1/30
As low as
the power consumption
(compared to previous models)
Waterproof
Construction
1000 Data
Points Memory
P30 Series
Features 8meters
including
pH
dual channel ION
modes
ORP
DO
EC
Lineup
pH
(If you want an electrode that is not tted as standard, please place separate orders for
the main unit only and the electrode you need.)
Temperature
Temperature
DO
Portable pH Meter
Portable DO Meter
DO-31P
HM-30P
pH
ORP
ch1
Temperature
ch2
Portable pH Meter
HM-31P
Electrical
Conductivity
pH
Electrical
Resistivity
ORP
Salinity
Temperature
Temperature
Portable Electrical
Conductivity/pH Meter
WM-32EP
c h D u al
an
t y p n el
e
ORP
Temperature
RM-30P
ch1
pH
ORP
Ion
Temperature
ch2
pH
ORP
Ion
Temperature
IM-32P
c h D u al
an
t y p n el
Electrical
Resistivity
Salinity
Temperature
ch1
Portable Electrical
Conductivity Meter
ch2
CM-31P
DO
pH
Temperature
ORP
Temperature
DM-32P
CM-31P
CM-31P-W
c h D u al
an
t y p n el
e
LEA
FREE SEN
D-
LEAD-FREE
R
SO
Corresponding sensors
GST-2729C CT-57101B
GST-2739C CT-57101A
PST-2729C CT-57101C
PST-2739C
Our original built-in oat for monitoring the internal solution concentration
allows the user to instantly recognize when the solution needs to be replaced.
(pH/ORP)
Cal-memo
pH/ORP
Electrode
Use
pH combination electrode
General environment/
Cal-memo (Calibration Memo)
immersion
GST-2729C
Waterproof type
Remarks
Electrode with HM-31P/WM-32EP
tted as standard (Lead length: 1 m)
Approval of type by Measurement Law
Electrode with HM-30P tted as
standard
(Lead length: 1 m)
Approval of type by Measurement Law
pH combination electrode
Organic solventCal-memo (Calibration Memo)
containing solution
ELP-031
pH014
0100
1m
pH combination electrode
Fluorinated acid
Cal-memo (Calibration Memo)
solution-resistance*1
ELP-040
pH212
050
1m
General environment/
ORP combination electrode
Cal-memo (Calibration Memo) immersion
PST-2729C
Waterproof type
0
2000mV
0100
General environment/
immersion
0
2000mV
0100
Waterproof type
Product Name
Code number
pH4.01 standard solution, 500 mL
143F191
pH6.86 standard solution, 500 mL
143F192
pH9.18 standard solution, 500 mL
143F193
Reference electrode internal
solution, 50 mL (4 bottles)
RE-4-20
(3.3 mol/L KCl solution)
ORP check solution
(pH4.01 standard solution, 500 mL
143F196
+ quinhydron powder, 5 packs)
Abrasive for ORP electrode, 10mL
AO-001
1mStandard
5m
11m
1mStandard Electrode with RM-30P tted as
5m
standard
(Lead length: 1 m)
11m
1 The glass electrode is affected by uorinated acid solution. However, because this product is a replaceable type glass
electrode tip, a reduction in operating costs can be expected. In regards to measuring the 1% uorinated acid solution (at
25 , for 1 min.), approximately 1000 measurements can be performed.
Electrical Conductivity
Cell
Use
Meas.Range
(Cell Constant)
0.1mS/m
Electrical conductivity cell
General environment/
Cal-memo (Calibration Memo)
10S/m250m -1
immersion
CT-27112B
080
Waterproof type
pure water
Electrical conductivity cell
Cal-memo (Calibration Memo) measurement/
CT-27111D
ow-through type
5S/m
20mS/m1m -1
080
100S/m
Electrical conductivity cell
General environment/
Cal-memo (Calibration Memo)
10S/m100m -1
tabletop use
CT-57101B
0100
High electrical
1mS/m
Electrical conductivity cell
Cal-memo (Calibration Memo) conductivity/tabletop 100S/m
1000m -1
CT-57101A
use
0100
Low electrical
5S/m
Electrical conductivity cell
Cal-memo (Calibration Memo) conductivity/tabletop 1S/m10m -1
CT-57101C
use
0100
Lead
Remarks
Length
Standard
1m
Cell with CM-31P/WM-32EP tted as
5m
standard (Lead length: 1 m)
11m
Cell with CM-31P-W tted as standard
1m
<Flow cell sold separately.>*2
Note: Cannot be connected to WM-32EP.
