Sie sind auf Seite 1von 80

MA_Siemens_Sitrans_FX300_R01_en_PRT.

book Page 2 Monday, June 16, 2008 1:56 PM

CONTENTS

Safety............................................................................................. 4
1.1

Intended use......................................................................................................... 4

1.2

Certifications ........................................................................................................ 5

1.3

Safety instructions from the manufacturer......................................................... 5

1.3.1

Notes about the documentation ................................................................................ 5

1.3.2

Display conventions ................................................................................................... 6

1.4

Safety instructions for the operator .................................................................... 6

INSTRUMENT DESCRIPTION ......................................................... 7


2.1

Scope of delivery .................................................................................................. 7

2.2

Instrument versions............................................................................................. 7

2.2.1

Versions with flanged connection ............................................................................. 8

2.2.2

Sandwich versions ..................................................................................................... 8

2.2.3

Instrument description.............................................................................................. 9

2.3

Nameplate.......................................................................................................... 10

2.4

Dual seal ............................................................................................................ 11

3.1

General installation instructions ....................................................................... 12

3.2

Storage ............................................................................................................... 12

3.3

Transport............................................................................................................ 13

3.4

Installation requirements .................................................................................. 14

Installation................................................................................... 12

3.4.1

General .................................................................................................................... 14

3.4.2

Liquid measurement ............................................................................................... 15

3.4.3

Vapour and gas measurement ................................................................................ 17

3.4.4

Piping with control valves........................................................................................ 17

3.4.5

Preferred installation location ................................................................................ 18

3.4.6

Turning the connection housing.............................................................................. 19

3.4.7

Turning the display .................................................................................................. 20

3.4.8

Thermal insulation .................................................................................................. 21

3.5

Inlet and outlet runs .......................................................................................... 22

3.5.1

Minimum inlet runs ................................................................................................. 22

3.5.2

Flow straightener .................................................................................................... 24

3.5.3

Minimum outlet runs ............................................................................................... 24

3.6

Installation ......................................................................................................... 25

3.6.1

General installation instructions............................................................................. 25

3.6.2

Installing instruments in sandwich design ............................................................. 27

3.6.3

Installing instruments in flange design .................................................................. 28

SITRANS FX300

Electrical connections ................................................................. 29


4.1

Safety instructions ............................................................................................. 29

4.2

Connecting the signal converter........................................................................ 30

4.3

Current output, electrical data .......................................................................... 31

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 3 Monday, June 16, 2008 1:56 PM

CONTENTS

SITRANS FX300

4.3.1

Power supply ........................................................................................................... 31

4.3.2

Load for HART communication............................................................................. 32

4.4

Pulse output, electrical data.............................................................................. 33

4.4.1

Connection pulse output.......................................................................................... 33

4.4.2

Configuring the pulse output................................................................................... 33

4.5

Protection category............................................................................................ 35

4.6

Earthing connections ......................................................................................... 36

5.1

Start-up .............................................................................................................. 37

Start-up and control .................................................................... 37


5.1.1

Start ......................................................................................................................... 37

5.1.2

Start-up and control ................................................................................................ 37

Operation ..................................................................................... 38
6.1

Display and operating elements ........................................................................ 38

6.2

Operating principles........................................................................................... 39

6.2.1

Description of the key functions.............................................................................. 39

6.2.2

Switch from measuring mode to menu mode ........................................................ 39

6.2.3

Navigation within the menu structure .................................................................... 40

6.2.4

Changing the settings in the menu ......................................................................... 40

6.2.5

Measures in the event of faulty indications ............................................................ 41

6.3

Overview of the most important functions and indicators ................................ 41

6.4

Error messages ................................................................................................. 43

Service ......................................................................................... 44
7.1

Replacing converter / LCD display .................................................................... 44

7.2

Returning the device to the manufacturer ........................................................ 46

7.2.1

General information ................................................................................................ 46

7.2.2

Form (for copying) to accompany a returned instrument ...................................... 47

Technical data ............................................................................. 48


8.1

Functional principle ........................................................................................... 48

8.2

Technical data .................................................................................................... 49

8.3

Dimensions and weights.................................................................................... 51

8.4

Flow tables ......................................................................................................... 59

9.1

Menu structure .................................................................................................. 62

Appendix ...................................................................................... 62
9.1.1

Overview of firmware versions ................................................................................ 62

9.1.2

Entering values in change mode ............................................................................. 63

9.1.3

Selecting characters in change mode..................................................................... 63

9.1.4

Quick Setup menu.................................................................................................... 64

9.1.5

Tests menu .............................................................................................................. 65

9.1.6

Menu item Setup (Firmware version v1 Basic) ....................................................... 66

9.1.7

Menu item Setup (Firmware version v6 Steam)...................................................... 69

9.1.8

Menu item Setup (Firmware version v7 Gas) .......................................................... 74

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

Safety

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 4 Monday, June 16, 2008 1:56 PM

SAFETY

SITRANS FX300

1.1 Intended use


The vortex flowmeters are made to measure the flow of gases, vapours and liquids.

These flowmeters are particularly suitable for measuring:

Clean liquids with low viscosity (< 10 cP)


Hydrocarbons with low viscosity (< 10 cP)
Water
Chemicals with low corrosiveness
Saturated steam
Superheated steam, including CIP and SIP applications in the food industry
Industrial gases

The meters are rated for the following flow velocities:

From 0.3 to 9.0 m/s (1 to 29.5 ft/s) for liquids


From 3.0 to 80 m/s (9.8 to 262 ft/s) for gases and vapours. 1

1
DN15 - 3,0 to 45 m/s (9.8 to 148 ft/s)
DN25 - 3,0 to 70 m/s (9.8 to 230 ft/s)
If the danger of waterhammers can occur in steam net works appropriate condendsate
separators have to be installed.
If the danger of water cavitations can occur appropriate measures have to meet.

WARNING!
The operator himself bears the sole responsibillty for the intended use of the flow meter
regarding the suitability and the comision resistance of the used materials against the measured
fluid.
The manufacturer is not liable for any damage resulting from improper use or use for other than
the intended purpose.
Do not use any abrasive or highly viscous media > 10 cP.

INFORMATION!
The sensors are fabricated from stainless steel 316 L (1.4404) or Hastelloy C22.
In your project planning, please observe the data given in the corrosion tables.
The pressure-bearing parts have been designed and rated for stationary operation taking
account of maximum pressure and temperature.
External forces and moments, caused e.g. by pipe stresses, have not been considered thereby.

Primarily, volumetric flow and temperature are measured, with pressure measurement as an
option. From these parameters the flowmeter calculates the mass flow or standard volumetric
flow using pre-programmed density data and then issues the measured values via various
communication interfaces.

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 5 Monday, June 16, 2008 1:56 PM

SAFETY 1

SITRANS FX300

1.2 Certifications

The flowmeter fulfills the statutory requirements of the following EC directives:

Pressure Equipment Directive 97/23/EC


Low Voltage Directive 73/23/EEC
EMC Directive 89/336/EC

As well as

EN 61010
EMC specification as per EN 61326/A1
NAMUR Recommendations NE 21 and NE 43

The manufacturer certifies successful testing of the product by applying the CE mark.

1.3 Safety instructions from the manufacturer


The flowmeter has been built and tested in accordance with the current state of the art, and
complies with the relevant safety standards.
However, dangers may arise from improper use or use for other than the intended purpose. For
this reason, observe all of the safety instructions in this document carefully.

1.3.1 Notes about the documentation


In addition to the safety rules and industrial safety regulations in this documentation, national and
regional safety rules and industrial safety regulations must also be observed.

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 6 Monday, June 16, 2008 1:56 PM

SAFETY

SITRANS FX300

1.3.2 Display conventions


The following symbols are used to help you navigate this documentation more easily:

WARNING!
These warning signs must be observed without fail. Even only partial disregarding such warnings
can result in serious health damage, damage to the device itself or to parts of the operators
plant.

DANGER!
This symbol designates safety advice on handling electricity.

CAUTION!
These warnings must be observed without fail. Even only partial disregarding such warnings can
lead to improper functioning of the device.

LEGAL NOTICE!
This symbol designates information on statutory directives and standards.

NOTE!
This symbol designates important information for the handling of the device.

HANDLING
This symbol designates all instructions for actions to be carried out by the operator in the
specified sequence.
i

CONSEQUENCE
This symbol designates all important consequences of the previous actions.

1.4 Safety instructions for the operator


CAUTION!
The flowmeters may only be installed, commissioned operated and maintained by properly
trained and authorized personnel.

This document must be read and understood in its entirety by all users prior to installation,
commissioning, operation and maintenance of the flowmeter.

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

INSTRUMENT DESCRIPTION

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 7 Monday, June 16, 2008 1:56 PM

SITRANS FX300

INSTRUMENT DESCRIPTION 2

2.1 Scope of delivery


INFORMATION!
Please check the contents of the consignment for completeness and intactness.

Figure 2-1: Scope of delivery

1
2
3
4

Device in ordered version


Operating instruction
Certificates, calibration report and parameter datasheet
Bar magnet

2.2 Instrument versions


The devices come in the following versions:

Converter with display


Blind device
Sensor in flanged design
Sensor in sandwich design

The following designs are available as options:

with pressure sensor


with isolation valve for the pressure sensor

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 8 Monday, June 16, 2008 1:56 PM

INSTRUMENT DESCRIPTION

SITRANS FX300

2.2.1 Versions with flanged connection


The compact unit consists of a measurement sensor and a signal converter. Together these form
a fixed mechanical unit.

Figure 2-2: Flanged instrument with display

1 Version with temperature sensor


2 Version with temperature sensor and optional pressure sensor
3 Version with temperature sensor, optional pressure sensor and shut-off valve

2.2.2 Sandwich versions


Sandwich-type flowmeters have 2 centering rings to facilitate installation.

Figure 2-3: Sandwich versions with display

1 Version with temperature sensor


2 Version with temperature sensor and optional pressure sensor
3 Version with temperature sensor, optional pressure sensor and shut-off valve

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 9 Monday, June 16, 2008 1:56 PM

SITRANS FX300

INSTRUMENT DESCRIPTION 2

2.2.3 Instrument description

Figure 2-4: Instrument description

1
2
3
4
5
6

Signal converter
Cable gland grey, standard design
Pressure sensor, optional
Shut-off valve, optional
Sensor
Centering ring

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 10 Monday, June 16, 2008 1:56 PM

INSTRUMENT DESCRIPTION

SITRANS FX300

2.3 Nameplate
INFORMATION!
Before installing the device, make sure that the information on the nameplate corresponds to the
ordering data.

Figure 2-5: Example of a nameplate

1
2
3
4
5
6
7
8

Device type
Manufacturer
Appointed ATEX &amp; DGRL body
Connection data: nominal diameter and pressure rating
DGRL data
Ex data
Electrical connection data
Manufacturer's website

Figure 2-6: Example of a nameplate

1
2
3
4
5
6
7

10

Serial number
Order number
Type code
Item number
fluid data
Software variant
Tag number

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 11 Monday, June 16, 2008 1:56 PM

SITRANS FX300

INSTRUMENT DESCRIPTION 2

2.4 Dual seal


To cover the requirements of ANSI/ISA-12.27.01-2003 Requirements for Process Sealing
Between Electrical Systems and Flammable or Combustible Process Fluid Dual seal, a
membrane vent is integrated in the neck of the instrument. This vent works in the improbable
case of a leakage as an annunciation valve between the primary seal (Process) and secondary
seal (electronic compartment).

1 Annunciation valve (Membrane vent)

The sealing between the pick-up and the Flow tube is considered as the primary seal. Its
construction material is always the same as the Flow tube itself (e.g. Stainless steel 316L or
Hastelloy C22). Chemicals used for the process, going through the Flow tube, must be
compatible with the designated material.
By using the membrane vent as annunciation valve, all requirements for a dual seal device in
accordance to above mentioned standard, are fulfilled.

It protects the electronic compartment against access by process media, due to possible
leakages.

The leakage of the primary seal is visible.

Regular maintenance should be performed by the customer to ensure this Dual Seal notification
system is working properly or that there is no leak.
If a leakage is noticed:
Siemens customer service should be contacted for servicing or replacement of the meter.

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

11

Installation

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 12 Monday, June 16, 2008 1:56 PM

INSTALLATION

SITRANS FX300

3.1 General installation instructions


CAUTION!
Installation, assembly, start-up and maintenance may only be performed by appropriately trained
personnel. The regional occupational health and safety directives must always be observed.

The following procedures have to be carried out before installing the device!
Check the packaging and the device itself for any damage.
Check the contents of the consignment for completeness.
Compare your order specification with the scope of delivery.

INFORMATION!
Comply with the application limits of the device with regard to pressure and temperature in
accordance with EN 1092-1 and/or ASME B 16.5. Additional information refer to Technical data
on page 49.

3.2 Storage

12

Store the flowmeter in a dry and dust-free location.


Avoid lasting direct exposure to the sun.
Store the flowmeter in its original packaging.
The permissible storage temperature is from -40C to +80C (-40F to +176F) for standard
meters.

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 13 Monday, June 16, 2008 1:56 PM

INSTALLATION 3

SITRANS FX300

3.3 Transport

Use carrying straps for transport and lay them around the two process connections.
Flowmeters must not be lifted by the converter housing for transport.
Never lift the flowmeter by the pressure sensor.
Do not use chains as these could damage the housing.

Figure 3-1: Transport instructions

CAUTION!
There is a risk of injury due to equipment that is not secured properly. The center of gravity of the
equipment is often higher than the lug for attaching the lifting belts.
During transport, prevent the flowmeter from unintentional slipping or turning.

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

13

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 14 Monday, June 16, 2008 1:56 PM

INSTALLATION

SITRANS FX300

3.4 Installation requirements


INFORMATION!
For accurate volumetric flow measurement the flowmeter needs a completely filled pipe and a
fully developed flow profile.
Please observe the instructions regarding inlet and outlet pipe runs as well as the installation
position.
In the case of vibrations on the piping choose the installation site in such a way that the
vibrations are at their lowest in transverse direction to the flowmeter.

CAUTION!
When installing the device in the piping, the following points must be observed:

Nominal diameter of connection pipe flange = nominal flange diameter of pipe!


Use flanges with smooth holes, e.g. welding neck flanges.
Carefully align the holes of the connecting flange and the flowmeter flange.
Check the compatibility of the gasket material with the process product.
Make sure that the gaskets are arranged concentrically. The flange gaskets must not project
into the pipe cross-section.
The flanges have to be concentric.
There must not be any pipe bends, valves, flaps or other internals in the immediate inlet run.
Sandwich-type flowmeters may only be installed using a centering ring.
Never install the flowmeter directly behind piston compressors or rotary piston meters.
Do not lay signal cables directly next to cables for the power supply.

3.4.1 General
INFORMATION!
A sufficient distance around the flowmeter in all directions must be kept free to allow servicing.

Figure 3-2: Temperature ranges

1 Max. medium temperature see tech. datasheet


2 Max. storage temperature and ambient temperature
Ambient temp.: Ex devices -20...+65C, non Ex devices -40...+85C (-40...+185F)
Storage temperature: -50...+85C (-58... +185F)
3 Protect device from direct exposure to sunlight

14

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 15 Monday, June 16, 2008 1:56 PM

INSTALLATION 3

SITRANS FX300

3.4.2 Liquid measurement


Prohibited installation

Figure 3-3: Upper pipe bend

CAUTION!
Prohibited: Installing the flowmeter in an upper pipe bend 1, because there is a risk of gas
bubbles 2 forming. Gas bubbles can lead to pressure surges and inaccurate measurement.

Figure 3-4: Downpipe and outlet

CAUTION!
Prohibited: Installing the device in a downpipe 3 or upstream of a run-out 4. There is the risk of
a partially-filled pipe.

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

15

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 16 Monday, June 16, 2008 1:56 PM

INSTALLATION

SITRANS FX300

Recommended installations

CAUTION!
It is absolutely necessary to maintain the required inlet runs and outlet runs!

1
2
3
4

16

If the device is installed in a downpipe, it is absolutely essential to install a riser pipe immediately after it.
Installing the device in an inclined riser pipe.
Installing the device in a vertical riser pipe.
Installing the device in the bottom pipe bend.

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 17 Monday, June 16, 2008 1:56 PM

INSTALLATION 3

SITRANS FX300

3.4.3 Vapour and gas measurement

Figure 3-5: Non-recommended installation

1 Lower pipe bends


2 Condensate

DANGER!
Do not install the instrument in a lower bend 1 because the risk of condensate 2 occurs.
Condensate can lead to cavitation and inaccurate measurement. Under certain circumstances
the instrument can be destroyed and the measured medium can leake.

3.4.4 Piping with control valves


INFORMATION!
To ensure trouble-free and accurate measurement, the manufacturer recommends not installing
the flowmeter downstream of a control valve. There is a risk of turbulence occurring that distorts
the measuring result.

Figure 3-6: Pipeline with control valve

1 Recommended: Installing the flowmeter upstream of the control valve, distance 5 DN


2 Not recommended: Installing the flowmeter directly downstream of control valves, due to vortex formation.

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

17

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 18 Monday, June 16, 2008 1:56 PM

INSTALLATION

SITRANS FX300

3.4.5 Preferred installation location

Figure 3-7: Preferred installation locations

1 Above a horizontal pipe


2 Below a horizontal pipe
3 On a vertical pipe

INFORMATION!
Depending on installation position you will have to turn round the display and/or the connection
housing. Additional information refer to Turning the display on page 20.

18

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 19 Monday, June 16, 2008 1:56 PM

INSTALLATION 3

SITRANS FX300

3.4.6 Turning the connection housing


DANGER!
All work on the device electrics may only be carried out by appropriately trained personnel. The
regional occupational health and safety directives must always be observed.

CAUTION!
Do not damage the electrical cable by over tightening.
Do not undo the electrical plug connection.

Figure 3-8: Allen screws on converter housing

Switch off the power supply to the flowmeter.


Loosen the four screws 1 on the underside of the connection housing.
Lift the connection housing and turn it to the required position in 90 stages.
Screw the connection housing back on tightly.

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

19

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 20 Monday, June 16, 2008 1:56 PM

INSTALLATION

SITRANS FX300

3.4.7 Turning the display


DANGER!
All work on the flowmeter electrics may only be carried out by appropriately trained personnel.
The regional occupational health and safety directives must always be observed.

INFORMATION!
If the flowmeter is installed in a vertical pipe run, you will have to turn the display by 90; if
installed below a pipe, turn 180.

Figure 3-9: Turning the display

Turn the display as follows:


Switch off the power supply to the flowmeter.
Unscrew the cover in front of the display 1 from the connection housing.
Pull the display 2 carefully a few centimetres out of the anchor fitting and turn it to the required
position 3.
Press the display back onto the spacer pins 4, until it clicks firmly into place.
Turn the cover with gasket 5 back onto the housing and tighten it by hand.

20

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 21 Monday, June 16, 2008 1:56 PM

INSTALLATION 3

SITRANS FX300

3.4.8 Thermal insulation


CAUTION!
The area above the converter support must not be thermally insulated.
The thermal insulation 3 may only extend to the maximum height 1 shown below up to the
connecting screws of the sensor.

Figure 3-10: Thermal insulation on connection piece and signal cable

1 Max. height of insulation up to intermediate piece between sensor and converter


2 Max. thickness of the insulation up to the bend of the pressure pipe
3 Insulation

CAUTION!
The thermal insulation 3 may only extend as far as the bend in the pressure measurement
piping 2.

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

21

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 22 Monday, June 16, 2008 1:56 PM

INSTALLATION

SITRANS FX300

3.5 Inlet and outlet runs


3.5.1 Minimum inlet runs

Figure 3-11: Minimum inlet runs

1 General inlet run without disturbing flow 20 DN


2 Behind a control valve 50 DN
3 After a pipe diameter reduction 20 DN

22

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 23 Monday, June 16, 2008 1:56 PM

INSTALLATION 3

SITRANS FX300

Figure 3-12: Minimum inlet run

1 After a single bend 90 20 DN


2 After a double bend 2x90 30 DN
3 After a double three-dimensional bend 2x90 40 DN

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

23

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 24 Monday, June 16, 2008 1:56 PM

INSTALLATION

SITRANS FX300

3.5.2 Flow straightener


If, due to the type of installation, the required inlet runs are not available, the manufacturer
recommends using flow straighteners. Flow straighteners are installed between two flanges
upstream of the flowmeter and shorten the required inlet run.

Figure 3-13: Flow straightener

1
2
3
4

Straight inlet run upstream of straightener 2 DN


Flow straightener
Straight pipe run between flow straightener and flowmeter 8 DN
Minimum straight outlet run 5 DN

3.5.3 Minimum outlet runs

Figure 3-14: Minimum outlet runs

1 Upstream of pipe expanders, pipe bends, control valves, etc. 5 DN


2 Upstream of metering points 5 6 DN

INFORMATION!
The interior of the pipe at the metering points must be free of burrs and other flow impediments.
The flowmeter has an internal temperature sensor. External temperature sensor must be located
5 DN behind the flow sensor outlet. Use sensors that are as short as possible to avoid
disturbing the flow profile.

24

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 25 Monday, June 16, 2008 1:56 PM

INSTALLATION 3

SITRANS FX300

3.6 Installation
3.6.1 General installation instructions
CAUTION!
Installation, assembly, start-up and maintenance may only be performed by appropriately trained
personnel. The regional occupational health and safety directives must always be observed.

The following procedures have to be carried out before installing the flowmeter in the piping!
Remove all transport safeguards and protective coverings from the flowmeter.
Make sure that the gaskets have the same diameter as the piping.
Make sure that the flowmeter is in the correct flow direction. This is indicated by an arrow on
the sensor housing.

CAUTION!
On metering points with varying thermal loads the flowmeters have to be mounted with stress
bolts (DIN 2509).
Stress bolts and/or nuts and bolts are not included in the scope of delivery.
Make absolutely certain that the measuring flanges are sitting concentrically.
When preparing the metering point, keep in mind the exact installed length of the flowmeter.
Additional information on the dimensions refer to Dimensions and weights on page 51.

