Beruflich Dokumente
Kultur Dokumente
.!
CHAPTER
1
Introduction
. .
"'
A ventilator is a device used to move gas into the lungs. The
design of the ventilator may be such that lung inflation is
accomplished by applying po~itive pressure to the airway or
by applying subatmospheric pressure around the chest.
In acute care institutions, lung expansion involving subatmospheric pressure is not as common as positive-pressure
ventilation. However, a brief review of the principles of
negative-pressure ventilation is essential to a comparison of
both methods of lung inflation .
Pressure Manometer
IM()IItM~ptJIIINo
Negative
Pressure
Control
' .
NEGATIVE-PRESSURE VENfiLATION
The use of negative pressure to expand the lungs dates
back to the start of the nineteenth century, 1 when it involved
enclosing all of the body but the head in a cylindrical tank
called the iron lung2 5 (Fig 1- 1). In subsequently developed
models the anterior portion of the chest and abdomen is
enclosed in an airtight shell (thoracoabdominal shell) (Fig
I-2,A and B) or the patient is fitted with a zippered gar-
INSPIRATION
EXHALATION
Bellows
Rotating
Wheel
FIG 1-1.
Iron lung. All but the head is enclost"~ in a sealed chamber. Slowly revolving wheel imparts reciprocal motion to bellows assembly connected
to chamber. When bellows expand, subatmospheric pressure generated within chamber causes chest to rise and inspiration to begin. During
upward motion of be llows a one-way valve opens and returns pressure within chamber to atmospheric. Chest recoils to normal position and
exhalation begins. Amount of positive and negative pressure can be controlled independently.
CHAPTER
1
Introduction
..
.
"
.,
' .
NEGATIVE-PRESSURE VENfiLATION
The use of negative pressure to expand the lungs dates
back to the start of the nineteenth century, 1 when it involved
enclosing all of the body but the head in a cylindrical tank
called the iron lung 2 5 (Fig 1-1). In subsequently developed
models the anterior portion of the chest and abdomen is
enclosed in an airtight shell (thoracoabdominal shell) (Fig
l-2,A and B) or the patient is fitted with a zippered gar-
A
INSPIRATION
' Negative
Pressure
Control
Bellows
Rotating
Wheel
FIG 1- 1.
Iron lung . All but the head is enclosf'~ in a sealed chamber. Slowly revolving wheel imparts reciprocal motion to bellows assembly connected
to chamber. When bellows expand, subatmospheric pressure generated within chamber causes chest to rise and inspiration to begin. During
upward motion of bellows a one-way valve opens and returns pressure within chamber to atmospheric. Chest recoils to normal position and
exhalation begins. Amount of positive and negative pressure can be controlled independently.
Suprasternal
Notch
CHEST SHELL
Flexible Hose
To Pump Unit
Straps
Pubic Crest
--
FIG 1-2.
A, position of chest shell used for negative-pressure ventilation. Inspiration is initiated when pump unit
generates subatmospheric pressure in airtight shell. When subatmospheric pressure is released, exhalation begins.
8 , cuirass shell used for negatve pressure ventilation. Patient i~ placed in supine position and cuirass is stabilized with
the use of straps and posts. Method of ventilation is identical to chest shell unit.
Basic Concepts
"'
abdomen
the body
I ro1 11 1he neck downward in an airtight bag (Fig 1-3, C). In
:.II 111odds the garment is held off the chest by a plastic shell
1 hr
t l ,,, I
: !) I
.r-
(
I
Arm
straps
Airtight
zipper--~
r
r
Garment shell
connection
(~
Pump unit
connection
Gannent shell
FIG 1-3.
Airtight garments used for negative pressure ventilation. A, garment is sealed at neck, arms, and legs. B, garment is sealed
at '!eck, arms, and waist. C, patient is placed in bag sealed at the neck and arms. To keep garment off patient' s chest and
enhance ventilation, a shell (D) fitted with a pump connection that extends through garment opening is used . Method of
ventilation is identical to shell units .
Introduction
PUMP UNIT
ToChe&Un'' ~ ,
'
~~ O
Expiratory
Pressure
Control
Inspiratory
- - - - Pressure
l l__~an--~
Control
Electric
Motor-
EXHALATION
INSPIRATION
FIG 1-4.
Schematic representation of pump unit used to provide negative pressure ventilation to shell units or garments.
Pump unit consists of piston connected off center to a slowly revolving wheel. The downward stroke of the
piston provides subatmospheric pressure to chest unit or garment to initiate inspiration. The upward stroke of the
piston rel~ases the subatmospheric pressure and allows chest to recoil to normal resting position and allow
eilialation. Amount of negative or positive pressure generated can be controlled independently.
POSITIVE-PRESSURE VENTILATION
The process of lung inflation by use of positive pressure
is similar in principle to mouth-to-mouth artificial ventilation. In this approach the rescuer exhales into the victim's
airway and directs positive pressure into the victim's lungs.
When the victim's chest has expanded to a suitable level the
rescuer stops exhaling and releases the pressure, and the
victim's lungs are allowed to empty. The procedure is then
repeated at a frequency appropriate to the victim's size.
When a device is used to inflate the lungs, the device is
called a ventilator. When positive pressure provides the
means for lung inflation, the process is called intermittent
2.
3.
4.
Basic Concepts
f'ltc
1> .. "
! I I }'
5 ):
.\
4.
Although the functional characteristics among ventilators are many, all ventilators must provide these four basic
l11nctions. Understanding the rationale behind this system of
classification enhances the clinician's ability to predict the
behavior of any ventilator in the clinical setting. In the
:'cclions that follow, the method used to determine the classification of each phase is closely examined.
BASIC CONCEPTS
Although current literature clearly indicates that the
availability of diverse ventilatory support is essential for the
intact survival of the critically ill,20 - 46 ventilator therapy is
not a perfect substitute for natural breathing. A ventilator
that can exactly mimic the normal physiological responses
of spontaneous ventilation has not been invented. We must
therefore concentrate on improving our knowledge of ventilators and the techniques involved with their use to mini-
Changeover from
insp'ir~ory phase
Changeover from.
to
expiratory phase
inspiratory phase
Inspiratory Phase - i --
expiratory phase
to
j
:-
- Expiratory Phase -
_....._
I
J
l
a:
:;)
"'"'a:w
a L---------~~-----------=~==-----J-------TIME - -- - -
FIG 1-5.
Four phases of respiratory cycle on controlled ventilation.
FIG 1-6.
Using balloons to describe the concept of compliance. (See text for de-
scription.)
Introduction
C = !J..V
!J..P
where
/, V
LP
C
Change in volume in li te rs
Ch ange in prcssu n.: in ern H2 0
Compli ance in L!c m H 2 0
Cr:
0.2
_
02
X 0.2
L
+
_ = 0 .1 /em H 20
02
Laminar Flow
Laminar flow is defined as a smooth, orderly flow in
which the particles constituting the fluid move in a path
parallel to the wall of the tube, as in Figure l-8 ,A. The
flow moves as concentric layers along the tube . The layer of
fluid next to the wall of the tube has zero velocity. Layers of
fluid at progressively greater distances from the wall have
higher velocities, with the maximum velocity occurring at
the very center of the tube.
Since the layer of fluid next to the wall of the tube is
essentially at rest, the frictional forces between the fluid and
the wall of the tube cannot be responsible for the drop in
pressure. In fact, the pressure drop along the tube during
laminar flow is caused by the viscosity of the gas. 65 - 66
. Viscosity can be thought of as the fluid's internal resistance
to flow and is caused by friction between the molecules of
the fluid. Different gases have different viscosities, as
shown in Table 1-1. It is interesting to note that if the fluid
is a liquid, the viscosity decreases with increases in tern- .
FIG 1-7.
When a steady flow of gas (\!) passes through tube , energy lost along tube
is demonstrated as drop in pressure(~) between ends of tube (P 1 - P2 ) .
Basic Concepts
Velocity head
Laminar flow
Velocity head
Turbulent flow
=Kx.!
=KX1
FIG 1-8.
Velocity profile of gas particles moving along tube during A, laminar flow,
and B, iurbulent flow.
TABLE 1-1.
Viscosities of Common Gases
'""
.....,
Oxygen
Viscosity
@ 20C
I0- 4 Poise*
Helium
Air
Nitrogen
Carbon
Dioxide
=K
l
0.0625
=K
16
.\
2.00
1.96
1.81
1.74
1.46
where
PI - P 2 = Pressure drop along the tube
K = Constant of proportionality, which
includes flowrate and viscosity of
the gas
L = Length of the tube
r = Radius of the tube
The relationship between the length of the tube and the
pressure drop along the tube is linear; that is, if the length
where
PI - P2 = Pressure drop along the tube
fJ- = Viscosity of the gas
L = Length of the tube
'lT = Pi (a mathematical constant)
r = Radius of the tube
V = Flowrate of the gas
In the human airway, conditions thai predispose to excesses in bronchial secretions and constriction oLthe smooth
muscles of the bronchi all contribute to the narrowing of the
airway and increase the effort required to move gas into and
Introduction
10
where
where
Turbulent Flow
A transition from laminar flow to turbulent flow takes
place above a certain flowrate.69 The point at which this
happens is known as the critical flowrate of the gas . In
turbulent flow the fluid particles no longer travel in straight
lines parallel to the wall of the tube but move in a haphazard
fashion across and opposite the general direction of flow, as
demonstrated in Figure 1-8 ,B. With a large number of
random particle fluctuations , gas in turbulent flow travels
along the tube in a uniform velocity, and the velocity of gas
closest to the wall of the tube is the same as the velocity at
its center. With turbulent flow the factor's responsible for the
pressure drop along the; tube include those described for
laminar flow but also involve gas density. In fact, the density of the gas becomes more important than its viscosity.
Airway Resistance
From the preceding discussion we conclude that the
pressure drop across a given tube is determined by the product of resistance and flowrate:
P1
P2
= resistance X flowrate
= em H 20 /Lisee x Lisee
=em H 20
'
.
Resistance
'
'
; ---t-
1'
P 1 - P2
fl
owrate
Airway resistance is defined as the pressure drop between the mouth and the alveoli divided by the flow-.
rate. 5 1 69 If P 1 and P2 are substituted for mouth pressure and
alveolar pressure respectively, the relationship that describes airway resistance is given:
where
.l
FIG 1-9.
Graph relating pressure drop along tube to type of flow present in tube.
PM
PA
V=
RAw
Flowrate in Lisee
Airway resistance in em H2 0 /Lisee
-~
'
Basic Concepts
r.~
..~
(~
r
r-
,r-
Mechanics.
,.--
r
.r-
r(
During a spontaneous inspiration the diaphragm contracts and moves downward. At the same time the rib cage
swings outward and elevates the sternum. This maneuver
enlarges the volume in the thoracic cavity and expands the
lungs. As the lungs expand, the pressure within them becomes less than atmospheric and gas is drawn down the
airway and into the lungs. Inspiration contil!ues until lung
pressure rises to equal atmospheric pressure.
Because the lungs and thoracic wall contain elastic tissue, exhalation is passive. When the diaphragm and the
muscles of the rib cage relax, the lungs and thoracic cage
recoil to their original size. As this happens the volume of
the thoracic compartment is reduced, the pressure within
the lungs becomes slightly greater than atmospheric pres~
sure, and gas moves out of the lungs until alveolar pressure-
and atmospheric pressure are again equal.
The lungs never completely empty;. the volume .,that
remains in the lungs at the end of a quiet spontan~ous exhalation is called the functional residual capacity (FRC).
Cardiovascular Responses to Spontaneous Ventilation
Cardiac output is defined as the product of stroke volume and heart rate. Stroke volume is the amount of blood
pumped from the heart with each beat. The more blood
returning to the heart by way of the veins (venous return),
the greater the stroke volume and cardiac output.
Blood pressure is the force of blood exerted against the
inner walls of an arterial vessel. Since blood pressure decreases markedly as blood travels through the arterial system
into venules, blood returning to the heart cannot entirely be
the result of heart action. 71 Spontaneous ventilation plays an
11
12
Introduction
SPONTANEOUS
VENTILATION
CONTROLLED
VENTILATION
Expiratory
pause
Volume
Expiratory
pause
Q
FRC+-------------~--
Alveolar
pressure
Intrapleural
pressure
O+-----------
Venous
return
4. The positive pressure also compresses the small pulmonary vessels, especially those in direct contact with alveolar pressure. This impedes pulmonary blood flow and
increases the work of the right side of the heart.
5. During positive-pressure ventilation, as the central
venous pressure (right atrial pressure) rises, venous return
to the heart falls.
Although the effects of these responses are not uniform , the reduction in the right heart output can lead to a
reduction in systemic blood pressure in- the hypovolemic
Basic Concepts
ZONE 1
....
Upright
Standing
ZONE2
ZONE3
Left Lateral
recumbent
Supine
FIG 1-11.
The concept of zones to illustrate lung perfusion to body position. In the
upright body, perfusion is least at the apices (zone 1) and greatest at the
bases (zone 3). Perfusion changes with body posjtion because of gravitational forces.
13
Ventilation
The functional residual capacity (FRC) is defined as the
gas volume that remains in the lungs at the end of a normal
exhalation. An inspired volume from the point of FRC is
called the tidal volume; it is the volume that enters the lungs
with each breath.
There are several reasons for the nonuniform distribution of ventilation throughout the lungs. The first is the
14
Introduction
APEX
ZONE f
CJ
ZON2
Pulmonary
Artery (Pa)
Alveolar
Pressure (PA = 0)
Pa > PA> Pv
c;
Pulmonary
Vein (Py)
ZON3
...
BASE
FIG 17 12.
Simplified zone inodel of the upright lung showing the factors responsible for the distribution of perfusion and alveolar diameter
at resting lung vol.ume. The vertical hydrostatic pressure results from gravitational forces. (See text for description.)
I . At FRC the uppermost alveoli are far more expanu~u than those of the lower regions. Therefore they are
Jess compliant than those of the lower regions because they
already contain a proportionately larger volume and will
resist further ex pansion.
2. The chest symmetry and movement of the ribs . The
lower ribs are more mobile and displace more volume than
the upper ribs.
3. The movement of the hemidiaphragms, which expands the lower regions more than the upper regions .
Basic Concepts
15
A.
...
B.
V 0 anat
V 0 alv
FIG 1-13.
Normal anterior view of lung scan (supine position) showing A, ventilation, and B, perfusion. (Courtesy Nuclear Medicine Department, Victoria
Hospital, London, Ontario, Canada.)
1ml/lb
150 ml
If this person has a tidal volume of 500 ml, then alveolar ventilation per breath is approximately
VA= VT- Yo
= 500 ml - 150 ml
= 350 ml
16
Introduction
sustained periods of hyperventilation the patient may experience dizziness , numbness, and tingling of the extremities.
If the Paco 2 is lowered even further, tetany may result. 69
This condition is characterized by increases in neuromuscular excitability, involuntary muscle contracti ons, and
spontaneous twitching .
100
90
80
70
60
PaC02
50 40
10
FIG 1-14.
Relationship between alveolar ventilation and Paco2 . Intersection of curve
and shaded area indicates normal range.
Basic Concepts
....
= --=-Cv
Cp
Vs
+where
VT
Vs
Cp
Cv
VT = - - X 500
1 + 50
50
=1~
=.!_X
500
= 250 ml
X 500
17
where
VT = Patient's tidal volume
Vs = Volume setting at the ventilator
PPK = Peak pressure
f = Compliance factor
It is important to note that the above equation provides
but a reasonable approximation of the actual volume that
reaches the patient. There are several factors that influence
the results. One is the location of the pressure sampling
port. Some ventilators sample the pressure at the airway
while others sample the pressure before the humidifier. One
mode of ventilator therapy provides two different sets of
pressure readings during the same inspiratory phase. These
18
Introduction
V~. N
III/11 /J II
II
rtl.ATOR
!! __:_~~
rJ_I~.
COCO DODD
fo-100ml-j
+--
SYRINGE
EXHAUST PORT
OCCLUDED
I'
.. ~
~---- 1
HUMIDIFIER
FIG 1-15.
Method used to determine tubing and ventilator compliance. (See text for description.)
...
will be evaluated under the heading Estimating Lung Mechanics.
There is yet another ventilatory technique that holds the
lungs inflated at a predetermined pressure during the expiratory phase (detailed in Section Four). This procedure,
called positive end-expiratory pressure or PEEP, must be
included in the calculations. When PEEP is used the equation should be modified to read
VT
where
VT =
Vs =
PPK =
PEEP =
f =
Minute volume
= Resprratory
.
rate
The more sophisticated microprocessor-controlled ventilator usually has a built-in VMD. This device takes the
ventilator and circuit compliances into account and displays
a plausible assessment of the patient's actual tidal volume.
Nevertheless , it is good practice to periodically measure the
PRESSURE SAMPLING
All ventilators must provide a means for sampling and
displaying the pressure developed during the process of
lung inflation. If the pressure displayed was the result of
only the patient's lung characteristics it would provide the
clinician with valuable information about the condition of
the conducting airways and lungs. Unfortunately, this is not
the case, because the pressure reading is also a reflection of
Basic Concepts
19
A
VENTID\TOR
FIG 1-16.
Placement of volume measuring device (VMD) to indicate patient's exhaled volume. (See text for description.)
Figure 1-17 ,C shows how weight is significantly reduced by extending the expiratory limb and exhalation manifold back to the ventilator. However, the circuit still requires a sensing line at the patient connection.
Figure 1-17 ,D displays how most of the problems are
solved by keeping the sensing line i~side the veptilator and
sampling the pressure in the expiratory limb pro_x.imal to the
exhalation valve. It is interesting to note that although the
pressure builds up in both the inspiratory and expiratory
limb during the positive-pressure breath, the position of the
sensing line does not sample the circuit resistance, because
gas does not flow in the expiratory limb during the inspiratory phase. The problem of moisture accumulation in the
sensing line still exists but has been minimized in most
ventilators by purging of the line with a small continuous
flow of dry gas. The sensing line can also be heated to
promote evaporation.
Introduction
20
A
' Exhalation
_; ManKold ,,
Expiratoty
Umb
Patient
Connection
FIG 1-17.
Common pressure-sampling sites found in ventilators. (See text.)
2.
3:
4.
Plateau Pres.sure
Basic Concepts
2.
