Modbus

Interface Description

SCADA
Supervisory Control and
Data Acquisation

Ethernet

Webserver

Modbus RTU

A-96.250.751 / 150312

Customer Support
SWAN and its representatives maintain a fully trained staff of technical specialists
around the world. For any technical question, contact your nearest
SWAN representative, or the manufacturer:
SWAN ANALYTISCHE INSTRUMENTE AG
Studbachstrasse 13
8340 Hinwil
Switzerland
Internet: www.swan.ch
E-mail: support@swan.ch

Document Status
Title:

AMI MODBUS Operators Manual

ID:

A-96.250.751

Revision

Issue

00

February 2012

First release

01

March 2012

Chapters AMI Phosphate-II and AMI pH/mV:pH/mV added

02

Sept. 2012

Chapter AMI Phosphate II updated

03

Mai 2013

Chapters AMI ISE and AMI Solicon4 added.


Chapter AMI Oxygen splitted into AMI Oxysafe & AMI Oxytrace

2012, SWAN ANALYTISCHE INSTRUMENTE AG, Switzerland, all rights reserved


subject to change without notice

AMI Profibus

1.
1.1.
1.2.

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Query-Response Cycle . . . . . . . . . . . . . . . . . . . . . . .
Data and Control Functions . . . . . . . . . . . . . . . . . . . . . . .

5
6
6

2.
Product Specific Data for AMI Codes-II . . . . . . . . . . . .
2.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.2.1
Read Input Status (Alarm &Status Bits), Function 02 .
2.2.2
Read Input Registers (Process data): Function 04 . . .
2.2.3
Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
2.2.4
Read Holding Registers (Parameter data): Function 03

9
9
9
9
10
11
11

3.
Product Specific Data for AMI Codes-II CC . . . . . . . . .
3.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.2.1
Read Input Status (Alarm &Status Bits), Function 02 .
3.2.2
Read Input Registers (Process data): Function 04 . . .
3.2.3
Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
3.2.4
Read Holding Registers (Parameter data): Function 03

13
13
13
13
14
15
15

4.
Product Specific Data for AMI Codes-II TC . . . . . . . . .
4.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.2.1
Read Input Status (Alarm &Status Bits), Function 02 .
4.2.2
Read Input Registers (Process data): Function 04 . . .
4.2.3
Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
4.2.4
Read Holding Registers (Parameter data): Function 03

17
17
17
17
18
19
19

5.
Product Specific Data for AMI Deltacon. . . . . . . . . . . .
5.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.2.1
Read Input Status (Alarm &Status Bits), Function 02 .
5.2.2
Read Input Registers (Process data): Function 04 . . .
5.2.3
Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
5.2.4
Read Holding Registers (Parameter data): Function 03

21
21
21
21
22
23
23

6.
Product Specific Data for AMI Hydrazine. . . . . . . . . . .
6.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.2.1
Read Input Status (Alarm &Status Bits), Function 02 .
6.2.2
Read Input Registers (Process data): Function 04 . . .
6.2.3
Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
6.2.4
Read Holding Registers (Parameter data): Function 03

25
25
25
25
26
26
27

7.
Product Specific Data for AMI Hydrogene . . . . . . . . . .
7.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7.2.1
Read Input Status (Alarm &Status Bits), Function 02 .
7.2.2
Read Input Registers (Process data): Function 04 . . .
7.2.3
Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
7.2.4
Read Holding Registers (Parameter data): Function 03

29
29
29
29
30
31
31

8.
Product Specific Data for AMI Inducon . . . . . . . . . . . .
8.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.2.1
Read Input Status (Alarm &Status Bits), Function 02 .
8.2.2
Read Input Registers (Process data): Function 04 . . .
8.2.3
Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
8.2.4
Read Holding Registers (Parameter data): Function 03

33
33
33
33
34
34
35

A-96.250.751 / 150312

AMI Profibus

9.
Product Specific Data for AMI ISE Universal . . . . . . . .
9.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9.2.1
Read Input Status (Alarm &Status Bits), Function 02. .
9.2.2
Read Input Registers (Process data): Function 04. . . .
9.2.3
Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
9.2.4
Read Holding Registers (Parameter data): Function 03

37
37
37
37
38
38
39

10.
Product Specific Data for AMI LineTOC . . . . . . . . . . . .
10.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10.2.1 Read Input Status (Alarm &Status Bits), Function 02. .
10.2.2 Read Input Registers (Process data): Function 04. . . .
10.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
10.2.4 Read Holding Registers (Parameter data): Function 03

41
41
41
41
42
42
43

11.
Product Specific Data for AMI Oxysafe . . . . . . . . . . . .
11.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
11.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
11.2.1 Read Input Status (Alarm &Status Bits), Function 02. .
11.2.2 Read Input Registers (Process data): Function 04. . . .
11.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
11.2.4 Read Holding Registers (Parameter data): Function 03

45
45
45
45
46
47
47

12.
Product Specific Data for AMI Oxytrace. . . . . . . . . . . .
12.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
12.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
12.2.1 Read Input Status (Alarm &Status Bits), Function 02. .
12.2.2 Read Input Registers (Process data): Function 04. . . .
12.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
12.2.4 Read Holding Registers (Parameter data): Function 03

49
49
49
49
50
51
51

13.
Product Specific Data for AMI Pharmacon . . . . . . . . .
13.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
13.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
13.2.1 Read Input Status (Alarm &Status Bits), Function 02. .
13.2.2 Read Input Registers (Process data): Function 04. . . .
13.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
13.2.4 Read Holding Registers (Parameter data): Function 03

53
53
53
53
54
54
55

14.
Product Specific Data for AMI Solicon4 . . . . . . . . . . . .
14.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
14.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
14.2.1 Read Input Status (Alarm &Status Bits), Function 02. .
14.2.2 Read Input Registers (Process data): Function 04. . . .
14.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
14.2.4 Read Holding Registers (Parameter data): Function 03

57
57
57
57
58
58
59

15.
Product Specific Data for AMI Phosphate . . . . . . . . . .
15.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
15.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
15.2.1 Read Input Status (Alarm &Status Bits), Function 02. .
15.2.2 Read Input Registers (Process data): Function 04. . . .
15.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
15.2.4 Read Holding Registers (Parameter data): Function 03

61
61
61
61
62
62
63

16.
Product Specific Data for AMI Phosphate-II . . . . . . . .
16.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
16.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
16.2.1 Read Input Status (Alarm &Status Bits), Function 02. .
16.2.2 Read Input Registers (Process data): Function 04. . . .
16.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
16.2.4 Read Holding Registers (Parameter data): Function 03

65
65
65
65
66
67
67

A-96.250.751 / 150312

AMI Profibus

17.
Product Specific Data for AMI pH-Redox. . . . . . . . . . .
17.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
17.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
17.2.1 Read Input Status (Alarm &Status Bits), Function 02 .
17.2.2 Read Input Registers (Process data): Function 04 . . .
17.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
17.2.4 Read Holding Registers (Parameter data): Function 03

69
69
69
69
70
70
71

18.
Product Specific Data for AMI pH/mV:pH/mV . . . . . . .
18.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
18.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
18.2.1 Read Input Status (Alarm &Status Bits), Function 02 .
18.2.2 Read Input Registers (Process data): Function 04 . . .
18.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
18.2.4 Read Holding Registers (Parameter data): Function 03

73
73
73
73
74
74
75

19.
Product Specific Data for AMI Powercon. . . . . . . . . . .
19.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
19.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
19.2.1 Read Input Status (Alarm &Status Bits), Function 02 .
19.2.2 Read Input Registers (Process data): Function 04 . . .
19.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
19.2.4 Read Holding Registers (Parameter data): Function 03

77
77
77
77
78
78
79

20.
Product Specific Data for AMI Rescon. . . . . . . . . . . . .
20.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
20.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
20.2.1 Read Input Status (Alarm &Status Bits), Function 02 .
20.2.2 Read Input Registers (Process data): Function 04 . . .
20.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
20.2.4 Read Holding Registers (Parameter data): Function 03

81
81
81
81
83
83
84

21.
Product Specific Data for AMI Silica . . . . . . . . . . . . . .
21.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
21.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
21.2.1 Read Input Status (Alarm &Status Bits), Function 02 .
21.2.2 Read Input Registers (Process data): Function 04 . . .
21.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
21.2.4 Read Holding Registers (Parameter data): Function 03

87
87
87
87
88
89
89

22.
Product Specific Data for AMI Soditrace . . . . . . . . . . .
22.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
22.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
22.2.1 Read Input Status (Alarm &Status Bits), Function 02 .
22.2.2 Read Input Registers (Process data): Function 04 . . .
22.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
22.2.4 Read Holding Registers (Parameter data): Function 03

91
91
91
91
92
93
93

23.
Product Specific Data for AMI Sodium. . . . . . . . . . . . .
23.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
23.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
23.2.1 Read Input Status (Alarm &Status Bits), Function 02 .
23.2.2 Read Input Registers (Process data): Function 04 . . .
23.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
23.2.4 Read Holding Registers (Parameter data): Function 03

95
95
95
95
96
97
97

24.
Product Specific Data for AMI Trides . . . . . . . . . . . . . .
24.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
24.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
24.2.1 Read Input Status (Alarm &Status Bits), Function 02 .
24.2.2 Read Input Registers (Process data): Function 04 . . .
24.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
24.2.4 Read Holding Registers (Parameter data): Function 03

99
99
99
99
100
101
101

A-96.250.751 / 150312

AMI Profibus

25.
Product Specific Data for Turbidity . . . . . . . . . . . . . . .
25.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
25.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
25.2.1 Read Input Status (Alarm &Status Bits), Function 02. .
25.2.2 Read Input Registers (Process data): Function 04. . . .
25.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
25.2.4 Read Holding Registers (Parameter data): Function 03

103
103
103
103
104
105
105

A-96.250.751 / 150312

Modbus
Introduction

1.

Introduction
AMI Transmitters can be operated as intelligent slave devices on the Modbus/JBUS using
twisted pair RS485 lines.

