Beruflich Dokumente
Kultur Dokumente
Abstract The introduction of Raman spectroscopy into new elds has been driven
largely by advances in the underlying technology. While the spectrometer is still
comprised of a light source, a wavelength selector, and a detector, the improvement
in functionality of each of these components has had dramatic impacts on areas
where Raman was once thought impractical, if not impossible. In addition, esoteric techniques once conned to academic spectroscopy labs are now nding wide
application.
This chapter will briey describe the basic function of a Raman spectrometer, while focusing on the enabling advances in spectrometer components including
capability, exibility, ease of use, and cost. Traditional laser sources have become
commodity items with air cooling, smaller forms, and lower cost in a variety of
wavelengths. Likewise, availability of high-power, tunable, and pulsed laser systems
have facilitated the use of higher-order techniques such as UV-resonance Raman
and CARS. The spectrograph form can be selected based on application, from traditional dispersive, to FT, to liquid crystal tunable lter, etc. Introduction of the
Raman signal to the spectrometer using ber optics has seen similar advances, such
as SORS and other ber bundle techniques. Detectors have become more sensitive,
with lower noise, better detection at longer wavelengths, and faster operation. Many
of these advances have resulted in miniaturization of the Raman system, so that systems once requiring an entire lab bench can now be handheld. Applications where
these advances have made signicant impact will be highlighted.
1.1 Introduction
The evolution of Raman spectroscopy from a spectroscopic novelty, to a complementary technique in niche applications, to an analytical powerhouse, has
closely paralleled the advancement of enabling technologies. While a simple block diagram of the components of a Raman spectrometer shown in
Figure 1.1 would still be comparable to the very early instruments built by
C.V. Raman [1], the improvement in functionality of each component has dramatically increased the impact of Raman spectroscopy in areas where it was
F. LaPlant
F. LaPlant
Light
Source
Sample
Detection
Dispersion
the phenyl ring would have a weak response, while the OH would be very
strong. However, in both infrared and Raman spectra, the exact frequency
and intensity of a given vibrational band will be eected by interactions with
the other vibrational modes present in the molecule. This means that the
spectrum for a given molecule is eectively a ngerprint of that molecule.
This is recognized in the US Pharmacopeia in that either the mid-IR or Raman spectrum can be used as conclusive identication of a pure substance.
This also has the important implication in that Raman spectroscopy can be
used to determine specic molecules that may be markers for disease states,
biothreats, or environmental contaminants.
The Raman eect has also been broadly applied to online and bench-top
quantitative applications, such as determination of pharmaceutical materials and process monitoring [46], in vivo clinical measurements [7], biological
materials [8, 9], to name only a few. Because the absolute Raman response is
dicult to measure accurately (sample presentation and delivered laser power
can vary), these measurements are almost always calculated as a percentage
with respect to the response from an internal standard. This standard is typically part of the sample matrix; in a drug product, the standard may be an
excipient; in a biological sample, it is commonly water.
This is represented graphically in Fig. 1.1. These spectra are from a calibration set consisting of diering levels of ethanol in decanol ranging from
0 to 10%. As described above, the spectra-to-spectra overall intensity varies
signicantly. A wide variety of approaches can be taken to normalize the spectra, suppress background interference, and enhance selectivity and accuracy.
Quantitative methods for Raman spectroscopy using univariate and multivariate methods are treated extensively elsewhere [10]. In this example, a second
derivative was applied to the spectra, which removes baseline variation and
enhances band resolution, followed by a univariate intensity calibration. Although limit of quantitation will vary widely depending on the analyte and
matrix, the results shown in Fig. 1.2 (quantitation in the 1% by weight
range) are typical.
