Beruflich Dokumente
Kultur Dokumente
I2
I1
I3
I4
1
Light Intensity I( )
I 0 I1 I 2 I 3
=
=
=
=n
I1 I 2 I 3 I 4
0.8
0.6
Ii = I o ni
0.4
Ii = I o10i logn 10
0.2
0
0
Light pathlength (l )
Io
Absorbance = A = log = cl
I
I
Transmittance = T = = 10 A
I0
A
0
1
2
3
T
1
0.1
0.01
0
0
A
c=
l
Absorption Spectra
Light absorption depends on wavelength of light
Light has dual nature
1. Wave
0.4
c
Absorbance
A = cl
wavelength
c speed of light wave propagation
= c/ frequency
2. Particle
Energy of light particle:
E = h =
0.3
0.2
0.1
0.0
400
500
600
700
Wavelength, nm
hc
Vertical transitions
are most probable
2) Franck-Condon principle:
electronic transition is much faster than
rearrangement of nuclei the distance
between nuclei is not able to change
during electronic transition
Vertical transition results in occupation of higher vibrational energy
levels Vibrational energy dissipates producing heat Light
absorption is always accompanied by sample heating
Absorption spectra
consist of broad
bands
N
N
H
O
O
Heme
Fe
N
N
N
N
His F8
Oxygenation of Hemoglobin
1.2
1.2
Hemoglobin
1
Oxyhemoglobin
Absorption
Absorbance
1
0.8
0.6
0.4
0.2
0
300
Isosbestic points
0.8
0.6
0.4
0.2
400
500
Wavelength, nm
600
700
0
300
400
500
Wavelength, nm
Analysis of a Mixture
Absorbance is an additive function Absorbance of a mixture is a
sum of the absorbances of the components:
A = X [ X ] l + Y [Y ] l + Z [ Z ] l + ...
1
Ab so rb an ce
1.2
O
0.8
0.6
0.4
0.2
0
300
A1 = [ H ]l + [O]l
Absorption of the mixture at wavelength 2:
H
1
O
1
350
400
450
500
Wav e le n g th , n m
A2 = 2H [ H ]l + 2O [O]l
To solve this system of equation with regards to [H] and [O] we
must:
1) measure the molar absorptivities of pure hemoglobin and
H
H
O
O
,
and
A1 1O [O]l
[H ] =
1H l
Substitute [H] in equation 2:
-1
Molar A bsorptivity, M cm
-1
100000
80000
60000
40000
20000
0
300
350
400
Wavelength, nm
A2 = 2H
A1 1O [O]l
O
l
+
2 [O ]l
H
1 l
A21H A1 2H
Solve last expression with regards to [O]: [O ] = O H
( 2 1 2H 1O )l
A1 2O A21O
Using expressions 3 and 4: [ H ] = O H
( 2 1 2H 1O )l
450
500
Spectrophotometers
Cuvet
Broad
spectrum
Narrow
line
Monochromator
Detector
I0
S2
Internal
conversion
10-12 s
2
1
0
S1
Inter-system conversion
The rate varies in a large range
The influence of heavy metals
Light
absorption
10-15 s
hA1
S0
2
1
0
Fluorescence
10-8 s
hA2
hF
2
1
0
T1
Phosphorescence
>10-6 s
Fluorescence Lifetime
Fluorescence intensity decays exponentially:
I = I0e-kt
where k is the monomolecular rate constant of fluorescence decay,
which can be determined experimentally from the kinetics of decay
Definition: lifetime of fluorescence:
= 1/k
do not confuse with half-life time, t1/2= ln2/k
/n = kr/(kr + knr)
Quantum Yield
Quantum yield:
Q=
Problem:
kr typical for fluorescence is 108 s-1
knr typical is ~ 109 s-1
Calculate Q typical for fluorescence
Solution: Q = kr/(kr+ knr) = 108/(108 + 109) 0.1
Quenching of Fluorescence
Two major mechanisms of quenching exist:
- Dynamic or collisional (depends on the diffusion of fluorophore
and quencher)
- Static (the formation of non-radiative complexes with quenchers
that does not depend on the diffusion of fluorophore and quencher)
Dynamic quenching is characterized by the Stern-Volmer eqn.:
I0/I = 1 + kq0[Q]
Where I0 and I are fluorescence intensities with and without
quencher, kq is a bimolecular rate constant of quenching (M-1s-1), 0
is the lifetime of the fluorophore in the absence of quenchers, [Q] is
the concentration of quencher.