Modbus/Jbus

The Modbus/Jbus Protocol communicates using a master-slave technique, in which only one
device (the master) can initiate transactions (queries). The other devices (the slaves) respond by
supplying the requested data to the master, or by taking the action requested in the query.
Typical master devices include host processors, PLCs etc. Typical slaves can be PLCs or
intelligent transmitter/controller like the AMI transmitters.
Instruments which only uses a subset of the available data from another instrument, are based
on the same data protocol.
For example, the AMI Deltacon DG and the AMI Deltacon Power provide exactly the same data,
but the AMI Deltacon Power does not provide any data concerning the degasser.
These instruments and the belonging chapters are listed in the following table:
Instruments based on the same data protocol.

See chapter

AMI Deltacon DG and AMI Deltacon Power

Product Specific Data for AMI Deltacon

AMI Oxysafe and AMI Oxytrace

Product Specific Data for Oxygen

AMI Sodium A and AMI Sodium P

Product Specific Data for AMI Sodium

AMI Turbiwell, AMI Turbitrace and AMI Turbitrack

Product Specific Data for Turbidity

AMI Phosphate-II and AMI Phosphate-II B

Product Specific Data for


AMI Phosphate-II

AMI pH/mV:pH/mV and AMI pH/mV:pH/mV Pool

Product Specific Data for


AMI pH/mV:pH/mV

A-96.250.751 / 140312

Modbus
Introduction

1.1.
The Query

The Response

Serial
Transmission
modes
RTU Mode

Coding
System

The Query-Response Cycle


The function code in the query tells the addressed slave device what kind of action to perform. The
data bytes contain any additional information the slave will need to function. For example, function
code 03 will query the slave to read holding registers and respond with their contents. The data field
must contain the information telling the slave which register to start at and how many registers to read.
If the slave makes a normal response, the function code in the response is an echo of the function
code in the query.
The data bytes contain the data collected by the slave.
Modbus supports two transmission modes: ASCII or RTU.
The AMI Instruments support Modbus RTU (Remote Terminal Unit) mode only.
In RTU mode, each 8-bit byte in a message contains two 4-bit hexadecimal characters. The main
advantage of this mode is that its greater character density allows better data throughput than ASCII
for the same baud rate. Each message must be transmitted in a continuous stream. Users select the
desired serial port communication parameters (baud rate, parity mode, etc). The serial parameters
must be the same for all devices on a Modbus network.
Eight bit binary
1 start bit
8 data bits, least significant bit sent first
1 bit for even / odd parity, no bit for no parity
1 stop bit if parity is used, 2 bits if no parity
Cyclic Redundancy Check (CRC)
RTU framing: messages start with a silent interval of at least 3 character times. A similar interval of at least 3 character times marks the end of the message.

Tab. 1-1 Typical message frame


Start

Slave
Address

Function

Data

CRC

END

T1-T2-T3-T4

8 Bits

8 Bits

n x 8 Bits

16 Bits

T1-T2-T3-T4

1.2.
Data address
in Modbus
Messages

Data and Control Functions


All data addresses in Modbus Messages are referenced to zero. The first occurrence of a data item is
addressed as item number zero. For example:
Holding Register 40001 is addressed as register 0000 in the data address field of the message. The function code field already specifies a holding register operation. Therefore the 4x
reference is implicit.

Modbus Function codes

The list below shows the function codes supported by the AMI Instruments. Codes are listed in
decimal. Function codes not supported return an exception response 02 to the master.
Tab. 1-2 Modbus function codes

Code

Name

Description

02

Read Input Status

Read Alarm & Status Bits

03

Read Holding Registers

Read Parameter Data

04

Read Input Registers

Read Process Data

05

Force Single Coil

Set Relays, take Keystrokes etc.

16

Preset Multiple Registers

Preset Parameter Data (SWAN use only)

A-96.250.751 / 140312

Modbus
Introduction

Data representation
Discrete
Inputs
Definition:
big-endian and
little-endian

The process- and parameter data of the AMI Instruments are represented by float (IEEE 754), signedor unsigned integer values. Depending on the range of the value, float, integer8, unsigned8,
integer16, unsigned16, integer32 or unsigned32 representation is used.
Alarm &Status bits are packed as one input per bit of the data field.
Big-endian and little-endian are terms that describe the order in which a sequence of bytes are stored
in computer memory. Big-endian is an order in which the "big end" (most significant value in the
sequence) is stored first (at the lowest storage address). Little-endian is an order in which the "little
end" (least significant value in the sequence) is stored first. For example, in a big-endian computer,
the two bytes required for the hexadecimal number 4F52 would be stored as 4F52 in storage. In a
little-endian system, it would be stored as 524F
ATTENTION
Big-endian format is used.

A-96.250.751 / 140312

Modbus
Introduction

A-96.250.751 / 140312

Modbus

Product Specific Data for AMI Codes-II

2.

Product Specific Data for AMI Codes-II

2.1.

Introduction
The AMI Codes-II is a complete monitoring system for the automatic, continuous measurement
and dosing control of chlorine and other disinfectants in potable water, swimming pools, cooling
water and effluent. It can be operated as an intelligent Slave device on the Modbus/JBUS using
twisted pair RS485 lines

2.2.

Function Description

2.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 32 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 240212

Addr.

Type

Description

Alarm

Dis. 1 Alarm high

Alarm

Dis. 1 Alarm low

Alarm

pH 1 Alarm high

Alarm

pH 1 Alarm low

Alarm

Dis. high

Alarm

Alarm

Sample Temp high

Alarm

Sample Temp low

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. shorted

11

Alarm

Temp. disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

Valve Defective

15

Alarm

16

Alarm

Control Timeout

17

Alarm

Reagent Pump

18

Alarm

Fotometer not connected

19

Alarm

Fotometer dirty

20

Alarm

Dis. invalid

21

Alarm

Reagent Empty

22

Alarm

Cleaning Reagent

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Cal. Recout

31

Alarm

Wrong Front-end

Modbus

Product Specific Data for AMI Codes-II

2.2.2

Addr.

Type

32

Alarm

33

Alarm

34

Alarm

35

Alarm

36

Alarm

37

Alarm

38

Alarm

39

Alarm

40

Alarm

Reagents low

41

Alarm

42

Alarm

43

Alarm

44

Alarm

45

Status

46

Status

47

Status

Description

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Codes-II.
Tab. 2-1
Byte

Input Registers
Data type

Description

Offset

10

Range of
physical value

Float

Disinfection

010 ppm

Float

pH

014 pH

Float

Temperature

12

Float

Flow

16

Float

Reserved

20

Float

Absorbance

24

Float

Raw Disinfection

28

Float

Raw pH

-30 to +120 C
0600 B/s
03
0300000 Hz
-5001500 mV
15066000 Ohm

32

Float

Raw Temp. (NTK5)

36

Float

Raw Flow

40

Float

Case Temperature

44

Float

Signal Output 1

020 mA

44

Float

Signal Output 2

020 mA

50

INSTRUMENT_STATUS

Instrument Display Status

0100 Hz
-55125 C

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Codes-II

2.2.3

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 2-2

2.2.4

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote Off

no

yes

15

Reset Instrument (like hardware reset)

reset

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 2-3
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History 1 Array

Array of CAL_HIST_1[15]

Calibration History Dis.

3001

1502

Cal_History 2 Array

Array of CAL_HIST_2[15]

Calibration History pH

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 240212

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

11

Modbus

Product Specific Data for AMI Codes-II

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish
4 = Italian

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

0100 %

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

12

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Codes-II CC

3.

Product Specific Data for AMI Codes-II CC

3.1.

Introduction
The AMI Codes-II CC is a complete monitoring system for the automatic, continuous
measurement and dosing control of chlorine and other disinfectants (e.g. monochloramine) in
potable water, swimming pools, cooling water, hot water loops and effluent. It can be operated as
an intelligent Slave device on the Modbus/JBUS using twisted pair RS485 lines.

3.2.

Function Description

3.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 48 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 240212

Addr.

Type

Alarm

Alarm high fac

Alarm

Alarm low fac

Alarm

Alarm high tc1

Alarm

Alarm low tc 1

Alarm

Alarm high tc 2

Alarm

Alarm low tc 2

Alarm

Sample Temp high

Alarm

Sample Temp low

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. shorted

11

Alarm

Temp. disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

Valve Defective

15

Alarm

Dis. invalid

16

Alarm

Control Timeout

17

Alarm

Reagent Pump

18

Alarm

Fotometer not connected

19

Alarm

Fotometer dirty

20

Alarm

Dis. invalid

21

Alarm

Reagent Empty

22

Alarm

Cleaning Reagent

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Calibration Recout

31

Alarm

Wrong Front-end

Description

13

Modbus

Product Specific Data for AMI Codes-II CC

3.2.2

Addr.

Type

Description

32

Alarm

Alarm high cmc

33

Alarm

Alarm low cmc

34

Alarm

Alarm high ccc

35

Alarm

Alarm low ccc

36

Alarm

Alarm high cdc

37

Alarm

Alarm low cdc

38

Alarm

Alarm high pH

39

Alarm

Alarm low pH

40

Alarm

DPD / Buffer

41

Alarm

Potassium Iodide

42

Alarm

Cleaning Solution

43

Alarm

44

Alarm

45

Status

46

Status

47

Status

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Codes-II CC.
Tab. 3-1
Byte

Input Registers
Data type

Description

Float

Free available chlorine (fac)

010 ppm

Float

Total chlorine 1 (tc1)

010 ppm

Float

Total chlorine 2 (tc2)

010 ppm

12

Float

Calculated monochloramine (cmc)

010 ppm

16

Float

Calculated combined chlorine (ccc)

010 ppm

20

Float

Calculated dichloramine (cdc)

010 ppm

24

Float

Temperature

28

Float

Flow

32

Float

pH Value

36

Float

Absorbance

40

Float

Photometer Frequency

44

Float

Remaining DPD/Buffer

0100 %

48

Float

Remaining Potassium Iodide

0100 %

52

Float

Remaining Cleaning Solution

0100 %

56

Float

Raw Temp. (NTK5)

60

Float

Raw Flow

64

Float

Raw pH

68

Float

Case Temperature

72

Float

Signal Output 1

020 mA

76

Float

Signal Output 2

020 mA

80

INSTRUMENT_STATUS

Instrument Display Status

Offset

14

Range of physical
value

-30 to +120 C
0100 B/s
014 pH
03
0300000 Hz

15066000 Ohm
0100 Hz
-5001500 mV
-55125 C

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Codes-II CC

3.2.3

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 3-2

3.2.4

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote Off

no

yes

15

Reset Instrument (like hardware reset)

reset

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 3-3
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History 1 Array

Array of CAL_HIST_1[15]

Calibration History Dis.