End-Expiratory Pressure
The end-expiratory pressure is defined as the pressure
maintained in the lungs during the expiratory pause. The
lungs are normally allowed to empty to atmospheric pressure, as in Figure 1-18. However, the pressure can be
clinician-selected to b~low-atmospheric pressure (negative
end-expiratory pressure, NEEP) or to above-atmospheric
End Expiratory
Pressure
Plateau
Pressure
Peak
Pressure
I.,_ inspiratory
1
i - -- -- - lnspiration time
Pause
End Expiratory
Pressure
r--
Expiratory Pause
- - - - - - . j..<---- - - - -
- - - - - - -- -- -1>1
1---- - - - -- -- - -- -
21
FIG 1-18.
Common pressure levels that may be observed during the respiratory cycle on controlled ventilation.
22
Introduction
a:
::>
12
w
a:
Q_
RESPIRATORY CYCLE
------------1-~
FIG 1-19.
Mean airway pressure is defined as area under pressure curve for duration of respiratory cycle . Area can be approximated
by drawing equally spaced vertical lines extending from the pressure curve to line of zero pressure on time axis. When
pressure curve reaches zero pressure , equally spaced dots are substituted for vertical lines. When heights of each line
(including dots) are measured against corresponding pressure scale, dividing sum of pressure readings by number of
readings results in mean airway pressure (shaded area) .
105
2.
3.
4.
5.
I : E ratio
Inspiratory hold
Expiratory resistance
Positive end-expiratory pressure
Negative end-expiratory pressure
I: E Ratio
Within each respiratory cycle a period of time is devoted to inspiration and a period of time is devoted to exhalation. A quantitative comparison of one with the other is
called the I:E ratio. For instance, an inspiration time of 2
seconds and an exhalation time of 4 seconds yields an I:E
ratio of I :2 .
Since the area under the pressure curve varies directly
with the length of inspiration and inversely with the length
of exhalation, the greater the I:E ratio, the greater the mean
airway pressure (Fig 1- 20). Simply stated, an I:E ratio of
I: I results in a greater increase in the mean airway pressure
than an I: E ratio of 1:3.
Some ventilators have inverse I:E ratio capabilities; for
example, 2:1 and 3:1. When ventilation techniques with
Inspiratory Plateau .
Inspiratory plateau-also cailed inspiratory hold, inflation hold, or inspiratory pause-is that portion of the inspiratory phase in which the lungs are held inflated at a fixed
level of volume (Fig 1- 21 ,B) or a fixed level of pressure
(Fig 1-21,C) until a predetermined time has elapsed (see
also Section Three). The longer the lungs are held inflated,
the greater the area under the pressure curve and the greater
the mean airway pressure.
Expiratory Resistance
Expiratory resistance, or expiratory retard , has been
used frequently during mechanical ventilation to mimic
pursed-lip breathing, which is commonly seen in patients
with chronic obstructive pulmonary disease. The procedure
supposedly splints the airways and facilitates uniform emptying of the lungs. The method employed during mechanical ventilation involves the placement of a variable resistance to the exhalation port of the circuit (see also Section
Four). The maneuver slows the exhaled flow from the lungs
and in the process contributes to an increase in mean airway
pressure (Fig 1-22,B).
Positive End-Expiratory Pressure
Positive end-expiratory pressure (PEEP) is a maneuver
that prevents the lung pressure from reaching atmospheric
pressure during the exhalation phase (details are found in
Basic Concepts
Standard conditions
I:E 1:2
23
Standard conalitions
Q)
::;
::J
(/)
(/)
(/)
(/)
~
0..
(j)
0::
'
~I
..
, ~
..
f---- 1
........
~
:J
(/)
(/)
, ~
-~ ,
~-.
::; .
"
Inspiration time
I:E 1:1
Inspiratory plateau
type 1
Q)
::;
(/)
Q)
f/l
0::
0..
~,
I~
..
, ~
.. ,
I~
.. ...
,
Exhalation time
I:E 1:1
..,
Inspiratory plateau
type 2
Q)
::J
::;
/ ct
(/)
f/l
f/l
f/l
a..
~I
E ~
Figure 1-2
FIG 1-20.
Effects of altering I:E ratio on mean airway pressure.- Stippled region
denotes normal area of p(!:ssure curve. Vertical hatched region represents
increase in area of curve and thus increase in mean airway pressure. A
shows normal area. In B, inspiration time was increased by decreasing
flowrate . In C, exhalation time was decrea8ed thus removing much of
expiratory pause at zero pressure.
Section 4). Because of its unquestionable value as a ventilatory adjunct, the use of positive end-expiratory pressure is
common. However, the elevated baseline pressure that results from its use also contributes to an increase in the mean
airway pressure (Fig l-22,C).
. - - - - 1-
r.
-j
. .
- .,...j.._-
- E - - --...~J
FIG 1-21.
Effects of inspiratory plateau on mean airway pressure. Stippled region
denotes normal area of pressure curve. Vertical hatched region represents
incre;se in area of curve and thus increase in mean airway pressure. A
shows normal area. In B, inspiratory plateau was established by holding a
fixed volume in the lungs for a predetennined time. In C, the plateau was
developed by holding a fixed pressure at the airway for a predetermined
time.
24
Introduction
A
Standard
conditions ~
I ~~ ~----E ----11.,...,1
Standard conditions
~ ~ ---..+j..
.--------E------~~1
B
Expiratory resistance
r~ - - 1
Negative end-expiratory
pressure (NEEP)
I~
-----~-----E -------~1
~I
c
Positive end-expiratory .
pressure (PEEP)
::J
"'"'
Q)
a::
--..~~....1 - - - - --
-------11~~
FIG 1-23.
Effects of negative end-expiratory pressure on mean airway pressure. Stippled region denotes normal area of pressure curve. Vertical hatched region
represents decrease in area of curve and decrease in mean airway pressure.
A shows normal area.
'""'
..
.... ,
~I -~~------E------11~1
FIG 1-22.
Effects of expiratory resistance (B) and positive end-expiratory pressure
(C) on mean airway pressure. Stippled region denotes normal area of
pressure curve. Vertical hatched region represents increase in area of curve
and thus increase in mean airway pressure. A shows normal area.
gas to the pleural space-a condition called tension pneumothorax. The probability of this disorder occurring increases in patients already suffering from lung damage. 108
The primary cause of lung rupture is overdistension of
the alveoli; it is the result of excessive volume rather than
pressure. To clarify this statement, say we have two balloons. The first balloon is inflated with air and allowed to
expand until it bursts. The second balloon is placed in a
container designed to limit its expansion. This balloon may
now be inflated using as much pressl}re as possible, but if it
cannot expand, it will never burst. This simple example
suggests that if the alveoli cannot be overinflated, lung rupture is unlikely to occur.
A common cause of alveolar overdistension is improper use of the ventilator and high levels of positive endexpiratory pressure.
CHAPTER2
\
Basic Concepts
31
Bellows
FIG 2-1.
Weighted bellows drive mechanism.
LINEAR-DRIVEN PISTON
Check valve
Check vaJve
FIG 2- 2.
Spring-loaded bellows drive mechanism.
32
pressure is generated during the forward stroke of the piston. However, because the connecting rod is attached offcenter to the large wheel, the piston does not move in the
cylinder at a constant speed. Consequently, the pressure and
flow developed by this drive mechanism vary with the motion of the piston.
This drive .mechanism has been in the field for quite
some time and is still found in at least one critical care
ventilator8 and in some ventilators specifically designed for
home care. The effects of a nonlinear-drive mechanism on
the pattern of flow and pressure entering the lungs are detailed in Section Two.
NONLINEAR-DRIVEN PISTONS
fn Figure 2-4 a constant-speed electric motor (not
shown) rotates a large wheel to which a connecting rod and
piston are attached. The arrangement causes the piston to
travel in a reciprocating motion in the cylinder, and positive
PRESSURE-REDUCING VALVES
Pressure-reducing valves (PRVs) are probably the most
popular type of drive mechanism and are used extensively
Check valve
FIG 2-3.
Linear-driven piston.
Rotating wheel
FIG 2-4.
Nonlinear-driven pistons.
Check valve
Basic Concepts
in microprocessor-controlled ventilators. As the name implies, a PRY reduces a high input pressure to a lower constant output pressure. 9 The high input pressure may originate from high-pressure cylinders, in which pressure in
excess of 2000 psig (13,790 kPa) is common, or from hospital station outlets, whei'e the pressure is usually maintained at 50 psig (344.75 kPa).
Regardless of the input pressure, the output pressure
from the PRY becomes the generated pressure. Some
PRYs have their output pressure preset as high as 50 psig,
while in others the pressure can be adjusted to only a few
em H20.
Figure 2-5 is a schematic representation of an adjustable PR V. The relationship that describes the functioning
principle of an adjustable PRY may be given by
p
_ (Fs I - Fs 2 ) - (PH X a)
L-
High pressure
(PH)
-------.. . .
FIG 2-5.
Adjustable pressure-reducing valve.
33
where
PL = Low or reduced pressure (generated pressure)
Fs, = Force of the large spring (adjusting spring)
Fs 2 = Force of the small spring (sealing spring)
PH = High input pressure (source pressure)
a = Area of the small seat
A = Area of large diaphragm
As mentioned earlier, pressure-reducing valves may be
adjustable or preset. The main difference between the two is
that in the former, the tension of the large spring (Fs) can
be regulated externally to adjust the output pressure. A sealing spring CFs) is used to prevent source pressure from
entering the PRY when the adjusting spring (Fs) is fully
relaxed. In the preset PRY, Fs, cannot be adjusted externally
and the output pressure is preset by the manufacturer.
Some ventilators do not have PRYs built within them and
are connected directly to the hospital station outlet. In such
'
34
Solenoid
Open
Solenoid
Closed
~3LOWERS
Fan
Inlet
FIG 2-6 .
Blower drive mechanism.
'
J CCtOC.
The behavior of injectors is similar to that of pressure-ducing valves in that the pressure of gas leaving the outlet
r
Air entrainment
Input
FIG 2-7.
Injector drive mechanism.
Output
Basic Concepts
Inial
35
Outlet
______ J______
U-tube manometer
Pressure Gradient
FIG 2-8.
Venturi tube, which was actually designed to meter flow. U-tube manometer indicates that when a gas flows through the tube, a pressure gradient is created
between the convergent inlet and the constriction of the tube. Pressure gradient can be translated to indicate flowrate.
When the secondary circuit contains bellows, the primary drive mechanism may be a pressure-reducing valve, a
blower, or an injector, such as shown in Figure 2-10.
The primary reason for the development of the double.circuit ventilator is that it provided a reliable method for the
delivery of predictable tidal volumes (see Volume Cycling
in Section Three). Present-day ventilators have returned to a
single-circuit mechanism but now incorporate electronic
measuring devices that can detect and respond to minute
changes in pressure, flow, and volume.
CIRCUITS
Direct Acting (Single Circuit)
A ventilator may be classified as having a direct acting
or an indireCt acting drive mechanism. If gas from the drive
mechanism is used directly as the source of pressure for
lung inflation, as assumed in the examples used thus far, the
ventilator is considered to be direct acting. Ventilators having a direct acting circuit have also been called single-circuit
ventilators_
Indirect Acting (Double Circuit) An indirect acting circuit is one that uses the direct
acting drive mechanism to operate another circuit. Ventilators having an indirect acting circuit are also called doublecircuit ventilators.
Secondary circuits may consist of a bag or a bellows
and may be powered by any drive mechanism. However,
when the secondary circuit comprises a bag, it is customary
to have the secondary circuit powered by a nonlinear-driven
piston, as in Figure 2-9.
PRESSURE-LIMITING MECHANISMS
the maximum pressure developed by the drive mechanism of the ventilator is called the generated pressure.
Since some drive mechanism can generate pressures as high
as 50 psig (= 3500 em H2 0), a level far beyond clinical
range, a pressure-limiting mechanism is built into every
ventilator to prevent excessive pressures from reaching the
lungs. The device can be preset or adjustable and is positioned between the drive mechanism of the ventilator and
the patient. Present-day ventilators have both, an adjustable
pressure-limiting mechanism, which is found on the control
panel of tbe ventilator, and a preset failsafe mechanism,
which forms an integral part of the ventilator circuit.
In Figure 2-11 ,A the mechanism consists of a springloaded valve assembly. Noimally the valve never opens, but
when the pressure within the patient circuit exceeds the
closing force of the spring, the disk lifts and the excess
pressure is vented to atmosphere (Fig 2-11 ,B). During this
time the lungs are held inflated at the set pressure level until
a cycling mechanism terminates the inspiratory phase.
\
36
Check valve
Primary circuit
Connecting rod
Piston
::::: ::.::: .:: :: :.: :: ...: ::.: :
r~
I
Rotating wheel
FIG 2-9.
Indirect acting ventilator (double circuit). Primary circuit is nonlinear-driven piston, and secondary circuit consists of large bag enclosed
within rigid chamber.
Check valve
'
-.
Secondary circuit
Primary circuit
Pressure-reducing valve
or
Blower
or
Injector
Primary drive mechanism
FIG 2-10.
Indirect acting ventilator (double circuit) . Primary drive mechanism may
be pressure-reducing valve, blower, or injector. Secondary circuit c,onsists
of bellows enclosed in rigid chamber.
Figure 2-12 demonstrates another type of pressurelimiting mechanism, which consists of two series resistarices and a parallel vent. Gas in the, exhalation' .valve line
travels to the exhalation manifold where it is used to power
the exhalation balloon valve during inspiration. The amou.nt
of pressure transmitted to the balloon valve is determined by
the back pressure created by the flow ~f gas through the
fixed resistance of the parallel vent. The tfow of gas through
the resistance of the parallel vent is in turn determined by
the position of the needle valve of the pressure-limit control. The greater the flow, the greater the pressure in the
balloon valve, and vice versa. When the pressure in the
patient circuit exceeds the sealing force of the balloon, the
excess circuit pressure squeezes past the balloon and spills
over to atmosphere. Once again the lungs are held inflated
until a cycling mechanism terminates the inspiratory phase.
The maximum pressur.e-lirnit control is preset but is otherwise identical to the pressure-limit control. It is designed as
a safety system and prevents the pressure from exceeding a
certain value.
In Figure 2::-13 ,A the pressure-limiting device comprises a disk, a rigid accumulator compressed by an adjustable force of a spring, and two electrical contacts . During
/
Basic Concepts
8
~ Control knob
Spring
Disk
From drive
mechanism
To patient
Patient circuit
FIG 2-11.
A, Pressure-limiting device consisting of spring-loaded valve. B, When pressure within patient circuit exceeds value set by tension of
spring, disk lifts and excess pressure is vented to atmosphere.
Pressure limit
control knob
EXHALATION
MANIFOLD
Exhalation
valve line
From drive
To exhalation
valve line
mechanism
PRESSURE LIMIT
MECHANISM
From drive
mechanism
FIG 2-12.
Pressure-limiting mechanism. See text for description.
Patient circuit
To
patient
37
- /
38
CONTROLLING FLOWRATE
When the generated pressure is high, it is not advisable
to connect the patient directly to the ventilator without some
means of controlling flowrate. In its simplest form the flowrate control is a high series resistance, which is interposed
between the drive mechanism of the ventilator and the patient (Fig 2-14). The relationship between the amount of
resistance offered by the control and the flowrate through
the patient circuit is inverse. Increasing resistance decreases
flowrate and vice versa.
In the chapters that follow, other methods of controlling
flowrate, including contributions made by microprocessor
technology, will be examined.
Pressure limit
control knob
SUMMARY
The drive mechanism of the ventilator provides the
source of pressure necessary to inflate the lungs. The maximum pressure the drive mechanism develops is called the
generated pressure of the ventilator.
The drive mechanism can be direct or indirect acting. A
ventilator is considered to be direct acting if the pressure
generated by the drive mechanism is used directly as the
source of lung inflation. These ventilators are called singlecircuit ventilators. A ventilator is said to be indirect acting
when the pressure generated from the primary drive mechanism is used to power a secondary circuit, which in tum
inflates the lungs. These are called double-circuit ventilators; they evolved to permit the deliv-ery of predictable volumes and the use of other ventilator adjuncts such as an
inflation hold.
Whether a ventilator has a direct or indirect acting
mechanism has no influence on the magnitude of the generated pressure and has little effect on the pattern of pressure and flow developed in the process of lung inflation.
State-of-the-art ventilators under microprocessor control are single-circuit and can be programmed to deliver any
waveform or ventilatory adjunct conceivable.
Since the pressure generated by some drive mechanisms can reach levels exceeding 3500 em H2 0, certain
-.
Electrical
contacts
Rigid
accumulator
From
patient circuit
FIG 2-13.
A, Pressure-limiting mechanism consisting of spring-loaded accumulator and electric contacts. B, When pressure in patient circuit exceeds
predetermined value, contacts meet and electric signal is transmitted to drive mechanism, which interrupts gas flow through patient circuit.
Basic Concept.,-
39
Flowrate
control
From drive
mechanism
To patient
Patient circuit
FIG 2-14.
Series resistance type of flowrate control. (See text.)
REFERENCES
1. Mushin WW, Rendall-Balcer L , Thompson PW, et a1: Automatic Ventilation of the Lungs, ed 3. Oxford , England,
Blackwell Scientific Publications Ltd , 1980.
2. Schreiber PJ: Ventilators. In Anesthesia equipment, Performance, Classification and Safety. Telford, Pa, North American Drager, 1972.
CHAPTER3
- INTRODUCTION
~ The only function of the ventilator in the inspiratory
TABLE 3-1.
,../"'
Flow
Generators
(High generated
pressure)
Microprocessor
Generated
(usually high generated
pressure)
Constant
Nonconstant
Increasing
Decreasing
Constant
Non constant
Decreasing
Constant flow
Nonconstant flow
Increasing flow
Decreasing flow
........
I-_
50
,..
!
The Initial Flow Waveform
( Figure 3-2 demonstrates that the initial flow into the
'-....lungs represents the flow of gas through the ventilator circuit and the patient's airway. Therefore the initial flow varies directly with
1.
= (Cp) L!cm
H 20
<Rv +
RAw) em H:iO/L!sec
= sec
'
Pressure
control
Pressure reducing
High inlet
pressure (50
MG'.Jih
~-
'j
Humidifier
~"
FIG 3-1.
Schematic represe:1tation, showing the essential components of a ventilator designed as a constant pressure generator.
51
-r---------------,
--
From
ventilator
Initial
flow (l./s)
.!-
10 . --
- - - - - - - --
--,
Initial
Mouth pressure
(em H 2 0)
AiJWays fill
lime
FIG 3-2.
Initial flow and initial mouth pressure levels. Note that only the airways are charged.
