Beruflich Dokumente
Kultur Dokumente
Bioengineering
AND
MIKHAIL SKLIAR*
entricular assist devices (VADs) have been used successfully for many years as a bridge to transplantation1, and they
hold the potential to become long-term alternatives to donor
heart transplantation (destination therapy). However, a control
system that automatically responds to physiologic cardiac demand for the continuous flow VAD does not exist. In hospitals,
From the *Department of Chemical Engineering, University of Utah,
Salt Lake City, Utah; and the Jewish Hospital Heart and Lung Institute,
University of Louisville, Department of Surgery, Louisville, Kentucky.
Submitted for consideration June 2003; accepted for publication in
revised form May 2004.
Presented at the 49th ASAIO Conference, June 19 21, 2003, Washington, DC.
Correspondence: Dr. Guruprasad A. Giridharan, Department of
Chemical Engineering, 50 South Central Campus Dr., Rm 3290 MEB,
University of Utah Salt Lake City, UT 84112.
DOI: 10.1097/01.MAT.0000136652.78197.58
403
404
GIRIDHARAN ET AL.
Table 1. Baseline Parameters of the Mock Circulatory System Used in Simulating the Human Equivalence of a Normal, Failing,
and Asystolic LV
Normal LV
Parameter
a
Failing LV
Asystolic LV
Rest
Exercise
Rest
Exercise
96
3.8
210
100
5.4
300
68
2.0
100
75
3.1
150
Rest
Exercise
0
0
0
0
LV, left ventricle; AoPr, aortic root pressure; CO, cardiac output; DLP, drive line pressure.
405
were constructed by plotting ventricular pressure against ventricular volume, in which each loop represents one complete
cardiac cycle (one beat). Characterizing hemodynamic parameters and pressure-volume loops were calculated for all experimental conditions.
Results
The hemodynamic parameters for a normal, failing, and
asystolic LV with and without continuous assist for each of the
three control strategies during rest and light exercise are listed
in Table 2. Without VAD assistance, the values of the total flow
rate, P, and Pa decrease during ventricular failure at rest
and exercise in comparison with the normal LV at rest and
exercise. The left ventricular end diastolic pressure for all the
control strategies remain within 3 mm Hg of the baseline
normal LV value. Because the left ventricular pressure and
volume sensor is introduced through the aortic valve (Figure
1), there is a back flow through that valve for baseline and all
VAD assist scenarios.
Table 2 indicates that all of the tested control strategies
increase the total flow, P, and Pa with failing and asystolic
LV. The Pa control strategy maintains or restores the total
flow rate to that of the physiologically normal heart during rest
and exercise and adapts best to the need for support. For
example, in the case of the normal heart during rest and
exercise, the average net VAD flow rate with this strategy is
close to zero, as expected, because the native LV provides all
the required cardiac output.
Figure 2 shows the comparison between the total flow rates
(sum of VAD flow rate and cardiac output) at baseline and
during assistance using constant rpm, P, and Pa control
strategies during rest and exercise scenarios for each clinical
test condition. The baseline cardiac output of 3.8 L/m at rest
and 5.4 L/m during light exercise are considered to be physiologic flows for the corresponding physical activities. The
comparison of total flow rates produced with different control
strategies shows that the Pa approach best matched the physiologic flow rate. The comparison of rest and exercise cases
shows that the control strategy of maintaining an average Pa
406
GIRIDHARAN ET AL.
Table 2. Comparison of Assisted and Unassisted Perfusion Under Different Scenarios
Total Flow
(L/m)
P
(mm Hg)
Pa
(mm Hg)
0
0
0
0
0
0
3.8
2.0
0
5.4
3.1
0
54.3
30.5
54.4
32.9
94.4
51.5
98.4
58.7
0.0
0.0
0.0
0.0
0.0
17.2
6.7
16.4
800
1440
1490
650
1490
1600
3.6
3.9
3.9
5.5
5.2
5.7
55.3
71.9
96.4
51.7
68.6
98.3
95.0
95.2
94.9
95.8
94.4
95.2
0.2
4.5
4.5
0.4
5.6
6.3
1.0
0.8
1
5.3
5.1
2.8
1300
1450
1320
1280
1600
1400
4.6
4.0
3.3
6.4
5.7
4.8
75.3
78.2
75.6
74.6
76.6
74.1
115.4
101.2
74.5
117.5
101.5
71.6
3.2
4.5
3.9
3.5
6.2
5.4
1.3
1.3
0.0
7.6
6.2
0.1
1450
1440
1450
1440
1440
1440
4.7
3.9
3.7
6.8
4.8
5.0
80.1
71.9
89.3
80.0
70.5
76.6
120.2
95.2
87.8
122.4
97.0
74.0
3.8
4.5
4.3
4.4
5.3
5.5
1.8
0.8
1
7.9
4.8
0.6
VAD rate
(rpm)
Baseline values
Normal LV, rest
Failing LV, rest
Asystolic LV, rest
Normal LV, exercise
Failing LV, exercise
Asystolic LV, exercise
Centrifugal VAD with Pa control
Normal LV, rest
Failing LV, rest
Asystolic LV, rest
Normal LV, exercise
Failing LV, exercise
Asystolic LV, exercise
Centrifugal VAD with P control
Normal LV, rest
Failing LV, rest
Asystolic LV, rest
Normal LV, exercise
Failing LV, exercise
Asystolic LV, exercise
Centrifugal VAD with rpm control
Normal LV, rest
Failing LV, rest
Asystolic LV, rest
Normal LV, exercise
Failing LV, exercise
Asystolic LV, exercise
VAD Flow
(L/m)
LVPeda
(mm Hg)
VAD, ventricular assist device; LVPed, left ventricular end diastolic pressure, LV, left ventricle.
LVP for the asystolic LV is a constant value
407
mand better than any alternative strategies. The Pa, being the
difference between the aortic and the pulmonary venous pressure (equal to left atrial pressure), is sensitive to changes both
in preload and afterload. The current in vitro study shows that
the proposed strategy of maintaining the desired average Pa
leads to an adequate adaptation for widely changing cardiac
demand and clinical conditions in a completely autonomous
way. The results show that, although some degree of physiologic adaptation is achieved with constant rpm and constant
P, these alternatives are less effective for a wide range of
physical activities and rapidly changing status of cardiac function (such as a sudden transition from failing to asystolic heart).
With the constant rpm control strategy, a broad range of
physical and clinical conditions would require an external
intervention to change the rpm setpoint according to some
expert rule, model prediction, or operator input. The constant
P approach does not perform well for a broad range of
clinical conditions of the native heart, which changed from
normal to asystolic LV in this study, but adapts well to the
changing cardiac demand due to different exercise levels. The
in vitro study is consistent with the results of computer simulations,17,26 28 which showed that the P control strategy
adapts better to widely varying cardiac output requirements
when compared with the traditional constant rpm control
approach. Because of the limitations of the mock circulatory
system, we were unable to test the higher cardiac demand
conditions to make an in vitro comparison between the different control strategies. In the limited range of cardiac demands
that could be tested in vitro, the performance of the Pa
control strategy is superior to P and constant rpm control
alternatives.
Using the proposed approach, both the natural heart and the
assist device are contributing to the pumping action of maintaining an average Pa. If cardiac function improves, the
native heart will increase its contribution to maintaining the
reference pressure difference, with the VAD controller autonomously and automatically responding to the decreased need
for assisted perfusion, as evidenced by the near zero net VAD
flow rate with the normal heart during rest and exercise. When
the net flow rate through the VAD is close to zero, blood does
not stagnate inside the VAD, although the residence time of
blood in the pump is higher, increasing the probability of
hemolysis. Because this scenario occurs only with a normal or
near normal heart, the patient could be weaned from the pump
at this stage.
The ability of the proposed control strategy to automatically
adjust its contribution towards maintaining Pa may prove to
be well suited to the application of the ventricular assist devices in cardiac recovery therapy14,15,29 of end-stage heart
failure (alone or in combination therapy), as well as in the
destination therapy.