3002

1502

Cal_History 2 Array

Array of CAL_HIST_2[15]

Calibration History pH

3004

1503

Cal_History 3 Array

Array of CAL_HIST_3[15]

Vfy History Dis.

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 240212

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

15

Modbus

Product Specific Data for AMI Codes-II CC

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish
4 = Italian

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

0100 %

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

16

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Codes-II TC

4.

Product Specific Data for AMI Codes-II TC

4.1.

Introduction
The AMI Codes-II TC is a complete monitoring system for the automatic, continuous
measurement of total chlorine and other disinfectants in potable water, swimming pools, cooling
water, hot water loops and effluent. It can be operated as an intelligent Slave device on the
Modbus/JBUS using twisted pair RS485 lines.

4.2.

Function Description

4.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 48 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 240212

Addr.

Type

Alarm

Alarm high tc1

Alarm

Alarm low tc 1

Alarm

Alarm high tc 2

Alarm

Alarm low tc 2

Alarm

Alarm high cdc

Alarm

Alarm low cdc

Alarm

Sample Temp high

Alarm

Sample Temp low

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. shorted

11

Alarm

Temp. disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

Valve Defective

15

Alarm

Dis. invalid

16

Alarm

Control Timeout

17

Alarm

Reagent Pump

18

Alarm

Fotometer not connected

19

Alarm

Fotometer dirty

20

Alarm

Absorbance too high

21

Alarm

Reagent Empty

22

Alarm

Cleaning Reagent

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Calibration Recout

31

Alarm

Wrong Front-end

Description

17

Modbus

Product Specific Data for AMI Codes-II TC

Addr.

Type

Description

32

Alarm

Alarm high pH

33

Alarm

Alarm low pH

34

Alarm

35

Alarm

36

Alarm

37

Alarm

38

Alarm

39

Alarm

40

Status

DPD / Buffer

Status

Cleaning Solution

41
42
43
44
45
46
47

4.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Codes-II TC.
Tab. 4-1
Byte

Input Registers
Data type

Description

Float

reserved

Float

Total chlorine 1 (tc1)

010 ppm

Float

Total chlorine 2 (tc2)

010 ppm

12

Float

reserved

16

Float

reserved

20

Float

Calculated dichloramine (cdc)

24

Float

Temperature

28

Float

Flow

32

Float

pH Value

36

Float

Absorbance

40

Float

Photometer Frequency

44

Float

Remaining DPD/Buffer

48

Float

reserved

52

Float

Remaining Cleaning Solution

56

Float

Raw Temp. (NTK5)

60

Float

Raw Flow

64

Float

Raw pH

68

Float

Case Temperature

72

Float

Signal Output 1

020 mA

76

Float

Signal Output 2

020 mA

80

INSTRUMENT_STATUS

Instrument Display Status

Offset

18

Range of physical
value

010 ppm
-30 to +120 C
0100 B/s
014 pH
03
0300000 Hz
0100 %
0100 %
15066000 Ohm
0100 Hz
-5001500 mV
-55125 C

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Codes-II TC

4.2.3

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 4-2

4.2.4

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote Off

no

yes

15

Reset Instrument (like hardware reset)

reset

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 4-3
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History 1 Array

Array of CAL_HIST_1[15]

Cal. History Dis.

3002

1502

Cal_History 2 Array

Array of CAL_HIST_2[15]

Cal. History pH

3004

1503

Cal_History 3 Array

Array of CAL_HIST_3[15]

Vfy History Dis.

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 240212

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

19

Modbus

Product Specific Data for AMI Codes-II TC

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish
4 = Italian

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

0100 %

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

20

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Deltacon

5.

Product Specific Data for AMI Deltacon

5.1.

Introduction
This description applies for the AMI Deltacon DG and the AMI Deltacon Power.
The AMI Deltacon DG and Deltacon Power are microprocessor controlled systems for the
conductivity measurement in power cycles. Both can be operated as an intelligent Slave device
on the Modbus/JBUS using twisted pair RS485 lines.

5.2.

Function Description

5.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 48 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 240212

Addr.

Type

Description

Alarm

Cond. 1 Alarm high

Alarm

Cond. 1 Alarm low

Alarm

Cond. 2 Alarm high

Alarm

Cond. 2 Alarm low

Alarm

Cond. 3 Alarm high

Alarm

Cond. 3 Alarm low

Alarm

Temp. 1 high

Alarm

Temp. 1 low

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. 1 shorted

11

Alarm

Temp. 1 disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

pH Error

15

Alarm

16

Alarm

Control Timeout

17

Alarm

Degasser disconnected

18

Alarm

Temp. 2 shorted

19

Alarm

Temp. 2 disconnected

20

Alarm

Temp. 3 shorted

21

Alarm

Temp. 3 disconnected

22

Alarm

Degasser Ctl Timeout

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Calibration Recout

31

Alarm

Wrong Front-end

21

Modbus

Product Specific Data for AMI Deltacon

Addr.

Type

Description

32

Alarm

pH Alarm high

33

Alarm

pH Alarm low

34

Alarm

Alkali Alarm high

35

Alarm

Alkali Alarm low

36

Alarm

Temp. 2 high

37

Alarm

Temp. 2 low

38

Alarm

Temp. 3 high

39

Alarm

Temp. 3 low

40

Alarm

Degasser Line Voltage

41

Alarm

Degasser Temperature

42

Status

Resin depleted (less than 10% remaining)

43
44
45
46
47

5.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Deltacon.
Tab. 5-1
Byte

Input Registers
Data type

Description

Float

Cond.1 (sc)

03000 S

Float

Cond.2 (sc)

03000 S

Float

Cond.3 (sc)

Offset

22

Range of physical
value

03000 S

12

Float

Temp. 1

-25 to +270 C

16

Float

Temp. 2

-25 to +270 C

20

Float

Temp. 3

-25 to +270 C

24

Float

Flow

050 l/h

28

Float

pH

711 pH

32

Float

Alkalyzer

36

Float

Cond.1 uc

03000 S

40

Float

Cond.2 uc

03000 S

44

Float

Cond.3 uc

03000 S

48

Float

Pt 1000 Temp. 1

5002400 Ohm

52

Float

Pt 1000 Temp. 2

5002400 Ohm

56

Float

Pt 1000 Temp. 3

5002400 Ohm

60

Float

Raw Flow

64

Float

Case Temperature

68

Float

Resin Capacity in %

72

Float

Degasser Temperature

-500 to +740 C

76

Float

Degasser Boiling Point

20120 C

80

Float

Degasser PWM

0100 mg

0100 Hz
-55 to +125 C
0100 %

0100 %

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Deltacon

Tab. 5-1
Byte

Input Registers
Data type

Description

Range of physical

Offset

5.2.3

value

84

Float

Degasser Line Voltage

0300 V

88

Float

Degasser Air Pressure

01500 hPa

92

Float

Degasser Case Temp

96

Float

Signal Output 1

020 mA

100

Float

Signal Output 2

020 mA

104

INSTRUMENT_STATUS

Instrument Display Status

-55 to +125 C

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 5-2

5.2.4

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote Off

no

yes

15

Reset Instrument (like hardware reset)

reset

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 5-3
Address

Parameter Table
Modbus

Parameter

Addr.
2300

Object

Data Type

Description

Acc.

FIRMWARE_REC

Firmware Version

Type

1151

Version

Record

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History 1 Array

Array of CAL_HIST_1[15]

Calibration History

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 7046 25013524

Log_Rec

Record

LOG_REC

Get Logger Record


1024 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 240212

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

23

Modbus

Product Specific Data for AMI Deltacon

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

0100 %

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

24

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Hydrazine

6.

Product Specific Data for AMI Hydrazine

6.1.

Introduction
The AMI Hydrazine is a Microprocessor controlled system for the determination and control of
hydrazine or carbohydrazide used as boiler feedwater oxygen scavengers. It can be operated as
an intelligent Slave device on the Modbus/JBUS using twisted pair RS485 lines.

6.2.

Function Description

6.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 32 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 240212

Addr.

Type

Description

Alarm

Hyd. Alarm high

Alarm

Hyd. Alarm low

Alarm

Alarm

Alarm

Alarm

Alarm

Sample Temp. high

Alarm

Sample Temp. low

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. shorted

11

Alarm

Temp. disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

Reference

15

Alarm

16

Alarm

17

Alarm

18

Alarm

19

Alarm

20

Alarm

21

Alarm

Control Timeout

22

Alarm

Cleaning Reagent

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Calibration Recout

31

Alarm

Wrong Front-end

25

Modbus

Product Specific Data for AMI Hydrazine

6.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Hydrazine.
Tab. 6-1
Byte

Input Registers
Data type

Description

Range of physical

Offset

6.2.3

value
01000 ppb

Float

Hydrazine

Float

Temperature

Float

Flow

12

Float

Counter Electrode Voltage

16

Float

Raw Hydrazine

20

Float

Raw Temp. (NTK5)

24

Float

Raw Flow

28

Float

Case Temperature

32

Float

Signal Output 1

020 mA

36

Float

Signal Output 2

020 mA

40

INSTRUMENT_STATUS

Instrument Display Status

-30 to +120 C
0600 rpm
mV
0100000 nA
15066000 Ohm
0100 Hz
-55 to +125 C

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 6-2

26

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote Off

no

yes

15

Reset Instrument (like hardware reset)

reset

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Hydrazine

6.2.4

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 6-3
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History 1 Array

Array of CAL_HIST_1[15]

Cal. History Hydrazine

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 240212

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

27

Modbus

Product Specific Data for AMI Hydrazine

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

0100 %

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

28

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Hydrogene

7.

