ECG Circuit Analysis and Design
Posted on: 01-11-2012 Posted in: PROJECTS
Abstract Of ECG circuit by:
1- Abdallah Ishbeata.
2- Mohammad Kalbouneh.
The aim of the project is to design ECG circuit that pick-up the signal from the heart using electrodes
actually we simulate the circuit in simulation software to sure that the circuit can work practically, and we
design circuit that detect any failure in the electrodes connection.
The main challenges include amplifying the desired weak signal in the presence of noise from other
muscles and electrical source so there is requirement needed in amplifier circuit to avoid any problem
may occur during the measurement I will mention it later.
ACKNOWLEDGMENTS
We cannot find the words that can covey the depth of our feeling towards many people who helped us.
All what we can say thanks very much and we will always remember and be grateful for them. This work
would have never come to be without the guidance of our supervisors and the support of our families.
We would like to express our deep thanks to DR. AMJED AL FAHOUM for Supervisions.
Last, we would like to express our appreciation and deep gratitude to our parents and friends for their support
and encouragement.
Chapter one
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 2/29
Physiological Fundamentals
1.1 Heart beat
Is the number of heartbeats per unit of time typically expressed as beats per minute (bpm) which can vary as
the bodys need to absorb oxygen and excrete carbon dioxide changes, such as during exercise or sleep, The
measurement of heart beat is used by medical professionals to assist in the diagnosis and tracking of medical
conditions. It is also used by individuals, such as athletes, who are interested in monitoring their heart beat to
gain maximum efficiency from their training.
1.2 Heart and Functions:
The heart is the organ responsible for pumping blood throughout the body. It is located in the middle of the
thorax, slightly offset to the left and surrounded by the lungs. The heart is composed of four chambers; two
atriums and two ventricles. The right atrium receives blood returning to the heart from the whole body. That
blood passes through the right ventricle and is pumped to the lungs where it is oxygenated and goes back to the
heart through the left atrium, then the blood passes through the left ventricle and is pumped again to be
distributed to the entire body through the arteries.
This is a list of events that occur in the heart on each heart beat. Figure1 shows heart behavior and part of the
generated signal.
Also known as QRS complex:
1. Atrium begins to depolarize.
2. Atrium depolarizes.
3. Ventricles begin to depolarize at apex. Atrium re polarizes.
4. Ventricles depolarize.
5. Ventricles begin to re polarize at apex.
6. Ventricles re polarize.
Figure 1- Electrical Activity of the Heart.
Figure 2 shows a typical heart signal. In this signal, the heart muscles generate different voltages. The P wave
represents the atrium contraction. QRS complex and the T wave represents the ventricles actions. Each time that
this signal is present, a heart beat is generated.
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 3/29
Figure 2.Typical heart signal.
Chapter two
ECG circuit design
2.1 Introduction:
The first stage of the ECG circuit include instrumentation amplifier it is the most important part in the circuit it
should provide high gain to amplify the weak of ECG signal and be able immunity the noise (common mode
signal ) and other signal in electromagnetic spectrum.
2.2 Noise:
Noise from the environment will easily swamp the tiny pulse signal from the heart. The leads connecting the
electrode to the amplifier will act like an antenna which will inadvertently receive unwanted radiated signals.
Such signals are for example the 50Hz from power lines and emfs from fluorescent lights will add a tiny
sinusoidal wave which is generally quite difficult to filter away, but in our project we will not concern this type
of noise (50Hz) since our range of signal is from 0.5-5 Hz. Noise and interference signals acquired in this type
of system are caused by the electric installation. The signals from the heart are too small and it is necessary to
amplify the signal and reduce the common-mode voltage on the system. Other aspects that generate noise are
muscle contractions, respiration, and electromagnetic emissions from electronic components.
2.3 Enhancements:
To address the issues above, the following measure will be taken:
A high gain instrumentation amplifier with a high Common Mode Rejection Ratio (CMRR) will be used to
receive the desired signal.
