Beruflich Dokumente
Kultur Dokumente
Medical Dosimetry
journal homepage: www.meddos.org
A R T I C L E I N F O
A B S T R A C T
Article history:
Received 2 January 2014
Received in revised form
6 March 2014
Accepted 10 April 2014
The purpose of this study was to perform delivery quality assurance with ArcCHECK and 3DVH system
(Sun Nuclear, FL) and to evaluate the suitability of this system for volumetric-modulated arc therapy
(VMAT) (RapidArc [RA]) verication. This software calculates the delivered dose distributions in patients
by perturbing the calculated dose using errors detected in uence or planar dose measurements. The
device is tested to correlate the gamma passing rate (%GP) and the composite dose predicted by 3DVH
software. A total of 28 patients with prostate cancer who were treated with RA were analyzed. RA
treatments were delivered to a diode array phantom (ArcCHECK), which was used to create a planned
dose perturbation (PDP) le. The 3DVH analysis used the dose differences derived from comparing the
measured dose with the treatment planning system (TPS)-calculated doses to perturb the initial TPScalculated dose. The 3DVH then overlays the resultant dose on the patient's structures using the resultant
PDP beams. Measured dose distributions were compared with the calculated ones using the gamma
index (GI) method by applying the global (Van Dyk) normalization and acceptance criteria, i.e., 3%/3 mm.
Paired differences tests were used to estimate statistical signicance of the differences between the
composite dose calculated using 3DVH and %GP. Also, statistical correlation by means of logistic
regression analysis has been analyzed. Dose-volume histogram (DVH) analysis for patient plans revealed
small differences between treatment plan calculations and 3DVH results for organ at risk (OAR), whereas
planning target volume (PTV) of the measured plan was systematically higher than that predicted by the
TPS. The t-test results between the planned and the estimated DVH values showed that mean values
were incomparable (p o 0.05). The quality assurance (QA) gamma analysis 3%/3 mm showed that in all
cases there were only weak-to-moderate correlations (Pearson r: 0.12 to 0.74). Moreover, clinically
relevant differences increased with increasing QA passing rate, indicating that some of the largest dose
differences occurred in the cases of high QA passing rates, which may be called false negatives. The
clinical importance of any disagreement between the measured and the calculated dose is often difcult
to interpret; however, beam errors (either in delivery or in TPS calculation) can affect the effectiveness of
the patient dose. Further research is needed to determinate the role of a PDP-type algorithm to
accurately estimate patient dose effect.
& 2014 American Association of Medical Dosimetrists.
Keywords:
3DVH
Quality assurance
ArcCHECK
Gamma index
Introduction
RapidArc (RA) (Varian Medical Systems, CA) is a form of
volumetric-modulated arc therapy (VMAT) that entered clinical
use in 2008. VMAT delivery requires a stricter mechanical
277
Fig. 1. Sagittal, axial, and coronal dose planes for a representative data set of the 28 data sets studied. (Color version of gure is available online.)
criteria were reported as those most commonly used by clinicians11 in per patient planar IMRT QA.
It was shown in a recent study that per-beam planar %GPs do
not predict the clinical effect on the patient in terms of changes in
dose-volume histogram (DVH) values for the clinical target volume
(CTV) and organs at risk (OARs).12
Indeed there have been several software systems that claim the
capability to estimate patient dose based on QA measurement,
such as the COMPASS system (IBA-Wellhofer, Bahnhofstrasse
590592 Schwarzenbruck, Germany), Dosimetry Check (Math Resolutions, LLC, Gales Lane, Columbia), and 3DVH (Sun Nuclear
Corporation, Melbourne, Florida). The potential of using such
products to perform patient DVHbased IMRT QA is in need of
further investigation.
In this study, we used the software program 3DVH with the
ArcCHECK to investigate the correlation among %GP obtained
during standard per-beam pretreatment QA tests.
The 3DVH software purports to be able to take beam measurements from a plan and reconstruct the full 3D dose distribution
within the patient for comparison with the TPS calculations.