Stern-Volmer Equation
*
f(t)
k = 0-1
I0 /I = 1 + kq0[Q] = 1 + KSV[Q]
KSV can be experimentally measured from the Stern-Volmer plot:
Higher
temp.
I0/I
Slope = KSV
1
0
[Q]
I0/I
1
0
[Q]
Efficiency of Quenching
The bimolecular rate constant of quenching, kq, is proportional to the
diffusion controlled bimolecular rate constant, kd. The coefficient of
proportionality is the efficiency of quenching :
kq = kd
1
Determination of : = kq / kd
kq can be calculated from kq = KSV/0
KSV can be determined from the SV plots
0 can be determined from the kinetics of fluorescence decay
kd can be calculated using the Smoluchovski equation:
4N A
kd =
( R f + Rq )( D f + Dq )
1000
where NA is Avogadro number, Rf and Rq are molecular radii of
the fluorophore and quencher, respectively, Df and Dq are diffusion
coefficients of the fluorophore and quencher, respectively
Step 2: finding kd
At 25C in water: DO2 = 2.5 10-5 cm2/s,
Assume that RO2 + Rtryp = 5
4N A
kd =
( RO + Rtryp )( DO + Dtryp ) = 1.2 1010 M 1 s 1
1000
2
Step 2: finding
= kq / kd = 1
E
Photon
Photon
ex
Polarization
unfavorable
for absorption
of photon
Photon
ex
ex
em
Note: 1. (I|| + 2I) is the total intensity of fluorescence, so that fluorescence anisotropy does
not depend on fluorescence intensity. 2 in the denominator shows that in a 3-D space there
are 2 indistinguishable directions perpendicular to ex
2. Maximum possible anisotropy of 1 (when I = 0) would be observed if: (i) the
fluorophore had ex = em , (ii) all fluorophore molecules were oriented favorably and
(ii) all fluorophore molecules were immobile
3. Maximum fluorescence anisotropy of randomly oriented solution of fluorophore is
0.4 due to excitation photoselection
4. Light scattering and reflection can produce r = 1
||
ex
em ||
em 2
2
em
em 1
45
Solution:
- To calculate the anisotropy we need to know I|| and I.
- I|| and I are dependent on the projections of em on the ||
(em ||) and axes (em 1 and em 2):
I1= a em 1
I2= a em 2
I|| = a em ||
where a is a constant
- The projections em 1 and em 2 are identical due to the
two axes being indistinguishable from each another,
thus:
I|| = a em || = a em cos
I = I1= I2 = a em sin cos 45 = 0.5 21/2 a em sin
I|| I
em
1. ex em
ex
t=0
em
ex
em
t = 5 ns
ex
em
t = 10 ns
Detector 1
B
B
E
E
E ||
E
Laser
Detector 2
P + F
PF
koff
Kd =
[ P ]eq [ F ]eq
[ P F ]eq
where [P]eq, [F]eq, and [PF]eq are equilibrium concentrations of free protein, free
fluorophore, and the complex, respectively. We will express [P]eq[F]eq [PF]eq
through fluorescence anisotropy and total concentrations of protein and
fluorophore, [P]0 and [F]0 using two principles:
1) Anisotropy is an additive function, that is
r = rF
2) Mass balance requires that:
and
[F]eq
[F]0
+ rPF
[P F]eq
[F]0
[F]eq
r = rF
+ rPF
= rF
+ rPF
= rF
+ rPF 1
[F]0
[F]0
[F]0
[F]0
[F]0
[F]0
[F]eq
[F]eq
[F]eq
[F]eq r rPF
rF
+ rPF rPF
= rPF +
=
=R
( rF rPF )
[F]0
[F]0
[F]0
[F]0 rF rPF
[F]eq
[F]eq
[F]0 -[F]eq
[F]eq
Kd =
[P]0 (1 R)
(1/R 1)
(1/R 1)
where R =
Kd =
[ P ]eq [ F ]eq
[ P F ]eq
where R =
r rPF
, 0 < R <1
rF rPF
r rPF
rF rPF
we get:
[F]eq
[F]0
+ rPF
[P F]eq
[F]0
and
0<
[F]eq
[F]0
< 1, preferably
[F]eq
[F]0
0.5
[F]eq
[F]0
+ rPF
[P F]eq
[F]0
= rF 0 + rPF 1 = rPF
[F]eq
[F]0
+ PF
[P F]eq
[F]0
1 1 [F]eq
1 [P F]eq
=
+
t tF [F]0 tPF [F]0
v
L
=
E tE
1 R0
k FRET ( r ) = D
0 r
where
0D is the lifetime of D in the absence of A,
R0 is the Frster radius, and
r is the distance between the D and A.