Product Name
Electrical conductivity cell check solution,
C solution, 100 mL (4 bottles)
Electrical conductivity cell check solution,
B solution,250 mL (2 bottles)
Flow cell (made of PP)
Flow cell (made of SUS)
Code number
0BI00001
0BI00002
CEF-22A
CEF-23A
1m
1m
1m
2 If you order the full CM-31P-W set, a ow cell is also tted as standard.
DO
Electrode
Use
Measuring range
If a standard membrane is
DO electrode
General environment/ used:
Cal-memo (Calibration Memo)
020mg/L
immersion
OE-270AA
If a high concentration
Waterproof type
membrane is used:
050mg/L
DO electrode
General environment/ 050
Cal-memo (Calibration Memo)
immersion
(High concentration membrane
OE-570BA
Waterproof type
set is sold separately.)
DO electrode
Cal-memo (Calibration Memo) Incubator bottle
OE-470AA
DO electrode
Cal-memo (Calibration Memo) Incubator bottle
OE-470BA
Remarks
Electrode with DO-31P/DM32P tted as standard
(Lead length: 3 m)
Can be used to conduct zero
ow rate measurements
1m
1m
020mg/L
Product Name
Code
Remarks
DO module
OEC-002
0CC00001
0CC00002
0CC00023
0CC00024
0CC00003
OCT-2502
0CC00022
0SM00002
0BG00007
143A030
Lead
3mStandard
5m
11m
3mStandard
5m
11m
Ion
Cal-memo
Notes: (1) The ion electrode is not provided for waterproof function and temperature measurement function. Measurable
solution temperature range is 0 50 .
(2) The batch measurement method is primarily used to conduct ion measurements. This method is conducted after
sampling, which uses beakers and other apparatuses.
(3) In addition to the electrode, the standard solution, ionic strength adjuster, and reference electrode external
solution are also required for conducting ion measurements.
(4) Make sure to contact us before you conduct ion measurements. We ask this because in certain cases it can be
difcult to conduct ion measurements, such as when there are coexisting ions in the sample.
Electrode name
Fluoride ion combination electrode
F-2021
Chloride ion combination electrode
CL-2021
Bromide ion combination electrode
BR-2021
Iodide ion combination electrode
I-2021
Cyanide ion combination electrode
CN-2021
Nitrate ion combination electrode
N-2031
Sulde ion combination electrode
S-2021
Sodium ion combination electrode
NA-2011
Potassium ion combination electrode
K-2031
Calcium ion combination electrode
CA-2031
Cadmium ion combination electrode
CD-2021
Copper ion combination electrode
CU-2021
Silver ion combination
electrodeAG-2021
Ammonia combination electrode
AE-2041
Carbon dioxide combination electrode
CE-2041
pH12 or more
Dissolved gas
1.491,490mg/L
Mg2+
Ca2+
Zn2+
NH4+
K+
Li+=103
H+=102 NH4+=3102 Na+=2103 Li+=104
Pb2+
Zn2+=101Mn2+=102
Cu2+
Mg2+
Cd2+
Ba2+
Fe2+=103 Ni2+=104
Hg2+
Ag+
Cu2+=Cannot coexist
Pb2+
Fe3+=100Cr3+=102 Na+
K+
Mg2+
Ca2+
Zn2+
Al3+=105
Ag+
Hg2+=Cannot coexist
Fe3+=101 Al3+=101 Cr3+=102 Ni2+=103Na+
Mg2+
Ca2+=104
Hg2+=Cannot coexistMg2+=103
Ca2+
Cu2+
Pb2+
Cd2+
Zn2+=104 Na+
K+=106
Volatile amines
Dissolved gas : Volatile weak acidAirborne gas : Acid gas
Note : A cell for calibration (CGC-202L) and an adapter for calibration
(6791140K) sold separately.
Effect of coexistent ion (selectivity coefcient for 0.1 mol/L ion concentration)
If an ion coexists in the solution, it can cause data errors when measuring the targeted ion. The effects of the coexistent ion are shown here.
A selectivity coefcient of 10x means that if the solution contains a coexistent ion that is 10 times greater than the value of the targeted ion that is measured, an error occurs in which the
value of the targeted ion equals the coexistent ion value.
If the concentration level of the coexistent ion is high enough to affect the measured values, we recommend conducting pretreatment in order to prevent interference.
Product Name
Exchange liquid junction for ion sensor (10 pieces)
Exchange membrane for ammonia electrode (10 sheets)
Membrane cartridge for carbon dioxide gas electrode (4 pieces)
Calibration cell for carbon dioxide electrode
Calibration adapter
Code
0LF00001
AE-FILM
CTC-211
CGC-202L
6791140K
Remarks
143F237
For the internal solutions of all ion combination electrodes (except AE /CE-2041).