Figure 3-15: Preparing the metering point

1 Installed length of flowmeter + thickness of gaskets.

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

25

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 26 Monday, June 16, 2008 1:56 PM

INSTALLATION

SITRANS FX300

CAUTION!
The inside diameters of the pipes, of the measuring sensor and the gaskets must be the same!
The gaskets must not project into the pipe cross-section.

Figure 3-16: Internal diameter

1 Internal diameter connection pipe


2 Internal diameter flange and gasket
3 Internal diameter measuring sensor

26

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 27 Monday, June 16, 2008 1:56 PM

INSTALLATION 3

SITRANS FX300

3.6.2 Installing instruments in sandwich design

Figure 3-17: Installation using centering ring

1
2
3
4
5
6

Sensor
Centering ring
Bolts with fixing nuts
Hole
Hole
Gasket

Install the flowmeter as follows:


Push the first bolt 3 through hole 5 of both flanges.
Screw on the nuts with washers at both ends of the bolt 3 but do not tighten them.
Mount the second bolt through the holes 4.
Place the measured value sensor 1 between the two flanges.
Insert the gaskets 6 between measured value sensor 1 and flanges and align them.
Check that the flanges are sitting concentrically.
Install the remaining bolts, nuts and washers. But do not tighten the nuts yet.
Turn the centering ring 2 anti-clockwise and align the flowmeter.
Check that the gaskets 6 are concentric; they must not project into the pipe.
Now tighten all nuts bit by bit alternately across the diagonal.

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

27

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 28 Monday, June 16, 2008 1:56 PM

INSTALLATION

SITRANS FX300

3.6.3 Installing instruments in flange design

Figure 3-18: Installing devices in flange design

1 Gasket
2 Bolts with fixing nuts

Install the flowmeter as follows:


Fix the flowmeter to one side of a flange using bolts and fixing nuts 2.
Insert the gaskets 1 between sensor and flange and align them.
Check that the gaskets are concentric; they must not project into the pipe.
Install the gasket, bolts and fixing nuts on the other flange side.
Align the meter and the gaskets so that they are concentric.
Now tighten all nuts bit by bit alternately across the diagonal.

28

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

Electrical connections

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 29 Monday, June 16, 2008 1:56 PM

ELECTRICAL CONNECTIONS 4

SITRANS FX300

4.1 Safety instructions


DANGER!
Only perform work on the electrical connections in the de-energized state.
Take note of the voltage data on the nameplate!

DANGER!
Observe national installation regulations!

CAUTION!
All work on the flowmeter electrics may only be carried out by appropriately trained personnel.
The regional occupational health and safety directives must always be observed.

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

29

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 30 Monday, June 16, 2008 1:56 PM

ELECTRICAL CONNECTIONS

SITRANS FX300

4.2 Connecting the signal converter

Figure 4-1: Signal converter housing with housing cover

1
2
3
4
5
6
7
8

Housing cover of the electrical terminal compartment


Electrical connection terminals with the housing cover open
Terminal A current output Terminal A+ current output +
Terminal B pulse output Terminal B+ Pulse output +
PE terminal in housing
PE terminal on connection piece between sensor and converter

The following procedures are to be performed:


Unscrew the housing cover 1 of the electrical terminal compartment.
Thread the connection cable through the leadthrough in the housing.
Connect the cable for the current output and the cable for the optional pulse output as shown in
the cable terminal diagrams below. To facilitate installation the connection plug can be
removed from the meter. The plug and socket are configured in such a way as to prevent
reverse polarity.
Connect the earthing to terminal 7 alternatively use the PE terminal on the connection piece
between sensor and converter 8.
Tighten he cable glands.
Manually screw on the housing cover with gasket.

30

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 31 Monday, June 16, 2008 1:56 PM

ELECTRICAL CONNECTIONS 4

SITRANS FX300

4.3 Current output, electrical data


4.3.1 Power supply
INFORMATION!
The supply voltage has to be between 14 VDC and 36 VDC. This is based on the total resistance
of the measuring loop. To determine this, add up the resistances of each component in the
measuring loop (not including the flowmeter).

The required supply voltage can be calculated using the formula below:

Ub = Rtot * 22mA + 14V

where
UB = the minimum supply voltage and
Rtot = the total measuring loop resistance

INFORMATION!
The power supply has to be able to supply a minimum of 22 mA.

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

31

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 32 Monday, June 16, 2008 1:56 PM

ELECTRICAL CONNECTIONS

SITRANS FX300

4.3.2 Load for HART communication


INFORMATION!
For HART communication a load of at least 230 ohm is required.

The maximum load resistance is calculated as follows:

R max =

UB - 14 V
22 mA

Figure 4-2: HART current output connection

1 Power supply
2 Optional amperemeter
3 Load

DANGER!
Use a twisted two-core cable to prevent electrical interference from impeding the DC output
signal.
In some cases a shielded cable may be necessary. The cable shield may only be earthed
(grounded) at one place (on the power supply unit).

32

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 33 Monday, June 16, 2008 1:56 PM

SITRANS FX300

ELECTRICAL CONNECTIONS 4

4.4 Pulse output, electrical data


INFORMATION!
When using the pulse output, two separate signal circuits are necessary if the pulse output is
utilized together with analog signals. Each signal circuit requires its own power supply.
The total resistance must be adapted so that the total current Itot does not exceed 100 mA.

4.4.1 Connection pulse output

Figure 4-3: Electrical connection pulse output

1
2
3
4

Power supply
Optional amperemeter
Load
e.g. counter

4.4.2 Configuring the pulse output


The pulse output is a passive frequency, "open collector" output, isolated from the current
interface and the sensor. It can be configured as a high current output or NAMUR output using a
jumper on the amplifier board.

Figure 4-4: Data pulse output

1
2
3
4
5

Pulse
closed
open
f<sub>max</sub> = 0.5 Hz
250 ms

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

33

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 34 Monday, June 16, 2008 1:56 PM

ELECTRICAL CONNECTIONS

SITRANS FX300

Jumper in NAMUR setting


Ri = 900

Umax = 36 VDC

The maximum frequency of the pulse output is: fmax = 0.5 Hz

Jumper in high current setting


Open:

Maximum voltage Umax = 36 VDC

Closed current IR < 1 mA

Closed:

Maximum current Imax = 100 mA

Voltage U < 2 VDC

The maximum frequency of the pulse output is: fmax = 0.5 Hz

Figure 4-5: Jumper settings pulse output

1
2
3
4

34

Electronic insert
Jumper on board
Jumper in high current setting
Jumper in NAMUR setting

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 35 Monday, June 16, 2008 1:56 PM

SITRANS FX300

ELECTRICAL CONNECTIONS 4

4.5 Protection category


The flowmeter meets all requirements of protection category IP 66/67.

DANGER!
After all servicing and maintenance work on the flowmeter, the specified protection category has
to be ensured again.

Figure 4-6: Cable feedthrough

Therefore it is essential to observe the following points:


Use only original gaskets. They must be clean and free of any damage. Defective gaskets have
to be replaced.
The electrical cables used must be undamaged and must comply with regulations.
The cables must be laid with a loop 1 upstream of the flowmeter to prevent water from getting
into the housing.
The cable glands 2 have to be tightened firmly.
Align the meter in such a way that the cable gland is never pointing up 3.
Close the unused cable glands using blanking plugs 4.
Do not remove the specified outer sheath from the cable gland.

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

35

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 36 Monday, June 16, 2008 1:56 PM

ELECTRICAL CONNECTIONS

SITRANS FX300

4.6 Earthing connections


The earthing can be done either by connecting the PE terminal in the housing or the PE terminal
on the connection piece between sensor and converter. Both of these electrical connections are
equally effective from a technical point of view.

Figure 4-7: Earth connection

1 Electrical earthing connection on piece between sensor and converter.


2 Electrical earthing connection earth on housing

DANGER!
The flowmeter has to be earthed properly to achieve accurate measurement. The earthing cabel
may not transfer any interference voltage.
Do not use this earth cabel to ground any other items of electrical equipment.

36

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

Start-up and control

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 37 Monday, June 16, 2008 1:56 PM

SITRANS FX300

START-UP AND CONTROL 5

5.1 Start-up
INFORMATION!
The device is always preset for the user and his application.
Compare the data in the factory settings report with your specifications!
The factory settings report is part of the scope of delivery.

5.1.1 Start
After the device is switched on, the display shows the following in sequence

Test,
The type of meter, and
The version number.

Afterwards the device performs a self-test and switches to measurement mode. Here all of the
parameters preset for the customer are analysed and checked for plausibility, and the last
measured value is displayed.

5.1.2 Start-up and control


INFORMATION!
The flowmeter is largely maintenance-free.
Observe the application limits in respect of temperature and medium. Additional information refer
to Technical data on page 49.

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

37

Operation

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 38 Monday, June 16, 2008 1:56 PM

OPERATION

SITRANS FX300

6.1 Display and operating elements


With the cover open the meter is operated by using the mechanical keys on the front; with the
cover closed a bar magnet is used.

CAUTION!
The switching point for the magnetic sensors is located directly under the glass above
corresponding symbol. Touch the symbol with the bar magnet only at right angles from the front.
Touching it sideways can cause incorrect operation.

Figure 6-1: Display and operating elements

1
2
3
4
5
6
7

38

Enter button (bar magnet)


Right button (bar magnet)
Up button (bar magnet)
Up button (mechanical)
Enter button (mechanical)
Right button (mechanical)
Display

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 39 Monday, June 16, 2008 1:56 PM

OPERATION 6

SITRANS FX300

The mechanical keys and keys for the bar magnet are identical in their function. In this
documentation the keys are represented as symbols to describe the operating functions:

Mechanical

Bar magnet

Symbol

Table 6-1: Description of the operating keys

6.2 Operating principles


6.2.1 Description of the key functions

Switch from measurement mode to menu mode


Switch to one menu level lower
Open menu item and activate change mode
In change mode: Move the input cursor one position to the right; after the last digit the
input cursor jumps back to the beginning.

In measurement mode: Switch between measured values and error messages


Change between the menu items within a menu level
In change mode: Changing parameters and settings; running through the available
characters; shifting the decimal point to the right.

Change one level up at the menu


Return to measuring mode with a query whether the data should be accepted

Table 6-2: Description of the opperation buttons

6.2.2 Switch from measuring mode to menu mode


Measuring mode

Operation

Menu mode

156.3
kg/h

1.1.1
Language

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

39

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 40 Monday, June 16, 2008 1:56 PM

OPERATION

SITRANS FX300

6.2.3 Navigation within the menu structure


Navigation within the menu is by means of the and ^ keys. Pressing the key takes you one
menu level lower, ^ takes you one menu level higher.
If you are already located at the lowest level (function level), you can use the key to go the the
change mode, which can be used to set data and values.
If you are located at the first level (main menu), you can use the ^ key to exit the menu mode
and return to the measuring mode.

Measuring
operation

Main menu

Submenu

Function

Edit
^

Table 6-3: Navigation menu structure

6.2.4 Changing the settings in the menu


Using the and keys, navigate to the menu in which you wish to change a setting or a value.
Use the key to activate change mode in the selected menu.
i

The current values or settings are displayed.

Change the value or setting using the and keys.