Product Specific Data for AMI Hydrogene

7.1.

Introduction
The AMI Hydrogene is a Microprocessor controlled system for the determination and control of
hydrazine or carbohydrazide used as boiler feedwater oxygen scavengers. It can be operated as
an intelligent Slave device on the Modbus/JBUS using twisted pair RS485 lines.

7.2.

Function Description

7.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 64 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 150513

Addr.

Type

Description

Alarm

Hydrogen Alarm high

Alarm

Hydrogen Alarm low

Alarm

Saturation Alarm high

Alarm

Saturation Alarm low

Alarm

Alarm

Alarm

Sample Temp. high

Alarm

Sample Temp. low

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. shorted

11

Alarm

Temp. disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

15

Alarm

16

Alarm

17

Alarm

Control Timeout

18

Alarm

19

Alarm

Quality Assurance

20

Alarm

21

Alarm

22

Alarm

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Calibration Recout

31

Alarm

Wrong Front-end

29

Modbus

Product Specific Data for AMI Hydrogene

Addr.

Type

16

Alarm

17

Alarm

18

Alarm

19

Alarm

20

Alarm

21

Alarm

22

Alarm

23

Alarm

24

Status

Description

Electrolyte depleted

25
26
27
28
29
30
31

7.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Hydrogene.
Tab. 7-1
Byte

Input Registers
Data type

Description

Offset

30

Range of physical
value
0.0120000 ppb

Float

Hydrogen

Float

Temperature

Float

Flow

12

Float

Saturation

16

Float

Air Pressure

20

Float

Raw Oxgen temp. compensated

020000 nA

24

Float

Raw Oxgen uncompensated

020000 nA

28

Float

Raw Temp. (Pt1000)

32

Float

Raw Flow

36

Float

Reserved

40

Float

Case Temperature

44

Float

Signal Output 1

020 mA

48

Float

Signal Output 2

020 mA

52

INSTRUMENT_STATUS

Instrument Display Status

-25 to +270 C
050 l/h
0200 %
5001200 hPa

5002400
0100 Hz
-55 to +125 C

A-96.250.751 / 150513

Modbus

Product Specific Data for AMI Hydrogene

7.2.3

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 7-2

7.2.4

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote Off

no

yes

15

Reset Instrument (like hardware reset)

reset

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 7-3
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History 1 Array

Array of CAL_HIST_1[15]

Cal. History Hydrazine

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 150513

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

31

Modbus

Product Specific Data for AMI Hydrogene

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

0100 %

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

32

A-96.250.751 / 150513

Modbus

Product Specific Data for AMI Inducon

8.

Product Specific Data for AMI Inducon

8.1.

Introduction
The AMI Inducon is a Microprocessor controlled electronic transmitter / controller for the
measurement of specific conductivity, concentration, salinity and TDS. It can be operated as an
intelligent Slave device on the Modbus/JBUS using twisted pair RS485 lines.

8.2.

Function Description

8.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 32 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 240212

Addr.

Type

Description

Alarm

Cond. Alarm high

Alarm

Cond. Alarm low

Alarm

Conc. Alarm high

Alarm

Conc. Alarm low

Alarm

Alarm

Alarm

Sample Temp. high

Alarm

Sample Temp. low

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. shorted

11

Alarm

Temp. disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

15

Alarm

16

Alarm

Control Timeout

17

Alarm

Temp. out of table

18

Alarm

Conc. out of table

19

Alarm

20

Alarm

21

Alarm

22

Alarm

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Calibration Recout

31

Alarm

Wrong Front-end

33

Modbus

Product Specific Data for AMI Inducon

8.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Inducon.
Tab. 8-1
Byte

Input Registers
Data type

Description

Range of physical

Offset

8.2.3

value
02000 mS

Float

Conductivity

Float

Temperature

Float

Flow

12

Float

Concentration

16

Float

Raw Value

20

Float

Pt 1000 Temp.

24

Float

Raw Flow

28

Float

Case Temperature

32

Float

Signal Output 1

020 mA

36

Float

Signal Output 2

020 mA

40

INSTRUMENT_STATUS

Instrument Display Status

-25 to +270 C
050 l/h
0100 %
02000 mS
5002400 Ohm
0100 Hz
-55 to +125 C

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 8-2

34

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote Off

no

yes

15

Reset Instrument (like hardware reset)

reset

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Inducon

8.2.4

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 8-3
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History

Array

Array of CAL_HIST_1[15]

Calibration History 1

3001

1502

Cal_History

Array

Array of CAL_HIST_2[15]

Calibration History 2

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 240212

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

35

Modbus

Product Specific Data for AMI Inducon

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Integer 16

0100 %
-1 = Disabled

AlarmsActive

Unsigned32 Alarms Pending

0100 %

Maintenance Value

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

36

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI ISE Universal

9.

Product Specific Data for AMI ISE Universal

9.1.

Introduction
The AMI ISE Universal is a Microprocessor controlled electronic transmitter / controller for the
measurement of specific conductivity, concentration, salinity and TDS. It can be operated as an
intelligent Slave device on the Modbus/JBUS using twisted pair RS485 lines.

9.2.

Function Description

9.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 32 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 261113

Addr.

Type

Description

Alarm

Alarm high

Alarm

Alarm low

Alarm

Alarm

Alarm

Alarm

Alarm

Sample Temp. high

Alarm

Sample Temp. low

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. shorted

11

Alarm

Temp. disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

15

Alarm

16

Alarm

17

Alarm

18

Alarm

19

Alarm

20

Alarm

21

Alarm

22

Alarm

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Calibration Recout

31

Alarm

Wrong Front-end

Control Timeout

37

Modbus

Product Specific Data for AMI ISE Universal

9.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI ISE Universal.
Tab. 9-1
Byte

Input Registers
Data type

Description

Range of physical

Offset

9.2.3

value

Float

Meas. value

0.011000 ppm

Float

Temperature

-30 to +130 C

Float

Flow

12

Float

Raw Value ISE2

-512 to + 512 mV

16

Float

Raw Value ISE1

-512 to + 512 mV

20

Float

NT5k Resistance

24

Float

Raw Flow

28

Float

Case Temperature

32

Float

Signal Output 1

020 mA

36

Float

Signal Output 2

020 mA

40

INSTRUMENT_STATUS

Instrument Display Status

050 l/h

15066000 Ohm
0100 Hz
-55 to +125 C

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 9-2

38

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote Off

no

yes

15

Reset Instrument (like hardware reset)

reset

A-96.250.751 / 261113

Modbus

Product Specific Data for AMI ISE Universal

9.2.4

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 9-3
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History

Array

Array of CAL_HIST_1[15]

Calibration History 1

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 261113

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

39

Modbus

Product Specific Data for AMI ISE Universal

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Integer 16

0100 %
-1 = Disabled

AlarmsActive

Unsigned32 Alarms Pending

Maintenance Value

0100 %

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

40

A-96.250.751 / 261113

Modbus

Product Specific Data for AMI LineTOC

10.

Product Specific Data for AMI LineTOC

10.1.

Introduction
The AMI LineTOC is a Microprocessor controlled complete monitoring system for the automatic,
continuous measurement of total organic carbon (TOC) in pure and high purity water. It can be
operated as an intelligent Slave device on the Modbus/JBUS using twisted pair RS485 lines.

10.2.

Function Description

10.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 32 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 240212

Addr.

Type

Description

Alarm

TOC Alarm high

Alarm

TOC Alarm low

Alarm

Cond. 1 Alarm high

Alarm

Cond. 1 Alarm low

Alarm

Cond. 2 Alarm high

Alarm

Cond. 2 Alarm low

Alarm

Temp. 1 Alarm high

Alarm

Temp. 1 Alarm low

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. 1 shorted

11

Alarm

Temp. 1 disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

Lamp

15

Alarm

Rovalve

16

Alarm

Control Timeout

17

Alarm

Periclip

18

Alarm

Temp. 2 shorted

19

Alarm

Temp. 2 disconnected

20

Alarm

Temp. 2 Alarm high

21

Alarm

Temp. 2 Alarm low

22

Alarm

EVG

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Calibration Recout

31

Alarm

Wrong Front-end

41

Modbus

Product Specific Data for AMI LineTOC

10.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI LineTOC.
Tab. 10-1
Byte

Input Registers

Data type

Description

Range of physical

Offset

10.2.3

value
02000 ppb

Float

TOC

Float

Conductivity 1

0100 S

Float

Conductivity 2

0100 S

12

Float

Temperature 1

-30 to + 130 C

16

Float

Temperature 2

-30 to + 130 C

20

Float

Concentration 1

02000 ppb

24

Float

Concentration 2

02000 ppb

28

Float

Ambient Temperature

25100 C

32

Float

Sample Temperature

25100 C

36

Float

Conductivity 1 uc

0100 S

40

Float

Conductivity 2 uc

0100 S

44

Float

Pt 1000 Temp. 1

15066000 Ohm

48

Float

Pt 1000 Temp. 2

15066000 Ohm

52

Float

Raw Concentration 1

02000 ppb

56

Float

Raw Concentration 2

02000 ppb

60

Float

Case Temperature

64

Float

Signal Output 1

020 mA

68

Float

Signal Output 2

020 mA

72

INSTRUMENT_STATUS

Instrument Display Status

-55 to +125 C

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 10-2

42

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

15

Reset Instrument (like hardware reset)

reset

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI LineTOC

10.2.4

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 10-3
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History_1 Array

Array of CAL_HIST_1[15]

Calibration History 1

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 7046 25013524

Log_Rec

Record

LOG_REC

Get Logger Record


1024 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 240212

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

43

Modbus

Product Specific Data for AMI LineTOC

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
9 = Bus

PercDos1

reserved

PercDos2

reserved

MaintValue

reserved

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

44

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Oxysafe

11.

Product Specific Data for AMI Oxysafe

11.1.

