REPAIR MANUAL

ISSUED............... AUG. 2008


REVISED ............

AUTO KERATO-REFRACTO-TONOMETER

TRK-1P

TOPCON CORPORATION
R-TRK1P-0-0809-02

R-TRK1P-0-0809-02

CONTENTS
Page

1. INTRODUCTION ......................................................................................................................................1
1-1 Features of product .................................................................................................................................1
1-2 Specifications and performance of product ............................................................................................1
1-3 Target users of this manual .....................................................................................................................2
1-4 How to use this manual ...........................................................................................................................2
1-5 Cautions ..................................................................................................................................................2

2. PRODUCT OUTLINE .............................................................................................................................3


2-1 Nomenclature ..........................................................................................................................................3
2-2 Optical system (Layout) .........................................................................................................................4
2-3 Electric system ........................................................................................................................................8
2-4 Mode setting switches ...........................................................................................................................10
2-5 Principles ..............................................................................................................................................11
2-6 Measurement procedure ........................................................................................................................22
2-7 Layout diagrams for PCB .....................................................................................................................31

3. REPAIR WORK ......................................................................................................................................40


3-1 Inspection and adjustment procedures ..................................................................................................40
3-2 Disassembly and assembly procedures .................................................................................................49
3-3 Trouble shooting ...................................................................................................................................79
3-4 Assembly and adjustment procedures ................................................................................................103

4. TOOL LIST .............................................................................................................................................187


4-1 Special tool list ...................................................................................................................................187
4-2 General repair tools .............................................................................................................................190

R-TRK1P-0-0809-02

1. INTRODUCTION
1-1 Features of product
(1) You can measure the refracting power, corneal shape, intraocular pressure and corneal thickness of the
patients eyes by easy operation.
(2) When measuring the intraocular pressure, air blows more gently than before.
(3) You can perform measurement quickly in the optimal condition because of the auto-alignment function
and auto-start function.

1-2 Specifications and performance of product


REF measurement
Measuring range

Minimum measurable pupil diameter


Target fixation
Measurement data display
Measurement data recording
Kerato measurement
Measuring range

Measurement data display


Measurement data recording

Ocular pressure measurement


Measuring range
Working distance
Measurement data display
Measurement data recording
Alignment
PD measurement
TV Monitor
External output terminal
Operating temperature
Main body travel
Chinrest travel distance

Hyperopia: 0 to +22D
0.25D step display (switchable to 0.12D step display)
Myopia: 0 to -25D
0.25D step display (switchable to 0.12D step display)
Astigmatism: 0 to 10D (+ or -)
0.25D step display (switchable to 0.12D step display)
Axial angle: 0 to 180
1 step display (switchable to 5 step display)
2.0mm
Auto fog system
Color LCD monitor screen
Built-in printer
(10 measurement data each of right and left eyes)
Radius of corneal curvature: 5.00 to 10.00mm 0.01mm step display
Corneal refractory power: 67.50 to 33.75D
0.25D step display (switchable to 0.12D step display)
(Proviso: corneal refractive index = 1.3375)
Corneal astigmatic power: 0 to 10D (+ or -)
Corneal astigmatic axial angle: 0 to 180
1 step display (switchable to 5 step display)
Color LCD monitor screen
Built-in printer
(10 measurement data each of right and left eyes)
(Proviso: printing only typical values for R/K mode)
0 to 30/0 to 60mmHg
11mm
Color LCD monitor screen (with average value)
Built-in printer (with average value)
Color LCD monitor screen
measuring range 85mm max., 1mm display unit
6.5 type VGA color LCD monitor
RS-2322, USB1
10 to 40C
distance 40mm back/forth, 90mm sides, 30mm up/down
60mm up/down (electric)

* For product improvements, the specification and appearance are subject to change without prior notice.
Essential performance
Measurement can be done.
Monitor display is not disturbed.

R-TRK1P-0-0809-02

1-3 Target users of this manual


TRK-1P is an optical precision inspection instrument to be used for ophthalmology and internal medicine.
Only a specialist for assembly and adjustment of medical machines can repair this instrument.
For others except the specialist, it is recommended to take the TOPCON repair technique training course.

1-4 How to use this manual


When performing initial inspection or inspecting, adjusting or repairing the TRK-1P for any problem, use this
manual according to the flow chart below.
Manual
Operations procedure

Initial inspection

Applicable item

Page

Trouble

Inspection

OK

Inspection and adjustment


procedures

P.40 - 48

Disassembly and assembly


procedures

P.49 - 78

Trouble shooting

P.79 - 102

NO
Trouble location
check
Searching for the
cause

Repair operations

Inspection

NO

Disassembly and assembly


procedures
Assembly and adjustment
procedures
Inspection and adjustment
procedures

P.49 - 78
P.103 - 186
P.40 - 48

Normal
YES

End

TRK-1P basic data

Measurement principle

P.11 - 21

1-5 Cautions
Refer to the illustrations of the parts list for the disassembly and assembly of the unit which is not described
in this manual. Refer to the instruction manual for replacing the consumables.
If the optical system is soiled when repairing, photographing is adversely affected. Be careful not to touch
or soil the optical system when disassembling, assembling or adjusting.
Order the repair parts by using the service parts list.

R-TRK1P-0-0809-02

2. PRODUCT OUTLINE
2-1 Nomenclature
Main body Section
Measuring head

Color LCD monitor


Control panel
Brightness control
Measurement switch
Control lever
Chinrest up/down button
Power lamp

External I/O terminal

Power supply Section

Forehead rest
Measuring opening
Height mark
Chinrest tissue stopper pin
Printer cover open button
Chinrest
POWER switch

Chinrest Section

Inlet cover

Measuring window at TONO/PACHO mode


Measuring window
Measuring opening

Measuring nozzle

Measuring window
lens cap

R-TRK1P-0-0809-02

Eye being
measured

Anterior eye section


illumination
LED4 950nm

Reflective
mirror 3

Measurement relay lens B

Anterior eye section illumination LED4


950nm

Measurement relay
lens A

Movement lens

Condenser
lens

Measurement
target

TV lens

Measurement
LED 860nm

Half
mirror 5

Tungsten
lamp

Alignment
relay lens B

Alignment relay
lens A

Reflective mirror 1

Pinhole prism

Parallel projection
LED2 950nm
Ring pattern LED26
950nm

Objective lens

Ring iris

Half mirror
3

SPOT LED
940nm

Half mirror
1

Condenser lens

Rotary
prism

Half mirror
2

Fixation
target relay
lens A

Fixation target
relay lens B

Fixation
target

Color
compensation
filter
Condenser
lens

Reflective
mirror 2

1/3" CCD
CAMERA

2-2 Optical system (Layout)

2-2-1 Measurement system

R-TRK1P-0-0809-02

Eye being
measured

Anterior eye section


illumination
LED4 950nm

Reflective
mirror 3

Measurement relay lens B

Anterior eye section illumination LED4


950nm

Measurement relay
lens A

Movement lens

Condenser
lens

Measurement
target

TV lens

Measurement
LED 860nm

Half
mirror 5

Tungsten
lamp

Alignment
relay lens B

Alignment relay
lens A

Reflective mirror 1

Pinhole prism

Parallel projection
LED2 950nm
Ring pattern LED26
950nm

Objective lens

Ring iris

Half mirror
3

SPOT LED
940nm

Half mirror
1

Condenser lens

Rotary
prism

Half mirror
2

Fixation
target relay
lens A

Fixation target
relay lens B

Fixation
target

Color
compensation
filter
Condenser
lens

Reflective
mirror 2

1/3" CCD
CAMERA

2-2-2 KR unit fixation target system

R-TRK1P-0-0809-02

Eye being
measured

Anterior eye section


illumination
LED4 950nm

Reflective
mirror 3

Measurement relay lens B

Anterior eye section illumination LED4


950nm

Measurement relay
lens A

Movement lens

Condenser
lens

Measurement
target

TV lens

Measurement
LED 860nm

Half
mirror 5

Tungsten
lamp

Alignment
relay lens B

Alignment relay
lens A

Reflective mirror 1

Pinhole prism

Parallel projection
LED2 950nm
Ring pattern LED26
950nm

Objective lens

Ring iris

Half mirror
3

SPOT LED
940nm

Half mirror
1

Condenser lens

Rotary
prism

Half mirror
2

Fixation
target relay
lens A

Fixation target
relay lens B

Fixation
target

Color
compensation
filter
Condenser
lens

Reflective
mirror 2

1/3" CCD
CAMERA

2-2-3 KR unit alignment system

R-TRK1P-0-0809-02

2-2-4 CT unit

Line sensor

Relay lens

Cylindrical lens
Z sensor

CCD camera

Relay lens

Split mirror

Objective
lens

Fixation LED

Cover glass

XY LED

Diffusion
plate
Dichroic mirror

Corrector lens

Applanation sensor

Collimator lens

Objective lens

Imaging lens
Anterior eye section
illumination LED
Diffusion plate

Split
mirror
Glass with
nozzle

Half mirror
Split mirror

Chamber window
glass
Split mirror

Eye being
measured

Anterior eye
section window
glass

PACHO LED
Iris

Objective lens

Condenser lens

Condenser lens
Relay lens
Z alignment/applanation LED

R-TRK1P-0-0809-02

2-3 Electric system

LCD Panel SW1 PCB

LCD

+12V
VIDEO
composition
circuit

CCD camera unit


+15V

Driver for lighting LED

LCD relay
PCB

LCD panel SW2 PCB


TF motor

Rotary solenoid
Open air motor

Measurement LED
S RING2 LED
SP LED
S RING1 LED
Fixation target lamp
SPOT LED

TRK zero detection sensor

2-3-1 Electric system block diagram

Motor drive circuit

FONT ROM

D/A conversion

Head changing rear limit sensor

Head changing front limit sensor

Head changing lower limit sensor

Head changing upper limit sensor

ANALOG
I/F PCB

MOTOR DRIVER PCB

+3.3V
+5V
+15V
+24V

Head changing back-and-forth motor

XY LED

Head changing up-and-down motor

Fixation LED

Z axis zero detection sensor

PACHO LED

X axis zero detection sensor

Driver for
lighting LED

Z LED

CPU

CONTROL PCB

Anterior eye section


illumination

Y axis lower limit sensor

Z sensor PCB

Y axis zero detection sensor

Applanation sensor PCB

Flash ROM
SDRAM
E2PROM
Clock generation circuit

Y axis upper limit sensor

Pressure sensor PCB

Analog signal
processing
circuit

LZM DRIVER PCB

USB

POWER PCB

+24V

RS232C IN

Voltage conversion circuit

Motor drive circuit


+5V

RS232C OUT
+5V

Z axis motor

Y axis motor

X axis motor

+24V
Switching
power supply

PRINTER PCB
Printer control circuit

Electric
chinrest PCB
DC motor

Fuse

Printer
SW control circuit

Plug

Commercial
power supply

Measurement
SW

Chinrest
up/down SW
CONTROL SW PCB

R-TRK1P-0-0809-02

P302 1 2
J302 1 2

P310 1 2 3
P304 1 2 P303 1 2
J310 1 2 3 J311 1 2 3 J304 1 2 J303 1 2

41840 6120
TF MOTOR ASSY

OPEN AIR MOTOR ASSY

ROTARY SORENOID ASSY

TRK ZEROKEN PCB ASSY

SPOT PCB ASSY

SRIING2 PCB ASSY

SR LED PCB ASSY

KOSHI LAMP ASSY

M LAMP PCB ASSY

CT OPTICAL HEAD

SRIING1 PCB ASSY

2-3-2 Wiring diagram

LCD UNIT

P312 1 2 3 P305 1 2 P308 1- 4 P307 1 - - 6


J312 1 2 3 J305 1 2 J308 1- 4 J307 1 - - 6

P306 1 2 P309 1- 4
J306 1 2 J309 1- 4

LCD PANEL SW CABLE 1 UNIT


LCD PANEL SW PCB 1

PROTECTIVE EARTH CABLE3

41850 7300

LCD RELAY PCB

LZM DRIVER PCB

INVERTER UNIT

LCD PANEL SW PCB 2

PROTECTIVE EARTH CABLE 4

J301 1 - - - - - - 30
P301 1 - - - - - - 30

LCD PANEL SW CABLE 2 UNIT

LCD INVERTER CABLE UNIT

CONDENSER UNIT

41850 7860
LZM CABLE UNIT

SOLENOID UNIT
P17 1 2
J17 1 2

FOR DEBUG

P16 1 2 3
J16 1 2 3

J99 1 - - 14

1 - - 10 J98

INVERTER POWER CABLE

LCD-CONTROL CABLE UNIT

P3 1 - - - - - - 30
J3 1 - - - - - - 30

PACHO CABLE 2 UNIT


CAMERA IMAGE CABLE UNIT
LINE SENSOR
PCB

PACHO CABLE 1 UNIT

VM_CABLE ASSY

1 2
1 2

CONTROL PCB

P9 J9

CCD CLOCK CABLE 1 UNIT

1 - - - 15
1 - - - 15

CCD UNIT

CCD CLOCK CABLE 1 UNIT

41850 7100

P8 J8

CCD POWER CABLE 1 UNIT

RM OPTICAL HEAD

1 2
1 2

PROTECTIVE EARTH CABLE 1

USB I/F CABLE 2 UNIT

CTPL-PWR CABLE 1 UNIT

CTPL-PWR CABLE 2 UNIT

HEAD CHANGING REAR LIMIT SENSOR UNIT

HEAD CHANGING FRONT LIMIT SENSOR UNIT

HEAD CHANGING LOWER LIMIT SENSOR UNIT

HEAD CHANGING UPPER LIMIT SENSOR UNIT

HEAD CHANGING BACK-AND-FORTH MOTOR UNIT

HEAD CHANGING UP-AND-DOWN MOTOR UNIT

PRINTER

Z AXIS JOG SENSOR UNIT

PRINTER PCB

MEASUREMENT SW
X AXIS JOG SENSOR UNIT

1 2
1 2

SWITCHING POWER SUPPLY

CONTROL SW
CABLE UNIT

CONTROL
SW PCB

PROTECTIVE EARTH CABLE 5

POWER PCB

POWER CABLE 2 UNIT

POWER CABLE 1 UNIT

3P_SW CABLE UNIT

Z AXIS ZERO DETECTION SENSOR UNIT

RS-232C IN CABLE UNIT

MOTOR DRIVER
PCB

P211 J211

XY LED UNIT

USB I/F CABLE UNIT

ENCODER SENSOR UNIT

1 2
1 2

P210 J210

FIXATION LED UNIT

M_DRV_POWER
CABLE UNIT

ROUGH/FINE MOVEMENT CHANGING SENSOR UNIT

1 2
1 2

P209 J209

PACHO LED UNIT

ANALOG PCB

Z LED UNIT

1 2
1 2

P208 J208

PRINTER POWER CABLE ASSY

P207 J207

41850 7200

ANTERIOR EYE SECTION


ILLUMINATION LED UNIT

P206 J206

1- 4
1- 4

Z SENSOR PCB

PRINTER I/F CABLE UNIT

P203 J203

Z SENSOR CABLE UNIT

Y AXIS LOWER LIMIT SENSOR UNIT

X AXIS MOTOR CABLE UNIT

APPLANATION SENSOR PCB

Y AXIS ZERO DETECTION SENSOR UNIT

P204 J204

APPLANATION SENSOR CABLE UNIT

Y AXIS UPPER LIMIT SENSOR UNIT

PRESSURE SENSOR PCB

X AXIS ZERO DETECTION SENSOR UNIT

1- 6
1- 6

J201

PROTECTIVE EARTH CABLE 2

STEPPING MOTOR Z AXIS

1 - - 12

P205 1 - - - 40
J205 1 - - - 40

Y AXIS MOTOR UNIT

ANALOG CABLE UNIT

Z AXIS MOTOR CABLE UNIT

J4 1 - - - 40
P4 1 - - - 40

STEPPING MOTOR X AXIS

P10 J10

CCD POWER CABLE 2 UNIT

CCD UNIT

RS-232C OUT CABLE UNIT

CHINREST CABLE UNIT

ELECTRIC
CHINREST PCB

R-TRK1P-0-0809-02

2-4 Mode setting switches


The mode setting switch is only SW5 on the CONTROL PCB.
The following table shows the functions of the mode setting switches. Refer to it.
No.

ON

Unused.

Compensation mode for China

PCB check mode

Unused.

Changes the R/K step. (0.01 display)

Displays the AD value among the CT measured values.

Displays the serviceman menu.

Adjustment mode

10

R-TRK1P-0-0809-02

2-5 Principles
2-5-1 Kerato-refractometer unit
1. Target image projection to eyeground (ref function)
The fundamental principle is the same as for previous models: the measurement target is projected from the
iris image of the pupil section and the light reflected from the eyeground is detected.
The following figure is a general diagram of this method.

Ring iris

Eyeground

Eye being measured


Measurement
target

The eyeground is observed from the center by


shooting from the pupil section ring iris image
into the eyeground.

As the figure shows, the ring pattern is used as the measurement target, the image is projected onto the eyeground from the pupil ring iris and the image reflected from the eyeground is detected with the CCD CAMERA. The detected image is image processed and the way in which the circle is converted shows the
refraction of the eye.
2. Measuring principle (kerato function)

Kerato ring image projected on the cornea.


Size: 3mm (in the case of R7.7)
Kerato ring

As shown in the figure, the kerato ring is projected on the cornea and the reflected image is detected with a
CCD CAMERA. The reflected image detected here undergoes image processing to compute the radius of
curvature and the refracting power of the cornea.

11

R-TRK1P-0-0809-02

3. Functional block diagram


Here is a flow chart for this process.

(In the case of REF.)


The target image is projected
onto the eyeground by the
optical system.

The image reflected back from


the eyeground is detected by
the REF CCD CAMERA.

Image processing.
Measurement of the center of
the reflected image

Round

Round
or
elliptical

Near shighted or
far sighted

Elliptical

Astigmatic
(In the case of KERATO)
A kerato ring image is
projected on the cornea with
the optical system.

The image reflected back


from the cornea is detected by
the KRT CCD CAMERA.

Image processing.
Measurement of the center of
the reflected image

12

R-TRK1P-0-0809-02

In the case of REF mode


The target image is projected
onto the eyeground by the
optical system.

The image reflected back from


the eyeground is detected by
the REF CCD CAMERA.

Rough measurement

Image processing.
Measurement of the center of
the reflected image

Calculation

NO

Focus
of the reflected
image OK?
YES

Measurement target motor is


moved.

The target image is projected


onto the eyeground by the
optical system.

The image reflected back from


the eyeground is detected by
the REF CCD CAMERA.

Measurement

Image processing.
Measurement of the center of
the reflected image

Calculation

S.C.A. (spherical, cylindrical,


axis) data output

13

R-TRK1P-0-0809-02

In the case of R/K mode


The target image is projected onto the
eyeground by the optical system.

The image reflected back from the eyeground


is detected by the REF CCD CAMERA.
Rough
measurement
of refractometer

Image processing. Measurement of the


center of the reflected image

Calculation

NO

Focus
of the reflected
image OK?
YES

Measurement target motor is moved.

The kerato ring image is projected on the


cornea by the optical system.

The image reflected back from the cornea is


detected by the KRT CCD CAMERA.

Kerato
measurement

The image data is written on the memory.

The target imgae is projected onto the


eyeground by the optical system.

The image reflected back from the eyeground is detected by the REF CCD CAMERA.
Ref
measurement
Image processing. Measurement of the
center of the reflected image.

Reads image data from memory.

Calculation of refractometer data.

Image processing. Measurement of


the center of the reflected image.

Calculation of kerato data.

Kerato
measurement

S.C.A. (spherical, cylindrical, axis) data


output. H.V. data output.

14

R-TRK1P-0-0809-02

In the case of KRT mode

The target image is projected


onto the eyeground by the
optical system.

The image reflected back from


the eyeground is detected by
the REF CCD CAMERA.

Image processing.
Measurement of the center of
the reflected image

Calculation

NO

Rough measurement
of refractometer

Focus
of the reflected
image OK?
YES

Measurement target motor is


moved.

The kerato ring image is


projected onto the cornea by
the optical system.

Only at the first measurement


The image reflected back from
the cornea is detected by the
KRT CCD CAMERA.

Kerato measurement

Image processing.
Measurement of the center of
the reflected image

Calculation

H.V. data output

15

R-TRK1P-0-0809-02

4. Image processing
(1) Objective
To find the centers of the inner diameter and outer diameter of the ring image reflected back from the eyeground.
This circle or ellipse
is measured.

(2) Method
1 As shown in the figure below, the image is divided up like a checker board. The right and left lines are at
regular intervals and parallel to the Y axis and the up and down lines are at regular intervals and parallel to
the X axis.
The axes are set up so that the intersection of the X-axis and Y-axis is the center of the ring.
2 Suppose that the difference of the coordinates for the inner diameter and the outer diameter of the ring is
only 7 units, as in the figure below. In this case, if the smallest unit that can be resolved is 1 unit, the center
between the outer diameter and the inner diameter is either Circle A or B. In other words, if the difference
between the inner diameter and the outer diameter is an even multiple of the minimum resolution, then the
center can be measured accurately, but if the difference is an odd multiple, then the center cannot be measured accurately.
3 If the reference units are made smaller than the measurement precision, an error of one reference unit
could be ignored.
Y axis

X axis

A
B

16

R-TRK1P-0-0809-02

The CCD CAMERA is a collection of extremely small such checker board squares. Normally, a CCD CAMERA comprises hundreds of thousands of squares.
In summary:
1 The reflected image detected with the CCD CAMERA is stored into memory with one data bit for each
checker board square.
2 The coordinates are calculated from the stored data and the center of the inner diameter and the outer
diameter of the ring image is found.
3 The pattern for the figure found in 2 is identified to obtain the change in the image.
(3) Calculation of the S.C.A. data

0D

S
TF MOTOR
movement distance
(S.E.=Spherical
Equivalent)

Deformation

S: Spherical
S : Spherical
C: Cylinder
C : Cylindrical
A: Axis
A : Axis

Measuring the change in the distance of the ring image from the origin of the coordinates (center of the ring
image) over a certain angle gives the S.C.A. data.

17

R-TRK1P-0-0809-02

5. Rotary prism method


There is fluctuation between the measured values because the eyeground condition is not always even due to a
disease, blood vessles, etc. Place a prism in front of the ring iris and rotate it during measurement. In a wider
area than the usual eyeground, the eyeground information is averaged to reduce the above fluctuation.
Measured target image projected
on the eyeground

Ring iris

Eyeground

Eye being measured


Prism
Ring iris

Projection condition on eyeground

Approx.
1.4d

Rotary prism method

Conventional method

18

R-TRK1P-0-0809-02

6. Principles of auto-alignment
For vertical or horizontal directions (X and Y axes), check the alignment by using the positional relation
between the alignment spot and alignment mark of the reflected image as the conventional RM/KR-8800
series. For the focus direction (Z axis), check the alignment by using the relation between the parallel projected spot and placido ring.

Measurable min. pupil diam.


2mm or equivalent

Eye being measured

Focus direction
(Z axis)

Horizontal direction
(X axis)
Vertical direction
(Y axis)

Alignment mark

Alignment spot

XY direction judgment

Alignment mark
NG
(Overlaps with the
alignment mark.)

OK
(Not in contact with
the alignment mark.)

OK
(In contact with the
alignment mark.)

