Beruflich Dokumente
Kultur Dokumente
0369
of Biomedical Engineering, Erasmus MC, P. O. Box 2040, Rotterdam, 3000 CA, the Netherlands
Cardiology Institute of the Netherlands, Utrecht, 3501 DG, the Netherlands
Department of Physics of Fluids, University of Twente, P. O. Box 217, Enschede, 7522 NB, the Netherlands
Interuniversity
AbstractThe impulse response of a transducer can be represented in the frequency domain by its complex analog, the
transfer function. The amplitude transfer function is measured
regularly in contrast to the phase transfer function (PTF).
Applications for the PTF range from adjusting the emitted pulse
shape for coding based imaging to the optimization of ultrasound
contrast imaging methods based on destructive interference. A
number of acoustic methods to measure a transducers PTF exists, but they usually require accurate distance and acoustic wave
speed measurements. Small discrepancies in these cause large
phase errors. We present a pulse-echo method to measure a transducers PTF without needing a measurement of the wave travel
distance and speed. We generalize it to rectangular transducers.
In our method the transducer is excited by a monofrequency
sine burst with a rectangular envelope. The transducer initially
vibrates at resonance (transient regime) prior to the forcing
frequency (steady state regime). The PTF value of the system is
the difference between the phases deduced from the transient and
the steady state regimes at different forcing frequencies. As the
PTF is calculated from a relative difference measuring the wave
travel distance or speed is unnecessary. The approach assumes
linear wave propagation and uses a pulse-echo setup. The method
was tested on a custom built single element transducer (square:
13 x 13 mm, center frequency 4 MHz, no backing or matching
layers). The results were compared with KLM model simulations.
Also, we phase calibrated a hydrophone, which was then used
to measure the PTF of the square transducer. The simulated
and measured resonance frequencies differed by 0.17 MHz. The
mean PTF difference between simulation and measurements was
7 - 14 . The methods reproducibility was 15 . The PTF of
the transducer was measured with good reproducibility, without
measuring the wave travel distance or speed of sound in the
medium. Our simple setup requires basic laboratory ultrasound
equipment.
Index Termsphase; pulse-echo; hydrophone; rectangular
transducers; transducer calibration.
I. I NTRODUCTION
The impulse response of a transducer can be represented in
the frequency domain by its complex analog, the transfer function. The amplitude transfer function is measured regularly in
contrast to the phase transfer function (PTF). Applications for
the PTF are for example adjusting the emitted pulse shape
for coding based imaging [1], the optimization of ultrasound
contrast imaging methods based on destructive interference
[2], [3] and dissociating the microbubble and setup responses
in single microbubble experiments [4]. A number of acoustic
methods to measure a transducers PTF exists, but they usually
require accurate distance and acoustic wave speed measurements. Small discrepancies in these cause large phase errors,
2 + 2
0 0 + 02 0
(0 , 0 , ) =
2 (
)
(
)
0
0
(1)
where
(0 , 0 , ) is the pressure at location (0 , 0 , ), is
.
(2)
2
By replacing the on-axis far field pressure produced by a
circular symmetric transducer in the derivation presented in
[5] by equation (2) the reciprocity relation connecting the
receive transfer function ( ) and the transmit transfer function
( ) for rectangular transducers can be derived analogously.
This results in the same relation as for circular symmetric
transducers:
2
=
,
(3)
0 0
1454
(0, 0, ) = 0
(4)
( ) =
C. Hydrophone measurement
Assuming linear propagation, the hydrophone signal in a
hydrophone experiment is described by:
= ,
(7)
1455
Fig. 2.
