Beruflich Dokumente
Kultur Dokumente
Chamber along
with the
detector
Pre-Amplified
Output Signal of Detector
12 V DC
Voltage
Cathode
Ray
Oscillosco
Spectroscopic
Amplifier
Cathode
Ray
Oscillosco
Multi
Channel
Analyzer
12V DC
POWER
VACUUM
CHAMBER
DETECT
0-60V DC
PRE-AMP
POWER
SPECTROSC
OPIC
AMPLIFIER
MULTI
CHANNEL
ANALYZER
CATHODE
RAY
CATHODE
RAY
Problems
Power Supply
We have encountered few problems regarding the power supply of the
detector ,In the specification of the detector it was given that the detector
power supply should be with in 6-12V DC .So as described we applied 6V DC
through a prebuilt variable power supply but the pre-amplified signal from
the detector was very poor as a result we started increasing the power
supply gradually to 9V then 12V.After increasing the power supply the preamplified signal quality has increased but still was not convincing. The Preamplified Signal after application of 12V DC supply from detector with Cs-137
as source has been given below
After this failure we decided to apply 9V power supply available in the NIM
BIN , but still the pre-amplified signal was not what we are expecting. Then
we started analyzing the problem and found that both the power supply that
we were using have some noise and some AC ripples as result the charge
collection in the detector was very poor and that is why we are not getting
desired pre-amplified signal from the detector .
After analyzing the problem we came to the conclusion that we need a pure
DC Voltage supply and we decided to use a 9V battery as a constant DC
power source. And the result after that drastically improved . The Preamplified Signal after application of 9V DC supply from the battery with Cs137 as source has been given below
Capacitor Problem
One of the major problems we encountered during this testing was related to
selection of the capacitor which has to be connected in series with the output
signal of the detector to eliminate the offset. In the specification sheet It was
advised to use dielectric type capacitor in the range of 1 microfarad to 100
Spectroscopy Amplifier
The output of the detector that is the pre-amplified signal of the detector
has to be amplified before it is feed to the Multi Channel Analyzer because
the output of the detector is very low and cannot be recognized by the Multi
Channel Analyzer. The amplifier we are using here is ORTEC model 672 .
pulse. This minimizes the operator skill and effort needed to achieve good
energy resolution and peak position stability at moderate to high counting
rates.When changing time constants or detectors, an accurate pole-zero (PZ)
adjustment is achieved by simply pushing the AUTO PZ button and waiting a
few seconds for the AUTO PZ BUSY LED to turn off. No oscilloscope is
required for this procedure. The AUTO PZ memory is protected against power
failures.
Choice of triangular and Gaussian filters effectively doubles the time
constants available for optimum resolution .Automatic noise discriminators
on both the pile-up rejecter and the
Baseline restorer eliminates all screwdriver adjustments. Automatic baseline
restorer rate for superior performance at both low and high counting rates
.Differential input for reduction of ground loop noise. Automatically
compensates for reset recovery with transistor-reset
Preamplifiers.
Settings we used during this experiment for this detector
FINE GAIN :
COARSE GAIN:
UNIPOLAR PULSE SHAPE: GAUSSIAN
SHAPING TIME:
POLE ZERO ADJUSTMENT: AUTO
INPUT:
BAL(Differential Input Gain Balance):
While we are testing the detector for Gamma radiation with Cesium-137 as
a source it was working properly after rectifying the earlier problem, but
when we started testing the detector for Alpha radiation with Amnesium-241
as a source we were getting very low count rate for the source though we
have taken all the necessary pre-caution like keeping the source and the
detector in a vacuum chamber so that the alpha particles from the source
could arrive at the detector surface and do not get attenuated in the air
because Alpha particles gets attenuated in the air with in few millimeter. And
even we kept the detector in dark room so that no light arrives at the
detector surface to provide proper light insulation.
After all this pre-cautionary measures the count rate was still very low
compared to the background noise. Then we observed that the front side of
the detector is having a black thin plastic cover but we were still not
convinced whether to peel off this plastic cover or not because in the
specification sheet it was mentioned that the detector had a 1-2 Micron
Aluminized Mylar to provide the detector light insulation and that black thin
cover could have been the Mylar. If we peel off that cover that cover then
detector could get damaged .
So we contacted the detector company to know what to do in this matter
and then they informed that the black plastic cover was there to protect the
Mylar while it was getting shipped from the company and we can peel off
that cover if it is necessary for the testing. After that we came to the
conclusion that the plastic cover was blocking most of the Alpha particle
thats why we were getting very low count rate .
As advised by the detector company after peeling off the plastic cover the
count rate increased drastically. The picture of the detector before and after
the peeling off is given below: