Beruflich Dokumente
Kultur Dokumente
The ISO 13320 standard explains how to calibrate and verify a particle size analyzer using Certified
Reference Materials CRMs. CRMs are measured using 5 primary methods to be sure of the powder size.
The results are traceable to major international standards (NIST).
Validation should be done using spherical CRMs such as glass beads and using the Mie theory.
Target values
Test Method
Repeatability
(6.4 in the standard)
Coefficient of variation:
< 3% for X50
< 5%* for X10 & X90
Accuracy
(6.5 in the standard)
info.particle@cilas.com
www.cilas.com
Available CRM
Certified Reference Materials (CRMs) (figure 3) are made of spherical
particles and the size distribution of these particles must be established
by using an absolute method.
There are 5 absolute methods for size characterization: Electroformed
Sieve, Gravitational Sedimentation, Centrifugal Sedimentation, Optical
Microscopy and Electrical Sensing Zone. The size distribution ratio of a
CRM is 1:10 (for example, a size distribution from 10 to 100 ms).
Several companies distribute CRMs : NIST and Whitehouse Scientific.
Whitehouse Scientific manufactures a single-shot glass microsphere
covering the complete range of measurement from 0.1 to 2,500 m.
Figure 3 : SEM Picture of
Certified Glass Beads
info.particle@cilas.com
www.cilas.com
The same ISO Calibration method is used for both wet and dry mode. This results in an excellent
correlation between the two different modes with respect to the demands of the ISO standard (figure 6
and 7).
For the same instrument, CILAS can guarantee the following results:
- Accuracy: <3%
- Reproducibility: less than 1% on D50
info.particle@cilas.com
www.cilas.com
Conclusion
CILAS is the only particle size analyzer manufacturer who calibrates the optical
bench and the dispersing system at the same time. Calibrating both systems at the
same time allows the user to eliminate measurement errors due to the powder
dispersion system.
Our instrument with DJD technology is completely ISO compliant.
The major advantage of CILAS technology is that both modes of dispersion
are calibrated and validated using the same procedure. This allows the user
good correlation between the two modes and a greater accuracy.
Bibliography
[1] The Development of New Reference Standards For Particle Size Instrument Calibration
Rideal, Dodds, Lechonski, Loyd, Mercus, World Congress on Particle Technology 3 (IchemE), 1998, Brighton,
England
[2] International Standard , ISO 13320-1, Particle Size Analysis-Laser Diffraction Methods, General Principles
info.particle@cilas.com
www.cilas.com