) Clamp
Arterial
cannula
Figure 3.7: Application of cardiopulmonary bypass, with venous cannulation to the left and arterial
cannulation to the right, where the arrows for SVC and IVC are the corresponding venous ows.
The points A)-D) on the right-hand side in the gure refer to the dierent cannulation techniques,
for example for cardioplegia inow 1) or left ventricle (LV) suction [129].
32
3.4 Application of Cardiopulmonary Bypass
and provides, with other advantages, better visibility conditions in the operating eld.
3.4.2 Maintenance Stage
Extensive monitoring of dierent patient values is necessary to guarantee a successful and safe
CPB. Among the most important values are blood-ow, blood-gases (pO
2
and pCO
2
), pressure
(arterial and venous), temperature, pH-value, coagulation/clotting status and urine excretion
rate. In addition to these values, the electroencephalography (EEG) can provide additional
information about anaesthetic conditions.
There is no international norm guideline for certain values, their maintenance and procedures.
However, recommendations that are based on clinical and physiological experiences exist (refer
to Table 2.2). Values such as oxygen partial pressure pO
2
and arterial blood-ow q
art
have to
be adapted to articial conditions depending on haemodilution and body surface (for example
pO
2
160 mmHg, pCO
2
40 mmHg, p
art
40-60 mmHg and q
art
2.4 l/min/m
2
body
surface are typical values). This adaption is achieved by the manual adjustment of control
variables, see Figure 1.1. The gas-ow to the oxygenator and the FiO
2
-value in the gas are for
example changed to achieve the right arterial blood-gas conditions. Additionally, the arterial
blood-ow may be changed to react appropriately to a change in the venous oxygen saturation
value.
For the blood-gas management during hypothermic CPB two methods exist. In the case of the
pH-stat method the measured values are corrected with translation tables to values at 37
C.
In contrast to the -stat method, the pH-stat method needs the additional mixture and control
of CO
2
gas in the oxygenator. The -stat method uses (measures) and controls the values at
37
C and tolerates the shifting pH-values at lower temperatures. -stat is the simpler and
more secure method because the blood-gases are given at 37
C and no correction is necessary.
Experimental measurements show [29], that -stat keeps the cerebral auto regulation intact
(in contrast to pH-stat) and the global cerebral perfusion seems to be adopted to the patients
metabolic needs.
The above-named vital variables have to be continually observed and small to large adjustments
in the control values are needed to maintain these values under changing conditions. Even
under automatically controlled blood-gas and perfusion the perfusion technician is essential to
determine control setpoints and react to articial disturbances (blood clotting, blood loss, etc.)
or unpredictable system failures (pump failure, blood-gas supply failure, etc.).
33
3 Extracorporeal Circulation
3.4.3 Weaning and Postoperative Stage
Weaning is the end of myocardial blood supply restriction (ischemia) up to the total discon-
nection of ECC. The oxygen consumption of the heart, represented by the heart muscle wall
tension, has to be kept at a minimum to avoid damage. A slow increase of heart ejection is
achieved by regulating the venous backow to the heart and during that stage the blood and
the body are gradually rewarmed. Unphysiological pressures should be avoided until the HLM
is nally disconnected. The remaining blood in the HLM can be reconditioned.
After thorax occlusion, the patient is subject to intensive care and anaesthetic treatment, to
stabilise the haemodynamics and reduce the risk of post-operative damage and inammatory
response.
During the weaning stage, the automatic control of the HLM can be further used by adapting
the control setpoint to required weaning changes.
34
4 Modelling of the System under Extracorporeal
Circulation
This chapter addresses system modelling under cardiopulmonary bypass (CPB) for the devel-
opment of control. Before introducing the models for haemodynamic and blood-gas processes,
a short section will address the control strategy. This control strategy section gives an overview
on control of CPB in literature, the suggested feedback variables, the control actuating princi-
ples and limitations, see Section 4.1 for the haemodynamic control strategy and Section 4.10
for the blood-gas control strategy.
The system is modelled in technical and physiological subsystems, which are interconnected
thereafter. The system modelling approach will consider haemodynamic modelling divided in
Sections 4.2-4.9 for each component rst and is followed by the modelling of blood-gases in
the oxygenator, Section 4.11. In the oxygenator, the technical subsystem is coupled in an ex-
tracorporeal state with the physiological organ blood. Therefore, modelling of technical and
physiological parts will be done in the same section.
Each modelled subsystem follows a short simulation and validation in Chapter 5 Model vali-
dation. Numerical constants are not included in this chapter and appear in Appendix B.
4.1 Haemodynamic Control Strategy
Dierent strategies on haemodynamic control of CPB exist in literature. The rst applications
of automatic control date back to the 1960-70s and were focused on the maintenance of an
appropriate pressure and ow in the venous return [130]. This was necessary since at that
time in most HLMs a venous withdrawal pump was used, which had to be monitored for safety
reasons (venous chatter). Several authors developed further control strategies, introducing PI
and rule-based control algorithms [17, 18, 35, 123]. Experimental implementation and in-vitro
tests produced good results. In these studies, ow in the central venous return was controlled
35
4 Modelling of the System under Extracorporeal Circulation
by a special withdrawal pump, and arterial perfusion was kept at a constant rate. Open-loop
control and safety mechanisms for aortic inow were successfully developed. Further feedback
control in CPB has been reported in recent years. Model predictive control algorithms were
developed and successfully tested in simulations for stationary and pulsatile perfusion with
roller pumps during CPB [115, 116]. Arterial blood-ow and venous partial oxygen pressure
were the primarily controlled variables in this study without in-vitro or in-vivo test series.
Other authors presented in-vivo experimental results with good feedback control performance
for arterial blood-ow control [5, 6, 95]. In these studies, stable but relatively slow fuzzy and
proportional control algorithms were developed for automatic perfusion with roller pumps.
When considering the literature on haemodynamic control of CPB, it becomes clear that the
control strategy and controlled variables also depend on the application of the HLM technique.
According to the HLM system used in this work (see Figure 3.1), the arterial haemodynamics
are to be controlled, where venous conditions and reservoir height have to be monitored. Table
4.1 summarises variables of main importance in the case of a CPB. However, normal conditions
as found in CPB literature (see Section 2.3.3) have to be maintained.
When considering haemodynamic arterial control, the reference values for both arterial pressure
and arterial ow have to be adjusted. Regarding the situation during CPB, this is only possible
with two dierent control actuators, the arterial blood pump and a vasoactive drug delivery
system (to adjust the TPR by changing the vessel tonus). The introduction of such a drug
delivery system was discarded in this work because of possible severe adverse eects and the
dicult testing methods (in-vivo tests series would be necessary). It has to be mentioned,
however, that such pressure control systems have already been developed and shown to work
successfully in animal experiments [102]. On the other hand, arterial blood-ow and pressure are
both of importance. The arterial blood-ow determines vital functions as for example transport
Table 4.1: Haemodynamic variables and conditions for control.
Variable Control conditions
arterial ow q
art
= q
aort
continually updated / may be pulsatile
2.4 l/min/m
2
mean ow / body surface area
aortic pressure p
aort
continually updated / may be pulsatile
40 p
aort
60 mmHg
venous ow q
ven
should be monitored
venous ow p
ven
should be monitored
reservoir height h
res
should be monitored
36
4.2 Centrifugal Blood Pump
of oxygen to and transport of carbon dioxide from the tissues, and related eects. The pressure
in the vascular system prevents the blood vessels from collapsing, therefore guaranteeing a
sucient micro-circulation. Furthermore, the pressure determines together with the blood-ow
the delivered static and dynamic energy to the tissues. A direct pressure control, such as in
physiological perfusion achieved by the various regulation mechanisms of the human body, is
also desirable in CPB conditions. Bearing this in mind, a third control strategy was developed
in addition to an arterial blood-ow and an arterial pressure control strategy. For this, the
arterial blood-ow control was augmented with a pressure boundary control, with the arterial
blood pump as the only actuator [84]. The three control strategies are explained in detail in
Sections 6.1-6.2. For the haemodynamic modelling the haemodynamics of the HLM and the
vascular dynamics have to be considered, see below. Measured variables for feedback control
are the arterial line ow and pressure.
4.2 Centrifugal Blood Pump
Today, more and more rotary or centrifugal blood pumps are used in ECC. Since the advantages
of centrifugal blood pumps outweigh the disadvantages (Section 3.1.2), a centrifugal blood
pump was chosen as the haemodynamic control actuator. The Medos DeltaStream pump
is a centrifugal (rotary) blood pump with diagonally streamed rotor, see Figure 3.3. The
DeltaStream was chosen for this work because of the good (low) time constant, low haemolysis
and good controllability.
In order to achieve better control performance, an accurate model of the blood pump including
nonlinearities should be known. Only a few models developed for rotary blood pumps exist in
literature [24]. The nonlinear model developed in this work is based on experiments and is able
to predict outow and pressure output over the whole range of static and dynamic conditions.
4.2.1 Brushless DC Motor
The rotary blood pump (DeltaStream, Medos AG, Germany) is driven by a brushless direct
current (BLDC) motor. The electrical and mechanical equations for the BLDC motor are
37
4 Modelling of the System under Extracorporeal Circulation
[71, 88]
(J
mot
+J
load
)
d
mot
dt
+b
mot
mot
= T
mot
b
load
mot
L
mot
di
mot
dt
+R
mot
i
mot
= u
in
u
emf
,
(4.1)
where J is an inertia, b is a damping, T is a torque, L is the inductance, R a resistance
and u a voltage. The indices mot, load, in, emf refer to the motor, motor-load, input and
electromagnetic feedback, respectively. The electro-mechanical interconnecting equations for
Eqs. 4.1 are
T
mot
= K
mot
i
mot
u
emf
=
2
3
K
emf
mot
,
(4.2)
where K
mot
is the motor torque constant and K
emf
the electromagnetic feedback gain. The
factor 2/3 is because of the three line coiling control of the motor. Figure 4.1 shows the electro-
mechanical connection of the BLDC motor elements. Rearranging Eqs. (4.1) and (4.2) gives
the following state space (SS) model in its linear representation
_
di
mot
dt
d
mot
dt
_
=
_
_
R
mot
L
mot
2
3
K
emf
L
mot
K
mot
J
mot
+J
load
b
mot
+b
load
J
mot
+J
load
_
_
_
i
mot
mot
_
+
_
1
L
mot
0
_
u
in
y = [0 1]
_
i
mot
mot
_
.
(4.3)
R
mot
J
load
J
mot
b
mot
L
mot
DC motor
U
emf
U
in
n
out
M
Figure 4.1: Equivalent electro-mechanical network diagram for the BLDC motor.
38
4.2 Centrifugal Blood Pump
Eqs. (4.3) are a linear second order SS system. However, certain elements comprise nonlinear
characteristics, which are either modelled to the SS representation or neglected, see below.
4.2.2 Centrifugal Pump and Nonlinear Motor Characteristics
A general nonlinear aspect of BLDC-motor systems is the temperature dependency of the
viscous damping b
mot
. Due to the cooling eect of the blood along the motor device, great
temperature changes in the DeltaStream centrifugal blood pump system do not occur and this
eect can be disregarded. The nonlinear eects following in this section were observed during
experiments and modelled in order to cope with experiments, where the experimental set-up
and methods are covered in Chapter 5.
In hydrodynamical centrifugal pump systems, the pressure output is dependent on ow and
rotational speed [24, 78]. Steady state gain measurements show that the hydrostatic pressure
dierence between pump pressure at outow and inow can be characterised by a 2
nd
-order
polynomial, additionally dependent on pressure at pump inlet
p
out
= f
1
(q
B
,
mot
, p
in
),
where p
in
is the pressure at pump inlet and assumed to be constant. Figure 4.2 shows the non-
linear static transfer function as a function of pump ow and rotary speed (data of the Medos
DeltaStream was used for this gure). The second nonlinearity observed during measurements
was the changing damping in the TF due to turbulent ow conditions in the impeller. This
eect was rst modelled as a changing damping of the load b
load
but later on neglected, since
there is no signicant impact on the model.
In addition to the hydrostatic pressure output, the motor current limit is modelled as a nonlin-
earity. Here, the generated motor torque is not proportional to motor current, but a nonlinear
saturation function.
T
mot
= f
2
(i
mot
)
39
4 Modelling of the System under Extracorporeal Circulation
1000 2000 3000 4000 5000 6000 7000 8000
rotary speed [R/min]
pump flow [l/min]
500
400
300
200
100
0
-100
-200
pressure [mmHg]
0 l/min
4 l/min
Figure 4.2: Nonlinear static pressure output.
These two nonlinear eects can be described by the following equations
p
out
= f
1
(q
b
,
mot
, p
in
) = b
2
2
mot
+b
1
mot
b
0
q
b
+p
in
T
mot
= f
2
(i
mot
) =
_
_
a
mot
i
mot
a
mot
i
mot
a
mot
< i
mot
< a
mot
a
mot
i
mot
a
mot
.
(4.4)
Eqs. (4.4) lead to the following nonlinear state space (SS) model, which is shown in block
diagram form in Figure 4.3 [110].
_
di
mot
dt
d
mot
dt
_
=
_
_
R
mot
L
mot
f
2
(i
mot
)
2
3
K
emf
L
mot
mot
+
1
L
mot
u
in
K
mot
J
mot
f
2
(i
mot
)
b
mot
J
mot
mot
_
_
p
out
= [f
1
(q
b
,
mot
, p
in
)]
(4.5)
Eq. (4.5) is now a two input one output state space system, with p
in
assumed to be constant.
Note that the blood-ow input q
b
depends on the hydrodynamic characteristics of the arterial
and venous line of the HLM circuit and the human vascular system. In the steady state case,
ow depends on the total peripheral resistance (TPR) extended by HLM ow resistance and in
the dynamic case on total hydraulic impedance of the system. Also note that nonlinear eects
40
4.2 Centrifugal Blood Pump
of air inclusion in the pump head (impeller) that highly degrade the ow in the blood pump
are neglected. However, in such a case the whole HLM system has to be stopped as a severe
system failure and air bubbles in the blood have to be removed.
4.2.3 External Rotary Speed Controller
Most centrifugal blood pumps come up with a device driver unit with a rotational speed con-
troller, which is also the case with the Medos DeltaStream system. For the modelling, the
external rotary speed controller has to be taken into the model, where it was implemented to
cope with experiments. For rotational motor speed control of the blood pump, the angular
velocity
mot
is fed back to form the error of angular motor velocities e
mot
e
mot
=
r
mot
with the control reference signal
r
. A standard proportional plus integral (PI)-controller with
anti-windup saturation was placed in the feedforward path
C(s) =
u
in
(s)
e
mot
(s)
= K
p
s +K
i
s
, (4.6)
where C(s) is the PI-controller transfer function, K
p
is the proportional and K
i
the integral
gain. The controller parameters were tuned for the analogue pump system case. In order to
get a fast control response, the system described by Eqs. (4.5) was linearised at minimum
system gain q
b
0 l/min and no current saturation. The PI-controller was tuned for that
case using the root locus procedure with the help of the Control System Toolbox (MATLAB)
[70, 94]. Tuning the controller at an operating point (OP) with very low system gain leads
([i ] ,
[u q ] )
w '
'
mot mot
in B
1
in = const.
( ,
q , p )
w
mot
B
[u , q ]'
in B
x x
p
out
Figure 4.3: Blockdiagram for the nonlinear state space system.
41
4 Modelling of the System under Extracorporeal Circulation
to underdamped system responses at system OPs of higher gain. However, the controller was
tuned to t to measurement data and stability was ascertained in simulation over the whole
operating range of the blood pump model. By using the backward Euler integration method
[97],
1
s
T
s
z
z 1
(4.7)
the system was discretised with a sampling time constant T
s
. The resulting dierence equation
for the PI-controller is
u
in
(k) = f
s
[u
in
(k 1) + K
p
([K
i
T
s
+ 1]e
mot
(k) e
mot
(k 1))] (4.8)
with k, the time of discrete sampling step instances and f
s
the anti-windup function. Anti-
windup is realised by means of a discrete integrator with saturation. With the substitution
x = u
in
(k 1) + K
p
([K
i
T
s
+ 1]e
mot
(k) e
mot
(k 1))
follows the anti-windup to
f
s
(x) =
_
_
a
, x a
x, a
< x < b
, x b
, (4.9)
with the PI-controller integrator limits a
, b
p +
v +f
f
, (4.11)
where v is the velocity of ow, is the uid density, p is the pressure, is the dynamic
viscosity and f
f
is the body force (units [N/m
3
]) applied to a volume element. , are the
Nabla and the Laplace operator, respectively. For homogeneous, incompressible Newton uids
holds = const., = const.. The continuity equation
t
+ v = 0 (4.12)
then becomes
v = 0.
45
4 Modelling of the System under Extracorporeal Circulation
If it is assumed that no external force is applied to the blood vessels
f
f
= 0.
The simplied, still nonlinear, NSE is then
v
t
+ (v )v =
1
p +v, (4.13)
with the kinematic viscosity
=
.
4.5.2 Simplied Electrical Analogue
To neglect the nonlinear terms of (4.11) and (4.12), which are the coupled wave velocities and
accelerations, it is assumed (and also shown, [103]) that the maximum and average velocities in
axial direction are small compared to the wave velocity and that the average radial velocity can
be neglected compared to the axial velocity. Linearising these negligible nonlinear terms, leads
to two joint second order partial dierential equations [96, 108]. For the further transforming
of the equations to cylindrical coordinates, a parabolic ow prole and homogeneity, elasticity,
isotropy and the validity of Hooks law for the vessel wall are assumed. Further neglecting the
second order partial derivatives and assuming an incompressible vessel wall results in two rst
order partial dierential equations,
p
x
v
= L
q
t
+R q
q
x
v
= C
p
t
,
(4.14)
where p and q describe the pressure and ow along the coordinate x
v
, which is the axis in
ow direction. Note that Eqs. (4.14) do not depend on the radial axis, which is because of
the assumed parabolic ow prole. By applying the nite dierence method using a Taylor
approximation [103, 139], Eqs. (4.14) are further simplied. Two location-dependent linear
dierence-dierential equations of rst order are the result, which describe the ow and pressure
46
4.5 The Vascular System
in a tube compartment of nite length x
v
,
p(t|x
v
) p(t|(x
v
+ x
v
)) = L x
v
dq(t|x
v
)
dt
+R x
v
q(t|x
v
)
q(t|x
v
) q(t|(x
v
+ x
v
)) = C x
v
dp(t|(x
v
+ x
v
))
dt
.
(4.15)
Note that the p(t|x
v
) and q(t|x
v
) are time-dependent functions at location x
v
. In Eqs. (4.15),
L is the uid inertance, C is the vascular compliance and R is the vascular resistance.
Finally the equations for the n
th
vascular compartment follow from Eq. (4.15) with the intro-
duction of subscript indices, substituting the length descriptions in brackets
p
n
p
n+1
= L
n
dq
n
dt
+R
n
q
n
q
n
q
n+1
= C
n
dp
n+1
dt
,
(4.16)
where x
v
is the the length of the discretised compartment. Variables p and q are time-
dependent. With x
v
n
, the parameters for compartment n are calculated
L
n
=
9x
v
n
4r
2
n
C
n
=
3r
3
n
x
v
n
2E
n
d
n
R
n
=
81x
v
n
8r
4
n
,
(4.17)
where E
n
is Youngs modulus. Figure 4.4 shows the electric analogue for a vascular element,
where the pressure p = u and the ow q = i are according to a voltage and a current respec-
tively. Eqs. (4.17) are obtained in the linearisation procedure as given in [108]. Note that
n
p
1 + n
p
n
q
1 + n
q
n
C
n
R
n
L
Figure 4.4: Electric analogue for a single vascular element.
47
4 Modelling of the System under Extracorporeal Circulation
the parameters for the computation of the elements depend on the patients physiological and
pathophysiological vascular conditions and can vary strongly.
4.5.3 Vascular Model Structure
Eqs. (4.16) were Laplace transformed for one compartment.
Q
n
(s) =
1
sL
n
+R
n
(P
n
(s) P
n+1
(s))
P
n+1
(s) =
1
sC
n
(Q
n
(s) Q
n+1
(s)),
(4.18)
where s is the Laplace operator. Eqs. (4.18) were implemented in a MATLAB/Simulink block
diagram as a transfer function. Figure 4.5 shows the Simulink block diagram, where p
n
and q
n
Transfer function
1
s
1
1
2
2
p
n
p
n+1
q
n
q
n+1
Gain Integrator
+
_
+
1
L s + R
n n
_
C
n
-1
Figure 4.5: MATLAB/Simulink implementation for a basic compartment.
are the pressure and ow at the compartments input, p
n+1
and q
n+1
are the pressure and ow
at the compartments output.
4.6 Vasoactive Drug Extension
During ECC, the vascular system is subject to dierent vasoactive drug infusions. Vasoac-
tive drugs are specically designed vasoactive agents, for either hypotensive or hypertensive
treatment. In most cases narcotic drugs also show an adverse vasoactive eect on the vascular
system, which is mostly of a dilative nature.
48
4.7 Volume Distribution Model
To describe the vasoactive inuence on the vascular system dependent on time, a volume dis-
tribution model is incorporated in the vascular model. Distribution of the drug is modelled by
perfect mixing chambers, coupled to each vascular element. The time-varying nonlinear inu-
ence of the vasoactive drug agent on the vascular resistance and compliance is not precisely
known up to the present. There exists, however, a huge body of scientic and clinic applica-
tion studies of dierent vasoactive and narcotic drug application and central arterial pressure
control, for example [33, 42, 45, 54, 61, 67, 69, 141].
For the modelling of the drug inuence, static TPR and compliance data from in-vivo mea-
surements, where the narcotic and vasoactive drug was applied, can be used. These results are
obtained from clinical studies and are used here to map the concentration of a certain vascular
compartment in a percentage change of the elements of that compartment.
4.7 Volume Distribution Model
In a vascular segment, the blood volume consists of unstressed and elastically stored volume.