Although both Raman and IR can be used to determine molecular identity
or quantify the presence of an analyte, there are critical dierences in the way
that they can be applied. The extremely strong absorption of OH in the midIR makes any measurement dicult in the presence of water a ubiquitous
material in most biological samples. Subtle spectral dierences that may be
apparent in the Raman spectrum tend to be obscured by the overwhelming
water absorption. This is shown in Fig. 1.3 for a solution of lactose in water;
1800
1600
1400
1000
800
Raman Shift (cm1)
600
400
200
Fig. 1.2. Comparison of raw and second derivative spectra for EtOH in decanol
while the water contributes only weakly to the Raman signal, with most of the
spectral features due to lactose. In general, the strength of the absorptions in
the mid-IR also limits its optical exibility; because almost all of the light is
absorbed within the top few microns of the sample, only the surface (or a very
thin section) of a sample can be measured. The capacity of the Raman probe
laser to penetrate into a sample and generate signal through a glass vial,
through layers of water, through layers of tissue, or through any nominally
transparent material is a signicant advantage. This advantage is possible
only because the interaction of the laser with the intervening material is weak
as is the resulting Raman signal.
The likelihood of a Raman photon being generated is dependent on a
variety of factors, but is generally on the order of one out of a million incident photons. The most directly controllable variable is the frequency of
excitation. Within a rst approximation, the intensity of the Raman signal
4
is proportional to 1/ () , where is the wavelength of the exciting photon.
This approximation assumes that the excitation is nonresonant; in the case
where a resonance exists, the signal can be orders of magnitude larger, and
this relationship does not hold. This wavelength dependence means that as
the photon wavelength becomes longer (shifted toward the red), the intensity
of the Raman signal decreases rapidly. In order to maximize the Raman signal, it would appear that the highest energy photon (or bluest laser) would be
the most desirable. However, uorescence is a competing process that can be
millions of times more ecient than the Raman eect. The transitions giving
rise to uorescence are more probable in the blue region of the spectrum, so
that increases in Raman signal at shorter excitation wavelengths are typically
more than oset by increases in background uorescence. Due to this inequal-
F. LaPlant
100% Decanol
.2
0
7
6
0.2
0.4
10% Ethanol
4
3
2
1
0.6
940
920
900
880
1
860
840
820
0.15 0.10
800
0.05
0.00
0.05
0.10
0.15
0.20
Raman Response
Fig. 1.3. Second derivative response for EtOH in decanol and calibration results
ity in signal strength, even trace quantities of uorescent materials can mask
the presence of a high-concentration analyte. Fluorescence can be minimized
by exciting at longer wavelengths, so that wavelength selection is a balance
between maximizing signal strength and minimizing interference.
The critical factor aecting every aspect of the Raman spectroscopy system is the weakness of the Raman eect: there are simply very few Raman
photons generated. All of the technical hallmarks of the modern Raman system, including lasers as an excitation source, high eciency collection geometries, high-throughput spectrometers with strong rejection of the laser
wavelength, and low noise, high-sensitivity detectors are all a result of the
necessity to absolutely maximize the sensitivity and selectivity of the Raman
response.
1.3 Lasers
The laser is very much the ideal light source for generation of Raman scattered photons. The light output from a laser is typically a single wavelength
(monochromatic) or a very narrow wavelength band. Because the width of an
observed Raman band is a convolution of the natural linewidth of the vibrational band with the laser linewidth, a broad frequency excitation source will
yield a spectrum with broad, poorly resolved Raman bands. Other practical
benets include ecient rejection of Rayleigh scattered photons (enabling very
low-frequency measurements) and high excitation intensity. Characteristically
low beam divergence enables optimization of a variety of sampling parameters. The laser can be focused to the diraction-limited spot size (smallest
eective spot diameter approximately equal to the laser wavelength) for high
photon ux at the measurement zone. This is the principle behind Raman
microscopy, where sample spectra can be obtained with very high spatial resolution. This also allows facile and ecient coupling to ber optics, which in
turn generates a variety of exibility advantages that will be discussed further in the next section. These advantages ensured that mercury arc lamps
were abandoned rapidly and universally soon after the introduction of the
laser [11].
Advances in laser technology, coupled with improvements in the supporting components, now enable a wide variety of lasers to be used. By far the most
common laser used in the modern Raman spectrometer is the diode laser operating at 785 nm. Diode lasers lack some of the advantages of gas lasers, such as
inherent wavelength stability and beam prole, although these dierences are
minimized in modern diode laser designs by using active temperature stabilization, dispersive intra- or extra-cavity optical elements, and beam-shaping
optics. However, there are multiple practical advantages that drive the use of
diode lasers in Raman applications. Although uorescence may not be totally
eliminated at 785 nm, it is signicantly depressed. As shown in Fig. 1.4, the
intensity of the Raman signal will be about a factor of 5 lower for a 785 nm
diode vs. the Ar-ion 514 nm line. However, the absorption of most uorescent
materials drops o rapidly over this region, so that samples that would generate sucient uorescent photons to saturate the detector with 514 nm or
632 nm excitation are often very easily measured at 785 nm. Another factor
is that the most common silicon-based detectors have a maximum detection
eciency in the region of 8001000 nm, which coincides with the region in
which the majority of the Raman signal of interest is generated (although the
quantum eciency falls of rapidly past 1000 nm, and the response to the CH
region of the spectrum at 3000 cm1 is diminished). The depth of penetration
of NIR light into biological tissue (many millimeters or more) can also be a
signicant advantage over visible excitation when trying to detect or visualize
subsurface materials.
There are several strategies to cope with samples that still exhibit unacceptable uorescence at 785 nm. The obvious direction would be to make the
excitation even redder; for example, some modern systems also operate at
830 nm to take advantage of further reduction of uorescence of biological tissues. More commonly, the solid-state Nd:YAG laser operating at the 1064 nm
fundamental is used in conjunction with a Fourier transform spectrometer
(this will be discussed in more detail in Section 1.5). At this wavelength, uorescence from all but the most intractable samples (e.g., charred materials,
heavy petroleum distillates) is negligible. However, the Raman signal is again
diminished by a factor of about 3.5 from excitation at 785 nm. To compensate,
the laser power is typically increased accordingly, where 10100 mW lasers are
common in the visible, 12 W lasers are the norm at 1064 nm. A variety of
sample protection schemes, such as sample spinning or wiggling, have been
implemented to prevent burning in these systems.
F. LaPlant
100
Mid-IR
80
60
Lactose + H2O
40
Raman
H2O
20
1800
1600
1400
1200
1000
800
600
400
10
F. LaPlant
[22]. This combination of dramatic cost and size reduction has resulted in
broad impact on Raman instrumentation. Raman microscopes, for example,
often now have multiple excitation lasers integrated into the microscope body
with little or no increase in the instrument footprint. There are a variety of
handheld Raman systems that can operate for many hours on battery packs
intended for handheld cameras. Instruments can be taken into the clinic for
in vivo measurements, or integrated into the manufacturing process on the
factory oor. This exibility also allows facile choice of laser wavelength for
specic measurements; for instance, uorescence from the heme complex can
be minimized by exciting at 830 nm rather than at 785 nm [23]. Laser wavelength can also be selected for resonant enhancement of specic vibrational
bands, or to maximize the enhancement from dierent SERS substrates. The
expense, availability, and cumbersome nature of the laser are no longer limiting factors in Raman spectroscopy; rather, selection of the laser is a key factor
in optimizing the results for a given application.
Diode - 785
Diode - 830
HeNe - 633
10
Ar-Ion - 488
Ar-Ion - 514
2x Nd: YAG - 532
100
3x Nd:YAG - 366
HeAg - 224
NeCu - 248
Although the other components of the Raman spectrometer are always highlighted as the most technically important, many important innovations have
been applied to optimizing and amplifying the signal by improving how
the sample interfaces between the excitation laser and the spectrometer. In
Fig. 1.1, this component would be represented by the arrow connecting the
light source with the sample, and the sample with the dispersion element.
Because dierent variations will be treated in detail in upcoming chapters on
0.1
200
400
600
800
1000
1200
11
ber probes, SORS, PhAT, etc., this section will only highlight these technologies and their relation with the other components.
As discussed previously, the relative weakness of the Raman eect has
always required that the Raman spectrophotometer be as ecient as possible
at generating, collecting, and detecting the scattered Raman photons. Because
the entrance to the spectrograph was a slit, and the slit needed to be as narrow
as possible to maintain spectral resolution, the focus of the laser on the sample
also had to be as small as possible (since the focus of the laser was imaged onto
the spectrometer slit). Because the Raman signal ideally scatters uniformly
in every direction, the signal photon ux drops o rapidly with distance from
the sample, and the collection optics needed a low f -number set as close to the
sample as possible for ecient signal collection. This made the presentation
of the sample to the Raman system extremely inexible; if the sample could
not be adapted to the constraints of the spectrometer, the experiment could
not be performed. In addition, because both the laser and the signal photons
were open beam, they needed to be steered by multiple lenses and mirrors
to keep the system properly aligned. This has some advantages; for instance,
maintaining the lasers Gaussian beam prole yields the smallest focal spot
at the sample (important for high spatial resolution in confocal microscopy),
and reective optics may be the only way to transmit very high peak powers.
However, maintaining stable geometry between the laser and spectrometer
is dicult; even minor thermal variations can cause signicant changes in
the angle at which the signal enters the spectrometer, shifting the apparent
wavelength and making accurate calibration challenging.
The introduction of ber optics to transmit the excitation laser to the
sample and the signal to the spectrometer yielded the double benet of both
improved geometric stability and sampling exibility. Although there are a
variety of specialty ber constructions for transmitting in the mid-IR, preserving polarization state, narrow wavelength transmission window, etc., the
most common type of optical bers are comprised of a silica core surrounded
by a material of higher index of refraction; this material may be either another
doped silica or a polymer. The dierence in the refractive index of the two
materials will set the critical angle at which light will be transmitted down
the length of the ber, expressed as a numerical aperture (NA). Coupled with
the diameter of the core, this will set the allowable modes that can propagate
through the ber. The eective result of increased numerical aperture is that
the angle at which light can be accepted from the sample increases. Typical
NA values will be between 0.12 and 0.22, although specialty bers may have
NAs as high as 0.40 or higher. The important issue is that for whatever ber
chosen, the NA is a critical parameter in the construction of the entire Raman
system. The laser launch conditions, the focusing and collection optics in the
ber probe, and the spectrometer f -number must all take the ber NA into
account to maximize signal throughput.
A multitude of dierent probes with varying ber orientation have been
proposed [24] although most are essentially similar to the rst fairly simple
12
F. LaPlant
13
1.5 Spectrometers
Although there are a variety of wavelength selection methods available, the
vast majority of Raman instruments utilize either dispersive or Fourier transform spectrometers. These are shown schematically in Fig. 1.6. The high
throughput and spectral resolution obtainable from these instruments make
them obvious choices for Raman spectroscopy; however, each has specic
strengths and drawbacks which make them more suitable in specic applications.
The primary function of the dispersive monochromator is to spatially separate photons of dierent energy coming from the sample. The degree of separation (dispersion) is a function of groove spacing in the grating; the narrower
the groove spacing, the higher the dispersion and the narrower the spectral
window observed. Coupled with dispersion is the resolution of the instrument,
which is how nely a narrow spectral feature can be observed. While this is
also a function of the grating dispersion, the input and output apertures are
limiting factors. While traditional monochromators used adjustable slits as
apertures, many modern dispersive instruments no longer have true slits but
use the input ber optic diameter as the input aperture, and the pixel size
on the CCD camera as the eective output aperture. This eectively xes
the limiting resolution of the instrument. Since most liquids and solids have
relatively broad natural linewidths, systems designed for routine applications
typically have a resolution of a few wave numbers.
