Interface Description
SCADA
Supervisory Control and
Data Acquisation
Ethernet
Webserver
Modbus RTU
A-96.250.751 / 150312
Customer Support
SWAN and its representatives maintain a fully trained staff of technical specialists
around the world. For any technical question, contact your nearest
SWAN representative, or the manufacturer:
SWAN ANALYTISCHE INSTRUMENTE AG
Studbachstrasse 13
8340 Hinwil
Switzerland
Internet: www.swan.ch
E-mail: support@swan.ch
Document Status
Title:
ID:
A-96.250.751
Revision
Issue
00
February 2012
First release
01
March 2012
02
Sept. 2012
03
Mai 2013
AMI Profibus
1.
1.1.
1.2.
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Query-Response Cycle . . . . . . . . . . . . . . . . . . . . . . .
Data and Control Functions . . . . . . . . . . . . . . . . . . . . . . .
5
6
6
2.
Product Specific Data for AMI Codes-II . . . . . . . . . . . .
2.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.2.1
Read Input Status (Alarm &Status Bits), Function 02 .
2.2.2
Read Input Registers (Process data): Function 04 . . .
2.2.3
Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
2.2.4
Read Holding Registers (Parameter data): Function 03
9
9
9
9
10
11
11
3.
Product Specific Data for AMI Codes-II CC . . . . . . . . .
3.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.2.1
Read Input Status (Alarm &Status Bits), Function 02 .
3.2.2
Read Input Registers (Process data): Function 04 . . .
3.2.3
Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
3.2.4
Read Holding Registers (Parameter data): Function 03
13
13
13
13
14
15
15
4.
Product Specific Data for AMI Codes-II TC . . . . . . . . .
4.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.2.1
Read Input Status (Alarm &Status Bits), Function 02 .
4.2.2
Read Input Registers (Process data): Function 04 . . .
4.2.3
Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
4.2.4
Read Holding Registers (Parameter data): Function 03
17
17
17
17
18
19
19
5.
Product Specific Data for AMI Deltacon. . . . . . . . . . . .
5.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.2.1
Read Input Status (Alarm &Status Bits), Function 02 .
5.2.2
Read Input Registers (Process data): Function 04 . . .
5.2.3
Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
5.2.4
Read Holding Registers (Parameter data): Function 03
21
21
21
21
22
23
23
6.
Product Specific Data for AMI Hydrazine. . . . . . . . . . .
6.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.2.1
Read Input Status (Alarm &Status Bits), Function 02 .
6.2.2
Read Input Registers (Process data): Function 04 . . .
6.2.3
Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
6.2.4
Read Holding Registers (Parameter data): Function 03
25
25
25
25
26
26
27
7.
Product Specific Data for AMI Hydrogene . . . . . . . . . .
7.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7.2.1
Read Input Status (Alarm &Status Bits), Function 02 .
7.2.2
Read Input Registers (Process data): Function 04 . . .
7.2.3
Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
7.2.4
Read Holding Registers (Parameter data): Function 03
29
29
29
29
30
31
31
8.
Product Specific Data for AMI Inducon . . . . . . . . . . . .
8.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.2.1
Read Input Status (Alarm &Status Bits), Function 02 .
8.2.2
Read Input Registers (Process data): Function 04 . . .
8.2.3
Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
8.2.4
Read Holding Registers (Parameter data): Function 03
33
33
33
33
34
34
35
A-96.250.751 / 150312
AMI Profibus
9.
Product Specific Data for AMI ISE Universal . . . . . . . .
9.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9.2.1
Read Input Status (Alarm &Status Bits), Function 02. .
9.2.2
Read Input Registers (Process data): Function 04. . . .
9.2.3
Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
9.2.4
Read Holding Registers (Parameter data): Function 03
37
37
37
37
38
38
39
10.
Product Specific Data for AMI LineTOC . . . . . . . . . . . .
10.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10.2.1 Read Input Status (Alarm &Status Bits), Function 02. .
10.2.2 Read Input Registers (Process data): Function 04. . . .
10.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
10.2.4 Read Holding Registers (Parameter data): Function 03
41
41
41
41
42
42
43
11.
Product Specific Data for AMI Oxysafe . . . . . . . . . . . .
11.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
11.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
11.2.1 Read Input Status (Alarm &Status Bits), Function 02. .
11.2.2 Read Input Registers (Process data): Function 04. . . .
11.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
11.2.4 Read Holding Registers (Parameter data): Function 03
45
45
45
45
46
47
47
12.
Product Specific Data for AMI Oxytrace. . . . . . . . . . . .
12.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
12.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
12.2.1 Read Input Status (Alarm &Status Bits), Function 02. .
12.2.2 Read Input Registers (Process data): Function 04. . . .
12.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
12.2.4 Read Holding Registers (Parameter data): Function 03
49
49
49
49
50
51
51
13.
Product Specific Data for AMI Pharmacon . . . . . . . . .
13.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
13.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
13.2.1 Read Input Status (Alarm &Status Bits), Function 02. .
13.2.2 Read Input Registers (Process data): Function 04. . . .
13.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
13.2.4 Read Holding Registers (Parameter data): Function 03
53
53
53
53
54
54
55
14.
Product Specific Data for AMI Solicon4 . . . . . . . . . . . .
14.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
14.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
14.2.1 Read Input Status (Alarm &Status Bits), Function 02. .
14.2.2 Read Input Registers (Process data): Function 04. . . .
14.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
14.2.4 Read Holding Registers (Parameter data): Function 03
57
57
57
57
58
58
59
15.
Product Specific Data for AMI Phosphate . . . . . . . . . .
15.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
15.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
15.2.1 Read Input Status (Alarm &Status Bits), Function 02. .
15.2.2 Read Input Registers (Process data): Function 04. . . .
15.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
15.2.4 Read Holding Registers (Parameter data): Function 03
61
61
61
61
62
62
63
16.
Product Specific Data for AMI Phosphate-II . . . . . . . .
16.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
16.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
16.2.1 Read Input Status (Alarm &Status Bits), Function 02. .
16.2.2 Read Input Registers (Process data): Function 04. . . .
16.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
16.2.4 Read Holding Registers (Parameter data): Function 03
65
65
65
65
66
67
67
A-96.250.751 / 150312
AMI Profibus
17.
Product Specific Data for AMI pH-Redox. . . . . . . . . . .
17.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
17.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
17.2.1 Read Input Status (Alarm &Status Bits), Function 02 .
17.2.2 Read Input Registers (Process data): Function 04 . . .
17.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
17.2.4 Read Holding Registers (Parameter data): Function 03
69
69
69
69
70
70
71
18.
Product Specific Data for AMI pH/mV:pH/mV . . . . . . .
18.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
18.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
18.2.1 Read Input Status (Alarm &Status Bits), Function 02 .
18.2.2 Read Input Registers (Process data): Function 04 . . .
18.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
18.2.4 Read Holding Registers (Parameter data): Function 03
73
73
73
73
74
74
75
19.
Product Specific Data for AMI Powercon. . . . . . . . . . .
19.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
19.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
19.2.1 Read Input Status (Alarm &Status Bits), Function 02 .
19.2.2 Read Input Registers (Process data): Function 04 . . .
19.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
19.2.4 Read Holding Registers (Parameter data): Function 03
77
77
77
77
78
78
79
20.
Product Specific Data for AMI Rescon. . . . . . . . . . . . .
20.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
20.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
20.2.1 Read Input Status (Alarm &Status Bits), Function 02 .
20.2.2 Read Input Registers (Process data): Function 04 . . .
20.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
20.2.4 Read Holding Registers (Parameter data): Function 03
81
81
81
81
83
83
84
21.
Product Specific Data for AMI Silica . . . . . . . . . . . . . .
21.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
21.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
21.2.1 Read Input Status (Alarm &Status Bits), Function 02 .
21.2.2 Read Input Registers (Process data): Function 04 . . .
21.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
21.2.4 Read Holding Registers (Parameter data): Function 03
87
87
87
87
88
89
89
22.
Product Specific Data for AMI Soditrace . . . . . . . . . . .
22.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
22.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
22.2.1 Read Input Status (Alarm &Status Bits), Function 02 .
22.2.2 Read Input Registers (Process data): Function 04 . . .
22.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
22.2.4 Read Holding Registers (Parameter data): Function 03
91
91
91
91
92
93
93
23.
