Beruflich Dokumente
Kultur Dokumente
4. The Maxwells bridge shown in the figure is at balance. The parameters of the
inductive coil are.
(A) R=R2R3/R4,L=C4R2R3
(B) L=R2R3/R4,R=C4R2R3
(C) R=R4/R2R3,L=1/(C4R2R3)
(D) L=R4/R2R3,R=1/(C4R2R3)
5. Two 8-bit ADCs, one of single slope integrating type and other of successive
approximate type, take TA and TB times to convert 5 V analog input signal to
equivalent digital output. If the input analog signal is reduced to 2.5 V, the
approximate time taken by the two ADCs will respectively, be
(A) TA,TB (B) TA/2,TB
(C) TA,TB/2 (D) TA/2,TB/2
Sol:Conversion time does not depend on input voltage so it remains same
for both type of ADCs.
Hence (A) is correct option.
6. Two sinusoidal signals p(1,t)=Asin1t and q(2t) are applied to X and Y inputs
of a dual channel CRO. The Lissajous figure displayed on the screen shown
below : The signal q(2t) will be represented as
(A) q(2t)=Asin2t,2=21
(B) q(2t)=Asin2t,2=1/2
(C) q(2t)=Acos2t,2=21
(D) q(2t)=Acos2t,2=1/2
Sol:-
7. The ac bridge shown in the figure is used to measure the impedance Z . If the
bridge is balanced for oscillator frequency f = 2 kHz, then the impedance Z will be
(B) (0 + j200)
(D) (260 + j200)
(A) A, B, C, A
(C) C, B, A, B
(B) A, B, C, B
(D) B, A, B, C
SOL:Since both the waveform appeared across resistor and inductor are
same so the common point is B. Signal Probe S1 is connecte with A,
S2 is connected with C and both the grount probes G1 and G2 are
connected with common point B.
Hence (B) is correct option.
9. A bridge circuit is shown in the figure below. Which one of the sequence given
below is most suitable for balancing the bridge ?
10. The time/div and voltage/div axes of an oscilloscope have been erased. A
student connects a 1 kHz, 5 V p-p square wave calibration pulse to channel-1 of
the scope and observes the screen to be as shown in the upper trace of the
figure. An unknown signal is connected to channel-2(lower trace) of the scope. It
the time/div and V/div on both channels are the same, the amplitude (p-p) and
period of the unknown signal are respectively
(A) 5 V, 1 ms
(C) 7.5 V, 2 ms
SOL:-
(B) 5 V, 2 ms
(D) 10 V, 1 ms
11. A sampling wattmeter (that computes power from simultaneously sampled values
of voltage and current) is used to measure the average power of a load. The
peak to peak voltage of the square wave is 10 V and the current is a triangular
wave of 5 A p-p as shown in the figure. The period is 20 ms. The reading in W
will be
(A) 0 W
(C) 50 W
SOL:-
(B) 25 W
(D) 100 W
14. The simultaneous application of signals x(t) and y(t) to the horizontal and vertical
plates, respectively, of an oscilloscope, produces a vertical figure-of-8 display. If
P and Q are constants and x(t)=Psin(4t+30c) , then y(t) is equal to
(A) Qsin(4t 30c)
(B) Qsin(2t + 15c)
(C) Qsin(8t + 60c)
(D) Qsin(4t + 30c)
SOL:-
15. A DC ammeter has a resistance of 0.1 and its current range is 0-100A. If the
range is to be extended to 0-500 A, then meter required the following shunt
resistance
(A) 0.010
(B) 0.011
(C) 0.025
(D) 1.0
SOL:-
16. The set-up in the figure is used to measure resistance R .The ammeter and
voltmeter resistances are 0.01 and 2000 , respectively. Their readings are 2 A
and 180 V, respectively, giving a measured resistances of 90 The percentage
error in the measurement is
(A) 2.25%
(C) 4.5%
(B) 2.35%
(D) 4.71%
17. A 1000 V DC supply has two 1-core cables as its positive and negative leads :
their insulation resistances to earth are 4 M and 6 M, respectively, as shown
in the figure. A voltmeter with resistance 50 k is used to measure the insulation
of the cable. When connected between the positive core and earth, then
voltmeter reads
(A) 8 V
(C) 24 V
(B) 16 V
(D) 40 V
18. Two wattmeters, which are connected to measure the total power on a threephase system supplying a balanced load, read 10.5 kW and 2.5 kW,
respectively. The total power and the power factor, respectively, are
(A) 13.0 kW, 0.334
(B) 13.0 kW, 0.684
(C) 8.0 kW, 0.52
(D) 8.0 kW, 0.334
19. A dc potentiometer is designed to measure up to about 2 V with a slide wire of
800 mm. A standard cell of emf 1.18 V obtains balance at 600 mm. A test cell is
seen to obtain balance at 680 mm. The emf of the test cell is
(A) 1.00 V
(B) 1.34 V
(C) 1.50 V
(D) 1.70 V
20. The circuit in figure is used to measure the power consumed by the load. The
current coil and the voltage coil of the wattmeter have 0.02 and 1000
resistances respectively. The measured power compared to the load power will
be
.
