Beruflich Dokumente
Kultur Dokumente
www.opticianonline.net
32
c o n t i n u i n g e d u c a t i o n
M , N V Optician
www.opticianonline.net
33
c o n t i n u i n g e d u c a t i o n
GLAUCOMA IS ONE of the worlds
leading causes of blindness. Intraocular
pressure (IOP) is an important criterion
for decision-making in both the screening
and management of this disease.
In practice, measurement of the IOP
is by tonometry. The force required to
applanate or flatten the outer coat of the
eye is the basis of applanation tonometry.
A true measure of IOP can only be
obtained using manometric methods
which require a probe to be inserted
directly into the anterior chamber.
Tonometry has evolved a long way
from the days of digital palpation and
the indentation tonometers, developed
in the late 1800s
1
(see Figure 1). Most
instruments now use the basic principle of
applanation tonometry, a method derived
from the Maklakov-Fick law, alternatively
known as the Imbert-Fick law.
IMBERT-FICK LAW
The Imbert-Fick law states that pressure
is equal to the force per unit area of
applanation for a spherical container,
assumed to be infinitely thin, dry and
perfectly elastic in form.
The cornea is, however, a rigid structure
which has an average thickness of 540m,
and a tear-moistened outer surface.
Nevertheless, by selecting a circular
zone 3.06mm in diameter the smallest
area of applanation (7.35mm
2
) is attained
while still giving accurate results,
introducing a relative error in IOP of only
2.5 per cent.
This is due to the balance between
the four forces acting on the cornea on
applanation:
Force of probe + Tear capillary action
= IOP + Ocular rigidity
Furthermore, once the density of
mercury (13.6g/cm
3
) is considered in the
final calculation, an easy conversion is
produced in which 1g of weight equates to
10mmHg of pressure.
Contact applanation
tonometry
Successful participation in each module
counts as one credit towards the GOC CET
scheme administered by Vantage and one
towards the AOIs scheme.
Priya Dabasia reviews the correct use of contact tonometers
and some modern contact and non-contact alternatives
(Module C3467, one standard point)
CONTACT APPLANATION
TONOMETRY
The two main contact applanation
tonometers used in the UK are the
Goldmann and the Perkins, which
use the basic principle of varying the
force applied to applanate a fixed area.
The basic instrumentation consists of a
flat applanating cone comprised of two
prisms with apices joined together. Its
main advantage over the first mechanical
indentation designs is the reduction
in induced aqueous outflow known as
ocular massage. Research has shown
there to be a maximum volume displace-
ment of 0.50mm
3
which has minimal
effects on the IOP.
Goldmann tonometry
This technique should only be performed
on completion of all other clinical tests
including refraction, as softening of
the cornea by anaesthesia in addition
to potential corneal abrasions can
cause blurred vision. Only dilation is
recommended post-Goldmann as the
softening action permits increased
absorption of the mydriatic across the
corneal surface.
1) The instrument should be calibrated
regularly depending on usage. The
Goldmann is mounted onto the slit lamp
with the cone in position. A calibration
key comprising a metal rod with three
grooves is inserted into the main body
of the instrument (Figure 2). The central
groove of the key is initially aligned with
the marking of the attaching device; in
this arrangement the tonometer cone
should rock forward at a weight of 0g.
The key is then moved towards the
Applanation
Indentation
(eg Schiotz)
Ocular
Response
Analyser (ORA)
Pascal DCT
Tonometers
Contact
Non-contact (eg
NT-3000, Pulsair)
Fixed weight, variable
area (eg Maklakoff)
Fixed area, variable
weight (Goldmann,
Perkins, Tonopen)
Dynamic Static
FIGURE 1. Summary of tonometer types
Optician M , N V
www.opticianonline.net
32
c o n t i n u i n g e d u c a t i o n
M , N V Optician
www.opticianonline.net
33
c o n t i n u i n g e d u c a t i o n
practitioner aligning the second (2g) and
third (6g) groove, repeating the same
procedure in turn. The instrument must
be sent for re-calibration if any non-linear
error or a linear error outside +/- 2.5mmHg
is recorded.
4
2) Ascertain any contraindications to
contact tonometry such as previous adverse
reactions to anaesthetic drops. The practi-
tioner must wash their hands thoroughly
using antibacterial soap. Further prelimi-
nary checks include the assessment of
visual acuity and corneal integrity, using
the slit lamp biomicroscope. Initially, a
wide parallelopiped with moderate to high
magnification is used, followed by a more
detailed examination using fluorescein
under cobalt blue light. Contact tonometry
is inadvisable if any signs of infections or
abrasions are found, as direct contact with
the eye inevitably risks further damage
even when conducted by an experienced
practitioner.