Product Specific Data for AMI Sodium. . . . . . . . . . . . .
23.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
23.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
23.2.1 Read Input Status (Alarm &Status Bits), Function 02 .
23.2.2 Read Input Registers (Process data): Function 04 . . .
23.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
23.2.4 Read Holding Registers (Parameter data): Function 03
95
95
95
95
96
97
97
24.
Product Specific Data for AMI Trides . . . . . . . . . . . . . .
24.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
24.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
24.2.1 Read Input Status (Alarm &Status Bits), Function 02 .
24.2.2 Read Input Registers (Process data): Function 04 . . .
24.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
24.2.4 Read Holding Registers (Parameter data): Function 03
99
99
99
99
100
101
101
A-96.250.751 / 150312
AMI Profibus
25.
Product Specific Data for Turbidity . . . . . . . . . . . . . . .
25.1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
25.2. Function Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
25.2.1 Read Input Status (Alarm &Status Bits), Function 02. .
25.2.2 Read Input Registers (Process data): Function 04. . . .
25.2.3 Force Single Coil: Function 05 . . . . . . . . . . . . . . . . . . .
25.2.4 Read Holding Registers (Parameter data): Function 03
103
103
103
103
104
105
105
A-96.250.751 / 150312
Modbus
Introduction
1.
Introduction
AMI Transmitters can be operated as intelligent slave devices on the Modbus/JBUS using
twisted pair RS485 lines.
Modbus/Jbus
The Modbus/Jbus Protocol communicates using a master-slave technique, in which only one
device (the master) can initiate transactions (queries). The other devices (the slaves) respond by
supplying the requested data to the master, or by taking the action requested in the query.
Typical master devices include host processors, PLCs etc. Typical slaves can be PLCs or
intelligent transmitter/controller like the AMI transmitters.
Instruments which only uses a subset of the available data from another instrument, are based
on the same data protocol.
For example, the AMI Deltacon DG and the AMI Deltacon Power provide exactly the same data,
but the AMI Deltacon Power does not provide any data concerning the degasser.
These instruments and the belonging chapters are listed in the following table:
Instruments based on the same data protocol.
See chapter
A-96.250.751 / 140312
Modbus
Introduction
1.1.
The Query
The Response
Serial
Transmission
modes
RTU Mode
Coding
System
Slave
Address
Function
Data
CRC
END
T1-T2-T3-T4
8 Bits
8 Bits
n x 8 Bits
16 Bits
T1-T2-T3-T4
1.2.
Data address
in Modbus
Messages
The list below shows the function codes supported by the AMI Instruments. Codes are listed in
decimal. Function codes not supported return an exception response 02 to the master.
Tab. 1-2 Modbus function codes
Code
Name
Description
02
03
04
05
16
A-96.250.751 / 140312
Modbus
Introduction
Data representation
Discrete
Inputs
Definition:
big-endian and
little-endian
The process- and parameter data of the AMI Instruments are represented by float (IEEE 754), signedor unsigned integer values. Depending on the range of the value, float, integer8, unsigned8,
integer16, unsigned16, integer32 or unsigned32 representation is used.
Alarm &Status bits are packed as one input per bit of the data field.
Big-endian and little-endian are terms that describe the order in which a sequence of bytes are stored
in computer memory. Big-endian is an order in which the "big end" (most significant value in the
sequence) is stored first (at the lowest storage address). Little-endian is an order in which the "little
end" (least significant value in the sequence) is stored first. For example, in a big-endian computer,
the two bytes required for the hexadecimal number 4F52 would be stored as 4F52 in storage. In a
little-endian system, it would be stored as 524F
ATTENTION
Big-endian format is used.
A-96.250.751 / 140312
Modbus
Introduction
A-96.250.751 / 140312
Modbus
2.
2.1.
Introduction
The AMI Codes-II is a complete monitoring system for the automatic, continuous measurement
and dosing control of chlorine and other disinfectants in potable water, swimming pools, cooling
water and effluent. It can be operated as an intelligent Slave device on the Modbus/JBUS using
twisted pair RS485 lines
2.2.
Function Description
2.2.1
A-96.250.751 / 240212
Addr.
Type
Description
Alarm
Alarm
Alarm
pH 1 Alarm high
Alarm
pH 1 Alarm low
Alarm
Dis. high
Alarm
Alarm
Alarm
Alarm
Alarm
10
Alarm
Temp. shorted
11
Alarm
Temp. disconnected
12
Alarm
13
Alarm
14
Alarm
Valve Defective
15
Alarm
16
Alarm
Control Timeout
17
Alarm
Reagent Pump
18
Alarm
19
Alarm
Fotometer dirty
20
Alarm
Dis. invalid
21
Alarm
Reagent Empty
22
Alarm
Cleaning Reagent
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Cal. Recout
31
Alarm
Wrong Front-end
Modbus
2.2.2
Addr.
Type
32
Alarm
33
Alarm
34
Alarm
35
Alarm
36
Alarm
37
Alarm
38
Alarm
39
Alarm
40
Alarm
Reagents low
41
Alarm
42
Alarm
43
Alarm
44
Alarm
45
Status
46
Status
47
Status
Description
Input Registers
Data type
Description
Offset
10
Range of
physical value
Float
Disinfection
010 ppm
Float
pH
014 pH
Float
Temperature
12
Float
Flow
16
Float
Reserved
20
Float
Absorbance
24
Float
Raw Disinfection
28
Float
Raw pH
-30 to +120 C
0600 B/s
03
0300000 Hz
-5001500 mV
15066000 Ohm
32
Float
36
Float
Raw Flow
40
Float
Case Temperature
44
Float
Signal Output 1
020 mA
44
Float
Signal Output 2
020 mA
50
INSTRUMENT_STATUS
0100 Hz
-55125 C
A-96.250.751 / 240212
Modbus
2.2.3
2.2.4
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote Off
no
yes
15
reset
Tab. 2-3
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History 1 Array
Array of CAL_HIST_1[15]
3001
1502
Cal_History 2 Array
Array of CAL_HIST_2[15]
Calibration History pH
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 240212
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
11
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
4 = Italian
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
0100 %
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
12
A-96.250.751 / 240212
Modbus
3.
3.1.
Introduction
The AMI Codes-II CC is a complete monitoring system for the automatic, continuous
measurement and dosing control of chlorine and other disinfectants (e.g. monochloramine) in
potable water, swimming pools, cooling water, hot water loops and effluent. It can be operated as
an intelligent Slave device on the Modbus/JBUS using twisted pair RS485 lines.
3.2.
Function Description
3.2.1
A-96.250.751 / 240212
Addr.
Type
Alarm
Alarm
Alarm
Alarm
Alarm low tc 1
Alarm
Alarm high tc 2
Alarm
Alarm low tc 2
Alarm
Alarm
Alarm
Alarm
10
Alarm
Temp. shorted
11
Alarm
Temp. disconnected
12
Alarm
13
Alarm
14
Alarm
Valve Defective
15
Alarm
Dis. invalid
16
Alarm
Control Timeout
17
Alarm
Reagent Pump
18
Alarm
19
Alarm
Fotometer dirty
20
Alarm
Dis. invalid
21
Alarm
Reagent Empty
22
Alarm
Cleaning Reagent
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Calibration Recout
31
Alarm
Wrong Front-end
Description
13
Modbus
3.2.2
Addr.