The blood volume describing equation is [107]
V
nT
= V
nU
+p
n
C
n
, (4.19)
where the total volume V
nT
of compartment n is the sum of unstressed volume V
nU
plus the
elastically stored volume p
n
C
n
. Depending on the pressure in a vascular compartment, the
vessel distends and more volume is stored. The compliance that determines this stored volume
is assumed to be linear in its operating range. The concentration in each compartment is the
ratio of vasoactive substance volume V
n
to total blood volume
c
n
=
V
n
V
nT
, (4.20)
where the substance concentration is c
n
. With the substance concentration in each compart-
ment, the substance ow q
n
depends on the direction of blood-ow in that compartment [107]
q
n
=
_
c
n1
q
n
q
n
0
c
n
q
n
q
n
< 0
. (4.21)
49
4 Modelling of the System under Extracorporeal Circulation
Finally the actual compartment substance volume is determined by the integral of the dierence
of substance volume inow and outow
V
n
=
t
_
0
(q
n
q
n1
)dt +V
n
(0). (4.22)
A time constant k
s,Drug
, corresponding to the substance half-life time (HLT), has to be intro-
duced and Eq. (4.22) becomes
V
n
=
t
_
0
(q
n
q
n1
k
s,Drug
V
n
)dt +V
n
(0). (4.23)
Eqs. (4.19), (4.20), (4.21) and (4.23) were implemented in parallel to the compartments of
the model and according to substance volume concentration of for example anesthetic or va-
soactive drugs the system parameters were changed. The parameters of the model therefore
change in dependence of drug application. Two functions, f
R
and f
C
map the vasoactive drug
volume in the dierent compartments to parameter changes of resistance R
n
and compliance
C
n
and are linear mappings tted to practically measured data. That is K
R,Drug
= R
n
/V
n
and
K
C,Drug
= C
n
/V
n
are constant gains, which cause percentage changes to the vascular para-
meters. Substance outow at the venous return is fed back to the arterial inow with a special
HLM compartment, describing the substance ow dynamics in the ECC circuit (tubing, venous
bag, etc., modelled with Eqs. (4.19), (4.20), (4.21) and (4.23)). Another variable parameter
is the HLT of the vasoactive substance. The integral time constant k
s,Drug
was obtained by
calculating the time constant of a rst order dierential equation for the HLT of the drug.
In the case of an injection of vasoactive drugs to the HLM, the substance distributes dynami-
cally over the vascular system, changing vascular resistance and compliance. Nonlinear eects
(e.g. saturation), as well as varying inuence of the vasoactive substance to certain vascular
regions (e.g. venous or arterial system) are disregarded, since typical ranges for the injection are
kept during simulations. Note that the vascular system (Eq. 4.16) extended with vasoactive
substance distribution is thereafter nonlinear and its parameters are time-varying. In addi-
tion to that, uncertainty has to be assumed for the vascular parameters of dierent patients,
including dierent pathophysiological changes.
50
4.8 Model Interconnection and Augmentation
4.8 Model Interconnection and Augmentation
The model of the vascular system and the vasoactive substance distribution equations are
interconnected and augmented with the subsystems centrifugal blood pump (CBP), oxygenator
and arterial catheter.
The overall system can be described as a nonlinear time-variant multiple input, single output
system, with system uncertainties, including disturbances
x = f(x, u) + f(x, u, t)
y = cx,
(4.24)
where the state vector x consists of systems ow, pressure, vasoactive drug volumes, pump
motor current, pump motor speed and controller output. Input vector u is the CBP input volt-
age for rotary speed reference and vasoactive drug or narcotic drug volume ow u = [
mot
V
in
].
Output vector c is a linear mapping of the system state aortic inow q
aort
or the system state
aortic pressure p
aort
, depending on control usage of the model. System dimensions are u R
2
(if only one vasoactive substance at a time is modeled) and y R. Note that the state dimen-
sion depends on the vascular compartment model, but is x R
18
with the six compartment
model, see Chapter 5. f() and f() are smooth nonlinear functions (i.e. f(), f() C
).
f() contains the system uncertainties and functions for time-variant parameters.
Figure 4.6 shows the block diagram of the interconnected model for aortic blood-ow control
(with aortic ow as the output in this case). Blocks in double brackets represent system dynam-
ics with nonlinearities. p
out
is the pressure at pump outlet, p
aort
the pressure at aortic input and
s
(Vasoactive Substances) the input vector of dierent substance volume ows. The transfer
functions
mot
-Control/M + BP/NPO, O + K and VS correspond to the rotary pump circuit
(speed controlled), the oxygenator and catheter and the vascular system augmented with vol-
ume distributions, respectively. NPO is the static nonlinear pressure output relation, described
by Eq. (4.4).
4.9 Modelling of Regulation Mechanisms
The regulation mechanisms of the body addressed in Section 2.5 react in a dierent way on
the vascular system than in physiological conditions. Local, neural, nerval and humoral control
51
4 Modelling of the System under Extracorporeal Circulation
-
(z) E(z)
mot
-
control
U (z)
in
H
S / H
M + BP O + K VS
Q (s)
Aort.
P (s)
out
P (s)
Aort.
mot
(s)
HLM Patient
Vasoactive
substances
NPO
Figure 4.6: Block diagram of the modelled system for haemodynamic control with dierent compo-
nents.
mot
-Control: rotational PI-controller, H: hold element, S / H: sample and hold element,
M + BP: motor plus blood pump, NPO: Nonlinear pressure output, O + K: oxygenator plus
cannula, VS: vascular system.
of haemodynamics are all subject to the articial environment during a CPB. The traumatic
stress, as induced by the opening of the ribcage, haemodilution and foreign surface contact of
the blood invoke dierent reactions of the control systems on the vascular system.
The nervous system can be assumed to be suciently damped to invoke strong reactions, but
the loss of vasoconstrictor tonus means an decrease of TPR. One of the most important in-
uences on the vascular system during ECC is the catecholamine release, which is suggested
to depend on humoral and in parts on nervous control mechanisms. Catecholamine release is
opposed to the initial drop of TPR on the onset of ECC, which can be explained with other
superimposed mechanisms. Local haemodynamic tissue control is assumed to have little eect
on the vascular resistance and compliance, since dierent concentrations (main inuence pH-
value) are kept at a physiological level (refer to Chapter 2).
Little is known about the dierent regulation mechanisms during ECC. Experimental stud-
ies measure the concentrations of dierent factors (e.g. catecholamines like epinephrine) but
reactions and interactions of the above-named mechanisms are hardly identiable or quanti-
able. Since general haemodynamics, like pressure and ow in the aortic arch, can be observed
during ECC with ease, the vascular parameters were assumed to have an extra uncertainty de-
pending on the haemodynamic regulation mechanisms. Figure 3.6 shows the haemodynamics
during CPB. According to the changes in pressure at constant blood-ow, TPR changes can
be determined. Uncertainties in resistance were assumed in the model and considered in the
robust control approach. Note that up to the present almost nothing is known about compli-
ance changes during ECC. Compliance uncertainties are assumed to exist and are based mainly
on observations of vasoactive substances, muscle relaxants or narcotics. These were of course
modelled as parameter uncertainties in the robust control approach.
52
4.10 Blood-Gas Control Strategy
4.10 Blood-Gas Control Strategy
For automatic control of blood-gases during CPB, only few approaches exist in literature.
Besides the control algorithms, dierent control inputs and sensors were used, as there are
certain dierent possibilities for this.
The control of oxygen delivery to the tissues, for example, was controlled with the blood-ow,
employed by [76, 77]. The process value hereby was the venous oxygen partial pressure. In
addition to that, venous carbon dioxide partial pressure was controlled with the total gas ow
through the oxygenator. This control strategy was avoided in this work since the change in the
blood-ow can lead to collapsing vessels (if too low) or cause vascular damage (if too high).
The control of venous blood-gas partial pressures was also avoided, since high and intoxicating
partial arterial pressures may occur, depending on the control input signal. A control of venous
blood-gas partial pressure with the oxygen fraction as the control input was developed in a
simulation study by [115].
With the control of arterial gas partial pressures in the blood an appropriate O
2
-supply and
CO
2
-removal can be guaranteed at a sucient arterial blood-ow. As the control input for
arterial oxygen partial pressure pO
2,a
the O
2
-fraction in the gas-ow through the oxygenator
is often used. In contrast to that, the arterial carbon dioxide partial pressure pCO
2,a
in the
blood can be controlled by either applying CO
2
-gas to the gas-ow through the oxygenator or
by the total gas-ow. The implementation of such a strategy for O
2
-control only, was reported
by [4]. This study included a linear-quadratic-gaussian self-tuning control system and showed
good performance in in-vitro experiments. It was an improvement on former control strategies,
which mainly relied on linear PI, PID or piecewise linear control. Instability problems in these
studies occurred with linear controllers not properly tuned, or when process dynamics changed
[74]. A control strategy, where pO
2,a
is controlled by the control input oxygen-fraction FiO
2
and
pCO
2,a
is controlled by the control input total gas-ow q
g
, seems to be most appropriate [46].
This control strategy is often used in cardiovascular surgery. The drawback of this method is the
coupled control input of gas-ow q
g
to the oxygen process input (see below) and the nonlinear
process characteristic. Despite these limitations that should be overcome with an advanced
control method, this control strategy has several advantages, such as: The good acceptance
and practical usage (up to the present this strategy is used for manual control); the saving
of an additional CO
2
-gas supply; a fast and reliable method to obtain almost physiological
blood-gas supply. For the modelling section below the control input for the oxygen process is
53
4 Modelling of the System under Extracorporeal Circulation
the oxygen fraction FiO
2
. For the carbon dioxide process the control input is the total gas-ow
q
g
. With these values the appropriate gas-ow through the oxygenator is adjusted by the gas
blender (see below). The arterial gas pressures pO
2,a
and pCO
2,a
are fed back as the process
values. Finally, reference values taken from CPB literature apply also for blood-gas control,
refer to Section 2.3.3.
4.11 Membrane Oxygenator Modelling
Blood-gas exchange in the lungs and transport of gas in the blood are dicult mathematical
processes incorporating nonlinearities and process time-delays. Blood-gases are exchanged in
the lung/oxygenator and the gases are transported to the body tissues via the blood stream.
The concentration of blood-gases in the venous blood depends on the metabolic activity of
the human body tissues. The oxygenation/carbon dioxide removal process in a membrane
oxygenator is similar to that in the human lung. For the modelling of the blood-gases during
ECC, a model that describes the dynamics of O
2
- and CO
2
-exchange [48, 49, 50] was adapted
to a membrane oxygenator system [47] and used for the development of an automatic control
strategy. Most important dierences to the physiological lung blood-gas exchange are the
changed diusion capacity of the oxygenator and the state of ECC. Hypothermia leads to a
left shift of the O
2
-binding curve and hypothermia together with anaesthesia reduce oxygen
consumption and carbon dioxide production in the tissues.
The modelling method for blood-gas exchange is compartmental modelling and is based on
the volume accounting equations of the compartments gas, plasma and red blood cells. Figure
4.7 shows a mixing chamber, as a generic compartment, with diusion along the oxygenator
membrane. The equation that describes the component i is [47, 57]
V
mc
d[C]
i
dt
= q
b,in
[C]
i,in
q
b,out
[C]
i,out
+D
i
(p
i,ext
p
i
) + R
i
, (4.25)
with V
mc
, the volume of the compartment/mixing chamber, d[C]
i
, the concentration of compo-
nent i, q
b,in
, q
b,out
, blood in- and outow and [C]
i,in
, [C]
i,out
, concentrations of component i at
in- and outow (inow concentrations of components correspond to venous conditions in the
membrane oxygenator). D
i
is the diusion capacity over the oxygenator membrane and nally
R
i
is a disturbance of the component i, corresponding to a loss or a chemical reaction. For the
54
4.11 Membrane Oxygenator Modelling
p
i,ext
D
i
p
i
[C]
i
V
mc
q
b,out
q
b,in
[C]
i,in
[C]
i,out
Blood
Membrane
Gas
Figure 4.7: Exchange of gases by diusion over a membrane between blood and gas compartment.
modelling below, equations similar to (4.25) will be ordered in gas, oxygen and carbon dioxide
(and related) equations and perfect mixing of components will be assumed, i.e. [C]
i,out
= [C]
i
.
4.11.1 Gas Mixing Strategy
Most often the ventilating gas used for the oxygenation process in a membrane oxygenator is
a mixture of pure oxygen (O
2
) and a carrier gas, which is either nitrogen (N
2
) or ambient air
(21 % O
2
, 79 % N
2
). By adjusting the fraction of the pure O
2
-volume ow in the gas mixer
the oxygen fraction of the total gas stream that ows through the oxygenator can be achieved.
Possible FiO
2
-values depend on the carrier gas that is used. Values of FiO
2
= [0.21 1] for
ambient air or FiO
2
= [0 1] for N
2
can be achieved. Higher oxygen fractions in the gas mean
higher partial pressures in the oxygenator gas compartment and therefore a higher diusion
gradient, which leads in turn to a higher O
2
-partial pressure in the arterial blood. The oxygen
fraction in the gas is determined by both, the ow of oxygen (q
O
2
) and the ow of carrier gas
(q
N
2
) to the oxygenator. At a given total mixed gas-ow to the oxygenator q
g
, the ow of pure
oxygen is q
O
2
= FiO
2
q
g
and the ow of inert gas is q
N
2
= q
g
q
O
2
, if nitrogen is the inert gas.
The oxygen fraction therefore is
FiO
2
=
q
O
2
q
O
2
+q
N
2
, (4.26)
which can range from 0 to 1 (or 0 - 100 %). Using Eq. (4.26) the setpoint values of the gas ows
of q
O
2
and q
N
2
are calculated according to the given values of FiO
2
and q
g
. The gas ows are
then adjusted automatically by the gas valves, with a control routine already implemented in
the gas valves. During this control process an error in the lower operating range occurs, which
is due to the gas ow valves. This control error in turn leads to a static gain error and has to
55
4 Modelling of the System under Extracorporeal Circulation
be regarded in model validation (see Section 5.6) and control (see Section 6.3).
4.11.2 The Gas Blender
The mixing of oxygen and carrier gases in the blender is modelled as a perfect mixing chamber
with incorporated dynamics [46]
_
FiO
2
(s)
q
g
(s)
_
=
_
1
T
g,b
s+1
e
T
d1
(q
g
)s
0
0
1
T
g,b
s+1
_
_
FiO
2,in
(s)
q
g,in
(s)
_
, (4.27)
where FiO
2,in
is the fractional oxygen input, q
g,in
is the gas-ow input command signal, T
g,b
is
the gas blender time-constant, T
d1
(q
g
) is the gas ow dependent time-delay and s is the Laplace-
transform variable. The time-delay T
d1
(q
g
) is a transport delay and is due to the distance from
gas blender to oxygenator. It applies to the oxygen fraction in the gas. The time-delay depends
on the gas-ow and the tubing system
T
d1
(q
g
) =
4
d
2
t,oxy
l
t,oxy
q
g
, (4.28)
where d
t,oxy
is the diameter and l
t,oxy
is the length of the gas tube between gas blender and
oxygenator. With higher gas-ows, the oxygen fraction time-delay drops to small values.
4.11.3 Gas Compartment
Two components are distinguished in the gas compartment, which are oxygen (O
2
) and carbon
dioxide (CO
2
). The gas fraction - partial pressure dependency (Henrys law) is
p
O
2
,g
= p
bar
FiO
2
p
CO
2
,g
= p
bar
FiCO
2
,
(4.29)
where p
O
2
,g
, p
CO
2
,g
are the gas partial pressures, p
bar
is the atmospheric pressure and FiO
2
, FiCO
2
are the mixing fractions of the respective gases. With the assumption q
g,in
= q
g,out
= q
q
and
56
4.11 Membrane Oxygenator Modelling
(4.29) the gas compartment describing equations are
V
g
FiO
2
dt
= q
g
(FiO
2,in
FiO
2,out
) D
O
2,m
(p
O
2
,g
p
O
2
,b
)
V
g
FiCO
2
dt
= q
g
(FiCO
2,in
FiCO
2,out
) D
CO
2,m
(p
CO
2
,g
p
CO
2
,pl
).
(4.30)
Note that FiCO
2,in
in (4.30) is assumed to be zero with the used control strategy (i.e. no
carbon dioxide is in the mixed gas). In the gas compartment model perfect mixing conditions
(FiO
2,in
= FiO
2
and FiCO
2,in
= FiCO
2
) and no ow dierences (q
g,in
= q
g,out
= q
g
) are
assumed. Substituting Eq. (4.29) into (4.30) and rearranging results in
V
g
dp
O
2
,g,out
dt
= q
g
(p
O
2
,g,in
p
O
2
,g,out
) D
O
2,m
p
bar
(p
O
2
,g,out
p
O
2
,b
)
V
g
dp
CO
2
,g,out
dt
= q
g
(p
CO
2
,g,in
p
CO
2
,g,out
) D
CO
2,m
p
bar
(p
CO
2
,g,out
p
CO
2
,pl
).
(4.31)
4.11.4 Oxygen Compartment
Total oxygen concentration [O
2
]
b
in the blood is the sum of dissolved and haemoglobin bound
oxygen,
[O
2
]
b
=
O
2
pO
2,b
+cap
b
S(pO
2,virt
), (4.32)
where pO
2,b
is the partial oxygen pressure in the blood. The dissolved oxygen is determined
by the O
2
-solubility
O
2
. Oxygen bound to the haemoglobin depends on the one hand on the
binding capacity
cap
b
= hct [Hb]
rbc
, (4.33)
in which hct is the haematocrit and [Hb]
rbc
is the haemoglobin concentration in the red blood
cells. On the other hand, the haemoglobin bound oxygen depends on the oxygen saturation
curve S(pO
2,virt
). For the oxygen saturation curve a virtual oxygen partial pressure pO
2,virt.
is
used, as the curve itself depends on a number of other variables (refer to Chapter 2)
pO
2,virt.
= x
x
= pO
2,b
10
0.024(37T
b
)+0.4(pH
virt
7.4)+0.06 log
40
pCO
2
. (4.34)
57
4 Modelling of the System under Extracorporeal Circulation
In Eq. (4.36), T
b
is the blood temperature, and pH
virt
is the virtual pH-value (see below). With
the substitution x
x
= pO
2,virt
, the oxygen saturation curve is
S(x
x
) =
a
1
x
x
+a
2
x
2
x
+a
3
x
3
x
+x
4
x
a
4
+a
5
x
x
+a
6
x
2
x
+a
7
x
3
x
+x
4
x
, (4.35)
where a
j
, j = 1, 2..7 are numerical constants tted to experimentally observed eects [63].
Assuming perfect mixing conditions, the oxygen compartment is described by
V
b
_
O
2
+cap
b
S(x
x
)
pO
2,virt
pO
2,virt
pO
2,b
_
dpO
2,b
dt
= q
b
([O
2
]
b,in
[O
2
]
b
) +D
O
2,m
(pO
2,g
pO
2,b
), (4.36)
where [O
2
]
b
is described by (4.32) and V
b
is the oxygenator total blood volume.
4.11.5 Carbon Dioxide Compartment
Carbon dioxide in the blood is transported in dissolved, bicarbonate ion (HCO
3
) and in car-
bamate (carb) form, refer to Chapter 2. Within the carbon dioxide compartment reactions in
the plasma and the red blood cells (erythrocytes) have to be considered and are separated for
ease. In the following equations the dehydration of bicarbonate
R
HCO
3
,pl
= k
u
CO
2
pCO
2,pl
+
k
v
k
[H]
pl
[HCO
3
]
pl
R
HCO
3
,rbc
= cat
_
k
u
CO
2
pCO
2,rbc
+
k
v
k
[H]
rbc
[HCO
3
]
rbc
_
(4.37)
is taken into account, in which k, k
u
and k
v
are the carbonic acid dissociation equilibrium, the
carbon dioxide hydration reaction forward rate and the carbon dioxide hydration backward
rate constant, respectively. In Eq. (4.37) cat is the carbonic anhydrase catalysis factor and
corresponds to the catalysed reaction by the enzyme carboanhydrase, which only exists in the
red blood cells. [H] is the hydrogen ion concentration. With the denitions for the plasma and
the red blood cell (erythrocytes) volumes and ows
V
rbc
= hctV
b
, V
pl
= V
b
(1 hct), q
rbc
= q
b
hct, q
pl
= q
b
(1 hct), (4.38)
58
4.11 Membrane Oxygenator Modelling
the equations for the carbon dioxide transfer are
V
pl
d[CO
2
]
pl
dt
= V
pl
CO
2
dpCO
2,pl
dt
= q
pl
([CO
2
]
pl,in
[CO
2
]
pl
) +
D
CO
2,m
(pCO
2,g
pCO
2,pl
) + D
CO
2,rbc
(pCO
2,rbc
pCO
2,pl
) + V
pl
R
HCO
3
,pl
(4.39)
for the plasma part and
V
rbc
d[CO
2
]
rbc
dt
= V
rbc
CO
2
dpCO
2,rbc
dt
= q
rbc
([CO
2
]
rbc,in
[CO
2
]
rbc
) +
D
CO
2,m
(pCO
2,pl
pCO
2,rbc
) + V
rbc
R
HCO
3
,rbc
V
rbc
d[carb]
dt
(4.40)
for the red blood cell part. Note that Eq. (4.39) is directly coupled to the gas compartment by
the diusion term D
CO
2,m
(pCO
2,g
pCO
2,pl
). The red blood cell compartment (4.40) is then
coupled to the plasma compartment by the interconnection term D
CO
2,m
(pCO
2,pl
pCO
2,rbc
).
In Eq. (4.40) [carb] is the carbamate concentration, interconnected to the plasma compart-
ment. The bicarbonate transfer equations for blood plasma ([HCO
3
]
pl
) and red blood cells
([HCO
3
]
rbc
) are
V
pl
d[HCO
3
]
pl
dt
= q
pl
([HCO
3
]
pl,in
[HCO
3
]
pl
)
D
HCO
3,rbc
_
[HCO
3
]
pl
[HCO
3
]
rbc
r
_
V
pl
R
HCO
3
,pl
(4.41)
and
V
rbc
d[HCO
3
]
rbc
dt
= q
rbc
([HCO
3
]
rbc,in
[HCO
3
]
rbc
) +
D
HCO
3,rbc
_
[HCO
3
]
pl
[HCO
3
]
rbc
r
_
V
rbc
R
HCO
3
,rbc
.