A band pass filter will be implemented to remove the noise. Because most of the noise types discussed are of
high frequency while the desired signal is relatively low.
Peak detection circuit to detect the failure in electrodes connection.
Oscillator to generate signal with frequency approximately 5okhz that pass through the instrumentation
amplifier when the loss connection occurred.
Analog to digital circuit to processing the signal using computer techniques.
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 4/29
Figure 3: block diagram of the circuit
2.4 instrumentation amplifier (INA 128):
The signal acquisition is the first consideration when an HRM is implemented. But the signal is too small and
contains a lot of added noise. As we said above the signal extracted from the heart has amplitude of
approximately 0.5mV.
Since, it is necessary to amplify the signal and filter the noise, and then extract the QRS complex.
An instrumentation amplifier is usually the very first stage in an instrumentation system. This is because of the
very small voltages usually received from the probes need to be amplified significantly to be proceeding stages.
We can summarize the reasons to use instrumentation amplifier:
1- Get differential signal.
2- High input impedance.
3- High CMRR.
Let us take some review about instrumentation amplifier:
Figure 4: general instrumentation op amp.
This intimidating circuit is constructed from a buffered differential amplifier stage with three new resistors
linking the two buffer circuits together. Consider all resistors to be of equal value except for R
gain
. The negative
feedback of the upper-left op-amp causes the voltage at point 1 (top of R
gain
) to be equal to V
1
. Likewise, the
voltage at point 2 (bottom of R
gain
) is held to a value equal to V
2
. This establishes a voltage drop across
R
gain
equal to the voltage difference between V
1
and V
2
. That voltage drop causes a current through R
gain
, and
since the feedback loops of the two input op-amps draw no current, that same amount of current through
R
gain
must be going through the two R resistors above and below it. This produces a voltage drop between
points 3 and 4 equal to:
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 5/29
The regular differential amplifier on the right-hand side of the circuit then
takes this voltage drop between points 3 and 4, and amplifies it by a gain of 1 (assuming again that all R
resistors are of equal value). Though this looks like a cumbersome way to build a differential amplifier, it has
the distinct advantages of possessing extremely high input impedances on the V
1
and V
2
inputs (because they
connect straight into the non inverting inputs of their respective op-amps), and adjustable gain that can be set by
a single resistor. Manipulating the above formula a bit, we have a general expression for overall voltage gain in
the instrumentation amplifier:
2.5 GAIN SELECTION:
The INA128 and INA129 are low power, general purpose instrumentation amplifiers offering excellent
accuracy. Their versatile 3-op amp design and small size make them ideal for a wide range of applications.
Current-feedback input circuitry provides wide bandwidth.
Even at high gain (200 kHz at G = 100). A single external resistor sets any gain from 1 to10, 000. INA128
provides an industry standard gain.
Equation; INA129s gain equation is compatible with the AD620. The INA128/INA129 is laser trimmed for
very low offset voltage (50mV), drift (0.5mV/C) and high common- mode rejection (120dB at G 100). It
operates with power supplies as low as 2.25V, and quiescent.
Current is only 700mAideal for battery operated systems. Internal input protection can withstand up to40V
without damage. The INA128/INA129 is available in 8-pin plastic DIP, and SO-8 surface-mount packages,
specified for the 40C to +85C temperature range. The INA128 is also available in dual configuration, the
INA2128.
FEATURES:
LOW OFFSET VOLTAGE: 50mV max.
LOW DRIFT: 0.5mV/C max.
LOW INPUT BIAS CURRENT: 5nA max.
HIGH CMR: 120dB min.
INPUTS PROTECTED TO 40V.
WIDE SUPPLY RANGE: 2.25 to 18V.
LOW QUIESCENT CURRENT: 700mA.
8-PIN PLASTIC DIP, SO-8.
Figure 5: circuit design of INA 128.
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 6/29
From our design we choose RG= 1k since the gain we get 51 v/v actually the gain not very large for reason
that although the instrumentation amplifier has high common mode rejection ratio but the noise still effect to
the output of the circuit according to this equation:
so this signal that depends on the frequency that will pass through instrumentation op amp.