Type
Description
32
Alarm
33
Alarm
34
Alarm
35
Alarm
36
Alarm
37
Alarm
38
Alarm
Alarm high pH
39
Alarm
Alarm low pH
40
Alarm
DPD / Buffer
41
Alarm
Potassium Iodide
42
Alarm
Cleaning Solution
43
Alarm
44
Alarm
45
Status
46
Status
47
Status
Input Registers
Data type
Description
Float
010 ppm
Float
010 ppm
Float
010 ppm
12
Float
010 ppm
16
Float
010 ppm
20
Float
010 ppm
24
Float
Temperature
28
Float
Flow
32
Float
pH Value
36
Float
Absorbance
40
Float
Photometer Frequency
44
Float
Remaining DPD/Buffer
0100 %
48
Float
0100 %
52
Float
0100 %
56
Float
60
Float
Raw Flow
64
Float
Raw pH
68
Float
Case Temperature
72
Float
Signal Output 1
020 mA
76
Float
Signal Output 2
020 mA
80
INSTRUMENT_STATUS
Offset
14
Range of physical
value
-30 to +120 C
0100 B/s
014 pH
03
0300000 Hz
15066000 Ohm
0100 Hz
-5001500 mV
-55125 C
A-96.250.751 / 240212
Modbus
3.2.3
3.2.4
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote Off
no
yes
15
reset
Tab. 3-3
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History 1 Array
Array of CAL_HIST_1[15]
3002
1502
Cal_History 2 Array
Array of CAL_HIST_2[15]
Calibration History pH
3004
1503
Cal_History 3 Array
Array of CAL_HIST_3[15]
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 240212
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
15
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
4 = Italian
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
0100 %
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
16
A-96.250.751 / 240212
Modbus
4.
4.1.
Introduction
The AMI Codes-II TC is a complete monitoring system for the automatic, continuous
measurement of total chlorine and other disinfectants in potable water, swimming pools, cooling
water, hot water loops and effluent. It can be operated as an intelligent Slave device on the
Modbus/JBUS using twisted pair RS485 lines.
4.2.
Function Description
4.2.1
A-96.250.751 / 240212
Addr.
Type
Alarm
Alarm
Alarm low tc 1
Alarm
Alarm high tc 2
Alarm
Alarm low tc 2
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
10
Alarm
Temp. shorted
11
Alarm
Temp. disconnected
12
Alarm
13
Alarm
14
Alarm
Valve Defective
15
Alarm
Dis. invalid
16
Alarm
Control Timeout
17
Alarm
Reagent Pump
18
Alarm
19
Alarm
Fotometer dirty
20
Alarm
21
Alarm
Reagent Empty
22
Alarm
Cleaning Reagent
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Calibration Recout
31
Alarm
Wrong Front-end
Description
17
Modbus
Addr.
Type
Description
32
Alarm
Alarm high pH
33
Alarm
Alarm low pH
34
Alarm
35
Alarm
36
Alarm
37
Alarm
38
Alarm
39
Alarm
40
Status
DPD / Buffer
Status
Cleaning Solution
41
42
43
44
45
46
47
4.2.2
Input Registers
Data type
Description
Float
reserved
Float
010 ppm
Float
010 ppm
12
Float
reserved
16
Float
reserved
20
Float
24
Float
Temperature
28
Float
Flow
32
Float
pH Value
36
Float
Absorbance
40
Float
Photometer Frequency
44
Float
Remaining DPD/Buffer
48
Float
reserved
52
Float
56
Float
60
Float
Raw Flow
64
Float
Raw pH
68
Float
Case Temperature
72
Float
Signal Output 1
020 mA
76
Float
Signal Output 2
020 mA
80
INSTRUMENT_STATUS
Offset
18
Range of physical
value
010 ppm
-30 to +120 C
0100 B/s
014 pH
03
0300000 Hz
0100 %
0100 %
15066000 Ohm
0100 Hz
-5001500 mV
-55125 C
A-96.250.751 / 240212
Modbus
4.2.3
4.2.4
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote Off
no
yes
15
reset
Tab. 4-3
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History 1 Array
Array of CAL_HIST_1[15]
3002
1502
Cal_History 2 Array
Array of CAL_HIST_2[15]
Cal. History pH
3004
1503
Cal_History 3 Array
Array of CAL_HIST_3[15]
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 240212
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
19
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
4 = Italian
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
0100 %
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
20
A-96.250.751 / 240212
Modbus
5.
5.1.
Introduction
This description applies for the AMI Deltacon DG and the AMI Deltacon Power.
The AMI Deltacon DG and Deltacon Power are microprocessor controlled systems for the
conductivity measurement in power cycles. Both can be operated as an intelligent Slave device
on the Modbus/JBUS using twisted pair RS485 lines.
5.2.
Function Description
5.2.1
A-96.250.751 / 240212
Addr.
Type
Description
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Temp. 1 high
Alarm
Temp. 1 low
Alarm
Alarm
10
Alarm
Temp. 1 shorted
11
Alarm
Temp. 1 disconnected
12
Alarm
13
Alarm
14
Alarm
pH Error
15
Alarm
16
Alarm
Control Timeout
17
Alarm
Degasser disconnected
18
Alarm
Temp. 2 shorted
19
Alarm
Temp. 2 disconnected
20
Alarm
Temp. 3 shorted
21
Alarm
Temp. 3 disconnected
22
Alarm
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Calibration Recout
31
Alarm
Wrong Front-end
21
Modbus
Addr.
Type
Description
32
Alarm
pH Alarm high
33
Alarm
pH Alarm low
34
Alarm
35
Alarm
36
Alarm
Temp. 2 high
37
Alarm
Temp. 2 low
38
Alarm
Temp. 3 high
39
Alarm
Temp. 3 low
40
Alarm
41
Alarm
Degasser Temperature
42
Status
43
44
45
46
47
5.2.2
Input Registers
Data type
Description
Float
Cond.1 (sc)
03000 S
Float
Cond.2 (sc)
03000 S
Float
Cond.3 (sc)
Offset
22
Range of physical
value
03000 S
12
Float
Temp. 1
-25 to +270 C
16
Float
Temp. 2
-25 to +270 C
20
Float
Temp. 3
-25 to +270 C
24
Float
Flow
050 l/h
28
Float
pH
711 pH
32
Float
Alkalyzer
36
Float
Cond.1 uc
03000 S
40
Float
Cond.2 uc
03000 S
44
Float
Cond.3 uc
03000 S
48
Float
Pt 1000 Temp. 1
5002400 Ohm
52
Float
Pt 1000 Temp. 2
5002400 Ohm
56
Float
Pt 1000 Temp. 3
5002400 Ohm
60
Float
Raw Flow
64
Float
Case Temperature
68
Float
Resin Capacity in %
72
Float
Degasser Temperature
-500 to +740 C
76
Float
20120 C
80
Float
Degasser PWM
0100 mg
0100 Hz
-55 to +125 C
0100 %
0100 %
A-96.250.751 / 240212
Modbus
Tab. 5-1
Byte
Input Registers
Data type
Description
Range of physical
Offset
5.2.3
value
84
Float
0300 V
88
Float
01500 hPa
92
Float
96
Float
Signal Output 1
020 mA
100
Float
Signal Output 2
020 mA
104
INSTRUMENT_STATUS
-55 to +125 C
5.2.4
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote Off
no
yes
15
reset
Tab. 5-3
Address
Parameter Table
Modbus
Parameter
Addr.
2300
Object
Data Type
Description
Acc.
FIRMWARE_REC
Firmware Version
Type
1151
Version
Record
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History 1 Array
Array of CAL_HIST_1[15]
Calibration History
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 240212
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
23
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
0100 %
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
24
A-96.250.751 / 240212
Modbus
6.
6.1.
Introduction
The AMI Hydrazine is a Microprocessor controlled system for the determination and control of
hydrazine or carbohydrazide used as boiler feedwater oxygen scavengers. It can be operated as
an intelligent Slave device on the Modbus/JBUS using twisted pair RS485 lines.
6.2.
Function Description
6.2.1
A-96.250.751 / 240212
Addr.
Type
Description
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
10
Alarm
Temp. shorted
11
Alarm
Temp. disconnected
12
Alarm
13
Alarm
14
Alarm
Reference
15
Alarm
16
Alarm
17
Alarm
18
Alarm
19
Alarm
20
Alarm
21
Alarm
Control Timeout
22
Alarm
Cleaning Reagent
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Calibration Recout
31
Alarm
Wrong Front-end
25
Modbus
6.2.2
Input Registers
Data type
Description
Range of physical
Offset
6.2.3
value
01000 ppb
Float
Hydrazine
Float
Temperature
Float
Flow
12
Float
16
Float
Raw Hydrazine
20
Float
24
Float
Raw Flow
28
Float
Case Temperature
32
Float
Signal Output 1
020 mA
36
Float
Signal Output 2
020 mA
40
INSTRUMENT_STATUS
-30 to +120 C
0600 rpm
mV
0100000 nA
15066000 Ohm
0100 Hz
-55 to +125 C
26
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote Off
no
yes
15
reset
A-96.250.751 / 240212
Modbus
6.2.4
Tab. 6-3
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History 1 Array
Array of CAL_HIST_1[15]
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 240212
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
27
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
0100 %
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
28
A-96.250.751 / 240212
Modbus
7.
