Service Manual
Tables of Contents
Charpter1 using This Manual.................................2
Charpter2 Hardware..............Error! Bookmark not
defined.
Chapter 3 liquid diagram analysis..........................9
Chapter 4 Histogram and pulse graph..................12
Charpter5 Troubleshouting....Error! Bookmark not
defined.
Chapter2 Hardware
1. Interface board
(1)the whole block diagram
P104 bus
Motor
control
and drive
unit
CPLD--->printer driver
Oneway
CPLD total
control unit
Valve drive
WBC-RBC enlarge
CPLD<--- WR
CPLD<---HGB
Two-way
Description:
(1) CPLD total control unit
Control every function module how to communicate with P104
2motor control and driver
Control motor working and amplify signal
3electromagnetic valve driver
3
Printer driver
HGB enlarge
J47
J39
6 36 4
2 1
J52
J29
3 2 1
2 1
2 1
P104 Bus
......
...
110V
1 2
J31
WBC-RBC Amplify
sign
J48
CPLD Control
Motor Driver
1 2 3
1 2 3 4
1 2 3
J38
V12
V1
2 1
1 2 3 4
1 2 3 4
1 2 3
3 2 1
WASTE
J45
2 1
J30
4 3 2 1
J50
DLN
CLEAN
V R8
4 3 2 1
Adjustable Resistor
J43
7 6 5 4 3 2 1 J24
SENS D Port
Adjustable Resistor
3 2 1
VR7
JP4
ASP REAGENT
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
SE N S C Po rt
Mo to r C Po rt
Motor D Port
POWER Control
JP5
2
4
6
8
V R9
J37
Valve Driver
1
3
5
7
Adjustable Resistor
J36
SE N S B Po rt
Mo to r B Po rt
J42
SE N S A Po rt
Mo to r A Po rt
J41
J27
J35
1 2 3
1 2 3
2 1
J40
Adjustable Resistor
3 2 1
VR6
Adjustable Resistor
POWER
LYSE
3 2 1
3 2 1
3 2 1
J32
J33
J34
J49
3 2 1
HGB Amplify
HGB
J44 5 4 3 2 1
VR10
Bexplanation
(1) Motor and sensor input port
(1) J40: motor A motor A in related mechanism fasten board
moduleconnection port
(2) J35: the related sensor A of motor A sensor A in related mechanism fasten
board moduleinput port
(3) J41: motor B motor B in related mechanism fasten board
moduleconnection port
(4) J : the sensor of motor B the sensor B in the mechanism fasten
boardinput port
(5) J42: motor Cmotor C in the mechanism fasten moduleconnection port
(6) J37: the sensor C of motor C the sensor of the related mechanism fasten
boardinput port
4
2. Control board
(1). The Connector Description
2. rinse analysis
3. lyse analysis
7
5. Waste Analysis
6. enhanced cleaning
Description: the space between the first line and the second line is lymphocyte; the
space between the second line and the third line is intermediate cell; the space between the
third line and the fourth line is granulocyte.
10
11
2. Pulse graph
After test, the pulse graph will be saved in the memory, it
can help you to analyze the problem.
(1) Normal Pulse Graph
Select special, then input the password: 6666, press F5 to
check WBC pulse graph; press F6 to check RBC pulse graph; press
F7to check PLT pulse graph; press F1 to return to main menu.
* The normal range of total pulse for normal people:
WBC< 1M
RBC< 600K
PLT< 1M
* the normal range of total pulse for normal QC:
WBC400 700K
RBC250450K
PLT300 600K
Below the normal pulse graph.
12
13
14
RBC
PLT Slight disturbance (normal sample) high concentration sample for PLT
Charpter5 Troubleshouting
Problem
description
WBCRBC
bubble
15
reason
Solution
WBC and
RBC clog
WBC or RBC
clog
16
HGB
failure HGB
bubble
Hemoglobin
unstable
17
A other
unstable,
B other
stable
parameters
are
parameters
are
PLT value is
too high in
ground test
PLT unstable
Vacuum pump
skid and make
a big noise
The value of A if the value of HGB is zero, A prime lyse, or change the piston
WBC is * or the lyse is not enough in the gasket ring of sampling room,
tube.
B if the total pulse of WBC less
0
B If the value of HGB is not
zero, after test QC or sample
18
RBC no value
PLT no value
distort no lyse
19
distort no
diluent
detector.
distort no rinse A check the voltage of 1th A if not, the sensor have broken,
pin of J33 it is about 1.2V
normally
B check the voltage of 3th
pin it is more than 3.0V
when the tube is filled with
reagent; it is less than 2.0V
when the tube is empty
change it
B if the voltage
or0V, change the
voltage is between
try to clean the
detector.
not be good at
classification
A inappropriate magnification
B inappropriate sampling
C inappropriate store of
blood
Only blank in
the screen
is OK
B no display and no other
function
Full waste
alarm
waste sensor
B it is OK when it is not use
the waste sensor
20
is about 5V
sensor. If the
2.0V and 3.0V,
white reagent