+
Z direction judgment
Within 0.75mm against the reference position

19

R-TRK1P-0-0809-02

2-5-2 Non-contact meter unit


The principle of this instrument is to blow an instantaneous jet of compressed air on the cornea, then to calculate the intraocular pressure from the numerical value of the cylinder pressure required to make the cornea to
a certain plane surface area.
Measurement method is as follows.
Measurements are made by blowing an air jet on the cornea of the patient's eye, determining the amount of the
change in shape from the change in the amount of the light reflected and detecting the condition at the instant
of applanation.
The jet of air blown on the patient's eye causes the cornea to change its shape in the sequence in the following
figure.
Cornea

Convex

Plane

Concave

Plane

Convex

The change in shape is measured from the difference between the received light from the infrared LED and
the reflected light detected by the intensitometer.
The reflected light to the applanation sensor is weak when the cornea is spherical (in the following figure at
the left). When the cornea is made to a certain plane surface area (in the following figure at the right), the
maximum reflected light is detected.
Applanation sensor

Eye
Nozzle
Alignment/applanation LED

20

R-TRK1P-0-0809-02

Amount of reflection light

The pressure of the air jet increases together with the chamber inner pressure (CP), which is continuously
measured by the pressure sensor. The intensity of the light reflected on the cornea (L) is detected by the intensitometer as the function of the chamber inner pressure (CP). (Refer to the following figure.) Using the function in the figure, L=f (CP), the chamber inner pressure (Pc) is determined for the maximum reflected light
intensity (Lmax). The intraocular pressure (IOP) is determined according to the obtained chamber inner pressure (Pc).

Chamber inner pressure

Corneal thickness measurement principles


Slit light is slantingly emitted for the patient's cornea and the light reflected by the front and backside of the
cornea is brought in by the line sensor.
The corneal thickness is calculated by using the interval (L) between the front reflection image and backside
reflection image on the line sensor.
Backside of cornea

Front of cornea

Backside reflection

L
Reflected
light
Front reflection
Line sensor
Slit light
Corneal
thickness

21

R-TRK1P-0-0809-02

2-6 Measurement procedure


2-6-1 General operation flow chart

Power switch ON

ON

Menu SW

Initial menu is set.

OFF

Hardware is initialized.

The title screen is displayed.

R/K

Mode is set.

T/P

R/K is changed to T/P.

The anterior eye section


observation screen is
displayed.

The anterior eye section


observation screen is displayed.
R/K mode
T/P mode

4
Mode selector SW

ON

ON

Mode selector SW

T/P is changed to R/K.

OFF

Measurement
mode SW

ON

OFF

Measurement
mode SW

OFF

ON

OFF

R/K

The measurement
mode is selected.

KRT

T/P

The REF mode screen


is displayed.

PACHO

TONO

REF
The R/K mode screen
is displayed.

The
measurement mode
is selected.

The T/P mode screen


is displayed.

The KRT mode screen


is displayed.

The TONO mode


screen is displayed.

The PACHO mode


screen is displayed.

22

R-TRK1P-0-0809-02

Measurement SW

ON
The patient's eye is measured.

OFF
RL selector SW

ON
The right and left positions in
the measuring unit are changed.

OFF

Print SW

The measurement data are displayed.

ON
Printer outputs the data.

OFF

External communication is set.

ON

Data are outputted.

OFF

Alignment SW

OFF

Target image SW

ON
Auto-alignment measurement/manual measurement is changed.

ON

OFF

Memory image

Not exist.

Menu SW

Menu mode

ON
Switch menu is changed.

OFF

IOL SW

ON
IOL mode is changed.

OFF

Fixation target SW

OFF

The target image is displayed.

ON

OFF

Switch selector SW

Exists.

ON
Fixation target brightness is changed.

23

R-TRK1P-0-0809-02

Clear SW

ON

OFF

Measurement data

Data are erased.

Not exist.

Corneal diameter SW

Exists.

ON
Corneal diameter measurement mode

OFF

Chinrest up/down SW

ON
Chinrest up/down movement

OFF

24

R-TRK1P-0-0809-02

Measurement SW

ON
The patient's eye is measured.

OFF
RL selector SW

ON

The measurement data are displayed.


The right and left positions in the
measuring unit are changed.

OFF

Print SW

ON
Printer outputs the data.

OFF

External communication is set.

ON
Data are outputted.

OFF

Alignment SW

ON
Auto-alignment measurement/manual measurement is changed.

OFF

Initial SW

ON
The measuring unit is reset to the original point.

OFF

Menu SW

ON
Menu mode

OFF

Switch selector SW

ON
Switch menu is changed.

OFF

Clear SW

ON

OFF

Measurement data
Data are erased.

30/60 selector SW

ON
The measurement range is changed.

OFF

Chinrest up/down SW

ON
Chinrest up/down movement

OFF

25

R-TRK1P-0-0809-02

2-6-2 Auto-alignment flow chart

T/P

Mode

R/K

2
3

The anterior eye section image is


captured to judge XY alignment.
Z distance: Within 0.75mm

YES

The center spot is extracted.

NO
The center of gravity for the
center spot is detected.

The center of
gravity is calculated.

Measurement

NG
-5mm

OK

YES

Z < -3.0mm

NO
"

The center of
gravity is calculated.

XY motor is driven.

YES

The center spot is rechecked.

NO

XY distance: Within 0.1mm

"TOO CLOSE"
" Dotted line bar is displayed.

NG
-3.0mm

Z < -0.75mm

YES

OK
NO

The XY movement distance is


calculated according to the pixels
to the center. (Unit: mm)

YES

NO

Z motor is driven.

The anterior eye section image is


captured to judge Z alignment.
The kerato ring LED lights.

Radius of the kerato No. 1 ring

XY distance: Within 0.1mm

XY motor is driven.

NG
+0.75mm < Z

The radius
calculation in
0/180 is OK.

+3.0mm

YES

NO
No. 1 ring is calculated.

The parallel
projection spot is
detected.

YES

XY distance: Within 0.1mm

NG
Z motor is driven.

OK

NO
XY motor is driven.

The ring spot ratio is calculated.

The Z distance is calculated according


to the ratio.

+3.0mm < Z

+5.0mm

YES
"FORWARD"
Solid line bar is displayed.

NO

26

R-TRK1P-0-0809-02

The anterior eye section image


is captured.

The bright spot area in the anterior



eye section illumination is detected.

Not exist.

Bright spot area

Exists.
The bright spot is judged.

The XY direction movement
is calculated.

X direction movement

XY rough alignment is finished.

Movement is finished.

Y direction movement
Measuring unit moves. (X direction)


Measuring unit moves. (Y direction)

The anterior eye section image is captured.

R/K

Mode

T/P
Kerato ring contrast is calculated.

X bright spot contrast is calculated.

Contrast is maximal.

YES

NO
Z direction movement is calculated.

Z rough alignment is finished.

Movement is finished.

Z direction movement
Measuring unit moves. (Z direction)

27

R-TRK1P-0-0809-02

The center spot is detected by


the two-dimensional PSD.

NG
(Out of range)

No spot

Within the XY rough


alignment range

The anterior eye section image is captured.

OK

The center spot position is detected.

NG
(Out of range)

No spot

Within the XY rough


alignment range

The anterior eye section image is captured.

OK

The anterior eye section illumination


position is detected.

NG
(Out of range)

Within the XY rough


alignment range

XY motor is driven.

No
illumination
image

OK

The center spot is detected by


the two-dimensional PSD.

NG (Out of range)

No spot

Within the XY rough


alignment range

OK

XY motor is driven.

Line sensor detection

NG (Out of range)

No spot

Within the Z alignment range

OK

The anterior eye section image is captured.

Z motor is driven.

The center spot position is detected.

YES
Does the spot exist?

NO

Spot shape

Vertical oval

Horizontal oval
Forward movement

NG (Out of range)

Backward movement

Within the X/Y/Z alignment range

OK
Measurement

28

R-TRK1P-0-0809-02

2-6-3 Manual alignment flow chart

Mode

T/P

R/K

The anterior eye section image is


captured to judge XY alignment.

Z distance:
-5.0mm Z < -3.0mm

The center spot is extracted.

"TOO CLOSE"
Dotted line bar is displayed.

NO

The center of gravity for


the center spot is detected.

The center of
gravity is calculated.

YES

NG
Z distance:
-3.0mm Z < -0.75mm

OK
The XY movement distance is
calculated according to the pixels to
the center. (Unit: mm)

YES

"TOO CLOSE" is erased.


Dotted line bar is displayed.

NO

The anterior eye section image is


captured to judge Z alignment.
The kerato ring LED lights.

Z distance:
+0.75mm < Z +3.0mm

YES

"FORWARD" is erased.
Solid line bar is displayed.

NO
Radius of the
kerato No. 1 ring

NG

OK
Z distance:
+3.0mm < Z

The No. 1 ring diameter is detected.

The parallel
projection spot is
detected.

NG

YES
+5.0mm
"FORWARD"
Solid line bar is displayed.

NO

OK
The ring/spot ratio is calculated.

The Z distance is calculated according


to the ratio.

Z distance:
+5.00mm or more

YES

Message and bar are erased.

NO

Measurement SW

ON

Measurement

OFF

The measurement result is displayed.

29

R-TRK1P-0-0809-02

The center spot is detected by


the two-dimensional PSD.

NG

Within the
XY rough alignment
range

OK


The center spot is detected by
the two-dimensional PSD.

Within the
XY fine alignment 
range

OK

Line sensor detection

The Z alignment
position is checked.

NG

OK

Alignment bar is displayed.

Measurement SW

Alignment bar is erased.

ON

OFF
Measurement

The measurement result is displayed.

30

R-TRK1P-0-0809-02

2-7 Layout diagrams for PCB


2-7-1 POWER PCB (41850 6200)
CONTROL SW PCB Y axis lower limit sensor
J611
Y axis zero detection sensor
PRINTER PCB PRINTER PCB
J642
J641

62

Switching
power
supply
P51

CONTROL PCB
J2

,
,

62

62

84

,

,
62

62

62
,

,

,
62

62

84

Y axis upper limit sensor

,

62

84

62

62
62

59

62
,

,
,

62

62
,

84
62 62
62
62
84
62
, , , , ,

84
,
,

,

,

,

,

Z axis zero
detection sensor
MOTOR DRIVER PCB
J632
MOTOR DRIVER PCB
J631

Encoder sensor

CHINREST PCB
J651

USB I/F CABLE

X axis zero
detection sensor

Rough/fine movement changing sensor


CONTROL PCB
J12

RS-232C IN

X axis jog sensor

RS-232C OUT
Z axis jog sensor

Volume
VR1: For adjusting the +7.5V power supply voltage VR4: For adjusting the +15V power supply voltage
VR2: For adjusting the +5V power supply voltage
VR5: For adjusting the X axis jog sensor voltage
VR3: For adjusting the +15V power supply voltage VR6: For adjusting the Z axis jog sensor voltage

Check pins
TP1:
TP2:
TP4:
TP5:
TP6:
TP7:
TP8:

GND
+24V
+15V
VDD (+5V)
+3.3V
CCD +15V
+5VB (+7.5V)

TP9:
TP10:
TP11:
TP13:
TP17:
TP18:
TP19:

Encoder output A
Encoder output B
X axis jog sensor adjustment voltage
Z axis jog sensor adjustment voltage
Rough/fine movement
+1.5V
Resetting

31

TP20:
TP21:
TP22:
TP23:
TP24:

Y upper limit
Y lower limit
Y zero
X zero
Z zero

R-TRK1P-0-0809-02

2-7-2 MOTOR DRIVER PCB (41850 6300)

62 62

62
62

62
62 62

62
,

,

62

62 62

,

,

POWER PCB
J604

POWER PCB
J605

62

62

62

,

62

,

Y axis motor

Z axis stepping motor

X axis stepping motor

TP1:
TP2:
TP3:
TP4:
TP5:
TP6:

GND
+24V
X_CLK
Z_CLK
Y_CLK
Y_DIR

TP7:
TP8:
TP9:
TP10:
TP11:
TP12:

X_DIR
Z_DIR
X_ENABLE
Y_ENABLE
Z_ENABLE
X_HOLD

TP13: Z_HOLD
TP14: Y_HOLD
TP15: VDD (+5V)

2-7-3 PRINTER PCB (41850 6400)


PRINTER

PRINTER

,

,

,
,

POWER PCB
J622

32

POWER PCB
J606

R-TRK1P-0-0809-02

2-7-4 CONTROL SW PCB (41850 6450)

59

59

,

.'&

,

POWER PCB
J613

Measurement SW

2-7-5 CHINREST PCB (41850 6500)

62
84
Motor

,
,

POWER PCB
J614

,

VR1: For adjusting the motor output


TP1: GND

33

R-TRK1P-0-0809-02

2-7-6 CONTROL PCB (41850 71000)


Head changing front limit sensor
LCD SW PCB2
J521

,

LCD relay PCB


J531

EEPROM

,

Head changing
back-and-forth motor

Head changing
rear limit sensor

LZM DRIVER PCB


J301

,

,

,
84

,

Solenoid

,

Condenser

84 ,

CCD unit (RM)


CN1

62

,

CCD unit (CT)


CN1

5-

59

,

,

+%
,
62A#&+0

62
62
+%

62

62
+%
62
CCD

,

,

62

59
59

unit (CT)
CN2

,

62

POWER PCB
J602

59

62

62

62

62 ,

64

CCD unit (RM)


CN2

,

CCD unit (CT)


CN5

59
62

CCD unit (RM)


CN5

59

62
62

62

62

62

25

LINE CCD PCB


J651

,
,

,

,
,

POWER PCB
J626

,

,

,

ANALOG I/F PCB


J205

POWER PCB
J603

LINE CCD PCB


J652

Head changing
up-and-down motor

Head changing
lower limit sensor

Head changing upper limit sensor

DIP switch
SW4: Reset switch

SW5: Mode selector switch

SW6: E2PROM writing switch

Volume
VR1: For adjusting the tilt angle of the pressure
rising curve (52)

VR2: For adjusting the tilt angle of the pressure


rising curve (65)

Test pins
TP1:
TP2:
TP3:
TP4:
TP5:
TP6:
TP7:

CCD 15V
VCC
5VA
AG
+1.5V
DG
3.3V

TP8:
TP9:
TP10:
TP11:
TP12
TP13:
TP14:

15V
+1.5V
+12V
-15V
VIDEO IN1
CLK16M
RESET

34

TP15:
TP16:
TP17:
TP20:

3.3VA
15VA
DG
VIDEO IN2

R-TRK1P-0-0809-02

2-7-7 ANALOG IF PCB (41850 7200)


XY LED
Z LED

CONTROL PCB
J4

Fixation LED

62
,

,

84 84

, ,
,

PACHO LED
84

84

,

Anterior eye section


illumination LED

84
84 84

,

,
62

62

62

62
62

84

84

PRESSURE
SENSOR PCB

62
84

62
)0&

,

,

APPLANATION SENSOR PCB

62 84

Z SENSOR PCB

Volume
VR1: For adjusting the reference voltage
VR2: For adjusting the offset voltage

VR5: For adjusting the LED light intensity to measure the corneal pressure
VR6: For adjusting the XY LED light intensity

VR3: For adjusting the pressure detection circuit VR7: For adjusting the light intensity of the Z sensensitivity
sor/applanation sensor light source
VR4: For adjusting the applanation detection circuit sensitivity

Check pins
TP201: For checking the applanation detection
TP406: AGND
circuit sensitivity voltage
TP202: APPLA_LVL
TP407: DGND
TP204: For checking the pressure detection circuit
sensitivity voltage
TP205: PRESSURE_LVL
TP210: For checking the reference voltage

35

R-TRK1P-0-0809-02

2-7-8 PRESS SENSOR PCB (41850 7310)

84

ANALOG I/F PCB


J204

VR1: For adjusting the offset pressure

2-7-9 APPLA SENSOR PCB (41850 7320)


ANALOG I/F PCB
J203

,

36

R-TRK1P-0-0809-02

2-7-10 Z SENSOR PCB (41850 7330)


Volume
VR1:
VR2:
VR3:
VR4:
84

For adjusting frequency


For adjusting gain
For adjusting frequency
For adjusting gain

84

Test pins
62

62

TP331: For adjusting frequency


TP333: For adjusting frequency
TP335: GND

84
84
62
,

62
62

ANALOG PCB
J206

2-7-11 LINE CCD PCB (41850 7340)

CONTROL PCB
J15

41)

,

%.-

8176

62
)0&

84

,
<A8+&'1

CONTROL PCB
J14

37

R-TRK1P-0-0809-02

2-7-12 LCD PANEL SW1 PCB (41850 7510)


LCD PANEL SW2 PCB
J523

LCD RELAY PCB


J532

,
59

59

,
59

59

59

59

2-7-13 LCD PANEL SW2 PCB (41850 7520)


CONTROL PCB
J11

LCD unit

84

,

,
59

59

LCD PANEL SW1 PCB


J512

59

,
59

59

2-7-14 LCD RELAY PCB (41850 7530)


CONTROL PCB
J13

,
,
LCD PANEL SW1 PCB
J511

LCD unit

,

38

R-TRK1P-0-0809-02

2-7-15 LZM DRIVER PCB (41840 7300)


Zero position detection sensor:
3 pins
Open air motor unit:
4 pins

Fixation target tungsten lamp:


3 pins
SPOT LED:
4 pins

S RING 1 LED:
2 pins

J309

Measurement LED:
2 pins

SPOT LED:
2 pins

J304

J312

J302

TF motor:
6 pins

Rotary shutter:
2 pins

J307

J305

J310

J308

J306

S RING 2 LED:
2 pins

VR4

VR6

VR2

VR3

VR1

J303
J301

VR5
J311

CONTROL PCB
J3

Volume
VR1: For adjusting the fixation target light
intensity (dark)
VR2: For adjusting the alignment spot light
intensity
VR3: For adjusting the fixation target light
intensity (bright)

VR4: For adjusting the parallel projection spot 1


light intensity
VR5: For adjusting the anterior eye section light
intensity
VR6: For adjusting the parallel projection spot 2
light intensity

2-7-16 CCD camera (41820 6200)

CONTROL PCB
J5 (RM), J8 (CT)

CN2

39

CN1

CONTROL PCB
J7 (RM), J10 (CT)

CN5

CONTROL PCB
J6 (RM), J9 (CT)

R-TRK1P-0-0809-02

3. REPAIR WORK
3-1 Inspection and adjustment procedures
These are the procedures for the adjustment carried out when the customer purchases a new auto keratorefracto-tonometer or before returning one to the customer after repairs. Of particular importance is the need
to check after repairs that new problems have not developed in areas other than that just repaired.

Procedure
25

Self-check

,
64

,

Operations check

59

59
,

Measurement
accuracy check

84 84

62 62

,
,

59

59

62
59

59
7

,

,
,
,

12345678

Timer check

Appearance check

3-1-1 Self-check
[Procedure]
(1) Set the DIP switch No. 3 of SW5 to ON on the CONTROL PCB as shown in the above figure. Turn on
the power of the instrument.
(2) The results and states are displayed on the monitor, so check each indicator according to the instructions.
When NO GOOD is displayed or there is a malfunction, carry out the following procedure and repair
the instrument in accordance with 3-3 Trouble shooting.

40

R-TRK1P-0-0809-02

Self-check initial image

INIT TEST

INIT TEST

1.
2.
3.
4.
5.
6.
7.
8.

9. RS232C TEST
10. PRINT TEST
11. ROTARY TEST
12. TF MOTOR SET
13. TARGET CHECK
14. USB TEST
15. XYZ TEST

FROM TEST
RAM TEST
EEROM USER
EEROM PRTCT
CAPTURE TV
CAPTURE VI
SWITCH TEST
TIME SET
NEXT -> MEAS.SW
EXEC -> PRINT SW

1. FROM TEST

NEXT

->

MEAS.SW

The FROM PCB and the CONTROL PCB address data bus line are checked.

FROM TEST

CHECK SUM
BA4A

Finished by Measurement switch.

2. RAM TEST

The RAM area in the CONTROL PCB is checked.

RAM TEST
IC24
NOW READING...

OK

A8000027:7F:75

41

R-TRK1P-0-0809-02

3. EEROM USER

The EEROM for initial setting in the CONTROL PCB is checked.

EEROM USER

NOW READING...

OK

NEXT -> MEAS.SW

4. EEROM PRTCT

The EEROM for compensation values in the CONTROL PCB is checked.

EEROM PRTCT
IC24
NOW READING...

OK

NEXT -> MEAS.SW


B4001007:4E:15

5. CAPTURE TV

Accumulation of the anterior eye section observed images (including the kerato
image) is checked. (8 images)

CAPTURE TV

BLK 0

* Whether 8 images are taken or not is checked.


If they are taken, it is normal.

NEXT MEAS.SW

42

R-TRK1P-0-0809-02

6. CAPTURE V1

Accumulation of the anterior eye section observed images (including the kerato
image) is checked. (8 images)
The difference from 5. CAPTURE TV is that the integral time is increased by
0.3.

CAPTURE V1

* It is normal if the screen is gradually brighter


because the images are taken.

BLK 0
NEXT MEAS.SW

7. SWITCH TEST

The control circuits of the CONTROL SW PCB, LCD PANEL SW1 PCB and
LCD PANEL SW2 PCB are checked.

F1

F2

F3

SWITCH TEST
F2:
F1:
F3:
SubF4:
SubF3:
SubF2:
SubF1:
SubF0:
DIP SW:3

CHIN_L:
CHIN_H:
MEAS.SW:
HEADER:
PAPER:
CUTTER:

X ZERO:
Y ZERO:
Z ZERO:
Y UP:
Y DOWN:

* It is normal if * is displayed when pressing


the selected switch.
* Keep pressing MEAS. SW until the buzzer
sounds to exit from this TEST.

SubF0 SubF1 SubF2 SubF3 SubF4

43

R-TRK1P-0-0809-02

8. TIME SET

The TIMER control circuit in the CONTROL PCB is checked.

TIME SET

INT

MEAS.SW

NO INIT PRINT.SW

9. RS232C TEST

The input terminal of RS232C, the cable and the RS232C control circuit in the
CONTROL PCB are checked.

RS232C TEST
OUTPUT DATA
INPUT DATA

NEXT -> MEAS.SW

10. PRINT TEST

The PRINTER UNIT is checked.

PRINT TEST

NEXT -> MEAS.SW

* Press MEAS. SW once, and the data are


printed.
Press it once more, and the instrument exits
from PRINT TEST.

44

R-TRK1P-0-0809-02

11. ROTARY TEST

The operation of the open air motor is checked.

ROTARY TEST
ROTARY OFF : The open air motor stops.
ROTARY ON : The open air motor rotates.

ROTARY OFF
PRINT SW -> ROTARY ON
MEAS.SW -> EXIT

12. TARGET CHECK

The target image is checked. (The rotary speed of the open air motor can be
changed.)

TARGET CHECK

T ON:080MS
T OFF:400MS
SubF0:UP,SubF4:DOWN
MEAS.SW:DATA SET/EXIT

13. USB TEST

The USB communication function is checked by using the special software.

USB TEST

ATTACH USB CABLE


NEXT -> MEAS.SW

45

R-TRK1P-0-0809-02

14. XYZ TEST

The operation of the auto-drive unit is checked.

XYZ TEST
X
Y
Z
Y
Y

MOTOR
MOTOR
MOTOR
MOTOR
MOTOR

DRIVE
OK
DRIVE
OK
DRIVE
OK
UP DRIVE
OK
DOWN DRIVE OK

PRINT SW > TEST START


MEAS.SW > EXIT

46

R-TRK1P-0-0809-02

3-1-2 Measurement accuracy check


(1) Fix the test eyes for tools on the chinrest and measure each of them in order, confirming that the values are
within the limits listed below.
(However, when the measured values exceed the accuracy limits, re-correct the setting.)
(2) Also check the diopter value of the accessory test eyes at this time.
Method of appraisal:
a. The typical values should be obtained after measuring the power and axial angle five times.
b. The fluctuation is the fluctuation width after measuring five times.
c. Measure by 0.12 steps.
Spherical test eye
Unit: diopter
Test eye

Spherical power

Fluctuation

Residual astigmatism

Fluctuation

0 to 10

0.25

0.12

0.12

0.12

10 to 15

0.37

0.12

0.12

0.12

-22
+25

0.50

0.12

0.25

0.12

15 to

Note: Use care concerning alignment for the spherical accuracy.


