Sie sind auf Seite 1von 38

Sr.

No
App.Cost* Qnt.
1 66,975 1
Bench top Microprocessor pH meter
should include Meter with integral
magnetic stirrer, pH electrode,
Temperature Sensor, Integral Sensor
holder/stand, Standard Solutions: pH 4,
7, 9 Calibration flasks with
magnetic stirring bar. Meter and Probe
should be water proof. Mode of
operation: fully automated / semi -
automated / manual: Automatic Buffer
recognition technical and up to 3
custom buffers chosen by the user.
CIVIL ENGG. DEPARTMENT
ENVIRONMENTAL ENGINEERING LAB.
Specification of Equipment
LABORATORY BENCH TOP P
H
METER
The instrument should includes:
Bench top pH Meter
pH electrode + ATC, low
maintenance, plastic body, gel for
general applications
Measured
Parameter :
pH, mV Temperature
Range 0 to 14pH
Resolution: Selectable from 0.1,
0.01 to 0.001 pH.
Temp Range: 0 to 80 C
Resolution : 0.1C
Accuracy: > 0.3 C
Automatic
recognition of
technical buffers, pH
2.00, 4.01, 7.00, 9.21
and 10.01 (values at
25C)(77F)
With 1, 2 or 3 buffers
selectable inside the
range.
Programmable
calibration validity
between 0 h and 7
days. Automatic
rejection of
electrodes in poor
condition.
Languages English.
Display Graphic, backlit
liquid crystal, 128x64
dots.
Data Logger: 330 measurements
Cable length: 1 meter
Display: LCD with backlight
Calibration beakers
Power supply with plug adapters
Technical specification
Workstation with built-in strirrer
and control
Calibration
Probe holder
Buffer and storage solutions
Probe protector tube
Data Storage : GLP features
(Date/Time, Sample
ID, User ID)
Temperature
Correction/
Compensation:
pH: Automatic with
temperature probe (pt
1000)
Power supply : External power
supply 240 V,
Connectable
Peripherals
Magnetic stirrer.
Materials Enclosure, ABS and
PC. Keypad, PET.
Physical
Parameters(Met
er)
Weight: 800 g. Size:
350 x 200 x 110 mm.
2 101,650 1
Warranty:- 1 Year
Training Include
LABORATORY BENCHTOP
CONDUCTIVITY/TDS METER
Conductivity
Range
0.2 S/cm to 200
mS/cm
Probe holder
storage solutions
Probe protector tube
Calibration beakers
Power supply with plug adapters
Bench Top Microprocessor conductivity
/TDS meter should be include: complete
meter, probe, Temperature Sensor
workstation with integrated stirrer, buffers,
standards, storage solutions, calibration
flasks. The system should be single probe
input display system suitable for
measurement of Conductivity, Resistivity,
Salinity, TDS & Temperature. Meter and
Probe should be water proof. The meter
should be able to connect with 220 Volts
through adapter. Mode of operation: fully
automated / semi -automated / manual with
automatic temperature compensation.
The equipment should includes:-
Benchtop Conductivity Meter
Condutivity/TDS electrode
Workstation with built-in strirrer and
control
Technical specifications
Resolution: 0.00 to 199.9 mS/cm
(Auto ranging, 5 digits
maximum)
Total Dissolved
Solids Range
0.1 mg/L to 128 g/L,
Resolution : 0 to 9.99, 99.9, 999
mg/L,9.99, 199.9 g/L
Salinity range 0 mg/L to 200 g/L
Resolution 0.0 to 1999 mg/L, 2.0 to
50.0 g/L
Resistivity Range 0 M Ohm to 500 M
Ohm
Resolution 0.0 to 1999 mg/L
Temp. 0 to 80 C
Resolution 0.1C
Accuracy > 0.3 C
Data Logging 400 measurements
Data Export sent to printer or PC
through the RS 232 C.
