Valid as of software version:
V 1.06.XX (amplifier)
V 1.03.XX (communication)
PROline promass 80
Coriolis Mass Flow
Mesuring System
MEASURING
VALUES
SYSTEMMEASURING
UNITS
VALUES
QUICKSYSTEM
SETUPUNITS
QUICK SETUP
OPERATION
OPERATION
USER INTERFACE
USER INTERFACE
TOTALIZER
TOTALIZER
MASS FLOW
UNIT MASS FLOW
MASS FLOW
COMMISSIONING
UNIT MASS FLOW
COMMISSIONING
LANGUAGE
LANGUAGE
ASSIGN
LINE 1
ASSIGN LINE 1
ASSIGN
TOTALIZER
ASSIGN
TOTALIZER
TEMPERATURE
UNIT MASS
UNIT
VOLUME FLOW
UNIT VOLUME
ASSIGN
STATUS
ASSIGN
STATUS
OUTPUT
OUTPUT
ON-VALUE
ON-VALUE
STATUS
STATUS
INPUT
INPUT
ASSIGN STATUS
ASSIGN
STATUS
INPUT
INPUT
PROCESS
PARAMETER
PROCESS
SYSTEM
PARAMETER
PARAMETER
SYSTEM
SENSOR DATA
PARAMETER
SUPERVISION
SENSOR
DATA
SIMULATION SYST.
SUPERVISION
SENSOR VERSION
SIMULATION SYST.
AMP. VERSION
TAG
LF-CUTOFF
ASSIGN
INSTALLATION
LF-CUTOFF
DIRECTION
INSTALLATION
K-FACTOR
DIRECTION
ACTUAL SYSTEM
K-FACTOR
CONDITION
ACTUAL
SYSTEM
FAILSAFE
MODE
CONDITION
SERIAL
NUMBER
FAILSAFE MODE
SW-REV.
LINE 1
OVERFLOW
VALUE 20 mA
END VALUE
F
FREQUENCYVALUE FVALUE
LOW
FREQUENCY
SIMULATION
SIMULATION
STATUS
MIN. PULSE WIDTH
STATUS
TAG
TAG
DESCRIPTION
BUS ADDRESSE
AKTIVE LEVEL
DESCRIPTION
HART PROTOCOL
BUS ADDRESSE
OFF-VALUE
SENSOR
SENSOR VERSION
AMP.
SERIAL
NUMBER
TYPE
SENSOR
TYPE
AMP. VERSION
SW-REV.
AMP.
I/O-MODUL
TYPE
HART PROTOCOL
EPD
O
OFF-VALUE
POSITVE
ZERO
SYSTEM
RETURN
DAMPING
PO
POSITVE ZERO
NOMINAL
RETURN
DIAMETER
EPD
SYSTEM
SENSOR
TYPE DAMPING
NOMINAL
ASSIGN SYSTEM
ERROR SENSOR
ERROR
D
DIAMETER CATEGORY TYPE
VALUE
SIMULATION
PREVIOUS
SYST. ASS
ASSIGN
SYSTEM
SIMULATION
MEASURAND
CONDITION
ERROR
SIMULATION
TYPE
MEASURAND
I/O-MODUL
F LOW
OFF-VALUE
TIME CONSTANT
O
OFF-VALUE
TIME CONSTANT
AKTIVE LEVEL
ZERO
PREVIOUS
SYST.
CONDITION
POINT
VALUE 20 mA
E
END VALUE
ASSIGN
MEASURING
ZERO
MODE
POINT
UNIT SUM
UNIT SUM
VA
VALUE 0_4 mA
VALUE 0_4 mA
CURRENT SPAN
ON-VALUE
MEASURING
MODE
100% VALUE
LINE 2
O
OVERFLOW
CURRENT SPAN
ON-VALUE
LINE 2
100% VALUE
LINE 1
SUM
STATUS
STATUS
OUTPUT
OUTPUT
STATUS ACCESS
DEFINE PRIVATE
1
100% VALUE
100% VALUE
STATUS ACCESS
CODE
SUM
ASSIGN
FREQUENCY
FREQUENCY
TAG
NAME
NAME
ASSIGN
ASSIGN LINE 2
OPERATION
OPERATION
MODE
MODE
COMMUNICATION
DEF
DEFINE PRIVATE
CODE
ACCESS CODE
ACCESS
ASSIGNCODE
LINE
PULSE/FREQ.PULSE/FREQ.OUTPUT
OUTPUT
COMMUNICATION
UNIT VOLUME
DENSITY
ASSIGN CURRENT
ASSIGN CURRENT
TEMPERATURE
UNIT
VOLUME FLOW
UNIT MASS
CURRENT
CURRENT
OUTPUT
OUTPUT
DENSITY
VOLUME FLOW
VOLUME FLOW
SW-REV.-NO.
VALUE
SENSOR
S
SIMULATION
SW-REV.
I/O-MODUL
S
SW-REV.-NO.
SENSOR
SW-REV.
II/O-MODUL
ERROR
CATEGORY
Endress+Hauser
Table of Contents
Table of Contents
1
Group OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
10
11
12
13
Group COMMUNICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
14
15
16
17
Group SUPERVISION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
18
19
20
21
Factory settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
21.1 SI units (not for USA and Canada) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
21.2 US units (only for USA and Canada) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
22
Endress+Hauser
Table of Contents
Registered trademarks
HART
Registered trademark of HART Communication Foundation, Austin, USA
S-DAT, T-DAT, F-CHIP
Registered trademarks of Endress+Hauser Flowtec AG, Reinach, CH
4
Endress+Hauser
1.1
The function matrix is a two-level construct: the groups form one level, functions
the other.
The groups are the highest-level grouping of the control options for the measuring
device.
Each group comprises a number of functions.
You select a group in order to access the individual functions for controlling or
parameterizing the measuring device.
You will find an overview of the groups in the table of contents on Page 3 and in
the graphical representation of the function matrix on Page 6.
You will also find an overview of the functions on Page 6, complete with the page
references of the detailed function descriptions.
The descriptions of the individual functions start on Page 7.
Example of how to parameterize a function (in this case changing the language for
the user interface):
Entry into the function matrix (F-key).
Select the OPERATION group.
Select the LANGUAGE function, change the setting from ENGLISH to DEUTSCH
with OS and save F (all texts displayed on the user interface appear in German).
Exit the function matrix (ESC > 3 seconds).
Esc
Esc
>3s
Esc
Esc
E
E
E
F-x0xxxxxx-19-xx-xx-xx-000
Endress+Hauser
MEASURING VALUES
(p. 7)
SYSTEM UNITS
(p. 8)
OPERATION
(p. 15)
USER INTERFACE
(p. 17)
TOTALIZER 1/2
(p. 20)
SIM. STATUS IN
(p. 42)
HART PROTOCOL
(p. 44)
EMPTY PIPE DET.
(p. 46)
ON-VALUE
(p. 38)
ACTIVE LEVEL
(p. 42)
TAG DESCRIPTION
(p. 44)
ASSIGN STATUS
(p. 38)
ASSIGN STATUS IN
(p. 42)
TAG NAME
(p. 44)
LANGUAGE GROUP
(p. 57)
SW-REV. AMP.
(p. 57)
AMP. HW VERSION
(p. 57)
SW-REV. S-DAT
(p. 57)
SENSOR TYPE
(p. 57)
SERIAL NUMBER
(p. 57)
SENSOR VERSION
(p. 57)
SIM. MEASURAND
(p. 56)
SIMULAT. SYSTEM
(p. 56)
SW-REV. I/O
(p. 57)
ERROR CATEGORY
(p. 54)
DENSITY COEF. C 5
(p. 53)
DENSITY COEF. C 4
(p. 53)
DENSITY COEF. C 3
(p. 53)
DENSITY COEF. C 2
(p. 53)
SUPERVISION
(p. 54)
TEMP. COEF. KM
(p. 52)
NOMINAL DIAMETER
(p. 52)
ZERO POINT
(p. 52)
K-FACTOR
(p. 52)
TEMP. COEF. KM 2
(p. 52)
FLOW DAMPING
(p. 51)
DENSITY DAMPING
(p. 51)
SENSOR DATA
(p. 52)
ACTUAL STATUS
(p. 39)
TIME CONSTANT
(p. 39)
OFF-VALUE
(p. 38)
SIMULATION FREQ.
(p. 34)
MEASURING MODE
(p. 49)
SYSTEM PARAMETER
(p. 49)
PULSE WIDTH
(p. 35)
PULSE VALUE
(p. 35)
ASSIGN PULSE
(p. 34)
ASSIGN FREQUENCY
(p. 29)
OPERATION MODE
(p. 29)
VALUE F HIGH
(p. 30)
VALUE F LOW
(p. 30)
CURRENT SPAN
(p. 23)
ASSIGN CURRENT
(p. 23)
PROCESS PARAM.
(p. 45)
TIME CONSTANT
(p. 27)
VALUE 20 mA
(p. 24)
VALUE 0_4 mA
(p. 24)
FAILSAFE MODE
(p. 22)
PRESSURE
(p. 48)
COMMUNICATION
(p. 44)
TOTALIZER MODE
(p. 21)
UNIT TOTALIZER
(p. 21)
OVERFLOW
(p. 20)
SUM
(p. 20)
ASSIGN TOTALIZER
(p. 20)
PRESSURE MODE
(p. 48)
STATUS INPUT
(p. 42)
FORMAT
(p. 18)
100% VALUE
(p. 18)
100% VALUE
(p. 18)
ASSIGN LINE 2
(p. 17)
ASSIGN LINE 1
(p. 17)
RESTORE ORIGINAL
(p. 48)
STATUS OUTPUT
(p. 38)
ACCESS CODE
COUNTER (p. 16)
STATUS ACCESS
(p. 16)
ACCESS CODE
(p. 15)
LANGUAGE
(p. 15)
QS COMMISSION
(p. 13)
DENSITY ADJUST
(p. 48)
PULSE/FREQ. OUT.
(p. 29)
UNIT VOLUME
(p. 9)
UNIT MASS
(p. 8)
UNIT PRESSURE
(p. 12)
REFERENCE DENSITY
(p. 7)
DENSITY
(p. 7)
CORR.VOL.-FLOW
(p. 7)
VOLUME FLOW
(p. 7)
MASS FLOW
(p. 7)
MEASURE FLUID
(p. 47)
HANDLING TOTALIZER
(p. 22)
QUICK SETUP
(p. 13)
SIMULATION PULSE
(p. 37)
FAILSAFE MODE
(p. 36)
ALARM DELAY
(p. 55)
CAL. COEF. KD 1
(p. 52)
EPD RESPONSETIME
(p. 46)
SYSTEM RESET
(p. 55)
CAL. COEF. KD 2
(p. 52)
FAILSAFE MODE
(p. 33)
TIME CONSTANT
(p. 32)
SIMULATION CURR.