3) Non-disposable tonometer cones
must be disinfected according to College
of Optometrists or ABDO guidelines to
prevent any risk of transmitting HIV,
vCJD or Hepatitis B. All have been
isolated in the tear film. It is advisable
to rinse the disinfected cone with saline,
before leaving it to air dry to prevent
any residual disinfectant causing a
toxic reaction. Alternatively, disposable
Tonosafe cones are available to use
with both the Goldmann and Perkins
tonometers, although at increased
expense (Figure 3).
4) The tonometer probe is inserted into
the main instrument body with prism
aligned horizontally, taking care not
to touch the sterile cone surface. This
can either be approximated visually, or
checked by aligning the white line of
the probe carrier with the 0/180 marking
of the body when using non-disposable
cones. However, if the corneal astigmatism
is 3DC or above, an elliptical applanation
area larger than 7.35mm
2
is produced,
leading to an IOP error of at least
1mmHg.
5
This is avoided by rotating the
prism 43 to the flattest corneal meridian.
Since the separation between white and
red markings of the probe carrier is 43,
the red marking simply needs to be
aligned with the prism for with the rule
or against the rule astigmatism.
5) One drop of anaesthetic (such as
benoxinate 0.4 per cent, or a combina-
tion such as proxymetacaine 0.5 per cent
with fluorescein) is instilled into the lower
conjunctival sac. Allow approximately 30
seconds for the drops to take effect, taking
this time to record the drug name, expiry
CONTACT APPLANATION TONOMETRY
FIGURE 2. Calibration rod
positioned such that
6g of force is needed
to move the probe
forward
Optician M , N V
www.opticianonline.net
34
c o n t i n u i n g e d u c a t i o n
M , N V Optician
www.opticianonline.net
35
c o n t i n u i n g e d u c a t i o n
date, dose, batch number and time of
instillation on the record card. The
corneal sensitivity can be checked using
either a sterile wisp of tissue or preferably
a wetted fluoret strip at the limbal edge.
6) The Goldmann is mounted onto the slit
lamp via a foot plate with two grooves. The
procedure is conducted through only one
eyepiece depending on whether the left or
right groove is used. Setting the base into
the nasal groove is advised as it compen-
sates for convergence of the eye. Initially,
the force is set to an average reading
13mmHg,
6
as this reduces contact time
with the cornea and prevents vibration of
the probe. The slit lamp is set to maximise
the light illuminating the cone head:
Wide angle of illumination (60)
High illumination to account for light
absorbance through cobalt blue filter
Widest beam width and height
Low to moderate magnification (10-
16X).
7) Applanation is initially attempted
without manipulating the lids as the
patient stares at a fixation target straight
ahead. If this is unsuccessful, the practi-
tioners free hand is used to carefully
part the lids using the forefinger to hold
the upper lid against the superior orbital
ridge and the thumb to draw the lower
lid down, taking care not to press on the
globe and cause a change in IOP.
8) The slit lamp is carefully advanced to
bring the cone within 2-3mm of the central
cornea. Final contact is made using the
joystick for fine control. This will cause a
slight backward tip of the cone and create
FIGURE 3. Tonosafe
disposable cone in
position
FIGURE 4. Goldmann dial this should be
set as close to the expected measurement
as possible to minimise movement during
contact (in this case 1.6g)
CONTACT APPLANATION TONOMETRY
FIGURE 5. Schematic
representation
of a variety of
appearances of the
meniscus rings; (a)
the ideal alignment,
(b) more force
needed, (c) too much
force used, reduce
force, (d) move cone
upward, (e) move
cone downward (f)
wait for fluorescein
to drain
(a)
(b)
(c)
(d)
(e)
(f)
Optician M , N V
www.opticianonline.net
34
c o n t i n u i n g e d u c a t i o n
M , N V Optician
www.opticianonline.net
35
c o n t i n u i n g e d u c a t i o n
ment) with cone in position. Initially,
calibration is checked with no weight in
place (0g), and then free weights of 2g
and 5g are placed in turn onto the cone
head. Correct calibration results in the
cone rocking forwards on reaching the
equivalent reading on the dial.
2) Follow Steps 2, 3 and 5 as for
Goldmann procedure. Perkins can be
conducted with the patient supine or
seated upright. Extend the instrument
headrest setting it firmly on the patients
forehead using the two middle fingers of
your free hand. Use the thumb of this
hand to lift the patients upper lid as
required (Figure 6).
3) Advance the cone carefully towards the
eye, with the patient fixating a target just
above the straight ahead position. Ensure
the rings are the correct thickness and
equal in size, making any minor adjust-
ments off the cornea, before reaching the
final endpoint using the same principles
as Goldmann (see Step 8).
4) Follow post-applanation Steps 9 and
10 as with Goldmann.
NON-CONTACT TONOMETRY
Although Goldmann tonometry
is considered the gold standard
tonometer, most general practices in the
CONTACT APPLANATION TONOMETRY