Introduction
The AMI Oxysafe is a complete monitoring system for the automatic, continuous measurement of
dissolved oxygen in sewage treatment water (aeration basins), fish rearing and surface water
such as lakes, rivers and ponds.
It can be operated as an intelligent Slave device on the Modbus/JBUS using twisted pair RS485
lines.

11.2.

Function Description

11.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 48 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 150513

Addr.

Type

Description

Alarm

Oxygen Alarm high

Alarm

Oxygen Alarm low

Alarm

Saturation Alarm high

Alarm

Saturation Alarm low

Alarm

Alarm

Alarm

Sample Temp. high

Alarm

Sample Temp. low

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. shorted

11

Alarm

Temp. disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

15

Alarm

16

Alarm

17

Alarm

Control Timeout

18

Alarm

19

Alarm

20

Alarm

21

Alarm

22

Alarm

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Calibration Recout

45

Modbus

Product Specific Data for AMI Oxysafe

Addr.

Type

Description

31

Alarm

Wrong Front-end

32

Alarm

33

Alarm

34

Alarm

35

Alarm

36

Alarm

37

Alarm

38

Alarm

39

Alarm

40
41
42
43
44
45
46
47

11.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Oxysafe.
Tab. 11-1
Byte

Input Registers

Data type

Description

Range of physical

Offset

46

value
020000 ppb

Float

Oxygen

Float

Temperature 1

Float

Flow

12

Float

Saturation

16

Float

Air Pressure

20

Float

Raw Oxygen Temp. compensated

050000 nA

24

Float

Raw Oxygen Temp. uncompensated

050000 nA

28

Float

Raw Temp. (NT5K)

32

Float

Raw Flow

36

Float

Remaining Electrolyte

-30 to + 120 C
050 l/h
0200 %
5001200 hPa

15066000 Ohm
0100 Hz
0100 %
-55 to +125 C

40

Float

Case Temperature

44

Float

Signal Output 1

020 mA

48

Float

Signal Output 2

020 mA

52

INSTRUMENT_STATUS

Instrument Display Status

A-96.250.751 / 150513

Modbus

Product Specific Data for AMI Oxysafe

11.2.3

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 11-2

11.2.4

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote off

no

yes

15

Reset Instrument (like hardware reset)

reset

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 11-3
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History_1 Array

Array of CAL_HIST_1[15]

Cal. History Oxytrace

3001

1502

Vfy_History_1

Array

Array of VFY_HIST_2[15]

Vfy. History Oxytrace

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 150513

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

47

Modbus

Product Specific Data for AMI Oxysafe

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

0100 %

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

48

A-96.250.751 / 150513

Modbus

Product Specific Data for AMI Oxytrace

12.

Product Specific Data for AMI Oxytrace

12.1.

Introduction
The AMI Oxytrace is a Electronic transmitter/controller for the measurement of dissolved oxygen
in high purity water.
Both can be operated as an intelligent Slave device on the Modbus/JBUS using twisted pair
RS485 lines.

12.2.

Function Description

12.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 48 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 150513

Addr.

Type

Description

Alarm

Oxygen Alarm high

Alarm

Oxygen Alarm low

Alarm

Saturation Alarm high

Alarm

Saturation Alarm low

Alarm

Alarm

Alarm

Sample Temp. high

Alarm

Sample Temp. low

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. shorted

11

Alarm

Temp. disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

15

Alarm

16

Alarm

Control Timeout

17

Alarm

Faraday Efficiency (QED)

18

Alarm

19

Alarm

20

Alarm

21

Alarm

22

Alarm

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Calibration Recout

31

Alarm

Wrong Front-end

49

Modbus

Product Specific Data for AMI Oxytrace

Addr.

Type

32

Alarm

33

Alarm

34

Alarm

35

Alarm

36

Alarm

37

Alarm

38

Alarm

39

Alarm

40

Status

Description

Electrolyte depleted

41
42
43
44
45
46
47

12.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Oxygen family.
Tab. 12-1
Byte

Input Registers

Data type

Description

Float

Oxygen

Float

Temperature 1

Float

Flow

Range of physical

Offset

50

value
020000 ppb
-30 to + 120 C
050 l/h
0200 %

12

Float

Saturation

16

Float

Air Pressure

20

Float

Raw Oxygen Temp. compensated

24

Float

Raw Oxygen Temp. uncompensated

5001200 hPa
050000 nA
050000 nA
15066000 Ohm

28

Float

Raw Temp. (NT5K)

32

Float

Raw Flow

0100 Hz

36

Float

Remaining Electrolyte

0100 %

40

Float

Case Temperature

44

Float

Signal Output 1

020 mA

48

Float

Signal Output 2

020 mA

52

INSTRUMENT_STATUS

Instrument Display Status

-55 to +125 C

A-96.250.751 / 150513

Modbus

Product Specific Data for AMI Oxytrace

12.2.3

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 12-2

12.2.4

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote off

no

yes

15

Reset Instrument (like hardware reset)

reset

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 12-3
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History_1 Array

Array of CAL_HIST_1[15]

Cal. History Oxytrace

3001

1502

Vfy_History_1

Array

Array of VFY_HIST_2[15]

Vfy. History Oxytrace

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 150513

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

51

Modbus

Product Specific Data for AMI Oxytrace

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

0100 %

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

52

A-96.250.751 / 150513

Modbus

Product Specific Data for AMI Pharmacon

13.

Product Specific Data for AMI Pharmacon

13.1.

Introduction
The AMI Pharmacon is a Microprocessor controlled electronic transmitter & controller for the
measurement of conductivity in purified water. It can be operated as an intelligent Slave device
on the Modbus/JBUS using twisted pair RS485 lines.

13.2.

Function Description

13.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 32 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 240212

Addr.

Type

Description

Alarm

Cond. Alarm high

Alarm

Cond. Alarm low

Alarm

Alarm

Alarm

Alarm

Alarm

Sample Temp. high

Alarm

Sample Temp. low

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. shorted

11

Alarm

Temp. disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

USP Error

15

Alarm

16

Alarm

Control Timeout

17

Alarm

Quality Assurance

18

Alarm

19

Alarm

20

Alarm

21

Alarm

22

Alarm

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Calibration Recout

31

Alarm

Wrong Front-end

53

Modbus

Product Specific Data for AMI Pharmacon

13.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Pharmacon.
Tab. 13-1
Byte

Input Registers

Data type

Description

Range of physical

Offset

13.2.3

value

Float

Conductivity

0300000 S

Float

Temperature

-25 to + 270 C

Float

Flow

12

Float

Conductivity uncompensated

16

Float

USP Value

20

Float

Pt 1000 Temp.

24

Float

Raw Flow

40

Float

Case Temperature

44

Float

Signal Output 1

020 mA

48

Float

Signal Output 2

020 mA

52

INSTRUMENT_STATUS

Instrument Display Status

050 l/h
0300000 S
0.63.1 S
5002400 Ohm
0100 Hz
-55 to +125 C

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 13-2

54

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote off

no

yes

15

Reset Instrument (like hardware reset)

reset

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Pharmacon

13.2.4

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 13-3
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3002

1502

History_2

Array

Array of HIST_2[15]

History Zero Calibration

3006

1503

History_3

Array

Array of HIST_3[15]

History Quality Assurance

3008

1504

History_4

Array

Array of HIST_4[15]

History Transmitter Test

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 240212

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

55

Modbus

Product Specific Data for AMI Pharmacon

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

0100 %

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

56

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Solicon4

14.

Product Specific Data for AMI Solicon4

14.1.

Introduction
The AMI Solicon4 is a Microprocessor controlled electronic transmitter & controller for the
measurement of conductivity in purified water. It can be operated as an intelligent Slave device
on the Modbus/JBUS using twisted pair RS485 lines.

14.2.

Function Description

14.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 32 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 150513

Addr.

Type

Description

Alarm

Cond. Alarm high

Alarm

Cond. Alarm low

Alarm

Conc. Alarm high

Alarm

Conc. Alarm low

Alarm

Alarm

Alarm

Sample Temp. high

Alarm

Sample Temp. low

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. shorted

11

Alarm

Temp. disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

15

Alarm

16

Alarm

Control Timeout

17

Alarm

Temp. out of table

18

Alarm

Conc. out of table

19

Alarm

20

Alarm

21

Alarm

22

Alarm

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Calibration Recout

31

Alarm

Wrong Front-end

57

Modbus

Product Specific Data for AMI Solicon4

14.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Solicon4.
Tab. 14-1
Byte

Input Registers

Data type

Description

Range of physical

Offset

14.2.3

value

Float

Conductivity

0300000 S

Float

Temperature

-25 to + 270 C

Float

Flow

12

Float

Concentration

16

Float

Contamination

20

Float

Pt 1000 Temp.

24

Float

Raw Flow

40

Float

Case Temperature

44

Float

Signal Output 1

020 mA

48

Float

Signal Output 2

020 mA

52

INSTRUMENT_STATUS

Instrument Display Status

050 l/h
0100 %
5002400 Ohm
0100 Hz
-55 to +125 C

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 14-2

58

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote off

no

yes

15

Reset Instrument (like hardware reset)

reset

A-96.250.751 / 150513

Modbus

Product Specific Data for AMI Solicon4

14.2.4

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 14-3
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History

Array

Array of CAL_HIST_1 [15]

Cal. History 1

3001

1502

Cal_History

Array

Array of CAL_HIST_2 [15]

Cal. History 2

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 150513

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

59

Modbus

Product Specific Data for AMI Solicon4

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %
-1 = Disabled

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

0100 %

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

60

A-96.250.751 / 150513

Modbus

Product Specific Data for AMI Phosphate

15.

Product Specific Data for AMI Phosphate

15.1.

Introduction
The AMI Phosphate is a complete monitoring system for the automatic, continuous measurement
of phosphate in potable water, effluents and cooling water. It can be operated as an intelligent
Slave device on the Modbus/JBUS using twisted pair RS485 lines.

15.2.

Function Description

15.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 48 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 240212

Addr.