Reference external solution for CA-2031 and I/S//F-2021.
143F238
143F239
0BG00005
143D042
143E031
143A281
143C483
143H091
CN-100
143B500
143B482
143B481
143A041
143A042
143C486
143C487
143F391
143F393
143F392
143D044
143A338
143A334
143A335
143A336
143A337
143A333
143A279
143A280
143A340
143A339
143D045
143A332
Note : We do not sell silver ion standard and sulde ion standard solutions. If you need these solutions, you must prepare yourself, following the steps listed in the instruction manual.
*Toxic : Exercise caution when handling.
P30 Series
Product Name
Portable pH
Meter
Portable
ORP Meter
Portable pH
Meter
Model Name
HM-30P
RM-30P
HM-31P
Portable Electrical
Conductivity Meter
CM-31P
DO-31P
WM-32EP
IM-32P
DM-32P
Membrane
type galvanic
cell method
pH : Glass electrode
method
Electrical conductivity :
AC two-electrode
method
DO
Electrical Conductivity
pH
ORPIon
DO
pH
ORP
pH
ORPIon
pH
ORP
CM-31P-W
Measuring
method
Glass
electrode
method
Platinum
electrode
method
pH
ORP
Custom LCD
Display
Sensor
Connection
ch1
pHORP
pH :
0.00-14.00
Temperature :
0-100.0
ORP :
0-2000mV
Temperature :
0-100.0
pH :
0.00-14.00
ORP :
0-2000mV
Temperature:
0-100.0
pH : 0.00-14.00
[If standard
ORP : 0-2000mV
membrane is used]
Temperature :
DO :
0-100.0
pH : 0.00-14.00
0-20.00mg/L
ORP :
Saturation rate :
[If
standard
cell
is
used]
0- 2000mV
0-200%
Electrical conductivity :
Temperature :
Temperature :
0.1mS/m-10 S/m
0-100.0
0-50.0
Electrical resistivity : Ion :
[If cell for pure water is
0.1
m-10k
m
[If high
differs
used]
Salinity (NaCl equivalent
according to
Electrical conductivity : concentration
obtained by electrical
membrane
is
used]
the electrode
5S/m-20mS/m
conductivity) : 0-4.00%
DO :
that is used.
Electrical resistivity :
0-50.0mg/L
Temperature : 0-80.0 (Temperature
50m-182km
measuring function
Saturation rate :
is not provided.)
Temperature : 0-80.0 0-500%
If the cell for tabletop
use is used, the
Temperature :
measuring range differs
If the cell for tabletop
0-50.0
according to the cell
use is used, the measuring
range differs according to
the cell that is used.
Display Range
pH :
-2.00-16.00
Temperature :
-5-110.0C
ORP :
0-2200mV
Temperature :
-5-110.0
pH :
-2.00-16.00
ORP :
0-2200mV
Temperature :
-5-110.0
Electrical conductivity :
0-200.0S/m
0-2.000mS/m
0-20.00mS/m
0-200.0mS/m
0-2.000S/m
0-20.00S/m
0-200.0S/m
Electrical resistivity :
0.005-2.000m
0-20.00m
0-200.0m
0-2.000km
0-20.00km
0-200.0km
0-2.000Mm
0-20.00Mm
Salinity(NaCl) :
0-4.04%
Temperature :
-5-110.0
that is used.