(4.42)
In Eqs. (4.41) and (4.42) r is due to a diusion of [HCO
3
] across the membrane of the red
blood cells and some complex biochemical eects [49]
r = (0.058pH
virt
0.437) S(x
x
) 0.529pH
virt
+ 4.6. (4.43)
59
4 Modelling of the System under Extracorporeal Circulation
These eects are responsible for the virtual pH-value
pH
virt
= log (r[H]
rbc
) (4.44)
used in Eq. (4.36). The diusion capacities D
CO
2,rbc
and D
HCO
3,rbc
are estimated based on
in-vitro measurements [49]
D
CO
2,rbc
= 0.693
CO
2
rbc
V
rbc
V
pl
V
rbc
+V
pl
, D
HCO
3,rbc
= 0.693
CO
2
HCO
3
V
rbc
V
pl
V
rbc
+V
pl
. (4.45)
In Eq. (4.45)
rbc
and
HCO
3
are the HLT of RBC membrane diusion and RBC membrane
chloride shift, respectively. Within the carbamate reaction the carbamate is directly bound
in the red blood cells to the haemoglobin by the carbamino reaction
V
rbc
d[carb]
dt
= q
rbc
([carb]
in
[carb]) +
k
a
[CO
2
]
rbc
V
rbc
([Hb] [carb])
_
k
zo
S(x
x
)
k
zo
+ [H]
rbc
+
k
zr
(1 S(x
x
))
k
zr
+ [H]
rbc
_
V
rbc
k
a
[carb][H]
rbc
k
c
,
(4.46)
where k
a
, k
zo
and k
zr
are CO
2
-Hb forward reaction-time constant, oxygenated Hb amino group
ionisation- and reduced Hb amino group ionisation-constant. The exchange of hydrogen
ions, described by concentrations takes place in plasma ([H]
pl
) and red blood cells ([H]
rbc
) and
is of importance for the other reactions. The hydrogen ion concentration in the plasma
V
pl
d[H]
pl
dt
= q
pl
([H]
pl,in
[H]
pl
) V
pl
2.303
rbc
[H]
rbc
R
HCO
3
,pl
(4.47)
is inuenced by the bicarbonate dehydration R
HCO
3
,pl
in the plasma, where in the red blood
cells hydrogen ion concentration
V
rbc
d[H]
rbc
dt
= q
rbc
([H]
rbc,in
[H]
rbc
)
V
rbc
2.303
rbc
[H]
rbc
_
R
HCO
3
,rbc
+ 1.5
d[carb]
dt
0.6cap
dS(x
x
)
dt
_
(4.48)
is inuenced additionally by the carbamate
d[carb]
dt
and the oxygenation of Hb cap
dS(x
x
)
dt
. In Eqs.
(4.47) and (4.48)
pl
and
rbc
are the plasma and the red blood cells buer capacities.
60
4.11 Membrane Oxygenator Modelling
4.11.6 The Blood-Gas Analyser
In contrast to former blood-gas analysis (BGA), where blood samples had to be taken to special
BGA machines, modern blood-gas analysers are able to observe the blood-gas values pO
2
and
pCO
2
continually, at a predened sampling time of T
s,BGA
= 6 s. The blood-gas analyser used
in this study (CDI 500, Terumo, Japan) was modelled with a rst-order lag dierential equation
in the Laplace-domain
_
p
O
2
(s)
p
CO
2
(s)
_
=
_
1
T
BGA
s+1
0
0
1
T
BGA
s+1
_
_
pO
2
(s)
pCO
2
(s)
_
, (4.49)
with T
BGA
= 20 s, the blood-gas analyser time constant for the partial pressure oxygen and
carbon dioxide input. The outputs p
O
2
and p
CO
2
are subject to an error due to amplitude
quantisation, because of discretisation, with quantisation interval 1 mmHg. A time-delay T
d2
(q
b
)
dependent on blood-ow and corresponding to the blood transport from arterial and venous
lines of the oxygenator to the BGA
T
d2
(q
b
) =
4
d
2
t,BGA
l
t,BGA
a
t1
(0.01q
b
+a
t0
)
, (4.50)
was introduced with the tubing length l
t,BGA
and the tubing diameter d
t,BGA
, where a
t0
is a
time-delay oset to prevent very large times-delays at zero ows. a
t1
is the conversion factor
from l/min to m
3
/s. With the time-delay (4.50) Eq. (4.49) becomes
_
pO
2,out
(s)
pCO
2,out
(s)
_
=
_
1
T
BGA
s+1
0
0
1
T
BGA
s+1
_
_
pO
2
(s)
pCO
2
(s)
_
e
T
d2
(q
b
)s
. (4.51)
4.11.7 Model Implementation and Generalisation
Eqs. (4.27) - (4.51) were implemented as subsystems in MATLAB/Simulink as shown in Figure
4.8. To get rid of the algebraic loop in Eq. (4.44), a rst order dierential equation with a fast
time constant was introduced in (4.44) to
dpH
virt
dt
= 10 (pH
virt
log (r[H]
rbc
)) . (4.52)
61
4 Modelling of the System under Extracorporeal Circulation
Oxygenator
D + NL
pO
2
GBD
GBD
GBTD
BGAD TD + Q
pO
2
pCO
2
FiO
2,in
Gas blender Oxygenator Blood gas analyser
q
b
q
g,in pCO
2
^
^
FiO
2
q
g
Figure 4.8: Block diagram of the oxygenator system with, GBT: Gas Blender Dynamics, GBDT:
Gas Blender Dynamics Delay-Time, Oxygenator D + NL: Dynamics and Nonlinearities, BGAD:
Blood-Gas Analysis Dynamics and TD + Q: Time-Delay + Quantisation (and discretisation).
The resulting system is a three input, two output system, with inputs composed of the oxygen
fraction FiO
2
, the gas ow q
g
and the blood-ow q
b
, outputs composed of oxygen partial
pressure pO
2
, carbon dioxide partial pressure pCO
2
.
The system of Eqs. (4.27) to (4.51) can be substituted to a state space (SS) model of the form
(without time-delay)
x = f
oxy
(x) + g
oxy
(x,u)
y = c
oxy
x,
(4.53)
where f
oxy
(x) R
15
is the new extended vector eld, g
oxy
(x,u) R
15
is the nonlinear in-
put vector eld, containing the excerpted input terms and c
oxy
R
215
is the linear output
mapping. In Eq. (4.53) the state vector is x R
15
and the output is y R
2
. f
oxy
() and
g
oxy
() are smooth nonlinear functions (f
oxy
(), g
oxy
() C
x=x
L
,u=u
L
+
f(x,u, t)
x
x=x
L
,u=u
L
B(t) =
f(x,u)
u
x=x
L
,u=u
L
+
f(x,u, t)
u
x=x
L
,u=u
L
,
(5.8)
76
5.5 Model Linearisation
where x
L
are the systems state conditions for linearisation and u
L
are the input linearisation
conditions for the input vector. The inuence of the vasoactive substances with time was ne-
glected (time dependency in the Jacobians), as the maximum change in the vascular parameters
at time t
vasc
was modelled as additional uncertainty. Eq. (5.7) therefore becomes
x = Ax +Bu
y = cx,
(5.9)
with time parameter in Eq. (5.8): t = t
vasc
. Nonlinearities in Eq. (4.24) were analysed in
a loop linearisation procedure, where the states that inuence the nonlinearities are changed
over a broad range. The vascular parameters were adapted to haemodynamics during ECC and
additional uncertainty was applied: TPR and compliance were changed up to 400 %, inertance
up to 100 %. Frequency responses of the linearised system at dierent operating points were
calculated and compared to each other. Figure 5.10 shows the result of the frequency domain
analysis, with variations induced by parameter changes. According to Figure 5.10, an upper
bound for uncertainty was dened for the development of robust control.
-30
-20
-10
0
10
20
30
amplitude [dB]
10
-2
10
-1
10
0
10
1
10
2
frequency [Hz]
Example for an amplitude response
~ 40dB/Decade
Worst case
Figure 5.10: Frequency response variations of the linearised system (4.24) with incorporated uncer-
tainty.
77
5 Simulation and Experimental Model Validation
5.6 The Oxygenator
The oxygenator model, that was adapted from the physiological lung model was rst validated
using real clinical data of a cardiopulmonary bypass surgery [47]. The model showed good
agreement in static gains, but a discrepancy with the transient response and the time-delay
was observed. The oxygenator model was therefore extended with gas mixer-, BGA-dynamics
and time-delay, see Section 4.11. Using real porcine blood, the transient response and the time-
delay were then validated in in-vitro experimental conditions (see Appendix D.3.1 for a detailed
description of the experimental methods). Step responses for the FiO
2
-fraction were applied
over the whole operating range and under various blood- and gas-ow conditions. Figure 5.11
shows an example of such a step-response. The venous conditions that were set up on the
pre-oxygenator side are pO
2,v
= 63 mmHg and pCO
2,v
= 46 mmHg. The measurement was
conducted at a blood-ow of q
b
= 4 l/min, a gas ow through the oxygenator of q
g
= 2 l/min,
a blood temperature of T
b
= 28
C, a pH = 7.35 and at a haematrocrit value of Hct = 21 %.
52
50
48
46
44
42
40
320
300
280
260
240
0 20 40 60 80 100 120 140 160 180 200
time [s]
FiO [%]
2
pO [mmHg]
2
experiment
model
Figure 5.11: Simulation and experimental step-response to an FiO
2
-change of 40 to 50 % of the
blood-gas oxygen process, at q
b
= 4 l/min, q
g
= 2 l/min.
78
5.6 The Oxygenator
Model and experiment show good transient response and time-delay agreement, but a static
gain error can be observed. This static gain error is due to an oset error in the gas valves
that occurs at lower gas-ow values (see Section 4.11.1). The error in the gas-ow leads to
a gain error of 7 % during these conditions but is higher at lower gas ows. Figure 5.12
shows such a step-response with venous conditions set to pO
2,v
= 45 mmHg and pCO
2,v
= 40
mmHg. The measurement was conducted at a blood-ow of q
b
= 2 l/min, a gas ow through
the oxygenator of q
g
= 1 l/min, a blood temperature of T
b
= 28
C, a pH = 7.35 and at a
haematrocrit value of Hct = 21 %. Figure 5.12 shows the step responses of the uncorrected
and the corrected model. For the correction of the model input the real FiO
2
-inow value was
measured during an experiment. The corrected curve in Figure 5.12 shows the system response
to the corrected oxygen fraction control input of FiO
2
= 26 to 35 %. Now the corrected model
shows a very good behaviour in transient response and steady-state conditions. Note that the
experimental results in Section 7.4 were obtained without any correction to the control input of
the gas mixers. For the experimental study it was assumed that the robustly tuned blood-gas
controllers had to handle those uncertainties.
experiment
model
model (corrected)
32
30
28
26
24
22
20
220
200
180
160
140
120
100
0 20 40 60 80 100 120 140 160 180 200
time [s]
FiO [%]
2
pO [mmHg]
2
Figure 5.12: Simulation, experimental and corrected step-response to an FiO
2
-change of 21 to 30 %
of the blood-gas oxygen process, at q
b
= 2 l/min, q
g
= 1 l/min.
79
6 Control Design
This chapter is divided into haemodynamic and blood-gas control (BGC).
For haemodynamic control three dierent control strategies were developed and compared to
each other. Controlled process values are hereby the arterial (aortic) blood-ow q
b
(Section 6.1)
or the aortic pressure p
aort
(Section 6.2.1). As a third control strategy a pressure boundary
control, superimposed on blood-ow control (Section 6.2.2), was suggested.
For the oxygen delivery and carbon dioxide removal in the blood, a simultaneous control strategy
was chosen. To control the partial pressures of oxygen (pO
2
) and carbon dioxide (pCO
2
) in the
blood, the gas oxygen inow fraction to the oxygenator FiO
2
was used for pO
2
- and the total
gas ow to the oxygenator q
g
was used for pCO
2
-control (Section 6.3).
6.1 Arterial Blood-Flow
Three dierent discrete controllers were designed and tuned for stationary and pulsatile arterial
blood-ow control. In order to guarantee stable control of the nonlinear circulatory system in
the presence of patient parameter uncertainties and disturbances, a proportional-plus-integral-
(PI) and a H
-controller were robustly tuned, using linearised models of Section 5.5, Eq.
(4.24).
Additionally, a self-tuning general predictive controller (GPC) with a parameter estimating
Kalman Filter (KF) was developed to adapt to slight nonlinearities and follow time-varying
parameters.
The principle control structure for the aortic blood-ow control is given in Figure 6.1. The
nonlinear transfer function in Figure 6.1 represents the system model of Eq. (4.24). Controlled
variable is the aortic inow q
aort
, control setpoint is r = q
ref
. The control block C
BF
refers to
one of the above-named controllers.
80
6.1 Arterial Blood-Flow
G
(s)
BF
u(nT) q (t)
aort
+
C
BF
_
q (nT)
ref
Figure 6.1: Principal control structure for aortic blood-ow control.
6.1.1 Robust PI - Blood-Flow Control
A worst-case approach of frequency responses depending on nonlinearities, time invariant pa-
rameter disturbances and multiplicative model parameter uncertainty leads to the robustly
tuned controller. The PI-controller of the form (4.8) was extended with discrete integrator
saturation (anti-windup), similar to Eq. (4.9)
f
PI,s
(u
I
) =
_
_
a
R
u
I
a
R
u
I
a
R
< u
I
< b
R
b
R
u
I
b
R
, (6.1)
where a
R
and b
R
are the integration saturation limits for the maximum and minimum rotational
speed and u
I
is the input to the integrator. The PI-controller with anti-windup saturation was
tuned with the worst case numerically estimated linearised system to have a gain margin (GM)
of GM = 31.2dB, a phase margin (PM) of PM = 57
- Blood-Flow Control
In the loop-shaping robust control synthesis, the system is augmented with the sensitivity
weightings W
1
(j) and W
3
(j) (Figure 6.3), which determine disturbance attenuation per-
formance and which are used to measure stability margins in the face of multiplicative plant
perturbations, respectively [145].
As a consequence of the small gain theorem [87], the stability can be specied in terms of the
81
6 Control Design
-2000 -1000 0 1000
-2000
-1500
-1000
-500
0
500
1000
1500
2000
2500
real axis
imag axis
-40
-20
0
20
40
60
80
magnitude [dB]
10
-2
10
-1
10
0
10
1
10
2
10
3
10
4
-225
-180
-155
-90
frequency [rad/s]
phase [deg]
GM
PM
Figure 6.2: Root locus of the open-loop compensated system G
BF
(s)C(s).
control system with the frequency-dependent weights, via the inequalities
|S(j)W
1
(j)| < 1
|T(j)W
3
(j)| < 1
(6.2)
with the corresponding sensitivity functions
S(j) =
1
G
BF
(j)C
BF
(j) + 1
T(j) =
G
BF
(j)C
BF
(j)
G
BF
(j)C
BF
(j) + 1
(6.3)
In Eq. (6.3) S(j) determines the disturbance attenuation. T(j) is used to measure multi-
plicative plant uncertainty [53, 87]. With the sensitivity functions (6.3), the closed-loop transfer
function matrix for the augmented plant, Figure 6.3, leads to the mixed-sensitivity weighted
cost function
G
Y
1,3
U
1
=
_
W
1
(j)S(j)
W
3
(j)T(j)
_
. (6.4)
82
6.1 Arterial Blood-Flow
U (jw)
1
Y (jw)
1
Y (jw)
3
C (jw)
BF
G (jw)
BF
E(jw) Y(jw)
U (jw)
2
W (jw)
3
W (jw)
1
Figure 6.3: Augmented system for robust control.
This in turn leads to the H
= sup
R
|G
Y
1,3
U
1
(j)| < 1. (6.5)
A 7
th
order H
2
= E{( )
2
}.
(6.7)
The polynomials for (6.6) are dened as
A(q
1
) = 1 + a
1
q
1
+a
2
q
2
+... +a
n
a
q
n
a
B(q
1
) = b
0
+b
1
q
1
+b
2
q
2
+... +b
n
b
q
n
b
,
C(q
1
) = 1 + c
1
q
1
+c
2
q
2
+... +c
n
c
q
n
c
(6.8)
85
6 Control Design
with n
a
n
b
, b
0
= 0, m > 1 and q
1
the backward shift operator [93]. Eq. (6.6) is rearranged
with the parameter vector
= [a
1
a
2
... a
n
a
; b
0
b
1
... b
n
b
]
T
(6.9)
and the observation vector
(k) = [y(k 1) y(k 2) ... y(k n
a
); u(k m) u(k m1) ... u(k mn
b
)]. (6.10)
Substituting Eqs. (6.9) and (6.10) into (6.6) gives
y(k) = (k) +(k)
(k) = C(q
1
)(k).
(6.11)
In further considerations, the noise colouring polynomial C(q
1
) is assumed to be one. Therefore
Eq. (6.11) simplies to
y(k) = (k) +(k). (6.12)
The parameter estimating KF equations for (6.12) were obtained from the well-known KF
observer equations by making the transition for state vector x and output mapping c ,
[59, 60, 72]. Recursive KF equations are separated in a prediction and a correction phase.
Prediction:
(k|k 1) =
(k 1|k 1)
(k|k 1) = (k 1|k 1) +R
w
(6.13)
Correction:
(k|k) =
(k|k 1) + (k)[y(k) (k)
(k|k 1)]
(k) = (k|k 1)
T
(k)[r
v
+(k)(k|k 1)
T
]
1
,
(k|k) = [I (k)(k)](k|k 1)
(6.14)
where
(k|k 1) is the estimated parameter vector, r
v
is the noise variance, R
w
is the process
noise covariance matrix, is the covariance matrix, is the Kalman gain vector, and I is
the identity matrix (all of appropriate dimensions). The notation
(k|k 1) indicates the
parameter estimate at time k, based on information of and up to k 1. Eqs. (6.13) and (6.14)
are implemented for online estimation of the parameters of a second order system.
86
6.2 Arterial Pressure Control
General Predictive Control
The GPC controller is designed and tuned using the general minimum variance cost function
law with integral control action u(k) = u(k) u(k 1) [26].
J
GPC
= E
min
_
h
p
j=k
c
[y(k +j) r(k +j)]
2
+
h
c
j=1
[
GPC
u(k +j 1)]
2
_
(6.15)
with control delay k
c
, prediction and control horizon h
p
and h
c
and
GPC
is the control input
cost weighting. There are two terms in the cost function (6.15) that are to be minimised for the
controller. The rst term is future deviations of the system output to the control set point and
the second term is the variance of future incremental control action. The incremental control
cost can thereby be tuned by the weighting
GPC
. Minimising the GPC cost function (6.15)
leads to the GPC controller [26], based on the ARMAX model (6.12). The GPC controller is
updated with estimated parameters at each sampling instance and is therefore self-tuning.
For simulation and experimental implementation, the sampling time T
s2
= 100 ms of the system
was chosen to t the time-delay. For parameter estimation, second order dynamics n
a
= 2,
n
b
= 1 were assumed in the KF, which gave the best results in control experiments. Since
a linear prediction series in a nonlinear system may lead to errors, the prediction horizon h
c
was chosen to h
c
= 1. The time series of future control reference values was assumed to be
unknown, therefore h
p
= 1. Another reason for the short prediction horizon are the system
inherent nonlinear characteristics, to which the self-tuning routine has to adapt (to prevent
system prediction errors due to nonlinearities, not predicted in the recursive prediction).
6.2 Arterial Pressure Control
6.2.1 Total Arterial Pressure Control
Similar to arterial blood-ow control, the arterial pressure was fed back and compared to the
pressure setpoint to form the error as input for control, see Figure 6.6. With regard to the
model, this means that the aortic pressure of the rst vascular compartment is used. Eq.
(4.24) was only changed in terms of the output mapping vector c.
Note that during CPB a direct aortic pressure measurement is in most cases not available.
87
6 Control Design
Pressure is measured in certain arteries and therefore subject to error, depending on the vascular
system. In Section 4.3 the ow pressure curve of the aortic cannula was shown to have a
quadratic behaviour (Figure 5.4). Therefore, the measured pressure has to be corrected by
the cannula curve, corresponding to the tip diameter of the cannula. According to (4.10), the
pressure drop over the cannula p
C
is dependent on blood-ow
p
C
= q
b
R
C
(q
B
). (6.16)
With this pressure drop over the cannula, the measured pre-cannula pressure was then corrected
to the aortic pressure
p
aort
= p
art
p
C
, (6.17)
see pressure correction (PC) block (Figure 6.6). For the tuning of a PI-controller, the model
(4.24) with aortic pressure output was linearised in the same way as in Section 6.1.1. The
resulting linear TF is
G
BPC
(s) =
P
aort
(s)
U
in
(s)
, (6.18)
where in the linearisation process a constant value for vasoactive drug injection (using Propofol
as one of the most inuencing substances) is used. With the root-locus method [71, 94], a
PI-controller of the form (4.6) - (4.9) was tuned with the worst case linearised model to have a
gain margin of GM = 17.9 dB and a phase margin of PM = 66.3
. Figure 6.10
shows the root-locus plot, together with the Bode-Plot of the open-loop compensated system
(C
cBPC
(s)G
cBFC
(s)), in which G
cBFC
is the closed blood-ow control circuit.
20 30 40 50 60 70 80
-20
-15
-10
-5
0
5
10
15
20
e [ ] = f (x)
pw
p [mmHg] = x
corr
Figure 6.9: Mean arterial pressure dierence mapping to a control error.
90
6.3 Blood-Gas
-2000 -1000 0 1000 2000
-2000
-1500
-1000
-500
0
500
1000
1500
2000
real axis
imag axis
-100
-50
0
50
magnitude [dB]
10
-2
10
-1
10
0
10
1
10
2
-225
-180
-135
-90
-45
frequency [rad/s]
phase [deg]
Figure 6.10: Root locus of the open-loop compensated system C(s)G
cBFC
(s).
6.3 Blood-Gas
Fortunately in physiological terms the transport of blood-gases in the human body is by
itself a highly nonlinear process. This nonlinear behaviour together with other physiological
control mechanisms prevents the body from under- or over oxygenation and keeps the carbon
dioxide content and the aliated pH-values within normal ranges in physiological conditions.
Unfortunately in control engineering terms the articial blood-gas circuit includes a number
of properties, complicating the development of control algorithms, such as
a) nonlinearities, which are mainly oxygen saturation and ow dependent dynamics and
result in process gain changes of 700 % and above,
b) uncertainties, for example the haemoglobin content of the blood, which depends on blood
priming and determines the binding capacities of oxygen and carbon dioxide and therefore
model gains (note that the haemoglobin content is measurable online during CPB),
c) time-varying parameters, for example the clotting of blood in the oxygenator and the
91
6 Control Design
resulting reduction of the diusion capacities (and therefore process gain) with time,
d) ow and tubing system dependent time-delays, which can vary up to and over 300 %.
Since classic linear feedback control theory performs only suboptimally on uncertain nonlinear
systems incorporating time-varying process characteristics and time-delays and therefore cannot
guarantee global stability, modern robust nonlinear control theory will be needed. A controller
for feedback control of blood-gases has to guarantee robust stability and performance under
the above-named system properties. The control approach, developed in this work, is described
below.
The Smith-predictor, as proposed in the 1950s [120], can improve the close-loop performance
of linear systems incorporating time-delays. However, special care has to be taken in Smith-
predictor design if performance and disturbance rejection in the closed-loop control circuit facing
uncertainties in the form of modelling errors is of concern. The Smith-predictor structure relates
to the internal model control (IMC) structure, which in turn oers performance advantages by
addressing the robustness problem explicitly [87]. With the help of modern geometric theory
of nonlinear control a large class of nonlinear systems can be linearised by a nonlinear state
feedback control law [44, 55, 56, 110]. In this way, the Smith-predictor structure can be extended
to this class of nonlinear systems by the use of open-loop observers (similar to IMC). By adding
an external linear controller, the linearised system compensated for time-delay can be made
robust with respect to time-delay errors and uncertainties [66].