Type

Description

Alarm

Phosphate Alarm high

Alarm

Phosphate Alarm low

Alarm

Alarm

Alarm

Alarm

Alarm

Alarm

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

11

Alarm

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

Valve defective

15

Alarm

16

Alarm

Control Timeout

17

Alarm

Reagent Pump

18

Alarm

Photometer not connected

19

Alarm

Photometer dirty

20

Alarm

Meas. invalid

21

Alarm

Reagent empty

22

Alarm

Detergent

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Calibration Recout

31

Alarm

Wrong Front-end

61

Modbus

Product Specific Data for AMI Phosphate

Addr.

Type

Description

Status

Reagents low

32
33
34
35
36
37
38
29
40
41
42
43
44
45
46
47

15.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Phosphate.
Tab. 15-1
Byte

Input Registers

Data type

Description

Float

Phosphate

Float

Flow

Float

Absorbance

Range of physical

Offset

15.2.3

value
010 ppm
0600 B/s
03
0300000 Hz

12

Float

Photometer Frequency

16

Float

Raw Flow

20

Float

Case Temperature

24

Float

Signal Output 1

020 mA

28

Float

Signal Output 2

020 mA

32

INSTRUMENT_STATUS

Instrument Display Status

0100 Hz
-55 to +125 C

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 15-2

62

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote off

no

yes

15

Reset Instrument (like hardware reset)

reset

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Phosphate

15.2.4

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 15-3
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History_1 Array

Array of CAL_HIST_1[15]

Cal. History Phosphate

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 240212

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

63

Modbus

Product Specific Data for AMI Phosphate

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

0100 %

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

64

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Phosphate-II

16.

Product Specific Data for AMI Phosphate-II

16.1.

Introduction
This description applies for the AMI Phosphate II and the AMI Phosphate II B.
The AMI Phosphate II is a complete monitoring system for the automatic, continuous
measurement of phosphate in potable water, effluents and cooling water.
The AMI Phosphate II B is a complete monitoring system for the automatic, continuous
measurement of ortho-phosphate in boiler water.
Both can be operated as an intelligent Slave device on the Modbus/JBUS using twisted pair
RS485 lines.

16.2.

Function Description

16.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 48 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 150312

Addr.

Type

Alarm

Description
Phos. 1 Alarm high

Alarm

Phos. 1 Alarm low

Alarm

Phos. 2 Alarm high

Alarm

Phos. 2 Alarm low

Alarm

Absorbance too high

Alarm

Alarm

Alarm

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

11

Alarm

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

Valve defective

15

Alarm

16

Alarm

Control Timeout

17

Alarm

Reagent Pump

18

Alarm

Photometer not connected

19

Alarm

Photometer dirty

20

Alarm

Meas. invalid

21

Alarm

Reagent empty

22

Alarm

Detergent

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Calibration Recout

31

Alarm

Wrong Front-end

65

Modbus

Product Specific Data for AMI Phosphate-II

Addr.

Type

32

Alarm

Description
Sample Flow 1 low

33

Alarm

Sample Flow 2 low

34

Alarm

Sample Flow 3 low

35

Alarm

Sample Flow 4 low

36

Alarm

Sample Flow 5 low

37

Alarm

Sample Flow 6 low

38

Alarm

29

Alarm

40

Status

Reagents low

41
42
43
44
45
46
47

16.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Phosphate II.
Byte

Data type

Description

Range of

Offset

010 ppm

Float

Phosphate 1

Float

Phosphate 2

010 ppm

Float

Phosphate 3

010 ppm

12

Float

Phosphate 4

010 ppm

16

Float

Phosphate 5

010 ppm

20

Float

Phosphate 6

010 ppm
0600 B/s

24

Float

Flow

28

Float

Absorbance

32

Float

Case Temperature

36

Float

Signal Output 1

020 mA
020 mA

40

Float

Signal Output 2

44

Unsigned8

Channel Status

45

50

66

physical value

Unsigned8

INSTRUMENT_STATUS

Seq. Input Status

03
-55125 C

Bit

Description

03

Channel Valve

47

Channel Meas.

Bit

Description

05

Seq. Inputs
Channel 16

reserved

No Flow

Instrument Display Status

A-96.250.751 / 150312

Modbus

Product Specific Data for AMI Phosphate-II

16.2.3

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 16-4

16.2.4

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote off

no

yes

15

Reset Instrument (like hardware reset)

reset

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 16-5
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History_1 Array

Array of CAL_HIST_1[15]

Cal. History Phosphate

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 150312

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

67

Modbus

Product Specific Data for AMI Phosphate-II

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

0100 %

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

68

A-96.250.751 / 150312

Modbus

Product Specific Data for AMI pH-Redox

17.

Product Specific Data for AMI pH-Redox

17.1.

Introduction
The AMI pH-Redox is a Microprocessor controlled Electronic transmitter/controller for the
continuous measurement of the pH value or Redox (ORP) in water. It can be operated as an
intelligent Slave device on the Modbus/JBUS using twisted pair RS485 lines.

17.2.

Function Description

17.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 32 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 240212

Addr.

Type

Description

Alarm

Alarm high

Alarm

Alarm low

Alarm

Alarm

Alarm

Alarm

Alarm

Sample Temp. high

Alarm

Sample Temp. low

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. shorted

11

Alarm

Temp. disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

15

Alarm

16

Alarm

17

Alarm

18

Alarm

19

Alarm

20

Alarm

21

Alarm

22

Alarm

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Calibration Recout

31

Alarm

Wrong Front-end

Control Timeout

69

Modbus

Product Specific Data for AMI pH-Redox

17.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI pH-Redox.
Tab. 17-1
Byte

Input Registers

Data type

Description

Range of physical

Offset

17.2.3

value
-3 to +15 pH
-500 to +1500 mV

Float

Ph
Redox

Float

Temperature

Float

Flow

12

Float

Raw Signal

16

Float

Pt 1000 Temp.

20

Float

Raw Flow

24

Float

Case Temperature

28

Float

Sensor Check

32

Float

Signal Output 1

020 mA

36

Float

Signal Output 2

020 mA

40

INSTRUMENT_STATUS

Instrument Display Status

-25 to +250 C
0200 l/h
-500 to +1500 mV
5002400
0100 Hz
-55 to +125 C
10400 M

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 17-2

70

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote off

no

yes

15

Reset Instrument (like hardware reset)

reset

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI pH-Redox

17.2.4

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 17-3
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History_1 Array

Array of CAL_HIST_1[15]

Cal. History Phosphate

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 240212

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

71

Modbus

Product Specific Data for AMI pH-Redox

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

0100 %

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

72

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI pH/mV:pH/mV

18.

Product Specific Data for AMI pH/mV:pH/mV

18.1.

Introduction
This description applies for the AMI pH/mV:pH/mV and the AMI pH/mV:pH/mV Pool
The AMI pH/mV:pH/mV is a complete monitoring system for continuous measurement of pH and
redox (ORP) in potable water and effluents.
The AMI pH/mV:pH/mV Pool is a complete monitoring system for continuous measurement of pH
and redox (ORP) potential in pool water.
Both can be operated as an intelligent Slave device on the Modbus/JBUS using twisted pair
RS485 lines.

18.2.

Function Description

18.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 32 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 150312

Addr.

Type

Alarm

Description
Alarm 1 high

Alarm

Alarm 1 low

Alarm

Alarm 2 high

Alarm

Alarm 2 low

Alarm

Temp.1 high

Alarm

Temp.1 low

Alarm

Temp. 2 high

Alarm

Temp. 2 low

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. 1 shorted

11

Alarm

Temp. 1 disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

Difference high

15

Alarm

Difference low

16

Alarm

Control Timeout

17

Alarm

18

Alarm

Temp. 2 shorted

19

Alarm

Temp. 2 disconnected

20

Alarm

21

Alarm

22

Alarm

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Calibration Recout

31

Alarm

Wrong Front-end

73

Modbus

Product Specific Data for AMI pH/mV:pH/mV

18.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI pH/mV:pH/mV.
Tab. 18-1
Byte

Input Registers

Data type

Description

Range of physical

Offset

18.2.3

value

Float

Meas. Value 1: pH
Redox

-315 pH
-5001500 mV

Float

Meas. Value 2: pH
Redox

-315 pH
-5001500 mV

Float

Temperature 1

-30 to +130 C

12

Float

Temperature 2

-30 to +130 C

16

Float

Flow

0200 l/h

120

Float

Difference

-315 pH

24

Float

Raw Signal 1

28

Float

Raw Signal 2

-5001500 mV

-5001500 mV

32

Float

Nt5K Temp. 1

15066000 Ohm

36

Float

Nt5K Temp. 2

15066000 Ohm

40

Float

Raw Flow

44

Float

Case Temperature

48

Float

Signal Output 1

020 mA

52

Float

Signal Output 2

020 mA

56

INSTRUMENT_STATUS

Instrument Display Status

0100 Hz
-55 to +125 C

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 18-2

74

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote off

no

yes

15

Reset Instrument (like hardware reset)

reset

A-96.250.751 / 150312

Modbus

Product Specific Data for AMI pH/mV:pH/mV

18.2.4

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 18-3
Address

Parameter Table
Modbus

Parameter

Object

Addr.
2300

Data Type

Description

Acc.

Firmware Version

Type

1151

Version

Record

FIRMWARE_REC

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History_1_pH
Cal_History_1_ORP

Array

Array of CAL_HIST_PH[15]
Array of CAL_HIST_ORP[15]

Cal. History 1

3002

1502

Cal_History_2_pH
Cal_History_2_ORP

Array

Array of CAL_HIST_PH[15]
Array of CAL_HIST_ORP[15]

Cal. History 2

4000 4126

20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070

25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 150312

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

75

Modbus

Product Specific Data for AMI pH/mV:pH/mV

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

0100 %

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

76

A-96.250.751 / 150312

Modbus

Product Specific Data for AMI Powercon

19.

Product Specific Data for AMI Powercon

19.1.

Introduction
The AMI Powercon is a complete monitoring system for the automatic, continuous measurement
of Silica in potable water, effluents and cooling water. It can be operated as an intelligent Slave
device on the Modbus/JBUS using twisted pair RS485 lines.