(p. 27)
BACKLIGHT
(p. 19)
ACTUAL CURRENT
(p. 27)
CONTRAST LCD
(p. 19)
UNIT DENSITY
(p. 11)
TEMP. COEF. KT
(p. 52)
DEVICE ID
(p. 44)
OUTPUT SIGNAL
(p. 36)
OUTPUT SIGNAL
(p. 32)
FAILSAFE MODE
(p. 27)
RESET TOTAL.
(p. 21)
DISPLAY DAMPING
(p. 19)
TEMPERATURE
(p. 7)
OPERATION HOURS
(p. 55)
DENSITY COEF. C 0
(p. 53)
VALUE SIMULATION
PULSE (p. 37)
FAILSAFE VALUE
(p. 33)
TEST DISPLAY
(p. 19)
UNIT TEMPERATURE
(p. 11)
DENSITY COEF. C 1
(p. 53)
ACTUAL FREQ.
(p. 33)
UNIT LENGTH
(p. 12)
Endress+Hauser
Note!
The engineering unit of the measured variable shown here can be set in the SYSTEM UNITS group
(see Page 8).
If the fluid in the pipe flows backwards, a negative sign prefixes the flow reading on the display.
MASS FLOW
In this function, the currently measured mass flow appears on the display.
Display shows:
5-digit floating-point number, including unit and sign
(e.g. 462.87 kg/h; 731.63 lb/min; etc.)
VOLUME FLOW
In this function, the currently measured volumetric flow appears on the display.
The volumetric flow is derived from the measured mass flowrate and the density of the fluid measured.
Display shows:
5-digit floating-point number, including unit and sign
(e.g. 5.5445 dm3/min; 1.4359 m3/h; 731.63 gal/d; etc.)
CORRECTED VOLUME
FLOW
The calculated corrected volume flow appears on the display. The calculated
corrected volume flow is derived from the measured mass flow and the reference density of the fluid (density at reference temperature, measured or fixed
entry).
User Interface:
5-digit floating-point number, including unit and sign
(e.g. 1.3549 Nm3/h; 7.9846 scm/day; etc.)
DENSITY
In this function, the currently measured density or the specific gravity appears
on the display.
Display shows:
5-digit fixed-point number, including unit
(e.g. 1.2345 kg/dm3; 993.5 kg/dm3; 1.0015 SG_20 C; etc.)
REFERENCE DENSITY
TEMPERATURE
Endress+Hauser
You can select the unit for the measured variable in this function group.
UNIT MASS FLOW
Use this function to select the unit for displaying the mass flow (mass/time).
The unit you select here is also valid for:
Current output
Frequency output
Switching points (limit value for mass flow, flow direction)
Low flow cut off
Options:
Metric:
gram g/s; g/min; g/h; g/day
kilogram kg/s; kg/min; kg/h; kg/day
Metric ton t/s; t/min; t/h; t/day
US:
ounce oz/s; oz/min; oz/h; oz/day
pound lb/s; lb/min; lb/h; lb/day
ton ton/s; ton/min; ton/h; ton/day
Factory setting:
Country dependent (kg/h or US lb/day)
UNIT MASS
Use this function to select the unit for displaying the mass.
The unit you select here is also valid for:
Pulse value (e.g. kg/p)
Options:
Metric g; kg; t
US oz; lb; ton
Factory setting:
Country dependent (kg or US lb)
Note!
The unit for the totalizer is independent of your choice here, it is selected
separately in the TOTALIZER group (see Page 20).
Endress+Hauser
Use this function to select the unit for the volume flow.
The unit you select here is also valid for:
Current output
Frequency output
Switching points (limit value for volume flow, flow direction)
Low flow cut off
Options:
Metric:
Cubic centimeter cm3/s; cm3/min; cm 3/h; cm 3/day
Cubic decimeter dm3/s; dm3/min; dm3/h; dm3/day
Cubic meter m3/s; m3/min; m3/h; m3/day
Milliliter ml/s; ml/min; ml/h; ml/day
Liter l/s; l/min; l/h; l/day
Hectoliter hl/s; hl/min; hl/h; hl/day
Megaliter Ml/s; Ml/min; Ml/h; Ml/day
US:
Cubic centimeter cc/s; cc/min; cc/h; cc/day
Acre foot af/s; af/min; af/h; af/day
Cubic foot ft3/s; ft3/min; ft3/h; ft3/day
Fluid ounce oz f/s; oz f/min; oz f/h; oz f/day
Gallon gal/s; gal/min; gal/h; gal/day
Million gallon Mgal/s; Mgal/min; Mgal/h; Mgal/day
Barrel (normal fluids: 31.5 gal/bbl) bbl/s; bbl/min; bbl/h; bbl/day
Barrel (beer: 31.0 gal/bbl) bbl/s; bbl/min; bbl/h; bbl/day
Barrel (petrochemicals: 42.0 gal/bbl) bbl/s; bbl/min; bbl/h; bbl/day
Barrel (filling tanks: 55.0 gal/bbl) bbl/s; bbl/min; bbl/h; bbl/day
Imperial:
Gallon gal/s; gal/min; gal/h; gal/day
Mega gallon Mgal/s; Mgal/min; Mgal/h; Mgal/day
Barrel (beer: 36.0 gal/bbl) bbl/s; bbl/min; bbl/h; bbl/day
Barrel (petrochemicals: 42.0 gal/bbl) bbl/s; bbl/min; bbl/h; bbl/day
Factory setting:
Country dependent (m3/h or US Mgal/day)
UNIT VOLUME
Note!
The unit for the totalizer is independent of your choice here, it is selected
separately in the TOTALIZER 1/2 group (see Page 20).
Endress+Hauser
Note!
The unit of the totalizers is independent of your choice here. The unit for each
totalizer is selected separately for the totalizer in question.
10
Endress+Hauser
Use this function to select the unit for displaying the fluid density.
The unit you select here is also valid for:
Current output
Frequency output
Switching points
Density adjustment value
Density responce value for EPD
Options:
Metric g/cm3; g/cc; kg/dm3; kg/l; kg/m3; SD 4 C, SD 15 C, SD 20 C;
SG 4 C, SG 15 C, SG 20 C
US lb/ft3; lb/gal; lb/bbl (normal fluids); lb/bbl (beer); lb/bbl (petrochemicals); lb/bbl (filling tanks)
Imperial lb/gal; lb/bbl (beer); lb/bbl (petrochemicals)
Factory setting:
kg/l
SD = Specific Density, SG = Specific Gravity
The specific density is the ratio of fluid density to water
(at water temperature = 4, 15, 20 C)
UNIT REFERENCE DEN- Use this function to select the unit for displaying the reference density.
SITY
The unit you select here is also valid for:
Current outputs
Frequency outputs
Relay switch points (limit value for density)
Fixed reference density (for calculation of corrected volume flow)
Options:
Metric:
kg/Nm3
kg/Nl
US:
g/Scc
kg/Sm3
lb/Scf
Factory setting:
kg/Nl
UNIT TEMPERATURE
Use this function to select the unit for displaying the temperature.
The unit you select here is also valid for:
Current output
Frequency output
Switching points
Options:
C (CELSIUS)
K (KELVIN)
F (FAHRENHEIT)
R (RANKINE)
Factory setting:
C (CELSIUS)
Endress+Hauser
11
Use this function to select the unit for the unit of length for nominal diameter.
The unit you select here is also valid for:
nominal diameter of the sensor (see the NOMINAL DIAMETER function on
Page 52).
Options:
MILLIMETER
INCH
Factory setting:
Country dependent (MILLIMETER or INCH)
UNIT PRESSURE
12
Endress+Hauser
QUICK SETUP
COMMISSION
Note!
The display returns to the QUICK SETUP COMMISSION cell if you press the ESC key
combination during programming of a parameter.
Only units not yet configured in the current Quick Setup are offered for selection
in each cycle. The unit for mass and volume is derived from the corresponding flow
unit.
The YES option remains visible until all the units have been parameterised.
NO is the only option displayed when no further units are available.
Only the outputs not yet configured in the current Quick Setup are offered for
selection in each cycle.
The YES option remains visible until all the outputs have been parameterised.
NO is the only option displayed when no further outputs are available.
The automatic parameterization of the display option contains the following basic
settings/factory settings:
YES: Main line = Mass flow; Additional line = Totalizer 1.
NO: The existing (selected) settings remain.
: see next page
Endress+Hauser
13
Esc
++
Quick Setup
Commission
E+
Quick Setup
HOME-POSITION
Language
Pre-setting
Selection System Units
Mass flow
Volume flow
Density
Temperature
Unit
Mass flow
Unit
Volume flow
Unit
Corr. Vol. flow
Unit
Density
Unit
Temperature
Unit
Totalizer
Unit
Totalizer
Unit
Ref. Density
Quit
Fix
Ref. Density
Configuration another unit ?
Yes
No
Measuring
Mode
Selection output type
Current output 1
Frequency/Pulse output
Current output 2
Quit
Operation mode
Frequency
Pulse
Assign
Current
Assign
Current
Assign
Frequency
Assign
Pulse
Current span
Current span
End value
frequency
Pulse value
Value
0/4 mA
Value
0/4 mA
Value
F Low
Pulse width
Value
20 mA
Value
20 mA
Value
F High
Output signal
Time constant
Time constant
Output signal
Failsafe mode
Failsafe mode
Failsafe mode
Time constant
Failsafe mode
No
No
Automatically
parameterization
of the display
F06-80xxxxx-19-xx-xx-en-000
14
Endress+Hauser
5 Group OPERATION
Group OPERATION
Function description OPERATION
LANGUAGE
Use this function to select the language for all texts, parameters and messages shown on the local display.
Note!
The displayed options depend on the available language group shown in the
LANGUAGE GROUP (8226) function.
Options:
Language group WEST EU / USA:
ENGLISH
DEUTSCH
FRANCAIS
ESPANOL
ITALIANO
NEDERLANDS
PORTUGUESE
Language group EAST EU / SCAND:
ENGLISH
NORSK
SVENSKA
SUOMI
POLISH
RUSSIAN
CZECH
Language group ASIA:
ENGLISH
BAHASA INDONESIA
JAPANESE (syllabary)
Factory setting:
Country-dependent (see Page 58 ff.)
Note!
If you press the OS keys at startup, the language defaults to ENGLISH.
ACCESS CODE
All data of the measuring system are protected against inadvertent change.
Programming is disabled and the settings cannot be changed until a code is
entered in this function.
If you press the OS key in any function the measuring system automatically
goes to this function and the prompt to enter the code appears on the display
(programming disabled).
You can enable programming by entering the personal code
(Factory setting = 80, see the DEFINE PRIVATE CODE function)
User input:
max. 4-digit number: 0...9999
Note!
The programming levels are disabled if you do not press a key within 60
seconds following return to the HOME position.
You can also disable programming in this function by entering any number
(other than the defined private code).
The Endress+Hauser service organization can be of assistance if you lose
your personal code.
Endress+Hauser
15
5 Group OPERATION
Use this function to define a personal code number for enabling programming.
User input:
max. 4-digit number: 0...9999
Factory setting:
80
Note!
If you define the personal code number = 0, the programming is always
enabled.
Programming has to be enabled before this code can be changed.
When programming is disabled the function cant be changed, this
precaution prevents others from changing your personal code without
your knowledge and consent.