7.1.
Introduction
The AMI Hydrogene is a Microprocessor controlled system for the determination and control of
hydrazine or carbohydrazide used as boiler feedwater oxygen scavengers. It can be operated as
an intelligent Slave device on the Modbus/JBUS using twisted pair RS485 lines.
7.2.
Function Description
7.2.1
A-96.250.751 / 150513
Addr.
Type
Description
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
10
Alarm
Temp. shorted
11
Alarm
Temp. disconnected
12
Alarm
13
Alarm
14
Alarm
15
Alarm
16
Alarm
17
Alarm
Control Timeout
18
Alarm
19
Alarm
Quality Assurance
20
Alarm
21
Alarm
22
Alarm
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Calibration Recout
31
Alarm
Wrong Front-end
29
Modbus
Addr.
Type
16
Alarm
17
Alarm
18
Alarm
19
Alarm
20
Alarm
21
Alarm
22
Alarm
23
Alarm
24
Status
Description
Electrolyte depleted
25
26
27
28
29
30
31
7.2.2
Input Registers
Data type
Description
Offset
30
Range of physical
value
0.0120000 ppb
Float
Hydrogen
Float
Temperature
Float
Flow
12
Float
Saturation
16
Float
Air Pressure
20
Float
020000 nA
24
Float
020000 nA
28
Float
32
Float
Raw Flow
36
Float
Reserved
40
Float
Case Temperature
44
Float
Signal Output 1
020 mA
48
Float
Signal Output 2
020 mA
52
INSTRUMENT_STATUS
-25 to +270 C
050 l/h
0200 %
5001200 hPa
5002400
0100 Hz
-55 to +125 C
A-96.250.751 / 150513
Modbus
7.2.3
7.2.4
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote Off
no
yes
15
reset
Tab. 7-3
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History 1 Array
Array of CAL_HIST_1[15]
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 150513
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
31
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
0100 %
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
32
A-96.250.751 / 150513
Modbus
8.
8.1.
Introduction
The AMI Inducon is a Microprocessor controlled electronic transmitter / controller for the
measurement of specific conductivity, concentration, salinity and TDS. It can be operated as an
intelligent Slave device on the Modbus/JBUS using twisted pair RS485 lines.
8.2.
Function Description
8.2.1
A-96.250.751 / 240212
Addr.
Type
Description
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
10
Alarm
Temp. shorted
11
Alarm
Temp. disconnected
12
Alarm
13
Alarm
14
Alarm
15
Alarm
16
Alarm
Control Timeout
17
Alarm
18
Alarm
19
Alarm
20
Alarm
21
Alarm
22
Alarm
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Calibration Recout
31
Alarm
Wrong Front-end
33
Modbus
8.2.2
Input Registers
Data type
Description
Range of physical
Offset
8.2.3
value
02000 mS
Float
Conductivity
Float
Temperature
Float
Flow
12
Float
Concentration
16
Float
Raw Value
20
Float
Pt 1000 Temp.
24
Float
Raw Flow
28
Float
Case Temperature
32
Float
Signal Output 1
020 mA
36
Float
Signal Output 2
020 mA
40
INSTRUMENT_STATUS
-25 to +270 C
050 l/h
0100 %
02000 mS
5002400 Ohm
0100 Hz
-55 to +125 C
34
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote Off
no
yes
15
reset
A-96.250.751 / 240212
Modbus
8.2.4
Tab. 8-3
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History
Array
Array of CAL_HIST_1[15]
Calibration History 1
3001
1502
Cal_History
Array
Array of CAL_HIST_2[15]
Calibration History 2
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 240212
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
35
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
Integer 16
0100 %
-1 = Disabled
AlarmsActive
0100 %
Maintenance Value
AlarmsActive
AlarmsLatched
AlarmsLatched
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
36
A-96.250.751 / 240212
Modbus
9.
9.1.
Introduction
The AMI ISE Universal is a Microprocessor controlled electronic transmitter / controller for the
measurement of specific conductivity, concentration, salinity and TDS. It can be operated as an
intelligent Slave device on the Modbus/JBUS using twisted pair RS485 lines.
9.2.
Function Description
9.2.1
A-96.250.751 / 261113
Addr.
Type
Description
Alarm
Alarm high
Alarm
Alarm low
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
10
Alarm
Temp. shorted
11
Alarm
Temp. disconnected
12
Alarm
13
Alarm
14
Alarm
15
Alarm
16
Alarm
17
Alarm
18
Alarm
19
Alarm
20
Alarm
21
Alarm
22
Alarm
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Calibration Recout
31
Alarm
Wrong Front-end
Control Timeout
37
Modbus
9.2.2
Input Registers
Data type
Description
Range of physical
Offset
9.2.3
value
Float
Meas. value
0.011000 ppm
Float
Temperature
-30 to +130 C
Float
Flow
12
Float
-512 to + 512 mV
16
Float
-512 to + 512 mV
20
Float
NT5k Resistance
24
Float
Raw Flow
28
Float
Case Temperature
32
Float
Signal Output 1
020 mA
36
Float
Signal Output 2
020 mA
40
INSTRUMENT_STATUS
050 l/h
15066000 Ohm
0100 Hz
-55 to +125 C
38
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote Off
no
yes
15
reset
A-96.250.751 / 261113
Modbus
9.2.4
Tab. 9-3
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History
Array
Array of CAL_HIST_1[15]
Calibration History 1
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 261113
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
39
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
Integer 16
0100 %
-1 = Disabled
AlarmsActive
Maintenance Value
0100 %
AlarmsActive
AlarmsLatched
AlarmsLatched
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
40
A-96.250.751 / 261113
Modbus
10.
10.1.
Introduction
The AMI LineTOC is a Microprocessor controlled complete monitoring system for the automatic,
continuous measurement of total organic carbon (TOC) in pure and high purity water. It can be
operated as an intelligent Slave device on the Modbus/JBUS using twisted pair RS485 lines.
10.2.
Function Description
10.2.1
A-96.250.751 / 240212
Addr.
Type
Description
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
10
Alarm
Temp. 1 shorted
11
Alarm
Temp. 1 disconnected
12
Alarm
13
Alarm
14
Alarm
Lamp
15
Alarm
Rovalve
16
Alarm
Control Timeout
17
Alarm
Periclip
18
Alarm
Temp. 2 shorted
19
Alarm
Temp. 2 disconnected
20
Alarm
21
Alarm
22
Alarm
EVG
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Calibration Recout
31
Alarm
Wrong Front-end
41
Modbus
10.2.2
Input Registers
Data type
Description
Range of physical
Offset
10.2.3
value
02000 ppb
Float
TOC
Float
Conductivity 1
0100 S
Float
Conductivity 2
0100 S
12
Float
Temperature 1
-30 to + 130 C
16
Float
Temperature 2
-30 to + 130 C
20
Float
Concentration 1
02000 ppb
24
Float
Concentration 2
02000 ppb
28
Float
Ambient Temperature
25100 C
32
Float
Sample Temperature
25100 C
36
Float
Conductivity 1 uc
0100 S
40
Float
Conductivity 2 uc
0100 S
44
Float
Pt 1000 Temp. 1
15066000 Ohm
48
Float
Pt 1000 Temp. 2
15066000 Ohm
52
Float
Raw Concentration 1
02000 ppb
56
Float
Raw Concentration 2
02000 ppb
60
Float
Case Temperature
64
Float
Signal Output 1
020 mA
68
Float
Signal Output 2
020 mA
72
INSTRUMENT_STATUS
-55 to +125 C
42
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
15
reset
A-96.250.751 / 240212
Modbus
10.2.4
Tab. 10-3
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History_1 Array
Array of CAL_HIST_1[15]
Calibration History 1
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 240212
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
43
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
9 = Bus
PercDos1
reserved
PercDos2
reserved
MaintValue
reserved
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
44
A-96.250.751 / 240212
Modbus
11.
11.1.
