Beruflich Dokumente
Kultur Dokumente
ELECTRODE SYSTEM
C-Electrode
The geometry of C-Electrode is based on previous
laboratory test performed by [4] and [5]. Some
characteristics of C-Electrode are shown in Fig.1.
Fig.1 Design planes and assembly picture of CElectrode, (a) Top view, notice that needles are turned
/4 rad, (b) needles detail (c), Front view, notice the
needles inclination on the z axis, (d) and (e) assembly
picture. Notice the rain water protection.
C-Electrode consists on a bronze rod of 14,3 mm
diameter and 2,3 m length, finished on a needle tip, with
100 steel needles placed around as can be seen in Fig.1.
4 needles were placed around the electrode each 0,04 m,
turned /4 rad on the horizontal plane and inclined /4
rad in z axis, as it is shown in Fig.1 (a) and (c).
Additionally the C-Electrode assembly has a water rain
protection (see Fig.1 (e)), in order to avoiding possible
unexpected electrode-support contacts (the support is
electrically connected to earth).
P-Electrode
The P-Electrode is a plate with 82 needles, with the
same dimensions that those on C-Electrode. The
minimum distance between needles is 100 mm to avoid
shielding effects between them [5]. Design and
assembly are shown in Fig.2.
I corona [mA]
1,0
LABORATORY TESTS
C- and P- Electrodes laboratory tests were performed in
order to characterize their corona current.
C-Electrode
In this test C-Electrode was placed (without support, as
it is shown in Fig.3) within a cylinder, with dimensions
= 0,50 m and l = 1 m, with purpose to simulate quasihomogenous electric field conditions in a coaxial
arrangement. The number of needles were 100, 50 and
25, to evaluate the shielding effect between them. Test
was performance for a background electric field varying
within 13,2 kV/m and 140 kV/m. Background electric
field is defined here as applied voltage divided by
cylinder interelectrode distance. Fig.3 shows the
schematic circuit and the arrangement test.
100 needles
50 needles
25 needles
0,6
0,4
0,2
0,0
0
20
40
60
80
100
120
140
I corona [uA]
(a)
(b)
Fig.2 (a) P-Electrode design. Notice 4 orthogonal
perforations for the supports and another central
perforation for the measurement signal, (b) assembly
picture, P-Electrode is placed 0,80 m over the building
ceiling level (10,5 m).
0,8
-2
Rs = 1M
Rs = 100 k
-3
-4
-5
-35
-30
-25
-20
-15
-10
-5
SPACE CONSIDERATIONS
The electrodes were located at 10,5 m over floor level
on the building 411 of the National University of
Colombia in Bogot D.C. (4 36' N, 74 5' W) at 2600
meters over sea level.
CONCLUSIONS
(1) Two different corona electrodes were developed: A
cactus-like electrode and a plate with needles. Both
electrode types were tested and corona currents were
measured. Higher corona current were obtained with CElectrode than with P-Electrode.
REFERENCES