Beruflich Dokumente
Kultur Dokumente
H I G H L I G H T S
G R A P H I C A L A B S T R A C T
A R T I C L E I N F O
A B S T R A C T
Article history:
Received 5 March 2014
Received in revised form 15 May 2014
Accepted 25 May 2014
Available online 29 May 2014
The capillary electrophoresis instrument developed for the concurrent determination of cations and
anions features two separate capillaries and individual detectors to allow independent optimization for
each group of ions. The capillaries are joined in a common injector block. The sample is drawn into the
injector with a small membrane pump and automated simultaneous injection into both capillaries is
achieved by pressurization of the uid with compressed air. Flushing of the injector and of the capillaries
with the background electrolyte is also carried out automatically by the same means. The buffer consisted
of 12 mM histidine and 2 mM 18-crown-6 adjusted to pH 4 with acetic acid and was suitable for the
contactless conductivity detection employed. The system was optimized for the determination of
cationic NH4+ and anionic NO
3 and NO2 , and linear calibration curves from about 20 mM up to about
1.5 mM were obtained for these ions. In a test run over 8 h, the reproducibility for the peak areas was
within 7%. For demonstration, the instrument was successfully applied to the concurrent monitoring of
the concentrations of the three ions during the biological removal of ammonium from contaminated
groundwater in a sequencing batch reactor, where NO
3 and NO2 are formed as intermediate products.
2014 Elsevier B.V. All rights reserved.
Keywords:
Dual-capillary electrophoresis
Capacitively coupled contactless
conductivity detection (C4D)
Simultaneous separations
Cations
Anions
1. Introduction
Capillary electrophoresis (CE) is a relatively simple method as
basically only a capillary and a high voltage power supply are
78
79
Fig. 1. Schematic drawing of the dual-capillary electrophoresis system. HV, high voltage; GND, electrical ground; V1, V2, electrically actuated 3-port valves; V3, V4,
electrically actuated stop valves; Pt: platinum electrodes.
80
Table 1
Typical operation sequence.
Step
Operation
Duration
1
2
3
4
5
6
7
8
15 s
180 s
4s
Variable
3s
12 min
15 s
180 s
and the injector block while both stop valves at the outlet are open
(designated as V3 and V4). Then the electrolyte is forced through
the capillaries for ushing by closing both stop valves at the outlet
of the injector block. Subsequently, valves V1 and V2 are turned
and sample is aspirated into the sample loop by activation of the
membrane pump. Valves V1 and V2 are then turned back to the
original position, and the sample plug is pushed into the injection
interface by the pressurized buffer. A split injection into both
capillaries is then performed by turning on the required backpressure, which is set with the needle valve, by closing only V3
while leaving V4 open. The interface is then ushed again to
replace the sample by background electrolyte before commencement of the separation. Both high voltage power supplies are
turned on at the same time for the concurrent separations of the
anionic and cationic analytes in the respective capillaries. The
common electrode in the interface remains grounded at all times.
Operation
V1, V2
V3
V4
High voltage
Membrane pump
1
1
2
1
1
2
1
1
Open
Closed
Open
Closed
Open
Open
Open
Closed
Open
Closed
Open
Open
Open
Open
Open
Closed
Off
Off
Off
Off
Off
On
Off
Off
Off
Off
On
Off
Off
Off
Off
Off
3.3. Monitoring of the concentrations of NH
4 , NO3 and NO2 during
biological removal of ammonium from contaminated groundwater
3.2. Performance
A slightly acidic background electrolyte (pH 4), which was
based on histidine, acetic acid and 18-crown-6 and had been used
successfully for the separation of inorganic cations as well as
anions by CE-C4D [9,35], was employed to investigate the
performance of the dual CE system. At the relatively low pH,
the electro-osmotic ow (EOF) is suppressed; therefore, no EOF
modication is needed. 18-crown-6 was included to facilitate
baseline separation of K+ and NH4+. An example of the concurrent
analysis of a standard mixture of cations and anions in the two
capillaries is shown in Fig. 3. As discussed, for example, in [35], a
CE-C4D system may be optimized either for fast separations, for
low limits of detection or for high separation efciency, and
compromises have to be made. In view of the application example
discussed below, the system was set up for high separation
efciency by injecting relatively short plugs of sample. The
calibration data for the three ions of interest (ammonium, nitrite
and nitrate) is given in Table 2. For NH4+ linearity up to 2000 mM
was achieved. For the anions nitrite and nitrate, the linear ranges
were somewhat shorter (up to 1500 mM). The correlation
coefcients obtained were better than 0.999 for all three ions.
