Beruflich Dokumente
Kultur Dokumente
The aim of this laboratory is to understand the importance of statistical distributions in the
operation of measurement equipment. Involves:
1. The measurement of the offset voltage and offset current from a large collection of
operation amplifiers
2. The analysis of the distribution of the measured data
3. The use of an operation amplifier to increase the signal level of a pressure sensor,
selecting the amplifier from the statistical population of amplifiers.
4. Calibration of the pressure sensor
PART ONE
MATERIALS USED IN THE EXPERIMENT
The operational amplifiers; LF353, TL072, and LM358.
Op- amp tester: to measure the offset voltage and the offset current
A Multimeter: used for measuring the voltage output
METHOD USED IN THE EXPERIMENT
Connect the Multimeter to the banana plug and set the meter to the voltage range
required.
The operational amplifiers used in this lab have amplifiers on two physical sides on
the chip (the left and the right)
The Switch S4 (in Figure 1) selects either the left side or the right side of an
operational amplifier. Once the left (or right) opamp is selected using Switch S4,
switch S1 on the left (or S2 on the right) selects the offset voltage or the offset
current. When the offset voltage is selected, the output voltage can then be
measured at the banana plug.
SAMPLE CALCULATIONS
To calculate the voltage offset we used this calculation when the offset voltage is selected
V
0
= -(
) V
offset
= -100V
offset
Example: for #1 :
V Therefore
0.4960mV
To calculate the offset current when the offset voltage and offset current is selected, the
output voltage (V
0
) is given by
) (
; Example: for #1
V,
0.4960mV
Therefore
=
= 2.4530A
The mean x =
The Gaussian distribution f(z)=
=
453
METHOD USED FOR CALIBRATION
Connect the circuit correctly as indicated as indicated in figure 7 using the best and
the worst voltage gotten from the LF353 op-amp.
We selected a long tube and measured the length of the tube as L1=1930mm and
the inner diameter of the tube D1=4mm the volume = 24.253cm
3
We connected one end of the tube to one sensor inlet and measured voltage V1
we filled the syringe with water and then connect it to the other end of the tube
and measure voltage V2
We marked the length of the rise of the water in the tube and placed the tube
horizontally along a ruler and measure the length of water pushed into the tube
at various increments (by pushing slowly on the syringe)
Measure the voltage V(L) at each value of water length L pushed into the tube
We determined the pressure change in the tube which results from the
compressed volume of air in the tube
SAMPLE CALCULAIONS
To calculate pressure we use P
0
V
0
= P
1
V
1
; P
0
= 1atm =101.3kPa ; V
0
=volume of the tube
V
water
=
; V
1
= V
0
- V
water
; P
1
=
Gauge pressure (pressure differential): P
gauge
=P
1
-P
0
To determine the output voltage from the the data sheet we used the formula
V
ouput
(mV) =
mV/KPa
Hence the sensitivity is obtained as 0.2426 mV/KPa using the best op-amp meanwhile the
sensitivity given in datasheet is typically at 0.4 mV / Kpa which is almost two times higher
than our group`s experimental result. This shows that the typical pressure sensor is more
sensitive. The sensitivity of the pressure sensor was not that accurate when compare to the
datasheet.
Problems observed with the calibration method
It wasnt easy using water as there was bound to be the leakage of water when it was put
into the tube.
Statistical uncertainty in parameters affects a measurement
There are several uncertainties due to the number of readings that were taken
Uncertainty due to the conditions of pressure, temperature , humidity and any drift
since factory calibration
There was also uncertainty associated with the calibration which is from the
manufacturers data
The uncertainty of input resistance may change the voltage gain of an op-amp and
thus brings systematic errors to the measurement with different amplifiers. Also
may increase the noise to the output signal.
There could be Uncertainty in the Common-Mode Rejection Ratio leading to a very
minimal reading in the output voltage
There was error due to resolution which is as a result of the readability of the data
Uncertainty as a result of the sensitivity to the changes in conditions of the room
However, we tried to minimize the Type B error by not putting it in autoranging and also by
using the same multimeter to measure the values
Contribution of each team member:
Thomas Aremu : the analysis of part one and the taking applying the pressure during the
lab he was involved in typing the report was his contribution
Buxin Teng : Set up circuits, reading the multimeter and recording in the lab together with
and carrying out calculations and analyzing part two was his contribution. We both had
equal percentage contribution
REFERENCES
Adrian keating (2013), Measurement and Noise, course note, University of western
Australia
Operational amplifiers, last updated 7/08/2013
http:///en.wikipedia.org/wiki/Operational_amplifiers last viewed 28/09/2013
APPENDIX