Introduction
The AMI Oxysafe is a complete monitoring system for the automatic, continuous measurement of
dissolved oxygen in sewage treatment water (aeration basins), fish rearing and surface water
such as lakes, rivers and ponds.
It can be operated as an intelligent Slave device on the Modbus/JBUS using twisted pair RS485
lines.
11.2.
Function Description
11.2.1
A-96.250.751 / 150513
Addr.
Type
Description
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
10
Alarm
Temp. shorted
11
Alarm
Temp. disconnected
12
Alarm
13
Alarm
14
Alarm
15
Alarm
16
Alarm
17
Alarm
Control Timeout
18
Alarm
19
Alarm
20
Alarm
21
Alarm
22
Alarm
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Calibration Recout
45
Modbus
Addr.
Type
Description
31
Alarm
Wrong Front-end
32
Alarm
33
Alarm
34
Alarm
35
Alarm
36
Alarm
37
Alarm
38
Alarm
39
Alarm
40
41
42
43
44
45
46
47
11.2.2
Input Registers
Data type
Description
Range of physical
Offset
46
value
020000 ppb
Float
Oxygen
Float
Temperature 1
Float
Flow
12
Float
Saturation
16
Float
Air Pressure
20
Float
050000 nA
24
Float
050000 nA
28
Float
32
Float
Raw Flow
36
Float
Remaining Electrolyte
-30 to + 120 C
050 l/h
0200 %
5001200 hPa
15066000 Ohm
0100 Hz
0100 %
-55 to +125 C
40
Float
Case Temperature
44
Float
Signal Output 1
020 mA
48
Float
Signal Output 2
020 mA
52
INSTRUMENT_STATUS
A-96.250.751 / 150513
Modbus
11.2.3
11.2.4
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote off
no
yes
15
reset
Tab. 11-3
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History_1 Array
Array of CAL_HIST_1[15]
3001
1502
Vfy_History_1
Array
Array of VFY_HIST_2[15]
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 150513
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
47
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
0100 %
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
48
A-96.250.751 / 150513
Modbus
12.
12.1.
Introduction
The AMI Oxytrace is a Electronic transmitter/controller for the measurement of dissolved oxygen
in high purity water.
Both can be operated as an intelligent Slave device on the Modbus/JBUS using twisted pair
RS485 lines.
12.2.
Function Description
12.2.1
A-96.250.751 / 150513
Addr.
Type
Description
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
10
Alarm
Temp. shorted
11
Alarm
Temp. disconnected
12
Alarm
13
Alarm
14
Alarm
15
Alarm
16
Alarm
Control Timeout
17
Alarm
18
Alarm
19
Alarm
20
Alarm
21
Alarm
22
Alarm
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Calibration Recout
31
Alarm
Wrong Front-end
49
Modbus
Addr.
Type
32
Alarm
33
Alarm
34
Alarm
35
Alarm
36
Alarm
37
Alarm
38
Alarm
39
Alarm
40
Status
Description
Electrolyte depleted
41
42
43
44
45
46
47
12.2.2
Input Registers
Data type
Description
Float
Oxygen
Float
Temperature 1
Float
Flow
Range of physical
Offset
50
value
020000 ppb
-30 to + 120 C
050 l/h
0200 %
12
Float
Saturation
16
Float
Air Pressure
20
Float
24
Float
5001200 hPa
050000 nA
050000 nA
15066000 Ohm
28
Float
32
Float
Raw Flow
0100 Hz
36
Float
Remaining Electrolyte
0100 %
40
Float
Case Temperature
44
Float
Signal Output 1
020 mA
48
Float
Signal Output 2
020 mA
52
INSTRUMENT_STATUS
-55 to +125 C
A-96.250.751 / 150513
Modbus
12.2.3
12.2.4
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote off
no
yes
15
reset
Tab. 12-3
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History_1 Array
Array of CAL_HIST_1[15]
3001
1502
Vfy_History_1
Array
Array of VFY_HIST_2[15]
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 150513
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
51
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
0100 %
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
52
A-96.250.751 / 150513
Modbus
13.
13.1.
Introduction
The AMI Pharmacon is a Microprocessor controlled electronic transmitter & controller for the
measurement of conductivity in purified water. It can be operated as an intelligent Slave device
on the Modbus/JBUS using twisted pair RS485 lines.
13.2.
Function Description
13.2.1
A-96.250.751 / 240212
Addr.
Type
Description
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
10
Alarm
Temp. shorted
11
Alarm
Temp. disconnected
12
Alarm
13
Alarm
14
Alarm
USP Error
15
Alarm
16
Alarm
Control Timeout
17
Alarm
Quality Assurance
18
Alarm
19
Alarm
20
Alarm
21
Alarm
22
Alarm
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Calibration Recout
31
Alarm
Wrong Front-end
53
Modbus
13.2.2
Input Registers
Data type
Description
Range of physical
Offset
13.2.3
value
Float
Conductivity
0300000 S
Float
Temperature
-25 to + 270 C
Float
Flow
12
Float
Conductivity uncompensated
16
Float
USP Value
20
Float
Pt 1000 Temp.
24
Float
Raw Flow
40
Float
Case Temperature
44
Float
Signal Output 1
020 mA
48
Float
Signal Output 2
020 mA
52
INSTRUMENT_STATUS
050 l/h
0300000 S
0.63.1 S
5002400 Ohm
0100 Hz
-55 to +125 C
54
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote off
no
yes
15
reset
A-96.250.751 / 240212
Modbus
13.2.4
Tab. 13-3
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3002
1502
History_2
Array
Array of HIST_2[15]
3006
1503
History_3
Array
Array of HIST_3[15]
3008
1504
History_4
Array
Array of HIST_4[15]
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 240212
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
55
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
0100 %
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
56
A-96.250.751 / 240212
Modbus
14.
14.1.
Introduction
The AMI Solicon4 is a Microprocessor controlled electronic transmitter & controller for the
measurement of conductivity in purified water. It can be operated as an intelligent Slave device
on the Modbus/JBUS using twisted pair RS485 lines.
14.2.
Function Description
14.2.1
A-96.250.751 / 150513
Addr.
Type
Description
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
10
Alarm
Temp. shorted
11
Alarm
Temp. disconnected
12
Alarm
13
Alarm
14
Alarm
15
Alarm
16
Alarm
Control Timeout
17
Alarm
18
Alarm
19
Alarm
20
Alarm
21
Alarm
22
Alarm
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Calibration Recout
31
Alarm
Wrong Front-end
57
Modbus
14.2.2
Input Registers
Data type
Description
Range of physical
Offset
14.2.3
value
Float
Conductivity
0300000 S
Float
Temperature
-25 to + 270 C
Float
Flow
12
Float
Concentration
16
Float
Contamination
20
Float
Pt 1000 Temp.
24
Float
Raw Flow
40
Float
Case Temperature
44
Float
Signal Output 1
020 mA
48
Float
Signal Output 2
020 mA
52
INSTRUMENT_STATUS
050 l/h
0100 %
5002400 Ohm
0100 Hz
-55 to +125 C
58
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote off
no
yes
15
reset
A-96.250.751 / 150513
Modbus
14.2.4
Tab. 14-3
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History
Array
Cal. History 1
3001
1502
Cal_History
Array
Cal. History 2
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 150513
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
59
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
0100 %
-1 = Disabled
AlarmsActive
AlarmsActive
0100 %
AlarmsLatched
AlarmsLatched
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
60
A-96.250.751 / 150513
Modbus
15.
15.1.
Introduction
The AMI Phosphate is a complete monitoring system for the automatic, continuous measurement
of phosphate in potable water, effluents and cooling water. It can be operated as an intelligent
Slave device on the Modbus/JBUS using twisted pair RS485 lines.
15.2.
Function Description
15.2.1
A-96.250.751 / 240212
Addr.
Type
Description
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
10
Alarm
11
Alarm
12
Alarm
13
Alarm
14
Alarm
Valve defective
15
Alarm
16
Alarm
Control Timeout
17
Alarm
Reagent Pump
18
Alarm
19
Alarm
Photometer dirty
20
Alarm
Meas. invalid
21
Alarm
Reagent empty
22
Alarm
Detergent
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Calibration Recout
31
Alarm
Wrong Front-end
61
Modbus
Addr.