Astigmatic test eye
Unit: diopter
Test eye

Spherical power

Fluctuation

Axial angle

Fluctuation

0.25 to 1

0.25

0.12

to 3

0.25

0.12

to 8

0.25

0.12

47

R-TRK1P-0-0809-02

3-1-3 Checking accuracy of kerato-use measurement


(1) Fix the test eyes for kerato tool on the chinrest and measure each of them in order, confirming that the values are within the limits listed below.
(However, when the measured values exceed the accuracy limits, re-correct the setting.)
(2) Also check the radius of curvature of the accessory test eye at this time.
Method of appraisal:
a. The typical values should be obtained after measuring the radius of curvature and the difference
between H and V axial angles five times.
b. The fluctuation is the fluctuation width after measuring five times.
(However, there should be a separate judgment for H and V.)
Spherical test eye
Unit: mm
Test eye

Radius of curvature

Fluctuation

Difference between H and V

R5 to R10

0.03

0.03

0.03

Astigmatic test eye


Test eye

Radius of curvature

Fluctuation

Astigmatic axial angle

Fluctuation

Radius of curvature:
0.5mm or more

0.05mm

0.03mm

3-1-4 Checking accuracy of tonometer-use measurement


(1) Fix the test eyes for tonometer tool on the chinrest and measure each of them in order, confirming that the
values are within the limits listed below.
(However, when the measured values exceed the accuracy limits, re-correct the setting.)
(2) Checking the intraocular pressure measurement
Method of appraisal:
Align the center of the test eye, measure it ten times and make sure that the measured values are within the
following standard values.
|Reference value - Average of the ten measured values| 2.5
Less than the reference value 30 : 0 - 30mmHg range
Reference value 30 or more
: 0 - 60mmHg range
(3) Checking the accuracy of the corneal thickness
Method of appraisal:
Measure each of the test eyes having five types of thickness five times and make sure that the measured
values are within the following standard values.
Within the reference value 0.01mm (The test eye thickness is less than 0.6mm.)
Within the reference value 0.025mm (The test eye thickness is 0.6mm or more.)

48

R-TRK1P-0-0809-02

3-2 Disassembly and assembly procedures


Removing the chinrest

Illustration

4
8

5
6

Proc.
1

Disassembly
Removal of chinrest

Removal method
1

Remove the cover cap 2 .


Remove the screw 3 .
Remove the chinrest cover 4 .
Remove the chinrest cable 5 .
Remove the screw 6 .
Remove the earth cable 7 .
Remove the four screws 8 .
Remove the chinrest 1 .

49

Point for assembly


Assemble the unit in the
reversed procedure of disassembly.

R-TRK1P-0-0809-02

Removing the covers (1)

9
9

7
6

Illustration

2
5
6
3

4
3

11

10

Proc.
1

Disassembly
Removal of
cover (right) 1

Removal of
cover (left) 4

middle

middle

Removal of front cover


7

Removal of monitor
back cover (rear) 10

Removal method

Point for assembly

Remove the two cover caps 2 .


Remove the two screws 3 .
Remove the middle cover (right)

Remove the two cover caps 5 .


Remove the two screws 6 .
Remove the middle cover (left)
Remove the two cover caps
Remove the two screws 9 .
Remove the front cover 7 .

Remove the four screws 11 .


Remove the monitor back cover

50

Assemble the units in the


reversed procedure of disassembly.

10 .

R-TRK1P-0-0809-02

Removing the covers (2)

13

16

14

Illustration

12

15

16
14

20

17

19

Proc.

Disassembly

Removal of head cover


(right) 12

Remove the two cover caps 13 .


Remove the four screws 14 .
Remove the head cover (right) 12 .

Removal of head cover


(left) 15

Remove the four screws 16 .


Remove the head cover (left)

15 .

Removal of base cover


(right) 17

Remove the three screws 18 .


Remove the base cover (right)

17 .

Removal of base cover


(left) 19

Remove the two screws 20 .


Remove the base cover (left)

Removal method

51

18
Point for assembly
Assemble the units in the
reversed procedure of disassembly.

19 .

R-TRK1P-0-0809-02

Removing the covers (3)

22
21

Illustration

22

22

Proc.

Disassembly

Removal of base cover


(front) 21

Removal method
Remove the three screws 22 .
Remove the base cover (front)

52

Point for assembly


21 .

Assemble the unit in the


reversed procedure of disassembly.

R-TRK1P-0-0809-02

Removing the POWER PCB/switching power supply

1
Illustration

2
6

3
7
4

4
5

Proc.

Disassembly

Removal method

Point for assembly


Assemble the units in the
reversed procedure of disassembly.

Removal
PCB 1

POWER

Remove the cables from the POWER


PCB.
Remove the five screws 2 .
Remove the POWER PCB 1 .

Removal of switching
power supply 3

Remove the four screws 4 .


Remove the POWER PCB mounting
plate 5 .
Remove the cables from the switching
power supply.
Remove the four screws 6 .
Remove the switching power supply 3
and the insulation sheet 7 .

of

53

R-TRK1P-0-0809-02

Removing the motor driver PCB

1
3

Illustration

Proc.

Disassembly

Removal method

Point for assembly

Removal of motor driver


PCB 1

Remove the cables from the motor


driver PCB.
Remove the five screws 2 .
Remove the motor driver fixing plate
3 .
Remove the six screws 4 .
Remove the motor driver PCB 1 .

Assemble the unit in the


reversed procedure of disassembly.

54

R-TRK1P-0-0809-02

Removing the KR unit (1/3)

2
4
1
5
6

Illustration

2
7

Proc.

Disassembly

Removal method

Point for assembly

Removal of outer frame

Remove the anterior eye section illumination LED cable.


Remove the eleven screws 2 .
Remove the outer frame 1 .

Assemble the units in the


reversed procedure of disassembly.

Removal of KR base
plate 3

Remove the snap pin 4 and then


remove the connecting pin 5 .
Remove the two bearing shafts 6 .
Remove the four screws 7 .
Remove the KR base plate 3 .

55

R-TRK1P-0-0809-02

Removing the KR unit (2/3)

9
8

Illustration

11

10

12

13
Proc.

Disassembly

Removal method

Point for assembly

Removal of KR measuring unit 8

Remove the cables from the KR measuring unit.


Remove the four screws 9 .
Remove the KR measuring unit 8 .

Assemble the units in the


reversed procedure of disassembly.

Removal of control PCB

Remove the cables from the control


PCB.
Remove the four screws 11 .
Remove the control PCB 10 .

10

Removal of analog PCB


12

Remove the cables from the analog


PCB.
Remove the four screws 13 .
Remove the analog PCB 12 .

56

R-TRK1P-0-0809-02

Removing the KR unit (3/3)

15
14
19
20

Illustration

19

14

16

17
18

Proc.
6

Disassembly
Removal
of
mounting plate

cover
14

Removal of KR cover
16

Removal of KR unit

18

Removal method

Point for assembly

Remove the two screws 15 .


Remove the KR cover mounting plate
14 .

Assemble the units in the


reversed procedure of disassembly.

Remove the three screws 17 .


Remove the KR cover 16 .
Remove the four screws 19 .
Remove the PCB plate 20 .
Remove the KR unit 18 .

57

R-TRK1P-0-0809-02

Removing the LZM PCB/CCD unit

Illustration

Proc.

Disassembly

Removal method

Point for assembly

Removal of LZM PCB

Remove the cables from the LZM PCB.


Remove the four screws 2 .
Remove the LZM PCB 1 .

Assemble the units in the


reversed procedure of disassembly.

Removal of CCD camera


3

Remove the cables from the CCD camera.


Remove the three screws 4 .
Remove the CCD camera 3 .

58

R-TRK1P-0-0809-02

Removing the CT unit

Illustration

Proc.
1

Disassembly
Removal of CT unit

Removal method
1

Remove the cables from the CT unit.


Remove the three screws 2 .
Remove the CT unit 1 .

59

Point for assembly


Assemble the unit in the
reversed procedure of disassembly.

R-TRK1P-0-0809-02

Removing the head changing sensors

Illustration

6
5
8
7

Proc.

Disassembly

Removal method

Point for assembly

Removal of head changing front limit sensor 1

Remove the two screws 2 .


Remove the head changing front limit sensor 1 .

Assemble the units in the


reversed procedure of disassembly.

Removal of head changing rear limit sensor 3

Remove the two screws 4 .


Remove the head changing rear limit sensor 3 .

Do not remove the sensor


fixing plate.

Removal of head changing upper limit sensor 5

Remove the two screws 6 .


Remove the head changing upper limit
sensor 5 .

Removal of head changing lower limit sensor 7

Remove the two screws 8 .


Remove the head changing lower limit
sensor 7 .

60

R-TRK1P-0-0809-02

Removing the head changing motors

Illustration

5
6

Proc.

Disassembly

Removal method

Point for assembly

Removal of head upand-down motor 1

Remove the two screws 2 .


Remove the three screws 3 .
Remove the head up-and-down motor
1 .

Assemble the units in the


reversed procedure of disassembly.

Removal of head backand-forth motor 4

Remove the two screws 5 .


Remove the three screws 6 .
Remove the head back-and-forth motor
4 .

61

R-TRK1P-0-0809-02

Removing the monitor unit

Illustration

Proc.

Disassembly

Removal method

Point for assembly

Removal of monitor unit

Remove the cables from the monitor


unit.
Remove the four screws 2 .
Remove the monitor unit 1 .

Assemble the units in the


reversed procedure of disassembly.

Removal of tilt unit

Remove the four screws


Remove the tilt unit 3 .

62

R-TRK1P-0-0809-02

Removing the monitor covers

Illustration

2
4

Proc.
1

Disassembly
Removal of
cover (rear) 1

monitor

Removal of monitor
cover (front) 3

Removal method
Remove the four screws 2 .
Remove the monitor cover (rear)
Remove the four screws 4 .
Remove the monitor cover (front)

63

Point for assembly

Assemble the units in the


reversed procedure of disassembly.

R-TRK1P-0-0809-02

Removing the LCD

4
2

Illustration

1
6

Proc.

Disassembly

Removal method

Point for assembly

Removal of LCD relay


PCB 1

Remove the two screws 2 .


Remove the LCD relay PCB 1 .
Remove the cables from the LCD relay
PCB.

Assemble the units in the


reversed procedure of disassembly.

Removal of inverter unit

Remove the cables from the inverter unit.


Remove the two screws 4 .
Remove the inverter unit 3 .

Removal of LCD unit

Remove the four screws 6 .


Remove the LCD unit 5 .

64

R-TRK1P-0-0809-02

Removing the LCD panel SW PCB's

Illustration

2
3

5
6

Proc.

Disassembly

Removal method

Point for assembly

Removal of LCD panel


SW PCB 1 1

Remove the cables from the LCD panel


SW PCB 1.
Remove the three key tops 2 .
Remove the two screws 3 .
Remove the LCD panel SW PCB 1 1 .

Assemble the units in the


reversed procedure of disassembly.

Removal of LCD panel


SW PCB 2 4

Remove the cables from the LCD panel


SW PCB 2.
Remove the five key tops 5 .
Remove the two screws 6 .
Remove the LCD panel SW PCB 2 4 .

65

R-TRK1P-0-0809-02

Removing the printer unit/printer PCB

Illustration

Proc.

Disassembly

Removal of printer unit


1

Removal of printer PCB


4

Removal method
Remove the three screws 2 .
Remove the four screws 3 .
Remove the printer unit 1 .
Remove the three screws
Remove the printer PCB

66

5
4

Point for assembly


Assemble the units in the
reversed procedure of disassembly.

.
.

R-TRK1P-0-0809-02

Removing the control lever unit

Illustration

Proc.

Disassembly

Removal method

Point for assembly

Removal of control lever


unit 1

Remove the cables from the control


lever unit.
Remove the three screws 2 .
Remove the control lever unit 1 .

Assemble the unit in the


reversed procedure of disassembly.

67

R-TRK1P-0-0809-02

Removing the changing mechanism unit

Illustration

Proc.

Disassembly

Removal method

Point for assembly

Removal of changing
mechanism unit 1

Remove the cables from the changing


mechanism unit.
Remove the four screws 2 .
Remove the changing mechanism unit
1 .

Assemble the unit in the


reversed procedure of disassembly.

68

R-TRK1P-0-0809-02

Removing the auto-drive unit

Reference surface for mounting

Illustration

Reference surface for mounting

Proc.

Disassembly

Removal method

Point for assembly

Removal of auto-drive
unit 1

Remove the cables from the auto-drive


unit.
Remove the five screws 2 .
Remove the auto-drive unit 1 .

Assemble the unit in the


reversed procedure of disassembly.
When mounting the unit,
push it against the reference surface and fix it.

69

R-TRK1P-0-0809-02

Removing the Y drive unit

2
Illustration

Proc.

Disassembly

Removal method

Point for assembly

Removal of Y drive unit

Remove the cables from the Y drive


unit.
Remove the four screws 2 .
Remove the Y drive unit 1 .

Assemble the unit in the


reversed procedure of disassembly.

70

R-TRK1P-0-0809-02

Removing the Y drive motor

1
Pull the motor with 24.5N.

Illustration

6
3

Proc.

Disassembly

Removal of Y drive
motor 1

Removal method

Remove the four screws 2 .


Loosen the two set screws 3 .
Remove the pulley 4 .
Remove the motor fixing plate
Remove the four screws 6 .
Remove the Y drive motor 1 .

71

Point for assembly


Assemble the unit in the
reversed procedure of disassembly.
5

.
When mounting the Y
drive motor, hang the timing belt on the pulley and
pull the motor with 24.5N,
using a spring balance.
Under this condition, fix
the Y drive motor.

R-TRK1P-0-0809-02

Illustration

Removing the upper/lower limit sensors/center sensor

1
2

6
4
5

Proc.

Disassembly

Removal of upper limit


sensor 1

Remove the two screws 2 .


Remove the upper limit sensor

Removal of lower limit


sensor 3

Remove the two screws 4 .


Remove the lower limit sensor

Removal of center limit


sensor 5

Remove the two screws 6 .


Remove the center sensor 5 .

Removal method

72

Point for assembly

Assemble the units in the


reversed procedure of disassembly.

R-TRK1P-0-0809-02

Illustration

Removing the nozzle unit

1
2

Special tool

Proc.

Disassembly

Removal method

Point for assembly

Removal of nozzle unit

Turn the special tool counterclockwise


and remove the retainer ring 2 .
Pull out and remove the nozzle unit 1 .

Assemble the unit in the


reversed procedure of disassembly.

73

R-TRK1P-0-0809-02

Illustration

Removing the CCD camera

Special tool

Proc.

Disassembly

Removal method

Point for assembly

Removal of CCD camera

Remove the cables from the CCD camera.


Remove the three screws 2 .
Remove the CCD camera unit 1 .

Assemble the unit in the


reversed procedure of disassembly.

74

R-TRK1P-0-0809-02

Removing the sensor PCB's (1/2)

Illustration

2
1
3
4
Proc.

Disassembly

Removal method

Point for assembly

Removal of Z SENSOR
PCB 1

Remove the cables from the Z SENSOR


PCB.
Remove the three screws 2 .
Remove the Z SENSOR PCB 1 .

Assemble the units in the


reversed procedure of disassembly.

Removal of LINE CCD


PCB 3

Remove the cables from the LINE CCD


PCB.
Remove the four screws 4 .
Remove the LINE CCD PCB 3 .

Removal of APPLE SENSOR PCB 5

Remove the five screws 6 .


Remove the shield plate 7 .
Remove the cables from the APPLE
SENSOR PCB.
Remove the two screws 8 .
Remove the APPLE SENSOR PCB 5 .

75

R-TRK1P-0-0809-02

Illustration

Removing the sensor PCB's (2/2)

Proc.
4

Disassembly
Removal of
SENSOR PCB

PRESS
1

Removal method

Point for assembly

Remove the cables from the PRESS


SENSOR PCB.
Remove the three screws 2 .
Remove the PRESS SENSOR PCB 1 .

Assemble the unit in the


reversed procedure of disassembly.

76

R-TRK1P-0-0809-02

Removing the LED unit

2
3
1
4

Illustration

5
6

Proc.

Disassembly

Removal of FIX LED


UNIT 1

Remove the two screws 2 .


Remove the FIX LED UNIT

Removal of XY LED
UNIT 3

Remove the two screws 4 .


Remove the XY LED UNIT

Removal of PACHO LED


UNIT 5

Remove the two screws 6 .


Remove the PACHO LED UNIT

Removal of Z LED UNIT

Remove the two screws 8 .


Remove the Z LED UNIT 7 .

Removal method

77

Point for assembly

Assemble the units in the


reversed procedure of disassembly.

When mounting the LED,


push it in the arrow direction.

R-TRK1P-0-0809-02

Removing the cylinder unit

Tool hole

Illustration

Tool hole

Proc.

Disassembly

Removal method

Removal of cylinder

Removal of piston unit


3

Remove the four screws


Remove the cylinder 1 .

Point for assembly


Assemble the units in the
reversed procedure of disassembly.

Using the tool hole made on the connecting shaft, turn the piston unit 3 counterclockwise to remove it.

78

R-TRK1P-0-0809-02

3-3 Trouble shooting


1. Troubles related with the power supply
The POWER lamp does not light up.
The LCD is not bright enough.

2. Troubles related with the LCD


Nothing is displayed.
The screen is moving. (Synchronization is impossible.)
The anterior eye section observed image is not displayed.

3. Troubles related with measurement


KR
NO TARGET is displayed.
OVER-SPH is displayed.
OVER-CYL is displayed.
ERROR is displayed.
OVER-R is displayed.
NO CENTER is displayed.
CT
The air does not blow automatically.
The data with ( ) or ERROR is displayed.
The measured value is not correct.
There is fluctuation between the measured values.
OVER is displayed.

4. Troubles related with printer


No printing
PAPER END is kept displayed.
CLOSE PRINTER COVER... is kept displayed.
The printer is noisy in operation.

5. Others
The fixation target is not visible.
The PD value is not printed (or is not correct).
The chinrest does not operate.
KR and CT are not changed to each other.
The auto-drive unit does not operate.
NG+ or NG- is displayed when checking air.
Initial operation is not done. (The message is displayed during the initial operation.)
The head does not operate though you tilt the control lever.
The driving speed is not changed though you slide the control lever.
The instrument keeps driving though you do not operate the control lever.
79

R-TRK1P-0-0809-02

The function switches do not work.


The chinrest up/down switch does not work.
The measurement switch does not work.

80

R-TRK1P-0-0809-02

3-3-1 Troubles related with the power supply


The POWER lamp does not light up.

+3.3V is not outputted.

CONTROL SW PCB

POWER PCB
62
5+5&
.'&

.'&

8

&)

8
8
&)
&)
&)

POWER LAMP

,

,
8

,

CHECK POINT
Check if LED 5 of the POWER
PCB lights up and the voltage
(TP6: +3.3V).

When it is OK: Check the wiring to the control SW


PCB.
When it is NG: Check the wiring or check the switching
power supply voltage (+24V).

81

R-TRK1P-0-0809-02

The LCD is not bright enough.

There is not the power supply (+12V) to the LCD.

POWER PCB

CONTROL PCB
62

.'&

8
8
8
&)
&)
&)

,

8
&)
8
8
8
&)
8
%%&

,

+55.(
.'&

8
8
&)
&)

TO LCD

,

,

CHECK POINT
1

Check No. 4, 5, 6 (+24V), GND


(+24V 5%) of No. 1, 2, 3 on
J601 and if LED 1 lights up.
Check the voltage (+24V)
between No. 3 and No. 6 on J603.

When it is NG: Check the wiring.


Replace the switching power supply.

Check if LED 1 of the


CONTROL PCB lights up and the
voltage (TP10: +12V).

When it is NG: Check the wiring of the CTPL-POWER


cable. Replace the CONTROL PCB.
When it is OK: Replace the LCD.

82

R-TRK1P-0-0809-02

3-3-2 Troubles related with the LCD


Nothing is displayed.

There is not the power supply (+15V) to the CCD camera.

The screen is moving.

The connection of the CCD CLOCK cable 1 unit and the camera
image cable unit is faulty or the CONTROL PCB is faulty.

POWER PCB

CONTROL PCB

,

,

&)
&)

,

8
&)

TO CCD (KR)

8
&)

TO CCD (CT)

.'&

8
8
&)
&)
&)

,

&)

8

+55.(

8
8
&)
8
%%&

&)

62

8

&)

8
&)

,

&)

CHECK POINT
Nothing is displayed.
1

Check No. 4, 5, 6 (+24V), GND


(+24V 5%) of No. 1, 2, 3 on
J601 and if LED 3 lights up.
Check the voltage (TP7: +15V).

When it is NG: Check the wiring.


Replace the switching power supply.

Check the voltage (+15V)


between No. 1 and No. 2 on J7
(KR) or on J10 (CT).

When it is NG: Check the wiring.


Replace the CONTROL PCB.

The screen is moving.


1

KR: Check the CCD CLOCK


cable 1 unit.
CT: Check the CCD CLOCK
cable 2 unit.

When it is OK: Replace the CONTROL PCB.

KR: Check the camera image


cable unit.
CT: Check the VM Cable Assy.

When it is OK: Replace the CONTROL PCB.

KR: Check the LCD-CONTROL


cable 1 unit.
CT: Check the LCD-CONTROL
cable 2 unit.

When it is OK: Replace the CONTROL PCB.

83

R-TRK1P-0-0809-02

The anterior eye section observed


image is not displayed.

The ANALOG PCB is faulty, the CONTROL PCB is


faulty or the CCD camera is faulty.

OPTICAL UNIT

ANALOG PCB

,

,





,

,





,

LZM PCB
,

52 .'& 2%$

IRIS
ROTARY SOLENOID

POWER PCB

ANTERIOR EYE
SECTION
ILLUMINATION
LED PCB

CCD
CAMERA

,

KR unit
* The LEDs and solenoids are controlled by SOFTWARE and are turned ON/OFF by the LZM PCB.
* The rotary solenoid closes the light path (iris) only at the moment of measurement.
(Normally, the light path is opened.)

CHECK POINT
The anterior eye section observed image is not displayed.
KR unit
1 Check the rotary solenoid operation
with the LED lighting tool.
2

Check if the anterior eye section


illumination LED unit lights up.

When it is OK: Replace the CONTROL PCB.


When it is NG: Replace the anterior eye section
illumination LED unit or LZM PCB.

Turn on the power of the instrument,


light the objective lens of the optical
head with a penlight, etc. and check
if the LCD senses the light.

When it is NG: Replace the CCD camera.

CT unit
1 Check if the anterior eye section
illumination LED unit lights up.

Turn on the power of the instrument,


light the objective lens of the optical
head with a penlight, etc. and check
if the LCD senses the light.

When it is OK: Replace the CONTROL PCB.


When it is NG: Replace the anterior eye section
illumination LED unit or ANALOG
PCB.
When it is NG: Replace the CCD camera.

84

R-TRK1P-0-0809-02

3-3-3 Troubles related with measurement


KR
NO TARGET is
displayed.

The objective lens in the optical head is dirty.


The measurement LED does not light.
The control of the CCD camera unit is faulty.
The image processing circuit is faulty.

OVER-SPH is
displayed.

The objective lens in the optical head is dirty.


The operation of the moving section in the optical head is faulty.
The operation of TF MOTOR is faulty.
The rotary solenoid operation is faulty.
The image processing circuit is faulty.

OVER-CYL is
displayed.

The objective lens in the optical head is dirty.


The rotary solenoid operation is faulty.
The image processing circuit is faulty.

ERROR is displayed.

The objective lens in the optical head is dirty.


The measurement LED does not light up.
The control of the CCD camera unit is faulty.

OVER-R is
displayed.

The objective lens in the optical head is dirty.


The operation of the moving section in the optical head is faulty.
The image processing circuit is faulty.

NO CENTER is
displayed.