The structure for blood-gas control can be divided into oxygen partial pressure pO
2
- and carbon
dioxide partial pressure pCO
2
-control. Figure 6.11 shows the structure of the suggested O
2
-
controller with Smith-predictor and open-loop state observer, linearisation by state feedback
() and external linear controller. The gure shows also the process input values gas ow
q
g
, blood-ow q
b
and oxygen fraction FiO
2
. Note that the process gain of the linearised plant
still varies with blood-ow q
b
, which can be seen as a predictable disturbance. The external
O
2
-controller is therefore gain-scheduled depending on blood-ow q
b
.
The process characteristics for pCO
2
-plant and controller are not so critical. An adaptive gain
controller, dependent on arterial blood-ow q
b
, as seen in Figure 6.12 was used for pCO
2
-control.
The structure of this section is as follows. First, the model of Eqs. (4.30) - (4.48) is substituted
to a suitable nonlinear ane state space model. After feedback linearisation the external linear
(pO
2
) controller and the adaptive gain (pCO
2
) controller are developed. The section ends up
with simulation results for setpoint tracking and disturbance rejection performance.
92
6.3 Blood-Gas
Y ( )
x
-
-
C
ELC
r
v1
^
y =
^
x
1 1
^
x
^
ss
^
ss
.
e
-s
^
T
t
ss
u = FiO
O 2
2
y = x
1
y
1
^
y = x e
d 1
-sT
d
Plant
v
q
g
q
b
(pO )
2
(pO )
2
^
y *
1
Figure 6.11: Oxygen partial pressure control structure with state linearisation, time-delay compen-
sation and external linear gain-scheduled controller, with () the linearisation routine, ss the
three input two output state space process (4.53), ss the state space estimation model with and
without time-delay (6.22).
C
CO
2
u =q
CO g
2
y
2
r
v2
q
b
_
(pCO )
2
(pCO )
2
FiO
2
x
ss y = x
2
^
Plant
Figure 6.12: Carbon dioxide partial control structure with external linear gain scheduled controller
and the three input two output state space process (4.53).
6.3.1 State Space Substitution
Eqs. (4.30) - (4.49) describe the transport processes in the oxygenator and the blood-gas
measurement. If the total system time-delay is neglected and the blood-ow is assumed to
be an external predictable disturbance, the model (4.53) is suited for oxygen partial pressure
linearisation and can be described as a nonlinear two-input two-output system. If furthermore
gas-ow for the rst time is assumed to be constant and pCO
2
as an output variable is neglected,
the system can be described as the following nonlinear ane state space system which is then
93
6 Control Design
suitable for input/output state linearisation (simplied from (4.53))
x = f(x) +g(x)u
y = cx,
(6.22)
where x R
13
, u R and y R. f() and g() are smooth nonlinear functions (f(), g() C
).
Note that from now on the venous conditions of the state variables are assumed to be constant
(otherwise the venous conditions could be seen as direct disturbance to the state variables of
the system). The substitutions for the state vector x are
x =
_
_
x
1
x
2
x
3
x
4
x
5
x
6
x
7
x
8
x
9
x
10
x
11
x
12
x
13
_
_
=
_
_
p
O
2
p
CO
2
pH
virt
[H]
rbc
pCO
2,rbc
pO
2,b
[HCO
3
]
rbc
pCO
2,pl
[H]
pl
[HCO
3
]
pl
p
CO
2
,g,out
[carb]
p
O
2
,g,out
_
_
, (6.23)
where values of Eq. (6.23) are now substituted into Eqs. (4.30) - (4.49). Derivatives of the
state variables are given in short hand form, denoting e.g.
dx
dt
= x.
The rst two equations for state variables x
1
and x
2
, which describe the blood-gas analyser,
are rearranged to
x
1
=
1
T
BGA
(x
6
x
1
)
x
2
=
1
T
BGA
(x
8
x
2
) .
(6.24)
94
6.3 Blood-Gas
Eqs. (4.30) - (4.48) become now, sorted in state order
x
3
=
1
T
pH
[x
3
log (x
4
((0.058x
3
0.437) S(x
x
) 0.529x
3
+ 4.6))]
x
4
=
1
V
rbc
_
q
rbc
([H]
rbc,in
x
4
) V
rbc
2.303
rbc
x
4
_
R
HCO
3
,rbc
+ 1.5 x
12
0.6cap
S(x
x
)
_
_
x
5
=
1
CO
2
V
rbc
_
q
rbc
([CO
2
]
rbc,in
CO
2
x
5
) +D
CO
2,m
(x
8
x
5
) + V
rbc
R
HCO
3
,rbc
V
rbc
x
12
_
x
6
=
q
b
([O
2
]
b,in
[O
2
]
b
) +D
O
2,m
(x
13
x
6
)
V
b
_
O
2
+cap
b
dS(x
x
)
dx
x
dx
x
x
6
_
x
7
=
1
V
rbc
_
q
rbc
([HCO
3
]
rbc,in
x
7
) + D
HCO
3,rbc
_
x
10
x
7
r
_
V
rbc
R
HCO
3
,rbc
_
x
8
=
1
V
pl
CO
2
_
q
pl
([CO
2
]
pl,in
CO
2
x
8
) +
D
CO
2,m
(x
11
x
8
) + D
CO
2,rbc
(x
5
x
8
) + V
pl
R
HCO
3
,pl
_
x
9
=
1
V
pl
_
q
pl
([H]
pl,in
x
9
) V
pl
2.303
rbc
x
9
R
HCO
3
,pl
_
x
10
=
1
V
pl
_
q
pl
([HCO
3
]
pl,in
x
10
) D
HCO
3,rbc
_
x
10
x
7
r
_
V
pl
R
HCO
3
,pl
_
x
11
=
1
V
g
_
q
g
(p
CO
2
,g,in
x
11
) +D
CO
2,m
p
bar
(x
8
x
11
)
_
x
12
=
1
V
rbc
_
q
rbc
([carb]
in
x
12
) +
k
a
CO
2
V
rbc
x
5
([Hb] x
12
)
_
k
zo
S(x
x
)
k
zo
+x
4
+
k
zr
(1 S(x
x
))
k
zr
+x
4
_
V
rbc
k
a
x
12
x
4
k
c
_
x
13
=
1
V
g
_
q
g
(p
O
2
,g,in
x
13
) + D
O
2,m
p
bar
(x
6
x
13
)
_
.
(6.25)
Note that the state equation for x
3
is rearranged from Eqs. (4.43) and (4.44). The state
equation for the Eqs. (4.43), (4.44) is introduced because of simulation problems with the
algebraic loop. The new time constant T
pH
is introduced to keep the transient responses of
state equation x
3
as minimal as possible with the goal of avoiding a sti system at the same
time.
95
6 Control Design
Substitution of the state variables for the dehydration of carbonic acid (4.37) gives
R
HCO
3
,pl
= k
u
CO
2
x
8
+
k
v
k
x
9
x
10
= f
HCO
3
,pl
(x
8
, x
9
, x
10
)
R
HCO
3
,rbc
= cat
_
k
u
CO
2
x
5
+
k
v
k
x
4
x
7
_
= f
HCO
3
,rbc
(x
4
, x
5
, x
7
).
(6.26)
The sum of solved and dissolved oxygen and the oxygen-binding curve correction (4.32) and
(4.36) nally become
[O
2
]
b
=
O
2
x
6
+cap
b
S(x
x
)
pO
2,virt.
= x
x
= x
6
10
0.024(37T
b
)+0.4(x
3
7.4)+0.06 log
40
x
5
.
(6.27)
The output vector c of the model (6.22) is a mapping of the O
2
-partial pressure, measured by
the blood-gas analyser. The second part of Eq. (6.22) can alternatively be written as
y = x
1
. (6.28)
The input of the model can be drawn from Eq. (6.25), where with substitution for the input
u = pO
2,g,in
it follows
(g(x)u)
T
=
_
0 0 0 0 0 0 0 0 0 0 0 0
uq
g
V
g
_
. (6.29)
Also note that the carbon dioxide partial pressure input, referring to the gas ow term in state
equation 11 in (6.25), is assumed to be zero, pCO
2,g,in
= 0, since carbon dioxide was not used
in this study. The nonlinear function f(x) is of course adapted according to those changes.
Note that input u to the model of (6.22) is the oxygen partial pressure in the gas, the output of
the model is the partial pressure in the blood and the blood-ow and the gas-ow are modelled
disturbances and are accounted in the state linearisation.
6.3.2 Linearisation by State Feedback
The linearisation control law by state feedback is [110]
u =
1
L
g
L
1
f
x
1
_
L
f
x
1
+
_
, (6.30)
96
6.3 Blood-Gas
where L
f
x
1
and L
g
L
f
x
1
are the Lie-dierentials dened by
L
f
x
1
=
x
1
x
f(x)
L
g
L
f
x
1
=
x
1
x
g(x).
(6.31)
The linearisation procedure can be described as to dierentiate the system output y until the
system input u appears in the term L
g
L
1
f
x
1
and is bounded away from zero for all x U,
where U is an open subset of R
13
, containing the equilibrium point of the undriven system x
0
.
The linearised system is then the
th
order linear system from the new substituted input to
the output y.
d
y
dt
= (6.32)
(with the strict relative degree , [110]).
If the system of (6.22) has stable zero dynamics and (6.22) has the strict relative degree , the
state feedback law [19, 110]
u =
1
L
g
L
1
f
x
1
_
j=0
j
L
j
f
x
1
+
_
(6.33)
with the polynomial
s
+
1
s
1
+ +
1
s +
0
,
= 1 (6.34)
which makes the system of (6.22) exponentially stable. The polynomial (6.34) is assumed to
be Hurwitz, i.e. has real positive coecients and roots which are either negative or pairwise
conjugate with negative real parts. The (locally) exponential stable linear system yields
j=0
j
y
(j)
= . (6.35)
Note that the coecients
j
, j = 0 . . . cannot be chosen arbitrarily, as can the time constant
of the asymptotically linearised system. Special care has to be taken of input saturation and
sampling-time.
For the linearisation law for asymptotic stabilisation (6.35), the Lie-dierentials L
g
L
1
f
x
1
and
L
j
f
x
1
up to j = are needed. This procedure with respect to the blood-gas system is described
97
6 Control Design
below. Dierentiating the system output of equation
x = f(x) +g(x)u
y = x
1
,
(6.36)
with respect to time yields
dy
dt
= y =
x
1
x
f(x) +
x
1
x
g(x)u, (6.37)
where
x
1
x
= [1 0 0 0 0 0 0 0 0 0 0 0 0] . (6.38)
Expansion of Eq. (6.37) leads to
y =
1
T
BGA
(x
6
x
1
) . (6.39)
The second term in Eq. (6.37) corresponding to the input is zero. Therefore, further dieren-
tiating of the output y gives
d y
dt
= y =
L
f
x
1
x
f(x) +
L
f
x
1
x
g(x)u
= L
2
f
x
1
+L
g
L
f
x
1
u,
(6.40)
where
L
f
x
1
x
=
1
T
BGA
(x
6
x
1
)
x
=
_
1
T
BGA
0 0 0 0
1
T
BGA
0 0 0 0 0 0 0
_
.
(6.41)
Incorporating (6.41) into (6.24)-(6.25) results in
L
2
f
x
1
=
_
1
T
BGA
0 0 0 0
1
T
BGA
0 0 0 0 0 0 0
_
f(x)
=
1
T
BGA
_
_
1
T
BGA
(x
1
x
6
) +
q
b
([O
2
]
b,in
[O
2
]
b
) + D
O
2,m
(x
13
x
6
)
V
b
_
O
2
+cap
b
dS(x
x
)
dx
x
dx
x
x
6
_
_
_
.
(6.42)
98
6.3 Blood-Gas
Since the second term of Eq. (6.40) L
f
L
g
x
1
is zero, Eqs. (6.40) and (6.42) combine to
y =
1
T
BGA
_
_
1
T
BGA
(x
1
x
6
) +
q
b
([O
2
]
b,in
[O
2
]
b
) + D
O
2,m
(x
13
x
6
)
V
b
_
O
2
+cap
b
dS(x
x
)
dx
x
dx
x
x
6
_
_
_
. (6.43)
For a third dierentiation of the output y and the partial derivative
L
2
f
x
1
x
, a few prerequisites
have to be made:
For the dierentiation of the oxygen saturation curve (4.35), numerator and denominator are
S
n
(x
x
) = a
1
x
x
+a
2
x
2
x
+a
3
x
3
x
+x
4
x
S
d
(x
x
) = a
4
+a
5
x
x
+a
6
x
2
x
+a
7
x
3
x
+x
4
x
.
(6.44)
Dierentiation with respect to x
x
gives
dS
n
(x
x
)
dx
x
= a
1
+ 2a
2
x
x
+ 3a
3
x
2
x
+ 4x
3
x
d
2
S
n
(x
x
)
dx
2
x
= 2a
2
+ 6a
3
x
x
+ 12x
2
x
,
(6.45)
and
dS
d
(x
x
)
dx
x
= a
5
+ 2a
6
x
x
+ 3a
7
x
2
x
+ 4x
3
x
d
2
S
d
(x
x
)
dx
2
x
= 2a
6
+ 6a
7
x
x
+ 12x
2
x
.
(6.46)
Further needed is the squared denominator polynomial
S
2
d
(x
x
) =
_
a
4
+a
5
x
x
+a
6
x
2
x
+a
7
x
3
x
+x
4
x
_ _
a
4
+a
5
x
x
+a
6
x
2
x
+a
7
x
3
x
+x
4
x
_
= a
4
+a
4
a
5
x
x
+a
4
a
6
x
2
x
+a
4
a
7
x
3
x
+a
4
x
4
x
+a
4
a
5
x
x
+a
2
5
x
2
x
+a
5
a
6
x
3
x
+a
5
a
7
x
4
x
+a
5
x
5
x
+a
4
a
6
x
2
x
+a
5
a
6
x
3
x
+a
2
6
x
4
x
+a
6
a
7
x
5
x
+a
6
x
6
x
+a
4
a
7
x
3
x
+a
5
a
7
x
4
x
+a
6
a
7
x
5
x
+a
2
7
x
6
x
+a
7
x
7
x
+a
4
x
4
x
+a
5
x
5
x
+a
6
x
6
x
+a
7
x
7
x
+x
8
x
= a
4
+ 2a
4
a
5
x
x
+ (2a
4
a
6
+a
2
5
)x
2
x
+ 2(a
4
a
7
+a
5
a
6
)x
3
x
+ (2a
4
+ 2a
5
a
7
+a
2
6
)x
4
x
+ 2(a
5
+a
6
a
7
)x
5
x
+ (2a
6
+a
2
7
)x
6
x
+ 2a
7
x
7
x
+x
8
x
(6.47)
99
6 Control Design
and its dierentiation
dS
2
d
(x
x
)
dx
x
= 2a
4
a
5
+ 2(2a
4
a
6
+a
2
5
)x
x
+ 6(a
4
a
7
+a
5
a
6
)x
2
x
+ 4(2a
4
+ 2a
5
a
7
+a
2
6
)x
3
x
+ 10(a
5
+a
6
a
7
)x
4
x
+ 6(2a
6
+a
2
7
)x
5
x
+ 14a
7
x
6
x
+ 8x
7
x
.
(6.48)
From Eqs. (6.44)-(6.48) and Eq. (4.35) result the dierentiation for Eq. (4.35) to
dS(x
x
)
dx
x
=
S
n
(x
x
)S
d
(x
x
) S
n
(x
x
)S
d
(x
x
)
S
2
d
(x
x
)
, (6.49)
where S
(x
x
) =
dS(x
x
)
dx
x
. Dierentiating (6.49) a second time, results in
d
2
S(x
x
)
dx
2
x
= [S
n
(x
x
)S
d
(x
x
) + S
n
(x
x
)S
d
(x
x
) S
n
(x
x
)S
d
(x
x
) S
n
(x
x
)S
d
(x
x
)] S
2
d
(x
x
)
[S
n
(x
x
)S
d
(x
x
) S
n
(x
x
)S
d
(x
x
)]
dS
2
d
(x
x
)
dx
x
1
S
4
d
(x
x
)
=
[S
n
(x
x
)S
d
(x
x
) S
n
(x
x
)S
d
(x
x
)] S
2
d
(x
x
) [S
n
(x
x
)S
d
(x
x
) S
n
(x
x
)S
d
(x
x
)]
dS
2
d
(x
x
)
dx
x
S
4
d
(x
x
)
.
(6.50)
Further prerequisites for the partial dierential
L
2
f
x
1
x
are the inner derivatives of the oxygen
saturation curve (4.35), which apply to the pO
2
-correction Eq. (4.36)
x
x
x
3
= 0.4 ln(10)x
6
10
0.024(37T
b
)+0.4(pH
virt
7.4)+0.06 log
40
pCO
2
x
x
x
5
= 0.06
x
6
x
5
10
0.024(37T
b
)+0.4(pH
virt
7.4)+0.06 log
40
pCO
2
x
x
x
6
= 10
0.024(37T
b
)+0.4(pH
virt
7.4)+0.06 log
40
pCO
2
,
(6.51)
and
x
3
_
x
x
x
6
_
= 0.4 ln(10)10
0.024(37T
b
)+0.4(pH
virt
7.4)+0.06 log
40
pCO
2
x
5
_
x
x
x
6
_
= 0.06
1
x
5
10
0.024(37T
b
)+0.4(pH
virt
7.4)+0.06 log
40
pCO
2
.
(6.52)
100
6.3 Blood-Gas
The derivation of the oxygenation saturation curve derivative with respect to the state variables
x
3
, x
5
and x
6
is for j = 3, 5, 6
x
j
_
S(x
x
)
x
x
_
=
S
2
(x
x
)
x
2
x
x
x
x
j
(6.53)
Using Eqs. (6.44)-(6.53), the dierent terms for the partial derivative
L
2
f
x
1
x
are
L
2
f
x
1
x
2
=
L
2
f
x
1
x
4
=
L
2
f
x
1
x
7
=
L
2
f
x
1
x
8
=
L
2
f
x
1
x
9
=
L
2
f
x
1
x
10
=
L
2
f
x
1
x
11
=
L
2
f
x
1
x
12
= 0, (6.54)
L
2
f
x
1
x
3
=
x
3
f
z
..
q
b
([O
2
]
b,in
[O
2
]
b
) + D
O
2,m
(x
13
x
6
)
T
BGA
V
b
_
O
2
+cap
b
dS(x
x
)
dx
x
dx
x
dx
6
_
. .
f
n
=
x
3
f
z
f
n
=
q
b
cap
S(x
x
)
x
x
x
x
x
3
f
n
T
BGA
V
b
cap
x
3
_
S(x
x
)
x
x
_
x
3
_
x
x
x
6
_
f
z
f
2
n
L
2
f
x
1
x
5
=
x
5
f
z
..
q
b
([O
2
]
b,in
[O
2
]
b
) + D
O
2,m
(x
13
x
6
)
T
BGA
V
b
_
O
2
+cap
b
dS(x
x
)
dx
x
dx
x
dx
6
_
. .
f
n
=
x
5
f
z
f
n
=
q
b
cap
S(x
x
)
x
x
x
x
x
5
f
n
T
BGA
V
b
cap
x
5
_
S(x
x
)
x
x
_
x
5
_
x
x
x
6
_
f
z
f
2
n
L
2
f
x
1
x
6
=
1
T
2
BGA
+
x
6
f
z
..
q
b
([O
2
]
b,in
[O
2
]
b
) + D
O
2,m
(x
13
x
6
)
T
BGA
V
b
_
O
2
+cap
b
dS(x
x
)
dx
x
dx
x
dx
6
_
. .
f
n
=
1
T
2
BGA
+
x
6
f
z
f
n
=
1
T
2
BGA
+
_ _
q
b
_
cap
S(x
x
)
x
x
x
x
x
6
+
O
2
_
D
O
2,m
_
f
n
T
BGA
V
b
cap
x
6
_
S(x
x
)
x
x
_
x
x
x
6
f
z
_
1
f
2
n
,
(6.55)
101
6 Control Design
and
L
2
f
x
1
x
1
=
1
T
2
BGA
L
2
f
x
1
x
13
=
D
O
2,m
V
b
_
O
2
+cap
b
dS(x
x
)
dx
x
dx
x
x
6
_.
(6.56)
From equation (6.54)-(6.56) follows
L
2
f
x
1
x
=
_
L
2
f
x
1
x
1
0
L
2
f
x
1
x
3
0
L
2
f
x
1
x
5
L
2
f
x
1
x
6
0 0 0 0 0 0
L
2
f
x
1
x
13
_
(6.57)
and
L
2
f
x
1
x
f(x) = L
3
f
x
1
. (6.58)
The Lie-derivative with respect to the input is
L
2
f
x
1
x
g(x) = L
g
L
2
f
x
1
=
D
O
2,m
q
g
V
g
V
b
_
O
2
+cap
b
dS(x
x
)
dx
x
dx
x
x
6
_, (6.59)
where during linearising control a gas ow of q
g
0 has to be prevented to guarantee the relative
degree = 3 (see Eq. (6.30)). This design issue is of special importance when designing the
pCO
2
-controller, which uses the gas ow q
g
as the control input. With
d
3
y
dt
3
=
L
2
f
x
1
x
f(x) +
L
2
f
x
1
x
g(x)u,
= L
3
f
x
1
+L
g
L
2
f
x
1
u,
(6.60)
follows the linearising feedback law to
u =
1
L
g
L
2
f
x
1
_
j=0
j
L
j
f
x
1
+
_
, (6.61)
with
L
g
L
2
f
x
1
= 0 x U. (6.62)
The parameters
0
. . .
3
were adapted to the input constraints and sampling time T
s,BGA
=
6 s of system (6.36) such that the closed-loop asymptotic linearised systems time constant is
102
6.3 Blood-Gas
placed at the triple pole p
1,2,3
= 0.25 (see Appendix B). Figure 6.13 shows the block diagram
of the linearisation loop, applied to the nonlinear O
2
-plant with y
1
= y as the output.
The linearisation model was implemented together with the nonlinear state-space system in
MATLAB/Simulink and showed asymptotic stabilisation over the whole operating range. However,
for practical implementation the problem of unavailability of the system states remains. There
are two ways to solve this problem guaranteeing robustness of the external linear control (which
is to be applied later).
1. The implementation of a robust discrete (exponential) observer.
2. The implementation of a robust internal model control (IMC) like structure, with the
estimation model as an open-loop state observer.