Historically, the second most important function of the Raman spectrometer was to reject scattered laser radiation. Because the laser scatter from
the sample can be many millions of times more intense than the signal and
cannot be rejected spatially, the power of the spectrometer to eciently and
Dispersive
Fourier Transform
Fixed Mirror
Beamsplitter
Detector
Moving
Mirror
CCD
Grating
Notch
Filter
Aperture
14
F. LaPlant
completely separate this light from the signal was critical. The traditional
dispersive solution to this problem was the use of a multi-stage monochromator. Large double or triple monochromators provide high spectral resolution
and excellent rejection; spectra can be measured as near as 10 cm1 from the
laser line. The introduction of the holographic notch lter in the early 1990s
provided a small, ecient method of selectively rejecting the laser photons
before they entered the spectrometer. This enabled the used of (relatively)
crude single monochromators like that shown in Fig. 1.6, and paved the way
for the current generation of handheld instruments employing very minimal
dispersive elements. Holographic notch lters were used because the dielectric
stack lters available at the time had poor throughput; signicant advances in
lter manufacturing technology now produce dielectric stack long-pass edge or
notch lters with >90% transmission eciency, allowing signal to be collected
as close as 50100 cm1 from the laser line, as well as operating well into the
UV Raman range.
Although single monochromators with external laser lters are the only
dispersive systems in common use, double and triple monochromators still
have utility in some specialty applications [31]. Very low-frequency features
arising from phonon modes such as Brillouin scattering, or rotational modes,
are still best observed using this technique, since lters do not cut o eciently enough to operate this close to the laser wavelength. For some resonance experiments where a tunable laser excitation source is used, it may
also be impractical to use a xed lter to reject the laser line, and wavelength
selectibility and laser rejection become more critical than throughput.
Monochromator design has also seen signicant advances, especially with
respect to improved gratings. Because many of the emerging applications are
associated with chemical imaging of the sample rather than single-point measurements, imaging spectrographs have been designed specically to generate
a at-eld output at the detector plane. Traditional gratings generate signicant optical aberration including curvature of the input image and nonplanar
focus at the detector. Aberration corrected holographic transmission gratings
or nonplanar reection gratings have largely eliminated these eects (imaging will be discussed at greater length at the end of this chapter). Reective
diraction gratings have also improved to the point that they are approaching
the eciency of transmissive holographic gratings, which may be as high as
90% throughput at peak eciency [32]. Signicant exibility in the footprint
of the resulting spectrometer is possible through careful grating selection. A
wide variety of miniature, fully integrated spectrometers are now commercially available as stand-alone units or as components integrated into systems
by OEMs. Echelle spectrometers are also available, which parse the dierent
regions of the spectrum such that they can be measured simultaneously in a
stack on the detector, allowing rapid measurement of spectra at high spectral
dispersion and wide wavelength spectrum in the coverage.
Fourier transform spectrometers measure the entire spectrum simultaneously in the form of an interferogram, rather than dispersing the spectrum
15
into separate wavelengths [33, 34]. As shown in Fig. 1.6, the interferogram
is generated by changing the path length of the signal by varying the position of the moving mirror. This position is monitored very accurately using
the interference fringes from the HeNe reference laser. The detected signal is
then transformed into frequency, and a spectrum generated that is essentially
equivalent to the dispersive response. The limiting resolution of the FT instrument is the distance that the moving mirror travels; even a fairly modest
mirror movement of a few centimeters can give tenths of a wave number resolution. Because the full spectrum is contained in the interferogram, a wide
spectral window can be obtained at high resolution. The spectrum also has
inherent wavelength accuracy, since the spectrum is internally calibrated by
the reference laser. Multivariate models can be more reliably applied to FT
data, and the data between instruments will be much more consistent. More
importantly, FT-Raman can work more eciently using a long-wavelength
1064 nm excitation laser, which eectively eliminates uorescence. This can
be important advantage when working with biological samples, although typical FT-Raman laser powers can be well above the damage threshold for many
samples [35].