The objective lens in the optical head is dirty.


The kerato LED and the parallel projection LED do not light up.
The CCD camera unit is faulty.
The image processing circuit is faulty.

LZM PCB

CONTROL PCB

MEASUREMENT LED
,

ON/OFF BY
SOFTWARE

CPU
CONTROL

, SIGNAL



,



IMAGE
PROCESSING
CIRCUIT
,
D/A
CONVERSION





LCD

CONTROL

, SIGNAL

GATE ARRAY

SDRAM A/D
CONVERSION





CCD
IMAGE CAMERA
, SIGNAL

85

R-TRK1P-0-0809-02

CHECK POINT
NO TARGET is displayed.
1

Check if the objective lens in the


optical head is dirty.

Check the M LED PCB with the


LED lighting tool.

When it is lit:
The LZM PCB is faulty. Replace it.
When it is not lit: The M LED PCB is faulty.
Return the optical head to the maker
for repair.

Self-check CAPTURE (VI mode)

When it is OK: The image processing circuit is faulty.


The CONTROL PCB is faulty.
So, replace the PCB.
When it is NG: The contact of the CCD CLOCK
cable 1 unit is faulty or the CCD
camera is faulty.

OVER-SPH is displayed.
1

Check if the objective lens in the


optical head is dirty.

Check the operation of the moving


section in the optical head.

Self-check FROM, RAM CHECK

Replace the LZM PCB.

When it is OK: Adjust the REF accuracy.


When it is NG: The CONTROL PCB is faulty.
Replace it.

OVER-CYL is displayed.
1

Check if the objective lens in the


optical head is dirty.

Self-check FROM, RAM CHECK

When it is OK: Adjust the REF accuracy.


When it is NG: The CONTROL PCB is faulty.
Replace it.

ERROR is displayed.
1

Check if the objective lens in the


optical head is dirty.

If there is unevenness in the light intensity, the optical


head is faulty. Return it to the maker for repair.

Check the M LED PCB with the


LED lighting tool. (Check if the
measured TAGET image has
unevenness in the light intensity.)

When it is NG: The CONTROL PCB is faulty.


Replace it.
The LZM PCB is faulty. Replace it.

Self-check FROM, RAM CHECK

When it is OK: The image processing circuit is faulty.


The CONTROL PCB is faulty.
So, replace the PCB.

Self-check CAPTURE (VI mode)

When it is NG: The contact of the CCD CLOCK


cable 1 unit is faulty or the CCD
camera is faulty.

86

R-TRK1P-0-0809-02

OVER-R is displayed.
1

Check if the objective lens in the


optical head is dirty.

Check the operation of the moving


section in the optical head.

Self-check FROM, RAM CHECK

Replace the LZM PCB.

If there is unevenness in the light intensity, the optical


head is faulty. Return it to the maker for repair.

When it is OK: Adjust the REF accuracy.


When it is NG: The CONTROL PCB is faulty.
Replace it.

NO CENTER is displayed.
1

Check if the objective lens in the


optical head is dirty.

If there is unevenness in the light intensity, the optical


head is faulty. Return it to the maker for repair.

Check the M LED PCB with the


LED lighting tool. (Check if the
measured TAGET image has
unevenness in the light intensity.)

When it is NG: The CONTROL PCB is faulty.


Replace it.
The LZM PCB is faulty. Replace it.

Self-check FROM, RAM CHECK

When it is OK: The image processing circuit is faulty.


The CONTROL PCB is faulty.
So, replace it.

Self-check CAPTURE (VI mode)

When it is NG: The contact of the CCD CLOCK


cable 1 unit is faulty or the CCD
camera is faulty.

CT
The air does not blow
automatically.

The nozzle window is dirty.


Readjust the alignment sensitivity circuit.

ANALOG PCB

Z SENSOR PCB

CHECK POINT
The air does not blow automatically.
1
2

The nozzle window is dirty.


Readjust the alignment sensitivity
circuit.

When it is NG: The CONTROL PCB, XY LED PCB


or Z sensor PCB is faulty.
CCD is faulty.
87

R-TRK1P-0-0809-02

The data with ( ) or ERROR is


displayed.

APPLANATION
SENSOR PCB

ANALOG PCB

The nozzle or nozzle mirror is dirty.


Readjust the applanation detection circuit sensitivity.
The ANALOG PCB or applanation sensor is faulty.
The nozzle is faulty.

Peak voltage

When jetting the air as using the test eye, etc., the peak
voltage of the TP201-TP407 applanation waveform is
as follows.
Data with ( )
: 30 range 2.5V
ERROR display : 60 range 4.0V

CHECK POINT
The data with ( ) or ERROR is displayed.
1

The nozzle or nozzle mirror is dirty.

Check the applanation peak voltage.


(TP201-TP407 of ANALOG PCB)

When the voltage is low:


Readjust the applanation detection circuit
sensitivity.
When the adjustment is impossible:
Measure the light intensity of the alignment LED.
When the light intensity is OK:
The ANALOG PCB or applanation sensor is
faulty. Replace it.

Check the peak applanation voltage.

When the applanation voltage in the periphery is


higher than the center, the nozzle is faulty. Replace it.

88

R-TRK1P-0-0809-02

The measured value is not correct.

The nozzle or nozzle mirror is dirty.


Readjust the intraocular pressure value conversion.
Readjust the reference voltage of the pressure detection
circuit.
Adjust the pressure with the ANALOG PCB or
CONTROL PCB.

CHECK POINT
The measured value is not correct.
1

The nozzle or nozzle mirror is dirty.

Clean the periphery of the nozzle.

Is the test eye for checking or


repairing dirty or deteriorated?

Clean the surface.

Readjust the intraocular pressure


value conversion.

Readjust the reference voltage of the


pressure detection circuit.
Adjust the pressure with the
CONTROL PCB (tilt).

There is fluctuation between the


measured values.

Replace the nozzle.

CHECK POINT
There is fluctuation between the measured values.
1

The nozzle or nozzle mirror is dirty.

Clean the periphery of the nozzle.

Is the test eye for checking or


repairing dirty or deteriorated?

Clean the surface.

Check if the nozzle is bent.

Replace the nozzle.

89

R-TRK1P-0-0809-02

3-3-4 Troubles related with printer


No printing

The PRINTER unit is faulty.

PAPER END is kept displayed.

The PRINTER unit is faulty.


The CONTROL PCB is faulty.

CLOSE PRINTER COVER is


kept displayed.

The PRINTER unit is faulty.


The CONTROL PCB is faulty.

The printer is noisy in operation.

The PRINTER unit is faulty.


The PRINTER ADJUST setting is faulty.

POWER PCB

PRINTER

PRINTER PCB
,

8
&)

,


,

,


,


*5
25









* PS (PAPER END) SIG. When a paper does not exist, it is HIGH.


* HS (CLOSE PRINTER COVER) SIG. When HEAD UP is done, it is HIGH.

CHECK POINT
No printing
1
2

Replace the PRINTER unit.


Replace the CONTROL PCB.

PAPER END is kept displayed.


Check the No. 22 connector on J644.
* PS (PAPER END sensor output)
Paper exists = LOW
Paper does not exist = HIGH

When the PS output is normal, the CONTROL PCB is


faulty. Replace it.
When the PS output is not normal, the PRINTER unit
is faulty. Replace it.

CLOSE PRINTER COVER is kept displayed.


Check the No. 26 connector on J644.
* HS (PAPER END sensor output)
HEAD DOWN = LOW
HEAD DOWN = HIGH

When the HS output is normal, the CONTROL PCB


is faulty. Replace it.
When the HS output is not normal, the PRINTER unit
is faulty. Replace it.

The printer is noisy in operation.


Check the value of PRINTER
ADJUST.

When the set value is 25000, the PRINTER unit is


faulty. Replace it.
When the set value is others except 25000, set the
value again and then check it again.
* If resetting is not possible or if the printer is still
noisy after resetting, the POWER PCB is faulty.
Replace it.
90

R-TRK1P-0-0809-02

3-3-5 Others
The fixation target is not visible.

The KOSHI PCB is faulty.


The LZM CABLE, CONTROL PCB or LZM PCB is
faulty.

ON/OFF BY
SOFTWARE

CHECK POINT
The fixation target is not visible.
1

Check the KOSHI PCB with the


LED lighting tool.

Adjust and check the light intensity


of the fixation target.

The PD value is not printed (or is


not correct).

When it is OK: The LZM CABLE or CONTROL


PCB is faulty. Replace it.
When it is NG: The KOSHI PCB is faulty.
Replace it.

The operation of the auto-drive unit is faulty.


The CONTROL PCB is faulty.
The POWER PCB is faulty.
The MOTOR DRIVER PCB is faulty.

CHECK POINT
The PD value is not printed.
1

Check the operation of the automechanism unit.

When it is OK: Check the CONTROL PCB, POWER


PCB or MOTOR DRIVER PCB.
When it is NG: Check the auto-drive unit.

91

R-TRK1P-0-0809-02

The chinrest does not operate.

Replace the POWER PCB or the chinrest/forehead rest


unit.

CHINREST PCB

POWER PCB

,

,

,

,

,

CONTROL SW PCB

CHECK POINT
The chinrest does not operate.
1

Check the J614 and J641 connectors.

Check the power supply voltage of


the +24V connector.

When the voltage is not outputted, replace the


POWER PCB. When the voltage is outputted, replace
the electric chinrest unit.

92

R-TRK1P-0-0809-02

KR and CT are not changed


to each other.

The CONTROL PCB, the FB/UD MOTOR or the sensor


is faulty.

467& /16
467& /16

46($ /16

+%
8TGH

46($ 176
46($ 176

6#()

($ /1614 #55;

,

+%
8TGH

467& 176
467& 176
,

&)

46($ /16

CONTROL PCB

6#()

7& /1614 #55;

8

%6-4 (4 5'0514

,

&)
8

%6-4 $# 5'0514

,

&)
8

%6-4 72 5'0514

,

&)
8

,

%6-4 &0 5'0514

&)

CHECK POINT
KR and CT are not changed to each other.
1

Check the connectors of J18, J21,


J19, J20, J22 and J23.

Check if the motor is disconnected.

Check if the sensor is disconnected.

Check the signals of the FB MOTOR


J18 and of the UD MOTOR J21.

When it is NG: The motor is faulty. Replace it.

Check the sensor signal.

When it is NG: The sensor is faulty. Replace it.

93

R-TRK1P-0-0809-02

The auto-drive unit does not


operate.

Replace the POWER PCB, MOTOR DRIVER or


CONTROL PCB.
Replace the X motor, Y motor, Z motor or sensor.
Replace the cable.

MOTOR DRIVER PCB

POWER PCB
8&&

: /1614

62

62

,

62

8

,

< /1614

62

8&&

,

8
8
8
&)
&)
&)

8&&

,

&)

62

&)

; /1614

%6-4 &0 5'0514

,

,

%6-4&05'0514

,

8&&

&)

62















,




; #:+5 722'4 .+/+6 5'0514

,

&)

8&& 62

8

8&&

,

8%%
176
)0&

: #:+5 <'41 &'6'%6+10 5'05'4

&)

,

&)

8&& 62

8&&

,

8%%
176
)0&

< #:+5 <'41 &'6'%6+10 5'05'4

&)
&)

CHECK POINT
The auto-drive unit does not operate.
1

Check XYZ TEST in self-check.

When NG is displayed, check the wiring for the


relevant motor and sensor. Replace the POWER
PCB. Replace the MOTOR DRIVER PCB.

Check the sensor signal.


X sensor: TP23
Y sensor: TP22
Z sensor: TP24

When the signal is not changed, the relevant sensor is


faulty. Replace it.
When the signal is changed, the POWER PCB is
faulty. Replace it.

94

R-TRK1P-0-0809-02

NG+ or NG- is displayed


when checking air.

The analog PCB or pressure sensor PCB is faulty.


Replace it.

51.'01+& 70+6

8

%10641. 2%$

8





24'55 5'0514 2%$

#0#.1) 2%$

CHECK POINT
NG+ is displayed when checking air.
1
2

Clean the nozzle.


Check the waveform between TP205 and TP406 (GND) on the analog PCB.
When the waveform is normal:
Check the reference voltage of the analog circuit:
The voltage between TP201 and TP407 (GND): 4.75V 0.02V
When the voltage is normal: Replace the analog PCB.
When the waveform shifts:
Remove the cables from J204.
Check the offset voltage of the pressure detection circuit.
The voltage between TP204 and TP407 (GND): 0.00V 0.02V (VR2)
Connect the cables to J204.
Check the offset voltage of the pressure sensor.
The voltage between TP204 and TP407 (GND): 0.0V 0.1V (VR1)
When resetting cannot be done after adjusting: Replace the analog PCB.
When peak of waveform is +2.8V or more:
Adjust the sensitivity of the pressure detection circuit.
When you cannot adjust the sensitivity: Replace the analog PCB or pressure sensor PCB.
When the sensitivity is normal, it is probable that there is a problem at the solenoid drive side.
Perform the cutoff adjustment for the 30 [mmHg] pressure rising curve.
When resetting cannot be done after adjusting: Replace the analog PCB.

95

R-TRK1P-0-0809-02

NG- is displayed when checking air.


1
2
3
4

Check if the piston is caught.


Check if there is 24V for the condenser.
Check if there is 24V for the control PCB.
Adjust the sensitivity of the pressure detection circuit.
When you cannot adjust the sensitivity: Replace the analog PCB or pressure sensor PCB.
When the sensitivity is normal, it is probable that there is a problem at the solenoid drive side.
Perform the cutoff adjustment for the 30 [mmHg] pressure rising curve.
When resetting cannot be done after adjusting: Replace the analog PCB.

Initial operation is not done. (The message is displayed during the initial operation.)
(a) NOT DETECT FRONT SIGNAL CALL SERVICE PERSON OR SWITCH ON POWER
SUPPLY AGAIN.
(b) NOT DETECT BACK SIGNAL CALL SERVICE PERSON OR SWITCH ON POWER
SUPPLY AGAIN.
(c) NOT DETECT UP SIGNAL CALL SERVICE PERSON OR SWITCH ON POWER SUPPLY
AGAIN.
(d) NOT DETECT DOWN SIGNAL CALL SERVICE PERSON OR SWITCH ON POWER
SUPPLY AGAIN.
(e) X INIT NG SORRY CANNOT DETECT ORG POSITION CALL SERVICE PERSON
(f) Y INIT NG SORRY CANNOT DETECT ORG POSITION CALL SERVICE PERSON
(g) Z INIT NG SORRY CANNOT DETECT ORG POSITION CALL SERVICE PERSON

CONTROL PCB


46($ /16

&)

46($ /16

+%
8TGH

46($ 176
46($ 176

6#()

467& /16

FB MOTOR ASSY

,

+%
8TGH

467& 176
467& 176
,

467& /16
6#()

UD MOTOR ASSY

8

CT/KR FR SENSOR

,
&)
8

CT/KR BA SENSOR

,
&)
8

CT/KR UP SENSOR

,
&)
8

CT/KR DN SENSOR

,
&)

&)

,



,



8
%%&

8
8

8
&)
8

,

,

8&&

X MOTOR

62

62

Z MOTOR

,

62

8

8
8
&)

8&&

,

&)
&)

,

8&&

,

&)

62

&)

Y MOTOR

CT/KR DN SENSOR

8

,

62

CT/KR DN SENSOR

,

8&&

,

















,




&)

62

Y AXIS UPPER LIMIT SENSOR

,
&)
8

8&& 62

8&&

,

8%%
176
)0&

8%%
176
)0&

X AXIS ZERO DETECTION


SENSOR

&)

,

&)

8&& 62

8&&

Z AXIS ZERO DETECTION


SENSOR

,
&)
&)

MOTOR DRIVER PCB

POWER PCB

96

R-TRK1P-0-0809-02

CHECK POINT
(a) NOT DETECT FRONT SIGNAL CALL SERVICE PERSON OR SWITCH ON POWER SUPPLY AGAIN.
1

Check the connectors J18 and J19 on


the CONTROL PCB.

Check if the motor or sensor is


disconnected.

Check the sensor signal.


When occluded (3.3)/When opened
(0V)

When the signal is not changed: Replace the sensor.

Check the back-and-forth changing


motor voltage.
Check the voltage of J18 on the
CONTROL PCB.

When it is NG: Replace the CONTROL PCB.

(b) NOT DETECT BACK SIGNAL CALL SERVICE PERSON OR SWITCH ON POWER SUPPLY AGAIN.
1

Check the connectors J18 and J20 on


the CONTROL PCB.

Check if the motor or sensor is


disconnected.

Check the sensor signal.


When occluded (3.3)/When opened
(0V)

When the signal is not changed: Replace the sensor.

Check the back-and-forth changing


motor voltage.
Check the voltage of J18 on the
CONTROL PCB.

When it is NG: Replace the CONTROL PCB.

(c) NOT DETECT UP SIGNAL CALL SERVICE PERSON OR SWITCH ON POWER SUPPLY AGAIN.
1

Check the connectors J21 and J22 on


the CONTROL PCB.

Check if the motor or sensor is


disconnected.

Check the sensor signal.


When occluded (3.3)/When opened
(0V)

When the signal is not changed: Replace the sensor.

Check the up-and-down changing


motor voltage.
Check the voltage of J21 on the
CONTROL PCB.

When it is NG: Replace the CONTROL PCB.

97

R-TRK1P-0-0809-02

(d) NOT DETECT DOWN SIGNAL CALL SERVICE PERSON OR SWITCH ON POWER SUPPLY AGAIN.
1

Check the connectors J21 and J23 on


the CONTROL PCB.

Check if the motor or sensor is


disconnected.

Check the sensor signal.


When occluded (3.3)/When opened
(0V)

When the signal is not changed: Replace the sensor.

Check the back-and-forth changing


motor voltage.
Check the voltage of J21 on the
CONTROL PCB.

When it is NG: Replace the CONTROL PCB.

(e) X INIT NG SORRY CANNOT DETECT ORG POSITION CALL SERVICE PERSON
1

Check the connectors of J631, J632


and J633 on the MOTOR DRIVER
PCB and the connector of J617 on
the POWER PCB.

Check if the motor or sensor is


disconnected.

Check the sensor signal.


When occluded (5V)/When opened
(0V)

When the signal is not changed: Replace the sensor.

Check the X axis motor signal.


Check if the signal is changed near
the X axis center with TP23 and TP1
(GND) on the POWER PCB.

When it is NG: Replace the POWER PCB.

(f) Y INIT NG SORRY CANNOT DETECT ORG POSITION CALL SERVICE PERSON
1

Check the connectors of J631, J632


and J633 on the MOTOR DRIVER
PCB and the connector of J619 on
the POWER PCB.

Check if the motor or sensor is


disconnected.

Check the sensor signal.


When occluded (5V)/When opened
(0V)

When the signal is not changed: Replace the sensor.

Check the Y axis motor signal.


Check if the signal is changed near
the Y axis center with TP22 and TP1
(GND) on the POWER PCB.

When it is NG: Replace the POWER PCB.

98

R-TRK1P-0-0809-02

(g) Z INIT NG SORRY CANNOT DETECT ORG POSITION CALL SERVICE PERSON
1

Check the connectors of J631, J632


and J633 on the MOTOR DRIVER
PCB and the connector of J621 on
the POWER PCB.

Check if the motor or sensor is


disconnected.

Check the sensor signal.


When occluded (5V)/When opened
(0V)

When the signal is not changed: Replace the sensor.

Check the Z axis motor signal.


Check if the signal is changed near
the Z axis center with TP24 and TP1
(GND) on the POWER PCB. (The
position which is backward by about
15mm from the center)

When it is NG : Replace the POWER PCB.

99

R-TRK1P-0-0809-02

The head does not operate though you tilt the control lever.

POWER PCB
Rough/fine movement
changing sensor unit


2%

8&&
,

62
+% +%


&)

X axis jog sensor unit






84
,

8





8

62

#)

#)

Z axis jog sensor unit

  

&)

,

84

8

#)
8

62

#)

#)

#)

CHECK POINT
The head does not operate though you tilt the control lever.
(a) The head does not operate in the X axis direction.
1

Check if the cable is disconnected.

Check if the X axis jog sensor shaft


is loosened.

Check the X axis jog sensor voltage.


Check if the voltage is changed with
TP11 and TP1 (GND) on the
POWER PCB.

When it is NG: Replace the control lever unit.

(b) The head does not operate in the Z axis direction.


1

Check if the cable is disconnected.

Check if the Z axis jog sensor shaft


is loosened.

Check the Z axis jog sensor voltage.


Check if the voltage is changed with
TP13 and TP1 (GND) on the
POWER PCB.

When it is NG: Replace the control lever unit.

The driving speed is not changed though you slide the control lever.
CHECK POINT
The driving speed is not changed though you slide the control lever.
Check the rough/fine movement changing sensor voltage.
Check if the voltage is changed by sliding the lever with
TP17 and TP2 (GND) on the POWER PCB.
100

When it is NG: Replace the control


lever unit.

R-TRK1P-0-0809-02

The instrument keeps driving though you do not operate the control lever.

CHECK POINT
The instrument keeps driving though you do not operate the control lever.
(a) The instrument keeps moving in the X direction.
Check the voltage of the X axis jog sensor initial
position.
Make sure that the voltage is +0.76 - +0.96V with
the TP11 and TP1 (GND) on the POWER PCB.

When it is NG: Adjust the potential


position.

(b) The instrument keeps moving in the Z direction.


Check the voltage of the Z axis jog sensor initial
position.
Make sure that the voltage is +0.76 - +0.96V with
the TP13 and TP1 (GND) on the POWER PCB.

When it is NG: Adjust the potential


position.

The function switches do not work.

41850 7700
LCD UNIT
      

2       
,       

41850 7550
LCD panel SW cable 1 unit
2 ,
    
    

 &)

    
    

, 2
 8




41850 7510
LCD panel SW PCB1
   ,
   2

41850 7560
LCD panel SW cable 2 unit
   2
   ,

 8

41850 7520
LCD panel SW PCB2

 &)
       ,
       2

41850 7540
LCD-CONTROL cable unit

2       
,       

41850 7100
CONTROL PCB

CHECK POINT
The function switches do not work.
1

Check the following connectors.


J13 on the CONTROL PCB
J511 and J532 on the LCD RELAY
PCB
J511 on the LCD PANEL SW PCB1
J523 on the LCD PANEL SW PCB2

Check if the connecting cable is


disconnected.

When it is OK: Replace the LCD PANEL SW PCB2.

101

R-TRK1P-0-0809-02

The chinrest up/down switch does not work.

CHINREST PCB

POWER PCB

J614

J652

J651

J613

POWER LAMP

J611

CONTRPL SW PCB

CHECK POINT
The chinrest up/down switch does not work.
1

Check if the POWER lamp is lit.

Check the following connectors.


J614 on the POWER PCB
J651 on the CHINREST PCB

Check if the cable is disconnected.

When it is NG: Replace the cable.

Check the voltage of J651 on the


CHINREST PCB.

When it is OK: Replace the chinrest unit.

The measurement switch does not work.

POWER PCB

CONTROL SW PCB

J613

J612

J611

CHECK POINT
The measurement switch does not work.
1

Check the following connectors.


J611 and J612 on the CONTROL
SW PCB
J613 on the POWER PCB

Check if the cable is disconnected.

When it is NG: Replace the cable.

Check the measurement switch.

When it is OK: Replace the CONTROL SW PCB.


102

R-TRK1P-0-0809-02

3-4 Assembly and adjustment procedures


3-4-1 Adjustments and cautions for unit replacements
The following table shows the necessary adjustments and cautions when replacing an electric unit in trouble
shooting, self-check or others.
Unit

Unit No.

Reference
page

Adjustments and cautions

POWER PCB

41850 6200

Power supply voltage adjustment and check

P.105

CCD camera unit


(RM)

41820 6200

Alignment spot focus/position adjustment


Accuracy adjustment
Auto-alignment adjustment

P.106
P.107
P.132-141

CONTROL PCB

41850 7100

P.34
P112-115
P.179
P.180

ANALOG IF PCB

41850 7200

Analog circuit reference voltage adjustment


Removal of offset voltage in the pressure detection
circuit
Pressure sensor offset voltage adjustment
Pressure detection circuit sensitivity adjustment
LED light intensity adjustment for measuring the
corneal thickness
Offset memory for measuring the corneal thickness
Light intensity adjustment of the light source for
XY alignment and applanation
CCD XY alignment offset memory
Light intensity adjustment of the light source for Z
alignment
Applanation detection circuit sensitivity adjustment

P.160
P.161

P.150
P.151-152
P.132-141

Replacement of EEPROM
Character display position adjustment
30 [mmHg] pressure rising curve cutoff adjustment
60 [mmHg] pressure rising curve cutoff adjustment

P.162
P.163
P.167
P.168
P.171
P.173
P.174
P.178

LZM DRIVER PCB

41840 7300

Center spot light intensity adjustment


Fixation target light intensity adjustment
Auto-alignment adjustment

CCD camera unit


(CT)

41820 6200

Position adjustment
P.155-156
CCD magnification adjustment
P.169-170
Light intensity adjustment of the light source for P.171
XY alignment and applanation
P.173
CCD XY alignment offset memory

PRESS SENSOR PCB 41850 7310

Pressure sensor offset voltage adjustment


Applanation detection circuit sensitivity adjustment

P.162
P.178

APPLA SENSOR
PCB

41850 7320

Applanation detection circuit sensitivity adjustment

P.178

Z SENSOR PCB

41850 7330

Z SENSOR PCB position adjustment


Z SENSOR PCB frequency adjustment
Z sensor circuit gain adjustment

P.175
P.176
P.177

LINE CCD PCB

41850 7340

LINE CCD PCB position adjustment


P.164-166
LED light intensity adjustment for measuring the P.167
corneal thickness
Offset memory for measuring the corneal thickness P.168

103

R-TRK1P-0-0809-02

Unit

Unit No.

Adjustments and cautions

Reference
page

XY LED UNIT

41850 7350

Light intensity adjustment of the light source for P.171


XY alignment and applanation

Z LED UNIT

41850 7360

Light intensity adjustment of the light source for Z P.174


alignment
Z sensor PCB position adjustment
P.175

PACHO LED UNIT

41850 7370

LINE CCD PCB position adjustment


P.164-166
LED light intensity adjustment for measuring the P.167
corneal thickness
Offset memory for measuring the corneal thickness P.185-186

104

R-TRK1P-0-0809-02

3-4-2 Power supply voltage adjustment and check


Illustration

Procedure
Power supply voltage adjustment and check
Check if the voltage meeting the standard is outputted at the test pins shown below. Adjust the
voltage.
TEST
PIN
POWER
TP5
PCB
TP6
TP4
TP8
TP7
TP18
TP1
CONTROL TP10
PCB
TP11
TP5
TP9
TP6
PCB

62

,

,

,

62

62
84

62

62

62

,

84

84
,
62

62

84

,
,

,

,

,

,

POWER PCB

,

,

,

62

,

Adjustment
part
VR2
--VR3
VR1
VR4
---------------

Standard
+5.0V 0.2V
+3.3V 0.13V
+15V 0.1V
+7.5V 0.5V
+14.7V - 15.5V
+1.6V 0.1V
GND
+12V 0.5V
-15V 0.5V
+1.5V 0.1V
+1.6V 0.1V
GND

5-

,
+%

62
,

,
62A#&+0
,

62

,

+%

62

62
+%

62

62
62
62

62

,
62

,

62

62

62

62

,

,

,

,

,

CONTROL PCB

105

R-TRK1P-0-0809-02

3-4-3 Alignment spot focus/position adjustment


Illustration

Procedure
1. Mount the tool holder.
2. Set the 0D test eye.
3. Remove the internal cover.
Focus adjustment
Loosen the screw at A of the lens unit.

Lens unit

Alignment
spot focus
adjustment

Focalize the alignment spot of the color LCD monitor by turning B of the lens unit.
* The spot is optimally focused when it is round.
This occurs just as it is about to switch from being
a horizontal oval to a vertical oval.
Tighten the screw at A.

Best focus

Position adjustment
Loosen the three fixing screws of the CCD camera
unit.

Monitor scale center tool


Eye
Color LCD
monitor

Adjust the position by moving the whole CCD camera unit so that the alignment spot may be at the center of the monitor scale center tool.
* Make sure that the position is perpendicular to the
color LCD monitor.
* Make sure visually that the CCD camera unit is
not inclined.
Tighten the three fixing screws securely.

Fixing screw

CCD camera
unit

Make sure that the alignment spot position is correct.


If it is deviated, repeat Procedure ** and after.

106

R-TRK1P-0-0809-02

3-4-4 Accuracy adjustment


The accuracy of TRK-1P can be adjusted with the test eyes for tools, tool holder and the test eyes of 0D and
10D as the conventional RM/KR series.
Accuracy adjustment is necessary when a measurement accuracy is out of the limit.
Perform the operations indicated by asterisks (*) in the table below.

Menu

When SPH. When CYL. When AXS. When autovalue is out of value is out of value is out of alignment
the limit
the limit
the limit
operation is
NG

When KRT. When corneal


value is out of diameter meathe limit
surement is
NG

GRAPHIC ADJUST
CENTER ADJUST
ZERO ADJUST

REF ADJUSTMENT GET S0

GET P2D

GET SPD

SPH LINEARITY

CYL LINEARITY

AXIS ADJUST
KRT ADJUSTMENT
PRALLEL LEVEL

PARALLEL BEND

ALPHA CALLCULATE

ZERO CALLCULATE

Z ALIGNMENT

CURVATURE ADJUST

AXIS ADJUST

CORNEA LENGTH ADJUST

107

R-TRK1P-0-0809-02

1. Checking before adjustment


Illustration

Procedure
1. Remove the instrument cover.

Tool holder

2. Mount the tool holder.

0D test eye

Optical head

3. Set the 0D test eye.

4. Turn on the power switch.

5. Check the following items.


The spot must be in the center of the alignment mark.
The alignment mark and the spot must be optimally focused.

6. Press Measurement switch once.

Alignment mark

Spot

7. Press the target image switch.

8. Observe the measurement target image.


There must be no dust or shadows.

Dust

Shadow

108

R-TRK1P-0-0809-02

2. E2PROM WRITE
The compensation value data are stored in E2PROM on the CONTROL PCB. When replacing the PCB, if
you change this E2PROM, adjustment is not necessary. Write the initial values onto E2PROM when you use
a compensation value which is not written yet.
Illustration

Procedure

25

,
64
62 62

,
,

,

59

59

59

59

59
7

,



62
59

84 84
,

Purpose
Before adjusting the electric units in the KR/RM
system, store the initial compensation values.

,
,

,

1. Set the DIP switch No. 8 of SW5 on the CONTROL PCB to ON. Then, turn on the power.

Set No. 8 to ON.

SW5

Print switch

2. The top page of the compensation menu is displayed.


Make sure that Common is selected. (The
selected item is displayed with yellow letters.)
Press Print switch.

Print switch

3. The contents of the COMMON menu are displayed.


Press Measurement switch and select 5.
E2ROM INITIALIZE. (The selected item is
displayed with yellow letters.)
Press Print switch.

ADJUST LIST
Common
KR
CT
ITEM SELECT -> MEAS SW
EXECUTE
-> PRINT SW

COMMON ADJUST
1.
2.
3.
4.
5.
6.
7.

DESTINATION SET
CHINA OFFSET SET
STAGE OFFSET
BASE CHECK
E2ROM INITIALIZE
PRINTER DURATION
DUMMT01

109

R-TRK1P-0-0809-02

Illustration

Procedure
4. The INITIAL WRITE screen is displayed.
Select KR with Chinrest up/down button.

INITIAL WRITE
ALL
KR
CT
MODEL :TRK-1P
PRINT SW: OK
CHIN SW :UP&DOWN
MEAS SW : CANCEL

5. Set the slide switch SW6 on the CONTROL


PCB to the U side.

25

,
64
62 62

,
,

,

59

59

59

59

84 84
,

62
59

59
7

,

,

,
,

SW6
Writing SW

Print switch

6. Press Print switch to perform initialization.


Make sure that SUCCESS is displayed on the
screen and press Measurement switch.

WRITE : SUCCEESS

PUSH : MAS SW

110

R-TRK1P-0-0809-02

Illustration

Procedure
7. The contents of the COMMON menu are displayed.
Press the right end switch among the function
switches to return to the top page of the compensation menu.

COMMON ADJUST
1.
2.
3.
4.
5.
6.
7.

DESTINATION SET
CHINA OFFSET SET
STAGE OFFSET
BASE CHECK
E2ROM INITIALIZE
PRINTER DURATION
DUMMT01

Right end switch

8. Set the slide switch SW6 on the CONTROL


PCB to the L side.

25

,
64

,

59

84 84

62 62

,
,
59

59

59

62
59

59
7

,

,

,
,
,

9. Set the DIP switch No. 8 of SW5 on the CONTROL PCB to OFF. Then, turn off the power.

SW6


SW5

Set No. 8 to OFF.

111

R-TRK1P-0-0809-02

3. GRAPHIC ADJUST
Illustration

Procedure
Purpose
Set the alignment frame nearly at the center of
the screen.

Tool holder

1. Mount the tool holder and 0D test eye.


0D test eye

Optical head

25

,
64
62 62

,
,

59

59

,

59

59

59
7

,

,

,
,
,

SW6



SW5

84 84

62
59

U
Set No. 8 to ON.

Print switch

KR ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.

2. Set the DIP switch No. 8 of SW5 on the CONTROL PCB to ON. Then, turn on the power.
Set the slide switch SW6 on the CONTROL
PCB to the U side.

3. The top page of the compensation menu is displayed.


Press Measurement switch and select KR.
(The selected item is displayed with yellow letters.)
Press Print switch.

4. The KR compensation menu is displayed.


Press Measurement switch and select 1.
GRAPHIC ADJUST.
Press Print switch.

GRAPHIC ADJUST
CENTER ADJUST
ZERO ADJUST
REF ADJUSTMENT
KRT ADJUSTMENT
LIMIT ADJUST
PRINTER ADJUST
KOSHI ADJUST
AUTO ALING ADJUST

112

R-TRK1P-0-0809-02

Illustration

GRAPHIC ADJUST

Spot
H POSITION:3
V POSITION:11

DOWN

UP

Alignment frame

LEFT

RIGHT RESET

Function switches

Procedure
5. The GRAPHIC ADJUST screen is displayed.
Press the following function switches and make
adjustment so that the spot may be displayed at
the center of the alignment frame.
DOWN switch: Moves the displayed alignment
frame downward.
UP switch:
Moves the displayed alignment
frame upward.
LEFT switch: Moves the displayed alignment
frame leftward.
RIGHT switch: Moves the displayed alignment
frame rightward.
RESET switch: Returns to the initial displayed
status.

6. After adjusting the alignment frame position,


press Measurement switch once.
GRAPHIC ADJUST

H POSITION:3
V POSITION:11

DOWN

UP

LEFT

RIGHT RESET

7. The set values are stored in E2PROM and the


compensation menu is displayed again.
KR ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.

GRAPHIC ADJUST
CENTER ADJUST
ZERO ADJUST
REF ADJUSTMENT
KRT ADJUSTMENT
LIMIT ADJUST
PRINTER ADJUST
KOSHI ADJUST
AUTO ALING ADJUST

113

R-TRK1P-0-0809-02

4. CENTER ADJUST
Illustration

Procedure
Print switch

KR ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.

GRAPHIC ADJUST
CENTER ADJUST
ZERO ADJUST
REF ADJUSTMENT
KRT ADJUSTMENT
LIMIT ADJUST
PRINTER ADJUST
KOSHI ADJUST
AUTO ALING ADJUST

1. Press Measurement switch on the KR compensation menu and select 2. CENTER


ADJUST. Press Print switch

CENTER ADJUST

Spot
H POSITION:477
V POSITION:258

DOWN

UP

Alignment frame

LEFT

RIGHT RESET

Function switches

CENTER ADJUST

H POSITION:477
V POSITION:258

DOWN

UP

LEFT

Purpose
Set the bright spot, which should be the temporary center for measurement, nearly at the center of the alignment frame.

RIGHT RESET

2. The CENTER ADJUST screen appears.


First, set the position in the horizontal direction.
Press the following function switches and make
adjustment so that the spot may be at the center
of the alignment frame in the right-and-left
direction.
LEFT switch: Moves the displayed alignment
frame leftward.
RIGHT switch: Moves the displayed alignment
frame rightward.
RESET switch: Returns to the initial displayed
status.
3. Next, set the position in the vertical direction.
Press the following function switches and make
adjustment so that the spot may be at the center
of the alignment frame.
DOWN switch: Moves the displayed alignment
frame downward.
UP switch:
Moves the displayed alignment
frame upward.
RESET switch: Returns to the initial displayed
status.
4. After adjusting the alignment frame position,
press Measurement switch once.
The set values are stored in E2PROM and the
compensation menu is displayed again.

114

R-TRK1P-0-0809-02

Illustration

Procedure
5. Set the slide switch SW6 on the CONTROL
PCB to the L side.

25

,
64
62 62

,
,

,

59

59

59

59

84 84

62
59

,

59
7

,

,
,
,

6. Set the DIP switch No. 8 of SW5 on the CONTROL PCB to OFF. Then, turn off the power.

SW6


SW5

U
Set No. 8 to OFF.

115

R-TRK1P-0-0809-02

5. ZERO ADJUST
Illustration

Procedure
Purpose
Set the fixation target at the 0D position, using
the diopter telescope.
1. Make sure that the volumes VR1 and VR3 on
the LZM DRIVER PCB have already been
turned fully counterclockwise.

25

,
64
62 62

,
,

59

,

59

59

59

,

59
7

,

,
,
,

SW6



SW5

84 84

62
59

U
Set No. 8 to ON.

Print switch

KR ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.

GRAPHIC ADJUST
CENTER ADJUST
ZERO ADJUST
REF ADJUSTMENT
KRT ADJUSTMENT
LIMIT ADJUST
PRINTER ADJUST
KOSHI ADJUST
AUTO ALING ADJUST

2. Set the DIP switch No. 8 of SW5 on the CONTROL PCB to ON. Then, turn on the power.
Set the slide switch SW6 on the CONTROL
PCB to the U side.

3. The top page of the compensation menu is displayed.


Press Measurement switch and select KR.
(The selected item is displayed with yellow letters.)
Press Print switch.
4. Press Measurement switch on the KR compensation menu and select 3. ZERO ADJUST.
Press Print switch.

116

R-TRK1P-0-0809-02

Illustration

Procedure
5. The ZERO ADJUST screen appears.
Press Print switch once.
Numerals are displayed beside FM.

Print switch

6. As watching the fixation target (Picture chart)


through the diopter telescope, press the [+]
switch and [-] switch among the function
switches to be in best focus.
(Before observation, fit the diopter telescope to
your eyes.)

ZERO ADJUST
1.FM +0.000MM

The value is displayed in yellow


when this is selected.

EXIT

[+] and [-] switches

KR ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.

GRAPHIC ADJUST
CENTER ADJUST
ZERO ADJUST
REF ADJUSTMENT
KRT ADJUSTMENT
LIMIT ADJUST
PRINTER ADJUST
KOSHI ADJUST
AUTO ALING ADJUST

[EXIT] switch

7. After adjusting focus, press Measurement


switch once. The compensation value is written on E2PROM.
8. Press the [EXIT] switch among the function
switches to return to the KR compensation
menu.

117

R-TRK1P-0-0809-02

6. REF ADJUSTMENT (REF accuracy adjustment)


6-1 Accumulation of S0 DATA
Illustration

Procedure
Print switch

KR ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.

GRAPHIC ADJUST
CENTER ADJUST
ZERO ADJUST
REF ADJUSTMENT
KRT ADJUSTMENT
LIMIT ADJUST
PRINTER ADJUST
KOSHI ADJUST
AUTO ALING ADJUST

1. Press Measurement switch on the KR compensation menu and select 4. REF ADJUSTMENT. Press Print switch once.

Print switch

REF ADJUSTMENT
1.
2.
3.
4.
5.
6.

GET
GET
GET
SPH
CYL
REF

Purpose
Measure S0 (ring radius on the CCD camera
at the focus position) and the ring center coordinate.

2. The REF ADJUSTMENT screen appears.


Press Measurement switch and select GET
S0.
Press Print switch once.

S0
P2D
SPD
LINEARITY
LINEARITY
AXIS ADJUST

EXIT

3. The set values of GET S0 are displayed.


Press Measurement switch.
GET S0
S0: +0.869300
H00 REF : 476
V00 REF : 276

EXIT

118

R-TRK1P-0-0809-02

Illustration

Procedure
4. The 0D address setting screen appears.
Make sure that the value in ( ) is displayed in
yellow.
Press the [+] switch and [-] switch among the
function switches to fit the value to the address
of the used test eye 0D.

GET S0
0D (-0.91)

5. After setting, press Measurement switch. The


set address is written on E2PROM.
EXIT

[+] and [-] switches

6. The SET TEST EYE 0D screen appears.


Set the 0D test eye. (Align the marking line.)
7. Press Measurement switch once.
Perform measurement five times.

SET TEST EYE 0D

EXIT

8. After measurement, the results of the GET S0


center coordinate measurement are displayed.
Press Measurement switch once.
The displayed measurement results are written
on E2PROM.
After writing, the REF ADJUSTMENT
screen appears again.

GET S0
S0 :+0.871130
H00 REF : 480
V00 REF : 258

EXIT

119

R-TRK1P-0-0809-02

6-2 Accumulation of P2D DATA


Illustration

Procedure
Print switch

Purpose
Measure P2D [Dptr/pls] (D quantity per
pulse).

REF ADJUSTMENT
1.
2.
3.
4.
5.
6.

GET
GET
GET
SPH
CYL
REF

1. Press Measurement switch on the REF


ADJUSTMENT screen and select GET P2D.
Press Print switch once.

S0
P2D
SPD
LINEARITY
LINEARITY
AXIS ADJUST

2. The set value of GET P2D is displayed.


Press Measurement switch once.
EXIT

3. The -10D/+10D address setting screen


appears.
Make sure that the value in ( ) for -10D is displayed in yellow.
Press the [+] switch and [-] switch among the
function switches to fit the value to the address
of the used test eye -10D.

GET P2D
P2D : 0.066320

EXIT

4. After setting, press Measurement switch. The


set address is written on E2PROM.

5. Make sure that the value in ( ) for +10D is displayed in yellow.


Press the [+] switch and [-] switch among the
function switches to fit the value to the address
of the used test eye +10D.

-10D (-10.97)
+10D (+ 9.00)

6. After setting, press Measurement switch. The


set address is written on E2PROM.

EXIT

[+] and [-] switches

120

R-TRK1P-0-0809-02

Illustration

Procedure
7. The SET TEST EYE -10D screen appears.
Set the -10D test eye. (Align the marking
line.)
8. Press Measurement switch once.
Measurement starts.

SET TEST EYE -10D

EXIT

9. The SET TEST EYE +10D screen appears.


Set the +10D test eye. (Align the marking
line.)
10. Press Measurement switch once.
Measurement starts.

SET TEST EYE +10D

EXIT

11. After measurement, the results of the P2D


measurement are displayed.
Press Measurement switch once.
The displayed measurement results are written
on E2PROM.
After writing, the REF ADJUSTMENT
screen appears again.

P1 : 151
P2 : 480
P2D : 0.060123

EXIT

121

R-TRK1P-0-0809-02

6-3 Accumulation of SPD DATA


Illustration

Procedure
Print switch

Purpose
Measure SPD (deviation of the target image on
CCD with "unit: diopter").

REF ADJUSTMENT
1.
2.
3.
4.
5.
6.

GET
GET
GET
SPH
CYL
REF

1. Press Measurement switch on the REF


ADJUSTMENT screen and select GET SPD.
Press Print switch once.

S0
P2D
SPD
LINEARITY
LINEARITY
AXIS ADJUST

2. The set value of GET SPD is displayed.


Press Measurement switch once.
GET SPD
SPD : -0.019700

EXIT

3. The 0D address setting screen appears.


Make sure that the value in ( ) is displayed in
yellow.
Make sure that the displayed value is fit to the
address of the 0D test eye.

GET SPD
0D (-0.91)

4. After checking, press Measurement switch.


The set address is written on E2PROM.

EXIT

122

R-TRK1P-0-0809-02

Illustration

Procedure
5. The SET TEST EYE 0D screen appears.
Set the 0D test eye. (Align the marking line.)
6. Press Measurement switch once.
Measurement starts.

SET TEST EYE 0D

EXIT

7. After measurement, the results of the SPD calculation are displayed.


Press Measurement switch once.
The displayed measurement results are written
on E2PROM.
After writing, the REF ADJUSTMENT
screen appears again.

SM : 0.843522
SP : 0.823266
SPD: -0.016830

EXIT

123

R-TRK1P-0-0809-02

6-4 SPH LINEARITY


Illustration

Procedure
Print switch

Purpose
Using the spherical test eye, adjust the linearity
and remove the residual astigmatism.

REF ADJUSTMENT
1.
2.
3.
4.
5.
6.

GET
GET
GET
SPH
CYL
REF

1. Press Measurement switch on the REF


ADJUSTMENT screen and select SPH LINEARITY.
Press Print switch once.

S0
P2D
SPD
LINEARITY
LINEARITY
AXIS ADJUST

2. The SPH compensation values are displayed.


Press Measurement switch six times.
SP-D00
SP-D01
SP-D02
SP-D03
SP-D04
SP-D05
SP-D00

:
:
:
:
:
:
:

-11.80411
- 9.37516
- 5.54806
+ 3.39273
+15.34058
+22.69850
+33.61048

EXIT

3. The test eye address setting screen appears.


Make sure that the value in ( ) for -20D is displayed in yellow.
Press the [+] switch and [-] switch among the
function switches to fit the value to the address
of the used test eye -20D.

-20D(-20.07)
-15D(-16.18)
-10D(-10.97)
- 0D(-0.91)
+10D(+9.00)
+15D(+13.81)
+20D(+20.39)

4. After setting, press Measurement switch.


The set address is written on E2PROM.
EXIT

[+] and [-] switches

5. Perform the same operation as Procedure 3 and


4 to fit the values to the addresses of -15D,
-10D, 0D, +10D, +15D and +20D.
The set addresses are written on E2PROM.

124

R-TRK1P-0-0809-02

Illustration

Procedure
6. The "SET TEST EYE -20D" screen appears.
Set the -20D test eye. (Align the marking
line.)
7. Press Measurement switch once.
Measurement starts.
8. After measurement, according to the instructions on the screen, measure the test eyes for
-15D, -10D, 0D, +10D, +15D and
+20D in the same way as Procedure 6 and 7.

SET TEST EYE -20D

EXIT

SP-D00
SP-D01
SP-D02
SP-D03
SP-D04
SP-D05
SP-D00

:
:
:
:
:
:
:

9. After measurement, the calculated compensation values are displayed.


Press Measurement switch once.
The calculated compensation values are written
on E2PROM.
Press Measurement switch six times to return
to the REF ADJUSTMENT screen.

-11.80411
- 9.37516
- 5.54806
+ 3.39273
+15.34058
+22.69850
+33.61048

EXIT

10. Set the slide switch SW6 on the CONTROL


PCB to the L side.

25

,
64

,

,

59

84 84

62 62

,
59

59

59

62
59

59
7

,

,

,
,

11. Set the DIP switch No. 8 of SW5 on the CONTROL PCB to OFF. Then, turn off the power.

,

SW6
 

SW5

Set No. 8 to OFF.

125

R-TRK1P-0-0809-02

6-5 CYL LINEARITY


Illustration

Procedure
Purpose
Using the astigmatic test eye, adjust linearity.