Since the rst approach means additional and unpredictable complexity and calculation cost,
the second approach was chosen for an application. By designing the external linear controller
(ELC) with regard to model and predicted delay-time uncertainty, robustness can be achieved.
6.3.3 Robust External Linear pO
2
-Controller Design
Following the Smith-predictor structure [120], a Smith-like predictor with linearisation feedback
for asymptotic stabilisation (Section 6.3.2) was combined with the prediction feedback
y
1
= y
1
+ (y
1
y
d
), (6.63)
to form the nonlinear control-loop for the real plant output y
1
=
pO
2
, see Figure 6.11. In Eq.
(6.63) y
1
is the predicted oxygen partial pressure system output without any time-delay, and
y
d
is the predicted time-delayed oxygen partial pressure output. An external linear controller
C
ELC
was gain-scheduled in dependance of the blood-ow through the oxygenator. This is
because of the changing linear system gain in dependance of the blood-ow q
b
.
Y ( )
v x
x f(x) g(x)
= + u y = x
1
y =
1
u =FiO
O 2
2
^
pO
2
^
^
Figure 6.13: Linearisation loop (6.61) for the nonlinear O
2
-plant without time-delay.
103
6 Control Design
The requirements for the external linear controller are robustness and performance in the face of
uncertainties, whether in the degrading plant or the prediction and linearisation routines. In a
rst step, uncertainty was assumed for the linearised system (6.35) in the form of unstructured
multiplicative and time-delay uncertainty.
The unstructured multiplicative uncertainty was modelled with the nominal model G
lin
(s),
which is the linear transfer function, obtained by the state linearisation (see above). The
multiplicative uncertainty was modelled such that
G
p
= G
lin
(s)(1 + l
m
(s)) (6.64)
describes a member of the family of plants, with l
m
(s) the unstructured multiplicative uncer-
tainty. The family of plants is then given by
=
_
G
p
(s) :
|G
p
(s) G
lin
(s)|
|G
lin
(s)|
l
m
()
_
, (6.65)
where
l
m
() |l
m
(s)| is the unstructured multiplicative uncertainty bound [87].
Multiplicative uncertainty was assumed in terms of bound l
m1
(s) and l
m2
(s). l
m1
(s) is composed
of uncertainties of the linearisation and the state estimation process and is described by the
worst case transfer function
G(s)
G(s) =
w0
w3
s
3
+
w2
s
2
+s
w1
+
w0
, (6.66)
which is the linearised TF with static gain and high frequency gain uncertainty, where it is
assumed that all TFs of the plant family are described by (6.66). Note that the linearised TF
is of third order, because of the relative degree of the linearisation process. Parameters for Eq.
(6.66) are given in Appendix B. The uncertainty l
m1
(s) follows to
l
m1
(s) =
G(s) G
lin
(s)
G
lin
(s)
. (6.67)
The multiplicative uncertainty l
m2
(s) is the error in the time-delay, when neglecting the input
time-delay at the linearisation (which is not large) and other time-delays in the prediction. If
the total time-delay is
T
dt
= T
d
+T
(6.68)
104
6.3 Blood-Gas
and composed of total process time-delay T
d
= T
d1
+ T
d2
and time-delay uncertainty T
, the
multiplicative uncertainty can be described as
l
m2
(s) =
e
sT
dt
e
sT
d
e
sT
d
= e
sT
1.
(6.69)
From Eq. (6.69) follows for the time-delay uncertainty to
|l
m2
(s)| =
_
_
_
2 sin
T
2
T
2
T
.
(6.70)
The time-delay uncertainty was modelled with a frequency-bound approximation
l
m2
(s) for
(6.70) of the form
l
m2
(s) = k
m2
s +
m2
s +
m2
, (6.71)
with constants to be found in Appendix B. The total system multiplicative unstructured
uncertainty is then lumped as
l
m
(s) = l
m1
(s)
l
m2
(s). (6.72)
With the requirement for robust stability, Eq. (6.2) becomes
T(j)l
m
(j)
< 1, (6.73)
where in this case the complementary sensitivity function is
T(j) =
C
ELC
(j)G
lin
(j)
C
ELC
(j)G
lin
(j) + 1
, (6.74)
as dened in (6.3).
To determine sucient disturbance rejection the sensitivity function (6.3)
S(j) =
1
C
ELC
(j)G
lin
(j) + 1
, (6.75)
can used, which leads to the nominal performance condition (6.2)
S(j)w
1
(j)
< 1, (6.76)
105
6 Control Design
where w
1
(s) is the performance weighting function yet to be specied. For integral control with
relatively high corner frequency (-3dB at 0.06 rad/sec), the control performance sensitivity
weighting was chosen to
w
1
(s) = k
p1
s +
p1
s
. (6.77)
An external linear PI-controller C
ELC
(s) was tuned with the linear system of (6.35) with re-
gard to the nominal performance (6.76) and the robust stability (6.73) condition. Figure 6.14
shows the sensitivity functions with the uncertainty, the performance weighting, and the robust
stability specication. The PI-pO
2
-controller performance to a step with the system, including
system uncertainty l
m1
(s) only is shown in Figure 6.15. The step response with the system
incorporating total uncertainty l
m
(s) is also shown in this gure.
As a second approach, a H
= sup
R
|G
Y
1,3
U
1
(j)| < 1. (6.79)
By using the MATLAB/Simulink robust control toolbox, a 4
th
-order H
-controller step response in Figure 6.15 refers to the system incorporating the total
uncertainty l
m
(s). The simulation shows the superior performance of the H
-controller, com-
pared to the PI-controller. Note that in system (6.64) time-delay uncertainty was modelled to
107
6 Control Design
20
10
0
-10
-20
-30
-40
magnitude [dB]
10
-3
10
-2
10
-1
10
0
10
1
frequency
rad
s
S(j ) w
T(j ) w
w (j )
1
w
l (j )
m
w
Figure 6.16: Sensitivity functions for robust stability and performance specication including the
frequency-dependent weightings for the H
-controller.
the uncertain linearised TF (6.66) with the frequency weight approximation (6.71).
With the linearisation routine, as given in Section 6.3.2, the process is linearised for in-
put/output behaviour, but the process gain changes occur at dierent blood-ows q
b
. Since
the pO
2
-controller was tuned with high static gains in the lower operating area of q
b
, control
performance will degrade with higher ows, as there are lower system gains. This degradation
of control performance can be overcome with a gain-scheduled control, depending on arterial
ow q
b
. Therefore, the controller gain k
p
of the PI- and the H
x
-
-
C
ELC
^
y =
^
x
1 1
^
x
^
ss
^
ss
.
e
-s
^
T
d
ss
^
y = x e
d 1
-sT
d
Simulated
Plant
v
C
CO2
u = q
CO g
2
c
oxy
q
b
-
y
(pCO )
2
2
^
^
y
(pO )
1
2
Linearisation
Loop
Discrete
Continuous
r
(
pCO
)
v2
2
r
(
pO
)
v1
2
u = FiO
O 2
2
T
s, BGA
y*
T
s, BGA
Figure 6.17: Control strategy for pCO
2
- (upper part) and pO
2
-control (lower part). The estimated
state space model ss for state linearisation is also implemented for the Smith-like predictor as
ss e
sT
d
using the full process model with time-delay. ss denotes the state space model of
the full time-delayed process (4.53), including time-delay and static gain (diusion capacity)
uncertainty. C
ELC
is either the PI- or the H
-controller is about half of the PI-controller. The GPC-controller gives the worst results,
which is because of the more sluggish control. As an indicator for haemolysis, the control
error variance of the H
-controller achieved the best results accompanied by the highest pressure variation in the
aortic arch. The lumped vascular system parameters that correspond to this simulation are
C
art
2 ml/mmHg and R
TPR
20 mmHg/(l/min). Simulation test series were repeated with
simulated vasoactive substance injections and pressure disturbances and the controllers showed
stability and sucient performance.
In experiments with the hydrodynamic system simulator, the pulsatile perfusion was validated.
Figure 7.5 shows an example of a pulsatile perfusion with the three BFCs. The values for
the pulsatile setpoint are HR = 40 BPM, PI
r
= 4 and MF = 3 l/min. In the lower part
of Figure 7.5, the pressure variation measured in the aortic arch of the MOCK is given. The
MOCK parameters that correspond to this experiment are C
art
2 ml/mmHg and R
TPR
20
mmHg/(l/min). In accordance with the measurements, the pressure variation generated by the
GPC-controller shows the worst results and is outmatched by the PI- and the H
-controller.
The control response of the H
-controller shows slightly more pressure variation in the corresponding pressure curve.
0 0.5 1 1.5 2 2.5 3 4
time [s]
3.5
5.5
5
4.5
4
3.5
3
2.5
2
1.5
80
75
70
65
60
55
pressure [mmHg]
flow [l/min]
setpoint
system output GPC
system output PI
system output H
inf
pressure output GPC
pressure output PI
pressure output H
inf
Figure 7.5: Experimental example of the closed-loop pulsatile perfusion with the three BFCs with
corresponding aortic pressure measurement. Values for the pulsatile control setpoint are HR =
40 BPM, PI
r
= 4 and MF = 3 l/min.
117
7 Simulation and In-vitro Control Study
The three BFCs were tested under dierent vascular conditions (7.1) and with various values
for the pulsatile setpoint. Stable control was achieved over these operating conditions, but the
control performance degraded with higher heart rates. The GPC-controller, in particular due
to the lower sampling time, cannot guarantee a pulsatile perfusion with higher HRs. In case of
a HR 60 BPM, the result of the GPC-controller is an almost stationary perfusion. PI- and
H
-controller can provide pulsatile perfusion at higher HRs. Figure 7.6 depicts the result of
a MOCK measurement with the pulsatile values HR = 70 bpm, PI
r
= 4 and MF = 3 l/min.
The pressure course has a good variation, but the eect of turbulent ows on the pressure can
be seen in the sharp peaks in both curves.
In Table 7.3, control performance indices are listed for two pulsatile perfusion examples with
the three blood-ow controllers. The IAE and control error variance in the measurements at
a HR of 40 BPM show a great dierence to that of a HR of 70 BPM. At 40 BPM the PI- and
the H
-controller are superior to the GPC-controller. That is because they are able to follow
the pulsatile control setpoint. At a HR of 70 BPM a larger phase shift can be observed (see
0 0.5 1 1.5 2 2.5 3 4
time [s]
3.5
setpoint
system output GPC
system output PI
system output H
inf
pressure output GPC
pressure output PI
pressure output H
inf
6
4
2
0
90
80
70
60
50
flow [l/min]
pressure [mmHg]
Figure 7.6: Experimental example of the closed-loop pulsatile perfusion with the three BFCs with
corresponding aortic pressure measurement. Values for the pulsatile control setpoint sinusoidal
half-wave are HR = 70 BPM, PI
r
= 4 and MF = 3 l/min.
118
7.2 Total Arterial Pressure Control
Figure 7.6), which leads to a large value in both performance indices. The IAE and the control
error variance are in that case no measure for the pulsatile performance of control.
Table 7.3: Performance of the three pulsatile blood-ow controllers at dierent operating points
under in-vitro conditions. Lumped MOCK parameters: TPR = 15 mmHg/(l/min), C = 1.5
ml/mmHg.
Performance index IAE [norm] 10
3
var(e
c
) 10
3
HR, [BPM] 40 70 40 70
GPC 1020 1840 1680 4080
PI 670 2240 840 8130
H
-controller
has more overshoot but the faster CRT (about 50 sec faster). The transient response of the PI-
pCO
2
-controller in both simulations is the same. On-switching of the controllers was simulated
time [s]
250
200
150
100
50
0
45
40
35
30
25
20
15
10
5
0
pO
2
[mmHg]
pCO
2
[mmHg]
0 20 40 60 80 100 120 140 160 180 200
pO setpoint
H process
Pi process
2
inf
pCO setpoint
pCO process (PI)
2
2
Figure 7.11: Simulation of blood-gas control switch-on.
125
7 Simulation and In-vitro Control Study
with dierent model conditions (see above) and at dierent ows. The results are similar to
those of Figure 7.11, with stability given in all simulations. Table 7.5 shows the performance
of the transient response after on-switching of the controllers. The simulation results shown in
Table 7.5 at a shunt value of 70 % refer to a reduction of the diusion capacities in the process
model of the oxygenator to 70 %. The values of Table 7.5 conrm the simulation results of
Figure 7.11 with the H
-controller is about 40 seconds faster in the partially occluded system, whereas the
total CRT of both controllers is a lot more sluggish in the in-vitro experiment, see Table 7.6.
Table 7.5: Performance of the PI- and the H
-controller has less overshoot (22.5 % in contrast to 45 %), whereas the control response
time of the PI-controller is about 20 seconds faster. The FiO
2
control-input of both controllers
can be seen in the lower part of Figure 7.13. Steps were applied in other operating areas over
the whole operating range (pO
2
= [100...250] mmHg) and similar results were obtained. The
simulations of reference value step changes were repeated with other blood-ow values (change
in linearised system gain) and the eect of the gain-scheduling was examined. However, results
are similar to those shown in Figure 7.13.
To simulate the result of a partially occluded oxygenator, the diusion capacities in the model
128
7.4 Blood-Gas Control
0 10 20 30 40 50 60 70 80 90
time [s]
180
170
160
159
140
130
120
40
35
30
25
20
pO setpoint
H process
PI process
2
inf
H control input
PI control input
inf
pO [mmHg]
2
FiO [%]
2
Figure 7.13: Simulation of a blood-gas control step-response from 120-160 mmHg.
were changed to 70 % (see above) and simulation was repeated. Figure 7.14 shows the result of
a step response with the occluded oxygenator. In this case, the H
-controller in simulation.
Simulation IAE [norm] OS [%] CRT [s] var(e
c
)
Normal PI 177 45 36 153.75
Normal H
-controllers. Figure 7.16 shows an example of a step-like reference value change of 120 to
160 mmHg of both controllers after four hours of continuous circulation. The H
-controller
has more overshoot, but the faster CRT. Table 7.8 lists the results of a reference value change
of pO
2,ref
= 120-160 mmHg directly after the start of an experiment and of one after four
hours. In contrast to the simulation study, the OS of the H
-pO
2
-controller in the in-vitro experi-
ment.
Experiment IAE [norm] OS [%] CRT [s] var(e
c
)
PI (start) 217.55 5 42 219.97
H
-controller in simulation.
Simulation IAE [norm] OS [%] var(e
c
)
Normal PI 1900 46.87 364.15
Normal H
-pO
2
- and PI-pCO
2
-controllers to a blood-ow disturbance in sim-
ulation with oxygenator diusion capacity of 70 %.
136
7.4 Blood-Gas Control
results to the performance of Table 7.9 were observed. During all simulations, the PI- and the
H
-controller were stable and the OS did not exceed the critical value of 62.5 %.
Figure 7.20 depicts the result of an in-vitro experiment with the H
-pO
2
-controller and PI-
pCO
2
-controller at a Hct value of 28 %. This measurement was conducted shortly after the
beginning of the experiment. The OS of the pO
2
-controller remains below the critical value
of 62.5 %, but the control response is a bit longer (as also the total time range of the experi-
ment is 1800 s in contrast to 1400 s in the simulation). The performance values IAE, OS and
variance of control error were calculated for four sample experiments directly at the onset of
simulation and after 4.5-5 hours at a temperature of 24
C and 32
-pO
2
- and PI-pCO
2
-controllers to a blood-ow disturbance in an
experiment with a Hct-value of 28 % after the beginning of circulation.
137
7 Simulation and In-vitro Control Study
Figure 7.20. As an example, the time series to a ow disturbance rejection at a temperature of
T = 24
-controller in the
in-vitro experiment.
Experiment IAE [norm] OS [%] var(e
c
)
PI (start, T = 28
(start, T = 28
(5 hours, T = 24
C) 3777.5 25 235.06
H
(3.5 hours, T = 32
C) 3118.8 25 162.62
pO setpoint
pO process
2
2
pCO setpoint
pCO process
2
2
200
150
100
50
0
50
40
30
20
10
0
6
5
4
3
2
1
q
b
[l/min]
pO [mmHg]
2
pCO [mmHg]
2
time [s]
0 500 1000 1500 2000 2500
Figure 7.21: Response of the PI-pO
2
- and PI-pCO
2
-controllers to a blood-ow disturbance in an
experiment with a Hct-value of 28 % and a temperature of T = 24
-pO
2
-control in-vitro experiment. The perfor-
mance values of this experiment are IAE = 980.53, OS = 31.25 % and var(e
c
) = 150.23 and are
similar to the performance obtained in simulation. Since the venous conditions are not expected
to change much on a blood-ow change, a result similar to Figure 7.22 is expected in in-vivo
application with a real patient, with process coupling similar to the simulation examples. Even
if slight to little changes to venous conditions occur the control is stable and shows sucient
performance (see examples with changing venous conditions above). A detailed description of
the de-oxygenation conditions and consequences can be found in Appendix D.3.2.
In Figure 7.22, it can be clearly seen, that the OS at lower blood-ows is higher than in other
blood-ow operating regions. This prolonged disturbance rejection is due to the higher time-
delay at lower blood-ows. In addition, the model-process mismatch, that derives from the
gas-valve error oset at lower gas-ows contributes to this eect.
250
200
150
100
50
0
pO [mmHg]
2
5.5
5
4.5
4
3.5
3
2.5
2
1.5
q
b
[l/min]
time [s]
0 100 200 300 400 500 600 700 800 900
pO setpoint
pO process
2
2
Figure 7.22: Response of the H
-pO
2
-controller to a blood-ow disturbance in an experiment after
four hours of continuous circulation.
140
8 Conclusion and Discussion
The goal of this work was to develop automatic control strategies and algorithms for the con-
trol of cardiopulmonary heart-lung support. Although modern integrated heart-lung machines
comprised of several safety systems are available on the market, none with a feedback control
strategy exist up to now and important vital variables are still adjusted manually during the
surgical procedure. The main reason for this is the safety requirement. A closed-loop controller
may become unstable, which could cause damage to the patient. A closed-loop control algo-
rithm has therefore to satisfy robust stability in addition to the desired performance. Robust
stability must be guaranteed in a complex biological system coupled to an articial organ, the
HLM. Diculties for control are inherent in the system and comprise nonlinearities, parameter
uncertainties for dierent patients and pathological vascular variances, changing transport time-
delays and time-varying parameters induced by various drugs or physiological bodys agents.
Although, all these diculties exist in the HLM and the human body, and some of these mech-
anisms are not well known up to the present, an automatic control is suggested to increase
the patients safety on the one hand and remove workload from the perfusionist technician on
the other hand. The well-known advantages of an automatic control, such as fast responses
to control reference changes and good disturbance rejection are expected to optimise the car-
diopulmonary bypass routine, thereby minimising the risk of tissue damage or inammatory
response.
In addition to the diculties that arise with the system that is to be controlled the right control
strategy has to be chosen. The question that arises in this context are mainly what should
be controlled and how concerning control inputs, reference variables and reference values. For
this reason, the physiological and technical background knowledge was presented (Chapter 2
and 3) and the literature on CPB control was critically examined. Among the most impor-
tant vital variables during CPB are certainly the haemodynamics and the blood-gas exchange,
bearing in mind that the HLM takes over the work of heart and lung (transport of blood and
oxygenation/carbon dioxide removal). Other control of vital variables, such as blood and body
141
8 Conclusion and Discussion
temperature or pH-value in the blood were not considered here. Blood and body temperature
are easily controlled by thermostats and in the case of pH-value anaesthetists prefer the manual
application of drugs. The haemodynamics (blood pressure and ow) and the blood-gas exchange
(oxygen and carbon dioxide partial pressures) were considered to be of main importance and
this work concentrated on the control of these vital variables. For control, a haemodynamic
and a blood-gas model were developed (Chapter 4), validated (Chapter 5) and controllers were
designed, based on these models (Chapter 6). The controllers were then validated in simulations
and in special in-vitro experiments (Chapter 7). In the in-vitro experiments, the physiological
human system was reproduced in parts by simulation circuits that simulate the physiological
behaviour of those systems. This is necessary before the application in in-vivo animal or human
test series.
The evident control actuator for haemodynamic control is the blood pump (articial heart).
Based on the blood pump as the control actuator, three dierent control strategies were de-
veloped and compared to each other for performance [85]. A feedback control is proposed and
introduced for arterial ow, arterial pressure and combined arterial ow/pressure. In contrast
to other authors [18, 115, 123], in this work only arterial pressure and ow were controlled. In
a HLM system with a buered venous bag the control of venous conditions plays a secondary
role. Furthermore, the use of vasoactive agents (vasoconstrictive or vasodilative) as an addi-
tional control actuator for arterial conditions, as for example used by [54, 61, 142] were avoided.
Vasoactive agents if applied over a longer time show severe adverse eects and are preferably
applied by anaesthetists manually.
For haemodynamic control, the system was divided in a technical and a physiological subsystem.
The technical subsystem, which comprised the haemodynamics of the HLMs components, was
implemented in MATLAB/Simulink and was validated in experiments. Particular attention was
laid to the modelling of a rotary blood pump with diagonally streamed impeller. Experimental
results show a good agreement in static and frequency domain measurements. The physiologi-
cal system comprised the vascular system described by the linearised Navier-Stokes equations.
Two models varying in complexity were implemented in MATLAB/Simulink. These two mod-
els were then compared in the frequency range to real vascular frequency responses taken from
literature. As a result, the low order model, that can adequately describe the patients vessel
dynamics, was preferred to the high order model. This low order model was then extended with
equations for vasoactive drug distribution. In a next step, technical and physiological models
were implemented and interconnected in MATLAB/Simulink. For the development of control,
142
the model was nally linearised in its worst case.
Based on this linearised model, three controllers were developed and tuned for arterial blood-
ow control. A PI- and a H
-controller was tuned in the same way as the external linear PI-controller but
using the H
-optimal approach. All of the developed blood-gas controllers were tested rst in
simulation routines [82] and then during several in-vitro experiments in alternating conditions
over the operating range [83]. During simulations and in-vitro experiments, the oxygen and the
carbon dioxide controllers were tested simultaneously with either the PI-pCO
2
- and PI-pO
2
-
or the PI-pCO
2
- and H
-pO
2
-controllers operating at a time. Stability and good performance
were observed during simulations and experiments [82, 83].
To test the controllers in in-vitro experiments, a test strategy was developed which allowed a
simulation of blood-gas exchange in a patient by a de-oxygenation and a carbonation routine
using three oxygenators in special gassing mode. By applying nitrogen and carbon dioxide
gases to the oxygenators, oxygen was successfully removed and carbon dioxide was added to
the blood to achieve an adequate oxygen consumption rate and venous gas partial pressure
conditions as in a normal cardiopulmonary bypass routine. The in-vitro experimental circuit
consisted in addition of a HLM and was lled with fresh primed porcine blood. Porcine blood
was used to simulate the highly nonlinear transport capabilities of human blood. The in-vitro
experimental circuit showed good results and normal oxygen ow rates, but also normal venous
conditions could be achieved and operated. A disadvantage, however, became obvious during
measurements. When changing the blood-ow in the circuit, the venous O
2
-condition behind
the de-oxygenators did not vary much. In contrast to that, the venous CO
2
-condition showed
great variance on varying blood-ow. This eect is due to the lower anity of carbon diox-
ide to the blood and the hence accelerated diusion process of CO
2
. In addition to that, the
oxygen and carbon dioxide saturation and binding curves given in Section 2.6 contribute to
this eect. Although the eect was damped with the implementation of an open-loop CO
2
-
de-oxygenation gas-ow control (Appendix D), variations in carbon dioxide partial pressure
still could be observed. Experiments were conducted for up to eight hours and above and so
especially long CPB routines could be simulated.
Two dierent tests were invoked on the blood-gas controllers during simulations and in-vitro
experiments. On the one hand step responses, in which after a step-like control reference change
the system response was recorded, were applied over the whole operating range. On the other
hand, the arterial blood-ow was changed and the control rejection to this disturbance was
recorded. Arterial blood-ow changes occur at the beginning (onset) of a CPB, in the main-
146
tenance stage (if necessary) and during the weaning of the patient from the machine. These
blood-ow disturbances were conducted with 1 l/min ow steps over a large ow range. Al-
though the simultaneous switch-on of the pO
2
- and pCO
2
-controllers without model and control
initialised states showed a slow response, inferior to manual control, good results were achieved
in step response tests. In step response tests the blood-gas controllers showed a fast control
response compared to manual control and an acceptable overshoot. The control behaviour of
the H
-pO
2
-controller was only slightly better than that of the PI-pO
2
-controller and the cou-
pling of the pO
2
- to the pCO
2
-controller and vice versa was minimal. The H
-pO
2
-controller
is only slightly better because of the similarly robust control approach (multiplicative model
uncertainty) and the appropriate and well-tuned PI-pO
2
-controller. During blood-ow distur-
bance test series, fast disturbance rejection was observed in simulation and in-vitro tests. The
H
-pO
2
-controller in these tests was also slightly superior to the PI-pO
2
-controller, as it showed
a higher degree towards stability and better performance. Compared to manual control and
to literature ([4] and others) the simultaneous control of pO
2
and pCO
2
showed a fast control
reference tracking and a very good disturbance rejection. The control remained stable even un-
der the more conservative in-vitro test conditions. The articially induced strong variations of
venous carbon dioxide pressure led to a stronger control action of the pCO
2
-controller, therefore
increasing the coupling eect to the pO
2
-controller. Simulation and in-vitro experiments were
conducted under dierent conditions. The in-vitro experimental tests for example were con-
ducted at varying temperatures, haematocrit values and over a long time. Although a control
degradation was observed in certain operating areas, the control remained stable and reliable.
During all disturbance rejection tests the predened operating area of pO
2
= 100-200 mmHg
and pCO
2
= 35-45 mmHg for gas partial pressures was not violated and therefore a sucient
oxygen supply under physiological pH-values could be guaranteed.
By attaching importance to the haemodynamics and the blood-gases during a cardiopulmonary
bypass routine, an automatic control strategy could be developed that is suggested to increase
the patients safety and remove workload from the perfusion technician. If both control ap-
proaches, haemodynamic and blood-gas are used simultaneously in a cardiopulmonary bypass
routine, the main life support function of the HLM is automated as follows: By guaranteeing a
sucient oxygen partial pressure in the blood, the oxygen ow to the tissues is guaranteed at
a certain and automatically controlled blood-ow. In addition, the carbon dioxide pressure in
the blood is controlled to keep the pH-value on a secure level avoiding body acidosis or alka-
losis. In order to get good control performance for reference changes and disturbance rejection
147
8 Conclusion and Discussion
in a complex biological system coupled to an articial organ, extensive system knowledge and
the use of modern control techniques are required. For that reason, two models for haemody-
namic and for blood-gas control were successfully developed and used for control. Controllers
were successfully developed and tuned for haemodynamic and blood-gas processes with regard
to nonlinearities, parameter variations/uncertainties and varying time-delays. The controllers
showed stability and good performance in simulations and in two specically designed mockup
in-vitro experiments.
Although the feasibility and performance of the developed control strategies and algorithms
were shown, numerous possibilities for future research arose during this work, of which a few
will be given here.
The controllers could be extended by several possible safety mechanisms, for example: the
automatic detection of a pressure or ow rise on occlusion of the arterial line and controller
reaction; the detection and response to a sensor failure.
Considering the in-vitro experiments, the haemodynamic vascular system simulator could be
extended to simulate other vessel compartments or branches. For the blood-gas exchange sim-
ulator, a control strategy could be developed to maintain the venous carbon dioxide partial
pressure. Finally, haemodynamic and blood-gas exchange simulators could be combined and
be driven with porcine blood to end up with a heart-lung machine coupled to a simulated pa-
tient as under real conditions. In addition, this circuit could serve for teaching purposes.
The oxygen partial pressure controller uses an input/output linearisation routine with a model
running in parallel to the process. A nonlinear blood-gas state observer could be developed
for state estimation to reduce the prediction error, therefore optimising control. Furthermore,
a simplied external linear PI-pO
2
-controller could be introduced and could be compared to
the complex one. The input/output linearisation routine could be substituted by a nonlinear
feedforward compensation [46]. This simplied controller could comprise a gain scheduling,
time-delay compensation and an either static or self-tuning/neural network based feedforward
compensation.
Finally, the control strategies have to be validated under real cardiopulmonary bypass con-
ditions, rst in animal experiments and at last in a normal cardiovascular surgery. During
these validation series, numerous medical research issues could evolve, dealing for example with
haemolysis reduction, pulsatile vs. stationary perfusion, vascular and blood-gas modelling or
the eect of automatic control on increasing safety and decreasing the mortality rate.
148
For further validation of automatic control in in-vivo experiments, the following strategy is
recommended:
- The implementation of the simultaneous control of the haemodynamics and the blood-gas
pressures in an animal experiment or in a real cardiopulmonary bypass routine.
- The use of the PI-arterial blood-ow controller with the extended PI-pressure boundary
controller, the PI-pO
2
- and the PI-pCO
2
-controller (with time-delay compensation, state
linearisation and gain-scheduling, as described above) in their real-time control environ-
ments, as in the in-vitro experiments.
- On-switching of control at the onset of CPB until the end of the weaning phase of the
patient from the machine.
- A comparison of the automatic control of the HLM to manual control, as achieved by
specialised perfusionist sta.
In order to compare the performance of an automatic to a manual control, not only the usual
control performance indices should be used. The introduction of appropriate variables may
help to determine the qualitative performance in physiological terms. Appropriate physio-
logical variables hereby may be [27, 124]: The rate of spontaneous cardiac conversion, in-
otropic drug use, urine output (hepatic, pancreatic and renal ow), skin temperature, platelet
count, brinogen concentration, plasma-free haemoglobin level, inammatory response, tissue
metabolism/toxicity and oxygen consumption. The use of indices for control performance and
physiological quality may then contribute towards the automatic control of the HLM, in order
to give a denite statement on increasing patients safety and decreasing the mortality rate.
149
A Abbreviations
Abbr. Meaning Abbr. Meaning
ACT Active Clotting Time KF Kalman Filter
ARMAX AutoRegressive Moving MAF Mean Arterial Flow
Average with eXogenous input MAP Mean Arterial Pressure
BGA Blood Gas Analysis MRAC Model Reference Adaptive
BGC Blood Gas Control Control
BLDC BrushLess Direct Current NPO Nonlinear Pressure Output
BFC Blood Flow Control PI Proportional plus Integral
BPC Blood Pressure Control PM Phase Margin
BPM Beats Per Minute SISO Single Input Single Output
BS Body Surface SNP Sodium NitroPrusside
BW Body Weight SS State Space
CPB CardioPulmonary Bypass TPR Total Peripheral Resistance
CRT ControlResponse Time VS Vascular System
CV CardioVascular
ECC ExtraCorporeal Circulation
ECMO ExtraCorporeal Membrane
Oxygenation
ELC External Linear Controller
GM Gain Margin
GPC General Predictive Control
HLM Heart-Lung Machine
HLT Half Life Time
IAE Integral of Absolute Error
IMC Internal Model Control
i
B Constants
Parameters for the arterial 128-compartment model can be found in Avolio [12].
CONSTANT VALUE CONSTANT VALUE
Rotary blood pump and rotational speed control:
b
0
15.33 b
mot
2.9 kg m
2
b
1
8.32 10
3
K
mot
6.5 mNm/A
b
2
6.7 10
6
K
emf
1/1464 (V min)
1
a
mot
, b
mot
7 A K
p
50 10
6
J
mot
8.83 kg/m
2
K
i
400 s
1
b
1.05 g/cm
3
p
in
39 mmHg
R
mot
360 m V
W
4 l
L
mot
49 H T
s,cpb
5 ms
a
0 V b
6 V
Vascular parameters:
R
1
10 g s/cm
4
L
3
3.1 g/cm
4
C
1
165 s
2
cm
4
/g R
4
25 g s/cm
4
L
1
0.8 g/cm
4
C
4
5 ms
2
cm
4
/g
R
2
90 g s/cm
4
L
4
1.7 g/cm
4
C
2
1.1 ms
2
cm
4
/g R
5
50 g s/cm
4
L
2
1.1 g/cm
4
C
5
0 s
2
cm
4
/g
R
3
1300 g s/cm
4
L
5
1.1 g/cm
4
C
3
10 ms
2
cm
4
/g
Vasoactive volume distribution:
V
1T
560 ml [A. element + k
s,Propofol
-ln(0.5)/(180 s)
a. line + volume oxy] K
R,Propofol
0.97
V
2T
413 ml K
C,Propofol
1.04
V
3T
1.081 l k
s,SNP
-ln(0.5)/(240 s)
ii
V
4T
4.387 l K
R,SNP
0.975
V
5T
1.01 l K
C,SNP
1
V
RT
942.5 ml [V. return]
Oxygenator, Arterial Filter, Tubing and Catheter:
R
oxy
477.12 g/(cm
4
s) L
oxy
56.56 g/cm
4
C
oxy
10.35 s
2
cm
4
/g R
fil
107.74 g/(cm
4
s)
C
fil
1.12 s
2
cm
4
/g R
tub
107.74 g/(cm
4
s)
L
tub
66.8 g/cm
4
a
C1
4.44 g/(cm
7
s)
a
C0
16 g/(cm
4
s) L
cath
10.4 g/cm
4
Hydrodynamic Vascular System Simulator:
T
s,1
10 ms T
s,MOCK
540 ms
T
s,2
100 ms f
gm
15 Hz
Gas blender:
T
g,b
0.3 s l
t,oxy
2 m
d
t,oxy
5 mm
Oxygenator:
V
g
0.1 l k
u
120 ms
1
V
b
0.25 l k
v
89 s
1
p
bar
760 mmHg k
a
5 kM
1
s
1
D
O
2
,m
11.291 l/(mmHg s) k
zo
8.4 nM
D
CO
2
,m
414.64 l/(mmHg s) k
zr
72 nM
a
1
-8532 cat 1310
3
a
2
2121
O
2
1.35 mM/mmHg
a
3
-67.07
CO
2
3 mM/mmHg
a
4
93610
3
pl
6 mM/pH
a
5
-31.3510
3
rbc
57.7 mM/pH
a
6
2.39610
3
rbc
1 ms
a
7
67.10
HCO
3
200 ms
k 559 10
6
M T
pH
100 ms
iii
B Constants
Oxygenator venous input conditions:
[O
2
]
b
6.8 mM [HCO
3
]
pl
26.3 mM
pO
2,b
40 mmHg [HCO
3
]
rbc
18.2 mM
[CO
2
]
pl
1.38 mM r 0.69
pCO
2,pl
46 mmHg [H]
pl
42.3 mM
[CO
2
]
rbc
1.38 mM [H]
rbc
61 nM
pCO
2,rbc
46 mmHg [carb] 2.35 mM
Blood-gas analyser:
T
BGA
20 s d
t,BGA
2 mm
T
s,BGA
6 s a
t1
0.05 (min m
3
)/(l s)
A
q
1 mmHg a
t0
16.66710
6
l
t,BGA
1 m
Blood-flow control:
K
p,BFC
15010
3
T
s,PI
10 ms
K
i,BFC
8 s
1
a
R
0 V ( 0 R/min)
b
R
2 V ( 10000 R/min)
GPC
0.1
r
v
0.1 R
w
diag(1, 1, 100, 1000)
h
p
1 h
c
1
Arterial Pressure Control:
K
p,BPC
3010
3
T
s,PI
10 ms
K
i,BPC
2.41 s
1
a
R
0 V
b
R
2 V
Arterial Pressure Boundary Control:
T
LPP
318.3 ms k
g
1/80
k
qb
5 mmHg/(l/min) c
g
50 mmHg
c
qb
3 l/min T
s,PI
10 ms
K
p,PBC
5010
3
K
i,PBC
0.714 s
1
k
m
5
Blood-gas control:
3
1 k
m2
2.5
iv
2
0.75
m2
0.08
1
0.1875
m2
0.4
0
0.0156 k
p1
0.5
0
0.125
p1
0.06
w0
27/1000 T
5 s
w3
1 k
CO
2
-8
w2
9/10 T
CO
2
25
w1
27/100 T
t,CO
2
5-10 s
w0
27/1000 a
O
2
0
b
O
2
a
CO
2
0.5 l/min
b
CO
2
7 l/min k
p0,O
2
0.05
k
I,O
2
0.12 s
1
k
p0,CO
2
-0.05
k
I,CO
2
0.11 s
1
v
C Notation and Symbols
Some conventions for symbols are made in this work. Scalars are represented by italic letters
(e.g k or K), vectors by lower case bold letters (e.g. v) and matrices by upper case BOLD
letters (e.g. A). Indices of vectors and matrices are italic lower case letters (e.g. x
i
, x
n
or a
ij
).
Unless otherwise stated, vectors are dened as column vectors.
Sets of numbers are represented by calligraphic letters, e.g. S. The nomenclature p(t|x) variable
p at time t, at location x. It follows a list with symbols and indices used in this work.
SYMBOL MEANING SYMBOL MEANING
General:
f
f
Force vector
G(s) Transfer function
|G(j)| Frequency-dependent magnitude
arg(G(j)) Frequency-dependent phase angle
h Height
m Mass
p Pressure
q Flow
r Radius
s Complex frequency, Laplace operator
t Time
T Time constant
v Velocity vector
V Volume
z Complex frequency z = e
sT
Angular frequency
j Complex frequency
Electro-Mechanical:
b
mot
Viscous motor damping R
mot
Winding resistance
i
mot
Motor current T
mot
Motor torque
vi
J Inertia u
emf
Back electro-magnetic
K
emf
Back EMF constant eld (EMF) voltage
K
mot
Motor torque constant u
in
Input voltage
L
mot
Motor inductance
mot
Rotational motor speed
Haemodynamics:
C Vascular compliance
c
n
Substance concentration of compartment n
C
n
Fluid compliance of compartment n
d
n
Diameter of vessel, compartment n
G
vasc
(s) Vascular transfer function
J
cl
Inertia of a cylinder
J
co
Inertia of a truncated cone
k
s
Half-life time constant of vasoactive substance
L Fluid inertance
L
n
Fluid inertance of compartment n
m
cy
Mass cylinder
m
co
Mass truncated cone
q
b
Blood-ow
q
n
Substance ow of compartment n
pCO
2
Carbon dioxide partial pressure
pO
2
Oxygen partial pressure
p
in
Pressure at pump inlet
p
out
Pressure at pump outlet
R Fluid resistance
R
C
Fluid resistance, cannula
R
cath
Fluid resistance, catheter
r
n
Radius of vessel, compartment n
R
n
Fluid resistance of compartment n
R
oxy
Fluid resistance, oxygenator
R
tub
Fluid resistance, tubing system
V
n
Substance volume of compartment n
V
nT
Total volume of compartment n
V
nU
Unstressed volume of compartment n
x
v
Longitudinal vascular axis
Z
vasc
(j) Vascular impedance
Dynamic viscosity
Kinematic viscosity
Density of blood
Shear stress
vii
C Notation and Symbols
blood-Gas:
[C] Concentration of component C
D Diusion constant
FiO
2
Oxygen fraction in the gas
Hb Haemoglobin
Hct Haematocrit value
pO
2
Oxygen blood-gas partial pressure
pCO
2
Carbon dioxide blood-gas partial pressure
p
O
2,g
Oxygen gas partial pressure
p
CO
2,g
Carbon dioxide gas partial pressure
q
g
Gas ow
R
i
Dehydration term associated to index i
S() Saturation function
Solubility
Buer capacity
i
Half-life time associated to a certain drug i
Control:
A System matrix
b, B Input vector, matrix
c, C Output vector, matrix
d, D Straight-way scalar, matrix
e, e Control error scalar, vector
G(s), G(s) Transfer function, matrix
h
c
Control horizon
h
p
Prediction horizon
I Unity matrix
k
c
Control delay
K
p
Proportional gain constant
K
i
Integral gain constant
r, r Control reference scalar, vector
r
v
Noise variance
R
w
Process noise covariance matrix
T
s
Sampling time
u, u System input scalar, vector
x System state vector
y, y System output scalar, vector
GPC
Control input cost weighting
viii
Covariance matrix
Math:
n, m, i, j Integer indices, Z
+
C Set of complex numbers
C
nm
Complex matrix of dimension n m
f(), f() Scalar or vector function
R Set of real numbers
R
n
n-dimensional vector space of real numbers
R
nm
Real matrix of dimension n m
Z
+
Set of non-negative integers
For all
INDICES MEANING INDICES MEANING
aort Aorta mot Motor
art Arterial m Membrane
b Blood n Referring to index
bar Barometric out Output
BGA Blood-Gas Analysis oxy Oxygenator
c Control, controller p Prediction
cath Catheter pl Plasma (blood)
cl Cylinder rbc Red blood cells
co Cone (erythrocytes)
emf Electro-magnetic eld ref Reference
fil Filter (arterial) s Sampling
f Force tub Tubing System
g Gas v Vascular
ht Heart vasc Vascular
in Input
ix
D Experimental Setup
D.1 Hydrodynamic Vascular System Simulator
The hydrodynamic vascular system simulation circuit (MOCK), developed at the Department of
Biomedical Engineering, Ruhr-University Bochum, Germany, consists of a ow resistance (tube
clamping), a compliance chamber and the tubing system [91]. To simulate the rheological
properties of blood, a water-glycol mixture (70 %/30 %) was used during all measurements
[106]. The structure of the ow resistance and the compliance chamber are shown in Figure
D.1. The total peripheral patients resistance (TPR) is simulated by the clamping of the
tube with a lever and a driving mechanism. The position of the lever determines the peripheral
resistance, which was also measured with the ow and the corresponding pressure drop over
the tube clamping. When stationary values are considered, the TPR of the MOCK is
R
TPR
=
p
TPR
q
aort
, (D.1)
with p
TPR
the pressure drop over the tube clamping. The total compliance of the patient was
simulated with a compliance chamber (Windkessel) with water level detection and controlled
air pressure. The hydrodynamic compliance is the change in the gas volume of the Windkessel
due to a given change in the uid pressure
C
art
=
V
p
. (D.2)
The compliance in the hydrodynamic system simulator was determined by the uid level in the
compliance chamber. The compliance was kept constant to changing haemodynamics by pres-
sure adoption to keep a constant volume. If stationary values of the compliance are considered,
Eq. (D.2) simplies to
C
art
=
V
air
p
. (D.3)
x
D.1 Hydrodynamic Vascular System Simulator
DM
L
TS
LE LD
A /A
in out
A
CC
B
CC
(a) (b)
q
b,out
q
b,in
Figure D.1: Schematic diagram of the hydrodynamic simulation circuit (MOCK) elements. (a) Fluid
resistance with driving mechanism DM, the tubing system TS and the lever L. (b) The compliance
chamber (Windkessel) with air in- and outow A
in
/A
out
, the air in the compliance chamber A
cc
,
the equivalent uid to blood in the compliance chamber B
cc
, the light emitter LE and the light
detector LD for the level detector.
Both elements, TPR and compliance of the MOCK were adjustable and computer-controlled
and showed almost linear behaviour over the whole operating range [91]. The hydrodynamic
vascular system elements were connected by a tubing system
1
(2m length, 3/8 diameter, 3/32
wall thickness), which was connected with a HLM, Figure D.2.
To simulate the total cardiopulmonary bypass case with HLM support, the human heart, which
is simulated by another blood pump, was disconnected and the HLM with components oxy-
genator
2
, arterial lter
3
(not shown in Figure D.2) and cannula
4
was connected to the MOCK.
A number of additional blood-ow
5
and pressure
6
sensors were connected to the circuit, for data
recording with the dSpace
7
real-time simulation and control environment. The rotary blood
pump
8
was connected to the automatic control system (Figure D.2) and the haemodynamic
controllers were initialised in the real-time software for the dSpace environment. The sampling
time for pressure and ow measurement and control with the dSpace control setup was T
s,1
for
the PI- and H
-controller and T
s,2
for the GPC-controller. Arterial compliance and TPR are
1
Tygon, Raumedic, Helmbrechts, Germany
2
Quadrox, Jostra, Hirrlingen, Germany
3
Anity, Medtronic, Minneapolis, USA
4
TPD 3439, Jostra, Hirrlingen, Germany
5
T110 ow meter, 9XL ow probe, Transonic Systems Inc., Ithaca, USA
6
Isotec pressure transducer, TAM-A amplier, Hugo Sachs Elektronik, Harvard Apparatus GmbH, March-
Hugstetten, Germany
7
DS1104, dSpace, Paderborn, Germany
8
DeltaStream, Medos AG, Stolberg, Germany
xi
D Experimental Setup
W
P P P
R
V H
PC
(parameter
control)
Arterial system
O C P
F
P
BP
P
dSpace control
A
F
Figure D.2: Setup of the hydrodynamic system circuit, with A: dSpace control unit, BP: blood pump,
C: cannula, F: ow sensor, H: heart (disconnected, R
H
), R: total peripheral resistance, O:
oxygenator, P: pressure transducer, V: venous system (level-controlled reservoir), W: compliance
chamber (Windkessel).
both computer-controlled by a personal computer with AD converter card
9
and serial interfaces,
at a sampling time of T
s,MOCK
. Arterial compliance and TPR could be changed in the ranges
C
art
= [0.8 . . . 2]
ml
mmHg
R
TPR
= [5 . . . 40]
mmHg
l/min
.