1456
were 14 and 8 for the transmit and receive PTFs respectively. The differences in mean value and shape between the
simulations and pulse-echo measurements are due to alignment
errors and diffraction induced phase bias in the measurements
- they are performed at a distance of 59 cm, but diffraction
is approximated by its far field limit. The mean differences
between the simulations and the hydrophone measurements
were respectively 13 and 7 . Part of the explanation for the
differences in mean value and shape between the simulations
and hydrophone measurements are alignment errors. The other
part has to do with the calibration of the hydrophone, which
was performed using a circular source at an axial distance
of 18 cm. At this axial distance and a 1 - 8 MHz frequency
range diffraction significantly affects the measured PTF of the
hydrophone. By also measuring the square transducers PTF at
an axial distance of 18 cm, these diffraction effects on the PTF
are mostly canceled out. This cancelation is not complete, as
the circular source has a slightly lower surface area compared
to the square transducer. The transducers resonance frequency
was measured to be 3.87 MHz and 3.89 MHz in the pulse-echo
and hydrophone experiments respectively. The KLM simulated
resonance frequency was 4.05 MHz, which differed from the
experimental results by 0.17 MHz. This was likely caused
by the variation of piezomaterial properties between batches.
The shape of the PSF is typical for a weakly damped, forced
harmonic oscillator. The reproducibility of the results was
15 .
[5] P. van Neer, H. Vos, and N. de Jong, Reflector-based phase calibration of ultrasound transducers, Ultrasonics, vol. Article in press,
doi:10.1016/j.ultras.2010.05.001, 2010.
[6] R. Cobbold, Foundations of Biomedical Ultrasound. Oxford, New York:
Oxford University Press, Inc., pp. 187, 2007.
[7] D. Leedom, R. Krimholtz, and G. Matthaei, Equivalent circuits for
transducers having arbitrary even- or odd-symmetry piezoelectric excitation, IEEE Trans. Sonics Ultrason., vol. SU-18, no. 3, pp. 128141,
1971.
[8] E. Merks, J. Borsboom, M. Voormolen, N. Bom, A. van der Steen, and
N. de Jong, A klm-circuit model of a multi-layer transducer for bladder
volume measurements, Ultrasonics, vol. 44, no. Suppl. 1, pp. 705710,
2006.
[9] E. Merks, A. Bouakaz, N. Bom, C. Lancee, A. van der Steen, and
N. de Jong, Design of a multi-layer transducer for acoustic bladder
volume assessment, IEEE Trans. Ultrason. Ferroelect. Freq. Control,
vol. 53, no. 10, pp. 17301738, 2006.
[10] S. Sherrit, H. Wiederick, and B. Mukherjee, A complete characterization of the piezoelectric dieelectric and elastic properties of motorola
pzt 3203 hd including losses and dispersion, in Proc. SPIE, vol. 3037,
Newport Beach, 1997, pp. 158169.
V. D ISCUSSION / CONCLUSIONS
Our pulse-echo method allows for the measurement of the
PTF of rectangular transducers, without measuring the wave
travel distance or speed of sound in the medium. Only basic
laboratory ultrasound equipment is required. Measured PTFs
of a square example transducer were presented showing only
a small difference between measurements and KLM model
simulations (between 7 and 14 ). The reproducibility of the
results was 15 . Moreover, weve shown the feasibility of
using a previously PTF calibrated transducer to measure the
receive PTF of a hydrophone, which was then used to measure
the PTFs of the square transducer.
ACKNOWLEDGMENT
The financial support of the Dutch Foundation for Technical
Sciences is greatly appreciated.
R EFERENCES
[1] T. Misardis and J. Jensen, Use of modulated excitation signals in
medical ultrasound. part ii: design and performance for medical imaging
applications, IEEE Trans. Ultrason. Ferroelectr. Freq. Control, vol. 52,
no. 2, pp. 192207, 2005.
[2] S. Krishnan, J. Hamilton, and M. ODonnell, Suppression of propagating second harmonic in ultrasound contrast imaging, IEEE Trans.
Ultrason. Ferroelectr. Freq. Control, vol. 45, no. 3, pp. 704711, 1998.
[3] C. Shen and Y. Hsieh, Optimal transmit phasing on tissue background
suppression in contrast harmonic imaging, Ultrasound Med. Biol.,
vol. 34, no. 11, pp. 18201831, 2008.
[4] F. Guidi, H. Vos, R. Mori, N. De Jong, and P. Tortoli, Microbubble
characterization through acoustically-induced deflation, IEEE Trans.
Ultrason. Ferroelectr. Freq. Control, vol. 57, no. 1, pp. 193202, 2010.
1457