25

,
64
62 62

,
,

59

59

,

59

59

62
59

59

,

,
,
,

,

1. Set the DIP switch No. 8 of SW5 on the CONTROL PCB to ON. Then, turn on the power.
Set the slide switch SW6 on the CONTROL
PCB to the U side.

SW6



SW5

84 84

Set No. 8 to ON.

Print switch

2. The top page of the compensation menu is displayed.


Press Measurement switch and select KR.
(The selected item is displayed with yellow letters.)
Press Print switch.

Print switch

3. The KR compensation menu is displayed.


Press Measurement switch and select 4. REF
ADJUSTMENT.
Press Print switch.

ADJUST LIST
Common
KR
CT
ITEM SELECT -> MEAS SW
EXECUTE
-> PRINT SW

KR ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.

GRAPHIC ADJUST
CENTER ADJUST
ZERO ADJUST
REF ADJUSTMENT
KRT ADJUSTMENT
LIMIT ADJUST
PRINTER ADJUST
KOSHI ADJUST
AUTO ALING ADJUST

126

R-TRK1P-0-0809-02

Illustration

Procedure
Print switch

REF ADJUSTMENT
1.
2.
3.
4.
5.
6.

GET
GET
GET
SPH
CYL
REF

4. The REF ADJUSTMENT screen appears.


Press Measurement switch and select CYL
LINEARITY.
Press Print switch once.

S0
P2D
SPD
LINEARITY
LINEARITY
AXIS ADJUST

5. The CYL compensation values are displayed.


Press Measurement switch once.
Fit the values to the addresses of -2D and
0D. The values are written on E2PROM.

CYL LINEARITY
CA : +0.000000
CB : +1.000000
CC1: +0.000000

EXIT

6. The test eye address setting screen appears.


Make sure that the value in ( ) for -8D is displayed in yellow.
Press the [+] switch and [-] switch among the
function switches to fit the value to the address
of the used test eye -8D.

-8D(-7.82)
-5D(-4.91)
-2D(-2.92)
0D(-0.00)

7. After setting, press Measurement switch.


The set address is written on E2PROM.
EXIT

[+] and [-] switches

8. Perform the same operation as Procedure 6 and


7 to fit the values to the addresses of -5D,
-2D and 0D. The set addresses are written
on E2PROM.

127

R-TRK1P-0-0809-02

Illustration

Procedure
9. The SET TEST EYE -8D screen appears.
Set the -8D test eye. (Align the marking line.)
10. Press Measurement switch once.
Measurement starts.
11. After measurement, according to the instructions on the screen, measure the test eyes for
-5D, -2D and 0D in the same way as Procedure 6 and 7.

SET TEST EYE -8D

EXIT

12. After measurement, the calculated compensation values are displayed.


Press Measurement switch once.
The calculated compensation values are written
on E2PROM.
After writing, the REF ADJUSTMENT
screen appears again.

CA1 : +0.000000
CB1 : +1.000000
CC1 : +0.000000

EXIT

13. Set the slide switch SW6 on the CONTROL


PCB to the L side.

25

,
64
62 62

,
,

,

59

59

59

59

84 84

62
59

,

59
7

,

,
,
,

14. Set the DIP switch No. 8 of SW5 on the CONTROL PCB to OFF. Then, turn off the power.

SW6


SW5

U
Set No. 8 to OFF.

128

R-TRK1P-0-0809-02

6-6 AXIS ADJUST


Illustration

Procedure

25

,
64
62 62

,
,

59

59

,

59

59

62
59

59

,

,
,
,

,

SW6



SW5

84 84

Purpose
Using the AXIS=90 test eye, adjust the astigmatic axis.
1. Set the DIP switch No. 8 of SW5 on the CONTROL PCB to ON. Then, turn on the power.
Set the slide switch SW6 on the CONTROL
PCB to the U side.

Set No. 8 to ON.

Print switch

2. The top page of the compensation menu is displayed.


Press Measurement switch and select KR.
(The selected item is displayed with yellow letters.)
Press Print switch.

Print switch

3. The KR compensation menu is displayed.


Press Measurement switch and select 3. REF
ADJUST.
Press Print switch.

ADJUST LIST
Common
KR
CT
ITEM SELECT -> MEAS SW
EXECUTE
-> PRINT SW

KR ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.

GRAPHIC ADJUST
CENTER ADJUST
ZERO ADJUST
REF ADJUSTMENT
KRT ADJUSTMENT
LIMIT ADJUST
PRINTER ADJUST
KOSHI ADJUST
AUTO ALING ADJUST

129

R-TRK1P-0-0809-02

Illustration

Procedure
Print switch

REF ADJUSTMENT
1.
2.
3.
4.
5.
6.

GET
GET
GET
SPH
CYL
REF

4. The REF ADJUSTMENT screen appears.


Press Measurement switch and select 6. REF
AXIS ADJUST.
Press Print switch once.

S0
P2D
SPD
LINEARITY
LINEARITY
AXIS ADJUST

5. The AXIS compensation value is displayed.


Press Measurement switch once.
AXIS ADJUST
DA : +0.000000

EXIT

6. The SET TEST EYE 90 DEG screen appears.


Set the -2D astigmatic test eye.
(Align the 90 marking line.)
7. Press Measurement switch once.
Measurement starts.

SET TEST EYE 90 DEG

EXIT

130

R-TRK1P-0-0809-02

Illustration

Procedure
8. After measurement, the calculated compensation values are displayed.
Press Measurement switch once.
The calculated compensation values are written
on E2PROM.
After writing, the REF ADJUSTMENT
screen appears again.

A( 5) : 74.000000
DA
:+15.000000

9. Press the right end switch among the function


switches.
The KR compensation menu is displayed again.

REF ADJUSTMENT
1.
2.
3.
4.
5.
6.

GET
GET
GET
SPH
CYL
REF

S0
P2D
SPD
LINEARITY
LINEARITY
AXIS ADJUST

Right end switch

10. Set the slide switch SW6 on the CONTROL


PCB to the L side.

25
,
64
62 62

,
,

,

59

59

59

59

84 84

62
59

59
7

,

,

,
,
,

11. Set the DIP switch No. 8 of SW5 on the CONTROL PCB to OFF. Then, turn off the power.

SW6


SW5

Set No. 8 to OFF.

131

R-TRK1P-0-0809-02

7. KRT ADJUSTMENT
7-1 PARALLEL LEVEL
Illustration

Procedure

25

,

Purpose
Using the R8 test eye, adjust the light intensity
of the parallel projection spot on CCD.

64

59

59

,

84 84

62 62

,

,

59

59

62
59

59
7

,

,

,
,
,

SW6


SW5

1. Set the DIP switch No. 8 of SW5 on the CONTROL PCB to ON. Then, turn on the power.
Set the slide switch SW6 on the CONTROL
PCB to the U side.

Set No. 8 to ON.

Print switch

2. The top page of the compensation menu is displayed.


Press Measurement switch and select KR.
(The selected item is displayed with yellow letters.)
Press Print switch.

Print switch

3. The contents of the KR ADJUST menu are


displayed.
Press Measurement switch and select 5. KR
ADJUSTMENT. (The selected item is displayed with yellow letters.)
Press Print switch.

ADJUST LIST
Common
KR
CT
ITEM SELECT -> MEAS SW
EXECUTE
-> PRINT SW

KR ADJUST
1.
2.
3.
4.
5.
6.
7.

GRAPHIC ADJUST
CENTER ADJUST
ZERO ADJUST
REF ADJUSTMENT
KRT ADJUSTMENT
LIMIT ADJUST
PRINTER ADJUST

132

R-TRK1P-0-0809-02

Illustration

Procedure
4. The "KRT ADJUSTMENT" screen appears.
Press Measurement switch and select 1.
PARALLEL LEVEL. (The selected item is displayed with yellow letters.)
Press Print switch.

Print switch

KRT ADJUSTMENT
1.
2.
3.
4.
5.
6.
7.
8.

PARALLEL LEVEL
PARALLEL BEND
ALPHA CALCULATE
Z ALIGNMENT
CURVATURE ADJUST
AXIS ADJUST
ZERO CALCULATE
CORNEA LENGTH ADJUST

5. The PARALLEL LEVEL screen appears.


Set the R8 test eye.
PARALLEL LEVEL

6. Press Measurement switch once.

SET TEST EYE R8

EXIT

Adjust with VR4. Adjust with VR6.

LEFT
225(26)
204(35)

7. The parallel projection spot light intensity


adjustment screen appears.

RIGHT
225(27)
204(35)

EXIT

133

R-TRK1P-0-0809-02

Illustration
VR4

Procedure
8. Adjust the light intensity of the parallel projection spot by turning VR4 and VR6 on the LZM
PCB.
Standard: 35 5

VR6

LZM PCB

LEFT
225(26)
204(35)

9. After adjustment, press the [EXIT] switch


among the function switches.
The KRT ADJUSTMENT screen appears
again.

RIGHT
225(27)
204(35)

EXIT

[EXIT] switch

134

R-TRK1P-0-0809-02

7-2 PARALLEL BEND


Illustration

Procedure
Purpose
Using the R8 test eye, check the bend of the
parallel projection spot.

Print switch

KRT ADJUSTMENT
1.
2.
3.
4.
5.
6.
7.
8.

1. Press Measurement switch on the KRT


ADJUSTMENT screen and select 2. PARALLEL BEND. (The selected item is displayed
with yellow letters.)
Press Print switch.

PARALLEL LEVEL
PARALLEL BEND
ALPHA CALCULATE
Z ALIGNMENT
CURVATURE ADJUST
AXIS ADJUST
ZERO SEARCH
CORNEA LENGTH ADJUST

2. The PARALLEL BEND screen appears.


Set the R8 test eye.
PARALLEL BEND

3. Press Measurement switch once.

SET TEST EYE R8

EXIT

4. The PARALLEL BEND check screen


appears.
Make sure that the right and left spots are set
between the two upper and lower lines (without
touching).
5. After checking, press the [EXIT] switch among
the function switches.
The KRT ADJUSTMENT screen appears
again.

EXIT

[EXIT] switch

135

R-TRK1P-0-0809-02

7-3 ALPHA CALCULATE


Illustration

Procedure
Print switch

KRT ADJUSTMENT
1.
2.
3.
4.
5.
6.
7.
8.

PARALLEL LEVEL
PARALLEL BEND
ALPHA CALCULATE
Z ALIGNMENT
CURVATURE ADJUST
AXIS ADJUST
ZERO SEARCH
CORNEA LENGTH ADJUST

1. Press Measurement switch on the KRT


ADJUSTMENT screen and select 3. ALPHA
CALCULATE. (The selected item is displayed
with yellow letters.)
Press Print switch.

2. The ALPHA CALCULATE setting value


screen appears.
Press Measurement switch.

ALPHA LALCUATE
RING
RING
RING
SPOT
SPOT

(0)
(1)
(2)
(0)
(1)

:
:
:
:
:

Purpose
Using the R8 test eye, measure the incidence
angle into the test eye for each ring and the incidence angle for the parallel projection spot.

0.098602
0.148386
0.203308
0.149289
0.149289

EXIT

3. The ALPHA CALCULATE measurement


start preparation screen appears.
Set the R8 test eye.
4. Press Measurement switch.

SET TEST EYE R8

EXIT

136

R-TRK1P-0-0809-02

Illustration

Procedure
5. After measurement, the measurement results are
displayed.
Press Measurement switch.

RING
RING
RING
SPOT
SPOT

ALPHA
(0) : 0.098602
(1) : 0.148386
(2) : 0.203308
(0) : 0.149289
(1) : 0.149289

EXIT

KRT ADJUSTMENT
1.
2.
3.
4.
5.
6.
7.
8.

PARALLEL LEVEL
PARALLEL BEND
ALPHA CALCULATE
Z ALIGNMENT
CURVATURE ADJUST
AXIS ADJUST
ZERO SEARCH
CORNEA LENGTH ADJUST

6. The displayed measurement results are written


on E2PROM.
After writing, the KRT ADJUSTMENT
screen appears again.

137

R-TRK1P-0-0809-02

7-4 ZERO SEARCH


Illustration

Procedure
Print switch

KRT ADJUSTMENT
1.
2.
3.
4.
5.
6.
7.
8.

PARALLEL LEVEL
PARALLEL BEND
ALPHA CALCULATE
Z ALIGNMENT
CURVATURE ADJUST
AXIS ADJUST
ZERO SEARCH
CORNEA LENGTH ADJUST

Purpose
Using the Z-R6, Z-R8 and Z-R9 kerato
test eyes, search for the reference position (defocus position) of each test eye.
1. Press Measurement switch on the KRT
ADJUSTMENT screen and select 7. ZERO
SEARCH. (The selected item is displayed with
yellow letters.)
Press Print switch.

2. The ZERO SEARCH measurement start preparation screen appears.


Set the Z test eye R8.
(Each time you press the print switch, the measurement start preparation screen is changed in
the order R9 R6 R8.)

ZERO SEARCH

3. Press Measurement switch once.


SET TEST EYER8

EXIT

HI:+0.100001
GOOD

EXIT

4. Measurement starts.
Adjust the micrometer of the test eye until
GOOD is displayed.
(The reference position for Z ALIGNMENT
will be obtained.)
(How to adjust)
When NEAR is displayed, adjust the
micrometer so that the objective lens may come
nearer to the test eye.
When FAR is displayed, adjust the micrometer in the reverse direction of the above.

138

R-TRK1P-0-0809-02

Illustration

Procedure
5. When GOOD is displayed, adjust the
micrometer so that the value of HI may be
closest to +1.

HI:+0.100001
GOOD

EXIT

[EXIT] switch

KRT ADJUSTMENT
1.
2.
3.
4.
5.
6.
7.
8.

PARALLEL LEVEL
PARALLEL BEND
ALPHA CALCULATE
Z ALIGNMENT
CURVATURE ADJUST
AXIS ADJUST
ZERO SEARCH
CORNEA LENGTH ADJUST

6. After deciding the reference position, press the


[EXIT] switch among the function switches.
The KRT ADJUSTMENT screen appears
again.
7. Search for the reference position for Z-R9 and
Z-R6 in the same way.

139

R-TRK1P-0-0809-02

7-5 Z ALIGNMENT
Illustration

Procedure
Print switch

KRT ADJUSTMENT
1.
2.
3.
4.
5.
6.
7.
8.

PARALLEL LEVEL
PARALLEL BEND
ALPHA CALCULATE
Z ALIGNMENT
CURVATURE ADJUST
AXIS ADJUST
ZERO SEARCH
CORNEA LENGTH ADJUST

Purpose
Using the Z-R8 test eye, measure the parallel
projection spot image height and ring diameter
by unit: 0.25mm in the range of Reference
position 0.75mm and calculate the ratio.
1. Press Measurement switch on the KRT
ADJUSTMENT screen and select 4. Z
ALIGNMENT. (The selected item is displayed
with yellow letters.)
Press Print switch.

2. The Z ALIGNMENT compensation value


screen appears.
Press Measurement switch once.

Z ALIGNMENT
A0 : 0.000000
A1 : 0.000000
A2 : 0.000000
B0 :+1.000000
B1 :+1.000000
B2 :+1.000000
RING H10:+1.000000

EXIT

3. The Z ALIGNMENT measurement start preparation screen appears.


Set the R8 Z test eye and change to -0.75
from the reference position on the scale.
4. Press Measurement switch once.

SET TEST EYE POSITION -0.75

EXIT

140

R-TRK1P-0-0809-02

Illustration

Procedure
5. After measurement, the left screen appears.
6. According to the instructions on the screen,
carry out measurement at the positions of
-0.50, -0.25, 0 (reference position),
+0.25, +0.50 and +0.75.

SET TEST EYE POSITION -0.50

EXIT

7. After the measurement at the +0.75 position,


the calculated compensation values are displayed.
A0: -7.346591
A1: +0.196969
A2: +5.562500

8. Press Measurement switch.

B0:+15.705222
B1: +0.655303
B2:-10.064394
RING H10:+0.990592

EXIT

KRT ADJUSTMENT
1.
2.
3.
4.
5.
6.
7.
8.

9. The compensation values are written on


E2PROM and the KRT ADJUSTMENT
screen appears again.

PARALLEL LEVEL
PARALLEL BEND
ALPHA CALCULATE
Z ALIGNMENT
CURVATURE ADJUST
AXIS ADJUST
ZERO SEARCH
CORNEA LENGTH ADJUST

141

R-TRK1P-0-0809-02

7-6 CURVATURE ADJUST


Illustration

Procedure
Print switch

Purpose
Using the kerato test eye, adjust its linearity and
resudual astigmatism.

KRT ADJUSTMENT
1.
2.
3.
4.
5.
6.
7.
8.

1. Press Measurement switch on the KRT


ADJUSTMENT screen and select 5. CURVATURE ADJUST. (The selected item is displayed with yellow letters.)
Press Print switch.

PARALLEL LEVEL
PARALLEL BEND
ALPHA CALCULATE
Z ALIGNMENT
CURVATURE ADJUST
AXIS ADJUST
ZERO SEARCH
CORNEA LENGTH ADJUST

2. The CURVATURE ADJUST compensation


value screen appears.
Press Measurement switch once.

CURVATURE ADJUST
KA1: +0.000000
KB1: +0.000000
KC1: +0.000000
KA2: +0.000000
KB2: +0.000000
KC2: +0.000000
KA3: +0.000000
KB3: +0.000000
KC3: +0.000000

EXIT

3. The CURVATURE ADJUST measurement


start preparation screen appears.
Set the R6 Z test eye.
4. Press Measurement switch once and carry out
measurement.
5. Measure the R8 and R9 test eyes in the
same way.

SET TEST EYE R6

EXIT

142

R-TRK1P-0-0809-02

Illustration

Procedure
6. After measuring the R9 test eye, the measurement results are displayed.
Press Measurement switch.

CURVATURE ADJUST
KA1: +0.005519
KB1: +0.912205
KC1: +0.339881
KA2: +0.235665
KB2: -0.123456
KC2: +0.123456
KA3: +0.000456
KB3: +0.000123
KC3: +0.000012

EXIT

KRT ADJUSTMENT
1.
2.
3.
4.
5.
6.
7.
8.

PARALLEL LEVEL
PARALLEL BEND
ALPHA CALCULATE
Z ALIGNMENT
CURVATURE ADJUST
AXIS ADJUST
ZERO SEARCH
CORNEA LENGTH ADJUST

7. The displayed measurement results are written


on E2PROM.
After writing, the KRT ADJUSTMENT
screen appears again.

143

R-TRK1P-0-0809-02

7-7 AXIS ADJUST


Illustration

Procedure
Print switch

Purpose
Using the 90 and 180 kerato astigmatic test
eyes, adjust the axial angle.

KRT ADJUSTMENT
1.
2.
3.
4.
5.
6.
7.
8.

1. Press Measurement switch on the KRT


ADJUSTMENT screen and select 6. AXIS
ADJUST. (The selected item is displayed with
yellow letters.)
Press Print switch.

PARALLEL LEVEL
PARALLEL BEND
ALPHA CALCULATE
Z ALIGNMENT
CURVATURE ADJUST
AXIS ADJUST
ZERO SEARCH
CORNEA LENGTH ADJUST

2. The AXIS ADJUST compensation value


screen appears.
Press Measurement switch once.

AXIS ADJUST
KAXIS: +0.000000

EXIT

3. The CURVATURE ADJUST measurement


start preparation screen appears.
Set the astigmatic test eye in 180.
4. Press Measurement switch once and carry out
measurement.
5. After measurement, set the astigmatic test eye in
90.

SET TEST EYE 180DEG

EXIT

6. Press Measurement switch once and carry out


measurement.

144

R-TRK1P-0-0809-02

Illustration

Procedure
7. After measurement, the measurement results are
displayed.
Press Measurement switch.

AXIS ADJUST
KAXIS: +1.000000

EXIT

KRT ADJUSTMENT
1.
2.
3.
4.
5.
6.
7.
8.

PARALLEL LEVEL
PARALLEL BEND
ALPHA CALCULATE
Z ALIGNMENT
CURVATURE ADJUST
AXIS ADJUST
ZERO SEARCH
CORNEA LENGTH ADJUST

8. The displayed measurement results are written


on E2PROM.
After writing, the KRT ADJUSTMENT
screen appears again.

145

R-TRK1P-0-0809-02

7-8 CORNEA LENGTH ADJUST


Illustration

Procedure
Print switch

1. Press Measurement switch on the KRT


ADJUSTMENT screen and select 8. CORNEA LENGTH ADJUST. (The selected item is
displayed with yellow letters.)
Press Print switch.

KRT ADJUSTMENT
1.
2.
3.
4.
5.
6.
7.
8.

Purpose
Adjust the iris meter.

PARALLEL LEVEL
PARALLEL BEND
ALPHA CALCULATE
Z ALIGNMENT
CURVATURE ADJUST
AXIS ADJUST
ZERO SEARCH
CORNEA LENGTH ADJUST

2. Turn the volume VR2 on the LZM DRIVER


PCB clockwise to the utmost limit. (Minimize
the light intensity of the center spot.)

DOT-13 : 123
D-SIZE : +0.105619

3. The CORNEA LENGTH ADJUST compensation value screen appears.


Press Measurement switch once.
EXIT

4. The CORNEA LENGTH ADJUST measurement start preparation screen appears.


Set the cornea diameter measurement checking
tool on the chinrest.

SET LENGTH
MEMORY OK?

5. Collimate the scale of the cornea diameter measurement checking tool.

EXIT

6. Press Measurement switch and capture the


image.

146

R-TRK1P-0-0809-02

Illustration

SET LENGTH

Procedure
7. Move the right and left bars so that the interval
between them may be 13 (mm).
(How to adjust)
To move the R bar:
Arrow switch at the lower left corner on the
screen
To move the L bar:
Arrow switch at the lower right corner on the
screen
To move the R and L bars at the same time:
Switch at the lower center on the screen

13(mm)

8. After setting 13 (mm) as the interval between


bars, press Measurement switch once.

9. The compensation value check screen appears.


Press Measurement switch once.
DOT-13 : 123
D-SIZE : +0.105619

EXIT

10. The displayed measurement results are written


on E2PROM.
After writing, the KRT ADJUSTMENT
screen appears again.

KRT ADJUSTMENT
1.
2.
3.
4.
5.
6.
7.
8.

PARALLEL LEVEL
PARALLEL BEND
ALPHA CALCULATE
Z ALIGNMENT
CURVATURE ADJUST
AXIS ADJUST
ZERO SEARCH
CORNEA LENGTH ADJUST

11. Press the right end switch among the function


switches.
The KR compensation menu is displayed again.

Right end switch

147

R-TRK1P-0-0809-02

8. AUTO ALIGN ADJUST


Illustration

Procedure
Print switch

KR ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.

GRAPHIC ADJUST
CENTER ADJUST
ZERO ADJUST
REF ADJUSTMENT
KRT ADJUSTMENT
LIMIT ADJUST
PRINTER ADJUST
KOSHI ADJUST
AUTO ALING ADJUST

Purpose
Measure the parallel projection spot interval and
No. 1 ring image height by unit: 0.5mm in the
range of Reference position 2mm. When the
distance from the reference position, which corresponds to the ratio between the interval and
height, is regarded as Y, calculate the coefficient of the approximation formula of the second
degree.
1. Press Measurement switch on the KR compensation menu and select 8. AUTO ALIGN
ADJUST. (The selected item is displayed with
yellow letters.)
Press Print switch.

Print switch

AUTOADJUSTMENT
1. ZERO SEARCH
2. Z ALIGNMENT [AUTO

2. The AUTO ADJUSTMENT menu is displayed.


Press Measurement switch and select 2. Z
ALIGNMENT [AUTO].
Press Print switch.

3. The AUTO ALIGN ADJUST compensation


value screen appears.
Press Measurement switch.