STATUS ACCESS
Use this function to check the access status for the function matrix.
Display shows:
ACCESS CUSTOMER (Parameterization enabled)
LOCKED (Parameterization disabled)
ACCESS CODE
COUNTER
Displays how often the customer code, service code or the digit 0 (codefree) has been entered to gain access to the function matrix.
Display:
max. 7-digit number: 0...9999999
Factory setting:
0
16
Endress+Hauser
ASSIGN LINE 1
Use this function to define the display value assigned to the main line (the
upper line of the local display) for display during normal measuring operation.
Options:
OFF
MASS FLOW
MASS FLOW IN %
VOLUME FLOW
VOLUME FLOW IN %
DENSITY
TEMPERATURE
TOTALIZER 1
TOTALIZER 2
CORRECTED VOLUME FLOW
CORRECTED VOLUME FLOW IN %
REFERENCE DENSITY
Factory setting:
MASS FLOW
ASSIGN LINE 2
Use this function to define the display value assigned to the additional line (the
bottom line of the local display) for display during normal measuring operation.
Options:
OFF
MASS FLOW
MASS FLOW IN %
VOLUME FLOW
VOLUME FLOW IN %
DENSITY
TEMPERATURE
TOTALIZER 1
TAG NAME
OPERATION/SYSTEM CONDITION
DISPLAY FLOW DIRECTION
MASS FLOW BARGRAPH IN %
VOLUME FLOW BARGRAPH IN %
TOTALIZER 2
CORRECTED VOLUME FLOW
CORRECTED VOLUME FLOW IN %
CORRECTED VOLUME FLOW BARGRAPH IN %
REFERENCE DENSITY
Factory setting:
TOTALIZER
Endress+Hauser
17
Note!
This function is not available unless one of the following was selected in the
ASSIGN LINE 1 function:
MASS FLOW IN %
VOLUME FLOW IN %
CORRECTED VOLUME FLOW IN %
Use this function to define the flow value to be shown on the display as the
100% value of the variable assigned to line 1.
User input:
5-digit floating-point number
Factory setting:
10 kg/s (if MASS FLOW IN % or MASS FLOW BARGRAPH IN % is selected)
10 l/s (if VOLUME FLOW IN % or VOLUME FLOW BARGRAPH IN % is
selected)
100% VALUE
(Line 2)
Note!
This function is not available unless one of the following was selected in the
ASSIGN LINE 2 function:
MASS FLOW IN %
VOLUME FLOW IN %
MASS FLOW BARGRAPH IN %
VOLUME FLOW BARGRAPH IN %
CORRECTED VOLUME FLOW BARGRAPH IN %
Use this function to define the flow value to be shown on the display as the
100% value of the variable assigned to line 2.
User input:
5-digit floating-point number
Factory setting:
10 kg/s (if MASS FLOW IN % or MASS FLOW BARGRAPH IN % is selected)
10 l/s (if VOLUME FLOW IN % or VOLUME FLOW BARGRAPH IN % is
selected)
FORMAT
Use this function to define the maximum number of places after the decimal
point displayed for the reading in the main line.
Options:
XXXXX. - XXXX.X - XXX.XX - XX.XXX -X.XXXX
Factory setting:
X.XXXX
Note!
Note that this setting only affects the reading as it appears on the display, it
has no influence on the accuracy of the system's calculations.
The places after the decimal point as computed by the measuring device
cannot always be displayed, depending on this setting and the engineering
unit. In these instances an arrow appears on the display between the measured value and the engineering unit (e.g. 1.2 kg/h), indicating that the
measuring system is computing with more decimal places than can be
shown on the display.
18
Endress+Hauser
Use this function to enter a time constant defining how the display reacts to
severely fluctuating flow variables, either very quickly (enter a low time constant) or with damping (enter a high time constant).
User input:
0...100 s
Factory setting:
1s
Note!
Setting the time constant to zero seconds switches off damping.
CONTRAST LCD
Use this function to optimize display contrast to suit local operating conditions.
User input:
10...100%
Factory setting:
50%
BACKLIGHT
Use this function to optimize the backlight to suit local operating conditions.
User input:
0...100%
Note!
Entering the value 0 means that the backlight is switched off. The display
then no longer emits any light, i.e. the display texts can no longer be read in
the dark.
Factory setting:
50%
TEST DISPLAY
Use this function to test the operability of the local display and its pixels.
Options:
OFF
ON
Factory setting:
OFF
Test sequence:
1. Start the test by selecting ON.
2. All pixels of the main line and additional line are darkened for
minimum 0.75 second.
3. Main line and additional line show an 8 in each field for
minimum 0.75 second.
4. Main line and additional line show a 0 in each field for
minimum 0.75 second.
5. Main line and additional line show nothing (blank display) for
minimum 0.75 second.
6. When the test completes the local display returns to its initial state and the
setting changes to OFF.
Endress+Hauser
19
ASSIGN TOTALIZER
Note!
The totalizer is reset to 0 as soon as the selection is changed.
SUM
Use this function to view the total for the totalizer measured variable
aggregated since measuring commenced. The value can be positive or
negative.
Display shows:
max. 7-digit floating-point number, including sign and unit (e.g. 15467.04 kg)
Note!
The totalizer response to faults is defined in the FAILSAFE MODE function
(see Page 22).
OVERFLOW
Use this function to view the overflow for the totalizer aggregated since
measuring commenced.
Total flow quantity is represented by a floating decimal point number
consisting of max. 7 digits. You can use this function to view higher
numerical values (>9,999,999) as overflows. The effective quantity is
thus the total of OVERFLOW plus the value returned by the SUM function.
Example:
Reading for 2 overflows: 2 E7 kg (= 20,000,000 kg)
The value returned by the SUM function = 196,845.7 kg
Effective total quantity = 20,196,845.7 kg
Display shows:
Integer with exponent, including sign and unit, e.g. 2 E7 kg
20
Endress+Hauser
Use this function to define the unit for the totalizer measured variable, as
selected beforehand.
Options (for the MASS FLOW assignment):
Metric g; kg; t
US oz; lb; ton
Factory setting:
Depends on nominal diameter and country,
[value] / [g ...kg or US oz...US ton]
corresponding to the totalizer unit factory setting (see Page 58 ff.)
Options (for the VOLUME FLOW assignment):
Metric cm3; dm 3; m3; ml; l; hl; Ml
US cc; af; ft3; oz f; gal; Mgal; bbl (normal fluids); bbl (beer); bbl (petrochemicals); bbl (filling tanks)
Imperial gal; Mgal; bbl (beer); bbl (petrochemicals)
Options (for the CORRECTED VOLUME FLOW assignment):
Metric Nl; Nm3
US Sm3; Scf
Factory setting:
Depends on nominal diameter and country,
[value] / [dm3 ...m3 or US gal...US Mgal]
corresponding to the totalizer unit factory setting (see Page 58 ff.)
TOTALIZER MODE
Use this function to define how the flow components are to be totalized.
Options:
BALANCE
Positive and negative flow components. The positive and negative flow components are balanced. In other words, net flow in the flow direction is registered.
FORWARD
Only positive flow components are totalized.
REVERSE
Only negative flow components are totalized.
Factory setting:
Totalizer 1 = BALANCE
Totalizer 2 = FORWARD
RESET TOTALIZER
Use this function to reset the sum and the overflow of the totalizer to zero
(= RESET).
Options:
NO
YES
Factory setting:
NO
Note!
If the device has a status input and is appropriately configured, a reset for the
totalizer can also be triggered by a pulse.
Endress+Hauser
21
Use this function to reset the totals (including all overflows) of the totalizers
(1...2) to zero (= RESET).
Options:
NO
YES
Factory setting:
NO
Note!
If the device has a status input and if it is appropriately configured, a reset for
the totalizer (1...2) can also be triggered by a pulse (see the ASSIGN STATUS
INPUT function ).
FAILSAFE MODE
Use this function to define the response of all totalizers (1...2) to a fault.
Options:
STOP
The totalizer is paused until the fault is rectified. The totalizer stops at the last
value prior to the occurrence of the error.
ACTUAL VALUE
The totalizer continues to count is based on the current flow measured value.
The fault is ignored.
HOLD VALUE
The totalizer continues to count the flow is based on the last valid flow value
(before the fault occurred).
Factory setting:
STOP
22
Endress+Hauser
ASSIGN CURRENT
Note!
If you select OFF, the only function shown in this group is these (ASSIGN
CURRENT) function.
CURRENT SPAN
Use this function to define the current span. The selection specifies the operational range and the lower and upper signal on alarm. For the current output 1
the option HART can be defined additionally.
Options:
020 mA
420 mA
420 mA HART (only current output1)
420 mA NAMUR
420 mA HART NAMUR (only current output1)
420 mA US
420 mA HART US (only current output1)
020 mA (25 mA)
420 mA (25 mA)
420 mA (25 mA) HART (only current output1)
Factory setting:
420 mA HART NAMUR (fr current output 1)
420 mA NAMUR (fr current output 2)
Note!
The option HART is only supported by the current output designated as current output 1 in the device software, (terminals 26 and 27).
When switching the hardware from an active (factory setting) to a passive
output signal select a current span of 420 mA (please refer to the Operating Instructions PROline promass 80, BA 057D/06/en)
Endress+Hauser
23
I[mA]
1
0-20 mA
0 - 20.5 mA
22
4-20 mA
4 - 20.5 mA
22
4-20 mA HART
4 - 20.5 mA
22
4-20 mA NAMUR
3.8 - 20.5 mA
3.5
22.6
3.8 - 20.5 mA
3.5
22.6
4-20 mA US
3.9 - 20.8 mA
3.75
22.6
4-20 mA HART US
3.9 - 20.8 mA
3.75
22.6
0 - 24 mA
25
4 - 24 mA
25
4 - 24 mA
25
F06-x3xxxxxx-05-xx-xx-xx-017
A = Current span
1 = Operational range (measuring information)
2 = Lower signal on alarm level
3 = Upper signal on alarm level
Note!
If the measured value exceeds the measuring range (as defined in the functions VALUE 0_4 mA and VALUE 20 mA a notice message is generated
(#351...352, current span).
In case of a fault the behaviour of the current output is according to the
selected option in the function FAILSAFE MODE. Change the error category
in the function ASSIGN SYSTEM ERROR to generate a fault message
instead of a notice message.
VALUE 0_4 mA
Note!
This function is not available unless the DENSITY, REFERENCE DENSITY or
TEMPERATURE option was selected in the ASSIGN CURRENT function.
Use this function to assign a value to the 0/4 mA current, (see Setting the
span by means of the 0_4 mA and 20 mA value on Page 25).
Option:
5-digit floating-point number (with sign for the TEMPERATURE measured
variable)
Factory setting:
0.5 [kg/l] or 50 [C]
VALUE 20 mA
Use this function to assign a value to the 20 mA current, (see Setting the span
by means of the 0_4 mA and 20 mA value on Page 25).