Generated pressure
Ventilator resistance
Airway resistance
10 em H2 0
4 em H2 0/Usec
6 em H2 0/Usec
.../ - ~
TC
=
=
52
-,
From
ventilator
Flow
(LJsec)
Mouth
pressure
(em H 2 0)
Lungs fill
1 TC
0.5
2 TC
3 TC
1.0
1.5
Time (sec)
FIG 3-3.
Initial and dynamic waveforms produced when a constant level of low pressure is applied to the airway. Broken diagonal lines
10 em H 2 0
4 em H20/Usec
0.05 L!cm H2 0
6 em H20/Lisee
+ RAw)
+ 6)
(Cp/2) X (Rv
= 0.02:5 X
= 0.25 sec
(4
TC = 3
0.25 = 0. 75 sec
Figure 3-5,B demonstrates that with the new time constant the flow waveform de~ays exponentially to approach
zen:~- from the same initial level in only half the previous
time. )
. -~ ..:
,..
= Cp
=
=
[Rv + (2RAw)]
0.05 X [4 + 2(6)]
0.8 sec
X
TC
0.8
2.4 sec
..
When the solenoid opens (see Fig 3-2), mouth pressuritn~es sharply but never quite high enough to reach the
level of the generated pressure for two reasons:
)
)
1~
1\
\
\
1\
\
1\.
"
.........,i'-.
'/
1/
TC
0.5
1.5 sec
TC
0 .25
= 0 .75
sec)
TC = 3
0 .8 = 2.4 sec
7
3
-~ -
--.
53
Time constants
_ ..G 3-4.
lll)iversal exponential curves. A, decay; B, rise. (See text for description .)
l
l
l
i
54
'
STANDARD CONDITlONS
COMPUANCE HALVED
RESISTANCE DOUBLED
l
1
,..-\
11
( -----.
Rowrate
(Usee)
-----.
10
Mouth pressure
(em
~ 0)
1 TC
2 TC
3 TC
1 TC
2TC
3TC
1 TC
2 TC
3 TC
0.5
1.0
1.5
0.25
0.5
0.75
0.8
1.6
2.4
Time (sec)
Time (sec)
..-...._.
Time (sec)
FIG 3-S.
Theoretical waveforms produced by a constant pressure generator. Diagonal lines of each set represent time constant of system.
Generated pressure
Ventilator resistance
Standard conditions:
Lung compliance
Airway resistance
10 em H2 0
4 em H2 0/Lisee
0 .05 L!cm H2 0
6 em H20/Lisee
SUMMARY
In the inspiratory phase a ventilator can be classified as
either a pressure generator or a flow generator. The classification is based entirely on the pattern of flow and pressure
developed during the process of lung inflation. The flow and
pressure pattern is determined by three factors:
. .
f~
1.
2.
3.
t____
REFERENCES
1. Mushin WW, Rendall-Baker L, Thompson PW, et a!: Automatic Ventilation of the Lungs, ed 3. Oxford, England,
Blackwell Scientific Publications Ltd , 1980.
2. Dupuis YG: A mathematical approach to the classification of
the inspiratory phase. Respir Tech 1978;. 14(4):9 .
3. Waters DJ, Mapleson WV.': Exponentials and the anaesthetist. Anaesthesia 1964; 19(2):274.
4. Spalding JMK: Pressure and duration of inspiration during
artificial_respiration .by intermittent positive pressure. Lancet
1955; 268:1099.
I '
L"
~
f
'
'
r=
.- CHAPTER4
In some ventilators it is possible for the drive mechanism tp___ be allowed to equalize with alveolar pressure. In Fig 4-1 a
generate pressures as high as 50 psig (= 3500 em H 20). ) high series resistance (flowrate control) is interposed be~
tween the generated pressure and the mouth and a solenc:~id
This situation is shown in Figure 4-1, where the generated
is used for initiating and terminating the mspiratory phase.)
pressure has been increased from 10 em H2 0 to 3500 em
In keeping as closely as possible with the conditioos'
H 2 0 ,-,
specified in Chapter 3, say that(the inspiratory phase is
.!Yith a generated pressure of this magnitude, it is
ended after a period of 1.5 secoillts-and that during this
clearly not advisable to connect the patient to the ventilator
time, a volume of 0 .5 L leaves the outJet of the ventilator.
~ without some means of controlling the flow. It should be
equally clear that pressures of this magnitude should never
If a volume of 0. 5 L is to leave the v~ntilator in 1. 5 seconds,
Pressure
control
Pressure reducing
'' .
I ,
r--,.
Humidifier
FIG 4-1.
Schematic representation of ventilator designed as constant flow generator. Note high generated pressure and position of flowrate control.
55
56
0.4 . - - - - - - - - - - - - - - - - - ,
From
ver1!ilator
Initial
flow (LJs)
)'
14,----------------,
'- t
------.
'-
Initial
Mouth pressure
'
(em H 20)
Airways
till
'
'-l
.......,
-!
0
lime
FIG 4-2.
Step rise in initial flow and mouth pressure generated by constant flow generator. Note that initial values represent airway filling .
51
From
venti lator
Flow
(Usee)
14 , . - - - - - - - - - - - - - - - ,
r
Mouth
pressure
(em H 2 0)
Lungs fill
~TC
530
1ime (sec)
1.5
FIG 4-3.
Dynamic waveforms generated by constant flow generator. Lungs fill after initial step rise in flow and mouth pressure.
= (Cp/2)
= 0.05/2
X
X
(Rv + RAw)
(10,600 + 6)
= 265 sec.
"
Figure 4-4,B clearly demonstrates that when the compliance is reduced b( one half, the flow waveform does not
,...--.., change appreciabl~)
l~ompared with the very high resistance of the ventilator (flowrate control), changes in airway resistance have
little influence over the total resistance of the system. Consequently, when airway resistance is doubled, the step rise
in 'the initial flowrate remains unchanged . However, the
time constant increases to
TC
Cp X [Rv + 2(RAw)J
0 .05 X [10,600 + 2(6))
= 531 sec
=
=
58
STANDARD CONDITIONS
COMPUANCE HALVED
RESISTANCE DOUBLED
04 1
r---------------------
Rowrate
(Usee)
25
rv\outh pressure
(em H20)
1.5
1.5
1ime (sec)
lime (sec)
1.5
1ime (sec)
FIG 4-4.
Theoretical wavefonns produced by constant flow generator showing effects of lung characteristics.
Generated pressure
Ventilator resistance
Inspiration time
Standard conditions
Lung compliance
Airway resistance
3500 em H20
10,600 em H20
1.5 sec
-
0.05 Ucm H2 0
6 em H2 0/Usec
=
=
i'
---....,
= (% rise
X generated pressure)
initial mouth pressure
= (0.006 X 3500) + 2
= 23 em H20
'_I,,
l
---....,
~
__J
-,
i
'i
~I
PM
= (%rise x
generated pressure)
initial mouth pressure
= (0.003 X 350Q) + 4
= 14.5 em H2 0 )
59
//
WAVEFORM ANALYSIS
In considering the waveforms produced when the generated pressure is high, it is evident that the waveforms
- differ significantly from those produced when the generated
pressure was low (see Chapter 3). When the generated pres~ ;ure was increased to 3500 em H 2 0, the flow waveform, for
all intents and purposes, remained constant regardless of
~. ung characteristics. This was due to the very large gradient
~that still existed between the generated pressure and alveoar pressure when inspiration ended. With no significant
_change in,the pressure gradient, there can be no changes in
.iow.
Mouth pressure could not rise to approach the level of
lhe generated pressure because of the resistance imposed by
'le flowrate control. Essentially, the mouth pressure waverorm was the result of this constant flow applied to the
- irway. Consequently, mouth pressure was influenced by
~ changes in lung characteristics.
. . . ._ The concept perhaps may be clarified with the use of a
_,...s.imple relationship:
From this relationship we can see that when the flowrate (V ) is held constant, any changes in alveolar pressure
- ,A) or in airway resistance (RAw) will have a corresponding effect on the mouth pressure (PM) - An increase in al:>~ .olar pressure as a result of a decrease in lung compliance
J.Xill cause an increase in mouth pressure. Furthermore, be....use the flow is not influenced by changes in lung condi, .tions, increases in airway .resistance must also cause an
uJCrease in the mouth pressure. Therefore when the gener___.<:d pressure is high, the mouth pressure waveform will
a.iter during the inflation process and also with changes in
. ~ 1g conditions.
From this discussion we conclude that aflow generator
.-..a ventilator that generates a pattern of flow that does not
Y1!fY during the inflation process, regardless of lung condi, ...ns, while the mouth pressure waveform is free to vary.
60
./
Most modern-day ventilators have high generated pressures and make use of a microprocessor-operated valve or
valves to replicate any waveform generated by a flow generator. It is interesting to note that because these ventilators
have high generated pressures they cannot reproduce the
waveforms of a pressure generator.
Although microprocessor technology is a complex
field , the way in which these ventilators shape the flow
waveforms is riot necessarily complicated.
Figure 4-5 shows the essential components of such a
mechanism. Once the clinician selects the desired waveform
from the control panel, the microprocessor circuit identifies
the option and then sets up a program for the proportional
solenoid. The proportional solenoid is not a simple solenoid
having only two positions, oN or oFF, but is capable of a
great number of intermediate steps, thus capable of shaping
any flow pattern. The action of the solenoid is analogous to
the clinician manually adjusting the flowrate control
throughout the inspiratory phase to achieve the desired
waveform. Of course, the microprocessor-operated solenoi<f .
is much quicker and far more accurate than the human hand.
Inspiration is declared when the solenoid opens. However, the solenoid does not open all the way as in CQnventional ventilators but just enough to accommodate the pre-
'>--;
Flow wavefonn
selector
Control panel
circuit board
Differential
transducer
pressur~
Microprocessor
circuit
!
I
'"""""'
From drive
mechanism
To patient
circuit
FIG 4-5.
Schematic representation showing essential components of microprocessor operated proportional solenoid. Flow pattern selected is compared
with actual flow to patient circuit by pneumotachometer connected to differential pressure transducer. When discrepancy occurs, microprocessor relays appropriate signals to proportional solenoid and condition is corrected. See Chapter 9 for description of pressure transducer.
"'"l ~ r; uurned
A microprocessor-controlled ventilator with an adjustable generated pressure has its limitations. Ifthe generated
pressure is inadvertently set too low and mouth pressure
approaches the level of the generated pressure, the feedback
loop mechanism would be unable to compensate for the
reduction in flow and the ventilator would beliave as a pressure generator.
SUMMARY
When the level of pressure generated by the drive
mechanism is far beyond what is needed in the alveoli to
::
);~; JH
.:!, __ , ,
Hr:
?:.i , .
~~
ffi'ri-8 -- f
'' ::~ :;
- ~~r ~~t ~"~i'':
:: : ;
. .,, ,
T;l:;c'l~iI-'-' I '~_,-11111111111111111111:1!1;11111111111111111111~-~H~~~~~-
Flowrate
A Final Note
;;::;
r r: .
(LPM)
-":::; ,, ,
40
..
:;:;
''
. .
;r:]:;
41
50
Pressure
(ern H o)
2
Time
61
..__ _ _ _ _ _ 3 sec. - - - - - - -
FIG 4-6.
Actual tracings showing action of microprocessor operated valve. Generated pressure reduced to 100 em H2 0 and lung
analogue adjusted until peak mouth pressure reached approximately 50 em H2 0. Note that flow pattern is maintained although
generated pressure is much Itiss than 5 times peak mouth pressure.
62
REFERENCES
1. Dupuis YG: A mathematical approach to the classification of
the inspiratory phase. Respir Tech 1978; 14(4):9.
2. Mushin WW, Rendall-Eaker L, Thompson PW, et al: Automatic Ventilation of the Lungs, ed 3. Oxford, England,
Blackwell Scientific Publications Ltd, 1980.
-.:....._
..
.1
!\
'
' 1
i
_.,_
..
,\
C--: 1APTER
5
The Nonconstant Flow Generator
,
f'l, ._,_, far it has been established that ventilators are classified
IS vlther flow generators or pressure generators. When the
~<- ~' rated pressure is high, the pattern of flow remains con;tant while the mouth pressure waveform is free to vary.
r\ '"' n {he generated pressure is low, the pattern of the
nouth pressure waveform remains constant while the flow
1../ ~ :form is free to vary.
_ Chapters 3 and 4 investigated the waveforms produced
>) . drive mechanism that generated a constant level of
>ressure. Consequently, when the generated pressure was
<h. , the ventilator was classified ' as a constant pressure
~e~rator, and when the generated pressure was high, the
leiutlator was classified as a constant flow generator.
' This chapter considers the waveforms produced by a
ype of flow generator whose drive mechanism consists of a
1c ~~ear-drive piston.
i__}~Figure 5-l a constant speed electric motor (not
:t~ .n) rotates a large wheel to which a connecting rod and
>istQn are attached. This arrangement causes the piston to
n._ . .! in a reciprocating motion in the cylinder. During the
'o~ard stroke of the piston, pressure developed in the cylnu-..r is directed to the patient circuit and inflates the lungs.
-A.lthough the wheel rotates at a constant speed, such a
nechanical linkage does not permit the piston to travel in
h<' -...ylinder at a constant speed.
This is demonstrated in Figure 5-2, where the upper
1c: -...of a large wheel is divided into equally spaced dots .
)i~ the wheel rotates at a constant speed, points a through
~ : Jng the circumference of the wheel pass a reference
)0 ~ (indicated by the ghosted arrow at the top of the
;vl,vd), in equal increments of time. However, when these
)C ts are synchronized to the position of the piston in the
;yhnder, it becomes evident that the motion of the piston is
lC ~ niform . In fact, the speed of the piston is lowest beween points a and band progressively increases to a maxIT ----. 1 speed between points d and e when the piston has
V
av
=Volume= 0.5
Time
1.5
= 0.33 Llset )
.--..
64
Mouth pressure
Humidifier
"FIG
5-l.
Schematic representation of a ventilator designed as a nonconstant flow generator.
- -
--
_.,..
..&
Wheel rotation
constant
H
ab
11
g
hi
Piston speed
not constant
FIG
5-2. associated wit.'l motion of piston connected away from center of rotating wheel. (See text for description.)
Phenomenon
ROTATING WHEEL
PRESSURE WAVEFORM
FLOW WAVEFORM
PISTON
65
.
_;___-'
L
-
LL
LL_
LL
~-'
LL
--'
__, LL
___.
_;__-1
L
__I
L
--+
------t
--+
L
___.
L
__I
FIG 5-3.
Aow and pressure waveforms produced by motion of piston connected off-center to wheel rotating at
uniform speed .
~r_
--
- - ---
:~~#:. ~-- ~;
u.
~~~ - -
------~----
---~-
--
----
. ,,
--
1-
'-~:-.'!~ti'::_.~
.L
66
vpk
o.637 x vpk
Effe~ts
.
vpk
X 1.57
0 .52 Lisee
of Lung Characteristics
= 0 .637
Vav
v.y
~esistance .hav\vj,rtually
Vpk
v.v x 1.57
= 0.33
v.v
1.57
In the preceding equation the average fiowrate was determined to be 0.33 Lisee, and if 0.5 Lis to be displaced by
the piston in 1.5 sec, the peak fiowrate must be
('
Flow
Vpk = Vav
1.57
"
0
1\
I
I
I
I
\
\
''
/
/
./
FIG 5-4.
Method of dete~pjng peak value of sine wave. During mechanical ventilation, peak value of wave represents
peak ftowrate, Vpk- (See text for de~cription.)
~-~
~j
~~
jj
~
t.~
O;c
PM
PA
= i;
STANDARD CONDITIONS
COMPLIANCE HALVED
RESISTANCE DOUBLED
RA
Pressure
(em H 2 0)
10
~7
oL
'
0 ;..--------. 6
0
~
('
Time (sec)
1.5
1.5
1.5
FIG S-6.
Wavefonns depicting effects of lung characteristics on mouth pressure wavefonn. In all instances lung inflation was accomplished by half
sine wave flow pattern .
SUMMARY
MICROPROCESSOR MECHANISM
The waveforms of the nonconstant flow generator can
be duplicated without the need of a wheel and piston as-
When the drive mechanism consists of a piston connected off-center to a wheel rotating at constant speed, the .
speed of the piston within the cylinder cannot be constant.
Because of this, the flowrate that develops during the forward stroke of the piston cannot be maintained at a constant
level. In fact , the motion of the piston generates a flow
waveform that assumes the shape of a half sine wave; because of the flow pattern, the pat_tern of pressure that develops_during lung inflation is sigmoid in shape .
Since the wheel rotates at a constant speed and is totally
uninfluenced by changes in lung conditions, the pattern of the
piston's speed must also be constant. Consequently, the pattern of flow generated during the forward stroke of the piston
remains equally unaffected by c~anges in lung conditions.
Furthermore, because the motion of the piston is maintained regardless of lung characteristics, the volume displaced
from the cylinder must also be constant. If the volume delivered
into the lungs is constant, then any changes to the lung compliance will have an effect on the mouth pressure waveform.
Moreover, any c~ange to the airway resistance while the pattern
of flow is maintained must also affect the mouth pressure.
( A flow gl!nerator has been defined as a ventilator that
generates a fixed pattern of flow into the lungs, regardless of
lung conditions, while the mouth pressure is free to vary)
.!
A
Flow waveform
STAND~&IOONDITIONS
0 50
t l{)wrate
(1/Ytr:)
m<>O~:Jll.....eJ.~J:roJm~
20
advent of microprocessors,
ventilators have as an option the
a half sine-wave flow waveform
wheel and
AI-
1.5
1.5
REFERENCES
1. Kacmarek RM, Mack CW, Dimas S: The Essentials of Respiratory Care, ed 3. StLouis, Mosby-Year Book Inc, 1990.
2. Scanlan, CL, Spearman CB, Sheldon RL: Egan's FuncU1mentals of Respiratory Care, ed 5. StLouis, Mosby-Year Book,
1990.
3. McPherson SP: Respiratory TherapyEquipment, ed 4 . St.
Louis, Mosby-Year Book, 1990.
4 . Deshpande VM, Pilbearn SP, Dixon RJ: A Comprehensive
Review in Respiratory Care, Norwalk, Conn, Appleton &
su~ (}ipy'HfuWffa&ill1~Y~~8f L, Thompson PW, et al: Auton/dtic Ventilation of the 'i~ngs, ed 3. Oxford, England,
is -~pressure because
Black\Yctl Scienti~ Publicati
td, 1980.
.
