Procedure 008
A) Tool is to be operated @ 90 PSI for the break-in procedure

B) Tool is to fasten 10 to 20 cycles on test fixture.

C) Tool is to be operated in clockwise and counter clockwise operation.

D) Check for excessive vibration, heat or gearbox noise when operating the tool.

E) If tool is operating correctly proceed into the calibration procedures.

F) If tool is not operating correctly, refer to the RAD Tool Repair Manual for diagnosis.
**Air used for calibration must be able to supply
100 PSI @ 30CFM.**

*New World Technologies Inc. Calibration
Lab is tested on Wika Model (213.54) 2.5 test
gauge with stated accuracy of +/-0.5% of span.*

*Compressed air for calibration at New World
Technologies Inc. is supplied by a Sullair ES-8
75 CFM twin screw compressor through a 3/4
air line to regulator and (3 meters / 12) of 1/2
air line hose from the regulator to the tool. The
air line is dried with a Sullair Air Dryer Model
SR-125.*


**The below tools are to be tested dynamically with a joint simulator rundown fixture.**
**All of the below tools are to be calibrated in 50 ft./lb. increments.**

RAD Classic Series
Imperial

RAD 350SL
RAD 350SL-2
RAD 550SL
RAD 550SL-2
RAD 800NG
RAD 800NG-2
RAD Classic Series
Metric

RAD 475SL
RAD 475SL-2
RAD 750SL
RAD 750SL-2
RAD 1100NG
RAD 1100NG-2
RAD Gen X Series
Imperial

RAD 6 GX-R
RAD 7GX
RAD 7 GX-R
RAD 10GX



RAD GEN X Series
Metric

RAD 8 GX-R
RAD 10GX
RAD 10 GX-R
RAD 14GX
Procedure:

A) Log date, serial number of gearbox, transducer and serial number of air motor in a database.

B) Adjust the PSI on the regulator accordingly so that the tool is achieving its maximum output torque.

C) Ensure that the target torque value achieved is within +/- 3% of desired value. Record PSI value.

D) Once the torque value has been achieved, remove the tool from the calibration system and free run
the tool to record the PSI reading in the database and proceed to the next torque value to be tested (50
ft./lb increments).

**Note** Tool must be free running in forward or reverse with no load when air pressure reading
is taken.

E) Once all torque settings have been calibrated and recorded, enter these values into a software
program of your choice and create a torque chart comparing TORQUE vs PSI.

F) Attach the torque chart to the tool to ensure proper air pressure
is used.
RAD Classic Series
Imperial

RAD 1400NG
RAD 1400NG-2
RAD 1400NGX-R
RAD 1800NG
RAD 1800NG-2
RAD 1800NGX-R
RAD 30
RAD 30-2
RAD 50
RAD Classic Series
Metric

RAD 1900NG
RAD 1900NG-2
RAD 1900 NGX-R
RAD 2400NG
RAD 2400NG-2
RAD 2400NGX-R
RAD 4000
RAD 4000-2
RAD 6800
RAD Gen X Series
Imperial

RAD 25GX
RAD 34GX
RAD 85GX







RAD Gen X Series
Metric

RAD 34GX
RAD 46GX
RAD 115GX
**The below tools are to be tested dynamically with a joint simulator rundown fixture.**
**All of the below tools are to be calibrated in 100 ft./lb. increments.**

Procedure:

A) Log date, serial number of gearbox, transducer and serial number of air motor in a database.

B) Adjust the PSI on the regulator accordingly so that the tool is achieving its maximum output torque.

C) Ensure that the target torque value achieved is within +/- 3% of desired value. Record PSI value.

D) Once the torque value has been achieved, remove the tool from the calibration system and free run
the tool to record the PSI reading in the database and proceed to the next torque value to be tested (100
ft./lb increments).

**Note** Tool must be free running in forward or reverse with no load when air pressure reading
is taken.

E) Once all torque settings have been calibrated and recorded, enter these values into a software
program of your choice and create a torque chart comparing TORQUE vs PSI.

F) Attach the torque chart to the tool to ensure proper air pressure is used.

Example of RAD 1800NG Torque Chart. (100 ft./lb increments)

Das könnte Ihnen auch gefallen