Option:
5-digit floating-point number (with sign for the MASS FLOW, VOLUME FLOW,
CORRECTED VOLUME FLOW and TEMPERATURE measured variables)
Factory setting:
Depends on nominal diameter [kg/h] or 2 [kg/l] or 200 [C]
24
Endress+Hauser
The span for the measured variable selected in the ASSIGN CURRENT
function is specified via the VALUE 0_4 mA and VALUE 20 mA functions.
The span is defined differently, depending on the measured variable selected:
MASS FLOW, VOLUME FLOW and CORRECTED VOLUME FLOW
The VALUE 0_4 mA function is not available; the value for the zero flow (0
kg/h or 0m3/h) is assigned to the 0/4 mA current.
The flow value for the 20 mA current is defined in the VALUE 20 mA function,
(input range 99999 to +99999). The appropriate unit is taken from the UNIT
MASS FLOW, UNIT VOLUME FLOW or UNIT CORR. VOL. FLOW function.
Example (for standard measuring mode):
I [mA]
6
F06-80xxxxx-05-xx-xx-xx-008
5
4
3
2
1
Q
A
A = Measuring range (the minimum measuring range has to exceed the value that correlates with a flow velocity of 0.3 m/s)
Endress+Hauser
25
3
2
1
Q
A
F06-80xxxxx-05-xx-xx-xx-009
A = Measuring range (the minimum measuring range has to exceed the value that correlates with a flow velocity of 0.3 m/s)
TEMPERATURE
The temperature value for the 0/4 mA current is defined in the
VALUE 0_4 mA function, (input range 99999 to +99999).
The appropriate unit is taken from the UNIT TEMPERATURE function.
The temperature value for the 20 mA current is defined in the
VALUE 20 mA function, (input range 99999 to +99999).
The appropriate unit is taken from the UNIT TEMPERATURE function.
Note!
Values with different signs cannot be entered for the 0_4 mA and 20 mA values if SYMMETRY is the option selected in the MEASURING MODE function,
(see Page 49), The message INPUT RANGE EXCEEDED appears on the display.
Example (for standard measuring mode)
I [mA]
6
5
3
2
1
Q
A
F06-80xxxxx-05-xx-xx-xx-010
A = Measuring range (the minimum measuring range has to exceed the value that correlates with a flow velocity of 0.3 m/s)
26
Endress+Hauser
Use this function to enter a time constant defining how the current output
signal reacts to severely fluctuating measured variables, either very quickly
(enter a low time constant) or with damping (enter a high time constant).
User input:
Fixed-point number 0.01...100.00 s
Factory setting:
1.00 s
FAILSAFE MODE
For safety reasons it is advisable to ensure that the current output assumes a
predefined state in the event of a fault. The setting you select here affects only
the current output. It has no effect on other outputs and the display (e.g. totalizers).
Options:
MIN. CURRENT
The current output adopts the value of the lower signal on alarm level (as
defined in the function CURRENT SPAN.
MAX. CURRENT
The current output adopts the value of the upper signal on alarm level (as
defined in the function CURRENT SPAN.
HOLD VALUE (not recommended)
Measuring value output is based on the last measuring value saved before the
error occurred .
ACTUAL VALUE
Measured value output is based on the current flow measurement.
The fault is ignored .
Factory setting:
MIN. CURRENT
ACTUAL CURRENT
Use this function to view the computed value of the output current.
Display shows:
0.00...25.00 mA
SIMULATION CURRENT
Note!
The SIMULATION CURRENT OUTPUT message indicates that simulation
is active.
The measuring device continues to measure while simulation is in progress,
i.e. the actual measured values are output correctly via the other outputs.
"
Caution!
The setting is not saved if the power supply fails.
Endress+Hauser
27
Note!
This function is not available unless the SIMULATION CURRENT function is
active (= ON).
Use this function to define a selectable value (e.g. 12 mA) to be output at the
current output. This value is used to test downstream devices and the flowmeter itself.
User input:
floating-point number: 0.00...25.00 mA
Factory setting:
0.00 mA
"
Caution!
The setting is not saved if the power supply fails.
28
Endress+Hauser
10
OPERATION MODE
ASSIGN FREQUENCY
Note!
This function is not available unless the FREQUENCY setting was selected in
the OPERATION MODE function.
Use this function to assign a measured variable to the frequency output.
Options:
OFF
MASS FLOW
VOLUME FLOW
CORRECTED VOLUME FLOW
DENSITY
REFERENCE DENSITY
TEMPERATURE
Factory setting:
MASS FLOW
Note!
If you select OFF, the only functions shown in this function group are ASSIGN
FREQUENCY and OPERATION MODE.
END VALUE
FREQUENCY
Note!
This function is not available unless the FREQUENCY setting was selected in
the OPERATION MODE function.
Use this function to define a full scale frequency for the frequency output.
You define the associated measured value of the measuring range in the
VALUE F HIGH function described on Page 30.
User input:
4-digit fixed-point number: 2...1000 Hz
Factory setting:
1000 Hz
Example:
VALUE F HIGH = 1000 kg/h, full scale frequency = 1000 Hz: i.e. a frequency
of 1000 Hz is output at a flow of 1000 kg/h.
VALUE F HIGH = 3600 kg/h, full scale frequency = 1000 Hz: i.e. a frequency
of 1000 Hz is output at a flow of 3600 kg/h.
Note!
In the FREQUENCY operating mode the output signal is symmetrical (on/off
ratio = 1:1). At low frequencies the pulse duration is limited to a maximum of
2 seconds, i.e. the on/off ratio is no longer symmetrical.
Endress+Hauser
29
Note!
This function is not available unless the DENSITY, REFERENCE DENSITY or
TEMPERATURE option was selected in the ASSIGN FREQUENCY function.
Use this function to assign a value to the start value frequency (0 Hz),
(see Setting the span by means of the f-min. and f-max. value on Page 30).
Option:
5-digit floating-point number (with sign for the TEMPERATURE measured
variable)
Factory setting:
0.5 [kg/l] or 50 [C]
VALUE F HIGH
The span for the measured variable selected in the ASSIGN FREQUENCY
function is specified via the VALUE F LOW and VALUE F HIGH functions.
The span is defined differently, depending on the measured variable selected:
MASS FLOW, VOLUME FLOW and CORRECTED VOLUME FLOW
The VALUE F LOW function is not available; the value for the zero flow
(0 kg/h or 0m3/h) is assigned to the start value frequency.
The flow value for the end value frequency is defined in the VALUE F HIGH
function, (input range 99999 to +99999). The appropriate unit is taken from
the UNIT MASS FLOW, UNIT VOLUME FLOW or UNIT CORRECTED VOLUME FLOW function.
Example (for standard measuring mode):
125
100
Value f min
Value f max
0
Measuring range
F06-80xxxxxx-05-xx-xx-en-011
Frequency [%]
30
Endress+Hauser
125
Value f min
Value f max
100
F06-80xxxxxx-05-xx-xx-en-012
Frequency [%]
Measuring range
= Density value and reference density at which the, in the function END
VALUE FREQUENCY defined, frequency should be output (entry in VALUE F
HIGH function).
TEMPERATURE
The temperature value for the 0/4 mA current is defined in the
VALUE F LOW function, (input range 99999 to +99999).
The appropriate unit is taken from the UNIT TEMPERATURE function.
The temperature value for the 20 mA current is defined in the
VALUE F HIGH function, (input range 99999 to +99999).
The appropriate unit is taken from the UNIT TEMPERATURE function.
Note!
Values with different signs cannot be entered for the VALUE F LOW and
VALUE F HIGH if SYMMETRY is the option selected in the MEASURING MODE
function (see Page 49). The message INPUT RANGE EXCEEDED appears
on the display.
Example (for standard measuring mode):
125
100
Value f min
Value f max
Measuring range
F06-80xxxxxx-05-xx-xx-en-013
Frequency [%]
Endress+Hauser
31
Note!
This function is not available unless the FREQUENCY setting was selected in
the OPERATION MODE function.
Use this function to select the polarity of the frequency signal.
Options:
PASSIVE - POSITIVE
PASSIVE - NEGATIVE
Factory setting:
PASSIVE - POSITIVE
Umax = 30 V DC
External power supply
Open
Collector
1 5 7 8 3
=
-
F-xxxxxxx-04-xx-xx-en-000
PASSIVE:
PASSIVE-POSITIVE
PASSIVE-NEGATIVE
transistor
TIME CONSTANT
F-xxxxxxx-05-xx-xx-en-002
Note!
For continuous currents up to 25 mA (I max = 250 mA / 20 ms)
transistor
conducting
conducting
non
conducting
non
conducting
Note!
This function is not available unless the FREQUENCY setting was selected in
the OPERATION MODE function.
Use this function to enter a time constant defining how the frequency output
signal reacts to severely fluctuating measured variables, either very quickly
(enter a low time constant) or with damping (enter a high time constant).
User input:
floating-point number: 0.00...100.00 s
Factory setting:
0.00 s
32
Endress+Hauser
Note!
This function is not available unless the FREQUENCY setting was selected in
the OPERATION MODE function.
For reasons of safety it is advisable to ensure that the frequency output
assumes a predefined state in the event of a fault. Use this function to define
this state. The setting you select here affects only the frequency output. It has
no effect on other outputs and the display (e.g. totalizer).
Options:
FALLBACK VALUE
Output is 0 Hz.
FAILSAFE LEVEL
Output is the frequency specified in the FAILSAFE VALUE function.
HOLD VALUE
Measured value output is based on the last measured value saved before the
fault occurred.
ACTUAL VALUE
Measured value output is based on the current flow measurement. The fault is
ignored.
Factory setting:
FALLBACK VALUE
FAILSAFE VALUE
Note!
This function is not available unless FREQUENCY was selected in the
OPERATION MODE function and FAILSAFE LEVEL was selected in the
FAILSAFE MODE function.
Use this function to define frequency that the measuring device outputs in the
event of an fault.
User input:
max. 4-digit number: 0...1250 Hz
Factory setting:
1250 Hz
ACTUAL FREQUENCY
Note!
This function is not available unless the FREQUENCY setting was selected in
the OPERATION MODE function.
Use this function to view the computed value of the output frequency.
Display shows:
0...1250 Hz
Endress+Hauser
33
Note!
This function is not available unless the FREQUENCY setting was selected in
the OPERATION MODE function.
Use this function to activate simulation of the frequency output.
Options:
OFF
ON
Factory setting:
OFF
Note!
The SIMULATION FREQUENCY OUTPUT message indicates that
simulation is active.
The measuring device continues to measure while simulation is in progress,
i.e. the currently measured values are output correctly via the other outputs.
"
Caution!
The setting is not saved if the power supply fails.
VALUE SIMULATION
FREQUENCY
Note!
This function is not available unless FREQUENCY was selected in the
OPERATION MODE function and the VALUE SIMULATION FREQUENCY
function is active (= ON).
Use this function to define a selectable frequency value (e.g. 500 Hz) to be
output at the frequency output. This value is used to test downstream devices
and the flowmeter itself.
User input:
0...1250 Hz
Factory setting:
0 Hz
"
Caution!
The setting is not saved if the power supply fails.