Although not directly addressed in this study, the overarching principle behind the proposed approach of maintaining
key pressure differences with mechanical blood pumps, while
relying on the natural regulation to adjust the resistances to the
blood flow to meet the physiologic demand, is also applicable
to the case of pulsatile ventricular assist devices, as well as the
total artificial heart. In the case of the TAH, the blood pump
should be controlled to maintain key pressure differences at
the average reference values, which, in the case of pulmonary
408
GIRIDHARAN ET AL.
Limitations
The performance of the mock circulation during normal,
failing, and asystolic heart test conditions is representative
of clinical observations from a purely hemodynamic/hydrodynamic viewpoint. Clearly, mock circulation experiments
are limited in replicating clinical conditions. For instance,
mock circulation cannot mimic neurohumoral responses,
tissue remodeling, or activation of regulatory proteins. The
particular mock circulatory system used in the experiment is
a single sided mock and can simulate the systemic (left side)
circulation only. In this mock system, the uptake cannula
inflow occlusion is physically impossible, and the reduced
LV pressure and volume are the only indications of suction.
The instrumentation used to record hemodynamic waveforms in the present study has inherent measurement errors
associated with each technique (i.e., pressure error 1
mm Hg, flow error 0.5 L/min), which we have attempted to minimize by using GLP compliant test equipment, calibration procedures, and documentation practices.
The gains and offsets assigned for the LV volume data may
have inherent errors associated with it, as it is based upon
the aortic root stroke volume, LVPed, and total flows
produced.
Despite these limitations, these results provide valuable insight into the differences of each control algorithm, which aids
in further development of control algorithms and experimental
protocols to identify optimal ventricular assist therapies that
can subsequently be validated in in vivo models.
Conclusions
The in vitro results show that maintaining an average pressure difference between the pulmonary vein and aorta (Pa)
provides an effective way to control a continuous flow LVAD
over a wide range of physiologic and cardiac demand conditions while reducing the probability of suction. Change in
vascular resistance is the dominant regulatory mechanism in
meeting the physiologic requirements for blood perfusion.
Maintaining the desired average Pa by adjusting the pump
rpm during changing cardiac demand, in effect, synchronizes
the assisted and natural perfusion. Therefore, the proposed
control strategy indirectly incorporates natural cardiovascular
regulation, which changes vascular resistance into VAD control. The comparison with the VAD control systems, which
maintain either constant reference pump rpm or constant
pump pressure head (P), shows that the proposed approach is
superior in autonomously maintaining an adequate perfusion
during changing cardiac demand for the test conditions simulated in vitro. Because the Pa control strategy automatically
adjusts its contribution to the total flow based upon the function of the native ventricle, the proposed approach may prove
to be well suited to the application of the ventricular assist
devices in recovery therapy.
Acknowledgment
This study was supported by the Established Investigator Award from
the American Heart Association.
References
1. Olsen D: Sixth international symposium for rotary blood pumps.
Artif Organs 26: 475 476, 1999.
2. Klute G, Tasch U, Geselowitz D: An optimal controller for an
electric ventricular assist device: Theory, implementation and
testing. IEEE Trans Biomed Eng 39: 394 403, 1992.
3. Boston J, Simaan M, Antaki J, Yu Y: Control issues in rotary heart
assist devices. Proceedings of the American Control Conference
34733477, 2000.
4. Schima H, Trubel W, Moritz A, et al: Noninvasive monitoring of
rotary blood pumps: Necessities, possibilities, and limitations.
Artif Organs 16: 195202, 1992.
5. Takatani S: Open-loop analysis of circulatory system in awake live
animals: Relations between mixed venous saturation (SvO2)
and hemodynamic parameters. Artif Organs 17: 79, 1992.
6. Takatani S, Noda H, Takano H, Akutsu T: Continuous in-line
monitoring of oxygen delivery to control artificial heart output.
Artif Organs 174: 458, 1990.
409