Display: LCD with backlight
Data Storage : GLP features
(Date/Time, Sample ID,
User ID)
Temperature
Correction/
Compensation:
Automatic with
temperature probe
Power supply : External power supply
240 V,
Cable Length 1 Meter
3 138,225 1
Warranty:- 1 Year
Training Include
Laboratory Multiparameter for LDO
Analysis
Range: 0.05 to 20.00
mg/L (0 to 200%)
Resolution: 0.01 mg/L
DO
Range/Resolution
Digital multi-parameter Meter having plug
and play operation to be supplied LDO
probes with necessary calibration and
maintenance accessories in a complete set.
The instrument shall have facility to plug
in any two probes and get simultaneous
measurement; simultaneous readout. The
instrument should feature Automatic
Temperature Compensation. The meters
should be able to store probe serial
numbers, calibration history, user ID, time,
date and location of readings. Should have
facility to transfer GLP/ISO data
effortlessly to a PC or USB flash stick. The
meter should be password protected for
accurate tamperproof readings.
The meter must be operable with normal
AA batteries and optionally with power.
The supplied probes should be rugged,
designed for harsh environments. Probe
should be able to store its calibration
history in the probe itself, minimizing
probe replacement and calibration, Should
feature Calibration Alert facility.
Technical specifications
Range: -10.0 to
110.0C
Resolution: 0.1C
Accuracy: 0.3C
(0.54F)
Result Display Display readings from
one probe or
simultaneous reading
from two probes
Memory 500 data points
Data Storage GLP/ISO compliant
reading data stored with
calibration details.
Calibration details and
check standard readings
documented as events in
log. Automatically store
in mode and interval
measurement mode.
Manually store in
mode.
Water Resistance Meter Casing: 1 meter
submersion for 30
minutes
4 56,414 1
Temperature
Range/Resolution
/Accuracy
Warranty:- 1 Year
Training Include
Digital Titrator:-
Digital Titrator, 0.00125 mL/digit, with
case, manual, and five straight delivery
tubes.Interchangeable cartridges which
allow to change titrants in seconds. No
cleaning or rinsing burets. More than 50
parameter/range combinations ,
Concentrated titrants should allow 40-100
typical titrations per cartridge.
5 95,000 1
Techanical Specification:-
Accuracy: 1 % (Uncertainty of readings is
1 digit. )
Weight:0.29 lbs. (0.132 kg)
Equipment should includes:-
a) Digital Titrator Qty 1.
b) Alkalinity Cartridge Approx
Test 200
c) Acidity Cartridge Approx
Test 200
d) Total Hardness Cartridge
Approx Test 200
Warranty:- 1 Year
Training Include
Portable Turbidity Meter:-
Equipment should includes:
Four AA alkaline batteries, a carrying case
with insert, primary calibration standards
in 1 sealed vials (20, 100, 800 NTU), a 10
NTU primary verification standard, 6
sample cells with screw-tops, instrument
manual (printed and on CD-ROM), quick
start guide, silicone oil and oiling cloth.
Technical Specification:-
Ratio turbidity metric determination using
a primary nephelo metric light scatter
signal (90) to the transmitted light scatter
signal.
Regulatory: 2100Q: Meets EPA Method
180.1
Light Source: 2100Q: Tungsten filament
lamp
Range: 0 to 1000 NTU (FNU)
Accuracy: 2% of reading plus stray light
from 0 to 1000 NTU
Repeatability: 1% of reading, or 0.01
NTU (FNU), whichever is greater
Resolution: NTU on lowest range
Stray Light: <0.02 NTU (FNU)
Signal Averaging: Selectable on/off
Detector: Silicon photovoltaic
Reading Modes (user selectable)
Normal (Push to Read)
Rapidly Settling Turbidity
Data Logger: 500 records
Power Requirement: 110-230 VAC, 50/60
Hz
4 AA alkaline batteries
Operating Conditions
Temperature: 0 to 50C (32 to 122F)
Relative Humidity: 0 to 90% @ 30C,
6 207,400 1
0 to 80% @ 40C, 0 to 70% @ 50C, no
condensing
Storage Conditions
-40 to 60C (-40 to 140F), instrument only
Warranty:- 1 Year
Training Include
Portable TSS Analyser
TSS Analyser should includes :
The instrument should measure Turbidity,
Suspended Solids, and Sludge Blanket
Level with multi-beam alternating light
method with infrared diode system.
Should be capable to store up to four
different calibration curves for suspended
solids and one for turbidity.
Instrument should not require need for
constant re-calibration for different
locations or types of sludge.
The sludge blanket level should be
measured by setting the units to mg/L, g/L
or percent.
The instrument software should
compensate for air bubbles to prevent its
influence on the measurement.
The probe should be made of good material
which should be scratch-resistant to
withstand harsh environments.
The internal memory of equipment should
be able to store at least 200 measuring
values, including time, date, measurement,
location, and homogeneity.
Technical Specifications
Accuracy for measurement of Turbidity:
less than 3 % or 0.02 NTU whichever is
greater
Accuracy for measurement of suspended
solids: less than 4 % or 0.001 g/L,
whichever is greater
Accuracy for measurement of TSS
Content: < 0.04 or 0.02 FNU (whichever
is greater)
Interface: RS485
Range for Suspended Solids : 0.001 to 400
gm/l(1 to 400,000 mg/l)
Measuring Range Turbidity: 0.001 to 9999
FNU
Warranty:- 1 Year
Training Include
7 433,320 1 BOD Meter & BOD Incubator with
Reagent:-
BOD Meter-
BOD Meter should measure BOD from 0
700 mg/l without any sample dilutions.
Meter shall give the results directly in ppm
and also shall show the BOD trend during
entire test length for each measurement
channel. Meter shall have at least 6
measuring channels. Meter shall be
compact enough to fit in to standard BOD
incubators available in the laboratory.
Meter shall be operated through soft touch
keypad for channel selection, control
settings etc., and shall shut off
automatically after the test length. Meter
shall store data of all measuring channels
for the entire test length. Stored data shall
be compatible to direct transfer to a PC.
Meter shall meet the following functional
specifications:
Specifications:
Range : Selectable, 0 to
35 mg/L, 0 to
70mg/L, 0 to 350
mg/L,
0 to 700 mg/L
Test Duration : Selectable, 5-
,7-, 10-, or 30 day test
Capacity : Six 473-ml BOD
bottles
Power : 230 Vac,
50/60 Hz
Computer interface : RS232
Display : Graphical LCD for
continuous
updates on
BOD test
results and
status.
Data collection and storage: Automated on
each test run.
Equipment should include:
Meter should supplied with 6 473ml amber
colored BOD bottles, Nutrient solution for
50 tests, CO
2
absorbing Power packs for
50 tests, 6 holding cups for CO
2
absorbing
powder, grease tube, Power adapter and
instrument operation manual.

BOD Incubator-
Incubating cultures of micro organisms,
storing samples, determining enzymatic
activity, and all uses requiring that
products be kept at a constant and precise
temperature. Continuous ventilation and
the special microprocessor controlled
AUTO-TUNING thermoregulation system
ensure temperature uniformity in all points
inside the chamber with continuous
monitoring of the room temperature so the
Set Point set is kept precisely aligned. The
internal temperature is programmed and
displayed digitally with selection ofthe
tenth degree Celsius from 3 to 50C
(stability 0.5C). Can be connected to a
PC using an RS232 connector for use with
dedicated software.

GENERAL FEATURES
Total volume:207 liters
Usable internal volume:196 liters (6.92
Cu Ft) Number of sockets:2 internal
electrical current socket Number of shelves
included:4 Max number of shelves:12
(distance of 40 mm between one shelf and
the next)
8 227,930 1
Electronic thermoregulation
system:AUTO - TUNING
Transparent internal door,yes
Power:350 W
Power supply:230 V / 50-60 Hz
Weight:40 Kg
Dimension (WxHxD):540x1300x560 mm
PERFORMANCES
Temperature range:,from 3.0 to 50.0C
Internal temperature stability: 0.5C
Internal temperature homogeneity: 0.5C
Internal temperature visualization:display
Temperature resolution:0.1C
Warranty:- 1 Year
Training Include
Portable Colorimeter:-
Colorimeter should includes:-
Auto-wavelength selection,
datal logging capabilities and user-
generated calibrations.