Type
Description
Status
Reagents low
32
33
34
35
36
37
38
29
40
41
42
43
44
45
46
47
15.2.2
Input Registers
Data type
Description
Float
Phosphate
Float
Flow
Float
Absorbance
Range of physical
Offset
15.2.3
value
010 ppm
0600 B/s
03
0300000 Hz
12
Float
Photometer Frequency
16
Float
Raw Flow
20
Float
Case Temperature
24
Float
Signal Output 1
020 mA
28
Float
Signal Output 2
020 mA
32
INSTRUMENT_STATUS
0100 Hz
-55 to +125 C
62
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote off
no
yes
15
reset
A-96.250.751 / 240212
Modbus
15.2.4
Tab. 15-3
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History_1 Array
Array of CAL_HIST_1[15]
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 240212
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
63
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
0100 %
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
64
A-96.250.751 / 240212
Modbus
16.
16.1.
Introduction
This description applies for the AMI Phosphate II and the AMI Phosphate II B.
The AMI Phosphate II is a complete monitoring system for the automatic, continuous
measurement of phosphate in potable water, effluents and cooling water.
The AMI Phosphate II B is a complete monitoring system for the automatic, continuous
measurement of ortho-phosphate in boiler water.
Both can be operated as an intelligent Slave device on the Modbus/JBUS using twisted pair
RS485 lines.
16.2.
Function Description
16.2.1
A-96.250.751 / 150312
Addr.
Type
Alarm
Description
Phos. 1 Alarm high
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
10
Alarm
11
Alarm
12
Alarm
13
Alarm
14
Alarm
Valve defective
15
Alarm
16
Alarm
Control Timeout
17
Alarm
Reagent Pump
18
Alarm
19
Alarm
Photometer dirty
20
Alarm
Meas. invalid
21
Alarm
Reagent empty
22
Alarm
Detergent
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Calibration Recout
31
Alarm
Wrong Front-end
65
Modbus
Addr.
Type
32
Alarm
Description
Sample Flow 1 low
33
Alarm
34
Alarm
35
Alarm
36
Alarm
37
Alarm
38
Alarm
29
Alarm
40
Status
Reagents low
41
42
43
44
45
46
47
16.2.2
Data type
Description
Range of
Offset
010 ppm
Float
Phosphate 1
Float
Phosphate 2
010 ppm
Float
Phosphate 3
010 ppm
12
Float
Phosphate 4
010 ppm
16
Float
Phosphate 5
010 ppm
20
Float
Phosphate 6
010 ppm
0600 B/s
24
Float
Flow
28
Float
Absorbance
32
Float
Case Temperature
36
Float
Signal Output 1
020 mA
020 mA
40
Float
Signal Output 2
44
Unsigned8
Channel Status
45
50
66
physical value
Unsigned8
INSTRUMENT_STATUS
03
-55125 C
Bit
Description
03
Channel Valve
47
Channel Meas.
Bit
Description
05
Seq. Inputs
Channel 16
reserved
No Flow
A-96.250.751 / 150312
Modbus
16.2.3
16.2.4
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote off
no
yes
15
reset
Tab. 16-5
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History_1 Array
Array of CAL_HIST_1[15]
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 150312
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
67
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
0100 %
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
68
A-96.250.751 / 150312
Modbus
17.
17.1.
Introduction
The AMI pH-Redox is a Microprocessor controlled Electronic transmitter/controller for the
continuous measurement of the pH value or Redox (ORP) in water. It can be operated as an
intelligent Slave device on the Modbus/JBUS using twisted pair RS485 lines.
17.2.
Function Description
17.2.1
A-96.250.751 / 240212
Addr.
Type
Description
Alarm
Alarm high
Alarm
Alarm low
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
10
Alarm
Temp. shorted
11
Alarm
Temp. disconnected
12
Alarm
13
Alarm
14
Alarm
15
Alarm
16
Alarm
17
Alarm
18
Alarm
19
Alarm
20
Alarm
21
Alarm
22
Alarm
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Calibration Recout
31
Alarm
Wrong Front-end
Control Timeout
69
Modbus
17.2.2
Input Registers
Data type
Description
Range of physical
Offset
17.2.3
value
-3 to +15 pH
-500 to +1500 mV
Float
Ph
Redox
Float
Temperature
Float
Flow
12
Float
Raw Signal
16
Float
Pt 1000 Temp.
20
Float
Raw Flow
24
Float
Case Temperature
28
Float
Sensor Check
32
Float
Signal Output 1
020 mA
36
Float
Signal Output 2
020 mA
40
INSTRUMENT_STATUS
-25 to +250 C
0200 l/h
-500 to +1500 mV
5002400
0100 Hz
-55 to +125 C
10400 M
70
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote off
no
yes
15
reset
A-96.250.751 / 240212
Modbus
17.2.4
Tab. 17-3
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History_1 Array
Array of CAL_HIST_1[15]
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 240212
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
71
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
0100 %
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
72
A-96.250.751 / 240212
Modbus
18.
18.1.
Introduction
This description applies for the AMI pH/mV:pH/mV and the AMI pH/mV:pH/mV Pool
The AMI pH/mV:pH/mV is a complete monitoring system for continuous measurement of pH and
redox (ORP) in potable water and effluents.
The AMI pH/mV:pH/mV Pool is a complete monitoring system for continuous measurement of pH
and redox (ORP) potential in pool water.
Both can be operated as an intelligent Slave device on the Modbus/JBUS using twisted pair
RS485 lines.
18.2.
Function Description
18.2.1
A-96.250.751 / 150312
Addr.
Type
Alarm
Description
Alarm 1 high
Alarm
Alarm 1 low
Alarm
Alarm 2 high
Alarm
Alarm 2 low
Alarm
Temp.1 high
Alarm
Temp.1 low
Alarm
Temp. 2 high
Alarm
Temp. 2 low
Alarm
Alarm
10
Alarm
Temp. 1 shorted
11
Alarm
Temp. 1 disconnected
12
Alarm
13
Alarm
14
Alarm
Difference high
15
Alarm
Difference low
16
Alarm
Control Timeout
17
Alarm
18
Alarm
Temp. 2 shorted
19
Alarm
Temp. 2 disconnected
20
Alarm
21
Alarm
22
Alarm
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Calibration Recout
31
Alarm
Wrong Front-end
73
Modbus
18.2.2
Input Registers
Data type
Description
Range of physical
Offset
18.2.3
value
Float
Meas. Value 1: pH
Redox
-315 pH
-5001500 mV
Float
Meas. Value 2: pH
Redox
-315 pH
-5001500 mV
Float
Temperature 1
-30 to +130 C
12
Float
Temperature 2
-30 to +130 C
16
Float
Flow
0200 l/h
120
Float
Difference
-315 pH
24
Float
Raw Signal 1
28
Float
Raw Signal 2
-5001500 mV
-5001500 mV
32
Float
Nt5K Temp. 1
15066000 Ohm
36
Float
Nt5K Temp. 2
15066000 Ohm
40
Float
Raw Flow
44
Float
Case Temperature
48
Float
Signal Output 1
020 mA
52
Float
Signal Output 2
020 mA
56
INSTRUMENT_STATUS
0100 Hz
-55 to +125 C
74
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote off
no
yes
15
reset
A-96.250.751 / 150312
Modbus
18.2.4
Tab. 18-3
Address
Parameter Table
Modbus
Parameter
Object
Addr.
2300
Data Type
Description
Acc.
Firmware Version
Type
1151
Version
Record
FIRMWARE_REC
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History_1_pH
Cal_History_1_ORP
Array
Array of CAL_HIST_PH[15]
Array of CAL_HIST_ORP[15]
Cal. History 1
3002
1502
Cal_History_2_pH
Cal_History_2_ORP
Array
Array of CAL_HIST_PH[15]
Array of CAL_HIST_ORP[15]
Cal. History 2
4000 4126
20012064
Event_Rec
Record
EVENT_REC
5000 8070
25014036
Log_Rec
Record
LOG_REC
A-96.250.751 / 150312
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
75
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
0100 %
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
76
A-96.250.751 / 150312
Modbus
19.
19.1.
Introduction
The AMI Powercon is a complete monitoring system for the automatic, continuous measurement
of Silica in potable water, effluents and cooling water. It can be operated as an intelligent Slave
device on the Modbus/JBUS using twisted pair RS485 lines.
