The installation steps of Universal hardness tester Vexus SHR187.5D


1. Prepare a instrument table (like below picture), there is hole on the table at diameter of 90mm, please check
the chapter 4.2.2 of the operation manual;

2. Take off the screws, open the plywood box of Vexus SHR187.5D, then place this hardness tester on the
instrument table and aim at the hole;

- 1/11 A 03 117
523080
Add: #117, 3-Dong, A-Qu, HongCheng Guoji Wujin ShangCheng, Nancheng District, DongGuan, China (523080)


DongGuan Sinowon Precision Instrument Co., Ltd
Tel: +86-769-23184144
Fax: +86-769-22854144
www.sinowon.com.cn
sinowon@sinowon.com.cn
3. Install four level bolts and adjust the levelness;
4. Take out the cover on the back and top of machine, then take out the bandage, elastic and cushion;

5. Install the weights and then install the dustproof cover, please pay attention to the number of weight;

Note: when you want to test 30Kg/HV30, you need take out this weight (No.5) in advance.

- 2/11 A 03 117
523080
Add: #117, 3-Dong, A-Qu, HongCheng Guoji Wujin ShangCheng, Nancheng District, DongGuan, China (523080)


DongGuan Sinowon Precision Instrument Co., Ltd
Tel: +86-769-23184144
Fax: +86-769-22854144
www.sinowon.com.cn
sinowon@sinowon.com.cn
6. Install the slipped testing table and indenter;

Note: please according to the hardness scale to choose a correct indenter and test force. For example, HRC
Rockwell hardness: diamond indenter and 150Kg test force.

- 3/11 A 03 117
523080
Add: #117, 3-Dong, A-Qu, HongCheng Guoji Wujin ShangCheng, Nancheng District, DongGuan, China (523080)


DongGuan Sinowon Precision Instrument Co., Ltd
Tel: +86-769-23184144
Fax: +86-769-22854144
www.sinowon.com.cn
sinowon@sinowon.com.cn
7. Install the digital eyepiece and objective;

- 4/11 A 03 117
523080
Add: #117, 3-Dong, A-Qu, HongCheng Guoji Wujin ShangCheng, Nancheng District, DongGuan, China (523080)


DongGuan Sinowon Precision Instrument Co., Ltd
Tel: +86-769-23184144
Fax: +86-769-22854144
www.sinowon.com.cn
sinowon@sinowon.com.cn

- 5/11 A 03 117
523080
Add: #117, 3-Dong, A-Qu, HongCheng Guoji Wujin ShangCheng, Nancheng District, DongGuan, China (523080)


DongGuan Sinowon Precision Instrument Co., Ltd
Tel: +86-769-23184144
Fax: +86-769-22854144
www.sinowon.com.cn
sinowon@sinowon.com.cn
8. Start to test it, first, please according to the hardness scale to choose a correct indenter and test force;

HBW: Brinell hardness scale (test result);


5x objective;
5.0:
Vickers hardness scale (convert result) you can setting it by yourself;
HV:
D1, D2: Length of indentation;
Dwell (s): dwell time (you can setting it by yourself);
Number: test times;
2011/04/23_11:46:55: current date and time you can setting it by yourself;
Indenter: d=2.5mm, Test force: F=187.5Kg correct indenter and test force;
Note: if you want to change the parameter, please press REVI into the setting interface.

- 6/11 A 03 117
523080
Add: #117, 3-Dong, A-Qu, HongCheng Guoji Wujin ShangCheng, Nancheng District, DongGuan, China (523080)


DongGuan Sinowon Precision Instrument Co., Ltd
Tel: +86-769-23184144
Fax: +86-769-22854144
www.sinowon.com.cn
sinowon@sinowon.com.cn

Main interface

Conversion Hardness Scale

Test Hardness Scale

HRC Rockwell Hardness Scale Interface

- 7/11 A 03 117
523080
Add: #117, 3-Dong, A-Qu, HongCheng Guoji Wujin ShangCheng, Nancheng District, DongGuan, China (523080)


DongGuan Sinowon Precision Instrument Co., Ltd
Tel: +86-769-23184144
Fax: +86-769-22854144
www.sinowon.com.cn
sinowon@sinowon.com.cn
9. Turn the working table to load the preliminary test force and listen the sound (or see LCD screen), when you
listen a drop noise, it show that the preliminary test force is complete, the machine will be load, unload test
force and display the test result automatically.

- 8/11 A 03 117
523080
Add: #117, 3-Dong, A-Qu, HongCheng Guoji Wujin ShangCheng, Nancheng District, DongGuan, China (523080)


DongGuan Sinowon Precision Instrument Co., Ltd
Tel: +86-769-23184144
Fax: +86-769-22854144
www.sinowon.com.cn
sinowon@sinowon.com.cn
Note:
You need fix the rod, leverage and weight before you send to your customers or move it. Please see the
below pictures.

- 9/11 A 03 117
523080
Add: #117, 3-Dong, A-Qu, HongCheng Guoji Wujin ShangCheng, Nancheng District, DongGuan, China (523080)


DongGuan Sinowon Precision Instrument Co., Ltd
Tel: +86-769-23184144
Fax: +86-769-22854144
www.sinowon.com.cn
sinowon@sinowon.com.cn

- 10/11 A 03 117
523080
Add: #117, 3-Dong, A-Qu, HongCheng Guoji Wujin ShangCheng, Nancheng District, DongGuan, China (523080)


DongGuan Sinowon Precision Instrument Co., Ltd
Tel: +86-769-23184144
Fax: +86-769-22854144
www.sinowon.com.cn
sinowon@sinowon.com.cn

- 11/11 A 03 117
523080
Add: #117, 3-Dong, A-Qu, HongCheng Guoji Wujin ShangCheng, Nancheng District, DongGuan, China (523080)

Das könnte Ihnen auch gefallen