Durable, hand-held, filter
photometers with rugged
components and waterproof,
dustproof, chemical-resistant
housings. Equipment should offer
simple, push-button program
selection and step-by-step
programme for the user guidance.
Colorimeter should preprogrammed to
measure COD, Aluminum, Ammonia,
Bromine, Chlorine, Chromium, Fluoride,
Iron, Manganese, nickel, nitrate, nitrite,
Ammonia-Nitrogen, Total Kjeldahl
Nitrogen, Phosphorous, Silica, Sulfides,
Sulfates, TOC, Suspended solids, Zinc etc.
Colorimeter should further meet the
following specifications:-
Wavelength Selection: Auto Select
Photometric Range: 0 to 2 Abs
Operational Modes: Concentration, Abs,
%T
External Outputs: Infrared to RS-232
Serial using the optional Data Transfer
Adapter
Data logging: 99 sample readings (each
reading includes date/time, parameter,
program number, concentration,
absorbance, % transmittance, sample
number, instrument serial number)
User Programs: 10 user-entered programs,
12 data points each
Display: Large liquid-crystal display
(LCD) shows results, parameter name,
program number and units
Weight: 470 g (1 lb)
Power: 4 AA alkaline or rechargeable
alkaline batteries, nominal life of six
months
Storage Temperature Range: -40 to 60C (-
40 to 140F)
Operating Temperature Range: 0 to 50C
(32 to 122F)
Environmental: Dustproof and waterproof
to one meter for 30 minutes
The equipment should include
a) COD Digestion Reactor
The digestion reactor shall be a bench top
instrument for heating prepared samples to
temperature for the designated time for the
COD test. The instrument shall have a
LCD display and shall have facility for
programmable temperature range of 37 to
165C in 1
o
C increment. The
programmable timer range should be from
0 to 480 minutes. Reactor shall emit an
audible signal automatically and shut down
at the end of the digestion time. The
instrument should have a heating rate of 20
to 150C in 10 minutes. The instrument
shall be equipped with stored digestion
programs for COD and TOC. The reactor
shall also be capable of storing 3 user
defined methods for digestion. The
instrument shall have Heating blocks to
accommodate 16-mm and or 20-mm vials
in total. Heating blocks shall be able to set
temperature and timers independently
allowing to use the reactor for two testing
methods. Heating blocks shall be covered
with transparent lids for operator safety.
The power requirements of the instrument
shall be 100-240 Vac, 50/60 Hz.
Technical Specification:-
Programmable Temperature Range: 37
to 165C (99 to 329F)
Operating Temperature: 10 to 45 C
(50 to 113F)
Programmable Timer Range: 0 to 480
minutes
Heating Rate: 20 to 150C in 10
minutes
Temperature Stability: 1C
Stored Programs
COD 150C 120 minutes
TOC 105C 120 minutes
Fixed 100, 105, 30, 60,
150, 165C 120 minutes
Power Requirements:
100 to 240 VAC, 50/60 Hz, 600 VA
Compliance: CE, GS, CTUVUS
Dimensions:
1326914

25.0 x 14.5 x 31.0 cm (9.8 x 5.7 x 12.2
in.)
Number of Vial Wells
Single block: 9 16-mm and 2 20-mm or
15 16-mm
Total cost
The equipment should be supplied with
following regents.
b) Reagent for Chloride Testing-
200 Test
c) Reagent for Sulphate Testing-
200 Test
d) Reagent for COD Testing-
300 Test Range- 0 to 1500mg/l
Warranty:- 1 Year
Training Include
VAT ( +5%) Total App. Cost*
(2012)
3348.75 70,324
CIVIL ENGG. DEPARTMENT
ENVIRONMENTAL ENGINEERING LAB.
5082.5 106,733
6911.25 145,136
2820.7 59,235
4750 99,750
10370 217,770
21666 454,986
11396.5 239,327
1393259.7