Unfortunately, Fourier Transform instrumentation has been overshadowed
to some extent by the advances made in dispersive technology. This is in no
small part due to the ability of dispersive manufacturers to leverage advances
in the rapidly evolving optoelectronics market, while FT-Raman depends
heavily on already mature FT-IR platforms. Although better detectors and
laser rejection lters have improved the noise characteristics of the system,
the fundamental throughput advantage that made FT a universally adopted
approach in the mid-IR was never suciently compelling to engender a similar dominance in Raman, principally because of the availability of low-noise
CCD detectors in the visible. The almost total suppression of background
uorescence is a great advantage; however, the recent introduction of lownoise multichannel detectors for the near-IR region (9501650 nm) has enabled
the development of dispersive systems operating with higher eciency with
1064 nm excitation. In spite of FT-Ramans intrinsic advantages, it is unclear
how well it will compete with the small, cheaper, more sensitive, exibly congured, and all-solid-state dispersive instruments as an enabling technology
driving the emerging applications described in this book.
Another wavelength selection mechanism of note is electronically tunable bandpass lter. There are two commercially available technologies, the
acousto-optic tunable lter (AOTF) and the liquid crystal tunable lter
(LCTF). Each uses dierent underlying mechanisms. The AOTF uses an
acoustic wave generated in an optically clear crystal to change the angle of
the input beam based on wavelength (essentially a variable transmission grating) [36], while the LCTF uses oriented liquid crystals to selectively retard
the beam, causing destructive interference for all wavelengths except the desired passband (referred to as a Lyot lter) [37]. There are a variety of other
practical dierences which impact the implementation of each technology; for
16
F. LaPlant
instance, the speed of switching is much greater with the AOTF, although
there are signicant input beam geometry restrictions; the LCTF has lower
throughput, but has few spatial input restrictions. LCTF units typically have
larger apertures with more uniform wavelength response across the aperture
area which have made them the unit of choice for imaging applications [38],
although AOTF units have been used for imaging as well [39]. Despite these
dierences, the ability of the user to either scan between selected wavelengths
or randomly access individual wavelengths is similar, so that they can be
grouped together for the purposes of this discussion.
There are several advantages to a tunable lter system. First, it is unnecessary to have a multichannel detector (for single-point measurements), since
only one wavelength is being selected at a time. The size of the detector is
also much more exible, since spectral resolution of the system is not a function of the detector and input aperture as it is in a classical monochromotor,
but rather limited only by the functional characteristics of the lter. Second,
since focusing and dispersive elements are minimized the spectrometer could
be made very small. Third, the entire spectrum does not need to be obtained;
the random access nature of the lter allows only the spectral features required for a measurement to be made. This can be a signicant advantage for
routine measurements.
Small, robust instruments have been produced based on this technology
[40, 41], and applications in biomedical, biothreat detection, and pharmaceutical analysis have been shown, especially for imaging [42]. However, there are
various disadvantages that have limited their widespread use. The eective
throughput of the tunable lters is not as high as for grating-based systems;
each is sensitive to input polarization, which is not always easy to control.
Because the Raman eect is comparatively weak, it is crucial that every component has the highest throughput possible. The spectral resolution of either
of these systems is not as good as that of a traditional system. Hence tunable lters have had better success in applications in near-IR and uorescence
spectrometers, where the measurement is typically less signal-limited, and
where the spectral features are broad. New generations of materials will no
doubt exhibit improved performance, as increases in throughput [43] and application to deep UV wavelengths [44] have been presented. Further advances
and decreased cost will be necessary before tunable lters will likely compete
favorably with competing technologies for general use.
1.6 Detectors
The charge-coupled device was rst used in Raman spectroscopic applications
in the late 1980s [45, 46], followed rapidly by the introduction of holographic
notch lters [47]. This combination, coupled with the visibility of FT-Raman
instruments introduced at around the same time, helped drive the growth
of Raman outside the academic lab. Although the throughput advantage of
17
Output
Shift Register
Input
Spectrometer
Readout
Amplifier
18
F. LaPlant
240000
200000
160000
400
600
800
1000
1200
1400
1600
somewhat variable depending on the quality of the CCD. The absolute S/N
level, on the other hand, is very dependent on chip material quality and construction. In practice, the target temperature for many spectroscopic CCDs
is around 40 C, although modern multistage Peltier cooling systems can
now routinely cool CCDs down to 80 C. The other principal noise source
is the noise induced in measuring the pixel charge, referred to as read noise.
