Beruflich Dokumente
Kultur Dokumente
analysis
Benefits and Experience
Donal Skelly
GE Energy
Lisburn, United Kingdom
donal.skelly@ge.com
PAS TECHNOLOGY
Photo Acoustic Spectroscopy (PAS) works along the
following principle: A gas substance absorbs light energy
following local heating by an IR light and transforms it into
kinetic energy (by the energy exchange process). Regularly
interrupting this process causes a series of pressure waves
(sound) that can be detected by microphones. By measuring
the sound at different wavelengths, the photo acoustic
spectrum of a gas sample can be recorded. This spectrum
can then be used to identify the absorbing components of the
sample (3).
Gas
Lower
detection
level in gas (ppm)
<1
<1
Methane (CH4)
<1
Ethane (C2H6)
<1
Ethylene (C2H4)
<1
Acetylene (C2H2)
<0.2
Methodology
Figure 1. Generic location of bond absorptions that
can lead, through vibrational excitation to the acoustic
effect in a Photoacoustic Spectrometer. When the exact
compounds of interest are known specific regions of the
spectrum can be focused upon for quantitative and
qualitative analysis.
RESULTS
Acetylene
60
165
162
111
142 103
19,4
58
60
58
62
61 -13,8
1,60
Gas Samples
Actual
Value
(ppm)
Min
Mean
Max
Std Dev
Ethylene
511.0
510.9
512.4
514.1
1.1
Ethane
509.0
513.8
514.3
514.7
0.4
Methane
493.0
495.0
495.3
495.7
0.3
Acetylene
499.0
504.2
504.5
504.8
0.2
Carbon
Dioxide
10000.0
10088.7
10113.1
10137.5
24.4
Carbon
Monoxide
506.0
509.7
510.4
510.8
0.4
Gas
69,4
143
Gas ID :
Quantification
Diff.
Std
Dev ppm
Ethylene
35,1 31
33
33
32
34
35
-5,98
Ethane
32,1 29
31
32
32
37
33
+0,73 2,66
Methane
36,3 31
34
35
33
36
38
-4,96
2,43
Carbon Monoxide
36,6 29
30
31
29
31
31
-17,6
0,98
7,12
Acetylene
34,7 29
1,17
31
31
30
32
32
-11,1
1,41
GAS CONCENTRATION
Range :
LOW
MEDIUM
HIGH
Unit :
(ppm)
(ppm)
(ppm)
Ethylene
17,4
35,1
70,2
Ethane
16,0
32,1
64,2
Methane
18,1
36,3
72,6
Carbon Monoxide
18,3
36,6
73,2
Carbon Dioxide
17,9
35,8
71,6
Acetylene
17,4
34,7
69,4
Gas ID :
Quantification
Std
Dev
(ppm)
Ethylene
17,4
17
17
17
17
17
16
-3,26
0,41
Ethane
16,0
17
17
18
15
15
16 +2,08
1,21
Methane
18,1
19
19
17
19
18
18 +1,29
0,82
Carbon
Monoxide
18,3
15
16
16
14
16
15
-16,2
0,82
Carbon Dioxide 17,9 110 109 131 124 105 111 +542
10,14
Gas ID :
Std.
Quantificatio
n
Std
Dev
(ppm)
Ethylene
70,2
62
61
63
60
64
65 -10,9
1,87
Ethane
64,2
59
60
61
59
59
63 -6,28
1,60
Methane
72,6
68
62
66
64
70
70 -8,17
3,27
Carbon
Monoxide
73,2
60
57
59
57
62
58 -19,7
1,94
Acetylene
17,4
16
16
16
16
15
15
-9,96
0,52
Field Experiences
In-service transformer oil PAS versus GC
The consistency of results for a random sample of oil from
an in-service transformer, were checked by PAS analysis
and also the Laboratory Head Space Analysis methods. The
results of several tests by each technique, on the same
specimen oil are shown in Table 7 and Table 8. A Standard
Deviation result in ppm is produced, together with the result
of the Standard Deviation as a percentage of the average
reading obtained. As there is no true standard value to
compare either method against, this is considered a test of
repeatability.
Gas ID :
Quantification
Ave
Test 1
Test 2
Test 3
(ppm)
Std
Dev
(ppm)
Diff
145
145
144
144,7
0,58
0,39
Ethane
(C2H6)
165
164
164
164,3
0,58
0,35
Methane
(CH4)
63
65
64
64
1,56
Carbon
Monoxi
de (CO)
32
31
33
32
3,13
Carbon
Dioxide
(CO2)
334
312
331
325,7
11,9
3,66
Acetyle
ne
(C2H2)
118
119
118,7
0,58
0,49
Gas Concentration
(PPM)
Gas ID :
Test 2
Quantification
Ave
Test 1
(%)
Ethylene
(C2H4)
119
Test 3
(ppm)
Std
Dev
(ppm)
Diff
(%)
Ethylene
161
161
164
162
1,73
1,07
Ethane
174
177
180
177
1,7
Methane
68
68
71
69
1,73
2,5
Carbon
Monoxide
38
41
44
41
7,3
Carbon
Dioxide
181
182
187
183,3
3,2
1,75
Acetylene
130
156
147
144,3
13,2
9,1
Table 8. Laboratory GC and headspace analysis of gas-inoil samples from field transformer
The PAS based analyser produced results which are in most
cases better than the Laboratory Head Space Analyser in
terms of repeatability, as would be expected from an online
unit not subject to sampling variance. In general the results
between both methods compared at a level that would not
CONCLUSION
PAS based DGA instruments have been developed with
the express purpose of addressing the shortcomings of
online GC based instruments. They provide a real
alternative to GC by matching their performance and
operating successfully in the field. Utilising a
technology historically designed for online application,
PAS instruments are very stable and repeatable
monitoring instruments suited for the tough
environmental and operational demands associated with
remote transformer monitoring. PAS is the new highend standard for monitoring critical transformers.
REFERENCES
[1]
[2]
[3]