19.2.
Function Description
19.2.1
A-96.250.751 / 240212
Addr.
Type
Description
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
10
Alarm
Temp. shorted
11
Alarm
Temp. disconnected
12
Alarm
13
Alarm
14
Alarm
15
Alarm
16
Alarm
17
Alarm
18
Alarm
19
Alarm
20
Alarm
21
Alarm
22
Alarm
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Calibration Recout
31
Alarm
Wrong Front-end
Control Timeout
77
Modbus
19.2.2
Input Registers
Data type
Description
Range of physical
Offset
19.2.3
value
Float
Conductivity
0300000 S
Float
Temperature
-25 to +270 C
Float
Flow
12
Float
Conductivity uncompensated
16
Float
Contamination
20
Float
Pt 1000 Temp.
24
Float
Raw Flow
28
Float
Case Temperature
32
Float
Signal Output 1
020 mA
36
Float
Signal Output 2
020 mA
40
INSTRUMENT_STATUS
050 l/h
0300000 S
5002400
0100 Hz
-55 to +125 C
78
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote off
no
yes
15
reset
A-96.250.751 / 240212
Modbus
19.2.4
Tab. 19-3
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History_1 Array
Array of CAL_HIST_1[15]
Cal. History 1
3002
1502
Cal_History_2 Array
Array of CAL_HIST_2[15]
Cal. History 2
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 240212
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
79
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
0100 %
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
80
A-96.250.751 / 240212
Modbus
20.
20.1.
Introduction
The AMI Rescon is a complete monitoring system for the automatic, continuous measurement of
Silica in potable water, effluents and cooling water. It can be operated as an intelligent Slave
device on the Modbus/JBUS using twisted pair RS485 lines.
20.2.
Function Description
20.2.1
A-96.250.751 / 240212
Addr.
Type
Description
Alarm
Alarm high
Alarm
Alarm low
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
10
Alarm
Temp. shorted
11
Alarm
Temp. disconnected
12
Alarm
13
Alarm
14
Alarm
15
Alarm
16
Alarm
17
Alarm
18
Alarm
Sensor shorted
19
Alarm
Sensor interrupted
20
Alarm
21
Alarm
22
Alarm
Control Timeout
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Calibration Recout
31
Alarm
Wrong Front-end
81
Modbus
Addr.
Type
32
Alarm
33
Alarm
34
Alarm
35
Alarm
36
Alarm
37
Alarm
38
Alarm
39
Alarm
40
Status
Description
Transmitter Check
41
42
43
44
45
46
47
82
A-96.250.751 / 240212
Modbus
20.2.2
Input Registers
Data type
Description
Float
Resistivity
Conductivity
Float
Temperature
Float
Flow
12
Float
Resistivity uncompensated
Conductivity uncompensated
16
Float
Cell Constant
20
Float
NT5k Temp.
24
Float
Raw Flow
28
Float
Case Temperature
32
Float
Signal Output 1
020 mA
36
Float
Signal Output 2
020 mA
40
INSTRUMENT_STATUS
Offset
20.2.3
0200 M
02000 S
-25 to +130 C
050 l/h
0200 M
02000 S
0.000000.99999 1/cm
15066000
0100 Hz
-55 to +125 C
A-96.250.751 / 240212
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote off
no
yes
15
reset
83
Modbus
20.2.4
Tab. 20-3
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History_1 Array
Array of CAL_HIST_1[15]
Cal. History 1
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
84
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
A-96.250.751 / 240212
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
0100 %
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
A-96.250.751 / 240212
85
Modbus
86
A-96.250.751 / 240212
Modbus
21.
21.1.
Introduction
The AMI Silica is a complete monitoring system for the automatic, continuous measurement of
Silica in potable water, effluents and cooling water. It can be operated as an intelligent Slave
device on the Modbus/JBUS using twisted pair RS485 lines.
21.2.
Function Description
21.2.1
A-96.250.751 / 240212
Addr.
Type
Description
Alarm
Si 1 Alarm high
Alarm
Si 1 Alarm low
Alarm
Si 2 Alarm high
Alarm
Si 2 Alarm low
Alarm
Alarm
Alarm
Sample Temp Hi
Alarm
Sample Temp Lo
Alarm
Alarm
10
Alarm
11
Alarm
Temp. disconnected
12
Alarm
13
Alarm
14
Alarm
Valve Defective
15
Alarm
16
Alarm
Control Timeout
17
Alarm
Reagent Pump
18
Alarm
19
Alarm
Fotometer dirty
20
Alarm
21
Alarm
Reagent Empty
22
Alarm
Sequencer
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
Rovalve
30
Alarm
Cal. Recout
31
Alarm
Wrong Front-end
32
Alarm
87
Modbus
21.2.2
Addr.
Type
Description
33
Alarm
34
Alarm
35
Alarm
36
Alarm
37
Alarm
38
Alarm
39
Alarm
40
Status
Reagents low
41
42
43
44
45
46
47
Input Registers
Data type
Description
Range of
Float
Temperature
Float
Flow
0600 B/s
Float
Si 1
05000 ppb
12
Float
Si 2
05000 ppb
16
Float
Si 3
05000 ppb
20
Float
Si 4
05000 ppb
24
Float
Si 5
05000 ppb
28
Float
Si 6
05000 ppb
32
Float
Absorbance
36
Float
Case Temperature
40
Float
Signal Output 1
020 mA
020 mA
Offset
44
Float
Signal Output 2
48
Unsigned8
Channel Status
49
50
88
physical value
Unsigned8
INSTRUMENT_STATUS
560 C
03
-55125 C
Bit
Description
03
Channel Valve
47
Channel Meas.
Bit
Description
05
Seq. Inputs
Channel 16
reserved
No Flow
A-96.250.751 / 240212
Modbus
21.2.3
21.2.4
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote Off
no
yes
15
1622
reset
dont care
dont care
Tab. 21-3
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History 1 Array
Array of CAL_HIST_1[15]
3004
1503
Cal_History 3 Array
Array of CAL_HIST_1[15]
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 240212
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
89
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
0100 %
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
90
A-96.250.751 / 240212
Modbus
22.
22.1.
Introduction
The AMI Soditrace is an Analyzer for the continuous measurement of sodium ions in trace
amounts in high purity water applications and steam generation. It can be operated as an
intelligent Slave device on the Modbus/JBUS using twisted pair RS485 lines.
22.2.
Function Description
22.2.1
A-96.250.751 / 240212
Addr.
Type
Description
Alarm
Alarm
Alarm
Alarm
Alarm
No Reagent
Alarm
No Sample
Alarm
Alarm
Alarm
Alarm
10
Alarm
Temp. shorted
11
Alarm
Temp. disconnected
12
Alarm
13
Alarm
14
Alarm
Reagent high
15
Alarm
16
Alarm
Control Timeout
17
Alarm
Sodium Offset
18
Alarm
Sodium Slope
19
Alarm
High Background
20
Alarm
Sodium Stability
21
Alarm
Regeneration Agent
22
Alarm
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Cal. Recout
31
Alarm
Wrong Front-end
91
Modbus
22.2.2
Input Registers
Data type
Description
Offset
92
Range of physical
value
020000 ppb
Float
Sodium
Float
Temperature
-30 to + 120 C
Float
Conductivity
010000 S
12
Float
Control
16
Float
Raw Sodium
-510 to +190 mV
20
Float
15066000
24
Float
Raw Conductivity
28
Float
Raw Control
32
Float
Case Temperature
36
Float
Signal Output 1
020 mA
40
Float
Signal Output 2
020 mA
44
INSTRUMENT_STATUS
0100 %
010000 S
0100 %
-55 to +125 C
A-96.250.751 / 240212
Modbus
22.2.3
22.2.4
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote Off
no
yes
15
reset
Tab. 22-3
Address
Parameter Table
Modbus
Parameter
Addr.
2300
Object
Data Type
Description
Acc.
FIRMWARE_REC
Firmware Version
Type
1151
Version
Record
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History 1 Array
Array of CAL_HIST_1[15]
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 240212
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
93
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
0100 %
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
94
A-96.250.751 / 240212
Modbus
23.
23.1.