AUTO ALIGN ADJUST


AL2 : +36.140530
AL1 : -134.558014
AL0 : +115.302132

EXIT

148

R-TRK1P-0-0809-02

Illustration

Procedure
4. The measurement start preparation screen
appears.
Set the Z-R8 test eye and change to the
+2.0mm position from the reference position
on the scale.

SET TEST EYE POSITON -2.0

EXIT

5. Press Measurement switch.


The calculated compensation values are displayed.
AL2 : +36.140530
AL1 : -134.558014
AL0 : +115.302132

6. Press Measurement switch. The calculated


compensation values are written on E2PROM
and the AUTO ADJUSTMENT menu is displayed again.

EXIT

7. Set the slide switch SW6 on the CONTROL


PCB to the L side.

25
,
64
62 62

,
,
,

59

59
,

59

59

84 84

62
59

59
7

,

,
,

8. Set the DIP switch No. 8 of SW5 on the CONTROL PCB to OFF. Then, turn off the power.

,

SW6


SW5

Set No. 8 to OFF.

149

R-TRK1P-0-0809-02

9. Center spot light intensity adjustment


Illustration

Procedure
Purpose
Make the center spot light intensity uniform.

,
62

,

,

1. Set the 0D test eye.

62A#&+0

2. Turn on the power of the instrument and start it


in the normal mode.

,
62
+%
62

,

CONTROL PCB

3. Connect the oscilloscope.


Observation probe:
TP ADIN (CONTROL PCB)
GND: TP4 (AG) (CONTROL PCB)

62

62
62
62
,

,

62

,

62
62
62

4. Find out the center spot with the TV mode of the


oscilloscope. Adjust VR2 of the LZM DRIVER
PCB so that the waveform width of the light
intensity saturation point in the center spot may
meet the standard.
Standard: 410nsec 20nsec

,

LZM DRIVER PCB

410nsec 20nsec

150

R-TRK1P-0-0809-02

10. Fixation target light intensity adjustment


Illustration

Procedure
1. Set the DIP switch No. 8 of SW5 on the CONTROL PCB to ON. Then, turn on the power.

25
,
64
62 62

,
,

,

59

59

59

84 84
59
7

,



59

62
59

,

,
,
,

Set No. 8 to ON.

SW5

Print switch

2. The top page of the compensation menu is displayed.


Press Measurement switch and select KR.
(The selected item is displayed with yellow letters.)
Press Print switch.

Print switch

3. The KR compensation menu is displayed.


Press Measurement switch and select 8.
KOSHI ADJUST.
Press Print switch.

KR ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.

GRAPHIC ADJUST
CENTER ADJUST
ZERO ADJUST
REF ADJUSTMENT
KRT ADJUSTMENT
LIMIT ADJUST
PRINTER ADJUST
KOSHI ADJUST
AUTO ALING ADJUST

KOSHI ADJUST
1. KOSHI_ADJUST_HIGH
2. KOSHI_ADJUST_LOW

4. Set the power meter on the tool holder.


5. The KOSHI ADJUST screen appears.
Press Measurement switch and select 1.
KOSHI ADJUST HIGH.
Press Print switch.

151

R-TRK1P-0-0809-02

Illustration

Procedure
6. The fixation target HIGH screen appears.
Adjust the light intensity to meet the standard
with VR3 on the LZM DRIVER PCB.
Standard: 15nW 0.3nW

Print switch

KOSHI ADJUST
1. KOSHI_ADJUST_HIGH
2. KOSHI_ADJUST_LOW

7. Press Measurement switch to return to the


KOSHI ADJUST screen.
Press Measurement switch and select 2.
KOSHI ADJUST LOW.
Press Print switch.

8. The fixation target LOW screen appears.


Adjust the light intensity to meet the standard
with VR1 on the LZM DRIVER PCB.
Standard: 5nW 0.3nW

152

R-TRK1P-0-0809-02

11. Printing the compensation values


Illustration

Procedure
1. Set the DIP switch No. 8 of SW5 on the CONTROL PCB to ON. Then, turn on the power.

25
,
64
62 62

,
,

,

59

59

59

59

59
7

,



84 84

62
59

,

,
,
,

Set No. 8 to ON.

SW5

2. The top page of the compensation menu is displayed. Press Measurement switch and select
KR. (The selected item is displayed with yellow letters.) Press Print switch on the control
panel.

Print switch

ADJUST LIST
Common

3. The KR compensation menu is displayed.

KR
CT
ITEM SELECT -> MEAS SW
EXECUTE
-> PRINT SW

4. Press Switch menu changing switch among


the function switches once. The compensation
values are printed and the top page of the compensation menu appears.

KR ADJUST
1.
2.
3.
4.
5.
6.
7.

GRAPHIC ADJUST
CENTER ADJUST
ZERO ADJUST
REF ADJUSTMENT
KRT ADJUSTMENT
LIMIT ADJUST
PRINTER ADJUST

5. Turn off the power.


Set the DIP switch No. 8 of SW5 on the CONTROL PCB to OFF.
Set No. 8 to OFF.

12345678
SW5
Switch menu
changing switch

153

R-TRK1P-0-0809-02

ADJUST DATA
HOFF
VOFF
KH00
KV00
CF
S0
H00
V00
P2D
SPD
A1
B1
C1
A2
B2
C2
A3
B3
C3
CA1
CB1
CC1
DA

: -1
: -4
:466
:251
:-0.057000
:+0.835155
:460
:251
:-0.000001
:-0.000001
:-0.000001
:-0.000001
:-0.000001
:-0.000001
:-0.000001
:-0.000001
:-0.000001
:-0.000001
:-0.000001
:-0.000001
:-0.000001
:-0.000001
:-0.000001

SP-M00
SP-M01
SP-M02
SP-M03
SP-M04
SP-M10
SP-M11
SP-M12
SP-M13
SP-M14
SP-M20
SP-M21
SP-M22
SP-M23
SP-M24

:+0.00000
:+0.00000
:+0.00000
:+0.00000
:+0.00000
:+0.00000
:+0.00000
:+0.00000
:+0.00000
:+0.00000
:+0.00000
:+0.00000
:+0.00000
:+0.00000
:+0.00000

RIG1
:0.099854
RIG2
:0.150269
RIG3
:0.205261
SPOT1 :0.149968
SPOT2 :0.149968
FOR_A1 :-0.000001
FOR_A2 :-0.000001
FOR_A3 :-0.000001
FOR_B1 :-0.000001
FOR_B2 :-0.000001
FOR_B3 :-0.000001
RINGHIO:-0.000001
A_DEF0 :-0.000001
A_DEF1 :-0.000001
A_DEF2 :-0.000001
A_DEF3 :-0.000001
A_DEF4 :-0.000001
A_DEF5 :-0.000001
A_DEF6 :-0.000001
KA1
:-0.000001
KB1
:-0.000001
KC1
:-0.000001
KA2
:-0.000000
KB2
:-0.000000
KC2
:-0.000000
KA3
:-0.000000
KB3
:-0.000000
KC3
:-0.000000
R6
:-0.000001
R8
:+1.643145
R9
:-0.000001
KAXIS :-0.000001
DOT-13 :490
D-SIZE :+0.026530
AL2
:-0.000010
AL1
:-0.000010
AL0
:-0.000010

Alignment frame
display position
Temporary measurement
center position
Fixation 0D position
Ring center coordinate
D quantity for one pulse
Target image displacement
per 1D on CCD
Compensation value for
calculating the spherical
power
Compensation value for
calculating the astigmatic
power
Compensation value for
calculating the axial angle

Spherical power linearity


compensation value

*******************
CIRCLE CENTER X:330
CIRCLE CENTER Y:236
CCD MAG
:2783
CCD CENTER X:480.17
CCD CENTER Y:249.47
X HOSEI
:5000
Y HOSEI
:5000
Z HOSEI
:5000
ENDOTHELIUM
: 612
PACHO MAG
:2961
SLND CNT 30STD: 8481
SLND CNT 30HI : 7961
SLND CNT 60
: 8637
AIR LOSS 30A : 500
AIR LOSS 30B : 2000
AIR LOSS 60A : 500
AIR LOSS 60B : 2000
PACHO TH2
:-153
PACHO TH1
:1092
PACHO INTERCEPT:
8
PACHO MODEL 0 : 394
PACHO MODEL 1 : 394
PACHO MODEL 2 : 394
PACHO MODEL 3 : 394
PACHO MODEL 4 : 394
IOP_CCT0
:65535
IOP_MAG0
:65535
IOP_CCT1
:65535
IOP_MAG1
:65535
IOP_CCT2
:65535
IOP_MAG2
:65535
IOP_CCT2
:65535
IOP_MAG2
:65535

Alignment frame position


CCD magnification
XY alignment reference
position on CCD
Endothelium signal reference
value on line sensor
Solenoid driving time
Compensation value for
calculating the intraocular
pressure
Compensation value for
calculating the corneal
thickness
Reference value of test eye
for compensating the corneal
thickness
Compensation value for
calculating the intraocular
pressure with corneal
thickness

TOPCON

Measured incidence angle

Compensation value for Z alignment

Compensation value for auto-start

Residual astigmatism compensation value


in linearity compensation

Cornea axial angle compensation value


Iris meter compensation value
Z alignment compensation value (AUTO)

154

R-TRK1P-0-0809-02

12. CCD camera position adjustment


Illustration

Procedure
1. Install the CCD alignment adjustment tool to the
CT optical head.

Tool pin

CCD alignment adjustment tool

Pushing tool

25
,
64
62 62

,
,

,

59

59

59

84 84
59
7

,



59

62
59

,

2. Set the DIP switches No. 7 and No. 8 of SW5 on


the CONTROL PCB to ON. Then, turn on
the power.

,
,
,

Set No. 7 and


No. 8 to ON.

SW5

Print switch

3. The left screen appears.


Press Print switch three times according to the
screen.

CONTINUE -> MEAS SW


BREAK
-> PRINT SW

155

R-TRK1P-0-0809-02

Illustration

Procedure
4. The top page of the compensation menu is displayed.
Make sure that CT is selected. (The selected
item is displayed with yellow letters.)
Press Print switch.

Print switch

ADJUST LIST
Common
KR
CT
ITEM SELECT -> MEAS SW
EXECUTE
-> PRINT SW

5. The CT compensation menu is displayed.


Press Measurement switch and select 1.
CCD XY SPOT.

CT ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.

CCD XY SPOT
DISPLAY ADJUST
CT CCD CENTER
CT CCD MAG
Z LEVEL
ALIGNMENT ADJUST
CORNEA ADJUST
SOLENOID TIMING
SAMPLE HOLD
MODEL EYE ADDRESS
IOP CONST INPUT
PACHO ADJUST

CCD RADIENT ADJUST

6. Install the monitor scale center tool on the monitor.


Illuminate the CCD alignment adjustment tool
from the back.

XY bright spot

7. Loosen the CCD camera fixing screw. Adjust


the CCD camera position so that the XY bright
spot may be at the center of the monitor scale
center tool and the vertical/horizontal chart may
be straight against the lines of the monitor scale
center tool. Then, tighten the fixing screw.

PEAK LVL:200
CONTRAST:
X:
Y:

Vertical/horizontal chart
Monitor scale center tool

156

R-TRK1P-0-0809-02

13. CT system: E2PROM WRITE


The compensation value data are stored in E2PROM on the CONTROL PCB. When replacing the PCB, if
you change this E2PROM, adjustment is not necessary. Write the initial values onto E2PROM when you use
a compensation value which is not written yet.
Illustration

Procedure

25
,
64
62 62

,
,

,

59

59

59

59

84 84
59
7

,



62
59

,

Purpose
Before adjusting the electric units in the CT system, store the initial compensation values.

,
,

,

1. Set the DIP switches No. 7 and No. 8 of SW5 on


the CONTROL PCB to ON. Then, turn on
the power.

Set No. 7 and


No. 8 to ON.

SW5

Print switch

2. The left screen appears.


Press Print switch three times according to the
screen.

Print switch

3. The top page of the compensation menu is displayed.


Make sure that Common is selected. (The
selected item is displayed with yellow letters.)
Press Print switch.

CONTINUE -> MEAS SW


BREAK
-> PRINT SW

ADJUST LIST
Common
KR
CT
ITEM SELECT -> MEAS SW
EXECUTE
-> PRINT SW

157

R-TRK1P-0-0809-02

Illustration

Procedure
4. The contents of the COMMON menu are displayed.
Press Measurement switch and select 5.
E2PROM INITIALIZE. (The selected item is
displayed with yellow letters.)
Press Print switch.

Print switch

COMMON ADJUST
1.
2.
3.
4.
5.
6.
7.

DESTINATION SET
CHINA OFFSET SET
STAGE OFFSET
BASE CHECK
E2ROM INITIALIZE
PRINTER DURATION
DUMMT01

5. The INITIAL WRITE screen is displayed.


Select TP with Chinrest up/down button.
INITIAL WRITE
ALL
KR/RM
TP
MODEL : TRK-1P
PRINT SW: OK
CHIN SW :UP&DOWN
MEAS SW : CANCE

6. Set the slide switch SW6 on the CONTROL


PCB to the U side.

25
,
64
62 62

,
,

,

59

59

59

59

62
59

84 84
59
7

,

,

,
,
,

SW6
Writing SW

158

R-TRK1P-0-0809-02

Illustration

Procedure
7. Press Print switch to perform initialization.
Make sure that SUCCESS is displayed on the
screen and press Measurement switch.

Print switch

WRITE:SUCCEESS

PUSH : MAS SW

8. The contents of the COMMON menu are displayed.


Press Right end switch among the function
switches to return to the top page of the compensation menu.

COMMON ADJUST
1.
2.
3.
4.
5.
6.
7.

DESTINATION SET
CHINA OFFSET SET
STAGE OFFSET
BASE CHECK
E2ROM INITIALIZE
PRINTER DURATION
DUMMT01

Right end switch

9. Set the slide switch SW6 on the CONTROL


PCB to the L side.

25
,
64
62 62

,
,
,

59

59
,

59

59

84 84

62
59

59
7

,

,
,
,

10. Set the DIP switches No. 7 and No. 8 of SW5 on


the CONTROL PCB to OFF. Then, turn off
the power.

SW6


SW5

Set No. 7 and


No. 8 to OFF.

159

R-TRK1P-0-0809-02

14. Analog circuit reference voltage adjustment


Illustration

Procedure

VR1:
Adjust to 4.75V 0.02V.

Purpose
Adjust the reference voltage for the A/D converter and pressure sensor.

62
,

84 84

, ,

,

,
84

84

84
84 84

,

1. Adjust VR1 on the ANALOG IF PCB to nearly


the center.

,
62
62

62
62
62
)0&

,

62
84

84

62
84
,

,

62 84

2. Tester connection
TP210: +" probe
TP407: - probe (GND)

ANALOG IF PCB

3. Turn on the power of the instrument.


4. Turn VR1 on the ANALOG IF PCB to adjust
the voltage.
Standard: 4.75 0.02V
5. Turn off the power of the instrument.

160

R-TRK1P-0-0809-02

15. Pressure detection circuit offset voltage removal


Illustration

Procedure
Remove the pressure
sensor PCB cable.

VR2:
Adjust to 0.00 0.02V.

1. Remove the pressure sensor PCB cable from


J204 of the ANALOG IF PCB.

62
,
,

84 84

, ,
,

84

84

84 84
,
62
62

62
62
62
)0&

,

84

,

62

84
,

2. Turn VR3 on the ANALOG IF PCB clockwise


to the utmost limit.

84

84

62

Purpose
Remove the offset voltage which is generated in
only the pressure detection circuit.

,

62 84

ANALOG IF PCB
VR3:
Turn it clockwise to
the utmost limit.

3. Tester connection
TP204: + probe
TP407: - probe (GND)
4. Turn on the power of the instrument.
5. Turn VR2 on the ANALOG IF PCB to adjust
the voltage.
Standard: 0.00 0.02V
6. Turn off the power of the instrument.

161

R-TRK1P-0-0809-02

16. Pressure sensor offset voltage adjustment


Illustration

Procedure

Connect the pressure


sensor PCB cable.

Purpose
Remove the offset voltage which is generated in
only the pressure sensor.

62
,

,
84

84

84 84

, ,

,

,

84
84 84

,

1. Turn VR3 on the ANALOG IF PCB clockwise


to the utmost limit.

,
62
62

62
62
62
)0&

62
84

84

62
84
,

,

2. Connect the pressure sensor PCB cable to J204


of the ANALOG IF PCB.

62 84

ANALOG IF PCB
VR3:
Turn it clockwise to
the utmost limit.

PRESS SENSOR PCB

3. Leave the PRESS SENSOR PCB in the air as it


is.
4. Tester connection
TP204: + probe
TP407: - probe (GND)
5. Turn on the power of the instrument.

84

VR1:
Adjust to 0.0 0.1V.

6. Turn VR1 on the PRESS SENSOR PCB to


adjust the voltage.
Standard: 0.0 0.1V
7. Turn off the power of the instrument.

162

R-TRK1P-0-0809-02

17. Pressure detection circuit sensitivity adjustment


Illustration

Procedure
62

,

,

84 84

, ,
,

84

84

,

84
84 84

,

Purpose
Adjust the sensitivity of the pressure detection
circuit (including the pressure sensor).

,
62
62

62
62

62

1. Turn VR3 on the ANALOG IF PCB approximately to the center.

84

84

62
84

62
)0&

,

,

62 84

2. Install the pressure tool to the pressure sensor.


VR3:
Turn approximately
to the center.

3. Tester connection
TP204: + probe
TP407: - probe (GND)

4. Give the air pressure of 200 2 mmHg to the


pressure sensor with the pressure tool.

J204

200 2mmHg

5. Turn on the power of the instrument.

Pressure sensor

6. Turn VR3 on the ANALOG IF PCB to adjust


the voltage.
Standard: 4.00 0.02V

Push the Pressure tool


rubber ball.

62
,

,

84 84

, ,
,

84

84

,

84
84 84

,

,
62
62

62
62

62
84

84

62
84

62
)0&

,

,

7. After adjusting, leave the pressure sensor as it is


in the air and check if its offset voltage meets
0.0 0.1V.
If not, repeat the procedure from 15. Pressure
detection circuit offset voltage removal.

62 84

8. Turn off the power of the instrument.

ANALOG IF PCB

Adjust VR3 to meet the following standard.


[When 200mmHg is given to the pressure sensor]
4.00 0.02V
[When the pressure sensor is left in the air]
0.0 0.1

9. Set the O-ring on the pressure sensor PCB sensor and install the sensor to the head.

163

R-TRK1P-0-0809-02

18. LINE CCD PCB position adjustment


Illustration

Procedure
Anterior eye section
window glass tool

Purpose
Make adjustment so that the Z alignment slit
light may be approximately at the center of Z
sensor.

Tool pin

1. Install the anterior eye section window glass


tool to the CT optical head.
2. Install the pushing tool and the Z adjustment
tool eye to the CT optical head.

Z adjustment test eye

Pushing tool

25
,
64
62 62

,
,

,

59

59

59

,



59

62
59

84 84
59
7

,

3. Set the DIP switches No. 7 and No. 8 of SW5 on


the CONTROL PCB to ON. Then, turn on
the power.

,
,
,

Set No. 7 and


No. 8 to ON.

SW5

Print switch

4. The left screen appears.


Press Print switch three times according to the
screen.

CONTINUE -> MEAS SW


BREAK
-> PRINT SW

164

R-TRK1P-0-0809-02

Illustration

Procedure
Print switch

ADJUST LIST
Common

5. The top page of the compensation menu is displayed.


Press Measurement switch and select CT.
(The selected item is displayed with yellow letters.)
Press Print switch.

KR
CT
ITEM SELECT -> MEAS SW
EXECUTE
-> PRINT SW

6. The CT compensation menu is displayed.


Press Measurement switch and select 7.
CORNEA ADJUST.

CT ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.

CCD XY SPOT
DISPLAY ADJUST
CT CCD CENTER
CT CCD MAG
Z LEVEL
ALIGNMENT ADJUST
CORNEA ADJUST
SOLENOID TIMING
SAMPLE HOLD
MODEL EYE ADDRESS
IOP CONST INPUT
PACHO ADJUST

,

41)

%.-

8176

7. Connect the oscilloscope's probes.


Observation probe: Z VIDEO
(LINE CCD PCB)
Trigger probe:
ROG (LINE CCD PCB)
Ground:
GND (LINE CCD PCB)

62
84

)0&
,
<A8+&'1

LINE CCD PCB

165

R-TRK1P-0-0809-02

Illustration

Procedure
62

,

,

84 84

, ,
,

84

84

,

84
84 84

,

,
62
62

62
62

62
84

84

62
84

62
)0&

,

,

41)

8. Turn VR5 on the ANALOG IF PCB counterclockwise to the utmost limit (minimum LED
light intensity).
Turn VR1 (CCD gain) on the LINE CCD PCB
counterclockwise to the utmost limit (minimum).

62 84

,

9. Loosen the LINE CCD PCB fixing screw.


Move the PCB in the arrow direction and, where
the valley part is maximum with the endothelium reflection signal of 3V or less, tighten the
screw to fix the PCB.

8176

%.-

62
84

)0&
,
<A8+&'1

LINE CCD PCB

Endothelium reflection signal


Maximum

3.0V or less

166

R-TRK1P-0-0809-02

19. Corneal thickness measurement LED light intensity adjustment


Illustration

Procedure
Purpose
Make the discrepancy in light intensity between
the instruments uniform.

Epithelium reflection signal


Endothelium reflection signal

1. Perform this adjustment in the condition just


after 18. LINE CCD PCB position adjustment.

3.0V 0.1V

VR5:
Adjust to 3.0V 0.1V.
62
,

,

84 84

, ,
,

84

84

,

2. Turn VR5 on the ANALOG IF PCB so that the


endothelium reflection signal level may meet
the standard.
Standard: 3.0V 0.1V

84
84 84

,

,
62
62

62
62

62
84

84

62
84

62
)0&

,

,

62 84

ANALOG IF PCB

167

R-TRK1P-0-0809-02

20. Corneal thickness measurement offset memory


Illustration

Procedure

25

,

Purpose
Store the reference position and calculate the
magnification on the line sensor.

64

,

84 84

62 62

,

,

59

59

59

59

62
59

59
7

,

,

,
,
,

1. Set the slide switch SW6 on the CONTROL


PCB to the U side (writing is permitted).

SW6
Writing SW

Print switch

CORNEA ADJUST
EPITHELIUM :
ENDOTHELIUM:
MAG_LINE_S :
COUNT
:
RESULT

2. Press Print switch.


The count on the screen is higher. The endothelium signal position and the calculated magnification on the line sensor are displayed.
3. Press Measurement switch.
The CT menu is displayed again.

4. Set the slide switch SW6 on the CONTROL


PCB to the L side (writing is not permitted).

SW6

168

R-TRK1P-0-0809-02

21. CCD magnification compensation


Illustration

Procedure
Purpose
When calculating the alignment deviation quantity with the image data, make the detected value
for all the instruments uniform.

Tool pin

1. Install the CCD magnification compensation


tool.

CCD magnification
compensation tool
Illuminate
with a bulb.

Pushing tool

2. Set the slide switch SW6 on the CONTROL


PCB to the U side (writing is permitted).

25

,
64

59

59

,

84 84

62 62

,

,

59

59

62
59

59
7

,

,

,
,
,

SW6
Writing SW

Print switch

3. Press Measurement switch on the CT compensation menu and select 4. CT CCD MAG.
Press Print switch.

CT ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.

CCD XY SPOT
DISPLAY ADJUST
CT CCD CENTER
CT CCD MAG
Z LEVEL
ALIGNMENT ADJUST
CORNEA ADJUST
SOLENOID TIMING
SAMPLE HOLD
MODEL EYE ADDRESS
IOP CONST INPUT
PACHO ADJUST

169

R-TRK1P-0-0809-02

Illustration

Procedure
4. Illuminate from the back of the magnification
compensation tool until the compensation value
is calculated (the measurement times are
counted to 20 automatically and the calculation is finished).