Although many CCDs read each pixel destructively (that is, the current from
the pixel can only be read once), it is possible to implement measuring techniques that read the charge multiple times, minimizing the read noise contribution. Speed and duty cycle can be improved by adding parallel registers to
the chip; although since the eciency of the detector is a function of the photoactive area of the chip, addition of non-active areas will reduce the number
of detected photons. Micro-lenses have been used to direct light to the active
regions of the chip to improve eciency for these kinds of devices.
There are a wide variety of variables on this basic theme that have been
used to enhance the operation of the CCD; since many techniques now exist
to minimize noise sources, most of the important current contributions are
focused on maximizing the signal through higher quantum eciency of the
detector, or amplifying the signal while keeping the read noise constant. Backthinning is one commonly used technique where the light enters what would
normally be the back side of the chip. Illuminating from the back increases
the eective photoactive area of the pixels since much of the front surface of
the chip is obscured by the transfer electronics. The eciency of a detector
with similar material construction can be improved from 40% for a frontilluminated chip to 80% for a back-thinned model. The detector can also be
made from specially doped silicon (referred to as deep depleted) that enhances
the sensitivity toward the red end of the spectrum. Although chip architecture
19
20
F. LaPlant
CCD
21
Camera
LCTF
Pinhole
Cylindrical
Lens
Point
Illumination
Line
Illumination
Global
Illumination
single spot, the spectral image dispersed onto the CCD is also a spot; there is
no spatial information, and most of the CCD area is left unused. An image of
the sample, if such a thing is desired, can be obtained by rastering the sample
under the laser focus, moving from point to point on the sample until the
desired area has been mapped out. Much has been made of the limitations
and potential pitfalls of this technique [57], although it has a long history of
being used to good eect in a wide variety of applications.
While the point-by-point mapping strategy will certainly work, the signal
throughput is very low and lower than necessary, since only a small fraction
of the CCD is used during any spectral acquisition. By focusing the laser to a
line rather than a spot, the resulting image can be focused on the slit, and subsequently imaged on the CCD. The x-direction on the CCD is still frequency,
while the y-axis is spatial variation in the sample. Alternately this conguration could employ various geometries of laser focus at the sample; since the
bers at the sample can be arranged in any conguration, with subsequent
ber rearrangement into a line at the slit, any conguration at the sample
could be used as long as the ber position can be decoded after collection of
the image spectrum. This highlights the necessity for an aberration corrected
imaging spectrometer/grating. If the image of the slit is distorted, the spectral
features and the image can be convolved with one another, and the resulting image will be impossible to resolve. The throughput of the spectrometer
is now greater than the increased ll factor of the CCD. Unfortunately the
spatial rejection is also compromised somewhat, since a pinhole spatial lter
22
F. LaPlant
cannot be used, although this eect is generally minimal. As with the confocal
imaging, the sample must be rastered in order to build up spectral data for
the areas of interest.
The third conguration illuminates the sample globally, using an LCTF
as the wavelength discrimination mechanism. A snapshot of the entire sample
is obtained at a given wavelength; the spectrum is built up one wavelength
at a time, rather than moving the sample and obtaining one position at a
time. The throughput on such a system could be very high, especially if only
a few wavelengths are required for spectral identication purposes. However,
as the illumination pattern increases, the spatial rejection decreases. With removal of the slit, there is no spatial rejection mechanism at all. For reasonable
images in a system such as this, either the sample needs to be very thin, or
completely opaque. Any subsurface contributions or scattering from photon
migration within the sample will contribute randomly to the collected signal,
smearing the resulting image. In addition, thermal problems are more likely
with wide-eld illumination. Although the ux may be higher in a point system (consider 50 mW in a 1 m diameter spot vs. 2 W in a 2 mm spot) heat
transport becomes an issue. While a spot (or a line) can dissipate heat into
the sample in three dimensions, the globally illuminated sample can only dissipate heat in one dimension into the depth of the sample. This makes it much
more dicult for the sample to remove the excess heat [58].