The 3DVH system was designed to incorporate the beam-bybeam phantom doses (measured and calculated) back into the
patient's images, structures, and TPS dose using planned dose
Fig. 2. Sample differences between the DVH obtained by TPS and the composite dose predicted by the 3DVH software. (Color version of gure is available online.)
278
Table 3
The DVH parameters calculated by the TPS Eclipse and by the software 3DVH for
PTVs and OARs and the relative p value
Patient
Mean %GP
1
2
3
4
5
6
7
98.1
99.0
98.5
98.0
98.4
97.6
97.0
Structure
Parameter
%DD
p value
PTV
PTV
Rectum
Bladder
Penile bulb
Femoral head dx
Femoral head sx
Dmean
D95%
Dmean
Dmean
Dmean
Dmean
Dmean
2.8
2.2
0.5
1.2
6
0.6
2.5
0.1
0.02
0.04
0.03
0.1
0.003
0.03
o 0.05
0.26
0.55
o 0.05
o 0.05
o 0.05
o 0.05
Methods
All patients with prostate cancer underwent radical radiotherapy up to 80 Gy to
the prostate and seminal vesicles with standard fractionation (2 Gy/die). Patients
were immobilized by customized devices with an empty rectum and a moderately
full bladder. Every day treatment was veried by cone-beam computed tomography scans. Gross tumor volume included the prostate and seminal vesicles
(proximal thirds), CTV was dened as gross tumor volume 3-mm circumferential
margin, and planning target volume (PTV) was dened as CTV 5-mm circumferential margin. The whole PTV received 95% to 105% of the prescribed dose. The
rectum, bladder, penile bulb, and femoral heads were set as OARs, and dose
constraints have been dened according to the quantitative analyses of normal
tissue effects in the clinic criteria.14
RA technique
Treatment planning was performed using photon beams of TrueBeam, with
nominal energies of 6 MV. RA plans were optimized for 2 arcs (rotation of 3581, from
1791 to 1811 clockwise (CW) and counterclockwise (CCW)), where multileaf collimator (maximum: 5 mm/deg and 2.5 cm/s), dose rate (maximum: 600 MU/min), and
gantry speed (maximum: 72 s/turn, i.e., 5 deg/s) are optimized simultaneously to
achieve the desired degree of modulation. All dose distributions were computed with
the anisotropic analytical algorithm implemented in the Eclipse planning system
with a calculation grid resolution of 2.5 mm.
A total of 28 DMLC prostate RA treatment plans were exported as DICOM
radiation therapy (RT) data sets from the planning system. These plans were
delivered to a 3D diode array (ArcCHECK, Sun Nuclear). Measured doses were
compared with the dose planes from the original plan, and dose differences were
used as input to 3DVH to calculate a delivered dose in the patient.
GI evaluation
The quality assurance on RA treatment plans is performed by creating a
verication plan in the Eclipse TPS. This was achieved by copying the treatment
plan onto volumetric phantom devices. The predicted dose to the detector planes
was calculated with a dose grid resolution of 2.5 mm, it is sufcient to keep the
dose error within 2%.15 This value was adopted for RA treatment plans.
Pretreatment verications were performed for all patient plans by acquiring
plane dose distributions of each treatment eld. Measurements were taken using
the diode array, routinely used in our institute, with absolute dose calibration, and
the software ArcCHECK.
Table 2
Evaluation of the %DD mean for 28 DVHs
Structure
Parameter
%DD
PTV
PTV
Rectum
Bladder
Penile bulb
Femoral head dx
Femoral head sx
Dmean
D95%
Dmean
Dmean
Dmean
Dmean
Dmean
0.07
0.02
0.04
0.03
0.10
0.01
0.03
2.8
2.2
0.5
1.2
6
0.6
2.5
Results
Planar dose distributions acquired during pretreatment verications, RT plan, structure set, dose, and computed tomography
images (the latter is optional) exported from TPS were loaded on
the software 3DVH. Overall, 2 kinds of comparisons were made,
the rst was a 2D/3D gamma evaluation, and the second was a
DVHs comparison.