.
mechanism
6. K~by \~,JneSR~ ;El;
,
: Mech~1cal v~ntlFIG fi'NI7r= PA + (V X RAw)
latwn, in~B.urton 00,
e
,1
JE (eds). RespiraSchemaiic BlPt'ffirtft~ ri?~o9es~%. 'ffr6cnt !:~~~ to generate a
tory Care : A Guide tifdlfhm Ph.fJtice, Philadelphia,
half sine wa.Ye ~wfi!l(jrii>, Once waveform IS sHedM Eohi control panel,
.TR T.inoincott Co 1977 .
microproc~or sets fip piograrn for proportional solenmd to nmruc the
. Tnrs Va1ue musf also represent the final value for mouth
mcffis ~~ijtyfu-c\WB{Jtffid 'f~ from cen~e;: of r~tating_ wheel. Exa.~t
pressure because when the piston has completed its forward
sif n of ~olenoidlsmonil6r~3"clbh<fin'!Ydb'~W<W\hlt!M\kd&~!Wei"cm
To patient
circuit
5ffid'tk\?
1"2HAPTER
6
The Nonconstant
Pressure Generator
The most significant difference between a pressure generafor and a flow generator is tj:J.e generated pressure. Therefore
it is possible to transform a nonconstant flow generator into
a nonconstant pressure generator~ providing a means of
reducing the generated pressure.;
Mouth pressure
Humidifier
FIG 6-1.
Schematic representation of a ventilator designed as a nonconstant pressure generator.
70
71
Time
...
r'
pressure rises to meet it (Fig 6-4). As the generated pressure decreases from its peak value, mouth pressure also
decreases but only to the point where it meets alveolar pressure. From this level any further reduction in mouth pressure is prevented by the position of a check valve, which
closes the moment the generated pressure drops slightly
below alveolar pressure. This results in an inflation hold
that is maintained until the piston has completed its forward
stroke) Under s~dard lung. conditions, the .mouth pressure
waveform resultmg from thts arrangement ts demonstrated
in Fig 6-3,A.
Effects of Lung Characteristics _
It was shown in Figure 6-2 that the rate of flow
through the resistance of the vent determines,~e pattern of
pressure generated by the drive mechanism (~ms.sunost of
the flow is vented to atmosphere, changes in lung conditions
cause an unappreciable increase over the total vented flow.
Hence the pattern of pressure drop across the resistance of
the vent, and the mouth pressure pattern remains unaffected
by changes in lung conditions1Fig 6-3,B and C).
___/
THE FLOW WAVEFORM
In considering the complexity of the drive mechanism,
the simplest method of assessing the flowrate into the lungs
is with the use .of the formula
PA
v. = PM__::_:_ _.:.::
l
,_..,.
_,--..,
F1G 6-2.
RAw
_ The flowrate (V) into the lungs is determined by the difference between mouth pressure (PM) and alveolar pressure
(PA), divided by the resistance of the airway CRAw) .
72
STANDARD CONDITIONS
A
10
10
r-
,-
I
I
I
Mouth pressure
(e m H 2 0)
_,--....._
I
~
0.34
1...
Flowrate
(Lisee)
~,
Time (sec)
~.
1.5
1.5
FIG 6-3.
Theoretical waveforms
1.5
i
produce~
Stroke volume
Inspiration time
Average ftowrate
Peak ftowrate
Resistance of vent
Standard conditions
Lung compliance
Airway resistance
..{'
0
~
<!'()r most of the inspiratory phase, mouth pressure follows the generated pressure and alveolar pressure rises exponentially to meet it. When alveolar pressure and mouth
pressure are equal, the check valve in the circuit closes and
flow into the lungs stops) Under standard lung conditions
the flow waveform that results is demonstrated in Figure
6-3,A.
I .
\
~
'
"
I~
WAVEFORM ANALYSIS
73
Mouth pressure
~~~~7
'
','
'
Generated
pressure
'
'\
\
\
\
\
\
\
lime - - - - - 1 1
...
~
FIG 6-4.
General shape of mouth pressure waveform. (See text for description.)
Figure 6-6 shows the simplified schematic of the actual ventilator where the secondary circuit consists of a
respiratory bag enclosed within a rigid chamber. 2 From this
schematic we note that one of the modifications made to the
drive mechanism is the placement of a shaft connected to
the center of the piston. The shaft is of uneven diameter, the
design of which permits the ratio of inspiration to exhalation
time to be maintained at 1 : 2.
In Figure 6-7 the piston is connected to the wheel not
by a simple connecting rod, but by an arm that extends to
the right side of the shaft, as indicated by the broken line.
To simplify the concept of this modification, the parallel
resistance (emptying pressure control) has been omitted
from Figure 6-7.
In 1 the rotation of the wheel pulls the shaft and piston
forward, causing a pressure buildup in the cylinder. Pressure from the cylinder is directed to a pressure chamber
where it is used to compress the respiratory bag. The volume of gas expelled from the bag inflates the expiration
valve, opens a one-way valve, and is then directed to the
patient outlet.
In 2 the bag has expelled all of its volume, but the
piston continues its positive stroke. The mouth pressure
waveform that develops is shown at the top right corner of
each illustration.
In 3 the piston is still in its forward stroke so the bag is
held compressed and prevented from reexpanding. An inflation hold results, and pressure travels down the airway
until equilibrium is reached between mouth pressure and
74
COMPLIANCE HALV ED
RESISTANCE DOUBLED
Generated pressure
(e rn H,.O:
II
12
Mouth pressure
(ern H20 )
0.5
Flowrate
(Lisee)
lu
,.
Time (sec)
1.5
1.5
1.5
FIG 6-5.
Actual tracings produeed by nonconstant flow generator fitted with parallel vent as described for Figure 6-1. Ventilator displaced
stroke volume of 2 L and was connected to a lung analogue under following conditions:
Standard condition
Lung compliance
Airway resistance
Compliance halved
Lung compliance
Resistance doubled
Airway resistance
0.052 Ucm H 2 0
6.6 em H2 0 /Usec
- - - -- - - -- - -
~1/\liA~ Pressure
vvrv.,----01imit
control
Water
lock
Respiratory
bag
Humidifier
Air inlet+
Air outlet+
Compressor
- -0- -
C)
Spirometer
Emptying pressure
control
FIG 6-6.
Simplified schematic of ventilator classified as increasing pressure generator. (Courtesy Gambro Engstrom AB, Sweden.)
-- -- -- -
- -,
76
r--
....... --
..-
__
,.
,....
I
I
---
@
4
r---
-----
---
77
Emptying
pressure
control
To
pressure
chamber
,... I
----1>~---- - l>
Inspiration ends
FIG 6-8.
Modified drive mechanism of increasing pressure generator. 1, rod; 2, slider; 3, lever; 4, control; 5, plunger; 6, spring; 7, disk.
SUMMARY
The main difference between a flow generator and a
pressure generator is the generated pressure. Therefore it is
possible to transform any flow generator into its equivalent
pressure generator by providing a means of reducing the
generated pressure. This was demonstrated with the nonconstant flow generator. The method involved the placement
of a low parallel resistance in the circuit of the ventilator. The
paral4!1 resistance allowed most of the flow to be vented to
atmosphere, and the pressure drop across the resistance generated the pressure necessary for lung inflation.
The pressure generated by this drive mechanism was
low, and because of this, the flowrate into the lungs did not
remain constant but decreased as the lungs filled. However,
the pattern of pressure generated at the mouth remained
constant regardless of lung conditions. Therefore the fundamental action of this ventilator must be regarded as a
pressure generator. Since the mouth pressure was not held
constant but the pattern of the mouth pressure waveform
was repeated within every stroke, the ventilator is classified
as a nonconstant pressure generator.
When the drive mechanism of a nonconstant pressure
generator is used to power a secondary circuit, the generated pressure waveform differs significantly from the one
produced by the original drive mechanism. Although both
ventilators are considered to be pressure generators, there
are no similarities between the pressure waveforms generated (Fig 6-10). Therefore the ventilator can no longer be
classified as a nonconstant pressure generator.
STANDARD CONDITIONS
COMPLIANCE HALVED
RESISTANCE DOUBLED
1.5
1.5
so I
~n er ated
press u re
(em HJO)
22
,.
Mouth pressure
(em H 2 0 )
Flowrate
(Lisee)
Time (sec)
1.5
--..FIG 6-9.
Actual pressure and flow tracings produced by ventilator classified as increasing pressure generator. Volume in respiratory bag was set at 0 .5 L (air)
--._ and frequency was adjusted to 12 breaths per minute. Emptying pressure control (spring-loaded parallel vent) was adjusted to generate pressure of
50 em H2 0. Ventilator was connected to lung analogue under following conditions:
0.052 Llcm H2 0
6 .6 em H20 /Lisee
14 em H20 /Lisee
INCREASING
PRESSURE GENERATOR
NONCONSTANT
PRESSURE GENERATOR
.,..... ..
I
!!'!
:l"
lime
;1
j
,
..
~
~
a:
Time
..
FIG 6-10.
Differences between waveforms generated by nonconstant pressure generator and increasing pressure generator.
r-
CHAPTER
9
Pressure Cycling
PRESSURE-CYCLING MECHANISMS
G _yentilator is said to be pressure cycled if the inspira-
Pneumatic-Cycling
Mechanism
......
~-....
/
l.~Q!lse
98
Pressure Cycling
17
/
/ '
10
A
13
6
..
t r
12 ~+
15~
14
...
FIG 9-1.
Schematic representation of pneumatically operated, pressure-cycled ventilator. (See
text for description.) 1, Pressure inlet; 2, flow rate control; 3, communication holes;
4, diaphragm; 5, sliding valve; 6, iron disk; 7, iron disk; 8 , ambient magnet; 9,
pressure magnet; 10, control knob; 11, control knob; 12, exhalation valve; 13, pressure manometer; 14, test lung; 15, patient circuit; 16, pressure compartment; / 7,
ambient compartment.
the two basic principles of a fluidic element: wall attachment and beam deflection. 8
When a jet of gas is forced through an opening, as in
Figure 9-2,A, the turbulent jet flow causes a localized drop
in lateral pressure and draws in ambient gas . If the entrainment of gas on one side of the jet is limited by the addition
of an adjacent surface such as in Figure 9-2,B, a low
,.
99
~-
Ps
Ps
FIG 9-2.
Wall attachment phenomenon, or Coanda effect. (See text for description.)
FIG 9-3.
Schematic representation of flip-flop valve showing what is known as beam deflection.
~
~.l
f./
.j~
100
Pressure Cycling
\7
\7
."
.
or
c ontrol signal
--t> (negative)
__..
Control sign al
(positive)
C,
c2
0,
\7
\]
c2
02
\7
'
~7
c,
o,
\7
\7
1\
1\
c2
02
~7
1\
\7
c,
/1\
o,+
FIG 9-4.
A, Schmitt trigger is an integrated fluidic circuit made up of 3 proportional amplifiers and 2
flip-flop valves. B shows how beam deflection is accomplished by applying positive pressure
at C 1 or negative pressure at C2 .
'
101
- ~ifference
in splitter configuration, wall attachment is prevented and the input pressure source (P 5 ) leaves from both
--..utlets simultaneously. The output from the amplifier is
taken as the difference between output 0 1 and 0 2.
"' When the control signal at C 1 is equal to the control
~ignal at C2> the differential output is zero (Fig 9-S,A) .
_iowever, a control signal applied to C 1 and of greater mag_..~ itude than c2 causes the beam to deflect to 01' increasing
ne output flow from that port and decreasing the flow from
- utput 0 2 (Fig 9-S,B). A similar control signal applied at
c 2 has the opposite effect (Fig 9-S,C).
For the sake of simplicity this text has adopted the
symbol given to the Schmitt trigger by the National Fluid
~ :>wer Association (illustrated in Fig 9-6,B). 1
~ (The Schmitt trigger module (refer to Fig 9-6,B) has art
Jetl''Ort, P5 , two output ports, 0 1 and 02> and two control
-"~Orts, c) and C2. But because of its design, in the absence
vi control signals at either ports, the output must always be
c,
Ps
Ps
Ps
Control
signal
1----+-,~,...----i c2
+t------,lr-----i
Control
signal
c2
c,
iero
Output
decreased
output differential
Output
increased
c, t-------J.:.--.......j+
c2
Output
increased
Output
decreased
FIG 9-5.
Proportional amplifier. (See text for description.)
L..
SCHMITT TRIGGER
OR/NOR
FLIP-FLOP
Ps
Ps
'7
C, 1--/~. ---tC2
02
o,
02
FIG 9-6.
Fluidic symbols for 3 fluidic devices.
('
o,
o,
02
102
"
'
......,
Pressure Cycling
Adjustable
reference pressure
c2
C,
------r-----~~----~~------~-----,
o 2 ..(,.~
o,
c2
02
c2
To fluidic
elements
(Ps)
Inlet
Power valve
Flowrate control
pres~ure-cycled
'
'
'
--"'.
---.._
'
......_
.......,
......_
~
---..,
103
--It
---th____.,.:L_r:::;..
_ _ _ _ _ _ _ _ _,.lt, To patient
"
_SJ
' ctrcutt
To patient circuit
FIG 9-8.
Pressure-cycling mechanism consisting of spring-loaded diaphragm and
microswitch. Silhouetted diagram below indicates position of cycling
mechanism in inspiratory limb of circuit.
\.J 1
~ zumatic/Electronic
Type I
,__..__ Many pressure-cycling mechanisms consist of an adJ"table spring-loaded diaphragm and microswitch, 14 simi!_,.,.,_ to that depicted in Figure 9-8. The mechanism is placed
IH !!J.e inspiratory limb of the ventilator circuit.
~:. During the inspiratory phase, as pressure in the patient
cir2illrificreases, the diaphragm bows upward, opposes the
',.....,Nnward force of the spring, and moves the pivoted lever
upward and closer to the microswitch. When the pressure in
"""" patient circuit is great enough to cause the pivotfd lever
!~ trip the micros witch, an electrical signal is sent to a
..~croprocessor (an electroni.s- logic circ;uit) and relayed to a
~"'-l enoid, which closes to end the inspiratory phase. The
n1aximum pressure allowed in the system is determined by
~ tension of the spring, which is adjusted by the control
knob. The greater the sQ.ti,_ng tension, the greater the cycling
.~' ssure and vice versa..:)
Although this system is used more commonly as a
"';kup safety system for other cycling mechanisms, it can
,
>
Pneumatic/Electronic Type II
Pressure cycling can also be accomplished with the use
of a pressure transducer. Although there are many types of
pressure transducers, basically pressure transducers convert
mechanical energy into electrical energy. 15 - 19
In Figure 9-9 the transducer is placed in the inspiratory
limb of the ventilator circuit and electrical means are used to
measure the motion of the diaphragm. During the inspira, tory phase, the diaphragm is deformed by the action of the
pressure building up in the system. When the pressure
against the diaphragm equals an electronic reference potential set by the operator, an electrical signal is transmitted to
the microprocessor and relayed to the solenoid, which
closes to end the inspiratory phase. The amount of pressure
allowed in the system is therefore governed electronically
104
(Pressure Cycling
i
./
t
Primary winding
1nput
<:::::=::::><
Secondary winding
<===><
o~tp ut
Diaphragm
_.If
---1-----L.::;._
_________+~
To patient
CirCUit
Control knob
Input
To patient circuit
the initial setting of the pressure-cycling mechanism because the patient's lung condition is not known at that moment. Clinical judgment dictates the cycling pressure selected and is customarily in the range of 10 to 20 em H2 0.
The exact setting for flowrate is also impossible to establish
but is initially adjusted to provide an adequate inspiration
time.
Once these preliminary settings are made, the patient is
connected to the ventilator. The volume delivered at that
pressure and flowrate is determined with a volume measuring device. The final adjustments to the cycling pressure
and flowrate are then performed based on the outcome of
the measured data. If the tidal volume is too low, the cycling pressure is increased. If the tidal volume is too high,
the cycling pressure is reduced.
Keeping the patient's lung characteristics constant, the
length of the inspiratory phase is also influenced by the cycling
pressure and the flowrate selected. Adjusting the cycling pressure alone may require further manipulation of the flowrate
control to maintain inspiration time.
When any pressure-cycled ventilator is confronted with
changes in lung conditions, the only parameter that remains
constant is the cycling pressure. Consequently, changes in
lung compliance or in airway resistance will have an adverse effect on the tidal volume and inspiration time, as
demonstrated in Figure 9-10.
FIG 9-9.
Pressure-cycling mechanism cons1stmg of pressure transducer. Type
shown schematically i~ a linear variable differential transformer (LVDT).
Iron core placed symmetrically between electrical windings monitors motion of diaphragm. Silhouetted diagram beloW shows mechanism in in, spiratory limb of patient .circuit.
'
CLINICAL CONSIDERATIONS
When pressure-cycled ventilators are used in the clinical setting, the volume delivered into the lungs and the time
taken to deliver the volume are related to the cycling pressure, the flowrate, and the -lung characteristics of the patient. 20 Unfortunately, there is no simple way of predicting
L,
.i......,
--..
CHAPTER10
Flow Cycling
clockwise (Fig 10-1 ,C) , which narrows the channel opening and further reduces the f,owrate. Once the flowrate is at
a level that is not high enough to offset the effects of the
metal rod, the valve rotates shut (Fig 10-1 ,D). At that
moment the ventilator is said to have flow cycl~
From this discussion it should be clear that-the flowcycled ventilator most operate as a constant pressure generator since the generated pressure and alveolar pressure
must approach equilibrium to achieve the desired volume
exchange. In the flow-cycled ventilator the reduction in
pressure gradient as the lungs fill provides the necessary
decay in flowrate for cycling to occur. The theoretical waveforms produced by the ventilatm: must approximate those of
the constant pressure generator discussed in Chapter 3.
Flow cycling must not be confused with pressure cycling since in the latter, a predetermined pressure set at the
ventilator must be met to end the inspiratory phase. With
flow-cycled ventilators, the preset pressure (which is in fact
the generated pressure) is never quite reached because at the
moment of flow cycling, a small pressure gradient still exists and provides the terminal flow required for cycling.
The terminal flow in these ventilations ranges between
1 and 4 Llmin. 4 In the examples that follow the inspiratory
phase is considered complete when the flow has decayed to
an arbitrary value of 3 Llmin (0.05 Lisee)/
CLiNICAL CONSIDERATIONS
With flow-cycled ventilators the vohime delivered into
the lungs and the time taken to deliver the yolume vary with
the level of pressure selected and lung characteristics of the
patient. Therefore the initial set up of the flow-cycled ventilator involves adjusting the generated pressure to a level
between 10 to 20 em H2 0. 5 The patient is then connected to
the ventilator and the volume exchange from the pressure
' ii i
'
113
F1G 10-1.