1.8 Conclusion
Raman spectroscopy has suered as the ugly, expensive, and nicky stepsister
of mid-IR spectroscopy for many years. The principal factor holding it back
from broader acceptance was the diculty and expense of performing the
experiments. It is instructive to note that almost every technological breakthrough was borrowed from another scientic discipline: the laser, gratings,
and spectrometers from physics, Fourier transform and the CCD detector from
the astronomers, and ber optics from telecommunications (holographic notch
lters may be a notable exception). But now that Raman instrumentation has
proved its ability to sustain multiple practical applications, the emerging applications described in this book will now no doubt drive the development of
new technologies and new Raman capabilities currently impossible to imagine.
References
1. C.V. Raman, K.S. Krishnan, Nature 121, 501 (1928)
2. Handbook of Raman Spectroscopy, ed. by I. Lewis, H.G.M. Edwards (Dekker,
New York, 2001)
3. M. Pelletier, Analytical Applications of Raman Spectroscopy (Blackwell, Oxford,
1999)
23
24
F. LaPlant
39. S.R. Goldstein, L.H. Kidder, T.M. Herne, I.W. Levin, E.N. Lewis, J. Microsc.
184, 35 (1996)
40. N. Gupta, R. Dahmani, Spectrochim. Acta A 56, 1453 (2000)
41. K. Chen, M.E. Martin, T. Vo-Dinh, Proc. SPIE, 5993(599307) (2005)
42. F. Yan, T. Vo-Dinh, Sens. Actuators B 121, 66 (2007)
43. X. Wang, T.C. Voigt, P.J. Bos, M.P. Nelson, P.J. Treado, Prog. Biomed. Opt.
Imag. Proc. SPIE 6378(637808) (2006)
44. N.S. Prasad, Int. J. High Speed Electron. Syst. 17, 857 (2007)
45. C.A. Murray, S.B. Dierker, J.D. Legrange, N.E. Schlotter, Chem. Phys. Lett.
137, 453 (1987)
46. J.M. Williamson, R.J. Bowling, R.L. McCreery, Appl. Spec. 43, 372 (1989)
47. M. Carrabba, K. Spencer, C. Rich, D. Rauh, Appl. Spec. 44, 1558 (1990)
48. J.M. Harnly, R.E. Fields, Appl. Spectrosc. 51, 334A (1997)
49. J. Janesick, Scientic charge-coupled devices. SPIE Press PM83 (2001)
50. A. OGrady, Prog.Biomed. Opt. Imag. Proc. SPIE, 6093(60930S) (2006)
51. T. Dieing, O. Hollricher, Vib. Spectrosc. 48, 22 (2008)
52. J.C. Carter, J. Scadi, S. Burnett, B. Vasser, S.K. Sharma, S.M. Angel, Spectrochim. Actat A 61, 2288 (2005)
53. E.V. Efremov, J.B. Buijs, C. Gooijer, F. Ariese, Appl. Spec. 61, 571 (2007)
54. M.A. Greenwood, Photonics Spectra 41, 33 (2007)
55. D. Renker, Nucl. Instrum. Methods Phys. Res., Sect. A 567, 48 (2006)
56. S. Schlucker, M.D. Schaeberle, S.W. Human, I.W. Levin, Anal. Chem. 75, 4312
(2003)
57. N. Everall, Appl. Spec. 62, 591 (2008)
58. D. Zhang, J.D. Hanna, Y. Jiang, D. Ben-Amotz, Appl. Spec. 55, 61 (2001)
http://www.springer.com/978-3-642-02648-5