Confirm the new value or setting using the ^ key.
Some menu items include several setting options. These are displayed one after the other by
pressing the ^ key.
i

You return to the main menu.

Press the ^ to save the settings you have made.


i

The following prompt is displayed: "Save Yes". Switch between "Yes" and "No" using the
key.

Switch between "YES" and "NO" using the key.

Quit Yes

^ The changes are applied, an update is performed and the display switches back to
measuring mode. Measuring operation.

Quit No

^ The changes are discarded and the display switches back to measuring mode.
Measuring operation.

CAUTION!
Each time parameters or settings are changed, the flowmeter carries out an internal plausibility
check.
If implausible inputs have been made, the indicator remains in the current menu, and the
changes are not accepted.

40

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 41 Monday, June 16, 2008 1:56 PM

OPERATION 6

SITRANS FX300

Example: Changing the default parameter from m3/h to l/min


Steps

Display

Step

107.2
m3/h

5 ^

Display
0000600.00
L/min

3x

1.1.1
Language

Display
unit

3x

1.1.4
Max. Flow

1.1.4
Max. Flow

m3/h
unit

Quit Yes

3x

L/min
unit

1787.00
L/min

6.2.5 Measures in the event of faulty indications


If the indications on the display or the responses to keypad commands are faulty, you have to do
a hardware reset. Switch the power supply OFF and ON again.

6.3 Overview of the most important functions and indicators


INFORMATION!
A complete list of all functions and short descriptions is provided in the appendix. All default
parameters and settings are adapted for the specific customer.

Level

Designation

Explanation

1.1.1

Language

Select the menu language

1.1.4

Max. Flow

Maximum flow rate


The value set is represented by a 20 mA current at the analog output.
If the current value exceeds the preset value, an alarm is indicated.

1.1.5

Min. Flow

Minimum flow rate


The set value does not represent the 4 mA value of the current output.

1.1.6

Timeconst.

Time constant, damping value [s]

2.1.1

Test I

Check current output

2.1.2

Test P

Pulse pulse output

3.1.1

Error Msg

Error indicator
Yes: Error messages are displayed
No: Error messages are suppressed, but a flashing cursor in the top left corner
shows that there are error messages.

Table 6-4: The most important functions

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

41

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 42 Monday, June 16, 2008 1:56 PM

OPERATION

SITRANS FX300

Measured variables Units

Media (equipment type)

Volume

m3/h
m3/min
m3/s
L/h
L/min
L/s
ft3/h
ft3/min
ft3/s
gal/h
gal/min
gal/s
ImpGal/h
ImpGal/min
ImpGal/s

Liquids
Vapours
Gases

Mass

kg/h
kg/min
kg/s
t/h
t/min
t/s
lb/h
lb/min
lb/s

Liquids
Vapours
Gases

Standard volume

m3/h (std.)
m3/min (std.)
m3/s (std.)
L/h (std.)
L/min (std.)
L/s (std.)
ft3/h (std.)
ft3/min (std.)
ft3/s (std.)

Gases

Table 6-5: Meter types, measured flow variables and units

42

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 43 Monday, June 16, 2008 1:56 PM

OPERATION 6

SITRANS FX300

6.4 Error messages


Error message

Cause

Corrective action

No Signal

No signal from the vortex amplifier

Check plug connection


At sensor problems contact Service

Low Freq

Sample frequency too low

Contact Service.

High Freq

Sample frequency too high

Contact Service.

Low Flow

Flow rate lower than minimum flow rate


qmin

Converter continues to show the current


flow rate, but the accuracy of the
measurement can be impaired.

High flow

Flow rate higher than maximum flow rate


qmax

Correction procedures are depending of


the application!
If flow exceeds the maximum value, the
sensor can be damaged physically.

Inv.Config

Configuration data in FRAM (non-volatile


memory) is not valid

Check entire configuration!


If error message persists, contact
Service

Amp. Fail

Pre-amplifier section has failed

Contact Service.

Chk.Instal

Flow signal quality is bad

1. Check qmin
2. Excessive pipe vibrations and
impeded flow profile

Low Signal

Vortex signal amplitude too low

1. qmin ok?
2. if qmin is ok, contact Service

Hi.Signal

Vortex signal amplitude too high

Occurs in high-density media


1. Check qmax
2. If qmax is ok, contact Service

L.Temp.Phy

Operating temperature lower than


physical limit

Take corrective action within the


process.

H.Temp.Phy

Operating temperature higher than


physical limit

Take corrective action within the process


as quickly as possible, otherwise
damage can occur to both sensor and
converter.

H.Pres.Phy

Operating pressure higher than physical


limit

Take corrective action within the process


as quickly as possible, otherwise
damage can occur to both sensor and
converter.

Tsens Shrt

Temperature sensor short-circuit

Shows error on temperature sensor!


Contact Service.

Tsens Open

Temperature sensor open circuit

Shows error on temperature sensor!


Contact Service.

P.Sen.Fail

Failure in pressure sensor

Shows error at pressure sensor!


Contact Service.

Table 6-6: Error messages

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

43

Service

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 44 Monday, June 16, 2008 1:56 PM

SERVICE

SITRANS FX300

7.1 Replacing converter / LCD display


The converter has to be replaced by a converter of the same type. The following parameters
have to be observed for it:

The item numbers must match: 2.143670.xxx


The software version 2 must match.

The basic version does not have any software identification


The gas version has the designation: "gas"
The steam version has the designation: "steam"

Figure 7-1: Board, example

1
2
3
4
5

44

Connector for display


Version number, software identification
Connector for sensor
Spacer pins
Fixing screw (Phillips)

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 45 Monday, June 16, 2008 1:56 PM

SERVICE 7

SITRANS FX300

Figure 7-2: Board, example

1
2
3
4
5

Connector for display


Version number, software identification
Connector for sensor
Spacer pins
Fixing screw (Phillips)

Perform the following steps:


Switch off the power supply
Unscrew front cover
Detach display from spacer pins.
Detach display cable 1.
Detach sensor cable 3.
Undo three fixing screws 5 (Phillips).
Remove converter.
Insert new converter.
Tighten three screws.
Attach sensor cable 3 (caution: the sensor cable must not be above the spacer pin 4 risk of
damage).
Attach display cable 1.
Mount display in required position, exerting uniform pressure over the entire surface.
Screw the cover on tightly by hand.

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

45

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 46 Monday, June 16, 2008 1:56 PM

SERVICE

SITRANS FX300

7.2 Returning the device to the manufacturer


7.2.1 General information
This device has been carefully manufactured and tested. If installed and operated in accordance
with these operating instructions, it will rarely present any problems.

CAUTION!
Should you nevertheless need to return a device for inspection or repair, please pay strict
attention to the following points:
Due to statutory regulations on environmental protection and safeguarding the health and
safety of our personnel, manufacturer may only handle, test and repair returned devices that
have been in contact with products without risk to personnel and environment.
This means that the manufacturer can only service this device if it is accompanied by the
following certificate (see next section) confirming that the device is safe to handle.

CAUTION!
If the device has been operated with toxic, caustic, flammable or water-endangering products,
you are kindly requested:
to check and ensure, if necessary by rinsing or neutralizing, that all cavities are free from such
dangerous substances,
to enclose a certificate with the device confirming that is safe to handle and stating the
product used.

46

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 47 Monday, June 16, 2008 1:56 PM

SERVICE 7

SITRANS FX300

7.2.2 Form (for copying) to accompany a returned instrument


Company:

Address:

Department:

Name:

Tel. no.:

Fax no.:

Manufacturer's order no. or serial no.:

The device has been operated with the following medium:

This medium is:

water-hazardous
toxic
caustic
flammable
We checked that all cavities in the device are free from such
substances.
We have flushed out and neutralized all cavities in the
device.

We hereby confirm that there is no risk to persons or the environment through any residual media
contained in the device when it is returned.

Date:

Signature:

Stamp:

Returning adress:
Siemens Flow Instruments A/S
Nordborgvej 81
DK - 6430 Nordborg
Att. Quality department E1-21

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

47

Technical data

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 48 Monday, June 16, 2008 1:56 PM

TECHNICAL DATA

SITRANS FX300

8.1 Functional principle


Vortex flowmeters are used to measure the volumetric flow of gases, vapours and liquids at
completly filled pipes.
The measuring principle is based on the Karman vortex street. The measuring tube contains a
bluff body, behind which vortex shedding occurs. The frequency f of the vortex shedding is
proportional to the flow rate v. The nondimensional Strouhal number S describes the relationship
between vortex frequency f, width b of bluff body and the mean flow velocity v:

S .v
b

The vortex frequency is recorded at the sensor and evaluated at the converter.

b b

Figure 8-1: Functional principle

48

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 49 Monday, June 16, 2008 1:56 PM

TECHNICAL DATA 8

SITRANS FX300

8.2 Technical data


Measuring system
Application range

Flow measurement of liquids and gases

Operating method / measuring principle

Karman vortex street

Measured value
Primary measured value

Number of separated vortices

Secondary measured value

Operating and standard volumetric flow, mass flow

Measuring accuracy
Accuracy

Re 20000 0.75% for liquids


Re 20000 1% for gases and vapours
10000 < Re < 20000 2% for liquids, gases and vapours
1

Repeatability

0.1%

Stability

0.1% over a periode of 1 year

Operating conditions
Ambient temperature

-40...+65C (-40...+149 F) Ex version


-40+85C (-40...+185 F) (non Ex version)

Storage temperature

-50+85C (-58...+185 F)

Product temperature

-40+240C (-40...+464 F)

Measured substances

Liquids, gases and vapour

Density

taken into consideration when rating

Viscosity

< 10 cP

Reynolds number

10000...2300000

Product pressure limit

max. 100 bar, higher pressures on request

Inlet conditions
Inlet run

20 x DN

Outlet run

5 x DN

Dimensions and weights

see chapter dimensions and weights

Materials
Sensor

1.4404/316L Hastelloy C22

Housing

Aluminium; 1.4404/316L in preparation

Sensor gasket

1.4435/316L / FPM; Hastelloy C22 / FFKM

1 Accuracy pressure- and temperature-compensated Re 20000 +/- 1.5% for gases and vapours; 10000 < Re < 20000 +/- 2.5% for gases
and vapours

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

49

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 50 Monday, June 16, 2008 1:56 PM

TECHNICAL DATA

SITRANS FX300

Power supply
Ex- version

1430 VDC

Non Ex- version

1436 VDC

Current output
Measuring range

4...20 mA

Over Range

20.8 mA +/- 1% (105% +/- 1%)

Load

Minimal 100 ; maximal Rges = ((UB - 14 V) / 22 mA)

Error signal

NAMUR NE 43
(-2.5 +/- 0.5%)
20.5 +/- 1.0% (105% +/- 1.0%)

Maximal

22 mA (112.5%)

Multidrop mode

4 mA

Digital output
HART
Physical layer

FSK

Device category

Transmitter

Pulse output
Pulse output

Pulse frequency max. 0.5 Hz

Power supply non- Ex

24 VDC as NAMUR or open < 1 mA, maximal 36 V, closed


100 mA, U < 2 V

Power supply Ex

24 VDC as NAMUR or open < 1 mA, maximal 30 V, closed


100 mA, U < 2 V

Display and operating interface


Local display

2 lines, 10 characters per line

Operating and display languages

German, English, French

Process connections
Process connection

EN or ASME flanges

Flange version

DN 15...300; ...12"