ASSIGN PULSE
Note!
This function is not available unless the PULSE setting was selected in the
OPERATION MODE function.
Use this function to assign a measured variable to the pulse output.
Options:
OFF
MASS FLOW
VOLUME FLOW
CORRECTED VOLUME FLOW
Factory setting:
MASS FLOW
Note!
If you select OFF, the only functions shown in this function group are ASSIGN
PULSE and OPERATION MODE.
34
Endress+Hauser
Note!
This function is not available unless the PULSE setting was selected in the
OPERATION MODE function.
Use this function to define the flow at which a pulse is triggered.
These pulses can be totaled by an external totalizer, and the total flow quantity
since measuring started can be registered in this way.
User input:
5-digit floating-point number
Factory setting:
Depends on nominal diameter and country,
[value] [dm3 ...m3 or US gal...US Mgal] / pulses
corresponding to the factory setting for the pulses value (see Page 58 ff.)
Note!
The appropriate unit is taken from the corresponding UNIT function in the
group SYSTEM UNITS (see Page 8).
PULSE WIDTH
Note!
This function is not available unless the PULSE setting was selected in the
OPERATION MODE function.
Use this function to enter the maximum pulse width of the output pulses.
User input:
0.5...2000 ms
Factory setting:
100 ms
Pulse output is always with the pulse width (B) entered in this function.
The intervals (P) between the individual pulses are automatically configured.
However, they must at least correspond to the pulse width (B = P).
B
B=P
B< P
P
F06-xxxxxxxx-05-xx-xx-xx-012
Note!
When entering the pulse width, select a value that can still be processed by an
external totalizer (e.g. mechanical totalizer, PLC, etc.).
"
Caution!
If the pulse number or frequency resulting from the pulse value entered, (see
function PULSE VALUE on Page 35), and from the current flowis too large to
maintain the pulse width selected (interval P is smaller than the pulse width B
entered), a system error message (pulse memory) is generated after buffering/
balancing time.
Endress+Hauser
35
Note!
This function is not available unless the PULSE setting was selected in the
OPERATION MODE function.
Use this function to configure the output in such a way that it for example
matches an external totalizer. According to the application, the direction of the
pulses can be selected here.
Options:
PASSIVE - POSITIVE
PASSIVE - NEGATIVE
Factory setting:
PASSIVE - POSITIVE
F-xxxxxxx-04-xx-xx-en-000
PASSIVE:
Umax = 30 V DC
External power supply
Open
Collector
1 5 7 8 3
=
-
PASSIVE-NEGATIVE
pulses
transistor
F-xxxxxxx-05-xx-xx-en-001
Note!
For continuous currents up to 25 mA (I max = 250 mA / 20 ms)
PASSIVE-POSITIVE
pulses
transistor
conducting
conducting
non
conducting
non
conducting
B = Pulse width
FAILSAFE MODE
Note!
This function is not available unless the PULSE setting was selected in the
OPERATION MODE function.
For reasons of safety it is advisable to ensure that the pulse output assumes a
predefined state in the event of a fault. Use this function to define this state.
The setting you select here affects only the pulse output. It has no effect on
other outputs and the display (e.g. totalizer).
Options:
FALLBACK VALUE
Output is 0 pulse.
HOLD VALUE
Measured value output is based on the last measured value saved before the
fault occurred.
ACTUAL VALUE
Measured value output is based on the current flow measurement. The fault is
ignored.
Factory setting:
FALLBACK VALUE
36
Endress+Hauser
Note!
This function is not available unless the PULSE option was selected in the
OPERATING MODE function.
Use this function to activate simulation of the pulse output.
Options:
OFF
COUNTDOWN
The pulses specified in the VALUE SIMULATION PULSE function are output.
CONTINUOUSLY
Pulses are continuously output with the pulse width specified in the PULSE
WIDTH function. Simulation is started once the CONTINUOUSLY option is
confirmed with the F key.
Note!
Simulation is started by confirming the CONTINUOUSLY option with the F key.
The simulation can be switched off again via the SIMULATION PULSE function.
Factory setting:
OFF
Note!
The notice message #631 SIM. PULSE indicates that simulation is active.
The on/off ratio is 1:1 for both types of simulation.
The measuring device continues to measure while simulation is in progress,
i.e. the current measured values are output correctly via the other outputs.
"
Caution!
The setting is not saved if the power supply fails.
VALUE SIMULATION
PULSE
Note!
This function is not available unless the COUNTDOWN option was selected in
the SIMULATION PULSE function.
Use this function to specify the number of pulses (e.g. 50) which are output
during the simulation. This value is used to test downstream devices and the
measuring device itself. The pulses are output with the pulse width specified in
the PULSE WIDTH function. The on/off ratio is 1:1.
Simulation is started once the specified value is confirmed with the F key.
The display remains at 0 if the specified pulses have been output.
User input:
0...10 000
Factory setting:
0
Note!
Simulation is started by confirming the simulation value with the F key.
The simulation can be switched off again via the SIMULATION PULSE function.
"
Caution!
The setting is not saved if the power supply fails.
Endress+Hauser
37
11
This group is not available unless the measuring device is fitted with a status output.
ASSIGN STATUS
Note!
The behaviour of the status output is of the quiescent-current type, in other
words the output is closed (transistor conductive) when normal, error-free
measuring is in progress.
Please read and comply with the information on the switching characteristics of the status output (see Page 40, 41).
If you select OFF, the only function shown in this function group is this function, in other words ASSIGN STATUS.
ON-VALUE
Note!
This function is not available unless LIMIT MASS FLOW, LIMIT VOLUME
FLOW, LIMIT CORRECTED VOLUME FLOW, LIMIT TEMPERATURE, LIMIT
DENSITY, LIMIT REFERENCE DENSITY or FLOW DIRECTION was selected in
the ASSIGN STATUS function.
Use this function to assign a value to the switch-on point (status output
conductive). The value can be greater or less than the switch-off point.
Positive and negative values are permissible.
User input:
5-digit floating-point number
Factory setting:
0 [kg/h] or 0 [m3/h] or 2 [kg/l] or 200 [C]
OFF-VALUE
Note!
This function is not available unless LIMIT MASS FLOW, LIMIT VOLUME
FLOW, LIMIT CORRECTED VOLUME FLOW, LIMIT TEMPERATURE, LIMIT
DENSITY, LIMIT REFERENCE DENSITY was selected in the ASSIGN STATUS
function.
Use this function to assign a value to the switch-off point (status output not
conductive). The value can be greater or less than the switch-on point.
Positive and negative values are permissible.
User input:
5-digit floating-point number
Factory setting:
0 [kg/h] or 0 [m3/h] or 2 [kg/l] or 200 [C]
38
Endress+Hauser
Note!
This function is not available unless one of the following was selected in the
ASSIGN STATUS function:
FLOW DIRECTION
LIMIT MASS FLOW
LIMIT VOLUME FLOW
LIMIT CORRECTED VOLUME FLOW
LIMIT DENSITY
LIMIT REFERENCE DENSITY
LIMIT TEMPERATURE
Use this function to enter a time constant defining how the status output reacts
to severely fluctuating measured variables, either very quickly (enter a low time
constant) or with damping (enter a high time constant).
The purpose of damping, therefore, is to prevent the status output changing
state continuously in response to fluctuations in flow.
User input:
5-digit floating-point number: 0.00...100.00 s
Factory setting:
0.00 s
ACTUAL STATUS
Use this function to check the current status of the status output.
Display shows:
NOT CONDUCTIVE
CONDUCTIVE
SIMULATION
SWITCH POINT
Note!
The SIMULATION STATUS OUTPUT message indicates that simulation is
active.
The measuring device continues to measure while simulation is in progress,
i.e. the current measured values are output correctly via the other outputs.
"
Caution!
The setting is not saved if the power supply fails.
VALUE SIMULATION
SWITCH POINT
Note!
This function is not available unless the ON setting was selected in the SIMULATION SWITCH POINT function.
Use this function to define the switching response of the status output during
the simulation. This value is used to test downstream devices and the flowmeter itself.
User input:
NOT CONDUCTIVE
CONDUCTIVE
Factory setting:
NOT CONDUCTIVE
"
Caution!
The setting is not saved if the power supply fails.
Endress+Hauser
39
F-xxxxxxx-05-xx-xx-xx-004
F-xxxxxxx-05-xx-xx-xx-005
40
Endress+Hauser
State
ON (operation)
Fault message
not
conductive
System OK
conductive
not
conductive
System OK
conductive
System OK
conductive
not
conductive
not
conductive
conductive
not
conductive
forward
conductive
reverse
not
conductive
Limit value
mass flow
volume flow
corrected volume flow
density
corrected density
temperature
Endress+Hauser
conductive
not
conductive
41
12
This group is not available unless the measuring device is fitted with a status input I/O module.
Note!
Positive zero return is active as long as the active level is available at the status
input (continuous signal). All other assignments react to a change in level
(pulse) at the status input.
ACTIVE LEVEL
Use this function to define whether the assigned function (see ASSIGN
STATUS INPUT function) is released when the signal level is present (HIGH)
or not present (LOW).
Options:
HIGH
LOW
Factory setting:
HIGH
MINIMUM
PULSE WIDTH
Use this function to define a minimum width which the input pulse must
achieve in order to trigger the defined switching function.
User input:
20...100 ms
Factory setting:
50 ms
SIMULATION STATUS
INPUT
Use this function to activate simulation of the status input, in other words to
trigger the function assigned to the status input (see the ASSIGN STATUS
INPUT function on Page 38).
Options:
OFF
ON
Factory setting:
OFF
Note!
The SIMULATION STATUS INPUT message indicates that simulation is
active.
The measuring device continues to measure while simulation is in progress,
i.e. the current measured values are output correctly via the outputs.
"
Caution!
The setting is not saved if the power supply fails.
42
Endress+Hauser
Note!
This function is not available unless the ON setting was selected in the
SIMULATION STATUS INPUT function.
Use this function to select the level to be assumed at the status input during
the simulation.
Options:
HIGH
LOW
Factory setting:
LOW
"
Caution!
The setting is not saved if the power supply fails.
Endress+Hauser
43
13 Group COMMUNICATION
13
Group COMMUNICATION
Function description COMMUNICATION
TAG NAME
Use this function to enter a tag name for the measuring device. You can edit
and read this tag name at the local display or via the HART protocol.
User input:
max. 8-character text, permissible: A-Z, 0-9, +,, punctuation marks
Factory setting:
_ _ _ _ _ _ _ _ (without text)
TAG DESCRIPTION
Use this function to enter a tag description for the measuring device. You can
edit and read this tag description at the local display or via the HART protocol.
User input:
max. 16-character text, permissible: A-Z, 0-9, +,, punctuation marks
Factory setting:
_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ (without text)
BUS ADDRESS
Use this function to define the address for the exchange of data with the HART
protocol.
User input:
0...15
Factory setting:
0
Note!
Addresses 1...15: a constant 4 mA current is applied.
HART PROTOCOL
Note!
The HART protocol can be activated with the selection 4-20 mA HART resp.