[If standard
membrane is used]
DO :
0-22.00mg/L
Saturation rate :
0-220%
Electrical Conductivity/
Resistivity Unit
Switching
Repeatability
(Main unit)
Auto/manual
Electrical
conductivity :
0.5%FS
pH : 0.02pH Electrical
resistivity :
ORP : 2mV
Temperature : 0.5%FS
Salinity :
0.2
0.5%FS
Temperature :
0.2
Auto/manual
[If standard
membrane is used]
DO : 0.03mg/L
Saturation rate :
2%
[If high concentration
membrane is used]
DO : 0.2mg/L
Saturation rate :
2%
Temperature :
0.2
pH : 0.02pH
ORP : 2mV
Electrical
conductivity :
0.5%FS
Electrical
resistivity:
0.5%FS
Salinity : 0.5% FS
Temperature :
0.2
pH : 0.02pH
ORP : 2mV
[If standard
pH : 0.02 pH membrane is used]
ORP : 2 mV DO: 0.03mg/L
Ion : 0.5%FS Saturation rate : 2%
Temperature : [If high concentration
membrane is used]
0.2
DO : 0.2mg/L
Saturation rate : 2%
Temperature : 0.2
pH : Auto/Manual
Electrical Conductivity/
Resistivity :
Switch setting between
Auto/Manual/None
Auto
Reference temperature :
25
Temperature coefcient
: 0-9.99%
Calibration
Capable of
three-point
calibration
Cell constant
calibration
Auto/Manual
Temperature
compensation method : Not applied to
ORP and ion
Linear method
Reference temperature :
25
Temperature coefcient
: 0-9.99%
pH :
Auto/Manual
DO : Auto
(optional setting)
(optional setting)
Capable of
three-point
calibration
pH :
-2.00-16.00
ORP :
0-2200mV
[If standard
membrane is used]
DO :
0-22.00mg/L
Saturation rate :
0-220%
Temperature
compensation method :
Auto/Manual
[If standard
membrane is used]
DO : 0-20.00mg/L
Saturation rate :
0-200%
Temperature :
0-50.0
Temperature
Compensation
pH : -2.00-16.00
ORP : 0-2200mV
Electrical conductivity :
0-200.0S/m
0-2.000mS/m
0-20.00mS/m
0-200.0mS/m
0-2.000S/m
0-20.00S/m
0-200.0S/m
Electrical resistivity :
0.005-2.000m
0-20.00m
0-200.0m
0-2.000km
0-20.00km
0-200.0km
0-2.000Mm
0-20.00Mm
Salinity (NaCl) : 0-4.04%
Temperature : -5-110.0
pH : 0.00-14.00
ORP :
0-2000mV
Temperature :
0-100.0
pH :
-2.00-16.00
ORP :
[If high
0-2200 mV
concentration
Ion :
membrane is used]
0.0g/L-999 g/L [If high
DO :
0-55.0mg/L
Temperature : concentration
membrane is used]
-5-110.0
Saturation rate :
DO :
0-550%
0-55.0mg/L
Temperature :
Saturation rate :
-5-110.0
0-550%
* In regards to the
Temperature :
In regards to the range,
range, the electrical
-5-110.0
the electrical conductivity/
conductivity/resistivity
resistivity differs according
to the cell that is used.
Electrical Conductivity/
Resistivity Range Switching
pH : Glass
pH : Glass
electrode method
electrode method
DO : Membrane
Ion : Ion electrode
type galvanic cell
method
method
ch2
Measuring
Range
pH : Glass
electrode method AC two-electrode
ORP : Platinum
method
electrode method
Zero/span
calibration
pH : Capable of three-point
pH/ion :
calibration
Capable of threeElectrical conductivity :
point calibration
Cell constant calibration
Product Name
Portable pH
Meter
Portable ORP
Meter
Portable pH
Meter
Model Name
HM-30P
RM-30P
HM-31P
Portable Electrical
Conductivity Meter
Portable DO
Meter
Portable Electrical
Conductivity/pH
Meter
DO-31P
WM-32EP
CM-31P
IM-32P
DM-32P
CM-31P-W
One-point calibration
Temperature Calibration
Correction function
(Called by Key Input)
Salinity correction
Atmospheric pressure
correction
(DO)
Salinity correction
Atmospheric pressure
correction
Data Memory
Clock Function
Provided (Interval: The interval can be specied between 1 sec.-99 min. 59 sec. or 2 sec.-99hr. 59 min.)
Printing Function
RS-232C
Interface
Number of Outputs/
Items
Analog
Output
Connectable
Devices
Communication
Specications
ch1
ch2
Output
Specications
Electrical conductivity/
electrical resistivity/
pH : 700mV salinity :
(pH0-14)
0-1V FS
(each range)
ORP : 1 V
(0-2000mV) Range :
Temperature : 100mV/range
0-1V (0-100 ) Temperature :
0-1 V (0-100 )
Number of outputs : 2
Measured
value and
temperature
Number of outputs : 3
Measured value,
temperature,
and range (only
for electrical
conductivity/
resistivity)
Parity: None
pH : 700mV
(pH0-14)
ORP : 1V
DO/saturation (0-2000mV)
rate :
resistivity/
0-1V FS
salinity :
(each range)
0-1V FS
Temperature : (each range)
0-1 V (0-100 ) Range :
100 mV/range
Temperature :
0-1V (0-100 )
pH : 700mV
(pH0-14)
pH : 700mV ORP : 1V
(pH0-14)
(0-2000mV)
ORP : 1V
DO/saturation
(0-2000mV) rate :
Temperature : 0-1V FS
0-1V (0-100 ) (each range)
Temperature :
0-1V (0-100 )
Waterproof Construction
IP67 (Enabled if the sensor is connected and if the external I/O portions are masked) (Can be immersed in water for 1m and 30 min.)