DVHs provided by the 3DVH (delivered dose distribution from
pretreatment verications) were compared with those calculated
by Eclipse (planned dose distribution) using the following dosevolume parameters: Dmean for all structures, and Dmean and D95
for PTVs.
Differences in the DVH parameters were calculated as
D
3DVH DTPS
%DD
100
DTPS
where D3DVH was the dose value shown by the software 3DVH, and
DTPS was the dose value reported by TPS. This analysis was
performed, for each patient, on PTVs and some OARs of the
anatomically treated district (the rectum, bladder, penile bulb,
femoral head dx [right], and femoral head sx [left]).
We calculated mean DVH values obtained using both 3DVH and
TPS. To verify whether our procedures were affected by systematic
errors, we performed a t-test statistical analysis between the mean
values.
279
Fig. 3. Correlation between the %GP calculated using the global 3%/3-mm criterion and the %DD for Dmean of the PTV, PTV 95%, rectum, bladder, penile bulb, femoral dx, and
femoral sx. (Color version of gure is available online.)
dened as the mean of differences between D3DVH and DTPS for all
organs.
In addition, we determined the degree to which the DVH values
are associated using the Pearson correlation coefcient (r). This
correlation test is used to measure the strength of a linear
association between 2 variables. The Pearson correlation coefcient, r, can take a range of values from 1 to 1. A value
of 0 indicates that there is no association between the 2 variables.
280
Table 4
Indexes of correlation between %GP and %DD for different structures
Structure
Parameter
PTV
PTV
Rectum
Bladder
Penile bulb
Femoral head dx
Femoral head sx
Dmean
D95%
Dmean
Dmean
Dmean
Dmean
Dmean
0.57
0.12
0.74
0.38
0.69
0.37
0.36
dx = right; sx = left.
Conclusions
The aim of this article was to evaluate the predictive meaning
of the GI, in terms of correlation between the %GP obtained
during standard pretreatment QA tests and the dose discrepancy
between planned DVH and patients' perturbed DVH. This article
evaluated whether the standard action levels used by most clinics
(3%/3 mm with a 95% passing rate and global normalization) are
justied.
The 3DVH is a useful tool to calculate delivered dose distributions during patient-specic IMRT and VMAT quality assurance. The intended use is to incorporate the ArcCHECK (or
electronic portal imaging device or MapCHECK) measurements
and patient's DICOM RT structure set, dose, and plan les to
reconstruct the 3D dose within the patient and to allow the
computation and comparison of the DVH curves, planned and
delivered, for the anatomical structures that were of interest
during the planning of the patient's treatment, rendering
phantom-based dose analysis obsolete. The process requires no
extra time when compared with standard QA with the
ArcCHECK.
Discussion
The 3DVH validity was demonstrated in previous
articles.8,10,12,20,21 In our study, we started by verifying the consistency of 3DVH with our clinical IMRT plans.
The patient dose QA methodology is used to accurately estimate the effect of errors on patient anatomy dose metrics.
There is a lack of correlation between conventional IMRT QA
performance metrics (%GPs) and dose differences in critical anatomical regions of interest. The most common acceptance criteria
and published actions levels therefore have insufcient, or at least
unproven, predictive power for per patient IMRT QA. However,
these criteria does not ensure that clinically acceptable dose errors.
There have not yet been correlation studies to prove, or
disprove, whether these accepted methods for IMRT QA and their
associated acceptance criteria are good predictors of clinically
relevant patient dose errors in per patient IMRT QA.
Recent experimental studies, which reveal the limited sensitivity of gamma analysis to patient dose deviation under different
IMRT errors, have been carried out.16,18,19,24 In fact, it was shown in
a recent study that per-beam planar %GPs do not predict the
clinical effect on the patient in terms of the changes in DVH values
for the CTV and OARs,12 which has questioned the feasibility of %
GP-based IMRT QA.