Sandwich version

DN 15...100; ...4"

Protection category
Protection category

IP 66/67

Approvals
ATEX

ATEX II 2G EEx d ia [ia] IIC T6

FM

Class 1 Div. 1 1

1 Pending

50

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 51 Monday, June 16, 2008 1:56 PM

TECHNICAL DATA 8

SITRANS FX300

8.3 Dimensions and weights


a = 133 mm /
5.24 inch

b = 105 mm /
4.13 inch
c = 179 mm /
7.05 inch

Flange version EN 1092-1


Pressure
rating

Size

Dimensions [mm]

Weight [kg]
With

DN

PN

Without

pressure sensor

15

40

17.3

95

200

265

144

6.1

5.5

15

100

17.3

105

200

265

144

7.1

6.5

25

40

28.5

115

200

265

144

7.9

7.3

25

100

28.5

140

200

265

144

9.9

9.3

40

40

43.1

150

200

270

144

10.8

10.2

40

100

42.5

170

200

270

144

14.8

14.2

50

16

54.5

165

200

275

144

12.7

12.1

50

40

54.5

165

200

275

144

12.9

12.3

50

63

54.5

180

200

275

144

16.9

16.3

50

100

53.9

195

200

275

144

18.4

17.8

80

16

82.5

200

200

290

154

17.4

16.8

80

40

82.5

200

200

290

154

19.4

18.8

80

63

81.7

215

200

290

154

23.4

22.8

80

100

80.9

230

200

290

154

27.4

26.8

100

16

107.1

220

250

310

164

22

21.4

100

40

107.1

235

250

310

164

25

24.4

100

63

106.3

250

250

310

164

30

29.4

100

100

104.3

265

250

310

164

36

35.4

150

16

159.3

285

300

325

174

35.8

35.2

150

40

159.3

300

300

325

174

41.8

41.2

150

63

157.1

345

300

325

174

59.8

59.2

150

100

154.1

355

300

325

174

67.8

67.2

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

51

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 52 Monday, June 16, 2008 1:56 PM

TECHNICAL DATA

Size

SITRANS FX300

Pressure
rating

Dimensions [mm]

Weight [kg]
With

DN

PN

Without

pressure sensor

200

10

206.5

340

300

350

194

38.4

37.8

200

16

206.5

340

300

350

194

38.4

37.8

200

25

206.5

360

300

350

194

47.4

46.8

200

40

206.5

375

300

350

194

55.4

54.8

250

10

260.4

395

380

370

224

58.0

57.4

250

16

260.4

405

380

370

224

59.0

58.4

250

25

258.8

425

380

370

224

75.0

74.4

250

40

258.8

450

380

370

224

93.0

92.4

300

10

309.7

445

450

395

244

76.3

75.7

300

16

309.7

460

450

395

244

82.8

82.2

300

25

307.9

485

450

395

244

99.3

98.7

300

40

307.9

515

450

395

244

128.1

127.5

Option: Version with two transmitters


Dimension H x 2
Specified weight + 2.80 kg

52

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 53 Monday, June 16, 2008 1:56 PM

TECHNICAL DATA 8

SITRANS FX300

a = 133 mm /
5.24 inch

b = 105 mm / 4.13 inch


c = 179 mm / 7.05 inch

Flange version ASME B16.5


Pressure
rating

Size

Dimensions [mm]

Weight [kg]
With

NPS

Class

Without

pressure sensor

1/2

150

15.8

90

200

265

144

5.1

4.5

1/2

300

15.8

95

200

265

144

5.5

4.9

1/2

600

13.9

95

200

265

144

5.7

5.1

150

26.6

110

200

265

144

6.8

6.2

300

26.6

125

200

265

144

7.8

7.2

600

24.3

125

200

265

144

8.1

7.5

1 1/2

150

40.9

125

200

270

144

8.9

8.3

1 1/2

300

40.9

155

200

270

144

11

10.4

1 1/2

600

38.1

155

200

270

144

12

11.4

150

52.6

150

200

275

144

11.6

11

300

52.6

165

200

275

144

13

12.4

600

49.3

165

200

275

144

14.5

13.9

150

78

190

200

290

154

20.4

19.8

300

78

210

200

290

154

23.4

22.8

600

73.7

210

200

290

154

24.4

23.8

150

102.4

230

250

310

164

24

23.4

300

102.4

255

250

310

164

32

31.4

600

97.2

275

250

310

164

41

40.4

150

154.2

280

300

325

174

36.8

36.2

300

154.2

320

300

325

174

51.8

51.2

600

146.3

355

300

325

174

76.8

46.2

150

202.7

345

300

350

194

50.6

50.0

300

202.7

380

300

350

194

75.4

74.8

10

150

254.5

405

380

370

224

75.0

74.4

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

53

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 54 Monday, June 16, 2008 1:56 PM

TECHNICAL DATA

SITRANS FX300

Pressure
rating

Size

Dimensions [mm]

Weight [kg]
With

NPS

54

Class

Without

pressure sensor

10

300

254.5

455

380

370

224

107.0

106.4

12

150

304.8

485

450

395

244

106.9

106.3

12

300

304.8

520

450

395

244

151.9

151.3

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 55 Monday, June 16, 2008 1:56 PM

TECHNICAL DATA 8

SITRANS FX300

a = 133 mm /
5.24 inch

b = 105 mm / 4.13 inch


c = 179 mm / 7.05 inch

Flange version ASME B16.5


Pressure
rating

Size

Dimensions [inches]

Weight [lb]
With

NPS

Class

Without

pressure sensor

1/2

150

0.62

3.54

7.87

10.43

5.67

11.24

9.92

1/2

300

0.62

3.74

7.87

10.43

5.67

12.13

10.8

1/2

600

0.54

3.74

7.87

10.43

5.67

12.57

11.24

150

1.05

4.33

7.87

10.43

5.67

14.99

13.67

300

1.05

4.92

7.87

10.43

5.67

17.2

15.87

600

0.96

4.92

7.87

10.43

5.67

17.86

16.53

1 1/2

150

1.61

4.92

7.87

10.63

5.67

19.62

18.3

1 1/2

300

1.61

6.1

7.87

10.63

5.67

24.25

22.93

1 1/2

600

1.5

6.1

7.87

10.63

5.67

26.46

25.13

150

2.07

5.91

7.87

10.83

5.67

25.57

24.25

300

2.07

6.5

7.87

10.83

5.67

28.66

27.34

600

1.94

6.5

7.87

10.83

5.67

31.97

30.64

150

3.07

7.48

7.87

11.42

6.06

44.97

43.65

300

3.07

8.27

7.87

11.42

6.06

51.59

50.26

600

2.9

8.27

7.87

11.42

6.06

52.79

52.47

150

4.03

9.06

9.84

12.21

6.46

52.91

51.59

300

4.03

10.04

9.84

12.21

6.46

70.55

69.22

600

3.83

10.83

9.84

12.21

6.46

90.39

89.07

150

6.07

11.02

11.81

12.8

6.85

81.13

79.81

300

6.07

12.6

11.81

12.8

6.85

114.2

112.88

600

5.76

13.98

11.81

12.8

6.85

169.31

101.85

150

7.98

13.58

11.81

13.78

7.64

146.39

145.65

300

7.98

14.96

11.81

13.78

7.64

190.32

189.65

10

150

10.02

15.51

14.96

14.57

8.82

197.09

195.77

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

55

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 56 Monday, June 16, 2008 1:56 PM

TECHNICAL DATA

SITRANS FX300

Pressure
rating

Size

Dimensions [inches]

Weight [lb]
With

NPS

56

Class

Without

pressure sensor

10

300

10.02

17.91

14.96

14.57

8.82

252.21

239.86

12

150

12

19.09

17.72

15.55

9.61

318.34

317.02

12

300

12

20.47

17.72

15.55

9.61

415.35

414.02

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 57 Monday, June 16, 2008 1:56 PM

TECHNICAL DATA 8

SITRANS FX300

a = 133 mm /
5.24 inch

b = 105 mm / 4.13 inch


c = 179 mm / 7.05 inch

Sandwich version EN
Pressure
rating

Size

Dimensions [mm]

Weight [kg]
With

DN

PN

Without

pressure sensor

15

100

16

45

65

265

144

4.1

3.5

25

100

24

65

65

265

144

4.9

4.3

40

100

38

82

65

270

144

5.5

4.9

50

100

50

102

65

275

144

6.6

80

100

74

135

65

290

155

8.8

8.2

100

100

97

158

65

310

164

10.1

9.5

Option: Version with two transmitters


Dimension H x 2
Specified weight + 2.80 kg

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

57

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 58 Monday, June 16, 2008 1:56 PM

TECHNICAL DATA

SITRANS FX300

a = 133 mm /
5.24 inch

b = 105 mm /
4.13 inch
c = 179 mm /
7.05 inch

Sandwich version ASME


Pressure
rating

Size

Dimensions [inches]

Weight [lb]
With

NPS

58

Class

Without

pressure sensor

1/2

150

0.63

1.77

2.56

10.43

5.67

9.04

7.72

1/2

300

0.63

1.77

2.56

10.43

5.67

9.04

7.72

1/2

600

0.55

1.77

2.56

10.43

5.67

9.04

7.72

150

0.94

2.56

2.56

10.43

5.67

10.8

9.48

300

0.94

2.56

2.56

10.43

5.67

10.8

9.48

600

0.94

2.56

2.56

10.43

5.67

10.8

9.48

1 1/2

150

1.5

3.23

2.56

10.63

5.67

12.13

10.8

1 1/2

300

1.5

3.23

2.56

10.63

5.67

12.13

10.8

1 1/2

600

1.5

3.23

2.56

10.63

5.67

12.13

10.8

150

1.97

4.02

2.56

10.83

5.67

14.55

13.23

300

1.97

4.02

2.56

10.83

5.67

14.55

13.23

600

1.97

4.02

2.56

10.83

5.67

14.55

13.23

150

2.91

5.31

2.56

11.42

6.1

19.4

18.08

300

2.91

5.31

2.56

11.42

6.1

19.4

18.08

600

2.91

5.31

2.56

11.42

6.1

19.4

18.08

150

3.82

6.22

2.56

12.21

6.46

22.27

20.94

300

3.82

6.22

2.56

12.21

6.46

22.27

20.94

600

3.82

6.22

2.56

12.21

6.46

22.27

20.94

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 59 Monday, June 16, 2008 1:56 PM

TECHNICAL DATA 8

SITRANS FX300

8.4 Flow tables


Measuring range limits
Size
DN
EN 1092-1

NPS
ASME B16.5

Qmin

Qmax

[m3/ h]

[m3/h]