1 R0
k FRET ( r ) = D
0 r
1 1
1
k FRET ( r = R0 ) = D = D
0 1 0
Typical R0 are 20 90
hv
k=
0D
k FRET ( r = R0 ) =
0D
heat
Calculation of R0
9000(ln 10) QD
4
R0 = 6
F
(
d
D
5
4
128 N A n
0
2
where
- QD is the quantum yield of D in the absence of A
- n is the refractive index of the medium (1.4 for biomolecules in aqueous solns.)
- NA is Avogadros number
- 2 is the orientation factor (2/3 for randomly oriented molecules)
- FD( ) is the normalized spectrum of fluorescence of D and () is the
extinction coefficient of A
R06
r=6
=
D
6
k FRET ( r ) 0
R06
R06
=
= 6 (1 / 1) R06
k
1
6
D
0
1/ 1
1/ 1
= kFRET/(kFRET + k)
k = 1/0D
Objectives
Magnification: The ratio between the sizes of the image and
object when object is at the working distance.
Working distance: Distance from the objective to the object at
which the image will be in focus
Angular Aperture:
AA = sin
Numerical Aperture:
NA = n(sin )
where n is the refractive index of the imaging medium between
the front lens of the objective and the specimen cover glass, a
value that ranges from 1.00 for air to 1.51 for specialized
immersion oils.
specimen
Working distance
dS
d
Objective lens
R
The part of light collected by a lens is the ratio between the solid body
S that defines the aperture and the solid body Ssph of the sphere:
S/Ssph = S/4l2
dS = l d Rd = l d (l sin )d = l 2 sin d d
2
S =l
(sin d )d = l
0 0
2
2
(cos
|
0 )d =
S/Ssph =
S/4l2
Angular aperture
Band-pass filters:
Transmittance, % Transmittance, %
Cut-off filters:
Filters
Wavelength
Absorption spectra are important
for estimation of quality of
cutting light
Wavelength
Mirrors
- In household mirrors, the reflecting surface is protected by
glass, we look through glass.
- In optical mirrors the reflective surface is exposed; therefore it
should be protected from oxidation by a layer of dielectric
coating
Domestic mirror
Optical mirror
Filter Cubes
Filter cubes are used in fluorescence microscopes and fluorescence
detection systems to allow excitation of fluorescence and collection
of fluorescence light with the same objective lens
specimen
exciter
EYE or detector
dichroic mirror
emitter
+1 kV
Modes of operation:
1) Current (higher noise, higher dynamic range)
2) Photon counting (noise reduction due to the discrimination of pulses that
correspond to single electrons born on the photocathode. Lower dynamic
range due to the need to resolve between single photons. The pulse width is
50 ns (5 10-8 s). The dynamic range is 0 105 counts/s.
Manufacturers: Hamamatsu (www.hamamatsu.com)