CCD MSGNIFICATION ADJUST

Compensation is finished.
Finish it with the measurement switch.
Compensation value:2735
Measurement times :20

5. After the compensation value is calculated,


press Measurement switch.
The CT compensation menu is displayed
again.

6. Set the slide switch SW6 on the CONTROL


PCB to the L side (writing is not permitted).

SW6
Writing is not permitted.

170

R-TRK1P-0-0809-02

22. Light intensity adjustment of the light source for XY alignment and applanation
Illustration

Procedure

Tool pin

Pushing tool

Purpose
Make the light intensity (parallel light) of the XY
LED, which is emitted to the patient's eye, for all
the instruments uniform.
1. Install the XYZ alignment adjustment tool.

XYZ alignment
adjustment tool

2. Press Measurement switch on the CT compensation menu and select 1. CCD XY SPOT.
Press Print switch.

Print switch

CT ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.

CCD XY SPOT
DISPLAY ADJUST
CT CCD CENTER
CT CCD MAG
Z LEVEL
ALIGNMENT ADJUST
CORNEA ADJUST
SOLENOID TIMING
SAMPLE HOLD
MODEL EYE ADDRESS
IOP CONST INPUT
PACHO ADJUST

,

, ,

62
,

84 84
,

84

84

,

84
84 84

,

,
62
62

62
62

62

3. Turn VR6 on the ANALOG IF PCB so that


TOTAL LEVEL on the monitor may be
within the standard.
Standard: 200 5 [digit]

84

84

62
84

62
)0&

,

,

62 84

4. After the adjustment, press Measurement


switch. The CT compensation menu is displayed again.
CCD RADIENT ADJUST

PEAK LVL:200
CONTRAST:
X:
Y:

171

R-TRK1P-0-0809-02

23. Alignment frame position adjustment


Illustration

Procedure
Purpose
Fit the alignment reference position to the alignment frame center.

Tool pin

Pushing tool

1. Install the XYZ alignment adjustment tool.

XYZ alignment
adjustment tool

Print switch

CT ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.

CCD XY SPOT
DISPLAY ADJUST
CT CCD CENTER
CT CCD MAG
Z LEVEL
ALIGNMENT ADJUST
CORNEA ADJUST
SOLENOID TIMING
SAMPLE HOLD
MODEL EYE ADDRESS
IOP CONST INPUT
PACHO ADJUST

3. Set the slide switch SW6 on the CONTROL


PCB to the U side (writing is permitted).
SW6
Writing is permitted.

ALIGN MARK ADJUST

XY bright spot

H POSITION : 329
V POSITION : 233

Marker

DOWN

RIGHT RESET

UP

LEFT

2. Press Measurement switch on the CT compensation menu and select 2. DISPLAY


ADJUST.
Press Print switch.

Function switches

4. The ALIGN MARK ADJUST screen appears.


Press the following function switches so that the
marker may be displayed at the XY bright spot
center.
DOWN switch: Moves the displayed
alignment frame downward.
UP switch:
Moves the displayed
alignment frame upward.
LEFT switch: Moves the displayed
alignment frame leftward.
RIGHT switch: Moves the displayed
alignment frame rightward.
RESET switch: Returns to the initial displayed
status.
5. After adjusting, press Measurement switch.
The CT compensation menu is displayed again.
172

R-TRK1P-0-0809-02

24. CCD XY alignment offset memory


Illustration

Procedure
Print switch

Purpose
Store the reference position, which is used for XY
alignment on the CCD, into E2PROM.

CT ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.

CCD XY SPOT
DISPLAY ADJUST
CT CCD CENTER
CT CCD MAG
Z LEVEL
ALIGNMENT ADJUST
CORNEA ADJUST
SOLENOID TIMING
SAMPLE HOLD
MODEL EYE ADDRESS
IOP CONST INPUT
PACHO ADJUST

CT CENTER ADJUST
HORIZONTAL PIX : 474.71
VERTICAL PIX : 275.76
SIGNAL LEVEL : 64

XY bright spot

1. Press Measurement switch on the CT compensation menu and select 3. CT CCD CENTER.
Press Print switch.

2. The ALIGN MARK ADJUST screen appears.


Press Measurement switch.
The reference position is automatically written
into E2PROM and the CT compensation
menu is displayed again.
3. Set the slide switch SW6 on the CONTROL
PCB to the L side (writing is not permitted).

CALIBRATION DATA READY

SW6
Writing is not permitted.

173

R-TRK1P-0-0809-02

25. Z alignment light source light intensity adjustment


Illustration

Procedure
Purpose
Set the light intensity of the Z alignment light
source uniform.

Anterior eye section


window glass tool

1. Install the anterior eye section window glass


tool to the optical head.

2. Press Measurement switch on the CT compensation menu and select 5. Z LEVEL.


CT ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.

CCD XY SPOT
DISPLAY ADJUST
CT CCD CENTER
CT CCD MAG
Z LEVEL
ALIGNMENT ADJUST
CORNEA ADJUST
SOLENOID TIMING
SAMPLE HOLD
MODEL EYE ADDRESS
IOP CONST INPUT
PACHO ADJUST

3. Contact the power meters probe with the anterior eye section window glass tool.

Power meter

62
,

,

84 84

, ,
,

84

84

84
84 84

,

4. Turn VR7 on the ANALOG IF PCB so that the


light intensity may meet the standard.
Standard: 150 4 [W] (Wavelength: 780 [nm])

,
62
62

62
62

62
84

84

62
84

62
)0&

,

,

,

62 84

* When measuring the light intensity with a commercial power meter, adjust the light intensity to
half the above standard value because the light
source blinks with the duty ratio 0.5.

174

R-TRK1P-0-0809-02

26. Z sensor PCB position adjustment


Illustration

Procedure
Anterior eye section
window glass tool

Tool pin

Pushing tool

Purpose
Make adjustment so that the Z alignment light
may be condensed at the Z sensor center.
1. Install the anterior eye section window glass
tool and XYZ alignment adjustment tool to the
optical head.

XYZ alignment
adjustment tool

84

84

62

2. Push the Z SENSOR PCB against the mounting


metal part. Under this condition, move the Z
SENSOR PCB up and down and adjust its position so that the bar displayed on the monitor
may be the longest.

62

84
84
62

,

62
62

Push it.

Z LEVEL
Z(1) 300 (2) 300
TOTAL LEVE : 768

Make this bar longest.

Z LED ILM LED

175

R-TRK1P-0-0809-02

27. Z sensor PCB frequency adjustment


Illustration

Procedure
Anterior eye section
window glass tool

Purpose
Fit the synchronizing frequency of the band-pass
filter circuit in the Z alignment circuit to the
LED modulating frequency (1.0kHz).

Tool pin

Pushing tool

1. Install the anterior eye section window glass


tool and XYZ alignment adjustment tool to the
optical head.

XYZ alignment
adjustment tool

84

2. Connection and setting of oscilloscope


Vertical axis : 200 [mV/div]
Horizontal axis : 500 [us/div]
CH1 : TP331
CH2 : TP333
GND : TP335

84

62

62

84
84
62

,

62
62

Z SENSOR PCB

Make the signal


peak maximum.

3. Turn VR1 on the Z SENSOR PCB clockwise so


that the signal peak may be maximum on the
oscilloscope.
Adjust VR3 in the same way. If the signal peak
has already passed, turn the trimmer counterclockwise again and make the adjustment once
more.

176

R-TRK1P-0-0809-02

28. Z sensor circuit gain adjustment


Illustration

Procedure
Purpose
Adjust the sensitivity so that the Z sensor detection signal may be uniform at Z1 and Z2 in the
condition that the Z LED spot is imaged at the
center.

84

84

62

62

84
84
62

,

62
62

1. Turn the volume VR2 and adjust the Z1 output


level to meet the standard value.
Turn the volume VR4 and adjust the Z2 output
level to meet the standard value.
If you have passed over the standard value,
return the volume and adjust it again.
Standard: 300 1 [Digit]

Z SENSOR PCB

Z LEVEL
Z(1) 300 (2) 300
TOTAL LEVE : 768

Z LED ILM LED

177

R-TRK1P-0-0809-02

29. Applanation detection circuit sensitivity adjustment


Illustration

Procedure

Anterior eye section


window glass tool

Purpose
Adjust the detection sensitivity to judge the real
measured value, the value with ( ) and the error
uniformly.

Tool pin

Pushing tool

1. Install the applanation system adjustment tool


and the anterior eye section window glass tool.

Applanation system
adjustment tool

62
,

,

84 84

, ,
,

84

84

,

84

2. Connection of oscilloscope
CH1 : TP201
GND : TP406

84 84

,

,
62
62

62
62

62
84

84

62
84

62
)0&

,

,

62 84

3. Press Measurement switch on the CT compensation menu and select 9. SAMPLE


HOLD.

CT ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.

CCD XY SPOT
DISPLAY ADJUST
CT CCD CENTER
CT CCD MAG
Z LEVEL
ALIGNMENT ADJUST
CORNEA ADJUST
SOLENOID TIMING
SAMPLE HOLD
MODEL EYE ADDRESS
IOP CONST INPUT
PACHO ADJUST

4. Press Chinrest up/down button to hold the signal. Under this condition, turn the volume VR4
on the ANALOG IF PCB so that the applanation
signal may meet the standard value.
Standard: 10 0.2 [V]p-p

10 0.2 [V]p-p

GND

178

R-TRK1P-0-0809-02

30. 30 [mmHg] pressure rising curve cutoff adjustment


Illustration

Procedure
Purpose
Make the 30 [mmHg] pressure rising curve for
all the instruments uniform.

CT ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.

CCD XY SPOT
DISPLAY ADJUST
CT CCD CENTER
CT CCD MAG
Z LEVEL
ALIGNMENT ADJUST
CORNEA ADJUST
SOLENOID TIMING
SAMPLE HOLD
MODEL EYE ADDRESS
IOP CONST INPUT
PACHO ADJUST

1. Press Measurement switch on the CT compensation menu and select 8. SOLENOID


TIMING ADJUST.

2. Check if the pressure is set in the 30 [mmHg]


range.

SOLENOID ADJUST
SOLENOID
PRESSURE
PULSE WIDTH
(SOLENOID_COUNT

:
:
:
:

3. Connection of oscilloscope
CH1 : TP205 (ANALOG PCB)
GND : TP406 (ANALOG PCB)
Trigger mode : DC trigger and trigger voltage
0.3 [V]
VOLTS
: 0.5V
TIMS
: 2ms

READY
30MMHG_STD
14.1 MSEC
0x1CA9)

COUNT UP -> CHINREST DOWN


COUNT_DOWN -> CHINREST UP

30 STD 30 HIGH 60 STD

DRIVE

[DRIVE] switch

5. Adjust the solenoid driving time with the chinrest up/down button and press the [DRIVE]
switch among the function switches to generate
air puff. Adjust the peak of the pressure waveform to be within the standard.

,
64
84 84
62
59

2.5

+0.1
0

,

59

4. Press the [DRIVE] switch among the function


switches and generate air puff. Turn the volume
VR1 on the CONTROL PCB to adjust the tilt
angle (52) of the pressure rising curve to fit to
the template.

,
,
,

6. After setting the pressure to the standard value,


register the solenoid driving time in the compensation ROM.
Pressure
waveform

[V]
52

Trigger voltage: 0.3V

179

R-TRK1P-0-0809-02

31. 60 [mmHg] pressure rising curve cutoff adjustment


Illustration

Procedure
Purpose
Make the 60 [mmHg] pressure rising curve for
all the instruments uniform.

SOLENOID ADJUST
SOLENOID
PRESSURE
PULSE WIDTH
(SOLENOID_COUNT

:
:
:
:

READY
60MMHG_STD
14.1 MSEC
0x1CA9)

1. Press the [60 STD] switch among the function


switches to change to the 60 [mmHg] range.
2. Press the [DRIVE] switch among the function
switches and generate air puff. Use the volume
VR2 on the CONTROL PCB to adjust the tilt
angle (65) of the pressure rising curve to fit to
the template.

COUNT UP -> CHINREST DOWN


COUNT_DOWN -> CHINREST UP

30 STD 30 HIGH 60 STD

DRIVE

[60 STD] switch


[DRIVE] switch

,
64

84 84

,

62
62
59

59
7

,

3. Adjust the solenoid driving time with the chinrest up/down button and press the [DRIVE]
switch among the function switches to generate
air puff. Adjust the peak of the pressure waveform to be within the standard.
4. After finishing the compensation, register the
solenoid driving time in the compensation
ROM.

,
,

5. After finishing this adjustment, return to 30. 30


[mmHg] pressure rising curve cutoff adjustment and make sure that the pressure waveform fits to the template.
Pressure
waveform

4.0 +0.2
0 [V]
65

Trigger voltage: 0.3V

180

R-TRK1P-0-0809-02

32. How to calculate the intraocular pressure conversion compensation value


Illustration

Procedure
1. Mount the repair test eye onto the chin rest.

25

,
64

59

59

,

84 84

62 62

,
,

59

59

62
59

59
7

,

,

,

2. Set the DIP switch No. 6 of SW5 on the CONTROL PCB to ON. Then, turn on the power.

,
,

Set No. 6 to ON.




SW5

3. Change to the TONO mode.


Make sure that the measurement mode 30 is
set.

01 0001
4 62 .
4

4. Measure the test eyes for 10mmHg,


20mmHg and 30mmHg twenty times.
(In the case of OVER for the 30mmHg test
eye, use the measurement mode 60.)

66


#8)#8)



 6101

OO*I

#761

5. Average the data of twenty measurements for


each test eye and calculate the constants A
and B by using Method of least squares.

123..

6. Measure the test eyes in the measurement mode


60 in the same way.

Reference
value seal

Test eye

Test eye holder

Reference value
No. 08001
TRK-1P

10
10.1

20
20.0

30
29.9

181

R-TRK1P-0-0809-02

Calculation by using Method of least squares in Excel


1. Input the standard value and the 1/10 value of A/D average.

2. Select two cells under A and B.

3. Input LINEST (C4:C6,D4:D6,TRUE).

4. Press the [Ctrl], [Shift] and [Enter] keys at the same time.
{} is added and so {LINEST (C4:C6,D4:D6,TRUE)} is displayed.

182

R-TRK1P-0-0809-02

33. How to input the intraocular pressure conversion compensation value


Illustration

Procedure

25

,
64

,

59

84 84

62 62

,
,
59

59

59

62
59

,

59

,
7

,

SW6



SW5

,
,

Purpose
Compensate the difference in the chamber volume and the difference between the pistons to
eliminate the discrepancy in the measured
intraocular pressure data.

1. Set the DIP switch No.8 of SW5 to ON and


set the slide switch SW6 to the U side (writing
is permitted) on the CONTROL PCB. Then,
turn on the power.

Set No. 8 to ON.

Print switch

ADJUST LIST
Common

2. The top page of the compensation menu is displayed.


Press Measurement switch and select CT.
(The selected item is displayed with yellow letters.)
Press Print switch.

KR
CT
ITEM SELECT -> MEAS SW
EXECUTE
-> PRINT SW

CT ADJUST

3. The CT compensation menu is displayed.


Press Measurement switch and select 11.
IOP CONST INPUT.

1. CCD XY SPOT
2. DISPLAY ADJUST
3. CT CCD CENTER
4. CT CCD MAG
5. Z LEVEL
6. ALIGNMENT ADJUST
7. CORNEA ADJUST
8. SOLENOID TIMING
9. SAMPLE HOLD
10.MODEL EYE ADDRESS
11.IOP CONST INPUT
12.PACHO ADJUST

183

R-TRK1P-0-0809-02

Illustration

Procedure
4. Input the calculated intraocular pressure compensation value.
Press Mode switch and select the item to be
changed. (The selected item is displayed with
yellow letters.)
Change the values with Function switches.

Mode switch

ITEM SELECT
APPLA CONSTANT INPUT
MODE 30
A :
564
B :
1593
MODE 60
A :
606
B :
2029

1000

100

10

Multiply the calculated


Then, count fractions of
and cast away the rest.
Multiply the calculated
Then, count fractions of
and cast away the rest.
Example)
A: 0.3224 332
B: -10.026 1003

Function switches

value A by 1000.
.5 and over as a unit
value B by 100.
.5 and over as a unit

5. After inputting the values, press Measurement


switch to return to the compensation menu.

25
,
64
62 62

,
,

,

59

59

59

59

84 84

62
59

59
7

,

7. Set the DIP switch No. 8 of SW5 to OFF and


set the slide switch SW6 to the L side on the
CONTROL PCB.



6. Turn off the power.

,

SW6
SW5

,
,

,

Set No. 8 to OFF.

184

R-TRK1P-0-0809-02

34. Corneal thickness compensation value memory


Illustration

Procedure
Purpose
To ensure the absolute accuracy of the corneal
thickness, store the compensation value, which
should be used in the final compensation calculation, into E2PROM.
1. Set the corneal thickness compensation test eye
onto the chinrest.
2. Set the DIP switch No. 8 of SW5 to ON and
set the slide switch SW6 to the U side (writing
is permitted) on the CONTROL PCB. Then,
turn on the power.
25

,
64

Print switch
,

59

84 84

62 62

,
,
59

59

59

62
59

,

Common



SW5

KR

ITEM SELECT -> MEAS SW


EXECUTE
-> PRINT SW

CT ADJUST
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.

,
,
,

Set No. 8 to ON.

ADJUST LIST

CT

,

59

SW6
U

3. The top page of the compensation menu is displayed.


Press Measurement switch and select CT.
(The selected item is displayed with yellow letters.)
Press Print switch.

4. Press Measurement switch on the CT compensation menu and select 12. PACHO
ADJUST.

CCD XY SPOT
DISPLAY ADJUST
CT CCD CENTER
CT CCD MAG
Z LEVEL
ALIGNMENT ADJUST
CORNEA ADJUST
SOLENOID TIMING
SAMPLE HOLD
MODEL EYE ADDRESS
IOP CONST INPUT
PACHO ADJUST

185

R-TRK1P-0-0809-02

Illustration

Procedure
5. First, acquire the data of the test eye equivalent
to 0.4mm. Set the value of MODEL EYE on
the screen to be the same as the real value of the
test eye, which should be measured, through the
panel switches at the lower section of the LCD.
(Measure the test eyes equivalent to 0.4, 0.5,
0.6, 0.7 and 0.8 ten times in order to acquire
data.)

PACHOADJUST

TARGET T=0.4 MODEL EYE

:0.381

DOWN

L_SHIFT R_SHIFT

6. After the preparation is OK, press Measurement switch to perform measurement. Carry
out alignment with the control lever until the
instrument reaches the auto-alignment start
position.

UP

7. After the instrument has finished measuring all


the test eyes, the compensation values are displayed and they are written in EEPROM.
If the primary compensation coefficient is deviated from 1 in a great degree (approx. 0.5),
clean the test eye with solvent and make adjustment again.

PACHOADJUST

TARGET T=0.4 MODEL EYE


L_SHIFT R_SHIFT

8. Press Print switch on the control panel, and


the CT ADJUST menu is displayed again.

:0.381

DOWN

UP

25
,
64
62 62

,
,

,

59

59

59

59

84 84

62
59

59
7

,

,
,

,

,

9. Set the slide switch SW6 on the CONTROL


PCB ASSY to the L side (writing is not permitted).
10. Set the DIP switch No. 8 of SW5 on the CONTROL PCB to OFF. Then, turn off the power.

SW6


SW5

Set No. 8 to OFF.

186

R-TRK1P-0-0809-02

4. TOOL LIST
4-1 Special tool list
Tool No.

Usage
rank

RM-A2000-03

0D test eye

Mounted on the tool test eye and


used for compensation of S0 and
SPD.

RM-7100-01

Cornea diameter measurement checking tool

Used for compensation of cornea


diameter measurement. (A normal
scale whose minimum unit is
0.5mm can be used as a substitute.)

KR-7100P-02

Auto-start tool

Used for checking the auto-start


measurement range and for compensation of auto-start.

KR-7100P-05

Special tool

Used to remove the kerato ring unit.

RM-8000-01

Tool test eyes for 0D,


10D, 15D,
20D, C-3D,
C-5D and C-8D

Used for confirming the measurement accuracy during assembly


inspection, for confirming the problem when the measured value is not
normal, and for inputting compensation values.

RM-8000-03

Tool holder

Mounted on the optical head and


used to mount the tools.

RM-8000-04

10D test eye

Mounted on the tool holder and


used for compensation of P2D.

Tool

Illustration

187

Application

R-TRK1P-0-0809-02

Usage
rank

Tool

RM-8000-07

Connector removal tool

Used for removing the connectors.

SL-01

Diopter telescope

Used to check and compensate


0D of the fixation target.

KR-01
+KR-8000-09

Kerato test eyes for


R6, R8, R9 and
two kinds of astigmatism

Used for confirming the measurement accuracy during assembly


inspection, for confirming the problem when the measured value is not
normal, and for inputting compensation values.

CT-10-06

Pressure tool

Used to adjust the CONTROL PCB


and pressure.

CT-50-05

Special tool

Used to remove/install the nozzle


unit.

CT-80-05

Alignment/applanation
tool fixing screw set

Used to fix the alignment tool and


the applanation light intensity
adjustment tool on the nozzle unit.

SP-3000P-01

Corneal thickness test


eyes for 0.4/0.5/0.6/
0.7/0.78mm

Used to check the corneal thickness


accuracy and to compensate the corneal thickness.

TRK-01

Anterior eye section


window glass tool

Used to compensate the accuracy.

Tool No.

Illustration

188

Application

R-TRK1P-0-0809-02

Usage
rank

Tool

TRK-02

Z compensation test eye

Used to adjust the corneal thickness


measurement LED light intensity
and to store the offset.

TRK-03

CCD
magnification
compensation tool

Used to compensate the CCD magnification.

TRK-04

XYZ alignment adjustment tool

Used to compensate the accuracy.

TRK-05

Applanation
system
adjustment tool

Used to adjust the applanation


detection circuit sensitivity.

TRK-07

CCD alignment adjustment tool

Used to adjust the CCD camera


position.

TRK-08

Attachment

Used to mount the test eyes of KR01+KR-8000-09, KR-8000-01 and


SP-3000-01 on the chinrest of TRK1P.

TRK-9

Checking test eye

Used to check the measurement


accuracy.
(After the conversion value is
decided, this will be common to CT60-05.)

TRK-10

Repair test eye

Used to check the measurement


accuracy, to compensate data and to
repair the instrument.
(After the conversion value is
decided, this will be common to CT60-06.)

Tool No.

Illustration

189

Application

R-TRK1P-0-0809-02

Tool No.

Usage
rank

Tool

Illustration

Application

TRK-11

Tool test eyes for 0D,


10D, 15D,
20D, C-2D,
C-4D and C-6D

Used for confirming the measurement accuracy during assembly


inspection, for confirming the problem when the measured value is not
normal, and for inputting compensation values.
(When this tool is used exclusively
for TRK-1P)

TRK-12

Kerato test eyes for


R6, R8, R9 and
two kinds of astigmatism

Used for confirming the measurement accuracy during assembly


inspection, for confirming the problem when the measured value is not
normal, and for inputting compensation values.
(When this tool is used exclusively
for TRK-1P)

TRK-13

Corneal thickness test


eyes for 0.4/0.5/0.6/
0.7/0.78mm

Used to check the corneal thickness


accuracy and to compensate the corneal thickness.
(When this tool is used exclusively
for TRK-1P)

TRK-14

CT unit extension cable


set

Used to adjust the CT unit.

4-2 General repair tools

Synchroscope
Multimeter
Optical power meter
Set of the six Phillips screwdrivers
Set of the four flat-head screwdrivers
3mm wooden-handle screwdriver
Hexagonal wrench set

190

R-TRK1P-0-0809-02

Printed in Japan 0808-300TH 1