Water
15

1/2

0.45

4.94

25

0.81

11.40

40

1 1/2

2.04

28.57

50

3.53

49.47

80

7.74

108.37

100

13.30

186.21

150

30.13

421.86

200

52.66

737.17

250

10

81.43

1140.02

300

12

114.83

1607.61

Values based on water at 20C

Air
15

1/2

6.79

32.56

25

10.20

113.94

40

1 1/2

25.35

326.63

50

43.89

565.49

80

96.14

1238.60

100

165.19

2128.27

150

374.23

4821.53

200

653.95

8425.53

250

10

977.16

13028.81

300

12

1377.95

18372.66

Values based on air at 20C and 1.013 bar abs

Flow rate limits


Nominal diameters

Minimum flow rates

Maximum flow rates

Product

EN

ASME

[m/s]

[m/s]

Liquids

DN15DN300

1/2"12"

0.5 x ( 998 / )0.5 1

7 x ( 998 / )0.47 1

Gas,
vapour

DN15DN300

1/2"12"

6 x ( 1.29 / )0.5 2

7 x ( 998 / )0.47 2

= operating density [kg/m3]


1 Minimum flow rates 0.4m/s - maximum flow rates 10m/s
2 Minimum flow rates 2m/s - maximum flow rates 80m/s; DN15 - 45 m/s and DN25 - 70 m/s

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

59

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 60 Monday, June 16, 2008 1:56 PM

TECHNICAL DATA

SITRANS FX300

Measuring range saturated steam: 1-7 bar


Overpressure [bar]

3.5

5.2

Density [kg/m]

1.13498

2.4258

3.27653

4.16732

Temperature [C]

120.6

148.2

160.4

170.6

Flow
DN
EN 1092-1

NPS
ASME B16.5

min

max

min

max

min

max

min

max

[kg/h]

[kg/h]

[kg/h]

[kg/h]

[kg/h]

[kg/h]

[kg/h]

[kg/h]

15

1/2

5.25

32.56

7.68

78.98

8.93

106.68

10.06

135.69

25

11.82

130.29

17.28

276.40

20.09

426.89

22.66

474.82

40

1 1/2

29.64

370.71

43.33

792.33

50.63

1070.2

56.8

1361.2

50

51.31

641.82

75.02

1371.8

87.19

1852.8

98.33

2356.6

80

112.41

1405.8

164.33

3004.7

191

4058.4

215.39

5161.8

100

193.14

2415.5

282.36

5162.7

328.16

6973.3

370.09

8869.2

150

437.56

5472.4

639.69

11696

743.45

15798

838.44

20093

200

764.62

9562.8

1117.8

20439

1299.2

27606

1465.1

35112

250

10

1177.07

14655.1

1716.4

31161.66

1993.6

42039.7

2247.44

53426.8

300

12

1659.85

20665.9

2420.39

43942.8

2811.29

58282.5

3169.24

75340.2

Measuring range saturated steam: 10.5 - 20 bar


Overpressure [bar]

10.5

14

17.5

20

Density [kg/m]

5.88803

7.60297

9.31702

10.5442

Temperature [C]

186.2

198.5

208.5

215

Flow

60

min

max

min

max

min

max

min

max

[kg/h]

[kg/h]

[kg/h]

[kg/h]

[kg/h]

[kg/h]

[kg/h]

[kg/h]

DN
EN
1092-1

NPS
ASME
B16.5

15

1/2

12.78

191.71

25

26.93

40

1 1/2

67.51

50

80

16.51

247.55

20.23

303.36

22.89

343.32

670.88

30.6

866.28

33.87

1061.58

36.04

1201.41

1878.2

76.72

2150.7

84.93

2395.3

90.35

2557.7

116.89

3251.7

132.82

3723.4

147.03

4147

156.42

4428.1

256.03

7122.4

290.93

8155.8

322.06

9083.7

342.62

9699.3

100

439.91

12238

499.9

14013

553.38

15608

588.69

16666

150

996.62

27725

1132.5

31747

1253.7

35359

1333.7

37756

200

1741.6

48449

1979

55478

2190.7

61789

2330.6

65977

250

10

2670.28

66065.16

3033.45

75626.77

3357.4

84214.04

3571

89910.45

300

12

3765.52

93162.2

4277.65

106645.6

4737.45

118754

5036.01

126787.8

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 61 Monday, June 16, 2008 1:56 PM

TECHNICAL DATA 8

SITRANS FX300

Measuring range saturated steam: 15 - 100 psig


Overpressure [psig]

15

50

75

100

Density [lbs/ft]

0.0719

0.1497

0.2036

0.2569

Temperature [F]

249.98

297.86

320.36

338.184

Flow

min

max

min

max

min

max

min

max

[lbs/h]

[lbs/h]

[lbs/h]

[lbs/h]

[lbs/h]

[lbs/h]

[lbs/h]

[lbs/h]

DN
EN
1092-1

NPS
ASME
B16.5

15

1/2

11.6

82.70

16.83

1720.50

25

26.25

289.40

37.86

40

1 1/2

65.81

829.61

94.92

50

113.94

1436.3

80

249.57

100

150

200

19.62

233.93

22.04

295.12

602.09

44.15

1726

110.68

818.63

49.59

1032.76

2346.7

124.32

2960.5

164.34

2988

191.63

4062.9

215.23

5125.6

3146.1

360

6545.3

419.74

8899.4

471.45

11227

428.81

5405.7

971.47

12246

618.51

11246

721.21

15291

810.06

19291

1401.2

25478

1633.9

34642

1835.2

43703

1697.6

21400

2448.6

44523

2855.2

60536

3206.9

76369

250

10

2562.72

32308.86

3777.85

68699.63

4371.7

92681.52

4946.03

117785.2

300

12

3613.84

45560.54

5327.61

96877.61

6164.78

130695.4

6974.68

166096.6

Measuring range saturated steam: 150 - 300 psig


Overpressure [psig]

150

200

250

300

Density [lbs/ft]

0.3627

0.4681

0.5735

0.6792

Temperature [F]

366.08

388.04

406.22

422.06

Flow

min

max

min

max

min

max

min

max

[lbs/h]

[lbs/h]

[lbs/h]

[lbs/h]

[lbs/h]

[lbs/h]

[lbs/h]

[lbs/h]

DN
EN
1092-1

NPS
ASME
B16.5

15

1/2

27.79

416.68

35.86

573.83

43.94

658.89

52.04

780.29

25

58.93

1458.16

66.94

1875.90

74.1

2089

80.63

2284.9

40

1 1/2

147.72

4107.2

167.83

4702.8

185.76

5237

202.15

5728

50

255.75

7111.9

290.56

8141.9

321.6

9066.8

350

9917

80

560.19

15578

636.44

17834

704.43

19860

766.6

21722

100

962.54

26766

1093.5

30643

1210.4

34124

1317.2

37324

150

2180.6

60639

2477.4

69421

2742.1

77307

2984

84556

200

3810.6

105960

4329.2

121310

4791.7

135090

5214.5

147760

250

10

5876.29

145648.6

6674.55

166728.3

7386.9

185660

7680.16

198218.4

300

12

8286.49

205387.3

9412.15

235112.9

10416.7

261809.6

10830.22

279518.9

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

61

Appendix

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 62 Monday, June 16, 2008 1:56 PM

APPENDIX

SITRANS FX300

9.1 Menu structure


9.1.1 Overview of firmware versions
There are three firmware versions, each of which is tailored to a different use of the device:

v1 (Basic): liquids and gases without compensation, saturated steam with temperature
compensation
v6 (Steam): saturated and superheated steam with pressure and temperature compensation,
gross heat meter
v7 (Gas): gas, gas mix and wet gas with pressure and temperature compensation, FAD (free
air delivery measurement)

The menu structure varies depending on the firmware version used.


The following table provides an overview of all the menu items in the first menu level.
For a complete description of the menus, please check the firmware version of your meter and
follow the corresponding references in the table.

Menu items

Firmware Version v1

Firmware Version v6

Firmware Version v7

1. Quick Setup

refer to Quick Setup menu on page 64

2. Tests

refer to Tests menu on page 65

3. Setup

refer to Menu item


Setup (Firmware version
v1 Basic) on page 66

5. Service

The Service menu can only be accessed by Service staff and is not described.

refer to Menu item


Setup (Firmware version
v6 Steam) on page 69

refer to Menu item


Setup (Firmware version
v7 Gas) on page 74

Table 9-1: Menu structure overview

INFORMATION!
The flowmeter has been set at the factory in accordance with the customer order. Therefore
subsequent configuration via the menu is only necessary if the intended use of the flowmeter is
changed.

62

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 63 Monday, June 16, 2008 1:56 PM

APPENDIX 9

SITRANS FX300

9.1.2 Entering values in change mode

Moves the insertion point one position to the right; after the last position, the insertion point
returns to the beginning.

Cycles through available values and characters; moves the decimal point to the right.
^
Accepting the entry.

9.1.3 Selecting characters in change mode


Depending on the menu function, a selection of the following characters is available:

Numbers
0

Letters (lowercase)
a

Letters (uppercase)
A

"

<

>

Symbols

1 "blank"

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

63

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 64 Monday, June 16, 2008 1:56 PM

APPENDIX

SITRANS FX300

9.1.4 Quick Setup menu


Level

Designation

1.1.1

Language

1.1.2

Location

1.1.3

Flow

1.1.4

Selection / input option

Select language for the menus


German ...

German

French

French

English ...^

English

0000000000
Location......^

Enter the name of the location (max.


10 characters)
Select the type of flow measurement

Volume

Volume measurement

Norm.Vol. 1

Standard volume measurement

Mass ...^

Mass measurement

Max. Flow

Explanation

Set maximum flow rate


m3/h
Unit ...^

Select unit for volume measurement

00600.0000 or other value


m3/h ......^

Enter value for the maximum


volumetric flow rate

Display Unit /
Display % Max Flow ...^

Display flow rate in units /


as a % of the maximum flow rate

m3/h norm
Unit ...^

Select unit for standard volume


measurement

00600.0000 or other value


m3/h norm ......^

Enter value for the maximum


standard volumetric flow rate

Display Unit /
Display % Max Flow ...^

Display flow rate in units /


as a % of the maximum flow rate

kg/h
Unit

Select unit for mass measurement

00600.0000 or other value


kg/h ......^

Enter value for the maximum mass


flow rate

Display Unit /
Display % Max Flow ...^

Display flow rate in units /


as a % of the maximum flow rate

1.1.5

Min. Flow

00300.0000
m3/h ......^

Enter value for the minimum flow rate


(> 0)

1.1.6

Timeconst.

00002.0000
s ......^

Enter time constant for output of


measured values in seconds
(0 20 s)
0: disabled

1
2
3
4

64

not available when Fluid = Liquid


available when FLOW = VOLUME (see menu item 1.1.3)
available when FLOW = NORM.VOL. (see menu item 1.1.3)
available when FLOW = MASS (see menu item 1.1.3)

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 65 Monday, June 16, 2008 1:56 PM

APPENDIX 9

SITRANS FX300

9.1.5 Tests menu


Level

Designation

2.1.1

Test I

Selection / input option

Explanation
Current output test

4 mA ^
8 mA ^
12 mA ^
16 mA ^
20 mA ^

2.1.2

Test P

0.5003 Hz ^

Pulse output test

Table 9-2: Tests menu

Every current value has to be confirmed with ^. Only after this the current output is set to the
respective value. The following is then displayed:

"Contin.Yes": continues with the current output test function


"Contin.No": returns to the menu.

The test current is maintained until you exit the menu.

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

65

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 66 Monday, June 16, 2008 1:56 PM

APPENDIX

SITRANS FX300

9.1.6 Menu item Setup (Firmware version v1 Basic)

Level

Designation

Selection / input
option

3.1
3.1.1

3.1.2

3.1.3

Configures the display


Error Msg

Cycle disp

Code 1

Show errors
Yes

...

No

...^

Shows error messages in measuring mode in plain text


alternating with the measured values
Shows a flashing cursor in the top left corner of the
display to indicated that there are errors.
Shows measured values

Yes

...

No

...^

Shows measured values cyclically in measuring mode


(Interval: 6 seconds)
Does not show measured values cyclically
Set security prompt 1 (Menu)

Yes

...

No

...^

Activate access code:


^^^
Deactivate access code

3.2
3.2.1

I/O settings
Range I

Current output 4-20 mA and error current


...

4-20mA

...

4-20/22E
4-20/3.55E
3.2.5

3.2.7

Function P

Tot.on/off

...^

Current output 4-20 mA


Current output 4-20 mA or error current 22 mA
Current output 4-20 mA or error current 3.55 mA
Pulse output

Yes

...

No

...^

Activate pulse output


Deactivate pulse output
Totalizer

Tot. on ...
Tot. off

66

Explanation

...^

Start totalizer
Stop totalizer (value is retained)

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 67 Monday, June 16, 2008 1:56 PM

APPENDIX 9

SITRANS FX300

Level

Designation

3.2.8

Tot.conf.

(available when FLOW =


VOLUME, see menu item
1.1.3)

(available when FLOW =


NORM.VOL., see menu
item 1.1.3)

(available when FLOW =


MASS, see menu item
1.1.3)

Selection / input
option

Explanation
Configures totalizer

m3
Unit ...^

Select totalizer unit for volume measurement

0000000000
m3 ......^

Enter preset totalizer value

Reset Yes /
Reset No ...^

Reset totalizer /
do not reset totalizer

Disp. Off /
Disp. On ...^

Display totalizer /
Do not display totalizer

m3 norm
Unit ...^

Select totalizer unit for standard volume measurement

0000000000
m3 norm ...
...^

Enter preset totalizer value

Reset Yes /
Reset No ...^

Reset totalizer /
do not reset totalizer

Disp. Off /
Disp. On ...^

Display totalizer /
Do not display totalizer /

kg
Unit ...^

Select totalizer unit for mass measurement

0000000000
kg ... ...^

Enter preset totalizer value

Reset Yes /
Reset No ...^

Reset totalizer /
do not reset totalizer

Disp. Off /
Disp. On ...^

Display totalizer /
Do not display totalizer

3.3

HART settings

3.3.1

Poll.Addr.

000
0 15
......^

Enter HART polling address for "Multidrop mode"

3.3.2

HART SV

Total Flow ^

HART secondary variable

3.3.3

HART TV

Temperature ^

HART tertiary variable

3.3.4

HART 4V

Density ^

HART quaternary variable

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

67

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 68 Monday, June 16, 2008 1:56 PM

APPENDIX

Level

Designation

SITRANS FX300

Selection / input
option

3.4
3.4.1

3.4.2

Set fluid and medium


Fluid

Set fluid type


Gas ...

Gas

Liquid ...

Liquid

Steam ...^

Steam

Medium

Set medium

(available when FLUID =


GAS or LIQUID, see
menu item 3.4.1)

Custom ...

Custom medium

(available when FLUID =


STEAM, see menu item
3.4.1)

Sat.Steam ...

Saturated steam

3.5
3.5.1

3.5.3

3.5.4

3.5.6

3.5.9

68

Explanation

Set pressure, temperature and density


T-Sensor

Sat. P/T

(available when FLUID =


STEAM, see menu item
3.4.1)
Temp. Opr.

Dens. Opr.

(available when FLOW =


NORM.VOL. or MASS,
see menu item 1.1.3)

Ref. density

Internal temperature sensor


No ...

No temperature sensor available

Yes ...^

Temperature sensor available


Density calculation with saturated steam

Sat. Temp. ^

Density calculation for saturated steam from the


saturated temperature
Operating temperature

C
Unit ...^

Set unit of temperature

0000000.0
C ^

Operating temperature

Disp. On /
Disp. Off ...^

Display temperature /
do not display temperature
Density with operating pressure and operating
temperature

kg/m3
Unit ...^

Set unit of density

00000.0000
kg/m3 ^

Operating density

00000.0000
kg/m3 ...
...^

Enter density for reference conditions (pressure and


temperature)

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 69 Monday, June 16, 2008 1:56 PM

APPENDIX 9

SITRANS FX300

9.1.7 Menu item Setup (Firmware version v6 Steam)

Level

Designation

Selection / input
option

3.1
3.1.1

3.1.2

3.1.3

Configures the display


Error Msg

Cycle disp

Code 1

Show errors
Yes

...

Shows error messages in measuring mode in plain


English alternating with the measured values

No

...^

Shows a flashing cursor in the top left corner of the


display to indicated that there are errors.
Shows measured values

Yes

...

No

...^

Shows measured values cyclically in measuring mode


(Interval: 6 seconds)
Does not show measured values cyclically
Set security prompt 1 (Menu)

Yes

...

No

...^

Activate access code:


^^^
Deactivate access code

3.2
3.2.1

I/O settings
Range I

Current output 4-20 mA and error current


...

4-20mA

...

4-20/22E
4-20/3.55E
3.2.2

3.2.3

3.2.4

3.2.5

3.2.6

Explanation

Variable I

(available when METER


TYPE = GROSS HEAT,
see menu item 5.3.1,
Service menu)
Power Unit

(available when METER


TYPE = GROSS HEAT,
see menu item 5.3.1,
Service menu)
FS Power

(available when
VARIABLE I = POWER,
see menu item 3.2.2)
Function P

Variable P

(available when METER


TYPE = GROSS HEAT,
see menu item 5.3.1,
Service menu)

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

...^

Current output 4-20 mA


Current output 4-20 mA or error current 22 mA
Current output 4-20 mA or error current 3.55 mA
Set output variable for current output

...

Flow

Flow

...^

Power

Power

Set heat output unit


kJ/h
Unit

Select unit for heat output measurement

...^

Disp. On /
Disp. Off ...^

Display measured value/


do not display measured value
Full-scale power value

7000000.00
kJ/h ^

Value of power with 20 mA current output

Pulse output
Yes

...

No

...^

Activate pulse output


Deactivate pulse output
Set output variable for pulse output

Total Flow
Energy

...

...^

Total flow
Energy

www.siemens.com/flow

69

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 70 Monday, June 16, 2008 1:56 PM

APPENDIX

Level

Designation

3.2.7

Tot.on/off

3.2.8

(available when FLOW =


NORM.VOL., see menu
item 1.1.3)

(available when FLOW =


MASS, see menu item
1.1.3)

3.2.10

70

Selection / input
option

E.Tot.on

(available when METER


TYPE = GROSS HEAT,
see menu item 5.3.1,
Service menu)
Energ.Unit

(available when METER


TYPE = GROSS HEAT,
see menu item 5.3.1,
Service menu)

Explanation
Totalizer

Tot. on

...

Start totalizer

Tot. off

...^

Stop totalizer (value is retained)

Tot.conf.

(available when FLOW =


VOLUME, see menu item
1.1.3)

3.2.9

SITRANS FX300

Configures totalizer
m3
Unit ...^

Select totalizer unit for volume measurement

0000000000
m3 ......^

Enter preset totalizer value

Reset Yes /
Reset No ...^

Do not reset totalizer/


reset totalizer

Disp. Off /
Disp. On ...^

Do not display totalizer


Display totalizer /

m3 norm
Unit ...^

Select totalizer unit for standard volume measurement

0000000000
m3 norm ...
...^

Enter preset totalizer value

Reset Yes /
Reset No ...^

Do not reset totalizer/


reset totalizer

Disp. Off /
Disp. On ...^

Do not display totalizer


Display totalizer /

kg
Unit ...^

Select totalizer unit for mass measurement

0000000000
kg ... ...^

Enter preset totalizer value

Reset Yes /
Reset No ...^

Do not reset totalizer/


reset totalizer

Disp. Off /
Disp. On ...^

Do not display totalizer


Display totalizer /
Energy totalizer

Tot. on

...

Start totalizer

Tot. off

...^

Stop totalizer (value is retained)

Set thermal energy unit


kJ
Unit

...^

Set thermal energy unit

0000000000
kJ ......^

Enter preset totalizer value

Reset No /
Reset Yes ...^

Do not reset thermal energy/


reset thermal energy

Disp. On /
Disp. Off ...^

Display thermal energy /


do not display thermal energy

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 71 Monday, June 16, 2008 1:56 PM

APPENDIX 9

SITRANS FX300

Level

Designation

Selection / input
option

3.3

Explanation
HART settings

3.3.1

Poll.Addr.

3.3.2

HART SV

000
0 15
......^

Enter HART polling address for "Multidrop mode"

HART secondary variable


Total Flow ...
Energy
...^
#available when
METER TYPE =
GROSS HEAT
(menu item 5.3.1,
Service menu)

3.3.3

HART TV

HART tertiary variable


Temperature ...
Pressure ...
...

Density
3.3.4

HART 4V

HART quaternary variable


Temperature ...
Pressure ...
...

Density

3.4

Set fluid and medium

3.4.1

Fluid

3.4.2

Medium

3.4.5

Dry.Fact.

(available when MEDIUM


= SAT.STEAM, see menu
item 3.4.2)

Fluid type
Steam ^

Steam
Process medium

Sat.Steam ...

Sat.Steam ...Saturated steam

Sup.Steam
...^

Superheated steam
Enter dryness fraction for saturated steam

0000001.00
0.85 TO 1
......^

3.5
3.5.1

Pressure, temperature and density


T-Sensor

Internal temperature sensor


No ...
Yes

3.5.2

Factor = 1 - %body of water

P-Sensor

No temperature sensor available


...^

Pressure sensor
Internal ...
---

Internal pressure sensor


...

External

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

Temperature sensor available

...^

External pressure sensor


No pressure sensor available

www.siemens.com/flow

71

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 72 Monday, June 16, 2008 1:56 PM

APPENDIX

Level

Designation

3.5.3

Sat. P/T

(available when MEDIUM


= SAT.STEAM, see menu
item 3.4.2)

3.5.4

3.5.5

Temp. Opr.

(available for saturated


steam with SAT.TEMP or
SUP.STEAM, see menu
items 3.4.2 and 3.5.3)

Pres. Opr.
(available for SAT.STEAM
with SAT.PRES. or
SUP.STEAM, see menu
items 3.4.2 and 3.5.3)

SITRANS FX300

Selection / input
option

Explanation
Density calculation with saturated steam

Sat. Temp.

...