22. A CRO probe has an impedance of 500 k in parallel with a capacitance of 10
pF. The probe is used to measure the voltage between P and Q as shown in
figure. The measured voltage will be
(A) 3.53 V
(C) 4.54 V
(B) 4.37 V
(D) 5.00 V
23. A moving coil of a meter has 100 turns, and a length and depth of 10 mm and 20
mm respectively. It is positioned in a uniform radial flux density of 200 mT. The
coil carries a current of 50 mA. The torque on the coil is
(A) 200 Nm
(B) 100 Nm
(C) 2 Nm
(D) 1 Nm
24. A dc A-h meter is rated for 15 A, 250 V. The meter constant is 14.4 A-sec/rev.
The meter constant at rated voltage may be expressed as
(A) 3750 rev/kWh
(B) 3600 rev/kWh
(C) 1000 rev/kWh
(D) 960 rev/kWh
25. A moving iron ammeter produces a full scale torque of 240 Nm with a deflection
of 120c at a current of 10 A . The rate of change of self induction (H/radian) of
the instrument at full scale is
(A) 2.0 H/radian
(B) 4.8 H/radian
(C) 12.0 H/radian
(D) 114.6 H/radian
26. A single-phase load is connected between R and Y terminals of a 415 V,
symmetrical, 3-phase, 4-wire system with phase sequence RYB. A wattmeter is
connected in the system as shown in figure. The power factor of the load is 0.8
lagging. The wattmeter will read
(A) 795 W
(C) +597 W
(B) 597 W
(D) +795 W
27. The simplified block diagram of a 10-bit A/D converter of dual slope integrator
type is shown in figure. The 10-bit counter at the output is clocked by a 1 MHz
clock. Assuming negligible timing overhead for the control logic, the maximum
frequency of the analog signal that can be converted using this A/D converter is
approximately
(A) 2 kHz
(C) 500 Hz
(B) 1 kHz
(D) 250 Hz
28. The items in Group-I represent the various types of measurements to be made
with a reasonable accuracy using a suitable bridge. The items in Group-II
represent the various bridges available for this purpose. Select the correct choice
of the item in Group-II for the corresponding item in Group-I from the following
List-I
List-II
P. Resistance in the milli1. Wheatstone Bridge
ohm range
Q. Low values of Capacitance
2. Kelvin Double Bridge
R. Comparison of resistance
3. Schering Bridge
which are nearly equal
S. Inductance of a coil with
4. Wiens Bridge
a large time-constant
5. Hays Bridge
6. Carey-Foster Bridge
Codes :
(A) P=2, Q=3, R=6, S=5
(B) P=2, Q=6, R=4, S=5
(C) P=2, Q= 3, R=5, S=4
(D) P=1, Q=3, R=2, S=6
29. A rectifier type ac voltmeter of a series resistance Rs , an ideal fullwave rectifier
bridge and a PMMC instrument as shown in figure. The internal. resistance of the
instrument is 100 and a full scale deflection is produced by a dc current of 1
mA. The value of Rs
required to obtain full scale deflection with an ac voltage of 100 V (rms) applied to the
input terminals is
(A) 63.56
(C) 89.93
(B) 69.93
(D) 141.3 k
30. A wattmeter reads 400 W when its current coil is connected in the R-phase and
its pressure coil is connected between this phase and the neutral of a
symmetrical 3-phase system supplying a balanced star connected 0.8 p.f.
inductive load. This phase sequence is RYB. What will be the reading of this
wattmeter if its pressure coil alone is reconnected between the B and Y phases,
all other connections remaining as before ?
(A) 400.0
(B) 519.6
(C) 300.0
(D) 692.8
34. Group-II represents the figures obtained on a CRO screen when thevoltage
signals Vx=Vxmsint and Vy=Vymsin(t+) are given to its X and Y plates
respectively and is changed. Choose the correct value of from Group-I to
match with the corresponding figure of
Group-II.
Group-I
P. = 0
Q. = /2
R. <<3/2
S. =3/2
Codes :
38. A 4 digit DMM has the error specification as 0.2% of reading+ 10 counts. If a DC voltage of
100V is read on its 200V full scale. The maximum error that can be expected in the reading is
(a) 0.1%
(b) 0.2%
(c) 0.3%
(d) 0.4%
SOL:- 0.3%
39. An average reading digital multimeter reads 10V when fed with a triangular wave, symmetric
about the time axis. For the same input, an rms reading meter will read
(a)20/3
(b) 10/3
(c) 203
(d) 103
Sol: for triangular wave,
Average value = Vm/3
Rms = Vm/3
Vmm = 30
Rms = 30/3 = 103
40. For a step input ei, the overshoot in the output eo will be
(a) Step Function
(b)square wave of frequency 50 Hz
(c) square wave of frequency 300 Hz (c) square wave of frequency 2.0 Hz
Sol: Step function
41. An energy meter connected to an immersion heater (resistive) operating on an AC 230V,50Hz,
AC single-phase source reads 2.3 unit (kWh) in 1h. the heater is removed from the supply and
now connected to a 400V peak to peak square wave source of 150Hz. The power in kW
dissipated by the heater will be
(a) 3.478
(b) 1.739
(c) 1.540
(d) 0.870
Sol:
Reading of energy meter
2.3 =V2/R *e = (230)2/R *1
R = 23
Power dissipated = V2/R = (200)2/23
= 1.739 kW
42. An ammeter has a current range of 0-5A and its internal resistance of 0.2. in order to change the
range of 0-25 A, we need to add a resistance of
(a) 0.8 in series with the ammeter
(b) 1.0 in series with the ammeter
(c) 0.04 in parallel with the ammeter
(d) 0.05 in parallel with the ammeter
Sol:
Multiplying power
m = I/Im = 25/5 = 5
Resistance of shunt
Rsh = Rm/m-1 = 0.2/5-1 = 0.05
43. A PMMC voltmeter is connected across a series combination of a dc voltage source V1 = 2V and
an AC voltage source V2(t) = 3sin(4t). the voltmeter reads
(a) 2V
(b) 5V
(c)(2+3/2)V
(d) (17/2)V
Sol:
The permanent magnet moving coil measures the DC value only. The torque will be reversed
if the current is reversed. If the instrument is connected to AC, the deflection responds to the
mean torque which is zero. Hence, voltmeter reads 2V.
44. The effect of stray magnetic fields on the actuating torque of a portable instrument is maximum
when the operating field of the instrument and the stray fields are
(a) Perpendicular
(b) parallel
(c) inclined at 600
(d) inclined at 300
Sol:
Actuating torque cos (angle between operating field and magnetic field)
For maximum actuating torque both are parallel
45. A current of -8+62(si nwt+300) is passed through three meters. They are a centre zero PMMC
meter, a true rms meter and a moving iron instrument. The respective readings (in amphere) will
be
(a) 8,6.10
(b) 8,6,8
(c) -8,10,10
(d) -8,2,2
Sol:
PMMC reads only DC value and since, it is centre zero type so, it will give -8 value. RMS
meter will read rms value of the current
So, RMS = (82 +(62/2)) = 10A
Moving iron also read rms value, so its reading will also be 10A, hence ultimately it will be
(-8, 10, 10)
46. A variable w is related to three other variables x, y, z as w = xy/z. the variables are measured with
meters of accuracy 0.5% reading, 1% of full scale value and 1.5 reading. The actual readings
of the three meters are 80, 20 and 50 with 100 being the full scale value for all three. The
maximum uncertainty in the measurement of w will be
(a) 0.5%
(b) 5.5%
(c) 6.7%
(d) 7.0%
Sol:
W=xy/z
Relative limiting error of w,
w/w = (w/x + w/y + w/z)
= (0.4/80 + 1/20 + 0.75/50 )
= 7% reading
47.The figure shows a three-phase delta connected load supplied from a 400V,50Hz,3-phase
balanced source. The pressure coil(PC) and current coil(CC) of a wattmeter are connected to
the load as shown,with the coil polarities suitably selected to ensure a positive deflection. The
wattmeter reading will be
A. 0
B. 1600Watt
C. 800Watt
D. 400Watt
Ans: C
48. An analog voltmeter uses external multiplier settings. With a multiplier setting of 20k, it reads
440V and-with a multiplier setting of 80k it reads 352V. For a multiplier setting of 40k,voltmeter
reads
A. 371V
B. 383V
C. 394V
D. 406V
Ans: D
49.the measurement of Hall coefficient of a semiconductor with one type of charge carrier gives the
information about
A. sign of charge carrier
B. density of charge carrier
C. both sign and density of charge carrier
D. mass of the charge carrier
Ans:C
50. The measurement system shown in the figure uses three sub systems in cascade whose gains are
specified as G1, G2 and 1/G3. The relative small errors associated with each respective subsystem
G1, G2,G3 are 1,2 and 3.The error associated with the output is
A.1+2+1/3
B.(1.2)3
C.1+2-3
D.1+2+3
Ans:C
51.
Ans. D
Sol:
(b) 1, 3 and 4
(c) 1 and 3
(d) 2 and 4
Sol: - Digital data acquisition systems can be used with analog output as well as digital output
Of the transducers. Analog data can be converted into digital data through A/D converter.
Further, Digital data acquisition systems are highly accurate due to better noise immunity.
So, (d) is the correct option.
53. Which one of the following statements correctly represents the post acceleration in a Cathode Ray
Tube?
(a) It provides deflection of the beam
(b) It increases the brightness of the trace if the signal frequency is higher than 10 MHz.
(c) It accelerates the beam before deflection
(d) It increases the brightness of the trace of low frequency signal.
Sol:- In an electrostatic system of deflection, a low value of accelerating voltage (Ea) should be used to
give high deflection sensitivity (S1/ Ea). This voltage is usually kept below 4kV. Although
This voltage gives good sensitivity but reduces brightness, which can be seriously impaired at high
frequencies. If, signals of frequencies higher than 10 MHz are to be displayed, post deflection
acceleration is necessary to increase the brightness of the trace which otherwise would be dim.
Hence, option (b) is correct.
54.
Sol: - The resistance of earth electrode should be low to give good protection. An earth electrode will
only be effective so long it has a low resistance to the earth and can carry large currents without
deteriorating.
Hence option (a) is correct.
55.
Which of the following electronic instruments can be used to frequency component of
a waveform and its higher harmonics?
1. CRO 2. Vacuum tube voltmeter
3. Spectrum Analyzer 4. Distortion factor
meter
Select the correct answer using the codes given below:
(a) 1 and 2
(b) 2 and 3
(c) 3 and 4
(d) 1 and 4
Sol: - Both spectrum analyzer and distortion factor meter are signal analyzers. A spectrum
analyzer sweeps the signal frequency band and displays a plot of amplitude versus frequency.
Distortion factor meter tunes out the fundamental signal and gives an induction of the
harmonics.
Hence option is correct.
56. In a single phase power factor meter, controlling torque is
(a) Provided by spring control
(b) Provided by gravity control
(c) Provided by stiffness of suspension
(d) Not required
Sol: - There will be two deflecting torques, one acting on the coil A and the other on coil B.
the coil windings are so arranged that the torques due to the two coils are opposite
in direction. So, there is no controlling torque.
57. A dc circuit can be represented by an internal voltage source of 50V with an output resistance of
100k. In order to achieve accuracy better than 99% for voltage measurement across its terminals,
the voltage measuring device should have a resistance of at least.
(a) 10 M
(b) 1 M
(c) 2 M
(d) 1k
Sol: - EL/E0 = 49.5/50 = 1/(1+100k/ZL)
ZL = 10 M
Hence, option (a) is correct.
58. In Electrodynamometer watt meters, the inductance of pressure coil produces error. The error is
(a) Constant irrespective of the power factor of the load
(b) Higher at high power factor load
(c) Higher at low power factor load
(d) Highest at unity power factor load
Sol: - the error caused by pressure coil inductance is VIsin tan. With low power factor,
the value of is large and therefore, the error is correspondingly large.
Hence, option is correct.
59. An analog voltage signal whose highest significant frequency is 1 kHz is to be coded with a resolution
of 0.01 percent for a voltage range of 0-10 V. the minimum sampling frequency and the minimum
number of bits should respectively be
(a) 1 kHz and 12 (b) 1kHz and 14
(c) 2 kHz and 12
(d) 2 kHz and 14
Sol: - Minimum sampling frequency = Nyquist sampling rate = 2 kHz and
Resolution = 0.01 = 1/(2n 1) 100 => n 1
4
60. An ac voltmeter using full-wave rectification and having a sinusoidal input has an ac
sensitivity equal to
(a) 1.414 times dc sensitivity
(b) Dc sensitivity
(c) 0.9 times dc sensitivity
(d) 0.707 times dc sensitivity
Sol: - For full-wave rectifier type of instruments
Vavg = 0.9V where V= Vm/2.
Hence option is correct.
61. Consider the following statements in connection with measurement of temperature:
1. A thermistor is highly sensitive as compared with platinum resistance
thermometer.
2. The resistance of a thermistor is solely a function of its absolute temperature
whether the source of heat is external, internal or both.
3. A thermistor has linear resistance temperature characteristics
4. Most thermistors exhibit negative resistance temeperature coefficient
Which of these statements are correct?
Sol:- Given that current through the branch AD is zero. This implies the case of bridge
balance. At the balance condition, voltage across AB will be equal to voltage across BD.
Also RAB=RBD => R = 120
Battery current IB = 1/(120||120+80||80)
IB = 1/ 100 = 10 mA
Hence, option is correct option
68. When a steady current is passed through a ballistic galvanometer, then the deflection will
Be
(a) maximum
(b) zero
(c) twice the normal value as it depends on Hibbert magnetic standard
(d) None of the above
Sol:- when a steady current is passed through a ballistic galvanometer, then the deflection
will be zero because, the net deflecting torque will be zero.
Hence, option (b) is correct.
69. The measurement of frequency can be carried out with
(a) Owens bridge
(b) Wiens bridge
(c) Maxwells inductance-capacitance bridge
(d) Scherings bridge
Sol:- measurement of frequency can be carried out with Wiens bridge
Hence, (b) is correct option
70. An angular deflection type indicating PMMC meter is provided with spring control and
(a) damping by air dashpot
(b) electromagnetic damping in the aluminium former only
(c) electromagnetic damping in the aluminium former and the moving coil as well
(d) no damping
Sol:- Generally, electromagnetic damping in the aluminium former is provided for a PMMC
type instrument with spring control.
Hence, (b) is correct option.
71. An advantage of PMMC instrument is that it
(a) is free from friction error
(b) has high torque-to-weight ratio of moving parts
(c) has low torque-to-weight ratio
(d) can be used on both AC and DC
Sol:- In a PMMC instrument, the deflecting torque is provided by the movement of coil of light
weight. Hence the torque to weight ratio of the moving parts in the instrument is high due to
which the sensitivity of the instrument is high.
Hence, (b) is correct option
72. The following is not essential for the working of an indicating instrument
(a) Deflecting torque
(b) Braking torque
(c) Damping torque
(d) Controlling torque
Sol:- For the working of an indicating instrument, braking torque is not essential.
Hence, (b) is correct option
(b) 50
(c) 100
(d) 1000
(b) 15
(c) 25
(d) 30
92. Which one of the following is used for the measurement of 3-phase
power factor ?
(A) Power factor meter
(B) Crossed-coil power factor meter
(C) Phase-angle watt hour meter
(D) Polarised-vane power factor meter
Sol:Power factor meter is used for measurement of 3-phase power factor.
Hence, (a) is correct option
93. What is the swamping resistance which is connected in series with the working coil
of a voltmeter to drastically reduce the error in measurement caused due to variation in
temperature, made of ?
(A) Constant
(B) Eureka
(C) Manganin
(D) Nichrome
Sol:Swamping resistance is made up of a material having high temperature coefficient of
resistance so that increase in temperature cannot affect its characteristics. Manganin is
generally used as swamping resistance which is connected in series with a voltmeter
pressure coil.
Hence, is correct option.
4
1
3
3
2
2
(B) 4
(D) 1
1
4
2
2
3
3
Sol:In a flux meter, a special type of ballistic galvanometer provided with very small controlling
torque and large electromagnetic damping.
Hence, (a) is correct option
102. Why is damping of a ballistic galvanometer kept small?
(A) To get minimum overshoot
(B) To make the system critically damped
(C) To make the system oscillatory
(D) To get first deflection large
Sol:To get minimum overshoot, damping of a ballistic galvanometer kept small.
Hence, (a) is correct option
103. Match list I with list II and select the correct answer using the code given below the lists:
List I
(Type of Instrument)
A.
B.
C.
D.
List II
(Example)
Indicating
Absolute
Recording
Integrating
1.
2.
3.
4.
Wattmeter
Tangent galvanometer
Aneroid barometer
Energy meter
Code:
(a)
(b)
(c)
(d)
1
4
1
4
2
2
3
3
3
3
2
2
4
1
4
1
= 0.01 ohm
Hence, is correct option
106. A 0 to 300 V voltmeter has an error oft: 2% of fsd. What is the range of readings if true
voltage is 30 V?
(a) 24V-36V
(b) 20V-40V
(c) 294 V - 306 V
d) 20V-30V
Sol:% error on 0-300v scale = 2/100* 300 = 6V
If the voltage is 30V, % error = 30+6 = 24 to 36V
Hence, (a) is correct option
107. ln a digital voltmeter, the oscillator frequency is 400 kHz. The ramp voltage falls from 8 V
to 0 V in 20 ms. what is the number of pulses counted by the counted?
(a) 8000
(b) 4000
(c) 3200
(d) 1600
Sol:No. of pulses = 400 1032010-3
= 8000
Hence, (a) is correct option
108. Which one of the following is frequency sensitive bridge?
(a) De-Sauty bridge (b) Schering bridge (c) Wiens bridge
(d)Maxwell's bridge
110. A 100 kV, 50 Hz supply is fed to a rectifier ammeter (using a bridge rectifier) through a
capacitor. The PMMC ammeter of the rectifier instrument reads 45 10-3 amp. What is the value
of the capacitor?
(a) 15.910-10 F
(b) 15.910-12 F
(c) 17.6610-9 F
(d) 17.6610-11 F
Sol:Bridge rectifier output is, V0 = 2Vm/
V0 = I Xc
2 Vm/ = I/C
2100103 2/ = 4510-3/ (250C)
C = 15.910-9 F
Hence, (a) is correct option
111. Which one of the following is not the criterion used to select potentiometer in a control
system?
(a) Accuracy (b) Noise
(c) Time response
(d) Frequency restiponse
Sol:Noise is not the criterion used to select potentiometer in a control system.
Hence, (b) is correct option
112. The resolution of DAC is governed by which one of the following (where n is the number
of digital inputs)?
(a)2n
(b) 2/n
(c) 2n
(d) 2n
Sol:Resolution = Vmax/(2n-1)
Hence, is correct option
113. Which one of the following statements is correct?
Spiral springs are used in instruments to
(a) Provide controlling torque
(b) provide damping torque
lead the current to moving coil as well as to provide the controlling torque
(d) Provide linear deflection
Sol:Spiral springs are used in instruments to lead the current to moving coil as well as to provide
the controlling torque.
Hence, is correct option
114. Which one of the following decides the time of response of an indicating instrument?
(a) deflecting system (b) controlling system (c) damping system (d) pivot and jewel bearing
Sol:Damping system decides the time of response of an indicating instrument.
Hence, is correct option
115. Which one of the following is measured by the loss of charge method?
(a) Low R
(b) High R
(c) Low L
(d) High L
Sol:High R is measured by the loss of charge method.
Hence, (b) is correct method.
116. The power of a three phase, three wire balanced system was measured by two wattmeter
method. The reading of one of the watt meters was found o be double that of the other. What is
the power factor of the system?
(a) 1
(b) 0.866
(c) 0.707
(d) 0.5
Sol:Cos = cos (tan-1(3(2P-P)/(2P+p)))
= 0.866
Hence, (b) is correct option
117. To measure 5 Volts, if one selects a 0-100 volt range voltmeter which is accurate within +
1%, then the error in this measurement may be up to
(a) + 1.5%
(b) + 2.5%
(c) + 7.5%
(d) + 20%
Sol:Magnitude of error = 1*100/100 1V
% of error = 1*100/5 = + 20%
Hence, (d) is correct option
\
118. The voltage coil of a single-phase house service energy meter
(a) is highly resistive
(b) is highly inductive
(c) is highly capacitive
a phase angle equal to load power factor angle.
(d) has
Sol:The voltage coil of a single-phase house service energy meter is highly inductive.
Hence, (b) is correct option
119. What is the series resistance required to extend the 0-100V range of a 20000/V meter to 01000 V?
(a) 10 M
(b) 16 M
(c) 18 M
(d) 20 M
Sol:Initial resistance = 20000100 = 2 M
Final resistance = 200001000 = 20 M
Therefore, series resistance = 20-2 = 18 M
Hence, is correct option
120. Hays bridge is suitable for the measurement of which one of the following?
3
1
2
4
List II
1. Self balancing bridge
2. Wien bridge
3. PMMC ammeter
4. Moving iron ammeter
(b)
(d)
A
2
2
B
1
4
C
3
3
D
4
1
(d) Wideband
(D) 10
1
1
2
2
3
4
4
3
(B)
(D)
4
4
1
1
2
3
3
2
Sol:Integrating type DVM measure true average using voltage to conversion. Basic ramp
type DVM does not use D/A converter. Staircase-ramp type DVM uses D/A converter.
Continuous balance type DVM is used in servo-driven potentiometer.
Hence, (b) is correct option
137. The measurement of very low and very high frequencies is invariably done using a
frequency counter in which one of the following?
(A) Frequency measurement mode only
(B) Period measurement mode only
(C) Frequency and period measurement modes, respectively
(D) Period and frequency measurement modes, respectively
Sol:Low frequency measurement in frequency mode gives low-accuracy. So, period
measurement is used.
Hence, (d) is correct option.
1
2
4
3
List II (Damping)
1. Air friction
2. Electromagnetic
3. Fluid friction
4. Eddy current
(b)
(d)
A
3
3
B
4
1
C
2
2
D
1
4
Sol:Air friction does not require the use of permanent magnet whose introduction may lead
to distortion of operating field. This damping is used whose operating field is weak. It is
generally used in moving iron, dynamometer, hotwire instruments.
Galvanometer uses electromagnet damping. PMMC type instrument uses eddy current
damping. Electrostatic instrument uses fluid friction damping.
Hence, is correct option.
140. In a vibrating reed type frequency meter, all the reeds
(a) are of identical dimensions and weight
(b) have different natural frequencies
have the same natural frequencies
(d) are not placed closed to an
electromagnet
Sol:Vibrating reed frequency meter indicates the supply frequency by means of individual
reeds. When rated voltage is applied across the terminals of the meter, the particular
reed whose natural frequency of vibration coincides with the supply frequency, vibrate
with full amplitude.
Hence, (b) is correct option
144. If an energy meter disc makes 10 revolutions in 100 seconds when a load of 450
W is connected to it, the meter constant (in rev/kWh) is
(a) 1000 (b) 500 (c) 1600 (d) 800