Figure 6: voltage noise VS frequency.
2.6 filtering stage:
The required band width for ECG signal (0.5 hz- 30 Hz) for normal heart human so we chosen the bandwidth of
the circuit near to this range, now if we choose the bandwidth (0.5-120Hz) notching filter required in design to
remove 5ohz noise from power line grid .
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 7/29
Figure 7: band pass filter.
2.7 dc offset stage:
Actually we use dc offset because some component of (QRSTU) waves in negative portion so if we want to
convert our signal to digital form we need to make dc offset or we can use bi polar analog to digital converter.
Figure 8: dc offset circuit.
Figure 9: ECG signal after apply dc offset.
Chapter three
3.1 Peak detector circuit:
Peak detectors are used when you have a rapidly changing AC input signal, and you want to obtain the peak
voltage the signal reaches. Peak detectors are really simple to make just a diode and a capacitor in their
simplest form. In our design where I use a peak detector to hold on to a peak signal that generated from
oscillator which is appear when there is failure connection at the input of instrumentation amplifier at the first
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 8/29
stage, simply the peak detector circuit contains capacitor at the anode of diode terminal for charging process
and second capacitor at cathode terminal for discharging process, the charging and discharging process depend
on = RC .
The comparator compare between the voltage at second capacitor and threshold voltage to detect the peak of 50
kHz signal.
Figure 10: peak detector circuit.
3.2 testing the circuit in failure connection of electrodes:
According to the impotency of the circuit in medical application we need to detect any failure may occur during
diagnostic, the most important problem that the doctor needed to know if there is any loss in electrode
connection to solve it and take the best measurements from devices.
Figure 11: failure connection in right arm.
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 9/29
Figure 12: failure connection in right leg.
3.3 oscillator circuit:
Oscillator circuit is required to generate signal for current circuit to detect the failure in electrode connection the
amplitude of generated signal depends on power supply of the circuit, the frequency of generated circuit can
calculate it according to the values of resistors and capacitor in the circuit.
Note : We change the oscillator circuit at the simulation with this circuit at Figure 13 when we connected our
circuit that we designed on the bread board .
Figure 13: oscillator circuit .
Chapter four
4.1 ECG signal simulator:
SPICE simulation, the language by which Multisim emulates circuit design behavior, does not run in real-
time. This means that if a real-world signal is acquired by a Lab VIEW instrument it cannot be directly injected
into simulation, since the measurements will be running at different rates (real time vs. simulated time).
In the growing world of Biomedical engineering, the need to be able to quickly design circuitry which
interfaces to a human signal is a common design problem. An amplifier for example can be designed in a circuit
simulation package, however the signals which will interface to that particular circuit will be simple simulation
stimuli (such as a sine wave, square wave etc). To truly test a biomedical amplifier, you need to be able to
interface the design to a real signal.
Using this unique Lab VIEW instrument you are able to define a human electrocardiogram (ECG) signal, and
have amplified it through a circuit created in Multisim.
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 10/29
Figure 14: ECG signal
4.2 steps of simulation:
This Lab VIEW instrument, built custom for NI Multisim produces a raw electrode ECG waveform. Since this
instrument is built in Lab VIEW we have a number of analysis, and signal altering functions available to us. In
this case we are able to add common mode voltage, as well as noise components to alter the signal to improve
the test and validation of our circuit.
Download the archived file to the desktop from Here
1- Close NI Multisim if open.
2- Open the attached 5925_ECG Amplifier.zip file.
3- Select My Computer.
4- Browse to C:\Program Files\National Instruments\Circuit Design Suite 10.1\lvinstruments.
5- Save the ECG Electrode.llb file from the zip file to this folder.
4.3 using the instrument ECG signal:
This instrument will use the example circuit that is in the attached 5925_ECG Amplifier.zip file folder.
1- Open Multisim by selecting Start > All Programs > National Instruments > Circuit Design Suite
10.1 > Multisim 10.1.
2- Go to File > Open.
3- Browse to the ECG Amp with ECG Signals.ms10 file from the 5925_ECG Amplifier.zip file.
4- You will see a complete amplifier based upon two common Op-Amps.
5- In Multisim select the Lab VIEW icon in the instrument toolbar (as seen below).
6- Select ECG Signals. This is the name of the ECG Electrode instrument (as seen below).
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 11/29

Figure 15: using ECG signal.
Appendix:
Sheet 1: over all circuit.
Part 1: ECG circuit and failure detection.
Sheet 2: over all results.
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 12/29
Sheet 3: 3D view.
References
1- (http://zone.ni.com/devzone/cda/epd/p/id/5925)
2- Wikipedia (ecg circuit )
3- All datasheet site download (www. Alldatasheet.com)
4- JOGN G.WEBSTER ,Medical Instrumentation, Application And Design.
5- http://www.analog.com/library/analogDialogue/archives/37-11/ecg.html.
6- Jessica ambourn, Portable ECG Logger, October 2003.
7- Leece Sofoklis Nikiforos, Heart Rate Monitor and Data Acquisition System
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 13/29
8- http://www.google.jo/search?sourceid=chrome&ie=UTF-8&q=ECG+ Instrumentation .
9- http://www.google.jo/search?sourceid=chrome&ie=UTF-8&q=Heart-
Rate+and+EKG+Monitor+Using+the+MSP430FG439
10- http://www.google.jo/search?sourceid=chrome&ie=UTF-8&q=Heart+Rate
11- http://www.news-medical.net/health/What-is-Heart-Rate.aspx.
12- http://www.topendsports.com/testing/heart-rate.htm.
13- http://en.wikipedia.org/wiki/Heart_rate.

Download this file ECG Circuit Analysis and Design
You can find the circuit , documentation and 5925_ecg_amplifier.
I hope you like this post
Thanks alot
About the Author
Ishbeata
A Electronics Engineer student Princess Sumya University in Jordan , seeking knowledge in Electronic Fields .
Popular Posts
Related Posts
Arduino Talks to Android application Via Bluetooth
ECG Circuit Analysis and Design
Electronic Switch By Mobile
Make a power supply at your home [ 0 to 24 volts ]
(6) Comments
(0) Trackbacks
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 14/29
1.
Bo Parshotam02-29-12
Thanks for helping out, good information.
(reply)
2.
Advanced Electrolysis Sydney03-03-12
Its truly a nice and helpful piece of info. Im satisfied that you shared this helpful information with us.
Please keep us informed like this. Thank you for sharing.
(reply)
3.
kataskevh04-01-12
That is a great tip particularly to those new to the blogosphere. Simple but very accurate information
Appreciate your sharing this one. A must read post!
(reply)
4.
EE/ME10-14-12
An excellent report! Well researched and very well presented.
(reply)
5.
haruka14102-22-13
Thanks information ^_^
(reply)
6.
Phan Hoai Nam06-25-13
thank you so much!
(reply)
Leave a Reply
Name *
Email *
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 15/29
Website
Submit Comment
Popular Posts
Recent Posts
Arduino Talks to Android application Via Bluetooth
05-20-2013
ECG Circuit Analysis and Design
01-11-2012
Electronic Switch By Mobile
11-17-2011
This online site is dedicated to you. Created in order to provide a place where you can post your questions,
ideas and comments around electronics.
2010 ENGINEERS LABS. All Rights Reserved
About
Contact Us
Tutorials
Projects
Courses
PIC
Expand
next previous
Close
Previous
0/0
Next
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 16/29
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 17/29
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 18/29
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 19/29
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 20/29
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 21/29
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 22/29
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 23/29
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 24/29
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 25/29
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 26/29
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 27/29
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 28/29
30/10/2014 ECG Circuit Analysis and Design Engineers Labs
http://engineerslabs.com/2012/01/ecg-circuit-analysis-and-design-simulation-by-multisim/ 29/29

Das könnte Ihnen auch gefallen