Introduction
This description applies for the AMI Sodium A and the AMI Sodium P.
The AMI Sodium A and P are complete monitoring systems for the continuous determination of
dissolved sodium in the ppb-range for steam, condensate and high purity water. Both can be
operated as an intelligent Slave device on the Modbus/JBUS using twisted pair RS485 lines.
23.2.
Function Description
23.2.1
A-96.250.751 / 240212
Addr.
Type
Description
Alarm
Alarm
Alarm
Alarm
pH 1 Alarm low
Alarm
Alarm
Alarm
Sample Temp Hi
Alarm
Sample Temp Lo
Alarm
Alarm
10
Alarm
Temp. shorted
11
Alarm
Temp. disconnected
12
Alarm
13
Alarm
14
Alarm
15
Alarm
16
Alarm
Control Timeout
17
Alarm
Reagent Empty
18
Alarm
19
Alarm
20
Alarm
21
Alarm
22
Alarm
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Cal. Recout
31
Alarm
Wrong Front-end
95
Modbus
23.2.2
Input Registers
Data type
Description
Offset
96
Range of physical
value
Float
Sodium Channel 1
Float
Temperature 1
Float
Flow (Sodium P)
Percent Dos (Sodium A)
12
Float
pH
16
Float
Sodium Channel 2
020000 ppb
-30 to + 120 C
0600 B/s
0100 %
014 pH
020000 ppb
20
Float
Raw Sodium
-420 to +120 mV
24
Float
15066000
28
Float
Raw pH
32
Float
Raw Flow
-500 to +1500 mV
0100 Hz
-55 to +125 C
36
Float
Case Temperature
40
Float
Signal Output 1
020 mA
44
Float
Signal Output 2
020 mA
48
INSTRUMENT_STATUS
80
Unsigned 16
Channel
82
Unsigned 16
Channel Timeout
84
Unsigned 16
Channel Interval
01
0600 s
115 min
A-96.250.751 / 240212
Modbus
23.2.3
23.2.4
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote Off
no
yes
Channel Select
channel 0
channel 1
15
reset
Tab. 23-3
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History 1 Array
Array of CAL_HIST_1[15]
3001
1502
Cal_History 2 Array
Array of CAL_HIST_2[15]
Cal. History pH
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 240212
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
97
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
0100 %
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
98
A-96.250.751 / 240212
Modbus
24.
24.1.
Introduction
The AMI Trides is a Microprocessor system for the automatic and continuous measurement of
disinfectants in potable water and swimming pools. It can be operated as an intelligent Slave
device on the Modbus/JBUS using twisted pair RS485 lines.
24.2.
Function Description
24.2.1
A-96.250.751 / 240212
Addr.
Type
Description
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
10
Alarm
Temp. shorted
11
Alarm
Temp. disconnected
12
Alarm
13
Alarm
14
Alarm
TRIDES Reference
15
Alarm
16
Alarm
Control Timeout
17
Alarm
18
Alarm
19
Alarm
20
Alarm
21
Alarm
22
Alarm
Cleaning Reagent
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Cal. Recout
31
Alarm
Wrong Front-end
99
Modbus
24.2.2
Input Registers
Data type
Description
Offset
100
Range of physical
value
Float
Disinfection
Float
pH
Float
Temperature
12
Float
Flow
16
Float
Hypochlorous Acid
020 ppm
-3 to + 16 pH
-500 to +1500 mV
-30 to + 120 C
0600 rpm
010 ppm
20
Float
24
Float
Raw Disinfection
mV
28
Float
32
Float
Raw pH
36
Float
Raw Flow
40
Float
Case Temperature
44
Float
Signal Output 1
020 mA
48
Float
Signal Output 2
020 mA
52
INSTRUMENT_STATUS
0100 A
15066000
-500 to +1500 mV
0100 Hz
-55 to +125 C
A-96.250.751 / 240212
Modbus
24.2.3
24.2.4
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote Off
no
yes
15
reset
Tab. 24-3
Address
Parameter Table
Modbus
Parameter
Addr.
2300
Object
Data Type
Description
Acc.
FIRMWARE_REC
Firmware Version
Type
1151
Version
Record
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History 1 Array
Array of CAL_HIST_1[15]
3001
1502
Cal_History 2 Array
Array of CAL_HIST_2[15]
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 240212
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
101
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
0100 %
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
102
A-96.250.751 / 240212
Modbus
25.
25.1.
Introduction
This description applies for the AMI Turbiwell, AMI Turbitrace and AMI Turbitrack.
The AMI Turbiwell, Turbitrace and Turbitrack are Nephelometric systems for the automatic and
continuous measurement of turbidity in water. All can be operated as an intelligent Slave device
on the Modbus/JBUS using twisted pair RS485 lines.
25.2.
Function Description
25.2.1
A-96.250.751 / 240212
Addr.
Type
Description
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
10
Alarm
11
Alarm
12
Alarm
13
Alarm
14
Alarm
15
Alarm
16
Alarm
Control Timeout
17
Alarm
Turbi Disconnected
18
Alarm
Turbi Overflow
19
Alarm
Laser off
20
Alarm
21
Alarm
22
Alarm
Cleaning Reagent
23
Alarm
Input Active
24
Alarm
IC MK41T56
25
Alarm
IC LM75
26
Alarm
IC PCF8574
27
Alarm
EEprom Microcon
28
Alarm
EEprom Motherboard
29
Alarm
EEprom Front-end
30
Alarm
Cal. Recout
31
Alarm
Wrong Front-end
103
Modbus
25.2.2
Input Registers
Data type
Description
Offset
104
Range of physical
value
0300 FNU
Float
Turbitity
Float
Float
Flow
12
Float
Turbitity raw
0250000 Hz
16
Float
5002400
20
Float
Raw Flow
24
Float
V-Signal
08 V
28
Float
V-Reference
08 V
32
Float
Range Gain
36
Float
Case Temperature
40
Float
Signal Output 1
020 mA
44
Float
Signal Output 2
020 mA
48
INSTRUMENT_STATUS
-25 to +250 C
0100 l/h
0100 Hz
2535
-55 to +125 C
A-96.250.751 / 240212
Modbus
25.2.3
25.2.4
Single Coil
Addr.
Description
OFF (0)
ON (1)
01
open
close
Hold
no
yes
Remote Off
no
yes
Start Calibration
no
yes
15
reset
Tab. 25-3
Address
Parameter Table
Modbus
Parameter
Addr.
Object
Data Type
Description
Acc.
Type
2300
1151
Version
Record
FIRMWARE_REC
Firmware Version
2400
1201
Num_Evt
Simple
Unsigned16
2402
1202
Num_Log
Simple
Unsigned16
3000
1501
Cal_History 1 Array
Array of CAL_HIST_1[15]
Calibration History
3001
1502
Cal_History 2 Array
Array of CAL_HIST_2[15]
Verification History
Event_Rec
Record
EVENT_REC
Log_Rec
Record
LOG_REC
A-96.250.751 / 240212
YYYYYY
MMMM
DDDDD
HHHHH
MMMMMM
SSSSSS
Year
063
Month
112
Date
131
Hour
023
Minutes
059
Seconds
059
105
Modbus
Parameter
INSTRUMENT_STATUS
Description
Range
32
Record
Language
Unsigned8
Language Selection
0 = German
1 = English
2 = French
3 = Spanish
RunMode
Unsigned8
Run Mode
0 = Run
1 = Hold
2 = Off
3 = Grab
AlaMode
Unsigned8
Alarm Mode
0 = Off
1 = on
2 = Blinking
MaintMode
Unsigned8
Maintenance Mode
0 = Off
1 = on
2 = Blinking
RelMode1
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
RelMode2
Unsigned8
0 = None
1 = Limit Hi On
2 = Limit Hi Off
3 = Limit Lo On
4 = Limit Lo Off
5 = Timer Off
6 = Timer On
7 = Percent Control
8 = Valve
9 = Bus
PercDos1
PercDos2
0100 %
MaintValue
0100 %
AlarmsActive
AlarmsActive
AlarmsLatched
AlarmsLatched
0100 %
AlarmsActiveE
AlarmsActiveE
AlarmsLatchedE
AlarmsLatchedE
TimeStamp
end Record
106
A-96.250.751 / 240212