(D.4)
The DeltaStream blood pump was monitored by a driving console and was placed about 40 cm
below the operating table (patient/MOCK). Noisy analogue pressure signals for exact MOCK
control were ltered with a variable lter
10
at a 3dB cut-o frequency of f
gm
. A venous cannula
(not shown in Figure D.2) was implemented in the venous return path of the HLM.
After lling the HLM and the MOCK with the substitute uid, the whole system was vented.
The controllers were initialised in the real-time code in the dSpace board and haemodynamic
control was started. Over the course of the experiments, the uid resistance of the oxygenator
and the arterial lter were checked as a matter of routine to detect a partial occlusion of these
components. In the case of any problems they were replaced.
9
PCI - 6035, National Instruments, Austin, USA
10
VDF 8 - 4, Kemo, Inc., Jacksonville, USA
xii
D.2 Pulsatile Control Setpoint
D.2 Pulsatile Control Setpoint
The setpoint for pulsatile control was made out of three variables: Heart rate (HR), mean ow
(MF) and pulsatility index (PI
r
). Sinusoidal half-waves with an additionally added DC-part
were used for the blood-ow control reference signal and were determined by MF and HR
(frequency). The pulsatility index PI
r
determines the ratio of sinusoidal peak value to DC part
ow, refer to Figure D.3. PI
r
is dened as PI
r
= A/MF, where A = peak - valley (see Figure
D.3). The given values MF and PI were used to calculate the amplitude A = MF PI and
the value a
0
= MF A/PI. With these values the pulsatile reference signal is
q
b,ref
(t) =
_
Asin(
2HR
60
t) 0 < t
T
2
a
0
T
2
< t T
. (D.5)
A simple control was implemented in the program to prevent the user from entering values that
would cause unphysiological ows. Also at a given PI
r
, MF and HR, the pulsatile setpoint
was calculated to satisfy all these values with a ow curve corresponding to the predened
boundaries (q
b
= [0 6] l/min). For that reason, appropriate boundaries for the values in Eq.
(D.5) were dened.
q
b
[l/min]
time [s]
5
4
3
2
1
0
0 0.5 1 1.5 2 2.5 3
1/HR
A
Positive
sinusoidal
half wave
MF (mean flow)
A/MF=PI
r
6
a
0
Figure D.3: Pulsatile control setpoint for the arterial blood-ow made out of heart rate (HR), pul-
satility index PI
r
and mean ow MF.
xiii
D Experimental Setup
D.3 In-vitro Blood-Gas Control
For in-vitro evaluation of automatic blood-gas control, the conditions during real CPB were
simulated by use of an articial blood-gas control circuit [81]. Since the experiments could
neither be accomplished in in-vivo conditions nor with real human blood, the following two
substitutions were made:
- To simulate the highly nonlinear transport capabilities of the human blood, fresh porcine
blood was used as a substitute. Porcine blood has similar haemodynamic and blood-gas
transport properties to human blood [113].
- The blood de-oxygenation and carbonation eect of the human body was simulated by the
use of an oxygenator in de-oxygenation mode with applied nitrogen and carbon dioxide
gas. Since one de-oxygenator cannot remove enough oxygen, three de-oxygenators had to
be used and were connected in series (blood), each having its own gas supply. Nitrogen
and carbon dioxide gases were then controlled to achieve venous conditions of blood-gases.
A detailed description of the in-vitro gas control setup and the materials and methods that
were used follows below.
D.3.1 Experimental Setup
As in the normal total CPB case, the venous return line was connected to an (open) reservoir,
see Figure D.4. From there the deoxygenated carbon dioxide rich blood was pumped through
the oxygenator
11
1 and an arterial lter
12
back to the simulated patient. The patient was
simulated by means of de-oxygenators 2,3,4, which were connected in series, see Figure D.5.
Blood for the BGA
13
was collected directly from the oxygenator (pre-oxygenator = venous
BGA, post-oxygenator = arterial BGA) and was fed back to the reservoir. Pressures in the loop
were measured with pressure sensors
14
, pre- and post-blood pump
15
, behind the oxygenator 1
and between oxygenators 2,3,4, and the reservoir. Both gas blenders
16
had O
2
-, CO
2
- and
11
Quadrox, Jostra, Hirrlingen, Germany
12
Anity NT Oxygenator, Medtronic, Minneapolis, USA
13
CDI 500, Terumo, Japan
14
Isotec pressure transducer, TAM-A amplier, Hugo Sachs Elektronik, Harvard Apparatus GmbH, March-
Hugstetten, Germany
15
Deltastream, Medos AG, Stolberg, Germany
16
GFC 17, Analyt GmbH & CO KG., M ulheim, Germany
xiv
D.3 In-vitro Blood-Gas Control
O
x
y
1
T
h
e
r
m
o
s
t
a
t
P
r
e
s
s
u
r
e
F
l
o
w
F
i
l
t
e
r
G
a
s
b
l
e
n
d
e
r
1
B
G
A
R
e
s
e
r
v
o
i
r
B
l
o
o
d
,
p
r
i
m
i
n
g
G
a
s
b
l
e
n
d
e
r
2
O
x
y
s
2
,
3
,
4
(
D
e
-
O
x
y
)
O
x
y
1
G
a
s
-
m
i
x
t
u
r
e
1
G
a
s
-
m
i
x
t
u
r
e
2
D
r
u
g
s
D
a
t
a
s
a
m
p
l
i
n
g
(
X
P
C
)
B
l
o
o
d
p
u
m
p
P
r
e
s
s
u
r
e
P
r
e
s
s
u
r
e
P
r
e
s
s
u
r
e
B
l
o
o
d
s
a
m
p
l
e
H
o
s
t
P
C
(
c
o
n
t
r
o
l
)
(
a
)
Figure D.4: Setup of the blood-gas control circuit with cooling water, blood-gas, arterial and venous
line, blood pump, oxygenator, de-oxygenator, reservoir, pressure/ow sensors, thermostat, data
sampling unit (XPC Target PC) and Host PC (control). A detailed description of the de-
oxygenators (a) is given in Figure D.5.
xv
D Experimental Setup
(a)
Gas
compartment
Water in Water in Water in
Water out Water out Water out
q
b
q
b
Gas
compartment
Gas
compartment
Figure D.5: Serial connection of the three oxygenators in de-oxygenation mode.
N
2
-gas inputs with automatically controlled gas-ow valves and operating range gas-ow of 0
to 10 l/min. Gas mixtures in each gas mixer could be changed from 0-100 % for each gas.
The blood of the circuit was cooled using a thermostat
17
connected to the oxygenators with
water as cooling medium. The blood-gas control circuit elements were connected by a tubing
system
18
(2m length, 3/8 diameter, 3/32 wall thickness) and additional Luer-Lock connections
were implemented for a drug (pre-oxygenator 1) and a blood sample (post-oxygenator 1) port.
The blood-ow was measured in the arterial line with an ultrasonic ow meter
19
. A PC with
XPC Target for data collection and control operating at a sampling time of T
s,XPC
= 1s was
connected to an AD-converter card
20
and communicated with a Host-PC over the serial port.
Data sampling and control were implemented on the Host-PC, where the controllers developed
in Chapter 6 operated with sampling time of the BGA, T
s,BGA
= 6s.
D.3.2 Materials and Methods
The fresh porcine blood was collected in a pre-heparinised
21
bowl (10000 IU, international
units), directly after regular slaughtering of the pig with high-voltage shock. The collected blood
(about 4-5 l) was bottled and depending on the amount of blood an additional heparin dosage
was applied. The heparin amount was chosen to be around 5000-10000 IU/Litre, where a normal
dosage of 8000 IE/Litre was applied, see sample protocol of an experiment in Table D.1. During
all measurements, additional heparin was applied (if necessary) to keep the active clotting time
(ACT) above 400 s. The ACT was measured
22
every four hours. Before starting the experiment,
17
FP 45, JULABO Labortechnik GmbH, Seelbach, Germany
18
Tygon, Raumedic, Helmbrechts, Germany
19
T110 ow meter, 9XL ow probe, Transonic Systems Inc., Ithaca, USA
20
PCL 712PG, Advantech, USA
21
Heparin-Sodium, Ratiopharm, Ulm, Germany
22
ACT II, Medtronic, Minneapolis, USA
xvi
D.3 In-vitro Blood-Gas Control
the online BGA control circuit was primed with Ringer-Lactate priming solution
23
and reference
measurements were made with a calibration BGA device
24
. The blood was then applied to the
circuit through a cardiotomy-lter and the circuit was vented. The amount of blood given to
the circuit was calculated using the formula
Hct
Haem
=
V
b
V
s
+V
b
Hct, (D.6)
with the actual haematocrit value of blood Hct, the desired haematocrit value of the haemo-
diluted uid Hct
Haem
, the blood volume V
b
and the priming solution volume V
s
. At a given
haematocrit value and a desired one of e.g. 21 %, Equation D.6 was rearranged to give the
blood volume to
V
b
=
V
s
Hct
Hct
Haem
1
. (D.7)
Before conducting the BGA control measurements, the base excess (BE) was set to zero, by
the application of sodium hydrogen carbonate
25
. The gas ows in the de-oxygenators were then
adjusted to agree with venous conditions. This was achieved with the following procedure:
1. An arterial blood-ow rate of q
b
= 4 l/min was set by the blood pump.
2. The gas ow rate of the oxygenator 1 was set to a total gas-ow 3 l/min with an O
2
-
concentration of 21 %. At the same time, the nitrogen gas-ow rates of the three de-
oxygenators were set in order to get an oxygen consumption rate of about 110-130 ml/min
(oxygen ow
V
O
2
). This oxygen consumption value corresponds to the total oxygen con-
sumption of a typical patient during anaesthesia with applied hypothermia of 28
C [68].
Arterial O
2
-saturation in these conditions was about 99 %, venous O
2
-saturation about
72 %.
3. To get a venous CO
2
-partial pressure of about 46 mmHg, the percentage carbon dioxide
ow of the three de-oxygenators was set to a value of FiCO
2
= 15-25 %.
After the procedure, venous conditions (pO
2,v
55 mmHg, pCO
2,v
46 mmHg) were obtained,
but a problem occured when changing the blood-ow. Since a total of three de-oxygenators
is needed to de-oxygenise the venous blood, the high gas-ow dramatically changes pCO
2,v
23
Ringer-Lactat, B. Braun AG, Melsungen, Germany
24
ABL 77 Series, Radiometer Copenhagen, Copenhagen, Denmark
25
Molar 8.4 % electrolyte concentrate, Serag-Wiessner GmbH & Co. KG, Naila, Germany
xvii
D Experimental Setup
Table D.1: Sample of an in-vitro BGA-control experimental protocol.
Date: 01/02/06 Experiment No. 8
Start time (killing) 6:10 h
Time to removal of blood 1 min
Initial amount of heparin 10,000 IU
Amount of blood 5 l
Total amount of heparin 8,000 IU/l
Start experiment 8:10h
Starting values
Hct: 26 %
Priming volume: 0.5 l
Temperature (Blood): 28
C
ACT: 420 s
Barometric pressure: 760 mmHg
BGA (reference): pO
2,a
= 31 mmHg, pCO
2,a
= 54 mmHg,
pH = 7.38
End of experiment 16:00h
Length experiment 7:50h
Ending values
Hct: 14 %
Priming volume: -
Temperature (Blood): 28
C
ACT: 400 s
Barometric pressure: 760 mmHg
BGA (reference): pO
2,a
= 140 mmHg, pCO
2,a
= 27 mmHg,
pH = 7.34
at blood-ow changes. In contrast to that, pO
2,v
does not change very much. This can be
explained by the following two eects.
1. Oxygen in the blood has a much higher binding anity than carbon dioxide. This is
because of the chemical binding of oxygen to haemoglobin. A large diusion pressure
xviii
D.3 In-vitro Blood-Gas Control
dierence is therefore needed to remove the oxygen from the blood. This is not the case
with carbon dioxide, which is removed at certain lower gas-ows. The de-oxygenators
were therefore vented with carbon dioxide gas (see above).
2. When considering the oxygen and carbon dioxide binding curves as given in Figure 2.3
and 2.5, it becomes clear that the carbon dioxide pressure varies stronger to a change of
content than the oxygen partial pressure. This is due to the shape of the curve at the
venous pO
2,v
- and pCO
2,v
-values.
These eects lead to very low carbon dioxide pressures at low blood-ows and higher carbon
dioxide pressures at higher ows. In order to overcome this highly unphysiological side ef-
fect, an open-loop control was proposed which directly changes the FiCO
2
-value depending on
ow. Measurements of pCO
2,v
show an almost quadratic relationship between blood-ow and
percentage of carbon dioxide gas-ow to obtain pCO
2,v
conditions. The open-loop control was
modelled by a fourth order polynomial and implemented in the real-time process control, Figure
D.6. However, even if pCO
2,v
is controlled in this way, it breaks down for about 20-30 s when
the blood-ow is changed. This is because the change in the blood-ow is a direct disturbance
to the states of the BGA system and reacts directly on the output (with the time-constant of
the BGA device). The control of the new FiCO
2
-value corrects pCO
2,v
after these 20-30 s.
2 2.5 3 3.5 4 4.5 5
32
30
28
26
24
22
20
18
FiCO [%]
2
q [l/min]
B
Figure D.6: Blood-ow - FiCO
2
relationship to keep a constant venous carbon dioxide partial pres-
sure, approximated for real-time control by a fourth order polynomial.
xix
D Experimental Setup
Right at the beginning of the experiment, a sample of the haemodiluted blood was taken for the
calibration of the blood-gas analysis and for ACT determination. This procedure was repeated
in steady-state blood-gas control conditions during the experiment and at the end of each ex-
periment. The values of the BGA calibration measurements were compared to the values of the
online BGA. During all in-vitro experiments, the temperature gradient between cooling water
and primed blood was kept below 5
C.
The online BGA measurements for automatic control were conducted in the -stat mode of
the online BGA device (see Section 3.4.2 for details). Before application of the controller,
the MATLAB/Simulink control model was updated with temperature, Hct, etc. values. Dur-
ing experimental in-vitro control measurements, variables were written on hard-disk from the
MATLAB program at dierent sampling times.
xx
Bibliography
[1] Maxon AG. Ec 22. Data Sheet, April, 2003.
[2] Maxon Motor AG. Maxon motor control. User Manual, July, 2001.
[3] Medos AG. Deltastream. Medos Medizintechnik, Product Information, 2003.
[4] J. Allen, A. C. Fisher, J. D. Gaylor, and A. R. Razieh. Development of a digital adaptive
control system for pO
2
regulation in a membrane oxygenator. J. Biomed. Eng., 14(5):404
11, 1992. 0141-5425 Journal Article.
[5] J. Anbe, H. Nakajima, Y. Ogura, M. Ozeki, T. Mitsuishi, T. Akasaka, and T. Tobi.
Development of a computer-regulated extracorporeal circulation system. Artif. Organs
Today, 2(2):117125, 1992.
[6] J. Anbe, T. Tobi, H. Nakajima, T. Akasaka, and K. Okinaga. Microcomputer-based
automatic regulation of extracorporeal circulation: A trial for the application of fuzzy
inference. Artif. Organs, 16(5):5328, 1992. 0160-564x Journal Article.
[7] J. J. E. Angell. The eects of altering mean pressure, pulse pressure and pulse frequency
of the impulse activity in baroreceptor bres from the aortic arch and right subclavian
artery in the rabbit. J. Physiol., 214:6588, 1971.
[8] American Heart Association. Heart disease and stroke - statistics 2004 update. 2004.
[9] G. Avanzolini, P. Barbini, and A. Cappello. Comparison of algorithms for tracking short-
term changes in arterial circulation parameters. IEEE Trans. Biomed. Eng., 39(8):8617,
1992. 0018-9294 Journal Article.
[10] G. Avanzolini, P. Barbini, A. Cappello, and G. Cevenini. Cadcs simulation of the closed-
loop cardiovascular system. Int. J. Biomed. Comput., 22(1):3949, 1988. 0020-7101
Journal Article.
xxi
Bibliography
[11] G. Avanzolini, P. Barbini, A. Cappello, and M. R. Massai. Sensitivity analysis of the
systemic circulation with a view to computer simulation and parameter estimation. J.
Biomed. Eng., 11(1):437, 1989. 0141-5425 Journal Article.
[12] A. P. Avolio. Multi-branched model of the human arterial system. Med. Biol. Eng.
Comput., 18(6):70918, 1980. 0140-0118 Journal Article.
[13] H. D. Baehr and K. Stephan. Warme- und Sto ubertragung. Springer, Berlin, 1996.
[14] L. Balmer. Signals and Systems. Prentice Hall, London, 1997.
[15] R. Bauernschmitt, E. Naujokat, H. Mehmanesh, S. Schulz, C. F. Vahl, S. Hagl, and
R. Lange. Mathematical modelling of extracorporeal circulation: Simulation of dierent
perfusion regimens. Perfusion, 14(5):32130, 1999. 0267-6591 Journal Article.
[16] W. Beitz and K. H. K uttner. Taschenbuch f ur den Maschinenbau. Springer-Verlag, Berlin
- Heidelberg - New York, 1995.
[17] T. Beppu, Y. Imai, and Y. Fukui. Computer-controlled cardiopumonary bypass system.
Systems and Computers in Japan, 23(11):7484, 1992.
[18] T. Beppu, Y. Imai, and Y. Fukui. A computerized control system for cardiopulmonary
bypass. J. Thorac. Cardiovasc. Surg., 109(3):42838, 1995. 0022-5223 Clinical Trial
Controlled Clinical Trial Journal Article.
[19] R. Berber and C. Kravaris, editors. Nonlinear Model Based Control. Kluwer Academic
Publishers, Dordrecht, 1998.
[20] F. Boschetti, F. M. Montevecchi, and R. Fumero. Virtual extracorporeal circulation
process. Int. J. Artif. Organs, 20(6):341351, 1997.
[21] H. Brandes, J. M. Albes, A. Conzelmann, M. Wehrmann, and G. Ziemer. Comparison of
pulsatile and nonpulsatile perfusion of the lung in an extracorporeal large animal model.
Eur. Surg. Res., 34(4):3219, 2002. 0014-312x Journal Article.
[22] G. Cattaneo, A. Strauss, and H. Reul. Compact intra- and extracorporeal oxygenator
developments. Perfusion, 19(4):251255, 2004.
xxii
Bibliography
[23] S. Choi, J. F. Antaki, R. Boston, and D. Thomas. A sensorless approach to control of a
turbodynamic left ventricular assist system. IEEE Trans. Contr. Sys. Tech., 9:473482,
2001.
[24] S. Choi, J. R. Boston, D. Thomas, and J. F. Antaki. Modelling and identication of an
axial ow pump. Proc. Am. Control Conference, 4:37143715, 1997.
[25] Jr. Clark, J. W., G. R. Kane, and H. M. Bourland. On the feasibility of closed-loop
control of intra-aortic balloon pumping. IEEE Trans. Biomed. Eng., 20(6):40412, 1973.
0018-9294 Journal Article.
[26] D. W. Clarke, C. Mohtadi, and P. S. Tus. Generalized predictive control. Automatica,
23(2):137160, 1987.
[27] R. de Vroege, P. M. Rutten, C. Kalkman, T. A. Out, P. G. Jansen, L. Eijsman, B. J.
de Mol, and C. R. Wildevuur. Biocompatibility of three dierent membrane oxygenators:
Eects on complement, neutrophil and monocyte activation. Perfusion, 12(6):36975,
1997. 0267-6591 Clinical Trial Journal Article Randomized Controlled Trial.
[28] D. E. Dick, J. E. Kendrick, G. L. Matson, and V. C. Rideout. Measurement of nonlinearity
in the arterial system of the dog by a new method. Circ. Res., 22(2):10111, 1968. 0009-
7330 Journal Article.
[29] A. Doenicke, D. Kettler, W. F. List, J. Radke, and J. Tarnow, editors. Anasthesiologie.
Springer, Berlin, 1995.
[30] J. J. Driessen, H. Dhaese, G. Fransen, P. Verrelst, L. Rondelez, L. Gevaert, M. van
Becelaere, and E. Schelstraete. Pulsatile compared with nonpulsatile perfusion using
a centrifugal pump for cardiopulmonary bypass during coronary artery bypass grafting.
Eects on systemic haemodynamics, oxygenation, and inammatory response parameters.
Perfusion, 10(1):312, 1995.
[31] V. B. Fiedler. Eects of pulsatile and non-pulsatile perfusion on the isolated canine heart.
Res. Exp. Med. (Berl.), 179(3):18398, 1981. 0300-9130 Journal Article.
[32] W. Forth, D. Henschler, and W. Rummel. Allgemeine und spezielle Pharmakologie und
Toxikologie. Bibliographisches Institut, Mannheim, 1983.
xxiii
Bibliography
[33] P. Francheteau, J. L. Steimer, H. Merdjan, M. Guerret, and C. Dubray. A mathematical
model for dynamics of cardiovascular drug action: Application to intravenous dihydropy-
ridines in healthy volunteers. J. Pharmacokin. Biopharm., 21(5):489513, 1993.
[34] O. Frank. Die Grundform des arteriellen Pulses. Z. Biol., 37:483526, 1899.
[35] Y. Fukui, K. Tsuchiya, and Y. Imai. Computer controlled extracorporeal circulation
(ECC) with pulsatile perfusion for an infant. Trans. Am. Soc. Artif. Intern. Organs,
28:1337, 1982. 0066-0078 Journal Article.
[36] Y. C. Fung. Biomechanics Circulation. Springer, New York - Berlin - Heidelberg, 1997.
[37] C. Gao, A. H. Stammers, R. L. Ahlgren, T. A. Ellis, H. B. Holcomb, B. T. Nutter, R. G.
Schmer, and L. Hock. The eects of preprimed oxygenators on gas transfer eciency. J.
Extra. Corpor. Technol., 35(2):1216, 2003. 0022-1058 Journal Article.
[38] J. F. Gardner, M. Ignatoski, U. Tasch, A. J. Snyder, and D. B. Geselowitz. Aortic
pressure estimation with electro-mechanical circulatory assist devices. J. Biomech. Eng.,
115(2):18794, 1993. 0148-0731 Journal Article.
[39] K. Gersten and H. Herwig. Stromungsmechanik. Vieweg, Braunschweig, 1992.
[40] C. Gobel, A. Arvand, R. Eilers, O. Marseille, C. Bals, B. Meyns, W. Flameng, G. Rau,
and H. Reul. Development of the MEDOS/HIA deltastream extracorporeal rotary blood
pump. Artif. Organs, 25(5):35865, 2001. 0160-564x Journal Article.