The gamma analysis did not give clear information on whether a
plan should be accepted. Kruse18 has recently shown that the
average passing rate that would be clinically acceptable for one
patient could be unacceptable for another. Consequently, pretreatment checks should be done taking into account the clinical
tolerances of the OAR and the PTV and not relying on %GPs. To do
this, there is a need for a tool, such as 3DVH software, that could
show the effect of the measured planar differences in the histogram.
References
1. Vanetti, E.; Clivio, A.; Nicolini, G.; et al. Volumetric modulated arc radiotherapy
for carcinomas of the oro-pharynx, hypo-pharynx and larynx: A treatment
planning comparison with xed eld IMRT. Radiother. Oncol. 92:1117; 2009.
2. Verbakel, W.F.; Cuijpers, J.P.; Hoffmans, D.; et al. Volumetric intensitymodulated arc therapy vs. Conventional IMRT in head-and-neck cancer:
A comparative planning and dosimetric study. Int. J. Radiat. Oncol. Biol. Phys.
74:2529; 2009.
3. Bedford, J.L.; Warrington, A.P. Commissioning of volumetric modulated arc
therapy (VMAT). Int. J. Radiat. Oncol. Biol. Phys. 73(2):537e45; 2009;
Otto, K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med.
Phys. 35(1):3107; 2008.
4. Guckenberger, M.; Richter, A.; Krieger, T.; et al. Is a single arc sufcient in
volumetric-modulated arc therapy (VMAT) for complex-shaped target volumes? Radiother. Oncol. 93(2):259e65; 2009.
5. Both, S.; Alecu, I.M.; Stan, A.R.; et al. A study to establish reasonable action
limits for patient-specic quality assurance in intensity-modulated radiation
therapy. J. Appl. Clin. Med. Phys. 2:18; 2007.
6. Basran, S.; Woo, M.K. An analysis of tolerance levels in IMRT quality assurance
procedures. Med. Phys. 35(6):23007; 2008.
7. Howell, M.; Smith, I.P.; Jarrio, C.S. Establishing action levels for EPID-based QA
for IMRT. J. Appl. Clin. Med. Phys. 9(3):1625; 2008.
8. Ezzell, A.; Burmeister, J.W.; Dogan, N; et al. IMRT commissioning: Multiple
institution planning and dosimetry comparisons, a report from AAPM Task
Group 119. Med. Phys. 36(11):535973; 2009.
281
17. Nelms, B.E.; Zhen, H.; Tom, W.A. Conventional IMRT QA analysis metrics
do not predict clinically relevant patient dose errors. Med. Phys. 38:1037;
2011.
18. Kruse, J.J. On the insensitivity of single eld planar dosimetry to IMRT
accuracies. Med. Phys. 37(6):251625; 2010.
19. Yan, G.; Liu, C.; Simon, T.A.; et al. On the sensitivity of patient-specic IMRT QA
to MLC positioning errors. J. Appl. Clin. Med. Phys. 10:2915; 2009.
20. Zhen, H.; Nelms, B.E.; Tom, W.A. Moving from gamma passing rates to patient
DVH-based QA metrics in pretreatment dose QA. Med. Phys. 38(10):547789; 2011.
21. Olch, A.J. Evaluation of the accuracy of 3DVH software estimates of dose to virtual
ion chamber and lm in composite IMRT QA. Med. Phys. 39(1):816; 2012.
22. Both, S.; Alecu, I.M.; Stan, A.R.; et al. A study to establish reasonable action
limits for patient specic quality assurance in intensity-modulated radiation
therapy. J. Appl. Clin. Med. Phys. 8(2):18; 2007.
23. Dempsey, J.; Romeijn, H.; Li, J.; et al. A Fourier analysis of the dose grid
resolution required for accurate IMRT. Med. Phys. 32:3808; 2005.
24. Feygelman, V.; Zhang, G.; Stevens, C.; et al. Evaluation of a new dose QA device,
or: The X's and O's of 3D dosimetry arrays. J. Appl. Clin. Med. Phys. 12:334668;
2011.