Beruflich Dokumente
Kultur Dokumente
pg 1 of 15
Table of Contents
1
Dive-Spread Installation
ROV Instalaltion
Cautionary Notes
Calibration
4.1
General
4.2
4.3
Subsea Calibrations
Operation
5.1
Wet Check
5.2
5.3
Recording CP Potentials
6.1
10
6.2
11
6.3
12
6.4
13
6.5
14
6.6
15
6.7
panel (optional)
pg 2 of 15
Fig. 1
pg 3 of 15
A male connector on a 60 whip is provided to facilitate splicing into the ROV junction box.
It is recommended that the two electrode wires are routed in a shielded twisted pair in the
ROV umbilical, the tip wire can be connected to a coaxial cable or any other cable that is
straight through to the control van with no breaks.
Topside the system can be wired directly into a video annotator (insure that it has an input
impedance of at least 10 Meg Ohm / Volt), or onto either a pair of multi-meters or a
Deepwater monitoring panel. See attached wiring schematics for hookup options.
Section 4 Calibration
4.1
General
If probes have been stored for a significant period, and have been allowed to dry out, it
may take 1-2 hours of immersion time before the electrodes stabilize.
The attached sheet shows the procedure; ensure that either native seawater or artificial
seawater (3% by weight salt) is used.
Procedures (forms are included for Cell to Cell, Master Silver/Silver Chloride, and zinc
coupon checks). The zinc calibration coupon is provided as standard, but the master
silver/silver chloride procedure is recommended). NEVER USE A METALLIC CONTAINER
WHEN PERFORMING CALIBRATIONS OR FOR STORAGE.
4.3
Subsea Calibrations
The subsea calibration procedures should be followed and the results logged at least once
per dive.
4.3.1 Cell To Cell Calibrations
This procedure checks one cell in the probe against the other, and should be
performed with all the cable immersed as it will quickly detect any cable damage.
If dual readouts are used as shown in the wiring diagrams, the cell to cell reading is
displayed continuously, this is the recommended procedure.
The cells should read within +/- 5 mV DC.
then one of the following has happened:
pg 5 of 15
1.
2.
3.
If in the middle of a dive it is possible to complete the dive, however the defective
cell should be isolated and the good cell used to finish the dive. The bad cell will
normally be the one, which displays the most positive potential. So read a potential
on the structure with both elements, the most negative cell should be used to
complete the dive, and the offset noted.
When recovering the umbilical cable from such a dive, monitor the cell-to-cell
value, if the problem is lead wire damage, the value will revert to within +/- 5mV
as soon as the damaged area clears the water, in this way the damage can be
quickly located and repaired.
4.3.2 Ground Check
This procedure verifies that the platform ground wire and the probe tip will yield the
same reading at any point on the main structure. This is particularly important
when the probe is being used in proximity mode. The POLASCAN READOUT has a
facility to switch between tip and platform ground. To verify the ground, stab the
steel structure under survey with the probe and record the value switching grounds
from tip to platform, the reading should be identical, if it varies by more than +/ 3
mVDC, check the platform ground connection.
Section 5 Operation
Most oil companies will have their own survey procedures that should be followed.
following guidelines are given as guidance only.
The
Wet Check
When interfaced to an ROV, it important to perform a wet check before every dive. Signal
errors caused by umbilical power cable interference could introduce reading errors.
With the probe wire to the ROV, immerse the system to at least (-) 30 feet, and with the
vehicle still in its TMS bring up one electrical system at a time and note the potential of
the tip (it will vary some depending on immersion time). If there is a problem, the
potential will change rapidly or will spike when the circuit is powered. Repeat for all
electrical and hydraulic systems. If a problem is detected it should be fixed, usually its a
poor ground connection or a faulty leaking underwater connector that cause the problem.
If the system at fault can be switched off without hindering operation, this is an acceptable
option.
pg 6 of 15
5.2
Changing Electrodes - Spare plug-in electrode elements can be provided with the ROV II
probe, these elements will also work in the CP Gun.
If the elements or one element needs to be changed, follow this procedure:
1.
Take the probe into a clean inside working area, with a clean grease free work
surface.
2.
3.
Remove the head assembly carefully by removing two screws on back of probe, pull
out gently until two electrode connectors are visible.
4.
Using an ohm meter measure between the connector pin and the electrode element
to verify which is which. (Do not handle electrode elements).
5.
6.
Apply a small smear of silicone grease into the open female connector and insert
the new electrode. (Avoid using excess grease, and any grease contact with
electrode element.)
7.
8.
Carefully re-insert the electrodes into the probe body and replace the head
assembly.
5.3
Recording CP Potentials
If wired correctly the CP potentials should have a negative polarity. Accurate readings will
be steady and not drifting so long as the probe is not moving. If readings are erratic this
is a sign of a connection problem, and the cause should be isolated and fixed.
To obtain accurate readings the probe apertures should be within 2 of the structure
surface. The further away the probe is held from the structure area under investigation,
the less accurate will be the reading.
On offshore structures protected with zinc or aluminum anodes the following potential
ranges should be observed.
(-) 0.750 - (-) 1.100 Volts
On Anodes
pg 7 of 15
Page 09
Page 10
Page 11
Page 12
Page 13
Page 14
Page 15
pg 8 of 15
pg 9 of 15
pg 10 of 15
pg 11 of 15
pg 12 of 15
pg 13 of 15
pg 14 of 15
pg 15 of 15