19.2.

Function Description

19.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 32 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 240212

Addr.

Type

Description

Alarm

Cond. Alarm high

Alarm

Cond. Alarm low

Alarm

Alarm

Alarm

Alarm

Alarm

Sample Temp. high

Alarm

Sample Temp. low

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. shorted

11

Alarm

Temp. disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

15

Alarm

16

Alarm

17

Alarm

18

Alarm

19

Alarm

20

Alarm

21

Alarm

22

Alarm

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Calibration Recout

31

Alarm

Wrong Front-end

Control Timeout

77

Modbus

Product Specific Data for AMI Powercon

19.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Powercon.
Tab. 19-1
Byte

Input Registers

Data type

Description

Range of physical

Offset

19.2.3

value

Float

Conductivity

0300000 S

Float

Temperature

-25 to +270 C

Float

Flow

12

Float

Conductivity uncompensated

16

Float

Contamination

20

Float

Pt 1000 Temp.

24

Float

Raw Flow

28

Float

Case Temperature

32

Float

Signal Output 1

020 mA

36

Float

Signal Output 2

020 mA

40

INSTRUMENT_STATUS

Instrument Display Status

050 l/h
0300000 S
5002400
0100 Hz
-55 to +125 C

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 19-2

78

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote off

no

yes

15

Reset Instrument (like hardware reset)

reset

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Powercon

19.2.4

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 19-3
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History_1 Array

Array of CAL_HIST_1[15]

Cal. History 1

3002

1502

Cal_History_2 Array

Array of CAL_HIST_2[15]

Cal. History 2

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 240212

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

79

Modbus

Product Specific Data for AMI Powercon

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

0100 %

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

80

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Rescon

20.

Product Specific Data for AMI Rescon

20.1.

Introduction
The AMI Rescon is a complete monitoring system for the automatic, continuous measurement of
Silica in potable water, effluents and cooling water. It can be operated as an intelligent Slave
device on the Modbus/JBUS using twisted pair RS485 lines.

20.2.

Function Description

20.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 32 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 240212

Addr.

Type

Description

Alarm

Alarm high

Alarm

Alarm low

Alarm

Alarm

Alarm

Alarm

Alarm

Sample Temp. high

Alarm

Sample Temp. low

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. shorted

11

Alarm

Temp. disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

15

Alarm

16

Alarm

17

Alarm

18

Alarm

Sensor shorted

19

Alarm

Sensor interrupted

20

Alarm

21

Alarm

22

Alarm

Control Timeout

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Calibration Recout

31

Alarm

Wrong Front-end

81

Modbus

Product Specific Data for AMI Rescon

Addr.

Type

32

Alarm

33

Alarm

34

Alarm

35

Alarm

36

Alarm

37

Alarm

38

Alarm

39

Alarm

40

Status

Description

Transmitter Check

41
42
43
44
45
46
47

82

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Rescon

20.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Rescon.
Tab. 20-1
Byte

Input Registers

Data type

Description

Range of physical value

Float

Resistivity
Conductivity

Float

Temperature

Float

Flow

12

Float

Resistivity uncompensated
Conductivity uncompensated

16

Float

Cell Constant

20

Float

NT5k Temp.

24

Float

Raw Flow

28

Float

Case Temperature

32

Float

Signal Output 1

020 mA

36

Float

Signal Output 2

020 mA

40

INSTRUMENT_STATUS

Instrument Display Status

Offset

20.2.3

0200 M
02000 S
-25 to +130 C
050 l/h
0200 M
02000 S
0.000000.99999 1/cm
15066000
0100 Hz
-55 to +125 C

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 20-2

A-96.250.751 / 240212

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote off

no

yes

15

Reset Instrument (like hardware reset)

reset

83

Modbus

Product Specific Data for AMI Rescon

20.2.4

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 20-3
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History_1 Array

Array of CAL_HIST_1[15]

Cal. History 1

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

84

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Rescon

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

0100 %

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

A-96.250.751 / 240212

85

Modbus

Product Specific Data for AMI Rescon

86

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Silica

21.

Product Specific Data for AMI Silica

21.1.

Introduction
The AMI Silica is a complete monitoring system for the automatic, continuous measurement of
Silica in potable water, effluents and cooling water. It can be operated as an intelligent Slave
device on the Modbus/JBUS using twisted pair RS485 lines.

21.2.

Function Description

21.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 48 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 240212

Addr.

Type

Description

Alarm

Si 1 Alarm high

Alarm

Si 1 Alarm low

Alarm

Si 2 Alarm high

Alarm

Si 2 Alarm low

Alarm

Alarm

Alarm

Sample Temp Hi

Alarm

Sample Temp Lo

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

11

Alarm

Temp. disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

Valve Defective

15

Alarm

16

Alarm

Control Timeout

17

Alarm

Reagent Pump

18

Alarm

Fotometer not connected

19

Alarm

Fotometer dirty

20

Alarm

21

Alarm

Reagent Empty

22

Alarm

Sequencer

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

Rovalve

30

Alarm

Cal. Recout

31

Alarm

Wrong Front-end

32

Alarm

Sample Flow 1 low

87

Modbus

Product Specific Data for AMI Silica

21.2.2

Addr.

Type

Description

33

Alarm

Sample Flow 2 low

34

Alarm

Sample Flow 3 low

35

Alarm

Sample Flow 4 low

36

Alarm

Sample Flow 5 low

37

Alarm

Sample Flow 6 low

38

Alarm

39

Alarm

40

Status

Reagents low

41

42

43

44

45

46

47

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Silica.
Tab. 21-1
Byte

Input Registers

Data type

Description

Range of

Float

Temperature

Float

Flow

0600 B/s

Float

Si 1

05000 ppb

12

Float

Si 2

05000 ppb

16

Float

Si 3

05000 ppb

20

Float

Si 4

05000 ppb

24

Float

Si 5

05000 ppb

28

Float

Si 6

05000 ppb

32

Float

Absorbance

36

Float

Case Temperature

40

Float

Signal Output 1

020 mA
020 mA

Offset

44

Float

Signal Output 2

48

Unsigned8

Channel Status

49

50

88

physical value

Unsigned8

INSTRUMENT_STATUS

Seq. Input Status

560 C

03
-55125 C

Bit

Description

03

Channel Valve

47

Channel Meas.

Bit

Description

05

Seq. Inputs
Channel 16

reserved

No Flow

Instrument Display Status

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Silica

21.2.3

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 21-2

21.2.4

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote Off

no

yes

15

Reset Instrument (like hardware reset)

1622

Channel Select 0-6 (Channel = 0 : Auto Select)

reset
dont care

dont care

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 21-3
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History 1 Array

Array of CAL_HIST_1[15]

Calibration History Silica

3004

1503

Cal_History 3 Array

Array of CAL_HIST_1[15]

Verification History Silica

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 240212

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

89

Modbus

Product Specific Data for AMI Silica

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

0100 %

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

90

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Soditrace

22.

Product Specific Data for AMI Soditrace

22.1.

Introduction
The AMI Soditrace is an Analyzer for the continuous measurement of sodium ions in trace
amounts in high purity water applications and steam generation. It can be operated as an
intelligent Slave device on the Modbus/JBUS using twisted pair RS485 lines.

22.2.

Function Description

22.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 32 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 240212

Addr.

Type

Description

Alarm

Sodium Alarm high

Alarm

Sodium Alarm low

Alarm

Cond. Alarm high

Alarm

Cond. Alarm low

Alarm

No Reagent

Alarm

No Sample

Alarm

Sample Temp high

Alarm

Sample Temp low

Alarm

Alarm

10

Alarm

Temp. shorted

11

Alarm

Temp. disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

Reagent high

15

Alarm

16

Alarm

Control Timeout

17

Alarm

Sodium Offset

18

Alarm

Sodium Slope

19

Alarm

High Background

20

Alarm

Sodium Stability

21

Alarm

Regeneration Agent

22

Alarm

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Cal. Recout

31

Alarm

Wrong Front-end

91

Modbus

Product Specific Data for AMI Soditrace

22.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Soditrace.
Tab. 22-1
Byte

Input Registers

Data type

Description

Offset

92

Range of physical
value
020000 ppb

Float

Sodium

Float

Temperature

-30 to + 120 C

Float

Conductivity

010000 S

12

Float

Control

16

Float

Raw Sodium

-510 to +190 mV

20

Float

Raw Temp. (NT5K)

15066000

24

Float

Raw Conductivity

28

Float

Raw Control

32

Float

Case Temperature

36

Float

Signal Output 1

020 mA

40

Float

Signal Output 2

020 mA

44

INSTRUMENT_STATUS

Instrument Display Status

0100 %

010000 S
0100 %
-55 to +125 C

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Soditrace

22.2.3

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 22-2

22.2.4

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote Off

no

yes

15

Reset Instrument (like hardware reset)

reset

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 22-3
Address

Parameter Table
Modbus

Parameter

Addr.
2300

Object

Data Type

Description

Acc.

FIRMWARE_REC

Firmware Version

Type

1151

Version

Record

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History 1 Array

Array of CAL_HIST_1[15]

Cal. History Sodium

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 240212

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

93

Modbus

Product Specific Data for AMI Soditrace

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

0100 %

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

94

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Sodium

23.

Product Specific Data for AMI Sodium

23.1.

Introduction
This description applies for the AMI Sodium A and the AMI Sodium P.
The AMI Sodium A and P are complete monitoring systems for the continuous determination of
dissolved sodium in the ppb-range for steam, condensate and high purity water. Both can be
operated as an intelligent Slave device on the Modbus/JBUS using twisted pair RS485 lines.

23.2.

Function Description

23.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 32 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 240212

Addr.