Sat. Pres. ...^

Density calculation for saturated steam from the


saturated temperature
Density calculation for saturated steam from the
saturated temperature
Operating temperature

C
Unit ...^

Set unit of temperature

0000000.0
C ^

Operating temperature

Disp. On /
Disp. Off ...^

Display temperature /
do not display temperature
Operating pressure

Pa
Unit ...^

Set unit of pressure

0000000.0
Pa ^

Operating pressure

Disp. On /
Disp. Off ...^

Display pressure /
do not display pressure

3.5.7

Temp.Norm.
(available when FLOW =
NORM.VOL., see menu
item 1.1.3)

00000020.0
C ......^

Enter reference temperature


Unit as in 3.5.4

3.5.8

Pres.Norm.
(available when FLOW =
NORM.VOL., see menu
item 1.1.3)

00000000.0
Pa ......^

Enter reference pressure


Unit as in 3.5.5

3.5.10

P-Excit. V

(available when PSENSOR = INTERNAL,


see menu item 3.5.2)

0005.00000
V ^

Excitation voltage of the pressure sensor

3.5.11

P-Sen.P1V1

(available when PSENSOR = INTERNAL,


see menu item 3.5.2)

Pressure sensor:
enter 1st calibration point
0001.00000
P1 kg/cm2g
......^

0002.00000
V1 mV ...
...^
3.5.12

P-Sen.P2V2

(available when PSENSOR = INTERNAL,


see menu item 3.5.2)

Pressure sensor:
enter 2nd calibration point
0005.00000
P2 kg/cm2g
... ...^

0048.00048
V2 mV ...
...^

72

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 73 Monday, June 16, 2008 1:56 PM

APPENDIX 9

SITRANS FX300

Level

Designation

3.5.13

Ext.P.Rng

(available when PSENSOR = EXTERNAL,


see menu item 3.5.2)

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

Selection / input
option

Explanation
Enter range of external pressure sensor

0000.00000
P. 4mA ^

Enter lower range value

0006.00000
P. 20mA ^

Enter upper range value

www.siemens.com/flow

73

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 74 Monday, June 16, 2008 1:56 PM

APPENDIX

SITRANS FX300

9.1.8 Menu item Setup (Firmware version v7 Gas)

Level

Designation

Selection / input
option

3.1
3.1.1

3.1.2

3.1.3

Configures the display


Error Msg

Cycle disp

Code 1

Show errors
Yes

...

No

...^

Shows error messages in measuring mode in plain text


alternating with the measured values
Shows a flashing cursor in the top left corner of the
display to indicate that there are errors.
Shows measured values

Yes

...

No

...^

Shows measured values cyclically in measuring mode


(Interval: 6 seconds)
Does not show measured values cyclically
Set security prompt 1 (Menu)

Yes

...

No

...^

Activate access code:


^^^
Deactivate access code

3.2
3.2.1

I/O settings
Range I

Current output 4-20 mA and error current


...

4-20mA

...

4-20/22E
4-20/3.55E
3.2.5

3.2.7

3.2.8

Function P

Tot.on/off

...^

Current output 4-20 mA


Current output 4-20 mA or error current 22 mA
Current output 4-20 mA or error current 3.55 mA
Pulse output

Yes

...

No

...^

Activate pulse output


Deactivate pulse output
Totalizer

Tot. on

...

Start totalizer

Tot. off

...^

Stop totalizer (value is retained)

Tot.conf.

(available when FLOW =


VOLUME, see menu item
1.1.3)

74

Explanation

Configures totalizer
m3
Unit ...^

Select totalizer unit for volume measurement

0000000000
m3 ... ...^

Enter preset totalizer value

Reset Yes /
Reset No ...^

Reset totalizer
Do not reset totalizer/

Disp. Off /
Disp. On ...^

Do not display totalizer


Display totalizer /

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 75 Monday, June 16, 2008 1:56 PM

APPENDIX 9

SITRANS FX300

Level

Designation
(available when FLOW =
NORM.VOL., see menu
item 1.1.3)

(available when FLOW =


MASS, see menu item
1.1.3)

Selection / input
option

Explanation

m3 norm
Unit ...^

Select totalizer unit for standard volume measurement

0000000000
m3 norm ...
...^

Enter preset totalizer value

Reset Yes /
Reset No ...^

Reset totalizer
Do not reset totalizer/

Disp. Off /
Disp. On ...^

Do not display totalizer


Display totalizer /

kg
Unit ...^

Select totalizer unit for mass measurement

0000000000
kg ......^

Enter preset totalizer value

Reset Yes /
Reset No ...^

Reset totalizer
Do not reset totalizer/

Disp. Off /
Disp. On ...^

Do not display totalizer


Display totalizer /

3.3

HART settings

3.3.1

Poll.Addr.

3.3.2

HART SV

000
0 15
......^

Enter HART polling address for "Multidrop mode"

HART secondary variable


Total Flow ...
FAD ...^
#available when
METER TYPE =
FAD Meter (see
menu item 5.3.1,
Service staff only)

3.3.3

HART TV

HART tertiary variable


Temperature ...
Pressure ...
Density

...

FAD
...^
#available when
METER TYPE =
FAD Meter (see
menu item 5.3.1,
Service staff only)

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

75

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 76 Monday, June 16, 2008 1:56 PM

APPENDIX

Level

Designation

3.3.4

HART 4V

SITRANS FX300

Selection / input
option

Explanation
HART quaternary variable

Temperature ...
Pressure ...
...

Density

FAD
...^
#available when
METER TYPE =
FAD Meter (see
menu item 5.3.1,
Service staff only)
3.4
3.4.1

Set fluid and medium


Fluid

Set fluid type


Gas ...

Gas

Gas Mix ...


3.4.2

Medium
(available when FLUID =
GAS / WET-GAS, see
menu item 3.4.1)

% Gas
(available when FLUID =
GAS-MIX, see menu item
3.4.1)

3.4.4

%Rel.Hum .
(available when FLUID =
WET-GAS, see menu item
3.4.1)

3.4.6

Fad Unit
(available when METER
TYPE = FAD METER, see
menu item 5.3.1, Service
staff only)

3.4.7

3.4.8

76

Wet gas
Set process medium

Air ...

Air

Ammonia ...

Ammonia

Argon ...

Argon

etc.
3.4.3

Gas mix
...^

Wet Gas

...^

Other gases not listed here


Define composition of gas mix

050.000000
Air ......^

Select gas and enter percentage

100.00
Total % ^

Total percentage

0000000.00
% Rel.Hum.
......^

Specify relative humidity

Specify unit for FAD


FAD
Unit

m3/h

Set unit for FAD


...^

Disp. On /
Disp. Off ...^

Suct.Temp.
(available when METER
TYPE = FAD METER, see
menu item 5.3.1, Service
staff only)

C
Unit

Atm.Press.
(available when METER
TYPE = FAD METER, see
menu item 5.3.1, Service
staff only)

Pa
Unit

Display measured value /


do not display measured value
Enter temperature at suction side of compressor

...^

0000200.00
C ......^

Select temperature unit


Enter temperature value
Atmospheric pressure

...^

00001.0000
Pa ......^

Select pressure unit


Select pressure unit

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 77 Monday, June 16, 2008 1:56 PM

APPENDIX 9

SITRANS FX300

Level

Designation

3.4.9

Fil.P.Drop
(available when METER
TYPE = FAD METER, see
menu item 5.3.1, Service
staff only)

Selection / input
option

Explanation
Filter pressure drop at compressor input

Pa
Unit

...^

Select pressure unit

00000.0000
Pa ......^

Select pressure unit

3.4.10

Inlet RH
(available when METER
TYPE = FAD METER, see
menu item 5.3.1, Service
staff only)

0000060.00
% Rel.Hum.
......^

Enter relative humidity at suction side of compressor

3.4.11

Actual Rpm
(available when METER
TYPE = FAD METER, see
menu item 5.3.1, Service
staff only)

0001500.00
RPM ^

Current speed of compressor motor as revolutions per


minute

3.4.12

Rated Rpm
(available when METER
TYPE = FAD METER, see
menu item 5.3.1, Service
staff only)

0001500.00
RPM ^

Rated speed of compressor motor as revolutions per


minute

3.4.13

Outlet Rh
(available when METER
TYPE = FAD METER, see
menu item 5.3.1, Service
staff only)

0000100.00
% Rel.Hum.
......^

Enter relative humidity at meter (compressor outlet)

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

www.siemens.com/flow

77

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 78 Monday, June 16, 2008 1:56 PM

APPENDIX

Level

Designation

SITRANS FX300

Selection / input
option

3.5
3.5.1

Pressure, temperature and density


T-Sensor

Internal temperature sensor


No ...
Yes

3.5.2

P-Sensor

No temperature sensor available


...^

---

3.5.5

3.5.6

Temp. Opr.

Pres. Opr.

Internal pressure sensor


...

...^

External pressure sensor


No pressure sensor available
Operating temperature

C
Unit ...^

Set unit of temperature

0000000.0
C ^

Operating temperature

Disp. On /
Disp. Off ...^

Display temperature /
do not display temperature
Operating pressure

Pa
Unit ...^

Set unit of pressure

0000000.0
Pa ^

Operating pressure

Disp. On /
Disp. Off ...^

Display pressure /
do not display pressure

Dens. Opr.
(for unknown gases with
standard volume or mass
measurement, see menu
items 1.1.3, 3.4.1 and
3.4.2)

Temperature sensor available


Pressure sensor

Internal ...
External

3.5.4

Explanation

Density with operating pressure and operating


temperature
kg/m3
Unit ...^

Set unit of density

00011.0000
kg/m3 ^

Operating density

3.5.7

Temp.Norm .
(available when FLOW =
NORM.VOL., see menu
item 1.1.3)

00000020.0
C ......^

Enter reference temperature


Unit as in 3.5.4

3.5.8

Pres.Norm .
(available when FLOW =
NORM.VOL., see menu
item 1.1.3)

00000000.0
Pa ......^

Enter reference pressure


Unit as in 3.5.5

3.5.9

Temp.Norm.
(for unknown gases with
standard volume
measurement, see menu
items 1.1.3, 3.4.1 and
3.4.2)

00001.2900
kg/m3 ^

Enter density for reference conditions (pressure and


temperature)

3.5.10

P-Excit. V
(available when PSENSOR = INTERNAL,
see menu item 3.5.2)

0005.00000
V^

Excitation voltage of the pressure sensor

78

www.siemens.com/flow

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

MA_Siemens_Sitrans_FX300_R01_en_PRT.book Page 79 Monday, June 16, 2008 1:56 PM

APPENDIX 9

SITRANS FX300

Level

Designation

3.5.11

P-Sen.P1V1
(available when PSENSOR = INTERNAL,
see menu item 3.5.2)

Selection / input
option

Explanation
Pressure sensor:
1st calibration point

0001.00000
P1 kg/cm2g
......^

0002.00000
V1 mV ...
...^
3.5.12

P-Sen.P2V2
(available when PSENSOR = INTERNAL,
see menu item 3.5.2)

Pressure sensor:
2nd calibration point
0005.00000
P2 kg/cm2g
... ...^

0048.00048
V2 mV ...
...^
3.5.13

Ext.P.Rng
(available when PSENSOR = EXTERNAL,
see menu item 3.5.2)

06/2008 SFIDK.PS.050.F1. 02 A5E02100423

Enter range of external pressure sensor


0000.00000
P. 4mA ...
...^

Enter lower range value

0006.00000
P. 20mA ...
...^

Enter upper range value

www.siemens.com/flow

79

Das könnte Ihnen auch gefallen