Performance Compensation
Temperature/humidity
Power Source
Power Consumption
Battery Life
Approx. 0.014W
Approx. 400hours4
Approx. 600hours
Approx. 1500hours
Approx. 0.014W
Approx. 400hours4
Outside Dimensions
Mass
Approx. 280g
(Including Batteries)
Approx. 300g
1) If the sample is earthed, make sure to use RS-232C and analog output in a insulated condition.
If you want to simultaneously (realtime) use RS-232C interface and analog output, you must have the special option cable. Please contact us for details.
2) The power consumption (consumption current) values shown are for when option devices (e.g. PC, printer) are not connected. If option devices are connected, the power consumption might
be approximately twice as high as the values shown, depending on the model.
3) Except for when the DO electrode with the stirring function is connected.
Standard Accessories
Product Name
Portable pH
Meter
Portable ORP
Meter
Portable pH
Meter
Model Name
HM-30P
RM-30P
HM-31P
Portable Electrical
Conductivity Meter
Portable DO
Meter
Portable Electrical
Conductivity/pH
Meter
Portable ion/
pH Meter
Portable DO/
pH Meter
DO-31P
WM-32EP
IM-32P
DM-32P
CM-31P
CM-31P-W
Standard
Accessories
[CM-31P]
Electrical
Electrical
conductivity
conductivity cell
DO electrode
CT-27112B
cell
ORP
OE-270AA
pH combination
pH combination (Lead length : 1m)
pH combination
CT-27112B
combination
DO electrode
(Lead length : 3m)
electrode
electrode
electrode
(Lead length : 1m)
[CM-31P-W]
electrode
OE-270AA
GST-2739C
GST-2729C
GST-2729C
pH electrode
Electrical
PST-2739C
(Lead length : 3m) pH combination
(Lead length : 1m)
(Lead length : 1m) conductivity cell
(Lead length : 1m)
GST-2729C
(Lead length : 1m)
electrode
(Lead length : 1m)
CT-27111D
GST-2729C
Flow cell made of PP
(Lead length : 1m)
CEF-22A
pH4.01 standard
solution (100mL)
pH6.86 standard
solution (100mL)
3.3 mol/L-KCl
solution (50mL)
Polybeaker
(50mL) (3pieces)
Polybeaker
(50mL) (3pieces)
pH4.01 standard
solution (100mL)
pH4.01 standard
solution (100mL)
pH4.01 standard
solution (100mL)
pH4.01 standard
solution (100mL)
pH6.86 standard
solution (100mL)
pH6.86 standard
solution (100mL)
pH6.86 standard
solution (100mL)
pH6.86 standard
solution (100mL)
3.3 mol/L-KCl
solution (50mL)
3.3 mol/L-KCl
solution (50mL)
3.3 mol/L-KCl
solution (50mL)
3.3 mol/L-KCl
solution (50mL)
Polybeaker
(50mL) (3pieces)
Polybeaker
Polybeaker
Polybeaker
(50mL) (3pieces) (50mL) (3pieces) (50mL) (3pieces)
AA alkaline batteries (for initial operation) ( 2 pieces), hand strap, instruction manual
Options
Code number
118N062
Remarks
For PC connection. Lead length: 2 m.
Code number
118N063
Remarks
External printer
Code number
External printer
(with connecting cable)
EPS-P30
P000119
0RD00001
118N061
Remarks
Compact sized printer with chart width of approx. 60 mm.
Ordinary printing level is sufcient for long-term data storage
(Cannot connect to HM-30P and RM-30P.)
Code number
Remarks
Ask
AC adapter
Electrode stand
(with column and stopper)
Electrode holder
6948810K
0IB00001
0IB00007
Electrode attachment(G)
0IB00004
Electrode stand/
holder/attachment
Carrying case
Code number
Stick holder
0IB00009
0IB00010
Anchor (AN-21P)
0IC00001
Remarks
This product provides a lead length of 5 m or more for waterproof
sensors that are immersed.
If you have trouble reaching a measurement point, you use this
product to safely measure from a position that is more accessible.
This product provides a lead length of 5 m or more for waterproof
sensors that are immersed.
Two sensors can be attached.
Can be used for waterproof sensors that are immersed. Anchor for
submerging.
1SUS rope
0IZ00002
This case allows you to store and carry the main unit, sensor, and
0DA00001 other accessories, such as the standard solution.
(comes with shoulder belt)
This portable soft case allows you to store the main unit when it is
SC-10P
connected to a sensor.
Soft case