Schematic representation of flow-sensitive valve. A, Expiratory phase. B, Inspiratory phase. C,
Terminal flow reached. D, End inspiration, begin expiration. 1, Gas inlet; 2, vane; 3, counterweight; 4, 5, channel openings.
selected is determined. The final adjustment to the generated pressure is then made based on the volume desired. 6
Increasing the generated pressure increases the tidal volume
and vice versa.
In the simplest form of the flow-cycled ventilator (such
as described thus far) the operator has no control over flowrate. When the generated pressure is held at a fixed level,
the length of the inspiratory phase depends entirely on the
behavior of the flow-cycling mechanism and lung characteristics of the paJient. This is demonstrated in Figure 10-2,
where the effects of lung characteristics on inspiration time
and tidal volume are presented. To co.nstruct the graph, the
generated pressure was adjusted to 10.5 em H20, and flow
cycling was assumed when the flowrate decayed to a level
of 0.05 Usee (3 Umin).
With changes in lung compliance, the flow-cycled ventilator behaves similarly to the pressure-cycled ventilator.
For instance, Figure 10-3,B demonstrates that when lung
compliance is reduced by one half, inspiration time and
tidal volume are also halved. Therefore, with changes in
lung compliance the pressure-cycled ventilator and the
flow-cycled ventilator differ only in the way in which the
inspiratory phase is ended.
In the flow-cycled ventilator the only method provided
to compensate for reduction in lung compliance is to increase the generated pressure until the tidal volume is restored. However, once the volume is restored, inspiration
time will always be less than observed during standard lung
114
Flow Cycling
0.7 5
A
Volume
(L)
0.5
Standard conditions
- - - - - - - - - - -
0 .25
1
x0 .25
i
i
f
~
..
INCREASING RESISTANCE
0.75
B
Volume
(L )
Standard conditions
_ _ _ _ _ xo
2:
x1
~
0.5
)(
)(
x5
""*---?<
I
I
f,
0.25
+
0
0.5
1.5
2
Inspiration time (sec)
2.5
FIG 10-2.
r
Effects of lung characteristics on volume and inspiration time with flow-cycled ventilator. Flow cycling occurs at 0.05 Lisee (3 L!min). A shows effects of lung compliance while maintaining airway resistance. B shows effects of airway resistance
while maintaining lung compliance.
Standard conditions
Lung compliance
Airway resistance
I
I
l
0.05 Ucm H2 0
6 em H2 0/Lisee
t
I
conditions. Since there is no flowrate control on the simplest form of the ventilator, nothing can be done to correct
the situation.
In the more sophisticated version -o f flow-cycled ventilator (Bennett PR-2 in Chapter 21) a variable series resistance (peak flow control) is positioned downstream to the
flow-sensitive valve (Fig 10-4). The peak flow control can
be adjusted to provide minimum or maximum resistance to
flow. Intermediate settings are also possible but the flowrate
cannot be completely shut off.
115
STANDARD CONDITIONS
COMPLIANCE HALVED
RESISTANCE DOUBLED
1.05
Flowrate
(Lisee)
10.31
Mouth pressure
(em H,O)
10
Alveolar pressure
(em H,O)
Volume
(L)
I
t
Time (sec)
1.5
1.5
1.5
FIG 10-3.
'
Theoretical waveforms produced by flow-cycled ventilator. Flow cycling
occurs at Of05 Usee (3 Umin).
10.5 em H 20
4. em H20 /Usee
Generated pressure
Ventilator resistance
Standard conditions
Lung compliance
Airway resistance
--
'
--
--
~::"'"'.a' .,.;.;.,;;;~-.~..;;.,,
~-
---
... ~- .
_ _, _
~ .. .
'
--
. . ;., "
----
--
. ~.
r~
!
116
Flow Cycling
TABLE 10-1.
Characteristics of a Flow-Cycled Ventilator
Flow Cycling
Parameter
Mouth
pressure*
Flow
Alveolar
pressure
Volume
Inspiration time
Compliance
t
t
Res istance
To
Compensate
for Cp
for RAwt
<t >
<t >
t
...
To patient
FIG 10-4.
Position of variables series resistance (peak flow control). (See text for
description.)
SUMMARY
A ventilator is said to be flow cycled if the inspiratory
phase is ended when the flow of gas through a flowsensitive valve decreases to a critical level. At the moment
of cycling, the pressure and volume in the lungs, along with
inspiration time, may all vary from one respiratory cycle to
the other. The only nonvariable is the terminal flowrate.
The flow-sensitive valve cannot compensate for changes
in lung compliance and in this respect behaves similarly to
the pressure-cycled ventilator. With reductions in lung compliance the volume in the lungs and inspiration time are decreased. To compensate, the generated pressure is increased
until the volume is restored . However, the maneuver does not
restore the length of-the inspiratory phase, which remains
shorter than originally observed . With the simplest form of
the flow-cycled ventilator nothing can be done to correct for
inspiration time.
In the more sophisticated versiop of the ventilator, a
variable series resistance (peak flow control) is positioned
downstream to the flow-sensitive valve and provides more
control over flowrate and the length of the inspiratory phase.
The flow-cycling mechanism reacts favorably to
changes in airway resistance. With increases in airway resistance the valve automatically compensates by decreasing
the flowrate. Although complete compensation is theoretically impossible, the reduction in flowrate allows more of
the generated pressure to reach the alveoli, and large fluctuations in tidal volume are minimized.
With changes in airway resistance the valve will attempt to maintain alveolar pressure at the expense of inspiration time. With minor increases in airway resistance the
increase in inspiration time may not be clinically significant. However, in the simplest form of the ventilator, the
operator has no control over flowrate, and the length of the
inspifatory phase cannot be fully corrected.
Some control is offered with the placement of a peak
flow control, but with pronounced increases in airway resistance, inspiration time can never be completely restored.
REFERENCES
1. Kacmarek RM, Mack CW, Dimas S: The Essentials of Respiratory Care, ed 3. StLouis, Mosby-Year Book Inc, 1990.
.
'
...
<.,/
r:
i
CHAPTER
11
I
j
I
I
Volume Cycling
1
I
II
I
i~
_../, ...
,.,
DOUBLE-CIRCUIT, VOLUME-CYCLING
MECHANISMS
Pneumatic/Fluidic
118
\ -;,.,_~.
"
' ,'
-,-
-~
The Change over from the Inspiratory Phase to the Expiratory Phase
'
119
-.''
Flip-flop
valve
'
I
c, t----k-----1
c,
::g:::::=::=::=:~:=::::::=:::~:=:=::E.
-
'
I'I
'
\
'
o, &
l[J
'
'
OR / NOR
c,
To patient
'
'
\
-----..
"
'
To exhalation valve
---._
Plate
'
\
:?
Volume control
crank
FIG 11-1.
Schematic representation of fluidically operated volume-cycled ventilator. (See text for description.)
__/
Pneumatic/Electronic
Microswitch
i'
II
/ "Figure 11-2 shows how volume cycling can be accomplished-with the use of a bellows and microswitch assembly.
The mechanism is found in the Ohio ventilators 4 described
in Chapters 37 and 38.
120
Volume Cycling
Electronic circuit
[~ l~- ----H
Volume
ooctml
jc'""'"
50 1
- - - - - l f -_ _ e_n_o_id_ _ _ _ _ _ _ _ _ _ _ _.Jj
From primary
drive mechanism
~stervalve
t1.-.
To exhalation va lve
_._----------------'--------__j
FIG 11-2.
1..
Potentiometer
In Figure 11-3 a potentiometer fitted with springloaded pulley is positioned outside and at the bottom of the
bellows canister. A spring-tensioned cord wrapped around
the pulley connects the bellows potentiometer to the bottom
of the bellows.
During the inspiratory phase, gas from the primary
drive mechanism inflates the exhalation valve, the canister
valve, and the bellows canister. As pressure in the canister
increases, the bellows rises and rotates the pulley of the
bellows potentiometer. This rotation is relayed electronically to the reference potentiometer on the control panel of
the ventilator. Once the bellows potentiometer has rotated to
a point where it matches the setting of the reference potentiometer, an electrical signal is transmitted to the solenoid,
which closes to end the inspiratory phase. At that moment
the ventilator is said to have volume cycled.
With this arrangement the bellows must always rise
from the bottom of the canister. The volume delivered into
the circuit is governed by the position of the reference potentiometer, which is calibrated to read volume. The higher
the reference setting, the greater the distance the bellows
must travel to match the reference settin.g and the greater the
The Change over from the Inspiratory Phase to the Expiratory Phase
121
To pat1 ent
\'
Canis ter
Volume contro l
Coed
Pulley
Bellows
potentiometer
Reference
potentiometer
ISI(canister valve
tLd~~~;;Hoo "'"
From pnmary
dnve mechanism _
-_
. ._ _ _ _ _ _ _ _ _ _ _ _
__..
_ _ _ _ _ _ _ _ _ _ _ ___J
FIG 11-3.
Use of potentiometers to accomplish volume cycling in double-circuit ventilator.
Volume control
- -
- I
I
I ,
Receiver
crystal
_g_
g
Rece ive r
-
~ Vo_rt_e_x_ _
--
-
..,.
Wedge
Tra nsmitter
crystal
Wedge
\..___)
To patient
A Transm itter
L_j
- J
Electronic
oscillator
FIG 11-4.
Ultrasonic flow transducer as volume-cycling mechanism.
- - -'
'-
'
an
..
:;:..:,"'.':::_ . : ...
--
' ..... : _.
: . : ......
~---.
.:..
---
- -- -
...
. . .,.. :
, 121
Volume Cycling
Volume coniro l
Electronic circuit
Strain gauge
Ventilator circuit
A
To patient
B
FIG 11-5.
Strain gauge used as volume-cycling mechanism.
:le CirCUit.
'""
During the inspiratory phase the flow of gas acts against
. .he flag, which by way of a small pin deforms the silicon
~rod (exaggerated in this schematic). This stretches the filament within the rod and thus changes its resistance. The
,.-- r:hange in resistance is monitored by an electronic circuit
built within the ventilator. Although the strain gauge actu~ally responds to flow,8 given an inspiration time, the signal
from the strain gauge is integrated electronically to measure
r"'volume. The solenoid closes when the volume measured
electronically through the circuit corresponds to a reference
"'setting on the control panel of the ventilator.
A similar volume-cycling mechanism is described in
Chapter 39 for the Siemens Servo 900C ventilator.
,....._, The Hot-Wire Anemometer
"
"
r-.
CLINICAL CONSIDERATIONS
Compared with the pressure- or flow-cycled ventilator,
the initial setup of the volume-cycled ventilator is straightforward. The operator merely has to determine what volume
is to be delivered and then adjust the specific dial on the
control panel of the ventilator.
As a working rule, the initial setting for the tidal volume is between 10 and 15 mllkg of body weight. 11 Once the
desired tidal volume is established, provisions must be
The Change over from the Inspiratory Phase to the Expiratory Phase
123
Volum e contro l
So leooid
o_____
E-Ie-c-tro n_i_c_
c -ir_c:_it_
F1G 11-6.
Hot-wire anemometer used as volume-cycling mechanism.
made to assure that the volume is delivered within a reasonable time . With most volume-cycled ventilators, a flowrate control is present and provides the necessary means of
regulating the length of the inspiratory phase.
The fact that all preliminary control adjustments can be
made before connecting the patient to the ventilator represents one of the many advantages of the volume-cycled
ventilator. .
Regardless of the cycling mechanism, as soon as the
patient is connected to the ventilator, the volume delivered
\O the patient must still be verified with a volume measuring
device (VMD). Not all of the volume selected at the ventilator will reach the patient. Some of the volume will be
lost in the ventilator circuit as a consequence of the system's
compliance . 12 - 14 To determine the volume that actually
reaches the patient, the VMD must be connected between
the ventilator outlet .and the patient. Lost volume can then
be obtained by subtracting the measured volume from the
control setting. The usual method of correcting for lost volume is by slowly increasing the control setting stepwise
until the desired volume. is displayed on the VMD. The
flowrate control may then require further adjustment to
maintain inspiration time.
With volume-cycled ventilators the only parameter remaining constant is the volume delivered by the ventilator.
The time required to deliver the volume and the pressure
developed during the process are greatly influenced by the
magnitude of the generated pressure and lung characteristics
of the patient.
In Figure 11-7 three levels of generated pressure are
used to demonstrate these effects. For each level, the ventilator was adjusted so that, under standard lung conditions,
a volume of 0 .5 L was delivered in 1.5 seconds . Figures
I i- 7 , A and II- 7, B show that when changes in lung
characteristics occur, the length of the inspiratory phase
..
Determination of Flowrate
Constant Flow Generator (Square Wave, Top Hat)
Microprocessor-controlled, volume-cycled ventilators
are capable of generating any flow waveform. When a constant flow is generated and a specific volume and
inspiration-to-exhalation ratio (I:E ratio) is required, a simple formula is used to determine the exact setting of the
flowrate control:
V = (VT x
Rate) x (I
E)
124
Volume Cycling
DECREASING COMPLIANCE
45
40
30
A
Mouth pressure
(em H20)
20
10
0 PG = 3500 em H20
1.5
2.5
1.5
.2
2.5
6. PG = 100 em H 20
0 PG = 25 em H 20
20
B
Mouth pressure
10
(em H 20)
FIG 11-7.
Volume cycling: Effects of generated pressure and lung characteristics on length of inspiratory
phase. Cycling occurs when volume delivered equals 0.5 L. For standard lung conditions, lung
compliance is 0.05 Lfcm H2 0 and airway resistance is 6 em H2 0/Usec. A. shows effects of lung
compliance while maintaining standard airway resistance. Range depicted for lung compliance is
from 0.0125 to 0.1 Ucm H2 0 . B demonstrates effects of airway resistance while maintaining
standard lung compliance. Range depicted for airway resistance is from 0 to 30 em H20 /Lisee.
where
IC
fr,,
'
V = (VT
=
Rate) X (I + E)
(0.5 X 10) X (1 + 3)
20 L/rnin (= 0.33 Lisee)
X
The Change over from the Inspiratory Phase to the Expiratory Phase
125
STANDARD CONDITIONS
COMPLIANCE HALVED
RESISTANCE DOUBLED
21 .96
Mouth pressure
(em H,O)
Flow rate
(Li see)
Alveolar pressure
(cmH,O)
Volume
(L)
~------~-------- -,
Time (sec)
1.5
1.5
1.5
FIG 11-8.
Theoretical waveforms produced by constant flow, volume-cycled ventilator. Moment of cycling is indicated by black dot.
3500 em H 2 0
10,479 em H2 0 /Usec
Generated pressure
Ventilator resistance
Standard conditions
Lung compliance
Airway resistance
0 .05 Ucm H2 0
6 em H20 /Usec
+ E) X 1.57
(1 + 3) X 1.57
(= 0.52 Lisee)
(VT X Rate) X (I
= (0.5 X 10)
= 31.4 Llmin
The theoretical waveforms produced by a volumecycled ventilator that generates a half sine wave flow pattern
are shown in Figure 11-9.
126
Volume Cycling
STANDARD CONDITIONS
COMPLIANCE HALVED
RESISTANCE DOUBLED
20.43 .
Mouth pressure
(em H,O)
0.52
Flow rate
(Li see)
20
Alveolar pressure
(em H,O)
Volume
(L)
t
0
Time (sec)
1.5
1.5
......"'---
---, -- -- - --
-,
1.b
FIG 11-9.
Theoretical waveforms produced by nonconstant flow; volume-cycled ventilator. Moment of volume cycling is indicated by black dot.
Tidal volume
Inspiration time
Average ftowrate
Peak ftowrate
Standard conditions
Lung compliance
Airway resistance
0.5 L
1.5 sec
0.33 Usee
0.52 Usee
0.05 UcmHp
6 em H 2 0/Usec
Vpk = lYT X
With respect to maintaining tidal volume a.'1d inspiration time, the behavior of the nonconstant flow, decreasing
or increasing flow, and the volume-cycled ventilator is identical to that of the constant flow generator. They differ only
in the shapes of the waveforms generated.
l
The Change over from the Inspiratory Phase to the Expiratory Phase
t
A
STANDARD CONDITIONS
COMPLIANCE HALVED
RESISTANCE DOUBLED
127
25
Mouth pressure
(em H 2 0)
0.40
Rowrate
(Usee)
.- .
~
~.
20
Alveoar pressure
(em H2 0)
0.5
Volume
(l)
1.5
1.5
Theoretical waveforms generated by microprocessor ventilator in the increasing flow (increasing ramp, accelerating
flow) mode.
";:. __ ,
Tidal volume
Inspiration time
Average flowrate
Peak flowrate
Standard conditions
Lung compliance
Airway resistance
~~,
'
1.5
FIG 11-10.
! , ;t,;.
1
,....--....
Time (sec)
1':"
ct..,
0.5 L
1.5 sec
0.33 Usee
0.44 Lisee
'
I~~~'..
.
.
,
--
. ,~
- ~
--
I ilK lung conditions, the length of the inspiratory phase depends t:ntirely on the level of pressure generated by the
ddvt: nlt:chanism. Therefore, when the generated pressure
in low, no simple equation can be used to determine the
llowrak rt:quired to maintain a given inspiration time or an
I I . !il tH> . Manipulation of the ftowrate control provides the
w d y method of regulating the length of the inspiratory
phn~c
ti l~ttor
SUMMARY
A ventilator is said to be volume cycled if the inspiratory phase is ended at the moment a predetermined volume
has been delivered into the patient circuit. At the moment of
cycling, the ftowrate, the time taken to deliver the volume,
and the pressure developed in the patient circuit may all
vary from one respiratory cycle to another. The only parameter remaining constant is the volume preset at the ventilator.
!t
'~
ll
It
128
Volume Cycling
i~
~~
'
STANDARD CONDITIONS
COMPLIANCE HALVED
RESISTANCE DOUBLED
25
Mouth pressure
(em H 2 0)
0.40
R cwrate
(Lisee)
20
~-=-- I
Alveoar pressure
(em H 2 0)
~
- - --- - - - -- .,
0.5
Volume
(L)
Time (sec)
1.5
1.5
1.5
FIG 11 - 11.
Theoretical waveforms generated by microprocessor ventilator in the decreasing flow (decreasing ramp, decelerating
flow) mode.