4-20 mA (25 mA) HART in the CURRENT SPAN function (see Page 23).
MANUFACTURER ID
DEVICE ID
44
Endress+Hauser
14
Use this function to assign the switching point for the low flow cut off.
Options:
OFF
MASS FLOW
VOLUME FLOW
CORRECTED VOLUME FLOW
Factory setting:
MASS FLOW
Use this function to assign the on value for the low flow cut off.
Low flow cut off is active if the setting is a value not equal to 0. The sign of the
flow value is highlighted on the display to indicate that low flow cut off is active.
User input:
5-digit floating-point number
Factory setting:
depends on nominal diameter
Note!
The appropriate unit is taken from the corresponding UNIT function in the
group SYSTEM UNITS (see Page 8).
Use this function to enter the switch-off point for low flow cut off.
Enter the switch-off point as a positive hysteresis value from the switch-on
point.
User input:
Integer 0...100%
Factory setting:
50%
Example:
2
F06-80xxxxxx-05-xx-xx-xx-007
H
1
Q = Flow [volume/time]
t = Time
H = Hysteresis
a = ON VALUE LOW FLOW CUT OFF = 200 g/h
b = OFF VALUE LOW FLOW CUT OFF = 10%
c = Low flow cut off active
1 = Low flow cut off is switched on at 200 g/h
2 = Low flow cut off is switched off at 220 g/h
Endress+Hauser
45
Use this function to activate the empty pipe detection (EPD). With empty
measuring tubes the density of the fluid falls below a specified value (see EPD
VALUE LOW function).
Options:
OFF
ON
Factory setting:
OFF
"
Caution!
Select a correspondingly low EPD VALUE LOW so that the difference to the
effective density of the fluid is sufficiently large enough.
This ensures that totally empty measuring tubes and not partially filled ones
are detected.
For gas measurement we strongly recommend to switch off empty pipe
detection.
Note!
This function is not available unless the ON selection was selected in the
EMPTY PIPE DETECTION (EPD) function.
Use this function to set an lower threshold for the measured density value, in
order to detect possible problems in the process indicated by too low density.
User input:
5-digit floating-point number
Factory setting:
0.2000 g/cc
Note!
This function is not available unless the ON selection was selected in the
EMPTY PIPE DETECTION (EPD) function.
Use this function to set an upper threshold for the measured density value.
User input:
5-digit floating-point number
Factory setting:
6.0000 g/cc
Use this function to enter the time span for which the criteria for an empty pipe
have to be satisfied without interruption before a notice message or fault message is generated.
User input:
fixed-point number: 1.0...60.0 s
Factory setting:
1.0 s
46
Endress+Hauser
In this function, a fixed value for the reference density can be entered, with
which the corrected volume flow or corrected volume is calculated.
User input:
5-digit floating-point number
Factory setting:
1 kg/Nl
"
Caution!
Before carrying out the calibration, please refer to BA 057D/06/en Promass 80
Operating Instructions where a detailed description of the zero point adjustment is given.
Note!
Programming is locked during zero point adjustment and the display shows:
ZERO ADJUST RUNNING.
If the zero point adjustment is not possible, e.g. with a flow velocity
> 0.1 m/s, or has been cancelled, then the alarm message ZERO ADJUST
NOT POSSIBLE is shown on the display.
If the Promass 80 measuring electronics are fitted with a status input, then
the zero point can also be activated by using this input.
After Zero point adjustment is completed, the new zero point can be called
up with theF key. If the F key is pressed again, you return to the ZERO
POINT ADJUST function.
In this function, enter the density set value of the particular fluid for which you
want to carry out a field density adjustment.
User input:
5 digit floating-point number, incl. units (corresponding to 0.1...5.9999 kg/l)
Note!
The preset density entered here should not vary from the actual fluid density
by a more than 10%.
The appropriate unit is taken from the corresponding UNIT function in the
group SYSTEM UNITS (see Page 8).
MEASURE FLUID
In this function the actual density of the fluid is measured for the density
adjustment.
Options:
CANCEL
START
Endress+Hauser
47
With this function a density adjustment can be carried out on site. The density
set value will thus be recalculated and stored. This ensures that the values
dependent on density calculations (e.g. volume flow) are as accurate as possible.
"
Caution!
Before carrying out a density adjustment, please refer to BA 057D/06/en
Promass 80 Operating Instructions where a detailed description of the
density adjustment is given.
Note!
The density adjustment can be executed if:
The sensor does not accurately measure the density which the operator
expects based on laboratory trials.
The characteristics of the fluid are outside the measuring points set at the
factory or reference conditions under which the flowmeter has been
calibrated.
The plant is used solely for measuring a fluid whose density is to be
determined very accurately under constant conditions.
User input:
CANCEL
DENSITY ADJUST
Factory setting:
CANCEL
RESTORE ORIGINAL
With this function the original density coefficient determined at the factory are
restored.
Options:
NO
YES
Factory setting:
NO
PRESSURE MODE
Use this function to configure an automatic pressure correction. In this way, the
effect of a pressure deviation between the calibration and process pressures
on the measured error for mass flow is compensated for, (see also Operating
Instructions PROline promass 80, BA 057D/06/en, Accuracy Chapter).
Options:
OFF
FIX (A fixed process pressure for pressure correction is specified)
Factory setting:
OFF
PRESSURE
Note!
This function is not available unless FIX was selected in the PRESSURE MODE
function.
Use this function to enter the value for the process pressure which should be
used during pressure correction.
User input:
7-digit floating-point number
Factory setting:
0 bar g
Note!
The appropriate unit is taken from the function group SYSTEM UNITS
(see Page 8).
48
Endress+Hauser
15
INSTALLATION
DIRECTION SENSOR
Use this function to reverse the sign of the measured variable, if necessary.
Options:
NORMAL (flow as indicated by the arrow)
INVERSE (flow opposite to direction indicated by the arrow)
Factory setting:
NORMAL
Note!
Ascertain the actual direction of fluid flow with reference to the direction
indicated by the arrow on the sensor (nameplate).
MEASURING MODE
Use this function to define the measuring mode for all outputs.
Options:
STANDARD
SYMMETRY
Factory setting:
STANDARD
The responses of the individual outputs in each of the measuring modes are
described in detail below:
Current and frequency output
STANDARD
The output signals of the current and frequency output are proportional to the
measured variable.
The flow components outside the scaled measuring range (between VALUE
0_4 mA or VALUE F LOW and the VALUE 20 mA or VALUE F HIGH ) are
not taken into account for signal output, but a message CURRENT RANGE AT
FULL SCALE VALUE or FREQUENCY RANGE AT FULL SCALE VALUE is
issued.
F-xxxxxxx-05-xx-xx-xx-003
mA
20
4
0
SYMMETRY
The output signals of the current and frequency output are independent of the
direction of flow (absolute amount of the measured variable). The VALUE 20
mA or VALUE F HIGH (e.g. backflow) corresponds to the mirrored VALUE
20 mA or VALUE F HIGH (e.g. flow).
mA
20
Endress+Hauser
4
0
F-xxxxxxx-05-xx-xx-xx-007
49
Pulse output
STANDARD
Only positive flow components are totaled. Negative components are not taken
into account.
SYMMETRY
Positive and negative flow components are taken into account.
Note!
The direction of flow can be output via the configurable status output.
Status output
Note!
Only if in the ASSIGN STATUS function the LIMIT option is selected.
STANDARD
The status output signal switches at the defined switching points.
SYMMETRY
The status output signal switches at the defined switching points, irrespective
of the sign. In other words, if you define a switching point with a positive sign
the status output signal switches as soon as the value is reached in the negative direction (negative sign) (see illustration).
Example for the SYMMETRY measuring mode:
Switch-on point: Q = 4
Switch-off point: Q = 10
= Status output switched on (conductive)
= Status output switched off (non-conductive)
Q
10
-4
- 10
POSITIVE ZERO
RETURN
F-xxxxxxx-05-xx-xx-xx-005
Use this function to interrupt evaluation of measured variables. This is necessary when a piping system is being cleaned, for example.
This setting acts on all functions and outputs of the measuring device.
Options:
OFF
ON (signal output is set to zero flow value, temperature and density are output
normally)
Factory setting:
OFF
50
Endress+Hauser
The density filter allows the sensitivity of the density measuring signal to be
lowered with respect to variations in the density of the fluid, e.g. with inhomogeneous liquids.
User input:
max. 5-digit number, including unit: 0.00...100.00 s
Factory setting:
0.00 s
Note!
The damping acts on all functions and outputs of the measuring device.
FLOW DAMPING
Using the interference blanking (= time constant for exponential filter) the
sensitivity of the flow measurement signal can be reduced with respect to
transient flows and interference peaks; e.g. with fluid containing solids or
gas bubbles, etc.
User input:
0...100 s
Factory setting:
0s
Note!
The damping acts on all functions and outputs of the measuring device.
Endress+Hauser
51
16
All sensor data, including calibration factor, zero point, nominal diameter, etc. are set at the factory. All the
sensor's parameter settings are saved on the S-DAT memory chip.
"
Caution!
Under normal circumstances you should not change these parameter settings, because changes affect
numerous functions of the entire measuring facility in general, and the accuracy of the measuring system
in particular. Consequently, most of the functions described below can be accessed only by entering a
special service code, which is not the same as your private code number.
Contact the E+H service organization if you have any questions about these functions.
K-FACTOR
This function shows the current calibration factor for the sensor.
Factory setting:
depends on nominal diameter and calibration
Note!
If the service code is used to call this function, this value can be edited.
ZERO POINT
This function shows the current zero-point correction value for the sensor. The
zero-point correction factor is calculated and set at the factory.
User input:
max. 5-digit number: 99999...+99999
Factory setting:
depends on calibration
NOMINAL DIAMETER
This function shows the nominal diameter for the sensor. The nominal diameter
depends on the size of the sensor and is set at the factory.
Factory setting:
depends on the size of the sensor
Note!
If the service code is used to call this function, this value can be edited.
52
TEMPERATURE
COEFFICIENT KM
TEMPERATURE
COEFFICIENT KM 2
TEMPERATURE
COEFFICIENT KT
CALIBRATION
COEFFICIENT KD 1
CALIBRATION
COEFFICIENT KD 2
Endress+Hauser
"
Caution!
A density adjustment can alter the calibration value of this coefficient.
DENSITY
COEFFICIENT C 1
"
Caution!
A density adjustment can alter the calibration value of this coefficient.
DENSITY
COEFFICIENT C 2
"
Caution!
A density adjustment can alter the calibration value of this coefficient.
DENSITY
COEFFICIENT C 3
"
Caution!
A density adjustment can alter the calibration value of this coefficient.
DENSITY
COEFFICIENT C 4
"
Caution!
A density adjustment can alter the calibration value of this coefficient.
DENSITY
COEFFICIENT C 5
"
Caution!
A density adjustment can alter the calibration value of this coefficient.