Type

Description

Alarm

Sodium 1 Alarm high

Alarm

Sodium 1 Alarm low

Alarm

pH1 Alarm high

Alarm

pH 1 Alarm low

Alarm

Sodium 2 Alarm high

Alarm

Sodium 2 Alarm low

Alarm

Sample Temp Hi

Alarm

Sample Temp Lo

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. shorted

11

Alarm

Temp. disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

15

Alarm

16

Alarm

Control Timeout

17

Alarm

Reagent Empty

18

Alarm

19

Alarm

20

Alarm

21

Alarm

22

Alarm

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Cal. Recout

31

Alarm

Wrong Front-end

95

Modbus

Product Specific Data for AMI Sodium

23.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Sodium A/P.
Tab. 23-1
Byte

Input Registers

Data type

Description

Offset

96

Range of physical
value

Float

Sodium Channel 1

Float

Temperature 1

Float

Flow (Sodium P)
Percent Dos (Sodium A)

12

Float

pH

16

Float

Sodium Channel 2

020000 ppb
-30 to + 120 C
0600 B/s
0100 %
014 pH
020000 ppb

20

Float

Raw Sodium

-420 to +120 mV

24

Float

Raw Temp. (NT5K)

15066000

28

Float

Raw pH

32

Float

Raw Flow

-500 to +1500 mV
0100 Hz
-55 to +125 C

36

Float

Case Temperature

40

Float

Signal Output 1

020 mA

44

Float

Signal Output 2

020 mA

48

INSTRUMENT_STATUS

Instrument Display Status

80

Unsigned 16

Channel

82

Unsigned 16

Channel Timeout

84

Unsigned 16

Channel Interval

01
0600 s
115 min

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Sodium

23.2.3

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 23-2

23.2.4

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote Off

no

yes

Channel Select

channel 0

channel 1

15

Reset Instrument (like hardware reset)

reset

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 23-3
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History 1 Array

Array of CAL_HIST_1[15]

Cal. History Sodium

3001

1502

Cal_History 2 Array

Array of CAL_HIST_2[15]

Cal. History pH

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 240212

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

97

Modbus

Product Specific Data for AMI Sodium

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

0100 %

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

98

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Trides

24.

Product Specific Data for AMI Trides

24.1.

Introduction
The AMI Trides is a Microprocessor system for the automatic and continuous measurement of
disinfectants in potable water and swimming pools. It can be operated as an intelligent Slave
device on the Modbus/JBUS using twisted pair RS485 lines.

24.2.

Function Description

24.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 32 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 240212

Addr.

Type

Description

Alarm

Dis. Alarm high

Alarm

Dis. Alarm low

Alarm

pH/Redox Alarm high

Alarm

pH/Redox Alarm low

Alarm

Alarm

Alarm

Sample Temp high

Alarm

Sample Temp low

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. shorted

11

Alarm

Temp. disconnected

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

TRIDES Reference

15

Alarm

16

Alarm

Control Timeout

17

Alarm

18

Alarm

19

Alarm

20

Alarm

21

Alarm

22

Alarm

Cleaning Reagent

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Cal. Recout

31

Alarm

Wrong Front-end

99

Modbus

Product Specific Data for AMI Trides

24.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Trides.
Tab. 24-1
Byte

Input Registers

Data type

Description

Offset

100

Range of physical
value

Float

Disinfection

Float

pH

Float

Temperature

12

Float

Flow

16

Float

Hypochlorous Acid

020 ppm
-3 to + 16 pH
-500 to +1500 mV
-30 to + 120 C
0600 rpm
010 ppm

20

Float

Counter Electrode Voltage

24

Float

Raw Disinfection

mV

28

Float

Raw Temp. (NT5K)

32

Float

Raw pH

36

Float

Raw Flow

40

Float

Case Temperature

44

Float

Signal Output 1

020 mA

48

Float

Signal Output 2

020 mA

52

INSTRUMENT_STATUS

Instrument Display Status

0100 A
15066000
-500 to +1500 mV
0100 Hz
-55 to +125 C

A-96.250.751 / 240212

Modbus

Product Specific Data for AMI Trides

24.2.3

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 24-2

24.2.4

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote Off

no

yes

15

Reset Instrument (like hardware reset)

reset

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 24-3
Address

Parameter Table
Modbus

Parameter

Addr.
2300

Object

Data Type

Description

Acc.

FIRMWARE_REC

Firmware Version

Type

1151

Version

Record

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History 1 Array

Array of CAL_HIST_1[15]

Cal. History Dis.

3001

1502

Cal_History 2 Array

Array of CAL_HIST_2[15]

Cal. History pH/Redox

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 240212

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

101

Modbus

Product Specific Data for AMI Trides

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

0100 %

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record

102

A-96.250.751 / 240212

Modbus

Product Specific Data for Turbidity

25.

Product Specific Data for Turbidity

25.1.

Introduction
This description applies for the AMI Turbiwell, AMI Turbitrace and AMI Turbitrack.
The AMI Turbiwell, Turbitrace and Turbitrack are Nephelometric systems for the automatic and
continuous measurement of turbidity in water. All can be operated as an intelligent Slave device
on the Modbus/JBUS using twisted pair RS485 lines.

25.2.

Function Description

25.2.1

Read Input Status (Alarm &Status Bits), Function 02


Up to 32 bits of Alarms are available as summarized in the table below:

A-96.250.751 / 240212

Addr.

Type

Description

Alarm

Turb. Alarm high

Alarm

Turb. Alarm low

Alarm

Alarm

Alarm

Alarm

Alarm

Sample Temp high (not implemented)

Alarm

Sample Temp low (not implemented)

Alarm

Sample Flow high

Alarm

Sample Flow low

10

Alarm

Temp. shorted (not implemented)

11

Alarm

Temp. disconnected (not implemented)

12

Alarm

Case Temp. high

13

Alarm

Case Temp. low

14

Alarm

15

Alarm

Zero Offset too high

16

Alarm

Control Timeout

17

Alarm

Turbi Disconnected

18

Alarm

Turbi Overflow

19

Alarm

Laser off

20

Alarm

21

Alarm

22

Alarm

Cleaning Reagent

23

Alarm

Input Active

24

Alarm

IC MK41T56

25

Alarm

IC LM75

26

Alarm

IC PCF8574

27

Alarm

EEprom Microcon

28

Alarm

EEprom Motherboard

29

Alarm

EEprom Front-end

30

Alarm

Cal. Recout

31

Alarm

Wrong Front-end

103

Modbus

Product Specific Data for Turbidity

25.2.2

Read Input Registers (Process data): Function 04


The process data are made up of the measuring values of the AMI Turbidity family.
Tab. 25-1
Byte

Input Registers

Data type

Description

Offset

104

Range of physical
value
0300 FNU

Float

Turbitity

Float

Temperature (future use fix 25C)

Float

Flow

12

Float

Turbitity raw

0250000 Hz

16

Float

Pt 1000 (future use)

5002400

20

Float

Raw Flow

24

Float

V-Signal

08 V

28

Float

V-Reference

08 V

32

Float

Range Gain

36

Float

Case Temperature

40

Float

Signal Output 1

020 mA

44

Float

Signal Output 2

020 mA

48

INSTRUMENT_STATUS

Instrument Display Status

-25 to +250 C
0100 l/h

0100 Hz

2535
-55 to +125 C

A-96.250.751 / 240212

Modbus

Product Specific Data for Turbidity

25.2.3

Force Single Coil: Function 05


This function is used to set relays and reset the instrument. Relays can only be forced if set to
fieldbus function
Tab. 25-2

25.2.4

Single Coil

Addr.

Description

OFF (0)

ON (1)

01

Relay Contacts 1 and 2

open

close

Hold

no

yes

Remote Off

no

yes

Start Calibration

no

yes

15

Reset Instrument (like hardware reset)

reset

Read Holding Registers (Parameter data): Function 03


Read Parameter data (see Parameter Table). All objects of type Record are defined in the
Manufacturer specific tables following the parameter table.
Note: Note: The Address is used in a proprietary way to select the designated record.

Tab. 25-3
Address

Parameter Table
Modbus

Parameter

Addr.

Object

Data Type

Description

Acc.

Type

2300

1151

Version

Record

FIRMWARE_REC

Firmware Version

2400

1201

Num_Evt

Simple

Unsigned16

No. of Event Records

2402

1202

Num_Log

Simple

Unsigned16

No. of Log Records

3000

1501

Cal_History 1 Array

Array of CAL_HIST_1[15]

Calibration History

3001

1502

Cal_History 2 Array

Array of CAL_HIST_2[15]

Verification History

4000 4126 20012064

Event_Rec

Record

EVENT_REC

Get Event Record


64 Records max.

5000 8070 25014036

Log_Rec

Record

LOG_REC

Get Logger Record


1536 Records max.

Manufacturer-specific Data Types & Records


DATE_STAMP Format

A-96.250.751 / 240212

YYYYYY

MMMM

DDDDD

HHHHH

MMMMMM

SSSSSS

Year
063

Month
112

Date
131

Hour
023

Minutes
059

Seconds
059

105

Modbus

Product Specific Data for Turbidity

Parameter
INSTRUMENT_STATUS

Size Data Type

Description

Range

32

Record

Instrument Display Status

Language

Unsigned8

Language Selection

0 = German
1 = English
2 = French
3 = Spanish

RunMode

Unsigned8

Run Mode

0 = Run
1 = Hold
2 = Off
3 = Grab

AlaMode

Unsigned8

Alarm Mode

0 = Off
1 = on
2 = Blinking

MaintMode

Unsigned8

Maintenance Mode

0 = Off
1 = on
2 = Blinking

RelMode1

Unsigned8

Relais 1 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

RelMode2

Unsigned8

Relais 2 Display Mode

0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus

PercDos1

Unsigned16 Percent Dosage Relais 1

PercDos2

Unsigned16 Percent Dosage Relais 2

0100 %

MaintValue

Unsigned16 Maintenance Value

0100 %

AlarmsActive

Unsigned32 Alarms Pending

AlarmsActive

AlarmsLatched

Unsigned32 Alarms no Acknowledged

AlarmsLatched

0100 %

AlarmsActiveE

Unsigned32 Extended Alarms Pending

AlarmsActiveE

AlarmsLatchedE

Unsigned32 Extended Alarms not


Acknowledged

AlarmsLatchedE

TimeStamp

DATE_STA Current Time


MP

end Record