Tidal volume
Inspiration time
Average flowrate
Peak flowrate
Standard conditions
Lung compliance
Airway resistance
0.5 L
1.5 sec
0.33 Usee
0.44 Lisee
0.05 L!cm H20
6 em H2 0 /Lisee
(/
When the ventilator generates a constant flow waveform and a specific I:E ratio and tidal v~lume are required,
a simple formula is used to determine the exact setting of
the flowrate control:
V = (VT x
Rate) x (I
+ E)
CHAPTER
12
J
Time Cycling
......-....
TIME-CYCLING MECHANISMS
Pneumatic
,.--Type I
:. A typical pneumatic time-cycling mechanism 3 is shown
in Figure 12-1 . The actual design may vary from one ventilator to another, but the principle of operation is the same.
The system consists of the placement of a balloon valve
between the drive mechanism of the ventilator and the patient circuit. During the inspiratory phase (Fig 12-1 ,A), the
filling of the balloon valve is controlled by the variable
resistance. The variable resistance, labeled "inspiration
time" on the control panel of the ventilator, controls the
time required to inflate the balloon . After a period of time,
the balloon becomes fully pressurized and source gas from
the drive mechanism to the patient circuit is cut off (Fig
12-1 ,B). Hence the changeover from the inspiratory phase
to the expiratory phase is time cycled~~\
132
Type II
Figure 12-2 demonstrates a simplified schematic of a
pneumatically operated time-cycled ventilator. The ventilator
represented schematically is the Bird Mark 24 (Fig 12-3).
The ventilator can be divided into two separate compartments: the exhalation timing compartment and the inspiration timing compartment. Each compartment consists
of a needle valve and a spring-loaded plunger and diaphragm assembly.
Figure 12-2,A shows the ventilator in the expiratory
phase. In this phase source gas is prevented from entering
the patient circuit by the position of the exhalation plunger
and diaphragm. Instead, source gas is directed to the right
of the diaphragm in the exhalation compartment by way of
a needle valve (exhalation time control) .
After a period of time, determined by the position of
the exhalation time control, the exhalation compartment
becomes fully pressurized and the diaphragm and plunger
are moved completely to the left (Fig 12-2,B). When this
happens, source gas is directed to three specific areas:
1. To a line that occludes a vent in the inspiratory timing
compartment
2. To the patient circuit by way of the ftowrate control
3. To another line extending from the ftowrate control to
the left side of the diaphragm in the inspiration timing compartment
During the inspiratory phase, pressure increases in the
inspiration timing compartment. This rate of increase is
governed by the position of the inspiration time needle
valve.
----._
133
The Change over from the Inspiratory Phase to the Expiratory Phase
Source gas
'
'
~Fro m low
pressu re so u rce
,,
Bllooo
==:::-:::J
_tW
_j
==
;
fl
--
,.
~I
INSPIRATION PHASE
-'\
Ball oo n
infl ated
./
'
.' ,
'
'
"-
'
,.
FIG 12-1.
Essential feature of simple pneumatic time-cycling mechanism.
1
\
.'
134
Time Cycling
Inspiratory timing
compartment
Vent
I===?
lvent
_ _ _j
L------------------ -,
To patient circuit
FIG 12-2.
\c hc""t c re prese ntation of Bi rd Mark 2 time cycled ven ti lator.
'
/~
FIG 12-3.
Bird Mark 2 time-cycled ventilator.
p~
Ps
Flip-Flop
vI
Gas pathway
when vent occluded
c2
c,
Cam
,--,
'
"
t"'
--Vent
OR/NOR
,_
Plunger
~ Spnng
'--l-1
Sliding
rod
o,.&
To
exhalation
valve
Canister
) ( Resistance
L__j
Relay valve
FIG 12-4.
Schematic diagram of ftuidicall y operated, time-cycling mechanism.
135
136
Time Cycling
Electromechanical
'
, / --.~
'
)' _f'
H
J
~.
Pressure
chamber
Cylinder
Rate control
ln sp1ra:ory phase;
.., 1
I
-r___,-..,
1.....
1/ 3 2/3 3/3
FIG 12-5.
Double-circuit, electromechanical, time-cycled ventilator.
\._,/ '
The Change over from the Inspiratory Phase to the Expiratory Phase
( ar;
60
5 sec
12
1.67 sec
E = 1
Using unity for the inspiration portion of the ratio, the exhalation portion must be:
E
3.33
1.67
=2
and therefore the I:E ratio is 1:2.
137
138
Time Cycling
w
--~,.,,, CI
Check
Inhale time
control
ExhaletimQ
control
Piston
~__Belt
----
Cylinder
Wheel
~~@
___ _.......-...-
--
Electric motor
lever
Slider
FIG 12-6.
Single-circuit, electromechanical, mixed-cycled ventilator. (See text for description.)
Electronic Timers
'-
Inspiration Timer 8
The simplest method in which time cycling can be accomplished electronically is shown schematically in Figure
12-7 . The mechanism consists of an electronic timer, circuit board, and solenoid. During the inspiratory phase the
solenoid is open and gas from the drive mechanism is directed to the patient circuit. After a period of time determined by the setting of the inspiration time control, an
electrical signal is sent to the solenoid, which closes to end
the inspiratory phase. To control the respiratory cycle, rate ,
and I:E ratio, an exhalation timer is also present.
60
= -R
ate
...
nsprrahon time
Exhalation time
= ~~ -
= 4 .5 sec
1.5
The Clumge over from the Inspiratory Phase to the Expiratory Phase
Inspiration time
control
139
Exhalation time
control
I
,.-
- ...,--I
-,
~
,- - <J-1
r<l-,
L-- _
_.
~.&
Electronic
circuit
' ,.
Solenoid
FIG 12-7.
Time-cycling mechanism consisting of simple electronic inspiration timer.
Inspiration time
control
L.------+1:1
____Q
_____.I El~~~.;'
L--
Soleooid
~
Source gas_,..
To patient circuit
FIG 12-8. Time-cycling mechanism consisting of electronic inspiration time and rate controls.
The l :E ratio in this time-cycling mechanism is a function of rate and inspiration time. Once exhalation time is
determined, the I:E ratio can be calculated quite easily with
lhc formula presented earlier:
E
= 1 X 4.5
1.5
3
'
140
Time Cycling
Rate
control
Percent of inspiration
time control
II
LL
+---
.....
-+
Electronic
circuit
1
Solenoid
Source gas . _ . .
FIG 12-9.
Electronic time-cycling mechanism consisting of rate and percent inspiration time controls.
= I+
- 1- E x
100
=-
1
-
x 100
= 0.75
= 25%
For a specific example, say that the rate control is set to
deliver 10 breaths per minute and the percent inspiration
time control is adjusted to 25% . The electronic circuit
"looks" at the control settings and sets up this relationship
to determine inspiration time:
.
. t"
I nsp1ra
1on time
= -60Rate
. . . time
.
Percent msp1rat10n
10
1.5 sec
The I:E ratio in this time-cycling mechanism is a function of the percent inspiration time setting. In the example,
the percent inspiration time setting is 25%. The remaining
~~ -
0. 75
2.25 sec
Substituting:
60
. .
I nsp1ration time = - X 0.25
sec
X 2.25
0.75
= 3
= 1:3
Rate and I:E Ratio Controls 11
In other time-cycled ventilators inspiration time is governed by the setting of the rate and I:E ratio controls (Fit_'
12-10). With this mechanism the length of the inspirato~,
~
The Change over from the Inspiratory Phase to the Expiratory Phase
Maxi mum inspiration
time con tro l
Rate
contro l
I:E ratio
co ntrol
II
f+--
141
I+~
,....
T1
s afety system
'
Electronic
circuit
1~
Source gas _ _ .
_ _ . To patient circuit
FIG 12-10.
Electronic time-cycling mechanism consisting of rate and I: E ratio controls. Presence of maximum inspiration time control prevents
inadvertent selection of prolonged inspiration time.
10
60 X 1
X ( + 3)
1
1.5 sec
60 X 1
X (1 + 3)
3 sec
Keeping the rate constant, increasing the I:E ratio from 1:3
to 1:1 would also double inspiration time.
CLINICAL CONSIDERATIONS
During controlled ventilation the single most important
parameter to consider is the volume exchange in the lungs.
With some time-cycled ventilators the tidal volume is not
controlled directly but determined by the product of flowrate and inspiration time. In others the tidal volume or
minute volume, rate, and I:E ratio are selected and the
ventilator automatically assigns the fiowrate.
However, on most time-cycled ventilators a direct
access to a fiowrate control is provided, and the setting
of this control and the setting of the time-cycling mechanism determine the volume delivered at the moment of
cycling.
With these time-cycled ventilators all parameters except for the fiowrate are selected at the discretion of the
operator. The fiowrate setting is not arbitrary and must be
calculated. The procedure is similar for all time-cycled ventilators.
In the case of the simple inspiration timer, assuming the
ftowrate control is calibrated in liters per minute (L!min),
the setting is determined by the following formula:
x
.
_
Volume (1)
Flowrate - In . .
.
( )
60(sec/mm)
sprratwn tlme sec
V = Volume
Rate
X (I
Flow x Time
Volume
(L)
0.4
0.3
...
0.2 +----.,.c---,r----,-----.40
10
20
30
0 PG = 3500 em H2 0
6 PG = 100 em H 2 0
0 Pr: = 25 em H 20
INCREASING RESISTANCE
Volume
(L)
+ E)
0.4
FIG 12-11.
Time cycling: Effects of generated pressure and lung characteristics on volume delivered at moment
of cycling. Cycling occurs when inspiration time equals 1.5 seconds. For standard lung conditions.
lung compliance is 0.05 Llcm H 20 and airway resistance is 6 em H 2 0/Lisee. A shows effects of lung
compliance while maintaining standard airway resistance. Range depicted for lung compliance is
from 0.0125 to 0.1 Llcm H20. B shows effects of airway resistance while maintaining standard lung
compliance. Range depicted for airway resistance is from 0 to 30 em H2 0 /Lisee.
CHAPTER14
Once the changeover from the inspiratory phase to the expiratory phase is complete, the ventilator must allow the
lungs to empty. This represents the only function of the
ventilator in the expiratory phase. 1
There are a few ways in which the ventilator can allow .
the lungs to empty, and therefore it is also possible to classify the ventilator during this phase. The three most common classifications are listed below:
1.
2.
3.
/
( To mimic "pursed-llp" breathing as seen with patients
suffering from chronic obstructive pulmonary disease
(COPD), some ventilators have the option of adding resistance to retard expiratory flow from the lungs. The procedure
is thought to facilitate emptying of the lungS-b-r ''splinting''
the airways and reducing airway collapse. 2___..,)
Methods
The method of providing expiratory resistance is simple and involves the application of mechanical resistance to
the exhalation port of the breathing circuit. In Figure 14-2
the variable resistance consists of a cap (also called "retard
cap" on Bird ventilators, described in Chapter 28), which
fits tightly on the exhalation port of the breathing manifold.
The cap has apertures of varying diameters , which are
aligned on the exhalation port so that the expired gas must
pass through one of the openings. Rotating the cap to the
smallest opening provides maximum expiratory resistance.
Figure 14-3 demonstrates another method of providing
expiratory resistance. Such a system is found in the Bennett
MA-J-,ventilator, presented in Chapter 24.
, During the exhalation phase the position of the pilot
val've-allews the exhalation balloon valve to empty to atmospheric pressure via the fixed and variable resistance
(labeled as expiratory resistance control). Normally, the expiratory resistance control is wide open and the rate at
which the exhalation valve empties is not impeded. In this
state no resistance is imposed on the exhaled flow and the
lungs empty passively to atmospheric pressure. Expiratory
resistance results when the control knob is adjusted to nar163
164
STANDARD CONDITIONS
COMPLIANCE HALVED
RESISTANCE DOUBLED
(em H 20 )
Flow rate
(Lisee)
Alveolar pressure
(em H 2 0)
0.5
I
Volume
(L)
I
I
I
I
2
Time (sec)
FIG 14- 1.
Theoretical wavef?rrns of expiratory phase produced by constant atmospheric pressure generator. Lungs were previously inflated with
volume of 0.5 L. Mouth pressure and alveolar pressure were at equilibrium at moment of cycling from inspiratory phase. Broken
diagonal line represents time constant of system:
Generated pressure
Ventilator resistance
Standard conditions
Lung compliance
Airway resistance
0 em H 2 0
2 em H 2 0 /Lisee
0.05 L!cm H2 0
6 em H2 0/Lisee
row the channel opening for the passage of gas from the
exhalation balloon valve. In such instances the exhalation
valve deflates slowly and delays the emptying of the lungs.
Maximum resistance is accomplished by rotating the control
knob and fully occluding one opening. When this is done,
all of the gas from the exhalation valve must pass through
the fixed resistance.
The capacitance (also called a booster) acts as a gas
reservoir and provides a means of regulating a smooth and
orderly . flow of gas from the exhalation balloon valve.
When fully charged, the exhalation balloon valve holds less
than a few cubic centimeters of gas : We can appreciate that
without the booster it would be impossible to control the
. Clinical Considerations
The use.of expiratory resistanc~ has met with varying
degrees of success. Expiratory resistance increases the mean
airway pressure, impedes cardiovascular function, and also'
promotes the dangers of air trapping. Air trapping occurs
when the lungs are reinflated before they have a chance to
completely empty (Fig 14-4). The likelihood of air trapping
increases with the amount of mechanical expiratory resistance applied . and on the lung characteristics of the
patient. 3 - 7
STANDARD CONDITIONS
COMPLIANCE HALVED
RESISTANCE DOUBLED
2G.,
Mouth pressure
(em H 2 0)
Fl owrate
(Lisee)
20
_ A lveo lar pressure
;
(em H 2 0)
0.5
Vo lume
\
(L)
\
\
I
\
2
Time (sec)
FIG 14-1.
Theoretical waveforms of expiratory phase produced by constant atmospheric pressure generator. Lungs were previously inflated with
volume of 0.5 L. Mouth pressure and alveolar pressure were at equilibrium at moment of cycling from inspiratory phase. Broken
diagonal line represents time constant of system:
Generated pressure
Ventilator resistance
Standard conditions
Lung compliance
Airway resistance
0 em H 20
2 em H20/Lisee
0.05 L!cm H2 0
6 em H2 0/Lisee
')
Clinical Considerations
The use of expiratory resistance has met with varying
degrees of success. Expiratory resistance increases the mean
airway pressure, impedes cardiovascular function , and also
promotes the dangers of air trapping. Air trapping occurs
when the lungs are reinflated before they have a chance to
completely empty (Fig 14-4). The likelihood of air trapping
increases with the amount of mechanical expiratory resistance applied and on the lung characteristics of the
patient. 3 - 7
Airway
pressure
Time--+-
Air trapping
FIG 14-4.
Air trapping occurs when inspiratory phase begins before lungs have completely emptied.
STANDARD CONDITIONS
COMPLIANCE HALVED
RESISTANCE DOUBLED
20
Mouth pressure
(em H 20)
\
\
Flowrate
(Lisee)
0.77
20
Alveolar pressure
(em H 20)
0.5
'
Volum e
(L)
' '\\
-...
Time (sec)
FIG 14-5.
Theoretical waveforms of expiratory phase produced by constant atmospheric pressure generator having a high resistance. Lungs were
previously inflated with volume of 0.5 L. Mouth pressure and alveolar pressure were at equilibrium at moment of cycling from inspiratory
phase. Broken diagonal line represents time constant of system:
Generated pressure
Ventilator resistance
Standard conditions
Lung compliance
Airway resistance
~.
0 em H20
20 em H 20/Usec
0.05 Ucm H2 0
6 em H20/Usec
161
lung pressure is held abo"~<,e atmospheric pressure throughout the respiratory eye~ The use of positive expiratory
pressure was reported as early as 1959 1 and continues to
play a major role in modern ventilatory techniques . 11 - 16
Positive expiratory pressure increases the mean airway
pressure and to some degree alters cardiovascular function.
The effects on the cardiovascular system are not uniform
and result when the intrapulmonary pressure is reflected
intrathoracically. Normally, increases in the mean airway
pressure cause a decrease in cardiovascular function, but
this is not always the case. Therefore the mean airway pressure should not be taken as an absolute indicator of intrathoClinical Considerations
racic pressure. In disease states associated with severe reThe use of NEEP is aimed at reducing the mean airway
ductions in lung compliance the transmission of
pressure and minimizing the cardiovascular effects of interintrapulmonary pressure to the intrathoracic space is not the
mittent positive-pressure ventilation. The mean airway pressame quantitatively.
Consider the simple lung thorax model in Figure 14sure decreases when negative pressure is applied to the airway
8,A, where a respiratory bag (lung) is suspended within a
during the exhalation phase. As demonstrated by the waverigid chamber (thoracic space). The bottom of the chamber
forms of Figure 14-7, the maneuver accelerates the emptying
of the lungs, regardless of the existing lung conditions.
is sealed with a flexible diaphragm (chest wall). The downAlthough NEEP has been shown to reduce cardiovascular . . ward force of a small weight connected to the center of the
effects of controlled ventilation, its use has also been associdiaphragm generates a subatmospheric pressure of - 5 em
ated with air trapping, 9 10 especially in patients with COPD.
H2 0 around the bag and represents intrathoracic pressure.
Although the content of the bag is at atmospheric pressure,
Consequently, the use ofNEEP has been abandoned.
it is partially expanded by the negative pressure surrounding
it. The volume present in the bag is considered to be the
functional residual capacity (FRC).
THE CONSTANT POSITIVE-PRESSURE
The compliance of the lung and the compliance of the
GENERATOR
chest wall are each given an arbitrary value of 0.2 Ucm
( _Jn contrast to the constant negative-pressure generator,
H 2 0. Lung compliance (CL) and chest wall compliance
a constant positive-pressure generator is one in which the
(Ccw) make up total compliance (Cr) and are related in
Method
One common method of generating NEEP is shown in
Figure 14-6. In" this schematic a Venturi is connected to the
exhalation port in such a way that negative pressure is created in the circuit and thus reflected in the lungs. The
amount of negative pressure generated is governed by the
position of the needle valve, which controls the flow of gas
to the jet of the Venturi. The greater the gas flow, the greater
the negative pressure. The mechanism operates during the
exhalation phase only. During the inspiratory phase gas to
the Venturi is interrup~
Pressure manometer
Negative pressure
control
Source gas
'
--~- ---
-...;;.;;......;;;'--,::--.....-~
U ---=-=-
connection
''
Exhaled gas
FIG 14-6.
Use of a Venturi to generate negative pressure during expiratory phase.