The reproducibilities of the measurements of peak areas and
migration times were better than 5% and around 1%, respectively.
The system was then set up for a supervised test run over a
period of 8 h, during which repeated measurements of the
standard mixture were carried out automatically at intervals of
15 min. The membrane pump also enabled automatic aspiration of
the sample for each measurement. The results for peak areas are
shown in Fig. 4. The maximum deviations are less than 7%, which
is deemed acceptable considering that these are due to the
accumulation of the errors of all operations, i.e. sample loading,
delivery, injection, separation and temperature uctuations. A drift
in peak areas is not evident from the data for this 8 h run, which
demonstrates the suitability of the system for unattended
operation.
Fig. 3. Concurrent separations of inorganic anions and cations. (A) Cations: NH4+,
50 mM; K+, 75 mM; Ca2+, 75 mM; Na+, 225 mM; Mg2+, 75 mM; Li+, 50 mM. (B) Anions:
2
Cl (400 mM); NO
(75 mM). Electrolyte: 12 mM
3 (75 mM); NO2 (75 mM); SO4
histidine and 2 mM 18-crown-6 adjusted to pH 4 with CH3COOH. Capillaries:
fused silica, 50 mm i.d., 40 cm effective length and 55 cm total length. Sample
loop: 50 mL. Gas pressure: 0.8 bar. Separation voltage: +15 kV for anion- and
15 kV for cation-separation.
81
Table 2
Calibration ranges, limits of detection and reproducibilities for the concurrent determination of NH
4 , NO3 and NO2 . Electrolyte: 12 mM histidine and 2 mM 18-crown-6
adjusted to pH 4 with acetic acid.
Ion
Correlation coefcient, r2
NO
3
NO
3
NH
4
151500
251500
152000
0.9995
0.9997
0.9991
6.0
7.5
5.0
4.5
3.9
3.3
1.1
1.2
1.2
a
b
c
5 Concentrations.
Concentrations corresponding to peak heights of 3 times the baseline noise.
Relative standard deviation in %, n = 6.
Fig. 4. Stability test. The concentrations of the ions and other conditions were as for
Fig. 3.
Fig. 5. Schematic drawing of the activated sludge reactor. During the treatment
process, samples were drawn from tap 2, and at the end, after the sludge had settled,
clean water was taken from tap 3.
82
Fig. 7. Concentration proles of NH
4 , NO3 and NO2 monitored by CE during
the biological treatment of ammonium-contaminated groundwater. Other
conditions as for Fig. 3.
4. Conclusions
A dual-channel CE system for the concurrent determination of
cations and anions was constructed and successfully demonstrated
for the monitoring of biological nitrogen removal from ammonium
contaminated groundwater. The instrument is inexpensive, simple
in construction and can therefore be assembled with little effort.
The state-of-the-art contactless conductivity detectors can be built
with modest expertise in electronics. The only item that required
engineering workshop facilities was the injection block, but it
should be possible to substitute this with commercial capillary
connectors [21]. Pneumatic actuation proved to be a facile
approach to the implementation of hydrodynamic injection, which
is essential in order to avoid a bias which otherwise occurs when
samples of varying background conductivity are to be analysed.
Further integration and miniaturization in order to obtain an
readily portable instrument and battery operation are possible.
Acknowledgements
83