[41] R. J. Gordon, M. Ravin, G. R. Daico, and R. E. Rawitscher. Eects of hemodilution on
hypotension during cardiopulmonary bypass. Anesth. Analg., 54(4):4828, 1975. 0003-
2999 Journal Article.
[42] W. Gundel, G. Cherry, B. Rajagopalan, L. B. Tan, G. Lee, and D. Schultz. Aortic input
impedance in man: Acute response to vasodilator drugs. Circulation, 63(6):130514,
1981. 0009-7322 Journal Article.
[43] A. C. Guyton. Textbook of Medical Physiology. W. B. Saunders Company, Philadelphia,
1986.
[44] M. A. Henson and D. E. Seborg. Critique of exact linearization strategies for process
control. J. Proc. Cont., 1:122139, 1991.
xxiv
Bibliography
[45] D. A. Hettrick, P. S. Pagel, and D. C. Warltier. Dierential eects of isourane and
halothane on aortic input impedance quantied using a three-element windkessel model.
Anesthesiology, 83(2):36173, 1995. 0003-3022 Journal Article.
[46] M. Hexamer, B. Misgeld, A. Prenger-Berningho, U. Sch utt, H. J. Knobl, R. Korfer, and
J. Werner. Automatic control of the extra-corporal bypass: System analysis, modelling
and evaluation of dierent control modes. Biomed. Technik, 49:316321, 2004.
[47] M. Hexamer and J. Werner. A mathematical model for the gas transfer in an oxygenator.
In D. Feng and E. Carson, editors, Modelling and Control in Biomedical Systems, pages
409414, Melbourne, Australia, 2003.
[48] E. P. Hill, G. G. Power, and L. D. Longo. A mathematical model of carbon dioxide
transfer in the placenta and its interaction with oxygen. Am. J. Physiol., 224(2):28399,
1973. 0002-9513 Journal Article.
[49] E. P. Hill, G. G. Power, and L. D. Longo. Mathematical simulation of pulmonary O
2
and
CO
2
exchange. Am. J. Physiol., 224(4):90417, 1973. 0002-9513 Journal Article.
[50] E. P. Hill, G. G. Power, and L. D. Longo. Kinetics of O2 and CO2 exchange. In J. B.
West, editor, Bioengineering aspects of the lung, pages 459514. Marcel Dekker Inc., New
York and Basel, 1977.
[51] MathWorks Inc. Control System Toolbox. The MathWorks, Natick, Mass., 2002.
[52] MathWorks Inc. Signal Processing Toolbox. The MathWorks, Natick, Mass., 2002.
[53] MathWorks Inc. Robust Control Toolbox. The MathWorks, Natick, Mass., 2004.
[54] S. Isaka and A. V. Sebald. Control strategies for arterial blood pressure regulation. IEEE
Trans. Biomed. Eng., 40(4):35363, 1993. 0018-9294 Journal Article Review.
[55] A. Isidori, editor. Nonlinear Control Systems. Springer-Verlag, Berlin, 1989.
[56] A. Isidori and C. I. Byrnes. Output regulation of nonlinear systems. IEEE Trans. Au-
tomat. Contr., 35(2):131140, 1990.
[57] J. A. Jacquez, editor. Compartmental Analysis in Biology and Medicine. University of
Michigan Press, Ann Arbor, 1985.
xxv
Bibliography
[58] D. Jaron, T. W. Moore, and P. He. Control of intraaortic balloon pumping: Theory and
guidelines for clinical applications. Ann. Biomed. Eng., 13(2):15575, 1985. 0090-6964
Journal Article.
[59] R. E. Kalman. A new approach to linear ltering and prediction problems. ASME - J.
Basic Eng., 82:3545, 1960.
[60] R. E. Kalman and R. Bucy. New results in ltering and prediction theory. J. Basic Eng.,
83:95108, 1961.
[61] H. Kaufman, R. Roy, and X. Xu. Model reference adaptive control of drug infusion rate.
Automatica, 20(2):205209, 1984.
[62] P. H. Kay and C. M. Munsch. Techniques in extracorporeal circulation. Arnold, London,
2004.
[63] G. R. Kelman. Digital computer subroutine for the conversion of oxygen tension into
saturation. J. Appl. Physiol., 21:13751376, 1966.
[64] T. Kitamura. Left atrial pressure controller design for an articial heart. IEEE Trans.
Biomed. Eng., 37(2):1649, 1990. 0018-9294 Journal Article.
[65] G. K. Klute, U. Tasch, and D. B. Geselowitz. An optimal controller for an electric
ventricular assist device: Theory, implementation and testing. IEEE Trans. Automat.
Contr., 39(4):394403, 1992.
[66] C. Kravaris and R. A. Wright. Deadtime compensation for nonlinear processes. AIChE
J., 35(9):15351542, 1989.
[67] T. W. Latson, W. C. Hunter, N. Katoh, and K. Sagawa. Eect of nitroglycerin on aortic
impedance, diameter, and pulse-wave velocity. Circ. Res., 62(5):88490, 1988. 0009-7330
Journal Article.
[68] G. Lauterbach. Handbuch der Kardiotechnik. Gustav Fischer Verlag, L ubeck, 1996.
[69] D. Lowe, D. A. Hettrick, P. S. Pagel, and D. C. Warltier. Propofol alters left ventricular
afterload as evaluated by aortic input impedance in dogs. Anesthesiology, 84(2):36876,
1996. 0003-3022 Journal Article.
xxvi
Bibliography
[70] K. Lu, J. W. Clark, F. H. Ghorbel, D. L. Ware, and A. Bidani. An integrated model
of the human cardiopulmonary system. In Proc. 23rd IEEE Eng. Med. Biol. Soc., pages
412414, Piscataway, NJ, 2001.
[71] J. Lunze. Regelungstechnik 1. Springer-Verlag, Berlin, 2002.
[72] J. Lunze. Regelungstechnik 2. Springer-Verlag, Berlin, 2002.
[73] J. Lunze. Automatisierungstechnik. Oldenbourg, M unchen, 2003.
[74] S. P. Marlow. A PO
2
regulation system for a membrane oxygenator. PhD thesis, Univer-
sity of Strathclyde, 1982.
[75] B. C. McInnis, Z.-W. Guo, P. C. Lu, and J.-C. Wang. Adaptive control of left ventricular
bypass assist devices. IEEE Trans. Automat. Contr., 30(4):322329, 1985.
[76] S. I. Merz. Automatic Control of Extracorporeal Life Support. PhD thesis, University of
Michigan, 1993.
[77] S. I. Merz, R. H. Bartlett, J. M. Jenkins, and P. T. Kabamba. Controller design for
extracorporeal life support. In Proc. 18th IEEE Eng. Med. Biol. Soc., pages 17331735,
Piscataway, NJ, 1996.
[78] B. J. E. Misgeld and M. Hexamer. Modellierung und Regelung des arteriellen Blutusses
wahrend der extrakorporalen Zirkulation. Automatisierungstechnik, 53(9):454461, 2005.
[79] B. J. E. Misgeld, J. Werner, and M. Hexamer. Automatic control of extracorporeal
circulation: Arterial blood ow control. Biomed. Technik, 50:857858, 2005.
[80] B. J. E. Misgeld, J. Werner, and M. Hexamer. Robust and self-tuning blood ow control
during extracorporeal circulation in the presence of system parameter uncertainties. Med.
Biol. Eng. Comput., 43:589598, 2005.
[81] B. J. E. Misgeld, J. Werner, and M. Hexamer. In-vitro experimental setup for the simu-
lation of blood-gas exchange during cardiopulmonary bypass. Biomed. Technik, 2006.
[82] B. J. E. Misgeld, J. Werner, and M. Hexamer. Nonlinear robust blood gas control by state
linearisation for the cardiopulmonary bypass. Control Eng. Practice, 2006. submitted.
[83] B. J. E. Misgeld, J. Werner, and M. Hexamer. Simultaneous control of O
2
- and CO
2
-blood
gases during cardiopulmonary bypass. Biomed. Technik, 2006.
xxvii
Bibliography
[84] B.J.E. Misgeld, J. Werner, and M. Hexamer. Strategies for haemodynamic control of ex-
tracorporeal circulation. In 6
th
IFAC Symposium on Modelling and Control in Biomedical
Systems, pages 351356, Reims, France, 2006.
[85] B.J.E. Misgeld, J. Werner, and M. Hexamer. Automatisierung der extrakorporalen
Zirkulation: Ein Vergleich verschiedener Regelansatze. In O. Simanski, editor, Automa-
tisierungstechnische Methoden und Systeme f ur die Medizin (AUTOMED), pages 5556,
Rostock, Germany, 2006.
[86] D. Moeller. Ein geschlossenes nichtlineares Modell zur Simulation des Kurzzeitverhaltens
des Kreislaufsystems und seine Anwendung zur Identikation. PhD thesis, Universitat
Bremen, Berlin, 1981.
[87] M. Morari and E. Zariou. Robust Process Control. Prentice-Hall International, Inc.,
Englewood Clis, NJ, 1989.
[88] P. Moreton. Industrial Brushless Servomotors. Reed Educational and Professional Pub-
lishing, Oxford, 2000.
[89] A. Mori, K. Watanabe, M. Onoe, S. Watarida, Y. Nakamura, T. Magara, R. Tabata, and
Y. Okada. Regional blood ow in the liver, pancreas and kidney during pulsatile and
nonpulsatile perfusion under profound hypothermia. Jpn. Circ. J., 52(3):21927, 1988.
0047-1828 Journal Article.
[90] M. H. Nadjmabadi, H. Rastan, M. T. Saidi, and E. Aftandelian. Hamodynamische Veran-
derungen nach akuter intraoperativer Hamodilution bei oener Herzchirurgie. Anaesthe-
sist, 27(8):364369, 1978.
[91] M. Nagel. Aufbau eines Versuchsstandes zur Simulation des menschlichen Kreislaufs unter
den Bedingungen der extrakorporalen Membranoxygenation (ECMO). Masters thesis,
Dortmund University, 2004.
[92] E. Naujokat. Ein Beobachtersystem f ur den Patientenzustand in der Herzchirurgie. PhD
thesis, Universitat Karlsruhe, Aachen, 2002.
[93] O. Nelles. Nonlinear System Identication. Springer-Verlag, Berlin, 2001.
[94] N. S. Nise. Control Systems Engineering. Wiley and Sons Inc., New York, 2004.
xxviii
Bibliography
[95] H. Nishida, T. Beppu, M. Nakajima, T. Nishinaka, H. Nakatani, K. Ihashi, T. Katsumata,
M. Kitamura, S. Aomi, M. Endo, and et al. Development of an autoow cruise control
system for a centrifugal pump. Artif. Organs, 19(7):7138, 1995. 0160-564x Journal
Article.
[96] A. Noordergraaf, A. P. D. Verdouw, and H. B. C. Boom. The use of an analog computer
in a circulation model. Prog. Cardiovasc. Dis., 5:41939, 1963.
[97] K. Ogata. Discrete - Time Control Systems. Prentice Hall, Englewood Clis, NJ, 1987.
[98] J. M. Orenstein, N. Sato, B. Aaron, B. Buchholz, and S. Bloom. Microemboli observed in
deaths following cardiopulmonary bypass surgery: Silicone antifoam agents and polyvinyl
chloride tubing as sources of emboli. Hum Pathol, 13(12):108290, 1982.
[99] M. Oshikawa, K. Araki, K. Nakamura, H. Anai, and T. Onitsuka. Detection of total assist
and sucking points based on the pulsatility of a continuous ow articial heart: In vivo
evaluation. Asaio J., 44(5):M7047, 1998. 1058-2916 Journal Article.
[100] M. T. OToole, editor. Encyclopedia and Dictionary of Medicine, Nursing and Allied
Health. Saunders, Philadelphia, 2003.
[101] A. D. Pacico, S. Digerness, and J. W. Kirklin. Acute alterations of body composition
after open intracardiac operations. Circulation, 41(2):33141, 1970.
[102] C. C. Palerm, B. W. Bequette, and S. Ozcelik. Robust control of drug infusion with time
delays using direct adaptive control: Experimental results. In Proc. Am. Contr. Conf.,
volume 5, pages 29722976, Chicago, U.S.A., 2000.
[103] L. De Pater. An Electrical Analogue of the Human Circulatory System. PhD thesis,
University of Groningen, Rotterdam, 1966.
[104] G. Pennati, G. B. Fiore, K. Lagan` a, and R. Fumero. Mathematical modeling of uid
dynamics in pulsatile cardiopulmonary bypass. Artif. Organs, 28(2):196209, 2004.
[105] C. J. Pepine, W. W. Nichols, Jr. Curry, R. C., and C. R. Conti. Aortic input impedance
during nitroprusside infusion. A reconsideration of afterload reduction and benecial ac-
tion. J. Clin. Invest., 64(2):64354, 1979. 0021-9738 Journal Article.
[106] H. Reul, H. Minamitani, and J. Runge. A hydraulic analog of the systemic and pulmonary
circulation for testing articial hearts. Proc. ESAO II, 2:120127, 1975.
xxix
Bibliography
[107] V. C. Rideout. Cardiovascular system simulation in biomedical engineering education.
Trans. Biomed. Eng., 19(2):101107, 1972.
[108] V. C. Rideout and D. E. Dick. Dierence-dierential equations for uid ow in distensible
tubes. IEEE Trans. Biomed. Eng., 14(3):1717, 1967. 0018-9294 Journal Article.
[109] W. C. Rose and A. A. Shoukas. Two-port analysis of systemic venous and arterial
impedances. Am J Physiol, 265(5 Pt 2):H157787, 1993. 0002-9513 Journal Article.
[110] S. Sastry, editor. Nonlinear Systems. Springer-Verlag, New York, 1999.
[111] H. Schima, J. Honigschnabel, W. Trubel, and H. Thoma. Computer simulation of the
circulatory system during support with a rotary blood pump. ASAIO Trans., 36(3):M252
4, 1990. 0889-7190 Journal Article.
[112] R. F. Schmidt, G. Thews, and F. Lang, editors. Physiologie des Menschen. Springer,
Berlin, 2000.
[113] K. Schmidt-Nielsen. Animal Physiology. Cambridge University Press, London, 1975.
[114] S. Schulz, R. Bauernschmitt, F. Maar, A. Schwarzhaupt, C. F. Vahl, and U. Kiencke. Mod-
ellierung des Barorezeptorreexes in einem pulsatilen Modell. Biomed. Techn. (Suppl.),
43:310311, 1998.
[115] A. Schwarzhaupt. Regelung der extrakorporalen Zirkulation auf der Basis eines Modells
des menschlichen Kreislaufes. PhD thesis, Universitat Karlsruhe, Gottingen, 2002.
[116] A. Schwarzhaupt, B. Qaqunda, and U. Kiencke. Entwurf eines pradiktiven MIMO-Reglers
f ur Herz-Lungen-Maschinen auf der Grundlage eines Modells der extrakorporalen Zirku-
lation. Biomed. Tech. (Suppl.), 43:336337, 1998.
[117] P. Segers, F. Dubois, D. De Wachter, and P. Verdonck. Role and relevancy of a cardio-
vascular simulator. CVE, 3(1):4856, 1998.
[118] T. Shimooka, Y. Mitamura, and T. Yuhta. Investigation of parameter estimator and
adaptive controller for assist pump by computer simulation. Artif. Organs, 15(2):11928,
1991. 0160-564x Journal Article.
xxx
Bibliography
[119] B. W. Smith, J. G. Chase, G. M. Shaw, and R. I. Nokes. Experimentally veried mini-
mal cardiovascular system model for rapid diagnostic assistance. Control Eng. Practise,
13:118393, 2004.
[120] O. J. M. Smith. Closer control of loops with dead time. Chem. Eng. Progress, 53(3):217
219, 1957.
[121] M. F. Snyder and V. C. Rideout. Computer simulation studies of the venous circulation.
IEEE Trans. Biomed. Eng., 16(4):32534, 1969. 0018-9294 Journal Article.
[122] M. F. Snyder, V. C. Rideout, and R. J. Hillestad. Computer modeling of the human
systemic arterial tree. J. Biomech., 1:341353, 1968.
[123] K. Soejima, Y. Nagase, K. Ishihara, Y. Takanashi, Y. Imai, K. Tsuchiya, and Y. Fukui.
Computer-assisted automatic cardiopulmonary bypass system for infants. In K. Atsumi,
M. Maekawa, and K. Ota, editors, Progr. Artif. Organs, pages 918922. ISAO Press,
Cleveland, OH, 1983.
[124] Z. Song, C. Wang, and A. H. Stammers. Clinical comparison of pulsatile and nonpulsatile
perfusion during cardiopulmonary bypass. J. Extra. Corpor. Technol., 29(4):1705, 1997.
0022-1058 Clinical Trial Journal Article Randomized Controlled Trial.
[125] K. M. Sutherland, D. T. Pearson, and L. S. Gordon. Independent control of blood gas
pO2 and pCO2 in a bubble oxygenator. Clin. Phys. Physiol. Meas., 9(2):97105, 1988.
[126] K. M. Taylor, W. H. Bain, and J. J. Morton. The role of angiotensin II in the development
of peripheral vasoconstriction during open-heart surgery. Am Heart J, 100(6 Pt 1):9357,
1980. 0002-8703 Journal Article.
[127] K. M. Taylor, W. H. Bain, M. Russell, J. J. Brannan, and I. J. Morton. Peripheral vascular
resistance and angiotensin II levels during pulsatile and no-pulsatile cardiopulmonary
bypass. Thorax, 34(5):5948, 1979. 0040-6376 Journal Article.
[128] K. M. Taylor, J. J. Brannan, W. H. Bain, P. K. Caves, and I. J. Morton. Role of
angiotensin ii in the development of peripheral vasoconstriction during cardiopulmonary
bypass. Cardiovasc Res, 13(5):26973, 1979. 0008-6363 Journal Article.
[129] R. J. Tschaut. Extrakorporale Zirkulation in Theorie und Praxis. Pabst Science Publish-
ers, Lengerich, 1999.
xxxi
Bibliography
[130] M. Turina, B. Litchford, I. Babotal, M. Intaglietta, and N. S. Braunwald. Servo-controlled
extended cardiopulmonary bypass. Trans. Am. Soc. Artif. Intern. Organs, 19:50410,
1973. 0066-0078 Journal Article.
[131] W. Saggau und I. Bac a, E. Ros, H. H. Storch, and W. Schmitz. Klinische und expe-
rimentelle Studie uber den pulsatilen und kontinuierlichen Flu wahrend des extrakorpo-
ralen Kreislaufs. Herz, 5(1):4250, 1980.
[132] V. Videm, T.E. Mollnes, P. Garred, and J.L. Svennevig. Biocompatibility of extracorpo-
real circulation. In vitro comparison of heparin-coated and uncoated oxygenator circuits.
J. Thorac. Cardiovasc. Surg., 101(4):654660, 1991.
[133] J. Wada, T. Hino, K. Kaizuka, and W. R. Ade. Automatic regulation of the cardiopul-
monary bypass. Perfusion, 1:117124, 1986.
[134] P. E. Wellstead and M. B. Zarrop. Self-Tuning Systems. Wiley and Sons Ltd., Chichester,
1991.
[135] C. Welp, J. Werner, D. Boehringer, and M. Hexamer. Ein pulsatiles Herz/Kreislauf-
Modell f ur die Herzschrittmachertechnik. Automatisierungstechnik, 50(7):326333, 2002.
[136] J. Werner, editor. Kooperative und autonome Systeme der Medizintechnik. Oldenbourg,
M unchen, 2005.
[137] J. Werner, D. Bohringer, and M. Hexamer. Simulation and prediction of cardiotherapeu-
tical phenomena from a pulsatile model coupled to the guyton circulatory model. IEEE
Trans. Biomed. Eng., 49(5):4309, 2002. 0018-9294 Journal Article.
[138] N. Westerhof and A. Noordergraaf. Arterial viscoelasticity: A generalized model. J.
Biomech., 3:35779, 1970.
[139] J. R. Womersley. An elastic tube theory of pulse transmission and oscillatory ow in
mammalian arteries. WADC Technical Report, pages 56614, 1957.
[140] G. Wright. The hydraulic power outputs of pulsatile and non-pulsatile cardiopulmonary
bypass pumps. Perfusion, 3:251262, 1988.
[141] T. Yaginuma, A. Avolio, M. ORourke, W. Nichols, J. J. Morgan, P. Roy, D. Baron,
J. Branson, and M. Feneley. Eect of glyceryl trinitrate on peripheral arteries alters left
ventricular hydraulic load in man. Cardiovas. Resarch, 20:153160, 1986.
xxxii
Bibliography
[142] Y.-C. Yu, J. R. Boston, M. A. Simaan, and J. F. Antaki. Estimation of systemic vascular
bed parameters for articial heart control. IEEE Trans. Automat. Contr., 43(6):765778,
1998.
[143] G. Zames. Feedback and optimal sensitivity: Model reference transformations, multiplica-
tive seminorms, and approximate inverses. IEEE Trans. Automat. Contr., 26(2):301320,
1981.
[144] G. Zames and B. A. Francis. Feedback, minimax sensitivity, and optimal robustness.
IEEE Trans. Automat. Contr., 28(5):585601, 1983.
[145] K. Zhou, J. C. Doyle, and K. Glover, editors. Robust and Optimal Control. Prentice-Hall,
New Jersey, 1996.
xxxiii
Curriculum Vitae
Personal Details
Name: Berno Johannes Engelbert Misgeld
Date of birth: 22. March, 1979
Place of birth: Euskirchen, Germany
Education
08/1996 - 06/1998: Technical High School, Euskirchen, Germany
Military Service
07/1996 - 05/1999: German Airforce, Fernmelderegiment 122, Cochem, Germany
Academic Education
10/2002 - 10/2003: Master of Science (M.Sc.) in Informatics and Control Engineering
Coventry University, Coventry, U.K.
10/1999 - 10/2003: Diplom Ingenieur (FH) in Electrical Engineering
Specialisation: Automation Engineering
Aachen University of Applied Sciences, Aachen, Germany
Work Experience
since 08/2006: Development engineer for automatic ight control/steering
system engineering at Diehl BGT Defence, N urnberg, Germany
02/2004 - 07/2006: Research associate at the Department for Biomedical Engineering,
School of Medicine, Ruhr-University Bochum, Bochum, Germany
06/2003 - 09/2003: Master Thesis at TRW LucasVarity, System modelling and control
of electric power steering systems, Birmingham, U.K.
04/2002 - 08/2002: Engineering student research project at HiTec Zang,
Automatic control for bio-chemical reactors, Aachen, Germany