168
18
r-
Mouth pressure
(em H 20)
STANDARD CONDITIONS
COMPLIANCE HALVED
RESISTANCE DOUBLED
-2 h~=
/.
Flowrate
(Lisee)
1.25
18
~=
Alveolar pressure
(em H 2 0)
-2
0.5
Volume
(L)
\
\
\
\
\
Time (sec)
FIG 14-7.
Theoretical waveforms of expiratory phase produced by constant negative-pressure generator. Lungs were previously inflated with
volume of 0.5 L. Mouth pressure and alveolar pressure were at equilibrium at moment of cycling from inspiratory phase. Broken
diagonal line represents time constant of system:
Generated pressure
Ventilator resistance
Standard conditions
Lung compliance
Airway resistance
- 2 em H2 0
10 em H2 0 /Usec
such a way that they are added reciprocally. Total compliance of this system is calculated as follows:
0.2
0.2
0.2
+ 0.2
PTx=c-
V =
PEEP
= 0.1 L!cm H2 0 x 5 em H2 0
= 0.5 L
CT X
cw
Substituting:
0.5 L
0.2 L!cm H20
= 2.5
em H2 0
169
Rigid chamber-
-5 em H2 0
Intrathoracic
pressure
FIG 14-8.
Lung/thc>rax model showing relationship between airway pressure and intrathoracic pressure. (See text for description.)
In Figure 14-S,C lung compliance (not total compliance) is reduced by one half, and the numerical values for
the FRC and pressure were obtained using the procedure
just described.
With this simple experiment we can see that although
the same amount of PEEP was applied to the airway, less
pressure was transmitted intrathoracically because the FRC
could not be maintained at its original level. In Figure 14S,D lung compliance is still halved, but the FRC was restored to 0.5 L by increasing PEEP to 7.5 em H2 0 . If we
compare Figure 14-S,C with Figure 14-S,D, we can fur-
ther appreciate that the level of pressure applied to the airway is not necessarily the pressure reflected intrathoracically. In fact, if the respiratory bag in Figure 14-8 were
substituted for a bag that could not expand, the pressure in
the rigid container would not change regardless of the
amount of pressure applied within the bag.
Clinically speaking, it is with patients suffering from reductions in lung compliance that positive expiratory pressure
has been shown to increase lung compliance, reduce intrapulmonary shunting, and improve arterial oxygenation without
significant compromises to the cardiovascular system.
The traditional range of positive expiratory pressure is
between 5 and 15 em H20. However, there are no hardand-fast rules for determining the exact level. The level
must be tailored to meet the needs of the individual
patient. 17 - 26 In this respect, positive expiratory pressures
_,
170
(~orms
7-
-.~
PEEP
CPAP
Airway
pressure
EPAP
c
"""'
IMVand PEEP
Airway
pressure
Time---
1- I -1-
-1- I -1-
FIG 14-9.
Four forms of positive expiratory pressure.
-1
1- I -1-
-1- I -1-
-1
n
n
d
g
e
171
i(
f
CsT
= P pit
VT
-PEEP
1wlwrc
~
Pulmonary Effects
Barotrauma. 44 - 46- The term suggests that physical
damage to the lungs occurs as a result of pressure. Despite this
definition, damage to the lungs is more likely associated with
alveolar overdistension rather than with pressure. 47 If the alveoli cannot be hyperinflated, lung rupture is unlikely to occur.
The consequences of barotrauma include the following:
1. Subcutaneous emphysema is a condition where leakage of air from the alveoli finds its way into the mediastinum and expands out beneath the tissues of the neck. Air in
the mediastinum (mediastinal emphysema) can lead to cardiac tamponade and decrease cardiac output. Apart from
alveolar rupture, subcutaneous emphysema can also result
from high levels of positive pressure delivered via a tracheostomy tube. Quite often subcutaneous emphysema is a sign
of a pneumothorax .
2. Pneumothorax is defined as the presence of air in the
pleural space. Alveolar rupture leading to air in the pleural
172
STANDARD CONDITIONS
COMPLIANCE HALVED
RESISTANCE DOUBLED
25
Mouth pressure
(em H20)
+5
''
'
Flowrate
(Lisee)
~
2.22
------
25
,~
,..----.,,
Alveolar pressure
(em H 2 0)
+5
0.5
,.--.
Volume
(L)
I
I
2
Time (sec}
FIG 14-10 . .
Theoretical waveforms of expiratory phase produced by constant positive-pressure generator. Lungs were previously inflated with volume
of 0.5 L. Mouth pressure and alveolar pressure were at equilibrium at moment of cycling from inspiratory phase. Broken diagonal line
represents time constant of system:
Generated pressure
Ventilator resistance
Standard conditions
+5 em HzO
3 em H2 0/Usec
0.05 UcmHzO
Lung compliance
Airway resistance
6 em H 20/Usec
GOOD PEEP VALVE
''
\,,E'P7 '''''""''
Airway pressure
''
PEEP-
Time-
-+I+
FIG 14-11.
Difference between good and poor
P~EP
-+I
crrcssure in the thoracic cavity beyond at' and causes varying degrees of lung colRESISTANCE
.ty of lung collapse depends on the amount
in the thoracic space. Pneumothorax is a
us complication than subcutaneous emphy:s immediate medical attention.
,.-,-,--~oJar Effects48-6I
+--"'------->ive levels of positive expiratory pressure
+--~-=o the intrathoracic space, cardiac output
)f a reduction in the venous return to the
ion63,64
\
I
ratory pr~ssure increases the level of ane (ADH) and changes renal -perfusion.
+-----.~..-~eduction in urine output is commonly as3
:itive expiratory pressure. A fall in urine
1 as 40% has been reported with a positive
37
.
. fl t re of 10 em H 2 0 .
Diuretics are often
1gs were prevwus1y m a < .
.
.
inspiratory phase. Brokeltam satisfactory unnary flow.
I
Exhaled gas
,,,
III
- R igid tube
173
Vent
; \--
~
______ t _
- c -:: --
FIG 14-12.
Water line
,-
,--.
174
the tube, bubbles through the water, and is vented to atmosphere. The exhalation process continues until lung pressure
empties to a level equal to the pressure exerted by the
weight of the water in the tube. At that moment bubbling is
no longer observed, and positive pressure (PEEP) remains
in the lungs .
The level of PEEP is controlled by the position of the
tube in the water. The deeper the tube is in the water, the
greater the level of PEEP. The PEEP level is displayed on
the ventilator manometer (not shown) or by connecting an
external manometer to sample the pressure in the flexible
hose.
The Emerson PEEP Valve. 66-Another mechanism
that uses water to generate positive expiratory pressure is
shown in Figure 14-13. In this system the PEEP mechanism is an integral part of the exhalation valve. Normally,
when no water is present in the column, the patient's exhaled gas is directed to the underside of the diaphragm,
which lifts and allows the lungs to empty passively to atmospheric pressure. When water is placed in the column,
the lungs do not completely empty and are held inflated to
a level that corresponds with the weight of the' water on the
exhalation diaphragm. The amount of PEEP is displayed on
the ventilator manometer and is controlled by varying the
amount of water in the column.
~agnetic
Pressurized
Water column
Diaphragm
Exhaled gas
Exhaled gas
FIG 14-13.
FIG 14-14.
175
t
Weighted ball
PEEP control
FIG 14-16.
Exhaled gas
FIG 14-15.
Dead-weight PEEP valve.
Filling port
Balloon valve
Disk
~------------~~~~-~---ln_le_t
Spring- loaded
ch eck valve
~,.---____.
! . .
Pilot valve
~. L,.
FIG 14-17.
Balloon valve PEEP device.
expiratory phase, lung pressure opposes the magnetic attraction and lifts the disk. The pathway for exhaled gas
remains open until lung pressure drops to a level where it
can no longer overcome the attractive forces between the
magnet and the rod. At that moment, the disk returns to its
normal resting position and PEEP results. The level of
PEEP is displayed on the ventilator manometer and is controlled by adjusting the position of the magnet relative to the
metal rod. The closer the magnet.is moved toward the rod
.
'
the greater the level of PEEP, and vice v:_9
176
PEEP control
Inlet
. . . - - - - Exhaled gas
Source gas
Exhalation port
FIG 14-18.
Venturi PEEP device.
Source gas
177
== =======:;-~
I
Venturi "T"
Exhalation
valve
Pressure manometer
Inlet
,'
....
~
Pressure manometer
Bellow
--
_l
Check valve
Canister
-~~
----1
PEEP control
Source gas.
~0,~
Exhalation valve
Variable
resistance
FIG 14-20.
Method of providing PEEP with continuous flow applied to exhalat!on balloon
v~ve
,,
~L
To"""'"''"
Fixed resistance
::
Exhaled
gas
Pressure manometer
Check valve
--~
PEEP control
To ventilator
0
:'
L
Exhaled
gas
-~
- .....-
Pressure
regulator
Source gas
Vent
Relay valve
'1'
I I
Vent
FIG 14-21.
Method of providing PEEP using pressure regulator. (See text for description.)
....
-....!
\C)
')
)l
....
00
Q
Pressure manometer
Check valve
To ventilator
PEEP control
j1
Venturi_
:==~
----:---~
Source gas
Vent
I
3-way solenoid
Qi
FIG 14-22.
Method of providing PEEP using a Venturi to generate pressure in exhalation valve.
Exhalation va lve
.-
Exhaled
gas
Fully opened
181
Fully closed
Actuating shaft
(plunger)
To
atmosphere
From
patient
FIG 14-23.
Schematic representation of electrodynamic motor used to generate positive end-respiratory pressure.
B.
A.
T
Airway pressure
Resistance
1
0
Time-
1..-
-+j.-
FIG 14-24.
Airway pressure waveforms showing resistance to spontaneous ventilation
during CPAP therapy.
2.
3.
4.
24 ,B, where the resistance of the CPAP system causes significant fluctuations in the airway pressure.
With both CPAP and EPAP resistance to spontaneous
ventilation increases the work of breathing, and resistance
to exhalation increases the mean airway pressure.
_)
_)
_)
.)
_)
...
~)
_)
_)
_)
_)
QC
Pressure manometer
~I
So"'"'" -
A
\J!
r-===:::::::::...=L......-,::::JII
T
I
bll "'"
Diaphragm
.
__.D .....
CPAP control
---...--- - ~
Source
gas
;:::::::J
Sensing \me
Demand valve
II
I I
L--~~J
Vent
l
~--
Humidifier
Spring-loaded
Venturi
FIG 14-25.
Method of providing CPAP with use of demand valve.
Patient
conn ection
_)
_)
_)
exhalation the check valve closes and the patient must exhale through the PEEP devices .
With a simple modification made to the system in Figure 14-26, CPAP can also be provided. In Figure 14-27
the aerosol T is replaced with a Bird Venturi T. 67 The
Venturi is supplied with gas from an oxygen flowmeter connected to a blender. The blender is adjusted to the corresponding 0 2 setting of the nebulizer to prevent fluctuations
in the inspired oxygen concentration. The Venturi is positioned to generate an equal amount of positive pressure
above the check valve to nullify the net pressure difference
across the valve and thus minimize fluctuations in airway
pressure. This is accomplished by observing the pressure
manometer during spontaneous ventilation and adjusting the
flowrate to the Venturi until the least amount of fluctuations
are seen. This method provides CPAP because the patient
no longer has to inspire below the CPAP pressure to open
the check valve and receive fresh gas. During exhalation the
check valve closes and the lungs empty through the PEEP
device to the prescribed level of CPAP.
With a further modification made to Figure 14-27,
CPAP can be provided without the need of check valves or
additional PEEP devices (Figure 14-28). The system consists of the same Venturi T piece described for Figure 1427; because of its position upstream from the exhalation
port, the procedure has been termed proximal CPAP. 67
During inspiration the patient draws in humidified gas
from the mainstream and some from the Venturi. During
exhalation the Venturi flow opposes the exhaled flow and
prevents lung pressure from emptying to atmospheric levels. At first, most of the exhaled gas is carried away by the
m~stream and vented to atmosphere. At equilibrium, only
the excess flow from the Venturi spills over into the mainstream. The level of CPAP is controlled by adjusting the
Flowmeter
__
0/
Aerosol tubing .
Ambient air
l I
Nebulizer
FIG 14-26.
Typical setup to deliver EPAP therapy.
183
::~~~:~r
heck
~~
L........;-
Mainstream
1
valve
I
Patient
connection
PEEP valve
gas
lUncts
Blender
02
flowmeter
11eter
Mainstream
Pressure
manometer
jebulizer
Patient
connection
Blender
02
flowmeter
CPAP control
Flowmeter
I I
I I
Vootoci T
Aom,~t~i ~
Nebulizer
r= =0
Mainstream
and
l j . l - - - - - 1_
1r
Patient
connection
. With this system no check valves or additio!la! PEEP devices are required.
==
.:::...:;:,.
flowrate to the Venturi while observing the pressure manometer. In this system the manometer must be connected
between the Venturi T and the patient connection. Otherwise , CPAP levels will not register on the manometer.
A popular method of providing CPAP is shown in Figure 14-29. Here a 3-L respiratory bag is filled continuously
by an oxygen flowmeter connected to a blender. Gas from
the bag opens the check valve, passes through the heated
humidifier, and is vented to atmosphere through any external PEEP valve (position indicated by the broken rectangle) . During inspiration the patient draws in gas from the
mainstream. The respiratory bag assures that inspiratory
flowrate demands of the patient are always met. During
exhalation the check valve at the humidifier closes and the
patient exhales through the PEEP valve. The spring-loaded
pressure relief valve is set slightly higher than the CPAP
level and is designed to prevent excessive pressure from
185
Blender
02
flowmete r
PEEP valve
,---,
I
---~...._ _ _j
Spring
loaded
valve
Failsafe
valve
Heated humidifer
Respiratory bag
. FIG 14-29.
Method of providing CPAP in continuous flow system.
Mainstream and
exhaled gas
186
Continuous
humidified air/ 0 2 mixture
CPAP control
Mainstream and
exhaled gas
Respiratory bag
Patient
connection
Pressure
manometer
II
Pressure
relief system
Water column
FIG 14-30.
Method of providing CPAP for neonates in continuous flow system.
.\
~\
fresh humidified gas from the inlet and from the larger tubing.
During exhalation the expired gas is washed out of the system
by the continuous flow of gas. The water column is a safety
relief system that prevents excessive pressure from building up
in the system.
The CPAP system in Figure 14-31 operates identically
to the one just described. The only difference is that a
coaxial tube70 is used instead of the modified elbow. A
water column can also be used instead of the spring-loaded
pressure relief valve.
Figure 14-32 is a simplified schematic of another
CPAP system proposed by Gregory and others. 69 In this
system the infant's head is positioned in a rigid transparent
chamber through an iris sleeve, which forms a partial seal
around the baby's neck. Warm humidified gas flows into the
chamber and is vented to atmosphere through a variable
resistance. The variable resistance controls the CPAP level
in the chamber and is displayed on the pressure manometer.
A spring-loaded valve acts as a safety system to prevent
excessive pressures from developing in the system. A water
column can be used instead of the spring-loaded valve.
Another port is used for emergency manual ventilation.
An oxygen sampling port and a trapdoor to provide
quick access to the infant's head are not shown in this
schematic.
SUMMARY
The only function of the ventilator in the expiratory
phase is to allow the lungs to empty. There are three com1
mon methods in which the ventilator can control the emptying of the lungs, and depending on the method selected, a
v~ntilator can be classified as a
Constant atmospheric-pressure generator
Constant negative-pressure generator
Constant positive-pressure generator
The Constant Atmospheric-Pressure Generator
187
Pressure
manometer
Coaxial tube
'11
Continuous humidified
air/ 0 2 mixture
Respiratory bag
Patient connection
. CPAP control
FIG 14-31.
Method of providing CPAP for neonates in c,ontinuous flow ~ystem using coaxial tube.
Blender
02
Pressure manometer
Mainstream and
exhaled gas
CPAP control
~~
\
''"
Iris sleeve
'\\ /
'\ f
flowmeter
Spring-loaded
.__Ve_n_
t - - - - ' .__ ____.:
LlL.....-____
.,_
.
.......-
cI
I I
I I
Ji
I
\
'
Jl
~:n~~=~ion
~/
-...:i
port
Heated humidifier
Transparent chamber
FIG 14-32.
Method of providing CPAP for neonates by placing infant's head in pressurized chamber. (See text for description.)
_ ______ , ,.
188
Positive expiratory pressure increases the FRC. In disease states associated with reductions in the FRC, the application of positive expiratory pressure aimed at returning
the FRC toward normal improves lung compliance, decreases intrapulmonary shunting, and increases arterial oxygenation without the need of high levels of inspired oxygen concentrations.
Positive expiratory pressure increases the mean airway
pressure, but the effects on the cardiovascular system are
not uniform. A reduction in cardiovascular function is observed when positive intrapulmonary pressure is transmitted
intrathoracically. The amount of pressure transmitted in-
3.
4.
CHAPTER15
.---.._;
The fourth and final function of the ventilator is to end
the expiratory phase and begin the process of lung inflation.
The final phase is called .
4 . The changeover from the expiratory phase to the
inspiratory phase
'\
The ventilator can be cycled into the inspiratory phase
m three ways:
PATIENT CYCLING
Patient cycling implies that the ventilator resp<'lnds to
an attempt at inspiration by the patient and delivers a controlled breath. This mode also has been called assisted ventilation. To accomplish patient cycling, the ventilator must
have an assist mechanism (discussed later). If such a mechanism is not present, the patient cannot initiate the inspiratory phase and the ventilator is thus classified as a strict
controller.
Two parameters are evaluated to establish the efficacy
of the assist mechanism: sensitivity and response time.
-----.,
~
1.
2.
3.
""
1. Patient cycling
2. Time cycling
3. . Manual cycling
Sensitiv.ity
This parameter determines the inspiratory effort required to trigger the ventilator. Normally, the patient inhales
a small volume of gas from the ventilator circuit, and the
corresponding drop in pressure activates the assist 'mechanism and a controlled breath is delivered. Some mechanisms may respond to volume or flow instead of pressure.
The sensitivity is usually adjustable and governed by the
setting of a knob at the ventilator. In its most sensitive
position only a small inspiratory effort is required to trigger
195
196
the ventilator. At the other extreme, the mechanism is usually disabled-and will no longer respond to any effort.