Endress+Hauser
MINIMAL
TEMPERATURE
MEASURED
MAXIMAL
TEMPERATURE
MEASURED
MINIMAL
TEMPERATURE
CARRIER TUBE
MAXIMAL
TEMPERATURE
CARRIER TUBE
53
17 Group SUPERVISION
17
Group SUPERVISION
Function description SUPERVISION
ACTUAL SYSTEM
CONDITION
PREVIOUS SYSTEM
CONDITION
Use this function to view the fifteen most recent fault and notice messages
since measuring last started.
Display shows:
The 15 most recent fault or notice messages.
ASSIGN SYSTEM
ERROR
Use this function to view all system messages and the associated error
categories (fault message or notice message). If you select a single system
fault you can change its error category.
Display shows:
List of system errors
Note!
Press the F key twice to call the ERROR CATEGORY function.
Use the OS key combination or select CANCEL in the system error list to
exit the function.
ERROR CATEGORY
Use this function to define whether a system fault triggers a notice message or
a fault message. If you select FAULT MESSAGES, all outputs respond to a
fault in accordance with their defined error response patterns.
Options:
NOTICE MESSAGES (display only)
FAULT MESSAGES (outputs and display)
Note!
Press the F key twice to call the ASSIGN SYSTEM ERROR function.
Use the OS key combination to exit the function.
ASSIGN PROCESS
ERROR
Use this function to view all process errors and the associated error categories
(fault message or notice message). If you select a single process error you
can change its error category.
Display shows:
List of process errors
Note!
Press the F key twice to call the ERROR CATEGORY function.
Use the OS key combination or select CANCEL in the process error list to
exit the function.
54
Endress+Hauser
17 Group SUPERVISION
Use this function to define whether a process error triggers a notice message
or a fault message. If you select FAULT MESSAGES, all outputs respond to a
fault in accordance with their defined error response patterns.
Options:
NOTICE MESSAGES (display only)
FAULT MESSAGES (outputs and display)
Note!
Press the F key twice to call the ASSIGN PROCESS ERROR function.
Use the OS key combination to exit the function.
ALARM DELAY
Use this function to define a time span for which the criteria for an error have
to be satisfied without interruption before an error or notice message is
generated.
Depending on the setting and the type of fault, this suppression acts on:
Displa
Current output
Frequency output
Status output
User input:
0...100 s (in steps of one second)
Factory setting:
0s
"
Caution!
If this function is activated fault and notice messages are delayed by the time
corresponding to the setting before being forwarded to the higher-order
controller (process controller, etc.). It is therefore imperative to check in
advance in order to make sure whether a delay of this nature could affect the
safety requirements of the process. If fault and notice messages cannot be
suppressed, a value of 0 seconds must be entered here.
SYSTEM RESET
OPERATION HOURS
Endress+Hauser
55
18
SIMULATION FAILSAFE
MODE
Use this function to set all inputs, outputs and totalizer to their defined fault
response modes, in order to check whether they respond correctly. During this
time, the words SIMULATION FAILSAFE MODE appear on the display.
Options:
OFF
ON
Factory setting:
OFF
SIMULATION
MEASURAND
Use this function to set all inputs, outputs and totalizer to their defined flowresponse modes, in order to check whether they respond correctly. During this
time, the words SIMULATION MEASURAND appear on the display.
Options:
OFF
MASS FLOW
VOLUME FLOW
CORRECTED VOLUME FLOW
DENSITY
REFERENCE DENSITY
TEMPERATURE
Factory setting:
OFF
"
Caution!
The measuring device cannot be used for measuring while this simulation is
in progress.
The setting is not saved if the power supply fails.
VALUE SIMULATION
MEASURAND
Note!
This function is not available unless the SIMULATION MEASURAND function is
active.
Use this function to define a selectable value (e.g. 12 kg/s). This value is used
to test downstream devices and the flowmeter itself.
User input:
5-digit floating-point number
Factory setting:
0 kg/h
(MASS FLOW)
0 m3/h
(VOLUME FLOW)
(CORRECTED VOLUME FLOW)
0 Nm 3/h
0 kg/l
(DENSITY)
0 kg/NI
(REFERENCE DENSITY)
0 C
(TEMPERATURE)
"
Caution!
The setting is not saved if the power supply fails.
56
Endress+Hauser
19
SERIAL NUMBER
SENSOR TYPE
Use this function to view the sensor type (e.g. Promass F).
SOFTWARE REVISION
NUMBER S-DAT
Use this function to view the software revision number of the S-DAT.
20
SOFTWARE
REVISION NUMBER
AMPLIFIER
Use this function to view the software revision number of the amplifier.
LANGUAGE GROUP
Note!
The language options of the available language group are displayed in the
LANGUAGE function.
You can change the language group via the configuration software
FieldTool. Please do not hesitate to contact your E+H sales office if you have
any questions.
Endress+Hauser
SOFTWARE
REVISION NUMBER
I/O
Use this function to view the software revision number of the I/O module.
57
21 Factory settings
21
Factory settings
Pulse value
[mm]
(approx. v = 2 m/s)
0.08
kg/h
kg/h
0.001
kg/p
0.40
kg/h
20
kg/h
0.010
kg/p
1.80
kg/h
90
kg/h
0.010
kg/p
8.00
kg/h
400
kg/h
0.100
kg/p
15
26.00
kg/h
1300
kg/h
0.100
kg/p
15 FB
72.00
kg/h
3600
kg/h
1.000
kg/p
25
72.00
kg/h
3600
kg/h
1.000
kg/p
25 FB
180.00
kg/h
9000
kg/h
1.000
kg/p
40
180.00
kg/h
9000
kg/h
1.000
kg/p
40 FB
300.00
kg/h
15000
kg/h
10.000
kg/p
50
300.00
kg/h
15000
kg/h
10.000
kg/p
80
720.00
kg/h
36000
kg/h
10.000
kg/p
100
1200.00
kg/h
60000
kg/h
10.000
kg/p
150
2600.00
kg/h
130000
kg/h
100.000
kg/p
Language
Country
58
Language
Australia
English
Austria
Deutsch
Belgium
English
Czech Republic
Czech
Denmark
English
England
English
Finland
Suomi
France
Francais
Germany
Deutsch
Hong Kong
English
Hungary
English
India
English
Indonesia
Bahasa Indonesia
Instruments International
English
Italy
Italiano
Japan
Japanese
Malaysia
English
Netherlands
Nederlands
Norway
Norsk
Poland
Polish
Portugal
Portuguese
Russia
Russian
Singapore
English
South Africa
English
Spain
Espanol
Sweden
Svenska
Switzerland
Deutsch
Thailand
English
Endress+Hauser
21 Factory settings
kg/l
mm
C
Reference density
kg/Nl
Pulse value
[mm]
(approx. v = 2 m/s)
0.003
lb/min
0.15
lb/min
0.002
lb/p
0.015
lb/min
0.75
lb/min
0.020
lb/p
0.066
lb/min
3.30
lb/min
0.020
lb/p
0.300
lb/min
15.00
lb/min
0.200
lb/p
15
1.000
lb/min
50.00
lb/min
0.200
lb/p
15 FB
2.600
lb/min
130.00
lb/min
2.000
lb/p
25
2.600
lb/min
130.00
lb/min
2.000
lb/p
25 FB
6.600
lb/min
330.00
lb/min
2.000
lb/p
40
6.600
lb/min
330.00
lb/min
2.000
lb/p
40 FB
11.000
lb/min
550.00
lb/min
20.000
lb/p
50
11.000
lb/min
550.00
lb/min
20.000
lb/p
80
26.000
lb/min
1300.00
lb/min
20.000
lb/p
100
44.000
lb/min
2200.00
lb/min
20.000
lb/p
150
95.000
lb/min
4800.00
lb/min
200.000
lb/p
Endress+Hauser
Language
Unit
English
Density
Length
Temperature
g/cc
INCH
F
Reference density
g/Scc
59
21 Factory settings
60
Endress+Hauser
15
42
27
33
39
54
48
47
55
23
17
17
29
45
54
34
42
38
54
20
B
Bus address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
C
Calibration coefficient
KD 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
KD 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Carrier tube temperature
Maximal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Minimal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Code
Access code counter . . . . . . . . . . . . . . . . . . . . .
Coefficient
Calibration
KD 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
KD 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Density
C0................................
C1................................
C2................................
C3................................
C4................................
C5................................
Temperature
KM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
KM 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
KT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contrast LDC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Current
Output
Current span . . . . . . . . . . . . . . . . . . . . . . . .
Endress+Hauser
52
52
53
53
16
52
52
53
53
53
53
53
53
52
52
52
19
24
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23, 24
Current Output
Actual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Assign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Failsafe mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Time constant . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Value Simulation . . . . . . . . . . . . . . . . . . . . . . . . . 28
Value 0_4 mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Value 20 mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
D
Damping
Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Define private code . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Coefficient
C 0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
C 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
C 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
C 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
C 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
C 5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Set value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Density adjust . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Density damping . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Device ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Display
Damping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
E
Empty pipe detection
EPD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
High value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Low value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Response time . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
End value frequency . . . . . . . . . . . . . . . . . . . . . . . . . 29
Error category
Process error . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
System error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
F
Factory setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Failsafe mode
Current output . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Pulse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Simulation system . . . . . . . . . . . . . . . . . . . . . . . . 56
Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Failsafe value
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Flow damping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Frequency
61
Actual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Assign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
End value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Failsafe mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Failsafe value . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Time constant . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Value F high . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Value F low . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Value Simulation . . . . . . . . . . . . . . . . . . . . . . . . . 34
Funktion matrix
Graphical illustration . . . . . . . . . . . . . . . . . . . . . . . 6
Layout and use . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
G
Group
Amplifier Version . . . . . . . . . . . . . . . . . . . . . . . . . 57
Communication . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Current output . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Measuring values . . . . . . . . . . . . . . . . . . . . . . . . . 7
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Process parameter . . . . . . . . . . . . . . . . . . . . . . . 45
Pulse/Frequency output . . . . . . . . . . . . . . . . . . . 29
Quick Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Sensor data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Sensor Version . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Simulation system . . . . . . . . . . . . . . . . . . . . . . . . 56
Status input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Status/Relais output . . . . . . . . . . . . . . . . . . . . . . 38
Supervision . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
System parameter . . . . . . . . . . . . . . . . . . . . . . . . 49
System units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
User interface . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
H
HART protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
I
I/O Modul type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Installation direction sensor . . . . . . . . . . . . . . . . . . . 49
K
K-Factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
L
Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
LCD Contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Low flow cut off
Assign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Off-value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
On-value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
47
49
53
53
42
N
Nominal Diameter . . . . . . . . . . . . . . . . . . . . . . . . . . 52
O
Off-value
Low flow cut off . . . . . . . . . . . . . . . . . . . . . . . . .
Status output . . . . . . . . . . . . . . . . . . . . . . . . . . .
On-value
Low flow cut off . . . . . . . . . . . . . . . . . . . . . . . . .
Status output . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operation
Mode Pulse-/Freq.-output . . . . . . . . . . . . . . . . .
Output signal
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pulse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Overflow
Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
45
38
45
38
29
32
36
20
P
Positive zero return . . . . . . . . . . . . . . . . . . . . . . . . .
Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pressure mode . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Previous system condition . . . . . . . . . . . . . . . . . . .
Process Error
Assign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Error category . . . . . . . . . . . . . . . . . . . . . . . . . .
Pulse
Assign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Failsafe value . . . . . . . . . . . . . . . . . . . . . . . . . . .
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . .
Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
50
48
48
54
54
55
34
36
36
35
35
Q
Quick Setup
Commission . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Commission (Diagram) . . . . . . . . . . . . . . . . . . . 14
15
19
Reset
System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Restore original . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
45
45
45
M
Manufacturer ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Mass flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Maximal
Carrier tube temperature . . . . . . . . . . . . . . . . . . 53
Temperature measured . . . . . . . . . . . . . . . . . . . 53
62
Measured fluid . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measuring mode . . . . . . . . . . . . . . . . . . . . . . . . . . .
Minimal
Carrier tube temperature . . . . . . . . . . . . . . . . . .
Temperature measured . . . . . . . . . . . . . . . . . . .
Minimum pulse width . . . . . . . . . . . . . . . . . . . . . . .
Sensor type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Serial number sensor . . . . . . . . . . . . . . . . . . . . . . .
Simulation
Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Failsafe mode . . . . . . . . . . . . . . . . . . . . . . . . . . .
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measurand . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Status input . . . . . . . . . . . . . . . . . . . . . . . . . . . .
57
57
27
56
34
56
42
Endress+Hauser
Switch point . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Software revision number
Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
I/O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
S-DAT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Status
Access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Status input
Active level . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Assign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Minimum pulse width . . . . . . . . . . . . . . . . . . . . .
Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Value Simulation . . . . . . . . . . . . . . . . . . . . . . . . .
Status output
Actual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Assign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Direction of flow . . . . . . . . . . . . . . . . . . . . . . . . .
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Limit value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Off-Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
On-Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Switching action . . . . . . . . . . . . . . . . . . . . . . . . .
Time constant . . . . . . . . . . . . . . . . . . . . . . . . . . .
Sum
Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Switch point
Simulation Status output . . . . . . . . . . . . . . . . . . .
Value Simulation Status output . . . . . . . . . . . . . .
System
Condition
Actual . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Previous . . . . . . . . . . . . . . . . . . . . . . . . . . .
Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
System Error
Assign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Error category . . . . . . . . . . . . . . . . . . . . . . . . . . .
39
57
57
57
16
42
42
42
42
43
39
38
40
40
40
38
38
41
39
Unit
Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Mass . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Mass flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Volume flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
20
39
39
54
54
55
54
54
T
Tag
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Temperature coefficient
KM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
KM 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
KT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Temperature measured
Maximal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Minimal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Test Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Endress+Hauser
Time constant
Current output . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Status output . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Totalizer
Assign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Failsafe mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Overflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Sum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Value
Pulse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Value F high . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Value F low . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Value Simulation
Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Measurand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Status input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Switch point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Value 0_4 mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Value 20 mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Version
Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Volume flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Z
Zero point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Zero point adjust . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Numerics
100% Value
Line 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Line 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
63
Europe
Austria Wien
Endress+Hauser Ges.m.b.H.
Tel. (01) 88 05 60, Fax (01) 88 05 63 35
Belarus Minsk
Belorgsintez
Tel. (017) 2 50 84 73, Fax (017) 2 50 85 83
Belgium / Luxembourg Bruxelles
Endress+Hauser S.A. / N.V.
Tel. (02) 2 48 06 00, Fax (02) 2 48 05 53
Bulgaria Sofia
Intertech-Automation Ltd.
Tel. (02) 9 62 71 52, Fax (02) 9 62 14 71
Croatia Zagreb
Endress+Hauser GmbH+Co.
Tel. (01) 6 63 77 85, Fax (01) 6 63 78 23
Cyprus Nicosia
I+G Electrical Services Co. Ltd.
Tel. (02) 48 47 88, Fax (02) 48 46 90
Czech Republic Praha
Endress+Hauser Czech s.r.o.
Tel. (02) 66 78 42 00, Fax (026) 66 78 41 79
Denmark Sborg
Endress+Hauser A/S
Tel. (70) 13 11 32, Fax (70) 13 21 33
Estonia Tartu
Elvi-Aqua O
Tel. (7) 30 27 32, Fax (7) 30 27 31
Finland Helsinki
Metso Endress+Hauser Oy
Tel. (204) 8 31 60, Fax (204) 8 31 61
France Huningue
Endress+Hauser S.A.
Tel. (389) 69 67 68, Fax (389) 69 48 02
Germany Weil am Rhein
Endress+Hauser Messtechnik
GmbH+Co. KG
Tel. (07621) 9 75 01, Fax (07621) 97 55 55
Norway Lierskogen
Endress+Hauser A/S
Tel. 32 85 98 50, Fax 32 85 98 51
Israel Netanya
Instrumetrics Industrial Control Ltd.
Tel. (09) 8 35 70 90, Fax (09) 8 35 06 19
Poland Wroclaw
Endress+Hauser Polska Sp. z o.o.
Tel. (071) 7 80 37 00, Fax (071) 7 80 37 60
Japan Tokyo
Sakura Endress Co. Ltd.
Tel. (0422) 54 06 11, Fax (0422) 55 02 75
Portugal Cacem
Endress+Hauser Lda.
Tel. (21) 4 26 72 90, Fax (21) 4 26 72 99
Ecuador Quito
Insetec Cia. Ltda.
Tel. (02) 2 26 91 48, Fax (02) 2 46 18 33
Jordan Amman
A.P. Parpas Engineering S.A.
Tel. (06) 5 53 92 83, Fax (06) 5 53 92 05
Romania Bucharest
Romconseng S.R.L.
Tel. (021) 41 12 50 1, Fax (021) 41 01 63 4
Kazakhstan Almaty
BEI Electro
Tel. (72) 30 00 28, Fax (72) 50 71 30
Russia Moscow
Endress+Hauser GmbH+Co
Tel. (095) 78 32 85 0, Fax (095) 78 32 85 5
Slovak Republic Bratislava
Transcom Technik s.r.o.
Tel. (2) 44 88 86 90, Fax (2) 44 88 71 12
Slovenia Ljubljana
Endress+Hauser (Slovenija) D.O.O.
Tel. (01) 5 19 22 17, Fax (01) 5 19 22 98
Spain Sant Just Desvern
Endress+Hauser S.A.
Tel. (93) 4 80 33 66, Fax (93) 4 73 38 39
Sweden Sollentuna
Endress+Hauser AB
Tel. (08) 55 51 16 00, Fax (08) 55 51 16 55
Switzerland Reinach/BL 1
Endress+Hauser Metso AG
Tel. (061) 7 15 75 75, Fax (061) 7 11 16 50
Turkey Levent/Istanbul
Intek Endstriyel lc ve Kontrol Sistemleri
Tel. (0212) 2 75 13 55, Fax (0212) 2 66 27 75
Ukraine Kiev
Photonika GmbH
Tel. (44) 2 68 81 02, Fax (44) 2 69 07 05
Yugoslavia Republic Beograd
Meris d.o.o.
Tel. (11) 4 44 29 66, Fax (11) 3 08 57 78
Africa
Asia
Greece Athens
I & G Building Services Automation S.A.
Tel. (01) 9 24 15 00, Fax (01) 9 22 17 14
Algeria Annaba
Symes Systemes et Mesures
Tel. (38) 88 30 03, Fax (38) 88 30 02
Azerbaijan Baku
Modcon Systems - Baku
Tel. (12) 92 98 59, Fax (12) 99 13 72
Hungary Budapest
Endress+Hauser Magyarorszg
Tel. (01) 4 12 04 21, Fax (01) 4 12 04 24
Egypt Heliopolis/Cairo
Anasia Egypt For Trading (S.A.E.)
Tel. (02) 2 68 41 59, Fax (02) 2 68 41 69
Iceland Reykjavik
Sindra-Stl hf
Tel. 5 75 00 00, Fax 5 75 00 10
Morocco Casablanca
Oussama S.A.
Tel. (02) 22 24 13 38, Fax (02) 2 40 26 57
Tunisia Tunis
CMR Controle, Maintenance et Regulation
Tel. (07) 17 93 07 7, Fax (07) 17 88 59 5
Pakistan Karachi
Speedy Automation
Tel. (021) 7 72 29 53, Fax (021) 7 73 68 84
Philippines Pasig City, Metro Manila
Endress+Hauser (Phillipines) Inc.
Tel. (2) 6 38 18 71, Fax (2) 6 38 80 42
Saudi Arabia Jeddah
Anasia Trading Est.
Tel. (02) 6 53 36 61, Fax (02) 6 53 35 04
Singapore Singapore
Endress+Hauser (S.E.A.) Pte. Ltd.
Tel. (65) 66 82 22, Fax (65) 66 68 48
Sultanate of Oman Ruwi
Mustafa & Sultan Sience & Industry Co. L.L.C.
Tel. 63 60 00, Fax 60 70 66
Taiwan Taipei
Kingjarl Corporation
Tel. (02) 27 18 39 38, Fax (02) 27 13 41 90
Thailand Bangkok 10210
Endress+Hauser (Thailand) Ltd.
Tel. (2) 9 96 78 11-20, Fax (2) 9 96 78 10
United Arab Emirates Dubai
Descon Trading L.L.C.
Tel. (04) 2 65 36 51, Fax (04) 2 65 32 64
Lithuania Kaunas
UAB Agava Ltd.
Tel. (03) 7 20 24 10, Fax (03) 7 20 74 14
Macedonia Beograd
Meris d.o.o.
Tel. (11) 44 42 96 6, Fax (11) 30 85 77 8
India Mumbai
Endress+Hauser (India) Pvt. Ltd.
Tel. (022) 56 93 83 33, Fax (022) 56 93 88 330
Moldavia Chisinau
S.C. Techno Test SRL
Tel. (02) 22 61 60, Fax (02) 22 83 13
Indonesia Jakarta
PT Grama Bazita
Tel. (21) 7 95 50 83, Fax (21) 7 97 50 89
Netherlands Naarden
Endress+Hauser B.V.
Tel. (035) 6 95 86 11, Fax (035) 6 95 88 25
Iran Tehran
Patsa Industry
Tel. (021) 8 72 68 69, Fax (021) 8 71 96 66
BA058D/06/en/10.03
50098512
FM+SGML 6.0
China Shanghai
Endress+Hauser (Shanghai)
Instrumentation Co. Ltd.
Tel. (021) 54 90 23 00, Fax (021) 54 90 23 03
America
05.03
Uzbekistan Tashkent
Im Mexatronika-Tes
Tel. (71) 1 91 77 07, Fax (71) 1 91 76 94
China Beijing
Endress+Hauser (Beijing)
Instrumentation Co. Ltd.
Tel. (010) 65 88 24 68, Fax (010) 65 88 17 25
Kuwait Safat
United Technical Services Est. For General
Trading
Tel. 2 41 12 63, Fax 2 41 15 93
Latvia Riga
Elekoms Ltd.
Tel. (07) 33 64 44, Fax (07) 33 64 48
http://www.endress.com