A ratio is used to determine the efficiency of the sensitivity mechanism, 1 and although the ratio was primarily
designed for neonatal application, it is useful in evaluating
the performance of ventilators in general:
Percent sensitivity ratio
----------'-'-- X
100
1.5 ml
500 ml
x 100
= 0.3%
Since this is well below 1% of the tidal volume , it
would be considered adequate for this patient. However, if .
the same mechanism were used with a pediatric patient
having a tidal volume of 50 rnl, the percent sensitivity ratio
would be
the assist mechanism. The ventilator then cycles to the inspiratory phase, and the pressure wave travels back to the
patient in the same way.
Pressure waves travel at the speed of sound, which
under normal conditions is approximately 1083 ft/sec (330
m/sec) or 1 msec (millisecond) per foot of ventilator tubing . While this may appear to be a very high velocity, it is not
necessarily so when we compare it with the patient' s inspiration time.
The comparison of the ventilator's response time with
the patient's inspiration time is called the percent response
time ratio:
Percent response time ratio
Ventilator
response
time
_
___
_:___ _
_ X
100
Inspiration time
Ideally, the ventilator's response time should not occupy more than 10% of the patient's inspiration time. 1 For
example, given a ventilator response time of 0.08 second
(80 msec) and an inspiration time of 1.5 seconds, the percent response time ratio would be
=
1.5 sec
= 5%
= !.:2 !IlJ
100
50 ml
3%
Response Time
Once tlie inspiratory effort is made, the ventilator
should respond and deliver a breath within a reasonable
time.
Response time is defined as the time lag between the
initial inspiratory effort and the moment the controlled
breath reaches the patient's airway. The factors responsible
for this time lag include the length of the ventilator circuit
and the characteristics of the assist mechanism.
An inspiratory effort generates a pressure wave that
travels along the circuit, enters the ventilator, and activates
0.08 sec
0.5 sec
100
= 16%
-~--------
- --
---
Assist Mechanisms
The assist mechanism is the device built within the
ventilator that is responsible for sensing an inspiratory effort and cycling the ventilator into the inspiratory phase .
The assist mechanism may operate pneumatically, electronically, or as a combination of both.
Pneumatic
Type I.-Figure 15-1 is a schematic representation of
the Bird pressure-cycled mechanism detailed in Chapter 9
and shown in Figure 9-1. For clarity, only the assist mechanism will be considered in this discussion. The reader
should consult Section Seven for the complete analysis of
the ventilators .
In Figure 15-1 a flexible diaphragm separates the pressure compartment from the ambient compartment, and a
sliding valve is connected to the center of the diaphragm.
197
INSPIRATORY EFFORT
Source gas
!
i. .
l
INSPIRATORY PHASE
FIG 15-1.
Pneumatic assist mechanism such as found in first generation Bird ventilators .
198
INSPIRATORY PHASE
Source gas
Patient connection
FIG 15-2.
Pneumatic assist mechanism such as found in Bennett PR series ventilators.
Pneumatic-Fluidic
In Figure 15-4 the assist mechanism consists of a
Schmitt trigger and a flip-f!op valve (see Chapter 9 for a
review of fluidic devices) . During the expiratory phase the
Schmitt trigger and the flip-flop valve are both at output 0 1 ,
An attempt at inspiration by the patient generates negative
Sensitivity control
Source gas
Ill
II
Patient connection
FIG 15-3.
Pneumatic sensitivity mechanism.
199
Flip-flop
valve
'.
~
Sensitivity
control
'I
)
To patient
circuit
Relay valve
FIG 15-4.
'I
~:
)
)
Electro-Pneumatic
In figure 15-5,A a spring-loaded diaphragm separates
}..,o electrical contacts. The superior surface of the dia--;rragm samples the pressure in the patient circuit by way of
a sensing line. During an inspiratory effort the diaphragm
~ws upward, moving the electrical contacts closer together.
When the inspiratory effort is sufficient to bring the contacts
\ gether, an electrical signal is transmitted to a valve that
opens to begin the inspiratory phase (Fig 15-5,B).
-----. Sensitivity is controlled by rotating a knob that adjusts
~e tension of the spring on the diaphragm. When the ten~on of the spring is increased, the contacts move further
Pressure Transducer
The second type of electro-pneumatic assist mechanism
is a pressure transducer (Fig 15-6). This pressure transducer is identical to the one introduced in Chapter 9, which
was then used as a pressure-cycling mechanism.
As stated in Chapter 9, the function of a pressure transducer is to convert mechanical energy into electrical energy.
In Figure 15-6 the transducer is positioned in the inspiratory circuit of the ventilator and electrical means are used to
measure the motion of the diaphragm. The mechanism consists of an iron core suspended on a diaphragm and 'placed
between two electrical windings. Moving the core within
the windings causes a net difference in output voltage. During an inspiratory effort the diaphragm moves the core
downward, and the net output voltage from the transducer
changes. This is recognized by the ventilator as an attempt
at inspiration by the patient. When the core is moved to a
position that generates a signal equal to an electronic reference potential set by sensitivity control, a controlled
breath is delivered.
Sensitivity is controlled by adjusting the reference control knob, which determines the distance the core must
travel downward to trigger the ventilator.
200
Sensitivity
control
. l_~,;ogUoe
S p n ng
:::-----=:
--
Diaphragm
'------1
Electrical
I contacts
__
:,____ _ _ _ _ _ _- - J .
FIG 15-5.
Electro-pneumatic assist mechanism.
Winding
lop~
Output
Iron core
.lnspirato.ry effort
Sen sitivity
control
Input
To patient circuit
FIG 15-6.
Use of pressure transducer as assist mechanism.
Photoelectric
Photoelectric cells are used to convert light energy into
electrical energy. Figure 15-7 demonstrates one practical
application of this principle in an assist mechanism. In this
system the photoelectric cell is separated from ~he light
source by a shutter connected to the bottom of a diaphragm.
The opposite side of the diaphragm monitors the pressure in
the patient circuit.
A sensitivity mechanism on the control piJ.nel of the
ventilator (not shown) supplies all but the necessary elec-
J
J
1
c:
if~ng line
t _
Diaphragm
Photoeleotcic cell
k T~oto~
201
eHoct
!=[] ~cx=J=!
Shutter
FIG 15-7.
Use of photoelectric cell as assist mechanism.
Sensitivity control
Electronic circuit
To patient circuit
Thermistor
FIG 15-8.
Use of heated thermistor bead as assist mechanism.
.~()ow
By
---._ The past few years have seen the development of a new
Jncept in patient cycling. The cycling mechanism, which
;Qe manufacturer calls flow by, 4 responds to the patient's
. _,spiratory flowrate rather than inspiratory pressure. 5 Al~pugh the flow triggering mechanism is straightforward,
me approach merits special consideration and is therefore
tailed in Chapter 19.
TIME
CYCLING
.......
~
A ventilator is said to be time cycled when the change~ . er from the expiratory phase -to the inspiratory phase is
"--"complished through a timing mechanism that is totally
lHdependent of the patient. If such a ventilator has no assist
.chanism, it is classified as a strict controller.
The time-cycling option is often used as a backup
ety system during assisted ventilation. During this mode
~e exhalation timer is usually set to cycle the ventilator
Time-cycling Mechanisms
The construction of the exhalation timer is essentially
the same as the inspiration timer discussed in Chapter 12.
Therefore they may be pneumatic, fluidic, electromechanical, or electronic.
Pneumatic
There are many types of pneumatically operated exhalation timers and far too many to consider in full detail.
However, all pneumatic timers behave the same way, and
for completeness a brief overview of the essential f~atures
of the most common types will be presented.
One of the simplest methods of timing the exhalation
phase is shown schematically in Figure 15-9. This mechanism is used in the first generation Bird ventilators detailed
202
Disk
Arm
Timer cartridge
Expiratory time
control
FIG 15-9.
Pnewnatic exhalation timer such as found in first generation Bird ventilators.
Electromechanical
Electromechanical mechanisms were described in
Chapter 12 and comprise a piston connected off-center to a
slower revolving wheel. Exhalation time in one such system
is determined by the speed of the piston's return stroke. In
the electromechanical system described for Figure 12-5 the
I:E ratio is fixed, and the rate control determines the length
Qf the expiratory phase.
In another electromechanical system described for Figure 12-6 the speed of the piston is controllable in both
directions, and therefore exhalation time is determined by
adjusting the speed of the motor during the piston's return
stroke.
Electronic
In these mechanisms the exhalation phase is timed by
an electronic circuit, just as in the inspiratory phase. Once
the expiration time has elapsed, an electrical signal is transmitted to a solenoid, which now opens to initiate the inspiratory phase. The reader should consult Chapter 12 for
complete details.
MANUAL CYCLING
The changeover from the inspiratory phase is said to be
manually cycled when the expiratory phase is ended by the
action of the operator.
203
Balloon valve
Rigid chamber
Spring
"'==:.=;'I r=--___J
~========~==-
-shaft
Expiratory
time control
Source gas
I
I
II
II
I
To patient circuit
FIG 15-10.
Pneumatic exhalation timer such as found in Bennett PR series ventilators.
,......,
~ The last function of the ventilator is to end the expiraLv'ry phase and begin the process of lung inflation. There are
t~ee ways in which the ventilator can be triggered into the
mspiratory phase:
1.
2.
3.
By the patient
After a predetermined time has elapsed
Manually
Patient cycled
Time cycled
Manually cycled
204
REFERENCES
1. Rodgers EJ: Physics vs physiology in infant ventilation .
Respir Ther 1972; 2(5):45 .
2. Epstein RA: The sensitivities and response times of ventilatory assistors . Anesthesiology 1971; 34(4):321.
3. Kirby RR: RDS and Infant Ventilation . Consumer Information Data. Palm Springs , Calif, Bird Corp, 1975 .
4. Puritan-Bennett Corp, Santa Monica, Calif.
5. Dupuis YG: Response time: Flow by vs pressure. RRT 1988;
42(3):5.
~\ CHAPTER 26
~,
'
~)
~
,-,
\
'
.,
~
~
The Bird Mark 7 (Fig 26-1) is a single-circuit, pneumatically powered and controlled ventilator that is designed
to operate with air or oxygen at a pressure of 50 psi. 1 2
In Figure 26-2 the ventilator consists of.an ambient
compartment (2) and a pressure compartment (6), separated
by a large diaphragm (5). Connected to the center of the
diaphragm is a sliding valve, called a ceramic switch (9).
The opposite ends of the ceramic switch are connected to
metal disks called clutch plates (3,8). Directly in line with
the clutch plate in the ambient compartment is the sensitivity magnet (15), and in line with the clutch plate in the
pressure compartment is the pressure magnet (7).
The ceramic switch has a channel opl!ning for the passage of source gas to the various pneumatic components
within the ventilator. During the expiratory phase (Fig 262,A) the ceramic switch is held stable in the oFF position (to
the left) by the magnetic attraction between the sensitivity
magnet (15) and the ambient clutch plate (3). In this position the channel opening in the ceramic switch is not aligned
with the source inlet, and no gas flows through the ventilator.
The ventilator operates in three modes:
1.
2.
3.
Assist
Assist/control
Manual
Assist Mode
During an inspiratory effort, subatmospheric pressure
is generated in the pressure compartment and is communicated to the right side of the large diaphragm by way of two
openings in the center body of the ventilator. When the
inspiratory effort is sufficient to overcome the attraction
330
FIG 26-1.
Bird Mark 7 ventilator.
Ventilators
A
EXPIRATORY PHASE
331
INSPIRATORY EFFORT
FIG 26-2.
Pneumatic diagram of Bird Mark 7 ventilator. 1, Aowrate control; 2, ambient compartment; 3 , ambient clutch plate; 4, air inlet filter; 5 , diaphragm;
6, pressure compartment; 7, pressure magnet; 8, pressure clutch plate; 9, ceramic switch; 10, exhalation valve; 11, micronebulizer; 12 , test lung
(patient); 13, sensitivity arm; 14, hand timer rod; 15, ambient magnet; 16, source gas inlet; 17, Venturi; 18, pressure limit arm; 19, air mix plunger;
20, inspiratory drive line; 21, mainstream line. (Courtesy 3m Canada Inc., London , Ontario.)
.
INSPIRATION
EXHALATION
FIG 26-3.
Pneumatic diagram of Bird Mark 7 ventilator. See text for description. (Courtesy 3M Canada Inc., London, Ontario.)
(10)
332
,-]Assist/Control Mode
The ventilator can be set to automatically cycle to the
'inspiratory phase without patient intervention by the action
~of the expiratory timer cartridge (Fig 26-4). The expiratory .
r--
A
\
\
,
3
~\
'
4 (l
\
\
-,
-,
....._
'
-~
........_
,--..,
I
FIG 26-4.
Expiratory timer cartridge. 1, source gas inlet; 2, ceramic switch; 3 , ambient clutch plate; 4, expiratory timer ann; 5, diaphragm; 6, spring;
7, check valve; 8, drain hole; 9, expiratory timer control. (Courtesy 3M Canada Inc . , London, Ontario.)
(
Ventilators
Flowrate Control
The flowrate control is a simple needle valve that is
positioned between the source gas inlet -amrthe ceramic
switch. The control has a scale between OFF and the number
40, and although source gas to the ventilator is cut off when
the fl.owrate control is turned to the OFF position, the scale is
for reference only, and the numbers have no quantitative
meaning with respect to actual flowrate .
The flowrate control determines the rate of pressure rise
before cycling occurs and thus provides control over the
4
10
333
11
FIG 26-5.
Air mix plunger assembly. 1, Venturi; 2, spring; 3, air mix plunger; 4, 0-ring seals; 5, hole (100% source gas);
6, source gas inlet; 7, Venturi jet; 8, Venturi gate; 9, reed ; 10, safety catch; 11 , pressure balance hole. (Courtesy
3M Canada Inc ., London, Ontario.)
---....
34
'
Venturi Gate
reservoir bag connected over the ambient air filter will eventually empty.
Sensitivity Arm
The amount of inspiratory effort required to trigger the
ventilator is controlled by adjusting the sensitivity arm. The
arm positions the sensitivity magnet closer to or away from
its corresponding clutch plate. The closer the sensitivity
magnet is moved toward the clutch plate, the greater the
effort required to trigger the ventilator, and vice versa. The
scale for the sensitivity arm i~ also used for reference, but
normally, when the arm is moved to the number 15 (on-the
scale), an inspiratory effort of - 1.5 em H 2 0 is required to
trigger the ventilator.
Pressure Manometer
The pressure manometer is situated in the ambient compartment but samples the pressure in the pressure compartment. In the Bird Mark 7 the manometer is calibrated in two
scales ranging from -10 to 60 em H 20, and -7 to approximately 45 mm Hg.
Classification
Inspiratory Phase
The classification of the ventilator in the inspiratory
phase depends on the position of the air mix plunger. With
the plunger rrioved to the ouT position, the Venturi is engaged, and the ventilator operates as a constant pressure
generator for most of the inspiratory phase. However, toward the end of the inspiratory phase when less gas is entrained by the Venturi, most of the gas delivered is from the
high pressure source, and the ventilator approaches the
characteristics of a flow generator.
With the air mix moved to the IN position, all of the gas
delivered through the patient circuit originates from the 50
Ventilators
335
TABLE 26-1.
FIG 26-6.
Bird Mark 8 ventilator.
Air
m:ix;IN
Time cycled
Manually cycled
Classification
Apart from the expiratory phase, the classification of
the Mark 8 is the same as the Mark 7. During the expiratory
phase the Mark 8 has the added option of generating negative pressure in the patient circuit. Therefore the ventilator
must also be classified as a constant negative-pressure generator. Table 26-2 summarizes the classification of the Bird
Mark 8 ventilator.
336
--4
.---.__
FIG 26-7.
Flow diagram of negative interrupter cartridge in Bird Mark 8 ventilator during inspiration (A) and exhalation (B). See text for
description. I , source gas inlet; 2, negative generator valve; 3, negative interrupter cartridge assembly; 4, expiratory drive line .
(Courtesy 3M Canada Inc ., London, Ontario.)
INSPIRATION
EXHALATION
.\
FIG 26-8.
Mark 8 or Mark 9 breathing head assembly. 1, inspiratory drive line; 2, exhalation valve; 3, micronebulizer; 4 , mainstream line;
5, . test lung (patient); 6, expiratory drive line; 7, negative Venturi. (Courtesy 3M Canada Inc ., London, Ontario.)
Ventilators
EXHALATION
13
INSPIRATION
---=:::::::::17
FIG 26-9.
Pneumatic diagram of Bird Mark 8 ventilator. 1, source gas inlet; 2, negative generator valve; 3, Venturi; 4, ceramic switch; 5, negative
generator cartridge; 6, flowrate control; 7, air mix plunger; 8, expiratory timer control; 9, pressure clutch plate; 10, pressure magnet; 11,
pressure compartment; 12, expiratory timer cartridge; 13, inspiratory drive line; 14, mainstream line; 15, negative Venturi; 16, expiratory
drive line; 17, test lung (patient). (Courtesy 3M Canada Inc., London, Ontario.)
337
..../
38--,.
l'A-.. \ E 26-2.
:Ias~ification-The
TABLE 26-3.
Bird Mark 8 Ventilator
- ~-----------------------------------------
----
: h. Jeover I to E
:::t ---..,eover E to I
'Air mix
IN
""""'
-{.; assification
----.
FIG 26-10.
Rird Mark 9 ventilator.
FIG 26-11.
Bird Mark 10 ventilator.
Ventilators
339
I
I
-'FWi
J l ~.if
l
I. !
FIG 26-12.
Flow accelerator mechanism of Bird Mark 10 ventilator. A, 'Inspiratory phase, terminal flow control off. B, Inspiratory phase, terminal flow control
on. 1, Source gas inlet; 2, terminal flow control; 3, Venturi; 4, flow accelerator cartridge; 5 , flow interrupter cartridge; 6, ceramic switch. (Modified
from original. Courtesy 3M Canada Inc., London, Ontario .)
Classification
The classification of the Bird Mark 10, as summarized
in Table 26-4, is similar to the Mark 7 with only one exception. Because the ventilator is on permanent air dilution,
the pressure generated by the Venturi is usually not very
high, and the ventilator operates as a constant pressure generator for most of the inspiratory phase.
BIRD CIRCUITS
Standard Breathing Head Assembly
The standard assembly, shown in Figure 26-15, consists of an exhalation valve and a rnicronebulizer. During
inspiration the inspiratory drive line (1) powers the micro nebulizer and pressurizes the exhalation valve. The mainstream
line (3) provides the bulk flow for lung inflation. During
exhalation gas to both lines is cut off, and the exhalation
valve empties through the jet of the rnicronebulizer.
.;